Next generation Single Photon Emission Computed Tomography

Dr Andy Boston

[email protected]

Page 2: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

UoL Nuclear Physics Research Group

• Research

• Study of the structure of the nucleus

• State-of-the-art semiconductor based instrumentation

• Teaching

• Modular MSc Post Graduate Taught programmes

• Knowledge exchange

• Collaborative work with industry

• Focus on Healthcare, Security & Energy

• CPD skills training

• Group

Page 3: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

3

Page 4: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Sensors: Knowledge Exchange

Direct application in medical security and energy areas as evidenced by funding from: CLASP/PNPAS, EPSRC/TSB, NERC, MRC, NHS, NNL (NDA), AWE

PROSPECTUS • Novel SPECT imaging system

NNL • Nuclear Decommissioning applications

NERC • Radionuclide Transport

All projects are collaborations some with industrial partners. All involve contributions from parts of STFC.

Page 5: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

What are the challenges?

• In Nuclear Medicine:

– Know the energy

– Want the location over a small field of view

– Need to cope with high count rates

– Multimodality applications (eg PET/CT)

– Image fusion

Page 6: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

What are the detector requirements?

• Need to know the location of the radiation:

– Use a mechanical collimator (Anger Camera)

– Use positron annihilation for LoRs

– Use other electronic collimation

• Range of energies:

– Medical 141 keV – 511 keV

– Security 60 keV – 2 MeV

• Operating environment:

– B-fields? Microphonics? High temperature?

Page 7: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

What is SPECT?

Functional imaging modality

8 mSv typical dose

Page 8: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk
Page 9: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Pulse Shape Analysis to decompose

recorded waves

Highly segmented HPGe detectors

· · · ·

Identified interaction

points

(x,y,z,E,t)i

Reconstruction of tracks e.g. by evaluation of

permutations of interaction points

Digital electronics to record and

process segment signals

g

1

2 3

4

reconstructed g-rays

Ingredients of g-Tracking

Page 10: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

From AGATA to Application

Page 11: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

ProSPECTus

Next generation Single Photon Emission Computed

Tomography Nuclear Physics Group, Dept of Physics, University of Liverpool, Nuclear Physics & Technology Groups, STFC Daresbury Laboratory, MARIARC &

Royal Liverpool University NHS Trust, CCC NHS Foundation Trust

Page 12: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

SPECT : Opportunities

Requirement • Want “A coupled SPECT/MRI system for simultaneous

Functional/Anatomical imaging” & need to see this demonstrated

The Market • 350-400 SPECT machines in the UK alone ~£0.5M each. CAGR

9%. Translating to ~100 new units pa. (US/EU similar 1 machine per 800k population).

• Camera ~50% of system value.

• Worldwide installed MRI base ~33000 machines, large potential retrofit and new build market.

• Main players GE (dominant), Siemens and Philips

TrueScatter

Other

Page 13: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

ProSPECTus : The Implication

Patient benefits:

• Earlier and more effective diagnosis of tumours (higher probability of effective treatment).

• Higher sensitivity offering the scope for shorter imaging time (more patients through one machine per day) or lower doses of radio pharmaceuticals.

• Cardiac and brain imaging / Image larger patients

• Monitor response to treatment

SPECT/MRI:

• Functional/Anatomical

• Image co-registration

Page 14: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

What’s different? Conventional SPECT Compton camera

• Gamma rays detected by a gamma camera

• Inefficient detection method

• Incompatible with MRI

• 2D information

• Gamma rays detected by a Compton camera

• Positions and energies of interactions used to locate the source

• 3D information.

Source

E0

Factors that limit the performance of a Compton Imager: Energy resolution, Detector position resolution, Doppler Broadening

Page 15: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Research : Compton Imaging

E1

E2

g

E1

E2

g

212

2 111cos

EEEcme

o Compton Cones of Response projected into image space

Page 16: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Research : Compton Imaging

E1

E2

g

E1

E2

g

212

2 111cos

EEEcme

o Compton Cones of Response projected into image space

Page 17: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Research : Compton Imaging

E1

E2

g

E1

E2

g

212

2 111cos

EEEcme

o Compton Cones of Response projected into image space

Page 18: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Research : Compton Imaging

E1

E2

g

E1

E2

g

212

2 111cos

EEEcme

o Compton Cones of Response projected into image space

Page 19: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Research : Compton Imaging

E1

E2

g

E1

E2

g

212

2 111cos

EEEcme

o Compton Cones of Response projected into image space

Page 20: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Image Reconstruction Algorithms

• Sensors have excellent energy & position information.

• Uniformity of sensor response

• Optimise existing:

– Analytical

– Iterative

– Stochastic

• Requirement for GPU acceleration

Page 21: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

FWHM ~ 8mm: full 3D

6 cm source to crystal

30 mm crystal to crystal

E = 1408 keV, 30 keV gate

Compton Camera measurements (Ge/Ge)

No PSA (5x5x20)

Iterative reconstruction

Page 22: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

MTRL- Medical Teaching and Research Laboratory Converting a lab to host a refurbished SPECT/CT scanner in partnership with STFC Daresbury Laboratory and Royal Liverpool University Hospital. Uses - training medical physics students and research.

SPECT/CT:

Funded by STFC Futures (refit laboratory) and University of Liverpool (purchase used scanner) Expert advice from Royal Liverpool Hospital

MTRL Timeline • Autumn 2012- discussions with Liverpool

University, Royal Hospital, STFC Futures. • Dec 2012- Futures offer funding for lab

refit. • Jan 2013- Clear equipment from lab T3 • Mar 2013- Install lead shielding on walls

and doors, new floor, rewire, paint, fit air con, fit personnel safety system.

• Early 2014 obtain & install scanner • SPECT/CT • 2014 first students

MTRL: Medical Teaching and Research Laboratory

Page 23: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

ProSPECTus: Next generation SPECT

• Detector head sensitivity maximised for 99mTc 141 keV gamma rays (also works at higher energies e.g. 131I 364keV).

• Sensitivity is a factor of 10 improvement over LEHR collimated SPECT detector heads.

• Multi-isotope imaging in single acquisition

• Wide energy range with one system

• 2 semiconductor detectors housed in 1 cryostat

• MRI-compatible

23

Page 24: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Gamma Ray Imaging Spectrometers

• 10 x imaging sensitivity

• Factor 2 -3 improvement in position resolution

• System locates sources in space (3D) and can identify the isotopes in the material

• Radioactive material found quickly, reducing cost, false alarms and search time – increasing cargo throughput

Page 25: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

Next generation Single Photon Emission Computed Tomography

Dr Andy Boston

[email protected]

Page 26: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

The potential: 3D Gamma & Optical Stereoscopic image fusion

26

1.5m standoff A Compton Camera provides 3D source location

Page 27: Next generation Single Photon Emission Computed … generation Single Photon Emission Computed Tomography Dr Andy Boston ajboston@liv.ac.uk

The potential: 3D Gamma & Optical Stereoscopic image fusion

27

1.5m standoff A Compton Camera provides 3D source location