Manual eddyvisor SC 29 May 2008 (v 01e) ibg Doc: 0713431e

EddyvisorSC v01 En

Embed Size (px)

Citation preview

  • ManualeddyvisorSC

    29 May 2008 (v 01e)

    ibg Doc: 0713431e

  • Software and Manual

    Copyright 2008 by ibg Prfcomputer GmbH,Pretzfelder Str. 27, D-91320 Ebermannstadt

    The copyright claimed covers all forms and types of materials and information fallingunder the copyright, which at the present time are legally admitted or granted in thefollowing, including material generated without restriction from the softwareprogramme appearing on the screen, such as symbols, screen displays and thelike.

    Exclusion of responsibility

    Great care was taken in compiling texts and figures. Nevertheless, errors cannot beruled out completely. Editors and authors cannot assume any legal responsibility orliability for erroneous information and resulting consequences. The editor will readilyaccept any proposals for improvement and hints at errors.

    The unit delivered incorporates the latest state of the art. The manufacturer declinesany legal responsibility and liability for inspections carried out with the apparatusand for resulting consequences. Successful implementation of an inspectionpresupposes, in addition to the control of the unit setting, also a knowledge of theproduct to be tested, its material properties, of possible errors and their underlyingcauses. If these prerequisites are not satisfied, the result will be questionable. Inorder to avoid faulty operation and other errors, training in the manufacturer'sfactory or by an independent training organisation is recommended.

    eddyvisor is a registered trademark of ibg Prfcomputer GmbH.

    HP and LaserJet are registered trademarks of Hewlett-Packard Corporation.

    Trademarks are used without warranting free usability.

  • ibg Headquarter:

    ibg Prfcomputer GmbHPretzfelder Strae 27D-91320 EbermannstadtTel: +49-9194-7384-0Fax: +49-9194-7384-10E-mail: [email protected]

    ibg International:

    ibg NDT Systems Corp.20793 Farmington RoadFarmington Hills, MI 48336Tel: +1-248-478 9490Fax: +1-248-478 9491E-mail: [email protected]

    ibg SWISS AGGalgenried 6CH-6370 StansTel: +41-41-612 2650Fax: +41-41-612 2651E-mail: [email protected]

    ibg UK Ltd.33 Park View RoadSutton ColdfieldGB-West Midlands B74 4PRTel/Fax: +44-121-352 1188E-mail: [email protected]

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 0-3

    Contents

    Exclusion of responsibility................................................................................ 0-1

    1. Introduction .................................................................................................... 1-11.1. What is an eddyvisor?..................................................................................... 1-11.2. The principle of eddy current structure testing .................................................. 1-21.3. Possibilities of eddy current structure test ......................................................... 1-3

    2. General information ....................................................................................... 2-12.1. Selection of structure test coils ......................................................................... 2-12.2. Creation of optimum test conditions .................................................................. 2-32.3. Wiring ............................................................................................................... 2-5

    2.3.1. Desktop instrument ................................................................................. 2-72.3.2. Switch panel version ............................................................................... 2-8

    2.4. Elements of the front panel ............................................................................... 2-92.4.1. Keys...................................................................................................... 2-102.4.2. Help function ......................................................................................... 2-112.4.3. Printer function...................................................................................... 2-112.4.4. Status bar.............................................................................................. 2-11

    2.5. Course of measurement ................................................................................. 2-12

    3. Short operating instructions ......................................................................... 3-13.1. Structure test .................................................................................................... 3-13.2. Crack detection................................................................................................. 3-5

    4. Operating modes and functions ................................................................... 4-14.1. Station and location bar .................................................................................... 4-14.2. Part type selection ............................................................................................ 4-14.3. Setup ................................................................................................................ 4-2

    4.3.1. Project & Language................................................................................. 4-24.3.1.1. Language.................................................................................... 4-2

    4.3.2. Test Assembly......................................................................................... 4-34.3.2.1. Station- and location bar ............................................................. 4-34.3.2.2. Station window............................................................................ 4-3

    4.3.3. Inputs and Outputs.................................................................................. 4-64.3.3.1. Selection of VCONI ..................................................................... 4-74.3.3.2. Monitor inputs, force outputs manually........................................ 4-74.3.3.3. Allow Signal Assignment ............................................................. 4-8

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 270-4 D-91320 Ebermannstadt Germany

    4.3.4. System Properties................................................................................. 4-104.3.4.1. Printer setting............................................................................ 4-104.3.4.2. Network setting of eddyvisor.................................................... 4-114.3.4.3. Date / Time ............................................................................... 4-124.3.4.4. Form ......................................................................................... 4-124.3.4.5. Network-Log.............................................................................. 4-124.3.4.6. Mouse pointer on/off ................................................................. 4-124.3.4.7. System monitor ......................................................................... 4-12

    4.3.5. Part Type Administration ....................................................................... 4-124.3.6. Frequencies & Locations (general)........................................................ 4-13

    4.3.6.1. Activate / deactivate location ..................................................... 4-144.3.6.2. Standard settings ...................................................................... 4-144.3.6.3. Discard changes ....................................................................... 4-144.3.6.4. Save settings............................................................................. 4-144.3.6.5. Save all settings ........................................................................ 4-14

    4.3.7. Frequencies and locations (structure test)............................................. 4-144.3.7.1. Tolerance zone shape (ellipse / rectangle) ................................ 4-144.3.7.2. Tolerance zone factor ............................................................... 4-154.3.7.3. Frequency ................................................................................. 4-154.3.7.4. Test time and settling delay....................................................... 4-164.3.7.5. AUTOSTART Configuration ...................................................... 4-16

    4.3.8. Frequencies and locations (crack detection) ......................................... 4-184.3.8.1. Band pass and tolerance zone .................................................. 4-184.3.8.2. Tolerance zone factor ............................................................... 4-194.3.8.3. Sender frequency...................................................................... 4-194.3.8.4. Tolerance zone factors commonly operated.............................. 4-194.3.8.5. Threshold .................................................................................. 4-194.3.8.6. Settling delay............................................................................. 4-194.3.8.7. End of test after .................................................................... 4-194.3.8.8. Lift-off compensation monitoring ............................................... 4-204.3.8.9. Monitoring of probe (noise threshold) ........................................ 4-204.3.8.10. Speed monitoring .................................................................... 4-204.3.8.11. Frequency range proposals..................................................... 4-214.3.8.12. Setup Lift-off compensation..................................................... 4-21

    4.4. Reference Data............................................................................................... 4-214.4.1. Data recording active ............................................................................ 4-224.4.2. Data recording stopped ......................................................................... 4-22

    4.4.2.1. Navigation in parts .................................................................... 4-224.4.2.2. Mark reference parts ................................................................. 4-234.4.2.3. Tolerance Zone Editor (structure test)....................................... 4-23

    4.5. PMFT - Preventive Multi-Frequency / Multi-Filter Test..................................... 4-244.5.1. Testing active........................................................................................ 4-244.5.2. Stop testing ........................................................................................... 4-24

    4.5.2.1. Navigation in parts .................................................................... 4-25

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 0-5

    4.5.2.2. Parts Counter............................................................................ 4-254.5.2.3. Screen off ................................................................................. 4-254.5.2.4. Add test part to Reference Data................................................ 4-264.5.2.5. Tolerance Zone Editor (structure test)....................................... 4-26

    4.6. Display options................................................................................................ 4-264.6.1. Structure test......................................................................................... 4-26

    4.6.1.1. Locus curve display................................................................... 4-264.6.1.2. Tolerance zone display - single ................................................. 4-264.6.1.3. Tolerance zone display - all....................................................... 4-274.6.1.4. Bargraph display ....................................................................... 4-27

    4.6.2. Crack detection ..................................................................................... 4-284.6.2.1. Outline bars............................................................................... 4-284.6.2.2. Tolerance zone display - single (xyt-display) ............................. 4-294.6.2.3. Tolerance zone display - all....................................................... 4-304.6.2.4. C-scan display........................................................................... 4-314.6.2.5. Bargraph display ....................................................................... 4-31

