Upload
byron-cuevas-mendoza
View
729
Download
201
Embed Size (px)
Citation preview
BC-5380 Auto Hematology Analyzer
Service Manual
I
Copyright
© 2008 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Service Manual, the issued Date is 2008-12 (Version: 1.0).
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray)
owns the intellectual property rights to this Mindray product and this manual. This manual may
refer to information protected by copyrights or patents and does not convey any license under
the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability
arising out of any infringements of patents or other rights of third parties.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure
of the information in this manual in any manner whatsoever without the written permission of
Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rent, adaption and translation of this manual
in any manner whatsoever without the written permission of Mindray is strictly forbidden.
, are the registered trademarks or trademarks owned by Mindray in China and
other countries. All other trademarks that appear in this manual are used only for editorial
purposes without the intention of improperly using them. They are the property of their
respective owners.
Responsibility on the Manufacturer Party
Contents of this manual are subject to changes without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for safety, reliability and performance of this product only in the
condition that:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
II
the electrical installation of the relevant room complies with the applicable national and
local requirements;
the product is used in accordance with the instructions for use.
This equipment must be operated by skilled/trained medical professionals.
It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.
Be sure to operate the analyzer under the situation specified in this manual; otherwise, the analyzer will not work normally and the analysis results will be unreliable, which would damage the analyzer components and cause personal injury.
III
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the
improper use or application of the product or the use of parts or accessories not approved by
Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
any Mindray product which has been subjected to misuse, negligence or accident;
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
any product of any other manufacturer.
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to Mindray,
the following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear on
the outside of the shipping container. Returned shipments will not be accepted if the
Mindray number is not clearly visible. Please provide the model number, serial number,
and a brief description of the reason for return;
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges);
3. Return address: Please send the part(s) or equipment to the address offered by Customer
Service department.
IV
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan,ShenZhen 518057, P.R.China,
Phone: +86 755 26582479 26582888
Fax: +86 755 26582934 26582500
1
Table of Contents
1 Using This Manual............................................................................................... 1-1 1.1 Introduction............................................................................................................ 1-1
1.2 Who Should Read This Manual............................................................................. 1-2
1.3 How to Find Information ........................................................................................ 1-3
1.4 Conventions Used in This Manual......................................................................... 1-4
1.5 Symbols ................................................................................................................. 1-5
2 System Structure................................................................................................. 2-1 2.1 Introduction............................................................................................................ 2-1
2.2 Fluidic System ....................................................................................................... 2-2
2.3 Hardware System.................................................................................................. 2-3
2.4 Main Structure ....................................................................................................... 2-4
2.5 Software structure ............................................................................................... 2-11 2.5.1 Menu structure ....................................................................................... 2-11 2.5.2 Passwords.............................................................................................. 2-31
3 Instrument Installation and Software Upgrade................................................. 3-1 3.1 Preparations .......................................................................................................... 3-1
3.1.1 Purpose .................................................................................................... 3-1 3.1.2 Tools ......................................................................................................... 3-1 3.1.3 Accessories .............................................................................................. 3-1
3.2 Installation Requirements ...................................................................................... 3-2 3.2.1 Space Requirements ................................................................................ 3-2 3.2.2 Power Requirements................................................................................ 3-2 3.2.3 Environment Requirements...................................................................... 3-2 3.2.4 PC Configuration Requirements .............................................................. 3-3
3.3 Package Checking and Unpacking ....................................................................... 3-7 3.3.1 Checks before unpacking......................................................................... 3-7 3.3.2 Unpacking the main unit........................................................................... 3-7 3.3.3 Checking packing list................................................................................ 3-9
3.4 Removal and Installation ..................................................................................... 3-10
3.5 Connections......................................................................................................... 3-23
3.6 Start-up ................................................................................................................ 3-27 3.6.1 Inspection before Startup ....................................................................... 3-27
3.7 Setup and Adjustment ......................................................................................... 3-35
3.8 Testing Other Functions....................................................................................... 3-41
3.9 Confirming Gain................................................................................................... 3-42
3.10 Calibration and Performance Test ....................................................................... 3-47
3.11 Software upgrade ................................................................................................ 3-53
4 Fluidic System ..................................................................................................... 4-1 4.1 Introduction of Fluidic Parts ................................................................................... 4-1
4.1.1 Piercing needle......................................................................................... 4-1 4.1.2 Probe Wipe............................................................................................... 4-1
Table of Contents
2
4.1.3 Pumps ...................................................................................................... 4-1 4.1.4 Syringes ................................................................................................... 4-1 4.1.5 Valves....................................................................................................... 4-1 4.1.6 Baths ........................................................................................................ 4-1 4.1.7 Volumetric tube......................................................................................... 4-2 4.1.8 Filters........................................................................................................ 4-2
4.2 Fluidic System ....................................................................................................... 4-3 4.2.1 Reagent volume required......................................................................... 4-3 4.2.2 Fluidic System Drawing............................................................................ 4-5 4.2.3 WBC Measurement Procedures .............................................................. 4-6 4.2.4 RBC/PLT Measurement Procedures........................................................ 4-6 4.2.5 HGB Measurement Procedures ............................................................... 4-7
4.3 Sequence of Whole Blood CBC+DIFF Measurement ........................................... 4-8 4.3.1 Sampling and dispensing procedures...................................................... 4-8 4.3.2 DIFF Channel ......................................................................................... 4-10 4.3.3 WBC/HGB Channel................................................................................ 4-13 4.3.4 RBC/PLT Channel .................................................................................. 4-15 4.3.5 Sequence of Predilute Mode CBC+DIFF Measurement........................ 4-18 4.3.6 Sequence of CBC Measurement ........................................................... 4-18
5 Hardware System ................................................................................................ 5-1 5.1 Mother board ......................................................................................................... 5-3
5.1.1 Introduction............................................................................................... 5-3 5.1.2 Board Composition................................................................................... 5-3 5.1.3 Adjustment and Test Points.................................................................... 5-26 5.1.4 Removal ................................................................................................. 5-28 5.1.5 Troubleshooting...................................................................................... 5-30
5.2 Data board........................................................................................................... 5-32 5.2.1 Introduction............................................................................................. 5-32 5.2.2 Board composition.................................................................................. 5-32 5.2.3 Adjustment and Test Points.................................................................... 5-45 5.2.4 Disassembly and assembly method....................................................... 5-49 5.2.5 Troubleshooting...................................................................................... 5-51
5.3 Drive board .......................................................................................................... 5-59 5.3.1 Introduction............................................................................................. 5-59 5.3.2 Board composition.................................................................................. 5-59 5.3.3 Adjustment and Test Points.................................................................... 5-75 5.3.4 Disassembly and assembly method....................................................... 5-78 5.3.5 Troubleshooting...................................................................................... 5-80
5.4 Autoloader board ................................................................................................. 5-99 5.4.1 Introduction............................................................................................. 5-99 5.4.2 Board Composition................................................................................. 5-99 5.4.3 Adjustment and Test Points.................................................................. 5-109 5.4.4 Disassembly and assembly method..................................................... 5-110 5.4.5 Troubleshooting.................................................................................... 5-112
5.5 Power board ...................................................................................................... 5-117 5.5.1 Introduction........................................................................................... 5-117 5.5.2 Board Composition............................................................................... 5-117 5.5.3 Adjustment and Test Points.................................................................. 5-122 5.5.4 Replacement and Connection.............................................................. 5-124 5.5.5 Troubleshooting.................................................................................... 5-127
5.6 Volumetric and pressure detecting board.......................................................... 5-129 5.6.1 Introduction........................................................................................... 5-129 5.6.2 Board Composition............................................................................... 5-129 5.6.3 Disassembly and assembly method..................................................... 5-132 5.6.4 Troubleshooting.................................................................................... 5-134
Table of Contents
3
5.7 Liquid-level detecting board .............................................................................. 5-140 5.7.1 Introduction........................................................................................... 5-140 5.7.2 Board composition................................................................................ 5-140 5.7.3 Replacement and Connection.............................................................. 5-143 5.7.4 Troubleshooting.................................................................................... 5-145
5.8 Laser Control Board .......................................................................................... 5-148 5.8.1 Introduction........................................................................................... 5-148 5.8.2 Board Composition............................................................................... 5-148 5.8.3 Adjustment and Test Points.................................................................. 5-150 5.8.4 Assembly and disassembly.................................................................. 5-151 5.8.5 Troubleshooting.................................................................................... 5-151
5.9 Pre-amplification board...................................................................................... 5-154 5.9.1 Introduction........................................................................................... 5-154 5.9.2 Board Composition............................................................................... 5-154 5.9.3 Adjustment and Test Points.................................................................. 5-158 5.9.4 Disassembly and assembly.................................................................. 5-158 5.9.5 Troubleshooting.................................................................................... 5-158
5.10 Indicator Board .................................................................................................. 5-160 5.10.1 Introduction........................................................................................... 5-160 5.10.2 Board Composition............................................................................... 5-160 5.10.3 Disassembly and assembly method..................................................... 5-161 5.10.4 Troubleshooting.................................................................................... 5-162
5.11 Key Board.......................................................................................................... 5-163 5.11.1 Introduction........................................................................................... 5-163 5.11.2 Board Composition............................................................................... 5-163 5.11.3 Disassembly and assembly method..................................................... 5-164 5.11.4 Troubleshooting.................................................................................... 5-165
6 Maintenance......................................................................................................... 6-1 6.1 Maintenance Modules and the Corresponding Settings ....................................... 6-1
6.2 General .................................................................................................................. 6-3
6.3 Disassembling the Panels ..................................................................................... 6-4 6.3.1 Removing the left door ............................................................................. 6-4 6.3.2 Removing the right door........................................................................... 6-5 6.3.3 Removing the top cover ........................................................................... 6-6 6.3.4 Removing the back panel......................................................................... 6-7 6.3.5 Removing the front door........................................................................... 6-8 6.3.6 Removing the front cover assembly......................................................... 6-9
6.4 Replacing the Valves, Pumps and Syringes........................................................ 6-11 6.4.1 Replacing valves .................................................................................... 6-13 6.4.2 Replacing the Pressure Chamber .......................................................... 6-15 6.4.3 Replacing the Vacuum Chamber ........................................................... 6-17 6.4.4 Replacing the Syringe Assembly............................................................ 6-18 6.4.5 Replacing the Waste Pump.................................................................... 6-24
6.5 Replacing the Bath/Aperture Assembly............................................................... 6-26
6.6 Replacing the Sampling Module and Adjusting Position ..................................... 6-29 6.6.1 Replacing the Sample Probe and Wipe ................................................. 6-29 6.6.2 Replacing the Optical Sensor................................................................. 6-31 6.6.3 Removing the Sampling Module ............................................................ 6-33
6.7 Maintaining and Replacing the DIFF Reaction Bath ........................................... 6-37
6.8 Autoloader Assembly (Closed-Tube)................................................................... 6-39 6.8.1 Removing and replacing the closed-tube sample compartment assembly 6-39 6.8.2 Removing and replacing the detection switches of the sample transport unit
Table of Contents
4
6-45 6.8.3 Replacing the sample transport unit....................................................... 6-46
6.9 Mix mechanism.................................................................................................... 6-58
6.10 Adjustment........................................................................................................... 6-66 6.10.1 Adjusting Sample Probe Position........................................................... 6-66 6.10.2 Adjusting Pincher Position ..................................................................... 6-73
7 Optical System..................................................................................................... 7-1 7.1 Optical System Adjustment and Troubleshooting.................................................. 7-1
7.2 Removing and Installing Optical System Assemblies............................................ 7-3 7.2.1 Laser driver board .................................................................................... 7-3 7.2.2 Front light assembly ................................................................................. 7-4 7.2.3 Flow cell assembly ................................................................................... 7-5 7.2.4 Rear light collimator assembly ................................................................. 7-7 7.2.5 Beam splitter assembly ............................................................................ 7-7 7.2.6 Rear light collector assembly ................................................................... 7-8 7.2.7 Rear light detector assembly.................................................................. 7-10
7.3 Adjustment........................................................................................................... 7-12 7.3.1 Coarse adjustment ................................................................................. 7-12 7.3.2 Preparation before fine tuning................................................................ 7-18 7.3.3 Fine tuning.............................................................................................. 7-19 7.3.4 Gain setup .............................................................................................. 7-24
7.4 Troubleshooting ................................................................................................... 7-27 7.4.1 Laser spot-deviation............................................................................... 7-27 7.4.2 Flow cell clog.......................................................................................... 7-27 7.4.3 Dirty flow cell .......................................................................................... 7-28
8 Troubleshooting .................................................................................................. 8-1 8.1 Error code and information.................................................................................... 8-1
8.2 Errors indicated by error messages ...................................................................... 8-9 8.2.1 Pressure errors......................................................................................... 8-9 8.2.2 Reagent errors ......................................................................................... 8-9 8.2.3 Hardware errors ..................................................................................... 8-10 8.2.4 Measurement errors............................................................................... 8-10 8.2.5 Temperature errors................................................................................. 8-12 8.2.6 Scattergram errors ................................................................................. 8-13
9 Appendixes ......................................................................................................... A-1 A. Accessories ...........................................................................................................A-1
B. List of Wearing Parts .............................................................................................B-1
C. Fluidic diagram ..................................................................................................... C-1
D. Pump and Valve Function Table ........................................................................... D-1
E. Tubing....................................................................................................................E-1
F. Method to identify cross network cable and direct-connected network cable .......F-1
1-1
1 Using This Manual
1.1 Introduction
The chapter explains how to use the BC-5380 service manual. This manual provides the
reference information and procedures needed in servicing the BC-5380. Before servicing the
BC-5380, read and understand the manual carefully for servicing the equipment properly and
for your safety.
This manual is to be used in conjunction with the operator’s manual of BC-5380.It does not
contain information and procedures already covered in the operator’s manual of BC-5380.
Be sure to operate and service the analyzer strictly as instructed in this manual and the operator’s manual.
Using This Manual
1-2
1.2 Who Should Read This Manual
To use this service manual effectively, the reader should posses:
A thorough understanding of
Basic electronic and fluidics principles and devices
Reagent systems
Quality control
Troubleshooting concepts
The ability to
Use basic mechanical tools and understand related terminology
Use a digital voltmeter (DVM) and an oscilloscope
Read fluidics schematics and understand related terminology
Read electronic schematics and understand related terminology
Using This Manual
1-3
1.3 How to Find Information
This service manual comprises 8 chapters and 5 appendices. Refer to the table below to find
the information you need.
If you want to … See …
learn about the system structure and the basic
principle of the BC-5380
Chapter 2 System Structure
learn about the installation requirements and how to
upgrade the BC-5380 software
Chapter 3 System Installation and Software Upgrade
learn about the fluidic system of the BC-5380 Chapter 4 Fluidic System
learn about the hardware system of the BC-5380 Chapter 5 Hardware System
learn about how to maintain the BC-5380 Chapter 6 Maintenance
learn about the optical system of the BC-5380 Chapter 7 Optical System
learn about how to troubleshoot the common errors
of the BC-5380
Chapter 8 Troubleshooting
learn about the main spare parts of the BC-5380 Appendix A Accessories
learn about the main wearing parts of the BC-5380 Appendix B List of Wearing Parts
learn about the schematic diagram of the fluidic
system of the BC-5380
Appendix C Fluidic Diagram
learn about the function of each valve and pump of
the BC-5380
Appendix D Pump and Valve Function Table
learn about the tubing connection of the BC-5380 Appendix E Tubing
learn about how to identify cross network cable and
direct-connected network cable
Appendix F Method to Identify Cross Network Cable and Direct-Connected Network Cable
Using This Manual
1-4
1.4 Conventions Used in This Manual
This manual uses certain typographical conventions to clarify meaning in the text:
all capital letters enclosed in [ ] indicate a key name (on the external keyboard), such as
[ENTER].
bold letters included in “ ” indicate text you can find on the screen, such as “Clean”.
bold letters indicate chapter titles, such as Chapter 1 Using This Manual.
All illustrations in this manual are provided as examples only. They may not necessarily reflect
your analyzer setup or data displayed.
Using This Manual
1-5
1.5 Symbols
You will find the following symbols in this manual:
When you see… Then…
read the statement below the symbol. The statement is
alerting you to an operating hazard that can cause
personnel injury.
read the statement below the symbol. The statement is
alerting you to a possibility of analyzer damage or
unreliable analysis results.
read the statement below the symbol. The statement is
alerting you to information that requires your attention.
read the statement below the symbol. The statement is
alerting you to a potentially biohazardous condition.
You may find the following symbols on the analyzer, reagents, controls or calibrators.
When you see… It means…
CAUTION, CONSULT ACCOMPANYING
DOCUMENTS.
BIOLOGICAL RISK
HIGH VOLTAGE
WARNING, LASER BEAM
WARNING, HOT SURFACE
PROTECTIVE EARTH (GROUND)
Using This Manual
1-6
EARTH (GROUND)
ALTERNATING CURRENT
FOR IN VITRO DIAGNOSTIC USE
BATCH CODE
USE BY
SERIAL NUMBER
CATALOG NUMBER (FOR CONTROLS)
USE BY (YYYY-MM-DD) (FOR
CONTROLS)
DATE OF MANUFACTURE
MANUFACTURER
TEMPERATURE LIMITATION
CONSULT INSTRUCTIONS FOR USE
IRRITATING SUBSTANCE
THE FOLLOWING DEFINITION OF THE
WEEE LABEL APPLIES TO EU MEMBER
STATES ONLY: THE USE OF THIS
Using This Manual
1-7
SYMBOL INDICATES THAT THIS
PRODUCT SHOULD NOT BE TREATED
AS HOUSEHOLD WASTE. BY ENSURING
THAT THIS PRODUCT IS DISPOSED OF
CORRECTLY, YOU WILL HELP PREVENT
BRINGING POTENTIAL NEGATIVE
CONSEQUENCES TO THE
ENVIRONMENT AND HUMAN HEALTH.
FOR MORE DETAILED INFORMATION
WITH REGARD TO RETURNING AND
RECYCLING THIS PRODUCT, PLEASE
CONSULT THE DISTRIBUTOR FROM
WHOM YOU PURCHASED THE
PRODUCT.
Be sure to observe the following precautions for the safety of patients, operators and yours
when you are servicing the analyzer.
Using This Manual
1-8
It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.
Never use combustible gas (e.g. anesthetic) or combustible liquid (e.g. ethanol) around the analyzer. Otherwise, the risk of explosion may exist.
Contacting exposed electronic components while the equipment is attached to power can cause personal injury from electric shock or damage to electronic components. Power down before removing covers to access electronic components.
Connect the analyzer to a socket having sole fuse and protective switch. Do not use the same fuse and protective switch with other equipment (e.g. life supporting equipment). Otherwise, the equipment failure, over current or impulse current that occurs at the startup moment may lead to tripping.
To prevent personal injury during the maintenance, keep your clothes, hairs and hands from the moving parts, such as sample probe, clipper and piercer.
Possible mechanical movement of the warned position may lead to personal injury during the normal operation, removal and maintenance.
Be sure to dispose of reagents, waste, samples, consumables, etc. according to government regulations.
The reagents are irritating to eyes, skin and diaphragm. Wear proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of water and if necessary, go see a doctor; if the reagents accidentally spill into your eyes, wash them off with plenty of water and immediately go see a doctor.
Using This Manual
1-9
Improper maintenance may damage the analyzer. Maintain the analyzer strictly as instructed by the service manual and inspect the analyzer carefully after the maintenance.
For problems not mentioned in the service manual, contact Mindray customer service department for maintenance advice.
To prevent personal injury or damage to equipment components, remove metal jewelry before maintaining or servicing electronic components of the equipment.
Electrostatic discharge may damage electronic components. If there is a possibility of ESD damage with a procedure, then do that procedure at an ESD workstation, or wear an antistatic wrist strap.
This equipment must be operated by skilled/trained medical professionals.
Samples, controls, calibrators and waste are potentially infectious. Wear proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling them in the laboratory.
All the analyzer components and surfaces are potentially infectious.
Take proper protective measures for operation or maintenance.
The sample probe tip is sharp and may contain biohazardous materials. Exercise caution to avoid contact with the probe when working around it.
2-1
2 System Structure
2.1 Introduction
Hardware
Hardware collects signals, controls and drives moving parts, and processes and displays
information. It mainly consists of the mother board, drive board, data board, power board, laser
control board, FS pre-amplification board, SS pre-amplification board, volumetric board,
indicator board, liquid-level detecting board, key board and autoloader control board.
Interface
The system mainly consists of two parts, the main unit (analyzer) and the external computer,
which are connected through the network port. Other connections are realized through the
interfaces of the external computer.
Software operating environment
The system software consists of the main unit software and the operation software. The main
unit software is operated on the data board inner the analyzer; the operation software is
operated under the WINDOWS platform of the external computer. The main unit software
analyzes the sequence, collects and identifies data. The operation software displays and prints
the results and stores them into the database, and realizes the interaction of the functions
including count, QC, calibration, maintenance, data management and communication, etc.
Fluidic System
The fluidic system indicates the tubing where reagents, samples and air pass in the analyzer.
The fluidic system controls the correlatively jointed fluidic parts in a set sequence by the
software and driving of the hardware to control the distribution and direction of the medium.
System Structure
2-2
2.2 Fluidic System Refer to Chapter 4 Fluidic System of this manual
System Structure
2-3
2.3 Hardware System Refer to Chapter 5 Hardware System of this manual
System Structure
2-4
2.4 Main Structure The BC-5380 auto hematology analyzer consists of the main unit (analyzer) and the
accessories.
Figure 2-1 Front of the analyzer
1---Front cover 2---Power/Status indicator
3---[OPEN] key 4---[RUN] key
5---Tube 6---Tube rack
7---Tube 8---Adapter
9---Sample compartment door
System Structure
2-5
Figure 2-2 Back of the analyzer
1 --- Network interface 2 --- M-53D diluent sensor connector
3 --- M-53D diluent inlet 4 --- M-53 cleanser inlet
5 --- M-53LH lyse inlet 6 --- M-53LEO(Ⅱ) lyse inlet
7 --- M-53LEO(Ⅰ) lyse inlet 8 --- AC input
9 --- Waste outlet 10 --- Waste sensor connector
System Structure
2-6
Figure 2-3 Left side of the analyzer
1--- Door lock 2---Power switch
System Structure
2-7
Figure 2-4 Right side of the analyzer
1---Door lock 2---Tube
3---Tube rack
System Structure
2-8
Figure 2-5 Front of the analyzer (front cover removed)
1 --- Sampling module 2 --- Tube
3 --- Tube rack 4 --- Sample transport assembly
5 --- Sample compartment assembly 6 --- Sheath fluid syringe assembly
7--- Sample injection syringe assembly 8 --- Sampling syringe assembly
9 --- Fluidic valve 10 --- Relieve valve assembly
System Structure
2-9
Figure 2-6 Left side of the analyzer (left door removed)
1 --- Front cover assembly 2 --- Diluent syringe assembly
3 --- Lyse syringe assembly 4 --- Pressure chamber assembly
5 --- Air pumps 6 --- Left door
7 --- liquid level detection unit 8 --- Valve assembly (2)
9 --- Valve assembly (1) 10 --- Power switch
11 --- Drive board 12 --- Data board
13 --- Mother board 14 --- Top cover
System Structure
2-10
Figure 2-7 Right side of the analyzer (right door removed)
1 --- Optical system 2 --- Isolation chamber
3 --- Sampling assembly 4 --- Fluidic valves
5 --- Fluidic valves 6 --- Volumetric unit
7 --- Vacuum chamber assembly 8 --- Right door
9 --- Waste pump assembly 10 --- Valve assembly (3)
11 --- Bath assembly 12 --- WBC reaction bath assembly
13 --- Pinch valve 14 --- Autoloader control board
15 ---Mix mechanism
System Structure
2-11
2.5 Software structure The software system consists of the main unit software which operates on the data board inner
the analyzer and the operation software which operates under the WINDOWS platform of the
external computer. The main unit software analyzes sequence, collects data and calculates the
parameters. The operation software realizes the interaction of the functions including count,
QC, calibration, maintenance, data management and parameter setup, etc.
2.5.1 Menu structure 1. Shortcut button area
When switching to any screen, a guidance bar will always be displayed on the left of the
screen containing 10 frequently used shortcut buttons. Click them to perform the
corresponding operations. The menu structure of the shortcut buttons is shown in Figure 2-8.
STAT
RunAny screen
Diluent
Worklist
Graph
QC
Table
Logout
Shutdown
Exit
Figure 2-8 Menu structure of the shortcut button area
2. “Run” message box
The menu structure of the “Run” message box is shown in Figure 2-9.
System Structure
2-12
RUN
CT-PD
AL-WB
CT-WB
Start
Ok
Cancel
Start
Ok
Cancel
Start
Ok
Cancel
Figure 2-9 Menu structure of the “Run” message box
3. “Worklist” screen
At the “Worklist” screen, you can click the function buttons to perform the corresponding
worklist-related operations.
The menu structure of the “Worklist” screen is shown in Figure 2-10.
System Structure
2-13
Figure 2-10 Menu structure of the worklist screen
4. “Review” screen
The “Review” menu has five submenus: Graph, Table, Data Backup, Data Export and
Compare. The Data Backup and Data Export functions can be enabled only when the analyzer
has just been started up or you are at the “Table” review screen; otherwise, they grey out. The
“Table” review includes the functions of auto-backup and auto-restore, which will be performed
automatically by the software when the condition of backup or restoring is fulfilled. The menu
structures of each screen of the “Review” screen are shown below.
Graph
The menu structure of the graph review screen is shown in Figure 2-11.
System Structure
2-14
Figure 2-11 Menu structure of the graph review screen
Table
The menu structure of the table review screen is shown in Figure 2-12.
TABLE
Result
Sample List
Microscopic Exam. and
Others
Optical
Edit Result
Restore Result
Store
Validate
Batch Validate
Search
Transmit
Pos./Total
Others Data Backup
Auto-backup
Auto-restore
Data Export
Delete
Deselect
Trend Graph
CV
Samples within Today
All Samples
Samples Found
Sample/Patient Info.
Figure 2-12 Menu structure of the table review screen
Data Export
The menu structure of the “Data Export” screen is shown in Figure 2-13.
System Structure
2-15
Figure 2-13 Menu structure of the data export screen
Data Backup
The menu structure of the “Data Backup” screen is shown in Figure 2-14.
Figure 2-14 Menu structure of the data backup screen
Compare
The menu structure of the “Compare” screen is shown in Figure 2-15.
System Structure
2-16
COMPARE Comparison Summary
Result Trend
Re-fill
Search
Re-fill
Search
Adjust Parameter Order
Adjust Parameter Order
Figure 2-15 Menu structure of the compare screen
5. QC screen
The QC programs include the L-J QC, X-B QC, X QC and X -R QC. Click the shortcut
button “QC” of the guidance bar on the left of the screen to enter the “L-J QC graph” screen.
Click “Menu” → “QC”, then the submenus of four QC programs will pop up for you to select.
L-J
The menu structure of the “L-J” QC program is shown in Figure 2-16.
System Structure
2-17
Figure 2-16 Menu structure of the “L-J” QC program
System Structure
2-18
X-B
The menu structure of the “X-B” QC program is shown in Figure 2-17.
X-B Settings
Graph
Have Preset Values
List
Restore Defaults
Set Limits
Delete
Calculate Preset ValuesSave Preset
Values
Pos./Total
Delete
Pos./Total
Save
Communication
Data Backup
Data Export
History
Figure 2-17 Menu structure of the “X-B” QC program
X
The menu structure of the “ X ” QC program is shown in Figure 2-18.
System Structure
2-19
X
Figure 2-18 Menu structure of the “ X ” QC program
System Structure
2-20
X -R
The menu structure of the “ X -R” QC program is shown in Figure 2-19.
-R
Run
Settings
Graph
Table
Start
Pos./Total
Delete
New Vial
Data Compare
Display order
Outliers
Pos./Total
Delete
Pos./Total
Save
Communication
Data Backup
Data Export
History
X
Figure 2-19 Menu structure of the “ X -R” QC program
6. ”Service” screen
The “Service” screen contains screens of “Maintenance”, “Status”, “Self-test”, “Debug”,
“Log”, “Counter” and “Version and Config. Information”. The menu structures of these
System Structure
2-21
screens are shown below.
“Maintenance” screen
The menu structure of the “Maintenance” screen is shown in Figure 2-20.
System Structure
2-22
Figure 2-20 Menu structure of the “Maintenance” screen
System Structure
2-23
“Status” screen
The menu structure of the “Status” screen is shown in Figure 2-21.
Status
Signal Collection
Export
Sensor
Temperature&Pressure
Voltage&Current
Export
Export
Export
Figure 2-21 Menu structure of the “Status” screen
“Self-test” screen
The menu structure of the “Self-test” screen is shown in Figure 2-22.
System Structure
2-24
Figure 2-22 Menu structure of the “Self-test” screen
System Structure
2-25
“Debug” screen
The menu structure of the “Debug” screen is shown in Figure 2-23.
Figure 2-23 Menu structure of the “Debug” screen
“Counter” screen
The menu structure of the “Counter” screen is shown in Figure 2-24.
Figure 2-24 Menu structure of the “Counter” screen
“Log” screen
The menu structure of the “Log” screen is shown in Figure 2-25
System Structure
2-26
Figure 2-25 Menu structure of the “Log” screen
“Version and Config. Information” screen
The menu structure of the “Version and Config. Information” screen is shown in Figure 2-26.
Figure 2-26 Menu structure of the “Version and Config. Information” screen
System Structure
2-27
7. “Setup” screen
The “Setup” screen contains screens of “Gain”, “Data Format”, “Reagent”, “Auxiliary”, “Para. Unit”, “Ref. Range”, “Print”, “Auto Maintenance”, “User and Password” and “Advanced”.
The menu structure of the “Setup” screen is shown in Figure 2-27.
System Structure
2-28
Para. Unit
Auxiliary
Reagent
Ref. Range
Communication
SETUP
RUO
Gain
User and Password
Maintenance
Date Format
Delete
Edit
New
Reset Password
Change Password
Close
Autoloader
Auto Maintenance
Barcode Info.
Lab Info.
Advanced
Cancel
Ok
Apply
Ok
Cancel
Figure 2-27 Menu structure of the “Setup” screen
8. “Statistics” screen
System Structure
2-29
The menu structure of the “Statistics” screen is shown in Figure 2-28.
Calculate Workload STATISTICS Statistical
Condition
Statistical Item
Workload Summary
Calculate Charge
Re-fill
Statistics
Adjust Order
Statistical Condition
Charge Summary
Re-fill
Statistics
Adjust Order
Figure 2-28 Menu structure of the “Statistics” screen
9. “Shortcut Code” screen
The menu structure of the “Shortcut Code” screen is shown in Figure 2-29.
Figure 2-29 Menu structure of the “Shortcut Code” screen
10. “Calibration” screen
The menu structure of the “Calibration” screen is shown in Figure 2-30.
System Structure
2-30
Figure 2-30 Menu structure of the “Calibration” screen
11. “Help” info.
The menu structure of the “Help” screen is shown in Figure 2-31.
ContentsHELP Service
Review
Daily Operations
Setup
Calibration
QC
Search
Figure 2-31 Menu structure of the “Help” screen
12. Shutdown
The menu structure of “Shutdown” is shown in Figure 2-32.
System Structure
2-31
SHUTDOWN Sleep
Shutdown
Figure 2-32 Menu structure of “Shutdown”
13. Exit
The menu structure of “Exit” is shown in Figure 2-33.
Figure 2-33 Menu structure of “EXIT”
2.5.2 Passwords The passwords are divided into three levels: common user, administrator (set by the user at
the “User and Password” screen) and service engineer (user name: “service”, and password:
“Se s700”. Note: there is a space between “e” and ”s”). An administrator is enabled to have all
the authorities of a common user, and a service engineer is enabled to have all the authorities
of an administrator. Tables below introduce the authorities enabled for each level of user.
Table 2-1 Authorities enabled for the common user level
Level Module Screen Authorities enabled
Common
User
Worklist Worklist Users of three levels are enabled with all
the authorities of the worklist screen, which
including: “New”, “Save”, “Insert”, “Delete”, “Search”, “Hide”,
“Select(Deselect)”, “Browse”, “Copy” and
“Print”.
System Structure
2-32
Count All screens
except the “QC
Run” and
“Service”
(Maintenance,
Status,
Self-test and
Debug); as
well as the
“Run” screen
Users of three levels are all enabled with
the authority of running samples.
Graph Data/Graph, Saving of DIFF, Print,
Pos./Total, Lock, Data Browse (the buttons
of “Validate”, “Edit Result” and “Restore Result” are available, but when you click
them, a message box will pop up for
authority identification.)
Saving of the “Microscopic Exam. and Others” screen, Print, Pos./Total, Lock,
Data Browsing and Setup of blood ESR ref.
range.
TABLE Save, Print, Communication, Deselect,
Search, Trend Graph (cannot be printed),
CV (cannot be printed), arrow buttons
above the “Pos./Total” control (the buttons
of “Validate”, “Batch Validate”, “Edit Result” and “Restore Result” are
available, but when you click them, a
message box will pop up for authority
identification.)
Data Export Data Export
Data Backup Data Backup
Compare Search and re-fill patient information,
adjust and print parameters in the
comparison summary, adjust and print
parameters in the result trend.
Auto-backup Auto-backup
Review
Auto-restore Auto-restore
System Structure
2-33
L-J L-J QC setup: browse file information of
different lot No.(but can not edit the QC
settings), Print
L-J QC count: Start QC Run, Print and
Pos./Total
L-J QC graph: New Vial, Data Compare,
Display Order, Calculate Preset Values,
Save Preset Values, Outliers, Print and
Pos./Total
L-J QC table: Print, Communication, Data
Backup, Data Export, History and
Pos./Total
X-B X-B QC setup: browse the information of
the screen but no edit allowed, Print
X-B QC graph: Calculate Preset Values,
Save Preset Values, Print and Pos./Total
X-B QC table: Print, Communication, Data
Backup, Data Export, History and
Pos./Total
QC
X X QC setup: browse file information of
different lot No.(but can not edit the QC
settings), Print
X QC count: Start QC Run, Print and
Pos./Total
X QC graph: New Vial, Data Compare,
Display Order, Calculate Preset Values,
Save Preset Values, Outliers, Print and
Pos./Total
X QC table: Print, Communication, Data
Backup, Data Export, History and
Pos./Total
System Structure
2-34
X -R X -R QC setup: browse file information of
different lot No.(but can not edit the QC
settings), Print
X -R QC count: Start QC Run, Print and
Pos./Total
X -R QC graph: New Vial, Data Compare,
Display Order, Outliers, Print and
Pos./Total
X -R QC table: Print, Communication,
Data Backup, Data Export, History and
Pos./Total
Maintenance Replace Reagent (“All Reagents”
unavailable), Clean, (“Empty WBC Bath”,
“Empty RBC Bath”, “Empty DIFF Bath”
all unavailable), Maintain the whole device
Status Temperature&Pressure ( “Analyzer Information” unavailable at the “Print” message box), Voltage&Current
(“Analyzer Information” unavailable at the
“Print” and “Export” message boxes)
Self-test Self-test for syringes and sampling
mechanism, valves and others
Counter Check and print all the information
Log Check, print and see details of logs
including “Set Paras”, “Other Logs” and
“All Logs”
Service
Version and
Config.
Information
Check, export and print the information of
the “Version and Config. Information”
screen
General
Format
Date Format, Reagent, Auxiliary
(“Authority” unavailable)
Setup
User and
Password
1. Check the information of the “User list” screen, but New, Edit, Delete and Reset
password are disabled.
2. When log in as a common user, the
password of the current login user can be
modified.
System Structure
2-35
Lab
Information
Check the Lab information, but not allowed
to edit
Calculate
Workload
Select statistical items, set statistical
conditions, adjust and print workload
summaries
Statistics
Calculate
Charge
Set calculate conditions, adjust and print
charge summaries
Department Check
Deliverer Check
Diagnosis Check
Shortcut Code
Gender Check
Calibration Calibration
Factor Browse and print user calibration factors
Help Help Check the help information
Shutdown Shutdown
Perform the Sleep and Shutdown
procedures
Exit Exit Log out and exit the system
Table 2-2 Authorities enabled for the administrator level
Level Module Screen Authorities enabled
Worklist Worklist
Run All screens
except the “QC Run” ,
“Service”
(Maintenance,
Status,
Self-test and
Debug) and
“Calibration”
Graph Delete, Histogram Adjust., Validate, Edit
Result and Restore Result
Table Validate, Batch Validate, Edit Result and
Delete
Administrator
Review
Data Export
System Structure
2-36
Data Backup
Compare
Auto-backup
Auto-restore
L-J L-J QC setup: Edit QC files, save and set
limits, get preset values
L-J QC count
L-J QC graph: Delete
L-J QC table: Delete
X-B X-B QC setup: Edit QC files, get preset
values, restore defaults, save
X-B QC graph: Delete
X-B QC table: Delete
X X QC setup: Edit QC files, save and set
limits, get preset values
X QC count
X QC graph: Delete
X QC table: Delete
QC
X -R X -R QC setup: edit QC files, save
X -R QC count
X -R QC graph: Delete
X -R QC table: Delete
Maintenance
Status Temperature&Pressure (“Analyzer Information” available at the “Print” message box), Voltage&Current
(“Analyzer Information” available at the
“Print” message box)
Self-test
Counter
Log Check, print and see the details of the
“Error Info.”
Service
Version and
Config.
Information
System Structure
2-37
General Setup
“Authority” setting items of the “Auxiliary”
screen
Check and set the parameters of screens
including Para. Unit, Ref. Range, Print,
Communication, RUO, Auto Maintenance,
Autoloader and Barcode Info.
User and
Password
New, Edit, delete user, reset the passwords
of all users and change the administrator’s
own password
Lab
Information Check and edit the Lab Information
Setup
Shortcut Code New, Edit and Delete the shortcut code
Calculate
Workload
Statistics
Calculate
Charge
Department New, Edit and Delete
Deliverer New, Edit and Delete
Diagnosis New, Edit and Delete
Shortcut Code
Gender New, Edit and Delete
Calibration
Factors
Save and restore the User Calibration
Factors (all the User Calibration Factors)
Calibrator
1. Calibration in the CT-CBC+DIFF mode
2. Calibrate one or more parameters
among WBC, RBC, HGB, MCV and PLT.
Calibration
Fresh Blood
1. Fresh Blood calibration in the CT-WB
and CT-PD Mode;
2. Calibrate one or more parameters
among WBC, RBC, HGB, MCV and PLT.
Help Help
Shutdown Shutdown
Exit Exit
Table 2-3 Authorities enabled for the service engineer level
Level Module Screen Authorities enabled
Service Worklist Worklist
System Structure
2-38
Run All screens
except the “QC Run” ,
“Service”
(Maintenance,
Status,
Self-test and
Debug) and
“Calibration”
Graph Special Info.
Table Trend graph (can be printed), CV (can be
printed), Optical
Data Export
Data Backup
Compare
Auto-backup
Review
Auto-restore
L-J
X-B
X
QC
X -R
Maintenance 1. Replace All Reagents
2. Empty WBC Bath, Empty RBC Bath and
Empty DIFF Bath at the “Maintenance”
screen
Status Temperature&Pressure (“Analyzer Information” available at the “Export” message box), Voltage&Current
(“Analyzer Information” available at the
“Export” message box), Sensor, Signal
collection and Export the Analyzer
Information
Self-test
Engineer
Service
Counter Initialize the number of times that each
item has been calculated except
the ”Sample Count Times” (samples in
QC, calibration, background counts and
reliability tests not included)
System Structure
2-39
Log Export, Save, Print and see Details of “Set Paras”, “Other Logs”, “All Logs”, “Error Info.” and “RunSequence”
Debug Debug
Version and
Config.
Information
General Setup
Edit FS, SS and SF values of the “Gain”
screen; set the value for “Auto prompt when n samples were run” at the “Auto Maintenance” screen
User and
Password
Lab
Information
Shortcut Code
Setup
Advanced Maintain
Calculate
Workload
Statistics
Calculate
Charge
Department
Deliverer
Diagnosis
Shortcut Code
Gender
Calibration
Factors
Import and Export User Calibration Factors
and Factory Calibration Factors, Restore
Defaults, Save, Restore (All the Factory
Calibration Factors and Factory Transfer
Factors), Print
Calibration
Calibrator
1. Factory calibrator calibration in the
WB-CBC, WB-CBC+DIFF, PD-CBC and
PD-CBC+DIFF mode.
2. Calibrate one or more parameters
among WBC, RBC, HGB, MCV, PLT and
MPV.
Help Help
System Structure
2-40
Shutdown Shutdown
Exit Exit
3-1
3 Instrument Installation and Software Upgrade
3.1 Preparations
3.1.1 Purpose Install the instrument and update the software properly as per the procedures introduced in this
chapter.
3.1.2 Tools Blade or clipper
USB flash drive
Pipette (200µl)
Cross-headed screwdriver
Thermometer with a probe
The software copied in the U-drive must be the specified one for the analyzer of the specified model. Otherwise, you can not proceed to install.
3.1.3 Accessories Plastic test tube
Glass Pipette
7µm standard particles
Gloves
Tissues
Controls (high and normal level)
Do prepare the 7µm standard particles firstly to proceed with the later confirmation work of the optical gain.
Instrument Installation and Software Upgrade
3-2
3.2 Installation Requirements 3.2.1 Space Requirements Check the site for proper space allocation. In addition to the space required for the analyzer
itself, arrange for:
at least 100 cm on both left and right sides;
at least 50 cm behind;
enough room on or below the countertop to accommodate the diluent, rinse and waste
containers
3.2.2 Power Requirements 1. Make sure the analyzer is properly grounded. Before turning on the analyzer, make sure
the input voltage meets the requirements.
2. Using pinboard may bring the electrical interference and the analysis results may be
unreliable. Please place the analyzer near the electrical outlet to avoid using the
plug-board.
3. Please use the original electrical wire shipped with the analyzer. Using other electrical wire
may damage the analyzer or cause unreliable analysis results.
Voltage: A.C. 100V-240V Input power: ≤300 VA Frequency: 50/60 Hz
3.2.3 Environment Requirements The installation environment of analyzer must meet the following requirements:
1. Optimal operating temperature: 15 ℃ - 30 ℃
2. Optimal operating humidity: 30 % - 85 %
3. Atmospheric pressure: 70 kPa - 106 kPa
4. The environment should be as free as possible from dust, mechanical vibrations, loud
noises, pollution and electrical interference.
5. It is advisable to evaluate the electromagnetic environment prior to the operation of this
analyzer. Make sure the electromagnetic interference is less than CLASS B. Do not use
this analyzer in close proximity to sources of strong electromagnetic radiation.
6. Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close.
7. Do not place the analyzer on a slope.
8. The environment should have good ventilation. Do not place the analyzer in direct sunlight
or in front of a source of heat or drafts.
Instrument Installation and Software Upgrade
3-3
Make sure the installation environment meets the above 8 requirements. Otherwise, the performance of the analyzer might be affected.
3.2.4 PC Configuration Requirements 1. Startup the computer, then right-click “My Computer” and select “Properties” to enter the
property screen shown in Figure 3-2. You can check the operation system and RAM of the
computer here. They should meet the following requirements:
Operation system : Windows 2000 Professional + SP4, Windows XP Home/XP
Professional + SP2 or Windows Vista.
RAM:256MB at least
Figure 3-1 Checking “Properties” of “My computer”
Instrument Installation and Software Upgrade
3-4
Figure 3-2 System properties
2. Right-click the desktop and select “Properties”, and then click the “Settings” tag to enter
the screen shown in Figure 3-4. You can check the resolution of the display here. It should
meet the following requirements:
Resolution:1024 x 768 or higher (adjust as shown in Figure 3-4);
Figure 3-3 Checking “Properties” of desktop
Operating system
RAM
Instrument Installation and Software Upgrade
3-5
Figure 3-4 Setting screen resolution
3. Double-click “My Computer” to enter the screen shown in Figure 3-5. You can check the
space available on each disk here. It should meet the following requirements:
Hard disk space:4GB available on one disk at least.
Figure 3-5 My computer
Then, close these 3 dialog boxes.
Disk space
Screen resolution
Instrument Installation and Software Upgrade
3-6
The PC hardware configuration and operation system must meet the foregoing requirements. Otherwise, the analyzer can not work properly.
If the operating system on the external PC is Windows 2000 Professional +SP4, the software for the analyzer can only be installed successfully after installing the Windows2000-KB835732 patch of the corresponding language.
To install the software on a PC without CD-ROM, you should copy the installation files from the installation CD to a USB flash drive on a PC with a CD-ROM, and then copy the files to the PC for installation through U drive. To install the software on a PC with a CD-ROM, install or update the software using the CD directly. To connect to the LIS, the PC should be configured with two network cards.
Instrument Installation and Software Upgrade
3-7
3.3 Package Checking and Unpacking
3.3.1 Checks before unpacking Please check if the package is intact before unpacking.
3.3.2 Unpacking the main unit 1. The appearance of the main unit is shown in Figure 3-6. Cut off the binding belt before
unpacking.
Figure 3-6 External package of main unit
2. Cut off the binding belt and remove the wooden cover. Then, remove the accessory box
shown in the figure below.
Figure3-7 Remove wooden cover
3. Lift the carton to expose the main unit.
Instrument Installation and Software Upgrade
3-8
Figure3-8 Remove accessory box
4. Remove the protection foam from both sides of the main unit.
Figure3-9 Remove carton
5. Remove the plastic bag from the main unit. Grab the bottom of the main unit and lift it onto a
countertop. Note that the main unit must be lifted by at least two persons.
Figure 3-10 Lift main unit
Instrument Installation and Software Upgrade
3-9
When lifting the main unit, keep it as level as possible and avoid strong impact.
3.3.3 Checking packing list Check the delivered components against the packing list to see if everything is delivered.
Instrument Installation and Software Upgrade
3-10
3.4 Removal and Installation 1. Open the right door of the main unit, remove the plastic cable tie vertically fixing the
sampling assembly and remove the clamp fixing the synchronous belt, as Figure 3-11
shows.
Figure 3-11 Remove the plastic cable tie and the clamp
2. When using the tubes of Ф14×75(mm)or Ф15×75(mm)model, you should disassembly
the adapter from the rack to load the tubes. Disassembly as shown in Figure 3-12.
Figure 3-12 Remove the adapter from the tube rack
3. To install the software on a PC without CD-ROM, the installation files should be copied
from the installation CD to a USB flash drive on a PC with a CD-ROM, and then plug the
flash drive to the USB interface of the PC for installation and open the folder with
Instrument Installation and Software Upgrade
3-11
installation files. To install the software on a PC with a CD-ROM, put the installation CD
directly into the CD-ROM.and open the folder with instaltion files. Copy the installation files
from the disk to the U drive. Then, plug the U drive to the USB interface and then open the
installation folder.
4. Check if there is any software of other hematology analyzer on the PC. If there is, uninstall
the software or database before installation, and then double-click “Setup.exe” to install
the software. Select the language from the pull-down list in the pop-up message box, and
then click “OK” to continue installing, as shown in Figure 3-13.
Figure 3-13 Language selecting
5. An installation directory box will pop up with the default directory: D:\Mindray\Auto
Hematology Analyzer. You can click “Browse” to change the directory if necessary. Then,
click “Next” to continue installing as shown in Figure 3-14.
Figure 3-14 Selecting directory
6. A progress bar shown in Figure 3-15 will pop up. Please wait while the software is being
installed.
Instrument Installation and Software Upgrade
3-12
Figure 3-15 Progress bar
7. When the installation of the software is finished, a message box shown in Figure 3-16 will
pop up. Click “Next” to install the necessary components and database.
Figure 3-16 Continuing installing
8. The message boxes shown in Figure 3-17 will pop up while installing, please click the
“Accept” to continue installing.
Instrument Installation and Software Upgrade
3-13
Figure 3-17 Installing .NET Framework 2.0
9. The message boxes shown in Figure 3-18 will pop up while installing, please click the
“Accept” button to continue installing.
Instrument Installation and Software Upgrade
3-14
Figure 3-18 Installing Windows Installer 3.1
10. Then, the message boxes shown in Figure 3-19 will pop up while installing, please click
the “Accept” to continue installing.
Instrument Installation and Software Upgrade
3-15
Figure 3-19 Installing SQL
11. Then, the message boxes shown in Figure 3-20 will pop up while installing, please click
the “Install” button to continue installing.
Figure 3-20 Installing VC++ R L
12. A progress bar “Installing XXX…” shown in Figure 3-21 will pop up while installing the
components. Please be patient and wait as it may take longer time.
Instrument Installation and Software Upgrade
3-16
Figure 3-21 Progress bar
13. After part of the components is installed, a message box shown in Figure 3-22 may pop up.
You should click “Yes” to restart the computer and the installation program may continue
automatically after restarting.
Figure 3-22 Restarting the PC
14. When finish installing the components, the database will be installed automatically as
shown in Figure 3-23 .
Figure 3-23 Installing database
Instrument Installation and Software Upgrade
3-17
15. When the installation is complete, a message box shown in Figure 3-24 will pop up. Click
“Close” to close the box and the entire software installation procedure is finished.
Figure 3-24 Installation complete
If it is found that the main unit software mismatches the PC software, you should upgrade the main unit software as per the procedures specified in 3.11.2 to match it with the software of the PC-end.
16. Install the driver program of the printer properly on the PC.
17. The BC-5380 consists of the IPU PC and the main unit as designed. Only basic interactive
keys like the [RUN] and [OPEN] keys are located on the main unit. The rest of the
operations are realized through the IPU PC. The IPU PC has two network interfaces, one
connecting the BC-5380 and the other one may connect to other information devices (e.g.
LIS system) or the LAN of the hospital, which are used to export data. There are two icons
of “Local Area Connection” for the two network interfaces existed. follow the procedures to
configure the network interfaces of the IPU PC:
(1) Right click “My Network Places” and select “Properties” from the pop-up dialog box.
Instrument Installation and Software Upgrade
3-18
Figure 3-25 Right clicking “Network neighbor”
(2) Right click one of the network interfaces and select “Properties”. In the pop-up message
box, select the check box of “Show icon in notification area when connected” and “Notify me when this connection has limited or no connectivity”, and then click “OK”. Repeat this
procedure to set the other network interface.
Figure 3-26 Connection properties
(3) At the “Local Area Connection Properties” screen, select “Internet Protocol (TCP/IP)” and then click “Properties”. At the pop-up screen, select “Use the following IP address”, and
then enter 10.0.0.1 into the “IP address” box and 255.0.0.0 into the “Subnet mask” box, as
shown in Figure 3-27 .
Instrument Installation and Software Upgrade
3-19
Figure 3-27 Setting IP address
(4) Then, click “OK” to save all the settings.
Never set the “Default gateway” and DNS server for the network interface (10.0.0.1) connecting the BC-5380. Otherwise, you may not access to the network connecting the other network interface.
If other computers or network are connected to the IPU PC (through the other network interface), be sure that the IP address 10.0.0.1 (used by the IPU PC) or 10.0.0.2 (used by the BC-5380) is not applied to other computers or computers in the other network.
18. The LIS computer is used to transmit data by connecting the IPU computer. Do as follows
to check and set the IP for the LIS computer.
(1) At the desktop of the LIS computer, right click “My Network Places” and select
“Properties” from the pop-up dialog box, as shown in Figure 3-28 .
Instrument Installation and Software Upgrade
3-20
Figure 3-28 “Network Neighbor” (LIS computer)
(2) Right click the “Local Area Connection” icon and select “Properties”, as shown in Figure
3-29 .
