GC Flow Path Inertness Ensuring Reliable Results Page 1

GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Embed Size (px)

Citation preview

Page 1: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

GC Flow Path Inertness Ensuring Reliable Results

Page 1

Page 2: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

What is GC Flow Path Inertness?

Page 2

Whatever goes in here…

Makes it all the way over here!

With no loss and no peak shape distortion (tailing)!!!

Page 3: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Peak Tailing – Why does it Happen?

Turbulent Flow Problem

-Dead volume, obstruction, poor installation, or severe column contamination

Active Sites in Injector, Column, and/or Flow Path

- Reversible adsorption of active compounds (-OH, -NH, -SH)

Page 3

Page 4: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Where does liner and column activity come from?

Isolated Geminal Vicinal Siloxane

- Sterically difficult to “cap” all of them(estimate 40-65% older traditional deactivations.)

- Oxidation of Stationary Phases from air leaks

- Injecting Reactive Samples or Non-volatile Contamination

Page 4

Page 5: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Agilent Inert Flow Path Solution

Ultimetal Plus– FPD

Ultimetal Capillary Flow Technology Devices Ultra Inert GC ColumnUltra Inert Gold Seal

Ultra Inert Inlet Liner

Flexible Metal Ferrules

Ultimetal Plus Inlet Weldments,Shell and Transfer Lines

Page 5

Page 6: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Top 5 Inertness Tips

1. Use Good Gas Purifiers2. Regular Inlet Maintenance3. Use Inert Liners, Seals,

Ferrules, etc. in Inlet

3. Select an Ultra Inert Column

4. Maintain an Inert Detector

Page 6

Page 7: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Agilent Ultra Inert Inlet Liners

Touchless PackagingEasy installation of new, clean liner without risk of contamination from touching.

Includes non-stick plasma treated o-ring

Page 7

Page 8: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Ultra Inert deactivation passes Endrin/DDT decomposition test after 100 injections

Endrin Decomposition Test : Robustness

1

2

3

5

6

1

2

3 4

5

6

1

2

3

5

6

2

34

5

6

Agilent Ultra Inert single taper liner (p/n 5190-2292)

%Breakdown

Endrin DDT Agilent UI liner 1st injection: 1.2 2.5

101st injection: 12.2 3.0

Competitor’s liner 1st injection: 1.1 1.7101st injection: 33.8 2.2

Competitor’s single taper deactivated liner

Peak identification: 1. DDE*2. Endrin3. DDD*4. Endrin aldehyde*5. DDT6. Endrin ketone*

* Breakdown products

1st injection101st injection

101st injection

1st injection

Page 8

Page 9: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

IS 1

12 3

4

5

IS2

IS1

12

3

4

5

IS2

Agilent Ultra Inert single taper liner with wool (p/n 5190-2293)

Competitor’s deactivated gooseneck liner with deactivated wool

Higher responses

Peaks:1. 2,4-Dinitrophenol 2. 4-Nitrophenol3. 4,6-Dinitro-2-methylphenol4. 4-Aminobiphenyl5. Pentachlorophenol IS1. Acenaphthene-d10IS 2. Phenanthrene-d10

Even with glass woolUltra Inert liners provide high responses for sensitive semivolatile acidic compounds.

Semi-Volatiles Suitability

Page 9

Page 10: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Drug of abuse are shown on GC/MS SIM chromatograms 5 ng of checkout standards on column

Basic Drug Suitability

Agilent UI single taper liner with wool (p/n 5190-2293)

Competitor’s deactivated gooseneck liner with deactivated wool

12

34

56

7

1

2

3 4

5 6

7

Higher response

Better peak shape

Peaks:1. Oxycodone2. Temazepam3. Flunitrazepam4. Heroin5. Nitrazepam6. Clonazepam7. Alprazolam

Page 10

Page 11: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

1. 1-Propionic acid

2. 1-Octene

3. n-Octane

4 4-Picoline

5. n-Nonane

6. Trimethyl phosphate

7. 1,2-Pentanediol

8. n-Propylbenzene

9. 1-Heptanol

10. 3-Octanone

11. n-Decane

DB-5ms Ultra Inert vs. Competitors

1

23

4 5

6 7

8

9 10 11

min0 2 4 6 8 10

pA

10

12.5

15

17.5

20

22.5

25

27.5 Agilent J&W DB-5msUI

30m x 0.25mm x 0.25um(P/N 122-5532UI)

1

2 34

5

67

8

9 1011

min0 2 4 6 8 10

pA

6

8

10

12

14

16

18

20

1

2 34

5

67

8

9 1011

min0 2 4 6 8 10

pA

6

8

10

12

14

16

18

20

1

2 3

4

5

7

8

9 1011

6min0 2 4 6 8 10

pA

8

10

12

14

16

18

CompetitorColumn

CompetitorColumn

Page 11

Ultra Inert Test Mix

Page 12: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Semivolatiles Prone to Peak Tailing(App. Note 5990-3416EN)

Page 12

Page 13: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Semivolatiles Prone to Peak Tailing(App. Note 5990-3416EN)

Page 13

Page 14: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Resolution of benzo-b & k fluoranthene isomers

19.00 19.500

100000

500000

1000000

1500000

Time-->

Abundance

Benzo-b-fluoranthene

Benzo-k-fluoranthene

Resolution of 1.2

As good as it gets…it only gets worse!

