BIOQUIMICA KOVALENT ADVIA1200 - Grupo .Protocolo de Automação – ADVIA 1200 Analytical conditions

  • View
    214

  • Download
    0

Embed Size (px)

Text of BIOQUIMICA KOVALENT ADVIA1200 - Grupo .Protocolo de Automação – ADVIA 1200 Analytical conditions

  • Rev: 3 02/2011

    cido rico

    Referncia Apresentao

    1010250K R1 1 x 200 mL + R2 1 x 50 mL + 1 x 3 mL padro de cido rico

    1010500K R1 2 x 200 mL + R2 1 x 100 mL + 1 x 3 mL padro de cido rico

    1010250T R1 10 x 20 mL + R2 2 x 25 mL + 1 x 3 mL padro de cido rico

    1010100M R1 2 x 40 mL + R2 2 x 10 mL + 1 x 3 mL padro de cido rico

    Protocolo de Automao ADVIA1200 Analytical conditions Sub-analy. conditions

    R1 volume 80 Name UA

    R2e volume 0 Digits 1

    R2 volume 20 M-wave L. 505

    R1 diluent vol 0 S-wave.L 694

    R2e diluent vol 0 Analy.Methd. EPA

    R2 diluent vol 0 Calc.Methd. STD

    Sample Vol (S) 2 Qualit. Judge Not do

    Sample Vol (U) 2 Calculation method setting

    Reagent 1 stir Weak M-DET.P.l 0

    Reagent 2e stir Weak M-DET.P.m 40

    Reagent 2 stir Weak M-DET.P.n 42

    Reaction time 10 S-DET.P.p 18

    Sample Type Serum S-DET.P.r 19

    Dilution specification Do Check D.P.I. 0

    Dilution correction Not Do Limit value 0,003

    Liquid volume correction Not Do Variance 10

    Reaction sample volume 2 Reac.typ Inc.

    Diluent method No dilution Cykle 2

    Undiluted sample volume 0 Factor 2

    Diluent volume 0,000 E2 corr Not Do

    Diluent position 0 Blank (u)

    Diluent volume from RPP 0,000 Blank (d)

    Sample Type Urine Sample (u)

    Dilution specification Do Sample (d)

    Dilution correction Do

    Liquid volume correction Do Endpoint method

    Reaction sample volume 2 Re.absorb (u)

    Diluent method No dilution Re.absorb (d)

    Undiluted sample volume 0 Rerun conditions set

    Diluent volume 0,000 Absorbance limit (u) D1

    Diluent position 0 Absorbance limit (d) D2

    Diluent volume from RPP 0,000 Effect.nbr.o.pnts (n)

    Abnormal val.limit (h) D1

    Abnormal val.limit (l) D2

    Reanalysis conditions

    Sample Type Serum Sample Type Urine

    Dilute 1 Dilute 1

    Serum reac.smp.vol (u) 2 Serum reac.smp.vol (u) 2

    Dilution method A dilution Dilution method A dilution

    Undiluted sample volume 20 Undiluted sample volume 20

    Diluent volume 40 Diluent volume 40

    Diluent position 1 Diluent position 1

    Diluent volume from RPP 0,000 Diluent volume from RPP 0,000

    Dilute 2 Dilute 2

    Serum reac.smp.vol (u) 10 Serum reac.smp.vol (u) 10

    Dilution method No dilution Dilution method No dilution

    Undiluted sample volume 0 Undiluted sample volume 0

    Diluent volume 0 Diluent volume 0

    Diluent position 0 Diluent position 0

    Diluent volume from RPP 0,000 Diluent volume from RPP 0,000

    Abnormal value setting

    Main Abnormal value (H1)

    Main Abnormal value (L1)