Figure 3-29 Connection properties
Then, the following box will pop up:
Instrument Installation and Software Upgrade
3-21
Figure 3-30 Connection properties (LIS computer)
(3) At the “Local Area Connection Properties” screen, double click “Internet Protocol (TCP/IP)” to enter the following screen.
Instrument Installation and Software Upgrade
3-22
Figure 3-31 Setting IP address
(4) Check and record the IP address shown in the screen. It will be used to set the IP
information on the BC-5380.
(5) If the IP address is not displayed here, you should set it manually (do not set the gateway
and DNS server). After setting the IP address, the subnet mask will be generated automatically.
Then, click “OK” to save the settings.
The IP address you set here can not be 10.0.0.1 or 10.0.0.2.
Instrument Installation and Software Upgrade
3-23
3.5 Connections
1. See Figure 3-32 for connections of reagent and waste containers. Note that the
M-53LEO(I) lyse, M-53LEO(II) lyse, M-53LH lyse and the analyzer should be placed in a
plane of the same level, and the M-53D diluent, M-53 cleanser, and waste containers
should be placed under the countertop.
Figure 3-32 Connections of reagent and waste containers
Make sure the M-53LEO (I) lyse, M-53LEO (II) lyse, M-53LH lyse and M-53 cleanser are placed on the same level as the analyzer.
Make sure the tubes are electrically isolated.
See Figure 3-33 for proper connections of reagent and waste containers. Note that the tube
connectors and their counterparts on the analyzer are of the same colors.
Instrument Installation and Software Upgrade
3-24
Figure 3-33 linking wires and lines
2. Locate the power inlet at the back of the main unit and connect the female end of the
power cable to the power inlet, and connect the three-pronged end of the power cable to a
power outlet.
Figure 3-34 Connection of PC and power supply
3. Connect the BC-5380 with the IPU computer by a cross network cable, as shown in Figure
3-35 .
Instrument Installation and Software Upgrade
3-25
Figure 3-35 Connecting the main unit and the IPU computer
Be sure to use the cross network cable to connect the IPU computer and the LIS computer.
The two ends of a cross network cable have no difference. You can plug them to IPU computer and the LIS computer respectively at will. Please see appendix H for method of how to distinguish a cross network cable from a direct-connected network cable.
4. There are two ways to connect the BC-5380 and the LIS computer.
(1) Connect the BC-5380 with the LIS computer by a cross network cable, as shown in Figure
3-36 .
Figure 3-36 Connecting the IPU computer and the LIS computer
Be sure to use the cross network cable to connect the analyzer and the LIS computer.
The two ends of a cross network cable can be randomly plugged into the two computers respectively.
(2) Connect the IPU computer to the LAN of the hospital, as shown in Figure 3-37 .
Instrument Installation and Software Upgrade
3-26
Figure 3-37 Connecting the IPU computer and LAN
Be sure to use the direct-connected network cable to connect the IPU computer to the LAN.
The two ends of a direct-connected network cable can be randomly plugged into the IPU computer and the network devices (HUB, Exchanger, Router, etc.) respectively.
Adopt one of the two ways to finish the network connection according to the condition of the
hospital or doctor’s requirement.
5. Connect the printer with the IPU computer with specified cable and connect the power
cord properly.
Instrument Installation and Software Upgrade
3-27
3.6 Start-up
3.6.1 Inspection before Startup Check and make sure the marks on the relieve valve are on the same line, as shown in Figure
3-38. Change the relieve valve if the marks are not on the same line.
Figure 3-38 Marks on relieve valve
1. Check and make sure the sampling tube and the two pipes connected the probe wipe are
not bended and clogged at the place of the flatten cable clamp (at top of the sampling
assembly), as shown in Figure 3-39.
Figure 3-39 Sampling and probe wipe tubing
2. Before starting up, remove the right door of the analyzer, and then check and make sure
the tubes connecting the V25, V26 and the tee-connector of the volumetric tube are not
bended and clogged, as Figure 3-40 shows.
Marks on relieve valve
Instrument Installation and Software Upgrade
3-28
Figure 3-40 Tubing free from bending and clogging
3. Check and make sure the thick 50 tube connecting the V27 is not bended and clogged, as
Figure 3-41 shows.
Figure 3-41 Tubing free from bending and clogging
4. Check and make sure the two-way connector and the tee connector connecting tube 1 are
in the same distance towards the pinch valve, as shown in Figure 3-42 .
Check and make sure the tubes connecting the V25, V26 and the tee-connector of the volumetric tube are not bended and clogged
Check and make sure the thick 50 tube connecting the V27 is not bended and clogged
Instrument Installation and Software Upgrade
3-29
Figure 3-42 Checking the pinch valve
3.6.2 Start up for the first time
1. Turn on the analyzer, wait a moment and then a prompt will appear
on the task bar. It means the local-connection-related network interface connects the
BC-5380 already. Similarly, if the other prompt appears on the
task bar, it means the local-connection-2-related network interface doesn’t connect the
BC-5380, and it is available to connect the LAN of the hospital. If the two prompts are all
then, you should unplug the network cable connecting the
BC-5380 from the IPU computer, and then plug it into the other network interface; then,
plug the network cable connecting the LIS computer into the rest network interface.
2. Double click the “BC-5380 Hematology Analyzer” icon on the desktop of the IPU
computer to start up the software. Then, enter the user name “service” and the password
“Se s700” (note: there is a space between “e” and ”s”) into the login box, as shown in
Figure 3-43 , and then click “OK” to initialize the analyzer.
Tube 1
Tee connector
Two-way connector
Instrument Installation and Software Upgrade
3-30
Figure 3-43 Logging in
3. During the initialization, a dialog box “Skip fluidic initialization?” will pop up. Click “No” to
continue the initialization.
Figure 3-44 “Skip fluidic initialization?”
4. When initialization is finished, click “Menu” “Setup” “Communication” to enter the
screen shown below. Set the IP address and Port here, and then click “Ok” to save the
settings.
Instrument Installation and Software Upgrade
3-31
Figure 3-45 Communication setup
The IP address you set here should be the one of the LIS computer connecting the IPU computer. The Port you set here should be the interception port of the LIS computer. The IP address of external server cannot be 10.0.0.1 and 10.0.0.2, and format like 10.0.0.* is not recommended.
5. Before transmitting data, make sure the LIS computer connecting the IPU computer is
working. Thus, the IPU program could locate the destination for transmitting.
6. If the connection is established between the IPU computer and the LIS computer, the “LIS”
icon in the IPU screen will turn to colorful, as shown in Figure 3-46, indicating you can start
transmitting data now.
Instrument Installation and Software Upgrade
3-32
Figure 3-46 LIS communication
The data communication will be terminated automatically once the IPU software is closed. The LIS computer is the server terminal and the IPU software is the customer terminal. The IPU software keeps trying to connect after starting up, if it is intercepted correctly by the server terminal, then the connection will be established and ready for communication.
The IPU computer can connect LIS computer through cross network cable; connect LIS computer in the same network through devices like HUBs or exchangers; connect LIS computer in a different network through devices like routers.
3.6.3 Inspection after Startup 1. Click “Menu→ Service→ Version and Config. Information”, as shown in Figure 3-47 to
enter the version and configuration information screen shown in Figure 3-48.
Instrument Installation and Software Upgrade
3-33
Figure 3-47 Menu option
Figure 3-48 Version and config. Info.
2. Check and make sure the following items are the correct and specified version: Boot
Software, Application Software, Operating System, Driver Board FPGA, Driver Board MCU,
Data board FPGA, Fluidics Sequence and Algorithm Library, as shown in Figure 3-48, and
then close the message box.
Instrument Installation and Software Upgrade
3-34
The specified version here means the PC end software should be complied with that of the main unit end. Upgrade the main unit end software to match the PC end software if necessary.
Instrument Installation and Software Upgrade
3-35
3.7 Setup and Adjustment
3.7.1 Setting Parameters 1. At the main screen, click “Menu”→“Setup”→“Reagent” to enter the reagent screen where
you can set the expiration date for the reagents, as shown in Figure 3-49.
Figure 3-49 Clicking “Reagent”
2. Set the expiration date for the following reagents: “Diluent”, “LEO (I) Lyse”, “LEO (II) Lyse”, “LH Lyse” and “Cleanser”.
Instrument Installation and Software Upgrade
3-36
Figure 3-50 Setting the expiration date
3. At the main screen, click “Menu”→“Setup”→“Advanced” to enter the screen shown in
Figure 3-52.
Figure 3-51 Entering “Advance” screen
Instrument Installation and Software Upgrade
3-37
Figure 3-52 “Advance” screen
4. Click “Ambient Temp.” at the “Advanced” screen, as shown in Figure 3-52, and the
“Ambient Temp. Calibration” screen will pop up. Put the thermometer to about 5cm from
the right side of the analyzer (do not touch the analyzer). Read the measured value after 1
minute and enter it in the blank box at the screen. Click “Calculate New Difference” to get
the new deviation, and then click “Ok” to go back to the “Advanced” screen, as shown in
Figure 3-53.
Figure 3-53 “Ambient Temp. Calibration” screen
5. Click “Ambient Temp.” at the “Advanced” screen to go to the “Ambient Temp. Calibration” screen;
6. Read the measured temperature after 1 minute and make sure the deviation of the value
measured by the analyzer from that by the thermometer should be less than ±1℃. If not,
readjust the ambient temperature until the deviation is less than ±1℃.
7. Input the ambient temperature measured by the analyzer and the thermometer, and then
Instrument Installation and Software Upgrade
3-38
click “Ok” at the “Ambient Temp. Calibration” screen to go to the “Advanced” screen.
Click “Ok” to close the “Advanced” screen.
3.7.2 Adjusting Position 1. Press the [RUN] key. Then, the analyzer will run a background count once.
2. At the graph review screen, click the “Special info.” button as shown in Figure 3-54 to
enter the special information screen shown in Figure 3-55.
Figure 3-54 Clicking “Special info.”
Figure 3-55 “Special info.” Screen
3. Firstly, make sure the DIFF pulse data length is less than 200. Otherwise, do the
background count till the DIFF pulse data length is less than 200, as shown in Figure 3-55.
Instrument Installation and Software Upgrade
3-39
Use the cross-headed screw driver to remove the shielding cover of the volumetric board.
Make sure the WBC Start signal delay is within [2.5, 3] and the RBC Start signal delay is
within [4, 5.5]. If they are found out of the limit, fine adjust the distance between the tee
connector under the volumetric tube and the lower fixing plate of the volumetric board
according to the WBC and RBC Start signal delay, as shown in Figure 3-56. When WBC
start signal delay is less than 2.5s, fine adjust the WBC volumetric tube downwards; When
WBC start signal delay is more than 3s, fine adjust the WBC volumetric tube upwards;
When RBC start signal delay is less than 4s, fine adjust the RBC volumetric tube
downwards; When RBC start signal delay is more than 5.5s, stop adjusting the RBC
volumetric tube upwards. When the WBC and RBC Start signal delay are within the limit,
do the background count three times consecutively to check and make sure the two values
are within the range every time, and then record the values of the last time. Meanwhile,
check the WBC and RBC count time and calculate the mean time. Then, close the special
info. message box.
Be sure to adjust the position of the volumetric tube by your hand directly. Never pull the tee connector hard to adjust, otherwise, the tee connector may leak and the volumetric tube may be broken.
Figure 3-56 Distance under the volumetric tube
4. At the main screen, click “Menu”→“Setup”→“Advanced” to enter the screen shown in
Figure 3-58.
Instrument Installation and Software Upgrade
3-40
Figure 3-57 Entering “Advance” screen
Figure 3-58 “Advance” screen
5. Confirm whether the WBC and RBC count mean time that you recorded previously
complies with the time displayed in the screen, as shown in Figure 3-58. Enter the
recorded mean time if the time displayed doesn’t comply with the recorded one.
Instrument Installation and Software Upgrade
3-41
3.8 Testing Other Functions Click “Menu”→“Setup”→“Print” to enter the print screen. You can set the connection between
the IPU computer and the printer here, as shown in Figure 3-59.
Figure 3-59 Print Setup
Select printer
Click “OK”
Instrument Installation and Software Upgrade
3-42
3.9 Confirming Gain 3.9.1 HGB blank voltage After the transportation and initialization, the HGB blank voltage required may differ from the
factory settings.
Thus, confirming and adjusting the HGB blank voltage through gain setup is necessary.
1. Entering the setup screen
At the main screen, click “Menu→ Setup→Gain” to enter the screen.
Figure 3-60 Setup screen
2. Confirming the HGB gain
You can adjust the HGB blank voltage by adjusting the HGB gain. The HGB blank voltage shall
be set to 4.48-4.52 V, 4.50V is favorable. Do as follows to adjust:
At the “Gain” screen, click the adjusting button of the “HGB current value” to set the HGB
blank voltage to 4.48-4.52 V, 4.50V is favorable.
3.9.2 Gain of Optical System After the transportation, the gain of the DIFF channel of the optical system may differ from the
factory settings.
Thus, adjusting the gain of the optical system is necessary. Do as follows to adjust:
1. At the “Graph” screen, select the work mode as “WB-CBC+DIFF” and then perform the
background count once and make sure the background results meet the specified
HGB gain
HGB blank voltage
Instrument Installation and Software Upgrade
3-43
requirement.
2. At the “Table” review screen, select the background record run previously, and then click
“Optical” to enter the screen shown in Figure 3-62 .
Figure 3-61 Setup screen
Figure 3-62 Setup screen
3. Click “Calculate”, and then make sure the calculated FS and SS 0.1max total (the larger
Click “Calculate”
SS 0.1max total
FS 0.1max total
Click “OK”
Instrument Installation and Software Upgrade
3-44
number) is less than 50. Otherwise, you should clean the flow cell and DIFF bath at the
“Maintenance” screen. Then, repeat steps 1 to 3 until the number of FS and SS 0.1max
particles are less than 50. Then, click “Ok” to back to the main screen, as shown in Figure
3-62.
4. Run the calibrator on the analyzer. Select the test results from the “Graph” review screen,
and click “Optical” to go to the “Optical” screen. Click “Calculate”, and then check if the
gravity center position of FS and SS meet the following requirements (choose the one
have more particles within the FS 0.1max range): the FS gravity center should locate
within ±0.57 of the low angle target peak; the SS gravity center should locate within ±0.63
of the high angle target peak
5. If the position of FS and SS gravity centers do not meet the requirements, input the FS and
SS gain target values of the corresponding particle at the “Optical” screen, and then get
the gain values, as shown in Figure 3-63
Figure 3-63 “Optical” screen
6. Click “Ok” to close the “Optical” screen, and then click “Menu→ Setup→Gain”. Confirm
the values in the FS and SS boxes are the same with the FS and SS gain values. If not,
input the recorded FS and SS gain values as current FS and SS values, and then click
“Ok” to exit the screen, as shown in Figure 3-64
Instrument Installation and Software Upgrade
3-45
Figure 3-64 Setting optical gains
7. At the “Graph” review screen, select the work mode as “WB-CBC+DIFF”. Then, prepare
the standard particle sample at the ratio that dispenses 3 drops of standard particles into
1ml distilled water. Then, well mix the sample and run the sample once.
Be sure to use the distilled water to prepare the standard particle sample. Prepared standard particle sample should be run within 12 hours.
8. When finish running, enter the “Table” screen and select the standard particle sample just
run, and then click the “Optical” button, as shown in Figure 3-65.
Instrument Installation and Software Upgrade
3-46
Figure 3-65 “Table” review screen
9. At the “Optical” screen, click the “Calculate” button. Check if the following requirements
are met: FS gravity center within (16.68, 19.42); SS gravity center within (100.56, 107.92);
SS 0.1max width ≤15.21, and then click “OK” to exit the optical screen.
10. If the gravity center and the 0.1max width of the standard particle do not meet the
requirements, fine tune the FS and SS gain values at the “Gain” screen. Run the calibrator
and the 7um standard particles, until the following requirements are met: for the calibrator,
the gravity center of the low angle should be within ±1.51; the gravity center of the high
angle should be within ±2.4 from the target values; for the 7um standard particles, the
gravity center of low angle should be 18.05±1.37 (16.68, 19.42); gravity center of high
angle should be 104.24±3.68 (100.56, 107.92); low angle width ≤6.50, high angle width of
7um standard particles ≤15.21. If these requirements are met, name and save the values.
Instrument Installation and Software Upgrade
3-47
3.10 Calibration and Performance Test 3.10.1 Background Count Back to the “Graph” screen and perform the background count in “WB-CBC+DIFF” mode. The
background results shall meet the specified requirement: WBC:≤0.1×109/L, RBC:≤0.03×1012/L,
HGB:≤1g/L, HCT:≤0.5%, PLT:≤5×109/L.
3.10.2 Carryover Follow the procedure 3 instructed in section 9.3.1 of the Operator’s Manual to make sure the
carryover of the analyzer meets the requirements listed in the table below:
Parameter Carryover
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.6%
HCT ≤0.5%
PLT ≤1.0%
3.10.3 Reproducibility In the “AL-WB-CBC+DIFF” mode, “AL-WB-CBC” mode, “CT-PD-CBC+DIFF” mode and
“CT-PD-CBC” mode, make sure the reproducibility of the analyzer meets the requirements
specified in Appendix B 5.4 of the Operator’s Manual
The CV requirements for whole blood mode:
Para. Condition CV% / absolute deviation d (WB)
WBC (6.0~15.0)×109/L ≤2.0%
RBC (4.00~6.00)×1012/L ≤1.5%
HGB (110~180)g/L ≤1.5%
MCV (80~110)fL ≤1.0%
PLT (150~500)×109/L ≤4.0%
The CV requirements for predilute mode:
Para. Condition CV% / absolute deviation d (PD)
WBC (6.0~15.0)×109/L ≤4.0%
RBC (4.00~6.00)×1012/L ≤3.0%
HGB (110~180)g/L ≤3.0%
MCV (80~110)fL ≤2.0%
PLT (150~500)×109/L ≤8.0%
Instrument Installation and Software Upgrade
3-48
3.10.4 Calibration 1. Click “Menu→Calibration”, and then click “Calibrator” to enter the screen. Select the
calibration mode as “Whole Blood”; enter the lot No. and the expiration date of the
calibrator, and then enter the WBC, RBC, HGB, MCV and PLT targets of the calibrator into
the table.
Figure 3-66 Calibration (WB)
2. Well mix the calibrator and then open the cap. Press the [OPEN] key to open the sample
compartment and put the calibrator into the compartment. Close the compartment door
manually, and then press the [RUN] key, as Figure 3-67 shows. When you hear the beep,
the analyzer is starting to run automatically.
Instrument Installation and Software Upgrade
3-49
Figure 3-67 Calibrator sampling
Be sure to immerse the tip of the sample probe into the calibrator (1 ml at least).
3. Repeat running the calibrator for 11 times as instructed by Step 2.Then, click the “Start”
button. The setup of whole blood calibration factors is completed. Click “Diluent” and a
message box will pop up. Put the tube into the tube holder and put the loaded tube holder
into the sample compartment, and then close the compartment door manually. Then press
the [RUN] key to start dispensing the diluent, as Figure 3-68 shows. Dispense the diluent
into the tube for 6 times consecutively. Then click the “Cancel” button to close the
message box. Press the [OPEN] key to open the sample compartment and take out the
tube and the tube holder.
Instrument Installation and Software Upgrade
3-50
Figure 3-68 Dispensing diluent
4. Well mix the calibrator. Set the pipette to 120µL and load the tip, and then aspirate the
calibrator, as Figure 3-69 shows.
Figure 3-69 Pipette
5. Clean the tip of the pipette by lint-free tissue, and then dispense the aspirated calibrator
into the tube loading the diluent and well mix it by the pipette, as Figure 3-70 shows.
Click “Diluent”
Then click “Cancel”
Instrument Installation and Software Upgrade
3-51
Figure 3-70 Mix the sample
6. Go back to the “Calibrator” screen, and then select the calibration mode as “Predilute”
and enter the lot No. and the expiration date of the calibrator, and then enter the WBC,
RBC, HGB, MCV and PLT targets of the calibrator into the table.
Figure 3-71 Calibration (PD)
7. Present the vial of calibrator well mixed as instructed by Step 6 into the tube holder, and
Instrument Installation and Software Upgrade
3-52
then put the loaded tube holder into the sample compartment. Close the compartment door,
and then press the [RUN] key. When you hear the beep, the analyzer is starting to run
automatically.
Be sure to immerse the tip of the sample probe into the calibrator (at least 1ml).
8. Running the calibrator consecutively for 11 times as instructed by Step 8. The setup of
predilute calibration factors is completed.
3.10.5 Linearity See installation verification report of the BC-5380.
Instrument Installation and Software Upgrade
3-53
3.11 Software upgrade 3.11.1 Upgrading the PC installed software We only update the IPU software provided that it has been completely installed on the PC. The
update will not change users’ custom configuration including settings and database. Follow the
following steps to update the software:
1. Double-click the setup. exe icon to launch the update program;
2. A window will pop up as shown in Figure 3-72;
Figure 3-72 Selecting the updating mode
3. Click “Next” to go to the installation screen shown in Figure 3-73;
Figure 3-73 Progress bar
4. When the program finishes the installation, a message box shown in Figure 3-74 will pop
up;
Instrument Installation and Software Upgrade
3-54
Figure 3-74 Prompting to update the database
5. Click the “Next” button to update the database, as shown in Figure 3-75;
Figure 3-75 Installing database files
6. When all the database files are installed, a message box shown in Figure 3-76 will prompt
Instrument Installation and Software Upgrade
3-55
the completion of installation.
Figure 3-76 Finishing installation
To update the Version 1.3 BC-5100 software, select the “Install” mode to install the new version in the previous folder rather than update the outdated software.
If the system prompts for not enough space on disk during installation, select another disk to install and make sure the free space on that disk is at least 4G.
3.11.2 Upgrading the software of the main unit (analyzer) 1. Click” Menu→All Programs→BC-5380 Auto Hematology Analyzer→Analyzer
Upgrade” to launch the update software;
Before perform this step, make sure that the analyzer is powered on for at least 30 seconds. Otherwise, the connection to the main unit may fail after reboot or the update server may be uploaded again after being connected to the main unit.
Instrument Installation and Software Upgrade
3-56
Figure 3-77 Client update software interface
For PC using Windows Vista, the IPU software or the update client may loose connection to
the main unit after reboot. Right click on the “Analyzer Upgrade” item and choose “Run as administrator” to solve this problem, as shown in Figure 3-78;
Figure 3-78 Running the update client in Windows Vista
2. Click “Analyzer Upgrade” shown in Figure 3-77 and a message box will pop up. Enter the
user name “service” and the password “Se s700” (note: there is a space between “e”
and ”s”) in the message box, as shown in Figure 3-79;
Instrument Installation and Software Upgrade
3-57
Figure 3-79 Logging in the client update software
3. Ensure proper connection between the PC (IPU software end) and the analyzer (main unit
end). If the connection fails, a message box shown in Figure 3-80 will pop up;
Figure 3-80 Connection failure message box
If the connection is proper, the update software will automatically check if the version of the
update server is suitable. If the version meets the requirement, Step 4 will be skipped;
otherwise, the update software will upload the latest update server files to the analyzer, as
shown in Figure 3-81;
Figure 3-81 Automatically uploading the update server to the analyzer
4. After the update server files have been uploaded successfully, you will be asked to restart
the main unit. Click “Ok” to restart, as shown in Figure 3-82;
Instrument Installation and Software Upgrade
3-58
Figure 3-82 Reboot message box
At this time, the update program has already been disconnect from the analyzer, so Step 1 and 2 should be performed again to launch the update program.
After the update server program being successfully updated and main unit being rebooted, wait 2 minutes before logging on to the client.
5. After logging on, a message box shown in Figure 3-83 will pop up, asking you to choose
the directory of the update files;
Figure 3-83 Selecting the directory of the update files (1)
6. Click “Browse…” in the message box shown in Figure 3-83 and select the update
package folder “step1” (or other names as defined) generated by the installation CD as the
target directory, as shown in Figure 3-84;
Instrument Installation and Software Upgrade
3-59
Figure 3-84 Selecting the directory of the update files (2)
7. Click “Ok” and the directory will be displayed in the box. Check if the displayed directory is
correct, as shown in Figure 3-85;
Figure 3-85 Selecting the directory of the update files (3)
8. Click “Ok” to start uploading update files to the main unit, as shown in Figure 3-86;
Figure 3-86 Uploading the update files
Instrument Installation and Software Upgrade
3-60
During uploading the update files, if the message box of “Analyzer disconnected! Can not upgrade the analyzer!” pops up, check if the selected folder is correct, and then perform the update procedures again.
9. After finishing uploading, a message box of successful uploading shown in Figure 3-87 will
pop up;
Figure 3-87 Uploading files successfully
The uploading process takes 1 to 2 minutes according to the size of the update files.
10. Click “Ok” and wait for a while. A message box will pop up to show the items to be updated,
as shown in Figure 3-88;
Instrument Installation and Software Upgrade
3-61
Figure 3-88 Items to be updated
11. Click “Ok” to start updating, as shown in Figure 3-89;
Figure 3-89 Progress bar
During installation, if a message box of “Disconnected from the analyzer. Unable to finish updating! Please connect the analyzer and the PC end software and try to update again.” pops up, do not switch off the power supply of the analyzer and restart the updating process from the beginning after 5 minutes.
Instrument Installation and Software Upgrade
3-62
12. Click “Ok” in the message box prompting successful update, and then switch off and
restart the analyzer;
13. Update the folder step2 (including the main unit software, configuration files and fluidic
sequence) according to the same process, and then switch off the analyzer;
14. Use the screwdriver to remove the three cross-headed screws to open the top cover, and
then open the right door. Connect the J3 interface of the data board and the J11 interface
of the autoloader board using a dedicated cord, and then restart the main unit.
Use the dedicated cord in the accessory box provided by the manufacturer.
15. Update the folder step4 (including the hardware writing software) according to the same
process, and then switch off the analyzer.
Do not delete the files under ftp://10.0.0.2/3101 before updating and keep the power supply of the main unit on. Always update the hardware after Step2 is updated successfully.
After replacing the data board (including rootfs, material code: 051-000014-00), remember to perform all the steps to update the software at the main unit end.
4-1
4 Fluidic System
4.1 Introduction of Fluidic Parts
4.1.1 Piercing needle Pierces the evacuated blood collection tube; aspirates and dispenses the blood sample.
4.1.2 Probe Wipe Washes the interior and exterior of the sample probe
4.1.3 Pumps P1: provides pressure for the pressure chamber.
P2: empties the DIFF bath and vacuum chamber and creates vacuum.
P3: empties the probe wipe, WBC bath and RBC bath.
4.1.4 Syringes Aspiration-Syringe: full volume 100 µl; it aspirates and dispenses samples and aspirates
the second diluted sample.
Diluent-Syringe: full volume 10ml; it dispenses diluent into the WBC and RBC bath and
supplies the diluent to the probe wipe.
Lyse-Syringe: full volume 2.5ml; 3 syringes are driven by a motor; it dispenses M-53Leo
(I), M-53Leo (II) and M-53H.
Sheath-Syringe: full volume 10ml; it dispenses the sheath into the flow cell, cleans the
flow cell and DIFF bath, and is also be used for sample preparation.
Sample-Syringe: full volume 250µl; it dispenses the sample into the flow cell.
4.1.5 Valves Fluidic valve: turns on and off by electromagnetic force; controls the fluidic or air flow
direction.
Pinch valve: turns on and off by electromagnetic force; starts/stops the fluidic flow.
Pressure switch: monitors the pressure of the DIFF channel.
4.1.6 Baths WBC bath: consists of a front bath, back bath and aperture. The WBC sample mixes and
reacts here and it is used for the measurement of HGB and WBC.
RBC bath: consists of a front bath, back bath and aperture. The RBC sample mixes and
Fluidic System
4-2
reacts here and it is used for the measurement of RBC/PLT.
DIFF bath: DIFF sample mixes and reacts here.
Vacuum chamber: creates and stores stable vacuum for counting the WBC and RBC by
impedance method; cleans the back bath and empties the volumetric tube.
Pressure chamber: creates and stores stable pressure to generate bubbles for the baths.
WBC isolation chamber: provides air buffer to isolate outside interference
RBC isolation chamber: provides air buffer to isolate outside interference
4.1.7 Volumetric tube WBC volumetric tube: volume 500ul; it measures the volume of the WBC sample
RBC volumetric tube: volume 300ul; it measures the volume of the RBC sample
4.1.8 Filters Filter of WBC volumetric tube: it filters the air entered the WBC volumetric tube
Filter of RBC volumetric tube: it filters the air entered the RBC volumetric tube
P1 filter: it filters the air entered the air pump.
Fluidic System
4-3
4.2 Fluidic System
4.2.1 Reagent volume required Sample analysis for a single sample each time
2 analysis modes: Open vial-whole blood mode and open vial-predilute mode.
Table 4-1 Reagent volume required when running samples
Volume of single analysis Reagents Name
CBC+DIFF CBC
M-50 Diluent M-53D Diluent ≤45mL ≤34mL
M-53LEO(I) Lyse 1.1ml 0
M-53LEO(II) Lyse 0.135ml 0 M-53 Lyse
M-53LH Lyse 0.5ml
Normal startup function (excluding background)
Reagents volume required:
Table 4-2 Reagents volume required when starting up normally
Reagents Name Volume of normal startup
M-50 Diluent M-53D Diluent 89ml
M-53LEO(I) Lyse 0
M-53LEO(II) Lyse 0 M-53 Lyse
M-53LH Lyse 2ml
Cleanser M-53 Cleanser 0
Abnormal startup function (excluding background)
Reagents volume required:
Table 4-3 reagents volume required when starting up abnormally
Reagents Name Volume of abnormal startup
M-50 Diluent M-53D Diluent 172ml
M-53LEO(I) Lyse 0
M-53LEO(II) Lyse 0 M-53 Lyse
M-53LH Lyse 2ml
Cleanser M-53 Cleanser 0
Starting up a drained system (after the Prepare to Ship procedure has be done)(excluding
startup background)
Reagents volume required:
Fluidic System
4-4
Table 4-4 reagents volume required for starting up a drained system (after the Prepare to Ship procedure has been done)
Reagents Name
Volume of starting up a
drained system (after the
prepare to ship procedure
has been done)
M-50 Diluent M-53D Diluent 207ml
M-53LEO(I) Lyse 12ml
M-53LEO(II) Lyse 12ml M-53 Lyse
M-53LH Lyse 12ml
Cleanser M-53 Cleanser 8ml
Normal shutdown function
Reagents volume required:
Table 4-5 reagents volume required when normally shutting down
Reagents Name Volume of normal shutdown
M-50 Diluent M-53D Diluent 53ml
M-53LEO(I) Lyse 0
M-53LEO(II) Lyse 0 M-53 Lyse
M-53LH Lyse 0
Cleanser M-53 Cleanser 29ml
Fluidic System
4-5
4.2.2 Fluidic System Drawing
Figure 4-1 Fluidic system (Whole Blood Mode)
Figure 4-2 Fluidic system (Predilute Mode)
Fluidic System
4-6
4.2.3 WBC Measurement Procedures
WBC Count/BASO channel
The reagents used include:
M-53LH Lyse: it lyses erythrocytes and platelets and separates basophiles from other WBCs.
Diluent: it is used for cleaning and provides the environment for reaction and measurement.
Principles: Electrical Impedance Method
Parameters: WBC, BASO# and BASO%
Diagrams: WBC histogram
Dilute ratio: 1:500
Metering volume: 500
Function description: 6µl blood sample, 2.5ml diluent and 0.5ml LH Lyse are mixed in the
bath, and then they react. After reaction, the sample flows to the back bath via the
aperture, driven by the vacuum, and the blood cells are counted when passing through
the aperture. The volume of the sample is measured by the volumetric tube.
DIFF channel
The reagents used include:
LEO (I) and LEO (II): they lyse RBCs and differentiate WBCs into 4 sub-populations.
Diluent: cleans the fluidic system and provides sheath.
Principles: flow cytometry and semiconductor laser scatter
Parameters: MONO#, MONO%, LYMPH#, LYMPH %, NEUT#, NEUT%, EOS# and
EOS%
Diagrams: 4-DIFF scattergram
Dilute ratio: 1:138
Metering volume: 127µl
Function description: 0.6ml Leo (I) is dispensed into the bath as the initial volume, and
then 9µl blood sample and 0.6ml Leo (I) are dispensed. After the sample is well mixed and
reacts with the reagent for a while, 0.135ml Leo (II) is dispensed. After the sample reacts
with the reagents for a while, the sample is driven to the bottom of the flow cell and the
sheath flow is enabled. Then the sample is dispensed into the flow cell by the sample
syringe for measurement.
4.2.4 RBC/PLT Measurement Procedures
Fluidic System
4-7
RBC/PLT channel
The reagents used include:
Diluent: dilutes samples, washes, provides a conductive environment and processes the cells
Principles: Electrical Impedance Method
Parameters: RBC and PLT
Diagrams: RBC histogram and PLT histogram
Dilute ratio: 1:20000
Metering volume: 300µl
Function description: 52.08µl sample dilution (1:417.7) is aspirated by the sample probe
from the WBC bath. The sample probe then moves to the RBC bath and mixes the
sample with 2.5ml diluent to dilute the sample to the ratio of 1:20000. Then, the mixed
sample is driven into the back bath through the aperture by the vacuum, and the blood
cells are counted when passing through the aperture. The volume of the sample is
measured by the volumetric tube.
4.2.5 HGB Measurement Procedures
HGB channel
The reagents used include:
Diluent: dilutes samples and washes
LH lyse: lyses the red blood cells and combines the hemoglobin
Principles: Colorimetry
Parameters: HGB
Dilute ratio: 1:500
The HGB unit calculates the concentration of HGB by comparing the light passing through
the sample with the blank reference reading.
Fluidic System
4-8
4.3 Sequence of Whole Blood CBC+DIFF Measurement
4.3.1 Sampling and dispensing procedures A. Aspirating samples and dispensing samples into the DIFF bath
20µl sample is aspirated by the aspirat① ion syringe from the sample tube, and then the
sample probe ascends while its exterior is being cleaned.
The sample probe moves to above the DIFF bath.②
The sample probe descends into the DIFF bath and dispenses ③ 9µl sample.
B. First Diluting
The sample pr① obe ascends from the DIFF bath, meanwhile the probe exterior is cleaned.
The sample probe moves to above the WBC bath②
The sample probe descends into the WBC bath, and then dispenses ③ 6µl sample. The
dispensed sample is diluted by the 2.5ml diluent that is already there to form the 1:417.7
dilution.
C. Cleaning the residual sample
The sample probe ascends from the WBC bath, meanwhile the probe exterior is cleaned.①
The sample probe dispenses diluent into the probe wipe from the top to wash off the residual ②
sample.
The sample probe descends again into the WBC bath and aspirates the diluted sample.③
D. RBC second diluting
The sample probe ascends from the WBC bath, meanwhile the probe exterior is cleaned.①
The sample probe moves above the RBC bath.②
The sample p③ robe descends into the RBC bath, and then dispenses 1ml diluted sample. The
dispensed sample is diluted by the diluent that is already in the bath to form the 1:20000
dilution.
E. Sample probe returning to home position
The sample probe ascends from the RBC① bath, meanwhile the probe exterior is cleaned.
The sample probe returns to the home position.②
Fluidic System
4-9
A B
C D
E
Figure 4-3 Dispensing procedures of sample probe
Fluidic System
4-10
4.3.2 DIFF Channel
Cleaning the reaction bath before dispensing samples(0 - 6s)
During this period:
The DIFF bath is drained. Then, 0.6ml LEO (I) Lyse is dispensed into the DIFF bath and then
the bath is drained again. The sheath syringe and sample syringe aspirate the diluent.
1. During 0-1.5s, valve V38 is energized and waste pump 2 operates to drain the DIFF bath.
2. During 2.8-3.8s, valve 05 is energized and the lyse syringe dispenses 0.6ml LEO (I) Lyse
into the DIFF bath to clean it.
3. During 4.5-6s, valve V38 is energized and waste pump 2 operates to drain the DIFF bath
and then the DIFF bath is ready to receive sample.
4. Valve 39 is pinched. The sheath syringe and sample syringe aspirate diluent respectively.
Sample dispensing, mixing and DIFF sample preparation (6-19.8s)
During this period:
1. Dispensing LEO (I) Lyse for the first time: valve V05 is energized and 1.1ml pre-heated
LEO (I) Lyse is dispensed into the DIFF bath via the 2.5ml lyse syringe.
2. Sample dispensing and mixing: The 100ul aspiration syringe dispenses 9ul sample into
the DIFF bath.
3. Mixing by bubbling: valve V09 is energized to mix the sample with the LEO(I) lyse by
bubbling.
4. Dispensing LEO (II) Lyse: valve V06 is energized and 0.135ml pre-heated LEO (II) Lyse is
dispensed into the DIFF bath via the 2.5ml Lyse syringe.
5. Mixing sample by bubbling: valve 09 is energized to mix the sample by bubbling.
6. DIFF sample preparation: valve V39 is de-energized to open the path between the DIFF
bath and the syringe. Valve 22 is energized and valve 23 keeps de-energized. The
sheath syringe aspirates 1.2ml sample from the DIFF bath.
When the sample preparation is done, the status of the fluidic system of the DIFF unit is shown
in Figure 4-4. The tubing highlighted in red is the part that holds the sample.
The sheath flow of the DIFF channel and the measurement (19.8-40.5s)
During this period:
1. Sheath fluid forming: V23 and V24 are energized. The sheath syringe dispenses the
diluent into the flow cell at a certain speed to form the sheath.
2. Dispensing the sample at a high speed: The sample syringe dispenses the sample into the
flow cell at a high speed.
3. Measurement: when the sample fluid is stable, the sample syringe dispenses the sample
Fluidic System
4-11
into the flow cell at a low speed while the sheath syringe dispenses the diluent into the flow
cell to surround the sample. Thus, the sheath flow passes through the flow cell at a stable
speed for measurement.
The fluidic system status of this period is shown in Figure 4-5.
Figure 4-4 Fluidic status of DIFF channel sample preparation
Fluidic System
4-12
Figure 4-5 Fluidic status of DIFF channel measurement
Cleaning of DIFF channel (40.5-59s)
During this period:
1. The sample surrounded by the sheath passes through the flow cell. Then, the flow cell and
the sample probe are washed.
2. Cleaning the sample preparation tubing: valve V22 are energized and valve V23 and V24
keep energized. The sheath syringe dispenses diluent to clean the sample preparation
tubing. Meanwhile, valve 38 is energized and waste pump 2 operates to drain the DIFF
bath.
3. Cleaning the DIFF bath
a. Dispensing 1.7ml diluent for cleaning: the sheath syringe dispenses 1.7ml diluent into
the DIFF bath through V21 and V22 to clean the bath.
b. Draining: valve 38 is energized and waste pump 2 operates to drain the DIFF bath.
Meanwhile, the sheath syringe aspirates 1.5ml diluent.
c. Dispensing 1.5ml diluent for soaking: valve V22, V23 and V24 are energized. The
sheath syringe dispenses 1.5ml diluent to soak the DIFF bath.
Fluidic System
4-13
4.3.3 WBC/HGB Channel
HGB background, cleaning, draining volumetric tube and sample distribution (0-18.2s)
1. 5s HGB background test.
2. Creating vacuum: the vacuum used for draining the volumetric tube and cleaning the back
bath is provided by the vacuum chamber. Thus, the waste pump is needed to create the
vacuum in the vacuum chamber.
3. Zapping and cleaning the back bath.
4. Draining the WBC bath: valve V36 is energized and waste pump 3 operates to drain the
WBC bath.
5. Draining the WBC volumetric tube: valve V25 and V31 are energized to drain the
volumetric tube for three times.
Figure 4-6 Fluidic status when draining WBC volumetric tube
6. WBC sample distribution:
a. At 9.4s, valve V02, V03 are energized; and at 9s, V27is energized and the waste
pump operates. The sample probe ascends while the diluent syringe dispenses the
diluent to clean the exterior of the sample probe.
b. Dispensing initial volume and sample probe moving to the WBC bath: valve V03 is
energized. The 10ml diluent syringe dispenses a small portion of the diluent into the
WBC bath as the initial volume. The sample probe moves to the WBC bath, and then
descends into the WBC bath.
c. Dispensing the blood sample: the aspiration syringe dispenses 6 µl sample into the
WBC bath while mixing the sample by bubbling.
Fluidic System
4-14
Dispensing lyse and mixing (18.3-25s)
1. The sample probe ascending: the sample probe ascends from the WBC bath, meanwhile
its exterior is washed.
2. Dispensing HGB lyse and mixing: valve V07 is energized. The 2.5ml lyse syringe
dispenses the LH Lyse into the WBC bath and the samples mixes with the lyse by bubbling.
The system waits for the sample becoming stable.
Creating vacuum (18.1-25s)
Vacuum is the power that drives the fluid movement. Waste pump 2 pumps out the air in the
vacuum chamber to create the vacuum between -300HPa and -260HPa.
Counting and bath cleaning
1. WBC counting: valve V33 is energized to start counting. Driven by the stable vacuum
(-300 to -260 HPa), the sample in the front bath flows to the back bath in the direction
shown in Figure 4-7. The sample volume measured is determined by the diluent volume
that flows between the two photocouplers of the WBC volumetric tube..
2. HGB measurement: HGB measurement starts at 41s.
3. Cleaning the front bath twice.
4. Cleaning the back bath: valve V29 and V31 are energized. Driven by the vacuum, the
diluent enters the volumetric tube via valve V29, and then passes through valve V31 and
enters the vacuum chamber. Then, valve V31, V33 and V13 are energized. The diluent
flows into the back bath through the volumetric tube to drain the waste in the back bath.
Driven by waste pump 2, the waste in the vacuum chamber is drained through V37.
Figure 4-7 Fluidic status when WBC counting
Fluidic System
4-15
Figure 4-8 Fluidic status when cleaning the WBC back bath
4.3.4 RBC/PLT Channel
Cleaning, volumetric tube draining (0-18.2s)
1. Creating vacuum: the vacuum used for draining the RBC volumetric tube and cleaning the
back bath is all provided by the vacuum chamber. Thus, waste pump P2 is used to create
the vacuum in the vacuum chamber.
2. Zapping and cleaning the back bath: zap the RBC aperture, and then open V30, V34 and
V14 to clean the back bath and drain the bubbles caused by zapping.
3. Draining the RBC bath and volumetric tube: valve V35 is energized and waste pump P3
operates to drain the RBC bath. Then, V26 and V32 are energized together to drain the
volumetric tube for three times.
4. First dilution and aspiration: during 0-8.7s, the sample probe finishes sampling and sample
is dispensed into the DIFF bath. Thus, the first dilution is formed during the 8.8-20.5s
period and then a portion of the diluted sample is aspirated back into the sample probe.
Fluidic System
4-16
Figure 4-9 Fluidic status when draining RBC volumetric tube
Second diluting and mixing (18.4-25s)
1. Aspirating the first dilution: the aspiration syringe aspirates the sample diluted at the ratio
of 1:417.6 from the WBC bath, and then ascends while its exterior is being cleaned.
2. Sample probe moving to RBC bath and dispensing initial volume and the sample: the
sample probe moves to the RBC bath. Valve V03 and V08 are energized at the same time,
and the diluent syringe dispenses a certain amount of diluent into the RBC bath and then
the sample probe descends into the RBC bath. Valve V01 and V03 are energized and the
10ml diluent syringe dispenses the first dilution into the RBC bath to form the second
dilution at the ratio of 1:20000.
3. Mixing: Valve V11 is energized. Driven by the pressure, the sample is mixed by bubbling
for three times.
Creating vacuum (18.1-25s)
The stable vacuum is created to drive the fluidics when running samples.
Waste pump 2 pumps out the air in the vacuum chamber to create the vacuum between
-300HP and -260HP.
Counting and bath cleaning
The fluidic status of RBC counting is shown in Figure 4-10.
1. RBC counting: valve V34 is energized to start counting. Driven by the stable vacuum
Fluidic System
4-17
(-300 to -260 HP), the sample in the front bath flows to the back bath in the direction
shown in the following figure. The sample volume measured is determined by the diluent
volume that flows between the two photocouplers of the WBC volumetric tube.
Figure 4-10 Fluidic status of RBC counting
Figure 4-11 Fluidic status when cleaning the back bath and draining the front bath
2. Draining the front bath and cleaning: Valve V35 is energized and waste pump pumps the
remaining sample out of the RBC bath. The diluent syringe dispenses 3ml diluent into the
RBC bath for soaking.
Fluidic System
4-18
3. Cleaning the back bath: The vacuum of -400 to -300HP is created in the vacuum chamber.
Valve V30 and V32 are energized. The volumetric tube is cleaned and bubbles are
discharged from the top of the volumetric tube. Vale V32, V14, V34 and V30 are energized
to clean the back bath, and then the waste is discharged.
4.3.5 Sequence of Predilute Mode CBC+DIFF Measurement In the predilute mode, the sample needs to be first manually diluted at the ratio of 1:10. The
sequence inside the analyzer has no great difference with the whole blood mode.
4.3.6 Sequence of CBC Measurement In the CBC mode, the DIFF channel is not used and the sample dispensing procedures are
also changed.
Sampling and sample dispensing procedures of CBC Mode
A. Aspirating sample and dispensing sample into the WBC bath
15µl sample is aspirated by the aspiration sy① ringe, and then the sample probe ascends
meanwhile its exterior is cleaned and the sample probe dispenses 3µl sample into the probe
wipe.
The sample probe moves to above the WBC bath②
The sample probe descends into the WBC bath and dispenses 6µl sample in③ to the WBC
bath.
B. Cleaning the sample remains
The sample probe ascends from the WBC bath, meanwhile its exterior is cleaned.①
The sample probe dispenses cleanser into the probe wipe from the top to wash off the ②
residual sample..
The sample probe descends③ again into the WBC bath and aspirates the diluted sample.
C. RBC second dilution
The sample probe ascends from the WBC bath meanwhile its exterior is cleaned.①
The sample probe moves to above the RBC bath.②
The sample probe descends into the RBC bath, and ③ then dispenses 1ml diluted sample into
RBC bath, where the sample is further diluted by the diluent that is already there to form the
second dilution at the ratio of 1:20000.
D. Sample probe returning to home position
The sample probe ascends from the RBC b① ath meanwhile its exterior is cleaned.
Fluidic System
4-19
The sample probe returns to the home position.②
A B
C D
Figure 4-12 Dispensing procedures of sample probe
5-1
5 Hardware System
The BC-5380 hardware system consists of the boards, power execution parts, sensors,
switches and cables. Their relationship is shown below.
Mother Board
Data Board
Drive BoardPower Supply Board
Volume mesure board and press
sensors
FSPre-amp
board
SSPre-amp
board
Liquid level sensor
Laser board
Auto-sample board
Indication Board
Key Board
Vavles and pumps
Fan
Sample drive
motors
Syringe drive
motors
WBC Pole
RBC Pole
Door driver
Optics couple sensor
Jam drive motor
Feed drive motor
AC Power Switch HGB Module
Heater for optics
systemAnd
Temp. sensors
Heater for Diff pool and
temp. sensors
Ambience
temp. sensor
Sensor on right door
Sensor on optics system
NetPort
BC-5380 Hardware System
Figure 5-1 BC-5380 hardware system structure
Note:
In the figure, the thin arrows represent the cable connections; the thick arrows (connecting the
data board and the driver board) represent the plugboard connections; the square boxes
Hardware System
5-2
represent the boards and the ellipses represent the components.
The boards of the BC-5380 and their functions are shown in the table below.
Table 5-1 Boards of the BC-5380 and their functions
No. Name Functions
1 Data board CPU basic system + FPGA + A/D + Analog
channel
2 Drive board MCU+FPGA+ Valves and pumps driving +
Photocoupler detecting + Temperature and
pressure monitoring
3 Mother board Power distribution and centralized wiring.
Plugboard supported
4 Laser control board Laser creating and controlling
5 FS pre-amplification
board
Collecting and pre-amplifying the low-angle
scattered light
6 SS pre-amplification
board
Collecting and pre-amplifying the high-angle
scattered light
7 Volumetric board Creating the volumetric signals of RBC and
WBC channels; the vacuum and pressure
sensors and amplifying the detected signals
8 Indicator board Buzzer and indicator
9 Power board Supply D5V,P12V,P24V,A±12V power
Supply AC120V/50Hz zapping power
10 Liquid-level board Liquid-level detection of reagents
11 Autoloader board Sample transport, mix control,
electromagnetic valve control
12 Key board [RUN] key and [OPEN] key
Hardware System
5-3
5.1 Mother board
5.1.1 Introduction The function of the mother board is to realize power distribution and centralized wiring of board
input/output.
5.1.2 Board Composition The mother board is composed of two slots and several connectors. The two slots are for
connecting the data board and the drive board respectively; while the connectors are for
various input/output signals.
Function
Power interface
Slot for data board
Slot for drive board
Connectors for board signals (autoloader board, laser control board, pre-amp board,
volumetric board, liquid level checker, key board and indicator board, etc.)
Connectors for component (valves, pumps, motors, heaters, fan, HGB module,
independent key, optics, etc.) signals
Communication interface (COM port and LAN port)
Block diagram
Hardware System
5-4
Sockets for Drive Board
Socket for
power( Autosample board)
Sockets for vavles,pumps,motors,heat
ers
Sockets for
PowerSupply
Socktes for Data Board
Socket for laser control board
Sockets for optics sensor
Socket for valume mesure and press
control board
Sockets for pre-amp. board
Socket for Indicator and key board
DB9(rs232)
Power filter& indicator
Isolate transfoemer
rs232 int. driver
Data borad power
Drive boatd power
Socket for Temp. sensor
Socket for HGB module
Socket for Auto-sample
Socket for fan
RS422 int driver
RJ45 Socket for suck key and right door chec
ker
Socket for Liq.level checker
Figure 5-2 Function of mother board
Interface definition
List of connectors
Table 5-2 List of connectors
SN. PIN Number Function Description
PA1 96 Data board connector, digital signals
PA2 96 Data board connector, analog signals
PB1 96 Drive board connector, digital signals and part of optics signals
PB2 96 Drive board connector, part of optics signals and motor drive
Hardware System
5-5
signals
PB3 96 Drive board connector, drive signals of motors, valves, pumps,
and heaters
J1 8 RS-232 LAN port
J2 9 Reserved for COM port debugging
J3 10 Multi-functional JTAG port, debugging or upgrading MCU
programs of autoloader board
J4 4 RS-422 COM port, COM port of autoloader board
J5 8 Connector for the key board and indicator board
J6 6 Right door switch, and optical shielding box switch
J7 6 Laser control board connector
J8 8 Reserved for side fluorescence channel (SF)
J9 8 side scatter channel (SS)
J10 8 Connector for forward scatter channel (FS)
J11 5 HGB module connector
J12 8 A+/- 12V connector
J13 3 AC120V connector
J14 6 Temperature sensor connector
J15 12 Volumetric board connector
J16 34 Connector for level sensors and associated photocouplers
J17 No in use
J18 24 Connector for photocouplers 1-6
J19 24 Connector photocouplers 7-12
J20 24 Connector photocouplers 13-18
J21 3 Reserved for DC mix motor
J22 8 Connector stepping motors 1-2
J23 8 Connector stepping motors 3-4
J24 8 Connector stepping motors 5-6
J25 8 Connector stepping motors 7-8
J26 No in use
J27 4 Connector for heater 2
J28 4 Connector for heater 1
J29 20 Connector for pumps 1-7
J30 20 Connector for valves 1-10
J31 20 Connector for valves 11-20
J32 20 Connector for valves 21-30
J33 20 Connector for valves 31-40
J34 2 Fan connector
J35 10 Connector for P12V and P24V
Hardware System
5-6
J36 6 Connector for autoloader board power
J37 4 D+5V connector
Data board slots (PA1 and PA2)
Data board slots are two 96pin DIN 41612 connectors.