    4.6.3. Parts counter and history ...................................................................... 4-314.7. Key switch - authorisation level ....................................................................... 4-32

    5. Data recording via network with eddyLogger.............................................. 5-1

    6. Connection with automatic sort systems .................................................... 6-16.1. Automatic test, program course ........................................................................ 6-16.2. Time diagram.................................................................................................... 6-26.3. Electronic diagram of interfaces........................................................................ 6-4

    6.3.1. Proposal for Signal Assignment .............................................................. 6-66.4. Example Opto-Interface .................................................................................... 6-7

    6.4.1. Current supply for in and outputs ............................................................ 6-76.4.2. Interface for universal sorting device....................................................... 6-86.4.3. Test on consecutive locations controlled by PLC .................................... 6-96.4.4. Part type switching controlled by PLC ................................................... 6-10

    6.5. Coil connection VCONS (structure test) .......................................................... 6-11

    7. Part type switching via PLC (structure test) ................................................ 7-17.1. Part type switching controlled by PLC............................................................... 7-17.2. Part type request controlled by PLC.................................................................. 7-2

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 270-6 D-91320 Ebermannstadt Germany

    8. Technical data.................................................................................................8-1

    9. Glossary ..........................................................................................................9-1

    10. Appendix .....................................................................................................10-1

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 1-1

    1. Introduction

    We congratulate you on the decision to employ the eddyvisor in your factory. Withthis instrument you have acquired the most advanced eddy current testing deviceworking according to the PREVENTIVE MULTI-FREQUENCY TEST as well as thePREVENTIVE MULTI-FILTER TEST (PMFT). Successful implementation of a testpresupposes, in addition to a full command of the instrument settings, also a solidknowledge of the product to be tested, its material properties, of the errors likely tooccur and their underlying causes.

    The tester must be conversant with the limitations and restrictions in errordetectability as well as with the efficiency of the equipment used and with its properfunction. It is the proper functioning that needs to be checked prior to every test andperiodically during operation in order to warrant reliable test results. If theserequirements are not satisfied, the result will be of questionable value.

    Should you have any further questions or need our advice for an optimum solutionof your testing problem, please do not hesitate to contact us. Our after-sales serviceis ready to help you at any time.

    Chapter "3. Short operating instructions" from page 3-1 is mainly intended to savethe users who are not constantly working with the eddyvisor the troublesomestudy of a detailed user manual.

    As a new user of the eddyvisor you should simply try to perform a test with thisunit under the guidance of the "3. Short operating instructions" before going throughthe detailed user manual and the fine adjustment of all parameters. You will besurprised how quick you will learn to handle the eddyvisor.

    Please read these concise operating instructions carefully in order to avoidoperating errors. We have tried hard to keep the instructions as short as possible inan effort not to burden you with unnecessary information.

    1.1. What is an eddyvisor?

    The eddyvisorS is an eddy current test instrument for Structure test using PMFTPreventive Multi-Frequency Test. This enables testing e.g. for material mix,hardness, case depth, material composition as well as material texture ofconductive materials.

    The eddyvisorC is an eddy current test instrument for Crack detection usingPMFT Preventive Multi-Filter Test. This enables testing e.g. for surface-opencracks or disturbances of surface texture of conductive materials.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 271-2 D-91320 Ebermannstadt Germany

    The eddyvisorSC combines both test methods (Structure & Crack test) in oneinstrument.

    1.2. The principle of eddy current structure testing

    When alternating current is sent through a transmitter coil, a magnetic field buildsup at the coil. If a conductive test specimen approaches the current-carrying coil, aneddy current is induced in the test specimen. This phenomenon is schematicallyillustrated in fig. 1.1.

    um ussi

    u

    u

    2

    1

    wi

    Figure 1.1

    The magnitude of the eddy current depends on the chemical composition (alloy) andon the structural condition (e.g. hardened). By changing the eddy current frequency,an electromagnetic "thumbprint" of the structural state is obtained in the coil system.This can be picked up by the instrument through the integrated receiver coil andevaluated.

    An alteration in the structural state, in the alloy and the like attributable to a mix-upof material, a batch mix or faulty thermal treatment causes also the eddy currentsignal to change. The eddyvisorS testing device recognizes such changes andindicates deviations.

    The eddyvisorS operates according to the method of PREVENTIVE MULTI-FREQUENCY TESTING (PMFT) developed by ibg. The merit of PREVENTIVEMULTI-FREQUENCY TESTING is seen in the high probability of detecting evenunknown flaws on account of the broad frequency band selected.

    In contrast to other procedures, ibg uses low field strength values. This means thatonly low currents flow through the coils and that low permeabilities develop inferromagnetic materials (initial permeability).

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 1-3

    This method affords the following advantages:

    It operates at the higher penetration depth of the eddy current so that theinfluence exerted by the marginal layers is reduced.

    The reproducibility of the measurements is guaranteed because no magneticalterations develop in the structure of the specimen.

    Improved testing stability because no heat builds up in the specimen, theinstrument and in the coil.

    The salient properties of PREVENTIVE MULTI-FREQUENCY TESTING (PMFT) arethe following:

    Maximum testing reliability - even unknown flaws are identified without fail.

    Only good parts must be presented to the instrument.

    The generation of tolerance zones is extremely simple and made in a very shorttime.

    1.3. Possibilities of eddy current structure test

    Eddy current inspection can be applied to any type of material being conductiveeither electrically or magnetically. This holds good for all metallic materials.

    The eddyvisorS takes care of all the fields of application summarily mentioned intable 1.1.

    Type of parts Parameters to be tested for

    Forged piecesBearing componentsVA-tubesScrewsBrake discsPower transmission elementsComponents of belt locksLinear guidesBalls

    heat treatment and material mix-uphardness and material mix-upmaterial mix-upheat treatmentchill depthpurity and core strengthhardness and texturepurity and grinding burnmaterial mix-up

    Table 1.1

    Please note that the eddy current inspection procedure is not an absolutemeasuring method but rather a comparative approach. The hardness of parts canbe measured with the eddyvisorS inspection device to within 2-3 HRC withoutany problem. However, as is the case with all eddy current tests, the inspection is

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 271-4 D-91320 Ebermannstadt Germany

    based on YES/NO answers. The eddy current testing instrument does recognizedeviations from existing tolerance zones, but cannot reveal the causes responsiblefor them. In order to obtain precise information on why the specimens wereclassified as not being in order, yet other testing procedures such as hardness testand metallographic studies must be employed to confirm the findings of the eddycurrent device.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 2-1

    2. General information

    The eddyvisor eddy current inspection device is multiprocessor-controlled. Theentire control software is configured hierarchically, so the selectable submenusautomatically result from each main menu. The eddyvisor is operated via a so-called soft key control system which enables the user to handle the device with arestricted number of keys.

    These operating instructions have been arranged so that each step of operation isexplained in chronological order. All you need to do is to execute the instructionsstep by step as indicated. While proceeding in this manner, you are guided throughthe complete software.

    Hint

    The [?]-is auxiliaryand serves to indicate a context-related help at any time.

    This means that the help needed is offered at any point of thesoftware.

    The key combinations to be actuated by the user are written inbold capital letters throughout the User Manual.

