Upload
ahmedgalalwaly1056
View
185
Download
4
Tags:
Embed Size (px)
DESCRIPTION
Physiologic Monitoring Systems, Cardiac Electrophysiology
Citation preview
5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA Tel +1 (610) 825-6000 Fax +1 (610) 834-1275 Web www.ecri.org E-mail [email protected]
UMDNS Information
This Product Comparison covers the following device term and product code as listed in ECRI Institutes Universal Medical Device Nomenclature System (UMDNS):
Monitoring Systems, Physiologic, Cardiac Electrophysiology [17-898]
Physiologic Monitoring Systems, Cardiac Electrophysiology
Scope of this Product Comparison
This Product Comparison covers physiologic monitoring and recording systems used for cardiac
electrophysiology studies (EPSs) and associated therapies. Included are dedicated units, as well as cardiac
catheterization physiologic monitoring systems that have EPS capabilities. For information on other physiologic
monitoring systems, see the following Product Comparisons:
Physiologic Monitoring Systems, Acute Care; Neonatal;
ECG Monitors; Monitors, Central Station
Physiologic Monitoring Systems, Cardiac Catheterization
Physiologic Monitoring Systems, Stress Exercise, Cardiac;
Pulmonary
Physiologic Monitoring Systems, Telemetric; ECG
Monitors, Telemetric; Monitors, Central Station
These devices are also called: cardiac electrophysiology
monitoring systems, cardiac monitors, electrocardiographic
monitors, operating room monitors, and OR monitors.
Purpose
Cardiac EPSs are used, in conjunction with cardiac
catheterization, for patients with arrhythmia or conduction
disorders. During an EPS, several diagnostic tests and therapeutic treatments are performed, including analysis
of the function of various components of the atrioventricular (AV) conduction system under certain stressors,
determination of the site of interventions, and assessment of sinus node function. EPSs also include induction and
termination of ventricular tachycardias by means of an electrical stimulator in a controlled setting for diagnostic
purposes or to evaluate responses to antiarrhythmia drug therapies. Other tests currently performed in EPS
laboratories include evaluation of automatic implantable cardioverter-defibrillators (AICDs) to ensure proper
functioning and tilt tests to diagnose presyncope.
EPSs are indicated for patients with sinus node dysfunction, AV block, and supraventricular and ventricular
tachycardias. They are also performed on patients undergoing
pharmacologic, electrical, or ablative treatment of arrhythmias;
survivors of cardiac arrest; and patients with Wolff-Parkinson-White
syndrome (anomalous AV excitation), unexplained syncope, or
heart palpitations.
Patients being considered for AICDs undergo EPSs before
implantation to determine, among other things, the cause of the
arrhythmia, the optimal site for shock delivery (i.e., lead placement),
Physiologic Monitoring Systems, Cardiac Electrophysiology
2 2009 ECRI Institute. All Rights Reserved.
and the most effective sequence of stimulation to achieve normal cardiac rhythm. (For more information, see the
Product Comparison titled Pacemakers, Cardiac, Implantable; Resynchronization.)
Principles of operation
EPSs are typically performed in cardiac
catheterization laboratories (cath labs) or dedicated
fluoroscopy suites. Before an EPS can be
performed, the patient must first undergo cardiac
catheterization, which allows endocardial catheter
electrodes capable of both electrical stimulation
and recording to be guided into the heart intra-
arterially or intravenously using fluoroscopy. (For
a detailed discussion of cardiac catheterization, see
the Product Comparisons titled Physiologic
Monitoring Systems, Cardiac Catheterization and
Radiographic/Fluoroscopic Systems,
Angiographic/Interventional; Cardiovascular.)
Before catheterization, a diagnostic 12-lead electrocardiogram (ECG) is usually performed to obtain baseline
values. This preliminary study can also disclose unsuspected abnormalities.
During catheterization, the electrophysiology (EP) monitoring/recording system amplifies, conditions, and
records signals obtained from the intracardiac electrodes as well as signals from surface electrodes and blood
pressure transducers. These monitoring systems also aid in the placement of catheters and help the physician
observe any changes resulting from treatment during catheterization.
A basic EP monitoring and recording system consists of a horizontal or vertical console with a monitor, a chart
recorder to print data and waveforms, a data recording system (e.g., magnetic or optical disk), and plug-in
modules to perform various amplification and conditioning functions. The recorder should be able to record 4 to
32 ECG signals, and the recording amplifiers should have high- and low-frequency filters for recording signals
from either surface or intracardiac electrodes. The recorder should also be able to respond to frequencies of up to
500 Hz. Most systems have between 6 and 38 channels (some may have up to 192 channels) and input switching
systems for selecting various combinations of electrodes. Waveforms are displayed on the screen and are
recorded correspondingly; most systems allow waveforms to be frozen on the monitors screen. Some can show
test protocols, results, and calculations on a separate video display unit, while others can annotate the chart
printouts, which can also feature a variety of reference lines, such as grids and event markers. Additional
workstations are available for analysis and manipulation of data and waveforms for data management purposes.
Final reports can be generated at the workstation, and waveforms and data can be stored on digital media (e.g.,
magnetic tape, optical disk).
The number of intracardiac electrodes used for EPSs varies. Generally, two to four catheters are used; the
number is determined by the complexity of the study being performed. Points of placement include high in the
right atrium, in the region of the sinus node; at the AV junction; in the apex of the right ventricle; across the
tricuspid valve ring to record His bundle activity; and in the coronary sinus. To perform tricuspid valve analysis,
the electrode at the catheter tip is drawn through the valve to detect a signal change. Catheters commonly
measure from 4 to 8 French (Fr), and each catheter can have from two to eight electrodes with interelectrode
distances of 1 to 10 mm. Externally, three to six leads may be used. During an open-chest procedure, the surgeon
may also touch the myocardium with a wand-like device that has electrodes at the tip.
