— For Fecal Immunochemical Test (FIT) —

Display

Two-rack Tray Ten-rack Tray

Sample Rack

Low Volume Sample Rack (M rack)

Calibrator and Control Rack (S rack)ibrator and Control RibrCalibr

Low Volume Sample Rack (M rac

Rack (S rack)

Sample Rack

Low Volume

Simply BetterWe have added many new features while retaining the most popular features of the old model.Advanced features include:

Easier Management with BarcodesCalibrators, controls and reagents all barcoded for quick, accurate hassle-free registration

Improved ReproducibilityMore accurate dispensing for much higher reproducibility

Ten-rack TrayTen-rack tray added as a standard feature allows simultaneous loading of 100 samples

Simultaneous Multiple AssayFor Multiple combination screening

Up to Five Days of Continued PlacementBuilt-in refrigeration allows reagents to be left in rack up to 5 days

Analytical Unit

Reagents

Hemoglobin Reagent Control

Specimen Diluent Hemoglobin Calibrator

Specimen Collection Container

Rack

CCCClClClClClicicicicickkkkClickClick

CCCClClClClClicicicicickkkkClickClick

CCCClClClClClicicicicickkkkClickClick

CCCClClClClClicicicicickkkkClickClick

CCCClClClClClicicicicickkkkClickClick

Starting Up

Loading a Sample Rack

Assay Start Screen

Measurement Result Screen

Shutdown Screen

Loading a Reagent

STEP

1

Power ON

Powering ON and Starting Up

Place a reagent on the reagent table.Place samples in a sample rack.

After performing and confirming calibration and control assay, place the sample on the rack and place the rack into a tray to start the assay.Consequently, the measurement is executed any time the tray is inserted into the rack transfer unit.

Calibration, Control, and Sample Assay

The assay screen allows checking in real time the assay status and the measurement result.The sample screen allows referring measurement result of each sample as well as detailed information.

Measurement Result

Click the Abort Assay icon (the software transitions to Waiting mode). Then, click the Exit button.The End Processing dialog box will be displayed. Select an option from the end processing menu.

Shutdown Operation

Start up the system main unit and the PC for control, and double click the NS-Prime User Program icon to automatically start up the system. The system executes self-diagnosis.When the system completes the startup process, a log-in screen will be displayed. Enter the user name and password to display the initial screen for assay.

Preparing a Reagent and Sample and Loading the Reagent

Simply Insert a Tray in the Rack Transfer Unit to Start the Assay.

2

3

4

5

The system is equipped with two-rack trays on six lanes as a standard for the analysis of a small number of samples. Of the six lanes, one lane is a priority lane, which allows priority interrupts (20 samples). The entire set of trays can be optionally changed to two-rack trays.

As a new feature, calibrators and controls can be registered through barcodes, eliminating the entire registering operation.

Ten racks (100 samples) can be simultaneously accommodated for a large number of samples. The system can be optionally equipped with ten-rack trays on both sides.

The 19-inch LED display provides greater visibility. The large and easy-to-understand icons facilitate checking the assay status, the remaining volume of reagents, the volume of waste, etc. The system also supports an optional touch panel.

Display

Two-rack Tray

Ten-rack Tray

Barcode Management of Calibrator and

Control

0 200 400 600 800 1000 1200Hemoglobin Concentration (ng/mL)

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

∆ A

bsor

banc

e

• Calibration Curve (Hemoglobin)

The reagent table can contain up to five sets of reagent bottles (max. 1500 tests, max. 5 items.) R1 and R2 are provided with a combination bottle managed with a barcode to save time and effort for installation.The interior of the reagent table is maintained at constant temperature (4–10°C), allowing reagents to be left in the reagent table for five days.

The metal probes on both R1 and R2 Dispense Arms provide improved dispensing accuracy and durability. In addition, the independent Reagent Probe Washing Unit enables the processing of 300 tests per hour, even when multiple items are assayed.

The DI-water tank capacity is increased to 10 L to reduce replacement time and effort.