    Should you nevertheless experience any problem with operation,please contact us under the following telephone number:

    ibg Prfcomputer GmbH, Tel. +49-9194-7384-0

    2.1. Selection of structure test coils

    Selection of the proper test coils and probes is of decisive importance for the testresult. In addition to a set of standard coils and probes, special coils adapted to thejob are available. The selection of suitable coils determine the type of job and theshape of the specimen. Here the following criteria should be considered:

    What is the job like?e.g. measurement of the case depth of shafts

    Which coil should be used?The suitable coil or probe can be taken from table 2.1.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 272-2 D-91320 Ebermannstadt Germany

    Field of application Frequency range

    Prob

    e

    Enci

    rclin

    gco

    il

    Enci

    rclin

    gco

    ilsh

    ield

    ed

    Rec

    tang

    ular

    coil

    Flat

    coil

    Material mix:

    ferritic materialaustenitic materialnon-ferrous metalshard metals

    25 Hz - 100 kHz100 Hz - 100 kHz100 Hz - 100 kHz

    50 Hz - 100 kHz

    +++++++

    ++++

    ++

    +

    +

    ++++

    ++

    +++++++

    Heat treatment:

    temperingsurface hardnesscase depth of shaftscase depth (carburized)carburizationdecarburization

    25 Hz - 25 kHz100 Hz - 25 kHz

    5 Hz - 400 Hz5 Hz - 4 kHz5 Hz - 25 kHz

    10 kHz - 100 kHz

    +++

    +

    ++

    ++++++

    +++

    ++++

    +

    ++

    ++++++

    +++

    +++

    +

    ++

    Cast iron:

    nodular cast ironledeburite (spherocast)

    5 Hz - 5 kHz25 Hz - 25 kHz

    +++

    +++

    ++

    +++

    +++

    Surface:

    grinding abuse 5 kHz - 100 kHz ++ ++

    ++ very good + good possible not possible

    Table 2.1

    Our standard coils are classified by frequency ranges. Whereas the frequencyrange of the encircling coils is directly indicated on the name plate, the probes arecoded as follows:

    1 ring 25 Hz to 10 kHz2 rings 250 Hz to 50 kHz3 rings 5 kHz to 250 kHz

    As far as encircling coils are concerned, a filling degree of less than 60 % shouldnot be selected as otherwise the measuring signals become too small and thedifferences, say, in the case depth are no longer recognized distinctly. When usingencircling coils, a second coil is required for compensation. This coil must feature a

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 2-3

    firmly (!!!) mounted compensation part. Removing or shifting the compensation partleads to wrong test results. The compensation part has no comparing functiondesigned to distinguish the OK-parts from the NotGood-parts.

    The basic design of the standard encircling coils is displayed in Fig. 2.1. Each coilcontains a transmitter and receiver winding. The receiver winding should always bearranged as close as possible to the area to be measured. It is very easy to find outon what side the receiver or transmitter windings are located. Just take an encirclingcoil and read the name plate. If you can read this plate and when the coil plug is onthe left of you, the receiver winding is below and the transmitter winding is above inthe coil.

    25Hz-25kHz

    P/NS/NDAT

    ibg

    Section A-A

    Coil plug

    Receiver winding

    Transmitter winding Frequency range

    A AA

    Figure 2.1

    For coil protection wearing bushes of abrasion-resistant plastic material e.g."Murtfeldt S green" should be used.

    2.2. Creation of optimum test conditions

    Be sure to pay attention to the following points when carrying out a testing job.Carefully read the following hints. Disregarding them is likely to lead to faultymeasurements. The eddy current procedure is no absolute method.

    The following parameters may cause the results to scatter:

    Inhomogeneities in material

    Dimensional tolerances in the specimen

    Dimensional tolerances in coil or probe positioning

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 272-4 D-91320 Ebermannstadt Germany

    Variations in temperature (structure test)

    For optimum results please pay attention to the following points:

    Make sure that the parts used for Reference Data really comply with the desiredspecifications.

    Show at least 5 OK parts to the instrument as Reference Dataat least 10 OK parts when structure testing.

    Select reasonable tolerance sizes.To avoid the risk of undesirable pseudo rejects.

    Do random checks during testing with OK and NG parts.

    Lock the instrument with the keyswitch to avoid unwanted access.

    Read the operating manual before initial operation.

    Provide constant and reproducible test conditions.

    Avoid interferences near the coil resp. probe such as current impulses, fluxsetups active during the measurement, electric welding and other magnetic orelectric interference fields.

    In addition for structure test:

    Different batches should be tested separately.The best results are obtained if for each material batch an own part type (withown reference data) is used.

    Please see to it that the test parts have constant test temperature.

    Please see to it that no metallic parts, which are not marked for testing, arelocated near the test coils.

    Carry out the test on a wooden table, if possible. Even steel girders underneath awooden board are likely to affect the test results.

    The following factors may adversely affect the measurement directly:

    Positioning of the specimen in the coil resp. in relation to the probe (distance).

    Dimensional tolerances of the specimens.

    Structure test: variation in batches or differences in temperature betweencalibration part and test part.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 2-5

    Use of coils resp. probes with a wrong frequency range.

    Lift-off effect due to dust and dirt.

    Edge effect caused by measuring near edge of part

    2.3. Wiring

    Prior to putting the eddyvisor into operation, all cables must be connected and allconnections to be checked.

    WARNING

    Please take care that all cables are plugged and screwed tightbefore the power switch is turned on.

    The instrument must be grounded properly (PE connector of themains plug or grounding screw on the instrument housing, seedrawing of the back side of the instrument below). Otherwise,voltage may constitute on the housing. If the instrument is not

    grounded properly, the safety standard protective ground are notobeyed. Furthermore, quality of testing is impaired so that

    optimum test results cannot be achieved.

    Also ensure that all plugs are firmly connected and screwed tight.

    If plugs are inserted after connection of the eddyvisor thevarious components are likely to be damaged!

    The eddyvisor consists of the operating unit eddyvisorHMI and the measuringunit for structure resp. crack test eddyvisorMSC (refer to "9. Glossary"). Both unitscommunicate via USB interface.

    Both units are connected to one unit as eddyvisor desktop instrument. Powersupply is effected via the measuring unit. USB coupling of the single parts is doneinternally without using the USB jack of the measuring unit.

    Operating and measuring unit are supplied separately when using the eddyvisor

    in a test system. Both units have to be supplied with line voltage. USB coupling ofthe single parts is done via one of the protected lower USB jacks of the operatingunit and the USB jack of the measuring unit.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 272-6 D-91320 Ebermannstadt Germany

    OS-CompactFlash(CF)

    Ethernet-jack

    mains plugfine-wire fuse T 2.0A

    USB-jacks(refer to text)

    Figure 2.2

    Figure 2.2 shows the bottom side of the operating unit eddyvisorHMI. An ibg CF-card has to be used in the CF-slot. The lower one of the two USB jacks is used forconnection of measuring unit of the eddyvisor. The Ethernet jack enablesconnection to a network printer, a company network resp. a PC/laptop.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 2-7

    2.3.1. Desktop instrument

    type plate mains plus and fine-wire fuse T 2.0A

    coil connection boardVCONS2 (structure test)

    I/O-boardVCONI

    probe connection boardVCONC2 (crack test)

    slot-no.

    Figure 2.3

    Figure 2.3 shows the connector side of the measuring unit. It is an eddyvisorDSCfor structure test with max. 2 coils and crack test with max. 2 probes. Slots 1 - 4include exchangeable boards.

    Slot 1 includes a coil connection board VCONS2 (for structure test) for use withmax. two coils - A1 and A2. Up to 8 coil connections per slot are possible(VCONS8).

    Slot 2 includes a probe connection board VCONC2 (for crack detection) for use withmax. two probes - A1 and A2. Up to 4 probe connections per slot are possible(VCONC4). Each probe connection board VCONCx requires internally a channelboard VKAN.

    Slot 3 includes an I/O-board VCONI for connection of 16 in- and 16 outputs 24 Vdcper D-sub plug. Power supply of the I/O jacks is effected either via D-sub plugs or

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 272-8 D-91320 Ebermannstadt Germany

    via bridges at the 24V power supply plug C. I/O-boards VCONI may be used onlyeither in slot 3 or in slot 6 or in slot 7.

    Slot 4 includes a mains supply board. Plug A shows the mains supply of themeasuring unit. Power supply of the operating unit of a desktop instrument iseffected internally. Plug B shows the USB jack of the measuring unit forconnection with a separate operating unit. The USB jack of the measuring unit ofthe desktop eddyvisor is not wired.

