Upload
nickan
View
70
Download
0
Embed Size (px)
DESCRIPTION
Measurements in Fluid Mechanics 058:180:001 (ME:5180:0001) Time & Location: 2:30P - 3:20P MWF 218 MLH Office Hours: 4:00P – 5:00P MWF 223B-5 HL. Instructor: Lichuan Gui [email protected] http:// lcgui.net. Lecture 4. Measurement systems and static response. - PowerPoint PPT Presentation
Citation preview
Measurements in Fluid Mechanics058:180:001 (ME:5180:0001)
Time & Location: 2:30P - 3:20P MWF 218 MLH
Office Hours: 4:00P – 5:00P MWF 223B-5 HL
Instructor: Lichuan [email protected]
http://lcgui.net
2
Lecture 4. Measurement systems and static response
3
Measuring systems and their components
Essential systems in fluid mechanics experiment
flowing fluids, flow-producing apparatus, test models etc. 1. Physical system:
2. Measuring system:
3. Experimenter(s):
sensors, electric and electronic circuits, data acquisition and processing devices, and software person(s) who plans, executes, and interprets the measurements
Response of measuring system - relationship between values of an input and an output
Inputs of measuring system
1. Desired inputs
2. Undesirable inputs
a. interfering inputs
b. modifying inputs
- add noise to desired inputs
- change response to desired inputs
Example: hot-wire anemometer used to measure air jet flow from a nozzle to lab room
- The draft of air produced by ventilation system in lab acts as interfering inputs.
- The room temperature change acts as modifying inputs.
4
Filtering, compensation, and output correction
Filters - used to reduce or eliminate undesirable input effects
- no-pass filters remove all fluctuations, permitting only a steady component
- low-pass filters remove fluctuations with frequencies above a cut-off value
- high-pass filters remove fluctuations with frequencies below a cut-off value
- band-pass filters remove all fluctuations except those with frequencies within a certain band
- band-reject filters remove all fluctuations with frequencies within a certain band
Filters classified according to frequency range:
Filters classified in terms of physical operation
- electrical-electronic filters, applied to electric signals;
- mechanical filters, designed to filter motion or force fluctuations, e.g. shock absorbers used to reduce vibration of an apparatus;
- thermal filters, designed to remove temperature fluctuations, e.g. thermal insulation
- electromagnetic filters, designed to remove the interfering effects of electric and magnetic fields- digital filters, applied to recorded signals
Rs1
Measuring gauge
V0
Filtering, compensation, and output correction
Compensation - introduce additional interfering or modifying inputs to partly or entirely cancel the original undesirable effects.
Example:
- sensitive to ambient-temperature fluctuationsBonded strain gauge
V0
Rs1
Measuring gauge
Rs2
Reference gauge
𝑅𝑠2=𝛼 (∆𝑇 )𝑅𝑠
Analytical correction - remove undesirable effects & errors from outputaccording to knowledge of undesirable inputs & system responses 5
6
Modes and functions of measuring system components
Operation modes:
Response modes:
1. Passive – output energy supplied by input
2. Active – output energy supplied by external excitation source
Functions of measuring system components:
1. Sensing used to produce output
2. Convection and conditioning used to transform output to a form, amplitude, or both more suitable for observation or further processing
- analogue, discrete (digital, binary, etc.), or hybrid
3. Transmission used to transfer signals or other information from one component to another
4. Processing and storage used display or store output
7
Static response and static calibration
Static system
- constant or slowly varying input and output
Static calibration:
1. Zero drift – parallel shift of primary calibration curve
2. Sensitivity drift – a change in the slope of the primary calibration curve
- determine input-output relationship (calibration curve) with standard system
- performed separated for each desired input
- accuracy depends on that of instruments used as standard
Effects of undesirable inputs:
Static response
1. Theoretically determined by physical law e.g. liquid manometer for gas pressure difference measurement, Fig. (a)
Hydrostatic law:
2. More commonly determined by static calibration, e.g. variable-reluctance pressure transducer, Fig. (b) calibrated with liquid manometer:
8
Static response and static calibration
Static performance characteristics:
• Static sensitivity
• Scale readability
• Span (input full-scale)
• Full-scale output
• Dynamic range
• Non-linearity
• Threshold
• Resolution
• Hysteresis
- slope of input-output relationship
- ratio of largest to smallest values of input
- smallest input level for detectable output
- difference between the output value corresponding to an input value reached from below and the output value corresponding to the same input value reached from above.
Elastic hysteresis of an band
Input
Out
put
- constant in linear system
linear
- range of input to be measured with acceptable accuracyspan
- range of output values measured from minimum to maximum input values
full-
scal
e ou
tput
- local sensitivity varies over input range in non-linear system
non-linear
- minimum change in output can be observed
- smallest input change for detectable output change
- maximum deviation of actual response from straight line determined by least-square fit of calibration measurements
9
Normality test and removal of outliers
Normality test
- Used to assess randomness of repeat measurement values
• Rearrange a number of repeat values xi, i=1,2,,N , so that xi xi+1
• Compute percentage of repeat values that are not more than xi , i.e.
• Compute mean value and variance as
𝜇𝑥=1𝑁∑
𝑖=1
𝑁
𝑥𝑖
𝜎 𝑥2= 1
𝑁−1∑𝑖=1
𝑁
(𝑥 𝑖−𝜇𝑥 )2
• Normalize the repeat values as
𝑥𝑖∗=
𝑥 𝑖−𝜇𝑥
𝜎𝑥
• Plot yi vs. xi* on probability
graph paper as right
• Assess deviation of plottedpoints from the Gaussian line
10
Normality tests and removal of outliers
Outliers
• Spurious values due to
(2) Temporary or intermittent undesirable input
(1) human error, e.g. misreading of an instrument’s output
• Identified by application of Chauvenet’s criterion
Value xi is a outlier if 𝜏 ∙𝜎𝑥 ≤|𝑥𝑖−𝜇𝑥|
Linear least-square fit (LLSF)
A set of calibration measurements:
LLSF line equation:
11
Homework
- Questions and Problems: 1, 4 on page 41
- Read textbook 2.1-2.2 on page 19-31
- Due on 08/31