Electrical stimulation is performed with a programmable electrical stimulator that has an isolated output. (EP
monitoring and recording systems may not include stimulators, which would need to be purchased from a
stimulator supplier.) Stimulation is introduced during sinus rhythm or atrial or ventricular pacingfor instance,
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 3
to induce ventricular tachycardias, the characteristics of which can guide AICD programming or reveal the
patients response to antiarrhythmic or other drugs. The stimulator must be able to deliver at least two precisely
timed impulses during spontaneous and paced rhythms with cycle lengths of 150 to 1,500 msec, as well as
multiple extra stimuli from two or more sites simultaneously. It is also advisable for the stimulator to have a
safety switch to prevent pacing at
Physiologic Monitoring Systems, Cardiac Electrophysiology
4 2009 ECRI Institute. All Rights Reserved.
Reported problems
Most problems associated with electrophysiologic monitoring and recording involve inserting, advancing, and
maintaining the position of the catheter. These problems include hematomas, infection, vascular damage,
thrombosis, mechanically induced electrical irritation and arrhythmias, catheter clotting, and balloon rupture.
Other common problems are poor electrode preparation and placement.
The hazard of electric shock to patients must be considered in any room where electrical equipment is used
particularly in settings where a catheter provides a direct pathway for current to travel to the conduction system
of the heart. A safe power-distribution system and its associated grounding are essential, as are isolated electrical
inputs on all equipment that may come in contact with the patient (along with periodic inspection and
measurement of interequipment leakage currents). In the United States, the electrical wiring and fittings must be
installed in accordance with the U.S. National Electrical Code and standards for safe use of electricity in
healthcare facilities.
Damaged muscle tissue interferes with and slows the ventricular depolarization wave, causing reentrant
currents, which may trigger sustained ventricular tachycardia. This condition can degenerate into ventricular
fibrillation or death.
Purchase considerations
ECRI Institute recommendations
Included in the accompanying comparison chart are ECRI Institutes recommendations for minimum
performance requirements for cardiac EP systems.
EP systems may be mobile (e.g., cart mounted), tabletop, or wall mounted. The maximum channels (i.e., the
maximum number of modules that can be plugged into the unit and operated at one time) should be selected by
the physician based on procedural requirements. The system should have at least 12 surface electrocardiogram
(SECG) channels and 24 intracardiac electrocardiogram (IECG) channels, as well as programmed lead switching.
The EP lab monitoring and recording systems should have a console with high-resolution color monitors
suitable for display of waveforms and data; an integrated computer with data-processing software and recording
media (e.g., DVD, optical disk); a chart recorder with printing capabilities; plug-in modules that acquire vital
signs and procedure monitoring functions; and a remote amplifier unit to process and condition cardiac electrode
signals.
The EP system should be able to annotate chart printouts using a variety of reference lines (e.g., grids, event
markers), be able to interface with RF-generating devices and cardiac cath lab imaging systems, and have
multitasking software with user-configurable parameters that allows real-time signal analysis.
The EP lab system should be capable of networking and interfacing with a hospital information system; such
networking should be bidirectional and Health Level 7 (HL7) compliant. The EP lab system should also have the
ability to interface with a national cardiology database (e.g., American College of Cardiology/National
Cardiovascular Data Registry [ACC/NCDR]) and cardiology data management system.
Other considerations
Cardiac EP systems are typically customized to meet the facilitys needs. A basic system is configured with
amplifiers and filters for signal acquisition and processing, a display screen, and a recorder. It is possible to use
preexisting amplifiers and interface them with a corresponding display screen and recorder.
Clinicians should determine the maximum number of ECG channels desired; the number of available ECG
channels varies among systems. Varying sweep speeds up to 100 mm/sec must be available when performing
EPSs. Display screens vary in the number of waveforms that are displayed; correspondingly, the recorders vary
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 5
in the number of traces recorded. Some EP systems do not offer recorders as standard equipment because all the
waveforms can be retrieved from storage and viewed onscreen for manipulation. However, if a recorder is
required, it can be interfaced with these systems. Semiautomated or automated input switching systems allow for
the selection of various electrode combinations. Some systems allow the addition of amplifiers as needed.
Monitors used during EPSs are modular, allowing additional capabilities to be added later as the needs of the
lab change. In addition, although each manufacturers system is configured differently, most allow some
flexibility in adapting their configurations to the particular needs of the hospital. Therefore, in choosing a system,
it is important that the hospital consider its present and future needs, space limitations, and patient volume.
The number of channels provided by the recording system should correspond to the institutions applications.
Current systems vary in the type of recorder used (i.e., thermal, fiberoptic, ink-jet, or laser).
Data communications and data storage capabilities vary among systems. Communications capabilities can
provide links to other EP labs, central workstations, personal computers, or hospital data management systems.
Storage capabilities include various hard disk capacities and can also include a variety of digital media for
archival purposes and easy retrieval of data. Hospitals should consider their communications and storage needs
before making a purchase decision. EP monitoring and recording systems also come with a variety of analysis
and data management software packages, which allow user-programmable or customized report generation.
Stimulators are not offered as standard components by most EP monitoring and recording system
manufacturers and must be purchased from a stimulator supplier. This must be factored into the cost of the EP
system.
Facilities should also consider whether they require separate rooms for EPSs only or catheterization/EPSs.
When planning, the time required for each test, type and number of tests, and patient volume should be
considered. Some cardiac cath lab monitoring systems are also used for EPSs. For more information on equipment
planning, facilities should contact ECRI Institutes Applied Solutions Group (ASG).
