Excellent for Basic compounds Improved separation for acidic or metal coordination compounds Simple mobile phase conditions Analysis of Basic Compounds Conventionally end-capped C18 columns still have many residual silanol groups on the silica surface of stationary phase that can form ionic bonds with basic compounds. The resulting peak tailing makes accurate quantification of analytes difficult, especially in trace analyses. COSMOSIL 3C18-EB with a better end-capping treatment offers improved peak shape and separation particu- larly for basic compounds. Comparison 1 (Basic Compounds) Comparison 2 (Basic Compounds) Octadecyl Bonded HPLC Column COSMOSIL 3C18-EB COSMOSIL Application Data COSMOSIL Application Data NACALAI TESQUE, INC Column: Column size: 4.6mmI.D. x 75mm Mobile phase: Acetonitrile/ 20mmol/l Phosphate buffer(pH7.0) = 60/40 Flow rate: 1.0 ml/min Temperature: 40°C Detection: UV254nm Sample: 1; Amitriptyline (0.2mg/ml) 2; Propylbenzene (I.S.) (2.0mg/ml) Inj. Vol: 2.0μl COSMOSIL 3C 18 -EB Competitor A C18 , 3 m N Amitriptyline Competitor B C 18 , 3 m Competitor C C 18 , 3 m COSMOSIL Application Data COSMOSIL Application Data NACALAI TESQUE, INC Column: Column size: 4.6mmI.D. x 75mm Mobile phase: Acetonitrile/ 20mmol/l Phosphate buffer(pH7.0) = 40/60 Flow rate: 1.0 ml/min Temperature: 40°C Detection: UV235nm Sample: 1; Dextromethorphan (0.5mg/ml) 2; Benzene (I.S.) (7.5mg/ml) Inj. Vol: 1.5μl H N H H 3 C O H 3 C Dextromethorphan COSMOSIL 3C 18 -EB Competitor A C18 , 3 m Competitor B C 18 , 3 m Competitor C C 18 , 3 m

Octadecyl Bonded HPLC Column COSMOSIL 3C18-EB › download › Cosmosil › COSMOSIL_3C18-EB.pdf · Dextromethorphan COSMOSIL 3C 18-EB Competitor A C 18, 3 @Ûm Competitor B C 18,

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

  • ・Excellent for Basic compounds・ Improved separation for acidic or metal coordination compounds・Simple mobile phase conditions

    Analysis of Basic CompoundsConventionally end-capped C18 columns still have many residual silanol groups on the silica surface of stationary phase that can form ionic bonds with basic compounds. The resulting peak tailing makes accurate quantification of analytes difficult, especially in trace analyses. COSMOSIL 3C18-EB with a better end-capping treatment offers improved peak shape and separation particu-larly for basic compounds.

    ・Comparison 1 (Basic Compounds)

    ・Comparison 2 (Basic Compounds)

    Octadecyl Bonded HPLC Column

    COSMOSIL 3C18-EB

    COSMOSIL Application DataCOSMOSIL Application Data

    NACALAI TESQUE, INC

    Column: Column size: 4.6mmI.D. x 75mmMobile phase: Acetonitrile/ 20mmol/l Phosphate

    buffer(pH7.0) = 60/40Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV254nm

    Sample: 1; Amitriptyline (0.2mg/ml)2; Propylbenzene (I.S.) (2.0mg/ml)

    Inj. Vol: 2.0µl

    COSMOSIL 3C18-EB Competitor A

    C18, 3 m

    N

    Amitriptyline

    Competitor B

    C18, 3 m

    Competitor C

    C18, 3 m

    COSMOSIL Application DataCOSMOSIL Application Data

    NACALAI TESQUE, INC

    Column: Column size: 4.6mmI.D. x 75mmMobile phase: Acetonitrile/ 20mmol/l Phosphate

    buffer(pH7.0) = 40/60Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV235nm

    Sample: 1; Dextromethorphan (0.5mg/ml)2; Benzene (I.S.) (7.5mg/ml)

