Parts and Repair Service GuideADVIA Centaur®and ADVIA Centaur® XP Immunoassay Systems
© 2006 Bayer HealthCare LLC. All rights reserved.No part of this service guide or the products it describes may be reproduced by any means or in any form without prior consent in writing from Bayer HealthCare LLC.Bayer, the Bayer cross, ADVIA Centaur, and ReadyPack are trademarks of Bayer. BayerCare Connect is a servicemark of Bayer.Corvac and Monoject are trademarks of Sherwood Medical Company. Courier and U.S. Robotics are trademarks of 3COM Corporation.GlidePoint is a trademark of Cirque Corporation.Lexmark is a trademark of Lexmark International, Inc. Microsoft Excel is a trademark of Microsoft Corporation.Microtainer and Vacutainer are trademarks of Becton, Dickinson and Company. Monovette is a trademark of Walter Sarstedt Kunststoff-Spritzgusswerk.Sparc is a trademark of Sparc International, Inc.Origin: Ireland
The information in this Parts and Repairs guide was correct at the time of printing. However, Bayer HealthCare LLC continues to improve products and reserves the right to change specifications, equipment, and maintenance procedures at any time without notice.If the system is used in a manner differently than specified by Bayer HealthCare LLC, the protection provided by the equipment may be impaired. See warning and hazard statements.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 1
078D0080-02, Rev. A
Contents
1 Parts, Parts Lists, Replacement Procedures
Cabinetry for the ADVIA Centaur SystemSystem Overview - Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1System Overview – Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2A/C Input Line Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4Monitor Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5Sample Entry Queue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6Sample Exit Queue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7Electronics Module Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8Sample Tip Waste Area Door. . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9Cuvette Waste Area Door. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10Lab Automation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12Parts List - Cabinetry for ADVIA Centaur System . . . . . . . . . . 1-13
Cabinetry for the ADVIA Centaur XP SystemSystem Overview - Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17System Overview - Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18A/C Input Line Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18Monitor Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19Sample Entry Queue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19Sample Exit Queue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20Electronics Module Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20Corner Door. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21Lab Automation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-23Parts List - Cabinetry for ADVIA Centaur XP System . . . . . . . 1-29
2 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1
078D0080-02, Rev. A
Cuvette Waste AreaCuvette Waste Area Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 1-33Parts List – Cuvette Waste Area . . . . . . . . . . . . . . . . . . . . . . . . 1-35
Cuvette Loader 113339Cuvette Loader Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-37Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-38Parts List – Cuvette Loader 113339 . . . . . . . . . . . . . . . . . . . . . 1-41Replacing the Cuvette Chute Solenoid. . . . . . . . . . . . . . . . . . . 1-43
Removing the Cuvette Chute Solenoid . . . . . . . . . . . . . . . . . . . . . . . . 1-43Installing the New Cuvette Chute Solenoid Assembly . . . . . . . . . . . . . 1-44Verifying the Operation of the Cuvette Chute Solenoid Using Diagnostics Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-44Verifying the Operation of the Cuvette Chute Solenoid Using Controller Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-45
Elevator 571023BCuvette Elevator PC Board Location . . . . . . . . . . . . . . . . . . . . 1-47Belt Clamp Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-48Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-49Parts List – Elevator 571023B . . . . . . . . . . . . . . . . . . . . . . . . . . 1-51
Incubation Chamber 571021B Incubation Chamber Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 1-53Wash Station Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-56Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-56TED Assembly Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-56Parts List – Incubation Chamber 571021B . . . . . . . . . . . . . . . . 1-59Replacing the Incubation Chamber Bearings . . . . . . . . . . . . . 1-63Replacing the Index Pin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-67Replacing the Incubation Chamber TEDs . . . . . . . . . . . . . . . . 1-69
Servo Motor (Quad B) or Wash Station Area (Quad A) . . . . . . . . . . . . 1-69Primary Reagent Compartment (Quad D) or Back Area (Quad C) . . . 1-71
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 3
078D0080-02, Rev. A
Index Mechanism 571024BIndex Mechanism Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-75Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-75Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-76Parts List – Index Mechanism 571024B . . . . . . . . . . . . . . . . . . 1-79
Luminometer 078-K122-01Luminometer Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-81Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-82Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-83Parts List – Luminometer 078-K122-01. . . . . . . . . . . . . . . . . . . 1-85Replacing the PMT TED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-89Replacing the PMT Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 1-95
Preheater 078-B059-01Preheater Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-99Cuvette Pusher Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-100Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-101Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-101TED Assembly Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-102Parts List – Preheater 078-B059-01. . . . . . . . . . . . . . . . . . . . . 1-105Replacing the Preheater TEDs . . . . . . . . . . . . . . . . . . . . . . . . 1-107
Ring Loader 572400BRing Loader Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-111Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-112Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-113Parts List – Ring Loader 572400B. . . . . . . . . . . . . . . . . . . . . . 1-115
Servo Motor 571730B (Old)Servo Motor Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-117Latch Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-118
4 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1
078D0080-02, Rev. A
Servo Motor (Old) Detail 571730B . . . . . . . . . . . . . . . . . . . . . . 1-118
Servo Motor 571022B (New)Servo Motor (New) Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 1-121Latch Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-122Servo Motor (New) Detail 571022B . . . . . . . . . . . . . . . . . . . . . 1-123Parts List – Servo Motor 571730B and 571022B . . . . . . . . . . 1-125
Wash Station 571025DWash Station Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-127Acid Probe Assembly Detail (572889). . . . . . . . . . . . . . . . . . . 1-128Friction Pad Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-129Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-129Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-130Parts List – Wash Station 571025D. . . . . . . . . . . . . . . . . . . . . 1-133
Electronics Module (fixed configuration) 121911DElectronics Module Assembly. . . . . . . . . . . . . . . . . . . . . . . . . 1-135Backplane PC Board 78-B047-01 (Fixed Configuration) . . . . 1-136RT SPARC I/O PC Board 575267 . . . . . . . . . . . . . . . . . . . . . . . 1-137Fixed Electronics Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-138
Electronics Module (slide out configuration) 571115BElectronics Module Assembly. . . . . . . . . . . . . . . . . . . . . . . . . 1-139BackPlane PC Board 575252F (Slide Out) . . . . . . . . . . . . . . . 1-140Slide Out Electronics Module . . . . . . . . . . . . . . . . . . . . . . . . . 1-141Parts List – Electronics Module 571115B and 121911D . . . . 1-143Replacing the Real Time SPARC PC Board . . . . . . . . . . . . . . 1-145Replacing the Real Time SPARC PC Board . . . . . . . . . . . . . . 1-147Replacing the User Interface SPARC PC Board (pre FCO 15) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-149
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 5
078D0080-02, Rev. A
User Interface Module 078-0103-XX for the ADVIA Centaur System
User Interface Module Assembly . . . . . . . . . . . . . . . . . . . . . . 1-151Parts List - User Interface Module 078-0103-XX* for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-153
User Interface Module 078-0330-XX for the ADVIA Centaur XP System
User Interface Module Assembly . . . . . . . . . . . . . . . . . . . . . . 1-155Parts List - User Interface Module 078-0330-XX* for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-157Replacing the Hard Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . 1-159
Removing the UI Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-159Removing the Hard Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-159Installing the New Hard Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-161Installing the UI Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-162
Application Module 078-0331-XX for the ADVIA Centaur System
Application Module Assembly. . . . . . . . . . . . . . . . . . . . . . . . . 1-163Parts List – Application Module 078-0331-XX* for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-165
Power SupplyPower supply 571258D (slide out electronics module). . . . . 1-167Power Supply 078–B047–01 (fixed electronics module). . . . 1-168Parts List – Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-169Replacing the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . 1-171
Thermal Control PC Module for the ADVIA Centaur System S575257E
Thermal Control PC Module Diagram. . . . . . . . . . . . . . . . . . . 1-173Parts List – Thermal Control PC Module for the ADVIA Centaur System 575257E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-175
6 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1
078D0080-02, Rev. A
Thermal Control PC Module for the ADVIA Centaur XP System XP 078-B109-XX
Thermal Control PC Module Centaur XP Diagram. . . . . . . . . 1-177Parts List – Thermal Control PC Module for the ADVIA Centaur XP System 078-B109-XX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-179Replacing the Thermal Control Module or the PCB . . . . . . . 1-181
Reagent Fluidics Module for the ADVIA Centaur System 078-B028-01
Reagent Fluidics Module Assembly . . . . . . . . . . . . . . . . . . . . 1-183Diluter 2 mL 078-B061-01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-184Tri–Continent Diluter 078-B061-01 . . . . . . . . . . . . . . . . . . . . . 1-185Parts List – Reagent Fluidics Module 078-B028-01 . . . . . . . . 1-187
Reagent Fluidics Module for the ADVIA Centaur XP System 078-B028-01
Reagent Fluidics Module Assembly . . . . . . . . . . . . . . . . . . . . 1-189Diluter 2 mL 115401C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-190Tri-Continent Diluter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-191Parts List – Reagent Fluidics Module 078-B028-01 . . . . . . . . 1-193
System Fluids Drawer 078-B055-01 for the ADVIA Centaur System
System Fluids Drawer Assembly . . . . . . . . . . . . . . . . . . . . . . 1-195Wash Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-198Diluter Detail 115401C (2 mL) 115402C (5 mL). . . . . . . . . . . . 1-199Tri-Continent Diluter Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-199Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-201Parts List – System Fluids Drawer 078-B055-01 for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-203
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 7
078D0080-02, Rev. A
System Fluids Drawer 078-081-01 for the ADVIA Centaur XP System
System Fluids Drawer Assembly . . . . . . . . . . . . . . . . . . . . . . 1-207Wash Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-210
Diluter Detail 115401C (2 mL) 115402C (5 mL). . . . . . . . . . . . 1-211Tri-Continent Diluter Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-211Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-213Parts List – System Fluids Drawer 078-B081-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-215
Vacuum Pump 572703BVacuum Pump Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-219
Vacuum Pump 078-B050-02Vacuum Pump Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-221Parts List – Vacuum Pump 572703B. . . . . . . . . . . . . . . . . . . . 1-223
Waste and Water Drawer 078-B005-01 for the ADVIA Centaur System
Waste and Water Drawer Assembly . . . . . . . . . . . . . . . . . . . . 1-225Water Reservoir Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-226Waste Reservoir Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-227Water Pressure Pump Detail . . . . . . . . . . . . . . . . . . . . . . . . . . 1-228Caster Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-229Parts List – Waste and Water Drawer 078-B005-01 for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-231
Waste and Water Drawer 078-B089-01 for the ADVIA Centaur XP System
Waste and Water Drawer Assembly . . . . . . . . . . . . . . . . . . . . 1-235Waste Reservoir Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-236Waste Reservoir Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-237Water Pressure Pump Detail . . . . . . . . . . . . . . . . . . . . . . . . . . 1-238
8 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1
078D0080-02, Rev. A
Caster Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-239Parts List – Waste and Water Drawer 078-B089-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-241
Primary Reagent Compartment 571111CPrimary Reagent Compartment Assembly. . . . . . . . . . . . . . . 1-245Lead Screw Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-251Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-252Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-252TED Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-253Parts List – Primary Reagent Compartment 571111C. . . . . . 1-255Replacing the Primary Reagent Compartment Leadscrew. . 1-259Replacing the Primary Reagent Compartment TEDs . . . . . . 1-269
Primary Reagent Compartment Door 571104BPrimary Reagent Compartment Door Assembly . . . . . . . . . . 1-279Parts List – Primary Reagent Compartment Door 571104B . 1-281
Reagent Barcode Scanner 571103Reagent Barcode Scanner Assembly . . . . . . . . . . . . . . . . . . . 1-283Belt Clamp Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-284Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-284Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-285Parts List – Reagent Barcode Scanner 571103C. . . . . . . . . . 1-287
Reagent Probe Guide 571110BReagent Probe Guide Assembly . . . . . . . . . . . . . . . . . . . . . . . 1-289Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-291Parts List – Reagent Probe Guide 571110B . . . . . . . . . . . . . . 1-293
Reagent Probe 1 571107CReagent Probe 1 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-295
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 9
078D0080-02, Rev. A
Reagent Probe 1 Tubing Routing . . . . . . . . . . . . . . . . . . . . . . 1-297Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-297Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-297Parts List – Reagent Probe 1 571107C . . . . . . . . . . . . . . . . . . 1-301
Reagent Probe 2 571108CReagent Probe 2 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-303Reagent Probe 2 Tubing Routing . . . . . . . . . . . . . . . . . . . . . . 1-304Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-305Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-305Parts List – Reagent Probe 2 571108C . . . . . . . . . . . . . . . . . . 1-309
Reagent Probe 3 571109CReagent Probe 3 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-311Reagent Probe 3 Tubing Routing . . . . . . . . . . . . . . . . . . . . . . 1-312Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-313Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-313Parts List – Reagent Probe 3 571109C . . . . . . . . . . . . . . . . . . 1-317
Ancillary Queue for the ADVIA Centaur System 078-B035-01 Ancillary Queue Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-319Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-321Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-323Parts List – Ancillary Queue 078-B035-01 for the ADVIA Centaur System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-325
Ancillary Queue for the ADVIA Centaur XP System 078-B035-01
Ancillary Queue Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-327Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-329Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-331Parts List – Ancillary Queue 078-B069-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-333
10 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1
078D0080-02, Rev. A
Replacing the Ancillary Queue TEDs . . . . . . . . . . . . . . . . . . . 1-335
Ancillary Entry 078-B036-01Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-339Ancillary Entry Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-340Leadscrew Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-340Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-341Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-341Parts List – Ancillary Entry 078-B036-01 . . . . . . . . . . . . . . . . 1-345
Inprocess and Exit Queue 571040BMotor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-347Motor Gap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-348Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-349Parts List – Inprocess and Exit Queue 571040B . . . . . . . . . . 1-351
Sample and Ancillary Probe for the ADVIA Centaur System 078-B045-01
Ancillary Motor and Belt Location . . . . . . . . . . . . . . . . . . . . . 1-353Sample Motor and Belt Location. . . . . . . . . . . . . . . . . . . . . . . 1-355Sample Probe Assembly Detail . . . . . . . . . . . . . . . . . . . . . . . . 1-356
Vertical Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-356
Ancillary Probe Assembly Detail. . . . . . . . . . . . . . . . . . . . . . . 1-357Vertical Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-357
Sample Diluter Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-358Ancillary Diluter Sample Air Pump Assembly Detail. . . . . . . 1-358Diluter 2 mL 115401 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-359Tri–Continent Diluter 572888 . . . . . . . . . . . . . . . . . . . . . . . . . . 1-361Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-361
Sample and Ancillary Probe for the ADVIA Centaur XP System 078-B123-01
Ancillary Motor and Belt Location . . . . . . . . . . . . . . . . . . . . . 1-363
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 11
078D0080-02, Rev. A
Sample Motor and Belt Location. . . . . . . . . . . . . . . . . . . . . . . 1-365Sample Probe Assembly Detail . . . . . . . . . . . . . . . . . . . . . . . . 1-366
Vertical Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-366
Ancillary Probe Assembly Detail. . . . . . . . . . . . . . . . . . . . . . . 1-366Vertical Motor Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-366
Sample Diluter Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-368Ancillary Diluter Sample Air Pump Assembly Detail. . . . . . . 1-368Diluter 2 mL 115401 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-369Tri–Continent Diluter 572888 . . . . . . . . . . . . . . . . . . . . . . . . . . 1-371Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-371Parts List – Sample and Ancillary Probe 078-B123-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-373Parts List – Sample and Ancillary Probe 078-B045-01 . . . . . 1-377
Sample Entry Queue for the ADVIA Centaur System 078-B002-01
Ancillary Motor and Belt Location . . . . . . . . . . . . . . . . . . . . . 1-381Belt Tensioner Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-382Belt Drive Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-382Magnet Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-383Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-383Parts List – Sample Entry Queue for the ADVIA Centaur System 078-B002-01. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-387
Sample Entry Queue for the ADVIA Centaur XP System 078-B078-01
Ancillary Motor and Belt Location . . . . . . . . . . . . . . . . . . . . . 1-389Belt Tensioner Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-389Belt Drive Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-390Magnet Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-391Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-391Parts List – Sample Entry Queue for the ADVIA Centaur XP System 078-B078-01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-395
12 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1
078D0080-02, Rev. A
Sample Exit Pusher 571037BSample Exit Pusher Assembly . . . . . . . . . . . . . . . . . . . . . . . . 1-397Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-397Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-397Parts List – Sample Exit Pusher 571037B . . . . . . . . . . . . . . . 1-401
Sample Tip Loader 571036C for the ADVIA Centaur SystemSample Tip Loader Assembly . . . . . . . . . . . . . . . . . . . . . . . . . 1-403Sample Tip Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-404Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-405Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-405
Sample Tip Loader 078-B136-01 for the ADVIA Centaur XP System
Sample Tip Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-408Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-409Motor and Pulley Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-409Parts List – Sample Tip Loader . . . . . . . . . . . . . . . . . . . . . . . . 1-411
Sample Tip Waste Area for the ADVIA Centaur System 078-B0009-02
Sample Tip Waste Area Assembly . . . . . . . . . . . . . . . . . . . . . 1-413Parts List – Sample Tip Waste Area 078-B009-02 for the ADVIA Centaur System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-417
Sample Tip Waste Area for the ADVIA Centaur XP System 078-B088-01
Parts List – Sample Tip Waste Area 078-B088-01 for the ADVIA Centaur XP System . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-423
Stat Sample Pusher 571035Stat Sample Pusher Assembly . . . . . . . . . . . . . . . . . . . . . . . . 1-425Sensor Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-426
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 13
078D0080-02, Rev. A
Motor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-427Parts List – Stat Sample Pusher 078-B026-01 . . . . . . . . . . . . 1-429
Universal Rack OptionSample Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-431Stat Sample Pusher . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-432Sample Exit Pusher . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-433Inprocess Queue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-434Rack Sensor Pusher Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-435Sample Exit Queue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-436Sample Exit Queue Rack Present Sensor . . . . . . . . . . . . . . . 1-437Sample Exit Queue Full Sensor. . . . . . . . . . . . . . . . . . . . . . . . 1-438Sample Entry Queue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-439Sample Entry Queue Work Surface . . . . . . . . . . . . . . . . . . . . 1-440Sample Entry Queue Tensioner . . . . . . . . . . . . . . . . . . . . . . . 1-441Parts List – Universal Rack Option . . . . . . . . . . . . . . . . . . . . . 1-443
Miscellaneous PartsParts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-445
2 Kits and ToolsKits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
3 Centaur PC BoardsSample and Ancillary Probes Horizontal PC Board 575212C . 3-1Inprocess Queue Tip Sensor PC Board 575217B . . . . . . . . . . . 3-2Ancillary Queue Sensor PC Board 575222B . . . . . . . . . . . . . . . 3-3Sample Probe Vertical Sensor PC Board for the ADVIA Centaur System 575227 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4Sample Probe Vertical Sensor PC Board for the ADVIA Centaur XP System 118365 . . . . . . . . . . . . . . . . . . . . . . 3-5Ancillary Entry PC Board 575232D . . . . . . . . . . . . . . . . . . . . . . 3-6
14 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1
078D0080-02, Rev. A
Rack Loaded Sensor PC Board 575237B . . . . . . . . . . . . . . . . . 3-8Entry Queue Sensor PC Board 575242B . . . . . . . . . . . . . . . . . . 3-9Low Level Controller PC Board 078-B042S01 . . . . . . . . . . . . . 3-10Backplane PC Board 575252F (slide out electronics module) 3-16Backplane PC Board S117937D (fixed electronics module). . 3-21Thermal Control Module PC Board for the ADVIA Centaur System S575257E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26Thermal Control Module PC Board for the ADVIA Centaur XP System 078-B111SXX. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29Servo Low Level Controller PC Board S575262C . . . . . . . . . . 3-32Fluid GIOB PC Board S575272E . . . . . . . . . . . . . . . . . . . . . . . . 3-34Wash Fluid Interconnect PC Board 575277B . . . . . . . . . . . . . . 3-36Ancillary Probe Vertical PC Board S575282C . . . . . . . . . . . . . 3-37Reagent Handling GIOB PC Board S575287C . . . . . . . . . . . . . 3-38Rack Present PC Board S575292B . . . . . . . . . . . . . . . . . . . . . . 3-40Reagent Fluid GIOB PC Board 078-B027S02 . . . . . . . . . . . . . . 3-41Exit Queue Interconnect PC Board 575302C . . . . . . . . . . . . . . 3-43Stat Area PC Board 575307C. . . . . . . . . . . . . . . . . . . . . . . . . . . 3-44Sample and Ancillary Probe GIOB PC Board S575322H. . . . . 3-45Sample and Ancillary Queue GIOB PC Board S575327D . . . . 3-47Cuvette Elevator PC Board 575337 . . . . . . . . . . . . . . . . . . . . . . 3-49Top Luminometer PC Board 575342C . . . . . . . . . . . . . . . . . . . 3-50Reagent Probe Pack-to-Pack PC Board 575347 . . . . . . . . . . . 3-51Sample Tip Loader PC Board 575352B. . . . . . . . . . . . . . . . . . . 3-52Servo Motor Interface PC Board 575357C . . . . . . . . . . . . . . . . 3-53Aspirate Probe PC Board 575362 . . . . . . . . . . . . . . . . . . . . . . . 3-54Bottom Luminometer PC Board 575367 . . . . . . . . . . . . . . . . . . 3-56Preheater PC Board 575372. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-57Cuvette GIOB PC Board S575377F . . . . . . . . . . . . . . . . . . . . . . 3-58Reagent Pack LED Array PC Board 575387 . . . . . . . . . . . . . . . 3-61Ancillary Entry LEDs PC Board 575392B . . . . . . . . . . . . . . . . . 3-62Tip Tray and Band PC Board S575397C. . . . . . . . . . . . . . . . . . 3-63
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 15
078D0080-02, Rev. A
Reagent Probe 1 Heater Control and Volume Detect Probe Interconnect PC Board 575402B . . . . . . . . . . . . . . . . . . . . . . . . 3-64Reagent Probe 2 Heater Control and Volume Detect Probe Interconnect PC Board 575407C . . . . . . . . . . . . . . . . . . . . . . . . 3-65Reagent Probe 3 Heater Control and Volume Detect Probe Interconnect PC Board 575412C . . . . . . . . . . . . . . . . . . . . . . . . 3-66Mixer Interconnect PC Board 575422C. . . . . . . . . . . . . . . . . . . 3-67Reagent Probe 1 Horizontal Interconnect PC Board 575432B 3-68Reagent Probe 2 Horizontal Interconnect PC Board 575437 . 3-69Reagent Probe 3 Horizontal Interconnect PC Board 575442B 3-70Reflective Chute Sensor Assembly PC Board 575447B . . . . . 3-71System Fluid Detect PC Board 575452C . . . . . . . . . . . . . . . . . 3-72Ancillary Pack Sensor PC Board 575457B. . . . . . . . . . . . . . . . 3-74Sample Pressure Level Sense PC Board 575462E . . . . . . . . . 3-75
Appendix A: Safety InstructionsProtecting Yourself from Biohazards . . . . . . . . . . . . . . . . . . . . . A-1
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Protecting Yourself from Barcode Scanner Lasers . . . . . . . . . A-2Laser Safety Classification of the ADVIA Centaur System. . . . . . . . . . .A-2Laser Safety Classification of the Barcode Scanners. . . . . . . . . . . . . . .A-3Reagent and Sample Barcode Scanners . . . . . . . . . . . . . . . . . . . . . . . .A-3Hand-held Barcode Scanner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
Appendix B: Service, Ordering, and WarrantyBayer Authorized Representative. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1Limited Instrument Warranty and Service Delivery Policy . . . . . . . . . . .B-1Warranty Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Information for Technical Assistance. . . . . . . . . . . . . . . . . . . . . B-4Printing the System Serial Number and Versions. . . . . . . . . . . . . . . . . .B-5Printing the Event Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5Printing Windows on the Workspace . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6Printing a Runtime Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6Printing Calibration Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6
Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6Bayer Authorized Representative. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6Addresses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6
078D0080-02, Rev. A
1 Parts, Parts Lists, Replacement Procedures
Cabinetry for the ADVIA Centaur System
System Overview - Front
1-2 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Overview – Back
078D0080-02, Rev. A
System Overview – Back
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 A/C Input Line Assembly 1-3
078D0080-02, Rev. A
A/C Input Line Assembly
1-4 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Top Cover
078D0080-02, Rev. A
Top Cover
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Monitor Stand 1-5
078D0080-02, Rev. A
Monitor Stand
1-6 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Entry Queue
078D0080-02, Rev. A
Sample Entry Queue
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Queue 1-7
078D0080-02, Rev. A
Sample Exit Queue
1-8 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Electronics Module Door
078D0080-02, Rev. A
Electronics Module Door
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area Door 1-9
078D0080-02, Rev. A
Sample Tip Waste Area Door
1-10 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Cuvette Waste Area Door
078D0080-02, Rev. A
Cuvette Waste Area Door
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Cuvette Waste Area Door 1-11
078D0080-02, Rev. A
1-12 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lab Automation
078D0080-02, Rev. A
Lab Automation
078D0080-02, Rev. A
Parts List - Cabinetry for ADVIA Centaur SystemItem Description REF Part Number
System Overview - Front
System Overview - Back5c Cover, Cable UI 12157513 Corner, User Interface Cap 572534
Box, Stop Switch 078-0111-01Cover, Cable Access 572381
Top Cover
Monitor Stand Assembly1 Clamp, Monitor 5513292 5727612 Holder, Barcode Scanner 1156703 Screw, Flat M4 x 0.7 x 12 mm 5717814 Handle, UI 5727575 Clamp, Keyboard 078–0100–016 Palmrest Assembly 1109578 078–B058–01
Glidepoint and Cable 8957558 078-0236-01External Glidepoint and Cable 478537 111215
7 Screw, Thumb M6 x 1.0 x 30 mm 1128648 Nut, Knurled M6 x 1.0 x 15 mm 572851
Entry Queue33 Cover, Entry Queue 5585684 57254134 Housing, Stat Area 572535C
PC Board, Ancillary Entry LEDs 1765254 575392B35 Tray, Entry Queue 078-0006-0136 Work Surface, Entry Queue 4253513 078-B001-0137 Panel, Entry Queue, Back Cover 11021438 Cover, Entry Queue Back 571996B
Electronic Module Door12 Latch, Adjustable Grip 57252513 Retainer 72426B14 Bracket, Door Hinge 572524
1-14 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List - Cabinetry for ADVIA
078D0080-02, Rev. A
15 Hinge, Door 57237016 Door, Electronics Compartment 572365D17 Adapter, Monitor Clamp 078–0075–0118 Screw, M4x.8 Phillips 9843467 60007x5519 Access Panel, Reagent Probes 9132080 57215520 Switch, Emergency Stop (red) 57215821 Switch, Computer Startup (green) 57251622 Filter, Air 3918457 572165
Retainer, Filter, Air 572164
Lab Automation23 Screw, Truss M5 x 12 mm 11710124 Cover, Rear Sample, Lab
Automation115867
25 Panel, Upper Back, Lab Automation
115866
26 Cover, Window, Lab Automation 115868
Exit Queue27 Work Surface, Exit Queue 57198528 Cover, Exit Queue Rack 571963C29 Cover, Exit Queue 5625414 57254230 Tray, Exit Queue 57100431 Bracket, Midlevel Front 110249B32 Block, Queue Cover Hinge 572537B
Cover, Back 078-0118-01
Sample Tip Waste Area39 Door, Sample Tip Waste 57236640 Handle, Pocket Pull41 Hinge, Upper Leaf 57237442 Hinge, Single Bolt 57237643 Hinge, Single Bolt 57237744 Hinge, Lower Leaf 57237545 Catch, Magnetic 572368
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List - Cabinetry for ADVIA Centaur
078D0080-02, Rev. A
Bracket, Magnet Mounting 11061146 Housing, Interior Tip Waste 572178
Cuvette Waste Area Door47 Housing, Cuvette Waste 572363B48 Bracket, Retaining 110408B49 Handle, Pocket Pull 57236950 Catch, Twin Ball 57253051 Support, Sample Entry Queue 57207952 Door, Cuvette Waste 57236453 Line Assy, A/C Input 572271B54 Not Used55 Hinge, Upper Leaf 57237256 Hinge, Single Bolt 57237757 Hinge, Single Bolt 57237658 Hinge, Lower Leaf 57237359 Bin, Tip Tray Waste 57242560 Cover, Fluid Module Access 572408
Bracket, Fluid Module Access Cover, Right
572409
Bracket, Fluid Module Access Cover, Left
572410
Fluids Bracket, Access Panel 57216061 Top Cover 12147062 Window, Diluter 57260563 Window, Front 572604B64 Cover, Sample Tip Loader 572608B65 Window, Reagent Probes 57260766 Window, Top 078–B038–0167 Handle, Top Window 572619
9 Shield, Air Deflector, Lab Automation 7072218 11715910 Washer, Flat M6 846-1021-0911 Cover, Ancillary Queue 572531
Item Description REF Part Number
078D0080-02, Rev. A
Cabinetry for the ADVIA Centaur XP System
System Overview - Front
1-18 header Melias Operating Manual
footer We never used this.....
System Overview - Back
A/C Input Line Assembly
Top Cover
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Monitor Stand 1-19
078D0080-02, Rev. A
Monitor Stand
Sample Entry Queue
1-20 header Melias Operating Manual
footer We never used this.....
Sample Exit Queue
Electronics Module Door
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Corner Door 1-21
078D0080-02, Rev. A
Corner Door
1-22 header Melias Operating Manual
footer We never used this.....
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lab Automation 1-23
078D0080-02, Rev. A
Lab Automation
1-24 header Melias Operating Manual
footer We never used this.....
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lab Automation 1-25
078D0080-02, Rev. A
1-26 header Melias Operating Manual
footer We never used this.....
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lab Automation 1-27
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Cabinetry for ADVIA Centaur XP System
Item Description REF Part Number
System Overview - Front1 Cover, Ancillary Queue 078-B091-012 Cover, Fluid Module Access 5724083 Box, Stop Switch 078-0371-01
Bracket, Fluid Module Access Cover, Right
572409
Bracket, Fluid Module Access Cover, Left
572410
Fluids Bracket, Access Panel 572160
System Overview - Back4 Switch, Computer Startup (green) 5725165 Cover, Service Back access 078-0313-016 Screw 60012x27 Filter, Air 3918457 5721658 Retainer, Air Filter 078-0315-019 Line Assy, A/C Input 572271B10 Cover, Right Back 078-0312-0110 Cover, Back Access, Electronics Module 078-0314-0112 Cover, Cable UI 078-0422-0113 Cover, Cable Access 078-0372-0114 Shield, Air Deflector, Lab Automation 7072218 11715915 Corner, User Interface Cap 078-0369-0116 Switch, Emergency Stop (red) 572158
Top Cover8 Cover Top, Rear 078-0347-019 Cover Top, Front 078-0348-01
Window, Top Access 078-0349-01
Monitor Stand Assembly 078-B122-01
10 Assembly, Arm not orderable separately
11 Clamp, Cable 509-0118-0112 Holder, Barcode Scanner 078-0457-0113 Nut, M4 Acorn Hex 543-0173-0814 Tray, Keyboard 078-0407-0115 Touchscreen 515-7247-01
Keyboard See Miscellaneous Section
Entry Queue16 Cover, Entry Queue 5585684 57254117 Housing, Stat Area 078-0368-0118 PC Board, Ancillary Entry LEDs 1765254 575392B19 Tray, Entry Queue 078-0006-0120 Work Surface, Entry Queue 9169294 078-0302-0121 Panel, Entry Queue, Back Cover 11021422 Cover, Entry Queue Back 078-0384-02
Electronic Module Door 078-0382-0123 Latch, Adjustable Grip 57252524 Bracket, Door Hinge 57252425 Hinge, Door 57237026 Door, Electronics Compartment 078-0382-0120 Screw, M4x.8 Phillips 9843467 60007x55
Lab Automation24 Screw, Truss M5 x 12 mm 11710125 Cover, Rear Sample, Lab Automation 11586726 Panel, Upper Back, Lab Automation 11586627 Cover, Window, Lab Automation 115868
Exit Queue28 Work Surface, Exit Queue 078-0366-01
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Cabinetry for
078D0080-02, Rev. A
29 Cover, Exit Queue 5625414 57254230 Tray, Exit Queue 57100431 Bracket, Midlevel Front 110249B32 Block, Queue Cover Hinge 572537B 572537B
Cover, Back 078-0365-01
Sample Tip Waste Area39 Door, Sample Tip Waste 57236640 Handle, Pocket Pull41 Hinge, Upper Leaf 57237442 Hinge, Single Bolt 57237643 Hinge, Single Bolt 57237744 Hinge, Lower Leaf 57237545 Catch, Magnetic 572368
Bracket, Magnet Mounting 11061146 Housing, Interior Tip Waste 572178
Cuvette Waste Area Door 078-B094-0146 Cover, Corner Door 078-0308-0147 Magnet Mount, door 078-0453-0148 Catch Plate, Corner Door 078-0452-0149 Magnet, Shielded 539-0010-0150 Screw, Countersink, Flat 846-1022-3451 Bin, Tip Tray Waste 572425
I THINK THE FOLLOWING ARE NOT NEEDED
62 Window, 57260563 Window, Front 572604B64 Cover, Sample Tip Loader 572608B65 Window, Reagent Probes 57260767 Handle, Top Window 572619
Item Description REF Part Number
078D0080-02, Rev. A
Cuvette Waste Area
Cuvette Waste Area Assembly
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Cuvette Waste Area Assembly 1-34
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Cuvette Waste AreaItem Description REF Part Number1 Collar, Cuvette Waste Bin 06907707 1101062 Screw, Retractable, M4 x 0.7 x 5.7 mm 00220165 5719643 Shutter, Waste Reservoir 5726224 Chute, Cuvette Waste 156351 5725445 Sensor, Optical Switch 00962374 1116746 Bin, Cuvette Waste 06412376 5726157 Liner, Cuvette Waste 0764029 078-0067-018 Clamp, Spring 04324275 1102429 Screw, Pan-Head M4 x 0.7 x 10 mm 02968655 60000x6510 Spring, Waste Bin Interlock 04215735 11502311 Not Used12 Screw, Pan-Head, M4 x 0.7 x 6 mm 60000x3513 Washer, Flat, M4 04563393 60028x514 Reservoir, Waste Chute 09580326 572547
078D0080-02, Rev. A
Cuvette Loader 113339
Cuvette Loader Assembly
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-38
078D0080-02, Rev. A
Motor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-39
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Cuvette Loader 113339Item Description REF Part Number1 Clamp, Torque Transfer 396069 265862 Cam, Wheel 264903 Bearing, Ball, 0.250 x 0.500 x 0.187 911338 205734 Ring, Retaining, E-ring, 0.250D 5785519 261075 Support, Cuvette Loader 03674086 571727B6 Spacer, 0.253 x 0.373 x 0.010 06373893 128617 Bearing, Journal, Flanged 0.250 x 0.375 x
0.2506068713 26082x7
8 Cam, Arm 7015400 1127019 Screw, Cap-Head M4 x 0.7 x 55 mm 60012x10710 Screw, Cap-Head M4 x 0.7 x 9 mm 00593158 60012x9
11 Washer, Lock 06913146 60026x512 Sensor, PC Board Chute Full 04162038 575447B13 Kit, Cuvette Chute Solenoid 05384913 12039014 Tube, Drop 01353061 57206915 Screw, Cap-Head M4 x 0.7 x 12 mm 01876943 60012x2316 Chute, Orientation 09678857 2714517 Fastener 01662811 27186x218 Gate, Lift, Cuvette 03536260 27187B19 Wedge, Cuvette Loader 09725693 2732520 Screw, Button-Head M4 x 0.7 x 6 mm 7904256 57250021 Ionizer, Air 09864405 57226822 Bracket, Ionizer Support 57241323 Bin, Cuvette Loader 572071B24 Block, Mixer 11270225 Screw, Cap-Head M4 x 0.7 x 16 mm 02345038 60012x3726 Block, Slotted 11270327 Shaft, Idler Sprocket 2637328 Support, Cuvette Loader 03674086 571727B29 Screw, Cap-Head M4 x 0.7 x 50 mm 60012x10030 Washer, Flat, M4 04563393 60028x531 Motor, Stepper, 6VDC 03828563 118498
1-42 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Cuvette Loader 113339
078D0080-02, Rev. A
32 Bushing, Sprocket, Mounting 2644433 Screw, Set M4 x 0.7 x 04 mm 60023x4034 Sprocket, 15T, 0.500P, Single Strand 2637635 Screw, Set 1/4-20 x 0.25 2657336 Spacer, Stepped 2650937 Block, Chain, Tension Adjuster 11668438 Screw, Pan-Head M4 x 0.7 x 20 mm,
Slotted60003x95
39 Spacer, Stepped 2650740 Bearing, Flanged, Journal 0.251 x 0.377 x
0.375, Bronze26572
41 Chain, Cuvette, Lift 2674642 Guide, Chain 26495
43 Support, Hopper, Bin, Back 2657044 Support, Feeder, Angled 2650545 Bearing, Clutch, One-Way 0.250 x 0.438 x
0.50019556
46 Partition, Cuvette Bin 9759016 123281
Item Description REF Part Number
078D0080-02, Rev. A
Replacing the Cuvette Chute Solenoid1. At the workspace, select System Status.