Connector type: AMP VME female connector 3*32PIN vertical connection
Model: 5535043-4
Table 5-3 Definition of PA1 for mother board
Column A Column B Column C
1 D+5V D+5V D+5V
2 D+5V D+5V D+5V
3 D+5V D+5V D+5V
4 D+5V D+5V D+5V
5 D+5V D+5V D+5V
6 DGND DGND DGND
7 DGND DGND DGND
8 DGND DGND DGND
9 DGND DGND DGND
10 DGND DGND DGND
11 DGND DGND DGND
12 TP_TX+ PWFBOUT TP_RX+
13 TP_TX- DGND TP_RX-
14 DGND DGND DGND
15 DGND DGND DGND
16 TMS_TO_DRV TMS_TO_AS UART3_TO_PC
17 TCK_TO_DRV TCK_TO_AS UART3_TO_DAT
18 DATA_TO_DRV DATA_TO_AS DGND
19 DATA_FROM_DRV DATA_FROM_AS DGND
20 DGND UART1_TO_DAT+ UART1_TO_AS+
21 DGND UART1_TO_DAT- UART1_TO_AS-
22 UART0_TO_DRV DGND DGND
23 UART0_TO_DAT DGND DGND
24 DGND DGND #SUCK_KEY
25 DGND #VM_ON DGND
26 DGND #RBC_START #BUZ
27 DGND #RBC_STOP #COUNT_KEY
28 DGND #WBC_START FAULT
29 DGND #WBC_STOP READY
Hardware System
5-7
30 DGND DGND OPTI_DOOR
31 DGND #LASER_ON #RIGHT_DOOR
32 DGND DGND #INSERT_DOOR
Table 5-4 Definition of PA2 for mother board
Column A Column B Column C
1 AGND AGND AGND
2 AGND LASER_CURT AGND
3 AGND AGND AGND
4 AGND AGND AGND
5 AGND AGND AGND
6 AGND AGND AGND
7 AGND AGND AGND
8 AGND AGND AGND
9 AGND AGND AGND
10 AGND AGND AGND
11 AGND NC AGND
12 AGND AGND A GND
13 AGND SS AGND
14 AGND AGND AGND
15 AGND FS AGND
16 AGND AGND AGND
17 AGND HGB_IN AGND
18 AGND AGND AGND
19 AGND HGB+ AGND
20 AGND HGB- AGND
21 AGND A-12V A-12V
22 AGND A-12V A-12V
23 AGND A+12V A+12V
24 AGND A+12V A+12V
25 AGND A+12V A+12V
26 NC NC NC
27 NC A+70V NC
28 NC NC NC
29 NC NC NC
30 NC AC120_B NC
31 NC NC NC
32 AC120_A NC NC
Hardware System
5-8
Connector J1 for LAN port
Connector type: RJ-45 female connector with iron casing
Model: 5406217-1
Table 5-5 Definition of J1 for mother board
Definition Description
1 ET_TP+ Transmit, positive
2 ET_TP- Transmit, negative
3 ET_RP+ Receive, positive
4 ET-NGP Receive, virtual ground
5 ET-NGP Receive, virtual ground
6 ET_RP- Receive, negative
7 ET-NGR Receive, virtual ground
8 ET-NGR Receive, virtual ground
Connector J2 for COM port
Connector type: DB9 male / female connector
Table 5-6 Definition of J2 for mother board (male connector)
Pin Definition Description
1 NC
2 RXD/TXD Send/receive
3 TXD/RXD Receive/send
4 NC
5 GND
6-9 NC
Table 5-7 Definition of J2 for mother board (female connector)
Pin Definition Description
1 GND
2 NC
3 TXD/RXD Send/receive
4 RXD/TXD Receive/send
5-9 NC
Connector J3 for JTAG
Connector type: HEADER WTB 2.54MM DIP2*5 TOP 70246SERIES
Model: 70246-1001, MOLEX
Hardware System
5-9
Table 5-8 Definition of J3 for mother board
Pin Definition Description
1 D3V3
2 GND
3 GND
4 TCK JTAG clock
5 TMS JTAG status select
6 DATA_FROM_AS JTAG data to data board
7 DATA_TO_AS JTAG data to drive board
8 NC
9 GND
10 NC
Connector J6 for aspirate key and door switch on/off checking
Connector type: HEADER WTB 2MM DIP1*6SIDE PHSERIES
Model: B6B-PH-K 2MM 'JST'
Table 5-9 Definition of J6 for mother board
Pin Definition Description
1 #SUCK_KEY NC
2 GND
3 #RIGHT_DOOR Right door switch on/off
4 GND
5 OPTI_DOOR optical shielding box switch
6 GND
Connector J4 for autoloader board signal
Connector type: HEADER WTB 2MM DIP1*4SIDE PHSERIES
Model: B4B-PH-K 'JST'
Table 5-10 Definition of J4 for mother board
Pin Definition Description
1 UART2_TO_AS+ 5474 serial port 2 to positive end of
sample board data
2 UART2_TO_AS- 5474 serial port 2 to negative end of
sample board data
3 UART2_TO_DAT+ sample board serial port to positive
end of 5474 serial port data
4 UART2_TO_DAT- sample board serial port to negative
end of 5474 serial port data
Hardware System
5-10
Connector J5 for indicator board
Connector type: HEADER WTB 2MM DIP1*8 TOP PHSERIES
Model: B8B-PH-K 'JST' 2MM
Table 5-11 Definition of J5 for mother board
Pin Definition Description
1 D+5V
2 #COUNT_KEY Start sampling switch, effective low
3 #INSERT_DOOR Insert STAT switch, effective low
4 FAULT Fault status indicator light, effective high
5 READY Ready status indicator light, effective high
6 NC
7 #BUZ Buzzer driven
8 GND
Connector J7 for laser board
Connector type: HEADER WTB 2.5MM DIP1*6SIDE XHSERIES
Model: B-6B-XH-A 'JST'
Table 5-12 Definition of J7 for mother board
Pin Definition Description
1 NC
2 AGND
3 LASER_CURT Laser current feedback, analog signal
4 A+12V
5 #LASER_ON Laser source on/off signal, effective low
6 D+5V
Connector J9 and J10 for pre-amplification board
Connector type: HEADER WTB 2.5MM DIP1*8SIDE XHSERIES
Model: B8B-XH-K 'JST'
Hardware System
5-11
Table 5-13 Definition of J9, J10 for mother board
Pin Definition Description
1 AGND
2 FS/SS FS/SS signal input, analog signal
3 AGND
4 AGND
5 A-12V
6 AGND Connecting inner shield of cable
7 A+12V
8 MHOLE Connecting external shield of cable
Connector J11 for HGB module
Connector type: HEADER WTB 2MM DIP1*5SIDE PHSERIES
Model: B5B-PH-K 'JST'
Table 5-14 Definition of J11 for mother board
Pin Definition Description
1 HGB+ HGB light drive,
positive
2 HGB- HGB light drive,
negative
3 HGB_IN HGB feedback
4 AGND
5 AGND
Connector J12 for analog A12V
Connector type: HEADER WTB 4.2MM DIP2*4 SIDE, 5566 serial
Model: 1-794509-1 'JST'
Table 5-15 Definition of J12 for mother board
Pin Definition Description
1 AGND
2 NC
3 NC
4 NC
5 A+12V
6 AGND
7 AGND
8 A-12V
Connector J13 for AC120V
Hardware System
5-12
Connector type: HEADER WTB 2.5MM DIP1*3 TOP XHSERIES
Model: B-3B-XH-A 'JST' 2.5MM
Table 5-16 Definition of J13 for mother board
Pin Definition Description
1 AC120V_A A end of 120V AC
2 NC
3 AC120V_B B end of 120V AC
Drive board slots: PB1, PB2 and PB3
DIN41612, 96PIN female
Definition of PB1-PB3 signals:
Note: On the mother board, PB1 is placed laterally, with column A under it, and row 1 at left
side. Since the drive board and mother board is connected vertically, column A should be
arranged at the nearest joint side of board.
Table 5-17 Definition of PB1 for mother board
column A column B column C
1 HT1_SA HT1_SB AMBI_SA
2 HT2_SA HT2_SB AMBI_SB
3 POSI_PRESS NEGA_PRESS DGND
4 DGND DGND DGND
5 DGND DGND DGND
6 DGND DGND DGND
7 DGND DGND DGND
8 DGND DGND DGND
9 DGND DGND DGND
10 DGND DGND DGND
11 DGND DGND DGND
12 D+5V D+5V D+5V
13 D+5V D+5V D+5V
14 D+5V D+5V D+5V
15 D+5V D+5V D+5V
16 D+5V D+5V D+5V
17 DGND DGND DGND
18 #LIQ1 #LIQ2 TMS_TO_DRV
19 #LIQ3 #LIQ4 TCK_TO_DRV
20 #LIQ5 #LIQ6 DATA_TO_DRV
21 #LYSE_SENSOR #WASTE_SENSOR DATA_FROM_DRV
22 #LS_ON DGND DGND
Hardware System
5-13
23 DGND DGND DGND
24 OP17_CTRL OP17_STS UART0_TO_DRV
25 OP18_CTRL OP18_STS UART0_TO_DAT
26 OP19_CTRL OP19_STS DGND
27 OP20_CTRL OP20_STS DGND
28 OP1_CTRL OP1_STS OP2_CTRL
29 OP2_STS OP3_CTRL OP3_STS
30 OP4_CTRL OP4_STS OP5_CTRL
31 OP5_STS OP6_CTRL OP6_STS
32 OP7_CTRL OP7_STS OP8_CTRL
Table 5-18 Definition of PB2 for mother board
column A column B column C
1 OP8_STS OP9_CTRL OP9_STS
2 OP10_CTRL OP10_STS OP11_CTRL
3 OP11_STS OP12_CTRL OP12_STS
4 OP13_CTRL OP13_STS OP14_CTRL
5 OP14_STS OP15_CTRL OP15_STS
6 OP16_CTRL OP16_STS DM_ON
7 DGND DGND DGND
8 NC NC NC
9 P+24V SM1_AP SM1_AP
10 P+24V SM1_AN SM1_AN
11 P+24V SM1_BP SM1_BP
12 P+24V SM1_BN SM1_BN
13 P+24V SM2_AP SM2_AP
14 P+24V SM2_AN SM2_AN
15 P+24V SM2_BP SM2_BP
16 P+24V SM2_BN SM2_BN
17 P+24V SM3_AP SM3_AP
18 P+24V SM3_AN SM3_AN
19 P+24V SM3_BP SM3_BP
20 P+24V SM3_BN SM3_BN
21 P+24V SM4_AP SM4_AP
22 P+24V SM4_AN SM4_AN
23 P+24V SM4_BP SM4_BP
24 P+24V SM4_BN SM4_BN
25 PGND SM5_AP SM5_AP
Hardware System
5-14
26 PGND SM5_AN SM5_AN
27 PGND SM5_BP SM5_BP
28 PGND SM5_BN SM5_BN
29 PGND SM6_AP SM6_AP
30 PGND SM6_AN SM6_AN
31 PGND SM6_BP SM6_BP
32 PGND SM6_BN SM6_BN
Table 5-19 Definition of PB3 for mother board
column A column B column C
1 PGND SM7_AP SM7_AP
2 PGND SM7_AN SM7_AN
3 PGND SM7_BP SM7_BP
4 PGND SM7_BN SM7_BN
5 PGND SM8_AP SM8_AP
6 PGND SM8_AN SM8_AN
7 PGND SM8_BP SM8_BP
8 PGND SM8_BN SM8_BN
9 PGND HT1_ON HT1_ON
10 PGND HT2_ON HT2_ON
11 PGND PGND PGND
12 PGND PUMP7 PUMP7
13 PGND PUMP6 PUMP6
14 P+12V PUMP5 PUMP5
15 P+12V PUMP4 PUMP4
16 P+12V PUMP3 PUMP3
17 P+12V PUMP2 PUMP2
18 P+12V PUMP1 PUMP1
19 V1 V2 V3
20 V4 V5 V6
21 V7 V8 V9
22 V10 V11 V12
23 V13 V14 V15
24 V16 V17 V18
25 V19 V20 V21
26 V22 V23 V24
27 V25 V26 V27
28 V28 V29 V30
Hardware System
5-15
29 V31 V32 V33
30 V34 V35 V36
31 V37 V38 V39
32 V40 NC NC
Connector J14 for temperature sensor
Connector type: HEADER WTB 2.5MM DIP1*6 TOP XHSERIES
Model: B-6B-XH-A 'JST'
Table 5-20 Definition of J14 for mother board
Pin Definition Description
1 HT1_SA
2 HT1_SB
both ends of thermistor of heater
1
3 HT2_SA
4 HT2_SB
both ends of thermistor of heater
2
5 AMBI_SA
6 AMBI_SB
both ends of thermistor of
environment temperature
Connector J15 for sensors of volumetric board and pressure control board
Connector type: HEADER WTB 2MM DIP1*12 TOP PHSERIES
Model: SIP12TD-B12B-PH-K 'JST'
Table 5-21 Definition of J15 for mother board
Pin Definition Description
1 P+12V
2 PGND
3 DGND
4 POSI_PRESS Pressure chamber (pressure) signal
5 NEGA_PRESS Vacuum chamber (vacuum) signal
6 DGND
7 #VM_ON Volumetric optics is on, effective low
8 #RBC_START RBC volume measurement starts, effective low
9 #RBC_STOP RBC volume measurement stops, effective low
10 #WBC_START WBC volume measurement starts, effective low
11 #WBC_STOP WBC volume measurement stops, effective low
12 D+5V
Connector J16 for liquid level sensor
Connector type: HEADER WTB 2MM DIP2*17 TOP
Model: B34B-PHDSS-B(LF)(SN) 'JST'
Hardware System
5-16
Table 5-22 Definition of J16 for mother board
Pin Definition Description
1 D+5V
2 #LIQ1 LIQ1status, effective low
3 #LIQ2 LIQ 2 status, effective low
4 #LIQ3 LIQ 3 status, effective low
5 #LIQ4 LIQ 4 status, effective low
6 #LIQ5 LIQ 5 status, effective low
7 #LS_ON Control signal for liquid level board
8 GND
9 NC
10 NC
11 #LIQ6 LIQ 6 status, effective low
12 GND
13 #LYSE_SENSOR LYSE level sensor, status, i, effective low
14 GND
15 #WASTE_SENSOR Waste level sensor status, effective low
16 GND
17 NC
18 NC
19 OP17_CTRL Optics 17 control
20 OP17_STS Optics 17 status
21 GND
22 GND
23 OP18_CTRL Optics 18 control
24 OP18_STS Optics 18 status
25 GND
26 GND
27 OP19_CTRL Optics 19 control
28 OP19_STS Optics 19 status
29 GND
30 GND
31 OP20_CTRL Optics 20 control
32 OP20_STS Optics 20 status
33 GND
34 GND
Connector J18, J19, J20 for optics of position sensor
Position sensor has three connectors, all using double-row and 2mm connectors.
Hardware System
5-17
Connector type: HEADER WTB 2.0MM DIP2*12 TOP
Model: B24B-PHDSS-B (LF) (SN) 'JST'
Table 5-23 Definition of J18 for mother board
Pin Definition Description
1 OP1_CTRL Optics 1 drive
2 OP1_STS Optics 1 status
3,4 DGND
5 OP2_CTRL Optics 2 drive
6 OP2_STS Optics 2 status
7,8 DGND
9 OP3_CTRL Optics 3 drive
10 OP3_STS Optics 3 status
11,12 DGND
13 OP4_CTRL Optics 4 control
14 OP4_STS Optics 4 status
15,16 DGND
17 OP5_CTRL Optics 5 drive
18 OP5_STS Optics 5 status
19,20 DGND
21 OP6_CTRL Optics 6 drive
22 OP6_STS Optics 6 status
23,24 DGND
Table 5-24 Definition of J19 for mother board
Pin Definition Description
1 OP7_CTRL Optics 7 drive
2 OP7_STS Optics 7 status
3,4 DGND
5 OP8_CTRL Optics 8 drive
6 OP8_STS Optics 8 status
7,8 DGND
9 OP9_CTRL Optics 9 drive
10 OP9_STS Optics 9 status
11,12 DGND
13 OP10_CTRL Optics 10 drive
14 OP10_STS Optics 10 status
15,16 DGND
17 OP11_CTRL Optics 11 drive
Hardware System
5-18
18 OP11_STS Optics 11 status
19,20 DGND
21 OP12_CTRL Optics 12 drive
22 OP12_STS Optics 12 status
23,24 DGND
Table 5-25 Definition of J20 for mother board
Pin Definition Description
1 OP13_CTRL Optics 13 drive
2 OP13_STS Optics 13 status
3,4 DGND
5 OP14_CTRL Optics 14 drive
6 OP14_STS Optics 14 status
7,8 DGND
9 OP15_CTRL Optics 15 drive
10 OP15_STS Optics 15 status
11,12 DGND
13 OP16_CTRL Optics 16
14 OP16_STS Optics 16 status
15,16 DGND
17-24 NC
Connector J21 (reserved) for DC mix motor
Connector type: HEADER WTB 2MM DIP1*3 TOP PHSERIES
Model: B3B-PH-K 'JST'
Table 5-26 Definition of J21 for mother board
Pin Definition Description
1 DM_ON DC Motor drive
2 NC
3 DGND
Connector J22, J23, J24, J25 for stepping motors
Each connector connects two motors.
Connector type: HEADER WTB 2.54MM DIP1*8 TOP XHSERIES
Model: B-8B-XH-A 'JST' 2.54MM
Hardware System
5-19
Table 5-27 Definition of J22 for mother board
Pin Definition Description
1 SM1_AP A-phase coil drive , motor 1
2 SM1_AN A-phase coil drive , motor 1
3 SM1-BP B-phase coil drive , motor 1
4 SM1_BN B-phase coil drive , motor 1
5 SM2_AP A-phase coil drive , motor 2
6 SM2_AN A-phase coil drive , motor 2
7 SM2-BP B-phase coil drive , motor 2
8 SM2_BN B-phase coil drive , motor 2
Table 5-28 Definition of J23 for mother board
Pin Definition Description
1 SM3_AP A-phase coil drive , motor 3
2 SM3_AN A-phase coil drive , motor 3
3 SM3-BP B-phase coil drive , motor 3
4 SM3_BN B-phase coil drive , motor 3
5 SM4_AP A-phase coil drive , motor 4
6 SM4_AN A-phase coil drive , motor 4
7 SM4-BP B-phase coil drive , motor 4
8 SM4_BN B-phase coil drive , motor 4
Table 5-29 Definition of J24 for mother board
Pin Definition Description
1 SM5_AP A-phase coil drive , motor 5
2 SM5_AN A-phase coil drive , motor 5
3 SM5-BP B-phase coil drive , motor 5
4 SM5-BN B-phase coil drive , motor 5
5 SM6_AP A-phase coil drive , motor 6
6 SM6_AN A-phase coil drive , motor 6
7 SM6-BP B-phase coil drive , motor 6
8 SM6_BN B-phase coil drive , motor 6
Hardware System
5-20
Table 5-30 Definition of J25 for mother board
Pin Definition Description
1 SM7_AP A-phase coil drive , motor 7
2 SM7_AN A-phase coil drive , motor 7
3 SM7-BP B-phase coil drive , motor 7
4 SM7_BN B-phase coil drive , motor 7
5 SM8_AP A-phase coil drive , motor 8
6 SM8_AN A-phase coil drive , motor 8
7 SM8-BP B-phase coil drive , motor 8
8 SM8_BN B-phase coil drive , motor 8
Connector J27, J28 for heaters
Connector type: HEADER WTB 2.5MM DIP1*4 TOP XHSERIES
Model: B-4B-XH-A 'JST'
Table 5-31 Definition of J27 for mother board
Pin Definition Description
1 P+24V
2 HT1_ON Heater 1 drive
3,4 Heater 1 control
Table 5-32 Definition of J28 for mother board
Pin Definition Description
1 P+24V
2 HT2_ON Heater 2 drive
3,4 Heater 2 control
Connector J29, J30, J31, J32, J33 for pumps and valves
Valves and pumps have five double-row PH-model connectors.
Connector type: HEADER WTB 2MM DIP2*10 TOP PHDSERIES
Model: B20B-PHDSS-B(LF)(SN)
Table 5-33 Definition of J29 for mother board
Pin Definition Description
1~6 NC
7 PUMP7 Pump 7 drive
8 P+12V
9 PUMP6 Pump 6 drive
10 P+12V
Hardware System
5-21
11 PUMP5 Pump 5 drive
12 P+12V
13 PUMP4 Pump 4 drive
14 P+12V
15 PUMP3 Pump 3 drive
16 P+12V
17 PUMP2 Pump 2 drive
18 P+12V
19 PUMP1 Pump 1 drive
20 P+12V
Table 5-34 Definition of J30 for mother board
Pin Definition Description
1 V1 Valve 1 drive
2 P+12V
3 V2 Valve 2 drive
4 P+12V
5 V3 Valve 3 drive
6 P+12V
7 V4 Valve 4 drive
8 P+12V
9 V5 Valve 5 drive
10 P+12V
11 V6 Valve 6 drive
12 P+12V
13 V7 Valve 7 drive
14 P+12V
15 V8 Valve 8 drive
16 P+12V
17 V9 Valve 9 drive
18 P+12V
19 V10 Valve 10 drive
20 P+12V
Table 5-35 Definition of J31 for mother board
Pin Definition Description
1 V11 Valve 11 drive
2 P+12V
Hardware System
5-22
3 V12 Valve 12 drive
4 P+12V
5 V13 Valve 13 drive
6 P+12V
7 V14 Valve 14 drive
8 P+12V
9 V15 Valve 15 drive
10 P+12V
11 V16 Valve 16 drive
12 P+12V
13 V17 Valve 17 drive
14 P+12V
15 V18 Valve 18 drive
16 P+12V
17 V19 Valve 19 drive
18 P+12V
19 V20 Valve 20 drive
20 P+12V
Table 5-36 Definition of J32 for mother board
Pin Definition Description
1 V21 Valve 21 drive
2 P+12V
3 V22 Valve 22 drive
4 P+12V
5 V23 Valve 23 drive
6 P+12V
7 V24 Valve 24 drive
8 P+12V
9 V25 Valve 25 drive
10 P+12V
11 V26 Valve 26 drive
12 P+12V
13 V27 Valve 27 drive
14 P+12V
15 V28 Valve 28 drive
16 P+12V
17 V29 Valve 29 drive
Hardware System
5-23
18 P+12V
19 V30 Valve 30 drive
20 P+12V
Table 5-37 Definition of J33 for mother board
Pin Definition Description
1 V31 Valve 31 drive
2 P+12V
3 V32 Valve 32 drive
4 P+12V
5 V33 Valve 33 drive
6 P+12V
7 V34 Valve 34 drive
8 P+12V
9 V35 Valve 35 drive
10 P+12V
11 V36 Valve 36 drive
12 P+12V
13 V37 Valve 37 drive
14 P+12V
15 V38 Valve 38 drive
16 P+12V
17 V39 Valve 39 drive
18 P+12V
19 V40 Valve 40 drive
20 P+12V
Connector J37 for D+5V
Connector type: HEADER WTB 4.2MM DIP2*2 TOP 5566SERIES
Model: 3928-1043
Table 5-38 Definition of J37 for mother board
Pin Definition
1 D+5V
2 D+5V
3 DGND
4 DGND
Connector J35 for P+12V, P+24V
Hardware System
5-24
Connector type: HEADER WTB 4.2MM DIP2X5 TOP 5566SERIES
Model: 39-31-0108
Table 5-39 Definition of J35 for mother board
Pin Definition
1 P+24V
2 P+24V
3 P+24V
4 P+12V
5 P+12V
6 PGND
7 PGND
8 PGND
9 PGND
10 PGND
Connector J34 for fan power
Connector type: HEADER WTB 2.5MM DIP1*2 TOP XHSERIES
Model: B-2B-XH-A 'JST'
Table 5-40 Definition of J34 for mother board
Pin Definition
1 P+12V
2 PGND
Connector J36 for power (autoloader board)
Connector type: HEADER WTB 4.2MM DIP2*3 TOP 5566SERIES
Model: 3928-1063
Table 5-41 Definition of J36 for mother board
Pin Definition
1 D+5V
2 P+12V
3 P+24V
4 DGND
5 PGND
6 PGND
Hardware System
5-25
Assembly drawing
Figure 5-3 Assembly drawing of the mother board
Hardware System
5-26
5.1.3 Adjustment and Test Points There are no test points designed specially for mother board. To test certain power or signal,
use the touch points of connectors directly.
On the mother board with the Debug Serial Communication Port (DSCP) configured, the
configuration of COM port is implemented by several connection matrices. See the following
introduction for connection method.
By adjusting JP1 to JP3, the J2 COM port can support the following 5 debugging modes (all via
PC COM port).
1. Data board COM 3: debugging the data board function
2. Data board COM 0: debugging the data board’s function of controlling the drive board
3. Drive board COM: debugging the drive board function
4. Data board COM 1: debugging the data board’s function of controlling the autoloader
board
5. Autoloader board COM: debugging the function of autoloader board
Notice: In case of functions 2 or 3 (4 and 5 are the same), do not connect the drive board and
the data board at the same time. Namely, in case of function 2, do not connect the drive board,
and in case of function 3, do not connect the data board.
JP1 and JP2 respectively connect UART1, 3, 0 to the input of the RS-232 transceiver. JP1 can
also be used for selecting the UART1 COM direction. JP2 connects interface signals of the
data board or the autoloader board to the RS-422 transceiver. (U2 in Figure 5-3)
See Table 5-42 for configuration details.
Table 5-42 Configuration details
Debugging COM port JP1 JP2
Data board UART0 Put jumpers on pins 1-2 and
3-4,
No jumpers
Drive board COM Put jumpers on pins 5-6
and 7-8
No jumpers
Data board UART1 Put jumpers on pins 9-10
and 11-12.
Put jumpers on pins
1-2,,3-4,,5-6, and,7-8..
Autoloader board
COM
Put jumpers on pins 9-10
and 11-12.
Put jumpers on pins
9-10,11-12,13-14,and 15-16.
Data board UART3 Put jumpers on pins 13-14
and 15-16.
No jumpers.
A male or female receptacle can be installed at J2. Four cable types can be supported, as
shown in the following table.
Hardware System
5-27
Table 5-43 COM port receptacle and cable selection
Connector cable JP3
Serial cable with female
connectors on both ends,
signals not intercrossed
3-4, 7-8, 13-14, 15-16
DB9 male
receptacle
(straight or 90°
installation)
Serial cable with female
connectors on both ends,
signals intercrossed; namely
the COM cable
3-4, 7-8, 9-10, 11-12
Serial cable with a female
connector on one end and a
male connector on the other
end, signals not intercross;
namely the COM extension
cable
1-2, 5-6, 13-14, 15-16
DB9 female
receptacle
(straight or 90°
installation)
Serial cable with a female
connector on one end and a
male connector on the other
end, signals intercross
1-2, 5-6, 9-10, 11-12
JP4 needs to be configured as shown in the following table when the MCU JTAG of the drive
board or the autoloader board is to be connected to J3.
Table 5-44 Connection matrix of multi-functional JTAG port
JTAG function JP4
Connect MCU JTAG of
autoloader board
No jumpers
Connect MCU JTAG of
drive board
Put jumpers on pins 1-2, 3-4, 5-6 and
7-8.
Note: the mother board does not provide a fixed connection to the JTAG port of the autoloader
MCU. If you need to update the autoloader MCU on the analyzer, connect the J11 and J3
receptacles using the cable dedicated for autoloader update before updating on PC using the
client update software. The dedicated cable is provided to the service and manufacture
personnel, not as an accessory of the analyzer.
Hardware System
5-28
5.1.4 Removal
Purpose
The mother board is unlikely to be damaged. But in case you need to replace it, follow the
steps below.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Removal
Follow these two procedures to disassemble the mother board:
To remove the mother board from the analyzer:
1. Power off the analyzer.
2. Unplug the power cable and the LAN cable from the back of the analyzer.
3. Remove the top cover and front cover of the analyzer
4. Remove the data board and the drive board.
5. Disconnect all cables from the mother board.
6. Use the 107 cross-headed screwdriver to remove the four M4X8 small panhead screws
fixing the bracket.
7. Pull the bracket to remove it.
Hardware System
5-29
Figure 5-4 Disassembly diagram of the mother board-1
1 ―――Bracket 2 ―――Mother board
3 ―――M4X8 panhead screw
To disassemble the mother board:
1. Use the 107 cross-headed screwdriver to remove the fourteen M3X6 cross-headed
panhead screws on the mother board.
2. Pull the mother board towards the panel by a certain distance, remove the LAN port
connector from the metal plate and take out the mother board.
Figure 5-5 Disassembly diagram of the mother board-2
1 ―――Bracket 2 ―――Mother board
3―――M3X6 cross-headed panhead screws
Hardware System
5-30
Installation
Install the mother board as per the above-mentioned procedures in the reverse order.
Note: Since the mother board has many connectors, pay attention to the number indicated on
the connector to avoid wrong connection. The design of the mother board has included certain
error-proof measures. Generally speaking, wrong connection will not lead to safety issues, yet
the whole unit cannot function properly
Verification
1. Power on the analyzer, the fan at the back of the analyzer should run normally. Use a
multimeter to measure the voltage of D+5V, P+12V, P+24V, A+/-12V, AC120V, and all
measurements should be within normal ranges.
2. Power indicators of the boards are on, showing that the mother board is functioning
correctly.
5.1.5 Troubleshooting Since the function of the mother board is to realize power distribution and centralized wiring, a
failure in the mother board usually takes the form of hardware system failure.
Table 5-45 Troubleshooting
Failure Cause Recommended Action
Abnormal voltage
detected
cable not properly
connected to power
connector
Check connectors J12, J13, J35,
J37;
If the connection is normal, use a
multimeter to check if the voltage
of each power source indicated
on the mother board is correct.
Fan not running. Fan damaged or no power
supply
Check connector J34;
Check the P12V indicator on the
drive board
Communication failure
of autoloader board
Connector J4 abnormal..
Connector J36 abnormal..
Check connector J4
Check connector J36
Pressure control
abnormal
Connector J15 abnormal Check connector J15
Volumetric abnormal Connector J15 abnormal Check connector J15
Liquid level inspection Connector J16 abnormal Check connector J16
Hardware System
5-31
abnormal, misinform
Laser not light or laser
current detected
abnormal
Connector J17 abnormal Check connector J7
Pre-amp board has no
power or work
abnormally.
Connectors J9 and J10
abnormal
Check connectors J9 and J10
Motor not active Motor drive cable not
properly connected .
Check connectors J22,J23,J24,
and 25
Valve not active Valve drive cable not
properly connected.
Check connectors J30,J31,J32
and J33
Pump not active Pump drive cable not
properly connected
Check connector J29
Temperature control
fails.
Temperature sensor not
properly connected.
Check connector J14
Hardware System
5-32
5.2 Data board
5.2.1 Introduction The “data” function of the data board consists of two aspects: analog part and digital part. The
analog part is responsible for obtaining, amplifying, and conditioning sensor signals and
converting the analog signals into digital signals. The digital part processes digital signals and
uploads them to the PC as well as provides the platform for the built-in software and the
access interface.
5.2.2 Board composition
Function
The analog part:
Conditioning and converting power
Monitoring voltage (including power voltage, WBC aperture voltage, RBC aperture voltage
and FS background voltage, also monitoring laser drive current and AD working status,
etc.)
Collecting and conditioning WBC/RBC volume signals
Detecting and amplifying HGB signals
Conditioning WBC DIFF signals
Driving constant-current supply
Controlling and setting amplification factors for all channels
Zapping apertures
AD conversion
The digital part:
Providing a platform for software operating
Controlling interfaces for the indicator board and the key board
Providing communication interfaces for the autoloader board and the drive board
Providing communication interfaces for PC
Controlling and monitoring interfaces for the volumetric board and the laser board
Processing digital signals and identifying pulse
Block diagram
Hardware System
5-33
Figure 5-6 Data board circuit
Analog power filter module
Monitoring module
RBC/PLT amplification
HGB amplification
HGB led constant current driver
Constant current source/zappingcontrol module of RBC/PLT hole
Digital circuits of data board
A/D module
SS amplification
FS amplification
SF amplification
HGB
Gain control
RBC/PLT aperture
Gain control
HGB
Driver control
Constant current source /burn control
Gain control
Gain control
SS
SF
FS
A±12V、AC120V
Gain control
WBC
Laser driver
RBC/PLTsensor
WBC sensor
HGB sensor
SS pre-amplification board
FS pre-amplification board
SS pre-amplification board
Power for analog circuits
WBC amplification
Constant current source/zappingcontrol module of WBC hole
RBC/PLT
WBC aperture
FS base currrent
signal monitoring of laser current driver
Gain control
Constant current source /zapping control
Figure 5-7 Functions of analog part on data board
Hardware System
5-34
Figure 5-8 Functions of digital part on data board
Description
The analog part:
Conditioning and converting voltages
The A±12V and AC120V analog power sent from the mother board are conditioned or
converted to be the required voltages for the board. The AC120V is used for zapping
apertures.
Monitoring voltage
The voltage of A±12V power and VCONST, WBC aperture voltage, RBC aperture voltage, FS
blank voltage, laser drive current and AD working status are all monitored via AD.
Collecting and amplifying WBC/RBC volume signals
WBC/RBC volume signals are collected and conditioned, and then sent to the AD converter.
Detecting and amplifying HGB signals
The HGB current signals are converted into voltage signals through colorimetric method, then
after conditioning, they are sent to the AD converter.
Conditioning WBC DIFF signals
The WBC DIFF signals (including FS, SS and SF signal; SF signal reserved )are first
collected and conditioned, and then sent to the AD converter.
Driving constant-current supply
The data board provides the constant current for the electrodes of the WBC bath and the
RBC/PLT bath and the HGB light.
Controlling setting amplification factors for all channels
Hardware System
5-35
The zapping voltage on-off, constant current supply/zapping switch and HGB light on-off are
controlled by the GPIO interface of CPU. The amplification factors of WBC DIFF, WBC,
RBC/PLT and HGB signal channels are set by the FPGA as required.
Zapping aperture
If the analyzer has been running for a certain period or the WBC/RBC bath aperture clogs, the
zapping function shall be performed to clean up the aperture by high temperature.
A/D conversion
The cell signals (WBC, RBC, PLT and WBC DIFF signals), voltage monitoring signals and
HGB signals are converted into digital signals, and then sent to FPGA for further processing.
The digital part:
Providing a platform for operating system and application software
CPU: MCF5474, 266MHz;
DDR:128M, 133MHz;
FLASH:16Mx16bit
Communicating with the TTL port of the driver board
Sending control command to the driver board and receiving its response.
Baud rate: 38400bps
Data bit: 8bit
Stop bit: 1bit
Start bit: 1bit
Parity: even
Communicating with the 422 port of the autoloader board
Sending control command to the autoloader board via serial port and receiving its response.
Baud rate: 38400bps
Data bit: 8bit
Stop bit: 1bit
Start bit: 1bit
Parity: even
The IO port for the driver board and autoloader board upgrading
By simulating the JTAG sequence through the CPU IO port, the driver board and the
autoloader board can be upgraded respectively.
Interface for the indicator board
The status indicator and the buzzer on the indicator board are controlled by the CPU IO port.
Interface for the volumetric board
Enabling the volumetric board and receiving its detecting signals
Interface for the laser control board
Enabling the laser control board and detecting laser diode current
Key input interface
Receiving the key events (including the aspiration key and the keypad).
Hardware System
5-36
Status detection interface
Detecting the open/closed status of the side door and laser shielding box
Analog signal input interface
Receiving the analog signals from the pre-amplification board and the aperture electrodes
10M/100M network interface
The network interface is introduced by the mother board and connected to PC by the cross
network cable, and capable of program downloading and data transmission, etc.
BDM debug interface
The BDM interface of CPU, which is used for detecting hardware, downloading program and
writing the FLASH, is not available for customers.
JTAG interface of FPGA
It debugs FPGA and downloads FPGA program. It is not available for customers.
Interface definition
The data board is a plug-in board. It provides 9 connectors: the connectors for aperture
electrodes, the connector to connect the digital part to the mother board, the connector to
connect the analog part to the mother board, the connectors for debugging and the reserved
connectors. See Figure 5-9 for the layout of the data board.
See the following table for the function of each connector.
Table 5-46 Interfaces of data board
Connector Function Number of
pins Note
J1 Connects the RBC/PLT aperture
electrode 3
/
J2 Connects the WBC aperture electrode 3 /
J3 Connects the digital part to the mother
board 96
/
J7 Connects the analog part to the
mother board 96
/
J8 JTAG interface of FPGA 10 /
J9 Reserved 20 /
J10 Reserved 40 /
J11 Reserved 14 /
J13 BDM interface of CPU 26 /
The layout of interfaces is shown in the figure below:
Hardware System
5-37
Figure 5-9 Interfaces layout of data board
Definition of J1
J1 is the signal interface of RBC/PLT.
Table 5-47 Definition of J1
Pin Name Note
1 SHELL Shielding ground
2 RHOLE_A RBC/PLT signals
3 RHOLE_B RBC analog ground
Definition of J2
J2 is the signal interface of WBC.
Table 5-48 Definition of J2
Pin Name Note
1 WHOLE _B WBC analog ground
2 WHOLE_A WBC signals
3 SHELL Shielding ground
Definition of J3
J3 is the signal interface for the digital part of data board and mother board, and is European
96 pin connector.
Hardware System
5-38
Table 5-49 Definition of J3
Pin Definition Function Pin Definition Function
A1 D+5V Power supply B17 JTAG_TCK_
AS
JTAG clock of autoloader
board
A2 D+5V Power supply B18 JTAG_TO_A
S
JTAG sending of
autoloader board
A3 D+5V Power supply B19 JTAG_FROM
_AS
JTAG receiving of
autoloader board
A4 D+5V Power supply B20 UART1_TO_
DAT+
Reserved debug serial
port sending
A5 D+5V Power supply B21 UART1_TO_
DAT-
Reserved debug serial
port sending
A6 DGND Signal ground B22 DGND Signal ground
A7 DGND Signal ground B23 DGND Signal ground
A8 DGND Signal ground B24 DGND Signal ground
A9 DGND Signal ground B25 #VMCTRL Volumetric board enabling
A10 DGND Signal ground B26 #RBC_STAR
T
Start RBC volume
metering
A11 DGND Signal ground B27 #RBC_STOP Stop RBC volume
metering
A12 TP_TX+ Positive sending
terminal of
network
B28 #WBC_STAR
T
Start WBC volume
metering
A13 TP_TX- Negative sending
terminal of
network
B29 #RBC_STOP Stop WBC volume
metering
A14 DGND Signal ground B30 DGND Signal ground
A15 DGND Signal ground B31 #LASER_CT
RL
Laser board enabling
A16 JTAG_TMS JTAG mode
signal
B32 DGND Signal ground
A17 JTAG_TCK JTAG clock signal C1 D+5V Power supply
A18 JTAG_TO_
DRV
JTAG sending of
drive board
C2 D+5V Power supply
A19 JTAG_TO_
DAT
JTAG receiving of
drive board
C3 D+5V Power supply
A20 DGND Signal ground C4 D+5V Power supply
A21 DGND Signal ground C5 D+5V Power supply
A22 UART0_TO Serial port C6 DGND Signal ground
Hardware System
5-39
_DRV sending of drive
board
A23 UART0_TO
_DAT
Serial port
receiving of drive
board
C7 DGND Signal ground
A24 DGND Signal ground C8 DGND Signal ground
A25 DGND Signal ground C9 DGND Signal ground
A26 DGND Signal ground C10 DGND Signal ground
A27 DGND Signal ground C11 DGND Signal ground
A28 DGND Signal ground C12 TP_RX+ Positive receiving
terminal of network
A29 DGND Signal ground C13 TP_RX- Negative receiving
terminal of network
A30 DGND Signal ground C14 DGND Signal ground
A31 DGND Signal ground C15 DGND Signal ground
A32 DGND Signal ground C16 UART3_TO_
PC
PC communication serial
port sending
B1 D+5V Power supply C17 UART3_TO_
DAT
PC communication serial
port receiving
B2 D+5V Power supply C18 DGND Signal ground
B3 D+5V Power supply C19 DGND Signal ground
B4 D+5V Power supply C20 UART1_TO_
AS+
autoloader board serial
port sending
B5 D+5V Power supply C21 UART1_TO_
AS-
autoloader board serial
port receiving
B6 DGND Signal ground C22 DGND Signal ground
B7 DGND Signal ground C23 DGND Signal ground
B8 DGND Signal ground C24 #
SUCK_KEY
Aspiration key control
input
B9 DGND Signal ground C25 DGND Signal ground
B10 DGND Signal ground C26 #BUZ Buzzer control output
B11 DGND Signal ground C27 #COUNT_KE
Y
Aspirate key control input
B12 PWFBOUT Network power
supply feedback
C28 FAULT_J Indicator control output
(red)
B13 DGND Signal ground C29 READY_J Indicator control output
(green)
B14 DGND Signal ground C30 OPTI_DOOR Laser shielding box
monitoring input
Hardware System
5-40
B15 DGND Signal ground C31 #RIGHT_DO
OR
Right door status
monitoring input
B16 JTAG_TMS
_AS
JTAG mode of
autoloader board
C32 #INSERT_KE
Y
Compartment door open
key control input
Definition of J7
J7 is the signal interface for the analog part of data board and mother board, and is European
96 pin plug.
Table 5-50 Definition of J7
Pin Name Note Pin Name Note
A1 AGND Analog ground B17 HGB_IN HGB signals input
A2 AGND Analog ground B18 AGND Analog ground
A3 AGND Analog ground B19 HGB+ HGB light drive positive
electrode
A4 AGND Analog ground B20 HGB- HGB light drive negative
electrode
A5 AGND Analog ground B21 A-12V -12V power supply
A6 AGND Analog ground B22 A-12V -12V power supply
A7 AGND Analog ground B23 A+12V +12V power supply
A8 AGND Analog ground B24 A+12V +12V power supply
A9 AGND Analog ground B25 A+12V +12V power supply
A10 AGND Analog ground B26 NC /
A11 AGND Analog ground B27 NC /
A12 AGND Analog ground B28 NC /
A13 AGND Analog ground B29 NC /
A14 AGND Analog ground B30 AC120V_
B
Zapping voltage input
A15 AGND Analog ground B31 NC /
A16 AGND Analog ground B32 NC /
A17 AGND Analog ground C1 AGND Analog ground
A18 AGND Analog ground C2 AGND Analog ground
A19 AGND Analog ground C3 AGND Analog ground
A20 AGND Analog ground C4 AGND Analog ground
A21 AGND Analog ground C5 AGND Analog ground
A22 AGND Analog ground C6 AGND Analog ground
A23 AGND Analog ground C7 AGND Analog ground
A24 AGND Analog ground C8 AGND Analog ground
A25 AGND Analog ground C9 AGND Analog ground
Hardware System
5-41
A26 NC / C10 AGND Analog ground
A27 NC / C11 AGND Analog ground
A28 NC / C12 AGND Analog ground
A29 NC / C13 AGND Analog ground
A30 NC / C14 AGND Analog ground
A31 NC / C15 AGND Analog ground
A32 AC120V_
A
Zapping voltage input C16 AGND Analog ground
B1 AGND Analog ground C17 AGND Analog ground
B2 LASER_I
N
Laser current
monitoring input
C18 AGND Analog ground
B3 AGND Analog ground C19 AGND Analog ground
B4 AGND Analog ground C20 AGND Analog ground
B5 AGND Analog ground C21 A-12V -12V power supply
B6 AGND Analog ground C22 A-12V -12V power supply
B7 AGND Analog ground C23 A+12V +12V power supply
B8 AGND Analog ground C24 A+12V +12V power supply
B9 AGND Analog ground C25 A+12V +12V power supply
B10 AGND Analog ground C26 NC /
B11 FS_IN FS signals input C27 NC /
B12 AGND Analog ground C28 NC /
B13 SS_IN SS signals input C29 NC /
B14 AGND Analog ground C30 NC /
B15 SF_IN SF signals input C31 NC /
B16 AGND Analog ground C32 NC /
Definition of J8
J8 is the JTAG interface of FPGA.
Table 5-51 Definition of J8
Pin Name Note
1 FPGA_TCK Clock
2 GND Signal ground
3 FPGA_TDO Data output
4 VDD Power supply
5 FPGA_TMS Testing mode selection
6 VDD Power supply
7 NC /
8 NC /
Hardware System
5-42
9 FPGA_TDI Data input
10 GND Signal ground
Definition of J9
Reserved. Not available for customers.
Definition of J10
Reserved. Not available for customers..
Definition of J11
Reserved. Not available for customers.
Definition of J13
J11 is the BDM interface of CPU.
Table 5-52 Definition of J13
Pin Definition Function
1 NC /
2 BKPT# Breakpoint setup
3 GND Signal ground
4 DSCLK synchronizing clock input
5 GND Signal ground
6 TCK Clock input of JATG
7 BDM_RSTI BDM reset
8 DSI Serial input
9 VDD Power supply
10 DSO Serial output
11 GND Signal ground
12 PSTDATA7
13 PSTDATA6
14 PSTDATA5
15 PSTDATA4
16 PSTDATA3
17 PSTDATA2
18 PSTDATA1
19 PSTDATA0
Configuration data
20 GND Signal ground
21 NC /
22 NC /
23 GND Signal ground
Hardware System
5-43
24 PSTCLK Processor clock
25 NC /
26 MCF_TA# Transmission response
Assembly drawing
Hardware System
5-44
Figure 5-10 Assembly drawing of data board
Hardware System
5-45
5.2.3 Adjustment and Test Points
All the adjustable parameters of analog part are adjusted as per the command of FPGA. Adjust
the parameters' settings in the software interface if necessary.
LED function definition
Table 5-53 LED function definition
No. Indicator Function description
1 D17 A-12V power indicator
2 D20 A+12V power indicator
3 D21 A-5V power indicator
4 D22 A+5V power indicator
5 D39 Reserved for software use.
6 D40 Reserved for software use.
7 D41 Reserved for software use.
8 D42 Reserved for software use.
9 D43 Reserved for software use.
10 D44 Network conflict indicator. It turns on when network
conflicts.
11 D45 100M network communication indicator. It turns on
when enabled.
12 D46 10M network communication indicator. It turns on
when enabled.
13 D47 Network connection indicator. It turns on when
enabled.
14 D48 Network full-duplex communication indicator. It
turns on when enabled.
15 D49 FPGA configuration complete indicator. It turns on
when the configuration is done.
16 D51 D+5V power indicator
Function definition of test points
Hardware System
5-46
Table 5-54 Test points on data board
No. Test point Signal
under testFunction description
1 +12V AVCC +12V power supply of analog part
2 -12V AVSS -12V power supply of analog part
3 +5V AVDD +5V power supply of analog part
4 -5V AVEE -5V power supply of analog part
5 VR_MON +2.5V +2.5V power supply of analog part
6 VR_H +2.5V +2.5V power supply of analog part
7 +56V VCONST Drive voltage of constant current source
of electrode
8 +12VM A+12V_MO
N
+12V power supply monitoring
9 -12VM A-12V_MO
N
-12V power supply monitoring
10 +56VM VCONSTM Monitor drive voltage of constant current
source of electrode
11 RHOLE RHOLE RBC Aperture Voltage monitoring
12 WHOLE WHOLE WBC Aperture Voltage monitoring
13 FSBASE FSBASE FS background voltage monitoring
14 LASER LASER Laser diode drive current monitoring
15 HGB1 HGB1 HGB channel voltage value after I/V
16 NHGB #HGBLED HGB light on-off control
17 HGB HGB HGB measurement voltage value
18 NBURN #BURN Zapping control
19 PULSE PULSE Pulse control signals that produce
zapping voltage
20 FS_IN FS_IN FS pre-amplification board output signals
21 FS1 FS1 Adjustable amplified FS signals
22 FS FS Analog signals outputted by FS channel
23 SS_IN SS_IN SS pre-amplification board output signals
24 SS1 SS1 Adjustable amplified SS signals
25 SS SS Analog signals outputted by SS channel
26 SF_IN SF_IN SF pre-amplification board output signals
27 SF1 SF1 Adjustable amplified SF signals
28 SF SF Analog signals outputted by SF channel
29 NCONST #CONST Constant current source control
30 NSEL_R #SELECT_ RBC aperture zapping/constant current
Hardware System
5-47
RBC selection control
31 NSEL_W #SELECT_
WBC
WBC aperture zapping/constant current
selection control
32 RBC1 RBC1 RBC channel output signals of adjustable
amplification grade
33 RBC2 RBC2 RBC channel output signals of direct
current restoring grade
34 RBC3 RBC3 RBC channel output signals of low-pass
filter grade
35 RBC RBC RBC channel signals of output grade
36 PLT1 PLT1 PLT channel output signals of direct
current restoring grade
37 PLT2 PLT2 PLT channel output signals of low-pass
filter grade
38 PLT PLT PLT channel signals of output grade
39 WBC1 WBC1 RBC channel output signals of adjustable
amplification grade
40 WBC2 WBC2 WBC channel output signals of direct
current restoring grade
41 WBC3 WBC3 WBC channel output signals of low-pass
filter grade
42 WBC WBC WBC channel signals of output grade
43 HCLK HCLK HGB and clock of monitoring voltage
44 HGND HGND Analog ground
45 RBCF RBCF Signals inputted into AD from RBC
46 VR_R VR_R AD reference voltage of RBC
47 RAGND RAGND Analog ground
48 RCLK RCLK AD clock signals of RBC
49 WBCF WBCF Signals inputted into AD from WBC
50 VR_W VR_W AD reference voltage of WBC
51 WAGND WAGND Analog ground
52 WCLK WCLK AD clock signals of WBC
53 PLTF PLTF Signals inputted into AD from PLT
54 VR_P VR_P AD reference voltage of PLT
55 PAGND PAGND Analog ground
56 PCLK PCLK AD clock signals of PLT
57 MCLK MCLK AD clock of monitoring voltage
58 FSF FSF AD input from FS
59 FSCLK FSCLK AD clock signals of FS
Hardware System
5-48
60 SSF SSF AD input from SS
61 SSCLK SSCLK AD clock signals of SS
62 SFF SFF AD input from SF
63 SFCLK SFCLK AD clock signals of SF
64 TP8 MCF_FBAD
1
Flexbus address data 1 of CPU
65 TP9 MCF_FBAD
2
Flexbus address data 2 of CPU
66 TP37 FPGA_DDR
_VRF
VREF power of DDR on FPGA, +1.25V
67 TP45 VCC +5V digital power supply
68 TP46 VDD +3.3V digital power supply
69 TP47 D_1V5 +1.5V digital power supply
70 TP48 D_1V8 +1.8V digital power supply
71 TP57 D_1V2_FG
PA
+1.2V digital power supply
72 TP71 D_2V5 +2.5V digital power supply
73 TP74 #MCF_CS1 Flexbus CS1 of CPU
74 TP75 MCF_R/#W Flexbus read/write signals of CPU
75 TP76 MCF_FBAD
6
Flexbus address data 6 of CPU
76 TP78 D_1V25VTT VTT power of DDR on CPU, +1.25V
77 TP79 #MCF_CS0 Flexbus CS0 of CPU
78 TP80 FPGA_DDR
_2V5
+2.5V DDR power on FPGA
79 TP81 FPGA_DDR
_VTT
VTT power of DDR on FPGA, +1.25V
80 TP82 MCF_FBAD
5
Flexbus address data 5 of CPU
81 TP83 D_1V25VR
F
VREF power of DDR on CPU, +1.25V
82 TP85 MCF_FBAD
5
Flexbus address data 4 of CPU
83 TP86 #MCF_OE Output enabling of Flexbus of CPU
84 TP87 MCF_TS Address latch indication of Flexbus of
CPU
85 TP95 MCF_FBAD
3
Flexbus address data 3 of CPU
86 TP96 #MCF_RST Reset input signals of CPU
Hardware System
5-49
I
87 TP101 #MCF_IRQ
1
Interrupt input signals 1 of CPU
88 TP102 #MCF_IRQ
2
Interrupt input signals 2 of CPU
89 TP103 #MCF_IRQ
3
Interrupt input signals 3 of CPU
90 TP104 #MCF_IRQ
5
Interrupt input signals 5 of CPU
91 TP105 #MCF_IRQ
6
Interrupt input signals 6 of CPU
92 TP106 #MCF_IRQ
7
Interrupt input signals 7 of CPU
AGND, TP10-21, TP38, TP44, TP70, TP73, TP77, TP88, TP90, TP92-94 and TP107-116 are
connected to the GND network and usually used as the connecting point of the oscilloscope
probe grounded end.
