26
5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA Tel +1 (610) 825-6000 Fax +1 (610) 834-1275 Web www.ecri.org E-mail [email protected] UMDNS Information This Product Comparison covers the following device term and product code as listed in ECRI Institute’s Universal Medical Device Nomenclature System™ (UMDNS™): Monitoring Systems, Physiologic, Cardiac Electrophysiology [17-898] Physiologic Monitoring Systems, Cardiac Electrophysiology Scope of this Product Comparison This Product Comparison covers physiologic monitoring and recording systems used for cardiac electrophysiology studies (EPSs) and associated therapies. Included are dedicated units, as well as cardiac catheterization physiologic monitoring systems that have EPS capabilities. For information on other physiologic monitoring systems, see the following Product Comparisons: Physiologic Monitoring Systems, Acute Care; Neonatal; ECG Monitors; Monitors, Central Station Physiologic Monitoring Systems, Cardiac Catheterization Physiologic Monitoring Systems, Stress Exercise, Cardiac; Pulmonary Physiologic Monitoring Systems, Telemetric; ECG Monitors, Telemetric; Monitors, Central Station These devices are also called: cardiac electrophysiology monitoring systems, cardiac monitors, electrocardiographic monitors, operating room monitors, and OR monitors. Purpose Cardiac EPSs are used, in conjunction with cardiac catheterization, for patients with arrhythmia or conduction disorders. During an EPS, several diagnostic tests and therapeutic treatments are performed, including analysis of the function of various components of the atrioventricular (AV) conduction system under certain stressors, determination of the site of interventions, and assessment of sinus node function. EPSs also include induction and termination of ventricular tachycardias by means of an electrical stimulator in a controlled setting for diagnostic purposes or to evaluate responses to antiarrhythmia drug therapies. Other tests currently performed in EPS laboratories include evaluation of automatic implantable cardioverter-defibrillators (AICDs) to ensure proper functioning and tilt tests to diagnose presyncope. EPSs are indicated for patients with sinus node dysfunction, AV block, and supraventricular and ventricular tachycardias. They are also performed on patients undergoing pharmacologic, electrical, or ablative treatment of arrhythmias; survivors of cardiac arrest; and patients with Wolff-Parkinson-White syndrome (anomalous AV excitation), unexplained syncope, or heart palpitations. Patients being considered for AICDs undergo EPSs before implantation to determine, among other things, the cause of the arrhythmia, the optimal site for shock delivery (i.e., lead placement),

Physiologic Monitoring Systems, Cardiac Electrophysiology

Embed Size (px)

DESCRIPTION

Physiologic Monitoring Systems, Cardiac Electrophysiology

Citation preview

  • 5200 Butler Pike, Plymouth Meeting, PA 19462-1298, USA Tel +1 (610) 825-6000 Fax +1 (610) 834-1275 Web www.ecri.org E-mail [email protected]

    UMDNS Information

    This Product Comparison covers the following device term and product code as listed in ECRI Institutes Universal Medical Device Nomenclature System (UMDNS):

    Monitoring Systems, Physiologic, Cardiac Electrophysiology [17-898]

    Physiologic Monitoring Systems, Cardiac Electrophysiology

    Scope of this Product Comparison

    This Product Comparison covers physiologic monitoring and recording systems used for cardiac

    electrophysiology studies (EPSs) and associated therapies. Included are dedicated units, as well as cardiac

    catheterization physiologic monitoring systems that have EPS capabilities. For information on other physiologic

    monitoring systems, see the following Product Comparisons:

    Physiologic Monitoring Systems, Acute Care; Neonatal;

    ECG Monitors; Monitors, Central Station

    Physiologic Monitoring Systems, Cardiac Catheterization

    Physiologic Monitoring Systems, Stress Exercise, Cardiac;

    Pulmonary

    Physiologic Monitoring Systems, Telemetric; ECG

    Monitors, Telemetric; Monitors, Central Station

    These devices are also called: cardiac electrophysiology

    monitoring systems, cardiac monitors, electrocardiographic

    monitors, operating room monitors, and OR monitors.

    Purpose

    Cardiac EPSs are used, in conjunction with cardiac

    catheterization, for patients with arrhythmia or conduction

    disorders. During an EPS, several diagnostic tests and therapeutic treatments are performed, including analysis

    of the function of various components of the atrioventricular (AV) conduction system under certain stressors,

    determination of the site of interventions, and assessment of sinus node function. EPSs also include induction and

    termination of ventricular tachycardias by means of an electrical stimulator in a controlled setting for diagnostic

    purposes or to evaluate responses to antiarrhythmia drug therapies. Other tests currently performed in EPS

    laboratories include evaluation of automatic implantable cardioverter-defibrillators (AICDs) to ensure proper

    functioning and tilt tests to diagnose presyncope.

    EPSs are indicated for patients with sinus node dysfunction, AV block, and supraventricular and ventricular

    tachycardias. They are also performed on patients undergoing

    pharmacologic, electrical, or ablative treatment of arrhythmias;

    survivors of cardiac arrest; and patients with Wolff-Parkinson-White

    syndrome (anomalous AV excitation), unexplained syncope, or

    heart palpitations.

    Patients being considered for AICDs undergo EPSs before

    implantation to determine, among other things, the cause of the

    arrhythmia, the optimal site for shock delivery (i.e., lead placement),

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2 2009 ECRI Institute. All Rights Reserved.

    and the most effective sequence of stimulation to achieve normal cardiac rhythm. (For more information, see the

    Product Comparison titled Pacemakers, Cardiac, Implantable; Resynchronization.)