2. Select Turn Mechanics Off, and then wait while the system mechanics power down.
Removing the Cuvette Chute Solenoid
1. Disconnect the chute full sensor cable from J4 on the Preheater PC board.2. Disconnect the cuvette chute solenoid cable from the cable extending through the
opening at the back of the luminometer.
NOTE: Push the waste probe arm down to allow easier access to the cuvette chute solenoid assembly.
3. Remove the four screws and eight washers that secure the cuvette chute solenoid assembly to the orientation chute. Save the screws and washers.
4. Remove the cuvette chute solenoid assembly.
Figure 1-1 Removing the Cuvette Solenoid Assembly
5. Remove the two screws that secure the chute full sensor to the cuvette chute solenoid assembly. Save the screws and sensor, and then discard the cuvette chute solenoid assembly.
CAUTIONObserve precautions for handling electrostatic–sensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Cuvette Chute Solenoid
078D0080-02, Rev. A
Installing the New Cuvette Chute Solenoid Assembly1. Install the chute full sensor to the new cuvette chute solenoid assembly using the two
screws removed previously.
2. Install the new cuvette chute solenoid assembly to the orientation chute using the four screws and eight washers removed previously.
Figure 1-2 Installing the Cuvette Chute Solenoid
3. Connect the cable from the chute full sensor to J4 on the Preheater PC board.
Figure 1-3 Preheater PC Board
4. Connect the cable from the cuvette chute solenoid to the cable extending through the opening at the back of the luminometer.Route these cables so they do not interfere with the cuvette loader cam.
5. Select Turn Mechanics On.
Verifying the Operation of the Cuvette Chute Solenoid Using Diagnostics ToolsUse this procedure to cycle the cuvette chute solenoid in diagnostics tools.
1. At the workspace, select System Status.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Cuvette Chute Solenoid
078D0080-02, Rev. A
2. Select Diagnostics Tools.
3. At the System-Diagnostics Tools window, select Empty & Fill Ring.
4. Select Perform.
5. As the system performs the empty & fill ring, watch the cuvettes as they slide down the orientation chute and enter the vertical cuvette chute. Ensure that the cuvette chute solenoid cycles as a cuvette loads into the incubation ring.This process takes approximately 12 minutes.
6. Close the System-Diagnostics Tools window.
Verifying the Operation of the Cuvette Chute Solenoid Using Controller DiagnosticsNOTE: Use this procedure if the Empty & Fill Ring procedure in Diagnostics Tools fails.
Use this procedure to cycle the cuvette chute solenoid in controller diagnostics.
1. At the workspace, select System Status.
2. Select Controller Diagnostics.
3. Insert the Controller Diagnostics diskette into the diskette drive.
4. Select Diskette in Drive.
5. If your Controller Diagnostics diskette has the cuv_sol sequence, use the following procedure to run the cuv_sol sequence:a. Select the cuv_sol sequence.b. Select Sequence Definition.c. Select Run Sequence.
WARNINGBefore selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury.
WARNINGBefore closing the System-Diagnostic Tools window, ensure that you are clear of subassemblies that can move. The system returns subassemblies to their home positions and can cause injury.
CAUTIONWait for the system to access the controller diagnostics sequences on the diskette. If you select the diskette drive twice, damage can occur to the diskette and cause an RT error.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Cuvette Chute Solenoid
078D0080-02, Rev. A
Verify that the cuvette chute solenoid cycles five times.
078D0080-02, Rev. A
Elevator 571023B
Cuvette Elevator PC Board Location
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Belt Clamp Detail 1-48
078D0080-02, Rev. A
Belt Clamp Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-49
078D0080-02, Rev. A
Motor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-50
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Elevator 571023BItem Description REF Part Number1 Screw, Pan-Head M4 x 0.7 x 12 mm
Phillips60013x23
2 PC Board, Cuvette Elevator 5753373 Screw, Cap-Head M4 x 0.7 x 16 mm 60013x374 Motor, Stepper 9864200 5729155 Screw, Retractable M5 x 0.8 x 10.5 mm 5720526 Washer, Flat M3 60056x37 Screw, Cap-Head #4-40 x 0.375 1184478 Guide, Linear 5720649 Screw, Cap-Head M2 x 0.4 x 8 mm 57241810 Pulley, Idler, 0.080P 25T 0.290W 0.432B 572046
11 Screw, Shoulder M4 x 0.7 x 14 mm 27014x712 Belt 0.08P 0.25W 190T 371015 57204413 Carriage, Cuvette Elevator 57188714 Clamp, Belt 57204515 Screw, Cap-Head M4 x 0.7 x 8 mm Phillips 60013x916 Pulley, 0.08P 20T 0.290W 0.188B 57204917 Screw, Cap-Head M2 x 0.4 x 10 mm 57241918 Pusher, Cuvette Elevator 57205019 Flag, Elevator EOT 572051
078D0080-02, Rev. A
Incubation Chamber 571021B
Incubation Chamber Assembly
1-54 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Incubation Chamber Assembly
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Incubation Chamber Assembly 1-55
078D0080-02, Rev. A
1-56 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Wash Station Cover
078D0080-02, Rev. A
Wash Station Cover
Sensor Detail
TED Assembly Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Assembly Detail 1-57
078D0080-02, Rev. A
1-58 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Assembly Detail
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Incubation Chamber 571021BItem Description REF Part Number1 Cover, Chamber Lumo Area 571197B2 Cover, Chamber Back Area 571198B3 Screw, Cap-Head M4 x 0.7 x 30 mm 60013x724 Cover, Chamber Servo Area 5711925 Cover, Chamber Sample Area 571193B6 Chute, Luminometer Entrance 571852B7 Cover, Luminometer Entrance 5319631 571854B8 Cover, Chamber Wash Separation Area 571194
Screw, Cap Head M4 x 70 mm 60013x1289 Magnet, Resuspension 11546310 Window, Incubation Chamber 57119511 Screw, Flat-Head, Phillips M4 x 0.7 x 5
mm60007x35
12 Cover, Chamber Index Area 57120013 Cover, Chamber R3 Area 57119614 Cover, Reagent Dispense Access 7205765 57120115 Screw, Retractable M4 x 0.7 x 8 mm 57153316 Screw, Cap-Head M4 x 0.7 x 65 mm 60013x12817 Screw, Cap-Head M4 x 0.7 x 16 mm 60013x3718 Screw, Flat-Head M3 x 0.5 x 20 mm
Phillips60001x93
19 Cover, Cuvette Ring Sensor Housing 57238020 Sensor, Optical Switch, Ring Tooth 571256C21 Housing, Cuvette Ring Sensor 57237922 Screw, Hex M4 x 0.7 x 10 mm 60013x1623 Ring, Cuvette Assembly 4450912 111676
Insulation 571251Activator, Loctite 111291Quick Metal, Loctite 111290Plate, Retainer 571253Bracket, Belt 771641 571187Plate, Magnet Mounting 115464B
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Incubation Chamber
078D0080-02, Rev. A
Magnet, Resuspension 115463Screw, Flat-Head, Phillips M4 x 0.7 x 5 mm
60007x35
Strip, Non-Reflective 111604Spring, Compression 571511Screw, Cap-Head M4 x 0.7 x 8 mm 60013x9Screw, Phillips M2.5 x 0.4 x 5 mm 571532Screw, Set 571531Pulley, Belt, Timing 1453686 571186Screw, Cap-Head M3 x 0.5 x 8 mm 60013x8
24 Pin, Index Detent 8232340 57118825 Spring, Compression 57151126 Screw, Phillips M2.5 x 0.4 x 5 mm 57153227 Bearing, Cuvette Ring 7285122 57117528 Screw, Set M5 x 0.8 x 8 mm 57153129 Screw, Cap-Head M3 x 0.5 x 8 mm 60013x830 Plate, Retainer 57125331 Pulley, Timing Belt 1453686 57118632 Screw, Flat-Head 60007x3533 Screw, Button 57250034 Not Used N/A35 Not Used N/A36 Plate, Magnet Mounting 115464B37 Screw, Cap-Head M4 x 0.7 x 8 mm 60013x938 Bracket, Belt 771641 57118739 Washer, Flat M5 60056x640 Screw, M3 x 0.5 x 12 mm 60013x2241 Sensor, Cuvette Present 7333968 11167242 Bracket, Cuvette Present Sensor 11307043 Screw, Cap-Head M3 x 0.5 x 8 mm 60007x5344 Gasket, Seal, Cuvette Elevator 57120245 Gasket, O-ring 57152146 Screw, Cap-Head M3 x 0.5 x 10 mm 60013x15
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Incubation Chamber
078D0080-02, Rev. A
47 Sensor, Optical Switch 7713183 11167148 Spacer 3.2 mm x 6.0 mm x 3.0 mm 57152949 Screw, Cap-Head M5 x 0.8 x 25 mm 60013x6650 Stop, Magnet Ring 57117751 O-ring, Metric 4 mm D x 15 mm wide 57152052 Fitting, M6 to 0.125 Barb 11197053 Block, TED Mounting 57118254 TED, Assembly 5483539 111677
Thermal Grease 12049155 Thermistor, Threaded Body 1849083 571184C
Shim 557-7282-01
Item Description REF Part Number
078D0080-02, Rev. A
Replacing the Incubation Chamber Bearings1. Install the set screws:
a. Locate the bearing holes on the magnet ring.b. Screw the set screw until the tip of the screw is flush with the outside of the magnet
ring.c. Repeat steps a and b for the five remaining bearing holes.
Figure 1-1 Incubation Ring Set Screw
2. Place the 6 bearings on the magnet ring.
Figure 1-2 Incubation Ring Bearings
3. Tilt the cuvette ring to place it on top of the bearings and the magnet ring.
1-64 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber
078D0080-02, Rev. A
Figure 1-3 Incubation Ring
4. Using a small screw driver, position the six bearings on the magnet ring until they align with the six set screws installed previously.
5. Install the screws into the bearings:a. Position the cuvette ring on top of the bearings.b. Partially install two screws into the bearing. Ensure the cuvette ring moves
without interference from the top of the screw.c. Repeat steps a and b for each bearing.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber
078D0080-02, Rev. A
Figure 1-4 Incubation Ring Screws
6. Adjust the set screws:a. Locate the six set screws that were installed previously.b. Adjust each set screw 1/4 turn clockwise.c. Rotate the cuvette ring and check for excessive side–to–side movement or
binding.d. If you feel side–to–side movements, adjust the set screws clockwise. If you feel
binding, adjust the set screws counterclockwise.
CAUTIONTo ensure consistent bearing tension, adjust all of the set screws equally each time you perform this procedure.
078D0080-02, Rev. A
Replacing the Index Pin1. Replace the index pin:
a. Locate the index pin hole on the cuvette ring assembly.You can view the index pin hole from the top of the incubation chamber through the inside ring opening. The hole is located on the magnet ring approximately 180 degrees from the incubation chamber drive belt clamp.
Figure 1-1 Index Pin Hole on Magnet Ring
b. Insert the compression spring and index pin into the index pin hole ensuring the pin opening is facing the outside of the ring.
Figure 1-2 Compression Spring
c. While pressing down on the index pin, insert the retainer through the outside of the magnet ring into the index pin.
1-68 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Index Pin
078D0080-02, Rev. A
d. Using the two screws, secure the retainer to the magnet ring.
Figure 1-3 Index Pin
078D0080-02, Rev. A
Replacing the Incubation Chamber TEDs
Servo Motor (Quad B) or Wash Station Area (Quad A)1. Turn off the system:
a. At the workspace, select System Status.b. Select Turn System Off.c. At the back of the system, turn off the power and unplug the power cord.
2. Remove the servo motor or the wash station assembly.
3. Remove the cuvette loader.
4. Remove the defective TEDs:a. Remove the 2 screws that secure the heatsink to the TED mounting block.
Figure 1-1 Heatsink Screws
b. Remove the 2 screws that secure the TED mounting block to the bottom of the incubation ring.
1-70 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber
078D0080-02, Rev. A
Figure 1-2 TED Mounting Block
c. Remove the TED mounting block.d. Remove the defective TEDs.
5. Install the TEDs:a. Apply a thin layer of thermal grease to both sides of the new TEDs.b. Install the new set of TEDs on the bottom of the incubation chamber ensuring the
red (R) and black (B) wires are in the same position as the old TEDs.
c. Install the TED mounting block and finger tighten the 2 screws.d. Install the 2 screws that secure the heatsink to the TED mounting block.
e. Tighten the 2 screws 0.9 N–m (8 in–lb) on the TED mounting block.f. Connect the cable for the TEDs.
6. Install the servo motor or the wash station assembly.
7. Calibrate the incubation chamber thermals.
CAUTIONEnsure the TEDs are installed flush with the bottom of the incubation ring or damage can occur to the TEDs.
CAUTIONFailure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber TEDs
078D0080-02, Rev. A
Primary Reagent Compartment (Quad D) or Back Area (Quad C)1. Turn off the system:
a. At the workspace, select System Status.b. Select Turn System Off.c. At the back of the system, turn off the power switch and unplug the power cord.
2. Remove the luminometer and the preheater assemblies.
3. Remove the cuvette loader.
4. Remove the defective TEDs:a. Remove the 2 screws that secure the heatsink to the TED mounting block.
Figure 1-3 Removing the Heatsink
b. Remove the 2 screws that secure the heatsink to the base of the system, and then remove the heatsink.
c. Remove the 2 screws that secure the TED mounting block.
1-72 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber
078D0080-02, Rev. A
Figure 1-4 Removing the TED Mounting Block
d. Remove the defective TEDs.
5. Install the TEDs:
a. Apply a thin layer of thermal grease to both sides of the new TEDs.b. Install the new set of TEDs on the bottom of the incubation chamber with the red
(R) and black (B) wires in the same position as the old TEDs.
c. Install the TED mounting block and finger tighten the 2 screws.d. Connect the cable for the TEDs.
6. Install the heat sink:a. Install the heatsink and tighten the 2 screws that secure the heatsink to the base of
the system.b. Install the 2 screws that secure the heatsink to the TED mounting block.c. Remove the 2 screws that secure the heatsink to the TED mounting block.
CAUTIONEnsure that no thermal grease is inside the TED or heating problems will occur.
CAUTIONEnsure the TEDs are installed flush with the bottom of the incubation ring or damage can occur to the TEDs.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Incubation Chamber TEDs
078D0080-02, Rev. A
d. Remove the 2 screws that secure the heatsink to the base of the system, and then remove the heatsink.
e. Tighten the 2 screws 0.9 N–m (8 in–lb) that secure the TED mounting block.f. Install the heat sink and tighten the 2 screws that secure the heatsink to the
Centaur baseplate.g. Install the 2 screws that secure the heatsink to the TED mounting block.
7. Install the luminometer and preheater assemblies.
8. Turn the system on:a. Plug in the power cord.b. At the back of the system, turn the power on.c. At the back of the system, press the computer startup switch.
9. Calibrate the incubation chamber thermals.
CAUTIONFailure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs.
078D0080-02, Rev. A
Index Mechanism 571024B
Index Mechanism Assembly
Sensor Detail
1-76 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail
078D0080-02, Rev. A
+
Motor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-77
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Index Mechanism 571024BItem Description REF Part Number1 Screw, Cap-Head M4 x 0.7 x 10 mm 60013x162 Motor, Stepper 4VDC 1184235 5729233 Shaft, Index Mechanism 5717434 Flag, EOT Sensor 5710065 Screw, Pan-Head M2.5 x 0.45 x 5 mm 60001x326 Foot, Index Plunger 9706931 234047 Housing, Index Mechanism 5710058 Sensor, Optical Switch 962374 1116749 Screw, Pan-Head, Self-Tap 571745
078D0080-02, Rev. A
Luminometer 078-K122-01
Luminometer Assembly
1-82 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
Sensor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-83
078D0080-02, Rev. A
Motor Detail
1-84 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Luminometer 078-K122-01Item Description REF Part Number1 Screw, Cap-Head M4 x 0.7 x 16 mm 60013x372 Belt, 0.080P .25W 175T Neoprene 8107910 5718223 Clip, Probe Retainer 4740554 571812B4 Pulley, Luminometer Rotor 5715485 Probe, Waste, Aspirate 1190170 571813B6 Wheel, Sensor 5711737 Screw, Pan-Head M2.5 x 0.4 x 5 mm
Phillips60001x32
8 Flag, EOT Sensor 5710069 PC Board, Luminometer, Top 9140830 575342C10 Screw, Cap-Head M4 x 0.7 x 6 mm 60013x2
11 Housing, Luminometer 571009CChute, Lumo Exit 571818BScrew, Cap-Head 60013x9
12 Rotor, LumoSegment, Lumo Rotor
571544B571010
Screw, Shoulder 27014x1013 Guide, Waste Probe 8947137 117109B14 Pulley, Timing, 0.080P 15T 0.290W 0.188 571821B15 Motor, Stepper 4VDC 571821B16 Latch, Over-Center Draw 571817
Screw, Cap- Head 572417Washer 60056x3
17 Motor, Stepper 4VDC 57291518 Gear, Spur 32P 24T 0.188W 0.187B Delrin 1116124 571815B19 Screw, Cap-Head M4 x 0.7 x 8 mm Phillips 60013x920 Plate, Motor Mounting 57181421 Screw, Cap-Head M3 x 0.5 x 5 mm Hex 57241722 Screw, Flat-Head #4-40 x 0.25 Phillips 57241523 Sensor, Optical Switch 962374 11167424 Spring, Comp 0.240 x 0.022 x 0.250 Stnl 57180725 Pad, Friction, Short 571808
1-86 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Luminometer
078D0080-02, Rev. A
26 Housing Assy, Waste 4805834 571030BHousing, Waste Probe 571805
27 Fitting, M6 to 3/16 OD 57153428 Tube, 0.250 x 0.375 Black Tygon29 Nut, Knurled M6 57152430 Cover, Top Luminometer 571719B
Clamp 12811x131 Screw, Flat M4 x 0.7 x 20 mm 60055x9532 PMT, Assembly 571029B33 O–Ring, 5.987x.103 Buna–N 57152734 Cover, Luminometer Top 6482145
Ball Bearing 571546Ring, Internal Retaining 9854647Screw, Cap-Head 60013x1
35 Retainer, Waste Probe Guide 11710836 TED, Assembly 111678
Thermal Grease 12049137 Gasket, Ted Mounting 2674038 Screw, Cap-Head M4 x 0.7 x 40 mm 60013x8639 Screw, Cap-Head M3 x 0.5 x 30 mm 60013x8640 Screw, Set M3 x 0.5 x 4 mm 57238541 PC Board, Luminometer Bottom 57536742 Screw, Pan-Head M4 x 0.7 x 12 mm
Phillips60013x23
43 Heatsink, PMT Ted 57180444 Kit, Base Probe Assembly 6102075 078-K081-01
Base Probe 078-0267-01Tubing, 75 in 572689Tubing, Heatshrink 5.5 in 26781Fitting, Βarbed 518-0883-01Fitting, 1/4-28 518-7076-01O-Ring 116828Teflon Tape 19720
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Luminometer 078-K122-01
078D0080-02, Rev. A
45 Not used46 Arm, Waste Probe 571811B47 Retainer, E-ring 57167248 Rod, Pusher 113744
Item Description REF Part Number
078D0080-02, Rev. A
Replacing the PMT TED1. Turn off the system:
a. At the workspace, select System Status.b. Select Turn System Off.c. At the back of the system, turn off the power and unplug the power cord.
2. Disconnect the PMT thermistor cable from connector J6 on the Luminometer Bottom PC board.
3. Disconnect the PMT cable from connector J1 on the Luminometer Bottom PC board.
4. Remove the luminometer.
5. Place the lens cover (part number 27127) over the lens of the PMT to protect it from light, scratches, and contamination.
6. Remove the defective TED:
CAUTIONDo not scratch or soil the PMT lens. Scratching or soiling the lens will interfere with the light rays and cause erroneous results.
CAUTIONEnsure that you select Turn System Mechanics Off and cover the PMT lens when exposing the inside of the luminometer to light. Low levels of light can saturate the PMT and it can take several hours for the PMT to recover. High levels of direct light can damage the PMT.
1-90 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT TED
078D0080-02, Rev. A
a. Remove the two screws that secure the heatsink to the back of the luminometer.
Figure 1-1 Luminometer Assembly
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT TED 1-91
078D0080-02, Rev. A
b. Remove the two screws that secure the heatsink to the PMT housing.
Figure 1-2 PMT Housing Assembly
c. Remove the TED gasket.d. Remove the defective TED.
7. Install the TED:
a. Apply a thin layer of thermal grease to both sides of the new TED.
b. Install the new TED on the bottom of the PMT housing, ensuring the red (R) and black (B) wires are in the same position as the old TED.
CAUTIONEnsure that no thermal grease is inside the TED or heating problems will occur.
CAUTIONEnsure the TED is installed flush with the bottom of the PMT housing or damage can occur to the TED.
1-92 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT TED
078D0080-02, Rev. A
c. Install the TED gasket.
d. Install the TED heatsink and tighten the two screws. Alternate the tightening of each screw on the TED mounting block to 0.9 N–m (8 in–lb).
e. Remove the lens cover (part number 27127) from the lens of the PMT.
f. Apply silicone oil (part number 27295) to a lint free tissue and then apply the oil to the O–ring.
g. Install the two screws that secure the heatsink to the back of the luminometer.
8. Install the luminometer.
9. Connect the PMT thermistor cable to connector J6 on the Luminometer Bottom PC board.
10. Connect the PMT cable to connector J1 on the Luminometer Bottom PC board.
11. Turn the system on:a. Plug in the power cord.b. At the back of the system, turn the power on.c. At the back of the system, press the computer startup switch.
12. Verify PMT performance:a. At the workspace, select System Status.b. Select Diagnostic Tools.c. At the System-Diagnostic Tools window, select Dark Count without Cuvette.
d. Select Perform.The System-Dark Count window opens.
e. Ensure that the RLU 1 values for Total Mean are less than 600. RLUs higher than 600 may indicate a defective PMT, electronic noise, or light leaks.
CAUTIONFailure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs.
CAUTIONDo not allow silicone oil on the PMT lens. Oil on the lens will Interfere with the light rays and cause erroneous results.
WARNINGBefore selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT TED 1-93
078D0080-02, Rev. A
f. Refer to the chart below for the Total SD limit. Match the RLU1 Total Mean with the SD Limit curve. Verify the Total SD value is at or below the SD Limit curve.
Figure 1-3 Total SD Limit Chart
13. Check the dark counts with a cuvette:a. At the System-Diagnostic Tools window, select Dark Count with Cuvette.
b. Select Perform.The System-Dark Count window opens.
c. Ensure that the difference between RLU 1 values for Total Mean with cuvettes and without cuvettes are less than 25. RLUs higher than 25 may indicate light leaks.
14. If the dark count tests do not meet specifications, refer to the ADVIA Centaur Service Troubleshooting Manual.
15. Calibrate the luminometer thermals.
WARNINGBefore selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury.
078D0080-02, Rev. A
Replacing the PMT Assembly1. Remove the two screws securing the PMT assembly to the luminometer.
NOTE: You may need to pry the PMT away from the luminometer.
2. Remove the PMT assembly from the luminometer.
Figure 1-1 Luminometer Assembly
3. Place the lens cover (part number 27127) over the lens of the PMT to protect it from scratches and contamination.
CAUTIONDo not scratch or soil the PMT lens. Scratching or soiling the lens will interfere with the light rays and cause erroneous results.
CAUTIONEnsure that you select Turn System Mechanics Off and cover the PMT lens when exposing the inside of the luminometer to light. Low levels of light can saturate the PMT and it can take several hours for the PMT to recover. High levels of direct light can damage the PMT.
1-96 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT Assembly
078D0080-02, Rev. A
4. Disconnect the PMT thermistor cable from connector J6 on the Luminometer Bottom PC board.
5. Disconnect the PMT cable from connector J1 on the Luminometer Bottom PC board.
6. Replace the PMT assembly:
a. Remove the lens cover.b. Apply silicone oil (part number 27295) to a lint–free tissue and then apply the oil
to the O–ring.
c. Press the PMT assembly into the luminometer so that no light can be seen between the heatsink and the PMT.
d. Ensure that the notch on the back of the PMT housing aligns with the driver motor.
7. Secure the PMT assembly to the luminometer using two screws.
8. Connect the PMT thermistor cable to connector J6 on the Luminometer Bottom PC board.
9. Connect the PMT cable to connector J1 on the Luminometer Bottom PC board.
10. Turn the system on:a. Plug in the power cord.b. At the back of the system, turn the power on.c. At the back of the system, press the computer start button.
11. Verify PMT performance:
12. Check the dark counts without a cuvette:a. At the workspace, select System Status.b. Select Diagnostic Tools.c. At the System-Diagnostic Tools window, select Dark Count without Cuvette.
CAUTIONDo not allow silicone oil on the PMT lens. Oil on the lens will Interfere with the light rays and cause erroneous results.
CAUTIONEnsure the PMT is seated onto the housing of the luminometer. Incorrect placement of the PMT will cause light leakage and erroneous results.
WARNINGBefore selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT Assembly 1-97
078D0080-02, Rev. A
d. Select Perform.The System-Dark Count window opens.
e. Ensure that the RLU 1 values for Total Mean are less than 600. RLUs higher than 600 may indicate a defective PMT, electronic noise, or light leaks.
f. Refer to the chart below for the Total SD limit. Match the RLU1 Total Mean with the SD Limit curve. Verify the Total SD value is at or below the SD Limit curve.
13. Check the dark counts with a cuvette:a. At the System-Diagnostic Tools window, select Dark Count with Cuvette.
b. Select Perform.The System-Dark Count window opens.
c. Ensure that the difference between RLU 1 values for Total Mean with cuvettes and without cuvettes are less than 25. RLUs higher than 25 may indicate light leaks.
NOTE: For both tests (Dark Counts With Cuvettes and Dark Count Without Cuvettes), record and plot the Total Mean and Total SD values. Find the point of intersection between the two plots using the graph above. If the two plots intersect above the curve, there is noise in the electronic circuitry (a defective PMT or a cable picking up interference).
14. If the dark count tests do not meet specifications, refer to the ADVIA Centaur Service Troubleshooting Manual.
NOTE: : Reset test (assay) calibrations to ensure that the system uses the calibration ranges that were defined by the manufacturer until the system calculates new observed ranges.
WARNINGBefore selecting Perform, ensure that you are clear of subassemblies that can move. The system moves subassemblies and can cause injury.
1-98 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the PMT Assembly
078D0080-02, Rev. A
15. Reset test calibrations:a. At the workspace, select Calibration.b. Select Reset Test Calibrations.c. Select Yes.d. Select Continue.
078D0080-02, Rev. A
Preheater 078-B059-01
Preheater Assembly
1-100 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Cuvette Pusher Detail
078D0080-02, Rev. A
Cuvette Pusher Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 1-101
078D0080-02, Rev. A
Sensor Detail
Motor Detail
1-102 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Assembly Detail
078D0080-02, Rev. A
TED Assembly Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Assembly Detail 1-103
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Preheater 078-B059-01Item Description REF Part Number1 Cover, Preheater Channel 7127594 078-B067-012 Screw, Cap-Head M4 x 0.7 x 16 mm 60013x373 Bracket, Tube Connection 5718584 Cover, Preheater Exit 5718555 Cover, Cuvette Entrance 8966808 571853B6 Screw, Retractable M4 x 0.7 x 8.9 mm 9867382 5724227 PC Board, Reflective Chute Sensor 4162038 575447B8 Screw, Cap-Head M4 x 0.7 x 5 mm 5724169 Rack, Preheater Pusher 3062064 57264310 Screw, Flat-Head #4-40 x 0.25 mm Phillips 57241511 Plate, Motor Mounting 57184112 Screw, Cap-Head M4 x 0.7 x 8 mm 60013x913 Gear, Spur 48P 28T 0.187W 0.187B 522–0309–0114 Motor, Stepper 4VDC 9864200 57291515 PC Board, Preheater 5051035 575372B16 Thermistor, Threaded Body 1849083 571184C
Thermal Grease 12105517 TED, Assembly 5483539 111677
Thermal Grease 12105518 Heatsink, Preheater 57185619 Washer, Flat M5 60056x620 Screw, Cap-Head M5 x 0.8 x 30 mm 60013x7321 Screw, Cap-Head M3 x 0.5 x 6 mm 7882579 60013x122 Sensor, Optical Switch 3048878 11167523 Bracket, Sensor 57186024 Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 60007x4325 Pusher, Block 57185926 Screw, Cap-Head M2 x 0.4 x 10 mm 57241927 Screw, Cap-Head M5 x 0.8 x 16 mm 60013x3828 Screw, Cap-Head M4 x 0.7 x 40 mm 60013x8629 Screw, Cap-Head M4 x 0.7 x 70 mm 60013x128
078D0080-02, Rev. A
Replacing the Preheater TEDs1. Turn off the system:
a. At the workspace, select System Status.b. Select Turn System Off.c. At the back of the system, turn off the main power.
2. Remove the luminometer.
3. Remove the preheater.
4. Removing the defective TEDs:a. Remove the bottom screw and washer that secure the TED mounting block to the
preheater.
Figure 1-1 Preheater Assembly
b. Remove the top screw and the washer that secure the TED mounting block to the preheater.
1-108 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Preheater TEDs
078D0080-02, Rev. A
c. Remove the TEDs.
Figure 1-2 Preheater/TEDs Assembly
5. Install the TEDs:
a. Apply a thin layer of thermal grease to both sides of the new TEDs.
CAUTIONEnsure that no thermal grease is inside the TED or heating problems will occur.
CAUTIONEnsure the TEDs are installed flush with the bottom of the preheater or damage can occur to the TEDs.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Preheater TEDs 1-109
078D0080-02, Rev. A
b. Install the new set of TEDs on the bottom of the preheater ensuring the red (R) and black (B) wires are in the same position as the old TEDs.
c. Install the TED mounting block and tighten the two screws. Alternate tightening each screw to 0.9 N–m (8 in–lb).
6. Install the preheater.
7. Install the luminometer.
8. Turn the system on:a. Plug in the power cord.b. At the back of the system, turn the power on.c. At the back of the system, press the computer startup switch.
9. Calibrate the preheater thermals.
CAUTIONFailure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs.