Data board key definition
Table 5-55 Key definition
No. Function
S12 Manually reset the main CPU
5.2.4 Disassembly and assembly method
Purpose
Damaged data board needs to be replaced by a new one..
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Disassembly
Data board disassembly consists of the following two main procedures:
Hardware System
5-50
Remove the data board from the main unit:
1. Shut down the analyzer.
2. Remove the left door.
3. Remove the top cover board
4. Use the 107 cross-headed screwdriver to remove the two cross-headed bolts (M3X6) from
the upper cover of the shielding box, and then remove the upper cover.
5. Disconnect the two power cables connected to the data board.
6. Use your hand or cross-headed screwdriver to remove the two hand-tightened screws
from the bracket.
7. Use both of your hands to flip outward the two latches on both sides of the data board, and
then draw the data board out.
Figure 5-11 Disassembly diagram of data board-1
1---Upper cover of the shielding box of
data board
2---Cross-recessed panhead screw M3x6
(with washer)
3---Data board 4---Hand-tightened screws
Further disassembly of the data board
1. Use the 107 cross-headed screwdriver to remove the 5 cross-headed panhead screws
(M3X6) from the data board.
Hardware System
5-51
2. Remove the data board from the bracket.
3. Use the 107 cross-headed screwdriver to remove the 2 panhead screws (M3X8) and 6
socket cap screws (M3X12) from the lower cover of the shielding box.
Figure 5-12 Disassembly diagram of data board-2
1---Cross-recessed panhead screw
M3x6 (with washer)
2---Data board
3---Bracket 4---Lower cover of the shielding box of data
board
5---Panhead screw M4X8 6---Socket cap screws M3X12
Assembly
Assemble the data board as per the above-mentioned steps in the reverse order.
Verification
1. Power indicators D20, D17, D22, D21 and D51 turns on.
2. Indicator D49 turns on after FPGA initialization.
3. Indicator D47 and D48 turn on and D45 flashes after network is enabled.
5.2.5 Troubleshooting
The errors analysis and the corresponding troubleshooting of the analog part on data board
are listed in the following table. Most of the errors can be tested and determined by testing the
corresponding test points with a multimeter. A measurement value within the normal range
indicates the circuit or component is in good working condition. Otherwise, be sure to find out
Hardware System
5-52
the particular error according to the certain symptom and remove it accordingly.
Table 5-56 Troubleshooting of analog part
No. Error Character Possible Cause Recommended action
Improper cable
connection with
the mother board
or disconnection
caused by a
broken cable
Reconnect the cable or
change a new cable
Improper 12V
power supply sent
by the mother
board or the
power board
Check whether the proper
power supply is sent by the
mother board or the power
board
1 12V power
supply error
The voltage of the
+12V test point
exceeds the range
of 12±0.6V; or the
ripple noise is
greater than 50mV.
Inductor L16 or
capacitor C111,
C95 and C88 are
damaged or not
properly soldered.
Re-solder or change the
component
Improper cable
connection with
the mother board
or disconnection
caused by a
broken cable
Reconnect the cable or
change a new cable
Improper -12V
power supply sent
by the mother
board or the
power board
Check whether the proper
power supply is sent by the
mother board or the power
board
2 -12V power
supply error
The voltage of the
-12V test point
exceeds the range
of -12±0.6V; or the
ripple noise is
greater than 50mV.
Inductor L18 or
capacitor C112,
C94 and C87 are
damaged or not
properly soldered
Re-solder or change the
component
Hardware System
5-53
3 +5V power
supply error
The voltage of the
+5V test point
exceeds the range
of +5±0.25 V
The +5V relating
circuit is rosin
jointed; the power
supply’s
impedance to the
ground is
decreased; or
power chip U14 is
damaged
First ensure the +12V power
supply works normally. Then,
check the +5V relating circuit,
re-solder or change the
component.
4 -5V power
supply error
The voltage of the
-5V test point
exceeds the range
of -5±0.25 V
The -5V relating
circuit is rosin
jointed; the power
supply’s
impedance to the
ground is
decreased; or
power chip U12 is
damaged
First ensure the -12V power
supply works normally. Then,
check the -5V relating circuit,
re-solder or change the
component.
5 +2.5V reference
power supply
error
The voltage of the
VR_MON or VR_H
test point exceeds
the range of
2.5±0.125 V
If the +5V (AVDD)
power supply
works normally,
then the problem
occurred probably
because power
chip U45 or U43,
and its peripheral
components are
damaged or not
properly soldered.
Check the relating circuit;
re-solder or change the
component
The boost circuit
works improperly
Check the chip U72, U70,
U71 and their peripherals for
any rosin joint or damage
6 +56 constant
current drive
power supply
error
The voltage of the
VCONST test point
exceeds the range
of 56±3V The 56V
voltage-regulator
diode is damaged
or not properly
soldered.
Re-solder or change the
component
Hardware System
5-54
7 +56V monitoring
voltage error
The voltage of the
+56VM test point
exceeds the range
of 1.37±0.2V
If the +56V power
supply works
normally, then the
problem occurred
probably because
something is
wrong with the
monitoring circuit..
Check the U39 and relating
components; re-solder or
change the component
8 +12V monitoring
voltage error
The voltage of the
+12VM test point
exceeds the range
of 1.8V-2.2V
The relating circuit
is not properly
soldered; the
power supply’s
impendence to the
ground is
decreased;
Something is
wrong with the
power IC or the
corresponding AD.
If the +12V power supply
works normally, check U40
and the relating components,
re-solder or change the
component.
9 -12V monitoring
voltage error
The voltage of the
-12VM test point
exceeds the range
of 2.2V-2.6V
The relating circuit
is not properly
soldered; the
power supply’s
impendence to the
ground is
decreased;
Something is
wrong with the
power IC or the
corresponding AD.
Check U40 and the relating
components, re-solder or
change the component under
the circumstance that the
-12V power supply works
normally
Constant current
source control
signals abnormal,
56V
voltage-regulator
diode abnormal or
boost circuit error
Check whether the constant
current source control signals
are normal. See the
foregoing troubleshooting to
remove the error of 56V
circuit.
10 RBC aperture
voltage
monitoring error
The value of the
aperture voltage
exceeds the range
of 10V-14V
Bath not properly
wired; clogging.
Check the bath wiring;
perform the unclogging
procedure
Hardware System
5-55
Monitoring circuit
error
Components of
the monitoring
circuit are not
properly soldered,
or the
corresponding AD
works abnormally.
Check whether the relating
circuit or AD works normally.
11 WBC aperture
voltage
monitoring error
Same as point 10 Same as point 10 Same as point 10
HGB light error or
the gain
Check whether HGB light is
on, or adjust the gain multiple
of HGB channel
12 HGB
background
voltage
abnormal
HGB value of the
test point exceeds
the normal range
(4.2V-4.8V) The
corresponding AD
works abnormally,
or the error
message is
accidentally
triggered by
improper wiring.
Check whether the
corresponding AD works
normally.
13 FS background
voltage
monitoring error
The value of the
FSBASE test point
exceeds the normal
range (0-400mv)
The flow cell is
dirty, or something
is wrong with the
corresponding
monitoring circuit
error.
Clean the flow cell in time, or
check whether the U38 and
the corresponding AD chip
U44 work normally.
14 Laser diode
drive current
monitoring error
The value of the
LASER test point
exceeds the range
of 1.0-3.0V.
Something is
wrong with the
current value sent
by the laser
control board
Check whether the cable is
properly connected to the
laser control board and
whether the monitoring value
sent from the laser control
board is normal
Hardware System
5-56
Data board
monitoring circuit
error, the
corresponding AD
error or
connection cable
abnormal
Check whether the operation
amplifier U38 and the
corresponding AD chip U44
work normally.
15 RBC aperture
zapping error
No sound is heard
and no bubbles are
produced when
zapping the bath
Something is
wrong with the
zapping voltage
supplied by the
power board, or
with the part of the
data board that
controls zapping.
Ensuring the wiring of the
bath is correct, then check
whether the AC120V zapping
power of the power board is
proper, or check whether the
zapping control signals are
effective.
16 WBC aperture
zapping error
Same as point 15 Same as point 15 Same as point 15
17 No
WBC/RBC/PLT
result
No WBC/RBC/PLT
result, or the result
displays “***”
Bath wiring error,
analog part
channel error, AD
collection error or
algorithm
identification error
Use a oscilloscope to check
whether the analog channel
can output correct signals. If
no signal is detected,,
check the amplification stage
by stage, and check the AD
collection and data uploading
as well.
18 WBC/RBC/PLT
background
abnormal
WBC/RBC/PLT
background
abnormal
Bad shielding,
constant current
source abnormal,
analog channel
resistor-capacitor
components
incorrect or
abnormal
First ensure the bath and the
fluidics work normally, and
then check the shielding and
grounding, check
components (such as
resistors) of the constant
current source, or check the
resistor-capacitor and
operation amplifier of the
signal conditioning channel.
Hardware System
5-57
Table 5-57 Troubleshooting of digital part
No. Error Causes and solutions
1 System startup failed Possible causes: input power supply error, including
improper power cord connection, abnormal power
supply voltage on each board, flash chip error that
causes the operating system to fail to be
downloaded into DDR SDRAM; DDR SDRAM error
that causes the system to fail to operate; Or system
clock abnormal.
Solutions: press the S12 key to reset the system,
and then check whether the error still exists; check
whether the power indicator D51 is on; check
whether the fuse F4 is broken; use a multimeter to
check the voltage of each power supply; use an
oscilloscope to check whether the clock crystal
oscillator output X2 is normal.
2 Network communication failed Possible causes: network connection abnormal;
using non-standard network cable; network chip
damaged, etc.
Solutions: directly connect the analyzer and PC with
the specified shielding-cross network cable. If
nothing is wrong with the network connection and the
settings, then check the crystal oscillator X3 output
by an oscilloscope. If you doubt the network chip
U36 is damaged, replace it with a new one.
3 Abnormal communication with
UART of autoloader board
Possible causes: improper connection; interface chip
U66 damaged.
Solutions: check and make sure the connection is
proper; check whether the interface signals are
normal by an oscilloscope.
4 volumetric board signals, laser
shielding box status and right
door status can not be detected
Possible causes: improper connection; interface chip
U74 damaged.
Solutions: check and make sure the connection is
proper; check the interface signals by an
oscilloscope.
5 Abnormal communication with
UART of drive board
Possible causes: improper connection; interface chip
U67 damaged.
Solutions: check and make sure the connection is
proper; check the interface signals by an
oscilloscope.
Hardware System
5-58
6 Events of Aspirate Key, Count
Key and Compartment Door
Open Key can not be detected,
or the buzzer and indicator on
the indicator board are out of
control.
Possible causes: improper connection; interface chip
U67 damaged.
Solutions: check and make sure the connection is
proper; check the interface signals by an
oscilloscope.
Hardware System
5-59
5.3 Drive board
5.3.1 Introduction The drive board realizes the following functions: 1) driving motors, pumps, valves and heaters
etc. 2) detecting such signals as the motor position and liquid levels; 3) collecting the signals of
temperature, pressure and voltage etc. for AD conversion.
5.3.2 Board composition
Function
Serial communication
Driving stepping motors
Driving valves and pumps
Temperature control
Pressure control
System status detection
Reagent status detection
Block diagram
EP1C6Q240C8C8051F020
A3979
MS1
MS2
STEP
SLEEP
A
AN
B
BN
MOTOR
8-channel Step-motor driver module
74HC595+
IRLML2502
IN
CLK
LATCH
EN
40
SV1~SV40
40-channel solenoid valvecontrol
HOA0880-T51
CD74HC14M_SEN
20-channel motor position sensor detection
Liquid level sensor or sensitive switch
74LVCR2245
HOLD
L-SEN/SWITCH
ALE
WR
RD
A/D[7:0]
A[15:8]
F_NCONFIG
NCSO
DATA
DCLK
ASDO
EPCS1S18N
PumpHeater
6
TemperaturePressure
Power Voltage (anolog signal)
8
Databoard UART
TXD
RXD F_INT
Databoard JTAG
TMS
TDI
CLK
TDO
7-channel high power pump & valve control2-channel heater control
2-channel temperature measurement2-channel pressure measurement3-channel power voltage measurement
8-channel liquid level sensor & sensitive switch detection
Figure 5-13 Block diagram of drive board
Hardware System
5-60
Description
Serial communication
The main control unit of the drive board is a MCU, which has two asynchronous serial
communication interfaces (UART). UART0 is used for asynchronous serial communication
with the data board. It receives and analyzes control commands sent by the data board, and
responds accordingly. Communication parameters:
Baud Rate: 38400bps
Start bit: 1bit
Data bit: 8 bits
Stop bit: 1bit
Parity: even
UART1 is reserved.
Stepping motor driving
The drive board controls stepping motors through an FPGA. After output by the FPGA, the
control signal is first sent to isolator 74AHCT245, and then to controller A3979 to drive the
motor. The drive board drives 8 stepping motors: the sample syringe motor, the sample
injection syringe motor, the sheath syringe motor, the diluent syringe motor, the lyse syringe
motor, the horizontal motor and the vertical motor.
Valve and pump driving
The drive board provides the drive control for 40 valves and 7 pumps and actually drives 34
valves, 1 pressure pump and 2 vacuum pumps (also be used as the waste pump). The MCU
writes control information of valve SV1-SV40 into the valve buffer of the FGPA. Based on the
data, the FPGA outputs valve control signals through the serial-parallel conversion chip
74HC595, and drive the valves through the MOS transistor IRLML2502. The MCU first sends
the control information of PUMP1-PUMP3 to isolator 74AHCT245, and then to the drive chip
ULN2068 to control the 3 pumps respectively.
Temperature Control
The temperature control circuit keeps temperature of the reaction bath (DIFF bath) and the
optical system constant. By comparing the temperature measurements and the target values,
the MCU turns on/off the heater to control the temperature of each channel. The MC sends
heater control signals to isolator 74AHCT245, then to driver FDS6670A to control the two
heaters.
Air pressure control
The pressure control circuit controls the pressure of the vacuum chamber and the pressure
chamber. The pressure signals (voltage signals) sent by the volumetric board are measured by
Hardware System
5-61
the drive board. Through voltage follower LM358, the signals are then sent to the ADC pin of
the MCU for AD conversion and corresponding operation. Then, the control signals are output
to turn on/off the pressure pump or the vacuum pump, thus, the pressure of each chamber is
adjusted.
System status detection
The drive board detects the ambient temperature and the voltage of power supply. The mother
board sends the inquiry command to obtain the corresponding system status parameter.
Ambient temperature sensor is located on the interface board. After I/V conversion and
amplification, the signals outputted by the temperature sensor are sent to the MCU for AD
conversion and operation, then the temperature value is obtained.
The voltage detection circuit detects all the power supplies (24V, 12V and 5V) on the drive
board.
Reagent status detection
The reagent status is detected by the passive float switch. There are totally 8 reagent detection
channels on the drive board as designed, and 6 channels of them are in use to detect the
following reagents: LEO1, LEO2, LH, cleanser, diluent and waste.
Interface definition
The drive board connects the motherboard via the 3 DIN41612 plugs and 3 DIN41612
receptacles located on the two boards respectively. All the drive-board-relating peripherals
(including valves, pumps, heaters, motors and sensors etc.) are connected to the motherboard
directly. The connectors’ layout and the corresponding connections are shown in the figure
below.
Hardware System
5-62
Driver board
PB3
+5V
Liquid level sensor
PB2 SocketPB1 Socket
UART
Position sensor
6-channel step-motor+24V +12V
2-channel step-motor
Valve
HeaterPUMP
PressureReaction pool temperatureAmbient temperature
插座(96Pin)
Support board
Board mounting bracket & guide
channel
Base board
Socket 96Pin
JTAG Optical train temperature
J9
J1 J10 J8 J2
J4 J5 J3
Socket
Figure 5-14 Interfaces layout of drive board
The functions of the interfaces are listed in the table.
Table 5-58 Interfaces on the driver board
Connector Function Number of
pins Note
J1 FPGA configuration interface
(AS mode) 10
Used when FPGA first
downloading
J2 FPGA Debug Interface 10 Not used on complete
device
J3 Interface with the motherboard 96
Control interface of
valves, pumps and
heaters
J4 Interface with the motherboard 96 Sensor detection and
communication interface
J5 Interface with the motherboard 96 Motor control interface
J8 Communication Debug
Interface 4
Not used on complete
device
J9 MCU Debug Interface 10 Used when MCU first
downloading
J10 Communication Debug
Interface 4
Not used on complete
device
Hardware System
5-63
Definition of J1
J1 is the FPGA configuration interface (AS mode).
Table 5-59 Definition of the interface for FPGA configuration
Pin Name Pin Name
1 F_DCLK 2 AGND
3 F_CONFDONE 4 VDD
5 F_CONFIG 6 F_NCE
7 F_DATA 8 F_NCSC
9 F_ASDO 10 GND
Definition of J2
J2 is the FPGA debug interface (JTAG mode).
Definition of J3
J3 connects to the PB3 connector of the motherboard, and is the interface used for valves,
pumps and heaters control.
Table 5-60 Valve/pump interface- Definition of J3
Pin Definition Note
1
2
3
4
5
6
7
8
9
10
11
12
13
PGND Power ground
14
15
16
17
18
12V0 12V power supply
19 SV1 Control signal for Valve 1
20 SV4 Control signal for Valve 4
Hardware System
5-64
21 SV7 Control signal for Valve 7
22 SV10 Control signal for Valve 10
23 SV13 Control signal for Valve 13
24 SV16 Control signal for Valve 16
25 SV19 Control signal for Valve 19
26 SV22 Control signal for Valve 22
27 SV25 Control signal for Valve 25
28 SV28 Control signal for Valve 28
29 SV31 Control signal for Valve 31
30 SV34 Control signal for Valve 34
31 SV37 Control signal for Valve 37
32 SV40 Control signal for Valve 40
33 M7-A Connecting end A of A-phase winding of the sample probe
X motor (SM 7)
34 M7-AN Connecting end AN of A-phase winding of the sample probe
X motor (SM 7)
35 M7-B Connecting end B of B-phase winding of the sample probe
X motor (SM 7)
36 M7-BN Connecting end BN of B-phase winding of the sample probe
X motor (SM 7)
37 M8-A Connecting end A of A-phase winding of the sample probe
Y motor (SM 8)
38 M8-AN Connecting end AN of A-phase winding of the sample probe
Y motor (SM 8)
39 M8-B Connecting end B of B-phase winding of the sample probe
Y motor (SM 8)
40 M8-BN Connecting end BN of B-phase winding of the sample probe
Y motor (SM 8)
41 HT1-DRV Control end of the reaction bath heater
42 HT2-DRV Control end of the optical system heater
43 PGND Power ground
44 PUMP7 Control signal for Pump 7 (Reserved. No output)
45 PUMP6 Control signal for Pump 6 (Reserved. No output)
46 PUMP5 Control signal for Pump 5 (Reserved. No output)
47 PUMP4 Control signal for Pump 4 (Reserved. No output)
48 PUMP3 Control signal for the Vacuum/Waste Pump (P3)
49 PUMP2 Control signal for the Vacuum/Waste Pump (P2)
50 PUMP1 Control signal for pressure pump (P1)
51 Sv2 Control signal for Valve 2
Hardware System
5-65
52 SV5 Control signal for Valve 5
53 SV8 Control signal for Valve 8
54 SV11 Control signal for Valve 11
55 SV14 Control signal for Valve 14
56 SV17 Control signal for Valve 17
57 SV20 Control signal for Valve 20
58 SV23 Control signal for Valve 23
59 SV26 Control signal for Valve 26
60 SV29 Control signal for Valve 29
61 SV32 Control signal for Valve 32
62 SV35 Control signal for Valve 35
63 SV38 Control signal for Valve 38
64 NC Empty
65 M7-A Connecting end A of A-phase winding of the sample probe
X motor (SM 7)
66 M7-AN Connecting end AN of A-phase winding of the sample probe
X motor (SM 7)
67 M7-B Connecting end B of B-phase winding of the sample probe
X motor (SM 7)
68 M7-BN Connecting end BN of B-phase winding of the sample probe
X motor (SM 7)
69 M8-A Connecting end A of A-phase winding of the sample probe
Y motor (SM 8)
70 M8-AN Connecting end AN of A-phase winding of the sample probe
Y motor (SM 8)
71 M8-B Connecting end B of B-phase winding of the sample probe
Y motor (SM 8)
72 M8-BN Connecting end BN of B-phase winding of the sample probe
Y motor (SM 8)
73 HT1-DRV Control end of the reaction bath heater
74 HT2-DRV Control end of the optical system heater
75 PGND Power ground
76 PUMP7 Control signal for Pump 7 (Reserved. No output)
77 PUMP6 Control signal for Pump 6 (Reserved. No output)
78 PUMP5 Control signal for Pump 5 (Reserved. No output)
79 PUMP4 Control signal for pinch valve V39
80 PUMP3 Control signal for the Vacuum/Waste Pump (P3)
81 PUMP2 Control signal for the Vacuum/Waste Pump (P2)
82 PUMP1 Control signal for pressure pump (P1)
Hardware System
5-66
83 SV3 Control signal for Valve 3
84 SV6 Control signal for Valve 6
85 SV9 Control signal for Valve 9
86 SV12 Control signal for Valve 12
87 SV15 Control signal for Valve 15
88 SV18 Control signal for Valve 18
89 SV21 Control signal for Valve 21
90 SV24 Control signal for Valve 24
91 SV27 Control signal for Valve 27
92 SV30 Control signal for Valve 30
93 SV33 Control signal for Valve 33
94 SV36 Control signal for Valve 36
95 SV39 Control signal for Valve 39
96 NC Empty
Definition of J4
J4 connects to the PB1 connector of the motherboard, and is the interface used for sensor
detection and communication.
Table 5-61 Sensors/Communication interfaces-Definition of J4
Pin Name Note
1 PRTO_T1 Temperature sensor signal end of heater 1 (reaction bath)
2 PRTO_T2 Temperature sensor signal end of heater 2 (optical system)
3 PRESSURE_1ST_A Input signal of pressure sensor 1 (Pressure)
4
5
6
7
8
9
10
11
GND GND
12
13
14
15
16
VCC 5V power supply
17 GND GND
18 LIQ_SIN1 Input signal of LEO1 lyse liquid-level sensor
Hardware System
5-67
19 LIQ_SIN3 Input signal of LH lyse liquid-level sensor
20 LIQ_SIN5 Input signal of liquid-level sensor5 (reserved)
21 LIQ_SIN7 Input signal of waste liquid-level sensor
22 LS_CTRL Control signal for liquid-level sampling
23 GND Digital ground
24 VC_SIN17 Sending end power signal of position sensor 17 (reserved)
25 VC_SIN18 Sending end power signal of position sensor 18 (reserved)
26 VC_SIN19 Sending end power signal of position sensor 19 (reserved)
27 VC_SIN20 Sending end power signal of position sensor 20 (reserved)
28 VC_SIN1 Sending end power signal of sampling syringe initial position
sensor (OP1)
29 M_SIN2 Receiving end feedback signal of sample injection syringe
initial position sensor (OP 2)
30 VC_SIN4 Sending end power signal of lyse syringe initial position
sensor (OP 4)
31 M_SIN5 Receiving end feedback signal of diluent syringe initial position
sensor (OP 5)
32 VC_SIN7 Sending end power signal of position sensor 7 (reserved)
33 PRTO_REF Reference end of heater temperature sensor signal
34 PRTO_REF Reference end of heater temperature sensor signal
35 PRESSURE_2ST_A Input signal of pressure sensor 2 (vacuum)
36
37
38
39
40
41
42
43
GND Digital ground
44
45
46
47
48
VCC 5V power supply
49 GND Digital ground
50 LIQ_SIN2 Input signal of LEO2 lyse liquid-level sensor
51 LIQ_SIN4 Input signal of cleanser liquid-level sensor
52 LIQ_SIN6 Input signal of diluent liquid-level sensor
Hardware System
5-68
53 LIQ_SIN8 Input signal of liquid-level sensor 8 (reserved)
54
55 GND Digital ground
56 M_SIN17 Receiving end feedback signal of position sensor 17
(reserved)
57 M_SIN18 Receiving end feedback signal of position sensor 18
(reserved)
58 M_SIN19 Receiving end feedback signal of position sensor 19
(reserved)
59 M_SIN20 Receiving end feedback signal of position sensor 20
(reserved)
60 M_SIN1 Receiving end feedback signal of sampling syringe initial
position sensor (OP 1)
61 VC_SIN3 Sending end power signal of sheath syringe initial position
sensor (OP 3)
62 M_SIN4 Receiving end feedback signal of lyse syringe initial position
sensor (OP 4)
63 VC_SIN6 Sending end power signal of relieve valve position sensor (OP
6)
64 M_SIN7 Receiving end feedback signal of position sensor 7 (reserved)
65 AMBINET_T_A Input signal of ambient temperature sensor
66
67
68
69
70
71
72
73
74
75
GND Digital ground
76
77
78
79
80
VCC 5V power supply
81 GND Digital ground
82 TMS TMS signal of JTAG interface
83 TCK Clock signal of JTAG interface
Hardware System
5-69
84 TDI Data input pin of JTAG interface
85 TDO Data output pin of JTAG interface
86
87 GND Digital ground
88 UP_TX Serial port sending data signal (UART0)
89 UP_RX Serial port receiving data signal (UART0)
90
91 GND Digital ground
92 VC_SIN2 Sending end power signal of sample injection syringe initial
position sensor (OP 2)
93 M_SIN3 Receiving end feedback signal of sheath syringe initial
position sensor (OP 3)
94 VC_SIN5 Sending end power signal of diluent syringe initial position
sensor (OP 5)
95 M_SIN6 Receiving end feedback signal of relieve valve position sensor
(OP 6)
96 VC_SIN8 Sending end power signal of position sensor 8 (reserved)
Definition of J5
J5 connects to the PB2 connector of the motherboard, and is the interface used for the
stepping motor control and the position sensor signal detection.
Table 5-62 Stepping motor control interface-definition of J5
Pin Definition Note
1 M_SIN8 Receiving end feedback signal of position sensor 8 (reserved)
2 VC_SEN10 Sending end power signal of autoloading position sensor (OP 10)
3 M_SIN11 Receiving end feedback signal of position sensor 11 (reserved)
4 VC_SEN13 Sending end power signal of WBC bath position sensor (OP 13)
5 M_SIN14 Receiving end feedback signal of RBC bath position sensor (OP
14)
6 VC_SEN16 Sending end power signal of sample probe Y lower position sensor
(OP 16)
7 GND Digital ground
8 NC Empty
9
10
11
12
13
24V0 24V power supply
Hardware System
5-70
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
PGND Power ground
33 VC_SEN9 Sending end power signal of close tube position sensor (OP 9)
34 M_SIN10 Receiving end feedback signal of autoloading position sensor (OP
10)
35 VC_SEN12 Sending end power signal of DIFF bath position sensor (OP 12)
36 M_SIN13 Receiving end feedback signal of WBC bath position sensor (OP
13)
37 VC_SEN15 Sending end power signal of sample probe Y upper position sensor
(OP 15)
38 M_SIN16 Receiving end feedback signal of sample probe Y lower position
sensor (OP 16)
39 GND Digital ground
40 NC Empty
41 M1_A Connecting end A of A-phase winding of the sample syringe motor
(SM 1)
42 M1_AN Connecting end AN of A-phase winding of the sample syringe
motor (SM 1)
43 M1_B Connecting end B of B-phase winding of the sample syringe motor
(SM 1)
44 M1_BN Connecting end BN of B-phase winding of the sample syringe
motor (SM 1)
Hardware System
5-71
45 M2_A Connecting end A of A-phase winding of the sample injection
syringe motor (SM 2)
46 M2_AN Connecting end AN of A-phase winding of the sample injection
syringe motor (SM 2)
47 M2_B Connecting end B of B-phase winding of the sample injection
syringe motor (SM 2)
48 M2_BN Connecting end BN of B-phase winding of the sample injection
syringe motor (SM 2)
49 M3_A Connecting end A of A-phase winding of the sheath syringe motor
(SM 3)
50 M3_AN Connecting end AN of A-phase winding of the sheath syringe motor
(SM 3)
51 M3_B Connecting end B of B-phase winding of the sheath syringe motor
(SM 3)
52 M3_BN Connecting end BN of B-phase winding of the sheath syringe motor
(SM 3)
53 M4_A Connecting end A of A-phase winding of the lyse syringe motor (SM
4)
54 M4_AN Connecting end AN of A-phase winding of the lyse syringe motor
(SM 4)
55 M4_B Connecting end B of B-phase winding of the lyse syringe motor
(SM 4)
56 M4_BN Connecting end BN of B-phase winding of the lyse syringe motor
(SM 4)
57 M5_A Connecting end A of A-phase winding of the diluent syringe motor
(SM 5)
58 M5_AN Connecting end AN of A-phase winding of the diluent syringe motor
(SM 5)
59 M5_B Connecting end B of B-phase winding of the diluent syringe motor
(SM 5)
60 M5_BN Connecting end BN of B-phase winding of the diluent syringe motor
(SM 5)
61 M6_A Connecting end A of A-phase winding of the reserved motor (SM 6,
reserved)
62 M6_AN Connecting end AN of A-phase winding of the reserved motor (SM
6, reserved)
63 M6_B Connecting end B of B-phase winding of the reserved motor (SM 6,
reserved)
64 M6_BN Connecting end BN of B-phase winding of the reserved motor (SM
Hardware System
5-72
6, reserved)
65 M_SIN9 Receiving end feedback signal of close tube position sensor (OP 9)
66 VC_SEN11 Sending end power signal of position sensor 11 (reserved)
67 M_SIN12 Receiving end feedback signal of DIFF bath position sensor (OP
12)
68 VC_SEN14 Sending end power signal of RBC bath position sensor (OP 14)
69 M_SIN15 Receiving end feedback signal of sample probe Y upper position
sensor (OP 15)
70 DCM DC motor control signal (reserved)
71 GND Digital ground
72 NC Empty
73 M1_A Connecting end A of A-phase winding of the sample syringe motor
(SM 1)
74 M1_AN Connecting end AN of A-phase winding of the sample syringe
motor (SM 1)
75 M1_B Connecting end B of B-phase winding of the sample syringe motor
(SM 1)
76 M1_BN Connecting end BN of B-phase winding of the sample syringe
motor (SM 1)
77 M2_A Connecting end A of A-phase winding of the sample injection
syringe motor (SM 2)
78 M2_AN Connecting end AN of A-phase winding of the sample injection
syringe motor (SM 2)
79 M2_B Connecting end B of B-phase winding of the sample injection
syringe motor (SM 2)
80 M2_BN Connecting end BN of B-phase winding of the sample injection
syringe motor (SM 2)
81 M3_A Connecting end A of A-phase winding of the sheath syringe motor
(SM 3)
82 M3_AN Connecting end AN of A-phase winding of the sheath syringe motor
(SM 3)
83 M3_B Connecting end B of B-phase winding of the sheath syringe motor
(SM 3)
84 M3_BN Connecting end BN of B-phase winding of the sheath syringe motor
(SM 3)
85 M4_A Connecting end A of A-phase winding of the lyse syringe motor (SM
4)
86 M4_AN Connecting end AN of A-phase winding of the lyse syringe motor
(SM 4)
Hardware System
5-73
87 M4_B Connecting end B of B-phase winding of the lyse syringe motor
(SM 4)
88 M4_BN Connecting end BN of B-phase winding of the lyse syringe motor
(SM 4)
89 M5_A Connecting end A of A-phase winding of the diluent syringe motor
(SM 5)
90 M5_AN Connecting end AN of A-phase winding of the diluent syringe motor
(SM 5)
91 M5_B Connecting end B of B-phase winding of the diluent syringe motor
(SM 5)
92 M5_BN Connecting end BN of B-phase winding of the diluent syringe motor
(SM 5)
93 M6_A Connecting end A of A-phase winding of the reserved motor (SM 6,
reserved)
94 M6_AN Connecting end AN of A-phase winding of the reserved motor (SM
6, reserved)
95 M6_B Connecting end B of B-phase winding of the reserved motor (SM 6,
reserved)
96 M6_BN Connecting end BN of B-phase winding of the reserved motor (SM
6, reserved)
Definition of J8
J8 is the interface used for communication debugging.
Definition of J9
J9 is the MCU debug interface (JTAG mode).
Table 5-63 FPGA debug interface-definition of J9
Pin Name Pin Name
1 VDD 2 GND
3 GND 4 TCK
5 TMS 6 TDO
7 TDI 8 NC
9 GND 10 NC
Definition of J10
J10 is the interface used for communication debugging.
Hardware System
5-74
Assembly drawing
Figure 5-15 Assembly drawing of the drive board
Hardware System
5-75
5.3.3 Adjustment and Test Points
LED function definition
Table 5-64 LED definition
Indicator Note
D1 FPGA configuration status indicator. It turns on during FPGA configuration.
D2 FPGA working status indicator. It flashes every 1s when FPGA works
normally.
D49 24V power supply (24V0) status indicator. It turns on when the 24V power
supply is normal.
D50 3.3V power supply (VDD) status indicator. It turns on when the 3.3V power
supply is normal.
D51 12V power supply (12V0) status indicator. It turns on when the 12V power
supply is normal.
D52 5V power supply (VDD) status indicator. It turns on when the 5V analog
power supply is normal.
D65 MCU working status indicator. It flashes every 1s when MCU works normally.
D66 Working status indicator of optical system heater. It turns on during heating.
D67 Working status indicator of reaction bath (DIFF bath). It turns on during
heating.
Function definition of test points
Table 5-65 Test point definition
No. Name Pin Note
1 TP1 PIN26 of U2 FPGA configuration signals F_NCONFIG
2 TP2 PIN28 of U2 FPGA 24.5454MHz clock input signals
FPGA_CLK
3 TP3 PIN228 of U2 Address latch enabling signals of MCU
external bus ALE
4 TP4 PIN50 of U2 Read signals of MCU external bus /RD
5 TP5 PIN20 of U2 FPGA interrupt request signals FPGA_INT
6 TP6 PIN73 of U2 Write signals of MCU external bus /WR
Hardware System
5-76
7 TP7 PIN11 of U2 FPGA reset signals F_RST
8 TP8 1V5 Voltage of +1.5VDC power supply
(1.5±0.075V)
9 TP9 SIO_IN Data input signals of MCU valve control chip
10 TP10 SIO_CLK Clock signals of MCU valve control chip
11 TP11 SIO_EN Enabling signals of MCU valve control chip
12 TP12 SIO_LATCH Data latch signals of MCU valve control chip
13 TP13 PIN17 of U10 Stepping pulse signals of sample probe X
motor STEP7
14 TP14 PIN15 of U9 Stepping pulse signals of sample probe Y
motor STEP8
15 TP16 PIN15 of U21 Stepping pulse signals of sample syringe
motor STEP1
16 TP17 PIN13 of U20 Stepping pulse signals of sample injection
syringe motor STEP2
17 TP18 PIN17 of U20 Stepping pulse signals of sheath syringe
motor STEP3
18 TP19 PIN17 of U25 Stepping pulse signals of lyse syringe motor
STEP4
19 TP20 PIN16 of U26 Stepping pulse signals of diluent syringe
motor STEP5
20 TP21 PIN11 of U10 Stepping pulse signals of reserved motor
STEP6 (not used)
21 TP22 VDD Voltage of +3.3VDC power supply
(3.3±0.15V)
22 TP23
23 TP24 GND Digital ground
24 TP25 1V5 Voltage of +1.5VDC power supply
(1.5±0.075V)
25 TP26 PIN1 of U32 Ambient temperature detection signals
(voltage) AMBINET_T
26 TP27 ADC_VCC Detection signals of VCC(+5VDC) after
voltage-dividing (2.00±0.15V)
27 TP28 ADC_12V Detection signals of 12V0(+12VDC) after
voltage-dividing (2.13±0.15V)
28 TP29 ADC_24V Detection signals of 24V0(+24VDC) after
voltage-dividing (2.23±0.15V)
29 TP30 PIN1 of U33 Pressure chamber detection signals
Hardware System
5-77
(voltage) PRESSURE_1ST
30 TP31 PIN7 of U33 Vacuum chamber detection signals
(voltage) PRESSURE_2ST
31 TP34 PIN2 of U37 Reference voltage of MCU analog-to-digital
module REFIN(2.50±0.006V)
34 TP35 PIN100 of U35 Voltage of MCU digital-to-analog output
channel 0
35 TP36 PIN26 of U35 MCU 25.5454MHz clock input signals
CLKIN
36 TP37 PIN99 of U35 Voltage of MCU digital-to-analog output
channel 1
37 TP38 PIN57 of U35 MCU system clock output signals SYSCLK
38 TP39 PIN6 of U45 Amplifier output signals of the temperature
collection module TEMPER_AD
39 TP40 PRTD_REF Basic voltage of the amplifier input signals
of the temperature collection module
40 TP41 PIN2 of U45 Reference voltage of temperature collection
module amplifier
41 TP42 PIN3 of U45 Input voltage of temperature collection
module amplifier
42 TP43 PIN8 of U22 Driving reference voltage of sample syringe
motor VREF1
43 TP44 PIN8 of U23 Driving reference voltage of sample
injection syringe motor VREF2
44 TP45 PIN8 of U24 Driving reference voltage of sheath syringe
motor VREF3
45 TP46 PIN8 of U29 Driving reference voltage of lyse syringe
motor VREF4
46 TP47 PIN8 of U28 Driving reference voltage of diluent syringe
motor VREF5
47 TP48 PIN8 of U27 Driving reference voltage of reserved motor
VREF6 (not used)
48 TP49 PIN8 of U11 Driving reference voltage of sample probe X
motor VREF7
49 TP50 PIN8 of U13 Driving reference voltage of sample probe Y
motor VREF8
Key definition
The reset switch S1 on the drive board is used to reset the drive board manually. Press the
Hardware System
5-78
reset switch to reset the MCU and FPGA.
5.3.4 Disassembly and assembly method
Purpose
The drive board is a key component. Please maintain or replace the drive board immediately if
any error or damage occurs. Please follow the procedures to have the drive board replaced.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Disassembly
Drive board disassembly consists of the following two main procedures:
Remove the drive board from the main unit:
1. Shutdown the analyzer and cut off the power supply;
2. Open the left door;
3. Use your hand or cross-headed screwdriver to remove the two hand-tightened screws
from the bracket.
4. Use both of your hands to flip outward the two latches on both sides of the drive board,
and then draw the drive board out.
Hardware System
5-79
Figure 5-16 Assembly drawing of drive board-1
1---Hand-tightened screws 2---Drive board
Further disassembly of the drive board
1. Use the 107 cross-headed screwdriver to remove the 5 cross-headed panhead screws
(M3X6) from the drive board.
2. Remove the drive board from the bracket.
Figure 5-17 Assembly drawing of drive board-2
1---Drive board 2---Bracket
3---Cross-recessed panhead screw
M3x6 (with washer)
Hardware System
5-80
Assembly
Please assemble the data board as per the above-mentioned steps in the reverse order.
Wear antistatic gloves before maintaining or removing the board.
Be sure to shut down the analyzer and cut off the power supply before disassembling/assembling the board.
Verification
1. Check and make sure that all the components are properly connected to the drive board.
2. First ensure the power supply is cut off, then assemble the drive board and motherboard
properly, then power on the analyzer.
3. Check if the indicators on the drive board are in the following status: 1) D49, D50 and D51
turn on; 2) D2 and D65 flash. You can also check the indicators in the “self-test” screen of
the analyzer. If the self-test is passed, it means the driver board is replaced successfully;
otherwise, you should troubleshoot the board.
4. Readjust the sample probe position as per Section 6.10.1 and save the result.
5.3.5 Troubleshooting
Error analysis and solutions
Errors related to the drive board are identified and handled by the host control software in the
form of error codes. Note that only some of the error codes reported to the host control
software from the dive board can be handled automatically, while some need to be judged with
the help of the users or service engineers. When error is reported, the error No. and error
names are provided by host control software. The recommended solutions may appear on the
screen as the software help information.
When you check for errors, first ensure the board is well fixed and all the error-related
connections are properly connected. For example, when “Sample probe action failed” is
reported, you should check: 1) whether the sample probe assembly is properly connected to
the board; 2) whether the connecters are well connected (in case of bad connection, shut
down the analyzer and cut off the power supply, and then re-plug the connecters); 3)whether
the connection inside the sample probe assembly is proper; 4) whether the symbol on the
sensor connecting line corresponds to the sensor position; 5) whether the symbol on the motor
connecting line corresponds to the motor position (See 6.6.1 for how to disassemble the
sample probe assembly if necessary).
Hardware System
5-81
Only when you make sure that nothing is wrong with the connections can you remove the error
according to the recommended action in the following table. If the error still exists after the
error causes are found and new components are replaced, you need to replace the drive
board.
The error codes, error names and recommended solutions are listed in the following table.
Table 5-66 Troubleshooting errors with the drive board
Error code Error name Error feature Recommended
action
0100 Valve output
failed
FPGA does not allow the MCU write
data into the valve-pump data
register within 50ms, i.e. the serial
output status indication bit in the
register (which is used to judge the
status of valves and pumps of
FPGA) is always 1. It might be FPGA
error or MCU read register error.
Replace drive board
0101
Valve NO. out
of limit (greater
than 40 or
equal to 0)
Download Valve NO. out of limit
(greater than 40 or equal to 0)
Communication
abnormal. Please try
again.
0201
X motor failed
to move to
target position
X motor does not move to the
pre-set target position after action
1. X motor error, action
is interfered.
2. X motor target
position photocoupler
error.
0202
X motor failed
to move to
home position
after
initialization
X motor is not at the home position
after initialization
1. X motor error, action
is interfered.
2. X motor initial
position photocoupler
error.
0203
X motor failed
to leave home
position after
initialization
X motor failed to leave home
position when initializing
1. X motor error, action
is interfered.
2. X motor initial
position photocoupler
error.
0204
Adjusting X
motor to target
position failed
X motor does not move to target
position after the adjusting command
is executed.
1. X motor error, action
is interfered.
2. X motor target
position photocoupler
Hardware System
5-82
error.
0205
Adjusting X
motor back to
home position
failed
X motor does not back to home
position after the adjusting command
is executed
1. X motor error, action
is interfered.
2. X motor initial
position photocoupler
error.
0206
Adjusting X
motor to leave
home position
failed
X motor does not leave home
position after the adjusting command
is executed
1. X motor error, action
is interfered.
2. X motor initial
position photocoupler
error.
0207
X motor
adjustment
error
X motor adjustment is out of limit [1,
9]
Re-send the X motor
adjusting command
0208
X motor
adjusting steps
out of limit
X motor adjusting steps at the target
position are out of limit [-20, 20]
Adjust the X motor
again
0209
X motor
adjusting end
position error
The position in the X motor adjusting
end command disagrees with the
pre-send adjusting position
Re-send the X motor
adjusting command
0211
Y motor failed
to move to
target position
Y motor does not move to the
pre-set target position after action
1. Y motor error, action
is interfered.
2. Y motor target
position photocoupler
error.
0212
Y motor failed
to initialize to
upper position
Y motor is not at the upper position
after initialization
1. Y motor error, action
is interfered.
2. Y motor upper
position photocoupler
error.
0213
Y motor failed
to initialize to
lower position
Y motor does not move to lower
position when initializing
1. Y motor error, action
is interfered.
2. Y motor lower
position photocoupler
error.
0214
Adjusting Y
motor to target
position failed
Y motor does not move to target
position after the adjusting command
is executed.
1. Y motor error, action
is interfered.
2. Y motor target
position photocoupler
error.
Hardware System
5-83
0215
Adjusting Y
motor to upper
position failed
Y motor does not back to upper
position after the adjusting command
is executed
1. Y motor error, action
is interfered.
2. Y motor upper
position photocoupler
error.
0216
Adjusting Y
motor to leave
upper position
failed
Y motor does not leave upper
position after the adjusting command
is executed
1. Y motor error, action
is interfered.
2. Y motor upper
position photocoupler
error.
0217
Y motor
adjustment
error
Y motor adjustment is out of limit [1,
6]
Re-send the Y motor
adjusting command
0218
Y motor
adjusting steps
out of limit
Y motor adjusting steps at the target
position are out of limit [-63,63]
Adjust the Y motor
again
0219
Y motor
adjusting end
position error
The position in the Y motor adjusting
end command disagrees with the
pre-send adjusting position
Re-send the Y motor
adjusting command
0220
Sample probe
assembly is
working
X motor or Y motor action is not
finished.
Send the sample
probe command after
the sample probe
action is finished
0221 Cannot adjust
sample probe
Can not send the command to move
the sample probe
upward/downward/leftward/rightward
by one step before sending the
adjusting command.
You can adjust the
sample probe only
after the adjusting
command is sent.
0222 X motor action
overtime
The action time is longer than the
download time.
0223 X motor action
forbidden
X motor is not at the photocoupler
position or Y motor is not at the
upper photocoupler position
Move the Y motor to
upper photocoupler
position
0224
X motor does
not match
current position
X motor current position does not
match the photocoupler position Initialize the X motor
0225 Y motor action
overtime
The action time is longer than the
download time.
0226 X motor action
forbidden
X motor is not at the photocoupler
position
Move the X motor to
the photocoupler
Hardware System
5-84
position
0227
Y motor does
not match
current position
Y motor current position does not
match the photocoupler position Initialize the Y motor
0228
Y motor failed
to pierce to
lower position
Y motor is not at the lower position
after piercing Pierce again
0229
Y motor failed
to pierce to
upper position
Y motor failed to back to upper
position when piercing Pierce again
0230
X motor
photocoupler
error
X motor photocoupler might be
damaged
Replace X motor
photocoupler
0231
Sample probe
RBC bath
photocoupler
error
RBC bath photocoupler might be
damaged
Replace RBC bath
photocoupler
0232
Sample probe
WBC bath
photocoupler
error
WBC bath photocoupler might be
damaged
Replace WBC bath
photocoupler
0233
Sample probe
DIFF bath
photocoupler
error
DIFF bath photocoupler might be
damaged
Replace DIFF bath
photocoupler
0234
Y motor
photocoupler
error
Y motor photocoupler might be
damaged
Replace X motor
photocoupler
0235
Sample probe
upper
photocoupler
error
Upper photocoupler might be
damaged
Replace the upper
photocoupler
0236
Sample probe
lower
photocoupler
error
Lower photocoupler might be
damaged
Replace the lower
photocoupler
0300
Sampling
syringe is
working
Sampling syringe does not finish the
previous action when it is requested
for initializing, resetting, aspirating or
dispensing.
Command is sent early
while the sampling
syringe is still working.
Hardware System
5-85
0301
Sampling
syringe
photocoupler
abnormal
The detected photocoupler signals
are without any change before and
after action.
1. Sampling syringe
photocoupler error (if
sampling syringe
motor moves)
2. Sampling syringe
motor error (if
sampling syringe
motor dose not move)
0302
Sampling
syringe
aspirating or
dispensing
action failed 1
After aspirating or dispensing, the
sampling syringe should be at the
home position, but the position
detected by the photocoupler
disagrees
1. Sampling syringe
photocoupler error (if
the sampling syringe
motor is at the home
position)
2. Sampling syringe
motor error (if the
sampling syringe
motor is not at the
home position)
0303
Sampling
syringe
aspirating or
dispensing
action failed 2
After aspirating or dispensing, the
sampling syringe should not be at
the home position, but the position
detected by the photocoupler
disagrees
1. Sampling syringe
photocoupler error (if
the sample syringe
motor is not at the
home position)
2. Sampling syringe
motor error (if the
sampling syringe
motor is at the home
position)
0304
Sampling
syringe
aspirating or
dispensing
action
forbidden
Before aspirating or dispensing, the
sampling syringe should be at the
home position, but the actual
position detected by the
photocoupler disagrees
Sampling syringe
photocoupler error (if
the sampling syringe
motor is at the home
position)
0305
Sampling
syringe
aspirating or
dispensing
action
forbidden 2
Before aspirating or dispensing, the
sampling syringe should not be at
the home position, but the actual
position detected by the
photocoupler disagrees
Sampling syringe
photocoupler error (if
the sampling syringe
motor is not at the
home position)
Hardware System
5-86
0306
Sample syringe
aspirate volume
over range
Requested volume + aspirated
volume > maximum aspirate volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase.
0307
Sample syringe
dispense
volume over
range
Requested volume > aspirated
volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase and
note the volume add
up error.
0308
Sampling
syringe action
overtime
The action time is longer than the
download time.
0310
Sample
injection
syringe is
working
Sample injection syringe does not
finish the previous action when it is
requested for initializing, resetting,
aspirating or dispensing.
Command is sent early
while the sample
injection syringe is still
working.
0311
Sample
injection
syringe
photocoupler
abnormal
The detected photocoupler signals
are without any change before and
after action.
1. Sample injection
syringe photocoupler
error (if sample
injection syringe motor
moves)
2. Sample injection
syringe motor error (if
sample injection
syringe motor does not
move)
0312
Sample
injection
syringe
aspirating or
dispensing
After aspirating or dispensing, the
syringe should be at the home
position, but the position detected by
the photocoupler disagrees
1. Sample injection
syringe photocoupler
error (if sample
injection syringe is at
the home position)
Hardware System
5-87
action failed 1 2. Sample injection
syringe motor error (if
sample injection
syringe is not at the
home position)
0313
Sample
injection
syringe
aspirating or
dispensing
action failed 2
After aspirating or dispensing, the
syringe should not be at the home
position, but the position detected by
the photocoupler disagrees
1. Sample injection
syringe photocoupler
error (if sample
injection syringe is not
at the home position)
2. Sample injection
syringe motor error (if
sample injection
syringe is at the home
position)
0314
Sample
injection
syringe
aspirating or
dispensing
action
forbidden 1
Before aspirating or dispensing, the
syringe should be at the home
position, but the position detected by
the photocoupler disagrees
Sample injection
syringe photocoupler
error (if sample
injection syringe is at
the home position)
0315
Sample
injection
syringe
aspirating or
dispensing
action
forbidden 2
Before aspirating or dispensing, the
syringe should not be at the home
position, but the position detected by
the photocoupler disagrees
Sample injection
syringe photocoupler
error (if sample
injection syringe is not
at the home position)
0316
Sample
injection
syringe aspirate
volume over
range
Requested volume + aspirated
volume > maximum aspirate volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase.
