38
Fluoroscopy Re- Credentialing William Robeson, Radiation Safety Officer North Shore University Hospital Radiology/Radiation Safety Office (516) 562- 3895 Presentation created by: Miyuki Yoshida-Hay

Fluoroscopy Re-Credentialing

  • Upload
    alaina

  • View
    87

  • Download
    0

Embed Size (px)

DESCRIPTION

Fluoroscopy Re-Credentialing. William Robeson, Radiation Safety Officer North Shore University Hospital Radiology/Radiation Safety Office (516) 562- 3895 Presentation created by: Miyuki Yoshida-Hay. Objectives. - PowerPoint PPT Presentation

Citation preview

Page 1: Fluoroscopy Re-Credentialing

Fluoroscopy Re-Credentialing

William Robeson, Radiation Safety OfficerNorth Shore University HospitalRadiology/Radiation Safety Office(516) 562- 3895

Presentation created by: Miyuki Yoshida-Hay

Page 2: Fluoroscopy Re-Credentialing

ObjectivesThe Joint Commission accreditation requires physicians who use or operate fluoroscopic x-ray systems be properly credentialed. This training shall include radiation safety, management of fluoroscopic radiation and operation of fluoroscopic x-ray system(s) used by the physician. Training in radiation safety and fluoroscopic radiation management is in addition to any clinical training or qualifications required to perform the specific clinical diagnostic or therapeutic procedures for which the fluoroscopic systems are used. The re-credentialing process is required in order to familiarize practitioners with the latest developments in technology and radiation safety.

Page 3: Fluoroscopy Re-Credentialing

Re-Credentialing RequirementsThe hospital requires documentation of appropriate training before granting fluoroscopic privileges Practitioners are required to be re-credentialed every two years.In order to become re-credentialed, a physician must do the following:– View this powerpoint presentation– Complete the self-assessment quiz– Upon completing the training, each physician will need to

complete an attestation. The attestation certificate is available for printing.

– Present this certificate to the department chairperson for recording the successful completion of the fluoroscopy safety requirement.

It is up to the each department to establish other training and education requirements needed to obtain fluoroscopy privileges.

Page 4: Fluoroscopy Re-Credentialing

TOPICSMobile C-armsFactors Influencing Fluoroscopy Exposure RateSources of Radiation ExposureRadiation ProtectionPersonnel MonitoringExample of a Skin Injury from FluoroscopyRadiation and PregnancySummary

Page 5: Fluoroscopy Re-Credentialing

Mobile C-armsMoved around ORs to visualize anatomy during surgeryMetallic c-arm contains x-ray tube at one end and image receptor at the otherAn associated unit contains the display

Page 6: Fluoroscopy Re-Credentialing

Modern fluoroscopy units produce images with an image intensifier (II) which brightens the image level sufficiently so that the image may be displayed on a TV screen. Fluoroscopy units are usually operated in an automatic brightness control (ABC) mode.These units will automatically adjust the brightness by first increasing the kVp to increase x-ray penetration and then adjusts the mA to increase intensity.Note: Exposure to a thick patient will be greater than to a thin patient and also abdominal fluoro will require a greater exposure than a chest fluoro due to increased thickness and tissue density in the abdomen.

Factors Influencing Fluoroscopy Exposure Rate

Page 7: Fluoroscopy Re-Credentialing

Factors Influencing Fluoroscopy Exposure Rate (cont.)

Recommendation # 1: The image intensifier input should be positioned as close to the patient as practicable. This results in a lower patient dose and sharper image.

Page 8: Fluoroscopy Re-Credentialing

Factors Influencing Fluoroscopy Exposure Rate (cont.)

Recommendation # 2: Use the exposure pedal as sparingly as possible.

Radiation exposure during fluoroscopy is also directly proportional to the length of time the unit is activated by the foot pedal. Depression of the foot pedal determines the length of exposure. The fluoroscopy time is an important determinant of patient and staff radiation dose. Fluoroscopy units are equipped with a timer and an alarm which sounds at the end of 5 minutes. The alarm serves as a reminder of the elapsed time.

Page 9: Fluoroscopy Re-Credentialing

Factors Influencing Fluoroscopy Exposure Rate (cont.)

