9
Electrospray ionization (ESI) University of Delaware Mass Spec. Facility (PAO)

Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

  • Upload
    others

  • View
    20

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

Electrospray ionization (ESI)

University of Delaware Mass Spec. Facility (PAO)

Page 2: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

How can ions be generated from charged liquid droplets?

Ion evaporation model:Electric field from a charged nanodroplet is sufficiently high to cause the ejection of small solvated ions from the droplet surface.• Low molecular Wight species

Charged residue model:Nanodroplets that contain a single analyte evaporate to dryness. As the last solvent shell disappears, the charge of the vanishing droplet is transferred to the analyte. • Large globular species

Chain ejection model:Unfolding switches the properties of the protein from compact/hydrophilic to extended/hydrophobic, which makes it unfavorable to reside in a droplet interior. They migrate to the droplet surface and get ejected.

Anal. Chem. 2013, 85, 2−9University of Delaware Mass Spec. Facility

(PAO)

Page 3: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

Factors affecting ESI signal intensity

• Hydrophobic proteins are ionized more efficiently than hydrophilic ones

• Folded globular proteins tend to generate ESI mass spectra with relatively low intensities, whereas unfolded polypeptide chains provide more intense signals

Folded

Unfolded

University of Delaware Mass Spec. Facility (PAO)

Page 4: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

Factors affecting ESI signal intensity

Sample concentration

have to be fulfilled to use mass spectra to determine theconcentration of components.

First, the concentrations have to remain below saturationlevels (see Fig. 4), and the relative ionization efficiencies haveto be determined. If the ligand is very small in comparison tothe protein, it might be acceptable to assume that the proteinand the (protein!ligand) complex have the same ionizationefficiency (25).

Second, to avoid in-droplet complex formation, the transi-tion from droplet to solvated molecule has to be fast andshould involve as little solvent evaporation as possible. Nano-electrospray produces very small primary droplets, whichevaporate in the micro-second range. True complexes cannotform in this time span. Loose associations between ligand andprotein in the form of a cluster may occur. However, suchclusters are not likely to withstand the desolvation process inthe transmission region of the mass spectrometer (25).

Third, the desolvation conditions in the transmission regionof the mass spectrometer when the last solvation shell isremoved should be gentle enough not to destroy correctlyformed ligand-protein complexes.

The stability of complexes in the gas phase is determinedby other forces than those in solution. In solution, hydropho-bic surfaces enhance the protein-ligand interaction becausethe surfaces avoid contact with the aqueous environment.Polar groups are often solvated and shielded by water mole-cules. In a vacuum, ionic interactions are much stronger thanin solution because there are no water molecules that canattenuate the Coulombic forces. Thus, protein complexeswhose stability is mostly based on hydrophobic surfacesmight be much less stable in vacuum and might fall apart inthe desolvation process (25).

In summary, if the experimental conditions are well chosen,the binding constants of binary complexes can be measuredusing electrospray mass spectrometry. However, there arecases in which the results do not reflect the in-solution kinet-ics of the complex formation despite the care taken; this isparticularly true for complexes stabilized by hydrophobicsurfaces.

Future Developments—The electrospray ionization processcan now be considered to be well understood. Changes willbe brought by using electrospray ionization sources to solve

FIG. 4. Change of the electrospray ion signal with analyte concentration. A, The standard behavior of the electrospray ion signal withincreasing analyte concentration. Over a range of three orders of magnitude, the signal grows linearly with concentration before it saturates.B, The signal dependence of a two component solution. The ion signals start to saturate simultaneously. They level off at the same or atdifferent total ion intensities. If the components differ considerably in hydrophobicity, the more hydrophobic component can even suppress thehydrophilic one at high concentrations (B 3) (16).

Principles of Electrospray Ionization

Molecular & Cellular Proteomics 10.7 10.1074/mcp.M111.009407–7

Signal increases over 3 orders

of magnitude and reaches

saturation

2 samples may saturate at

different intensities

More hydrophobic sample

suppresses hydrophilic

sample

Molecular & Cellular Proteomics 10: 10.1074/mcp.M111.009407, 1–8, 2011.

University of Delaware Mass Spec. Facility (PAO)

Page 5: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

For two adjacent peaks (in a multiple-

charge ion mass spectrum) m1/z1 and

m2/z2 where z1 = z2 + 1

z 2 = ((m1/z 1 )– 1) / (( m2/z 2 )– (m1/z 1 ))

Z2 =1301 - 1/ (1431-1301) = 10

m/z m/z700 750 800 850 900 950 1000 1050 1100 1150 1200 1250 1300 1350 1400 1450 1500 1550 1600 1650 1700 1750 1800 1850 1900 1950 2000 2050

%

0

100012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+

5.00e51431

1431

1431

1431

1431

13011301

1301

1301

13011193

1193 1300

14311302

1302

14301302

1302 14291302

1432

1432

1590

1432 1590

1590

1432

1590

1432

14321590

1432 1588

1590

1591

1591

1591

1591

1591

178915911789

15961789

m1/z1

m2/z2

Calculating the mass of protein from and ESI spectrum

m/z = (mass + z)/z

Mass = (m/z * z)-z

14300 = (1431*10)-

10

University of Delaware Mass Spec. Facility

(PAO)

Page 6: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

University of Delaware Mass Spec. Facility (PAO)

Δ = 0.10Z = 1/ Δ = 10

Determining charge sate from isotopic distribution

Page 7: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

MaxEnt

• MaxEnt1 uses an entropy based deconvolution algorithm that utilizes probabilistic distance as a function of mass of parent ion.

• This requires translating mass spectra into vectors and probabilities

• A probability distribution is constructed from the mass spectrum and is compared to a model of multiply charged ion of the parent mass.

• A plot of this difference as a function of parent mass is the deconvoluted spectrum

J Am Soc Mass Spectrom 1992, 3, 207-215University of Delaware Mass Spec. Facility

(PAO)

Page 8: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

MaxEnt1 deconvolution

1. Ranges: This is the mass range which will be examined .2. Resolution: Low numbers mean lots of resolution. ie The answer will be more precise but the processing will take longer. High resolution (long processing time) is approx 0.01. Low resolution (fast processing time) is approx20.3. Damage model: This is how Max Ent will evaluate the data. Uniform Gaussian is typically used.

Width at half height: Find a peak in the center of the charge state envelope and measure the width at half height. This is the value to be entered.4. Minimum intensity ratios: This is the height of adjacent peaks relative to the most intense peak which will be considered as part of the charge state envelope. 33 – 33 usually works well.

MaxEnt result is quantitave: it reflects the peak intensities of species in the mass spectrumUniversity of Delaware Mass Spec. Facility (PAO)

Page 9: Electrospray ionization (ESI)...012616_Lyso 557 (3.106) Cm (554:582) TOF MS ES+ 1431 5.00e5 1431 1431 1431 1431 1301 1301 1301 1301 1301 1193 1193 1300 1431 1302 1302 1302 1430 1302

Consideration for running modified protein samples

• Be aware of signal intensity saturation.

• Modification or breaking of disulfide bonds may cause protein to be extend à CEM, more efficiently ionized. Check charge state distribution.

• Some artifacts /peak harmonics may occur when using MaxEnt à check raw data to rule out artifacts.

• In a complex mixture, hydrophobic analytes can suppress the signal of hydrophilic analytes.

University of Delaware Mass Spec. Facility (PAO)