0317
Sample
injection
syringe
Requested volume > aspirated
volume
Check and make sure
nothing is wrong with
the sampling syringe
Hardware System
5-88
dispense
volume over
range
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase and
note the volume add
up error.
0318
Sample
injection
syringe action
overtime
The action time is longer than the
download time.
0320
sheath fluid
syringe is
working
Sheath fluid syringe does not finish
the previous action when it is
requested for initializing, resetting,
aspirating or dispensing.
Command is sent early
while the sheath fluid
syringe is still working.
0321
Sheath fluid
syringe
photocoupler
abnormal
The detected photocoupler signals
are without any change before and
after action.
1. Sheath fluid syringe
photocoupler error (if
sheath fluid syringe
motor moves)
2. Sheath fluid syringe
motor error (if sheath
fluid syringe motor
dose not move)
0322
Sheath fluid
syringe
aspirating or
dispensing
action failed 1
After aspirating or dispensing, the
syringe should be at the home
position, but the position detected by
the photocoupler disagrees
1. Sheath fluid syringe
photocoupler error (if
sheath fluid syringe
motor is at the home
position)
2. Sheath fluid syringe
motor error (if sheath
fluid syringe motor is
not at the home
position)
0323
Sheath fluid
syringe
aspirating or
dispensing
action failed 2
After aspirating or dispensing, the
syringe should not be at the home
position, but the position detected by
the photocoupler disagrees
1. Sheath fluid syringe
photocoupler error (if
sheath fluid syringe
motor is not at the
home position)
2. Sheath fluid syringe
motor error (if sheath
Hardware System
5-89
fluid syringe motor is
at the home position)
0324
Sheath fluid
syringe
aspirating or
dispensing
action
forbidden 1
Before aspirating or dispensing, the
syringe should be at the home
position, but the position detected by
the photocoupler disagrees
Sheath fluid syringe
photocoupler error (if
sample syringe motor
is at the home position)
0325
Sheath fluid
syringe
aspirating or
dispensing
action
forbidden 2
Before aspirating or dispensing, the
syringe should not be at the home
position, but the position detected by
the photocoupler disagrees
Sheath fluid syringe
photocoupler error (if
sample syringe motor
is not at the home
position)
0326
Sheath fluid
syringe aspirate
volume over
range
Requested volume + aspirated
volume > maximum aspirate volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase.
0327
Sheath fluid
syringe
dispense
volume over
range
Requested volume > aspirated
volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase and
note the volume add
up error.
0328
Sheath fluid
syringe action
overtime
The action time is longer than the
download time.
0330 Lyse syringe is
working
Lyse syringe does not finish the
previous action when it is requested
for initializing, resetting, aspirating or
dispensing.
Command is sent early
while the lyse syringe
is still working.
0331 Lyse syringe The detected photocoupler signals 1. Lyse syringe
Hardware System
5-90
photocoupler
abnormal
are without any change before and
after action.
photocoupler error (if
the lyse syringe motor
moves)
2. Lyse syringe motor
error (if the lyse
syringe motor does not
move)
0332
Lyse syringe
aspirating or
dispensing
action failed 1
After aspirating or dispensing, the
syringe should be at the home
position, but the position detected by
the photocoupler disagrees
1. Lyse syringe
photocoupler error (if
the lyse syringe motor
is at the home position)
2. Lyse syringe motor
error (if the lyse
syringe motor is not at
the home position)
0333
Lyse syringe
aspirating or
dispensing
action failed 2
After aspirating or dispensing, the
syringe should not be at the home
position, but the position detected by
the photocoupler disagrees
1. Lyse syringe
photocoupler error (if
the lyse syringe motor
is not at the home
position)
2. Lyse syringe motor
error (if the lyse
syringe motor is at the
home position)
0334
Lyse syringe
aspirating or
dispensing
action
forbidden 1
Before aspirating or dispensing, the
syringe should be at the home
position, but the position detected by
the photocoupler disagrees
Lyse syringe
photocoupler error (if
the lyse syringe motor
is at the home position)
0335
Lyse syringe
aspirating or
dispensing
action
forbidden 2
Before aspirating or dispensing, the
syringe should not be at the home
position, but the position detected by
the photocoupler disagrees
Lyse syringe
photocoupler error (if
the lyse syringe motor
is not at the home
position)
0336
Lyse syringe
aspirate volume
over range
Requested volume + aspirated
volume > maximum aspirate volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
Hardware System
5-91
overflow at the fluidics
debugging phase.
0337
Lyse syringe
dispense
volume over
range
Requested volume > aspirated
volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase and
note the volume add
up error.
0338 Lyse syringe
action overtime
The action time is longer than the
download time.
0340 Diluent syringe
is working
Diluent syringe does not finish the
previous action when it is requested
for initializing, resetting, aspirating or
dispensing.
Command is sent early
while the diluent
syringe is still working.
0341
Diluent syringe
photocoupler
abnormal
The detected photocoupler signals
are without any change before and
after action.
1. Diluent syringe
photocoupler error (if
the diluent syringe
motor moves)
2. Diluent syringe
motor error (if the
diluent syringe motor
dose not move)
0342
Diluent syringe
aspirating or
dispensing
action failed 1
After aspirating or dispensing, the
syringe should be at the home
position, but the position detected by
the photocoupler disagrees
1. Diluent syringe
photocoupler error (if
the diluent syringe
motor is at the home
position)
2. Diluent syringe
motor error (if the
diluent syringe motor is
not at the home
position)
0343
Diluent syringe
aspirating or
dispensing
action failed 2
After aspirating or dispensing, the
syringe should not be at the home
position, but the position detected by
the photocoupler disagrees
1. Diluent syringe
photocoupler error (if
the diluent syringe
motor is not at the
Hardware System
5-92
home position)
2. Diluent syringe
motor error (if the
diluent syringe motor is
at the home position)
0344
Diluent syringe
aspirating or
dispensing
action
forbidden 1
Before aspirating or dispensing, the
syringe should be at the home
position, but the position detected by
the photocoupler disagrees
Diluent syringe
photocoupler error (if
the diluent syringe
motor is at the home
position)
0345
Diluent syringe
aspirating or
dispensing
action
forbidden 2
Before aspirating or dispensing, the
syringe should not be at the home
position, but the position detected by
the photocoupler disagrees
Diluent syringe
photocoupler error (if
the diluent syringe
motor is not at the
home position)
0346
Diluent syringe
aspirate volume
over range
Requested volume + aspirated
volume > maximum aspirate volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase.
0347
Diluent syringe
dispense
volume over
range
Requested volume > aspirated
volume
Check and make sure
nothing is wrong with
the sampling syringe
aspirating and
dispensing action; then
consider programming
overflow at the fluidics
debugging phase and
note the volume add
up error.
0348 Diluent syringe
action overtime
The action time is longer than the
download time.
0390
Automatic
testing
command
forbidden
Start testing the motors before the
automatic testing command is
received.
First start the
automatic testing
command, and then
test the motors.
0401 Reaction bath The reaction bath temp. setup is out Re-set the reaction
Hardware System
5-93
temp. setup out
of limit
of the MCU pre-set range
[30 ,50 ]℃ ℃
bath temperature
0402
Optical system
temp. setup out
of limit
The optical system temp. setup is
out of the MCU pre-set range
[30 ,50 ]℃ ℃
Re-set the optical
system temperature
0403 Reaction bath
temp. error
Reaction bath temperature is out of
the set range
1. Check whether the
heater of the reaction
bath is well connected.
2. Check whether the
temperature collection
channel of the reaction
bath is normal.
0404 Optical system
temp. error
Optical system temperature is out of
the set range
1. Check whether the
heater of the optical
system is well
connected.
2. Check whether the
temperature collection
channel of the optical
system is normal.
0405
Reaction bath
temp.
compensation
error
Compensation value and ones’
complement code don’t match
Reset the reaction
bath temp.
compensation
0406
Optical system
temp.
compensation
error
Compensation value and ones’
complement code don’t match
Reset the optical
system temp.
compensation
0407
Ambient temp.
compensation
error
Compensation value and ones’
complement code don’t match
Reset the ambient
temp. compensation
0408
Reaction bath
temp.
compensation
out of limit
Reaction bath temp. compensation
out of the set range [-12.7℃, 12.7
℃]
Reset the reaction
bath temp.
compensation
0409
Optical system
temp.
compensation
out of limit
Optical system temp. compensation
out of the set range [-12.7℃, 12.7
℃]
Reset the optical
system temp.
compensation
Hardware System
5-94
040a
Ambient temp.
compensation
out of limit
Ambient temp. compensation out of
the set range [-12.7℃, 12.7℃]
Reset the ambient
temp. compensation
0501 Ambient temp.
abnormal
Ambient temperature is out of the set
range
1. The actual ambient
temperature is out of
the working range.
2. Check whether the
temperature
transducer and the
measurement circuit
work normally.
0601
Controlling
pressure
chamber
pressure
failed
The pressure in the chamber can not
reach the target range within the
specified fluidics time.
1. Check whether the
pneumatic connection
is well connected and
whether the pressure
pump works normally.
2. Check whether the
pressure sensor works
normally and whether
the pressure detection
signals of the MCU are
normal.
0602
Controlling
vacuum
chamber
vacuum failed
The vacuum in the chamber can not
reach the target range within the
specified fluidics time.
1. Check whether the
pneumatic connection
is well connected and
whether the vacuum
pump and valve 30
and 31 works normally.
2. Check whether the
pressure sensor works
normally and whether
the pressure detection
signals of the MCU are
normal.
0603
Adjusting
pressure
chamber
pressure failed
The pressure in the chamber is
being controlled. Can not send the
adjusting command
Wait and then send the
adjusting command
again after the
previous pressure
controlling is done.
0604 Adjusting The vacuum in the chamber is being Wait and then send the
Hardware System
5-95
vacuum
chamber
vacuum failed
controlled. Can not send the
adjusting command
adjusting command
again after the
previous vacuum
controlling is done.
0605
Pressure
chamber
pressure out of
limit
The pressure in the chamber is out
of the set range
1. Check whether the
pneumatic connection
is well connected and
whether the pressure
pump works normally.
2. Check whether the
pressure sensor works
normally and whether
the pressure detection
signals of the MCU are
normal.
0606
Vacuum
chamber
vacuum out of
limit
The vacuum in the chamber is out of
the set range
1. Check whether the
pneumatic connection
is well connected and
whether the vacuum
pump and valve 30
and 31 works normally.
2. Check whether the
pressure sensor works
normally and whether
the pressure detection
signals of the MCU are
normal.
0607
Pressure
chamber
pressure
control range
error
The pressure control of the pressure
chamber is of negative range.
Re-send the control
command of positive
range
0608
Vacuum
chamber
vacuum control
range error
The vacuum control of the vacuum
chamber is of positive range
Re-send the control
command of negative
range
0701
FPGA
download mark
error
Wrong download mark. It is not ':'
start Re-download
0702 FPGA The download command is not Re-download
Hardware System
5-96
download
command error
defined.
0703
Read all
configuration
chip content
finished
Finish reading data
0704 Data write out
of limit
The data written for a single time
exceeds the maximum of 32 digits Re-download
0705
Data write and
head-address
over range
The data written plus the
head-address exceed the maximum
content of one page (256 bytes).
Re-download
0706 Write forbidden Failed to write the configuration chip Send the enabling
command first
0707 Write overtime Write delay exceeds 10ms Re-download
0731 Write-in
EEPROM error Failed to write data into EEPROM
1. Check the EEPROM
circuit for any
missing-soldered
component; check the
EEPROM pin for short
circuit and missolder,
etc.
2. If short circuit is not
found and all
components are
properly soldered, it
indicates the EEPROM
is damaged.
0801
Functional code
in the control
code error
The functional code is not defined in
the communication rules.
Communication
abnormal. Please try
again.
0802 Data length
error
The data length mismatches the
corresponding data length of the
current command label
Communication
abnormal. Please try
again.
0803 Command label
error
The command label is not defined in
the communication rules.
Communication
abnormal. Please try
again.
0804 Check code
error
The check code mismatches the
calculated one
Communication
abnormal. Please try
again.
0805 End code error The end code is not the one defined Communication
Hardware System
5-97
in the communication rules. (016) abnormal. Please try
again.
0806
Identity code in
the control code
error
The identity code is not defined in
the communication rules.
Communication
abnormal. Please try
again.
0817
None-sequence
pack analysis
error
Unable to analyze the
none-sequence pack command
Re-send the correct
command
Troubleshooting special errors
1. Troubleshooting the abnormal pneumatic pressure
If a certain pressure value is abnormal, check the related pumps, valves and airway
for leakage.
If the pneumatic system is normal, then the error should be caused by the damaged
pressure sensor on the drive board. Please replace the sensor or the drive board.
The pressure signals are outputted by the volumetric board. See the following table
for the voltage and pressure.
Table 5-67 Voltage signals and the corresponding pressures
Pressure
channel
Pressure
range Test point
Relation between sensor
output voltage and pressure Note
Pressure
chamber 2.5~4kPa TP30
Vacuum
chamber
-8 ~
-40kPa TP31
( ) 25025.1 ×−= VoutP
P represents
pressure, and
its unit is kPa.
Volt
represents the
voltage of the
test point, and
its unit is V
2. Troubleshooting the abnormal mechanism action
To observe and troubleshoot the abnormal mechanism actions, the service engineers are
usually needed. When errors of this kind happen, please be sure to troubleshoot them as per
the above-mentioned recommended action and the following procedures.
Check whether the error-related mechanisms (such as motors) can move. If the
mechanisms do not move, then the error should be caused by the abnormal executive
components, such as disconnected or damaged motor. If nothing is wrong with them, then
Hardware System
5-98
the error should be caused by the damaged drive board, so the driver board should be
replaced.
If the mechanisms can move, check whether they move in position and whether they are
jammed by lines or other mechanisms during the movement. For transmissive
photocoupler, if the baffle enters above 2/3 photocoupler, it is baffled and moves is in
position.
If nothing is wrong with the mechanism action, consider the photocoupler error. Replace
the photocoupler.
If errors still remain after the photocoupler is replaced, then it should be caused by the
error components on the driver board. Replace the driver board.
Hardware System
5-99
5.4 Autoloader board
5.4.1 Introduction The autoloader board is to feed and mix blood samples, as well as scan barcode information
on tubes.
5.4.2 Board Composition
Function
Stepping motor control
Electromagnet control
Position sensor and sensitive switch status detection
Scanner control
Parameter storage
Block diagram
BUS
C8051F020UA
RT
0
IRQ
Step M
otor Control
EP1C3T144
1
2
3
4
Interface
RS422
D 5VP 12VP 24V
LDODigital 5V 3.3V
FPGA Configuration
Device
Config
Config
5
LDO3.3V 1.5V
Reset
Reset
FPGA Reset
Sensor
Sensor
Status
Mother B
oard
JTA
G
Pow
er B
oard
Sole
noid
Scanner
UART1
Clock
Clock
SMXM
SMZM
SMRM
Sample Mix
SFXM
SFYM
Sample Feed
EE
PR
OM
Figure 5-18 Block diagram of the autoloader board
Description
Stepping motor drive
The autoloader board provide the drive control to the following five stepping motors:
X-co motor of the mix mechanism
Z-co motor of the mix mechanism
Mix motor
Hardware System
5-100
X-co loading motor
Y-co feeding motor
The stepping motor drive is a type of bipolar consist-current half-step drive, and there are
position sensors corresponding to the stepping motors to detect their positions.
Electromagnet control
The electromagnet is used to control the process of opening the sample compartment door
(manual close).
Built-in scanner control
The built-in scanner control is used to read barcodes on tubes and tube racks. The autoloader
board controls the built-in scanner through MCU, and the UART communication between MCU
and the built-in scanner through Serial port 1.
Position sensor and sensitive switch status detection
There are 13 position sensor and 1 sensitive switch on the autoloader. The status of the
sensors and switch is detected by FPGA, and MCU will check the status value of the FPGA
internal storage when necessary.
Parameter storage
As there may be manufacture or installation errors, it is necessary to adjust the initial positions
for the X-co motor of the mix mechanism and the mix motor. The adjusted position parameters
will be stored on the autoloader board.
Interface definition
The autoloader is configured with 14 external interfaces, shown as follows.
Table 5-68 External ports on the autoloader board
Interface Function Number of pins Description
J1 FPGA configuration interface (AS mode)
10 FPGA downloading
J2 FPGA debugging interface (JTAG)
6 FPGA debugging
J3 Electromagnet interface 2 / J4 Reserved 2 J5 Power interface 6
J6 Mix mechanism sensor interface
20
J7 Feeding unit sensor interface 1
20
Not applied to the unit
Hardware System
5-101
J8 Feeding unit sensor interface 2
20
J10 Reserved 5 /
J11 MCU configuration interface
10 /
J12 Communication interface 4 / J14 Feeding motor interface 8 /
J15 Control the X-co motor and Z-co motor of the mix mechanism
8 /
J16 Mix motor interface 4 / P1 Built-in scanner interface 15 /
Definition of J1:
J1 is the FPGA configuration interface (AS mode).
Table 5-69 Definition of the FPGA configuration interface
Pin Name Pin Name
1 F_DCLK 2 AGND 3 F_CONFDONE 4 VDD 5 F_CONFIG 6 F_NCE 7 F_DATA 8 F_NCSC9 F_ASDO 10 GND
Definition of J3:
J3 is the electromagnet interface.
Table 5-70 Definition of the electromagnet interface
Pin Name Description Mark on Wire
Level Note
1 12V0 Working power supply for the electromagnet
12V Current:1A
2 solenoid Control signal for the electromagnet
SOLENOID
Definition of J5:
Table 5-71 Definition of the power interface
Pin Name Description 1 D+5V 5V power supply (digital) 2 P+12V 12V power supply (power) 3 P+24V 24V power supply (power) 4 GND Digital ground 5 PGND Power ground 6 PGND Power ground
Hardware System
5-102
Definition of J6:
J6 is the mix mechanism sensor interface.
Table 5-72 Definition of the mix mechanism sensor interface
Pin Name Description Mark on Wire Note 1 V_X_START Sending end power signal of
X-co initial position 2 X_START Photocoupler signal of X-co
initial position 3 GND Ground 4 GND Ground
MX_START Transmissive sensor
5 V_X_END Sending end power signal of X-co end position
6 X_END Photocoupler signal of X-co end position
7 GND Ground 8 GND Ground
MX_END Transmissive sensor
9 V_Z_START Sending end power signal of Z-co initial position
10 Z_START Photocoupler signal of Z-co initial position
11 GND Ground 12 GND Ground
MZ_START Transmissive sensor
13 V_Z_END Sending end power signal of Z-co end position
14 Z_END Photocoupler signal of Z-co end position
15 GND Ground 16 GND Ground
MZ_END Transmissive sensor
17 V_R_START Sending end power signal of mix mechanism initial position
18 R_START Photocoupler signal of mix mechanism initial position
19 GND 20 GND
MR_START Transmissive sensor
21 V_SYRINGE Sending end power signal of the sampling position
22 SYRINGE Photocoupler signal of the sampling position
23 GND Ground 24 GND Ground
Preserved Transmissive sensor
Definition of J7:
J7 is the feeding sensor interface.
Hardware System
5-103
Table 5-73 Definition of the feeding sensor interface 1
Pin Name Description Mark on Wire
Note
1 V_X_START Sending end power signal of the photocoupler for X-co feeding motor initial position
3 GND Ground 5 X_START Receiving end signal of the
photocoupler for X-co feeding motor initial position
FX_START
Transmissive sensor
2 V_ X_END Sending end power signal of the photocoupler for X-co feeding motor end position
4 GND Ground 6 X_END Receiving end signal of the
photocoupler for X-co feeding motor end position
FX_END
Transmissive sensor
7 V_Y_START Sending end power signal of the photocoupler for Y-co feeding motor initial position
9 GND Ground 11 Y_START Receiving end signal of the
photocoupler for Y-co feeding motor initial position
FY_START
Transmissive
sensor
8 V_Y_RIGHT Sending end power signal of the right photocoupler for Y-co feeding
10 GND Ground 12 Y_RIGHT Receiving end power signal of the right
photocoupler for Y-co feeding
FY_RIGHT Transmissive
sensor
13 V_Y_LEFT Sending end power signal of the left photocoupler for Y-co feeding
15 GND Ground 17 Y_LEFT Receiving end signal of the left
photocoupler for Y-co feeding
FY_LEFT Transmissive
sensor
14 V_UL_START Sending end power signal of the photocoupler for unloading initial position
16 GND Ground 18 UL_START Receiving end signal of the
photocoupler for unloading initial position
F_UL_START
Transmissive sensor
19 GND Ground 20 X_READY Sensitive switch for X-co loading ready
position FX_READY Sensitive
switch
Definition of J8:
J8 is the other feeding sensor interface.
Hardware System
5-104
Table 5-74 Definition of the feeding sensor interface 2
Pin Name Description Mark on Wire
Note
1 V_Y_READY Sending end power signal of the Y-co feeding ready photocoupler
3 GND Ground 5 VCC Receiving end power signal of the Y-co
feeding ready photocoupler 7 Y_READY Receiving end signal of the Y-co feeding
ready photocoupler
Reserved Reflective photocoupler
2 V_UL_FULL Sending end power signal of photocoupler detecting full unloading platform
4 GND Ground 6 VCC Receiving end power signal of
photocoupler detecting full unloading platform
8 UL_FULL Signal of photocoupler detecting full unloading platform
F_UL_FULL
Reflective photocoupler
9 V_TUBE Sending end power signal of tube detecting photocoupler
11 GND Ground
13 VCC Receiving end power signal of tube detecting photocoupler
15 TUBE Tube detecting photocoupler signal
F_TUBE Reflective photocoupler
10 VT_RACK Sending end power signal of sample volume sensor
12 GND Ground 14 VCC Receiving end power signal of sample
volume sensor 16 RACK Sample volume sensor signal
Reserved Reflective photocoupler
17 VC_DOOR Sending end signal of sample compartment door detecting sensor
18 DOOR Receiving end signal of the photocoupler for initial unloading position
19 GND Ground
DOOR
Reflective photocoupler
20 NC Floating
Definition of J11:
J11 is the MCU configuration interface.
Hardware System
5-105
Table 5-75 Definition of the MCU configuration interface
Pin Name Description 1 VDD Power signal 2 GND Ground 3 GND Ground 4 TCK JTAG clock 5 TMS JTAG mode selection signal6 TDO JTAG output data signal 7 TDI JTAG input data signal 8 NC Floating 9 GND Ground 10 NC Floating
Definition of J12:
J12 is the communication interface.
Table 5-76 Definition of the communication interface
Pin Name Description 1 RX+ Serial data receiving end 2 RX- Serial data receiving end 3 TX+ Serial data receiving end 4 TX- Serial data receiving end
Definition of J14:
J14 is the feeding motor interface.
Table 5-77 Definition of the feeding motor interface
Pin Name Description Mark on Wire
1 X_BN Control signal of Phase A, X-co loading motor
2 X_B Control signal of Phase AN, X-co loading motor
3 X_AN Control signal of Phase B, X-co loading motor
4 X_A Control signal of Phase BN, X-co loading motor
SFXM
5 Y_BN Control signal of Phase A, Y-co feeding motor
6 Y_B Control signal of Phase AN, Y-co feeding motor
7 Y_AN Control signal of Phase B, Y-co feeding motor
8 Y_A Control signal of Phase BN, Y-co feeding motor
SFYM
Hardware System
5-106
Definition of J15:
J15 is the control interface of X-co motor and Z-co motor for the mix mechanism.
Table 5-78 Definition of the interface of X-co motor and Z-co motor for the mix mechanism
Pin Name Description Mark on Wire
1 X_BN Control signal of Phase BN, X-co manipulator motor
2 X_B Control signal of Phase B, X-co manipulator motor
3 X_AN Control signal of Phase AN, X-co manipulator motor
4 X_A Control signal of Phase A, X-co manipulator motor
SMXM
5 Z_BN Control signal of Phase BN, Z-co manipulator motor
6 Z_B Control signal of Phase B, Z-co manipulator motor
7 Z_AN Control signal of Phase AN, Z-co manipulator motor
8 Z_A Control signal of Phase A, Z-co manipulator motor
SMZM
Definition of J16:
J16 is the mix motor interface.
Table 5-79 Definition of the mix motor interface
Pin Name Description Mark on Wire
1 R_BN Control signal of Phase BN, mix motor
2 R_B Control signal of Phase B, mix motor
3 R_AN Control signal of Phase AN, mix motor
4 R_A Control signal of Phase A, mix motor
SMRM
Definition of P1:
P1 is the built-in scanner interface.
Hardware System
5-107
Table 5-80 Definition of the built-in scanner interface
Pin Name Description 1 VCC Power signal 2 TXD Serial data receiving end 3 RXD Serial data sending end 4 GND Signal ground 5 NC Floating 6 CTS Sending “ready” signal 7 OUTPUT1 Scanner out put signal 1 8 \ \ 9 Trigger Trigger signal 10 RTS Sending request signal 11 OUTPUT3 Scanner out put signal 3 12 \ \ 13 Chassis Shielding ground 14 OUTPUT2 Scanner out put signal 2 15 NC Floating
Hardware System
5-108
Assembly drawing
Figure 5-19 Assembly drawing of the autoloader
Hardware System
5-109
5.4.3 Adjustment and Test Points
LED function definition
Table 5-81 LED definition
Indicator Description
D3 12V power supply status indicator. It turns on when the 12V power supply is normal
D4 5V power supply status indicator. It turns on when the 5V power supply is normal
D5 3.3V power supply status indicator. It turns on when the 3.3V power supply is normal
D25 MCU working status indicator. It flashes every 0.5s when MCU works normally
D26 FPGA working status indicator. It flashes every 0.5s when FPGA works normally
Function definition of test points
Table 5-82 Test point definition
No. Name Pin Note
1 TP1 PIN3 of X1 24.4545M crystal oscillator clock output
signal, FPGA working clock
2 TP2 PIN47 of U2 RD signals of MCU
3 TP3 PIN62 of U2 WR signals of MCU
4 TP4 PIN10 of U2 ALE signals of MCU
5 TP5 PIN28 of U2 FPGA reset signals
6 TP6 PIN5 of J1 FPGA configuration start signal
7 TP7 VCC +5VDC voltage(5±0.25V)
8 TP8 12V0 +12VDC voltage (12±1.2V)
9 TP9 24V0 +24VDC voltage(24±1.2V)
10 TP10 VDD +3.3VDC voltage(3.3±0.15V)
11 TP11 GND
12 TP12 GND
13 TP13 1V5 +1.5VDC voltage(1.5±0.05V)
14 TP15 TUBE Tube detection photocoupler output signal
(0.4V when blocked, and 0.2V unblocked)
15 TP18 PIN3 of X2 24.4545M crystal oscillator clock output
signal, MCU working clock
Hardware System
5-110
16 TP21 PIN12 of U16 Stepping pulse signals of Y-co feeding
motor
17 TP23 PIN12 of U18 Stepping pulse signals of X-co feeding
motor
18 TP24 PIN12 of U22 Stepping pulse signals of Z-co mix motor
19 TP25 PIN12 of U23 Stepping pulse signals of X-co mix motor
20 TP26 PIN12 of U24 Stepping pulse signals of the mix motor
Key definition
The reset switch S1 on the drive board is used to reset the drive board manually. Press the
reset switch to reset the MCU and FPGA.
5.4.4 Disassembly and assembly method
Purpose
The autoloader board is a key component. You should maintain or replace the board
immediately if any error or damage occurs. Follow the procedures to have the autoloader
board replaced.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Disassembly
1. Shutdown the analyzer and cut off the power supply;
2. Open the left door;
3. Use the 107 cross-headed screwdriver to remove the two M4X8 panhead screws fixing the
shielding box, and then remove the box;
4. Unplug all the cords and connectors connected to the autoloader control board;
5. Use the 107 cross-headed screwdriver to remove the five M3X6 cross-headed panhead
screws fixing the autoloader control board, and then remove the board.
Hardware System
5-111
Figure 5-20 Removing the autoloader control board
1 ---Shielding box 2 --- Panhead screw M4X8
3 --- Autoloader control board 4 ---Cross-headed panhead screw M3X6
Assembly
Assemble the autoloader control board as per the above-mentioned steps in the reverse order.
Wear antistatic gloves before maintaining or removing the board.
Be sure to shut down the analyzer and cut off the power supply before disassembling/assembling the board.
Verification
1. Check and make sure that all the components are properly connected to the autoloader
control board.
2. After ensuring the power supply is cut off, assemble the drive board and motherboard
properly, and then power on the analyzer.
3. Check if the indicators on the autoloader control board are in the following status: 1) D3,
D4 and D5 turned on; 2) D25 and D26 flashing. You can also check the indicators in the
“self-test” screen of the analyzer. If the self-test is passed, it means the autoloader control
board is replaced successfully; otherwise, you should troubleshoot the board.
Hardware System
5-112
4. Readjust the pincher position as per Section 6.10.2
5.4.5 Troubleshooting
Error analysis and solutions
Errors related to the autoloader board are identified and handled by the host control software
in the form of error codes. Note that only some of the error codes reported to the host control
software from the autoloader board can be handled automatically, while some need to be
judged with the help of users or service engineers. When an error is reported, the error No.
and error names are provided by host control software. The recommended solutions may
appear on the screen as the software help information.
Most errors of the autoloader board are identified by checking the photocoupler information, so
check the photocouplers first in case of errors. To check the photocouplers: first make sure that
the photocouplers are not blocked, and check if the status information of the photocouplers
from the software matches the actual status; hide the photocouplers with lighttight material,
and check if the status information matches the actual status. Inconsistency found in either one
of the two checks indicates invalidation of the photocouplers; otherwise, the error comes from
the board itself.
Typical errors and solutions:
1. Y-co left or right photocoupler error
The cause of such an error: there two photocouplers mounted on the Y-co feeding direction of
the autoloader, which are called the left photocoupler and right photocoupler. In the
autoloading mode, the current tube feeding status is identified based on the number of times
when the level of the photocouplers jumped and the status of the photocouplers. Therefore,
the left and right photocouplers should be related with each other based on certain logical
requirements. If the requirements are not met, the “Y-co left photocoupler or right photocoupler
error” will be reported.
Troubleshooting measures: 1) check the photocouplers for errors (according to the steps
mentioned above); 2) Check if there is any tube rack blocked along the Y-co feeding line which
occurs when the Y-co feeding motor is not able to push the tube rack due to the tube rack out
of shape or assembly error of the sample transport unit. In such a case, shut down the
analyzer and switch off all the power supplies, and then pull the tube rack along the Y-co
feeding line to identify if the tube rack is out of shape and if the unit is mis-assembled. Replace
with a qualified tube rack if the former one is out of shape; disassemble and repair the sample
transport unit in case of mis-assembly; 3) check if there is any tube rack blocked because
there are tube racks slanting at the unloading area or the unloading platform is not qualified.
Check if the length of the tube rack meets the corresponding requirement and if the metallic
plate of the unloading unit is properly manufactured. 4) Check if the autoloader assembly
(including the buffer plates for the left and right photocouplers, Y-co pallet and Y-co tube rack
backboard) is properly assembled. If there is any mis-assembly, re-adjust the installation
Hardware System
5-113
according to the part related to mechanical structure of the analyzer in this manual, and make
sure the mechanism is out of error.
2. The compartment door cannot be opened
Possible causes for this error: a) compartment door photocoupler error; b) compartment spring
error or door structure error.
Troubleshooting measures: a) check if there is any error with the photocouplers; b) press the
[OPEN] key and check if the electromagnet moves. If the electromagnet acts properly, then
there may be errors with the spring or the mechanical structure of the door and replacement or
adjustment may be necessary.
Error codes and troubleshooting measures
Table 5-83 Troubleshooting errors with the autoloader board
ERR Error name Error feature Recommended action
5A00 EEPROM write error
Failed to write data to EEPROM
1. Check if there are EEPROM elements not welded, EEPROM pin short or mis-welded
2. If there are no such errors, then it is identified as an EEPROM error and board
replacement is needed
5201 SMXM action failed
Current action forbidden while the mix mechanism is
in motion
Confirm the action status of the mix mechanism and resend the command
5202 SMXM action failed
SMXM should not at the home position, but the
home position photocoupler is blocked
1. SMXM home position photocoupler error: replace the photocoupler if it is confirmed to be a photocoupler error
5203 SMXM action failed
SMXM should at the home position, but the home
position photocoupler is not blocked
5204 SMXM action failed
SMZM or SMRM not at home position
Check the Z-co and R-co photocouplers replace the photocoupler if it is confirmed to be
a photocoupler error
5205 SMXM action failed
SMXM not initialized, action forbidden Initialize the X-co motor
5206 SMXM action failed SMXM action overtime
Check if the predefined action time is reasonable and the actual action speed of the
motor meets the requirement of the design
5207 SMXM action failed
SMXM should not at the end position, but the end position photocoupler is
blocked
1. SMXM end position photocoupler error: replace the photocoupler when it is confirmed to be an photocoupler error
5208 SMXM action failed
SMXM should at the end position, but the end
position photocoupler is not blocked
5209 SMXM
photocoupler error
Both the home position and end position photocouplers are blocked, then at least one of them has an error
1. Check the SMXM home position and end position photocoupler, and replace the photocoupler when it is confirmed to be an photocoupler error
Hardware System
5-114
5401 SMZM action failed
Current action forbidden while the mix mechanism is
in motion
1. Confirm the action status of the mix mechanism and resend the command
5402 SMZM action failed
SMZM should not at the home position, but the
home position photocoupler is hide
1. Home position photocoupler error: replace the photocoupler if it is confirmed to be a photocoupler error
5403 SMZM action failed
SMZM should at the home position, but the home
position photocoupler is not blocked
5404 SMZM action failed
SMXM not at the end position or SMRM not at the
home position
1. Check the SMXM and SMRM photocouplers, and replace the photocoupler when it is confirmed to be an photocoupler error
5405 SMZM action failed
SMZM not initialized, action forbidden Initialize the Z-co motor
5406 SMZM action failed SMZM action overtime
Check if the predefined action time is reasonable and the actual action speed of the
motor meets the requirement of the design
5407 SMZM action failed
SMZM should not at the end position, but the end position photocoupler is
blocked
1. SMZM end position photocoupler error: replace the photocoupler when it is confirmed to be an photocoupler error
5408 SMZM action failed
SMZM should at the end position, but the end
position photocoupler is not blocked
5409 SMZM
photocoupler error
Both the home position and end position photocouplers are blocked, then at least one of them has an error
1. Check the SMZM home position and end position photocoupler, and replace the photocoupler when it is confirmed to be an photocoupler error
5501 SMRM action failed
Current action forbidden while the mix mechanism is
in motion
1. Confirm the action status of the mix mechanism and resend the command
5502 SMRM action failed
SMRM should not at the home position, but the
home position photocoupler is blocked
1. SMRM home position photocoupler error: replace the photocoupler if it is confirmed to be a photocoupler error
5503 SMRM action failed
SMRM should at the home position, but the home
position photocoupler is not blocked
5504 SMRM action failed
SMXM or SMZM not at the end position
2. Check the SMXM and SMZM photocouplers, and replace the photocoupler when it is confirmed to be an photocoupler error
5505 SMRM action failed
SMRM not initialized, action forbidden Initialize the R-co motor
5506 SMRM action failed SMRM action overtime
Check if the predefined action time is reasonable and the actual action speed of the
motor meets the requirement of the design
5508 SMRM action failed
Not reach the target position after finish moving Check the SMRM motor and photocouplers
5601 Mix mechanism Action forbidden while the Confirm the action status of the mix
Hardware System
5-115
action failed current mechanism is in motion
mechanism and resend the command
5604 Mix mechanism action failed
Number of mixing times out of predefined range
Confirm the number of mixing times and resend the correct command
5605 Mix mechanism action failed
Mix mechanism not initialized Initialize the mix mechanism
5606 Mix mechanism action failed
At least one of the 3 motors is not at the home position
Check the home position photocouplers of the mix motors. and replace the photocoupler when it is confirmed to be an photocoupler
error
5700 SFXM action failed SFXM motor in motion Confirm the SFXM action status and resend
the command
5701 SFXM action failed
SFXM home position photocoupler status does not match the position of
the current motor
1. Self-test if the SFXM home position photocoupler is broken
2. When the photocoupler is working properly, send a SFXM action command to check if the SFXM motor is normal
5702 SFXM action failed
Tube rack along Y-co; X-co motor forbidden to move to
the end position
1. Remove the tube rack long Y-co, and then resend the command
5703 SFXM action failed
SFXM end position photocoupler status does not match the position of
the current motor
1. Self-test if the SFXM end position photocoupler is broken;
2. When the photocoupler is working properly, send a action command to check if the SFXM motor is normal
5704 SFXM
photocoupler error
Both the SFXM end position and home position
photocouplers are blocked
1. Check if both the SFXM end position and home position photocouplers are blocked. If
not, it is the photocoupler at fault.
5705 SFXM loading error
Both the SFXM end position photocoupler signal and the X-co ready sensitive switch invalid after finish loading
1. When there is a ready tube rack, if it is determined that the X-co ready sensitive switch does not function properly, then replace the switch
2. When there is no tube racks, SFXM at end position indicates a SFXM end position photocoupler error. Replace the photocoupler.
5706 SFXM action failed
SFXM motor action overtime
Check if the defined action time is reasonable and the actual action speed of the motor meet
the requirement of the design
5708 SFXM action failed
FX_READY sensitive switch error: should be
switched off, but is detected to be switched on
Sensitive switch error: replace the sensitive switch after confirming the switch is broken
5801 Functional
control code error
Functional code not defined in the communication
protocol Communication error. Please resend
5802 Data length errorData length differs from that
of the current command label
Communication error. Please resend
5803 Command label error
Command label not defined in the communication
protocol Communication error. Please resend
5804 Check code error The check code differs from the calculated check code Communication error. Please resend
Hardware System
5-116
5805 End code error Not the end code defined in the communication protocol
(016) Communication error. Please resend
5806 Identity control code error
Identity code not defined in the communication protocol Communication error. Please resend
5817 Non-sequence pack analysis
error
Unable to analyze the non-sequence pack Resend the correct command
Hardware System
5-117
5.5 Power board
5.5.1 Introduction The power board provides six stable power supplies, including D5V, A+12V, A-12V, AC130V,
P12V and P24V.
5.5.2 Board Composition
Function
The power board provides the following six stable power supplies for each functional board
and module of the BC-5380: D5V, A+12V, A-12V, AC130V, P12V and P24V.
Block diagram
Figure 5-21 Block diagram of the power board
Description
The power board works with the 90V-264V AC input voltage (50-60Hz).
Hardware System
5-118
Once the AC power switch is turned on, all the circuits work and the
D5V,A+12V,A-12V,AC130V,P12V and P24V voltage are outputted.
Circuit loads of the power board are described in the table below:
Table 5-84 Output voltage features
Voltage Minimum current
Nominal current
Output voltage range
Voltage adjustment
rate
Load adjustment
rate
Noise
D5V 2A 5A 4.85/5.25V ±1% ±5% 100mV
+A12V 0mA 1A 11. 5/12. 5V ±1% ±5% 100mV
-A12V 0mA 650mA -11.5/-12.5V ±1% ±5% 100mV
P12V 0.3A 5.5A 11.5/12.5V ±1% ±5% 150mV
P24V 0A 4A 22/27V ±1% ±10% 150mV
AC120V 0mA 60mA 115/145V(RMS) ±1% ±10% /
For the P24V, the 7.8A maximum current with the approximate lasting time of 2S and the
period of 60S may occur when working.
Interface definition
The power board is configured with 7 external interfaces. The J1 and J2 interfaces are
connectors of the socket; the rest of the interfaces are connectors of the plug terminal leading
from the board edges. The No. of the PCB board is TP1-TP20.The location of the interfaces on
the power board is shown below.
Hardware System
5-119
Figure 5-22 Interface layout of the power board
See the following table for the function of each interface.
Table 5-85 Interfaces on the power board
Interface Function Number of pins Number of lands Note
J1 AC input 3 / /
J2
Power supply output
of the fans 6 / /
A-J12
Power supply output
of the analog part 6 4
PCB
soldering
output
A-J13 AC output 3 2
PCB
soldering
output
Hardware System
5-120
A-J35
Power supply output
of the power part 10 10
PCB
soldering
output
A-J37
Power supply output
of the digital part 4 4
PCB
soldering
output
Definition of J1:
Table 5-86 Definition of AC input interface
Pin Definition
1 L, Live wire
2 PG, Earth wire
3 N, Neutral wire
Definition of J2:
Table 5-87 Definition of the fan interface
Pin Definition
1/3/5
P12V,connects the positive end of the P12V output
voltage
2/4/6
PGND, connects the negative end of the P12V
output voltage
Definition of A-J12:
Table 5-88 Definition of analog board output
Land Definition Pin
TP16AGND\TP17AGND
AGND, connects the reference
ground of the analog board output
voltage 6,7
TP15A12V
+A12V,connects the positive end of
the +A12V output voltage
5
TP18A-12V
-A12V,connects the negative end of
the -A12V output voltage
8
/ NC 1/2/3/4
Definition of A-J13:
Hardware System
5-121
Table 5-89 Definition of AC output
Land Definition Pin
TP19\TP20
Two output ends of
AC120
1/3
/ NC 2
Definition of A-J35:
Table 5-90 Definition of power supply output
Land Definition Pin
TP1P12V/TP2P12V positive end of the P12V
output voltage 4/5
TP12PGND/TP13PGND/TP3P
GND\TP4PGND\TP14PGND
negative end of the P12V
output voltage 6/7/8/9/10
TP9P24V\TP10P24V\TP11P24
V
Positive end of the P24V
output voltage 1/2/3
Definition of A-J37:
Table 5-91 Definition of D5V output
Land Definition Pin
TP5D5V/TP6D5V Positive end of the 5V output
voltage 1/2
TP7DGND/TP8DGND DGND, negative end of the
5V output voltage
3/4
Assembly drawing
Hardware System
5-122
Figure 5-23 Assembly drawing of the power board
5.5.3 Adjustment and Test Points The important test points when maintaining the power board are listed below:
Hardware System
5-123
Table 5-92 Test points on the power board
No. Pin Function Reference value
1 Q101.2 Q101 switching waveform /
2 Q101.1 Q101 drive waveform /
3 C114.+ PFC output voltage 390±20V
4 U101.9 Reference output voltage of the chip 7.5V
5 U101.14 U101 oscillating waveform /
6 C206.+ VCC voltage 17V~22.5V
7 C204 VDD voltage 12±1V
8 C223 supply voltage of U201 12±1V
9 Q201.1 Q201 drive waveform /
10 Q201.2 Q201 switching waveform /
11 U201.8 U201 reference voltage 5V
12 U201.4 U201 oscillating waveform /
13 Q201.3 R235 voltage waveform, Q201 current
waveform
/
14 C232.+ D5V output voltage and ripple voltage 5±0.5V
15 C311.+ U301 supply voltage 12±1V
16 U301.4 U301 oscillating waveform /
17 U301.8 U301 reference voltage 8V
18 Q303.1 Q303 drive waveform /
19 Q303.2 Q303 switching waveform /
20 Q303.3 Q303 current waveform, R313 voltage
waveform
/
21 C326.+ 12VB voltage 12±0.5V
22 C328.+ 130V DC output 115V~150V
23 C332.+ A+12V output 12±0.5V
24 C338.- A-12V output -12±0.5V
25 U351.15 U351 supply voltage 12±0.5V
26 U351.5 U351 oscillating waveform /
27 U351.11 Q351 drive waveform /
28 U351.14 Q351 drive waveform /
29 U351.16 U351 reference voltage 5V
30 Q361.1 Q361 drive waveform /
31 Q362.1 Q362 drive waveform /
32 Q363.1 Q363 drive waveform /
33 Q364.1 Q364 drive waveform /
34 C409.+ U401 supply voltage 12±1V
Hardware System
5-124
35 U401.5 U401 oscillating waveform /
36 U401.14 U401 reference voltage 5V
37 U401.9 Q401 drive waveform /
38 Q406.1 Q406 drive waveform /
39 Q406.2 Q406 switching waveform /
40 C423.+ P24V output voltage, ripple 22~27V
41 C442.+ P12V output voltage, ripple 12±0.5V
5.5.4 Replacement and Connection
Purpose
Damaged power assembly or power board needs to be replaced by a new one.
Tools
107 cross-headed screwdriver
Sharp-nose pliers
Multimeter
Disassembly
Disassembly of the power board consists of the following two main procedures (removing the
whole power assembly and further disassembly of the inner components):
Remove the whole power assembly from the main unit:
1. Shutdown the analyzer and cut off the power supply;
2. Open the left door;
3. Remove the data board(see section 5.2.4 for details);
4. Remove the drive board(see section 5.3.4 for details);
5. Unplug the 3 cables connecting the power assembly from the mother board; loose the
cable fixer and remove the cables.
6. Use the 107 cross-headed screwdriver to remove the cross-headed panhead screws
(M4X8) fixing the earth wire from the ground pole.
7. Use the 107 cross-headed screwdriver to remove the 6 cross-headed panhead screws
(M4X8) fixing the power assembly.
8. Draw out the power assembly for 50-100mm and unplug the 2 cable connectors
connecting the power switch.
Hardware System
5-125
9. Draw out and remove the power assembly.
Figure 5-24 Removing the power assembly 1
1 ---Mother board 2 --- Power switch
3 --- Ground pole 4 --- Power assembly
5 --- Panhead screw M4X8
Further disassembly of the power assembly
1. Use the 107 cross-headed screwdriver to remove the 2 cross-headed panhead screws
(M4X8) fixing the power cover, and then remove the cover together with the power socket
from the power assembly.
2. Use the 107 cross-headed screwdriver to remove the 2 cross-headed panhead screws
(M3X8) fixing the power socket, and then disassembly the power socket.
3. Use the 107 cross-headed screwdriver to remove the 4 cross-headed panhead screws
(M3X8) fixing the shielding box, and then remove the shielding cover.
4. Use the shape-nose pliers to loose the 6 M3 nuts fixing the fan and unplug the cable plug
from the power board, and then remove the fan.
5. Use the 107 cross-headed screwdriver to remove the 9 cross-headed panhead screws
(M3X6) from the power board, and then remove the power board.
Hardware System
5-126
Figure 5-25 removing the power assembly 2
1---Panhead screw M3X8 2 --- Power socket
3---Panhead screw M4X8 4 ---Fan
5---M3 Nut 6---Power board
7 ---fixing board for power board 8---Cross-recessed panhead screw M3x6
9---Power board shielding cover 10 ---Power cover
Assembly
Assemble the power board as per the above-mentioned steps in the reverse order.
Wear antistatic gloves before maintaining or removing the board.
Be sure to shut down the analyzer and cut off the power supply before disassembling/assembling the board.
Verification
1. Check and make sure the earth wire of the AC input socket is properly connected to the
conductive cover of the main unit.
2. Connect the power cord and turn on the AC switch, then the fan works.
Hardware System
5-127
When assembling, before fixing the upper cover of the power assembly, be sure the AC plug and the plug of the fan are connected with the power board.
Be sure the screws are tightened to firmly fix the power assembly with the main unit.
5.5.5 Troubleshooting
Causes and recommended action
Refer to the flow chart below to troubleshoot the power board.
Hardware System
5-128
Figure 5-26 Power board troubleshooting flow chart
Be sure the screws are tightened to firmly fix the power assembly with the main unit.
Hardware System
5-129
5.6 Volumetric and pressure detecting board
5.6.1 Introduction The volumetric board provides the metric time signals for the RBC/PLT and WBC count.
Principles for the volumetric time signals: the detected volumetric start and end signals are
processed by the volumetric sensor, through current-voltage transferring, level-transforming
and de-bouncing, the drive-board-required TTL level is obtained and outputted. Principles for
the vacuum and pressure detecting: the detected vacuum and pressure signals are transferred
to the current signals by the vacuum and pressure sensor, and then through the signal
transferring, the AD-required the voltage signals are outputted to the drive board.
5.6.2 Board Composition
Function
Provides the time signals of RBC/PLT and WBC count, i.e. the start and end time of
running.
Provides the vacuum and pressure intensity signals of the vacuum chamber and pressure
chamber, sends the signals to the drive board.
Block diagram
Figure 5-27 Block diagram of the volumetric and pressure detecting board
Hardware System
5-130
Description
WBC count start/end signals
The WBC count start/end signals detected by the WBC volumetric sensor are I-V transferred,
level-transformed and de-bounced by the volumetric board, and then sent to the drive board.
RBC/PLT count start/end signals
The RBC/PLT count start/end signals detected by the RBC/PLT volumetric sensor are I-V
transferred, level-transformed and de-bounced by the volumetric board, and then sent to the
drive board.
Drive control the volumetric sensor
The working status of the volumetric sensor is controlled by the drive board through J1
connector. When the level of the control signal is high, the volumetric sensor doesn't work;
when the level is low, the volumetric sensor works.
Vacuum/pressure voltage signals
The vacuum/pressure intensity signals of the vacuum/pressure chamber are transferred to the
voltage signals by the vacuum/pressure sensor on the volumetric board. Then, they are further
transferred to the A/D-required voltage signals and then sent to the drive board through the J1
connector.
Interface definition
The volumetric board has only one 12PIN connector connecting the drive board. The location
of the connector is shown below.
Hardware System
5-131
Figure 5-28 Interface layout of the volumetric board
Table 5-93 Pin definition
No. Pin Signal description Signal Type
1 P+12V +12V power supply +12 power supply
2 PGND Power ground /
3 DGND Digital ground /
4 POSI_PRESS Pressure voltage signals output Analog
5 NEGA_PRESS Vacuum voltage signals output Analog
6 DGND Digital ground /
7 #VM_CTRL Volumetric sensor control signals TTL level
8 #RBC_START RBC count start signals TTL level
9 #RBC_STOP RBC count stop signals TTL level
10 #WBC_START WBC count start signals TTL level
11 #WBC_STOP WBC count stop signals TTL level
12 D+5V Digital +5V power supply +5V power supply
Hardware System
5-132
Assembly drawing
Figure 5-29 Assembly drawing of the volumetric board-Top
Figure 5-30 Assembly drawing of the volumetric board-Bottom
5.6.3 Disassembly and assembly method
Purpose
The volumetric board is the front component of the whole device. Damaged volumetric board
needs to be replaced by a new one. Please follow the procedures to have the volumetric board
replaced.
Hardware System
5-133
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Sharp-nosed pliers
Disassembly
1. Shut down the analyzer;
2. Open the right door;
3. Use the 107 cross-headed screwdriver to remove the 2 cross-headed panhead screws
(M3X8) fixing the shielding box, and then remove the shielding cover.
4. Disconnect the 4 pipes connecting the volumetric tube.
5. Use the 107 cross-headed screwdriver to remove the 2 cross-headed panhead screws
(M3X8), and use the sharp-nosed pliers to remove the 2 socket cap screws (M3X16) fixing
the volumetric board.