078D0080-02, Rev. A
Ring Loader 572400B
Ring Loader Assembly
Figure 1-1 Ring Loader Assembly
1-112 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
Sensor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-113
078D0080-02, Rev. A
Motor Detail
078D0080-02, Rev. A
Parts List – Ring Loader 572400BItem Description REF Part Number1 Screw, Flat-Head M5 x 0.8 x 12 mm 5724232 Arm, Ring Loader 572057B3 Screw, Cap-Head M4 x 0.7 x 8 mm 60013x94 Screw, Retractable M4 x 0.7 x 8.9 mm 846-0039-105 Plate, Ring Loader Mounting 078-0061-016 Housing, Waste Probe 5718057 Spring, Compression 0.240 x 0.022 x 0.250 5718078 Pad, Friction 5718089 Screw, Flat-Head #4-40 x 0.25 Phillips 57241510 Plate, Mounting, Waste Probe Motor 57181411 Gear, Spur 32P 24P 0.188W 0.187B 1116124 571815B12 Motor, Stepper 4VDC Bipolar 9864200 57291513 Sensor, Optical Switch 962374 11167414 Screw, Cap-Head M3 x 0.5 x 5 mm 57241715 Screw, Cap-Head M4 x 0.7 x 8 mm 60013x9
078D0080-02, Rev. A
Servo Motor 571730B (Old)
Servo Motor Assembly
1-118 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Latch Detail
078D0080-02, Rev. A
Latch Detail
Servo Motor (Old) Detail 571730B
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Servo Motor (Old) Detail 571730B 1-119
078D0080-02, Rev. A
078D0080-02, Rev. A
Servo Motor 571022B (New)
Servo Motor (New) Assembly
1-122 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Latch Detail
078D0080-02, Rev. A
Latch Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Servo Motor (New) Detail 571022B
078D0080-02, Rev. A
Servo Motor (New) Detail 571022B
Figure 1-1 Servo Motor Assembly Detail
078D0080-02, Rev. A
Parts List – Servo Motor 571730B and 571022BItem Description REF Part Number1 Screw, Cap-Head M5 x 0.8 x 16 mm 60013x382 Screw, Cap-Head M4 x 0.7 x 12 mm 5717533 Bracket, Servo Motor 5717254 Plate, Servo Motor Mounting 078-0060-015 Latch, Compression 5717466 Motor, Servo (old) 1592244 571730B7 PC Board, Servo Motor 7617141 575357C8 Pulley, Timing 5717489 Motor, Servo (new) 4611886 571022B10 Adjuster, Belt 571752
078D0080-02, Rev. A
Wash Station 571025D
Wash Station Assembly
Figure 1-1 Wash Station Assembly
1-128 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Acid Probe Assembly Detail
078D0080-02, Rev. A
Acid Probe Assembly Detail (572889)
Figure 1-2 Acid Probe Assembly Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Friction Pad Detail 1-129
078D0080-02, Rev. A
Friction Pad Detail
Figure 1-3 Friction Pad Detail
Sensor Detail
Figure 1-4 Wash Station Sensor Detail
1-130 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail
078D0080-02, Rev. A
Motor Detail
Figure 1-5 Wash Station Motor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-131
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Wash Station 571025DItem Description REF Part Number1 Wash Manifold 01080626 078-0353-012 Tube, Connection, Arm 03310289 1171163 Assembly, Aspirate Probe Connector 05061677 078-B076-01
Plate 03649057 571871Screw, Captive 02280653 571533Fitting 00123534 518-0891-01Tubing 06291994 562-3070-03
4 Screw, Cap-Head M4 x 0.7 x 8 mm 06291994 60013x95 Probe, Wash Aspirate 03040281 571843
Grounded Probe, Wash Aspirate 02873484 1256696 Clip, Probe Retainer 04740554 571812B7 Screw, Cap-Head M4 x 0.7 x 20 mm 08373815 60013x518 Clamp, Cable 07960792 212339 Block, Wash Separation Assembly 03944946 116699B
Guide, Probe (Quantity 4) 03548633 112405BO-ring (Quantity 8) 04483748 571518Retainer, Probe Guide (Quantity 4) 02026323 112404Screw, Cap-Head M3 x 0.5 x 5 mm 08102390 572417Block, Wash Separation 03944946 116699B
10 Probe, Acid Dispense 03204667 110607C11 PC Board, Aspirate Probe 09571793 57536212 Screw, Cap-Head M4 x 0.7 x 6 mm 0189043 60013x213 Sensor, Optical Switch0 962374 11167414 Screw, Cap-Head M3 x 0.5 x 6 mm 06892408 60013x115 Screw, Flat-Head #4-40 x 0.25 Phillips 57241516 Plate, Motor Mounting, Wash Probe 57184117 Gear, Spur 01116124 571815B18 Motor, Stepper 4VDC 09864200 57291519 Screw, Cap-Head M4 x 0.7 x 5 mm 57241620 Arm, Wash Probe 03329850 57121221 Cover, Fixed Probe 2628622 Base, Fixed Probe 3490449 26285
134 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Wash Station 571025D
078D0080-02, Rev. A
23 Spring, Comp. 0.240 x 0.022 x 0.250 01112021 57180724 Pad, Friction, Long 03058539 57182925 Gasket, Junction Block 05150238 078-0275-0226 Retainer, Probe Guide 02026323 11240427 Screw, Cap-Head M3 x 0.5 x 5 mm 08102390 57241728 Housing, Wash Probe 08242214 57121329 Assembly, Wash Tubing 05097795 121587
Block, Junction 05247469 078-0505-01Tubing 01296947 572689Retainer, Tubing 111972Clamp 26699Fitting 05271787 571826Plate, Manifold (Quantity 1) 03649057 571871Plate Manifold (Quantity 2) 571827Fitting, (Quantity 14) 05271787 571826Screw, Captive (Quantity 3) 02280653 571533Screw, M4 x 0.7 x 16 mm
(Quantity 2)02960484 60013x37
O-rings, Junction Block (Quantity 20)
07737546 116828
30 Screw, Cap M4 x 16 mm 02960484 60013x3731 Guide, Aspirate Probe (single guide) 04913823 078–0088–01
Kit, Guide, Aspirate Probe (4 guides) 07254626 078–K110–01
Item Description REF Part Number
078D0080-02, Rev. A
Electronics Module (fixed configuration) 121911D
Electronics Module Assembly
Figure 1-1 Electronics Module Assembly
1-136 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Backplane PC Board 78-B047-01
078D0080-02, Rev. A
Backplane PC Board 78-B047-01 (Fixed Configuration)
Figure 1-2 Backplane PC Board
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 RT SPARC I/O PC Board 575267 1-137
078D0080-02, Rev. A
RT SPARC I/O PC Board 575267
Figure 1-3 RT SPARC I/O PC Board
1-138 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Fixed Electronics Module
078D0080-02, Rev. A
Fixed Electronics Module
Figure 1-4 Fixed Electronics Module
078D0080-02, Rev. A
Electronics Module (slide out configuration) 571115B
Electronics Module Assembly
Figure 1-1 Electronics Module (slide out configuration) Assembly
BackPlane PC Board 575252F (Slide Out)
Figure 1-2 Back Plane PC Board (slide out configuration)
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Slide Out Electronics Module 1-141
078D0080-02, Rev. A
Slide Out Electronics Module
Figure 1-3 Slide Out Electronics Module
078D0080-02, Rev. A
Parts List – Electronics Module 571115B and 121911DItem Description REF Part Number1 PC Board, Servo, Low Level Controller S575262C
Fuse, 2AMP 5728082 PC Board, Sequence Controller 117154B3 PC Board, Low Level Controller 078-B042S01
Fuse, 2AMP 5728084 PC Board, RT SPARC 572345C
IC, Memory 16MB Dram Simm 5751915 PC Board, Multi Master Bus 117149B6 PC Board, Backplane (slide out module) 575252F7 Standoff, #4-40 x 0.188 5718898 Assembly, PC Photon Counter 5753179 PC Board, RT SPARC I/O 57526710 PC Board, Backplane (fixed module) 078–B047–0111 PC Board, Power On/Off Control (fixed
module)117947
12 Power Supply (fixed module) 11907913 RT SPARC Service Assembly 9136329 078-B060R-01
078D0080-02, Rev. A
Replacing the Real Time SPARC PC Board1. Power down the system:
a. At the Workspace, select the System Status button.b. Select Turn System Off.c. At the back of the system, turn the power off.
2. Pull out the electronics module.
3. Remove the Real Time (RT) SPARC PC board from the Backplane PC board.
4. Disassemble the RT SPARC PC board:a. Separate the multiport S-BUS PC board (575207) from the SPARC PC board
(572345).b. Remove the six memory SIMMs (575191) plugged into the SPARC PC board
(572345). c. Remove the two screws securing the SPARC PC board to its metal bracket.d. Remove the SPARC PC board from the bracket and from the Interface PC board. e. Remove the three screws securing the Interface PC board to the metal bracket and
remove the PC board from the bracket.
5. Remove NVRAM the chip from the old board.If the chip is not damaged, save it for installation into the new board.
6. Assemble a new RT SPARC PC board:a. Obtain a new SPARC PC board.b. Mount the multiport S-BUS board on the SPARC PC board.c. Install the six memory SIMMs into the SPARC PC board. d. Install the NVRAM chip into the board.
7. If the old NVRAM chip was damaged and you installed a new chip, place the system ethernet address on the new chip:a. Obtain the evaporating password.b. Turn the system on.c. At the Welcome window, log in as root.d. Enter the evaporating password.e. Move the cursor off the Command Tool window.
CAUTIONObserve precautions for handling electrostatic–sensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.
1-146 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Real Time SPARC
078D0080-02, Rev. A
f. While holding down the right glidepoint button, move the cursor to the right arrow at Programs.
g. Select Texteditor and release the right glidepoint button.h. Select File.i. Open the etc folder.j. Double-click the ethers file.k. You will see a number like 8:0:20:80:5e:2b m0rt. The underlined portion is the
NVRAM ID. l. Delete these numbers and enter the new address found at the U0707 on the new
Sparc board.Ensure that you do not delete the colons. Verify that you entered the numbers correctly.
m. Select File for save.
File changes to Save for 1 second.
n. Move the cursor to the top bar of the text editor window, hold down the right glidepoint button and select Quit.
o. In the cmdtool window, enter init 5.p. Reboot the system.q. To confirm the RT NVRAM ID, open a new xterm window and enter cat /etc/
ethers.The system will display the address you entered.
r. Enter exit.If you entered the NVRAM ID incorrectly, the system will have a 700 00 07 error.
078D0080-02, Rev. A
Replacing the Real Time SPARC PC Board1. Power down the system:
a. At the Workspace, select the System Status button.b. Select Turn System Off.c. At the back of the system, turn the power off.
2. Pull out the electronics module.
3. Remove the Real Time (RT) SPARC PC board from the Backplane PC board.
4. Disassemble the RT SPARC PC board:a. Separate the multiport S-BUS PC board (575207) from the SPARC PC board
(572345).b. Remove the six memory SIMMs (575191) plugged into the SPARC PC board
(572345). c. Remove the two screws securing the SPARC PC board to its metal bracket.d. Remove the SPARC PC board from the bracket and from the Interface PC board. e. Remove the three screws securing the Interface PC board to the metal bracket and
remove the PC board from the bracket.
5. Remove NVRAM the chip from the old board.If the chip is not damaged, save it for installation into the new board.
6. Assemble a new RT SPARC PC board:a. Obtain a new SPARC PC board.b. Mount the multiport S-BUS board on the SPARC PC board.c. Install the six memory SIMMs into the SPARC PC board. d. Install the NVRAM chip into the board.
7. If the old NVRAM chip was damaged and you installed a new chip, place the system ethernet address on the new chip:a. Obtain the evaporating password.b. Turn the system on.c. At the Welcome window, log in as root.d. Enter the evaporating password.e. Move the cursor off the Command Tool window.
CAUTIONObserve precautions for handling electrostatic–sensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.
1-148 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Real Time SPARC
078D0080-02, Rev. A
f. While holding down the right glidepoint button, move the cursor to the right arrow at Programs.
g. Select Texteditor and release the right glidepoint button.h. Select File.i. Open the etc folder.j. Double-click the ethers file.k. You will see a number like 8:0:20:80:5e:2b m0rt. The underlined portion is the
NVRAM ID. l. Delete these numbers and enter the new address found at the U0707 on the new
Sparc board.Ensure that you do not delete the colons. Verify that you entered the numbers correctly.
m. Select File for save.
File changes to Save for 1 second.
n. Move the cursor to the top bar of the text editor window, hold down the right glidepoint button and select Quit.
o. In the cmdtool window, enter init 5.p. Reboot the system.q. To confirm the RT NVRAM ID, open a new xterm window and enter cat /etc/
ethers.The system will display the address you entered.
r. Enter exit.If you entered the NVRAM ID incorrectly, the system will have a 700 00 07 error.
078D0080-02, Rev. A
Replacing the User Interface SPARC PC Board (pre FCO 15)1. Power down the system:
a. At the Workspace, select the System Status button.b. Select Turn System Off.c. At the back of the system, turn the power off.
2. Pull out the electronics module.
3. Remove the User Interface (UI) SPARC PC board.a. Remove the two small pan head screws that secure the UI SPARC PC board
connector to the backplate.b. Unplug the UI SPARC PC board from the I/O PC board that is secured to the
backplane. Do not remove the UI SPARC PC board from the electronics module.
c. Remove the three cables from J27, J30, and J32 on the Backplane PC board that are connected to the UI SPARC PC board.
d. Partially remove the UI SPARC PC board.e. Disconnect the remaining cable from the UI SPARC PC board.f. Remove the UI SPARC PC board.
4. Disassemble the UI SPARC PC board:a. Separate the SBUS to MMBUS PC board (571263) from the SPARC PC board
(572345) and from the Interface PC board (575267).
5. Assemble a new UI SPARC PC board:a. Secure the Interface PC board to the metal bracket with three screws. b. Obtain a new SPARC PC board.c. Insert the SPARC PC board into the Interface PC board. d. Secure the SPARC PC board to the metal bracket with two screws.e. Mount the SBUS to MMBUS PC board on the SPARC PC board and the Interface
PC board. f. Remove the six memory SIMMs (575191) plugged into the old SPARC PC board.g. Install the six memory SIMMs on the new SPARC PC board.
6. Reconnect the User Interface (UI) SPARC PC board.a. Partially insert the UI SPARC PC board in the electronics module.
CAUTIONObserve precautions for handling electrostatic–sensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.
1-150 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the User Interface
078D0080-02, Rev. A
b. Reconnect the cable to the UI SPARC PC board that was removed in step 3e.c. Plug the UI SPARC PC board into the I/O PC board that is secured to the
backplane. d. Secure the UI SPARC PC board connector to the backplate with two small pan
head screws.e. Reconnect the three cables removed from the UI SPARC PC board in step 3c.
078D0080-02, Rev. A
User Interface Module 078-0103-XX for the ADVIA Centaur System
User Interface Module Assembly
152 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 User Interface Module Assembly
078D0080-02, Rev. A
CAUTIONObserve precautions for handling electrostatic–sensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.
078D0080-02, Rev. A
Parts List - User Interface Module 078-0103-XX* for the ADVIA Centaur System
**Last 2 digits of part number correspond to module version.
Item Description REF Part Number1 Power Supply, 180W Medical 1234222 Kit, Hard Disk Drive, 32 GB, SCSI 9211398 078-K016-023 Drive, Diskette 5578939 515-7173-124 Drive, DVD ROM 2378602 078–K186–015 PC Board, Audio Card 1234266 PC Board, Video Card 6811254 078-K157-017 PC Board, Power Control 9469166 1234288 Fan, Ultra Axi CPU 8876019 1234299 Fan, Enclosure 12343010 DIMM, Memory, 64 MB 4887059 123431
11 Speaker 7326511 12343212 Cable, SCSI Interface, with Terminator 078-0016-0113 Cable, Diskette Drive 12343414 Cable, Serial Splitter A-modem (B-LIS) 724929 12343515 Adapter, Serial Port 1758746 123014
078D0080-02, Rev. A
User Interface Module 078-0330-XX for the ADVIA Centaur XP System
User Interface Module Assembly
CAUTIONObserve precautions for handling electrostatic–sensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.
078D0080-02, Rev. A
Parts List - User Interface Module 078-0330-XX* for the ADVIA Centaur XP System
**Last 2 digits of part number correspond to software version.
Item Description REF Part Number1 Power Supply, 180W Medical 1234222 Kit, Hard Disk Drive, 32 GB, SCSI 9211398 078-K016-023 Drive, Diskette 5578939 515-7173-124 Drive, DVD ROM 078–0211–025 PC Board, Audio Card 1234266 PC Board, Video Card 515–7316–027 PC Board, Power Control 9469166 1234288 Fan, Ultra Axi CPU 8876019 1234299 Fan, Enclosure 12343010 DIMM, Memory, 64 MB 4887059 123431
11 Speaker 7326511 12343212 Cable, SCSI Interface, with Terminator 078-0016-0113 Cable, Diskette Drive 12343414 Cable, Serial Splitter A-modem (B-LIS) 724929 12343515 Adapter, Serial Port 1758746 123014
078D0080-02, Rev. A
Replacing the Hard Disk Drive
Use these procedure to replace the hard disk drive in the User Interface (UI) Module.
Removing the UI Module 1. Turn the system off.
a. At the workspace, select System Status. b. Select Turn System Off. c. Select Yes. d. At the back of the system, turn the power switch off. e. Disconnect the power cord from the system.
2. Loosen the two captive screws that secure the UI Module to the support rails.
3. Partially slide the UI Module forward allowing you access to the cable connections.
4. Disconnect the cables from the back of the UI Module.
5. Remove the UI Module from the support rails.
Removing the Hard Disk Drive 1. Remove the cover from the UI Module.
2. Remove the CD-Rom drive. a. Disconnect the two cables from the back of the CD-Rom drive.
NOTE: If you cannot remove the two inside mounting screws on the CD-Rom, remove the diskette drive module. The diskette drive module is secured to the bottom of the UI module with four screws.
b. Remove the four screws the secure the CD-Rom drive.
c. Push and remove the CD-Rom drive through the front of the UI Module.
3. Remove the SCSI interface cable from the back of the hard disk drive.
CAUTIONObserve precautions for handling electrostatic–sensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.
CAUTIONDo not pull the CD-Rom drive to the inside of the UI module or damage will occur to the EMI shielding strips.
1-160 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Hard Disk Drive
078D0080-02, Rev. A
NOTE: Perform step 4 to remove the SCSI interface cable from the UI module only if the old cable does not include an SCSI terminator.
4. If required, remove the cable for the hard drive and SCSI:
NOTE: The ferrite block is attached to the base of the UI module with double-sided tape. Remove the ferrite block from the base if you cannot open the ferrite block. a. Remove the ferrite block from the base of the UI module. b. Open the ferrite block surrounding the cable. Save the ferrite block. c. Disconnect the SCSI interface cable from the CPU board, CD-Rom drive, and
hard disk drive. d. Remove the cable adapter for the CD-Rom drive. Save the cable adapter and
discard the cable.
5. Remove the hard disk drive:a. Disconnect the two cables from the back of the hard disk drive.b. Disconnect the 2-pin connector from the hard disk drive.c. Remove the five screws the secure the mounting bracket and hard drive to the
bottom of the UI Module. d. Remove the mounting bracket and hard disk drive from the UI Module.
Figure 1-1 Hard Disk Drive Assembly
e. Remove the four screws the secure the hard drive to the mounting bracket.f. Remove the hard disk drive.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Hard Disk Drive 1-161
078D0080-02, Rev. A
Installing the New Hard Disk Drive 1. Install the Hard disk drive (part number 078-0017-01).
NOTE: Some hard disk drives have a jumper next to the LED OUT pins. You may need to trim the side of the LED connector to ensure that the LED connector attaches to the pins securely.
NOTE: Check the label on the hard drive for the LED OUT position. a. Connect the 2-pin connector to the LED OUT pins on the hard disk drive. b. Secure the hard disk drive to the mounting bracket using four screws. c. Secure the mounting bracket and hard disk drive to the bottom of the UI Module
using five screws. d. Connect the power cable to the hard disk drive. e. If you are not replacing the SCSI interface cable, connect the old cable to the hard
disk drive and continue with step 3.
NOTE: Perform step 2 to install the SCSI interface cable from the UI module only if the old cable does not include an SCSI terminator.
2. If required, install the new cable for the hard disk and SCSI: a. Connect the SCSI interface cable (part number 078-0016-01) to the CPU PC
Board.
NOTE: The ferrite block is attached to the base of the UI module with double-sided tape. Remove the ferrite block from the base if is not positioned as shown above.
Figure 1-2 User Interface Drive Component
b. Route the cable through the ferrite block. c. Close the ferrite block. d. Position the ferrite block in the UI module base.
e. Remove the backing from the tape on the bottom of the terminator. f. Secure the terminator to the UI module base close to the hard drive.
CAUTIONTo prevent damage to the connectors, do not twist the ribbon cable.
1-162 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Hard Disk Drive
078D0080-02, Rev. A
g. Connect the SCSI interface cable to the back of the hard disk drive.
NOTE: Ensure that you install the CD-Rom drive upside down. This position properly orients the drive on the system. The CD-Rom drive faces the same direction as the diskette drive.
3. Install the CD-Rom drive.
a. Gently push the CD-Rom drive into position through the front of the UI Module. b. Install the CD-Rom drive using the four screws that secure it. c. Connect the power cable to the back of the CD-Rom drive. d. Using the cable adapter removed previously, connect the SCSI interface cable to
the back of the CD-Rom drive.
4. Install the cover to the UI Module.
Installing the UI Module 1. Partially slide the UI Module forward onto the two support rails allowing you access
to the cable connections.
2. Reconnect and secure the cables to the back of the UI Module.
3. Slide the UI Module to the end of the support rails.
4. Secure the UI Module using the two captive screws.
5. Turn the system on. a. Connect the power cord to the back of the system and then turn the power switch
on. b. Turn the system on by pressing the green computer startup button located at the
back of the system.
6. Load the correct version of software on the system.
For more information, refer to the current version of software installation instructions.
CAUTIONBe careful that you do not damage the EMI shielding strip when you install the CD-Rom drive. The strip reduces EMI emissions from the UI Module.
078D0080-02, Rev. A
Application Module 078-0331-XX for the ADVIA Centaur System
Application Module Assembly
CAUTIONObserve precautions for handling electrostatic–sensitive devices. Use an antistatic mat and wrist strap when performing this procedure to prevent damage to the PC board and components.
1-164 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Application Module Assembly
078D0080-02, Rev. A
078D00??-01, Rev. A
Parts List – Application Module 078-0331-XX* for the ADVIA Centaur XP System
**Last 2 digits of part number correspond to software version.
Item Description REF Part Number1 Application Module 078-0331-**2 Isolation Transformer, ADVIA Centaur
System01654657 078-0402-01
2 Isolation Transformer, ADVIA Centaur XP System
01654657 078-0402-01
3 Right Side Panel Assembly, ADVIA Centaur System
07580108 122241
3 Right Side Panel Assembly, ADVIA Centaur XP System
08203286 078-B087-01
4 Centaur XP System Application PC Software 04297030 078-S046-015 Cable, Ethernet, Shielded 6203920 527-5153-03
ADVIA Centaur XP Operator’s Guide: 1 Parts List – Application Module 078-0331-XX* for the ADVIA Centaur XP
078D00??-01, Rev. A
078D0080-02, Rev. A
Power Supply
Power supply 571258D (slide out electronics module)
Figure 1-1 Power Supply
1-168ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Power Supply 078–B047–01 (fixed
078D0080-02, Rev. A
Power Supply 078–B047–01 (fixed electronics module)
Figure 1-2 Power Supply
078D0080-02, Rev. A
Parts List – Power SupplyItem Description REF Part Number1 Power Supply (slide out electronics
module)3628599 571258D
Fuse, 10A SB 60044x282 Power Supply (fixed electronics module) 8335220 078-B047-013 Transformer, Isolation (fixed electronics
module)9312976 078-B020–01
4 AC Input Line Assembly 7820265 078-B048-01
078D0080-02, Rev. A
Replacing the Power Supply
1. If the power is on, power down the system:a. At the workspace, select System Status.b. Select Turn System Off.
2. At the back of the system, turn the power off and unplug the power cord.
3. Remove the front cover of the electronics module.
4. Disconnect the AC power cable from the left side of the power supply module.
5. Disconnect the green (top) wire and white (bottom) wire from the DC Filter.
6. Remove the two Phillips head screws that secure the front of the power supply module to the electronic module base.
ELECTRICAL HAZARDInstrument operation with the power supply cover removed or open presents a potentially lethal electrical hazard. Do not operate the system with the power supply exposed.
1-172 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Power Supply
078D0080-02, Rev. A
7. Disconnect the two power supply cables J26 and J32 from the Backplane PC board.
8. Pull the power supply module out of the electronics module.
9. Install the new power supply into the electronics module.
10. Install the two Phillips head screws that secure the front of the power supply module to the electronic module base.
11. Connect the two power supply cables J26 and J32 to the Backplane PC board.
12. Connect the green (top) wire and white (bottom) wire to the DC Filter.
13. Connect the AC power cable to the left side of the power supply module.
14. Install the front cover of the electronics module.
15. Power up the system:a. Plug in the power cord.b. At the back of the system, turn the power on.c. At the back of the system, press the green computer startup switch.
078D0080-02, Rev. A
Thermal Control PC Module for the ADVIA Centaur System S575257E
Thermal Control PC Module Diagram
F1 Primary Reagent Compartment Loop AF2 Primary Reagent Compartment Loop CF3 Incubation Chamber Loop B F4 Incubation Chamber Loop A F5 Incubation Chamber Loop C F6 Incubation Chamber Loop DF7 Primary Reagent Compartment Loop BF8 Ancillary QueueF9 PreheaterF10 PMT1F11 PMT2
078D0080-02, Rev. A
Parts List – Thermal Control PC Module for the ADVIA Centaur System 575257E
Item Description REF Part Number1 PC Board, Thermal Control 575256C
Fuse, 5ASB (F3, F4, F5, F6, F9, F10, F11) 03841330 60044x14Fuse, 6.3ASB (F1, F2, F7, F8) 575194
078D0080-02, Rev. A
Thermal Control PC Module for the ADVIA Centaur XP System XP 078-B109-XX
Thermal Control PC Module Centaur XP Diagram
F1 Primary Reagent Compartment Loop AF2 Primary Reagent Compartment Loop CF3 Incubation Chamber Loop B F4 Incubation Chamber Loop A F5 Incubation Chamber Loop C F6 Incubation Chamber Loop DF7 Primary Reagent Compartment Loop BF8 Ancillary QueueF9 PreheaterF10 PMT1F11 PMT2
078D0080-02, Rev. A
Parts List – Thermal Control PC Module for the ADVIA Centaur XP System 078-B109-XX
Item Description REF Part Number1 PC Board, Thermal Control 05229977 078-B111SXX
Fuse, 5ASB (F3, F4, F5, F6, F9, F10, F11) 03841330 60044x14Fuse, 6.3ASB (F1, F2, F7, F8) 575194
078D0080-02, Rev. A
Replacing the Thermal Control Module or the PCB1. Turn off the system:
a. At the workspace, select System Status.b. Select Turn System Off.
2. At the back of the system, turn off the power and unplug the power cord.
3. Open the Electronics Module door.
4. Remove all cables from the Module.
5. Remove the Module from the system.
• If the PCB needs to be replaced, open the Module.
6. Remove the PCB.
7. Install the new Thermal Control Module PCB.
8. Reassemble the Module.
9. Reinstall the Module into the Centaur.
10. Turn the system on:a. Plug in the power cord.b. At the back of the system, turn the power on.c. At the back of the system, press the computer startup switch.
11. Sign in as an FSE.
NOTE: The following steps must be completed in order to restore the thermal calibration settings.
12. Go to the Thermals Window and select Perform Calibration.
13. At the bottom of the Thermal Calibration window you will see three buttons:
• First select Restore Calibration – When the Restore Calibration button is selected, the system will go to the hard drive and retrieve the thermal calibration values and store them in ram on the Thermal PCB.
• Now select the Save Calibration button – When the Save Calibration button is selected, the calibration values are Flashed to the EEProm on the Temperature Control Module and saved to the hard drive.
• Thermal Calibrations have now been restored to the Temperature Control Module.
Reagent Fluidics Module for the ADVIA Centaur System 078-B028-01
Reagent Fluidics Module Assembly
1-184 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 078-B061-01
078D0080-02, Rev. A
Diluter 2 mL 078-B061-01
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri–Continent Diluter 078-B061-01 1-185
078D0080-02, Rev. A
Tri–Continent Diluter 078-B061-01
078D0080-02, Rev. A
Parts List – Reagent Fluidics Module 078-B028-01Item Description REF Part Number1 Syringe, Tri-Continent, 2 mL 078-0502-01
Plunger 7121359 112248
2*
*May not be present
Plate, Mounting 115808
3 Valve, Pinch, V37 2386559 1125814 Gauge, Vacuum Indicator 2637217 193955 Regulator, Vacuum 9691438 5716176 Fitting, Barb, #10-32_5/32 UNF 2508468 1166377 PC Board, Reagent Fluid GIOB 1939228 078-B027S028 Manifold, Reagent 6561061 5728279 Valve, Two-Way NC Manifold,
Mounted,V19–22, V28–31, V47, V63, V64264987 571567B
10 Transducer, Pressure 1846351 57288611 Diluter, 2 mL, 078-B061-0112 Fitting (two piece insert with barb) 7325388 11008013 Chamber, Fluid 3275688 11539814 O-Ring 0.614ID x 070W Ethylene
Propylene1926711 115945
15 Seal, 2.0 mL, 0.500 ID 7026771 11539516 PC Board, Diluter 114830D17 Screw #4-40 x 187 60030x40318 Nut, Chamber 1718884 115633B19 Clamp, Tubing 6601055 11467920 Valve, Relief 4169504 566-0911-0121 O-ring, Fitting 7737546 11682822 Screw, Cap M4 x 8 mm 60013x923 Tubing, Tygon 6971510 2676624 Connector 6509019 26899x625 Tubing, Silicon 6845434 122166
078D0080-02, Rev. A
Reagent Fluidics Module for the ADVIA Centaur XP System 078-B028-01
Reagent Fluidics Module Assembly
Diluter 2 mL 115401C
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri-Continent Diluter 1-191
078D0080-02, Rev. A
Tri-Continent Diluter
078D0080-02, Rev. A
Parts List – Reagent Fluidics Module 078-B028-01Item Description REF Part Number1 Syringe, Tri-Continent, 2 mL 078-0502-01
Plunger 7121359 112248
2*
*May not be present
Plate, Mounting 115808
3 Valve, Pinch, V37 2386559 1125814 Gauge, Vacuum Indicator 2637217 193955 Regulator, Vacuum 9691438 5716176 Fitting, Barb, #10-32_5/32 UNF 2508468 1166377 PC Board, Reagent Fluid GIOB 1939228 078-B027S028 Manifold, Reagent 6561061 5728279 Valve, Two-Way NC Manifold,
Mounted,V19–22, V28–31, V47, V63, V64264987 571567B
10 Transducer, Pressure 1846351 57288611 Diluter, 2 mL, 115401C
Note: The new style diluter is interchangeable with the Tri-Continent diluter.