    At desktop eddyvisor, only slots 1 - 4 of the measuring unit are accessible.

    2.3.2. Switch panel version

    At test systems, eddyvisor with separate operating and measuring unit, all slots (1- 7) of the measuring unit are accessible. Slots 1 - 4 are at the outside of the switchcabinet, slots 5 - 7 are inside.

    Slots 1 - 4 includes coil connection boards VCONSx and crack detection probeboards VCONCx and point to the test mechanic. Slot 5 is equipped with the mainssupply board and slots 6 - 7 with I/O-boards and point to the PLC.

    Important note!The eddyvisor must be properly earthed: PE-connector of the

    mains plug or earthing screw on the housing of the measuring unitin test systems. Otherwise, voltage may constitute on the housing.

    If the instrument is not earthed properly, the safety standardprotective ground are not obeyed. Furthermore, quality of testing

    is impaired so that optimum test results cannot be achieved.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 2-9

    2.4. Elements of the front panel

    The eddyvisor is operated on the front panel of the operating unit. Figure 2.4shows the front panel.

    key switch

    (b)

    (a) (c)

    ON-function lamp 2x USB jack touch panel

    Figure 2.4

    Touch panelThe touch panel displays information and serves as key keyboard. The menusare explained in the following chapter.

    USB-jacksThere are two USB jacks behind a sliding cover in order to save or recallsettings.

    ON function lampA blue LED is on when the instrument is on.

    Key switchThe key switch has 3 positions. The key may be taken out at all 3 positions.

    (a) The instrument is switched off (separated from the power line).

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 272-10 D-91320 Ebermannstadt Germany

    (b) The instrument is switched on. All functions are accessible.

    (c) The instrument is switched on. Restricted access to functions of theinstrument. Use: No unauthorised access at running test possible.

    When turning the key to position (c) a window appears. Here the rangeof access restrictions may be chosen.

    2.4.1. Keys

    The operation of the eddyvisor is carried out via the displayed keys on the touchscreen. The meaning of the keys is explained by the screen text. Further help ispossible by pressing the [?]-key and then the key to be explained.

    At the left side of the touch screen, in many operating modes, stations and locationsare displayed resp. selected.

    At the upper edge of the touch screen, the main navigation bar is displayed. Thecurrent operating condition can be seen by the pressed key.

    At the right side of the touch screen, different function keys are displayeddepending on operating mode.

    Figure 2.5 shows (as example) the help for the reference data menu.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 2-11

    printer help

    date, timetest-infodemo-modeserial numberev software version

    help text

    status bar

    Activate Ref.-Data Recording /Data Recording activeHere, a station is set to condition reference data recor-ding.The first activated position of a station is nowsensitive for incoming start signals, as soon as it isstarted all other activated positions are sensitive forstart signals as well.When pressing the key "Data Recording active" isdisplayed within the key.When "station states are operated together" is set inSETUP - test assembly all stations are activated.

    Stop Ref.-Data Recording /Data Recording stoppedAfter the reference data for the parts are recorded

    the reference data recording for the station may be

    Figure 2.5

    2.4.2. Help function

    The eddyvisor provides a context-relating help. By pressing the [?]-key and thenthe key to be explained help is called. Figure 2.5 shows the help for reference datamenu.

    2.4.3. Printer function

    The eddyvisor enables documentation by using screen prints. By pressing theprinter-key the menu for printing of the current screen is called. Printing to a printer(set in "Setup" / "System settings") is possible, or a screenshot can be stored to amemory stick, storable in different formats.

    2.4.4. Status bar

    The status bar is at the lower edge of the screen. The following information aredisplayed:

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 272-12 D-91320 Ebermannstadt Germany

    EV XX.XXXSoftware version of eddyvisor

    S/N: XXXXXSerial number of the eddyvisor

    in DEMO-modeThe word "DEMO" is red-blinking. All functions can be used in demo mode. Anymodifications of settings, however, are not saved. Any changes get lost whenswitching off the instrument.

    "Neg: xx Id: xx Err: xx"Debug information for ibg.

    "Authorisation: ..." in horizontal, right-hand keyswitch position:The current authorisation level e.g. "authorisation: basic position" is displayed.

    Activity display (structure test at reference data recording or testing):Display of the current measuring e.g. "Calibration" for automatic calibration;"S1/P2 Settling Delay", "S1/P2 Testing" for station 1 / location 2 in settlingdelay resp. measuring.

    Date and time in format year-month-day hour:minute:second.

    2.5. Course of measurement

    The eddyvisorS operates on the ibg-developed principle of PREVENTIVEMULTI-FREQUENCY TESTING (PMFT). After the measurement has beentriggered, the test frequencies are applied successively to the coil system, startingat the lowest test frequency. At each test frequency the response of the specimento the coil system is recorded and the electric voltage vector is broken down into itsX and Y components and stored.

    Upon completion of the test, the values stored are, depending on the operatingmode, either compared with the tolerance zones (mode SORTING) or assigned tothe material data (mode RECORDING O.K. PARTS).

    STARTx (start pulse)

    settling delay

    BUSY-S(structure test runs)OKx (sort output Gut)

    ENDx (end of measuring)

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 2-13

    Measurement starts with the "START-Signal":

    Measurements are started with the START-signal and need a certain time (BUSY-S-signal). A measurement can only be triggered in mode REFERENCE DATA orPREVENTIVE MULTI-FREQUENCY TEST.

    Settling delay:

    After starting a test, but before the test proper begins, the operator may select anadjustable settling delay. This delay allows the specimen to "settle inside the testcoil", so the test will only be started as soon as the part has taken up its finalposition in the coil.

    The "BUSY-S-signal":

    If the test unit functions in one of the two operating modes REFERENCE DATA orPREVENTIVE MULTI-FREQUENCY TEST the BUSY-S-signal (BUSY, structuretest) will appear immediately following the START-signal. The settling delay startsrunning and upon expiration the test proper commences at the first frequency.

    If the start of the following location is set to AFTER the settling delay of this location is immediately started, the BUSY-S-signal remains active.

    If it concerns, however, the last channel within a test block the BUSY-S-signaldisappears. Disappearing of the BUSY-S-signal means that the eddyvisorS isready to start a new measurement.

    The sorting decision with the "OK-i-signal":

    After testing a location at all frequencies the sorting decision is made andtransmitted to the sort outputs (OK-i-signal, NOK-i-signal). In mode REFERENCEDATA no sort signal appears.

    The sum sorting decision with the "OK-S-" and "NOK-S-signal":

    After testing all locations of one station the sorting decision for the station is madeand transmitted to the sort output (OK-S-signal and NOK-S-signal). For this, theformation of the sum signal OK-S must be set. In mode REFERENCE DATA no sortsignal appears.

    The END-i-signal:

    After testing a location at all test frequencies the allocated END-i-signal appears.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 272-14 D-91320 Ebermannstadt Germany

    The sum signal END-S:

    After testing all locations of one station the sum signal for the station is made andtransmitted to e.g. the output END-S. For this, the formation of the sum signal END-S must be set.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 3-1

    3. Short operating instructions

    3.1. Structure test

    These concise operating instructions are intended to show you, that theeddyvisorS enables you to obtain test results in a short time, whenever projectsare available. A standard project 'screw test' (DEMO mode) is used here. It isadjustable in the menu SETUP, PROJECT & LANGUAGE.

    Below are the sequence of steps to be executed:

    (a) The probe resp. the coils are directly connected to the measuring unit ofthe eddyvisorS, coil board VCONS in slot 1 and jack A. Usingencircling coils: one part is put into the compensation coil(s). Using self-compensated coils or probes: no compensation part needed.

    (b) If operating and measuring unit are separate they have to be connectedvia a USB-cable.

    (c) Connect eddyvisorS to mains supply. If operating and measuring unitare separate both units must be supplied with mains voltage.

    (d) Prepare at least 10 (better some more) OK parts (as reference).