6. Pull out the volumetric board for 30-50mm; unplug the cable connectors on its back and
disconnect the 2 pipes, and then remove the board.
Figure 5-31 Disassembly of the volumetric board
1---volumetric tube bracket 2---Panhead screw M3X8
3---Volumetric tube 4---Rubber ring
5---Socket cap screws (M3X16) 6---Shielding box of volumetric board
7---Volumetric board
Hardware System
5-134
Assembly
Assemble the power board as per the above-mentioned steps in the reverse order.
Verification
1. The volumetric board is assembled properly.
2. Start up the analyzer properly.
3. Running samples in the whole blood mode. If the RBC/WBC clogging, bubbles and
abnormal pressure/vacuum are not reported by the analyzer, the replacement is
succeeded. If one or more above-mentioned error(s) is reported, you should remove the
error accordingly.
5.6.4 Troubleshooting
Table 5-94 Troubleshooting the volumetric board
No. Error name Error
type Error feature Causes Probability
Recommended
action
Improper cable
connection with
the mother board
or disconnection
caused by a
broken cable
Medium
Reconnect the
cable or
change a new
cable
1 +12V power
supply error
System
error
The voltage of
the P+12V test
point is not
equal to
12V±0.6V.
Improper power
supply of the
main unit or
Improper +12V
power supply
sent by the
mother board
Low
See
troubleshooting
of the drive
board and
power board
2 +5V power
supply error
Board
error
The voltage of
the P+5V test
point is not
equal to
5V±0.25V.
L2 is damaged
(or rosin jointed)
or something is
wrong (rosin
jointed) with the
C20 and C13.
Low
Replace or
re-solder the
components
Hardware System
5-135
Improper cable
connection with
the mother board
or disconnection
caused by a
broken cable
Medium
Reconnect the
cable or
change a new
cable
Something is
wrong with the
+5V power
supply.
Low See No.2 in the
table
3
Comparator
reference power
supply wrong
Board
error
The TP6
output voltage
≠ 2.5V±0.25V Something is
wrong with the
U8 chip or it is
rosin jointed.
Low
Replace or
re-solder the
U8
Improper cable
connection with
the mother board
or disconnection
caused by a
broken cable
Medium
Reconnect the
cable or
change a new
cable
4 Control signal
error
System
error
Voltage of the
# VM_CTRL
signal ≥ 0.8V Control signals
sent by the drive
board through
the mother board
is wrong
Low
See
troubleshooting
of the drive
board
Control signal
error Medium
See No.4 in the
table
5
Constant current
source of the
volumetric
sensor error
Board
error
Standard
voltage (TP5) ≠
2.5V±0.25V
Control circuit
error (Q1 or U2 is
wrong or rosin
jointed; resistors
R10, R11, R18,
R23, R25 and
R24 is wrong or
rosin jointed)
Medium
Replace or
re-solder the
components
Hardware System
5-136
VWST, VWSP,
VRST or VRSP
are not adjusted
well.
Medium
Adjust VWST,
VWSP, VRST
and VRSP to
make the
voltage of the
point WT, WP,
RT and RP
equals to
2.7V±0.1V
respectively.
Volumetric
sensor U15, U21,
U16 or U17 is
wrong (or rosin
jointed).
Low
Replace or
re-solder the
components
Constant current
source of the
volumetric
sensor error
Medium See No.5 in the
table
6
Voltage of the
point WT, WP, RT
and RP is not the
expected value
Board
error
Voltage of the
point WT, WP,
RT and RP ≠
2.7V±0.1V
+5V power
supply error Low
See No.2 in the
table
D1, D2, D3 or D4
is damaged (or
rosin jointed)
Low
Re-solder or
change the
component
+5V power
supply error Low
See No.2 in the
table
Something is
wrong with the
TP6 standard
power supply
Low See No.3 in the
table 7
Indicator
abnormal
Board
error
Indicators D1,
D2, D3 and D4
are off when
fluid flows or
are still on
when no fluid
flows through
the volumetric
tube Voltage of the
point WT, WP, RT
and RP ≠
2.7V±0.1V
Medium See No.6 in the
table
8 Constant current
source error of
Board
error
Standard
voltage TP10 ≠
+12V power
supply error Low
See No.1 in the
table
Hardware System
5-137
the vacuum
sensor
2.5V±0.25V C14, R37, U11,
R47 or U4 is
damaged (or
rosin jointed).
Medium
Re-solder or
change the
component
+12V power
supply error Low
See No.1 in the
table
9
Constant current
source error of
the pressure
sensor
Board
error
Standard
voltage TP7 ≠
2.5V±0.25V
C21, R36, R35,
U3 or U10 is
damaged (or
rosin jointed).
Medium
Re-solder or
change the
component
+12V power
supply error Low
See No.1 in the
table
10
Pressure/vacuum
unit +2.5V
reference power
supply error
Board
error
Standard
voltage TP9 ≠
2.5V±0.25V
R39, C11, R38,
R48, VR9, U9 or
U6 is damaged
(or rosin jointed)
Medium
Re-solder or
change the
component
VR9 is not
adjusted well. Medium
Adjust VR9,
making the
relative voltage
of the TP9 to
TP8 ≠
1.250±0.002V
Pressure/vacuum
2.5V reference
power supply
error
Medium See No.10 in
the table
11
Pressure/vacuum
1.25V voltage
error
Board
error
The relative
voltage of the
TP9 to the TP8
≠1.250±0.002V
U6 is damaged
or rosin jointed Medium
Replace or
re-solder the
components
+12V power
supply error Low
See No.1 in the
table
Constant current
source error of
the pressure
sensor
Medium See No.9 in the
table
12 Pressure output
voltage abnormal
Board
error
When the input
pressure is
0HPa or
450HPa, the
output voltage
of the point PP
is not equal to
1.250±0.002V 1.25V voltage
error Medium
See No.11 in
the table
Hardware System
5-138
PPZ or PPG is
not adjusted wellMedium
Adjust PPZ to
make the
voltage of the
PP is
1.250±0.002V
when the input
pressure is 0.
Then, adjust
PPG to make
the voltage of
the PP is
1.250±0.002V
when the input
pressure is
450HPa.
or
2.250±0.002V.
U12, U14, R34,
R44, R45, R46,
R50, R52, C10 or
C23 is damaged
(or rosin jointed)
Medium
Re-solder or
change the
component
+12V power
supply error Low
See No.1 in the
table
Constant current
source error of
the vacuum
sensor
Medium See No.8 in the
table
13 Vacuum output
voltage abnormal
Board
error
When the input
pressure is
0HPa or
-450HPa, the
output voltage
of the point NP
is not equal to
1.250±0.002V 1.25V voltage
error Medium
See No.11 in
the table
Hardware System
5-139
NPZ or NPG is
not adjusted wellMedium
Adjust NPZ to
make the
voltage of the
NP is
1.250±0.002V
when the input
pressure is 0.
Then, adjust
NPG to make
the voltage of
the NP is
0.250±0.002V
when the input
pressure is
-450HPa.
or
0.250±0.002V.
U12, U13, R33,
R40, R41, R42,
R43, R51, C7 or
C22 is damaged
(or rosin jointed)
Medium
Re-solder or
change the
component
Hardware System
5-140
5.7 Liquid-level detecting board 5.7.1 Introduction The liquid level board detects the status (sufficient or insufficient) of the LEO(I), LEO(II), LH
and cleanser and sends the detected results to the drive board. Principles for the liquid level
detecting: the sufficient or insufficient signals detected by the liquid sensor are processed by
current-voltage transferring, level-transforming and de-bouncing. Finally, the
drive-board-required TTL level is obtained and outputted.
5.7.2 Board composition
Function
The liquid level board detects the status (sufficient or insufficient) of the LEO(I),LEO(II), LH and
cleanser, transfers the signals and sends them to the drive board.
Block diagram
Liquid Level Board
LEO(I) SensorLEO
Itube
I-V transform Level transform
LEO(II) SensorLEO
IItube
I-V transform Level transform
LH SensorLH tube I-V transform Level transform
CL SensorCL tube I-V transform Level transform
J1 MotherBoard
Figure 5-32 Block diagram of the liquid-level detecting board
Hardware System
5-141
Description
For the liquid level of the LEO(I), LEO(II), LH and cleanser are detected by the same principle,
so the liquid level detection of LEO(I) is used just to be the example.
Detecting the status (sufficient/insufficient) of LEO(I)
The status (sufficient/insufficient) signals of the LEO(I) detected by the sensor is conducted the
I-V converting, level-transforming and de-bouncing by the liquid-level detecting board and then
sent to the drive board.
Drive control of the detecting sensors of the four reagents
The working status of the detecting sensors is controlled by the drive board through J1
connector. When the level of the control signal is high, the detecting sensors do not work;
when the level is low, the detecting sensors work.
Interface definition
The liquid level detecting board has only one connector of 8PIN connecting the drive board.
The location of the connector is shown below.
J1
Figure 5-33 Interface definition of the liquid-level detecting board
The definition of pins is shown in the table below.
Hardware System
5-142
Table 5-95 Pin definition of the liquid-level detecting board
No. Pin Signal description Signal Type
1 D+5V Digital +5V power supply +5V power supply
2 #LIQ1 LEO(I) status output signals TTL level
3 #LIQ2 LEO(II) status output signals TTL level
4 #LIQ3 LH status output signals TTL level
5 #LIQ4 Cleanser status output signals TTL level
6 / / /
7 #D_CTRL Drive control signals of detecting sensor TTL level
8 DGND Digital ground /
Assembly drawing
Figure 5-34 Assembly drawing of the liquid-level detecting board-Top
Figure 5-35 Assembly drawing of the liquid-level detecting board-Bottom
Hardware System
5-143
5.7.3 Replacement and Connection
Purpose
Damaged liquid-level detecting board needs to be replaced by a new one to ensure the
operation of the analyzer. Please follow the procedures to have the board replaced.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Disassembly
Liquid-level detecting board disassembly consists of the following two main procedures:
Remove the board from the main unit:
1. Shut down the analyzer.
2. Open the left door.
3. Use the 107 cross-headed screwdriver to remove the 2 cross-headed panhead screws
(M4X8) fixing the shielding box, and then remove the shielding cover.
4. Pull the 4 pipes out from the pipe clamp of the liquid-level detecting board.
5. Use the 107 cross-headed screwdriver to remove the 4 cross-headed panhead screws
(M3X6) from the liquid-level detecting board.
6. Pull out the liquid-level detecting board for 30-50mm, unplug the cable connectors on its
back, and then remove the board.
Hardware System
5-144
Figure 5-36 Disassemble the liquid-level detecting board-1
1---Cross-recessed panhead screw
M3x6 (with washer)
2---Liquid-level detecting board shielding
cover
3---Panhead screw M3X8 4---Liquid-level detecting board
Further disassembly of the drive board
1. Use the 107 cross-headed screwdriver to remove the 4 cross-headed panhead screws
M3X8 fixing the pipe clamp.
2. Remove the 2 pipe clamps.
Figure 5-37 Disassemble the liquid-level detecting board-2
1---Cross-recessed panhead screw M3X8 2---Liquid-level detecting board
3---Pipe clamp
Hardware System
5-145
Assembly
Assemble the power board as per the above-mentioned steps in the reverse order.
Verification
1. The liquid-level detecting board is assembled properly.
2. The tubing is connected properly.
3. Start up the analyzer properly. Running samples in the whole blood mode. If “No LEO(I)
(/LEO(II)/LH /cleanser)” or bubbles are not reported by the analyzer, the replacement is
succeeded. If one or more above-mentioned error(s) is reported, you should remove the
error accordingly.
5.7.4 Troubleshooting
Table 5-96 Troubleshooting the liquid-level detecting board
No. Error name Error
type
Error
features Causes Probability
Recommended
action
Improper cable
connection with
the mother board
or disconnection
caused by a
broken cable
Medium
Reconnect the
cable or change a
new cable
Improper power
supply of the main
unit or Improper
+5V power supply
sent by the
mother board
Low
See
troubleshooting of
the drive board
and power board
1 +5V power
supply error
System
error
The voltage
of the +5V
test point is
not equal to
5V±0.25V.
L2 is damaged (or
rosin jointed) or
something is
wrong (rosin
jointed) with the
C7 and C11.
Low
Replace or
re-solder the
components
2
Comparator
reference
power
Board
error
The TP1
output
voltage ≠
Something is
wrong with the
+5V power supply.
Low See No.1 in the
table
Hardware System
5-146
supply
wrong
2.5V±0.25V Something is
wrong with the
chip U9 or R10 or
they are rosin
jointed.
Low Replace or
re-solder the U9
Improper cable
connection with
the mother board
or disconnection
caused by a
broken cable
Medium
Reconnect the
cable or change a
new cable
3 Control
signal error
System
error
Voltage of
the #
D_CTRL
signal ≥
0.8V
Control signals
sent by the drive
board through the
mother board is
wrong
Low
See
troubleshooting of
the drive board
Control signal
error Medium
See No.3 in the
table
4
Abnormal
driving of
the
detecting
sensor
Board
error
Voltage to
ground of
collector
electrode of
Q1 is not
between
4.4V-5.25V.
Control circuit
error (Q1, U1, U2,
U3 or U4 are
wrong or rosin
jointed; resistors
R1, R2, R9, R40,
R41, R42 or R43
are wrong or rosin
jointed).
Medium
Replace or
re-solder the
components
VLL1, VVL2,
VVL3 or VVL4 are
not adjusted well.
Medium
Adjust VL1, VL2,
VL3 and VL4 to
make the voltage
of the point VVL1,
VVL2, VVL3 and
VVL4 equals to
2.2V±0.1V
respectively.
5
The voltage
of point
VLL1,
VVL2,
VVL3 and
VVL4 is not
the
expected
value.
Board
error
voltage of
point VLL1,
VVL2,
VVL3 and
VVL4 ≠
2.2V±0.1V
Detecting sensors
U1, U2, U3 or U4
are wrong (or
rosin jointed).
Low
Replace or
re-solder the
components
Hardware System
5-147
Abnormal driving
of the detecting
sensor
Medium See No.4 in the
table
+5V power supply
error Low
See No.2 in the
table
LED1, LED2,
LED3 or LED4 are
damaged (or rosin
jointed)
Low
Re-solder or
change the
component
+5V power supply
error Low
See No.1 in the
table
Something is
wrong with the
TP1 standard
power supply
Low See No.2 in the
table
voltage of point
VLL1, VVL2,
VVL3 and VVL4 ≠
2.2V±0.1V
Medium See No.5 in the
table
6 Indicator
abnormal
Board
error
Indicators
LED1,
LED2,
LED3 and
LED4 are
still on
when the
tube is
filled with
fluid or are
off when no
fluid exists
in the tube.
U7, R15, R19,
R32, R17, R13,
R14, R18, R21,
R5, R11, R30,
R22, R12, R31,
R4, R23, R25,
R26, R27 or R28
are damaged (or
rosin jointed).
Medium
Re-solder or
change the
component
Hardware System
5-148
5.8 Laser Control Board 5.8.1 Introduction The laser control board controls the laser to ensure it is stable and moderate.
5.8.2 Board Composition
Function
Power conditioning
Laser drive current monitoring
Constant-laser power controlling
Block diagram
Figure 5-38 Block diagram of the laser control board
Description
Power conditioning
The power conditioning module filters the ±12V power sent by the data board. After filtering,
the ripple is less than 100mV.
Laser drive current monitoring
The working current of the laser is measured and sent to the data board for monitoring.
Constant-laser power controlling
The laser control board controls the laser by constant-power method. I.e. the laser output
Hardware System
5-149
power is monitored by the inner photoelectrical detector, and the results obtained forms a
closed loop system through negative feedback to ensure the constant-power output. The
power is controlled (3mW-5mW) by adjusting the potentiometer VR1 on the board.
Interface definition
The laser control board is configured with 2 external connectors. One is the J1 connecting the
mother board (connecting data board through the mother board), the other one is the J2
connecting the laser, as the figure shows.
Figure 5-39 Interface layout of the laser control board
Table 5-97 Definition of J1 (connecting the mother board)
PIN Definition Note Direction I/O Level
1 AVSS -12V Analog power
supply
I -12V
2 AGND Analog ground O 0
3 LASER Monitoring voltage of
laser drive current O No greater than 5V
4 AVCC 12V Analog power supply I 12V
5 #CONTROL Control signals produced
by standard voltage
I/O Low level: ≤0.8V
High level: OC door
of 5V
6 VCC 5V digital power supply I 5V
Hardware System
5-150
Table 5-98 Definition of J2 (connecting the laser)
No. Name Note Direction I/O
Level
1 LDA Anode of semi-conductor laser diode O -
2 LDC Cathode of semi-conductor laser diode I -
3 PDA Anode of semi-conductor laser receiver I -
Assembly drawing
Figure 5-40 Assembly drawing of the laser control board
5.8.3 Adjustment and Test Points
Adjust the slide rheostat VR1 clockwise, then the resistance of the VR1 connecting circuit, the
voltage of the test point TP2VREF and the laser intensity increases. Adjust the slide rheostat
VR1 anticlockwise, then the resistance of the VR1 connecting circuit, the voltage of the test
point TP2VREF and the laser intensity decreases.
Adjust the voltage of TP2VREF to 3.9-4.1V (adjustable as per the situation; it is the voltage
difference relative to A-12V), then the voltage of TPILD is 0.8-2.0V (it is also the voltage
difference relative to A-12V; note: the laser works normally).
The voltage defined in the laser control board testing process should be used as the latest
standard voltage of TP2VREF and TPILD.
Hardware System
5-151
Never try to plug or unplug the semi-conductor laser before the power supply is cut off.
Do adjust the slide rheostat gradually increased.
See the test points in the table below and their locations on the board in the assembly drawing.
Table 5-99Test points of the laser control board
No. Test point
Signal under test
Function
1 A+12V AVCC The +12V power from the data board
2 A-12V AVSS The -12V power from the data board
3 D5V VCC The 5V power from the data board. It enables the
upper photocoupler.
4 AGND
Digital
ground on
board
\
5 TPVREF VREF The 2.5V standard voltage produced by the U1 chip
(relative to A-12V).
6 TP2VREF 2VREF
Double voltage to ground of the rheostat
sliding-end. It controls the intensity of the laser in
direct ratio.
7 TPIPD PDA
The step-down voltage on R1, R21 or R1/R21
caused by the feedback current signals from the
photoelectrical receiving part of the laser.
8 TPILD LASER
The working current signals of the diode part of the
laser. It reflexes the working current of the diode
part through the voltage of TPILD, detecting
whether the laser is working under the expected
status.
5.8.4 Assembly and disassembly See Chapter 7 Optical System for details.
5.8.5 Troubleshooting The troubleshooting information of the laser control board is listed in the table below.
Hardware System
5-152
Table 5-100 Troubleshooting the laser control board
No. Error Error
Features Possible Causes
Recommended action
Improperly connected
or disconnected with
the mother board
Reconnect the cable or
change a new cable
Improper power sent
to the laser control
board
Check and make sure the
power sent by the mother
board and the data board
is proper. 1
+12V power
supply error
The voltage of
the A+12V test
point exceeds
the range of
12V±0.6V
Something is wrong
with the power
control circuit on the
board. Grounding
short-circuits. The
power is lower or
higher.
Check the circuit on the
board.
2 -12V power
supply error
The voltage of
the A-12V test
point exceeds
the range of
-12V±0.6V
Something is wrong
with the power
control circuit on the
board. Grounding
short-circuits. The
power is lower or
higher.
Check the circuit on the
board.
Improperly connected
or disconnected with
the mother board
Reconnect the cable or
change a new cable
3 5V power
supply error
The voltage of
the D5V test
point exceeds
the range of
5V±0.25V
Improper power is
sent to the laser
control board by the
data board or mother
board.
Check and make sure the
power sent by the mother
board and the data board
is proper.
Improperly connected
or disconnected with
the mother board
Reconnect the cable or
change a new cable
4 Control signal
error
The voltage of
the signal
#CONTROL ≤
0.8V
Improper control
signals sent by the
data board or mother
board
Check and make sure the
signals sent by the mother
board and the data board
are proper.
Hardware System
5-153
5 Reference
voltage error
The voltage of
TPVREF ≠
2.5V±0.1V
(regards -12V
as reference)
First ensure the -12V
power works
normally. Then, the
chip U1 is rosin
jointed or damaged.
Re-solder or change the
U1
6
Laser
intensity
control
reference
voltage error
The voltage of
TP2VREF is
abnormal
(should be
3.9V-4.1V)
First ensure the
TPVREF is normal.
Then, the slide
rheostat is rosin
jointed or damaged.
Or, something is
wrong with the
soldering of the
operation amplifier
U2 and its
peripherals.
Re-solder or change the
components
Laser diode
damaged or
connected improperly
Re-connect or change the
connection cable, or
change the laser diode
5V voltage error See No.3 in the table
Control signals error
or succeeding control
circuit error
See No.4 in the table and
check the succeeding
circuit.
7
Laser
intensity
monitoring
signals TPILD
output error
The voltage of
TPILD
exceeds the
range of
0.8-2V The operation
amplifier U3 and its
peripherals are rosin
jointed or damaged
Re-solder or change the
components
The voltage defined in the laser control board testing process should be used as the latest
standard voltage of TP2VREF and TPILD.
Hardware System
5-154
5.9 Pre-amplification board 5.9.1 Introduction The pre-amplification board conducts the optical-electrical conversion to the two scattered
lights (FS-low-angle scattered light; SS-high-angle scattered light) come from the flow cell and
amplifies the signals. The amplifying boards of the low-angle signal and the high-angle signal
share one PCB. The amplification factor is obtained by soldering the resistors.
5.9.2 Board Composition
Function
Power conditioning
Optical-electrical converting
Signal conditioning
Block diagram
Figure 5-41 Block diagram of the pre-amplification board
Description
Power conditioning
The power conditioning module filters the ±12V power sent by the data board. After filtering,
the ripple is less than 50mV.
Optical-electrical converting
The optical-electrical converting module converts the optical signals into the electrical signals
by the photodiode.
Hardware System
5-155
Signal conditioning
The current signals are transferred into the voltage signals through I/V and then sent to the
succeeding amplifying-conditioning module for further processing. Then, the data board input
required signals are obtained.
Interface definition
The FS/SS pre-amplification board is configured with one external connector respectively, i.e.
the J1 connector connecting the mother board (connecting data board through the mother
board). The layout of the board is shown below.
Figure 5-42 Interface layout of the pre-amplification board
Table 5-101 Definition of J1 (connecting the mother board)
PIN Definition Note Direction I/O
Level
1 AGND Analog ground O 0
2 FS/SS Pre-amplification board output voltage signals <1V
3 SHELL Shielding ground - 0
4 AVSS A-12V analog power supply I -12V±5%
5 AGND Analog ground O 0
6 AVCC A+12V analog power supply I 12V±5%
The laser control board is configured with 2 external connectors. One is the J1 connecting the
Hardware System
5-156
mother board (connecting data board through the mother board), the other one is the J2
connecting the laser.
Figure 5-43 Interface layout of the laser control board
Table 5-102 Definition of J1 (connecting the mother board)
PIN Definition Note Direction I/O Level
1 AVSS -12V Analog power supply I -12V
2 AGND Analog ground O 0
3 LASER Monitoring voltage of
laser drive current O ≤5V
4 AVCC 12V Analog power supply I 12V
5 #CONTRO
L
Control signals produced
by standard voltage
I/O Low level: ≤0.8V
High level: OC
door of 5V
6 VCC 5V digital power supply I 5V
Table 5-103 Definition of J2 (connecting the laser)
No. Name Note Direction I/O Level
1 LDA Anode of semi-conductor laser diode O -
2 LDC Cathode of semi-conductor laser diode I -
3 PDA Anode of semi-conductor laser receiver I -
Hardware System
5-157
Assembly drawing
(a) Top
(b) Bottom
Figure 5-44 Assembly drawing of the FS/SS pre-amplification board
Hardware System
5-158
5.9.3 Adjustment and Test Points No adjustment components. See the test points in the table below and their location on the
board in the assembly drawing.
Table 5-104 Test points definition of the pre-amplification board
No. Test point
Signal under test
Function
1 AGND
Digital
ground on
board
\
2 +12V AVCC The +12V power from the data board
3 -12V AVSS The -12V power from the data board
4 OUT OUT FS/SS pre-amplification board output signals
5 D5V VCC The 5V power from the data board. It enables the
upper photocoupler.
5.9.4 Disassembly and assembly See Chapter 7 Optical System for details.
5.9.5 Troubleshooting
Table 5-105 Troubleshooting the pre-amplification board
No. Error Error Features Possible Causes Recommended action
Improper cable connection
with the mother board or the
data board, or disconnection
caused by a broken cable
Reconnect the cable or
change a new cable
Improper 12V power sent by
the data board or mother
board
Check whether the proper
power supply is sent by
the mother board or the
data board
1
12V
power
supply
error
The voltage of the
+12V test point
exceeds the range of
12±0.6V; or the
ripple noise is
greater than 50mV. Rosin joint or damage
happens to the +12V test
point relating inductor L1 or
the capacitors C24, C27, C20
and C23.
Re-solder or change the
component
Hardware System
5-159
Improper cable connection
with the mother board or the
data board, or disconnection
caused by a broken cable
Reconnect the cable or
change a new cable
Improper -12V power sent by
the data board or mother
board
Check whether the proper
power supply is sent by
the mother board or the
data board
2
-12V
power
supply
error
The voltage of the
-12V test point
exceeds the range of
-12±0.6V; or the
ripple noise is
greater than 50mV. Rosin joint or damage
happens to the -12V test point
relating inductor L2 or the
capacitors C22, C26, C25 and
C21.
Re-solder or change the
component
The FS board is installed in
the location of the SS board
by mistake.
Change the board SS output signals
too weak (supposing
it is caused only by
the circuit problems) R4, R5 or R6 resistance error,
rosin jointed or damaged
Re-solder or change the
component
The SS board is installed in
the location of the FS board
by mistake.
Change the board
FS output signals
too strong
(supposing it is
caused only by the
circuit problems) R4, R5 or R6 resistance error,
rosin jointed or damaged
Re-solder or change the
component
The photodiode D1 or R1 is
rosin jointed or damaged.
Re-solder or change the
component
The
pre-amplification
board (FF/SS)
output signals are
too weak to be
detected.
The U2 or U3 is rosin jointed
or damaged.
Re-solder or change the
component
3
Channel
output
value
error
Something is wrong
with the band width
of the
pre-amplification
board (FF/SS).
Check whether the relating
resistors and capacitors R4,
R5, R6, C1, C5 and C4 are
rosin jointed or damaged.
Re-solder or change the
component
Hardware System
5-160
5.10 Indicator Board 5.10.1 Introduction The indicator board realizes the functions of analyzer status indicating and buzzer alarming.
5.10.2 Board Composition
Function
Analyzer working status indicating
Buzzer alarming
Description
Analyzer working status indicating
The indicator board indicates the working status of the analyzer by a two-color indicator. When
the system works normally, the indicator displays green. When error happens, the indicator
turns red.
Buzzer alarming
The alarming function is realized by the buzzing of the buzzer.
Interface definition
The indicator board is configured with one external connector J1.
Table 5-106 Interface definition of the indicator board
PIN Name Note Direction
I/O Level
1 VCC Power supply - +5V
4 FAULT Error indicator (red) I TTL
5 WORK Working indicator
(green) I TTL
7 BUZ Buzzer I TTL
8 GND Ground - 0
Hardware System
5-161
Assembly drawing
Figure 5-45 Assembly drawing of the indicator board
5.10.3 Disassembly and assembly method
Purpose
Damaged indicator board needs to be replaced by a new one.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Disassembly
1. Shut down the analyzer.
2. Open the front cover assembly.
3. Use the 107 cross-headed screwdriver to remove the 2 tapping screws (PT3X6) fixing the
indicator board, and then remove the board.
Hardware System
5-162
Figure 5-46 Disassemble the indicator board and the key board
1---Key board 2--- Tapping screw (PT3X8)
3---Front cover 4--- Indicator board
5--- Tapping screw (PT3X6)
Assembly
Assemble the indicator board as per the above-mentioned steps in the reverse order.
Verification
Turn on the analyzer and start up the PC when finish replacing the indicator board. During the
process from “startup self-test” to “ready”, closely observe the indicator and the buzzer, if they
work normally, the replacement is successful.
5.10.4 Troubleshooting
Table 5-107 Troubleshooting the indicator board
No. Error Causes and recommended action
1 Indicator doesn't work If the indicator could work off and on when you shaking
the wiring on the J1 connector, then, it might be caused
by the improper connection. If nothing is wrong with the
connection and the system works normally, then, the
diode of the indicator might be damaged.
2 The buzzer is mute when
system alarming.
If the buzzer could buzz off and on when you shaking
the wiring on the J1 connector, then, it might be caused
by the improper connection. If nothing is wrong with the
connection or the buzzer buzzes abnormally (e.g. a
hoarse buzzing), then, the buzzer might be damaged.
Hardware System
5-163
5.11 Key Board 5.11.1 Introduction The key board enables the operator to interact with the main unit through the [RUN] and
[OPEN] keys during the autoloading/manual loading process.
5.11.2 Board Composition
Function
[RUN] key
[OPEN] key
Description
Use the [RUN] key
The operator can press the [RUN] key to start autoloading.
Use the [OPEN] key
The operator can press the [OPEN] key to open the sample compartment door.
Interface definition
The indicator board is configured with one external connector J1.
Table 5-108 Interface definition of the key board
PIN Name Note Direction
I/O Level
2 COUNT_KEY [RUN] key - TTL
3 INSERT_KEY [OPEN] key I TTL
4 GND Ground I 0
Hardware System
5-164
Assembly drawing
Figure 5-47 Assembly drawing of the key board
5.11.3 Disassembly and assembly method
Purpose
Damaged key board needs to be replaced by a new one.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Disassembly
1. Shut down the analyzer.
2. Open the front cover assembly.
3. Use the 107 cross-headed screwdriver to remove the 4 tapping screws (PT3X8) fixing the
key board, and then remove the board.
Hardware System
5-165
Figure 5-48 Disassemble the key board and the indicator board
1---Key board 2--- Tapping screw (PT3X8)
3---Front cover 4--- Indicator board
5--- Tapping screw (PT3X6)
Assembly
Assemble the key board as per the above-mentioned steps in the reverse order.
Verification
Turn on the analyzer and start up the PC when finish replacing the key board. Wait for the PC
interface to show “ready”, and then press the [OPEN] key in the front cover. If the sample
compartment door opens, then the key board is successfully replaced.
5.11.4 Troubleshooting As the structure of the key board is quite simple, it seldom has errors. If there is no response
after pressing the keys, connect the 2 and 4 pins, as well as the 3 and 4 pins of the J1 interface.
If the circuit is closed after pressing one of the keys, the key board is out of error and you
should check other components for errors; if the circuit is open, the key is not working properly.
6-1
6 Maintenance
6.1 Maintenance Modules and the Corresponding Settings
No. Type Maintenance Settings to be reset
1 Replace the drive board Re-adjust the position of the sampling assembly
by software
2 Replace the autoloader board
Re-adjust the position of the clipper by software
3 Replace the data board Re-adjust the gains, calibration factors and count-related settings
4 Replace the power board
Re-conduct the safety test 5
Boards
Replace the volumetric board Reset the count time and the start signal delay
6
Remove the sample feeding assembly from the analyzer and then re-assemble it
Manually adjust the horizontal position of the sample feeding assembly Re-adjust the position of the sampling assembly to the sample compartment and the autoloading position by software.
7
Remove the sample compartment assembly from the analyzer and then re-assemble it
1. Manually adjust the position of the sample compartment on the sample feeding assembly. 2. Re-adjust the position of the sampling assembly to the sample compartment by software.
8 Disassemble and assemble the sample feeding assembly
Re-assemble the sample feeding assembly as per the steps specified in Section 6.8
9
Disassemble and assemble the DIFF bath assembly or have it upgraded
1. Re-adjust the position of the sampling assembly to the DIFF bath. 2. If the temperature and the mixing are greatly changed, then re-adjusting the optical gain is necessary.
10
Remove the DIFF bath assembly from the analyzer and then re-assemble it
Re-adjust the position of the sampling assembly to the DIFF bath by software
11
Remove the mix assembly from the analyzer and then re-assemble it
1. Manually adjust the horizontal position of the mix assembly. 2. Re-adjust the position of the clipper by software.
12
Disassemble and assemble the clipper or have it upgraded
1. Locate the clipper to the right position by clamp. 2. Re-adjust the position of the clipper by software.
13 Disassemble and assemble the mix assembly
Re-assemble the mix assembly as per the specified process.
14
Mechanism
Replace and disassemble/assemble the front cover
1. Manually adjust the horizontal position of the mix assembly. 2. Re-adjust the position of the clipper by software.
Maintenance
6-2
15
Replace and disassemble/assemble the bottom cover
1. Manually adjust the horizontal position of the sample feeding assembly. 2. Re-adjust the position of the sampling assembly to the sample compartment and the autoloading position by software.
16 Replace the aperture 1. Re-adjust the gains and the calibration factors.2. Reset the count time and the start signal delay.
17 Replace the bath 1. Re-adjust the gains and the calibration factors.2. Reset the count time and the start signal delay.
18 Replace the volumetric tube
1. Re-enter the volume of the volumetric tube at the setting screen. 2. Reset the count time and the start signal delay.
19
Baths
Replace HGB Re-adjust the gain and calibration of HGB.
20 The vacuum created in the sequence is changed Reset the count time and the start signal delay
21
Sequence Sample results are changed greatly Re-adjust the gain and the calibration.
Maintenance
6-3
6.2 General You can service the analyzer as per Section 6.3 – 6.7 . After repairing, follow the verification
procedures to verify the analyzer’s status.
Maintenance
6-4
6.3 Disassembling the Panels
6.3.1 Removing the left door
Purpose
To disassemble the electric and fluidic parts inside the left side of the analyzer, follow this
procedure to remove the left door.
Tools
107 flat-headed screwdriver
Removal
1. Power off the analyzer and unplug the power cord
2. Insert 107 the flat-headed screwdriver into the slot of the door lock, and turn the lock
tongue 90° counterclockwise.
3. Pull the side door outward from the bottom.
Figure 6-1 Removing the left door
Installation
Install the left door as per the above-mentioned steps in the reverse order.
Maintenance
6-5
6.3.2 Removing the right door
Purpose
To disassemble the electric and fluidic parts inside the right side of the analyzer, follow this
procedure to remove the right door.
Tools
107 flat-headed screwdriver
Removal
1. Power off the analyzer and unplug the power cord.
2. Insert 107 flat-headed screwdriver into the slot of the door lock, and turn the lock tongue
90° counterclockwise.
3. Pull the side door outward from the bottom.
Make sure that the power supply is shut off when you remove the right door; otherwise, the analyzer will alarm and stop running.
Figure 6-2 Removing the right door
Maintenance
6-6
Installation
Install the right door as per the above-mentioned procedures in the reverse order.
6.3.3 Removing the top cover
Purpose
To disassemble the optical components, boards, cables and connectors, follow this procedure
to remove the top cover.
Tools
107 cross-headed screwdriver
Removal
1. Power off the analyzer and unplug the power cord.
2. Use the 107 cross-headed screwdriver to remove the three M4X8 small panhead screws.
3. Pull out the top cover from back upward.
Maintenance
6-7
Figure 6-3 Removing top cover
1 ―――M4X8 small panhead screw 2 ―――top cover
Installation
Install the top cover as per the above-mentioned procedures in the reverse order.
6.3.4 Removing the back panel
Purpose
To maintain the fluidic connectors and electric circuits at the center of analyzer, follow this
procedure to remove the back panel.
Tools
107 cross-headed screwdriver
Maintenance
6-8
Removal
1. Power off the analyzer and unplug the power cord.
2. Use the 107 cross-headed screwdriver to remove the two M4X8 small panhead screws
used for fixing the back panel of power supply。
3. Pull the back panel outward.
Figure 6-4 Removing the back panel of power supply
1 ―――M4X8 small panhead screw 2 ―――back panel of power supply
Installation
Install the back panel as per the above-mentioned procedures in the reverse order.
6.3.5 Removing the front door
Purpose
To maintain the fluidic system inside the front cover, follow this procedure to remove the front
door.
Tools
107 cross-headed screwdriver
Maintenance
6-9
107 flat-headed screwdriver
Removal
1. Power off the analyzer and unplug the power cord.
2. Remove the right door as per section 6.3.3 .
3. Turn the handspike handle 90° (clockwise and counterclockwise) to make it slide into the
slot in the front board, and the front door will be pushed open by the handspike.
4. Lift the front door upwards.
Figure 6-5 Removing the front door
1 ―――Handspike handle 2 ―――Handspike
3 ―――Back panel 4 ―――Mix mechanism
5 ―――Front panel 6 ―――Front door
6.3.6 Removing the front cover assembly
Purpose
To maintain the fluidic system, syringes and sample transport unit inside the front cover, follow
this procedure to remove the front cover.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Maintenance
6-10
Removal
1. Power off the analyzer and unplug the power cord.
2. Remove the left door as per section 6.3.1 .
3. Remove the right door as per section 6.3.2 .
4. Remove the top cover board as per section 6.3.3 .
5. Make sure there are no tubes or tube racks left at the sample transport area (if there are,
remove them).
6. Use the 107 cross-headed screwdriver to remove the six M4X8 countersunk head screws
on the left, right and top side of the front cover.
7. Pull the front cover outward by 50 - 80 mm, and disconnect the indicator connector from
the indicator board.
8. Remove the front cover forward.
Figure 6-6 Removing the front cover assembly
1 ―――M4X8 small panhead screw 2 ―――top cover
3 ―――Key board 4 ―――indicator light board
5 ―――right door 6 ―――M4X8 countersunk head screw
7 ―――front cover assembly 8 ―――left door
Installation
Install the front cover as per the above-mentioned procedures in the reverse order.
Maintenance
6-11
6.4 Replacing the Valves, Pumps and Syringes (Reagent container and reagent inlet, waste bump, fluidic valve, pinch valve, syringe)
Figure 6-7 Removing fluidic system inside the left door
1 ―――valve assembly I 2 ―――M4X12 small panhead screws
3 ―――plain washer Ф3 4 ―――M3X12 cross-headed panhead screw
(with lock washer)
5 ―――diluent syringe 6 ―――lyse syringe
7 ―――M4X8 small panhead screws 8 ―――pressure chamber
9 ―――air pump 10―――valve assembly II
Maintenance
6-12
Figure 6-8 Removing fluidic system inside the right door
1 ―――valve 2 ―――M3X8 small panhead screw
3 ―――M4X12 small panhead screw 4 ―――vacuum chamber
5 ―――waste pump assembly 6 ―――valve assembly III
7 ―――pinch valve
Maintenance
6-13
Figure 6-9 Removing fluidic system inside the front door
1 ―――valve 2 ―――pressure-leakage valve
3 ―――M3X8 small panhead screw 4 ―――sheath syringe
5 ―――injection syringe 6 ―――sampling syringe
7 ―――plain washer Ф3 8 ―――M3X12 cross-headed panhead screw
(with lock washer)
6.4.1 Replacing valves
Purpose
In case of malfunction of a valve(s), replace the valve.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Removal
Follow these two procedures to disassemble the valve assemblies.
To remove the valve assembly (valve assembly I, valve assembly II, or valve assembly III)
Maintenance
6-14
1. To remove valve assembly I and II, remove the left door (as per section 6.3.1 ); to remove
valve assembly III, remove the right door (as per section 6.3.2 ).
2. Disconnect all tubes connected to the valve(s) as shown in Figure 6-7, Figure 6-8, and
Figure 6-9.
3. Use the 107 cross-headed screwdriver to remove the M4X12 small panhead screws
fastening the valve assembly.
4. Pull the valve assembly, outward and disconnect all lines from the assembly..
5. Take out valve assembly.
To remove a single valve (including pinch valve, or single valve on the valve assemblies)
1. To remove the valve on the left side, remove the left door (as per section 6.3.1 ); to remove
the valve on the right side, remove the right door (as per section 6.3.2 ); to remove the
valve on the front side, remove the front cover assembly. (as per section 6.3.2 )
2. Disconnect all pipes connected to the valves as shown in Figure 6-7, Figure 6-8, and
Figure 6-9 (if a pinch valve needs to be removed, remember to pinch the corresponding
tube).
3. Use the 107 cross-headed screwdriver to remove the two M3X8 small panhead screws
fastening the valve assembly.
4. Pull the valve outward, and disconnect the lines connected to the valve. Then take the
valve out.
Figure 6-10 Valve assembly I
Maintenance
6-15
Figure 6-11 Valve assembly II
Figure 6-12 Valve assembly III
1 ―――valve 2 ―――M3X8 small panhead screw
3 ―――valve bracket
Installation
Install the valve assembly as per the above-mentioned procedures in the reverse order.
Verification
1. Check the fluidic connections.
2. Power on and verify the installation.
6.4.2 Replacing the Pressure Chamber
Purpose
In case of malfunction, replace the pressure chamber.
Maintenance
6-16
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Removal
Follow these two procedures to disassemble the pressure chamber.
To remove the pressure chamber:
1. Remove the left door (as per section 6.3.1 ).
2. Disconnect all tubes connected to the assembly as shown in Figure 6-7.
3. Use the 107 cross-headed screwdriver to remove the four M4X8 small panhead screws.
4. Pull the pressure chamber out.
To disassemble the pressure chamber:
1. Use the 107 cross-headed screwdriver to remove the four M3X12 panhead washer screws,
then take out the end cap and mounting plate;
2. Use the 107 flat-headed screwdriver to remove the airproof screw fixed on the end cap,
then take out the small O ring (9.5x1.8).
3. Take out the big O ring (48.7x1.8).
Figure 6-13 Removing pressure chamber
1―――M3X12 cross-headed panhead
screw (with lock washer)
2 ―――end cap
3 ―――O ring 48.7x1.8 4 ―――O ring 9.5 x 1.8
5 ―――airproof screw 6 ―――mounting plate
7 ―――air pressure chamber
Maintenance
6-17
Installation
Install the pressure chamber as per the above-mentioned procedures in the reverse order.
Verification
1. Check if all parts are correctly assembled and tightened.
2. Check the fluidic connections..
3. Power on and verify the installation.
6.4.3 Replacing the Vacuum Chamber
Purpose
In case of malfunction, replace the vacuum chamber.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Removal
Follow these two procedures to disassemble the vacuum chamber.
To remove the vacuum chamber:
1. Remove the right door (as per section 6.3.2 )
2. Disconnect all tubes connected to the assembly as shown in Figure 6-8.
3. Use the 107 cross-headed screwdriver to remove the two M4X12 small panhead screws.
4. Pull vacuum chamber outward.
To disassemble the vacuum chamber:
1. Use the 107 cross-headed screwdriver to remove the four M3X12 panhead washer screws,
then take out the end cap and mounting plate.
2. Use the 107 flat-headed screwdriver to remove the air tightening screw fixed on the end
cap, then take out the small O ring (9.5x1.8).
3. Take out the big O ring (48.7x1.8).
Maintenance
6-18
Figure 6-14 Removing vacuum chamber
1―――M3X12 cross-headed panhead
screw (with lock washer)
2 ―――end cap
3 ―――O ring (48.7x1.8) 4 ―――O ring (9.5x1.8)
5 ―――airproof screw 6 ―――mounting plate
7 ―――air pressure chamber
Installation
Install the vacuum chamber as per the above-mentioned procedures in the reverse order.
Verification
1. Check if all parts are correctly assembled and tightened.
2. Check the reagent connections.
3. Power on and verify the installation.
6.4.4 Replacing the Syringe Assembly
Purpose
In case of malfunction, replace the syringe assembly.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
2 mm hexagon wrench
Maintenance
6-19
2.5 mm hexagon wrench
150 mm monkey wrench
Removal
Follow these two procedures to disassemble the syringe.
To remove the syringe assembly
1. To the remove lyse syringe or diluent syringe, remove the left door (as per section 6.3.1 );
to the remove sheath syringe, injection syringe or sampling syringe, remove the front
cover assembly (as per section 6.3.5 ).
2. Disconnect all tubes connected to the syringe as shown in Figure 6-7 and Figure 6-9.
3. Use the 107 cross-headed screwdriver to remove the four M3X12 panhead washer screws
fixing syringe and Ф3 plain washer. .
4. Incline the assembly outward, and pull it out by 30 - 50mm. Disconnect the wire of the
motor and the photocoupler, and then take the assembly out. (* Note that the four rubber
washers of the syringe should be kept well for reassembly.)
To disassemble the syringe:
1. Remove the lyse syringe, injection syringe, and sample syringe
a. To remove the glass syringe unit on the lyse syringe (three 2.5ml syringes on the
assembly), remove the left door (as per section 6.3.1 ).
b. To remove the glass syringe unit on the injection syringe or on the sample syringe (250ul
and 100ul respectively), remove the front cover (as per section 6.3.5 or 6.3.6 ).
c. The syringe assembly needs not to be disassembled wholly. Use the 107 flat-headed
screwdriver to remove the M4X8 special screw fixing the syringe. Rotate the left glass
syringe unit with hand, and then take out the washer of syringe (as per Figure 6-15).
Apply some glue to the threads. Do not proceed to the next step until the glue is dry.
Maintenance
6-20
Figure 6-15 Removing glass syringe unit
1 ―――M4X8 special screw 2 ―――glass syringe unit
3 ―――washer
d. Use the 107 cross-headed screwdriver to remove the two M3X8 socket head screws on
the electric switch and remove the switch. Use the 107 cross-headed screwdriver to
remove the two M3X6 cross-headed panhead screws, and then remove the electric switch
(Figure 6-16).
Maintenance
6-21
Figure 6-16 Removing syringe assembly
1――MGN9-C1-R95-Z0CM Linear
rolling guide
2 ―――syringe slider
3 ―――M3X8 socket head screw 4 ―――43F4J-05-010 linear stepping motor
5 ―――M3X6 cross-headed panhead
screw
6 ―――PHOTOELEC Optical Sensor
7 ―――syringe photocoupler bracket 8 ―――M3X8 cross-headed panhead screw
9 ―――Syringe protection shield 10 ―――M4X8 retaining screw
e. Use the 107 cross-headed screwdriver to remove the syringe protection shield. Use
2.5mm hexagon wrench to remove the four M3X8 socket head screws on syringe slider.
Use 2 mm hexagon wrench to remove M4X8 retaining screw on syringe slider and remove
the syringe slider. Use 2.5 mm hexagon wrench to remove the four M3X8 socket head
screws fastening the linear stepping motor , then remove the linear stepping motor (Figure
Maintenance
6-22
6-16);
Make sure to insert the retaining screw into the corresponding hole, and apply some glue to the threads.
f. Use 2.5mm hexagon wrench to remove the three M3X8 socket head screws on linear
rolling guide, and then remove the linear rolling guide. (*Note: The slider should not be off
its guide way.)
2. Remove diluent syringe and sheath syringe (*Note: Two syringes are the same, both being
the 10ml Mindray syringe.)
a. To remove the Mindray syringe unit on the diluent syringe, remove the left door (as per
section 6.3.1 )
b. To remove the Mindray syringe unit on the sheath syringe, remove the front cover
assembly (as per section 6.3.5 or 6.3.6 )
c. The syringe assembly should not be removed wholly. Use the 107 cross-headed
screwdriver and flat-headed screwdriver to remove the 10ml syringe presser the two
tailor-made M3X12 socket head screws, and remove the Mindray syringe unit. (Figure
6-17)
Figure 6-17 Removing Mindray syringe unit
1 ―――10ml syringe presser 2 ―――M3X12 socket head screw
3 ―――Mindray syringe unit 4 ―――M4X8 tailor-made screws
Maintenance
6-23
d. Use the 107 cross-headed screwdriver to remove the four M3X10 panhead screws. The
ceramic plug, flange, lock ring and O ring can be disassembled in the axis direction.
(Figure 6-18)
Figure 6-18 Removing Mindray syringe unit
1 ―――10ml ceramic plug 2 ―――PT3X10 cross-headed panhead
screw
3 ―――10ml syringe flange 4 ―――10ml syringe lock ring
5 ―――O ring 15X3 6 ―――10ml syringe body
Before replacing the lock ring, make sure relative parts (especially the operation surface of the ceramic plug) are clean, and then insert the plug into the flange, lock ring and O ring in turn and make it fixed to the syringe body.
Installation
Install the Mindray syringe unit as per the above-mentioned procedures in the reverse order.
Verification
1. Check if all parts are correctly assembled and tightened.
2. Check wire connections.
3. Check reagent connections.
4. Power on and verify the installation.
Maintenance
6-24
6.4.5 Replacing the Waste Pump
Purpose
In case of malfunction, replace the waste pump.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
100 mm monkey wrench
Removal
Follow these two procedures to disassemble the waste pump.
To remove the waste pump:
1. Power on the analyzer to empty the fluidic system. Power off the analyzer and unplug the
power cord from the wall outlet.
2. Remove the right door (as per section 6.3.2 ).
3. Disconnect the four tubes connected to the pump;
4. Disconnect the two wires connected to the pump;
5. Use the 107 cross-headed screwdriver to remove the two M4X12 small panhead screws
on fastening assembly as shown in Figure 6-8.
6. Move the assembly left, make the fastening hole out of the bolt, and then remove it.
To disassemble the waste pump:
1. Use the 107 cross-headed screwdriver to remove the four M3X8 small panhead screws
fixing the pump.
2. Remove the waste pump
3. Use the 100 mm monkey wrench to remove the three nuts and take out the three washers.
4. Remove the pump mounting plate.
5. Remove three shock pads via hand.
Maintenance
6-25
Figure 6-19 Removing waste pump
1 ―――M3X8 small panhead screws 2 ―――rotary pump
3 ―――pump mounting plate 4 ―――8-32UNC-2B nut
5 ―――GB93 4 washer 6 ―――shock pad
7 ―――bracket
Installation
Install the waste pump as per the above-mentioned procedures in the reverse order.
Verification
1. Check if all parts are correctly assembled and tightened.
2. Check the electrical connection of the rotatory pump.
3. Check the reagent connections.
4. Power on and verify the installation.
Maintenance
6-26
6.5 Replacing the Bath/Aperture Assembly
Purpose
In case of malfunction of the WBC or RBC reaction bath assembly, replace the corresponding
assembly.
Tools
107 cross-headed screwdriver
107 flat-headed screwdriver
Removal
1. Power on the analyzer to empty the fluidic system. Power off the analyzer and unplug the
power cord from the wall outlet.
2. Remove the left door (as per section 6.3.1 ).
3. Remove the right door (as per section 6.3.1 ).
4. Remove the top cover (as per section 6.3.1 ).
5. Use the 107 cross-headed screwdriver to remove the two M3X6 cross-headed washer
screws on the top cover of data board shielding box, and then remove the top cover.
6. Disconnect the bath signal lines of the WBC reaction bath or RBC reaction bath that needs
to be removed, and then loosen all buckles fastening the lines in the direction of wiring,
until the wire is drawn to the bath.
7. Use the 107 cross-headed screwdriver to remove M3X8 small panhead screw on bath
shield and take it out, then remove the shield. (Figure 6-21)
8. Use the 107 cross-headed screwdriver to remove the two M3X8 small panhead screws on
bath presser and take it out, then remove the bath presser. (Figure 6-21)
9. Pull the bath out with a distance of about 50-80 mm. Disconnect all tubes connected to the
bath. Get the signal wire from wire-protective ring and then remove the WBC bath
assembly or the RBC bath assembly.