Recommendation # 3: Use “last-image” hold and pulsed fluoro whenever possible.

Most modern fluoro units are equipped with “last-image” hold, which stores the last fluoro image and allows viewing without having to expose the patient again. Many fluoro units also offer a “pulsed fluoro mode”, in which the x-ray beam is pulsed rapidly on and off and results in a lower radiation dose without significantly degrading the appearance of the image on the display.

Page 11: Fluoroscopy Re-Credentialing

Factors Influencing Fluoroscopy Exposure Rate (cont.)

Recommendation # 4: Use the smallest field of view practicable.Radiation exposure also depends on x-ray field size and keeping the x-ray field as small as possible (by using collimators) which will decrease the dose to BOTH the patient and staff in the fluoroscopy suite. Restricting the field size not only decreases radiation dose but will also produce a better image. The contrast in the image between various tissue types will be greater for the smallest field of view that encompasses the desired anatomy.

Page 12: Fluoroscopy Re-Credentialing

Factors Influencing Fluoroscopy Exposure Rate (cont.)

Recommendation # 5: High dose or detail modes should be used only sparingly.Many fluoro units will have various dose modes, such as low dose, medium dose and high dose mode.

It is important to recognize that fluoroscopic image quality can be adversely affected by too few x-rays in the image; the image is noisy for low dose. More tissue contrast is produced by the “high dose” mode which will improve the image quality at the expense however of increased patient dose.

Page 13: Fluoroscopy Re-Credentialing

Factors Influencing Fluoroscopy Exposure Rate (cont.)

Recommendation # 6: Magnification should be used only when necessary.

Fluoroscopy units are capable of using different magnification modes. Image resolution is improved with magnification but field size is reduced and patient radiation dose is increased. Patient dose is minimized by using the lowest magnification (largest field size) appropriate for the image procedure being performed.

Under Normal mode, there is little magnification with the whole beam used to generate a bright image. Under Mag 1 mode, a smaller beam area is projected to the same II output.  The resulting object size is larger, but the image is dimmer due to the less beam input. The ABC system senses the brightness loss and either boosts machine X-ray output, increases tube voltage, or a combination of both.

Page 16: Fluoroscopy Re-Credentialing

Factors Influencing Fluoroscopy Exposure Rate (cont.)

Recommendation # 7: For C-arm type fluoroscopy units the patient should be positioned as far from the x-ray tube as practicable to minimize patient entrance dose. To reduce personnel exposure the x-ray tube should be positioned beneath the patient.

Page 17: Fluoroscopy Re-Credentialing

In the case of portable C-Arm systems, eliminating the air gap between the I-I and the patient ensures that the table top is as far away as possible from the X-ray tube, minimizing radiation exposure to the patient’s skin.

Note: The separator cone should always be utilized before commencing fluoroscopy on portable C-arm systems, as depicted below on the image at right.

   

Page 18: Fluoroscopy Re-Credentialing

Factors Influencing Fluoroscopy Exposure Rate (cont.)

In conventional under-table x-ray tube fluoroscopic units, the x-ray tube is located at a fixed distance from the patient’s skin.

In C-arm fluoroscopy, where the distance between the x-ray tube and image intensifier is fixed, the patient can be positioned in close proximity to the x-ray tube which increases the entrance skin dose and reduces image sharpness.

It is preferable to locate the C-arm x-ray tube underneath the patient. Since the radiation transmitted through the patient is typically only 5 – 10 % of the entrance dose, inadvertent exposure to the operator hand on the exit side of the patient will result in a smaller dose compared to the dose to the hand on the entrance side of the patient. Also the amount of scatter radiation the operator is exposed to on the beam exit side of the patient is significantly less than on the beam entrance side.

Page 19: Fluoroscopy Re-Credentialing

Note: The benefit is exaggerated - some operator dose occurs on the image intensifier (I-I) side.

Sorenson, 2000.

X-ray tube

X-ray tube

Page 20: Fluoroscopy Re-Credentialing

Sources of Radiation ExposureDIRECT EXPOSURE– Entrance Skin Exposure (ESE) rates (where the x-rays enter the

patient) are limited to less than 10 R/minute(NOTE: At ESE rates of 10 R/min, 30 minutes of fluoroscopy can deliver 300 R in skin dose.)