    (e) The characteristic eddy current data are recorded by "showing" the knownOK-parts. The eddyvisorS automatically forms tolerance zones.

    (f) The specimens are tested by comparing their eddy current characteristicswith the tolerance zones formed by the material data and sorting based onGood (OK) and NotGood (NOK) criteria.

    Now proceed step by step as indicated in the short operating instructions.

    The standard configuration of the eddyvisorS includes already a testconfiguration. This standard configuration is accessible via menu SETUP,PROJECT & LANGUAGE, STANDARD PROJECT ... (DEMO MODE). Any changesto this demo project will get lost when switching off the eddyvisorS or leaving thedemo mode.

    Switch on the instrument with the key switch, key in vertical (middle) position.

    Select the standard project 'screw test' (DEMO mode) in the menu SETUP,PROJECT & LANGUAGE.

    Then select the main menu REFERENCE DATA.

    The eddyvisorS now displays the MAIN MENU and has loaded a projectavailable in the internal memory. In the upper left-hand corner, the currently active

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 273-2 D-91320 Ebermannstadt Germany

    part type can be seen, here type 001, generated on 2005-06-01 (year-month-day),name "type", "no. 01". REFERENCE DATA is selected in the main menu.

    main menupart typeselection active menu, condition

    sub menu

    station,state

    Figure 3.1

    press REFERENCE DATA (if necessary)The mode REFERENCE DATA becomes active. In this mode, OK parts are readin and tolerance zones are formed. A tolerance zone with REFERENCE DATARECORDING STOPPED appears and on the right-hand side the relating menu.

    press ACTIVATE DATA RECORDINGThe mode changes to REFERENCE DATA RECORDING RUNS. The stationkeys get a small yellow arrow symbol which means the release for referenceparts recording.

    Put one OK part resp. the coil to test position and press the START-keyon the eddyvisorS. Measuring values via the set frequency arerecorded for the part and displayed in the tolerance zone.

    Change the OK parts and press the START-key on the eddyvisorS.Further measuring values are recorded. The value of the latest referencepart is specially displayed. At the same time tolerance zones are formedrelating to the measuring values. These tolerance zones are used for thesucceeding test for comparison.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 3-3

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 273-4 D-91320 Ebermannstadt Germany

    PREVENTIVE MULTI-FREQUENCY TEST

    latest recorded parts

    tolerance fieldsstation,release

    displayoptions

    history

    location

    Figure 3.2

    Press PREVENTIVE MULTI-FREQUENCY TEST in the main menu. Themenu PREVENTIVE MULTI-FREQUENCY TEST becomes active.Several small tolerance zones appear one below the other and the modePMFTesting active is displayed. The station number has a small greenarrow symbol which means the release for parts testing.

    Put the part to be tested resp. the coil into test position and press theSTART-key on the eddyvisorS. If the part is good (OK) a green OKsymbol goes on. If the part is bad (NOK) a red NOK symbol goes on. Atthe same time bargraphs are displayed for each measuring frequencymirroring the relative position of the measuring values regarding thetolerance zone. The position of the measuring value is indicated in thetolerance zones.

    Change the part to be tested and again press the START-key on theeddyvisorS. The maximum deviations of the latest recorded partsappear as bargraphs in the lower part of the screen besides the smallhistory of the test results.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 3-5

    The display options are bargraph diagram, local material curve andseveral tolerance zones. When displaying the results of a singlefrequency the bargraph changes to a display of values of this frequency.NOK parts are marked in red.

    As per this course all parts to be tested are verified at all locations.

    PREVENTIVE MULTI-FREQUENCY TEST, press STOP TESTING.The mode changes to STOP TESTING. The station numbers get a smallblack double-bargraph symbol which means lock regarding parts testing.The submenu changes and two yellow arrow keys appear besides theparts statistic.

    Pressing "Reset parts counter" opens a window with a safety request ifparts counter and sorting data are really to be deleted. In doubtful casesay no.

    Between the latest recorded parts may be changed by pressing the yellowarrow keys. For a quicker change the selection may be limited to OKparts and NOK parts. The selection is done via a switch besides thestatistic.

    Pressing "Add tested part to ref data" enables adding of a part to thereference data. A safety request avoids adding of parts by mistake. Partswhich have been added for trial purposes may be marked as invalid anddeleted in REFERENCE DATA menu.

    Press the PRINTER symbol on the eddyvisorS in order to print a testprotocol. An inquiry window appears. You can select between printing thescreen as picture file in a certain format or printing the protocol on aprinter. The just running sort test is not influenced by the printing process.

    3.2. Crack detection

    These concise operating instructions are intended to show you, that theeddyvisorC enables you to obtain test results in a short time. Starting point is thefactory setting. For this, "Standard settings" in menu "Setup" / "Project & Language"are called.

    Below are the sequence of steps to be executed:

    (a) The crack detection probe is connected with the measuring unit of theeddyvisorC, to the probe connection board VCONC in slot 1, jack A1.

    (b) If operating and measuring unit are separate they have to be connectedvia a USB-cable.

    (c) Connect eddyvisorC to mains supply. If operating and measuring unitare separate both units must be supplied with mains voltage.

    (d) Prepare at least 5 (better some more) OK parts (as reference).

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 273-6 D-91320 Ebermannstadt Germany

    (e) Recording of characteristic noise of the surface (as eddy current data) by"showing" of these OK parts. The eddyvisorC automatically generatestolerance zones.

    (f) The specimens are tested by comparing their eddy current characteristicswith the tolerance zones and sorting based on Good (OK) and NotGood(NOK) criteria.

    Now proceed step by step as indicated in the short operating instructions.

    The standard configuration of the eddyvisorC includes already a testconfiguration. This standard configuration is accessible via menu SETUP,PROJECT & LANGUAGE, STANDARD SETTING.

    Switch on the instrument with the key switch, key in vertical (middle) position.

    Select the "standard setting" in the menu SETUP, PROJECT & LANGUAGE.

    Then select the main menu REFERENCE DATA.

    The eddyvisorC now displays the MAIN MENU and has loaded the standardsetting. In the upper left-hand corner, the currently active part type can be seen,here type 001, generated on 2007-05-09 (year-month-day), name "type", "no. 01".REFERENCE DATA is selected in the main menu.

    main menupart type functionkeys

    station,state

    Figure 3.3

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 3-7

    press REFERENCE DATA (if necessary)The mode REFERENCE DATA becomes active. In this mode, OK parts are readin and tolerance zones are formed. The minimum tolerance zone of one of 30band passes is shown. The station condition "stopped" is visible on the left-handside in the station key and on the right-hand side at condition of function key.

    press ACTIVATE DATA RECORDINGThe mode changes to REFERENCE DATA RECORDING RUNS. The stationkeys get a small yellow arrow symbol which means the release for referenceparts recording.

    Put one OK part resp. the probe to test position and start rotation of partresp. probe. Press the START-key on the eddyvisorC. Now the noisesignals of the surface are recorded and displayed live. If testing includesa surface scan do the advance movement (the scan) now. It is normalthat the signals of the first reference part are displayed very large asdisplay is done with the largest amplification.

    After the scan press the STOP-key on the eddyvisorC. A "polar field"is automatically formed around the signals taken between START andSTOP and displayed in orange. This happens in the xy-display for each ofthe 30 band passes. The green tolerance zones are calculated of thepolar fields of all reference data plus the tolerance zone factors and aredisplayed.

    Now change OK part and do the same procedure; start rotation, pressSTART on eddyvisorC, do scan, press STOP on eddyvisorC andstop rotation.So the measuring values for further reference parts are taken. Themeasuring value of the latest reference part are displayed. At the sametime, tolerance zones to the measuring values are formed. Thesetolerance zones are used for the following test for comparison.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 273-8 D-91320 Ebermannstadt Germany

    PREVENTIVE MULTI-FILTER TEST

    latest recorded parts (history)

    tolerance fieldsstation,release

    displayoptions

    parts counter

    location

    Figure 3.4

    press Preventive Multi-Filter Test in main menuThe menu Preventive Multi-Filter Test becomes active. Several small tolerancezones appear one under the other and the station condition changes to "testactive". The station numbers get a small green arrow symbol indicating releasefor parts testing.