12 Fitting (two piece insert with barb) 7325388 11008013 Chamber, Fluid 3275688 11539814 O-Ring 0.614ID x 070W Ethylene
Propylene1926711 115945
15 Seal, 2.0 mL, 0.500 ID 7026771 11539516 PC Board, Diluter 114830D
17 Screw #4-40 x 187 60030x40318 Nut, Chamber 1718884 115633B19 Clamp, Tubing 6601055 11467920 Valve, Relief 4169504 566-0911-0121 O-ring, Fitting 7737546 11682822 Screw, Cap M4 x 8 mm 60013x923 Tubing, Tygon 6971510 2676624 Connector 6509019 26899x625 Tubing, Silicon 6845434 122166
078D0080-02, Rev. A
System Fluids Drawer 078-B055-01 for the ADVIA Centaur System
System Fluids Drawer Assembly
196 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly 197
078D0080-02, Rev. A
198 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Wash Manifold
078D0080-02, Rev. A
Wash Manifold
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter Detail 115401C (2 mL) 115402C
078D0080-02, Rev. A
Diluter Detail 115401C (2 mL) 115402C (5 mL)
Tri-Continent Diluter Detail
200 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri-Continent Diluter Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 201
078D0080-02, Rev. A
Sensor Detail
078D0080-02, Rev. A
Parts List – System Fluids Drawer 078-B055-01 for the ADVIA Centaur System
Item Description REF Part Number1 Screw, Flat M3 x 0.5 x 6 mm 60007x432 Tube, Formed 7526731 5719573 Lid Cover Connector Kit 4594191 078-K027-01
Lid, Bulk Bottle 571017Connector, Fluid Bottle 571600B
4 Screw, Pan-Head M3 x 0.5 x 6 mm 60000x435 Sensor, Optical Switch 1116746 Screw, M3 x 0.5 x 10 mm 60000x637 Valve, 3-Way, V43–46, V61 272658 PC Board, Wash Fluid Interconnect 575277B
9 Guide, Tube 11034810 Sensor, Capacitance 11326311 Stand, Tube 11283912 Reservoir 57101113 Gasket, System Fluid Reservoir 57283014 Connector, System Fluid Reservoir 078-0273-0115 Retainer, Holder Reservoir 57277716 Drip Tray, Lower Module 57159517 Pump, Acid, Kit 113747
Pump, Acid 1458x69Fitting 110080
Tubing, Acid Kit 572889Probe, Acid Dispense 110607Cover, Fixed Probe 26286Tubing, 3/8 OD x 3/16 ID 27341Tubing, Teflon, 0.035 ID 21347Fitting, Quick Disconnect 26519x6
18 Valve, Pinch, V38–41 11258119 PC Board, Fluid GIOB S575272E20 Pump, Base, Kit 113746
Pump, Base 1458x252
204 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – System Fluids Drawer
078D0080-02, Rev. A
Clamp, Wire 12811x1Tubing, Silicon, 0.194 OD 26861Tubing, LDPE, 0.058 ID 26601Fitting, Y-, Barbed 26599x6Fitting, Checkvalve 26860Fitting 110080
Tubing, Base, Kit 078-K081-01Tubing, Base Probe 12286Tubing, 30.4 cm (12 in) 116573Tubing, 190.5 cm (75 in) 572689Tubing, 15.2 cm (6 in) 7256777 572489Cover, Fixed Probe 26286Probe, Dispense 110608Fitting, Quick Disconnect 110080
21 Syringe, Tri-Continent, 5 mL 078-0503-0122 Plunger, 5 mL 11224723 Syringe, Tri-Continent, 2 mL 078-0502-0124 Plunger, 2 mL 7121359 11224825 Valve, 2-way, normally closed, V70
andV715224002 115870
26 Valve, 2-Way, NC Manifold, Mounted,V1–4, V6–17, V32, V59 –60, V68 – 69,and V72 – 76
264987 571567B
27 Fitting, With Insert, Barb, 2-Piece 7325388 11008028 O-ring, Fitting 7737546 11682829 Valve, 3-Way, Manifold Mounted, 4 inch,
V48 – 573352291 571568
30 Wash Manifold 3096287 11850631 Valve, 3-Way, V66 – 67, and V77 –78 7841092 078-K185-0132 Clip, Plastic, Snap-In 2713533 Connector, Tubing Male 113390
Fitting (Quantity 4) 571826Plate 571827Screw, Captive 571533
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – System Fluids Drawer
078D0080-02, Rev. A
34 Connector, Tubing, Male 3273316 113919Fitting (Quantity 7) 571826Screw, Captive 571533Plate 571871
35 DiluterDiluter, 2 mL 078-B061-01Diluter, 5 mL 078-B062-01
36 Fitting (two piece insert with barb)
7325388 110080
37 Chamber, FluidChamber, Fluid, 2 mL 3275688 115398Chamber, Fluid, 5 mL 2409877 115399
38 O-Ring, Ethylene propyleneO-Ring, 2 mL, 0.614ID x 070W 1926711 115945O-Ring, 5 mL, 0.864ID x 070W 1129145 115946
39 SealSeal, 2.0 mL, 0.500 ID 7026771 115395Seal, 5.0 mL, 0.750 ID 1571786 115396
40 PC Board, Diluter 114830D41 Screw #4-0 x 187 60030x40342 Nut, Chamber 1718884 115633B43 Clip, Valve 9654265 11430844 Plate, Mounting 11580845 Screw, Cap M4 x 4 mm 60013x946 Not used47 Guard, Diluter Thumbwheel 03948003 078-0485-01
Handle, System Fluid 078-0443-01Door, Diluter 078-0178-03Tray, System Fluids 571123Spacer, Lid 078-0322-01Slides 571688
48 Reservoir, Large 08411342 078-0270-01
Item Description REF Part Number
078D0080-02, Rev. A
System Fluids Drawer 078-081-01 for the ADVIA Centaur XP System
System Fluids Drawer Assembly
208 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly 209
078D0080-02, Rev. A
210 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 System Fluids Drawer Assembly
078D0080-02, Rev. A
Wash Manifold
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter Detail 115401C (2 mL) 115402C
078D0080-02, Rev. A
Diluter Detail 115401C (2 mL) 115402C (5 mL)
Tri-Continent Diluter Detail
212 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri-Continent Diluter Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 213
078D0080-02, Rev. A
Sensor Detail
078D0080-02, Rev. A
Parts List – System Fluids Drawer 078-B081-01 for the ADVIA Centaur XP System
Item Description REF Part Number1 Screw, Flat M3 x 0.5 x 6 mm 60007x432 Tube, Formed 5559055 078-0316-013 Lid Cover Connector Kit 078-K192-01
Lid, Bulk Bottle 7395645 078-0317-01Connector, Fluid Bottle 078-B082-01
4 Screw, Pan-Head M3 x 0.5 x 6 mm 60000x435 Sensor, Optical Switch 1116746 Screw, M3 x 0.5 x 10 mm 60000x637 Valve, 3-Way, V43–46, V61 272658 PC Board, Wash Fluid Interconnect 575277B
9 Guide, Tube 11034810 Sensor, Capacitance 11326311 Stand, Tube 11283912 Reservoir 08411342 078-0270-0113 Gasket, System Fluid Reservoir 57283014 Connector, System Fluid Reservoir 07395645 078-B082-0115 Retainer, Holder Reservoir 57277716 Drip Tray, Lower Module 078-0321-0117 Pump, Acid, Kit 113747
Pump, Acid 1458x69Fitting 110080
Tubing, Acid Kit 572889Probe, Acid Dispense 110607Cover, Fixed Probe 26286Tubing, 3/8 OD x 3/16 ID 27341Tubing, Teflon, 0.035 ID 21347Fitting, Quick Disconnect 26519x6
18 Valve, Pinch, V38–41 11258119 PC Board, Fluid GIOB S575272E20 Pump, Base, Kit 113746
Pump, Base 1458x252
216 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – System Fluids Drawer
078D0080-02, Rev. A
Clamp, Wire 12811x1Tubing, Silicon, 0.194 OD 26861Tubing, LDPE, 0.058 ID 26601Fitting, Y-, Barbed 26599x6Fitting, Checkvalve 26860Fitting 110080
Tubing, Base, Kit 078-K081-01Tubing, Base Probe 12286Tubing, 30.4 cm (12 in) 116573Tubing, 190.5 cm (75 in) 572689Tubing, 15.2 cm (6 in) 7256777 572489Cover, Fixed Probe 26286Probe, Dispense 110608Fitting, Quick Disconnect 110080
21 Syringe, Tri-Continent, 5 mL 078-0503-0122 Plunger, 5 mL 11224723 Syringe, Tri-Continent, 2 mL 078-0502-0124 Plunger, 2 mL 7121359 11224825 Valve, 2-way, normally closed, V70
andV715224002 115870
26 Valve, 2-Way, NC Manifold, Mounted,V1–4, V6–17, V32, V59 –60, V68 – 69,and V72 – 76
264987 571567B
27 Fitting, With Insert, Barb, 2-Piece 7325388 11008028 O-ring, Fitting 7737546 11682829 Valve, 3-Way, Manifold Mounted, 4 inch,
V48 – 573352291 571568
30 Wash Manifold 3096287 11850631 Valve, 3-Way, V66 – 67, and V77 –78 7841092 078-K185-0132 Clip, Plastic, Snap-In 2713533 Connector, Tubing Male 113390
Fitting (Quantity 4) 571826Plate 571827Screw, Captive 571533
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – System Fluids Drawer
078D0080-02, Rev. A
34 Connector, Tubing, Male 3273316 113919Fitting 571826Screw, Captive 571533Plate 571871
35 DiluterDiluter, 2 mL 078-B061-01Diluter, 5 mL 078-B062-01
36 Fitting (two piece insert with barb) 7325388 11008037 Chamber, Fluid
Chamber, Fluid, 2 mL 3275688 115398Chamber, Fluid, 5 mL 2409877 115399
38 O-Ring, Ethylene propyleneO-Ring, 2 mL, 0.614ID x 070W 1926711 115945O-Ring, 5 mL, 0.864ID x 070W 1129145 115946
39 SealSeal, 2.0 mL, 0.500 ID 7026771 115395Seal, 5.0 mL, 0.750 ID 1571786 115396
40 PC Board, Diluter 114830D41 Screw #4-0 x 187 60030x40342 Nut, Chamber 1718884 115633B43 Clip, Valve 9654265 11430844 Plate, Mounting 11580845 Screw, Cap M4 x 4 mm 60013x946 Not used47 Guard, Diluter Thumbwheel 03948003 078-0485-01
Handle, System Fluid 078-0318-01Door, Diluter 078-0319-01Tray, System Fluids 078-0320-01Spacer, Lid 078-0322-01Slides 078-0323-01Valve, Ancillary Queue Drain 01449808 078-0363-02Bracket, 3-Way Valve 078-0360-01
Item Description REF Part Number
078D0080-02, Rev. A
Vacuum Pump 572703B
Vacuum Pump Assembly
078D0080-02, Rev. A
Vacuum Pump 078-B050-02
Vacuum Pump Assembly
078D0080-02, Rev. A
Parts List – Vacuum Pump 572703BItem Description REF Part Number1 Screw, Cap-Head M4 x 0.7 x 30 mm 60013x722 Fan, 24V 2586418 23287
Connector 572317Lug 111288
3 Screw, Button M4 x 0.7 x 6 mm 7904256 5725004 Screw, Cap-Head M4 x 0.7 x 8 mm 60013x95 Pump, Vacuum and Controller 7861530 571616D6 Screw, Pan-Head #8-32 x 0.625 Phillips 25042x927 Guard, Fan 200698 Fitting 4586245 1166369 Latch, Compression 572065
Spacer 572712BSpring, Compression 572068Pawl 572067Nut 60035X5
10 Washer, Flat 846-1021-0711 Standoff 558-0867-1612 Stem, Adjustment 653-0034-0113 Cover, Terminal Block 078-0036-0114 Cover, Vacuum Pump 078-0035-0115 Nut, Hex 10848816 Mount, Vibration 12420917 Washer, Lock 25042x9218 Plate, Vacuum Mount 572711B19 Not Used20 Block, Terminal 571420
Screw, Cap-Head M4 x 0.7 x 6 mm 60013x921 Grommet 556-0383-0122 Holder, Fuse 3387753 117121
Fuse, 5A 60044x14Washer, Lock 25049x92Nut, Hex 60038x5
078D0080-02, Rev. A
Waste and Water Drawer 078-B005-01 for the ADVIA Centaur System
Waste and Water Drawer Assembly
226 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Water Reservoir Detail
078D0080-02, Rev. A
Water Reservoir Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Waste Reservoir Detail 227
078D0080-02, Rev. A
Waste Reservoir Detail
228 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Water Pressure Pump Detail
078D0080-02, Rev. A
Water Pressure Pump Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Caster Detail 229
078D0080-02, Rev. A
Caster Detail
230 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Caster Detail
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Waste and Water Drawer 078-B005-01 for the ADVIA Centaur System
Item Description REF Part Number1 Lid, Water, ADVIA Centaur System 571128F
Pin 571625ARing, Retaining 3/16 inch Shaft 572439
2 Lid, Waste 571127FPin 571625ARing, Retaining 3/16 inch Shaft 572439
3 PC Board, System Fluid Detect 5511044 575452CScrew, Cap M4 x 8 SST 60013x9
4 Trap, Water Assembly 1295053 549-8072-01Bracket, Water Trap 078-0042-01
Gasket 7940767 078-0065-01Tape, Teflon 590-0042-01
5 Fitting, Elbow, 1/4 NPT-3 Flow 3024294 518-0876-016 Filter, Water 2757433 26970
Strap 1143067 Kit, Water Pressure Pump Assembly 3842272 078-K002-01
Pump, Water Pressure 571615Tape, Teflon 19720Clamp 6601055 114679
8 Pump, Bellows 5317728 114424Bracket, Mounting 2341512 121253Screw, Cap M4 x 8 SST 60013x9
9 Sensor, Optical Switch 962374 11167410 Screw, Set M4 x 0.7 x 4 mm Hex 60023x4011 Screw, Shoulder 5 mm x 12 SST 846-0021-1112 Washer, M5, SST 846-1021-0813 Spring, Comp 0.390 x 0.043 x 0.500 8686287 57287014 Elbow, 1/4 inch Flow, 1/4 Barb, Shutoff 7542990 57286315 Fitting, Waste Transfer Tube 078-0026-01
Tube, Waste Transfer 4150501 078-0028-01
232ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Waste and Water Drawer
078D0080-02, Rev. A
16 Sensor, Optical, Water Reservoir Full, 5.08 cm (2 in)
4935924 078-B128-01
17 Reservoir, Water 3514445 57110118 Cap, Water Bottle 9962296 57259619 Clip, Optical Sensor 078-0024-0120 Elbow, 1/8 inch Flow, 1/8 Barb, Thru 7471368 57177121 Elbow, 1/4 inch Flow, 1/4 Barb, Thru 4255656 57177322 Coupling, 1/8 inch Flow, 1/8 NPT, Shutoff 57176723 Reservoir, Waste 078-0025-0124 Cap, Waste 1712924 118359
Cap 29058 114218Baffle 306833 114630Seal 114303Fitting 571766O-Ring 545-0196-01
25 Waste Bottle Fluid Sensor Operator Kit 0946549 078-K074-02Sensor, Optical, Waste Bottle,
7.62 cm (3 in)078-B029-01
Language Pack, Replacing the Waste Bottle Fluid Sensor
078D0271-01
25 Waste Bottle Fluid Sensor and Cable Kit (Service)
8768658 078-K075-01
Sensor, Optical, Waste Bottle, 7.62 cm (3 in)
078-B029-01
Cable, sensor 078-B030-01Clamp “P”Screw, S.H.C.S, M4 12, SST
26 Coupling, 1/4 inch Flow, Thru 57177227 Bottle, Waste 12436528 Spigot 57161829 Bottle, Water 113850B30 Filter, Air Vent 11631931 Sensor, Optical, Waste Reservoir Low,
10.16 cm (4 in)6298441 078-B130-01
32 Screw, Pan-Head M3 x 0.5 x 6 mm 60000x43
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Waste and Water Drawer
078D0080-02, Rev. A
33 Sensor, Optical, Waste Reservoir Full, 7.62 cm (3 in)
8134047 078-B129-01
34 Elbow, 1/8 inch Flow, 1/8 Barb, Shutoff 11459135 Coupling, 1/8 inch Flow, 1/8 Barb, Thru 11502236 Valve, Check 116151
Clamp, Tubing 11467937 Nut, M3 60024x438 Isolator, Vibration 11726639 Coupling, 1/4 inch Flow 11207540 Screw, M4 x 0.7 x 20mm Stainless Socket 60055x9541 Plate, Caster Mounting 115486B42 Caster 11548743 Washer, Flat M4 60028x544 Screw, Pan M4 x .08 Phillips 60000x5545 Kit, Degasser Assembly 078-K049-02
Clamp, Tubing, 7/16 OD 19947x746 Manifold, Water Reservoir 11711547 Bracket 571621C48 Slide 57168949 Clamp, Tubing 11467950 Fitting, Pressure Pump Output 518-7075-0151 Gasket, Optical Sensor 57193752 Clip, Water Trap 509-0331-0153 Screw, Cap M4 x 8 SST 60013x954 Drawer, Waste Water 57159655 Fitting, Waste Reservoir 518-0818-0556 Screw, Pan M4 x 6 60003x4557 Wall, Back 571129D
Screw, Cap M4 x 14 Hex. 60013x3058 Clamp 23491
Assembly, Sensor Cable 078-B030-01
Item Description REF Part Number
078D0080-02, Rev. A
Waste and Water Drawer 078-B089-01 for the ADVIA Centaur XP System
Waste and Water Drawer Assembly
236 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Waste Reservoir Detail
078D0080-02, Rev. A
Waste Reservoir Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Waste Reservoir Detail 237
078D0080-02, Rev. A
Waste Reservoir Detail
238 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Water Pressure Pump Detail
078D0080-02, Rev. A
Water Pressure Pump Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Caster Detail 239
078D0080-02, Rev. A
Caster Detail
240 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Caster Detail
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Waste and Water Drawer 078-B089-01 for the ADVIA Centaur XP System
Item Description REF Part Number1 Lid, Water 078-0358-012 Lid, Waste 078-0359-01
Pin 571625ARing, Retaining 3/16 inch Shaft 572439
3 PC Board, System Fluid Detect 5511044 575452CScrew, Cap M4 x 8 SST 60013x9
4 Trap, Water Assembly 1295053 549-8072-01Bracket, Water Trap 0831660 078-0042-02Gasket 7940767 078-0065-01Tape, Teflon 590-0042-01
5 Fitting, Elbow, 1/4 NPT-3 Flow 3024294 518-0876-016 Filter, Water 2757433 26970
Strap 114306Clip 0803047 571714B
7 Kit, Water Pressure Pump Assembly 3842272 078-K002-01Pump, Water Pressure 571615Tape, Teflon 19720Clamp 6601055 114679
8 Pump, Bellows 5317728 114424Bracket, Mounting 2341512 121253Screw, Cap M4 x 8 SST 60013x9
9 Sensor, Optical Switch 962374 11167410 Screw, Set M4 x 0.7 x 4 mm Hex 60023x4011 Screw, Shoulder 5 mm x 12 SST 846-0021-1112 Washer, M5, SST 846-1021-0813 Spring, Comp 0.390 x 0.043 x 0.500 8686287 57287014 Elbow, 1/4 inch Flow, 1/4 Barb, Shutoff 7542990 57286315 Fitting, Waste Transfer Tube 078-0026-01
Tube, Waste Transfer 4150501 078-0028-0116 Sensor, Optical, Water Reservoir Full,
5.08 cm (2 in) 4935924 078-B128-01
242ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Waste and Water Drawer
078D0080-02, Rev. A
17 Reservoir, Water 3514445 57110118 Cap, Water Bottle 9962296 57259619 Clip, Optical Sensor 078-0024-0120 Elbow, 1/8 inch Flow, 1/8 Barb, Thru 7471368 57177121 Elbow, 1/4 inch Flow, 1/4 Barb, Thru 4255656 57177322 Coupling, 1/8 inch Flow, 1/8 NPT, Shutoff 57176723 Reservoir, Waste 04329595 078-B090-0124 Cap, Waste 1712924 118359
Cap 29058 114218Baffle 306833 114630Seal 114303Fitting 571766O-Ring 545-0196-01
25 Waste Bottle Fluid Sensor Operator Kit 0946549 078-K074-02Sensor, Optical, Waste Bottle,
7.62 cm (3 in)078-B029-01
Language Pack, Replacing the Waste Bottle Fluid Sensor
078D0271-01
25 Waste Bottle Fluid Sensor and Cable Kit (Service)
8768658 078-K075-01
Sensor, Optical, Waste Bottle, 7.62 cm (3 in)
078-B029-01
Cable, sensor 078-B030-01Clamp “P”Screw, S.H.C.S, M4 12, SST
26 Coupling, 1/4 inch Flow, Thru 57177227 Bottle, Waste 00004926 078-0166-0128 Spigot 57161829 Bottle, Water 078-B046-0130 Filter, Air Vent 11631931 Sensor, Optical, Waste Reservoir Low,
10.16 cm (4 in)6298441 078-B130-01
32 Screw, Pan-Head M3 x 0.5 x 6 mm 60000x4333 Sensor, Optical, Waste Reservoir Full,
7.62 cm (3 in)8134047 078-B129-01
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Waste and Water Drawer
078D0080-02, Rev. A
34 Elbow, 1/8 inch Flow, 1/8 Barb, Shutoff 11459135 Coupling, 1/8 inch Flow, 1/8 Barb, Thru 11502236 Valve, Check 116151
Clamp, Tubing 11467937 Nut, M3 60024x438 Isolator, Vibration 11726639 Coupling, 1/4 inch Flow 11207540 Screw, M4 x 0.7 x 20mm Stainless Socket 60055x9541 Plate, Caster Mounting 115486B42 Caster 11548743 Washer, Flat M4 60028x544 Screw, Pan M4 x .08 Phillips 60000x5545 Kit, Degasser Assembly 078-K049-02
Clamp, Tubing, 7/16 OD 19947x746 Manifold, Water Reservoir 1338682 11711547 Bracket 571621C48 Slide 57168949 Clamp, Tubing 11467950 Fitting, Pressure Pump Output 518-7075-0151 Gasket, Optical Sensor 57193752 Clip, Water Trap 509-0331-0153 Screw, Cap M4 x 8 SST 60013x954 Drawer, Waste Water 078-0356-0155 Fitting, Waste Reservoir 518-0818-0556 Screw, Pan M4 x 6 60003x4557 Wall, Back, ADVIA 078-0356-0158 Clamp 23491
Assembly, Sensor Cable 078-B030-0159 Valve, Monthly Cleaning 078-B014-0260 Diaphragm, Water Pump 04457976 549-B073-0161 Kit, Water Pump Poppet Valve 00732506 566-1237-01
Item Description REF Part Number
078D0080-02, Rev. A
Primary Reagent Compartment 571111C
Primary Reagent Compartment Assembly
Figure 1-1 Reagent Compartment Assembly
1-246 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment
078D0080-02, Rev. A
Figure 1-2 Reagent Compartment Assembly
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment
078D0080-02, Rev. A
Figure 1-3 Reagent Compartment Assembly
1-248 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment
078D0080-02, Rev. A
Figure 1-4 Reagent Compartment Assembly
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment
078D0080-02, Rev. A
Figure 1-5 Reagent Compartment Assembly
1-250 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment
078D0080-02, Rev. A
Figure 1-6 Reagent Compartment Assembly
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Lead Screw Detail 1-251
078D0080-02, Rev. A
Lead Screw Detail
Figure 1-7 Reagent Compartment Lead Screw Detail
1-252 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
Sensor Detail
Figure 1-8 Reagent Compartment Sensor Detail
Motor Detail
Figure 1-9 Reagent Compartment Motor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 TED Detail 1-253
078D0080-02, Rev. A
TED Detail
Figure 1-10 Reagent Compartment TED Detail
078D0080-02, Rev. A
Parts List – Primary Reagent Compartment 571111CItem Description REF Part Number1 Screw, Button M4 x 0.7 x 12 mm Socket 1099032 Holder, Rinse Station 2 & 3 2841884 572657B3 PC Board, Mixer Interconnect 7483250 575422C4 Screw, Cap-Head M4 x0.7 x 05 mm 5724165 Holder, Rinse Station 1 8653699 572656B6 Shutter, Reagent Probe 8523078 5717857 Screw, Cap-Head M4 x 0.7 x 16 mm 60013x378 Washer, Flat M4 Stnl 60056x59 Cover, Reagent Box, ADVIA Centaur
System2825080 571636C
9 Cover, Reagent Box, ADVIA Centaur XP System
078-0377-01
10 Assembly, Crossflow Blower, ADVIA Centaur System
3360812 572934
11 Screw, Cap-Head M4 x 0.7 x 10 mm 60013x1612 Motor, Stepper 4VDC 9864200 57291513 Screw, Flat-Head #4-40 x .38 Phillips 60031x40614 Sensor, Optical Switch 7713183 11167115 Screw, Cap-Head M3 x 0.5 x 06 mm 7882579 60013x116 Plate, Interrupter 57246117 Screw, Set M4 x 0.7 x 4 mm 60022x3918 Cam 57245019 Plate Adapter, Probe Shutter 57247220 Screw, Flat-Head M4 x 0.7 x 10 mm Socket 60055x6521 Screw, Flat-Head M4 x 0.7 x 12 mm Socket 57178122 Plate, Motor Mounting BC Shutter 57245123 Screw, Flat-Head M3 x 0.5 x 10 mm
Phillips60007x63
24 Plate, Top Reagent Tray 571506C25 Plate, Mixer 571482C26 Slide, Reagent Pack (1-30) 0009995 124356
Slide #1 571892BSlide #2 571893B
1-256 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Primary Reagent
078D0080-02, Rev. A
Slide #3 571894BSlide #4 571895BSlide #5 571896BSlide #6 571897BSlide #7 571898BSlide #8 571899BSlide #9 571900BSlide #10 571901BSlide #11 571902BSlide #12 571903BSlide #13 571904BSlide #14 571905BSlide #15 571906BSlide #16 571907BSlide #17 571908BSlide #18 571909BSlide #19 571910BSlide #20 571911BSlide #21 571912BSlide #22 571913BSlide #23 571914BSlide #24 571915BSlide #25 571916BSlide #26 571917BSlide #27 571918BSlide #28 571919BSlide #29 571920BSlide #30 571921B
27 Shaft, Reagent Tray 571495B28 Retainer, Shaft, Reagent Tray 571847B
Grease, Lithium 57266229 Screw, Cap-Head M4 x 0.7 x 12 mm 60013x2330 Screw, Flat-Head M4 x 0.7 x 16 mm Socket 60055x85
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Primary Reagent
078D0080-02, Rev. A
31 Connector, Drive, Reagent Tray 57150132 Screw, Cap-Head M4 x 0.7 x 14 mm 60013x3033 Screw, Cap-Head M4 x 0.7 x 22 mm 60013x5834 Washer, Shoulder, #8 Nylon 5286512 57210635 Spacer, TED, Reagent Box 571498B36 Thermistor, Threaded Body 1849083 571184C
Thermal Grease 12105537 Coupling, Flexible 0.25 x 0.75 x 1.025 571509B38 Motor Mount, Reagent Drive 571488B39 Screw, Cap-Head M4 x 0.7 x 8 mm 60013x940 Motor, Stepper 6VDC 2124384 57291341 Sensor, Slotted 962374 11167442 Clamp, Split-Hub 0.251D x 0.187 57222843 Spacer, 0.250 x 0.375 x 0.063 Stainless 2653844 Bearing, Ball Flanged 0.250 x 0.500 x
0.1884308474 26541
45 Support, Leadscrew Assembly 57244946 Ring, Retaining 0.168 x 0.300 x 0.15 57147447 Bearing, Journal 0.19 x 0.315
x0.375-Delrin5266031 571473
48 Leadscrew, Assembly 572661B49 Shaft, Drive Reagent Mixer 57147250 Flag, Reagent Drive 57147551 Shutter, Bar Code 571786C52 Retainer, Bearing 57148053 TED, Assembly 5518057 112586
Thermal Grease 12105554 Insulation 57223255 Insulation 5518057 57247756 Kit, Reagent Door Seal 8570068 572460B57 Bracket, Reagent Drive 078-0097-0158 Guide, Reagent Shaft 078-0098-0159 Washer, Lock 25042x9260 Nut, Hex 108488
Item Description REF Part Number
1-258 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Primary Reagent
078D0080-02, Rev. A
61 Mount, Isolator 12420962 Magnet 57199163 Screw, Shoulder 9408892 57223764 Seal, Gasket 2793456 572235
Item Description REF Part Number
078D0080-02, Rev. A
Replacing the Primary Reagent Compartment Leadscrew1. Remove the primary reagent packs and store them in a refrigerator.
2. At the workspace, select System Status.
3. Select Turn Mechanics Off, and then wait while the system mechanics power down.
4. Remove the three screws that secure the back left cover to the frame.
5. Remove the back left cover.
Figure 1-1 Primary Reagent Compartment – ADVIA Centaur System
1-260 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
6. Access the leadscrew assembly through the back of the system.
Figure 1-2 ADVIA Centaur System – Back View
7. Move the reagent probe 1 and reagent barcode scanner to the right allowing access to the back left corner of the primary reagent compartment.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
8. Remove the reagent mixer drive shaft from the leadscrew housing.
Figure 1-3 Primary Reagent Compartment Mixer Assembly
1-262 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
9. Using a 7/64 inch ball driver, loosen the two set screws that secure the mixer leadscrew to the mixer motor coupling.
Figure 1-4 Reagent Compartment Mixer Assembly
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
10. Using a 3-mm hex driver, remove the two screws that secure the leadscrew mounting block to the back of the primary reagent compartment. Save the screws.
Figure 1-5 Primary Reagent Compartment Mixer Assembly
1-264 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
11. Remove the leadscrew assembly from the primary reagent compartment.
Figure 1-6 Primary Reagent Compartment Mixer Assembly
12. Remove the clamp, the spacer, the leadscrew mounting block, and the reagent drive flag from the leadscrew assembly and save the parts.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
13. Discard the leadscrew housing and the leadscrew.
Figure 1-7 Primary Reagent Compartment Leadscrew Assembly
1-266 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
14. Install the clamp, the spacer, the leadscrew mounting block, and the reagent drive flag on the new leadscrew assembly.
Figure 1-8 Primary Reagent Compartment Leadscrew Assembly
15. Press up on the bottom of the leadscrew until the top edge of the leadscrew is flush against the bearing.
16. Press down and tighten the screw on the clamp.
17. Apply NSK Grease No. 1 to the middle of the leadscrew and leadscrew mounting block arm.
18. Move the probe mechanism manually until a thin layer of grease spreads to the entire leadscrew.Remove any excess grease.NOTE: Do not tighten the set screws on the motor mixer coupling at this time.
19. Place the leadscrew assembly in the motor mixer coupling that is in the primary reagent compartment.
CAUTIONFailure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
20. Ensure that the motor mixer coupling moves freely on the shaft.
21. Using a 3-mm hex driver, install the two screws that secure the leadscrew mounting block to the back of the primary reagent compartment.
22. Use a small flatblade screwdriver to lift the bottom of the rubber coupling from the mixer motor, and then insert an edge of a standard piece of paper (the width is approximately 0.102 mm [0.004 in]). Remove the screwdriver.
Figure 1-9 Primary Reagent Compartment Leadscrew Assembly
23. Push down on the coupling and the sheet of paper.
24. Ensure that the upper and lower set screws on the rubber coupling are aligned, and then use a 7/64 inch ball driver to tighten the bottom set screw.
25. Tighten the top set screw, and then remove the sheet of paper.
26. Manually move the leadscrew:
CAUTIONEnsure that the rubber coupling moves freely on the motor shaft or system damage can occur.
CAUTIONEnsure that the reagent drive flag does not hit the sensors or system damage can occur.
1-268 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
a. Ensure that the bottom of the rubber coupling is not rubbing against the motor.b. Ensure that the set screws on the clamps are not rubbing against the mounting
plate.c. Ensure that the reagent drive flag is not hitting the mixer movement sensors.
27. Install the reagent mixer drive shaft:a. Tilt the reagent mixer to the back of the primary reagent compartment.b. Rotate the leadscrew and lower the leadscrew housing to the bottom of travel.
c. Install the new reagent mixer drive shaft into the leadscrew housing.If the drive shaft is not going into the leadscrew housing smoothly, remove the drive shaft and install it again.Ensure that the drive shaft engages the reagent mixer plate.
NOTE: Ensure that the rinse station tubing does not interfere with the movement of the leadscrew.
28. Install the back left cover to the frame.
29. Select Turn System Mechanics On.
30. Calibrate the reagent mixer.
31. Verify that the reagent probes bottom calibrations are correct.
32. Install the primary reagent packs.
CAUTIONDo not apply Loctite to the new reagent mixer drive shaft. If you apply Loctite, the leadscrew housing can crack.
CAUTIONDo not force the new reagent mixer drive shaft into the leadscrew housing. If you force the drive shaft into the leadscrew housing, you can damage the threads.
078D0080-02, Rev. A
Replacing the Primary Reagent Compartment TEDs1. Turn off the system:
a. At the workspace, select System Status.b. Select Turn System Off.c. At the back of the system, turn off the main power and unplug the power cord.d. Remove the primary reagent packs from the primary reagent compartment and store
them in a refrigerator.
2. Remove the three screws that secure the back left cover to the frame.
3. Remove the back left cover.
Figure 1-1 Primary Reagent Compartment
1-270 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
4. Remove the primary reagent compartment cables:
Figure 1-2 Primary Reagent Compartment Assembly
NOTE: Label all cable connections before removing the cables.a. Disconnect the sensor motor interconnect cable (P10) and the thermistors cable
(P14) from the primary reagent compartment PC board.b. Disconnect the fan cable.c. Disconnect the LED array upper cable (P19) and the LED array lower cable (P22)
from the Reagent GIOB PC board.d. Move the cables away from the primary reagent compartment.
5. Remove the five screws and washers that secure the top cover and reagent probe shutter to the primary reagent compartment.
6. Remove the primary reagent compartment:
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
a. Remove the two rinse stations.b. Remove the four screws that secure the primary reagent compartment to the
frame.c. Move the reagent compartment out and away from the Incubation Chamber
approximately 152 mm (6 in).d. Remove the tubing from the two drain ports located at the bottom left and right
side of the primary reagent compartment.e. Disconnect the TED power cable from the connector.f. Remove the primary reagent compartment and place it on a flat surface.
7. Remove the two screws that secure the TED wire cover to the primary reagent compartment.
8. Pull the TED wire cover away from the reagent compartment.
9. Remove the TED connector from the cover.
10. Remove the top cover and reagent probe shutter.
1-272 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
11. Loosen the two screws and retainers that secure the reagent mixer to the mixer bearing.
Figure 1-3 Primary Reagent Compartment Assembly
12. Remove the reagent mixer.
13. Remove the two screws that secure the coldsink cover to the coldsink.
14. Remove the coldsink cover.
15. Remove the two screws that secure the fan to the reagent compartment.
16. Remove the fan.
17. Disconnect the three thermistor cables from the primary reagent compartment PC board and route the cables through the left side of the reagent compartment.
CAUTIONBe careful not to kink the thermistor wires or damage can occur to the wires.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
18. Remove the bottom plate to the primary reagent compartment: a. Remove the 11 screws and washers that secure the bottom plate to the reagent
compartment.
Figure 1-4 Primary Reagent Compartment TEDS Assembly
b. Simultaneously pull and twist the mixer bearing until you break the tight seal between the bottom plate and the reagent compartment.
c. Guide the TED wires through the reagent compartment and remove the bottom plate.
19. Remove the TED heatsinks and insulation:a. Remove the 12 screws that secure the six TED spacer blocks to the bottom plate. b. Remove the two pieces of insulation surrounding the TED mounting blocks.
CAUTIONDo not pry the bottom plate from the reagent compartment or damage will occur to the reagent compartment.
1-274 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
20. Remove the TEDs:
Figure 1-5 Primary Reagent Compartment TEDS Assembly
a. Remove the two screws that secure each TED mounting block to the heatsink.b. Remove the six TEDs. c. Clean the thermal grease from the heatsinks, TED mounting blocks, and bottom
plate.
21. Install the TEDs:
CAUTIONApply only enough thermal grease to fill any gaps or holes on the metal surface of the TED mounting blocks. Excess thermal grease acts as a insulator and reduces the performance. Use only 40 to 60 µL of thermal grease. Thermal grease on the edges of the TEDs can cause cooling problems.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
a. Apply a thin layer (20 – 30 µL) of thermal grease to the TED side of each TED spacer block.
b. Apply a thin layer (20 – 30 µL) of thermal grease to the two TED areas of each heatsink.
c. Remove any excess thermal grease from the TED spacer blocks and heatsinks with a straight edge.
d. Install the TEDs on the bottom of each heatsink ensuring the red (R) and black (B) wires are in the same position as the old TEDs.
Figure 1-6 Primary Reagent Compartment Connector Back View
e. Install the TED spacer blocks and torque the two screws. Alternate tightening each screw, a little at a time to 0.68 N–m (6 in–lb) on the TED spacer block.
f. Wait three minutes and then tighten each screw to 0.68 N–m (6 in–lb).
CAUTIONEnsure the TEDs are installed flush with the bottom of the heatsink or damage can occur to the TEDs.
CAUTIONFailure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs.
1-276 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
22. Install the heatsinks:a. Install the two pieces of insulation over the TED mounting blocks.b. Apply a thin layer of thermal grease to each TED spacer block.c. Apply a thin layer of thermal grease to each TED spacer area of the bottom plate.d. Remove any excess thermal grease from the blocks with a straight edge.e. Place the bottom plate over the TED mounting blocks.
f. Install the 12 screws that secure the six TED spacer blocks to the bottom plate.Alternate tightening each screw, a little at a time to 1.7 N–m (15 in–lb) on the TED spacer block.
g. Wait three minutes and then tighten the 12 screws to 1.7 N–m (15 in–lb).
23. Assemble the reagent compartment: a. Guide the TED wires through the reagent compartment. b. Pull the TED wires away from the reagent compartment.c. Ensure the TED wires are routed inside the reagent compartment and install the
bottom plate into the reagent compartment.Tighten the 11 screws to 0.68 N–m (6 in–lb) on the TED spacer block.
d. Install the TED connector into the TED wire cover.e. Fold the TED wires inside the cover and secure the TED wire cover to the reagent
compartment.f. Route the thermistor cables through the left side of the reagent compartment and
connect them to the primary reagent compartment PC board.g. Install the fan into the primary reagent compartment.h. Install the coldsink cover onto the cold sink. i. Clean the grease from the mixer bearing.j. Apply a thin layer of lithium grease (part number 572662) on the mixer bearing.k. Install the primary reagent mixer to the mixer bearing.l. Install the top cover and the probe shutter to the primary reagent compartment.
24. Install the primary reagent compartment:a. Place the primary reagent compartment on the frame leaving approximately
152 mm (6 in) behind the primary reagent compartment.b. Install the two rinse stations.c. Install the tubing from the two drain ports.d. Connect the TED power cable from the connector.
CAUTIONEnsure the TED mounting blocks are installed flush with the bottom plate or cooling problems will occur.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Primary Reagent
078D0080-02, Rev. A
e. Install the four screws that secure the primary reagent compartment to the frame.f. Connect the sensor motor interconnect cable, thermistor cable, fan cable, LED
upper cable, and LED lower cable.g. Install the back left cover.
25. Turn the system on:a. Plug in the power cord.b. At the back of the system, turn the power on.c. At the back of the system, press the computer startup switch.