Maintenance
6-27
Figure 6-20 Removing signal lines of bath
1 ―――M3X6 cross-headed panhead
washer screw
2 ―――top cover of data board shielding box
3 ―――data board
Figure 6-21 Removing bath assembly
1 ―――shielding plate 2 ―――WBC bath
3 ―――bath presser 4 ―――M3X8 small panhead screw
5 ―――RBC bath 6 ―――shield cover
7 ―――isolation chamber clamp 8 ―――isolation chamber
Maintenance
6-28
Installation
Install the bath assembly as per the above-mentioned procedures in the reverse order.
Verification
1. Check if all parts are mounted to the former position and are fastened.
2. Check if the connection of signal lines of WBC or RBC is correct or not.
3. Check if the connection of power line and reagent pipe is correct or not.
4. Power on and verify the installation.
Maintenance
6-29
6.6 Replacing the Sampling Module and Adjusting Position
6.6.1 Replacing the Sample Probe and Wipe
Purpose
To remove or replace the sample probe and probe wipe, follow the steps in this section.
Tools
107 cross-headed screwdriver (107X75)
Replace parts according to requirements:
Probe wipe
open vial sample probe / piercing needle
metal wipe clip
sample probe presser
Preparation
1. Power off the analyzer if it is running.
2. Turn off the main power switch on the right side of the analyzer and unplug the power cord.
3. Remove the right door (as per section 6.3.2 ).
4. Move the sampling module to the back of analyzer.
Removal
1. Remove the probe wipe from the sampling module (Figure 6-22)
a. Disconnect the two tubes from the probe wipe.
b. Pull the wipe downward to remove the probe wipe, as indicated by the arrow in the
figure below, and remove the wipe clip.
c. Separate the wipe from sampling module in the direction of arrow
d. Get out the inlet and outlet pipe from the wipe.
Maintenance
6-30
Figure 6-22 Removing wipe
1 ―――pipe clamp 2 ―――presser under the guiding rod
3 ―――probe wipe 4 ―――wipe clip
2. Remove the sample probe from the sampling module. (Figure 6-23)
a. Pull out the sample aspiration tube from the above of sample probe.
b. Use the cross-headed screwdriver to remove the retaining screw (M3X8 small
panhead screw) on probe presser, and take out the fixing flake.
c. Get out the probe from sideward.
Maintenance
6-31
Figure 6-23 Removing sample probe
1 ―――slider 2 ―――sample probe
3 ―――presser 4 ―――M3X8 small panhead screw
When remove the sample probe, get probe wipe out of the presser under guiding rod. The probe tip may contain biohazardous materials. Exercise caution when handling it and keep it to the right position.
Installation
To install a new sample probe and wipe, repeat the above “removal” step 1 or 2 in the reverse
order.
6.6.2 Replacing the Optical Sensor
Purpose
To remove or replace the optical sensor, follow the steps in this section.
Maintenance
6-32
Tools
107 cross-headed screwdriver(107X75)
Replace parts according to requirements:
Sensor module on motor position(3003-21-34925)
Photocoupler fixer
Preparation
1. Power off the analyzer and unplug the power cord.
2. Turn off the main power switch on the right side of the analyzer and unplug the power cord.
3. Remove the right door (as per section 6.3.2 ).
Removal
1. Remove the vertical optical sensor module (Figure 6-24)
a. Disconnect the connector of the sensor to be replaced ( not shown in the figure).
b. Use the cross-headed screwdriver to remove the M3x6 screw on the sensor.
c. Remove the sensor module from the bracket.
Figure 6-24 Removing vertical optical sensor
1 ―――M3x6 cross-headed screw 2 ―――sensor module on motor position
3 ―――bracket
2. Remove the horizontal optical sensor module (Figure 6-25)
a. Use the cross-headed screwdriver to remove the M3x6 screw on the photocoupler
Maintenance
6-33
presser to be replaced
b. Gently pull out the photocoupler presser (with the sensor module on motor position)
from the sample mother board and get out the plug of sensor module from the
corresponding hole.
c. Disconnect the sensor connector (not shown in figure). Take out the photocoupler
presser and the sensor module on motor position.
d. Use the cross-headed screwdriver to remove the M3x6 screw fixing the sensor module.
Separate the photocoupler presser from the sensor module on the motor position.
Figure 6-25 Removing optical sensor on horizontal position
Installation
To install a new optical sensor, repeat the above “removal” step 1 or 2 in the reverse order.
6.6.3 Removing the Sampling Module
Purpose
To remove the sampling module, follow steps in this section.
Tools
107 cross-headed screwdriver (107X75)
wire cutter
Replace parts according to requirements:
Sampling module
Maintenance
6-34
Preparation
1. Power off the analyzer.
2. Turn off the main power switch on the right side of the analyzer and unplug the power cord.
3. Move the analyzer to turn its right side outward, and expose the back side as much as
possible.
4. Open the back cover of the analyzer (as per section 6.3.2 )
5. Remove the right side door (as per section 6.3.2 )
6. Move the sampling module to the back of the analyzer.
Removal
1. Disconnect all tubes from on the sampling module, and cut the binding belt with the wire
cutter. Remove the tubes from the whole sampling module.
2. Disconnect all the connectors of the vertical part of sampling module, and unplug the
connector from the motor directly.
3. Disconnect all the connectors of the horizontal part of sampling module from the space left
by opened back cover at the back of the analyzer. (BC-5380 has four connectors, and
BC-5380 has five connectors). Unplug the connectors from the motor directly.
4. Use the cross-headed screwdriver to remove the two M3X8 small panhead screws fixing
the joint.
5. Use the cross-headed screwdriver to remove the four M4X12 cross panhead screws fixing
the sampling module.
6. Hold the sampling module and move it gently outward. Pull out the optical sensor module
from corresponding hole with caution. Be careful not to damage or break the line.
7. Check if all connectors of optical sensor module are drawn out from the corresponding
holes, then hold out the sampling module.
Maintenance
6-35
Figure 6-26 Removing the sampling module
1 ―――M3X8 small panhead screw 2 ―――link
3 ―――spacer 4 ―――sampling module
5 ―――M4X12 panhead screw
Watch for the sampling probe when removing sampling module. If necessary, remove sampling probe first as introduced in section 6.6.1 to avoid potential biologic pollution.
Maintenance
6-36
Installation
To replace new sampling module, repeat the above step 1 to 7 in the reverse order.
Maintenance
6-37
6.7 Maintaining and Replacing the DIFF Reaction Bath
Purpose
To remove DIFF reaction bath, follow steps in this section.
Tools
107 cross-headed screwdriver (107X75)
wire cutter
Replace parts according to requirements:
DIFF reaction bath
Preparation
1. Power off the analyzer.
2. Turn off the main power switch on the right side of the analyzer and unplug the power cord.
3. Remove the right side door (as per section 6.3.2 )
Removal
1. Use the wire cutter to cut off the binding belts fixing the tubes and remove them from the
DIFF bath assembly.
2. Use the cross-headed screwdriver to remove the two M4X8 small panhead screws fixing
the DIFF bath assembly.
3. Hold the DIFF bath assembly and move it gently outward. Expose the connectors and pipe
connectors at the back side, and then disconnect them.
4. Take out the DIFF bath assembly.
Maintenance
6-38
Figure 6-27 Removing DIFF bath assembly
1 ―――DIFF bath assembly 2 ―――M4X8 small panhead screw
Installation
To replace new DIFF bath assembly, repeat the above removal step 1 to 4 in the reverse order.
Remember to readjust the position of the sample probe over the DIFF bath as per Step 11 to
14 in Section 6.10.1 .
Maintenance
6-39
6.8 Autoloader Assembly (Closed-Tube) 6.8.1 Removing and replacing the closed-tube sample compartment assembly
Purpose
To remove or replace the closed-tube sample compartment, follow steps in this section.
Tools
107 cross-headed screwdriver (107X75)
1.5mm hexagonal wrench
sharp-nose pliers or small flat-headed screwdriver
Replace parts according to requirements:
Damp gear
PHOTOELEC switch
Gear
Electromagnet
Preparation
1. Power off the analyzer.
2. Turn off the main power switch on the right side of the analyzer and unplug the power cord.
3. Remove the right side door (as per Section 6.3.2 )
4. Remove the left side door (as per Section 6.3.1 )
5. Remove the top cover (as per Section 6.3.3 )
Removal
Follow these two procedures to disassemble the waste pump.
To remove the sample compartment assembly:
1. Remove the front cover assembly (as per Section 6.3.6 )
2. Unplug the connector of the photoelectric switch on the sample compartment assembly
and the cable connector of the electromagnet.
3. Mark the position of the sample compartment assembly on the autoloader board (make
sure the sample compartment assembly can be installed back to the right position).
4. Use the 107 cross-headed screwdriver to remove the 3 M4X8 small panhead screws, and
then move out the closed-tube sample compartment assembly in the direction as indicated
by the arrowhead in Figure 6-28.
Maintenance
6-40
Figure 6-28 Removing the sample compartment assembly
1 ―――Autoloader 2 ―――Sample compartment assembly
3 ―――M4X8 small panhead screw
To disassembly the closed-tube sample compartment assembly:
1. Use the 107 cross-headed screwdriver to remove the two M3X6 cross-headed panhead
screws, and then remove the electromagnet assembly; remove the two M3X6
cross-headed panhead screws in the electromagnet assembly, and then remove the
electromagnet body; remove the E ring to take out the spring, as shown in Figure 6-29
Maintenance
6-41
Figure 6-29 Removing the electromagnet assembly
1 ―――M3X6 cross-headed panhead screw 2 ―――Electromagnet assembly
3 ―――E ring 4 ―――Spring
5 ―――M3X6 cross-headed panhead screw 6 ―――Electromagnet body
2. Use the 107 cross-headed screwdriver to remove the M4X12 cross-headed panhead
screw fixing the photocouplers, and then remove the photocouplers, as shown in Figure
6-30
Figure 6-30 Removing the photocouplers
1 ―――PHOTOELEC switch 2 ―――M4X12 cross-headed panhead screw
Maintenance
6-42
3. Use the 1.5mm hexagonal wrench to remove the two M3X5 set screws, and then remove
the gear; use the 107 cross-headed screwdriver to remove the two M3X6 cross-headed
panhead screws fixing the damp gear, and then remove the damp gear, as shown in
Figure 6-31.
Figure 6-31 Removing the gear and the damp gear
1 ―――M3X5 set screw 2 ―――Gear
3 ―――Damp gear 4 ―――M3X6 cross-headed panhead screw
4. Use the sharp-nose pliers to remove the spring from either of the two fixing ends, as
shown in Figure 6-32.
5. Use the sharp-nose pliers or the small flat-headed screwdriver to remove either of the two
D2 E rings on the shaft, and loosen the two M3X5 set screws in the tube holder
compartment fixing the shaft. Then take out the shaft and remove the tube holder
compartment body and the helical torsion springs.
During installation, make sure the M3X5 set screws are fixed in the same level of the notches in the shaft, and one of the screws should be plunged into the cone-shaped hole.
Maintenance
6-43
Figure 6-32 Removing the helical torsion springs
1 ―――D2 E ring 2 ―――Shaft
3 ―――M3X5 set screw 4 ―――Tube holder compartment
5 ―――Pulling spring 6 ―――Helical torsion springs
6. Use the 1.5mm hexagonal wrench to remove the M3X5 set screw fixing the shaft, and then
you can take out the shaft, deep groove ball bearing, wheel, bush and gasket in turn, as
shown in Figure 6-33.
During installation, the two bearings should be put into the wheel (the bearings are supposed to be cling to the inner of the wheel) until their ends reaches those of the wheel; the set screws should be fixed in the level of the notches in the shaft.
Maintenance
6-44
Figure 6-33 Disassembling the shaft assembly
1 ―――Shaft 2 ―――3x6x2.5 deep groove ball bearing
3 ―――Wheel 4 ―――Bush
5 ―――Gasket 6 ―――M3X5 set screw
Installation
To replace the sample compartment assembly with a new one, repeat the first removal
procedure (step 1 to 4) in the reverse order. During installation, take into account the
following items:
1. One side of the sample compartment assembly should be set close to the positioning step
raised on the bottom board (autoloader bottom board)
2. Be sure to install the assembly back to the position marked in disassembly.
To replace components like the photocouplers, electromagnet, damp gear or springs,
repeat the second procedure above in the reverse order. During installation, take into
account the following items:
1. The tube holder compartment body should be put right in the center
2. Always install the damp gear after the gear is installed
3. Turn the core of the electromagnet 360° and make sure there is no interference
4. Move the guiding axis of the electromagnet with you hand and make sure the sample
compartment can turn properly. Push the compartment body back to the initial position and
make sure it can be locked up.
Maintenance
6-45
6.8.2 Removing and replacing the detection switches of the sample transport unit
Purpose
To remove or replace the detection switches (detecting the sample transport unit for proper
loading of tube rack and empty/full status of the unloader), follow the steps in this section.
Tools
101 cross-headed screwdriver
107 cross-headed screwdriver (107X75)
Replace parts according to requirements:
Sensitive switch
PHOTOCOUPLER (reflective)
Preparation
1. Power off the analyzer.
2. Turn off the main power switch on the right side of the analyzer and unplug the power cord.
3. Remove the right side door (as per Section 6.3.2 )
4. Remove the left side door (as per Section 6.3.1 )
5. Remove the top cover (as per Section 6.3.3 )
Removal
1. Remove the front cover assembly (as per Section 6.3.6 )
2. Unplug the cable connector corresponding to the switch to be replaced
3. Use the 101 cross-headed screwdriver to remove the two M2X8 cross-headed panhead
screws fixing the sensitive switch detecting the proper loading of tube racks, and then
remove the sensitive switch.
During installation, remember to check the orientation of the sensitive switch and make necessary adjustment. To make adjustment, move the tube rack manually to where the switch locates until it cannot be moved any more; the clicking sound made by the sensitive switch indicates proper installation (move the tube rack repeatedly to identify).
4. Use the 107 cross-headed screwdriver to remove the M2X8 small panhead screws fixing
the switches detecting the empty/full status of the unloader, and then remove the switch,
as shown in Figure 6-34.
Maintenance
6-46
Install the switch as level as possible.
Figure 6-34 Disassembling the switches of the sample transport assembly
1 ―――M3X8 small panhead screw 2 ―――PHOTOCOUPLER (reflective)
3 ―――M2X8 cross-headed panhead screw 4 ―――Sensitive switch
Installation
To replace the switch with a new one, repeat Step 1 to 4 above in the reverse order.
6.8.3 Replacing the sample transport unit
Purpose
As there may be errors with the sample transport unit (especially the switches and driving
units) during operation, you need to replace these components according to the steps in this
section.
Tools
101 cross-headed screwdriver
Maintenance
6-47
107 cross-headed screwdriver (107X75)
1.5mm hexagonal wrench
2.5mm hexagonal wrench
3mm hexagonal wrench
Preparation
1. Power off the analyzer.
2. Turn off the main power switch on the right side of the analyzer and unplug the power cord.
3. Remove the right side door (as per Section 6.3.2 )
4. Remove the left side door (as per Section 6.3.1 )
5. Remove the top cover (as per Section 6.3.3 )
Removal
Follow the two procedures to remove and disassemble the sample transport unit.
Remove the sample transport unit:
1. Remove the front cover assembly (as per Section 6.3.6 )
2. Remove the shielding box from the autoloader board, and unplug the cable connectors
between the sample transport assembly and the autoloader control board
3. Use the 107 cross-headed screwdriver to remove the three M4X12 cross-headed panhead
screws fixing the sample transport unit, and then remove the sample transport unit, as
shown in Figure 6-35.
Note: You should readjust the position of the sample transport unit during installation. See
Chapter 3 for how to make adjustment.
Maintenance
6-48
Figure 6-35 Disassembling the sample transport unit (1)
1 ―――Sample transport unit 2 ―――M4X12 panhead screw
Disassemble the sample transport assembly:
1. See Figure 6-36 for components of the sample transport compartment
Maintenance
6-49
Figure 6-36 Disassembling the sample transport unit (2)
1 ―――Loading unit (forth-and-back) 2 ―――Sample transport frame
3 ―――Closed-tube sample compartment assembly 4 ―――Riser unit
5 ―――Unloader unit 6 ―――Platform unit
7 ―――Loading unit (left-and-right) 8 ―――Backboard
2. Unplug the connector of the stationary barcode scanner; use the 107 screwdriver to
unscrew the four M3X8 small panhead screws fixing the stationary barcode scanner and
its bracket on the backboard unit, and then remove the stationary barcode scanner and its
bracket, as shown in Figure 6-36 and Figure 6-37
3. Use the 107 cross-headed screwdriver to remove other units on the backboard like the
tube detecting photocouplers and the tube rack locating plate as per Figure 6-37
During installation, make sure the back of the tube locating block and that of the tube rack locating plate are at the same plane; adjust the distance between the tube rack block and the tube rack to be around 0.2mm.
Maintenance
6-50
Figure 6-37 Disassembling the sample transport unit (3)
1 ―――Stationary barcode scanner 2 ―――Barcode scanner bracket
3 ―――Backboard bracket 4 ―――Sensor fixing board
5 ―――M3X8 small panhead screw 6 ―――Tube rack locating plate
7 ―――Rubber baffle plate 8 ―――Rubber baffle plate presser
9 ―――M2X4 cross-headed panhead screw 10―――PHOTOCOUPLER (reflective)
11―――Tube locating block
4. Use the 107 cross-headed screwdriver to remove the two M3X8 small panhead screws
fixing the unloader unit and remove the connector of the photoelectric sensor on the
unloader unit, and then remove the unloader unit, as shown in Figure 6-36 and Figure 6-37
5. Use the 107 cross-headed screwdriver to remove the unloader unit as per Figure 6-38
Maintenance
6-51
Figure 6-38 Disassembling the sample transport unit (4)
1 ―――M3X8 small panhead screw 2 ―――M4X10 cross-headed panhead screw
3 ―――GP1A05A photoelectric sensor 4 ―――Unloader pulling spring
5 ―――Sensory plate 6 ―――Top cover of the gear holder
7 ―――Gear shaft 8 ―――Short rack
9 ―――Gear holder 10―――Long rack
11―――Bottom cover of the gear holder 12―――M3X8 cross-headed countersunk screw
13―――Unloader bracket
6. Use the 107 cross-headed screwdriver to remove the two M3X8 small panhead screws
fixing the spring presser, and then remove the spring presser, as shown in Figure 6-36 and
Figure 6-39
Maintenance
6-52
Figure 6-39 Disassembling the sample transport unit (5)
1 ―――Riser unit 2 ―――Large plain washer GB96 3
3 ―――Spring presser 4 ―――M3X8 small panhead screw
5 ―――Shaft sleeve 6 ―――Sensory plate 3
7. Use the 107 cross-headed screwdriver to remove the two M3X8 small panhead screws
fixing the front supporting board, and then remove the front supporting board, as shown in
Figure 6-36 and Figure 6-40
8. Unplug the connectors of the sensitive switch and the PHOTOCOUPLER (reflective) on
the platform unit; use the 107 cross-headed screwdriver to remove the six M3X8 small
panhead screws fixing the platform unit, and then remove the platform unit, as shown in
Figure 6-36 and Figure 6-40
Maintenance
6-53
Figure 6-40 Disassembling the sample transport unit (6)
1 ―――M3X8 small panhead screw 2 ―――Sample transport platform
3 ―――M4X8 panhead screw 4 ―――Front supporting board
9. Unplug the connectors of the stepping motor and the three photoelectric sensor in the
left-and-right loading unit; use the 107 cross-headed screwdriver to remove the three
M3X8 small panhead screws fixing the left-and-right loading unit, and then remove the
left-and-right loading unit, as shown in Figure 6-36 and Figure 6-41
10. Use the 1.5mm hexagonal wrench, 2.5mm hexagonal wrench and the 107 cross-headed
screwdriver to remove the left-and-right loading unit as per Figure 6-41
Maintenance
6-54
Figure 6-41 Disassembling the sample transport unit (7)
1 ―――Stepping motor PK244M-01B 2 ―――M3X5 hexagon set screw
3 ―――Washer 4 ―――Driven belt shaft
5 ―――Synchronous pulley, P16MXL6.4 6 ―――E ring GB/T896-86.4
7 ―――Guiding rail supporting block 8 ―――Photoelectric sensor GP1A05A
9 ―――Guiding rail MGN9C1R95ZOCM 10―――M3X8 cheese head hexagon screw
11―――Belt presser 12―――Left-and-right loading slid plate
13―――M3X5 cheese head hexagon screw 14―――Shaft sleeve
15―――M3X12 cross-headed panhead
screw (with washer)
16―――Pallet
17―――Plain washer GB97.1 3 18―――M4X10 cross-head panhead screw
19―――Synchronous belt 20―――M3X8 small panhead screw
21―――Left-and-right loader bracket 22―――Synchronous pulley, P16MXL6.4
11. Unplug the connectors of the stepping motor and the two photoelectric sensor in the
forth-and-back loading unit; use the 107 cross-headed screwdriver to remove the four
M3X8 small panhead screws fixing the left-and-right loading unit, and then remove the
forth-and-back loading unit, as shown in Figure 6-36 and Figure 6-42
12. Use the 1.5mm hexagonal wrench, 2.5mm hexagonal wrench, 101 cross-headed
screwdriver and the 107 cross-headed screwdriver to remove the forth-and-back loading
unit as per Figure 6-42
Maintenance
6-55
Figure 6-42 Disassembling the sample transport unit (8)
1 ―――Driven belt shaft 2 ―――Washer
3 ―――Synchronous belt TBN160MXL025 4 ―――Synchronous pulley, P16MXL6.4
5 ―――Stepping motor PK244M-01B 6 ―――M2X8 cross-headed panhead screw
7 ―――Large plain washer GB96 3 8 ―――Right pusher
9 ―――M3X5 hexagon countersunk head
set screw
10―――Belt presser
11―――M3X5 cheese head hexagon screw 12―――Pulling spring
13―――M3X12 cross-headed panhead
screw (with washer)
14―――Shaft sleeve 2
15―――Left pusher 16―――Forth-and-back loading pusher board
17―――Photoelectric sensor GP1A05A 18―――M3X8 small panhead screw
19―――M4X10 cross-headed panhead
screw
20―――Guiding rail MGN9C1R155ZOCM
21―――M3X8 cheese head hexagon screw 22―――Forth-and-back loader bracket
23―――E ring GB/T896-86.4 24―――Synchronous belt TBN160MXL02
(two positioning plate)
Maintenance
6-56
Installation
Repeat the removal steps in the reverse order to install.
Take into account the following items during installation and adjustment:
Requirements for installing the forth-and-back loading unit:
Install the guiding rail close to one side, and put in the positioning screws before installation
Before installing the synchronous pulley of the driven belt shaft, apply some lubricator on the
surface of the shaft.
Make sure the shaft end of the stepping motor and the end of the synchronous pulley are at
the same plane.
While installing the stepping motor, adjust the tensility of the synchronous belt before locking
up screws.
Requirements for installing the left-and-right loading unit:
Install the guiding rail close to one side, and put in the positioning screws before installation
Before installing the synchronous pulley of the driven belt shaft, apply some lubricator on the
surface of the shaft.
Make sure the shaft end of the stepping motor and the end of the synchronous pulley are at
the same plane.
While installing the stepping motor, adjust the tensility of the synchronous belt before locking
up screws.
Requirements for installing the unloader unit:
Make sure the two racks can move freely with little resistance after installation. If the resistance
is large, repair is necessary; otherwise, the unit may not be able to restore the initial position.
See the initial positions of the short rack in Figure 6-43 with one tooth hanging over the end.
Figure 6-43 Initial position of the short rack
Requirements for installing the sample transport unit:
While installing the assembled sample transport unit to the main unit, readjust tbe sample
proble position as per Step 1 to 10 in Section 6.10.1
Maintenance
6-57
Verification
1. Readjust the position of the sample probe as per Step 10 to 14 in Section 6.10.1 .
2. Check if all parts are mounted to the former position and are fastened.
3. Check if the connection of signal lines of WBC or RBC is correct or not.
4. Power on and verify the installation.
Maintenance
6-58
6.9 Mix mechanism
Purpose
As there may be errors with the mix mechanism during operation, you need to replace the
mix mechanism according to the steps in this section.
Tools
107 cross-headed screwdriver (107X75)
1.5mm hexagonal wrench
2.5mm hexagonal wrench
3mm hexagonal wrench
Replace parts according to requirements:
BC-5380 mix mechanism
Preparation
1. Power off the analyzer.
2. Turn off the main power switch on the right side of the analyzer and unplug the power cord.
3. Remove the right side door (as per Section 6.3.2 )
4. Remove the left side door (as per Section 6.3.1 )
5. Remove the top cover (as per Section 6.3.3 )
Removal
Follow the two procedures to remove and disassemble the mix mechanism.
Remove the mix mechanism:
1. Remove the shielding box on the autoloader board, and unplug the cable connectors
between the mix mechanism and the autoloader control board
2. Remove the front cover assembly (as per Section 6.3.6 )
3. Use the 107 cross-headed screwdriver to remove the four M4X8 panhead screws, and
then remove the mix mechanism, as shown in Figure 6-44.
Maintenance
6-59
Figure 6-44 Removing the mix mechanism
1 ―――Mix mechanism 2 ―――M4X8 small panhead screw
Disassemble the mix mechanism:
1. Unfasten the plastic cable tie fixing the components and pull out the plugs from the five
photocouplers and three motors to the cable, and then remove the cable
2. Use the 107 cross-headed screwdriver to remove the three M3X6 cross-headed panhead
screws fixing the small motor protection cover, and then remove the cover, as shown in
Figure 6-45
3. Use the 2.5mm hexagonal wrench to remove the three M3X8 cheese head hexagon
screws connecting the rotation and X-co motion unit and the Z-co motion unit, and then
remove the two units, as shown in Figure 6-45
Maintenance
6-60
Figure 6-45 Disassembling the mix mechanism
1 ―――Rotation and X-co motion unit 2 ―――M3X8 cheese head hexagon screw
3 ―――Small motor protection cover 4―――M3X6 cross-headed panhead screw
5 ―――Z-co motion unit
4. Use the 107 cross-headed screw driver to remove the four M3X6 cross-headed panhead
screws fixing the protection cover, and then remove the protection cover, as shown in
Maintenance
6-61
Figure 6-46
5. Use the 107 cross-headed screw driver to remove the four M3X6 cross-headed panhead
screws fixing the two photocouplers on the 43 motor installation board, and then remove
the motor installation board and the elements on the board (draw out the motor shaft from
the motor nut), as shown in Figure 6-46
6. Use the 107 cross-headed screw driver to remove the three M3X6 cross-headed panhead
screws fixing the motor nut, and then remove the nut, as shown in Figure 6-46
7. Use the 2.5mm hexagonal wrench to remove the four M3X5 cheese head hexagon screws
fixing the connection block, and then remove the block, as shown in Figure 6-46
8. Use the 107 cross-headed screw driver to remove the two M3X6 cross-headed panhead
screws fixing the Z-co sensory plate, and then remove the two Z-co photocoupler sensory
plates, as shown in Figure 6-46
9. Use the 2.5mm hexagonal wrench to remove the five M3X5 cheese head hexagon screws
fixing the linear guiding rail MGN9C1R175Z0CM, and then remove the guiding rail, as
shown in Figure 6-46
10. Use the 2.5mm hexagonal wrench to remove the two M3X6 cross-headed panhead
screws fixed in the slider connecting board for baffling
Figure 6-46 Disassembling the mix mechanism
Maintenance
6-62
1 ―――Protection cover 2 ―――Photocouplers
3 ―――M3X6 cross-headed panhead screw 4―――Connecting block
5 ―――Motor nut 6 ―――Z-co sensory plate
7 ―――Linear guiding rail MGN9C1R175Z0CM 8 ―――Bracket
9 ―――M3X5 cheese head hexagon screw 10 ―――M3X8 small panhead screw
11 ―――43 motor installation board 12 ―――Sensor installation bracket
11. Use the 107 cross-headed screw driver to remove the two M3X6 cross-headed panhead
screws fixing the sensor installation bracket, as shown in Figure 6-47
12. Use the 2.5mm hexagonal wrench to remove the four M3X8 cheese head hexagon screws
fixing the linear stepping motor E43H4Q, and then remove the linear stepping motor
E43H4Q, four M3 large washer and 8 O rings, as shown in Figure 6-47
Figure 6-47 Disassembling the mix mechanism
1 ―――Linear stepping motor E43H4Q 2 ―――O ring
3 ―――43 motor installation board 4―――M3X6 cross-headed panhead
screw
5 ―――Sensor installation bracket 6 ―――M3 large washer
7 ―――M3X8 cheese head hexagon screw 8 ―――Sleeve
13. Use the 107 cross-headed screw driver to remove the six M3X6 cross-headed panhead
screws fixing the photocouplers, X-co sensory plate, X-co sensor installation board, and
then remove the photocouplers, X-co sensory plate, X-co sensor installation board, as
shown in Figure 6-48
14. Use the 107 cross-headed screw driver to remove the four M2.5X6 cross-headed panhead
screws fixing the linear stepping motor, and then remove the stepping motor E28H41, as
shown in Figure 6-48
15. Use the 2.5mm hexagonal wrench to remove the two M3X8 cheese head hexagon fixing
the 28 motor installation board, and then remove the installation board, as shown in Figure
6-48
Maintenance
6-63
16. Use the 2.5mm hexagonal wrench to remove the four M3X5 cheese head hexagon screw
fixing the connecting board, and then remove the connecting board
17. Use the 2.5mm hexagonal wrench to remove the four M3X5 cheese head hexagon screw
fixing the linear guiding rail MGN9C1R75Z0CM, and then remove the guiding rail
18. Use the 107 cross-headed screw driver to remove the two M3X6 cross-headed panhead
screws fixed in the slider connecting board for baffling
Figure 6-48 Disassembling the mix mechanism
1 ―――Linear motor E28H41 2 ―――M3X8 cheese head hexagon screw
3 ―――28 motor installation board 4―――M2.5X6 cross-headed panhead
screw
5 ―――Slider connecting board 6 ―――linear guiding rail
MGN9C1R75Z0CM
7 ―――Connecting board 8 ―――M3X5 cheese head hexagon screw
9 ―――M3X6 cross-headed panhead screw 10 ―――X-co sensory plate
11 ―――Photocouplers 12 ―――X-co sensor installation board
19. Use the 2.5mm hexagonal wrench to remove the M3X8 hexagon screw fixing the presser,
Maintenance
6-64
and then remove the presser, as shown in Figure 6-49
20. Use the 2.5mm hexagonal wrench to remove the M3X8 hexagon screw fixing the rotation
photocoupler plate, and then remove the rotation photocoupler plate, as shown in Figure
6-49
21. Use the 2.5mm hexagonal wrench to remove the M3X8 hexagon screw fixing the swing
arm and use the 2mm hexagonal wrench to remove the M4X8 set screw fixing the swing
arm, and then remove the swing arm
22. Use the 107 cross-headed screw driver to remove the two M3X6 cross-headed panhead
screws fixing the spring tube pincher, and them remove the M3 large washer and the
spring tube pincher, as shown in Figure 6-49
23. Use the 2.5mm hexagonal wrench to remove the four M3X10 cheese head hexagon
screws fixing the rotation stepping motor, and then remove the four M3 large washer, the
rotation stepping motor and the synchronous belt, as shown in Figure 6-49
24. Use the 1.5mm hexagonal wrench to remove the four set screws fixing the two small
pulleys, and then remove the pulleys, as shown in Figure 6-49
25. Use the 2.5mm hexagonal wrench to remove the two M3X8 screws fixing the bearing
holder, and then remove the bearing holder, pulley shaft and the two deep groove ball
bearings, as shown in Figure 6-49
26. Use the 107 cross-headed screw driver to remove the three M3X6 cross-headed panhead
screws fixing the linear stepping motor E28H41 nut, and then remove the linear stepping
motor E28H41 nut, as shown in Figure 6-49
27. Use the 107 cross-headed screw driver to remove the M3X6 cross-headed panhead screw
fixing the rotation sensor bracket, and then remove the rotation sensor bracket and the
photocouplers
Figure 6-49 Disassembling the mix mechanism
Maintenance
6-65
1 ―――Presser 2 ―――Spring tube pincher
3 ―――Rubber washer 4―――M3X8 cheese head hexagon screw
5 ―――Swing arm 6 ―――M4X8 set screw
7 ―――M3X5 set screw 8 ―――Small pulley
9 ―――Pulley shaft 10 ―――Rotation motor installation board
11 ―――Rotation sensor bracket and
photocouplers
12 ―――Bearing holder
13 ―――Deep groove ball bearing 14 ―――Rotation sensory plate
15 ―――Rotation stepping motor 16 ―――Linear stepping motor E28H41 nut
17 ―――M3X6 cross-headed panhead screw 18 ―――M3 large washer
19 ―――M3X10 cheese head hexagon screw 20 ―――Synchronous belt
Installation
Perform the steps above in the reverse order to install.
During installation or position adjustment, take into consideration the following items:
1. Install the guiding rail close to one side, and put in the positioning screws before
installation
2. Make sure the motion parts can move properly
3. Keep the synchronous belt fully tightened
4. Adjust the initial position of the pincher (for forth-and-back movement, left-and-right
movement, and rotation) while installing the assembled unit to the main unit. See Section
6.10.2 for how to make adjustment.
Maintenance
6-66
6.10 Adjustment
6.10.1 Adjusting Sample Probe Position 1. Switch off the power of the analyzer and open the sample compartment manually. Put the
sample probe in-compartment locating fixture in the sample compartment and close the
compartment door, as shown in Figure 6-50. Loosen the three screws on the sample
transport unit shown in Figure 6-51, Figure 6-52 and Figure 6-53. Move the sample probe
assembly and sample transport unit horizontally and enable the sample probe to aim at the
central hole in the locating fixture, and then fix the three screws.
Figure 6-50 Sample transport unit
Figure 6-51 Left screw Figure 6-52 middle screw Figure 6-53 Right screw
2. Connect the reagents properly and connect the main unit and the PC with a network cable.
Start up the PC software on the main unit and on the PC. Then, enter the user name
“service” and the password “Se s700” (note: there is a space between “e” and ”s”) into the
pop-up box, and then click “Ok” to initialize the analyzer. During initialization, a message
box will pop up asking you whether to “Ignore fluidic initialization?”. Click “Yes” to skip
the initialization.
3. Press the [OPEN] key on the front cover, and the sample compartment will be opened
automatically. Put the sample probe in-compartment locating fixture in the sample
compartment, as shown in Figure 6-54.
Maintenance
6-67
Figure 6-54 Placing the sample probe in-compartment locating fixture
4. Click “Menu” → “Service” → “Debug” to enter the “Debug” screen, as shown in Figure
6-55.
Figure 6-55 Clicking “Debug”
5. Click “CT Piercing Position” and “Start” in turn, and the sample probe will move to the
position above the sample compartment automatically, as shown in Figure 6-56.
Maintenance
6-68
Figure 6-56 Starting CT piercing position adjustment
6. Click “Forward”, “Backward”, “Up” and “Down” to adjust the position of the sample probe,
making the probe inserted into the slot, and then click “Up” to move out the probe, as
shown in Figure 6-57.
Click “Start”
Maintenance
6-69
Figure 6-57 Adjusting the CT piercing position using the software
7. Put a piece of paper on the locating fixture and click “Down” to make the sample probe
touch the paper. Then click “Stop” to save the position of the probe over the sample
compartment. Open the sample compartment manually to remove the sample
in-compartment locating fixture and close the compartment door, as shown in Figure 6-58.
Figure 6-58 Placing a piece of paper on the fixture
8. Put the sample probe autoloading locating fixture at the autoloading position of the sample
Click “Forward”, “Backward”, “Up” and “Down”
Click “Stop”
Maintenance
6-70
transport assembly, as shown in Figure 6-59
Figure 6-59 Placing the locating fixture at the piercing position
9. Click “AL piercing Position” and “Start” in turn, and the sample probe will move onto the
autoloading position, as shown in Figure 6-60
Figure 6-60 Starting AL piercing position adjustment
10. Click “Forward”, “Backward”, “Down” and “Up” to adjust the position of the sample probe,
Click “Start”
Maintenance
6-71
making the probe inserted into the slot of the locating fixture. Click “Up” and put a paper on
the AL positioning fixture, and then click “Down” until the sample probe reaches the paper,
as shown in Figure 6-61. Click “Stop” to save the autoloading position of the sample probe,
and then remove the sample probe AL positioning fixture, as shown in Figure 6-62.
Figure 6-61 Adjusting the sample probe AL piercing position using the software
Figure 6-62 Sample probe reaching the paper
11. Put the sample probe DIFF bath up position locating fixture in the aperture of the DIFF
Click “Forward”, “Backward”, “Down” and “Up”
Click “Stop”
Maintenance
6-72
bath, as shown in Figure 6-63.
Figure 6-63 Placing the DIFF bath up position locating fixture
12. Click “DIFF Bath Up Position”, and then click “Start” to move the sample probe onto the
DIFF bath, as shown in Figure 6-64.
Figure 6-64 Starting DIFF bath position adjustment
13. Click “Forward”, “Backward”, “Up” and “Down” to adjust the sample probe, making it
inserts in the position in front of the round slot. Click “Stop” to save the position of the
sample probe, and then remove the fixture, as shown in Figure 6-65.
Click “Start”
Maintenance
6-73
Figure 6-65 Adjusting the sample probe DIFF bath up position using the software
14. Click “Ok” to exit the “Debug” screen.
6.10.2 Adjusting Pincher Position 1. Switch off the power of the analyzer and put the tube rack at the piercing position of the
sample transport unit, making the recess of the tube rack lap over the presser, as shown in
Figure 6-66, Figure 6-67 and Figure 6-68.
Figure 6-66 The recess of the tube rack
Click “Forward”, “Backward”, “Up” and “Down”Click “Stop”
Click “Ok”
Recess
Maintenance
6-74
Figure 6-67 The presser of the sample transport unit
Figure 6-68 Placing the tube rack
2. Move the pincher to the tube grabbing position and push upwards. Then put the tube
locating fixture into the pincher and make the pincher upright. Move the pincher
downwards slowly, and check if the tube locating fixture can be inserted to the tube rack
properly. If not, remove the four set screws and the left-and-right mixing assembly to insert
the tube locating fixture to the tube rack. Secure the screws after adjustment, as shown in
Figure 6-69.
Figure 6-69 Adjusting the manipulator position manually
3. Connect the reagents properly and connect the main unit and the PC with a network cable.
Start up the main unit and the PC software. Then, enter the user name “service” and the
Presser
The presser laps over the recess of the tube rack
Four screws
Tube locating fixture
Maintenance
6-75
password “Se s700” (note: there is a space between “e” and “s”) into the pop-up box, and
then click “Ok” to initialize the analyzer. During initialization, a message box will pop up
asking you whether to skip the initialization. Click “Yes” to skip the initialization
4. Click “Menu” → “Service” → “Debug” to enter the “Debug” screen
5. Click “Manipulator” and “Start” in turn, and the pincher will move onto the tube grabbing
position, as shown in Figure 6-70.
Figure 6-70 Starting manipulator position adjustment
6. Put the tube rack at the piercing position of the sample transport unit, and then put the
tube locating fixture into the pincher, as shown in Figure 6-71
Figure 6-71 Placing the pincher locating fixture
7. Click “Left” and “Right” to adjust the pincher to be upright, and then click “Forward”,
Click “Start”
Maintenance
6-76
“Backward”, “Up” and “Down” to make the tube locating fixture able to insert in the hole of
the AL piercing position locating fixture. Click “Up” and remove the two fixtures, as shown
in Figure 6-72 and Figure 6-73.
Figure 6-72 Adjusting the pincher position using the software
Figure 6-73 The hole in the piercing position locating fixture
8. Click “Stop” to save the position of the pincher, as shown in Figure 6-72.
9. Click “Ok” to exit the “Debug” screen, as shown in Figure 6-72.
Click “Left”, “Right” “Forward”, “Backward”, “Up” and “Down”
Click “Stop”
Click “Ok”
7-1
7 Optical System
7.1 Optical System Adjustment and Troubleshooting
Purpose
To replace and repair the optical system component in case of malfunction.
Tools
Antistatic gloves
Hexagon wrench
Cross-headed screwdriver
Flat-headed screwdriver
Horologic driver
Duke (4K-07) 7µm standard particles
Oscilloscope or oscillometer
Removal and installation
Before removing the optical system component, remove the right side door and the top cover
of the analyzer first (refer to Chapter 6 Maintenance), then remove all the screws on them, as
Figure 7-1 shows. Remove the protective cover of the optical system. After component
installation, reinstall the protective cover and secure it. The above-mentioned steps will not be
repeated in the following procedures.
Optical System
7-2
Figure 7-1 Remove the protective cover of the optical system
Screws
Optical System
7-3
7.2 Removing and Installing Optical System Assemblies The optical system consists of the following assemblies: the front light assembly, flow cell
assembly, rear light collimator assembly, beam splitter assembly, rear light collector assembly,
and rear light detector assembly.
Wear antistatic gloves when removing boards.
Be sure to turn off the analyzer and disconnect the power cord before removing boards.
Do not plug/unplug the laser generator unless the power cord is disconnected. Otherwise, the laser generator will be damaged.
7.2.1 Laser driver board
The laser driver board supplies power to the laser generator and controls the output power of
the laser.
1. Turn off the analyzer, open the laser cover and remove the connecting line from the driver
board.
2. Remove the fixing screws, and remove the laser driver board support.
3. Replace the laser driver board and reinstall it in the reverse order, then secure the fixing
screws.
Optical System
7-4
Figure 7-2 Laser driver board installation
1——fixing screws of laser driver board
support
2——fixing screws of laser driver board
7.2.2 Front light assembly The front light assembly generates the laser beam that enters the flow cell. Remove and install
the assembly strictly as instructed below. See Figure 7-3 for the assembly drawing.
1. Turn off the analyzer and disconnect its power cord.
2. Disconnect the laser power connector from the laser control board. Disconnect the heater,
temperature sensor and over temperature protection switch connectors.
3. Use the hexagon wrench to remove the fixing screws at the two sides of the front light
assembly and remove the assembly.
4. Use the hexagon wrench and the cross-headed screwdriver to remove the screws fixing
the laser generator assembly and the cylindrical mirror assembly separately, and then
remove the assemblies from the front light guide assembly.
Optical System
7-5
Figure 7-3 Front light assembly
1——Laser assembly 3——Cylindrical mirror assembly
2——Front guide assembly
5. According to the malfunction of the front light assembly, replace the whole laser assembly
or the whole cylindrical mirror assembly, install the new one on the front light assembly and
secure the assembly to the optical support.
6. Reconnect the laser power connector, heater, temperature sensor and the connector of
the over temperature protection switch.
7.2.3 Flow cell assembly
The flow cell assembly is used to create a stable sheath fluid flow, which enables the particle
tested to pass through the flow cell in a certain speed and to interact with the incident laser to
create scattered light. Remove and install the flow cell assembly strictly as instructed below.
1. Power on the analyzer to empty the fluidic pipes, then power off the analyzer and
disconnect the power cord.
2. Disconnect the sheath charging tube, sample charging tube and waste outlet tube (at the
outlet of the flow cell). Use the hexagon wrench to remove the two fixing screws and take
out the flow cell assembly carefully, as Figure 7-4 shows.
Optical System
7-6
Figure 7-4 Flow cell fluidic connecting tube
1―――Sample charging tube 2―――Sheath charing tube
3. Use the hexagon wrench to remove the two connecting screws and divide the flow cell
assembly into two parts: rectifier assembly and detecting assembly, as Figure 7-5 shows.
4. Clean or replace the detecting assembly and liquid transporting assembly as needed.
Figure 7-5 Flow cell structure
1——Rectifier assembly 3——Detecting assembly
2——O ring 4——Hexagon screw
5. Install the flow cell assembly as stated above in the reverse order.
Optical System
7-7
7.2.4 Rear light collimator assembly The rear light collimator assembly collects the scattered light created by the particle tested and
allows the light to be tested by the rear light detector assembly. Remove the rear light
collimator assembly strictly as instructed below. See Figure 7-6 for the assembly drawing of
the rear light collimator assembly.
1. Power off the analyzer and disconnect its power cord.
2. Use the hexagon wrench to remove the fixing screws at two sides of the rear light
assembly and remove the assembly.
3. Use the hexagon wrench to remove the two screws fixing the diaphragm lens assembly to
remove the lens assembly.
Figure 7-6 Rear light collimator assembly
1―――Diaphragm lens assembly 2―――Rear light guide assembly
4. Install the new rear light diaphragm lens assembly on the rear light guide assembly to form
a new rear light collimator assembly. Ensure that the superface edge of the diaphragm
lens assembly parallels the end face edge of backward oriented assembly. Install and
secure the assembly to the optical support.
7.2.5 Beam splitter assembly The beam splitter assembly splits the scattered light into two paths: one for the forward low
angle PD assembly and the other for the forward high angle PD assembly. See Figure 7-7 for
the assembly drawing of the beam splitter assembly.
1. Power off the analyzer and disconnect its power cord.
2. Use the hexagon wrench to remove the fixing screws at two sides of the beam splitter
assembly and remove the assembly.
3. Use the flat-headed screwdriver to remove the two clamping rings. Use the cross-headed
Optical System
7-8
screwdriver to remove the screws at the top board and the side boards to remove the
boards.
4. Remove the two protective washers to remove the beam-splitting prism.
5. When replacing or installing a new beam-splitting prism, pay attention to the direction of
the prism (see Figure 7-7 ). Install the prism following steps above in the reverse order.
Figure 7-7 Beam splitter assembly
1——Base of spectroscope 4——Clamping ring
2——Side board 5——Washer
3——Top board 6——Beam-splitting prism
7.2.6 Rear light collector assembly The rear light collector assembly enhances the collection efficiency by focusing the scattered
light, which has passed the beam splitter assembly, on the PD detecting target. It consists of a
LAS assembly and a MAS assembly.
LAS assembly
1. Power off the analyzer and disconnect its power cord.
2. Use the hexagon wrench to remove the two fixing screws on the LAS assembly and
remove the LAS assembly from the optical support.
3. Use the cross-headed screwdriver to remove the two screws fixing the low angle
diaphragm and remove the diaphragm. Use the cross-headed screwdriver to remove the
two screws fixing the rear light focus lens assembly and remove the assembly.
Optical System
7-9
Figure 7-8 LAS assembly
1——Low angle diaphragm 2——Rear light focus lens assembly
4. Install and secure the low angle diaphragm and the rear light focus lens assembly as
needed.
5. Install the LAS assembly to the optical support and screw it with the hexagon wrench.
During securing, ensure that the diaphragm assembly is close to the M4 hexagon screws.
MAS assembly
1. Power off the analyzer and disconnect its power cord.
2. Use the hexagon wrench to remove the two screws fixing the MAS assembly and then
remove the MAS assembly from the optical support.
3. Use the cross-headed screwdriver to remove the two screws fixing the diaphragm and
then remove the diaphragm. Use the cross-headed screwdriver to remove the two screws
on the rear light focus lens assembly and then remove the assembly.
Figure 7-9 MAS assembly
1——High angle diaphragm 2——Rear light focus lens assembly
Optical System
7-10
4. Install and secure the diaphragm and the rear light focus lens assembly as needed.
5. Install the MAS assembly to the optical support and screw it with the hexagon wrench.
During securing, ensure that the diaphragm assembly is close to the M4 hexagon screws.
7.2.7 Rear light detector assembly The rear light detector assembly detects the scattered light signal and converts it into electrical
signals for further processing. It consists of two parts: a forward low angle PD assembly and a
forward middle low angle PD assembly.
Forward low angle PD assembly
1. Power off the analyzer and disconnect its power cord.
2. Use the hexagon wrench to remove the screws fixing the forward low angle PD assembly
and then remove the assembly from the optical support.
3. Use the cross-headed screwdriver to remove the two screws fixing the PD board and
remove the shielding box from the forward low angle PD assembly.
Figure 7-10 Forward low angle PD assembly
1——Shielding box 3——PD
2——Low angle pre-amplification board 4——PD cover
4. Use the cross-headed screwdriver to loosen the low angle pre-amplification board, remove
the PD board, replace the corresponding parts and reinstall them orderly.
5. Assemble the shielding box to the forward low angle PD assembly and then install the
forward low angle PD assembly to the optical support.
Forward middle low angle PD assembly
1. Power off the analyzer and disconnect its power cord.
2. Use the hexagon wrench to remove the screws fixing the forward middle low angle PD
Optical System
7-11
assembly and remove the assembly from the optical support.
3. Use the cross-headed screwdriver to remove the two screws fixing the PD board and then
remove the shielding box from the forward middle low angle PD assembly.
Figure 7-11 Forward middle low angle PD assembly
1——Shielding box 3——PD
2——High angle pre-amplification board 4——PD cover
4. Use the cross-headed screwdriver to loosen the low angle pre-amplification board, remove
the PD cover and the PD board, replace the corresponding parts and install them orderly.
5. Assemble the shielding box to the forward middle low angle PD assembly and then install
the forward middle low angle PD assembly to the optical support.
Pre-amplification board
1. Power off the analyzer and disconnect its power cord.
2. Use the cross-headed screwdriver to loosen the two screws securing the PD board, and
remove the shielding box from the PD assembly.
3. Use the cross-headed screwdriver to remove the pre-amplification board from the PD
assembly and replace it with a new pre-amplification board, then reinstall it orderly.
Note: The rear light PD assembly needs not to be removed from the optical system during the
replacement of the pre-amplification board assembly.
Optical System
7-12
7.3 Adjustment Any assembly in the optical system should be adjusted after the installation once it has been
disassembled. The adjustment procedure can be simplified according to how the assembly
has been disassembled.
The adjustment to the optical system consists of coarse adjustment and fine tuning. Coarse
adjustment roughly aligns every optical assembly and makes them approximately on the same
optical axis, so that the signal can be created rapidly when standard particles are run. Fine
tuning ensures the best position of the optical system. An oscilloscope is used to observe the
signal waveform and the corresponding optical assembly position is finely tuned so that the
signal generated meets the adjustment standard.
7.3.1 Coarse adjustment According to each optical assembly, the coarse adjustment can be divided into: front light
assembly adjustment, flow cell adjustment, rear light collimator assembly adjustment, rear light
collector assembly diaphragm adjustment and backward detecting PD assembly adjustment.
Front light assembly adjustment
Purpose:
Horizontally, the beam is upright to the rear light collimator lens. The collimator emergent far
field spot is in the vertical direction.
Principle:
The laser beam distributes symmetrically along the optical axis; the reflex fully covers the
incident light when the light enters vertically.
Procedure:
Install the rear light assembly first, and cling the rear light assembly closely to the pin stop at
the rear light position, as Figure 7-12 shows. Aim the pin stop of the front light assembly at the
optical support, secure the fixing screw (make sure the forward optical system attaches the
optical support and is still movable). Turn on the power of the laser, and turn the laser
assembly to make the far field spot of the emitting laser vertical. Attach the rear light collimator
assembly to the pin stop, use a hexagon wrench to check whether the light reflected from the
surface of the rear light collimator lens is on the same axis with the incident light and slowly
turn the front light assembly to make them on the same axis, and then secure the assembly, as
Figure 7-13 shows.
Optical System
7-13
Figure 7-12 Rear light assembly installation
Figure 7-13 Front light assembly adjustment
Optical System
7-14
Flow cell adjustment
Purpose:
Adjusting the flow cell to irradiate the incident light to the surface of the flow cell vertically.