    Put the test part resp. the probe in test position, start rotation, press the START-key on the eddyvisorC, scan, press the STOP-key on the eddyvisorC.If the part is good (OK) a green symbol goes on. If the part is bad (NG) a redsymbol goes on. At the same time, bars for each band pass are faded in whichshow the size of the signal in relation to the tolerance zone.

    Change the part and do the same procedure (rotation, start, scan, stop) for eachpart. The maximum deviation of the latest recorded part appears as barsequence in the lower part of the screen as small history to the test results.

    Pressing the display options shifts between bargraph display, xyt-display (withpolar- and tolerance zone) and C-scan-display. Displaying the results of a singleband pass, the history does only show this one band pass, too.The red/green indication of result left ahead always refers to the sorting result ofthe complete station, i.e. the result of all bad passes.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 3-9

    As per this procedure all parts to be tested are checked at all locations.

    press Stop testingThe station condition changes to testing stopped. The station numbers get asmall black double-bargraph symbol indicating lock regarding parts testing. Thefunction keys change and two yellow arrow keys appear beside the partsstatistic.

    Pressing "reset parts counter" opens a window with a safety request if the partscounter and the sorting data are really to be deleted. In doubtful case say no.

    The yellow arrow keys enable switching between the latest recorded parts. Forquicker change, the selection may be limited to OK or NG parts only. Selection isdone via a key beside the statistic.

    Pressing "Add test part to RefData" allows adding of a part to the reference data.A safety request avoids erroneous adding of parts. In reference data menu,added parts can be marked as invalid and removed for test purpose.

    Press printer symbol on eddyvisorCTo print a test protocol when a printer is connected. An inquiry window appears.You may decide whether the current screen is printed as picture file in a certainformat or whether a protocol is printed. The running sorting is not influenced byprinting.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 4-1

    4. Operating modes and functions

    The main menu of the eddyvisor includes:

    SetupProject settings of the eddyvisor

    Reference dataRecording of reference parts to generate tolerance zones

    Preventive Multi-Frequency / Multi-Filter TestTesting of parts and sorting to OK and NG

    4.1. Station and location bar

    A station bar on the left-hand side of the screen allows to select single or allstations. The single station keys get a coloured icon to mark the current condition ofthe station:

    two black rectangles on the lower left-hand side:station stopped.

    yellow triangle on the upper left-hand side:station in reference data recording

    green triangle on the upper right-hand side:station in Preventive Multi-Frequency Test

    A location bar is positioned beside the station bar. Depending on the selectedstation the corresponding locations are displayed. Single or all locations areselectable via the locations keys.

    If a station / location and bargraph display are selected, a list of test frequencieswith their letters A - H appears beside the location bar.

    4.2. Part type selection

    The key on the upper left-hand side of the touch screen shows the current part type.

    Press this key and a menu "Overview of saved Part Types" appears. A list ofavailable part types is shown. The list may be sorted according to number of part,caption 1, caption 2 or date by pressing the key in the headline. Repeatedlypressing the key inverts sorting sequence.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 274-2 D-91320 Ebermannstadt Germany

    Part types are selectable via the arrow keys on the left-hand side of the list. Aselected part type is loaded into the current resp. all stations by pressing Loadonly in this Station and "Load in all stations" and the menu is left. The key"Cancel" leaves the menu without any changing.

    For further information see chapter SETUP - Part Type.

    4.3. Setup

    All settings of the eddyvisor are done in menu SETUP. It is subdivided into thefollowing points:

    Project & LanguageSetting of language, storage and restorage of settings of the instrument to resp.from the USB storage medium (USB-stick) as projects.

    Test assemblyTest assembly describes the stations with their locations and the relating coilsand logical signals.

    In- and outputsAllocation of logical signals and physical in- and outputs.

    System settingsBasic settings like printer, integration of eddyvisor into a network anddate/time.

    Part typeCreation of individually named part types for use with test assembly.

    Frequencies & LocationsSetting of band passes, settling delay and test times and versatile monitoringfunctions. If used, also setting of lift-off compensation.

    4.3.1. Project & Language

    Menu of eddyvisor to select language and to load and store the current project.

    4.3.1.1. Language

    The language of the user interface is selected. Pressing a key causes immediateswitching of language.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 4-3

    4.3.2. Test Assembly

    Test Assembly describes the stations with their locations and the relating coils andlogical signals.

    Pressing the key "Default Settings" sets test assembly to standard settings. Thistest assembly includes a station with one location.

    Modifications of the complete test assembly are taken by key "Save all stations" orthe latest conditions are restored by key "Undo". If menu is changed a safetyrequest asks if to use the modified test assembly resp. to undo or to cancel menuchange.

    4.3.2.1. Station- and location bar

    Stations and locations are selected, added or deleted via the station and locationbar on the left-hand side of the screen.

    Pressing the key "Add Structure STATION" completes the station bar by astructure test station. Pressing the key "Add Crack STATION" completes thestation bar by a crack test station. 8 stations are the maximum.

    A selected station is deleted by pressing the key "Delete STATION". The remainingstations are numbered anew.

    Locations may be added or deleted to resp. from a station by pressing the key "AddLOCATION" or "Delete LOCATION". Added locations get automatically the basicsettings. Maximum 32 locations are totally possible for all stations.

    4.3.2.2. Station window

    The test assembly of the selected station is shown in the station window. The nameof the station is shown in the upper left-hand corner.

    The key "Station states will be operated together / individually" determines if allstations switch between the modes reference data recording running resp....stopped, Preventive Multi-Frequency Test running resp. ...stopped at thesame time or individually.

    In the upper part of the station window each location has a bar with parameters. Ifthere are more locations than can be displayed a scroll bar is shown. Below thereare sum information on the station and the amount of locations.

    Coil resp. probe

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 274-4 D-91320 Ebermannstadt Germany

    In general, a pair of coils is used for structure testing, one test coil and the relatingcompensation coil.

    Crack detection uses probe w/o or with lift-off compensation.

    The pair of coils / the probe is connected to the VCONS / VCONC. Connection plugis shown in format:

    VCONS-slot (1 - 4), connector VCONS (A or B), pair of coil (1 or 2)

    Example: coil 1A2 means VCONS in slot 1, connector A, pair of coil 1.

    The coil is set in the same way for all locations of a station in the location barindividually or via the key "coil". When selecting a coil, an overview appearsshowing all possible coil connections. Touching same key again closes theoverview. Touching a marked coil proposal in the overview takes it.

    Coil monitoring is set either "ACTIVE" or "not active". If coil monitoring is activethe test coils are monitored for short-circuit or break. In case of any failure testing isstopped and a failure message appears on the screen.

    Start Signal

    Start of reference data recording or Preventive Multi-Frequency Test is triggered bya logical start signal. The logical start signal of a location knows different startmodes:

    START 1..32Triggering of start via rising edge at one input of a VCONI-board. It isrecommended to use START-i signals of standard pin assignment only.

    Conti (structure test only)Continuous triggering - the start signal is triggered automatically in shortintervals (approx. 20 msec.). Application: triggering of the first test of a stationwhich permanently tests.

    AFTER, symbol (structure test only) Automatic start trigger after testing of the previous location. Application:triggering of the succeeding measuring in a coil head with several locations.

    AutoStart (structure test only)Automatic triggering when a test part approaches the test coil. AutoStartfunction has to be set individually in menu "Frequencies & Locations.

    External, Ext. (structure test only)Use of an external START-key which is connected to the coil plug e.g. structuretest probes with integrated START-key.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 4-5

    Manual, short Manu (structure test only)Triggering of start only by START-key on the eddyvisor. The START-key ofthe eddyvisor may be used always for alternative triggering of a test unlesslocked by the key switch.