26. Calibrate the reagent probe shutter.
27. Calibrate the reagent mixer.
28. Calibrate the reagent probes.
29. Calibrate the primary reagent compartment thermals.
078D0080-02, Rev. A
Primary Reagent Compartment Door 571104B
Primary Reagent Compartment Door Assembly
Figure 1-1 Primary Reagent Compartment Door Assembly
1-280 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Primary Reagent Compartment
078D0080-02, Rev. A
Figure 1-2 Primary Reagent Compartment Door Assembly
078D0080-02, Rev. A
Parts List – Primary Reagent Compartment Door 571104BItem Description REF Part Number1 Handle, Reagent Door 5724572 Door, Reagent Hinged, ADVIA Centaur
System571564B
2 Door, Reagent Hinged, ADVIA Centaur XP System
078-0380-01
3 Screw, Cap-Head M4 x 0.7 x 20 mm 60012x514 Cover, LED Display 5717355 Plate, Reagent Door Closer, ADVIA
Centaur System7835548 572256B
5 Plate, Reagent Door Closer, ADVIA Centaur XP System
078-0379-01
6 Screw, Button M4 x 0.7 x 6 mm 572500
7 Screw, Cap-Head M4 x 0.7 x 6 mm Hex 60013x28 Switch, Tilt 267709 Bracket, Tilt Switch 078-0029-0110 Panel, Rear Reagent Door, ADVIA Centaur
System572253
10 Panel, Rear Reagent Door, ADVIA Centaur XP System
078-0378-01
11 Screw, Button-Head M4 x 0.7 x 12 mm 57249912 Seal, Hinge Housing 572480B13 Display, LED 5255676 571955B
PC Board, Reagent Pack LED Array
575387
Panel, LED Display 571943Screw, Cap-Head M3 x 0.5 x 6 mm 60012x1
14 Fastener, Loop 57245615 Fastener, Hook 57247916 Screw, Flat-Head M3 x 0.5 x 16 mm 60055x8517 Thermometer 12019818 Clamp 12811x1
078D0080-02, Rev. A
Reagent Barcode Scanner 571103
Reagent Barcode Scanner Assembly
1-284 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Belt Clamp Detail
078D0080-02, Rev. A
Belt Clamp Detail
Sensor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-285
078D0080-02, Rev. A
Motor Detail
1-286 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Reagent Barcode Scanner 571103CItem Description REF Part Number1 Motor, Stepper 4VDC 9864200 5729152 Plate, Barcode Idler 078-0252-013 Scanner, Barcode Head 6912824 571259E4 Bracket, Barcode Head 078-0085-015 Screw, Cap-Head 6/32 x 0.31 Slotted Steel 800-0178-016 Screw, Flat-Head M3 x 0.5 x 8 mm 300150 60006x437 Screw, Cap-Head M3 x 0.5 x 8 mm 60012x88 Screw, Cap-Head M3 x 0.5 x 6 mm 60012x1559 Clamp, Belt 57179010 Screw, Shoulder #10-0.25 x 0.375 x 0.125 57268611 Pulley, Idler 7316923 114473
Pulley 571683Bearing (Quantity 2) 6335851 23254
12 Sensor, Optical Switch 7713183 11167113 Screw, Cap-Head M3 x 0.5 x 6 mm 60012x114 Plate, Motor Mount 57179715 Washer, Flat M4 Zinc 60028x516 Screw, Cap-Head M4 x 0.7 x 35 mm 60014x7917 Guide, Linear Barcode Assembly 078-0084-0118 Belt, Timing 3 mm 57179119 Pulley, Timing Belt 57179220 Screw, Pan-Head #4-40 x 0.438 Phillips 57250221 Guide, Chain 11098922 Screw, Thumb 9867382 572422
078D0080-02, Rev. A
Reagent Probe Guide 571110B
Reagent Probe Guide Assembly
1-290 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe Guide Assembly
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-291
078D0080-02, Rev. A
Motor Detail
078D0080-02, Rev. A
Parts List – Reagent Probe Guide 571110BItem Description REF Part Number1 Screw, Cap-Head M4 x 0.7 x 12 mm 60012x232 Motor, Stepper 6VDC 2124384 5729133 Pulley, Timing 6398047 5715884 Chain Guide, Probe 1, Upper 2511825 1109795 Chain Guide, Probe 3, Upper 9804062 1109826 Chain Guide, Probe 2, Upper 1056989 1109817 Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 300150 60006x438 Screw, Cap-Head M4 x 0.7 x 18 mm 60012x449 Not Used10 Not Used11 Not Used12 Pulley, Idler 7102842 078-B006-0113 Not Used14 Not Used15 Guide, Linear 8274116 57157416 PC Board, Reagent Probe, Pack-to-Pack 5175982 57534717 Plate, Belt Mounting 11902818 Screw, Flat-Head M3 x 0.5 x 10 mm
Phillips60006x63
19 Screw, Cap-Head M3 x 0.5 x 8 mm 60012x820 Plate, Adapter 57157521 Screw, Cap-Head M4 x 0.7 x 6 mm 60012x222 Guide, Linear 57157323 Plate, Adapter, Probe 2 57158924 Belt, 0.200P x 0.375W 2387431 571576
078D0080-02, Rev. A
Reagent Probe 1 571107C
Reagent Probe 1 Assembly
1-296 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe 1 Assembly
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe 1 Tubing Routing 1-297
078D0080-02, Rev. A
Reagent Probe 1 Tubing Routing
Motor Detail
Sensor Detail
1-298 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 1-299
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Reagent Probe 1 571107CItem Description REF Part Number1 Screw, Pan-Head M3 x 0.5 x 6 mm Phillips 60000x432 Spacer 6362506 266263 PC Board, Probe 1 Horizontal Interconnect 9489841 575432B4 Screw, Cap-Head M4 x 0.7 x 12 mm 60012x235 Motor, Stepper 4VDC 6300918 5729056 Plate, Motor Mounting 5712247 Screw, Flat-Head #4-40 60031x4068 Sensor, Optical Switch 5566949 1125859 Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 300150 60006x4310 PC Board, Reagent Probe 1 Heater Control
and Volume Detect9841251 575402B
11 Screw, Pan-Head #4-40 x 0.438 Phillips 57250212 Gear, Spur 24P 16T 0.250W 0.187B 57122613 Pulley, 40DP 10T 4692398 57123014 Screw, Special Hex 7817930 57123615 Spacer, 0.125 x 0.187 x 0.010 Stainless 2662516 Bearing, Ball Flanged 0.124 x 0.312 x
0.1406571222
17 Not Used18 Not Used19 Shaft, Guide 57121920 Leadscrew & Follower 5483601 57121821 Guide, Block Follower 1096417 57121722 Probe, Reagent 4042237 57123323 Sleeve, Probe 078-0087-0224 Bracket, Cable Carrier 571239B
Grommet Strip 1229825 Block, Cable Bracket Mounting 57124026 Guide, Linear Reagent Probe 1873480 571214B27 Washer, Lock M4 Zinc (only use with item
29 on older systems with the reagent probe assembly mounted with two screws)
60026x5
28 Washer, Flat M4 Zinc 60028x5
1-302 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Reagent Probe 1
078D0080-02, Rev. A
29 See note below:NOTE: Screw, Cap-Head M4 x 0.7 x 5 mm (only use on older systems with the reagent probe assembly mounted with two screws)
60012x2
NOTE: Screw, Cap-Head M3 x 0.5 x 5 mm (only use on systems with the reagent probe assembly mounted with four screws with Loctite part number 17414x1)
572417
30 Screw, Pan-Head M3 x 0.5 x 8 mm Phillips 60000x5331 Belt, 40DP 0.188W 6034835 57122332 Pulley, 40DP 36T 0.188W 0.188B 4806784 2608333 Plate, R1 Pulley Guard 078-0063-0134 Flags-Sensor, Side/Side Probe 1 078-0019-0135 Rack, 24DP 571215B
36 Mount, Rail and Rack 57121637 Bracket, Upper Chain 571231B38 Guide, Chain, Probe 1 Lower 11098339 Kit, Heater Coil w/ Bubble Detector 2368984 078-K015-03
Sensor, Volume Detector 113692Insulator 571238Gasket, Volume Detection 9051595 116314Loctite, Adhesive 18006Bracket, Reagent Bubble Detector 078-0051-01
40 Casting, Machined Probe 1 Carriage 11526941 Fitting, Barbed 518-0877-0142 Tubing, Silicon 9991636 562-3070-01
Fastener, Ty-wrap 615801 25029x85NOTE: For replacement of the reagent probe heater coil for Reagent Probe 1, 2 or 3, refer to eBRITE for Reagent Probe Heater Coil Assembly Installation Instructions.
Item Description REF Part Number
078D0080-02, Rev. A
Reagent Probe 2 571108C
Reagent Probe 2 Assembly
1-304 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe 2 Tubing Routing
078D0080-02, Rev. A
Reagent Probe 2 Tubing Routing
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-305
078D0080-02, Rev. A
Motor Detail
Sensor Detail
1-306 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 1-307
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Reagent Probe 2 571108CItem Description REF Part Number1 Screw, Cap-Head M4 x 0.7 x 6 mm 60012x22 Washer, Lock M4 (only use with item 15 on
older systems with the reagent probe assembly mounted with two screws)
60026x5
3 Washer, Flat M4 60028x54 Bracket, Cable Carrier, Probe 2 571244B5 Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 300150 60006x436 Guide, Chain, Probe 2 Lower 1109847 Screw, Pan-Head M3 x 0.5 x 6 mm 60000x438 Guide, Linear 1873480 571214B9 Bracket, Upper Chain 571247B10 Mount, Rail & Rack 57121611 Clamp, Cable 7295993 26870
Loctite, Adhesive 18006x112 Rack 24 DP 571215B13 Flags, Sensor Side/Side 078-0020-0114 Not Used15 See note below:
NOTE: Screw, Cap-Head M4 x 0.7 x 6 mm (only use on older systems with the reagent probe assembly mounted with two screws)
60012x2
NOTE: Screw, Cap-Head M3 x 0.5 x 5 mm (only use on systems with the reagent probe assembly mounted with four screws with Loctite part number 17414x1)
572417
16 Spacer 6362506 2662617 PC Board, Probe 2 Horizontal Interconnect 1363407 57543718 Gear, Spur 24P 16T 0.250W 0.187B 571226B19 Screw, Pan-Head #4-40 x 0.438 Phillips 57250220 PC Board, Reagent Probe 2 Heater Control
and Volume Detect Sensor2565887 575407C
21 Screw, Flat-Head #4-40 x 0.38 Phillips 60031x40622 Plate, Motor Mounting 57122423 Motor, Stepper 4VDC 6300918 572905
1-310 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Reagent Probe 2
078D0080-02, Rev. A
24 Pulley, 40DP 10T 4692398 57123025 Screw, Special Hex 7817930 57123626 Spacer, 0.125 x 0.187 x 0.010 Stainless 2662527 Bearing, Ball Flanged, 0.124 x 0.312 x
0.1406571222
28 Not Used29 Not Used30 Shaft, Guide 57121931 Leadscrew and Follower 5483601 57121832 Guide, Block Follower 6362506 2662633 Probe, Reagent 57123334 Sleeve, Probe 078-0087-0235 Belt 6034835 57122336 Pulley, 40 DP 4806784 2608337 Sensor, Optical Switch 11258538 Kit, Heater Coil w/ Bubble Detector 078-K015-03
Sensor, Volume Detector 113692Insulator 571238Gasket, Volume Detection 9051595 116314Loctite, Adhesive 18006Bracket, Reagent Bubble Detector 078-0051-01
39 Not Used40 Casting, Machined Probe 2 Carriage 11527241 Fitting, Barbed 518-0877-0142 Tubing, Silicon 9991636 562-3070-01
Fastener, Ty–wrap 615801 25029x85NOTE: For replacement of the reagent probe heater coil for Reagent Probe 1, 2 or 3, refer to eBRITE for Reagent Probe Heater Coil Assembly Installation Instructions.
Item Description REF Part Number
078D0080-02, Rev. A
Reagent Probe 3 571109C
Reagent Probe 3 Assembly
1-312 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Reagent Probe 3 Tubing Routing
078D0080-02, Rev. A
Reagent Probe 3 Tubing Routing
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 1-313
078D0080-02, Rev. A
Motor Detail
Sensor Detail
1-314 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 1-315
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Reagent Probe 3 571109CItem Description REF Part Number1 Screw, Cap-Head M4 x 0.7 x 12 mm 60012x232 Washer, Lock M4 Zinc (only use with item
16 on older systems with the reagent probe assembly mounted with two screws)
60026x5
3 Washer, Flat M4 Zinc 60028x54 Screw, Pan-Head M3 x 0.5 x 6 mm Phillips 60000x435 Plate, R3 Pulley Guard 078-0064-016 Bracket, Cable Carrier Probe 3 571250
Grommet Strips 122987 Screw, Flat-Head M3 x 0.5 x 6 mm Phillips 300150 60006x438 Bracket, Upper Chain 571247B9 Mount, Rail & Rack 57121610 Rack, 24DP 571215B11 Guide, Linear Reagent Probe 1873480 571214B12 Flags, Sensor Side/Side 078-0021-0113 Pulley, 40DP 36T 0.188W 0.188B 4806784 2608314 Belt, 40DP 0.188W 6034835 57122315 Screw, Pan-Head M3 x 0.5 x 8 mm Phillips 60000x5316 See note below:
NOTE: Screw, Cap-Head M4 x 0.7 x 6 mm (only use on older systems with the reagent probe assembly mounted with two screws)
60012x2
NOTE: Screw, Cap-Head M3 x 0.5 x 5 mm (only use on systems with the reagent probe assembly mounted with four screws with Loctite part number 17414x1)
572417
17 Spacer 6362506 2662618 PC Board, Probe 3 Horizontal Interconnect 1591744 575442B19 Gear, Spur 24P 16T 0.250W 0.187B 571226B20 Screw, Pan-Head #4-40 x 0.438 Phillips 57250221 PC Board, Reagent Probe 3 Heater Control
Volume Detect1153232 575412C
22 Screw, Flat-Head #4-40 x 0.38 Phillips 60031x40623 Plate, Motor Mounting 571224
1-318 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Reagent Probe 3
078D0080-02, Rev. A
24 Motor, Stepper 4VDC 6300918 57290525 Screw, Special Hex Head #8-32 Nylon 7817930 57123626 Not Used27 Not Used28 Leadscrew & Follower 5483601 57121829 Probe, Reagent 4042237 57123330 Sleeve, Probe 078-0087-0231 Bearing, Ball Flanged 0.125 x 0.3125 x
0.1406571222
32 Guide, Block Follower 1096417 57121733 Shaft, Guide 57121934 Spacer, Inner Race 0.125 x 0.187 x 0.010 2662535 Pulley, 40DP 10T 0.188W 0.125B 4692398 57123036 Guide, Chain, Probe 3 Lower 11098637 Sensor, Optical Switch 5566949 11258538 Kit, Heater Coil w/ Bubble Detector 2368984 078-K015-03
Sensor, Volume Detector 113692Insulator 571238Gasket, Volume Detection 9051595 116314Loctite, Adhesive 18006Bracket, Reagent Bubble Detector 078-0051-01
39 Not Used40 Casting, Machined Probe 3 Carriage 11527541 Fitting, Barbed 518-0877-0142 Tubing, Silicon 9991636 562-3070-01
Fastener, Ty-wrap 615801 25029x85NOTE: For replacement of the reagent probe heater coil for Reagent Probe 1, 2 or 3, refer to eBRITE for Reagent Probe Heater Coil Assembly Installation Instructions.
Item Description REF Part Number
078D0080-02, Rev. A
Ancillary Queue for the ADVIA Centaur System 078-B035-01
Ancillary Queue Assembly
1-320 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Queue Assembly
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 1-321
078D0080-02, Rev. A
Sensor Detail
1-322 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail 1-323
078D0080-02, Rev. A
Motor and Pulley Detail
078D0080-02, Rev. A
Parts List – Ancillary Queue 078-B035-01 for the ADVIA Centaur System
Item Description REF Part Number1 Screw, Pan-Head M5 x 0.8 x 16 mm 60004x862 Washer, Flat, M5 60056x63 Cover, Ancillary Queue 571924B4 Screw, Flat-Head M5 x 0.8 x 16 mm 60055x865 Screw, Cap-Head M4 x 0.7 x 10 mm 06903884 60013x236 Washer, Flat M4 09269183 60056x57 Block, Mounting 5722088 Screw, Cap-Head M4 x 0.7 x 16 mm 02960484 60013x29 Mount, Sensor 572115B10 Sensor, Optical Switch 00962374 111674
11 Not used12 Spring, Ancillary Clamp 09397904 078-0038-0113 Clamp, Bearing 078-0077-0114 Holder, Ancillary Pack 078-0079-0115 Gear, Spur 57210816 Base, Ring 02052758 078-0080-0117 Thermistor, Threaded Body 1849083 571184C
Thermal Grease 09770826 12105518 Spacer, TED 572118B19 TED, Assembly 07583492 111679
Clamp 08823500 1831220 Insulator, Ancillary Queue, ADVIA
Centuar System572124C
21 Sink, Heat, ADVIA Centaur System 57211022 Washer, Shoulder 05286512 57210623 Screw, Cap-Head M4 x 0.7 x 20 mm 08373815 60013x5124 Motor, Stepper 4VDC 01943640 57292425 Plate, Motor Mounting 078-0038-0126 Screw #4-40 x .31 835-2127-0127 Screw, Cap M4 x 06 Hex 01689043 60013x228 Screw, Pan-Head M4 x 0.7 x 6 mm 04380558 60003x45
29 Scanner, Barcode 03369739 571259E30 PC Board, Ancillary Queue Sensor 01195296 575222B31 Not used32 Screw, Self-Tapping, #4-20 x 0.38 57277333 Bracket, Barcode 572202D
Tube, Drainage 572206B
Item Description REF Part Number
078D0080-02, Rev. A
Ancillary Queue for the ADVIA Centaur XP System 078-B035-01
Ancillary Queue Assembly
1-328 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Queue Assembly
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 1-329
078D0080-02, Rev. A
Sensor Detail
1-330 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail 1-331
078D0080-02, Rev. A
Motor and Pulley Detail
078D0080-02, Rev. A
Parts List – Ancillary Queue 078-B069-01 for the ADVIA Centaur XP System
Item Description REF Part Number1 Screw, Pan-Head M5 x 0.8 x 16 mm 60004x862 Washer, Flat, M5 60056x63 Cover, Ancillary Queue 00474949 078-0279-014 Screw, Flat-Head M5 x 0.8 x 16 mm 60055x865 Block, Mounting 5722086 Screw, Cap-Head M4 x 0.7 x 20 mm 08373815 60013x517 Washer, Flat M4 09269183 60056x58 Mount, Sensor 01988466 078-0278-019 Sensor, Optical Switch 03818185 21470x110 Screw, Self-Tapping, #4-20 x 0.38 572773
11 Screw, Button M3 x 0.5 x 6 mm 02683391 57249612 Bracket, Flag 01654495 078-0277-0113 Spring, Ancillary Pack Clamp 07015613 078-0283-0114 Standoff, M3 x 25 mm 558-0869-0615 Clamp, Bearing 01166571 078-0291-0116 Holder, Ancillary Pack 06088064 078-0281-0117 Gear, Spur 078-0387-0118 Motor, Stepper 4VDC 03159785 078-0388-0119 Screw, Cap-Head M4 x 0.7 x 10 mm 06903884 60013X1620 Screw, Flat-Head, M 6 x 20 846-1032-6221 Base, Ring 08497077 078-0276-0122 TED, Assembly 08827956 571505B
Clamp, Cable, Flat 08823500 1831223 Sink, Heat 03714916 078-0280-0124 Screw, Cap M4 x 06 Hex 01689043 60013x225 Scanner, Barcode 3369739 571259E26 Bracket, Ancillary Queue Motor 078-0430-0127 Bracket, Barcode 572202D28 Bracket, PC Board 57212329 PC Board, Ancillary Queue Sensor 01195296 575222B30 Screw, Pan-Head M4 x 0.7 x 6 mm 03545693 60000x45
1-334 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Ancillary Queue
078D0080-02, Rev. A
31 Standoff, Ring Base 00576148 078-0284-0132 Thermistor, Threaded Body 01849083 571184C
Thermal Grease 12049133 Support, Thermistor, Insulation 00217709 078-0285-0134 Washer, Shoulder #8 05286512 572106
Tube, Drainage 03214786 078-0286-01
Item Description REF Part Number
078D0080-02, Rev. A
Replacing the Ancillary Queue TEDs
1. Remove the ancillary reagent packs:a. At the workspace, select System Status.b. Select Diagnostic Tools.c. Select Empty Ancillary Queue. d. Remove the ancillary reagent packs and store them in a refrigerator.
2. Turn off the system:a. At the Workspace, select the system symbol and then select Turn System Off.b. At the back of the system, turn off the main power and unplug the power cord.
3. Remove the ancillary reagent queue.
NOTE: Label all cable connections before removing the cables.
4. Remove the four screws that secure the ancillary entry to the baseplate.
5. Remove the ancillary queue drive motor.
6. Remove the ancillary entry.
7. Remove the defective TEDs:
CAUTIONFailure to remove the ancillary reagent packs before replacing the TEDs will damage the system.
CAUTIONFailure to remove the ancillary drive motor before removing the baseplate will damage the system.
CAUTIONDisconnect ancillary queue motor and sensor cables before removing the bottom plate or damage can occur to the system.
336ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Ancillary Queue TEDs
078D0080-02, Rev. A
a. Remove the three screws that secure the bottom plate to the standoffs.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Ancillary Queue TEDs
078D0080-02, Rev. A
b. Remove the four screws and nylon washers that secure the heatsink to the bottom of the ancillary queue.
c. Remove the two TEDs.d. Remove the two TED spacer plates.e. Clean the old thermal grease from each spacer plate.f. Apply new thermal grease to the chamfered side of each spacer plate.g. Install the spacer plates with the chamfered side flush with the bottom of the
ancillary queue.
8. Loosen the eight screws that secure the ducting to the heatsink.
338ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Replacing the Ancillary Queue TEDs
078D0080-02, Rev. A
9. Install the TEDs:
a. Apply a thin layer of thermal grease to both sides of the new TEDs.
b. Install the new set of TEDs on the bottom of the preheater ensuring the red (R) and black (B) wires are in the same position as the old TEDs.
c. Install the TED heatsink and tighten the four screws. Tighten the two inside screws first and then the two outside screws to 0.68 N-m (6 in-lb) on the TED heat sink.
d. Wait three minutes and then tighten the two inside screws first and then the two outside screws to 0.68 N-m (6 in-lb).
10. Install the three screws that secure the baseplate to the standoffs.
11. Place the ducting flat against the baseplate.
12. Tighten the eight screws that secure the ducting to the heatsink.
13. Align and install the ancillary entry.
14. Connect the motor and sensor cables.
15. Align and install the ancillary queue motor.
16. Install the ancillary queue.
17. Turn the system on:a. Connect the power cord.b. At the back of the system, turn the power on.c. At the back of the system, press the computer startup switch.
18. Calibrate the ancillary probe.
19. Calibrate the ancillary queue thermals.
CAUTIONEnsure that no thermal grease is inside the TED or heating problems will occur.
CAUTIONEnsure the TEDs are installed flush with the bottom of the ancillary queue or damage can occur to the TEDs.
CAUTIONFailure to alternate the tightening of each screw or overtightening the screws on the mounting block will damage the TEDs.
078D0080-02, Rev. A
Ancillary Entry 078-B036-01
Housing
340 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Entry Motor
078D0080-02, Rev. A
Ancillary Entry Motor
Leadscrew Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 341
078D0080-02, Rev. A
Sensor Detail
Motor and Pulley Detail
342 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail 343
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Ancillary Entry 078-B036-01Item Description REF Part Number1 Screw, #4-20 Self-Tap 5727732 Top Guide, Sliding Door 572189B3 Door, Ancillary Infeed 078-0082-014 Insulation 078-0171-015 Gear, Spur 5721926 Screw, #4-40 x 0.625 5727747 Washer, Flat 0.125 x 0.280 x 0.032 25042x638 Screw, Cap-Head M4 x 0.7 x 6 mm 01689043 60013x29 Washer, Flat M4 04563393 60028x510 Screw, Cap-Head M3 x 0.5 x 10 mm 60013x1511 PC Board, Pack Sensor 05981962 575457B12 Bracket, Pack Sensor Board 572200D13 Rod, Anti-pivot 57219614 Housing, Infeed 572188B15 Spacer, 0.194 x 0.500 x 0.031 05762977 57218716 Pulley, Timing Belt, 40DP 6043184 571628B17 Pulley, Timing Belt, 40P 57219718 Belt 6301817 57219819 Motor, Stepper 4VDC 57219620 Foam, Strip 572201B 21 Bearing, Ball 0.188 x 0.375 x 0.125 07236865 57201922 Sensor, Optical Switch 00962374 11167423 Screw, Pan-Head M3 x 0.5 x 5 mm 60001x3324 PC Board, Ancillary Entry 02792883 575232D25 Screw, Flat-Head M3 x 0.5 x 10 mm 60007x6326 Leadscrew Follower Assembly 01487173 078-B019-01
Leadscrew follower bar nut27 Arm 078-0052-0128 Washer, #4 05094672 818-0053-0129 Screw, #4-40X 848-0128-01 848-0128-0130 Bracket, Infeed 078-0385-01
078D0080-02, Rev. A
Inprocess and Exit Queue 571040B
Motor Location
348 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Gap
078D0080-02, Rev. A
Motor Gap
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail 349
078D0080-02, Rev. A
Motor and Pulley Detail
078D0080-02, Rev. A
Parts List – Inprocess and Exit Queue 571040BItem Description REF Part Number1 Screw, Cap-Head M5 x 0.8 x 8 mm 60013x102 Support, Tip Tray 7871447 571699B3 Screw, Flat-Head M3 x 0.5 x 8 mm 60055x534 Plate, Positioning, Sample Rack 571968B5 Screw, Flat-Head M4 x 0.7 x 10 mm 60055x656 Spring, Positioning 7946811 5719767 Screw, Button M3 x 0.5 x 6 mm Socket
Steel572496
8 Bracket, Flag Strip 571971B9 Washer, Lock Int Tooth M3 Zinc 60027x310 Screw, Cap-Head M3 x 0.5 x 6 mm 7882579 60013x1
11 Screw, Pan-Head M4 x 0.7 x 12 mm 60003x7512 PC Board, Inprocess Queue Tip Sensor 8744694 575217B13 Washer, Lock Int Tooth M4 Zinc 60027x514 Screw, Cap-Head M4 x 0.7 x 16 mm 60013x3715 Washer, Flat M4 Zinc 60056x516 Screw, Cap-Head M4 x 0.7 x 8 mm 60013x917 Guide, Linear, Inprocess Queue 57196518 Mounting, Support 571969B19 Ring, Retaining 7937995 1177420 Spacer, 0.2525 x 0.310 x 0.051 2327721 Clamp, Belt Support 57197022 Clamp, Timing Belt 57197423 Pulley, Timing, 0.200P 30G 0.438W
0.500B916127 114477
Pulley 571967Bearing (Quantity 2) 4308474 26541
24 Pulley, Timing, 0.200P 20G 0.438W 0.250B
6811300 571973
25 Plate, Motor Mounting 57197726 Belt, Timing, 0.200P 0.38W 296G
Neoprene6506508 571966
27 Spacer, 0.250 x 0.375 x 0.375 Stainless 571978
352 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Inprocess and Exit
078D0080-02, Rev. A
28 PC Board, Rack Present 6438634 S575292B29 Latch, Draw Adjustable Non-corrosive 57210130 PC Board, Exit Queue Interconnect 1346863 575302C31 Motor, Stepper 6VDC 4616756 57290932 Screw, Cap-Head M4 x 0.7 x 10 mm 60013x1633 Cover, Exit Queue Back 57196334 Tray, Exit Queue 571004
Item Description REF Part Number
078D0080-02, Rev. A
Sample and Ancillary Probe for the ADVIA Centaur System 078-B045-01
Ancillary Motor and Belt Location
354 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Motor and Belt Location
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Motor and Belt Location 355
078D0080-02, Rev. A
Sample Motor and Belt Location
356 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Probe Assembly Detail
078D0080-02, Rev. A
Sample Probe Assembly Detail
Vertical Motor Location
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Probe Assembly Detail 357
078D0080-02, Rev. A
Ancillary Probe Assembly Detail
Vertical Motor Location
358 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Diluter Detail
078D0080-02, Rev. A
Sample Diluter Detail
Ancillary Diluter Sample Air Pump Assembly Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 115401 359
078D0080-02, Rev. A
Diluter 2 mL 115401
360 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 115401
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri–Continent Diluter 572888 361
078D0080-02, Rev. A
Tri–Continent Diluter 572888
Motor and Pulley Detail
362 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail
078D0080-02, Rev. A
078D0080-02, Rev. A
Sample and Ancillary Probe for the ADVIA Centaur XP System 078-B123-01
Ancillary Motor and Belt Location
364 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Motor and Belt Location
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Motor and Belt Location 365
078D0080-02, Rev. A
Sample Motor and Belt Location
366 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Probe Assembly Detail
078D0080-02, Rev. A
Sample Probe Assembly Detail
Vertical Motor Location
Ancillary Probe Assembly Detail
Vertical Motor Location
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Ancillary Probe Assembly Detail 367
078D0080-02, Rev. A
368 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Diluter Detail
078D0080-02, Rev. A
Sample Diluter Detail
Ancillary Diluter Sample Air Pump Assembly Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 115401 369
078D0080-02, Rev. A
Diluter 2 mL 115401
370 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Diluter 2 mL 115401
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Tri–Continent Diluter 572888 371
078D0080-02, Rev. A
Tri–Continent Diluter 572888
Motor and Pulley Detail
372 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Sample and Ancillary Probe 078-B123-01 for the ADVIA Centaur XP System
Item Description REF Part Number1 Pulley, Timing Belt, 0.080P 15T 0.41W
0.188B5232986 572209B
2 Motor, Stepper 4VDC 5192771 5729143 PC Board, Sample and Ancillary Probe
GIOB9822435 S575322H
4 Belt, Timing, 0.080P 0.375W 487T Neoprene
4513760 572224
5 Ring, Retaining 7937995 117746 Spacer, 0.2525 x 0.310 x 0.051 232777 Plate, Sample Ancillary Probe Pulley 7879650 078-B039-018 PC Board, Sample and Ancillary Probe
Horizontal Travel2070780 575212C
9 Motor, Stepper 4VDC 6300918 57290510 Belt, Timing, 0.080P 0.375W 235T
Neoprene6902853 572223
11 Bearing, Ball, Flanged 0.312 x 0.500 x 0.156 Shielded
4811788 572242
12 Leadscrew and Follower 078-0039-01Flag, Follower 571464BWasher, Flat M4 60056x5Screw, Cap M4 x 10 60013x16
13 Screw, Cap-Head M5 x 0.8 x 10 mm Hex 60013x1714 Sensor, Bubble Detector 9457664 571954C15 Support, Ancillary Probe 9580040 078-0040-0116 Probe, Ancillary 4042237 57123317 Sleeve, Ancillary Probe 078-0087-0218 Pulley, Timing Belt, 40DP 16G 0.312W
0.188B6043184 571628B
19 Belt, Timing, 40P 0.312W 105T Polyurethane
9327949 572215B
20 Pulley, Timing, 40DP 48G 0.35W 0.188B 4737391 57159821 Motor, Stepper 4VDC 572924
374 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample and Ancillary
078D0080-02, Rev. A
22 Screw, Flat-Head #4-40 x 0.25 Slotted 25040x45
23 PC Board, Ancillary Probe Vertical S575282C
24 Bearing, Ball, Flanged 0.188 x 0.375 x 0.125 Shielded
7236865 572019
25 Screw, Pan-Head M2.5 x 0.4 x 10 mm Phillips
60000x62
26 PC Board, Sample Pressure Level Sense 5545666 575462E27 Plate, Mounting 11580828 Not Used29 Plunger, Sample, Assembly 3570019 117735
Body, Plunger 6663735 118345Spring, Comp SS Lee LC-029C-14 8999323 572463BPlunger 1337473 572217CRing, Internal Retaining E-RIX-7 5047542 572218Grease 20920
30 Pulley, 40DP 24T 0.312W 0.188B 3398496 118344PC Board, Sample Probe Vertical 6821152 118365
32 Screw, Pan-Head #6-32 x 0.20 2328133 Spacer 0.380 OD x 0.252 ID 2328034 Gasket, O-ring, 0.380 x 0.060 8793466 1157035 Diluter, Sample 250 µL1458x1736 Syringe, Sample 250 µL 1776434 2613437 Guide, Linear, Probe Upper 5341874 572032B38 Guide, Linear, Ancillary Probe 2447566 572030B39 Guide, Linear, Probe Lower 9129802 572031B40 Plunger, 2 mL, Tri-Continent 7121359 11224841 Not Used42 Filter, Level Sense 8675129 11209543 Connector, T 20GA 3544034 57222044 Pump, Sample Probe Fluid Sense Kit 8255995 11166945 Kit, Valve, Solenoid, Two-Way, V23, V24 078-K013-0146 Damper, Pulsation 1630340 571042
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample and Ancillary Probe
078D0080-02, Rev. A
47 Plate, Motor Mounting 571465B48 Not Used49 Carriage 572120B50 Fitting, Quick Disconnect 1179975 2355951 Fitting, Luer-20GA 1879357 11026852 Spacer 0.187 x 0.250 x 0.063 11026753 Carriage, Sample and Ancillary 7041363 113764
Frame, Carriage 572129CBearing 4811788 572242Bearing (Quantity 2) 7236865 572019Leadscrew and Follower 572130Rod 571570B
54 Rinse Station, Ancillary Probe 2807554 11026555 Screw, Soc. Head Cap M4 x 8 SST 60013x956 Bracket, Motor Mounting 078-0007-0157 Screw, Binding #4-40x 0.31 13002x158 Screw, Flat M4 x 8 mm 60055x5559 Screw, Flat M3 x 0.5 x 12 mm 60007x7360 Clamp, Belt 571466
Bracket 572138B61 Clamp, Belt 572136B
Bracket 57213762 Screw, Cap M3 x 0.5 x 10 mm 60013x1563 Screw, M3 x 0.5 x 8 mm 60013x864 Screw, Cap M4 x 0.7 x 10 mm 60013x1665 Washer, Flat M4 60056x566 Motor, Diluter 1458x29567 Clamp, Rinse Station 2669968 Washer, Flat M4 60056x569 Washer, Lock M4 60027x570 Screw, Cap M4 x 0.7 x 16 mm 60013x3771 Washer, Flat M3 60056x3
