Irradiator Specification allation Qualification (IQ) ftware validation Operational Qualification (OQ) Performance Qualification ( nstrate irradiator ifications are met tallation in accordance with ed upon specifications een vendor and operator he irradiator) ma dosimetry mples: sources installed, rce rack moves properly, veyance eam racterisation of beam Characterize dose delivery of the Irradiator under baseline conditions: Dose magnitude, distribution and reproducibility Gamma 1)Fully loaded irradiator with single density 2)Mixed densities in the irradiator 3)Incomplete load configurations 4)Process interruptions E-beam 1)Surface dose profile 2)Dose distribution (3 dimensions) 3)Dose versus process parameters 4)Process interruptions Establish process for irradia of product within predescribed limits Process specification

Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Embed Size (px)

Citation preview

Page 1: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Irradiator Specification

Installation Qualification (IQ)+ Software validation

Operational Qualification (OQ) Performance Qualification (PQ)

Demonstrate irradiatorspecifications are met(installation in accordance withagreed upon specificationsbetween vendor and operatorof the irradiator)

Gamma

No dosimetryExamples: sources installed,Source rack moves properly,conveyance

E-beam

Characterisation of beam

Characterize dose delivery of theIrradiator under baseline conditions:

Dose magnitude, distribution and reproducibility

Gamma

1)Fully loaded irradiator with single density2)Mixed densities in the irradiator3)Incomplete load configurations4)Process interruptions

E-beam

1)Surface dose profile2)Dose distribution (3 dimensions)3)Dose versus process parameters4)Process interruptions

Establish process for irradiation of product within

predescribed limits

Process specification

Page 2: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Installation Qualification of electron beam irradiator

Characterisation of the electron beam

Position and shape of the beam spot (ensure product surface is covered) Electron energy Scan width and uniformity

Beam current

Involve dosimetry

Methods inISO/ASTM 51649

Note: beam profile needs to be characterised at multiple distances from the conveyor

Page 3: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

electrons

EXAMPLE: Beam width characterisation; Setpoint 60 cm

77 cm

59 cm

32 cm

Mark with pen at distance 25 cm (so peak not related to dose delivery)

Page 4: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

electrons77 cm

32 cm

EXAMPLE: Beam spot characterisation

Mark with pen (so peak not related to dose delivery)

Page 5: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

EXAMPLE E-Beam energy measurement

electrons

Based on descending slope energy can be calculated (ASTM51649)

Al

Page 6: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Operational Qualification of Gamma irradiator

1) “Standard operating conditions” i.e. Irradiator filled to design capacity with a material of homogeneous density

2) “At or beyond Standard operating conditions” of the irradiator

At least 2 different runs, each with a different density (“low” and “high”) At least 3 irradiation containers in each run 3-dimensional dosimeter grid

DUR as function of density (compare with specification)Min and Max dose zone identified as function of density

Dose distribution study

Dose magnitude study (scalability with timer setting)

Variability of dose delivery

Impact of density mixing on Dose magnitude and distribution

Incompletely filled irradiation containers

Process interruptions

Compare resultsWith those of 1)

Dose rate as function of density (throughput, compare with specification)

In the same run or even in the same irradiation container

Page 7: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Transition Dose Map: Low density (or empty)

followed by high density loads

1st and last dose map Compare to

homogeneous map

Partial Load Map: May use compensating

dummy to center load product

Map at maximum dose locations and compare to fully loaded adjacent containers.

Low density (or empty as worst case)followedby high density

Dose map at least their interfase

Impact of density mixing on dose magnitude and distribution

Page 8: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Partially filled irradiation container

Dose map their interfase

Incompletely filled irradiation containers

Page 9: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Process interruption Cycle source between fully up and fully down (multiple times) Single material Irradiation container adjacent to the source plane Response of dosimeter might be influenced by fractionated exposure

Page 10: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Operational Qualification of electron beam irradiator

1) “Standard operating conditions” i.e. Irradiator filled to design capacity with a material of homogeneous density

2) “At or beyond Standard operating conditions” of the irradiator

At least 1 run At least 3 irradiation containers in each run 3-dimensional dosimeter grid

DUR (as function of density)Min and Max dose zone identified as function of density

Dose distribution study

Dose magnitude study (scalability with process parameters)

Variability of dose delivery

Surface dose map Demonstrate whole surface is effectively irradiated

Process interruption study

All scenarios for interruption need to be included in the test program

For all pathways (single – double sided exposure, different beam energy)

Edge effects Product irradiated as a single entity

Page 11: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

EXAMPLE: Dose versus process parameters at an electron beam facility

Page 12: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

EXAMPLE: Dose distribution in reference product (double sided exposure)

40cm high 0.15 g/cc stack of homogeneous material

10 MeV electron beam

Page 13: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

EXAMPLE: Dose distribution in reference product (single sided exposure)

0.15 g/cc stack of homogeneous material

10 MeV electron beam

Page 14: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Process interruption study

Location closest to the scanhorn 1-dimensional dosimeter strip in direction of product conveyance Response of dosimeter might be influenced by fractionated exposure Test all possible sources of interruptions e.g. Accelerator, conveyor, safety, electrical Parameters at the extremes of operating conditions

EXAMPLE: Accelerator interruption at maximum of conveyor speed

Page 15: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Performance Qualification, GOAL

Determining dose delivery characteristics to product in a defined load configuration and for a specified process flow

Minimum and maximum dose zone determined

Variation of dose delivery characterized

Process capability for meeting specified dose range is assessed

Process monitoring practice established

Acceptance criteria

Page 16: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Select a Product Load Configuration and a Process Flow

Select a Performance Qualification dosimeter grid

• Product Dose Requirements• Irradiator Operational Qualification• Processing Efficiency

• Operational Qualification Dose Distribution Characteristics• Additional Locations to Account for Product Heterogeneity• Product Load Configuration / Orientation (partial fill of irradiation container)• Consider Routine Monitoring Practices

Performance Qualification, METHOD

Write protocol and get it approved by all parties

Execute

Analyse data

Write report and get it approved by all parties

Page 17: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Process specification includes

Description of packaged product (dimensions, weight, orientation in shipper) and acceptable variations Loading pattern in the irradiation container Conveyor path, process flow including lead / trail product if required Setpoints and tolerances on irradiator parameters Dose range of the product Temperature restrictions Restrictions on time between end of manufacturing and completion of irradiation Process monitoring position, frequency and acceptance criteria Any required re-orientation for multiple exposure

Page 18: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

Questions:

-Do you have a signed record on file stating all the characteristics of the irradiator?

-Have you during IQ / OQ verified characteristics of the irradiator as they are mentioned in that signed record with specifications?

-Did PQ cover all scenarios of processing the product (partially filled cartons, partially filled irradiation containers, ...) ?

-Is the outcome of PQ detailed in a process specification? Is it sufficiently detailed and signed by contract irradiator and product manufacturer?

Page 19: Irradiator Specification Installation Qualification (IQ) + Software validation Operational Qualification (OQ) Performance Qualification (PQ) Demonstrate

References:

-IAEA Guidelines for development, validation and routine control of industrial radiation processes

-ASTM 51649:2005 Practice for dosimetry in an electron beam facility for radiation processing at energies between 300 keV and 25 MeV

-ASTM 2303:2003 Standard guide for absorbed dose mapping in radiation processing facilities

-AAMI TIR 29:2012 Guide for process characterization and control in radiation sterilization of medical devices

- ISO 11137 series