Fab 10 Laboratory Analysis Services

Intro to Focused Ion Beam (FIB) Chip Circuit Edit Chris Kang (Presenting)Steve Herschbein, Carmelo Scrudato,George Worth & Edward HermannMgr: Ray Wagner

FIB-SEM WorkshopLaurel, MD2/19/2016

Page 2: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Why an “Intro” Talk to This Audience?

Educational: Some of you aren’t familiar with this application.

Self-serving! We need your help…

Microchip process technology is moving faster than advances in FIB tools, process gases, and sample prep.

Tool platforms, columns & ion species, e-beam, gas chemistry & delivery, in-situ light optics, navigation & automation (remove the human factor), damage anneal, 3D sample prep, etc. – All are in need of improvement and investment!

2

Page 3: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Intro to Focused Ion Beam (FIB) Editing

FIB Chip Editing / FIB Circuit Modification / FIB Microsurgery

Most people in typical analysis fields deal with “dead things”.Even the biologists… If it isn’t dead when it arrives, it soon will be…

By contrast, we are asked to do brain surgery on “live patients”.

With up to 5 Billion transistors, will a design be ‘first time right’??

New masks & a wafer Fab run to test a theory can take months. Is there a better way to model a change, before making a change?

3

Page 4: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Why Perform FIB Chip Edit?

FIB circuit editing is used to produce a limited number of hand-built prototype devices that will roughly mimic what a mask change would accomplish.

Concept to first working parts in 1- 2 weeks, versus up to 3 months by new masks and a Fab turn.

Used to insure that the proposed change will indeed fix the issue, that the identified design deficiency is the one & only, and allows software development or hardware demonstrations to remain on schedule. It can have a major impact on Time-to-Market.

Prototyping is also commonly used by the Development, Yield, Test, and the Failure Analysis Community to better understand the physical layout, design manual vs. speed sort anomalies, test coverage assurance, defect site isolation, high speed TX/RX tuning, etc.

4

Page 5: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Chip Edit Origin: The “Yellow Wire” EC

Simple “yellow wire” edit on a two layer PC board.

Board features fully visible, no blind navigation required.

Solder in yellow wire jumpers, scratch out unwanted Cu traces.

Cut

Connect

5

Page 6: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Basic Frontside FIB Edit

Simple “yellow wire” edit on a two layer metal process.

Chip features fully visible, no blind navigation required.

FIB Process: Mill viasto metal lines & add jumper straps, mill 3 cuts.

Cut

Connect

6

Page 7: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Like an SEM, but primary beam is traditionally Ga+ (30 keV typical).

Use beam to sputter, with image created by collecting secondary electrons or ions.

Utilize transferred energy to activate precursor gases for selective etch, or deposition of conductors & insulators.

Perform high volume or precision nanoscale materials modification.

A Chip Edit FIB is a “direct write” tool, a whole Fab in a box…Imaging / Litho / Implant / Etch / Deposition

Sample

Ion column

Ion beam Ga+

(2-50 keV)

Secondary electrons

Secondary ions

Tunable Gas Jets

Chip Edit Tool Basics: What is a FIB?

7

Page 8: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Features Specific to a Chip Edit Tool

Sophisticated gas delivery systemVariety of gas precursorsMotor driven precision placement nozzlesMass flow controllers & mixing manifoldSpecial high volume XeF2 nozzle - gas concentration (some tools)

In-Situ Optical System100x FOV IR (thru silicon) &/or broadband imaging

CAD Navigation Overlay & Local Coordinate Lock FilesCamelot / GDS / Oasis / Chipview / others

High Precision StagesLaser tracked or tight tolerance with correction mapping

Leading Edge Ion ColumnsHigh current, gas immune, high resolutionPrecise pattern generation & landing energy controlFuture: SEM column for non-destructive imaging & gas processingFuture: Alternate ion species – select by mass, conductivity, etc.

8

Page 9: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Example of Drill & Fill – Create a New Via

9

M4

M2

M3

M7

M6

M5

M2

M4

XeF2 gas assisted FIB mill of a new contact via, then filled with tungsten metal.

This could be an access point for an internal electrical analysis, or a step in a functional circuitry change.

Page 10: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Dynamic Shape Changes During Si Thinning

As Received: Packaged Module Delidded Module: 100um+ Bowing

IR monitoring for thickness

Reflectance metrology thickness mapping generates corrected mill contour map for next round of shaping – down to 10 um remaining.

Center of chip is thinner than perimeter due to board-dominated CTE stress.

10

Y X

ZChip-Package

Interaction!

Page 11: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

11

FIB Backside Chip Circuit Edit Process

Backside Edit Plan:Cut at the M1- M2 contacts, connect M1 latch input to Vdd.M2 = Blue, M1 = Red, Poly = Green, S-D = Pink

Page 12: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

12

CAD Layout vs. FIB Image upon Touchdown:

Make access points for cut & contacts:

FIB Backside Chip Circuit Edit (Bulk)

Page 13: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

13

Insulate, connect the wires & cover the cut:

FIB Backside Chip Circuit Edit (Bulk)

Page 14: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Typical FIB Edit Request Format

Request to remove a pair of spurious inverters

14

Page 15: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Alternate to typical “Wiring EC Edit” Method

Option 1: Cut M2 & M3 wiring as per customer request. Option 2: Delete the dangling devices by milling them away.

15

Page 16: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Alternate to typical “Wiring EC Edit” Method

Look for “equivalent circuit’ opportunities that allow for logic removal (or fixing an output state within a driver) rather than the need to gain access to M2-M3 wiring between logic blocks.

Delete a logic circuit by milling away the transistors until only the M1 wiring remains, then bridge input to output.

16

Page 17: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Flip Chip Module Reconstruction

17

Page 18: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Chip Edit FIB of the Near Future?

Maybe(?) based around a dual-beam (plus optical) Electron & Optical for imaging.Ions & electrons (possibly photons) for processing.Ideally all coaxially arranged – Same perspective view.Full complement of gases w/gas box regulation.Alternative or complement to Ga+ – an ‘any ion’ column.

18

Page 19: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Next Gen FIB Edit Tools

Commercially Available FIB Tools:Gallium (LMIS): High brightness, small spot, moderate mass. He+ & Ne+ Gas Field Ion Sources (GFIS): Selected for high resolution imaging & light sputtering (available as a dual beam with Ga+).Xe+ (Plasma): Heavy ion selected for mass removal (dual beam w/e-).

Papers at ISTFA, FEBIP, FIB/SEM Workshop, etc.Electron Beam chemical activation & processing (dual beam w/Ga+).(83 papers & posters last year at FEBIP)

Beryllium, Silicon, Gold, Cr, AuSi, AuBeSi, etc. (LMAIS – metal alloy)being investigated for specific applications.

Lithium & Cesium cold ion beam sources in research phase.

19

Page 20: Fab 10 Laboratory Analysis Services Intro to ... - FIB SEM€¦ · FIB-SEM Workshop. Laurel, MD. 2/19/2016. ... Test, and the Failure Analysis Community to better understand the physical

Closing Remarks

FIB Chip Circuit Editing has been used by the semiconductorindustry quite effectively for over 20 years to emulate redesigned hardware.

It’s getting tough…. Chip geometries are shrinking faster than new edit tool technology is evolving.

Modern packaging (3D stacking) will bring new challenges.

Our customers still want this edge, so we are committed to working with the vendor & the research community to bring next generation tooling to the semiconductor lab.

20