Cost containment
Because EP is a dynamic technology, manufacturers frequently make software changes to enhance
performance or to meet current clinical needs; therefore, facilities should negotiate for software upgrades and
enhancements for no cost for the life of the equipment. Service contract costs should be specified in all purchase
contracts; they are usually negotiable. If a hospitals clinical engineering department plans to service the
equipment after the warranty period, the cost of service training should be included in the purchase contract, and
the hospital should negotiate for an inventory of frequently needed parts to be placed in the facility on
consignment. All clinical training should be included at no charge.
A life-cycle cost (LCC) analysis can be used to compare high-cost alternatives and/or to determine the positive
or negative economic value of a single alternative. For example, hospitals can use LCC analysis techniques to
examine the cost-effectiveness of leasing or renting equipment versus purchasing the equipment outright.
Because it examines the cash-flow impact of initial acquisition costs and operating costs over a period of time,
LCC analysis is most useful for comparing alternatives with different cash flows and for revealing the total costs
of equipment ownership. One LCC techniquepresent value (PV) analysisis especially useful because it
accounts for inflation and for the time value of money (i.e., money received today is worth more than money
received at a later date). Conducting a PV/LCC analysis often demonstrates that the cost of ownership includes
more than just the initial acquisition cost and that a small increase in initial acquisition cost may produce
significant savings in operating costs. The PV is calculated using the annual cash flow, the dollar discount factor,
and the lifetime of the equipment (in years) in a mathematical equation.
Physiologic Monitoring Systems, Cardiac Electrophysiology
6 2009 ECRI Institute. All Rights Reserved.
The following represents a sample seven-year PV/LCC analysis for an EP system.
Present Value/Life-Cycle Cost Analysis
Assumptions
Dollar discount factor is 6%
Inflation rate is 6% for a full-service contract
Service contract is based on 7% of the purchase price
Capital costs
Basic system = $105,000
Additional slave monitor = $4,000
Fluoroscopy image storage = $10,000
Total Capital Costs = $119,000
Operating costs
Service contract, years 2 through 7 = $8,330/year
Total Operating Costs = $0 in year 1; $8,330/year for years 2 through 7
PV = ($168,980)
As illustrated by the above sample PV/LCC analysis, the initial acquisition cost is only a fraction of the total
cost of operation over seven years. Therefore, before making a purchase decision based solely on the acquisition
cost, buyers should consider operating costs over the lifetime of the equipment.
Other costs not included in the above analysis, but important to consider before making a final purchasing
decision, include the costs of purchasing and operating EP imaging equipment. The fluoroscopic imaging system
used during the catheterization procedure should be, at minimum, a single-plane configuration with either a
tilting table or a C-arm. A biplane configuration is preferable and is often necessary for larger patients.
Maintenance costs for the imaging equipment (e.g., x-ray tube replacement) must also be considered.
EPSs can be performed in a cardiac cath lab, a dedicated or mobile fluoroscopy suite, or a dedicated EP
laboratory. Because an EPS can take three to six hours or more, hospitals should analyze an EP laboratorys
impact on existing cardiac catheterization or fluoroscopy services. Scheduling problems in cath labs may result
due to increasing numbers of EPSs being performed. As a result, hospitals may consider upgrading or adding a
new lab facility dedicated to EPSs. Installing a dedicated EP lab can cost between $300,000 and $500,000, while a
new cath lab will cost twice that. Depending on laboratory setup and utilization issues, hospitals must choose
between a dedicated EP system or a cath lab system with an EP option.
For further information on PV/LCC analysis, customized analyses, and purchase decision support, readers
should contact ECRI Institutes SELECTplus Group.
Stage of development
Since the late 1960s, cardiac EPSs have come into wide clinical acceptance for evaluating cardiac arrhythmias.
Over time, there has been increased computerization of the cath labs in which EPSs are typically performed. This
enables automated calculations and storage of waveforms and data on digital media.
Bibliography
Bevilacqua LM, Berul CL. Advances in pediatric electrophysiology. Curr Opin Pediatr 2004 Oct;16(5):494-9.
Darling EJ. Overview of cardiac electrophysiologic testing. Crit Care Nurs Clin North Am 1994 Mar;6(1):1-13.
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 7
Determinants of predicted efficacy of antiarrhythmic drugs in the electrophysiologic study versus
electrocardiographic monitoring trial. The ESVEM investigators. Circulation 1993 Feb;87(3):323-9.
Jalife J. Basic cardiac electrophysiology for the clinician. 2nd ed. Hoboken (NJ):Wiley-Blackwell; 2009.
Mason JW. A comparison of electrophysiologic testing with Holter monitoring to predict antiarrhythmic-drug
efficacy for ventricular tachyarrhythmias. N Engl J Med 1993 Aug 12;329(7):445-51.
Scheinman MM, Keung E. The year in review of clinical cardiac electrophysiology. J Am Coll Cardiol 2008 May
27;51(21):2075-81.
Strong growth in EP boosts cardiac rhythm market. Clinica 1996 Feb 5;691:16.
Ward DE, Camm AJ. Dangerous ventricular arrhythmiascan we predict drug efficacy? [letter]. N Engl J Med
1993 Aug 12;329(7):498-9.
Willems S, Weiss C, Ventura R, et al. Catheter ablation of atrial flutter guided by electroanatomic mapping
(CARTO): a randomized comparison to the conventional approach. J Cardiovasc Electrophysiol 2000
Nov;11(11):1223-30.
Zickler P. Electrophysiology. Med Imaging 1995 Jul;10(7):44-51.