Item Description REF Part Number
376 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample and Ancillary
078D0080-02, Rev. A
72 Diluter, 2 mL, Note: The new style diluter is interchangeable with the Tri-Continent diluter.
115401
73 Fitting (two piece insert with barb) 11008074 Chamber, Fluid 3275688 11539875 O-Ring 0.614ID 070W Ethylene propylene 1926711 11594576 Seal, 2.0 mL, 0.500 ID 7026771 11539577 PC Board, Diluter 11483078 Screw #4-40 x 187 60030x40379 Nut, Chamber 1718884 11563380 Screw, Cap-Head M4 x 12 60013x23
Item Description REF Part Number
078D0080-02, Rev. A
Parts List – Sample and Ancillary Probe 078-B045-01Item Description REF Part Number1 Pulley, Timing Belt, 0.080P 15T 0.41W
0.188B5232986 572209B
2 Motor, Stepper 4VDC 5192771 5729143 PC Board, Sample and Ancillary Probe
GIOB9822435 S575322H
4 Belt, Timing, 0.080P 0.375W 487T Neoprene
4513760 572224
5 Ring, Retaining 7937995 117746 Spacer, 0.2525 x 0.310 x 0.051 232777 Plate, Sample Ancillary Probe Pulley 7879650 078-B039-018 PC Board, Sample and Ancillary Probe
Horizontal Travel2070780 575212C
9 Motor, Stepper 4VDC 6300918 57290510 Belt, Timing, 0.080P 0.375W 235T
Neoprene6902853 572223
11 Bearing, Ball, Flanged 0.312 x 0.500 x 0.156 Shielded
4811788 572242
12 Leadscrew and Follower 078-0039-01Flag, Follower 571464BWasher, Flat M4 60056x5Screw, Cap M4 x 10 60013x16
13 Screw, Cap-Head M5 x 0.8 x 10 mm Hex 60013x1714 Sensor, Bubble Detector 9457664 571954C15 Support, Ancillary Probe 9580040 078-0040-0116 Probe, Ancillary 4042237 57123317 Sleeve, Ancillary Probe 078-0087-0218 Pulley, Timing Belt, 40DP 16G 0.312W
0.188B6043184 571628B
19 Belt, Timing, 40P 0.312W 105T Polyurethane
9327949 572215B
20 Pulley, Timing, 40DP 48G 0.35W 0.188B 4737391 57159821 Motor, Stepper 4VDC 57292422 Screw, Flat-Head #4-40 x 0.25 Slotted
25040x45
378 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample and Ancillary
078D0080-02, Rev. A
23 PC Board, Ancillary Probe Vertical S575282C
24 Bearing, Ball, Flanged 0.188 x 0.375 x 0.125 Shielded
7236865 572019
25 Screw, Pan-Head M2.5 x 0.4 x 10 mm Phillips
60000x62
26 PC Board, Sample Pressure Level Sense 5545666 575462E27 Plate, Mounting 11580828 Not Used29 Plunger, Sample, Assembly 3570019 117735
Body, Plunger, ADVIA Centaur System
6661600 572133B
Spring, Comp SS Lee LC-029C-14 8999323 572463BPlunger 1337473 572217CRing, Internal Retaining E-RIX-7 5047542 572218Grease 20920
30 Pulley, 40DP 24T 0.312W 0.188B 5977175 57220531 PC Board, Sample Probe Vertical 539781 575227B32 Screw, Pan-Head #6-32 x 0.20 2328133 Spacer 0.380 OD x 0.252 ID 2328034 Gasket, O-ring, 0.380 x 0.060 8793466 1157035 Diluter, Sample 250 µL1458x1736 Syringe, Sample 250 µL 1776434 2613437 Guide, Linear, Probe Upper 5341874 572032B38 Guide, Linear, Ancillary Probe 2447566 572030B39 Guide, Linear, Probe Lower 9129802 572031B40 Plunger, 2 mL, Tri-Continent 7121359 11224841 Not Used42 Filter, Level Sense 8675129 11209543 Connector, T 20GA 3544034 57222044 Pump, Sample Probe Fluid Sense Kit 8255995 11166945 Kit, Valve, Solenoid, Two-Way, V23, V24 078-K013-0146 Damper, Pulsation 1630340 57104247 Plate, Motor Mounting 571465B
Item Description REF Part Number
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample and Ancillary Probe
078D0080-02, Rev. A
48 Not Used49 Carriage 572120B50 Fitting, Quick Disconnect 1179975 2355951 Fitting, Luer-20GA 1879357 11026852 Spacer 0.187 x 0.250 x 0.063 11026753 Carriage, Sample and Ancillary 7041363 113764
Frame, Carriage 572129CBearing 4811788 572242Bearing (Quantity 2) 7236865 572019Leadscrew and Follower 572130Rod 571570B
54 Rinse Station, Ancillary Probe 2807554 11026555 Screw, Soc. Head Cap M4 x 8 SST 60013x956 Bracket, Motor Mounting 078-0007-0157 Screw, Binding #4-40x 0.31 13002x158 Screw, Flat M4 x 8 mm 60055x5559 Screw, Flat M3 x 0.5 x 12 mm 60007x7360 Clamp, Belt 571466
Bracket 572138B61 Clamp, Belt 572136B
Bracket 57213762 Screw, Cap M3 x 0.5 x 10 mm 60013x1563 Screw, M3 x 0.5 x 8 mm 60013x864 Screw, Cap M4 x 0.7 x 10 mm 60013x1665 Washer, Flat M4 60056x566 Motor, Diluter 1458x29567 Clamp, Rinse Station 2669968 Washer, Flat M4 60056x569 Washer, Lock M4 60027x570 Screw, Cap M4 x 0.7 x 16 mm 60013x3771 Washer, Flat M3 60056x372 Diluter, 2 mL, Note: The new style diluter
is interchangeable with the Tri-Continent diluter.
115401
Item Description REF Part Number
380 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample and Ancillary
078D0080-02, Rev. A
73 Fitting (two piece insert with barb) 11008074 Chamber, Fluid 3275688 11539875 O-Ring 0.614ID 070W Ethylene propylene 1926711 11594576 Seal, 2.0 mL, 0.500 ID 7026771 11539577 PC Board, Diluter 11483078 Screw #4-40 x 187 60030x40379 Nut, Chamber 1718884 11563380 Screw, Cap-Head M4 x 12 60013x23
Item Description REF Part Number
078D0080-02, Rev. A
Sample Entry Queue for the ADVIA Centaur System 078-B002-01
Ancillary Motor and Belt Location
382 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Belt Tensioner Detail
078D0080-02, Rev. A
Belt Tensioner Detail
Belt Drive Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Magnet Detail 383
078D0080-02, Rev. A
Magnet Detail
Motor Detail
384 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 385
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Sample Entry Queue for the ADVIA Centaur System 078-B002-01
Item Description REF Part Number1 Screw, Cap-Head M4 x 0.7 X 8 mm 02784961 60013x9 2 Cover, Entry Queue, Back 05561920 078-0117-013 Screw, Button-Head M3 x 0.5 X 5 mm
Phillips 02683391 572496
4 PC Board, Rack Present 06438634 S575292B 5 PC Board, Sample Entry Queue Sensor 08726947 575242B 6 Motor, Stepper 4VDC 05192771 572914 7 Bearing, Ball 0.250 x 0.375 x 0.125 06068713 26082x7 8 Bearing 06776297 116258 9 Gear, Spur 09157865 571988 10 Gear, Spur 03120951 572770
Loctite, 222 03904871 17414 11 Shaft, Input Queue Drive 00995779 078-0023-01
Loctite, Moly-paste 03338817 590-0331-0412 Ring, Retaining E-ring 0.250D 05785519 26107 13 Screw, Cap-Head M3 x 12 mm 08057921 60013x22 14 Washer, Flat 02073054 846-1000-06 15 Screw, Cap-Head #4-40 x 0.375 572421 16 Belt, Entry Queue 09697398 571001 17 Panel, Entry Queue, Back Cover 05581174 110214 18 Spring 00155614 116259 19 Block, Tensioner Mounting 03105235 571995 20 Block, Tensioner Bearing 00307384 571993 21 Shaft, Tensioner 00188148 572003 22 Pulley, Tensioner 05314990 571994
Loctite, Thread Locker 03131961 590-0014-02 23 Washer, Flat M3 60056x3 24 Queue, Magnet Input 00237246 078-B011-01
Cover, Magnet 571990 Magnet 08249375 571991
25 Kit, Entry Queue Belt Guide 03931585 078-K071-01
388 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample Entry Queue for
078D0080-02, Rev. A
Plate, Base 571999Item Description REF Part Number
078D0080-02, Rev. A
Sample Entry Queue for the ADVIA Centaur XP System 078-B078-01
Ancillary Motor and Belt Location
Belt Tensioner Detail
390 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Belt Drive Detail
078D0080-02, Rev. A
Belt Drive Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Magnet Detail 391
078D0080-02, Rev. A
Magnet Detail
Motor Detail
392 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 393
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Sample Entry Queue for the ADVIA Centaur XP System 078-B078-01
Item Description REF Part Number1 Not used 2 Not used3 Screw, Button-Head M3 x 0.5 X 5 mm
Phillips 572496
4 PC Board, Rack Present 6438634 S575292B 5 PC Board, Sample Entry Queue Sensor 575242B 6 Motor, Stepper 4VDC 5192771 572914 7 Bearing, Ball 0.250 x 0.375 x 0.125 26082x7 8 Bearing 116258 9 Gear, Spur 9157865 571988 10 Gear, Spur 3120951 572770
Loctite, 222 17414 11 Shaft, Input Queue Drive 995779 078-0023-01
Loctite, Moly-paste 590-0331-0412 Ring, Retaining E-ring 0.250D 5785519 26107 13 Screw, Cap-Head M3 x 12 mm 60013x22 14 Washer, Flat 2073054 846-1000-06 15 Screw, Cap-Head #4-40 x 0.375 572421 16 Belt, Entry Queue 9697398 571001 17 Not used18 Spring 155614 116259 19 Block, Tensioner Mounting 571995 20 Block, Tensioner Bearing 307384 571993 21 Shaft, Tensioner 188148 572003 22 Pulley, Tensioner 5314990 571994
Loctite, Thread Locker 590-0014-02 23 Washer, Flat M3 60056x3 24 Queue, Magnet Input 237246 078-B011-01
Cover, Magnet 571990 Magnet 571991
25 Kit, Entry Queue Belt Guide 078-K071-01
396 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample Entry Queue for
078D0080-02, Rev. A
Plate, Base 078-0327-01Item Description REF Part Number
078D0080-02, Rev. A
Sample Exit Pusher 571037B
Sample Exit Pusher Assembly
Sensor Detail
Motor Detail
398 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 399
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Sample Exit Pusher 571037BItem Description REF Part Number1 Motor, Stepper 4VDC 572914 2 Gear, Spur 48P 28T 0.187W 0.187B 522-0309-01
Loctite, Adhesive 590-0014-02 3 Housing, Pusher 572009B 4 Screw, Cap-Head #4-40 x 0.375 5724215 Sensor, Optical Switch, Wide Gap 3048878 111675 6 Screw, Cap-Head M3 x 0.5 x 6 mm 7882579 60013x1 7 Gear, Rack, Exit Pusher 572007 8 Washer, Pusher End 572595 9 Screw, Flat-Head M3 x 0.5 x 6 mm 60055x43 10 Washer, Flat 60056x3
078D0080-02, Rev. A
Sample Tip Loader 571036C for the ADVIA Centaur System
Sample Tip Loader Assembly
404 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Loader
078D0080-02, Rev. A
Sample Tip Loader
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 405
078D0080-02, Rev. A
Sensor Detail
Motor and Pulley Detail
078D0080-02, Rev. A
Sample Tip Loader 078-B136-01 for the ADVIA Centaur XP System
408 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Loader
078D0080-02, Rev. A
Sample Tip Loader
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail 409
078D0080-02, Rev. A
Sensor Detail
Motor and Pulley Detail
410 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor and Pulley Detail
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Sample Tip LoaderItem Description REF Part Number 1 Screw, Cap-Head M3 x 0.5 X 10 mm 60013x15 2 Washer, Flat M3 60056x3 3 PC Board, Tip Tray and Band Sensor S575397C 4 Cover, Sensor 572434 5 Screw, Set M3 x 0.5 x 4 mm 60022x39 6 Guide, Tip Tray Cover 571653 7 Guide, Tip Tray, Indexing Left 572213 8 Bearing, Plastic Flanged 0.125 x 0.172 x
0.078 4179763 571646
9 Ring, Retaining, E-Ring 0.095 x 0.240 x 0.017 Stainless
571672
10 Spring, Extension 0.18 x 1.12 3.08 lb/In 856027 572682B 11 Screw, Pan M3 x 6 mm 6004x43 12 Block, Bearing 078–0009–01 13 Screw, Cap-Head M4 x 0.7 X 10 mm 846–1010–30 14 Washer, Flat M4 SST 846–1000–04 15 Loader, Sample Tip Tray Assembly 1794238 113762C
Shaft, Square 571676C Shaft, Arm Pivot 571675E Ring, Retaining (Quantity 2) 14458x4 Arm (Quantity 2) 571644C Block (Quantity 2) 572765B Spring 571693B Screw, Set M3 x 8 (Quantity 2) 572771B Loctite, Adhesive 222 17414
16 Spring, Tip Tray Latch 571693B 17 Screw, Flat-Head M3 x 0.5 x 6 mm Socket 60055x43 18 Spring, Guide, Front 5144580 571692B 19 Support, Spring Guide, Back 571695 20 Support, Spring Guide, Front 571696 21 Sensor, Optical Switch 962374 111674 22 PC Board, Tip Loader 3909210 575352B
412 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample Tip Loader
078D0080-02, Rev. A
23 Screw, Pan-Head #4-40 x 0.187 Phillips 60030x403 24 Screw, Cap-Head M4 x 0.7 x 8 mm Hex 60013x9 25 Gear, Spur 32P 20T 0.187W 0.187B 8676982 572022B 26 Motor, Stepper 4VDC 5192771 572914 27 Spring, Guide, Back 571691B 28 Linkage, Tip Tray, Guide Lock 204046 078-0010-01 29 Guide, Tip Tray, Indexing Right 572214 30 Plate, Vertical Base 571669B 31 Plate, Bearing 078-0008-01 32 Bracket, Tip Loader Support 572433A 33 Plate, Front, Vertical 571670B 34 Plate, Back, Vertical 571668B 35 Guide, Tip Tray Orientation 571657 36 Kit, Sample Tip Loader or 078-K001-01
Guide, Tip Tray Exit 6455776 571655B 37 Rod, Support 571673A 38 Screw, Pan M3 x 6 mm 60004x43 39 Linkage, Tip Tray Indexing 571660 40 Bar, Tray Guide Right 571666
Item Description REF Part Number
078D0080-02, Rev. A
Sample Tip Waste Area for the ADVIA Centaur System 078-B0009-02
Sample Tip Waste Area Assembly
414 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area Assembly
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area Assembly 415
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Sample Tip Waste Area 078-B009-02 for the ADVIA Centaur System
Item Description REF Part Number1 Kit, Cover, Connector (Includes lid and
connector)078-K027–01
2 Nut, Hex-Head, M8 60024x93 Sensor, Optical Switch 111674 21470x14 Screw, Flat-Head M3 x 0.5 x 5 mm 60006x335 Screw, Cap-Head M3 x 0.5 x 6 mm 60012x16 Chain Guide, Cleaning Solution Reservoir,
ADVIA Centaur System110228
6 Chain Guide, Cleaning Solution Reservoir, ADVIA Centaur XP System
078-0340-01-
7 Sensor, Proximity 113261 572432
8 Slide, Ball Bearing, 11 – 18 lbs., ADVIA Centaur System
8848643 572523
8 Slide, Ball Bearing, 11 - 18 lbs., ADVIA XP Centaur System
3828105 078-0344-01
9 Screw, Pan-Head, M4 x 0.7 x 6 mm 60003x3510 Liner, Tip Waste, ADVIA Centaur System 0878489 078-0066-0110 Liner, Tip Waste, ADVIA Centaur XP
System078-0343-01
11 Bin, Tip Waste 57261612 Collar, Tip Waste Bin 11010713 Screw, Cap-Head M4 x 0.7 x 40 mm 60013x8614 Carrier, Cleaning Solution Reservoir,
ADVIA Centaur System572519
14 Carrier, Cleaning Solution Reservoir ADVIA Centaur XP System
078-0342-01
15 Bottle, Cleaning Solution Reservoir 2908954 57252016 Tube, Formed 7526731 57195717 Screw, Pan-Head M8 x 90 mm 60003x21818 Connector See Item 119 Shutter, Waste Reservoir 57262220 Screw, Retractable M4 x 0.7 x 5.7 mm,
Knurled, Snap Fit220165 571964
418 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample Tip Waste Area
078D0080-02, Rev. A
21 Clamp, Spring 4324275 11024222 Screw, Pan-Head M4 x 0.7 x 10 mm 60000x6523 Spring, Waste Bin Interlock 57263824 Detent, Waste Bin Interlock 57263925 Screw, Pan-Head M4 x 0.7 x 6 mm 3545693 60000x4526 Washer, Flat, M4 60028x527 Chute, Tip Waste 7109448 57254828 Clamp, Cable 2362447 12811x129 Clamp, Cable 8461889 2349130 Screw, Cap-Head M3 x 0.5 x 6 mm 60012x131 Reservoir, Waste Chute 9580326 572547
Item Description REF Part Number
078D0080-02, Rev. A
Sample Tip Waste Area for the ADVIA Centaur XP System 078-B088-01
420 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area for the
078D0080-02, Rev. A
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Tip Waste Area for the
078D0080-02, Rev. A
078D0080-02, Rev. A
Parts List – Sample Tip Waste Area 078-B088-01 for the ADVIA Centaur XP System
Item Description REF Part Number1 Kit, Cover, Connector (Includes lid and
connector)078-K027–01
2 Nut, Hex-Head, M8 60024x93 Sensor, Optical Switch 111674 21470x14 Screw, Flat-Head M3 x 0.5 x 5 mm 60006x335 Screw, Cap-Head M3 x 0.5 x 6 mm 60012x16 Chain Guide, Cleaning Solution Reservoir,
ADVIA Centaur System110228
6 Chain Guide, Cleaning Solution Reservoir, ADVIA Centaur XP System
078-0340-01-
7 Sensor, Proximity 113261 572432
8 Slide, Ball Bearing, 11 – 18 lbs., ADVIA Centaur System
8848643 572523
8 Slide, Ball Bearing, 11 - 18 lbs., ADVIA XP Centaur System
3828105 078-0344-01
9 Screw, Pan-Head, M4 x 0.7 x 6 mm 60003x3510 Liner, Tip Waste, ADVIA Centaur System 57217710 Liner, Tip Waste, ADVIA Centaur XP
System078-0343-01
11 Bin, Tip Waste 57261612 Collar, Tip Waste Bin 11010713 Screw, Cap-Head M4 x 0.7 x 40 mm 60013x8614 Carrier, Cleaning Solution Reservoir,
ADVIA Centaur System572519
14 Carrier, Cleaning Solution Reservoir ADVIA Centaur XP System
078-0342-01
15 Bottle, Cleaning Solution Reservoir 2908954 57252016 Tube, Formed 7526731 57195717 Screw, Pan-Head M8 x 90 mm 60003x21818 Connector See Item 119 Shutter, Waste Reservoir 57262220 Screw, Retractable M4 x 0.7 x 5.7 mm,
Knurled, Snap Fit220165 571964
424 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Sample Tip Waste Area
078D0080-02, Rev. A
21 Clamp, Spring 4324275 11024222 Screw, Pan-Head M4 x 0.7 x 10 mm 60000x6523 Spring, Waste Bin Interlock 57263824 Detent, Waste Bin Interlock 57263925 Screw, Pan-Head M4 x 0.7 x 6 mm 3545693 60000x4526 Washer, Flat, M4 60028x527 Chute, Tip Waste 7109448 57254828 Clamp, Cable 2362447 12811x129 Clamp, Cable 8461889 2349130 Screw, Cap-Head M3 x 0.5 x 6 mm 60012x131 Reservoir, Waste Chute 9580326 572547
Item Description REF Part Number
078D0080-02, Rev. A
Stat Sample Pusher 571035
Stat Sample Pusher Assembly
426 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sensor Detail
078D0080-02, Rev. A
Sensor Detail
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Motor Detail 427
078D0080-02, Rev. A
Motor Detail
078D0080-02, Rev. A
Parts List – Stat Sample Pusher 078-B026-01Item Description REF Part Number1 Screw, Cap-Head M4 x 0.7 x 6 mm 60013x22 PC Board, Rack Loader 8575213 575237B3 Belt, Rack, Infeed 4904972 5710024 Block, Infeed 3048541 113765
Block 572020Magnet 572025Loctite 572501
5 Bracket, Infeed, Block, Support 5720216 Screw, Cap-Head M3 x 0.5 x 6 mm 7882579 60013x17 Scanner, Barcode, ADVIA Centaur System 4965793 115027C7 Scanner, Barcode, ADVIA Centaur XP
System7343254 078-0390-01
8 Bracket, Barcode Scanner 5720169 Screw, Cap-Head #6-32 x 0.250 57249410 Cable Tie, Ty-Wrap 57299211 Screw, Cap-Head M5 x 0.8 x 25 mm 60013x6612 Pulley, Infeed Idler Assembly 5520752 113766
Housing 572024Pulley 572012Bearing (Quantity 2) 7236865 572019
13 Pulley, Infeed Drive Assembly 924782 113767Housing 572024Pulley 572013Bearing 7236865 572019
14 Screw, Cap-Head M4 x 0.5 x 25 mm 60013x64Screw, Cap-Head M3 x 0.5 x 30 mm 60013x71
15 Housing, Infeed Mechanism 57202416 Block, Pusher, Stat Exit 57201517 Bracket, Rack Stop 57276418 Not Used19 Sensor, Slotted Optical Switch 111674 21470x120 Screw, Cap-Head #4-40 x 0.250 572493
430 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Stat Sample Pusher
078D0080-02, Rev. A
21 Plate, Stat Top 57202622 Screw, Captive M4 x 0.7 x 15.75 mm 57202723 PC Board, Rack Present 6438634 S575292B24 PC Board, Stat Area 8924552 575307C25 Gear, Spur 32P 20T 0.187W 0.187B Acetal 8676982 572022B26 Motor, Step 4VDC BP 44.4 Ht .80 Shaft 572914 57163427 Washer, Flat M4 60056x528 Washer, Flat M3 60056x329 Housing, Stat Exit Pusher 078-0043-0130 Label, Laser Warning 078-0003-013132
Item Description REF Part Number
078D0080-02, Rev. A
Universal Rack Option
Sample Probe
432 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Stat Sample Pusher
078D0080-02, Rev. A
Stat Sample Pusher
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Pusher 433
078D0080-02, Rev. A
Sample Exit Pusher
434 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Inprocess Queue
078D0080-02, Rev. A
Inprocess Queue
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Rack Sensor Pusher Cable 435
078D0080-02, Rev. A
Rack Sensor Pusher Cable
436 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Queue
078D0080-02, Rev. A
Sample Exit Queue
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Queue Rack Present
078D0080-02, Rev. A
Sample Exit Queue Rack Present Sensor
438 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Exit Queue Full Sensor
078D0080-02, Rev. A
Sample Exit Queue Full Sensor
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Entry Queue 439
078D0080-02, Rev. A
Sample Entry Queue
440 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Entry Queue Work Surface
078D0080-02, Rev. A
Sample Entry Queue Work Surface
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Sample Entry Queue Tensioner 441
078D0080-02, Rev. A
Sample Entry Queue Tensioner
078D0080-02, Rev. A
Parts List – Universal Rack OptionItem Description REF Part Number1 Body, Sample Probe Plunger 06663735 1183452 Pulley, Leadscrew 03398496 1183443 PC Board, Universal Normal Rack Present 08176629 1183624 Standoff 1185245 Belt, Rack Loader 02062737 1187736 PC Board, Stat Rack Present 0118368C 1183687 Housing, Stat Exit Pusher 03538387 1185478 Standoff, Pusher Extension 1185039 Bar, Spring Mounting 11840010 Screw, Cap M4 x 20 Stainless 02683391 57249611 Spring, Rack Positioning 04406301 11840212 Screw, Button Head M3 x 6 Stainless 02683391 57249613 Plate, Positioning 03873089 11840114 Cable, Rack Pusher Sensor 09255077 11972815 Cable, Hitachi Exit Queue 01603491 11847916 PC Board, Hitachi Exit Queue Full 08990202 11837517 Sensor, Rack Present 05672579 11972918 Screw, Pan M3 x 0.5 x 6mm 07908944 60000x4319 Tray, Exit Queue 00843898 11878020 Screw, Flat M3 x 0.5 x 6mm Stainless Steel 60007x4321 Washer, Lock M3 03617287 60026x322 Sensor, Exit Queue Full 07311808 11848123 Washer, Flat M3 05551992 60028x324 Nut, M3 x 0.5 06552356 60024x425 Screw, Pan M3 x 0.5 x 5mm Stainless Steel 60001x3326 PC Board, Rack Pusher/Entry Queue Home 9644596 11966827 Screw, M3 x 0.5 x 6mm Stainless Steel 06892408 60013x128 Bracket 11965729 Bracket 11878830 Screw, Flat M4 x 0.7 x 12mm Stainless
Steel60007x85
31 Block, Sample Rack Pusher, Assembly 01691579 119825
444 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List – Universal Rack Option
078D0080-02, Rev. A
32 Tray, Entry Queue, ADVIA Centaur System
06848190 118781
32 Tray, Entry Queue, ADVIA Centaur XP System
06848190 078-0441-01
33 Work Surface, Entry Queue 02254393 11878934 Washer, Flat #8 06734691 25042x6735 Screw, Cap M4 x 0.7 x 12mm 01876943 60012x2336 Pulley 119656
Item Description REF Part Number
078D0080-02, Rev. A
Miscellaneous Parts
Parts ListItem Description REF Part Number
Blower, Ancillary, Assembly 09694747 572925Blower, Primary Reagent Compartment, Assembly
07318179 572931
Cord, Power Line 05430311 572334Fan, Electronics Module 571878Keyboard, English, ADVIA Centaur System
01103618 571266D
Keyboard, English, ADVIA Centaur XP System
02000952 078-0393-01
Keyboard, German, ADVIA Centaur System
03010811 109955D
Keyboard, German, ADVIA Centaur XP System
078-0394-01
Keyboard, Italian, ADVIA Centaur System 00814111 109956CKeyboard, Italian, ADVIA Centaur XP System
04719040 078-0395-01
Keyboard, French, ADVIA Centaur System 03036438 109958CKeyboard, French, ADVIA Centaur XP System
078-0397-01
Keyboard, Spanish, ADVIA Centaur System
07012150 109957C
Keyboard, Spanish, ADVIA Centaur XP System
02992998 078-0396-01
Modem, Kit 04227660 115519Monitor,Kit 03376425 078-K070-01Printer, North America Check eBRITE for
latest informationScanner, Barcode Assembly, Hand Held 02865821 118099Switch, Fast S top (red) 572158Cord, Power Line, European 572866Switch, Computer Startup (green) 572516
1-446 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 1 Parts List
078D0080-02, Rev. A
Labels, Rack 1-100, ADVIA Centaur System
00424364 112249B
Labels, Rack 101-200, ADVIA Centaur System
09270017 112250B
Labels, Rack 201-300, ADVIA Centaur System
00294533 112251B
Labels, Rack 301-400, ADVIA Centaur System
07421808 112252B
Labels, Rack 401-500, ADVIA Centaur System
01135889 112253B
Labels, Rack 1-100, ADVIA Centaur XP System
09391302 078-0333-01
Labels, Rack 101-200, ADVIA Centaur XP System
03828105 078-0334-01
Labels, Rack 201-300, ADVIA Centaur XP System
08579502 078-0335-01
Labels, Rack 301-400, ADVIA Centaur XP System
01160808 078-0336-01
Labels, Rack 401-500, ADVIA Centaur XP System
02115415 078-0337-01
Kit, Mouse, 3-button SUN 05749490 078-K080-01
Item Description REF Part Number
078D0080-02, Rev. A
2 Kits and Tools
KitsDescription Part REFKit, Routine Certification 115835Kit, Extended Certification 115836Kit, Direct Plumbing 078-K087-01Kit, Lab Automation 115869Kit, Maintenance 112712Kit, Labcell/Workcell 078-K077-01
2-2 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 2 Tools
078D0080-02, Rev. A
ToolsDescription Part Number REFConnector, GIOB Bypass 116325Diskette, Controller Diagnostics Sequences 078–S019–01Flowmeter 1458x144Tool, Reagent Pack 112749Tool, Elevator Up 113907Tool, Flow Adapter 078–B024–01Tool, Alignment, Reagent/Anc Probe to Cuvette 117112Tool, Ancillary Pack Calibration 111010Tool, Aspirate Probe Home 113906Tool, Barcode Shutter 114046Tool, Reagent Bottom 114565Tool, Sample Probe Calibration 117883Tool, Sample Tip Tray Assembly Alignment 114348Tool, Sample Tip Tray Alignment 114347Tool, Vacuum Calibration 117736Wrench Set, Bondhus Balldriver Hex 1.5 to 10mm
110864
Terminator, GIOB 111009YSI M4610 Temperature Meter 110367YSI M4611 Probe 110368Pliers, Groove Joint 113958
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Probes Horizontal
078D0080-02, Rev. A
3 Centaur PC Boards
Sample and Ancillary Probes Horizontal PC Board 575212C
Connector Cable Connected to . . .J1 571280 J6 of Sample and Ancillary Probe GIOB PC board
S575322H
Sensor FunctionU1 Tip 1 PositionU2 Tip 2/Tube 1 PositionU3 Tip 3 PositionU4 Tip 4/Tube 2 PositionU5 Tip 5 PositionU6 Tip 6/Tube 3 PositionU7 Tip 7 PositionU8 Tip 8/Tube 4 PositionU9 Tip 9 PositionU10 Tip 10/Tube 5 PositionU11 Tip Eject PositionU12 Sample Cuvette PositionU13 Lab Auto PositionU14 Ancillary Horizontal Home PositionU15 Ancillary Cuvette PositionU16 Ancillary Wash PositionU17 Ancillary Pack Position
3-2 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Inprocess Queue Tip Sensor PC
078D0080-02, Rev. A
Inprocess Queue Tip Sensor PC Board 575217B
Connector Cable Connected to . . .J1 571296 J6 of Sample and Ancillary GIOB PC board S575327D
Sensor FunctionU1 Inprocess Queue HomeU2 Rack PositionU3 Rack DirectionU4 Tip PositionU5 Tip Direction
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Queue Sensor PC Board
078D0080-02, Rev. A
Ancillary Queue Sensor PC Board 575222B
Connector Cable Connected to . . .J1 571292 J4 of Exit Queue Interconnect PC board 575302CJ2 Ancillary Queue Home sensorJ3 Ancillary Queue Position sensorJ4 Ancillary Queue Direction sensorJ5 Ancillary Queue motorJ6 Ancillary Queue thermistor 1J7 Not usedJ8 Not used
3-4 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample Probe Vertical Sensor PC
078D0080-02, Rev. A
Sample Probe Vertical Sensor PC Board for the ADVIA Centaur System 575227
Connector Cable Connected to . . .J1 571284 J7 of Sample and Ancillary Probe GIOB S575322H J2 Sample Vertical motor
Sensor FunctionU1 Vertical HomeU2 Above Tip EjectU3 Below Tip Eject
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample Probe Vertical Sensor PC Board
078D0080-02, Rev. A
Sample Probe Vertical Sensor PC Board for the ADVIA Centaur XP System 118365
Connector Cable Connected to . . .J1 571284 J7 of Sample and Ancillary Probe GIOB S575322H J2 Sample Vertical motor
Sensor FunctionU1 Vertical HomeU2 Above Tip EjectU3 Below Tip Eject
3-6 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Entry PC Board 575232D
078D0080-02, Rev. A
Ancillary Entry PC Board 575232D
Connector Cable Connected to . . .J1 571287 J5 of Exit Queue Interconnect PC board 575302CJ2 Ancillary Entry motorJ3 Ancillary Door motorJ4 Ancillary Door Home sensorJ5 Ancillary Door Position sensorJ6 Ancillary LED PC board 575392B
J7 Ancillary Pack PC board 575457B
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Entry PC Board 575232D 3-7
078D0080-02, Rev. A
Sensor FunctionU1 Door HomeU2 Door Extend
3-8 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Rack Loaded Sensor PC Board
078D0080-02, Rev. A
Rack Loaded Sensor PC Board 575237B
Connector Cable Connected to . . .J1 571342 J8 of Stat Area PC board 575307C
Sensor FunctionU1 Stat Load PositionU2 Normal Load Position
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Entry Queue Sensor PC Board 575242B
078D0080-02, Rev. A
Entry Queue Sensor PC Board 575242B
Connector Cable Connected to . . .J1 571291 J3 of Stat Area PC board 575307C
Sensor FunctionU1 Entry Queue Home
3-10 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board
078D0080-02, Rev. A
Low Level Controller PC Board 078-B042S01
There are four fuses on each PC board; their functions are described below:
Volts Label Rating Location Protects +24 F0 2A 250V
5x20 mm FB 572808
Pumps 2 LLC PC board 078-B042S01
resuspend dispense, motor 0 (P1)
+24 F1 2A 250V 5x20 mm FB 572808
Pumps 2 LLC PC board wash 1, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Pumps 2 LLC PC board wash 2, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Pumps 2 LLC PC board acid, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 57280
Pumps 1 LLC PC board wash displacement, motor 0 (P1)
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board
078D0080-02, Rev. A
+24 F1 2A 250V 5x20 mm FB 572808
Pumps 1 LLC PC board wash 3, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Pumps 1 LLC PC board not used
+24 F3 2A 250V 5x20 mm FB 572808
Pumps 1 LLC PC board base, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Aspirate Probes LLC PC board
probe 1, motor 0 (P1)
+24 F1 2A 250V 5x20 mm FB 572808
Aspirate Probes LLC PC board
probe 2, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Aspirate Probes LLC PC board
probe 3, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Aspirate Probes LLC PC board
probe 4, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Preheater & Lumo LLC PC board
cuvette conveyor, motor 0 (P1)
+24 F1 2A 250V 5x20 mm FB 572808
Preheater & Lumo LLC PC board
cuvette pusher, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Preheater & Lumo LLC PC board
luminometer rotor, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Preheater & Lumo LLC PC board
waste probe, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Reagent Pack LLC PC board
reagent shelf, motor 0 (P1)
Volts Label Rating Location Protects
3-12 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board
078D0080-02, Rev. A
+24 F1 2A 250V 5x20 mm FB 572808
Reagent Pack LLC PC board
reagent barcode, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Reagent Pack LLC PC board
reagent barcode shutter, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Reagent Pack LLC PC board
reagent probe shutter, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Reagent Probe 3 LLC PC board
pack to pack, motor 0 (P1)
+24 F1 2A 250V 5x20 mm FB 572808
Reagent Probe 3 LLC PC board
pack to cuvette, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Reagent Probe 3 LLC PC board
vertical, motor 2(P1)
+24 F3 2A 250V 5x20 mm FB 572808
Reagent Probe 3 LLC PC board
diluter, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Reagent Probe 2 LLC PC board
pack to pack, motor 0 (P1)
+24 F1 2A 250V 5x20 mm FB 572808
Reagent Probe 2 LLC PC board
pack to cuvette, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Reagent Probe 2 LLC PC board
vertical, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Reagent Probe 2 LLC PC board
diluter, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Reagent Probe 1 LLC PC board
pack to pack, motor 0 (P1)
Volts Label Rating Location Protects
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board
078D0080-02, Rev. A
+24 F1 2A 250V 5x20 mm FB 572808
Reagent Probe 1 LLC PC board
pack to cuvette, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Reagent Probe 1 LLC PC board
vertical, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Reagent Probe 1 LLC PC board
diluter, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Sample Tip LLC PC board
tip loader, motor 0 (P1)
+24 F1 2A 250V 5x20 mm FB 572808
Sample Tip LLC PC board
entrance queue, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Sample Tip LLC PC board
not used
+24 F3 2A 250V 5x20 mm FB 572808
Sample Tip LLC PC board
not used
+24 F0 2A 250V 5x20 mm FB 572808
Ancillary LLC PC board
ancillary entry, motor 0 (P1)
+24 F1 2A 250V 5x20 FB mm 572808
Ancillary LLC PC board
ancillary queue, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Ancillary LLC PC board
ring loader, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Ancillary LLC PC board
ancillary door, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Sample Racks LLC PC board
rack loader, motor 0 (P1)
Volts Label Rating Location Protects
3-14 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board
078D0080-02, Rev. A
+24 F1 2A 250V 5x20 mm FB 572808
Sample RLC PC board inprocess queue, motor 1 (P1)
+24 F2 2A 250V 5x20 mm FB 572808
Sample Racks LLC PC board
sample exit pusher, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Sample Racks LLC PC board
stat sample pusher, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Ancillary Probe LLC PC board
pack to cuvette, motor 0 (P1)
+24 F1 2A 250V 5x20 mm FB 572808
Ancillary Probe LLC PC board
not used
+24 F2 2A 250V 5x20 mm FB 572808
Ancillary Probe LLC PC board
vertical, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Ancillary Probe LLC PC board
diluter, motor 3 (P1)
+24 F0 2A 250V 5x20 mm FB 572808
Sample Probe LLC PC board
Sample to cuvette, motor 0 (P1)
+24 F1 2A 250V 5x20 mm FB 572808
Sample Probe LLC PC board
Not used
+24 F2 2A 250V 5x20 mm FB 572808
Sample Probe LLC PC board
vertical, motor 2 (P1)
+24 F3 2A 250V 5x20 mm FB 572808
Sample Probe LLC PC board
diluter, motor 3 (P1)
LED FunctionCR1 On PC board red LEDCR2 On PC board, green LED
Volts Label Rating Location Protects
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Low Level Controller PC Board
078D0080-02, Rev. A
NOTE: Green LED lit and red LED not lit indicates normal operation of PC board. Red LED lit and green LED not lit indicates fault condition on PC board. Red LED lit and green LED flashing indicates reset condition.