    Method of start is set individually in the location bar or via the superset Start-keylogical Input for location 2 or higher. When selected an overview of the methodof start appears, pressing again closes the overview. Pressing a method of starttakes it.

    Sort signal

    During Preventive Multi-Frequency Test an OK-i signal (OK-1..32 resp. OK-S) andNOK-i signal (NOK-1..32 resp. NOK-S) is created after the sorting decision of oneor all locations. The sum signal "Result LOCATION 1-" indicates that all locationsare OK.

    The logical signals are assigned to a pin in menu "In and outputs". Description is asfollows:

    VCONI-slot (3,6,7), connector on VCONI (A or B), pin (1..37)

    Example: 3A11 means VCONI in slot 3, connector A, pin 11.

    If several outputs are assigned to one signal only one pin is given. It isrecommended to use signals of the standard pin assignment only.

    The signal used is set for all locations individually in the location bar or via thesuperset key sort signal "logical Output". The sum signal is determined in the bar"result LOCATION 1-". When selected an overview of all possible sort signalsappears. Touching again closes the overview. Pressing a sort signal in the overviewtakes it. The symbol "---" removes the assignment.

    The time of sort signals is set individually for each location or via the superset keysort signal "Time" for all locations the same. Duration of sum signal is set in thebar "result LOCATION 1-". When selecting a menu appears with the followingpossibilities:

    Arrow keys and allow to set a duration between 10 ms and 1,000 msec.

    SYNC_0the sort signals disappear when start trigger (the start signal) disappears orwhen the station is stopped.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 274-6 D-91320 Ebermannstadt Germany

    SYNC_1the sort signals disappear when the next start trigger (the next start signal)arrives or when stopping the station.

    SYNC_2the sort signals disappear just before the next sorting decision or when stoppingthe station. Thus a maximum duration of sort impulse length is achieved.

    End Signal

    After measuring of one resp. all locations, an END-i signal (END-1..32 resp. END-S)may be created. The signal used is set individually in the location bar or via thesuperset key end signal "logical Output" for all locations the same. Whenselecting an overview appears of all possible end signal. Touching again closes theoverview. Pressing an end signal in the overview takes it. The symbol "---" removesthe assignment.

    The time of end signal is set individually for each location or via the superset keyend signal "Time" for all locations the same. The duration of sum signal is set inthe bar "result LOCATION 1-". When selecting a small menu appears to set:

    arrow keys and allow to set a duration between 10 ms and 1,000 msec.

    Monitor sorting gate is set either ACTIVE or not active. If active, then:

    - connect a sensor at input GatePos-i which is allowed to give a +24V-signalonly when sorting gate is in NOK position.There is a direct assignment between the sum signals END-i from resultLOCATION 1-... and GatePos-i: At END_1, GatePos_1 is used, at END_2,GatePos_2 follows a.s.o.

    - the sum sort signal OK-i from result LOCATION 1-... is supplied to control asorting gate.

    - the sum end signal END-i from result LOCATION 1-... is supplied to control astopper and in case of NOK-parts delayed as long as the input "GatePos-i"gives a +24V-signal. If delay is more than 1 second a failure message isgenerated and the station stopped.

    If monitor sorting gate is set to "not active" monitoring is off.

    4.3.3. Inputs and Outputs

    The assignment of internal logical signals of the eddyvisor (Start-1, GatePos-S,etc.) to the physical inputs and outputs of the VCONI takes place.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 4-7

    It is set in test assembly which logical signals are really used. In addition, there aresignals which are always present (e.g. StatActive-1).

    It is recommended to work with the standard assignment and to do individualsettings only as an exception.

    The inputs and outputs work with 24Vdc-Logik and may be directly connected withswitches, sensors (PNP-type, +24V switching) resp. sorting gates and other loadconsumers. Control of the eddyvisor via a PLC is feasible as well.

    In order to increase the current load capacity of the outputs several output signals ofone VCONI may be connected with the same logical signal and switched in parallel.

    4.3.3.1. Selection of VCONI

    All available connectors are sorted according to the slot of the VCONI in the upperline and connector designator A or B; they may be selected by this identification,built of slot number and connector designator. Possible slots of the VCONI are slot3 in desktop instrument model, slots 6 and 7 in switchboard integration model.

    4.3.3.2. Monitor inputs, force outputs manually

    All input signals are asked every approx. 200 msec. and displayed. "ON" means+24Vdc input voltage. "OFF" means no input voltage.

    All outputs may be operated manually to check electrical wiring and function.

    Remark:When entering menu all stations must be stopped. Already started measurementsare finished and may be looked at, only then the outputs may be switched manuallyand the set function of inputs and outputs is unassigned until leaving the menu.

    Pressing the key "Force Outputs (manual)" releases the outputs for manualoperation after a safety request. The name of the key changes to "DeforceOutputs.

    Attention:If the outputs are manually operated this may affect any connected instruments! Ifseveral outputs of a VCONI are switched in parallel in order to increase current loadcapacity, consider this when manual switching the outputs, danger of short circuit,however, does not exist as the outputs of the VCONI are pure +24Vdc-driver (High-Side Driver).

    The released outputs appear as key with name "OFF" for 0V resp. "ON" for +24Vdc.Pressing the key changes the condition of the key and the output. When there is novoltage at the corresponding pin of the connector, the voltage supply of the VCONI-

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 274-8 D-91320 Ebermannstadt Germany

    board has to be checked: The bridge connector has to be put for supply frominternal voltage source or the VCONI has to be supplied externally with +24Vdc.

    4.3.3.3. Allow Signal Assignment

    Each logical signal can be assigned to a physical input resp. output via the menu"Allow Signal Assignment". Logical input signals may be assigned only to onephysical input. Logical output signals may be assigned to any physical outputs. Afterpressing the key, a safety request appears and the name of the key changes to"Quit Signal Assignment".

    It is recommended to work with the standard assignment and to choose otherproposals only as an exception. In very special cases only, an individual setting isnecessary.

    Modified assignments of signals are displayed in blue colour. Pressing the key"Discard changes" restores the previous assignments and pressing "Savesettings" takes any changes. When leaving the menu a safety request asks if totake of discard assignments.

    Standard Signal assignment

    A standard assignment of signals for each constellation of instrument isrecommended. Pressing this key restores this constellation.

    Show Proposal

    A menu shows different pre-defined connector assignments. Should the standardassignment not cover the case of application, other typical assignments are givenhere. The text right beside the selection field informs on the planned application.Touching the names displays the new assignment with blue changes. Selection hasto be confirmed with key "Load Assignment" or leave the menu with key "DiscardProposal". An overview of all proposals is given in chapter 6.3.1. Proposal forSignal Assignment.

    Unassign all Pins

    All assignments between logical signals and physical inputs and outputs on thisconnector are unassigned and marked with "---".

    Logical Signal Assignment - Selection window

    Pressing the keys for logical signals assigned to the inputs and outputs opens awindow "Logical Signal Assignment". There are separate menus for inputs andoutputs. The following signals are available at the moment:

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 4-9

    Global signals, inputs

    o GlobalSelfCalibOff - Stop automatic calibration.ResetCnt-1 - reservedDisableStart - Stop START-signals.

    o PLC-IN - signals for remote control by PLCStrobe-TypeStrobe-MenuStrobe-CmdData-IN-0..7

    Signals for structure test, inputs

    o Start-SignalsStart-1..32 - start signal for structure test.

    o GatePos-SignalsGatePos-1..32, -S - message from sorting gate.