Supplier information
BARD
Angiomed GmbH & Co Medizintechnik KG Sub C R Bard Inc [332055]
Wachhausstrasse 6
Karlsruhe D-76227
Germany
Phone: 49 (721) 94450 Fax: 49 (721) 9445111
Internet: http://www.crbard.com
E-mail: [email protected]
Bard China Sub C R Bard Inc [343164]
Swissotel Beijing Hong Kong Macau Center Dongsishitiao
Beijing 100027
People's Republic of China
Phone: 86 (10) 65014225
Bard Electrophysiology Div CR Bard Inc [435644]
55 Technology Dr
Lowell, MA 01851
Phone: (978) 441-6202, (800) 824-8724 Fax: (978) 323-2260
Internet: http://www.bardep.com
Bard Nederland bv Sub C R Bard Inc [343215]
Weverstedehof 10
Nieuwegein NL-3430 BD
The Netherlands
Phone: 31 (30) 6000570 Fax: 31 (30) 6022443
Internet: http://www.crbard.com
E-mail: [email protected]
Physiologic Monitoring Systems, Cardiac Electrophysiology
8 2009 ECRI Institute. All Rights Reserved.
GALIX
Galix Biomedical Instrumentation Inc [176265]
2555 Collins Ave Suite C-5
Miami Beach, FL 33140
Phone: (305) 534-5905 Fax: (305) 534-8222
Internet: http://www.galix-gbi.com
E-mail: [email protected]
GE HEALTHCARE
GE Healthcare Asia (Japan) [300443]
4-7-127 Asahigaoka Hino-shi
Tokyo 191-8503
Japan
Phone: 81 (3) 425826820 Fax: 81 (3) 425826830
Internet: http://japan.gehealthcare.com/cwcjapan
E-mail: [email protected]
GE Healthcare France [171319]
11 avenue Morane Saulnier
Velizy Cedex F-78457
France
Phone: 33 (6) 07453493 Fax: 33 (1) 34495202
Internet: http://www.gehealthcare.com
E-mail: [email protected]
GE Healthcare Technologies Diagnostic Cardiology [441472]
8200 W Tower Ave PO Box 414
Milwaukee, WI 53223-3219
Phone: (414) 355-5000, (800) 643-6439 Fax: (414) 355-3790
Internet: http://www.gehealthcare.com
GE Healthcare UK Ltd [401906]
Amersham Place
Little Chalfont HP 94SP
England
Phone: 44 (870) 6061920 Fax: 44 (1494) 544350
Internet: http://www.gehealthcareeurope.com
MENNEN MEDICAL
Mennen Medical Corp [101498]
950 Industrial Blvd
Southampton, PA 18966
Phone: (215) 259-1020, (800) 223-2201 Fax: (215) 675-6212
Internet: http://www.mennenmedical.com
E-mail: [email protected]
Mennen Medical Ltd [171234]
4 Ha-Yarden Street PO Box 102
Rehovot IL-76100
Israel
Phone: 972 (8) 9323333 Fax: 972 (8) 9328510
Internet: http://www.mennenmedical.com
E-mail: [email protected]
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 9
SIEMENS MEDICAL
Siemens Canada Ltd [174735]
2185 Derry Rd W
Mississauga, ON L5N 7A6
Canada
Phone: (905) 819-8000, (888) 303-3353 Fax: (905) 819-5777
Internet: http://www.siemens.ca
E-mail: [email protected]
Siemens Medical Solutions USA Inc Health Service Corp [399199]
51 Valley Stream Pkwy
Malvern, PA 19355
Phone: (610) 219-6300, (888) 826-9702 Fax: (610) 219-3124
Internet: http://www.siemensmedical.com
E-mail: [email protected]
Siemens SA de CV [339105]
Poniente 116 No 590
Cd de Mexico 02300
Mexico
Phone: 52 (5) 3282000 Fax: 52 (5) 3282017
Internet: http://www.siemens.com.mx
E-mail: [email protected]
ST JUDE MEDICAL
St Jude Medical UK Ltd [301608]
Stratfrod Business & Technology Parl Banbury Road
Stratford upon Avon CV37 7GY
England
Phone: 44 (1789) 207600 Fax: 44 (1789) 207601
Internet: http://www.sjm.com
E-mail: [email protected]
St Jude Medical Japan KK [382507]
F-Nisei Ebisu Building 6/Fl 3-16-3 Higashi Shibuya-ku
Tokyo 150-0011
Japan
Phone: 81 (3) 57780702 Fax: 81 (3) 57780704
Internet: http://www.sjm.com
E-mail: [email protected]
St Jude Medical Inc Cardiology/Atrial Fibrillation Div [452237]
14901 DeVeau Pl
Minnetonka, MN 55345-2126
Phone: (952) 933-4700, (800) 328-3873 Fax: (952) 933-0307
Internet: http://www.sjm.com
E-mail: [email protected]
TEB
TEB (Tecnologia Eletronica Brasileira Ltda) [174429]
Avenida Diederichsen 1057-1059 Vila Guarani
Sao Paulo-SP 04310-000
Brazil
Physiologic Monitoring Systems, Cardiac Electrophysiology
10 2009 ECRI Institute. All Rights Reserved.
Phone: 55 (11) 50178555 Fax: 55 (11) 50176472
Internet: http://www.teb.com.br
E-mail: [email protected]
Note: The data in the charts derive from suppliers specifications and have not been verified through
independent testing by ECRI Institute or any other agency. Because test methods vary, different products
specifications are not always comparable. Moreover, products and specifications are subject to frequent changes.
ECRI Institute is not responsible for the quality or validity of the information presented or for any adverse
consequences of acting on such information.
When reading the charts, keep in mind that, unless otherwise noted, the list price does not reflect supplier
discounts. And although we try to indicate which features and characteristics are standard and which are not,
some may be optional, at additional cost.
