Chapter 8 The X-ray Imaging System

• The Console is the part of the machine that the operator controls the operation of the x-ray machine.

• All machine console are a little different but there are always similarities. The console is where we control x-ray tube current and voltage.

Page 2: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

The Console Controls

• The console will have controls for:

• mA and time or mAs

• kVp

• Focal Spot

• Line Voltage Compensation

• Automatic Exposure Control

Page 3: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Symbols Used to Draw Circuits

• We will be using the symbols to define the circuits in the x-ray machine

Page 4: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Console Circuits

Page 5: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Line Compensation

• At the bottom left is the controls for line voltage compensation.

• Most machine are designed to operate at 220 volts while some will work with 110 volts or 440 volts

Page 6: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Line Compensation

• The power company often cannot provide exactly 220 volts at all times.

• Elevators and Air Conditioners may reduce the voltage available for the x-ray unit.

Page 7: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Line Compensation

• Older machine have a meter to monitor the line voltage attached to the autotransformer.

• The operator can adjust the taps on the transformer to account for low or high incoming voltage.

Page 8: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Line Compensation

• More modern units automatically adjusts for the incoming power so a meter is not provided.

• Often over looked by the operator.

• Results in improper exposure.

Page 9: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Autotransformer

• The autotransformer is designed to supply voltage of varying magnitude to several different circuits of the x-ray machine including both the filament circuit and high voltage circuits.

Page 10: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Autotransformer

• The autotransformer has only one winding and one core.

• The single winding has a number of connection or electric taps.

Page 11: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

kVp Adjustment

• Most consoles will have one or two knobs that change the taps on the autotransformer for major and minor kVp.

• Modern units have a LED readout of kVp.

Page 12: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

kVp Adjustment

• Setting the desired kVp will determine the voltage applied to the step-up transformer in the high voltage section of the machine.

Page 13: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

kVp Adjustment

• If a meter is provided, it is placed across the output terminals of the autotransformer and therefore it reads voltage and not kVp. The scale will read in kVp.

Page 14: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mA Control

• The tube current, the number of electrons crossing from the cathode to anode per second is measured in milliapmeres (mA).

• The quantity of electrons is determined by filament temperature.

Page 15: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mA Control

• The filament normally operates at currents between 3 and 6 A.

• The Tube Current is controlled through a separate circuit called the filament circuit

Page 16: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mA Control

• Voltage is provided by taps of the autotransformer. This voltage is reduced with precise resisters to a value corresponding to the mA stations available.

Page 17: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mA Control

• Tube current is usually not continuously variable, usually only currents of 50, 100, 150, 200 & 300 mA and higher are provided.

• Newer units are continuously variable.

Page 18: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mA Control

• The voltage is then delivered to the filament transformer. The filament transformer lowers the voltage so it is called a step down transformer.

Page 19: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mA Control

• The selection of the small or large filament is connected to the mA selection or as a separate control.

Page 20: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Exposure Timers

• For any given radiographic examination, the number of x-rays reaching the image receptor is directly related to the tube current and the time that the tube in energized.

• The timer circuit is separate from the other main circuits.

Page 21: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Exposure Timers

• It consists of a mechanical or electronic device whose action is to make and break the high voltage across the tube on the primary side of the high voltage section.

Page 22: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Types of Timers

• There are five types of timers:

• Mechanical Timers

• Synchronous Timers

• Electronic Timers

• mAs Timers

• Phototimers

Page 23: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Mechanical Timers

• Very simple device that has a clock mechanism.

• Operator turns the dial to the desired time. As it unwinds, the exposure is made.

• Can be used for exposure time longer than 250 milliseconds.

• Very old machine and dental units.

Page 24: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Electronic Timers

• Most sophisticated, complicated and most accurate timer.

• Consists of complex circuit based upon the time required to charge a capacitor through a variable resister.

• Depending upon the incoming power accurate to 1 ms. Most units have this type timer.

Page 25: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mAs Timers

• Most modern machine are designed to accurately control the tube current and exposure time.

• The product of mA and time (mAs) determines the number of x-ray photons emitted and the density on the film.

Page 26: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mAs Timer

• A special type of timer monitors the product of mA and terminates the exposure when the desired mAs has been attained.

• This is a mAs timer.

Page 27: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mAs Timer

• Designed to provide the shortest exposure and the highest safe tube current for the given filament.

• Some have the ability to change mA manually.

Page 28: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mAs Timer

• Since it monitors the actual tube current, it is on the secondary side of the H.V. Circuit

• Units here have mAs timers.