Analytical Unit

Reagent Table

Reservoir Tank

40

35

30

25

20

15

10

5

0

--5

Mea

sure

d va

lue

(Hb

ng/

mL)

• Minimum Detection Limit (2SD method)

0 10/109/108/107/106/105/104/103/102/10

7.6

1/10

3.6

Dilution

• Carryover

1,250,000625,000312,500156,25078,125

00000

10012

* * * ** * * ** * * ** * * ** * * *

Assay Value by Sample ProbeCarryover

by CellHb Conc. (ng/mL)

• Within-run Reproducibility (Hemoglobin)

123456789

1011121314151617181920

MeanSDCV

58 60 60 60 59 59 59 60 58 58 58 59 60 58 59 58 59 60 58 59

59.0 0.83 1.4%

114 116 116 115 114 112 113 114 114 112 114 113 113 116 112 112 112 112 112 111

113.4 1.53 1.4%

249 250 251 252 247 251 251 250 252 248 255 254 249 255 255 252 251 252 252 248

251.2 2.35 0.9%

483 492 491 484 498 484 494 477 495 484 484 488 479 504 481 474 475 470 486 484

485.4 8.51 1.8%

Sample 1 Sample 2 Sample 3 Sample 4

Unit: ng/mL

• On-Board Stability (Hemoglobin Reagent)

2223

219218222222

0125

Day Blank

122120122122

LowControl

High

Unit: Hb ng/mL

Dilution

Mea

sure

d Va

lue

(Hb

ng/

mL)

• Linearity (Hemoglobin)

0/10 1/10 2/10 3/10 4/10 5/10 6/10 7/10 8/10 9/10 10/100

200

400

600

800

1000

1200

1400

Dispensing and Liquid Feeding Syringes

Top Cover

Syringe Unit

Reagent Table

Sampling Stage

Reaction Table

Sample Dispense ArmR2 Dispense Arm

R1 Dispense Arm

Ten-rack Tray

Two-rack Tray

LED Sample Status Indicator

Space for Placing Reagent Caps

Rack Transfer Unit

Power Button

Name of Each Part

Your Distributor

Caution

Basic Specifications

Unit Dimensions

805mm 620mm

400mm

Trade nameMeasurement principleProcessing capacitySample capacityPriority sample capacity

Sample dilution

Reaction cells

SamplingReagent dispensingMixing system

Temperature control systemLight sourceSensor

Display and input

Data memoryOutput

Barcode

DimensionsWeightRequired powerMarketing notification no. (Japan)JMDN codeMain accessories

Discrete Clinical Chemistry Analyzer NS-PrimeColloidal gold agglutination300 tests per hour max.220 samples max. (22 racks)20 samples max. simultaneously (2 racks)

Manual setting: Dilution ratio 1/1, 1/10, 1/100, 1/200Automatic setting: Dilution ratio 1/1, 1/10, 1/100

5 semi-disposable cells x 10 (resin-molded cell)Automatic cell cleaning with detergent and DI-waterNozzle type (liquid level detection function, automatic cleaning with detergent and DI-water)Nozzle type (liquid level detection function, automatic cleaning with detergent and DI-water)Mixer: Rotation type (automatic cleaning with detergent and DI-water)

Reaction table: Air bath-heating blockReagent table: Cooling with Peltier cooler

LED (wavelength: 540, 630, and 660 nm)PhotodiodeColor LED (19-inch), keyboard, mouse *Option: Touch panel display (19-inch)

240,000 tests *Option: Page printer (size A4, laser type)Reading of rack barcodes and sample barcodesReading of reagent, calibrator, and control barcodes

W805mm x D620mm x H400mm (System main unit)Approx. 70 kg (System main unit)AC100-240V 50/60Hz, 400VA (System main unit)27B1X0005500000835743010

Main options*Ten-rack tray, *Printer, *Touch panel display

◆ Sample rack: 22 pcs ◆ Calibrator and control rack (S rack): 1 pc ◆ Low volume sample rack (M rack): 1 pc ◆ Ten-rack tray: 1 pc ◆ DI-water tank: 2 pcs ◆ Washing solution tank: 2 pcs ◆ Waste tank: 1 pc ◆ Master PC (mouse, keyboard, LED display) ◆ Software (CD) ◆ LAN cable ◆ User manual for the analyzer ◆ Package insert for the reagents

For correct and safe use of the product, be sure to read the package insert and user manual before using the product.