  • Rev: 3 02/2011

    Notas: 1. Por favor, recorra a bula do produto para informaes detalhadas sobre os seguintes testes: Relevncia clnica Mtodo e Princpio Composio e Estabilidade dos Reagentes Amostras Calibradores e Controles Desempenho e Caractersticas considerando: - Faixa de medio - Especificidade/Interferentes - Sensibilidade/Limite de Deteco - Preciso (Reprodutibilidade, Repetibilidade) - Comparao de mtodo - Valores de referncia - Literatura 2. A estabilidade do reagente a bordo do analisador pelo menos um ms contanto que sejam evitadas a contaminao e a evaporao. 3. Fabricado por: Kovalent do Brasil Ltda. Rua Cristovo Sardinha, 110 Jardim Bom Retiro So Gonalo

  • Rev: 3 - 02/2011

    Albumina

    Protocolo de Automao ADVIA 1200

    Analytical conditions

    R1 volume 100.0 Sub Param. 46 1 Up Down

    R2e volume 0.00 Sub-analy-conditions Standards setting

    R3 volume 0.00 Name Alb FV *** BLK-L -9.999 STD H -9.999 Multi-STD Setting

    Digits 1 BLK-H 9.9999 STD L 9.9999 Error judge rare

    R1 Extra vol H-wave.L. 596nm

    R2 Extra vol S-wave.L. 694nm

    R3 Extra vol Anal.mthd EPA

    R1 diluent vol 0.00 Calc.mthd STD

    R2e diluent vol 0.00

    Quali.judg Not do

    R3 diluent vol 0.00 CCalculation method setting

    Sample Vol( S ) 1,0 M-DET.P.1 0 S-DETP.p. 0 Reac.typ Inc

    Sample vol (U) 1,0 Qualit. Judgment set M-DET.P.m 41 S-DET.P.r 0 Reaction rate method

    Real time correct form M-DET.P.n 42 Cycle 2

    Reanalysis conditions Check D.P.I 0 Factor 2.0

    Rerun conditions set Limit value 0.003 E2 corre Not do

    Reagent 1 stir Weak Mark setting Variance 10.0 Blank() 9.9999

    Reagent 2e stir Weak Normal Value setting * Prozone Blank(d) -9.9999

    Reagent 2 stir Weak Abnormal value setting Prozone form None Sample() 9.9999

    Reaction Time 10 min. Prozone limit 9.999 Sample

    (d)

    -9.9999

    Prozone judge Upper limit

    Anaysis item

    Condition setting (M)

    Judge limit 9.999 * Endpoint method

    M-DET.P.m 0 S-DET.P.p 0 Re.absorb() 9.9999

    M-DET.P.n 0 S-DET.P.r 0 Re.absorb(d) -9.9999

    IMA setting

    Setting

    Allowance

    Referncia Apresentao

    1110060K R 1 x 60 mL + 1 x 3 mL Padro de Albumina

    1110250K R 1 x 250 mL + 1 x 3 mL Padro de Albumina

    1110300T R 12 x 25 mL + 1 x 3 mL Padro de Albumina

    1110100M R 2 x 50 mL + 1 x 3 mL Padro de Albumina

  • Rev: 3 - 02/2011

    Notas: 1. Por favor, recorra a bula do produto para informaes detalhadas sobre os seguintes testes: Relevncia clnica Mtodo e Princpio Composio e Estabilidade dos Reagentes Amostras Calibradores e Controles Desempenho e Caractersticas considerando: - Faixa de medio - Especificidade/Interferentes - Sensibilidade/Limite de Deteco - Preciso (Reprodutibilidade, Repetibilidade) - Comparao de mtodo

    - Valores de referncia - Literatura

    2. A estabilidade do reagente a bordo do analisador pelo menos um ms contanto que sejam evitadas a contaminao e a evaporao. 3. Fabricado por: Kovalent do Brasil Ltda. Rua Cristovo Sardinha, 110 Jardim Bom Retiro So Gonalo