For those models whose prices were supplied to us in currencies other than U.S. dollars, we have also listed the
conversion to U.S. dollars to facilitate comparison among models. However, keep in mind that exchange rates change
often.
Need to know more?
For further information about the contents of this Product Comparison, contact the HPCS Hotline at +1 (610)
825-6000, ext. 5265; +1 (610) 834-1275 (fax); or [email protected] (e-mail).
Last updated June 2009
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 11
Policy Statement
The Healthcare Product Comparison System (HPCS) is published by ECRI Institute, a nonprofit organization.
HPCS provides comprehensive information to help healthcare professionals select and purchase diagnostic and
therapeutic capital equipment more effectively in support of improved patient care.
The information in Product Comparisons comes from a number of sources: medical and biomedical
engineering literature, correspondence and discussion with manufacturers and distributors, specifications from
product literature, and ECRI Institutes Problem Reporting System. While these data are reviewed by qualified
health professionals, they have not been tested by ECRI Institutes clinical and engineering personnel and are
largely unconfirmed. The Healthcare Product Comparison System and ECRI Institute are not responsible for the
quality or validity of information derived from outside sources or for any adverse consequences of acting on such
information.
The appearance or listing of any item, or the use of a photograph thereof, in the Healthcare Product Comparison
System does not constitute the endorsement or approval of the products quality, performance, or value, or of
claims made for it by the manufacturer. The information and photographs published in Product Comparisons
appear at no charge to manufacturers.
Many of the words or model descriptions appearing in the Healthcare Product Comparison System are
proprietary names (e.g., trademarks), even though no reference to this fact may be made. The appearance of any
name without designation as proprietary should not be regarded as a representation that is not the subject of
proprietary rights.
ECRI Institute respects and is impartial to all ethical medical device companies and practices. The Healthcare
Product Comparison System accepts no advertising and has no obligations to any commercial interests. ECRI
Institute and its employees accept no royalties, gifts, finders fees, or commissions from the medical device
industry, nor do they own stock in medical device companies. Employees engage in no private consulting work
for the medical device industry.
About ECRI Institute
ECRI Institute, a nonprofit organization, dedicates itself to bringing the discipline of applied scientific research
in healthcare to uncover the best approaches to improving patient care. As pioneers in this science for over 40
years, ECRI Institute marries experience and independence with the objectivity of evidence-based research.
More than 5,000 healthcare organizations worldwide rely on ECRI Institutes expertise in patient safety
improvement, risk and quality management, healthcare processes, devices, procedures, and drug technology.
ECRI Institute is one of only a handful of organizations designated as both a Collaborating Center of the World
Health Organization and an Evidence-based Practice Center by the U.S. Agency for Healthcare Research and
Quality. For more information, visit http://www.ecri.org.
Physiologic Monitoring Systems, Cardiac Electrophysiology
12 2009 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1
BARD GALIX GE HEALTHCARE
Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT
WHERE MARKETED Worldwide Asia, Latin America Worldwide
FDA CLEARANCE Yes No Yes
CE MARK (MDD) Yes No Yes
CONFIGURATION Mobile (e.g., cart mounted), tabletop, wall mount
Cart mounted, horizontal, portable
PC-based software, fixed desk, mobile cart
Stationary, mobile
MAXIMUM CHANNELS By user requirements 16-80 16 IECG, 12 SECG Up to 128 with 224 inputs, 12-lead ECG, SpO2, up to 4 IBP, NIBP, TDCO, optional ETCO2, analog output synchronization
PLUG-IN MODULES SECG, IBP, device interface(s), IECG; others as required
SECG plus 2 IBP, V lead plus 2 IBP, IECG, STIM interface for cardiac stimulator, HLII1
Integrated 12-lead ECG, 16 HIS ECG, built-in multiprogrammable stimulator
CLab II plus amplifier, TRAM 451, ETCO2
SECG CHANNELS 12 12 12 Up to 12
IECG CHANNELS 24 Up to 72, 144 inputs 16 Up to 224
STIMULATORS INCLUDED
Optional No Yes Optional MicroPace
PROGRAMMED LEAD SWITCHING
Yes Yes Yes Yes
DISPLAY SCREEN Yes Type High-resolution, 16 colors
(1280 x 1024) High-resolution SVGA color LCD
Two or three 20" flat-panel, 1600 x 1200 resolution
Diag size, cm (in) 50.8 (20) 43.2 (17), 48.3 (19), 53.3 (21)
50.8 (20)
Data displayed Not specified Not specified Not specified User selectable Number of traces 32, user-configurable Up to 80 simultaneously 16 64/page Sweep speeds, mm/sec 5, must be user-selectable 25, 50, 100, 200, 400, 800 25, 50, 100, 200, 400, 800 5, 10, 25, 50, 100, 200, 400,
user-entered value
This is the first of five pages covering the above model(s). These specifications continue onto the next four pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 13
Product Comparison Chart
MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1
BARD GALIX GE HEALTHCARE
Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT
RECORDER Type Thermal array, laser printer High-resolution laser Any Windows-compatible
laser printer NA (optional laser printer)
