BD FACS Canto II
The BD FACS Canto II offers three lasers and is able to detect up to 8 parameters ( fluorochromes) at the same time!
For the analysis at the Canto II you have the choice between two different configurations:
Configuration 4-2-2 (BD default)
A photodiode is used to detect the FSC signal!
Detector Array
Laser Parameter PMT- Position
BP-Filter
LP-Filter
Examples for possible
Fluorochromes Trigon 405 nm
-violett- 1 2
A B
510/50 450/50
502 LP
AmCyan Pacific Blue
Octagon 488 nm -blue-
3 4
5 6
SSC (Granularity)
A B C D E F G H
780/60 670 LP
585/42 530/30 488/10 Empty Empty
735 LP 655 LP 610 LP 556 LP 502 LP
PE-Cy7 PerCP, PerCP-Cy5.5
PE
FITC
Trigon 633 nm -red-
7
8
A B C
780/60 685 LP 660/20
735 LP APC-Cy7
APC
Configuration Copy of 4-2-2
A photodiode is used to detect the FSC signal!
The samples can only be injected via 5mL Round-Bottom Tubes with a maximum filling volume of 4mL !
Technical Highlights Lasers 488-nm solid state, 20-mW laser output
633-nm HeNe, 17-mW laser output 405-nm solid state diode, 30-mW fiber
Flow cell 180 x 430µm rectangular quartz flow cell FSC Detector Photodiode SSC Detector Photomultiplier with 488/10 BP Maximum sample acqusition Rate
10,000 events per second
Parameters Area (A), Width (W), Height (H) for all channels with up to two ratios and Time (T)
Compensation Full inter-beam matrix during or post acquisition
Software BD FACS DivaTM Software v.6.1.3
Detector Array
Laser Parameter PMT- Position
BP-Filter
LP-Filter
Examples for possible
Fluorochromes Trigon 405 nm
-violett- 1 2
A B
C
529/24 457/50
Empty
502 LP v-GFP, v-YFP CFP, Hoechst, BFP, Pacific
Blue, Cerulan
Octagon 488 nm -blue-
1 2
3 4 5
SSC (Granularity)
A B
C D E F G H
780/60 670 LP
529/24 509/20 488/10 Empty Empty
735 LP 610 LP
556 LP 525 LP 502 LP
PE-Cy5, PE-Cy7 PerCP, PerCP-Cy5.5, PI,
7AAD, PE-Texas Red, e-RFP
GFP, Venus, YFP FITC, eGFP, eYFP
Trigon 633 nm -red-
1
2
A B C
780/60
660/20
735 LP 685 LP
APC-Cy7
APC