    Inj. Vol: 1.5µl

    HN

    HH3C

    OH3C

    Dextromethorphan

    COSMOSIL 3C18-EB Competitor A

    C18, 3 m

    Competitor B

    C18, 3 m

    Competitor C

    C18, 3 m

  • Analysis of Acidic or Metal Coordination CompoundsAcidic or metal coordination compounds can also form bonds with residual silanol groups and metal impurities causing peak tail-ing. COSMOSIL 3C18-EB with a better end-capping treatment and high purity silica gel offers improved peak shape and separa-tion for acidic or metal coordination compounds.

    ・Acidic Compounds ・Metal Coordination Compounds

    Synthesis ReproducibilityIf you use insufficiently end-capped C18 columns for basic compound analyses, you may need to spend much time adjusting mo-bile phase conditions. You may need

    1) more than 3 or more organic solvents or buffers2) ion-pair reagents or additives3) buffers in different pH

    In addition, the complexity of the mobile phase often is detrimental to reproducibility. COSMOSIL 3C18-EB with a new end-cap-ping treatment allows good peak shape and separation using simple mobile phase conditions.

    ・Analysis with Simple Condition

    COSMOSIL Application DataCOSMOSIL Application Data

    NACALAI TESQUE, INC

    Column: 3C18-EBColumn size: 4.6mmI.D. x 100mmMobile phase:Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV237nm

    Sample: 1; Amlodipine (0.0077mg/ml)2; Benazepril (0.0225mg/ml)

    Inj. Vol: 50µl

    Acetonitrile/ 20mmol/l Phosphate buffer(pH2.5) = 30/70

    Methanol/ Acetonitrile/ 20mmol/l KH2PO4, 0.2%Triethylamine(pH3.0 with H3PO4) = 35/15/50(*)

    (*): USP 38(1) Amlodipine and Benazepril Hydrochloride Capsules, Performance Tests

    Complicated Condition Simple Condition

    COSMOSIL Application DataCOSMOSIL Application Data

    NACALAI TESQUE, INC

    Column: 3C18-EBColumn size: 4.6mmI.D. x 75mmMobile phase: Acetonitrile/ 20mmol/l Phosphate

    buffer(pH2.5) = 30/70Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV254nm

    Sample: 1; Salicylic Acid (1.0mg/ml)2; Ethyl p-Hydroxybenzoate (I.S.)(0.05mg/ml)

    Inj. Vol: 1.0µl

    HO

    O OH

    Salicylic Acid

    l)ml)

    COSMOSIL Application DataCOSMOSIL Application Data

    NACALAI TESQUE, INC

    Column: 3C18-EBColumn size: 4.6mmI.D. x 75mmMobile phase: Acetonitrile/ 20mmol/l Phosphoric Acid = 10/90Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV254nm

    Sample: 1; Oxine-copper (0.03mg/ml)2; Caffeine (I.S.) (0.3mg/ml)

    Inj. Vol: 2.0µl

    N

    OCu

    O

    NOxine-copper

  • Routine Analysis3C18-EB has excellent synthesis reproducibility and column-to-column reproducibility, so it is applicable in routine analysis suchas quality control of drugs.

    ・Excellent Synthesis Reproducibility

    Analysis under Ammonium Acetate BufferCOSMOSIL 3C18-EB produces sharp peaks using ammonium acetate buffer in LC-MS.