Number of traces 12, user-configurable 80 16 Laser printer, based on screen layout
Paper speed, mm/sec 5, 12.5, 25, 50; variable Not specified 25, 50, 100, 200, 400, 800 NA Annotations Name, ID, date, time;
clinical values as required Yes Name, operator, date/time,
speed, ECG grid, position, comments
Name, ID, paper speed, ECG and waveform grids, text header, date/time, free drawing
Line selections Timing, millimeter grid/ruler Grid for amplitude and timing reference measurements, 12-lead ECG available on demand to print
Amplitude and time grids Laser printer, based on screen layout
INTEGRAL COMPUTER Any IBM Pentium Best available PC 2.33 GHz dual-core Xeon, Windows XP SP2 operating system
Applications Not specified Not specified Not specified Data management Yes Yes Yes Yes Statistical analysis Yes See Other Specifications No Yes, Centricity CVIS Report generator Yes Yes Microsoft Word-based MS Word-based Other None specified None specified Orders, scheduling, ADT,
charge capture, physician reporting, Holter, macros
Storage Not specified (DVD) 500 GB hard drive, CD-R/DVD long-term
Dual 80 GB HDD, 9.4 GB DVD/RAM, 144-900 GB RAID server
NETWORKING BETWEEN LABS
Yes No Windows Yes
INTERFACES Biosense Webster's CARTO XP, ablation generators and stimulators
HIS Yes Optional No Optional (ADT, orders, billing/charge capture, material management, results)
Suppliers Not specified NA Optional
This is the second of five pages covering the above model(s). These specifications continue onto the next three pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
14 2009 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1
BARD GALIX GE HEALTHCARE
Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT
Cardiology data management system
Yes Optional No Integrated with GE Centricity Cardiology CVIS, optional interface to third-party CVIS
Suppliers Not specified NA Optional Cath lab imaging system Not specified Not specified Not specified Optional, GE and third party
via DICOM National Cardiology Database
Not specified Not specified Not specified GE and third party
This is the third of five pages covering the above model(s). These specifications continue onto the next two pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 15
Product Comparison Chart
MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1
BARD GALIX GE HEALTHCARE
Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT
POWER REQUIREMENTS VAC, Hz 100/120/220/240, 50/60 110/220, 50/60 Hz 100-240, 50/60 Hz Consumption, W Not specified 500 1,800 maximum
H x W x D, cm (in) 86.3 x 74 x 160 (34 x 29 x 63)
9.6 x 30.2 x 27.2 (3.8 x 11.9 x 10.7), console without PC
82.5 x 81 x 76 (32.5 x 32.1 x 30)
WEIGHT, kg (lb) 272 (600) 2 (4.4) without PC 113 (249.2)
PURCHASE INFORMATION
Price range $50,000-150,000 $16,000-25,000 Not specified Warranty 1 year, parts, service, and
installation 18 months, parts and labor 1 year, parts and labor
Delivery time, ARO 2-4 weeks 4 weeks 45 days Service contract ~8% of purchase price Yes Optional Year first sold 2003 1998 2003, 2000 (CardioLab
2000, 4000, and 7000), 1989 (DOS-based)
Number installed USA Not specified Not specified >1,000 (combined IT
platform) Worldwide Not specified Not specified >1,500
Fiscal year January to December April to March January to December
GREEN FEATURES None specified None specified None specified
This is the fourth of five pages covering the above model(s). These specifications continue onto the next page.
Physiologic Monitoring Systems, Cardiac Electrophysiology
16 2009 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1
BARD GALIX GE HEALTHCARE
Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT
OTHER SPECIFICATIONS Windows 2000 operating system; customizable user interface; modular STAMP amplifier; template matching; T-wave subtraction; waveform analysis; data centralization (stimulator, RF generator, imaging); MS Word reports; DVD archiving; MicroPace stimulator interface; optional image capture.
Programmable and preconfigured protocols; selection of stimulation channel by software; real-time and capture screens; triggering by any channel and stimulus synchronization; automatic rate measurement; amplitude and time measurement with unlimited number of calipers; event markers; storage of long-term studies; multiprogrammable stimulator with up to 3 extrastimuli; 3 methods of tachycardia termination; backup pacing and capability to work with independent automatic increment/decrement of coupling extrastimuli intervals.
EP-only application; 3-D mapping interface option with Biosense Webster CARTO XP system; bidirectional interface; auto storage of mapping points; transfer of color snapshot and comprehensive report generation; auto control of CardioLab from CARTO XP system; CardioImage fluoroscopy image management system with up to 4 inputs of imaging; live and frozen snapshots display; centralized waveform archive with near real-time review of raw waveforms from any office PC connected to server; custom forms; interactive statistic, user-defined macros to automate tasks; full-disclosure storage of waveforms; remote modem support; dual hemo/EP capabilities; Nellcor OxySmart or Masimo SpO2; interactive administrative reporting; inventory management; billing/charge capture; online structured physician reporting.
UMDNS CODE(S) 17898 17898 17898 17898
LAST UPDATED May 2005 June 2009 June 2008
Supplier Footnotes 1These recommendations are the opinions of ECRI Institute's technology experts. ECRI Institute assumes no liability for decisions made based on this data.