Page 29: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

mAs Timer

• APR or Anatomically Programs Timers have computers that store the technical factors in the machine.

• Select the view and enter the patient size and the machine is ready!!!!

Page 30: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Phototimers

• A phototimer that measures the quantity of radiation reaching the receptor and terminates the exposure when sufficient radiation needed to produce the correct density on the film.

• Offered in addition to a manual timer.

Page 31: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Phototimers

• There are two types of phototimers:

• 1. Photomultiplier tube that reads a fluorescent screen behind the film.

• 2. Ion chamber between the grid and film.

Page 32: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Phototimers

• Ion Chambers is used on most modern x-ray units.

• It is flat and radiolucent so it will not interfere with the image. Multiple chambers can be used to optimize the image.

Page 33: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Phototimers

• Commonly referred to as Automatic Exposure Control or AEC.

• Widely used in Medical Radiography.

• Used at our Benton Clinic.

Page 34: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

AEC Console

• With AEC, the operator can select:

• Where to read the radiation.

• The desired film density

• kVp and backup mAs

Page 35: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

AEC Console

• Many operators do not measure the patient and set a arbitrary back up mAs or time.

• Ideally, the patient is measured and the back up mAs is set at 2X the normal mAs.

Page 36: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

AEC Console

• This allows the AEC to adjust exposure for the patient’s habitus and area density.

• Radiation is measured at the center of the film or off to the sides of the film.

Page 37: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

AEC Console

• The center is read for most radiography and especially for the spine.

• The sides are read for PA chest, abdomen and rib radiography.

Page 38: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Other functions on the Control Console.

• The console will also have the exposure button or buttons.

• The prep button is depressed to prepare the tube for exposure.

• The rotor will spin up to 3400 RPM.

Page 39: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Exposure Button

• A green light will let you know that the machine is ready to make the exposure.

• The exposure button is then depressed and the exposure is initiated.

Page 40: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Exposure Button

• The button must be held down until the exposure is complete.

• If your finger slips off the button, the exposure is terminated.

Page 41: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Exposure Button

• The exposure control buttons are referred to as a”Dead man Switch”

• After the buttons are released, the rotor motor reverses and the rotor reduces speed.

Page 42: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Exposure Button

• During the exposure you will hear an audible tone so you will know that the exposure is in progress.

Page 43: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Chapter 8 High Voltage Section

• The high voltage section converts low voltage from incoming power to kilo-voltage of the correct wave form.

• It is usually enclosed in a large metal container in the x-ray room.

Page 44: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

High Voltage Section

• It consists of three primary sections:– High voltage step up

transformer– Filament Transformer– Rectifiers ( Diodes)

• All components immersed in oil.

Page 45: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

High Voltage Transformer

• The high voltage transformer is a step-up transformer.

• There will be more winding on the secondary side compared to the primary side.

• The ratio of windings is referred to as the turns ratio.

Page 46: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

High Voltage Transformer

• The only difference between the primary and secondary waveforms is the amplitude.

• The turn ratio for most x-ray high voltage transformers is between 500 and 1000.

• Incoming Volts converted to output: Kilovolts.

Page 47: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Voltage Rectification

• Transformers operate with alternating current.

• Remember that x-ray tubes operate on direct voltage ( electron moving in one direction).

• To convert AC to DC we use rectifiers.

Page 48: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Half-Wave Rectification

• Sometimes the x-ray tube alone will work as the diode this is called self-rectification.

• When one or two diodes are placed in the circuit that stops the negative flow of electrons it is called Half Wave Rectification.

• 60 pulses per second.

Page 49: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Full-Wave Rectification

• Full wave rectified x-ray machines contain at least four diodes.

• It changes the polarity of the negative half of the wave.

• This allows 120 pulses of x-ray per second.

• The exposure time can be cut in half compared to half-wave systems.

Page 50: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Three-Phase Power

• If three phases of power are combines with the phase off by one step, the normal reduction of voltage back to zero is removed. Commonly called the Ripple.

• Technical factor cut in half due to more efficient power.

• Too expensive got office use.

Page 51: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

High Frequency Generator

• By changing the frequency from 60 Hz to a higher frequency of 500 to 1000, the ripple is reduced to less than 1%.

• Single phase machine operating on 220 volts and even 110 volts are more efficient that machine operating on three-phase power.

Page 52: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Types of X-ray Generators

• The type of generator will determine the efficiency of the machine.

Page 53: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Wave Forms of Different Generator Types

• As the ripple effect decreases, the efficiency increases.