3-16ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide
078D0080-02, Rev. A
Backplane PC Board 575252F (slide out electronics module)
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide out
078D0080-02, Rev. A
Connector Cable Connected to . . .J1 Sample Probe low level controller PC board
078-B042S01J2 Ancillary Probe low level controller PC boardJ3 Sample Racks low level controller PC boardJ4 Ancillary low level controller PC boardJ5 Sample Tip low level controller PC boardJ6 Not UsedJ7 Reagent Probe 1 low level controller PC boardJ8 Reagent Probe 2 low level controller PC boardJ9 Reagent Probe 3 low level controller PC boardJ10 Reagent Pack low level controller PC boardJ11 Not UsedJ12 572302 J101 Fluid GIOB PC board 078-B027S01J13 Real Time SPARC PC boardJ14 Sample Handling SC1 sequencer PC boardJ15 Reagent Handling SC2 sequencer PC boardJ16 Cuvette Handling SC3 sequencer PC boardJ17 571411 J101 Sample & Ancillary Queue GIOB PC board
S575327DJ18 Preheater and Luminometer low level controller PC
boardJ19 Aspirate Probes low level controller PC boardJ20 Pumps 1 low level controller PC boardJ21 Pumps 2 low level controller PC boardJ22 Not UsedJ23 Not usedJ24 572298 J101 Sample & Ancillary Probe GIOB PC board
S575322HJ25 Ring and Elevator Servo low level controller PC boardJ26 Power Supply Power ConnectorJ27 Not UsedJ28 572301 J101 Cuvette GIOB PC board S575377FJ29 121151 Ethernet
3-18ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide
078D0080-02, Rev. A
J30 Not UsedJ31 572300 J101 Reagent Handling GIOB PC board S575287CJ32 121152 User Interface ModuleJ33J34 572315 J5 Thermal Control Module SS575257EEJ35 External barcode scannerJ36 571344 J102 Sample & Ancillary Queue GIOB PC board
S575327DJ37 571320 J3 Bottom Luminometer Assembly 575367J38 571319 or
119161J3 Servo Motor 575357C NOTE: Use cable part number 119161 only on systems that have white connectors for both J38 on the Backplane PC board and J3 on the Servo Motor Interconnect PC board.
J39 571313 J5 Sample & Ancillary Probe GIOB PC board S575322H
J40 571312 J1 Sample & Ancillary Queue GIOB S575327DJ41 571314 J7 Reagent Handling GIOB PC board S575287CJ42 571316 J13 Reagent Handling PC board S575287CJ43 571318 Fluid GIOB PC board 078-B027S01J44 571317 J3 Cuvette GIOB PC board S575377FJ45 572299 J101 Reagent Fluid GIOB PC board 078-B027S02J46 571450 J9 Exit Queue Interconnection PC board 575302CJ47 571449 Reagent Pack Barcode ScannerJ48 Lab Automation RS-232
Test Point Sparc SpecificationsTP1 RT –12.00 VDC +/– 2% (–11.76 to –12.24)TP2 RT 24.00 VDC +/– 3% (23.28 to 24.72)TP3 RT 12.10 VDC +/– 2% (11.86 to 12.34)TP4 RT 5.10 VDC +/– 2% (4.99 to 5.20)TP5 RT GroundTP6 UI –12.30 VDC +/– 2% (–12.05 to –12.55)TP7 UI 12.10 VDC +/– 2% (11.86 to 12.34)TP8 UI 5.30 VDC +/– 2% (5.19 to 5.41)
Connector Cable Connected to . . .
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide out
078D0080-02, Rev. A
Volts Label Rating Location Protects+5V PTC1 0.5A Backplane PC board
575252Fencoder (J38)
+24V PTC2 0.5A Backplane PC board 575252F
luminometer solenoid (J44)
+5V PTC3 2.5A Backplane PC board 575252F
sample and ancillary barcode reader (J46)
+5V PTC4 2.5A Backplane PC board 575252F
reagent barcode reader (J47)
+24V PTC5 0.5A Backplane PC board 575252F
Ionizer (J44)
+5V PTC6 2.5A Backplane PC board 575252F
disk enclosure module (J30)
-12V PTC7 2.5A Backplane PC board 575252F
not used (J45)
+24V PTC8 7A Backplane PC board 575252F
Sample and Ancillary Queue GIOB PC board S575327D (J24)Sample and Ancillary Probe GIOB PC board S575322H (J17) Reagent Fluid GIOB PC board 078-B027S02 (J45)
+5V PTC9 7A Backplane PC board 575252F
Reagent Handling GIOB PC board S575287C (J28) Cuvette GIOB PC board S575377F (J17)Fluid GIOB PC board 078-B027S01 (J12)
+12V PTC10 2.5A Backplane PC board 575252F
not used
+5V PTC11 2.5A Backplane PC board 575252F
bottom luminometer assembly 575367 (J37)
3-20ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board 575252F (slide
078D0080-02, Rev. A
+5V PTC12 7A Backplane PC board 575252F
Sample and Ancillary Queue GIOB PC board S575327D (J24) Sample and Ancillary Probe GIOB PC board S575322H (J17) Reagent Fluid GIOB PC board 078-B027S02 (JJ45)
+12V PTC13 2.5A Backplane PC board 575252F
disk enclosure module (J30)
+24V PTC14 7A Backplane PC board 575252F
Reagent Handling GIOB PC board S575287C (J28) Cuvette GIOB PC board S575377F (J31)Fluid GIOB PC board 078-B027S01 (J12)
+5V PTC15 2.5A Backplane PC board 575252F
handheld barcode reader (J35)
+12V PTC16 2.5A Backplane PC board 575252F
PMT (J37)
Volts Label Rating Location Protects
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D (fixed
078D0080-02, Rev. A
Backplane PC Board S117937D (fixed electronics module)
Connector Cable Connected to . . .J1 Sample Probe low level controller PC boardJ2 Ancillary Probe low level controller PC boardJ3 Sample Racks low level controller PC boardJ4 Ancillary low level controller PC boardJ5 Sample Tip low level controller PC boardJ6 Servo DecoderJ7 Reagent Probe 1 low level controller PC board
J8 Reagent Probe 2 low level controller PC boardJ9 Reagent Probe 3 low level controller PC board
3-22 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D
078D0080-02, Rev. A
J10 Reagent Pack low level controller PC boardJ11 121152 User Interface ModuleJ12 572302 J101 Fluid GIOB PC board 078-B027S01J13 Real Time SPARC PC boardJ14 Sample Handling SC1 sequencer PC boardJ15 Reagent Handling SC2 sequencer PC boardJ16 Cuvette Handling SC3 sequencer PC boardJ17 571411 J101 Sample & Ancillary Queue GIOB PC board
S575327DJ18 Preheater and Luminometer low level controller PC
boardJ19 Aspirate Probes low level controller PC boardJ20 Pumps 1 low level controller PC board
J21 Pumps 2 low level controller PC boardJ22 Not UsedJ23 121152 User Interface ModuleJ24 572298 J101 Sample & Ancillary Probe GIOB PC board
S575322HJ25 Ring and Elevator Servo low level controller PC boardJ26 Power Supply Power ConnectorJ27 Not UsedJ28 J101 Cuvette GIOB PC board S575377FJ29 121151 EthernetJ30 Not UsedJ31 572300 J101 Reagent Handling GIOB PC board S575287CJ32 Power Supply Sense and Control ConnectorJ33 Not UsedJ34 572315 J5 Thermal Control Module S575257EJ35 572328 External barcode scannerJ36 575344 J102 Sample & Ancillary Queue GIOB PC board
S575327DJ37 571320 J3 Bottom Luminometer Assembly 575367J38 Servo MotorJ39 571313 J5 Sample & Ancillary Probe GIOB PC board
S575322H
Connector Cable Connected to . . .
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D (fixed
078D0080-02, Rev. A
J40 571312 J1 Sample & Ancillary Queue GIOB S575327DJ41 571314 J7 Reagent Handling GIOB PC board S575287CJ42 571316 J13 Reagent Handling PC board S575287CJ43 571318 Fluid GIOB PC board 078-B027S01J44 571317 J3 Cuvette GIOB PC board S575377FJ45 572299 J101 Reagent Fluid GIOB PC board 078-B027S02J46 57145 J9 Exit Queue Interconnection PC board 575302CJ47 571449 Reagent Pack Barcode ScannerJ48 572327 Lab Automation RS-232
Test Point Sparc Specifications TP1 RT -12.13 VDC +/- 4% (-11.65 to -12.61)
TP2 RT 24.20 VDC +/- 3% (23.47 to 24.93) TP3 RT 12.12 VDC +/- 3% (11.76 to 12.48) TP4 RT 5.18 VDC +/- 1.5% (5.10 to 5.25) TP5 RT Ground TP6 UI -12.16 VDC +/- 2% (-11.91 to -12.40) TP7 UI 12.10 VDC +/- 2% (11.85 to 12.34) TP8 UI 5.075 VDC +/- 2% (4.97 to 5.18)
Volts Label Rating Location Protects +5V PTC1 0.5A Backplane PC board
S117937D encoder (J38)
+24V PTC2 0.5A Backplane PC board S117937D
luminometer solenoid (J44)
+5V PTC3 2.5A Backplane PC board S117937D
sample and ancillary barcode reader (J46)
+5V PTC4 2.5A Backplane PC board S117937D
reagent barcode reader (J47)
+24V PTC5 0.9A Backplane PC board S117937D
Ionizer (J44)
-12V PTC7 2.5A Backplane PC board S117937D
not used (J45)
Connector Cable Connected to . . .
3-24 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D
078D0080-02, Rev. A
+24V PTC8 7A Backplane PC board S117937D
Sample and Ancillary Queue GIOB PC board S575327D (J24)Sample and Ancillary Probe GIOB PC board S575322H (J17)Reagent Fluid GIOB PC board 078-B027S02 (J45)
+5V PTC9 7A Backplane PC board S117937D
Reagent Handling GIOB PC board S575287C (J28)Cuvette GIOB PC board S575377F (J17)Fluid GIOB PC board 078-B027S01 (J12)
+12V PTC10 2.5A Backplane PC board S117937D
not used
+5V PTC11 2.5A Backplane PC board S117937D
bottom luminometer assembly 575367 (J37)
+5V PTC12 7A Backplane PC board S117937D
Sample and Ancillary Queue GIOB PC board S575327D (J24)Sample and Ancillary Probe GIOB PC board S575322H (J17)Reagent Fluid GIOB PC board 078-B027S02 (JJ45)
+24V PTC14 7A Backplane PC board S117937D
Reagent Handling GIOB PC board S575287C (J28)Cuvette GIOB PC board S575377F (J31)Fluid GIOB PC board 078-B027S01 (J12)
+5V PTC15 2.5A Backplane PC board S117937D
handheld barcode reader (J35)
+12V PTC16 2.5A Backplane PC board S117937D
PMT (J37)
+24V PTC17 2.5A Backplane PC board S117937D
J50 Cardcage Fan
+24V PTC18 2.5A Backplane PC board S117937D
J50 Cardcage Fan
Volts Label Rating Location Protects
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Backplane PC Board S117937D (fixed
078D0080-02, Rev. A
+24V F1 10ASB Backplane PC board S117937D
Vacuum PumpVolts Label Rating Location Protects
3-26 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board
078D0080-02, Rev. A
Thermal Control Module PC Board for the ADVIA Centaur System S575257E
Connector Cable Connected to . . . J1 571374 Primary Reagent Compartment TEDs, heat sink fan,
circulation fan J2 572311 Ancillary Queue TEDs, heat sink fan J3 572312 Preheater TEDs and PMT TEDs J4 572313 Incubation Chamber TEDs J5 572315 J34 of Backplane PC board 575252F J7 571360 J2 of Cuvette GIOB PC board S575377F J8 572265 J14 of Mixer Interconnect PC board 575422C and J12
of Sample and Ancillary Queue GIOB PC board S575327D
J9 571352 Power Supply 571258
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board for
078D0080-02, Rev. A
LED Function CR42 On PC board, red LED (lit when ancillary compartment starts cooling) CR43 On PC board, green LED (blinking is normal during running) CR44 On PC board, red LED (inhibit, power supply is out of tolerance low,
watchdog trip) CR45 On PC board, red LED (lit when reagent compartment starts cooling) CR65 On PC board, LED, P2 sign bit for bi-polar loops (lit = heating; not lit =
cooling) CR67 On PC board, LED, Preheater sign bit for bi-polar loops (lit = heating;
not lit = cooling) CR68 On PC board, LED, Ring D sign bit for bi-polar loops (lit = heating; not
lit = cooling) CR69 On PC board, LED, Ring C sign bit for bi-polar loops (lit = heating; not
lit = cooling) CR70 On PC board, LED, PMT1 sign bit for bi-polar loops (lit = heating; not
lit = cooling) CR71 On PC board, LED, Ring A sign bit for bi-polar loops (lit = heating; not
lit = cooling) CR72 On PC board, LED, Ring B sign bit for bi-polar loops (lit = heating; not
lit = cooling)
Volts Label Rating Location Protects+24V F1 6.3A 250V 5x20
mm SB 575194Thermal Control Module S575257E
primary reagent compartment (front of instrument) thermal electric device (J1)
+24V F2 6.3A 250V 5x20 mm SB 575194
Thermal Control Module S575257E
primary reagent compartment (middle of instrument) thermal electric device (J1)
+24V F3 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
ring B thermistor via Cuvette GIOB PC board S575377F (J7)
+24V F4 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
ring A thermistor via Cuvette GIOB PC board S575377F (J7)
+24V F5 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
ring C thermistor via Cuvette GIOB PC board S575377F (J7)
3-28 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board
078D0080-02, Rev. A
+24V F6 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
ring D thermistor via Cuvette GIOB PC board S575377F (J7)
+24V F7 6.3A 250V 5x20 mm SB 575194
Thermal Control Module S575257E
primary reagent compartment (middle of instrument) thermal electric device (J1)
+24V F8 6.3A 250V 5x20 mm SB 575194
Thermal Control Module S575257E
ancillary refrigerator thermal electric device (J2)
+24V F9 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
preheater thermal electric devices (J3)
+24V F10 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
PMT 1 thermal electric device (J3)
+24V F11 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
PMT 2 thermal electric device (J3)
+24V PTC2 0.9A Thermal Control Module S575257E
ancillary heatsink fan and fan wiring (J2)
+24V PTC 525 1.35A Thermal Control Module S575257E
reagent heatsink fan and fan wiring (J1)
+24V PTC1 0.9A Thermal Control Module S575257E
reagent circulation fan and fan wiring (J1)
Volts Label Rating Location Protects
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board for
078D0080-02, Rev. A
Thermal Control Module PC Board for the ADVIA Centaur XP System 078-B111SXX
Connector Cable Connected to . . . J1 571374 Primary Reagent Compartment TEDs, heat sink fan,
circulation fan J2 572311 Ancillary Queue TEDs, heat sink fan J3 572312 Preheater TEDs and PMT TEDs J4 572313 Incubation Chamber TEDs J5 572315 J34 of Backplane PC board 575252F J7 571360 J2 of Cuvette GIOB PC board S575377F J8 572265 J14 of Mixer Interconnect PC board 575422C and J12
of Sample and Ancillary Queue GIOB PC board S575327D
J9 571352 Power Supply 571258
3-30 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board
078D0080-02, Rev. A
LED Function CR42 On PC board, red LED (lit when ancillary compartment starts cooling) CR43 On PC board, green LED (blinking is normal during running) CR44 On PC board, red LED (inhibit, power supply is out of tolerance low,
watchdog trip) CR45 On PC board, red LED (lit when reagent compartment starts cooling) CR65 On PC board, LED, P2 sign bit for bi-polar loops (lit = heating; not lit =
cooling) CR67 On PC board, LED, Preheater sign bit for bi-polar loops (lit = heating;
not lit = cooling) CR68 On PC board, LED, Ring D sign bit for bi-polar loops (lit = heating; not
lit = cooling) CR69 On PC board, LED, Ring C sign bit for bi-polar loops (lit = heating; not
lit = cooling) CR70 On PC board, LED, PMT1 sign bit for bi-polar loops (lit = heating; not
lit = cooling) CR71 On PC board, LED, Ring A sign bit for bi-polar loops (lit = heating; not
lit = cooling) CR72 On PC board, LED, Ring B sign bit for bi-polar loops (lit = heating; not
lit = cooling)
Volts Label Rating Location Protects+24V F1 6.3A 250V 5x20
mm SB 575194Thermal Control Module S575257E
primary reagent compartment (front of instrument) thermal electric device (J1)
+24V F2 6.3A 250V 5x20 mm SB 575194
Thermal Control Module S575257E
primary reagent compartment (middle of instrument) thermal electric device (J1)
+24V F3 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
ring B thermistor via Cuvette GIOB PC board S575377F (J7)
+24V F4 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
ring A thermistor via Cuvette GIOB PC board S575377F (J7)
+24V F5 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
ring C thermistor via Cuvette GIOB PC board S575377F (J7)
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Thermal Control Module PC Board for
078D0080-02, Rev. A
+24V F6 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
ring D thermistor via Cuvette GIOB PC board S575377F (J7)
+24V F7 6.3A 250V 5x20 mm SB 575194
Thermal Control Module S575257E
primary reagent compartment (middle of instrument) thermal electric device (J1)
+24V F8 6.3A 250V 5x20 mm SB 575194
Thermal Control Module S575257E
ancillary refrigerator thermal electric device (J2)
+24V F9 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
preheater thermal electric devices (J3)
+24V F10 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
PMT 1 thermal electric device (J3)
+24V F11 5A 250V 5x20 mm SB 60044x14
Thermal Control Module S575257E
PMT 2 thermal electric device (J3)
+24V PTC2 0.9A Thermal Control Module S575257E
ancillary heatsink fan and fan wiring (J2)
+24V PTC 525 1.35A Thermal Control Module S575257E
reagent heatsink fan and fan wiring (J1)
+24V PTC1 0.9A Thermal Control Module S575257E
reagent circulation fan and fan wiring (J1)
Volts Label Rating Location Protects
3-32 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Servo Low Level Controller PC
078D0080-02, Rev. A
Servo Low Level Controller PC Board S575262C
There are three fuses on the PC board; their functions are described below:
Volts Label Rating Location Protects+24V F0 2A 250V
5x20 mm FB 572808
Ring and Elevator Servo LLC PC board S575262C
index pin, motor 0 (P1)
+24V F1 2A 250V 5x20 mm FB 572808
Ring and Elevator Servo LLC PC board S575262C
cuvette elevator, motor 1 (P1)
+24V F2 2A 250V 5x20 mm FB 572808
Ring and Elevator Servo LLC PC board S575262C
ring, servo 0 (P1)
Connector Cable Connected to . . . P1 Plugs into J25 of Backplane PC board 575252F J3 Expansion PC board I/O J5 Expansion PC board I/O
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Servo Low Level Controller PC Board
078D0080-02, Rev. A
NOTE: Green LED lit and red LED not lit indicates normal operation of PC board. Red LED lit and green LED not lit indicates fault condition on PC board. Red LED lit and green LED flashing indicates reset condition.
LED FunctionCR1 On PC board, red LEDCR2 On PC board, green LED
3-34 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Fluid GIOB PC Board S575272E
078D0080-02, Rev. A
Fluid GIOB PC Board S575272E
Connector Cable Connected to . . . J10 571318 J43 Backplane PC board 575252F J11 571328 Diluter and FMI Pump Motors and Switches J12 571327 Diluter and FMI Pump Motors and Switches J13 571321 J11 Wash Fluid Interconnect PC board 575277B J14 571322 J12 Wash Fluid Interconnect PC board 575277B J15 571351 Manifold Valves J16 571350 Manifold Valves J17 571349 Pinch Valves J18 572314 Slotted sensor (cleaning solution Lid) J19 571304 Capacitive Level sensor (cleaning solution) J101 572302 J12 Backplane PC board 575252F
J102 571457 Reagent Fluid GIOB PC board 078-B027S02
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Fluid GIOB PC Board S575272E 3-35
078D0080-02, Rev. A
J103 Not Used
Volts Label Rating Location Protects+24V F1 2A 250V 5x20
mm FB 572808Fluid GIOB PC board PROT +24A to valve
drive circuitry (J13 and J15)
+24V F2 2A 250V 5x20 mm FB 572808
Fluid GIOB PC board PROT +24B to valve drive circuitry (J16 and J17)
+24V PTC1 0.9A Fluid GIOB PC board PROT VCC to diluters (J11 and J12) and to wash fluid interconnect PC board (J13 and J14)
Connector Cable Connected to . . .
3-36 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Wash Fluid Interconnect PC Board
078D0080-02, Rev. A
Wash Fluid Interconnect PC Board 575277B
Connector Cable Connected to . . . J1 Wash 2 Lid sensor J2 Wash 1 Lid sensor J3 Acid Level sensor J4 Base Lid sensor J5 Base Level sensor J6 Acid Lid sensor J7 Wash 1 Level sensor J8 Wash 3 Lid sensor J9 Wash 2 Level sensor J10 Wash 3 Level sensor J11 571321 J13 of Fluid GIOB PC board J12 571322 J14 of Fluid GIOB PC board J13 Wash 2 Vent J14 Acid Vent J15 Wash 3 Vent J16 Wash 1 Vent J17 Base Vent
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Probe Vertical PC Board
078D0080-02, Rev. A
Ancillary Probe Vertical PC Board S575282C
Connector Cable Connected to . . . J1 571283 J8 Sample and Ancillary Probe GIOB PC board
S575322H J2 Ancillary Probe Vertical motor J3 Ancillary Probe Volume Detector 571954
Sensor FunctionU4 Ancillary Vertical Home
3-38 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Handling GIOB PC Board
078D0080-02, Rev. A
Reagent Handling GIOB PC Board S575287C
Connector Cable Connected to . . .J1 571297 J1 of Reagent Probe 1 Pack-to-Pack PC board 575347J2 571300 J1 of Reagent Probe 2 Pack-to-Pack PC board 575347J3 571299 J1 of Reagent Probe 3 Pack-to-Pack PC board 575347J4 Reagent Probe 1 Pack-to-Pack motorJ5 Reagent Probe 2 Pack-to-Pack motorJ6 Reagent Probe 3 Pack-to-Pack motorJ7 571314 J41 of Backplane PC board 575252FJ8 571298 J3 of Reagent Probe 1 Heater Control/Volume Detect/
Probe Interconnect PC board 575402BJ9 571301 J3 of Reagent Probe 2 Heater Control/Volume Detect/
Probe Interconnect PC board 575407CJ10 571302 J3 of Reagent Probe 3 Heater Control/Volume Detect/
Probe Interconnect PC board 575412CJ12 571311 J1 Reagent Probe 1 Diluter, J1 Reagent Probe 2
Diluter, and J1 Reagent Probe 3 Diluter,J13 571316 J42 of Backplane PC board 575252F
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Handling GIOB PC Board
078D0080-02, Rev. A
J14 Reagent Barcode Scanner motorJ15 Reagent Barcode Scanner Home sensorJ16 571341 Reagent Barcode Scanner End of Travel sensorJ18 571366 J10 of Reagent Mixer Interconnect PC boardJ19 571367 J1 Upper Reagent Pack LED Indicators PC board
575387J22 571447 J1 Lower Reagent Pack LED Indicators PC board
575387J101 572300 J28 of Backplane PC board 575252F (Reagent Power)J102 571455 J103 of Cuvette GIOB PC board S575377FJ103 571456 J102 of Fluid GIOB PC board
LED Function CR1 Reagent Probe 1 Heater Status (Green LED indicates the temperature of
the probe is in tolerance) CR2 Reagent Probe 2 Heater Status (Green LED indicates the temperature of
the probe is in tolerance) CR3 Reagent Probe 3 Heater Status (Green LED indicates the temperature of
the probe is in tolerance) CR4 Reagent Probe 1 Volume Check (Amber LED indicates that air is in
front of the volume detector) CR5 Reagent Probe 2 Volume Check (Amber LED indicates that air is in
front of the volume detector) CR6 Reagent Probe 3 Volume Check (Amber LED indicates that air is in
front of the volume detector)
Volts Label Rating Location Protects+24V F2 2A 250V 5x20
mm FB 572808Reagent Handling GIOB PC board S575287C
reagent probe 1 heater drive (J8)
+24V F3 2A 250V 5x20 mm FB 572808
Reagent Handling GIOB PC board S575287C
reagent probe 2 heater drive (J9)
+24V F4 2A 250V 5x20 mm FB 572808
Reagent Handling GIOB PC board S575287C
reagent probe 3 heater
Connector Cable Connected to . . .
3-40 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Rack Present PC Board S575292B
078D0080-02, Rev. A
Rack Present PC Board S575292B
NOTE: Same PC board used three places.
Connector Cable Connected to . . . Direct J6 of Exit Queue PC board 575302CDirect J4 of Stat Area PC board 575307C Direct J5 of Stat Area PC board 575307C
Sensor Function U1 External Queue Full (J6 of 575302C) U1 Normal Rack Present (J4 of 575307C) U1 Stat Rack Present (J5 of 575307C)
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Fluid GIOB PC Board
078D0080-02, Rev. A
Reagent Fluid GIOB PC Board 078-B027S02
Table 3-1
Connector Cable Connected to . . . J2 572305 +24V from power block J7 572306 Water pump J8 572303 J5 of System fluid sensor PC board 575452C J15 571348 Reagent Fluids Manifold valves J101 572299 Power from J45 on Backplane PC board 575252F J102 571456 J103 on Reagent Handling GIOB PC board S575287C J103 571457 J102 on Fluid GIOB PC board P1 571764 Water pressure transducer
3-42 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Fluid GIOB PC Board
078D0080-02, Rev. A
P8 572304 Vacuum pump enable
Volts Label Rating Location Protects+24V F1 2A 250V 5x20
mm FB 572808Reagent Fluid GIOB PC board 078-B027S02
valve drive circuitry for all pressurized manifold valves and pinch valves (J15)
+24V PTC1 1.35A Reagent Fluid GIOB PC board J8 078-B027S02
System Fluid Sensor PC board 575452C (see F1 on this PC board) (J8)
+24V PTC2 1.35A Reagent Fluid GIOB PC board J8 078-B027S02
water pump (J7)
Connector Cable Connected to . . .
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Exit Queue Interconnect PC Board
078D0080-02, Rev. A
Exit Queue Interconnect PC Board 575302C
Connector Cable Connected to . . . J1 571285 J3 Sample and Ancillary Queue GIOB PC board
S575327D J2 571286 J4 Sample and Ancillary Queue GIOB PC board
S575327D J3 575012
575011Female on 575302C (J3) Male on 575307C (J2)
J4 571292 J1 Ancillary Queue PC board 575222B J5 571287 J1 Ancillary Entry PC board 575232D J6 Exit Queue Full sensor PC board S575292B J7 Spare motor J8 Spare sensor J9 571450 J46 of Backplane PC board 575252F
J10 Ancillary Barcode Head 571259
3-44 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Stat Area PC Board 575307C
078D0080-02, Rev. A
Stat Area PC Board 575307C
Connector Cable Connected to . . . J1 575012
575011J3 of Interconnect Exit Queue 575302C (Female Connector) Male connector J1 on this PC board
J2 Sample Barcode Reader 571259 J3 571291 Entry Queue PC board 575242B J4 Normal Rack Present PC board S575292B J5 Stat Rack Present PC board S575292B J6 Stat Pusher Home sensor J7 Stat Pusher Position sensor J8 571342 Rack loader PC board 575237B J9 110334 Start Stop PC board 575312 110333J10 Stat Pusher Motor
J11 Rack Loader Motor J12 Entry Queue Motor
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Probe GIOB PC
078D0080-02, Rev. A
Sample and Ancillary Probe GIOB PC Board S575322H
Connector Cable Connected to . . . J1 572947 Sample Level Sense PC board 575462E J2 Sample Pump 3-way valve J3 Bleed 3-way valve J4 Sample Pump Drive J5 571313 J39 of Backplane PC board 575252F J6 571280 J1 Sample and Ancillary Horizontal PC board
575212C J7 571284 J1 Sample Vertical PC board 575227 J8 571283 J1 Ancillary Vertical PC board S575282C J9 571281 J1 Ancillary Diluter and J1 Sample Diluter J10 Sample Probe Horizontal motor J11 Ancillary Probe Horizontal motor
3-46 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Probe GIOB
078D0080-02, Rev. A
J12 571290 Sample Diluter motor J13 Spare Sample motor J14 Spare Ancillary motor J101 572298 J17 of Backplane PC board 575252F J102 571453 J103 of Sample/Ancillary Queue PC Board S575327D J103 571454 J102 Cuvette GIOB S575377F
Volts Label Rating Location ProtectsVCC PTC1 1.35A Sample and
Ancillary Probe GIOB PC board S575322H
Sample Level Sense PC board 575462E (J1)Sample Probe Vertical Sensor PC board 575227 (J7) Auxiliary Probe Vertical PC board S575282C (J8)ancillary diluter, sample diluter (J9)
+24V PTC2 1.85A Sample and Ancillary Probe GIOB PC board S575322H
Sample Level Sense PC board 575462E (J1)pump valve (J2) bleed valve (J3) level sense pump (J4)Sample and Ancillary Probes Horizontal PC board 575212C (J6)
Connector Cable Connected to . . .
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Queue GIOB PC
078D0080-02, Rev. A
Sample and Ancillary Queue GIOB PC Board S575327D
Connector Cable Connected to . . . J1 571312 J40 Backplane PC board 575252F J2 Spare sensor J3 571285 J1 of Sample Interconnect PC board 575302C J4 571286 J2 of Sample Interconnect PC board 575302C J5 571330 J1 of Sample Tip Loader PC board 575352B J6 571296 J1 of Sample Queue PC board 575217B J7 571294 Inprocess Queue motor J8 Exit Pusher Motor J9 Spare motor J10 Exit Pusher Home sensor J11 Exit Pusher Position sensor J12 572265 J8 of Thermal Control PC board S575257E
J13 571282 Cuvette Loader motor J14 572297 Stop Action switch
3-48 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample and Ancillary Queue GIOB
078D0080-02, Rev. A
J15 571288 Cuvette Loader Up/Home sensor J16 571288 Cuvette Loader Down sensor J101 571411 J24 of Backplane PC board 575252F J102 571453 J102 Sample and Ancillary Probe GIOB PC board
S575322HJ103 571344 J36 of Backplane PC board 575252F
Volts Label Rating Location ProtectsVCC PTC1 1.85A Sample and
Ancillary Queue GIOB PC board S575327D
spare sensor (J2)Exit Queue Interconnect PC board 575302C (J3and J4)Sample Tip Loader PC board 575352B (J5)Inprocess Queue Tip Sensor PC board 575217B (J6)exit pusher home sensor (J10)exit pusher position sensor (J11) ring loader motor (J13)ring loader home sensor (J15)ring loader sensor (J16)
+24V PTC2 1.35A Sample and Ancillary QueueGIOB PC board S575327D
Exit Queue Interconnect PC board 575302C (J3)Inprocess Queue Tip Sensor PC board 575217B (J6)
Connector Cable Connected to . . .