Model Footnotes
Data Footnotes
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 17
Product Comparison Chart
MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP
System1 EP WorkMate Recording System
WHERE MARKETED Worldwide Worldwide Worldwide Worldwide
FDA CLEARANCE Yes Yes Yes Yes
CE MARK (MDD) Yes Yes Yes Yes
CONFIGURATION Stationary, mobile PC-based software, movable console, mobile cart, combo with Horizon SE
Cabinet, table Cart or permanent install, 2 primary monitors, slave monitors, 120 or 56 catheter inputs
MAXIMUM CHANNELS Up to 128 with 224 inputs, 12-lead ECG, SpO2, up to 4 IBP, NIBP, TDCO, optional ETCO2, analog output synchronization
32 or 64, 12 SECG, 2 IBP, 18/50 IECG
128 448 stored from up to 140 total inputs
PLUG-IN MODULES CLab II plus amplifier, TRAM 451, ETCO2
Integrated 12-lead ECG, 18/50 IECG, dual output multiprogrammable stimulator
Inputs available in integrated input box and Hemopod (12-channel ECG, HR, cycle length, 4 IBP, thermal dilution CO, SpO2, NIBP, 1 QRS sync, respiration/CO2 module)
Up to 4 pressure-monitoring channels, multimodality image acquisition, remote nurses charting station, navigation interfaces, FFT analysis
SECG CHANNELS Up to 12 12 12 12
IECG CHANNELS Up to 224 18/50 0, 32 (not available in USA), 64, or 128
56, 120
STIMULATORS INCLUDED
Optional MicroPace Dual-output multiprogrammable
Stimulator interface Yes
PROGRAMMED LEAD SWITCHING
Yes Yes Yes Yes
DISPLAY SCREEN Type Two or three 20" flat-panel,
1600 x 1200 resolution 2 high-resolution SVGA LCD flat-panel color
21" (1600 x 1200) flat-panel color (EP and combo unit)
High-resolution LCD flat-panel
Diag size, cm (in) 50.8 (20) 50.8 (20) 53.3 (21), Sensis XP configuration dependent
53.3 (21)
Data displayed User selectable HR, systolic, diastolic, mean pressure
User-configurable data parameters, including 12-lead ECG, 4 IBP, NIBP, SpO2, respiration, HR, cycle length, ablation data
SECG, IECG pressure tracings, heart rate, ablation parameters
Number of traces 64/page 2 panels, 32 traces each Unlimited, user configured, theoretical end at 80 traces
32/page, 6 pages
Sweep speeds, mm/sec 5, 10, 20, 25, 50, 100, 200, 400, user-entered value
25, 50, 100, 150, 200, 300 12.5, 25, 50, 100, 200, 400 (real-time and review mode)
10-600
This is the first of five pages covering the above model(s). These specifications continue onto the next four pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
18 2009 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP
System1 EP WorkMate Recording System
RECORDER Type NA (optional laser printer) Any network laser printer Laser printer
(network/parallel port) LaserJet
Number of traces Laser printer, based on screen layout
32 Unlimited, user configured 64
Paper speed, mm/sec NA 25, 50, 100, 150, 200, 250, 300
12.5-400 real-time mode 10-300
Annotations Name, ID, paper speed, ECG and waveform grids, text header, date/time, free drawing
Name, operator, date/time, speed, ECG grid, position, comments
Yes Annotations are user-configurable; text, characters, or numbers
Line selections Laser printer, based on screen layout
Amplitude and time grids Millimeter grid, raster is user configurable
Lines or grid format
INTEGRAL COMPUTER 2.33 GHz dual-core Xeon, Windows XP SP2 operating system
2.8 GHz Pentium 4 Pentium 4, 3.0 GHz Dual-core Xeon processors
Applications Not specified Not specified Microsoft XP Professional on ACQ and workstations, Microsoft Server 2003 on high-end server
Not specified
Data management Yes Optional Yes (with information system)
EP monitoring and stimulation
Statistical analysis Yes, Centricity CVIS Yes (with optional data management)
Yes (with information system)
No
Report generator MS Word-based MS Word-based MS Word-based online and offline
Yes
Other Orders, scheduling, ADT, charge capture, physician reporting, Holter, ST segment, macros
None specified Dedicated AXIOM Sensis server
None specified
Storage Dual 80 GB HDD, 9.4 GB DVD/RAM, 144-900 GB RAID server
80 GB hard drive, optional CD-R/DVD-RW
160 GB hard disk (RAID 5 available)
160 GB RAID redundant backup
NETWORKING BETWEEN LABS
Yes Windows 2000 Ethernet, TCP/IP, DICOM, HL7, open architecture
Optional
INTERFACES Biosense Webster's CARTO XP, ablation generators and stimulators
Not specified Not specified Biosense Webster CARTO system, Phillips Allura system, most approved ablation generators, HL7, LUMEDX, Xcelera
HIS Optional (ADT, orders, billing/charge capture, material management, results)
Optional DICOM, HL7, ASCII, bidirectional
Yes
Suppliers Optional Not specified; standard HL7 protocol for ADT, HL7 results or XML, RTF, and FTP
Proprietary Optional
This is the second of five pages covering the above model(s). These specifications continue onto the next three pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 19
Product Comparison Chart
MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP System1 EP WorkMate Recording
System
Cardiology data management system
Integrated with GE Centricity Cardiology CVIS, optional interface to third-party CVIS
Optional Yes Yes
Suppliers Optional Mennen Medical All Optional Cath lab imaging system
Optional, GE and third party via DICOM
Not specified Siemens proprietary Not specified
National Cardiology Database
GE and third party Not specified ACC-NCDR modules (data collected for submission to third party), Heartlab/LUMEDX/Apollo/AXIS/Goodroe
Not specified
This is the third of five pages covering the above model(s). These specifications continue onto the next two pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
20 2009 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP
System1 EP WorkMate Recording System
POWER REQUIREMENTS VAC, Hz 100-240, 50/60 Hz 110/220, 50/60 Hz 100-120, 230-240, 50/60 Hz 85-265, 47-63 Consumption, W 1,800 maximum 400 1,200 VA maximum 1,500
H x W x D, cm (in) 82.5 x 81 x 76 (32.5 x 32.1 x 30)
74.5 x [80.5-96] x 82 (29.3 x [31.7-37.8] x 32.3) console with monitor; 103 x 60 x 60 (40.6 x 23.6 x 23.6) mobile cart
72 x 120 x 80 (28.3 x 47.2 x 31.5) with cabinet
183 x 121.9 x 91.4 (72 x 48 x 36)
WEIGHT, kg (lb) 113 (249.2) 80 (176) console without transformer, 50 (110) mobile cart without transformer, 29 (63.9) transformer
~85 (187.4), including monitors
272 (600)
PURCHASE INFORMATION
Price range Not specified $66,750-82,000 Not specified $208,215-238,140 Warranty 1 year, parts and labor 1 year, parts and labor 1 year 1 year, parts and labor Delivery time, ARO 45 days 90 days FOB 6-8 weeks 30-60 days Service contract Optional Yes Optional 1-year increments, full
service and biomedical second call-in
Year first sold 2003, 2000 (ComboLab 2000 and 7000), 1989 (DOS-based)
2003 2003 1996
Number installed USA >1,000 (combined IT
platform) Not specified >700 700
Worldwide >1,500 Not specified >1,000 400 Fiscal year January to December January to December October to September January to December
GREEN FEATURES None specified None specified None specified None specified
This is the fourth of five pages covering the above model(s). These specifications continue onto the next page.
