KISS Internal Hex Abutment System

Internal Hex KISS Abutment Procedure

Embed Size (px)

DESCRIPTION

No Description

Citation preview

Page 1: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Page 2: Internal Hex KISS Abutment Procedure

Healing abutment in place

Choose matchingabutment

The presentation that follows lists only one combination of parts. Obviously the clinical situation may call for substitution of another part on this slide.

KISS Internal Hex Abutment System

Page 3: Internal Hex KISS Abutment Procedure

Prosthetic instruments needed

KISS Internal Hex Abutment System

Page 4: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Remove healing abutmentfrom implantAssemble screw driver with

Thumb Knob

Page 5: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Insert assembled screw driver into impression transfer

Page 6: Internal Hex KISS Abutment Procedure

Insert the KISS Abutment into implant lining up the lobes of the abutment with the lobes of the implant

KISS Internal Hex Abutment System

Use only light finger pressure to torque screw

Page 7: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Remove the screwdriver thumb wheel assembly

Page 8: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Insert driver into the abutment screw and turn torque ratchet until torque of 30 Ncm is reached and the head of the ratchet releases

Insert the hex driver within the 30Ncm torque ratchet

Page 9: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Lobes within the impression matches flats on the abutment

Plastic snap impression has three ribs that match the lobes inside the impression

Page 10: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Line up the lobes of the KISS Abutment with the internal lobes of the snap impression coping

Seat snap impression coping on the solid abutment

The outside should snap on the bevel of the implant.

Page 11: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Inject impression material around the impression transfer

Page 12: Internal Hex KISS Abutment Procedure

Insert tray with impression material

KISS Internal Hex Abutment System

Seat impression to capture dental arch

Page 13: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Remove the impression from the mouth when material has set

Impression snap is captured and picked up within the impression

Page 14: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Use a crown form to reline over the sleeve to create a well fitting temporarySeat the temporary resin

sleeve over the abutment

Page 15: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment SystemLaboratory abutment analog is inserted into the impression coping. Make sure to match the lobes on the abutment analog to the lobes of the impression snap

Page 16: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Apply soft tissue replica material around abutment level analog Pour dental stone into impression

Page 17: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Dental stone model with analog in place after separation of impression from dried stone

Page 18: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Apply engaging waxing sleeve for a single crown, non-engaging waxing sleeve for a splinted multi-unit restoration.

Engaging waxing sleeve has internal lobes to match the lobes on the abutment analog

Page 19: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Apply wax over the waxing sleeve and complete the undercasting

Page 20: Internal Hex KISS Abutment Procedure

Cast wax up with the usual procedures and try in the casting on the stone model

Stack and fire the ceramic in the usual manner

Ceramic crown on model

KISS Internal Hex Abutment System

Page 21: Internal Hex KISS Abutment Procedure

KISS Internal Hex Abutment System

Cement restoration on abutment