Page 14

Page 15: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Agilent J&W DB-624 UI vs. Competitor 624ms Organic acid performance at 200 ppm

Column: Agilent J&W DB-624UI 30 m x 0.32 mm x 1.8 um vs Rxi-624Sil MSOven: 35C 7.45 min hold, 6.72/min to100C (2.23min hold),10.08 /min to 220C(4.47 min hold), 16.79 /min (4.17 min hold)Carrier : Helium 39.6 cm/s (approx. 2.6 mL/min) set at 35C, EPC-Constant FlowInlet: Split, 20:1 at 250C (total flow approx 51 mL/min, and 11.2 psi)Inlet liner: Ultra Inert with wool Detector: FID at 280C, H2 @ 40 mL/min, Air @ 400 mL/min, N2 makeup @ 30 mL/min

min10 12 14 16 18 20 22 24 26

pA

2

4

6

8

10

12

14

16

18

Acetic acid

propionic acidbutanoic acid Octanoic acid

No detection on Competitor’s 624ms

Severe tailing on competitor’s 624ms

Page 15

Page 16: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

New Flexible Metal Ferrule - Design FeaturesGroovesreduce bulk stiffness during compression and improves ID constriction.

Angle eliminates wall frictional contact to allow the material to flow into the ID area for larger range of constriction.

Tighter ferrule ID specification and tolerances (30%)

Wider opening allows for easier column insertion

1

2

3

4

Page 16

Page 17: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Current vs. New Agilent Flexi-Metal Ferrule

Current Ferrule

Stiff- prone to cause damage

Poor installation Can lead to: column crushing leakage thread damage

Small ID compression (40 to 60um) seals at tip

Tight pinch point

Agilent Flexi-Metal Ferrule

More forgiving

Easier to use

Large ID compression (100 to 120um)

Broad lamp shadesealing surface

Forgiving pinch point

Page 17

Page 18: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

New Touchless Dial Packaging

Easy column insertion

No lost ferrules

No contamination risks

Page 18

Page 19: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

1 ng on Column Forensic/Toxicology Check Out Mix Flexible metal Ferrules vs. Siltite installed post column

min15 16 17 18 19

pA

6

6.2

6.4

6.6

6.8

7

7.2

7.4

7.6

7.8 12

13

14

15

16

17

18

19

20

21

22

23

24 25

26

Blue = Flexible Metal Ferrules in Ultimate Union Red = Siltite Ferrules in Ultimate Union

Improved Response forBenzodiazepines

12 SKF‐525a, proadifen13 Oxazepam14 triphenyl phosphate 15 Codeine16 Lorazepam17 Diazepam18 Hydrocodone19 Dronabinol20 Oxycodone21 Temazepam22 Flunitrazepam23 Diacetylmorphine24 Nitrazepam25 Clonazepam26 Alprazolam

Page 19

Page 20: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Enlarged chromatograms for sensitive pesticides

UltiMetal Plus on Flexible Metal FerruleIs deactivation needed for such small surface?

4x10

0

1

2

4x10

0

1

2

Counts vs. Acquisition Time (min)4.5 5 5.5 6 6.5 7 7.5

Untreated FMFPost column Std 500ppb

FMF + UltiMetal PlusPost column Std 500ppb

AcephateOmethoate

Demeton-S

AcephateOmethoate

Demeton-S

Page 20

Page 21: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Agilent UI Gold Seal: Deactivated Gold Surface

• Soft gold plating is essential for proper sealing

• Ultra Inert chemistry blocks active sites (gold is NOT inert)

• Smooth surface doesn't leak

Agilent MIM seal Alterative vendor machined seal

Reliable ppb and ppt measurements require

attention to the little things!