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 21
Product Comparison Chart
MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP
System1 EP WorkMate Recording System
OTHER SPECIFICATIONS EP-only application; 3-D mapping interface option with Biosense Webster CARTO XP system; bidirectional interface; auto storage of mapping points; transfer of color snapshot and comprehensive report generation; auto control of CardioLab from CARTO XP system; CardioImage fluoroscopy image management system with up to 4 inputs of imaging; live and frozen snapshots display; centralized waveform archive with near real-time review of raw waveforms from any office PC connected to server; custom forms; interactive statistic, user-defined macros to automate tasks; full-disclosure storage of waveforms; remote modem support; Nellcor OxySmart or Masimo SpO2; interactive administrative reporting; inventory management; billing/charge capture; online structured physician reporting.
12-lead ECG and IECG interface for third-party mapping systems; interface to RF ablation systems.
Multitasking, multiuser system; DICOM, HL7, ASCII communication optional. Meets requirements of CAN/CSA C22.2 No. 601-1M90; EN 865; IEC 60601-1, 60601-1-2, 60601-1-25, 60601-2-30, 60601-2-34, and 60601-2-49; and UL 2601-1.
None specified.
UMDNS CODE(S) 17898 17898 17898 17898
LAST UPDATED June 2008 June 2009 June 2009 June 2009
Supplier Footnotes 1St Jude Medical recently purchased EP MedSystems.
Model Footnotes 1Hemo, EP, and combo system.
Data Footnotes
Physiologic Monitoring Systems, Cardiac Electrophysiology
22 2009 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL TEB SP12/32
WHERE MARKETED Mexico, South America
FDA CLEARANCE No
CE MARK (MDD) No
CONFIGURATION Cart mounted
MAXIMUM CHANNELS 38
PLUG-IN MODULES None, all functions are integral; 12 SECG, 18 IECG, 4 IBP, 1 thermodilution, 4 external inputs, 3 outputs
SECG CHANNELS 12
IECG CHANNELS 18
STIMULATORS INCLUDED
Yes
PROGRAMMED LEAD SWITCHING
Yes
DISPLAY SCREEN Type 2 high-resolution color LCDs Diag size, cm (in) 43 (17) Data displayed Not specified Number of traces 32 Sweep speeds, mm/sec 12.5, 25, 50, 100, 200, 300
This is the first of five pages covering the above model(s). These specifications continue onto the next four pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 23
Product Comparison Chart
MODEL TEB SP12/32
RECORDER Type Laser printer Number of traces 32 Paper speed, mm/sec 12.5, 25, 50, 200, 300, 400 Annotations Date, time, lead, patient ID,
gain, paper speed Line selections Time grid
INTEGRAL COMPUTER AMD SEMPROM 64 3200 Applications Not specified
Data management No Statistical analysis No Report generator Yes Other No
Storage 80 GB
NETWORKING BETWEEN LABS
No
INTERFACES HIS No
Suppliers NA
This is the second of five pages covering the above model(s). These specifications continue onto the next three pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
24 2009 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL TEB SP12/32
Cardiology data management system
No
Suppliers NA Cath lab imaging system Not specified National Cardiology Database
Not specified
This is the third of five pages covering the above model(s). These specifications continue onto the next two pages.
Physiologic Monitoring Systems, Cardiac Electrophysiology
2009 ECRI Institute. All Rights Reserved 25
Product Comparison Chart
MODEL TEB SP12/32
POWER REQUIREMENTS VAC, Hz 110/220, 50/60 Consumption, W 380
H x W x D, cm (in) 140 x 90 x 63 (55 x 35 x 25)
WEIGHT, kg (lb) 82 (180)
PURCHASE INFORMATION
Price range $44,000 Warranty 1 year, parts and labor Delivery time, ARO 4-6 weeks Service contract No Year first sold 1994 Number installed
USA NA Worldwide 452
Fiscal year January to December
GREEN FEATURES None specified
This is the fourth of five pages covering the above model(s). These specifications continue onto the next page.
Physiologic Monitoring Systems, Cardiac Electrophysiology
26 2009 ECRI Institute. All Rights Reserved.
Product Comparison Chart
MODEL TEB SP12/32
OTHER SPECIFICATIONS Continuous storage in hard disk; DVD recorder included to save stored data at end of procedure; USB port; Windows-based software included on PC. Registered by the Brazilian Ministry of Health. TEB is ISO 9001 and ISO 13485 certified; IEC 601 certified.
UMDNS CODE(S) 17898
LAST UPDATED June 2009
Supplier Footnotes
Model Footnotes
Data Footnotes
Scope of this Product ComparisonPurposePrinciples of operationSecurityExternal databases and registries
Reported problemsPurchase considerationsECRI Institute recommendationsOther considerationsCost containmentPresent Value/Life-Cycle Cost Analysis
Stage of developmentBibliographySupplier informationPolicy StatementAbout ECRI InstituteProduct Comparison Chart