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Cuvette Elevator PC Board 575337 3-49
078D0080-02, Rev. A
Cuvette Elevator PC Board 575337
Connector Cable Connected to . . . J1 Cuvette Motor J2 571308 of Cuvette GIOB PC board S575377F
Sensor Function U1 Elevator lower End of Travel sensor U2 Elevator upper End of Travel sensor
3-50 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Top Luminometer PC Board
078D0080-02, Rev. A
Top Luminometer PC Board 575342C
Connector Cable Connected to . . . J1 Luminometer motor J2 Waste Probe motor J3 Waste Probe Home sensor J4 J5 571307 J5 Bottom Luminometer PC board 575367
Sensor Function U1 Luminometer Direction sensor U2 Luminometer Position sensor U3 Luminometer Home sensor
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe Pack-to-Pack PC Board
078D0080-02, Rev. A
Reagent Probe Pack-to-Pack PC Board 575347
Connector Cable Connected to . . . J1 571297 J1 Reagent Handling GIOB PC board S575287C
Sensor Function U1 End of Travel sensor U2 Position sensor U3 Direction sensor
3-52 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample Tip Loader PC Board
078D0080-02, Rev. A
Sample Tip Loader PC Board 575352B
Connector Cable Connected to . . . J1 571330 J5 Sample and Ancillary GIOB PC board S575327D J2 Tip Tray and Band PC board S575397C (Tray Present) J3 110854 Sample Tip Loader Door sensor J4 Sample Tip Loader Home sensor J5 Sample Tip Loader Position sensor J6 Sample Tip Tray High sensor J7 Tip Tray and Band PC board S575397C (Band Present) J8 Sample Tip Loader motor J9 572296 Stop Action switch
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Servo Motor Interface PC Board
078D0080-02, Rev. A
Servo Motor Interface PC Board 575357C
Connector Cable Connected to . . . J1 Servo motor J2 Encoder J3 571319 or
119161J38 of Backplane PC board 575252FNOTE: Use cable part number 119161 only on systems that have white connectors for both J38 on the Backplane PC board and J3 on the Servo Motor Interconnect PC board
3-54 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Aspirate Probe PC Board 575362
078D0080-02, Rev. A
Aspirate Probe PC Board 575362
Connector Cable Connected to . . . J1 Aspirate Probe 1 motor J2 Aspirate Probe 2 motor J3 571286 J13 of Cuvette GIOB PC board S575377F J4 Aspirate Probe 3 motor J5 Aspirate Probe 4 motor J6 Aspirate Probe 1 Home sensor J7 Aspirate Probe 2 Home sensor
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Aspirate Probe PC Board 575362 3-55
078D0080-02, Rev. A
J8 Aspirate Probe 3 Home sensor J9 Aspirate Probe 4 Home sensor J10 Index Mechanism motor J11 Index Mechanism Pin Up sensor J12 Index Mechanism Pin Down sensor J13 Incubation Chamber Ring Tooth sensor
Connector Cable Connected to . . .
3-56 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Bottom Luminometer PC Board
078D0080-02, Rev. A
Bottom Luminometer PC Board 575367
Connector Cable Connected to . . . J1 112254 PMT 1 J2 112254 PMT 2 J3 571320 J37 Backplane PC board 575252F J4 571309 J6 of Cuvette GIOB PC board S575377F J5 571307 J5 Top Luminometer PC board 575342C J6 Thermistor PMT 1 J7 Thermistor PMT 2
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Preheater PC Board 575372 3-57
078D0080-02, Rev. A
Preheater PC Board 575372
Connector Cable Connected to . . . J1 Preheater Pusher motor J2 Cuvette Conveyor motor (cuvette loader) J3 Reflective Chute Empty sensor J4 Reflective Chute Full sensor J5 Cuvette Present sensor J6 Preheater thermistor J7 571310 J5 Cuvette GIOB PC board S575377F J8 Preheat Pusher In sensor (home) J9 Preheat Pusher Out sensor (end of travel)
3-58 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Cuvette GIOB PC Board S575377F
078D0080-02, Rev. A
Cuvette GIOB PC Board S575377F
Connector Cable Connected to . . . J1 116810 J5 of Fluids Heater Control PC board J2 571360 J7 of Thermal Module PC board S575257E
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Cuvette GIOB PC Board S575377F 3-59
078D0080-02, Rev. A
J3 571317 J44 of Backplane PC board 575252F J4 571308 J2 of Cuvette Elevator PC board 575337 J5 571310 J7 of Preheater PC board 575372 J6 571309 J4 of Bottom Luminometer PC board 575367 J7 571362 Ring A Thermistor 1 Incubation Chamber J8 Ring B Thermistor 2 Incubation Chamber J9 Ring C Thermistor 3 Incubation Chamber J10 Ring D Thermistor 4 Incubation Chamber J11 Home sensor J12 not used J13 571286 J3 of Wash Station and Index Mechanism PC board J14 571357 Tip Waste Bin sensorJ15 571358 Cuvette Waste Bin sensorJ16 572267 IonizerJ101 572301 J31 of Backplane PC board 575252F (GIOB Power)J102 571454 J103 Sample Probe GIOB PC board S575322HJ103 571455 J102 Reagent Handling GIOB PC board S575287C
Volts Label Rating Location Protects+5V PTC1 1.85A Cuvette GIOB
PC board S575377F
PRT-VCC to: bottom luminometer (J6)cuvette lift (J4)preheater (J5)ring home sensor (J11)ring EOT sensor (J12)Wash Station and Index Mechanism PC board (J13)tip waste bin (J14)cuvette waste bin (J15)
LED Function CR1 Green LED indicates the acid probe heater is working CR2 Green LED indicates the base probe heater is working CR3 Green LED indicates the system fluid heater is working
Connector Cable Connected to . . .
3-60 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Cuvette GIOB PC Board S575377F
078D0080-02, Rev. A
CR4 Not Used CR5 Not Used CR101 Red LED indicates a parity error or the system is in the reset mode
LED Function
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Pack LED Array PC Board
078D0080-02, Rev. A
Reagent Pack LED Array PC Board 575387
NOTE: Same PC board used two places.
NOTE: LEDs are numbered from left to right, CR1 through CR30.
Connector Cable Connected to . . . J1 571367 J19 on Reagent Handling GIOB S575287C PC board
(Upper LED Display) J1 571447 J22 on Reagent Handling GIOB PC Board S575287C
(Lower LED Display)
LED Function CR1 Green LED, column 0, row o CR2 Green LED, column 0, row 1 CR3 Green LED, column 0, row 2 CR4 Green LED, column 0, row 3 CR5 Green LED, column 0, row 4 CR6 Green LED, column 0, row 5 CR7 Green LED, column 0, row 6 CR8 Green LED, column 0, row 7
3-62 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Entry LEDs PC Board
078D0080-02, Rev. A
Ancillary Entry LEDs PC Board 575392B
Connector Cable Connected to . . . Hard Wired J6 Ancillary Entry PC board 57523
LED FunctionCR1 Ancillary Pack Yellow LED (from J6 of 575232D) CR2 Ancillary Pack Green LED (from J6 of 575232D)
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Tip Tray and Band PC Board S575397C
078D0080-02, Rev. A
Tip Tray and Band PC Board S575397C
NOTE: Same PC board used twice.
Connector Cable Connected to . . . Direct Wired 27146 J2 Tip Loader PC board 575352B (Tray Present) Direct Wired 27146 J7 Tip Loader PC board 575352B (Band Present)
Sensor FunctionU1 Tip Tray Present (J2 on 575352B) U1 Tip Band Present (J7 of 575352B)
3-64ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 1 Heater Control and
078D0080-02, Rev. A
Reagent Probe 1 Heater Control and Volume Detect Probe Interconnect PC Board 575402B
Connector Cable Connected to . . . J1 Reagent Probe 1 Horizontal motor J2 Reagent Probe 1 Vertical motor J3 571298 J8 Reagent Handling GIOB PC board S575287C J4 571954 Probe 1 Volume Detector J5 Probe 1 Heater Coil J6 571372 Horizontal Slotted Sensors PC board 575432B J7 Vertical Slotted Sensors
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 2 Heater Control and
078D0080-02, Rev. A
Reagent Probe 2 Heater Control and Volume Detect Probe Interconnect PC Board 575407C
Connector Cable Connected to . . . J1 Probe 2 Horizontal motor J2 Probe 2 Vertical motor J3 571301 J9 Reagent Handling GIOB PC board S575287C J4 571954 Probe 2 Volume Detector J5 Probe 2 Heater coil J6 571372 Horizontal Slotted Sensors PC board 575437 J7 Vertical Slotted sensors
3-66ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 3 Heater Control and
078D0080-02, Rev. A
Reagent Probe 3 Heater Control and Volume Detect Probe Interconnect PC Board 575412C
Connector Cable Connected to . . . J1 Probe 3 Horizontal Motor J2 Probe 3 Vertical Motor J3 571302 J10 Reagent Handling GIOB PC board S575287C J4 571954 Volume Detector 571954 J5 Heater Coil 571235 J6 571372 Horizontal Slotted Sensor PC board 575442B J7 Vertical Slotted sensor
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Mixer Interconnect PC Board 575422C
078D0080-02, Rev. A
Mixer Interconnect PC Board 575422C
Figure 3-1 Mixer Interconnect PC Board 575422C
Connector Cable Connected to . . . J1 Reagent Probe Shutter motor J2 Reagent Probe Shutter Open sensor J3 Reagent Probe Shutter Closed sensor J4 Primary Compartment Door tilt switch J5 Reagent Mixer motor J6 Reagent Mixer Upper sensor J7 Reagent Mixer Middle sensor J8 Reagent Mixer Lower sensor J10 571366 J18 Reagent Handling GIOB PC board S575287C J11 571326 Bar Code Scanner Shutter Closed sensor J12 Bar Code Scanner Shutter Open sensor J13 Bar Code Scanner Shutter Motor sensor J14 572265 J8 Temperature Control Module PC board S575257E J15 Primary Reagent Compartment, Loop A thermistor J16 Primary Reagent Compartment, Loop B thermistor J17 Primary Reagent Compartment, Loop C thermistor J18 Spare J19 Spare
3-68 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 1 Horizontal
078D0080-02, Rev. A
Reagent Probe 1 Horizontal Interconnect PC Board 575432B
Connector Cable Connected to . . . J1 571372 J6 Heater Control and Volume Detect Probe
Interconnect PC board 575402B
Sensor Function U1 Horizontal Position sensor U2 Horizontal Direction sensor U3 Horizontal End of Travel sensor
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 2 Horizontal Interconnect
078D0080-02, Rev. A
Reagent Probe 2 Horizontal Interconnect PC Board 575437
Connector Cable Connected to . . . J1 571372 J6 Heater Control and Volume Detect Probe
Interconnect PC board 575407C
Sensor Function U1 Horizontal Position sensor U2 Horizontal Direction sensor U3 Horizontal End of Travel sensor
3-70 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reagent Probe 3 Horizontal
078D0080-02, Rev. A
Reagent Probe 3 Horizontal Interconnect PC Board 575442B
Connector Cable Connected to . . . J1 571372 Heater Control and Volume Detect Probe Interconnect
PC board 575412C
Sensor Function U1 Horizontal Position sensor U2 Horizontal Direction sensor U3 Horizontal End of Travel sensor
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Reflective Chute Sensor Assembly PC
078D0080-02, Rev. A
Reflective Chute Sensor Assembly PC Board 575447B
Same assembly used twice.
Connector Cable Connected to . . . Direct wired J3 Preheater Assembly 575372 Direct wired J4 Preheater Assembly 575372
Sensor FunctionU1 Chute empty (J3 on 575372)U1 Chute full (J4 on 575372)
3-72 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 System Fluid Detect PC Board
078D0080-02, Rev. A
System Fluid Detect PC Board 575452C
Connector Cable Connected to . . . J1 Water Transfer Pump motor J2 Water Overflow sensor J3 Waste Working sensor J4 Waste Bulk Overflow sensor J5 572303 J8 Reagent Fluid GIOB PC board 078-B027S02 J6 Water Lid sensor J7 Waste Lid sensor J8 Water Low sensor J9 Waste Transfer Pump motor
LED Function CR2 Water Overflow CR3 Water Low CR4 Waste Working CR5 Waste Overflow
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 System Fluid Detect PC Board 575452C
078D0080-02, Rev. A
Volts Label Rating Location Protects +24V F1 2A 250V
5x20 mm FB 572808
System Fluid Sensor/Pump PC board 575452C
Reagent Fluid GIOB PC board 078-B027S02 (J5) water transfer pump (J1) and waste transfer pump circuitry (J9)
3-74 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Ancillary Pack Sensor PC Board
078D0080-02, Rev. A
Ancillary Pack Sensor PC Board 575457B
Connector Cable Connected to . . . Direct Wired 27146 J7 Ancillary Entry PC board 575232D (Ancillary Pack
Present)
Sensor Function U1 Ancillary pack present (J7 on 575232D)
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: 3 Sample Pressure Level Sense PC Board
078D0080-02, Rev. A
Sample Pressure Level Sense PC Board 575462E
Connector Cable Connected to . . . J1 572947 J1 Sample and Ancillary Probe GIOB PC board
S575322H
Sensor Function U4 Pressure sensor
078D0080-02, Rev. A
Appendix A: Safety Instructions
This information summarizes the established guidelines for handling laboratory biohazards. This summary is based on the guidelines developed by the Centers for Disease Control, the Clinical and Laboratory Standards Institute Document M29-A3, Protection of Laboratory Workers from Occupationally Acquired Infections, and the Occupational Safety and Health Administration’s Bloodborne Pathogens Standard.1–3
Protecting Yourself from BiohazardsUse this summary for general information only. It is not intended to replace or supplement your laboratory or hospital biohazard control procedures.
By definition, a biohazardous condition is a situation involving infectious agents biological in nature, such as the hepatitis B virus, the human immunodeficiency virus, and the tuberculosis bacterium. These infectious agents may be present in human blood and blood products and in other body fluids.
The following are the major sources of contamination when handling potentially infectious agents:
• needlesticks
• hand-to-mouth contact
• hand-to-eye contact
• direct contact with superficial cuts, open wounds, and other skin conditions that may permit absorption into subcutaneous skin layers
• splashes or aerosol contact with skin and eyes
To prevent accidental contamination in a clinical laboratory, strictly adhere to the following procedures:
• Wear gloves while servicing parts of the instrument that have contact with body fluids such as serum, plasma, urine, or whole blood.
• Wash your hands before going from a contaminated area to a noncontaminated area, or when you remove or change gloves.
• Perform procedures carefully to minimize aerosol formation.
• Wear facial protection when splatter or aerosol formation are possible.
• Wear personal protective equipment such as safety glasses, gloves, lab coats or aprons when working with possible biohazard contaminants.
• Keep your hands away from your face.
2 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: A Protecting Yourself from Barcode
078D0080-02, Rev. A
• Cover all superficial cuts and wounds before starting any work.
• Dispose of contaminated materials according to your laboratory’s biohazard control procedures.
• Keep your work area disinfected.
• Disinfect tools and other items that have been near any part of the instrument sample path or waste area with 10% v/v bleach.
• Do not eat, drink, smoke, or apply cosmetics or contact lenses while in the laboratory.
• Do not mouth pipet any liquid, including water.
• Do not place tools or any other items in your mouth.
• Do not use the biohazard sink for personal cleaning such as rinsing coffee cups or washing hands.
To prevent needlestick injuries, needles should not be recapped, purposely bent, cut, broken, removed from disposable syringes, or otherwise manipulated by hand.
References1. Centers for Disease Control. 1988. Update: Universal precautions for prevention of
transmission of human immunodeficiency virus, hepatitis B virus and other bloodborne pathogens in healthcare settings. MMWR, 37:377–382, 387, 388.
2. Clinical and Laboratory Standards Institute (formerly NCCLS). Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline - Third Edition. CLSI Document M29-A3.[ISBN 1-56238-567-4]. Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2005).
3. Federal Occupational Safety and Health Administration. Bloodborne Pathogens Standard. 29 CFR 1910. 1030.
Protecting Yourself from Barcode Scanner LasersTo avoid damage to the eyes, never look directly at the laser beam or at its reflection from a shiny surface. Never point a hand-held barcode scanner at anyone.
Laser Safety Classification of the ADVIA Centaur SystemDuring normal operation with all of the protective housings in place, the ADVIA Centaur system is classified as CDRH Class I and EN 60825-1 Class 1. No direct exposure to laser hazard exists for persons in the immediate area.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: A Protecting Yourself from Barcode
078D0080-02, Rev. A
.
Figure A-1 CDRH Class 1 Label
Some field service procedures require the removal of the protective housings that prevent human access to the laser radiation. The removal of the protective housings may change the classification of the system to CDRH Class 2 and EN60825-1 Class 2. All field service procedures must be followed precisely. Only Bayer HealthCare-trained field service personnel should perform procedures related to laser assemblies.
Figure A-2 CDRH Class 2 Label
The laser labels are positioned on the instrument as shown below.
Laser Safety Classification of the Barcode ScannersThe laser safety classification of the reagent and sample barcode scanners when they are unprotected by the system housings is CDRH Class 2 and EN 60825-1 Class 2. The laser safety classification of the hand-held barcode scanner used with the system is CDRH Class 2 and EN 60825-1 Class 2.
Reagent and Sample Barcode ScannersADVIA Centaur XP Operator’s Guide procedures for testing the reagent or sample barcode scanners contain the following laser warning:
LASER WARNING: To avoid damage to the eyes, never look directly at the laser beam or at its reflection from a shiny surface. Only Bayer-trained field service personnel should perform procedures related to laser assemblies. Refer to Appendix A, Protecting Yourself from Barcode Scanner Lasers, for more information.
The specifications for the laser optical assemblies in the ADVIA Centaur ancillary reagent, primary reagent, and sample barcode scanners are summarized in the following table:
The locations of the reagent and sample barcode scanners and their associated laser safety labels are shown below. The laser apertures for the sample and ancillary reagent barcode scanners are directed into the system away from the operator.
During normal operation, reflections from the sample barcode scanner laser beam may be visible to persons in the immediate area, but no laser safety hazard is associated with this exposure. The laser aperture for the primary reagent barcode scanner is directed toward the barcode labels at the end of the reagent packs inside the primary reagent compartment.
Characteristic SpecificationMaximum Power Output 1.0 mWWavelength 670 nmPulse Duration Continuous Wave (cw)Units of Beam Divergence 0.7 mr
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: A Protecting Yourself from Barcode
078D0080-02, Rev. A
During normal operation, the barcode scanner does not scan the reagent packs until the primary reagent compartment door is closed. When the primary reagent compartment door is closed, no operator exposure to the laser exists.
Figure A-3 Class 1 and Class 2 Laser Hazard Labels
Hand-held Barcode ScannerADVIA Centaur XP Operator’s Guide procedures that use the hand-held barcode scanner are introduced with the following laser warning:
1 Primary Reagent Barcode Scanner2 Sample Barcode Scanner3 Ancillary Reagent Barcode Scanner
LASER WARNING: The hand-held barcode scanner emits low-power visible laser radiation. To avoid damage to the eyes, never look directly at the laser beam or at its reflection from a shiny surface. Also, never point the scanner at anyone. Only Bayer-trained field service personnel should perform procedures related to laser assemblies. Refer to Appendix A, Protecting Yourself from Barcode Scanner Lasers, for more information.
A-6 header Melias Operating Manual
footer We never used this.....
This table contains a summary of the specifications for the laser optical assembly in the ADVIA Centaur hand-held barcode scanner:
The laser aperture and laser safety label for the hand-held barcode scanner are on the other side of the barcode scanner window:
Figure A-4 Handheld Barcode Scanner Laser Warning Label
Characteristic SpecificationMaximum Power Output 1.0 mWWavelength 650 nmPulse Duration 7.4 msUnits of Beam Divergence 2mr
1 Barcode scanner window.
078D0080-02, Rev. A
Appendix B: Service, Ordering, and Warranty
This section provides the following information:
• address of the Bayer authorized representative, which is the Bayer contact within the European community
• addresses for obtaining service and technical information and for ordering supplies
• system warranty and service delivery policy information
Bayer Authorized RepresentativeBayer Diagnostics Europe LimitedChapel Lane, Swords, Co. Dublin, Ireland
Limited Instrument Warranty and Service Delivery PolicyBayer HealthCare and its authorized distributors may provide customers who acquire new Bayer HealthCare instruments with a limited warranty either in a specific agreement or in standard language on their invoices. This limited warranty is designed to protect customers from the cost associated with repairing instruments that exhibit malfunctions due to defects in materials and/or workmanship during the warranty period.
Bayer Diagnostics, at its election, provides warranty service either by providing repair service of the instrument on site, or by exchanging the defective instrument or component, subject to the limitations and exclusions set forth in Replacement of Parts and Warranty and Service Exclusions below. Repairs, replacements or exchanges of instruments or components provided during the warranty or any additional service period, does not extend the warranty or service period beyond the initially agreed upon period.
When the customer calls for service, the Bayer HealthCare representative or authorized distributor informs the customer of the type of service available for the customer’s instrument, and instructs the customer as to how to obtain that service.
Warranty PeriodThe limited warranty period generally commences upon installation of the original instrument at the customer’s location and extends for a period of one year thereafter, unless otherwise specifically agreed to by and between Bayer HealthCare (or its authorized distributors) and customer in a writing signed by duly authorized representatives of both parties (sales representatives are generally not authorized representatives of Bayer HealthCare for these purposes).
B-2 ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B
078D0080-02, Rev. A
Additional Service Period
The customers, with some exceptions, may purchase additional service coverage beyond any initial warranty period as part of the original instrument acquisition for second or subsequent years beyond the original installation date. The customer’s original Purchase Invoice or appropriate Agreement Addendum must indicate the term in months for additional service coverage.
Service During Normal Hours
The customer may obtain service for instruments during normal business hours by contacting the nearest Bayer HealthCare location or authorized distributor. Refer to the list of Bayer HealthCare locations in this section.
Extent of a Service Call
Warranty or service calls generally include onsite repair or exchange of instruments or components, travel to the location of the instrument, and onsite labor during normal business hours. A warranty or service call is initiated by the customer by following the instructions on how to obtain service for the customer’s instrument. The service call is considered complete when any defects in material or workmanship have been corrected by repair or replacement and the instrument conforms to the applicable specifications. When service is complete, the customer receives a copy of the documentation detailing all work performed by the Bayer HealthCare representative or authorized distributor.
Service Outside Normal Hours
Customers, with some exceptions, may also request service to be delivered or an exchange to be initiated outside normal business hours, including evenings, weekend days, or nationally observed holidays, by contacting the nearest Bayer HealthCare location or authorized distributor. Service performed outside normal hours is subject to a surcharge unless the customer has in place a service product option that provides service at the time requested.
Replacement of Parts
In performing service, Bayer HealthCare or its authorized distributors provide appropriate parts to repair the instrument, or arranges for the exchange of the instrument or affected parts, at no charge with the exception of certain parts or subassemblies that are considered Customer Maintenance Items. Customer Maintenance Items include, but are not limited to, the following items: lamps, electrodes or sensors (which are covered by a separate warranty), reagents, calibrators, controls, paper, and pens. Consult the appropriate system operator’s manuals for a complete list of Customer Maintenance Items for any specific model of instrument.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B B-3
078D0080-02, Rev. A
Design Changes and Retrofitting of Instruments
Bayer HealthCare reserves the right to change the design or construction of specific models of instruments at any time without incurring any obligation to make such changes available to individual customers or instruments. If Bayer HealthCare notifies customers of a change that improves the performance or reliability of their instrument, and requests to retrofit that instrument, the customer must agree to allow Bayer HealthCare or an authorized distributor, at Bayer HealthCare’ expense, to retrofit components or make design changes, which does not adversely affect the instrument’s performance characteristics.
Key Operator Designation
Each customer designates a key operator who is available to Bayer HealthCare representatives to describe instrument malfunctions by telephone and/or to perform simple adjustments and corrections as requested. If a key operator is not designated or is unavailable when the customer requests service, the delivery of service may be delayed.
OSHA Requirements (US only)
When service is required at a customer location, the customer must provide the Bayer HealthCare representative with adequate facilities that comply with the regulations of the Secretary of Labor under the Occupational Safety and Health Act (OSHA) of 1970, as amended.
Warranty and Service Exclusions
The following exclusions are in addition to any exclusions provided for in any written warranty or service agreement.
IF ANY OF THE FOLLOWING EVENTS OCCUR, THE WARRANTY OR SERVICE PROVISIONS DO NOT APPLY:
1. Repairs or modifications have been made to the instrument by someone other than an authorized Bayer HealthCare representative.
2. The instrument has been operated using accessories and supplies other than Bayer HealthCare brand accessories, or consumable supplies and/or reagents not having the same grade, quality, and composition as defined in the system operator’s manuals.
3. Bayer HealthCare has notified customers of a change that improves the performance or reliability of their instrument and customer has not agreed to retrofit or make design changes to the instrument.
4. Customer did not purchase the instrument from Bayer HealthCare or one of its authorized distributors.
5. The instrument has not been installed within 90 days of shipment to the customer’s facility unless otherwise specified.
6. The customer has not performed appropriate customer maintenance procedures, as outlined in the system operator’s manuals.
7. The instrument has been misused or used for a purpose for which it was not intended.
8. The instrument has been damaged in transit to the customer or damaged by the customer while moving or relocating it without supervision by a Bayer HealthCare representative.
9. Damage was caused by floods, earthquakes, tornados, hurricanes, or other natural or man-made disasters.
10. Damage was caused by Acts of War, vandalism, sabotage, arson, or civil commotion.
11. Damage was caused by electrical surges or voltages exceeding the tolerances outlined in the system operator’s manuals.
12. Damage was caused by water from any source external to the instrument.
13. The customer has purchased an alternative agreement whose terms of warranty or service supersede these provisions.
Bayer HealthCare or its authorized distributors can invoice customers, at current standard labor and parts rates, for instruments repaired to correct damage or malfunctions due to any of the reasons listed above.
OTHER THAN AS STATED ABOVE, THERE ARE NO OTHER WARRANTIES, EXPRESS OR IMPLIED, WITH RESPECT TO THE INSTRUMENT, ITS SALE TO THE CUSTOMER, ITS LEASE TO THE CUSTOMER, OR THE SALE OF THE INSTRUMENT TO THE CUSTOMER AT THE EXPIRATION OR TERMINATION OF THE LEASE AGREEMENT.
BAYER HEALTHCARE SPECIFICALLY DISCLAIMS ANY AND ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE OR PURPOSE. BAYER HEALTHCARE’S LIABILITY FOR BREACH OF ANY WARRANTY OR SERVICE AGREEMENT SHALL BE LIMITED ONLY TO THE REPAIR OR REPLACEMENT OF DEFECTIVE EQUIPMENT AND SHALL NOT INCLUDE ANY DAMAGES OF ANY KIND, WHETHER DIRECT, INDIRECT, INCIDENTAL, CONTINGENT, OR CONSEQUENTIAL. BAYER DIAGNOSTICS SHALL NOT BE LIABLE FOR DELAY FROM ANY CAUSE IN PROVIDING REPAIR OR EXCHANGE SERVICE.
ANY LIMITATIONS OR OTHER PROVISIONS NOT CONSISTENT WITH APPLICABLE LAW IN PARTICULAR JURISDICTIONS OR A SPECIFIC WRITTEN AGREEMENT DO NOT APPLY TO CUSTOMERS IN THOSE JURISDICTIONS OR SUBJECT TO THOSE AGREEMENTS.
Information for Technical AssistanceRefer to the procedures in this appendix to provide system information that you may need when you call for technical assistance.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B Information for Technical Assistance B-5
078D0080-02, Rev. A
Printing the System Serial Number and Versions Use this procedure to print information about the system, such as the serial number and the current version of software used on your ADVIA Centaur XP system.
1. At the workspace, select System Status.
2. Select About the System.
3. At the workspace, select Print.
4. Select Print Workspace.The serial number is displayed on the System Status button. The version information is displayed at the About the System window.
Printing the Event LogUse this procedure to print the list of system events displayed at the System – Event Log window.
You can display specific information at the System – Event Log window by selecting the appropriate View and Sort combinations:
• The View that you select determines the information that the system displays. For example, if you select Errors, the system displays only the events that are errors.
• The Sort that you select determines the order in which the system displays the events. For example, if you select Severity, the system displays the events in order of most severe to least severe.
The following table describes the View and Sort combinations you can use to perform some routine tasks:
1. At the workspace, select Event Log.
2. Select the appropriate view and identify the problem you want to solve.To search for a specific event code in the event log, select Event Code in Search for and then enter the event code.
3. Select Print.
4. Select Print Workspace.
Task View SortDisplay the events in reverse chronological in order of when the events occurred
Events Date
Display the critical and warning events in order of most severe to least severe
Errors Severity
Display the events by subsystem Events Subsystem
B-6 header Melias Operating Manual
footer We never used this.....
Printing Windows on the WorkspaceUse this procedure to print the workspace, including any windows that are open. Windows that can be helpful in resolving problems include Quality Control – Data, Calibration –Summary, Calibration – Data, and Worklist – Summary. Open the appropriate windows
before you print the workspace.
1. At the workspace, select Print.
2. Select Print Workspace.
Printing a Runtime ReportUse this procedure to print the report of all results generated (including replicates, ratios, and means) for all samples run.
1. At the workspace, select Print.
2. Select Report Options.
3. At the Print – Report Options window, select Runtime Report.
4. Select Print Report.
Printing Calibration DataYou can print detailed information on specific calibration data as described in Section 5, Calibration.
ContactsThis section provides the following information:
• the address of the Bayer authorized representative, which is the Bayer contact within the European community
• the Bayer addresses for obtaining service and technical information and for ordering supplies
Bayer Authorized RepresentativeBayer Diagnostics Europe LimitedChapel Lane, Swords, Co. Dublin, Ireland
AddressesFor technical assistance contact your local technical support provider. For customer service or additional information contact your local technical support distributor.
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B Contacts B-7
078D0080-02, Rev. A
Bayer S.A.División DiagnósticosRicardo Gutiérrez 3652 B1605EHD Munro - Buenos AiresArgentina+54 (0) 11-4762-7000Bayer Australia LimitedABN 22 000 138 714Diagnostics Division2 Keith Campbell CourtScoresby Victoria 3179Australia+61 (0) 3-9212-8444Bayer Austria GesmbH Geschäftsbereich DiagnostikaLerchenfelder Gürtel 9-11 A-1164 Wien, Austria+43 (0) 1-711-46-2424Bayer s.a.-n.v.Division DiagnosticsAvenue Louise 143 Louizalaan1050 Bruxelles-Brussel,Belgium+32 (0) 2-535-66-81
Bayer S.A.Produtos Diagnósticos Rua Domingos Jorge 110004779-900 - São Paulo - SPBrazil+55-11-5694-5166Bayer Inc.77 Belfield RoadToronto, OntarioCanada M9W 1G6416-248-0771Bayer s.r.o.Oddìlení DiagnostikaZastoupení pro Èeskou a Slovenskou republikuLitvínovská 609/3CZ-190 21 Praha 9Èeská republika+420 266101442Bayer S.A.División DiagnósticaA.A. 80387Av. De las Américas No 57-52Santafé de Bogotá, D.C. - Colombia+571 (09) 423-4199Bayer HealthCare - DiagnosticsNørgaardsvej 32DK-2800 Kgs. Lyngby, Denmark+45-45-23-50-00
B-8 header Melias Operating Manual
footer We never used this.....
Bayer OYSuomalaistentie 7FIN 02270 Espoo, Finland+358-9-887-887Bayer Diagnostics Tour Horizon 52, quai de Dion Bouton 92807 Puteaux Cedex, France +33 (0) 1-49-06-56-00Bayer Vital GmbHDiagnostikaSiemensstraße 3D-35463 Fernwald, Germany+49 (0) 641 4003-0Bayer Hellas AG 18-22 Sorou str.GR-151 25 MarousiAthens, Greece+30 (0) 210 6187 500Bayer HealthCare Co. Ltd.20/F Gee Chang Hong Centre65 Wong Chuk Hang RoadHong Kong+852-28147337Bayer Hungária Kft.Alkotás PointH-1123 Budapest, HungaryAlkotás u. 50+36 (06) 1 4874 100Bayer Diagnostics India Limited589, SayajipuraAjwa RoadBaroda - 390 019Gujarat, India+91 (0) 26-5562720 Bayer S.p.A.Divisione DiagnosticiVia Grosio 10/420151 Milano, Italia+39-2-3978-1
Bayer Medical Ltd.Marunouchi Kitaguchi Bldg.1-6-5, Marunouchi, Chiyoda-kuTokyo 100-8263, JapanBayer Diagnostics Korea LimitedKye Myung Bldg. 4FMyungil-Don, 48-7Kangdong-Ku Seoul 134070Korea+82 (0) 2-428-5987Bayer (Malaysia) Sdn. Bhd11th FloorWisma DamansaraJalan Semantan50490 Kuala LumpurMalaysia+603 (0) 62093088Bayer de México, S.A. de C.V.División DiagnósticosVía Morelos 330-ECol. Santa Clara C.P. 55540Ecatepec, Estado de MéxicoCentro de Atención Telefónica (CAT)+52 55 57 28 33 06Health CareDivision DiagnosticsEnergieweg 13641 RT MijdrechtThe Netherlands+31 (0) 297-280666Bayer New Zealand LtdDiagnostics Business Group3 Argus PlaceGlenfieldAuckland, New Zealand+64 9-441-8549
ADVIA Centaur and ADVIA Centaur XP Parts and Repair Service Guide: B Contacts B-9
078D0080-02, Rev. A
Bayer B.V.Bayer ASDrammensvn. 147 BNO-0277 Oslo, Norway+47 24 11 18 00Bayer Sp. Z o.o.Al. Jerozolimskie 15802-326 Warszawa, Polska+48 (0) 22-572-3500Bayer Diagnostics Europe Ltd(sucursal em Portugal) Rua Quinta do Pinheiro, 52795-653 Carnaxide, Portugal+351-21-416-4227Bayer Puerto Rico Inc. Diagnostics DivisionVictoria Industrial Park Building 1 Carolina, Puerto Rico787-752-8989Bayer (South East Asia) Pte Ltd.No. 9 Benoi SectorSingapore 629844+65 6261 3389Bayer (Pty) Ltd.Healthcare Division27 Wrench RoadIsando 1600, South Africa+27 (0) 11-921-5048
Química Farmacéutica Bayer, S.A.División Diagnósticosc/ Pau Claris, 19608037 Barcelona, España+34-93 4956500Bayer ABDrakegatan 1S-402 24 Göteborg, Sweden+46-31-83-98-00Bayer (Schweiz) AGGB DSGrubenstrasse 6CH-8045 Zürich Switzerland+41 (0) 1-465-81-11Bayer plc Diagnostics DivisionBayer HouseStrawberry HillNewbury, RG14 1JAUnited Kingdom+44 (0) 1635-563000