– ESE rates for typical fluoroscopy procedures are usually less than 5 R/min.

– On some machines an operator can deliberately choose a setting that will increase the output. The use of higher radiation rates or "boost" modes are useful in situations requiring high video image resolution. ESE of up to 20 R/min is permitted for short duration. Special operator reminders, such as audible alarms, are activated during "boost" modes.

Page 21: Fluoroscopy Re-Credentialing

Sources of Radiation ExposureSCATTER EXPOSURE TO PERSONNEL– Most of the radiation exposure

received by the operator or other personnel in the fluoroscopy suite is due to scatter radiation from the patient.

– The operator will be exposed to a dose rate of approximately one one-thousandth (1/1000) of the ESE rate at a distance of 1 meter from the center of the fluoroscopy field. Sorenson, 2000.

Page 22: Fluoroscopy Re-Credentialing

Sources of Radiation ExposureFactors which increase the dose from scatter radiation: – Large patients which will cause the automatic brightness

control (ABC) to adjust the kVp and mA to higher values causing greater amounts of scatter radiation.

– A large x-ray field, a result of not restricting field size will increase scatter radiation

– The length of time the fluoroscopy unit is on. Complex interventional cases will require greater procedure time, increasing dose to both the patient and operator

Page 23: Fluoroscopy Re-Credentialing

Other sources of exposure to the operator may be associated with the following:– A small percentage of exposure to the operator may

be due to leakage radiation through the x-ray tube housing.

– C-arm operators should be aware that the shielding built in to “fixed” fluoroscopy systems is not available for protection against backscatter. This may be of greater concern if the C-arm is rotated out of the normal vertical plane.

Sources of Radiation Exposure

Page 24: Fluoroscopy Re-Credentialing

Radiation ProtectionThe three most productive means of reducing radiation dose is:– Time: Minimize time spent in the radiation field.

Use of “last-image-hold” and pulse fluoro features are technical advantages in reducing the total

time x-rays are produced

– Distance: Radiation dose rates increase or decrease according to the inverse square law

Ex: Double your distance from the source and decrease your exposure by a factor of 4

– Shielding: Use of lead garments, lead gloves, thyroid shields, leaded eyeglasses, lead drapes

and clear leaded glass barriers between the

patient and operator

Page 25: Fluoroscopy Re-Credentialing

PPE and Radiation Monitoring

Page 26: Fluoroscopy Re-Credentialing

Personnel MonitoringEven when radiation protection techniques and engineering controls are in place to reduce personnel exposure, individual dose monitoring is required.

Various types of dosimeters are available (i.e., film badges, thermo-luminescent (TLD) and optically-stimulated luminescent (OSL) badges).

Badges are assigned to an individual and must never be shared.

A badge designed to measure the whole body (torso including head) should be worn at the collar – OUTSIDE the lead apron.

Page 27: Fluoroscopy Re-Credentialing

Note that the time to expression of symptoms is long enough that the patient may no longer be in the hospital when symptoms appear. The physician performing the fluoroscopy cannot discern the damage by observing the patient immediately following the procedure.

FDA Specification of Radiation-Induced Skin InjuriesThreshold Dose Typical Fluoro-On Time in Minutes

Skin Effect rem Sv Normal mode @ 10 R/min

High Dose mode @ 20 R/min

Time to Onset

Early Transient Erythema 200 2 20 minutes 10 minutes Hours

Temporary Epilation 300 3 30 minutes 15 minutes 20 days

Basal Cell Erythema 600 6 60 minutes 30 minutes 10 days

Permanent Epilation 700 7 70 minutes 35 minutes 20 days

Dry Desquamation 1000 10 100 minutes 50 minutes 30 days

Page 28: Fluoroscopy Re-Credentialing

Example of a Skin Injury from Fluoroscopy

This case, patient A is that of a 40-year-old male who underwent coronary angiography, coronary angioplasty and a second angiography procedure due to complications, followed by a coronary artery by-pass graft, all on March 29, 1990. The area of injury six to eight weeks following the procedures. The injury was described as "turning red about one month after the procedure and peeling a week later."