  • Rev: 3 - 02/2011

    Alfa Amilase G7 Referncia

    Apresentao

    2080075K

    R1 3x20mL + R2 1x15mL

    2080250T

    R1 10x20mL + R2 2x25mL

    Protocolo de Automao ADVIA 1200

    Analytical conditions

    R1 volume 80.00 Sub Param. 46 1 Up Down

    R2e volume 0.00 Sub-analy-conditions Standards setting

    R3 volume 20.00 Name Ami

    FV

    6300 BLK-L -9.999 STD H -9.999 Multi-STD Setting

    Digits 0 BLK-H 9.9999 STD L 9.9999 Error judge rare

    R1 Extra vol H-wave.L. 410nm

    R2 Extra vol S-wave.L. 694nm

    R3 Extra vol Anal.mthd RRA

    R1 diluent vol 0.00 Calc.mthd ABB

    R2e diluent vol 0.00

    Quali.judg Not do

    R3 diluent vol 0.00 CCalculation method setting

    Sample Vol( S ) 6,0 M-DET.P.1 25 S-DETP.p. 0 Reac.typ Inc

    Sample vol (U) 6,0 Qualit. Judgment set M-DET.P.m 34 S-DET.P.r 0 Reaction rate method

    Real time correct form M-DET.P.n 42 Cycle 2

    Reanalysis conditions Check D.P.I 21 Factor 2.0

    Rerun conditions set Limit value 0.003 E2 corre Not do

    Reagent 1 stir Weak Mark setting Variance 10.0 Blank() 0,2270

    Reagent 2e stir Weak Normal Value setting * Prozone Blank(d) 0,0110

    Reagent 2 stir Weak Abnormal value setting Prozone form None Sample() 1,2000

    Reaction Time 10 min. Prozone limit 9.999 Sample

    (d)

    0,0110

    Prozone judge Upper limit

    Anaysis item

    Condition setting (M)

    Judge limit 9.999 * Endpoint method

    M-DET.P.m 0 S-DET.P.p 0 Re.absorb() 9.9999

    M-DET.P.n 0 S-DET.P.r 0 Re.absorb(d) -9.9999

    IMA setting

    Setting

    Allowance

  • Rev: 3 - 02/2011

    Notas 1. Por favor, recorra a bula do produto para informaes detalhadas sobre os seguintes testes: Relevncia clnica Mtodo e Princpio Composio e Estabilidade dos Reagentes Amostras Calibradores e Controles Desempenho e Caractersticas considerando: - Faixa de medio - Especificidade/Interferentes - Sensibilidade/Limite de Deteco - Preciso (Reprodutibilidade, Repetibilidade) - Comparao de mtodo

    - Valores de referncia - Literatura

    2. A estabilidade do reagente a bordo do analisador pelo menos um ms contanto que sejam evitadas a contaminao e a evaporao. 3. Fabricado por: Kovalent do Brasil Ltda. Rua Cristovo Sardinha, 110 Jardim Bom Retiro So Gonalo

  • Rev: 3 - 02/2011

    Bilirrubina Direta

    Referncia Apresentao

    1080075K R1 3 x 20 mL + R2 1 x 15 mL

    1080250K R1 5 x 40 mL + R2 1 x 50 mL

    1080250T R1 10 x 20 mL + R2 2 x 25 mL

    1080100M R1 2 x 40 mL + R2 2 x 10 mL

    Protocolo de Automao ADVIA 1200 Analytical conditions Sub-analy. conditions

    R1 volume 80 Name DBIL

    R2e volume 0 Digits 1

    R2 volume 20 M-wave L. 545

    R1 diluent vol 0 S-wave.L 694

    R2e diluent vol 0 Analy.Methd. EPA

    R2 diluent vol 0 Calc.Methd. STD

    Sample Vol (S) 6 Qualit. Judge Not do

    Sample Vol (U) 6 Calculation method setting

    Reagent 1 stir Weak M-DET.P.l 0

    Reagent 2e stir Weak M-DET.P.m 40

    Reagent 2 stir Weak M-DET.P.n 42

    Reaction time 10 S-DET.P.p 1