• There is one more type of generator. It uses is called stored energy.

Page 54: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Stored Energy Generators

• If 220 volt power is not available, the operator may choose a stored energy machine.

• A battery charger is powered by typical house hold current.

• If produces direct current.

Page 55: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Stored energy or Capacitor Discharge Generators

• There is a short charging time before the exposure can be made.

• The disadvantage to the design is a drop in power at the end of the exposure of about 1 kV/mAs. This is called a falling load generator.

Page 56: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Generator Types Pros & Cons

• Single phase half or self rectified: Cheap but not efficient. Full wave rectified better.

• Three phase: Expensive to install but cheaper to maintain. Too costly for most offices. 6 pulse less costly than 12 pulse

• High Frequency: very efficient and works with single or three phase power.

• Stored energy: works on conventional 110 volt power but batteries must be replaced.

Page 57: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

The Basic X-ray Circuits

• Circuits that make up the basic x-ray machine.

Page 58: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Other Parts of the X-ray Room

• The tube is suspended on the tube stand.

• The tube stand may be wall and floor mounted or ceiling suspended.Locks are provided for horizontal and vertical movement.

Page 59: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Other Parts of the X-ray Room

• When the tube is angled toward the wall grid holder, the horizontal lock allow us to set the distance between the tube and the film (SID).

Page 60: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Other Parts of the X-ray Room

• When the tube is aimed at the table, the vertical lock allows us to set the SID.

• Hanging on the wall grid cabinet is the non-Bucky film holder.

• It allows erect non-grid films.

Page 61: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Other Parts of the X-ray Room

• X-ray tables may be bolted to the floor or mobile. The table will also have a grid cabinet for grid radiography.

• We will discuss grids in greater detail next week.

Page 62: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Collimator and Angle Indicator

• The tube stand also has an angle indicator attached parallel to the tube.

• There are views that will require tube angles.

Page 63: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Collimator and Angle Indicator

• The Collimator is attached to the x-ray tube below the glass window where the useful beam is emitted.

• Lead shutters are used to restrict the beam.

Page 64: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Collimator and Angle Indicator

• A mirror and light source allows us to restrict the beam to the area of interest.

• Collimation is our greatest tool in keeping patient exposure as low as possible.

Page 65: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Other items that may be in a x-ray room.

• Fluoroscopy Equipment: Allows dynamic imaging of the body.

• Consists of:

• Image intensifier with television camera and monitor.

• Spot-film device for making radiographs or

• Motion picture camera or digital imaging.

Page 66: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Image Intensifier & Fluoroscopy

• Thomas Edison invented the fluoroscope in 1896. Early units consisted of a fluorescent hand held viewer that the doctor held in from of the patient during continuous exposure.

• This resulted in the first x-ray death.

• Dose is still relatively high compared to plain film radiography.

Page 67: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Image Intensifier & Fluoroscopy

• Plain film radiography uses up to several hundred mA and fractions of seconds.

• Fluoroscopy tubes operate at less than 5 mA but for minutes. 2 to 4 mA is normal.

• In California Fluoroscopy is beyond the chiropractic scope of practice.

Page 68: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Image Intensifier & Fluoroscopy

• Shortly after WW2, Bell Laboratories invented the photomultiplier tube. This was developed into the modern image intensifier.

• The multiplication of the light emitted by a input fluorescent screen is picked up by a cesium photocathode and converted into electrons.

Page 69: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Image Intensifier & Fluoroscopy

• A potential of about 25,000 volts is maintained between the photocathode and the anode.

• There are electronic optics and electrostatic focusing lenses between the photocathode and output phosphor.

Page 70: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Image Intensifier & Fluoroscopy

• The output phosphor can be viewed via mirror optics or a video monitoring system.

• A Videotape recorded can be placed into the video chain.

• Fluoroscopy allows the evaluation of the internal structures in motion. Normal uses include:

Page 71: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Uses of Fluoroscopy

• Dynamic spinal imaging of range of motion and with contrast called myelograms.

• Dynamic studies of joints with or without contrast media.

• Studies of the digestive system.

• Studies of arteries and blood flow called angiography.

Page 72: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

Uses of Fluoroscopy

• When connected to a computer, for digital fluoroscopy and spot films.

• With digital fluoroscopy, digital angiography is possible.

• By over-lapping an image without contrast, digital subtraction angiography is performed where the bone is removed.

Page 73: Chapter 8 The X-ray Imaging System The Console is the part of the machine that the operator controls the operation of the x-ray machine. All machine console

End of Lecture