    Global signals, outputs

    o GlobalReqSelfCalib - message of eddyvisor: calibration due.Counter-1 - reservedBusy-S - structure test is running.WatchDog - reserved.

    o PLC-OUT - signals for remote control by PLC.OUT-TypeOUT-MenuOUT-CmdData-OUT-0..7

    Signals of structure test, outputs

    o End-SignalsEnd-1..32, -S - test finished

    o OK-SignalsOK-1..32, -S - test result OK.

    o NOK-SignalsNOK-1..32, -S - test result not OK.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 274-10 D-91320 Ebermannstadt Germany

    o StatActive-SignalsStatActive-1..8 - station in reference data recording or in PreventiveMulti-Frequency Testing.

    unused, unassignment of single signal assignments, symbol ---.

    4.3.4. System Properties

    Basic settings like printer, integration of eddyvisor into a local network anddate/time are managed in menu "System Properties" as well as some statusvariables are shown.

    Any changes may be discarded by key "Discard changes" unless they areexplicitly saved by the key "Save settings". When changing menu a safety requestasks if to store or discard changes or cancel change of menu.

    Pressing "Standard settings" sets all settings in this menu to standard value. Dateand time are not influenced.

    4.3.4.1. Printer setting

    Driver (printer language)

    Set the language of the printer which is used and with it the printer driver. Thefollowing languages are available:

    PS Level 3Suitable for Post Script capable printers.

    PCL 6 (Laserjet)Suitable for laser printers which support PCL6

    PCL 5C (Laserjet Color)Suitable for colour laser printers which support PCL5

    PCL 5 (Laserjet)Suitable for laser printers which support PCL5e

    PCL 3 (Laserjet + Deskjet)Suitable for many HP Deskjet printers

    HP printers resp. HP (PCL) compatible printers are exclusively supported (exceptfrom PS). The language of the printer can be found under technical data of theprinter.

    Attention: HP printers with "LIDL" (Lightweight Imaging Device Interface Language)are not supported.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 4-11

    Interface (USB or TCP/IP)

    Selects the interface of used printer. A local printer will be connected via USB, anetwork printer will be connected via Ethernet using TCP/IP.

    USB printer type

    One USB printer type only is allowed to be connected. The first printer model whichis connected after switching on resp. already connected will be used. Furtherconnected printers are ignored.

    TCP/IP Network printer

    In case of a network printer (TCP/IP) the IP-address of this network printer must bespecified. A "standard-TCP/IP port" with following settings is used:

    - Protocol: Raw- Port number: 9100- SNMP: No

    Paper size

    Determines the paper size. It may be selected between letter and (DIN) A4.

    4.3.4.2. Network setting of eddyvisor

    Obtain address, IP-address, Subnet-Mask

    Determines where the eddyvisor gets its network address from. This setting isneeded when the eddyvisor is connected to a local network.

    DHCP (Dynamic Host Configuration Protocol) means that the network addressesare obtained automatically from a DHCP-server. In this case, no further settings arerequired.

    If "static" is selected, a valid IP-address in the network and the networks subnet-mask have to be set. Remark: The subnet mask divides the IP-address into aforepart indicating the network and a back part indicating the user-ID.

    Host-name of eddyvisor

    The host name of the eddyvisor in the network is derived from the serial numberof the instrument: e.g. ev00123 for eddyvisor S/N 00123.

    Remark: Many operating systems provide the command "ping" to check the networkconnection. Access to the eddyvisor S/N 00123 is tested with ping ev00123.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 274-12 D-91320 Ebermannstadt Germany

    4.3.4.3. Date / Time

    Change date and time of internal clock.

    Changes are taken immediately.

    4.3.4.4. Form

    Company's name to be used in printouts.

    4.3.4.5. Network-Log

    The measuring results of "reference data recording" and "Preventive Multi-Frequency Testing" " or Preventive Multi-Filter Testing may be sent continuouslyto a network user. This taking down of measuring results may be switched on andoff.

    The target address of log data is given either as hostname or IP-address. Remark:Not every network supports usage of a broadcast-ID.

    4.3.4.6. Mouse pointer on/off

    A mouse pointer may be faded in for presentations with an external monitor.

    4.3.4.7. System monitor

    The values of "CPU temperature", "Motherboard temperature", "CPU core voltage"and speed of fan 1 and 2 are displayed for diagnosis of system. These values aremeaningless for operation of eddyvisor.

    4.3.5. Part Type Administration

    The standard project includes part type 1 with name "standard" and "part number01", refer to part type field on the eddyvisor in the upper left-hand side. This menuallows to create and individually name of part types.

    The available part types are listed in a table.

    Selection key Allows to select a part type.

    Type-No.Opens a window with two arrow keys and to set the number of part type. Part type 1 cannot be renumbered.

  • Manual eddyvisorSC Error! Utilice la ficha Inicio para aplicar berschrift 1 al textoque desea que aparezca aqu.ibg Doc: 0713431e, 29 May 2008 Error! Utilice la ficha Inicio para aplicar berschrift 2al texto que desea que aparezca aqu.

    Pretzfelder Strae 27D-91320 Ebermannstadt Germany 4-13

    Caption 1Opens a window with a keyboard to type individual descriptions.

    Caption 2Opens a window with a keyboard to type individual descriptions.

    DateDate when part type was created.

    Touching the headlines sorts the list of part types according to the relatingcriterion. Several touching reverts order.

    The following actions are available to create and delete part types:

    Key "Add new Part TypeAdds a part type to the list at the position with lowest number.

    Key "Copy marked Part TypeCopies the selected part type into the list at the position with lowest number.Does not copy reference and sorting data.

    Key "Copy marked Part Type with PartsCopies the selected part type into the list at the position with lowest number.Does also copy reference and sorting data.

    Key "Delete Part TypeDeletes selected part type from the list. A different part type is automaticallyselected.

    Key "Load in all stationsLoads selected part type to all test stations of the eddyvisor. Does not leavethe menu.

    If part types are set up then the key Part type selection in the upper left-handside of the screen may be used without calling the setup. It also opens the"Overview of saved Part Types", however, the parts may only be sorted andloaded. A selected part type is load into the current resp. all stations with key"Load only in this station" and "Load in all stations" and the menu is left.The key "Cancel" leaves the menu without changes.

    4.3.6. Frequencies & Locations (general)

    The test procedure of the individual stations and locations is significantlycharacterised by the test frequencies / band passes, settling delay and test time.Menu "Frequencies & Locations" allows to set all part type specific parameters.

    When changing menu a safety request asks if to store or discard changes or cancelchange of menu.

  • Error! Utilice la ficha Inicio para aplicar berschrift 1 al texto que desea que aparezcaaqu. Manual eddyvisorSCError! Utilice la ficha Inicio para aplicar berschrift 2 al texto que desea que aparezcaaqu. 29 May 2008, ibg Doc: 0713431e

    Pretzfelder Strae 274-14 D-91320 Ebermannstadt Germany

    4.3.6.1. Activate / deactivate location

    A test block of a station usually includes several locations. The eddyvisor supportstesting of different part types. Often it is the case that a certain part type has to betested at a certain location, another not.

    Locations may be excluded resp. included again by pressing the key "DeactivateLOCATION" resp. "Activate LOCATION". Name of key changes accordingly.Deactivated stations and locations are marked grey crossbred in the stations andlocations bar and are not considered at testing.

    The information which locations are active resp. inactive for a certain part type issaved with the relating part type.

    4.3.6.2. Standard settings

    This key resets all settings in this menu to standard settings.

    4.3.6.3. Discard changes

    This key discards any changes before they were explicitly saved.

    4.3.6.4. Save settings

    This key saves the setting for the currently selected location(s).

    4.3.6.5. Save all settings

    This key saves the settings for all stations and locations.

    4.3.7. Frequencies and locations (structure test)

    4.3.7.1. Tolerance zone shape (ellipse / rectangle)

    The eddyvisorS may create both elliptic and rectangular tolerance zones and inPreventive Multi-Frequency Testing it checks the test results for keeping thetolerance.

    Switching between elliptic and rectangular tolerance zone shape is done by thetolerance zone keys A - F for each frequency. Pressing TOL zone changes alltolerance zone shapes at the same time. The selected shape is symbolically shownin the key.

  • Manual eddyvisorSC Error! Utilice la