Page 21

Page 22: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Response Comparison for Sensitive Pesticides

3x10

0

1

Standard gold seal

3x10

0

1

Ultra Inert Gold Seal

3x10

0

1

Counts vs. Acquisition Time (min)4 4.5 5 5.5 6 6.5 7 7.5

3x10

0

4

8

treaded SS

3x10

0

4

8

3x10

0

4

8

Standard gold seal

Counts vs. Acquisition Time (min)4 4.5 5 5.5 6 6.5 7 7.5

Acephate Omethoate

Acephate Omethoate

Acephate Omethoate

Acephate Omethoate

Acephate Omethoate

Acephate Omethoate

100 ppb STD 500 ppb STD

Ultra Inert Gold Seal

treaded SS

Page 22

Page 23: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

New Product: Agilent Inert Inlet

UltiMetal Plus treatment forinert surface 7890 inlet shell & top weldments

•Limits adsorption/degradation of active analytes

•GC/MS and GC-ECD pesticide analysis

or Any Active Analyte Analyses

Agilent’s proven proprietary Ultimetal

Plus surface treatment

Page 23

Page 24: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Putting It All Together –Inert Flow Path vs. Standard Flow Path

20.20 20.40 20.60 20.80 21.00 21.20 21.40 21.60

10000

20000

26000

Green = Inert Flow PathBlue = Standard Flow Path

Column: Agilent J&W HP-5ms UI 30 m x 0.25 mm x 0.25 um Oven: 100C 4 min hold, 10/min to 280C, 6 /min to 300C (4.67 min hold),Carrier : Helium 52.7 cm/s (2 mL/min) set at 100C, EPC-Constant FlowInlet: Pulsed Splitless 35 PSI pulse until 0.73 min, 0.75 min purge 50 ml/min, gas saver 20, ml/min at 2 minInlet liner: Ultra Inert with wool / Standard single taper liner with wool (p/n 5190-3165)Gold Seal: UI Gold Seal / Standard gold sealDetector: MSD Scan mode 40 t0 450 m/z, 230 C source temp, 150 C Quad temp, 310 C transfer line

1

23

4

5

1. Dronabinol2. Oxycodone3. Temazepam4. Flunitrazepam5. Diacetyl morphine

Drugs of Abuse 250 ppb using HP-5ms UI

Page 24

Page 25: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

So Remember…

1. Use Good Gas Purifiers2. Do Regular Inlet Maintenance3. Use Inert Liners, Seals,

Ferrules, etc. in Inlet

3. Select an Ultra Inert Column

4. Maintain an Inert Detector

Page 25

Page 26: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

AND Don’t Touch Anything!!! French Fries!

Column: DB-5ms, 30m x 0.25mm, 0.25umCarrier: H2, 60 cm/sec, constant flowInjector: split 1:20, 250oCDetector: FID, 320oC, N2 makeup gasOven: 40oC for 0.75 min, 40-325oC at 20oC/min,

325oC for 30 min

Red: French Fry

Blue: system blank

Procedure:(1) Held French fry for 5 seconds.(2) Fingertip was wiped with paper towel to remove as much of the offending material as possible.(3) Lightly touched the part of the column sticking up above the ferrule.(4) Installed column into injector.(5) Set oven temperature to 40oC.(6) Started oven temperature program as soon as oven reached 40oC.

Page 26

Page 27: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Contamination from Liquid Soap

Column: DB-5ms, 30m x 0.25mm, 0.25umCarrier: H2, 60 cm/sec, constant flowInjector: split 1:20, 250CDetector: FID, 320C, N2 makeup gasOven: 40C for 0.75 min, 40-325C at 20C/min,

325C for 30 min

Red: Liquid Soap

Blue: system blank

Procedure:(1) One very small drop of liquid soap placed on one fingertip.(2) Fingertip was wiped with paper towel to remove as much of the offending material as possible.(3) Lightly touched the part of the column sticking up above the ferrule.(4) Installed column into injector.(5) Set oven temperature to 40C.(6) Started oven temperature program as soon as oven reached 40C.

Page 27

Page 28: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Contamination from Hand Lotion

Column: DB-5ms, 30m x 0.25mm, 0.25umCarrier: H2, 60 cm/sec, constant flowInjector: split 1:20, 250CDetector: FID, 320C, N2 makeup gasOven: 40C for 0.75 min, 40-325C at 20C/min,

325C for 30 min

Procedure:(1) One very small drop of hand lotion placed on one fingertip.(2) Fingertip was wiped with paper towel to remove as much of the offending material as possible.(3) Lightly touched the part of the column sticking up above the ferrule.(4) Installed column into injector.(5) Set oven temperature to 40C.(6) Started oven temperature program as soon as oven reached 40C.

Red: Hand Lotion

Blue: system blank

Page 28

Page 29: GC Flow Path Inertness.ppt - Agilent€¦ · What is GC Flow Path Inertness? ... Agilent UI liner 1st injection: ... Contamination from Liquid Soap Column: DB-5ms, 30m x 0.25mm,

Page 29

ANY QUESTIONS?Technical Support