Principle:
That the flow cell is vertical to the incident laser uses the principle of that the reflex fully covers
the incident light when the light enters vertically.
Procedure:
Turn the flow cell into the optical path (do not position the square aperture detection area in the
optical path). Use a hexagon wrench to check whether the light reflected from the flow cell
surface is on the same axis with that of the incident light and adjust the flow cell slowly to
harmonize them and then secure the flow cell.
Figure 7-14 Flow cell coarse adjustment
Rear light collimator assembly adjustment
Rear light horizontal position adjustment Purpose:
To make sure the front and the rear light is on the same axis approximately.
Principle:
The beam that goes through the main point of the lens will not change the transmitting
direction.
Procedure:
1. Remove the rear light collimator assembly; check whether the beam spot that passes
through the beam-splitting prism and the diaphragm on the MAS assembly is symmetrical.
Adjust the MAS assembly to make the beam spot symmetrical, as shown in Figure 7-15.
Optical System
7-15
Figure 7-15 Adjustment of the horizontal position of the rear light collimator assembly
2. Move horizontally the rear light collimator assembly along the pin stop and observe the
beam spot position on the diaphragm of the MAS assembly. When the spot is in the center
of the diaphragm (as shown in Figure 7-16), secure the fixing screw of the rear light
collimator assembly.
Figure 7-16 The adjustment of the horizontal position of the rear light collimator assembly is finished
The rear light assembly must cling to the pin stop when it is moved.
Rear light collection assembly adjustment
LAS assembly diaphragm adjustment
Optical System
7-16
Purpose:
To obtain the effect of selecting low angle scattered light by positioning the LAS assembly
diaphragm symmetrically along the light axis.
Procedure:
1. Place a piece of white paper with rough surface at the back of the LAS assembly
diaphragm, and adjust the horizontal adjusting screw cap until the spot is symmetrical
horizontally (as shown in Figure 7-17).
2. Adjust the vertical adjusting screw cap until the spot is symmetrical vertically (as shown in
Figure 7-17).
Figure 7-17 Adjustment of the LAS assembly diaphragm
MAS assembly diaphragm adjustment Purpose:
To obtain the effect of selecting middle low angle scattered light by positioning the MAS
assembly diaphragm symmetrically along the light axis.
Note: After the adjustment of the horizontal position of the rear light collimator assembly is
finished, the position adjustment of the MAS assembly diaphragm is finished. You can secure
the fixing screw.
Rear light detector assembly adjustment
Forward low angle PD assembly adjustment Purpose:
To ensure the scattered light from the forward low angle is completely collected by the PD.
Procedure:
Optical System
7-17
Adjust the horizontal and vertical positions of the forward low angle PD assembly until the
center of the spot emitted from the LAS assembly diaphragm is completely received by the
aperture (as shown in Figure 7-18).
Figure 7-18 Adjustment of the forward low angle PD assembly
1―――Aperture diaphragm of eliminating
veiling glare
2―――Low angle beam spot
Forward middle low angle PD assembly adjustment Adjust the forward middle low angle PD assembly in the same way as adjusting the forward
low angle PD assembly. Ensure that the center of the emitted spot is completely received by
the aperture (as shown in Figure 7-19).
Figure 7-19 Spot from the forward middle low angle PD assembly
1―――Aperture diaphragm of eliminating
veiling glare
The above description is a complete procedure of coarse adjustment. By the adjustment above,
each assembly is roughly positioned symmetrically along the light axis. The following
describes the fine tuning of the optical system.
Optical System
7-18
7.3.2 Preparation before fine tuning
Adjustment of the horizontal position of the flow cell
Before fine tuning, the detecting square aperture of the flow cell should be turned into the
optical path. Check the MAS assembly diaphragm. When the sideline shades of the square
aperture distribute symmetrically on the diaphragm, the flow cell is turned into the optical path,
as shown in Figure 7-20.
Figure 7-20 Flow cell position adjustment
1——Flow cell horizontal adjusting knob
Adjustment of diaphragm against direct light
After the flow cell horizontal position adjustment is finished, turn the diaphragm against direct
light into the center of the optical path, so that the light intensity shielded by the diaphragm is
the largest and brightest. The following two methods show how to judge whether the
diaphragm is right in this position.
Visual check method:
Place a card (with a drawing line on it) on the MAS assembly diaphragm so that the two
sideline shades of the flow cell distribute symmetrically along the drawing line, as shown in
Figure 7-21. Adjust the position of the diaphragm against direct light. When the shielded spot
distributes symmetrically along the drawing line, the adjustment is finished.
Optical System
7-19
Figure 7-21 Visual check method
1―――Beam spot 3―――Drawing line
2―――Inner aperture of the flow cell 4―――High angle diaphragm
DC background method:
Output the pre-amplification board TPOUT test point of the forward low angle PD assembly (or
the TP43 LASIN test point connected to the signal processing board) to the oscilloscope and
select DC gear. Adjust the position of the diaphragm against direct light until minimum LAS DC
background is obtained, and then the adjustment is finished.
After the adjustment, secure the two screws of the diaphragm against direct light (as shown in
Figure 7-22).
Figure 7-22 Diaphragm against direct light
1——Lock screw of the diaphragm
against direct light
7.3.3 Fine tuning The purpose of fine tuning is to ensure the optical system is in the optimal position when it
detects the scattered light signal from the designed angle. The fine tuning purpose mainly
consists of the following three parts: (1) to focus the laser on the cell flow; (2) To ensure that
the center of the sample flow basically covers the spot center; and (3) To adjust the scattered
Optical System
7-20
light of the rear light assembly from the designed angle. Therefore, the tuning is mainly on the
forward optical system axial position, the flow cell horizontal position and the position of the
rear light collimator assembly.
Before the particle adjustment, at the “Main”→ “Menu”→ “Setup”→ “Gain” screen, set the
values of SS and FS to be 100.
Adjustment of front light assembly axial position
Purpose:
Ensure that the laser beam focus is on the cell flow.
Procedure:
1. Test the voltage of the tp2vref test point of the laser control board to see whether the
voltage reaches 4V. If not, adjust the variable resistance of the laser control board until the
voltage is 4V.
2. Output the pre-amplification board TPOUT test point of the forward low angle PD
assembly to the oscilloscope channel 1 (CH1), set the oscilloscope to be: voltage 50mv/div,
time 500ns/div and AC coupling.
3. Adopt standard particles as sample for test. Add 3 drops of Duke (4K-07) 7µm standard
particles into a 1.5mL centrifugal tube, and then add diluent or deionized water into the
centrifugal tube to 1mL. Use the diluted standard particle solution within 3 hours.
4. Run the standard particles in the open-vial mode normally, check the signal in the
oscilloscope channel 1 and adjust the adjusting knob of the front light assembly (as shown
in Figure 7-23) in the axial position. When the maximum signal amplitude and stability is
obtained and distributes normally, the pulse width is the narrowest and about 1.3us, as
shown in Figure 7-24, secure the fixing screws of the front light guide assembly.
Optical System
7-21
Figure 7-23 Fine tuning of the front light axial position
1――Front light axial adjustment knob 2——laser axial adjustment lock screws
Figure 7-24 Standard particle signal graph output by forward optical system
Note: The sampling time of the DIFF channel is about 20s. To simplify the adjustment and
generate signals quickly, you can adjust the forward optical axis adjusting knob in advance
until the distance between the front end surface of the front light guide assembly and that of
the guide assembly support is about 4mm.
Fine tuning of the rear light collimator assembly
Purpose:
To ensure that what the PD assembly detects is the scattered light signals of the designed
collection angles.
Optical System
7-22
Procedure:
1. Output the pre-amplification board TPOUT test point of the forward middle low angle PD
assembly (or the TP44 MASIN test point connected to the signal processing board) to the
oscilloscope channel 2 (CH2), set the oscilloscope to: voltage 100mv/div, time 500ns/div
and AC coupling.
2. Run standard particles in the open vial mode normally, adjust the adjusting knob of the
rear light collimator assembly and check the signal pulses of the oscilloscope channel 2.
Secure the set screws of the rear light collimator assembly when the pulse peaks reach
the maximum value, the signal are stable and the shake is little.
3. Switch to the channel 1 (CH1) and check whether the signal peaks are 190mv±10mv. If not,
adjust the adjustable resistance on the laser control board until the signal peaks of the
CH1 are 190mv±10mv.
4. Check whether the signal peaks of the channel 2 (CH2) are about 800mv. If not, adjust the
adjusting knob of the rear light collimator assembly and confirm whether the pulse signal
peaks of the CH2 are at the maximum.
Fine tuning of the flow cell horizontal position
Purpose:
To ensure that the center of the sample flow is positioned in the center of the beam spot.
Procedure:
1. Run standard particles in the open vial mode, and check whether the signal peaks of the
oscilloscope channel 1 (CH1) are stable and shake little.
2. Adjust the flow cell horizontal position until the particle signal peaks displayed on the
oscilloscope are at the maximum and uniform.
Optical System
7-23
Figure 7-25 Adjustment of the flow cell horizontal position
1―――Horizontal adjusting knob of the
flow cell
3. Step 3: After the test, standard particles scattergrams will be obtained, as shown in Figure
7-26:
Figure 7-26 Standard particles scattergrams
Click the “Optical” in the “Review” menu to enter the “Optical” screen (the default screen is
the DIFF channel calculation screen), then click “Calculate” to obtain the particle scattergram
parameters of the DIFF channel and then check whether the parameters meet the following
standards:
MAS 0.1max width < 15.21
LAS 0.1max width < 6.5
4. Step 4: When all the standard particle parameters meet the standards above, secure the
flow cell lock screws (as shown in Figure 7-27) and secure the protective cover of the
optical system.
Optical System
7-24
Figure 7-27 the secured flow cell
1―――Lock screws of the flow cell
Now, the adjustment of the optical system is finished.
7.3.4 Gain setup After the optical system adjustment, you need to reset the DIFF channel gains. The gain setup
consists of two steps:
Calibration using calibrators
At the count screen, ensure that the WBC background is under 0.1, and then run the calibrator
in the CT-WB-CBC+DIFF mode normally. (If the WBC background exceeds the limit, please
clean the flow cell or do a background check).
After the test, Click the “Optical” in the “Review” menu to enter the “Optical” screen (the
default screen is the DIFF channel calculation screen), and then click “Calculate” to obtain the
calibrator scattergram parameters of the DIFF channel.
Optical System
7-25
Figure 7-28 Optical system output calibrator signal
Enter into the Peak target boxes:
LAS low angle gravity center target
MAS high angle gravity center target
Calculate the gains, enter the gains into the corresponding boxes of the
“Main”→”Setup”→”Gain” screen and save the results according to the prompt on the screen.
After the gain setup, retest the calibrator and 7µm standard particles to verify the gain and
check if the following requirements are met. For the calibrator: the low angel gravity center of
the calibrator should be within ±0.57 from the target; the high angel gravity center within ±0.63
from the target. For the 7um standard particles: the low angel gravity center of the calibrator
should be 18.05±1.37 (16.68, 19.42); high angel gravity center 104.24±3.68 (100.56, 107.92);
low angel width ≤6.50; high angel width ≤15.21. Otherwise, continue calibrating until the
following requirements are met. For the calibrator: the low angel gravity center of the calibrator
should be within ±1.51 from the target; the high angel gravity center within ±2.4 from the target.
For the 7um standard particles: the low angel gravity center of the calibrator should be
18.05±1.37 (16.68, 19.42); high angel gravity center 104.24±3.68 (100.56, 107.92); low angel
width ≤6.50; high angel width ≤15.21.
Sample verification
After the above standard particle gain setup is finished, get several normal fresh blood
samples for test. The scattergram of a normal fresh blood sample should be like the images
shown in Figure 7-29:
Optical System
7-26
Figure 7-29 Scattergram of a normal fresh blood sample
If the WBC scattergram results of the normal fresh blood samples are significantly different
from the figure above, please check whether the reagents, reaction system and the fluidic
system work normally.
Optical System
7-27
7.4 Troubleshooting
7.4.1 Laser spot-deviation The SS and FS gain setup is not changed during the use of the analyzer. Yet the scattergram
is compressed to the left down side, namely, the two signals collecting angle direction turn
smaller, as shown in Figure 7-30. This is mainly because the focusing spot at the flow cell
position move forward or backward, causing the diminishment of laser intensity.
Figure 7-30 Compressed scattergram
Settlement:
Make standard particle fist, and record the peak value position of standard particle. Loose the
laser axial adjustment lock screws (as shown in Figure 7-20). Turn the adjustment screw of
front light assembly counterclockwise for a quarter turn. Make standard particle and see the
peak value position. If the peak value position becomes greater, keep turning the adjustment
screw counterclockwise with 1/8 circle each time till the peak value position becomes smaller,
then turn back with 1/8 circle. If the peak value position becomes smaller, turn the adjustment
screw in the reverse order with the same method. After finding the maximum position of
standard particle, lock the laser axial adjustment fixing screws.
7.4.2 Flow cell clog This is because a big blood clot or impurity blocks the flow cell or sampler inlet, which makes
the sheath flow unable to form. The scattergram is accordingly abnormal and scattered, as
Figure 7-31 shows.
Optical System
7-28
Figure 7-31 Scattered scattergram
Settlement:
Soak with dilute probe cleanser. If it does not work, replace the flow cell assembly.
7.4.3 Dirty flow cell It can be classified into: dirty inner wall and dirty outer surface. Dirty outer surface might due to
the pollution from waste, and dirty inner wall is caused by the pollution of blood clot on the
inner wall of flow cell.
Settlement:
Dirty outer surface: wipe the outer surface gently with dustless cloth soaked in alcohol, and in
the same direction
Dirty inner wall: refer to the cleaning method for the inner wall of liquid flow cell.
8-1
8 Troubleshooting
8.1 Error code and information
Table 8-1 Error code and information
Error Code Error Name Error Code Error Name
0x01000100 Valve output failed 0x01003011 High pressure
0x01000101 Valve No. out of range 0x01003020 WBC volumetric tube filter
clog
0x01000201 Sampling assembly horizontal
motor failed to move to target
position
0x01003021 RBC volumetric tube filter
clog
0x01000202 Sampling assembly horizontal
motor failed to return to home
position after initialization
0x01003030 DIFF reaction bath temp.
error
0x01000203 Sampling assembly horizontal
motor failed to leave home
position after initialization
0x01003031 Temperature out of working
range
0x01000204 Adjusting sampling assembly
horizontal motor to target
position failed
0x01003032 Temperature out of operating
range
0x01000205 Adjusting sampling assembly
horizontal motor to home
position failed
0x01003033 Optical system temp. error
0x01000206 Adjusting sampling assembly
horizontal motor to leave
home position failed
0x01003050 FS blank voltage abnormal
0x01000207 Sampling assembly horizontal
motor adjustment error
0x01003060 HGB blank voltage abnormal
0x01000208 Sampling assembly horizontal
motor adjusting steps out of
limit
0x01003061 Laser diode current abnormal
0x01000209 Sampling assembly horizontal
motor adjusting end position
error
0x01003062 HGB channel error
0x01000211 Sampling assembly vertical 0x01003070 Diluent expired
Troubleshooting
8-2
motor failed to move to target
position
0x01000212 Sampling assembly vertical
motor failed to initialize to
upper position
0x01003071 HGB Lyse expired
0x01000213 Sampling assembly vertical
motor failed to initialize to
lower position
0x01003072 LEO(I) Lyse expired
0x01000214 Adjusting sampling assembly
vertical motor to target
position failed
0x01003073 LEO(II) Lyse expired
0x01000215 Adjusting sampling assembly
vertical motor to upper
position failed
0x01003074 Cleanser expired
0x01000216 Adjusting sampling assembly
vertical motor to leave upper
position failed
0x01003080 No Diluent
0x01000217 Sampling assembly vertical
motor adjustment error
0x01003081 No LH lyse
0x01000218 Sampling assembly vertical
motor adjusting steps out of
limit
0x01003082 No LEO(I) lyse
0x01000219 Sampling assembly vertical
motor adjusting end position
error
0x01003083 No LEO(II) lyse
0x01000220 Sampling assembly is working 0x01003084 No Cleanser
0x01000221 Cannot adjust sampling
assembly
0x01003085 Waste is full
0x01000222 Sampling assembly horizontal
motor action overtime
0x01003090 Emptying WBC bath failed
0x01000223 Sampling assembly horizontal
motor action forbidden
0x01003091 Emptying RBC bath failed
0x01000224 Sampling assembly horizontal
motor does not match current
position
0x010030A0 Constant current source 55V
voltage abnormal
0x01000225 Sampling assembly vertical
motor action overtime
0x010030A1 +12V analogue voltage
abnormal
0x01000226 Sampling assembly vertical
motor action forbidden
0x010030A2 -12V analogue voltage
abnormal
Troubleshooting
8-3
0x01000227 Sampling assembly vertical
motor mismatch current
position
0x010030A3 Drive board 24V voltage
abnormal
0x01000228 Sampling assembly vertical
motor failed to pierce to lower
position
0x010030A4 Drive board 12V voltage
abnormal
0x01000229 Sampling assembly vertical
motor failed to pierce to upper
position
0x010030A5 Drive board 5V voltage
abnormal
0x01000230 Sampling assembly horizontal
motor photocoupler error
0x010030B0 Data board DDR abnormal
0x01000231 Sampling assembly RBC bath
photocoupler error
0x010030B1 Data board FLEXBUS
abnormal
0x01000232 Sampling assembly WBC
bath photocoupler error
0x010030B2 Data board FLASH abnormal
0x01000233 Sampling assembly DIFF bath
photocoupler error.
0x010030B3 Data board AD7908
abnormal
0x01000234 Sample assembly vertical
motor photocoupler error
0x010030B4 Data board AD7928
abnormal
0x01000235 Sampling assembly upper
photocoupler error
0x010030C0 Drive board communication
error
0x01000236 Sampling assembly lower
photocoupler error
0x010030D0 Side door open
0x01000300 Sampling syringe busy 0x010030D1 Optical assembly cover open
0x01000301 Sampling syringe
photocoupler abnormal
0x010030D2 Sample compartment door
open
0x01000302 Sampling syringe failed to
return to home position after
action
0x01004200 Software protocol resolution
0x01000303 Sampling syringe failed to
leave home position after
action
0x01004201 Response overtime
0x01000304 Sampling syringe failed to
return to home position before
action
0x01004202 Heartbeat disconnected
0x01000305 Sampling syringe failed to
leave home position before
action
0x01004401 DIFF data FIFO overflow
0x01000306 Sample syringe aspirate 0x01004402 DIFF data RAM overflow
Troubleshooting
8-4
volume over range
0x01000307 Sample syringe dispense
volume over range
0x01004403 WBC data FIFO overflow
0x01000308 Sample syringe action
overtime
0x01004404 WBC data RAM overflow
0x01000310 Sample injection syringe busy 0x01004405 RBC data FIFO overflow
0x01000311 Sample injection syringe
photocoupler abnormal
0x01004406 RBC data RAM overflow
0x01000312 Sample injection syringe
failed to return to home
position after action
0x01004407 PLT data FIFO overflow
0x01000313 Sample injection syringe
failed to leave home position
after action
0x01004408 PLT data RAM overflow
0x01000314 Sample injection syringe
failed to return to home
position before action
0x01004409 HGB data FIFO overflow
0x01000315 Sample injection syringe
failed to leave home position
before action
0x0100440a HGB data RAM overflow
0x01000316 Sample injection syringe
aspirate volume over range
0x01004600 Analyzer communication
disconnected
0x01000317 Sample injection syringe
dispense volume over range
0x01004601 Analyzer network
configuration fail
0x01000318 Sample injection syringe
action overtime
0x01005200 Loading motor busy
0x01000320 Sheath fluid syringe busy 0x01005201 Manipulator pinch motor busy
0x01000321 Sheath fluid syringe
photocoupler abnormal
0x01005202 Manipulator pinch motor
failed to leave home position
0x01000322 Sheath fluid syringe failed to
return to home position after
action
0x01005203 Manipulator pinch motor
failed to return to home
position
0x01000323 Sheath fluid syringe failed to
leave home position after
action
0x01005204 Manipulator pinch motor
action forbidden
0x01000324 Sheath fluid syringe failed to
return to home position before
action
0x01005205 Manipulator pinch motor
initialization failed
0x01000325 Sheath fluid syringe failed to 0x01005206 Manipulator pinch motor
Troubleshooting
8-5
leave home position before
action
action overtime
0x01000326 Sheath fluid syringe aspirate
volume out of range
0x01005207 Manipulator pinch motor
failed to leave end position
0x01000327 Sheath fluid syringe dispense
volume out of range
0x01005208 Manipulator pinch motor
failed to return to end position
0x01000328 Sheath fluid syringe action
overtime
0x01005400 Manipulator pinch motor
initialization forbidden
0x01000330 Lyse syringe busy 0x01005401 Manipulator elevation motor
busy
0x01000331 Lyse syringe photocoupler
abnormal
0x01005402 Manipulator elevation motor
failed to leave home position
0x01000332 Lyse syringe failed to return to
home position after action
0x01005403 Manipulator elevation motor
failed to return to home
position
0x01000333 Lyse syringe failed to leave
home position after action
0x01005404 Manipulator elevation motor
action forbidden
0x01000334 Lyse syringe failed to return to
home position before action
0x01005405 Manipulator elevation motor
initialization failed
0x01000335 Lyse syringe failed to leave
home position before action
0x01005406 Manipulator elevation motor
action overtime
0x01000336 Lyse syringe aspirate volume
out of range
0x01005407 Manipulator elevation motor
failed to leave end position
0x01000337 Lyse syringe dispense volume
out of range
0x01005408 Manipulator elevation motor
failed to return to end position
0x01000338 Lyse syringe action overtime 0x01005500 Manipulator elevation motor
forbid mix motor action
0x01000340 Diluent syringe busy 0x01005501 Mix motor busy
0x01000341 Diluent syringe photocoupler
abnormal
0x01005502 Mix motor failed to leave
home position
0x01000342 Diluent syringe failed to return
to home position after action
0x01005503 Mix motor failed to return to
home position
0x01000343 Diluent syringe failed to leave
home position after action
0x01005504 Mix motor action forbidden
0x01000344 Diluent syringe failed to return
to home position before action
0x01005505 Mix motor initialization failed
0x01000345 Diluent syringe failed to leave
home position before action
0x01005506 Mix motor action overtime
0x01000346 Diluent syringe aspirate 0x01005507 Mix motor failed to leave end
Troubleshooting
8-6
volume out of range position
0x01000347 Diluent syringe dispense
volume out of range
0x01005508 Mix motor failed to return to
home position
0x01000348 Diluent syringe action
overtime
0x01005600 Action forbidden when the
mix motor is not at home
position
0x01000401 Optical system setting out of
limit
0x01005601 Mix mechanism busy
0x01000402 DIFF bath temperature setting
out of limit
0x01005602 Mix mechanism initialization
failed
0x01000601 Setting pressure failed 0x01005603 Mix mechanism action
overtime
0x01000602 Setting vacuum setting failed 0x01005604 Mix time setting out of limit
0x01000603 Adjusting pressure chamber
pressure failed
0x01005605 Mix mechanism initialization
failed
0x01000604 Adjusting vacuum chamber
vacuum failed
0x01005606 Mix mechanism current
action forbidden
0x01000605 Pressure test failed 0x01005700 Feeding motor busy
0x01000606 Vacuum test failed 0x01005701 Feeding motor initial position
photocoupler error
0x01000607 Pressure chamber pressure
out of control limit
0x01005702 Feeding position tube rack
error
0x01000608 Vacuum chamber vacuum out
of control limit
0x01005703 Feeding motor end position
photocoupler error
0x01000611 Flow cell clog 0x01005704 Feeding motor photocoupler
error
0x01000701 Driver board FPGA download
symbol error
0x01005705 Feeding motor loading error
0x01000702 Driver board FPGA download
command error
0x01005706 Feeding motor action
overtime
0x01000703 Driver board FPGA load
address out of range
0x01005708 Feeding position switch error
0x01000704 Drive board single data written
out of limit
0x01005801 Autoloader board functional
code error
0x01000705 Drive board data write out of
limit
0x01005802 Autoloader board data length
error
0x01000706 Drive board write forbidden 0x01005803 Autoloader board command
label error
0x01000707 Drive board FPGA write 0x01005804 Autoloader board check code
Troubleshooting
8-7
overtime error
0x01000731 Drive board EEPROM write
error
0x01005805 Autoloader board end code
error
0x01000801 Drive board functional code
error
0x01005806 Autoloader board identity
code error
0x01000802 Drive board data length error 0x01005820 Autoloader board FPGA
download symbol error
0x01000803 Drive board command label
error
0x01005821 Autoloader board FPGA
download command error
0x01000804 Drive board check code error 0x01005822 Autoloader board FPGA load
address out of range
0x01000805 Drive board end code error 0x01005823 Autoloader board single data
write out of limit
0x01000806 Drive board identity code error 0x01005824 Autoloader board data write
out of limit
0x01000807 Sequence pack analysis busy 0x01005825 Autoloader board write
forbidden
0x01000808 No complete sequence pack 0x01005826 Autoloader board write
overtime
0x01000809 Sequence pack head frame
error
0x01005900 Loading motor receive
command error
0x01000810 Sequence pack start analysis
number error
0x01005901 Loading motor initialization
forbidden
0x01000811 Sequence pack total frames
error
0x01005902 Unloading tray full, unloading
forbidden
0x01000812 Sequence pack save error 0x01005903 Loading motor initial position
photocoupler error
0x01000813 Sequence pack end frame
error
0x01005904 Loading motor unloading
position photocoupler error
0x01000814 Sequence pack body frame
error
0x01005905 Unloading position right
photocoupler error
0x01000815 Sequence pack frame order
error
0x01005906 Unloading position left
photocoupler error
0x01000816 Sequence pack execution
overtime
0x01005907 Loading motor not initialized
0x01000817 None-sequence pack analysis
error
0x01005908 Loading motor feeding not at
initial position
0x01002000 Background abnormal 0x01005909 Loading motor unloading not
at initial position
Troubleshooting
8-8
0x01002010 WBC count starts too slow 0x0100590A Loading motor action
overtime
0x01002011 WBC count time too long 0x0100590B Autoloader receive command
error(not in sampling mode)
0x01002020 WBC count starts too fast 0x0100590C Autoloader receive command
error(not in stop status)
0x01002021 WBC count time too short 0x0100590D Autoloader horizontal loading
photocoupler match error
0x01002022 WBC bubbles 0x0100590E Autoloader unloading tubes
out of limit when initialization
0x01002030 RBC count starts slow 0x0100590F Photocoupler blocked after
autoloader initialization
0x01002031 RBC count time too long 0x01005a00 Autoloader board EEPROM
chip error
0x01002040 RBC count starts too fast 0x01005a01 Opening sample
compartment door failed
0x01002041 RBC count time too short 0x01005a02 Autoloader action forbidden
when autoloader is paused
0x01002042 RBC bubbles 0x01005a03 Autoloader adjust forbidden
when autoloader is not at
adjusting status
0x01003000 Low vacuum 0x01005a04 Autoloader adjustment
parameter setting out of
range
0x01003001 High vacuum 0x01005a05 Scanner read error
0x01003010 Low pressure
Troubleshooting
8-9
8.2 Errors indicated by error messages When operating the analyzer, if error(s) is detected, the corresponding error message will be
displayed on the bottom of the screen. In the error message area, the severity levels are
discriminated from high to low by background colors, with red for the highest and green for the
lowest.
Click the error message on the screen and the corresponding help information will pop up. You
can click the "Remove error" button as instructed by the help information to remove the error,
or, remove some errors as instructed by the troubleshooting procedures of this manual.
8.2.1 Pressure errors
Table 8-2 Pressure errors
Error Message Description Possible causes
Setting pressure failed The pressure created in
the specified time can
not meet the
requirement.
Pressure pump abnormal
Tubing connects the pressure chamber
or pressure pump leaks
Pressure filter leaks
Pressure sensor abnormal, immoderate
error
Related detecting circuit abnormal
Setting vacuum setting
failed
The vacuum created in
the specified time can
not meet the
requirement.
Vacuum pump abnormal
Tubing connects the vacuum chamber or
vacuum pump leaks
Vacuum sensor abnormal, immoderate
error
Related detecting circuit abnormal
8.2.2 Reagent errors
Table 8-3 Reagent errors
Error Message Description Possible causes
No Diluent Diluent is used up or the
sensor of the diluent
pick up tube is
connected improperly.
The Diluent is used up.
The sensor of the diluent pick up tube is
connected improperly or damaged and
needs to be replaced.
No LH lyse LH lyse is used up or
excessive bubbles exist
in the LH lyse channel
The LH lyse is used up.
LH lyse channel leaks, leading to
excessive bubbles.
Troubleshooting
8-10
No LEO(I) lyse LEO(I) lyse is used up or
excessive bubbles exist
in the LEO(I) lyse
channel
The LEO(I) lyse is used up.
LEO(I) lyse channel leaks, leading to
excessive bubbles.
No LEO(II) lyse LEO(II) lyse is used up
or excessive bubbles
exist in the LEO(II) lyse
channel
The LEO(II) lyse is used up.
LEO(II) lyse channel leaks, leading to
excessive bubbles.
No Cleanser Cleanser is used up or
excessive bubbles exist
in the Cleanser channel
The cleanser is used up.
Cleanser channel leaks, leading to
excessive bubbles.
Waste is full The waste container is
full or the sensor
connecting the waste
container is connected
improperly
Waste container is full.
The sensor of the waste pick up tube is
connected improperly or damaged and
needs to be replaced.
Diluent expired Diluent expired Diluent expired
Exp. date setting is wrong
LEO(I) Lyse Expired LEO(I) lyse expired LEO(I) Lyse expired
Exp. date setting is wrong
LEO(II) Lyse expired LEO(II) lyse expired LEO(II) Lyse expired
Exp. date setting is wrong
LH Lyse expired LH lyse expired LH Lyse expired
Exp. date setting is wrong
Cleanser expired Cleanser expired Cleanser expired
Exp. date setting is wrong
8.2.3 Hardware errors Refer to Chapter 5 Hardware System for how to troubleshoot each board.
8.2.4 Measurement errors
Table 8-4 Measurement errors
Error Message Description Possible causes
Background
abnormal
One or more background
results are out of the
specified range
1. Bath, sample probe, probe wipe or
aperture are contaminated.
2. Diluent is expired or contaminated
3. Diluent tubing and back bath tubing
leak, leading to excessive bubbles
Troubleshooting
8-11
4. Bubbles or clog is alarmed during the
background count.
5. Outside interference caused by poor
shielding
LAS background
voltage abnormal
Value is out of the range
of 0-400mv
1. LAS voltage signal is improperly
inducted.
2. Flow cell is contaminated.
3. Related detecting circuit abnormal
HGB background
voltage abnormal
HGB background voltage
is out of the range
[3.2,4.9]V
1. No diluent in the WBC/HGB bath;
HGB light is not turned on.
2. WBC/HGB bath is contaminated.
3. Reagent is expired or contaminated.
4. Improper HGB gain settings
5. HGB light assembly abnormal
6. Related detecting circuit abnormal
RBC clog RBC count time too long
RBC starts too slow
1. Improper settings for reference count
time and start time.
2. The bath is contaminated or the
aperture is clogged by impurity.
3. The photocoupler of the volumetric
tube is damaged.
4. Vacuum abnormal
5. Volumetric-tube-related tubing clogs.
6. Poor connections between the valves
and the driver board.
RBC bubbles RBC count time too short
RBC starts too fast
1. Improper settings for reference count
time and start time.
2. The photocoupler of the volumetric
tube is damaged.
3. Volumetric tube is dirty.
4. Bubbles exist at the tee connector of
the volumetric tube.
5. Fluidic error that leads to the failed
emptying of the volumetric tube.
6. Vacuum abnormal
7. Poor connections between the valves
and the driver board.
WBC clog WBC count time too long
WBC starts too slow
1. Improper settings for reference count
time and start time.
2. The bath is contaminated or the
Troubleshooting
8-12
aperture is clogged by impurity.
3. The photocoupler of the volumetric
tube is damaged.
4. Vacuum abnormal
5. Volumetric-tube-related tubing clogs.
6. Poor connections between the valves
and the driver board.
WBC bubbles WBC count time too short
WBC starts too fast
1. Improper settings for reference count
time and start time.
2. The photocoupler of the volumetric
tube is damaged.
3. Volumetric tube is dirty.
4. Bubbles exist at the tee connector of
the volumetric tube.
5. Fluidic error that leads to the failed
emptying of the volumetric tube.
6. Vacuum abnormal
7. Poor connections between the valves
and the driver board.
8.2.5 Temperature errors
Table 8-5 Temperature errors
Error Message Description Possible causes
Optical system temp.
error
Temperature value read
is out of the range of [30,
40] .℃
The temperature transducer inner the
optical system error
The heater inner the optical system error
DIFF reaction bath
temp. error
Reaction bath abnormal
or reaction bath
temperature transducer
error
The bath temperature is out of the
reference range of 35℃-37 . Please ℃
adjust the temperature to meet the
requirement.
The temperature transducer of reaction
bath is damaged and needs to be
replaced.
Ambient temperature
out of working range
Ambient temperature
abnormal or ambient
temperature transducer
error
The ambient temperature is out of the
working range of 15℃-30 . Please ℃
adjust the temperature to meet the
requirement
The ambient temperature transducer is
damaged and needs to be replaced.
Troubleshooting
8-13
Ambient temperature
out of operating range
Ambient temperature
abnormal or ambient
temperature transducer
error
The ambient temperature is out of the
operating range of 10℃-40 . Please ℃
adjust the temperature to meet the
requirement
The ambient temperature transducer is
damaged and needs to be replaced.
8.2.6 Scattergram errors
Table 8-6 Scattergram errors
Error Message Description Possible causes
Fuzzy division
between MON and
NEU
Fuzzy division between
MON and NEU occur on
normal samples, which
leads to abnormal or no
MON differential result.
Inaccurate volume of DIFF differential
reagents is added in. LEO(I) should be
1.2ml and LEO(II) should be 0.15ml.
The DIFF bath is contaminated.
Flow cell is dirty.
Reaction bath temperature is out of the
normal range. The reaction is not
thoroughly processed.
LYM too high
For normal samples, the
DIFF scattergram
differentiates but fuzzy
division occur between
LYM and Ghost, which
leads to abnormal or no
LYM differential result.
RBCs are not thoroughly hemolyzed.
Inaccurate volume of DIFF differential
reagents is added in. LEO(I) should be
1.2ml and LEO(II) should be 0.15ml.
The DIFF bath is contaminated.
Reaction bath temperature is out of the
normal range. The reaction is not
thoroughly processed.
Spots scatter all over
the scattergram
Spots obtained from
normal patient samples
are randomly scattered
all over the scattergram.
Sheath flow is not formed. Excessive
bubbles enter the WBC channel. Check
whether the diluent container is empty to
make sure if something is wrong with the
diluent sensor.
Compressed
scattergram
For normal patient
samples, the
scattergrams obtained is
compressed either
evenly or to one side.
Calculate the gain of the analyzer
according to the standard scattergram. If
the newly set gain is normal and the
scattergram differentiates properly
without the compressed problem, then no
further adjustment is necessary.
If the gain setup cannot solve the
Troubleshooting
8-14
problem, refer to the optical system
section in the service manual to remove
the error.
A-1
9 Appendixes
A. Accessories
A.1 Spare parts and assemblies No. Code/Model Name
1 3001-30-68501 Data Board
2 3101-30-68503 Drive Board
3 3101-30-68505 Mother Board
4 3101-30-68507 Liquid Detection Board
5 3101-30-68509 Power Supply Board
6 3101-30-68511 Volumetric Board
7 3101-30-68513 Laser Control Board
8 3101-30-68515 FS Pre-amplify Board
9 3101-30-68517 SS Pre-Amplify Board
10 3101-30-68519 Indicator Board
11 3102-30-69199 Autoloader board
12 0033-30-74615 10ml syringe
13 3100-10-49438 100ul syringe
14 3101-10-69301 250ul syringe
15 M90-100032--- 2.5ml syringe
16 3003-20-34942 2-way valve(Asco)
17 3003-20-34941 3-way valve(Asco)
18 3101-30-68611 Release Valve Assembly
19 M07-00014S-00 Pinch Valve Assembly
20 3101-30-68351 Waste Pump Assembly
21 3001-10-07252 Waste Pump
22 530B-10-05275 Pressure Pump
23 3001-10-07057 Volumetric Tube
24 3100-30-40745 tube rack assembly
25 3101-30-68320 Diff Bath Assembly
26 3102-30-69230 HGB assembly
27 3102-30-69253 WBC bath assembly
Appendixes
A-2
28 3102-30-69223 RBC bath assembly
29 3003-20-53969 RBC Ruby (D70nm)
30 3003-20-53970 WBC Ruby (D100nm)
31 3102-20-69171 Sampling probe
32 3101-30-68756 Diluent Cap Assembly
33 3101-30-68765 LEO (I) Lyse Cap Assembly
34 3101-30-68766 LEO (II) Lyse Cap Assembly
35 3101-30-68767 LH Lyse Cap Assembly
36 3101-30-68768 Cleanser Cap Assembly
37 3101-30-68776 Waste Cap Assembly
38 3006-20-74804 Probe Wipe
39 3003-20-34949 Isolation Chamber
40 3102-20-69165 Pressure Chamber
41 3101-30-68547 Vacuum Chamber
42 3001-10-07054 Air filter
43 2800-21-28878 Syringe Motor Position Sensor Assembly
44 3003-21-34925 Sampling Motor Position Sensor Assembly
45 0030-10-13064 Syringe Motor (100ul 250ul 2.5ml 10ml)
46 BA30-10-15115 Sampling motor
47 0000-10-11272 motor 2.33V 1.8'for sample mixing assembly
48 0000-10-11271 motor 2.1V 1.8'for sample mixing assembly
49 3102-30-69231 autoloader assembly
50 3101-30-68651 Optical System Assembly
51 3100-30-49556 Front Laser Assembly
52 3100-30-49559 Back Laser Assembly
53 3100-30-49560 Splitter Assembly
54 3101-30-68662 Las Stop Assembly
55 3101-30-68663 Mas Stop Assembly
56 3101-30-68664 Las PD Assembly
57 3101-30-68665 Mas PD Assembly
58 3100-30-49568 Flow Cell Assembly
59 3100-10-49436 7um particle (15ml)
60 M90-100031--- Tubing, PTFE, 0.066"X0.098"
61 M90-000026--- Tubing, PTFE,1/32"X1/16", 3000074
62 M90-000025--- Tubing, 1/8"X1/4",R-3603AAC02007, Tygon
Appendixes
A-3
63 M6G-020011--- Tubing, PharMed, 1/16"ODX1/8"ID
64 M6G-020009--- Tubing, 0.031"ID0.156"OD
65 M6G-020007--- Tubing, OD3mm ID1mm EVA
66 M6G-020006--- Tubing, Silicone, 1/16"X3/16", TYGON 3350
67 A21-000007--- Tubing, Silicone,3/32"X7/32"2800392-100
68 3001-10-07069 Tubing,1/16"X1/8",S-50-HLAAX02002,Tygon
69 0040-10-32301 Tubing,PTFE,.040"ID+/-.002"X.066"OD+/-.002”
70 3101-20-68590 Fan of Power Supply Assembly
71 3101-20-68572 Fan of Main Unit
72 M07-00097S--- Switch
73 3001-10-07060 Aspirate key
A.2 Spare pipes No. Code/Model Name
1 0040-10-32301 Tubing.PTFE,.040""ID+/-.00
2 3001-10-07069 Tubing.1/16"X1/8",S-50-HLAAX02002,Tygon
3 A21-000002--- Tubing.Silicone,1/8"X1/4" X100ft,2800546-100
4 M6G-020006--- Tubing.Silicone,1/16""X3/1
5 M6G-020007--- Tubing OD3mm ID1mm EVA
6 M6G-020009--- Tubing. Silicone 0.031""ID
7 M90-000025--- Tubing.1/8""X1/4"",R-3603
A.3 Spare PCBA No. Code/Model Name
1 051-000014-00 Data board (including rootfs) 2 3101-30-68503 Drive board 3 3101-30-68505 Mother board 4 3101-30-68507 Liquid-level board 5 3101-30-68509 Power board 6 3101-30-68511 Volumetric board 7 3101-30-68513 Laser control board 8 3101-30-68515 FS pre-amplification board 9 3101-30-68517 SS pre-amplification board
10 3101-30-68519 Indicator board 11 3102-30-69197 Key board 12 3102-30-69199 Autoloading board
Appendixes
A-4
A.4 Spare cables No. Code/Model Name
1 3100-20-49018 Network cable (for connection between the analyzer and the computer)
2 3100-20-49040 Connecting cable of Micro-switch 3 3101-20-68574 Connecting cable of reagent sensor 4 3101-20-68575 Connecting cable of volumetric board 5 3101-20-68578 Connecting cable of power switch 6 3101-20-68581 Connecting cable of laser control 7 3101-20-68584 SS signal line 8 3101-20-68585 FS signal line 9 3101-20-68589 Connecting cable from switch to power board
10 3101-20-68590 Power box fan and connecting cable 11 3101-20-68591 Outlet earth wire 12 3101-20-68592 Connecting cable of pump 13 3101-21-68588 Connecting cable of AC input 14 3101-21-68593 Connecting cable of laser 15 3102-20-69101 Power cord of the autoloading unit 16 3102-20-69102 Connecting cable of autoloader control 17 3102-20-69104 Wiring harness of the mixing motors 18 3102-20-69105 Connecting cable of DIFF bath 19 3102-20-69106 Connecting cable of the indicator board 20 3102-20-69107 Connecting cable of the switch control 21 3102-20-69108 Connecting cable of the autoloading motors and sensors 22 3102-20-69103 of the autoloading motors and sensors
A.5 Spare connectors No. Code/Model Name
1 3001-10-07066 Tee connector T220/230-1
2 3102-20-69219 Waste tube connector
3 M90-100009--- Connector. Female Luer,Lug,Panel,1/4-28UNF,1/8"ID
4 M90-100010---
Connector. Lock Nut, Panel Mount,1/4-28UNF,Orange
Nylon
5 M90-100012--- Connector. Lock Nut, Panel Mount,1/4-28UNF,White Nylon
6 M90-100013--- Connector. Lock Nut, Panel Mount,1/4-28UNF,Black Nylon
7 M90-100015--- Lock Nut
8 M90-100016--- Connector. Lock Nut, Panel Mount,1/4-28UNF,Blue Nylon
9 M90-100017---
Connector. Male Luer Lock Ring, For MTLP or
LC23,Orange
10 M90-100019--- Connector. Male Luer Lock Ring, For MTLP or LC23,White
11 M90-100020--- Connector. Male Luer Lock Ring, For MTLP or LC23,Black
Appendixes
A-5
12 M90-100021--- Connector. Male Luer Lock Ring, For MTLP or LC23,Red
13 M90-100022--- Connector. Male Luer Lock Ring, For MTLP or LC23,Green
14 M90-100023--- Connector. Male Luer Lock Ring, For MTLP or LC23,Blue
15 M90-100024--- Connector. Male Luer,1/16"Barb,White Nylon,MTL210-1
16 M90-100025--- Connector. Male Luer,1/8"Barb,White Nylon,MTL230-1
17 M90-100028--- Connector.Tee,400Barb,3/32"ID,White Nylon
18 M90-100028-01 Connector. Straight Through,40 0Barb,3/32"ID,White
19 M90-100028-03 Connector.Y,400Barb,3/32"ID,White Nylon
20 M90-100030--- Connector.Y,200Barb,1/8"ID,White Nylon
21 M90-100046--- Connector. Coded Lock Ring, For FTLLB or FTLB, Orange
22 M90-100048--- Connector. Coded Lock Ring, For FTLLB or FTLB, White
23 M90-100049--- Connector. Coded Lock Ring, For FTLLB or FTLB, Black
24 M90-100051--- Connector. Coded Lock Ring, For FTLLB or FTLB, Green
25 M90-100052--- Connector. Coded Lock Ring, For FTLLB or FTLB, Blue
26 M90-100065--- Connector. Tee Reduction,400Barb,1/8"&3/32"ID,White
27 M90-100069--- Connector.Thread,10-32UNF,1/4"Hex,400Barb,3/32"ID
A.6 Consumables No. Name Code/Model Package
Specifications
1 M-53LEO(I)
Lyse(Chinese/4L) A11-000105--- 4LX1
2 M-53LEO(I)
Lyse(Chinese/1LX4) A11-000113--- 1LX4
3 M-53LEO(I)
Lyse(English/1LX4) A11-000128--- 1LX4
4 M-53LEO(I)
Lyse(English/4L) A11-000129--- 4LX1
5 M-53LEO(II)
Lyse(Chinese/200mlX4)A11-000114--- 200mLX4
6 M-53LEO(II)
Lyse(Chinese/400mlX4)A11-000115--- 400mLX4
7 M-53LEO(II)
Lyse(English/200mlX4) A11-000130--- 200mLX4
8 M-53LEO(II)
Lyse(English/400mlX4) A11-000131--- 400mLX4
9 M-53LH
Lyse(Chinese/500mlX4)A11-000116--- 500mLX4
Appendixes
A-6
10 M-53LH
Lyse(Chinese/1LX4) A11-000117--- 1LX4
11 M-53LH
Lyse(English/500mlX4) A11-000132--- 500mLX4
12 M-53LH
Lyse(English/1LX4) A11-000133--- 1LX4
13 M-53
Cleanser(Chinese/1LX4)A11-000118--- 1LX4
14 M-53
Cleanser(English/1LX4) A11-000134--- 1LX4
15 M-53D
Diluent(domestic/20L) A12-000140--- 20LX1
16 M-53D
Diluent(export/20L) A12-000167--- 20LX1
17 M-50P Probe Cleanser
(domestic/50mL) A12-000151---
50mLX1
18 M-50P Probe Cleanser
(export/50mL) A12-000178---
50mLX1
B-1
B. List of Wearing Parts
No. Code/Model Name
1 3001-10-07054 Air Filter
2 3003-20-34941 Tee valve ASCO458284
3 3003-20-34942 Two-way valve ASCO458283
4 3006-20-74804 Probe Wipe
5 M07-00014S-00 VAL S305 07-Z030H 12VDC
6 M07-00076F--- FUSE Slow-Blow
7 M07-00077F--- FUSE Slow-Blow 125V 5A
8 M07-00079F--- FUSE Slow-Blow 125V 7A
9 M08-000789--- PHOTOELEC switch
10 M6G-020011--- Tubing.PharMed,1/16""ODX1/
C-1
C. Fluidic diagram
D-1
D. Pump and Valve Function Table
D1. Valve Function Table
No. Function Two
way/Tee
V01 Provides fluid for the sample probe; V01 opens, the waste discharging
path of the diluent syringe connects the sample probe. 3
V02 Provides fluid for the probe wipe; V02 opens, the waste discharging
path of the diluent syringe connects the inlet of the probe wipe. 3
V03 Aspirating/discharging switch of diluent syringe 3
V04 Cleans the interior of the sample probe; V04 opens, the sample probe
connects the sheath fluid syringe. 2
V05 Aspirating/discharging switch of LEO (I) lyse 3
V06 Aspirating/discharging switch of LEO (II) lyse 3
V07 Aspirating/discharging switch of LH lyse 3
V08 Provides fluid for RBC bath; V08 opens, the waste discharging path of
the diluent syringe connects the fluidic line of RBC bath 3
V09 Opens and closes the pressure chamber and the waste outlet of the
DIFF bath; bubbles the DIFF bath 2
V10 Opens and closes the path between the vacuum chamber and the
pressure chamber 2
V11 Opens and closes the pressure chamber and the isolation chamber t of
the RBC bath; bubbles the RBC bath 2
V12 Opens and closes the pressure chamber and the isolation chamber t of
the WBC bath; bubbles the WBC bath 2
V13 Together with V33 and V29 to control the cleaning of the WBC back
bath 2
V14 Together with V34 and V30 to control the cleaning of the RBC back bath 2
V15 Controls the aspiration of the cleanser 3
V21 Opens and closes the sheath fluid syringe and the sample dispensation
syringe; assists the sheath fluid dispensing the sample 2
V22 Opens and closes the sheath fluid syringe and DIFF bath; cleans DIFF
bath 2
V23 Aspirating/discharging switch of sheath fluid syringe 3
V24 Opens and closes the sheath fluid syringe and flow cell; dispenses
sheath fluid 2
V25 Empties the WBC volumetric tube 2
Appendixes
D-2
V26 Empties the RBC volumetric tube 2
V27 Controls the waste discharging of the probe wipe 2
V28 Controls the waste discharging of the flow cell 2
V29 Controls the cleaning of the WBC back bath 3
V30 Controls the cleaning of the RBC back bath 3
V31 Empties the WBC volumetric tube 3
V32 Empties the RBC volumetric tube 3
V33 Controls the WBC counting 2
V34 Controls the RBC counting 2
V35 Controls the draining of the RBC bath 2
V36 Controls the draining of the WBC bath 2
V37 Controls the draining of the vacuum chamber 2
V38 Controls the draining of the DIFF bath 2
V39 Controls the preparation of the DIFF sample 2
D2. Pump function table Name Function
P1 Creates pressure in the pressure chamber, which is used to bubble and flash
the aperture
P2
Discharges the waste of the DIFF bath; empties and creates vacuum in the
vacuum chamber to drive the counting by impedance method, clean the back
bath and drain the volumetric tube.
P3 Discharges the waste of the WBC bath, RBC bath and probe wipe
D3. Syringe function table Name Function
Diluent syringe Dispenses diluent into the WBC bath and RBC bath; cleans the sample
probe and dispenses fluid for the probe wipe
Sheath syringe Prepares samples; forms sheath fluid; cleans the DIFF bath and the
sample preparation tubing
Aspiration
syringe
Aspirates and dispenses samples
Sample syringe Dispenses the sample into the flow cell
Lyse-Syringe Aspirates and dispenses the LEO (I) Lyse, LEO (II) Lyse and LH Lyse
E-1
E. Tubing
E1. Components
F-1
F. Method to identify cross network cable and direct-connected network cable
F.1. Identify cross network cable 1. With the metal contactors’ side up, mark the 8 lines in the connector from left to right with 1-8.
Accordingly, the colors of the lines are: white and orange, orange, white and green, blue, white
and blue, green, white and brown, and brown.
Connector of the network cable
2. Define one end of the cable to be A randomly. Then, the other end is B. way to connect the
end A and B is shown in the Figure below.
End A and End B
End A:white and green, green, white and orange, blue, white and blue, orange, white and
brown, and brown ( 1 and 3 is exchanged; 2 and 6 is exchanged)
End B:white and orange, orange, white and green, blue, white and blue, green, white and
brown, and brown
F.2. Identify direct-connected network cable 1. With the metal contactors’ side up, mark the 8 lines in the connector from left to right with 1-8.
Accordingly, the colors of the lines are: white and orange, orange, white and green, blue, white
and blue, green, white and brown, and brown.
Appendixes
F-2
Connector of the direct-connected network cable
2. Define one end of the cable to be A randomly. Then, the other end is B. way to connect the
end A and B is shown in the Figure below.
End A and End B of the direct-connected network cable
End A:white and orange, orange, white and green, blue, white and blue, green, white and
brown, and brown
End B:white and orange, orange, white and green, blue, white and blue, green, white and
brown, and brown
P/N: 046-000142-00(V1.0)