Page 29: Fluoroscopy Re-Credentialing

In mid-May 1990, it had the appearance of a second-degree burn. The condition in late summer 1990, exact date unknown, with the appearance of a healed burn, except for a small ulcerated area present near the center.

Example of a Skin Injury from Fluoroscopy (cont.)

Page 30: Fluoroscopy Re-Credentialing

Skin breakdown continued over the following months with progressive necrosis.

Example of a Skin Injury from Fluoroscopy (cont.)

Page 31: Fluoroscopy Re-Credentialing

The injury eventually required a skin graft. The magnitude of the skin dose received by this patient is not known. However, from the nature of the injury, it is probable that the dose exceeded 20 Gy.

Example of a Skin Injury from Fluoroscopy (cont.)

Wagner LK, Eifel PJ, Geise RA. Potential Biological Effects Following High X-ray Dose Interventional Procedures. JVIR 1994; 5:71-8

Page 32: Fluoroscopy Re-Credentialing

Pregnancy and RadiationThe decision to perform a radiological procedure on a patient who may be pregnant is a medical decision and shall be made by a physician in consultation with the patient. If the procedure is to be performed, the physician must explain the risks to the patient, provide informed consent and the appropriate consent form shall be signed.Shielding shall be used to shield the abdomen from radiation provided it does not interfere with the procedure.Every attempt must be made to minimize direct exposure to the fetus according to the principles described in this presentation.Medical emergency radiological procedures however take precedence over pregnancy status.

Page 33: Fluoroscopy Re-Credentialing

PAUSE to properly plan and prepare for study Activate dose saving features of equipment No exposures unless necessary Depress last image hold and last image grab instead PULSE at lowest possible rate

Principles for Fluoroscopy

Page 34: Fluoroscopy Re-Credentialing

Quiz1. Which parameter effects the penetration of an x-ray beam through

a patient?a) kVpb) mAc) Both

2. Automatic brightness control (ABC) _______ the quantity of radiation for large patientsa) Increasesb) Decreasesc) Has no effect

3. Correct fluoroscopy technique requires which of the following to be close to the patient:a) X-ray tubeb) Image intensifierc) Does not matter

Page 35: Fluoroscopy Re-Credentialing

Quiz

4. Which of the following reduces patient dose?a) Last image holdb) Pulsed fluoroscopyc) Both of the above

5. Which of the following is better technique to reduce patient dose?a) Use of collimationb) Use of magnification modec) Neither has an effect on dose

6. For c-arm fluoroscopy, which is better?a) Position x-ray tube under tableb) Position x-ray tube over tablec) It does not matter

Page 36: Fluoroscopy Re-Credentialing

Quiz

7. When wearing a single radiation badge during a fluoroscopy procedure, it must be worn:a) Attached to the thyroid collarb) Underneath the lead apron at the waistc) There is no need for radiation badges in fluoroscopy

8. The biggest concern for patients receiving high dose fluoroscopy is:a) Skin damageb) Cancer inductionc) Both concerns are equal

9. Fluoroscopy procedures should never be performed on a pregnant woman.a) Trueb) False

Page 37: Fluoroscopy Re-Credentialing

Quiz10. The most productive means of reducing radiation exposure during

fluoroscopy is:a) Wearing lead garmentsb) Reducing fluoroscopy timec) Both of the above

Answers1 a 6 a2 a 7 a3 b 8 a4 c 9 b5 a 10 c

Page 38: Fluoroscopy Re-Credentialing

I have read the material on radiation safety and fluoroscopy and understand the operation and radiation safety features of the fluoroscopic units that I will use.

Print Name ___________________________

Department ____________________________

The information presented in this training document is designed to give a practitioner a basic knowledge of radiation safety principles as they apply to the use of fluoroscopy. In addition, the practitioner must have knowledge and experience in the use of the specific radiographic systems which they will utilizing.This attestation must be submitted and filed with the department chairman who privileges physicians to perform specific fluoroscopic procedures. The responsibility for delineation of clinical privileges ultimately lies with the department chairman.

Signature _______________________________________ Date ___________

PHYSICIAN ATTESTATIONFLUOROSCOPY RE-CREDENTIALING