    Analysis of standards in USP

    COSMOSIL Application DataCOSMOSIL Application Data

    NACALAI TESQUE, INC

    Column: 3C18-EBColumn size: 4.6mmI.D. x150mmMobile phase: Acetonitrile / H2O = 60/40 Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV254nm

    Sample: 1; Acetophenone (0.05mg/ml)2; Methyl Benzoate (0.5mg/ml)3; Benzene (2.0mg/ml)4; Toluene (2.0mg/ml)

    Inj. Vol: 1.5µl

    Column: 3C18-EBColumn size: 4.6mmI.D.x 150mmMobile phase: Acetonitrile/ 20mmol/l Phosphate

    buffer(pH7.0) = 60/40Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV254nm

    Sample: 1; Amitriptyline (0.2mg/ml)2; Propylbenzene (I.S.) (2.0mg/ml)

    Inj. Vol: 2.0µl

    COSMOSIL Application DataCOSMOSIL Application Data

    NACALAI TESQUE, INC

    Column: 3C18-EBColumn size: 4.6mmI.D. x 75mmMobile phase:Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV254nm

    Sample: 1; Amitriptyline (0.2mg/ml)2; Propylbenzene (I.S.) (2.0mg/ml)

    Inj. Vol: 1.0µl

    Acetonitrile/ 10mmol/l CH3COONH4 = 60/40Acetonitrile/ 20mmol/l Phosphate buffer(pH7.0) = 60/40

    N

    Amitriptyline

    COSMOSIL Application DataCOSMOSIL Application Data

    NACALAI TESQUE, INC

    Column: 3C18-EBColumn size: 4.6mmI.D. x 100mmMobile phase: A; Acetonitrile/ 36.7mmol/l Phosphate buffer(pH7.6) = 10/90

    B; Acetonitrile/ 36.7mmol/l Phosphate buffer(pH7.6) = 80/20 B conc. 0 20%(0 10min), 20 100%(10 30min)

    Flow rate: 1.0 ml/min Temperature: 40°CDetection: UV302nm

    Sample: 1; Omeprazole related compound A (0.001mg/ml)2; Omeprazole (0.001mg/ml)

    Inj. Vol: 20µl

    USP 37(3) Esomeprazole Magnesium Delayed-Release Capsules, Impurities

  • 1205

    For research use only, not intended for diagnostic or drug use.

    NACALAI TESQUE, INC.Nijo Karasuma, Nakagyo-ku, Kyoto 604-0855 JAPANTELFAXWebsiteE-mail

    : +81-(0)75-251-1730: +81-(0)75-251-1763: www.nacalai.com: [email protected]

    Specifications

    Silica Gel High Purity Porous Spherical SilicaAverage Particle Size 3 μmAverage Pore Size approx. 120 ÅSpecific Surface Area approx. 300 m2/gStationary Phase Octadecyl GroupBonding Type Monomeric Main Interaction Hydrophobic InteractionEnd-capping Treatment Near-perfect TreatmentpH Range 2-10*Carbon Content approx. 14.5%*Optimum pH range of columns based on silica gel is between 2 and 7.5. Extreme pH may decrease column lifetime.

    Ordering Information

    Other size may be available. Please enquire.

    Product Name Column Size Product NumberCOSMOSIL 3C18-EB Packed Column

    2.0 mm I.D. x 50 mm 09794-212.0 mm I.D. x 75 mm 09795-112.0 mm I.D. x 100 mm 09796-012.0 mm I.D. x 150 mm 09797-912.0 mm I.D. x 250 mm 09798-813.0 mm I.D. x 50 mm 09799-713.0 mm I.D. x 75 mm 09800-213.0 mm I.D. x 100 mm 09811-81

    Product Name Column Size Product NumberCOSMOSIL 3C18-EB Packed Column

    3.0 mm I.D. x 150 mm 09814-513.0 mm I.D. x 250 mm 09827-914.6 mm I.D. x 50 mm 09840-014.6 mm I.D. x 75 mm 09841-914.6 mm I.D. x 100 mm 09842-814.6 mm I.D. x 150 mm 09843-714.6 mm I.D. x 250 mm 09844-61

    COSMOSIL 3C18-EB Guard Column 4.6 mm I.D. x 10 mm 09839-41

    SiCH3H3C

    Octadecyl Group