    Principles of operation

    EPSs are typically performed in cardiac

    catheterization laboratories (cath labs) or dedicated

    fluoroscopy suites. Before an EPS can be

    performed, the patient must first undergo cardiac

    catheterization, which allows endocardial catheter

    electrodes capable of both electrical stimulation

    and recording to be guided into the heart intra-

    arterially or intravenously using fluoroscopy. (For

    a detailed discussion of cardiac catheterization, see

    the Product Comparisons titled Physiologic

    Monitoring Systems, Cardiac Catheterization and

    Radiographic/Fluoroscopic Systems,

    Angiographic/Interventional; Cardiovascular.)

    Before catheterization, a diagnostic 12-lead electrocardiogram (ECG) is usually performed to obtain baseline

    values. This preliminary study can also disclose unsuspected abnormalities.

    During catheterization, the electrophysiology (EP) monitoring/recording system amplifies, conditions, and

    records signals obtained from the intracardiac electrodes as well as signals from surface electrodes and blood

    pressure transducers. These monitoring systems also aid in the placement of catheters and help the physician

    observe any changes resulting from treatment during catheterization.

    A basic EP monitoring and recording system consists of a horizontal or vertical console with a monitor, a chart

    recorder to print data and waveforms, a data recording system (e.g., magnetic or optical disk), and plug-in

    modules to perform various amplification and conditioning functions. The recorder should be able to record 4 to

    32 ECG signals, and the recording amplifiers should have high- and low-frequency filters for recording signals

    from either surface or intracardiac electrodes. The recorder should also be able to respond to frequencies of up to

    500 Hz. Most systems have between 6 and 38 channels (some may have up to 192 channels) and input switching

    systems for selecting various combinations of electrodes. Waveforms are displayed on the screen and are

    recorded correspondingly; most systems allow waveforms to be frozen on the monitors screen. Some can show

    test protocols, results, and calculations on a separate video display unit, while others can annotate the chart

    printouts, which can also feature a variety of reference lines, such as grids and event markers. Additional

    workstations are available for analysis and manipulation of data and waveforms for data management purposes.

    Final reports can be generated at the workstation, and waveforms and data can be stored on digital media (e.g.,

    magnetic tape, optical disk).

    The number of intracardiac electrodes used for EPSs varies. Generally, two to four catheters are used; the

    number is determined by the complexity of the study being performed. Points of placement include high in the

    right atrium, in the region of the sinus node; at the AV junction; in the apex of the right ventricle; across the

    tricuspid valve ring to record His bundle activity; and in the coronary sinus. To perform tricuspid valve analysis,

    the electrode at the catheter tip is drawn through the valve to detect a signal change. Catheters commonly

    measure from 4 to 8 French (Fr), and each catheter can have from two to eight electrodes with interelectrode

    distances of 1 to 10 mm. Externally, three to six leads may be used. During an open-chest procedure, the surgeon

    may also touch the myocardium with a wand-like device that has electrodes at the tip.

    Electrical stimulation is performed with a programmable electrical stimulator that has an isolated output. (EP

    monitoring and recording systems may not include stimulators, which would need to be purchased from a

    stimulator supplier.) Stimulation is introduced during sinus rhythm or atrial or ventricular pacingfor instance,

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 3

    to induce ventricular tachycardias, the characteristics of which can guide AICD programming or reveal the

    patients response to antiarrhythmic or other drugs. The stimulator must be able to deliver at least two precisely

    timed impulses during spontaneous and paced rhythms with cycle lengths of 150 to 1,500 msec, as well as

    multiple extra stimuli from two or more sites simultaneously. It is also advisable for the stimulator to have a

    safety switch to prevent pacing at

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    4 2009 ECRI Institute. All Rights Reserved.

    Reported problems

    Most problems associated with electrophysiologic monitoring and recording involve inserting, advancing, and

    maintaining the position of the catheter. These problems include hematomas, infection, vascular damage,

    thrombosis, mechanically induced electrical irritation and arrhythmias, catheter clotting, and balloon rupture.

    Other common problems are poor electrode preparation and placement.

    The hazard of electric shock to patients must be considered in any room where electrical equipment is used

    particularly in settings where a catheter provides a direct pathway for current to travel to the conduction system

    of the heart. A safe power-distribution system and its associated grounding are essential, as are isolated electrical

    inputs on all equipment that may come in contact with the patient (along with periodic inspection and

    measurement of interequipment leakage currents). In the United States, the electrical wiring and fittings must be

    installed in accordance with the U.S. National Electrical Code and standards for safe use of electricity in

    healthcare facilities.

    Damaged muscle tissue interferes with and slows the ventricular depolarization wave, causing reentrant

    currents, which may trigger sustained ventricular tachycardia. This condition can degenerate into ventricular

    fibrillation or death.

    Purchase considerations

    ECRI Institute recommendations

    Included in the accompanying comparison chart are ECRI Institutes recommendations for minimum

    performance requirements for cardiac EP systems.

    EP systems may be mobile (e.g., cart mounted), tabletop, or wall mounted. The maximum channels (i.e., the

    maximum number of modules that can be plugged into the unit and operated at one time) should be selected by

    the physician based on procedural requirements. The system should have at least 12 surface electrocardiogram

    (SECG) channels and 24 intracardiac electrocardiogram (IECG) channels, as well as programmed lead switching.

    The EP lab monitoring and recording systems should have a console with high-resolution color monitors

    suitable for display of waveforms and data; an integrated computer with data-processing software and recording

    media (e.g., DVD, optical disk); a chart recorder with printing capabilities; plug-in modules that acquire vital

    signs and procedure monitoring functions; and a remote amplifier unit to process and condition cardiac electrode

    signals.

    The EP system should be able to annotate chart printouts using a variety of reference lines (e.g., grids, event

    markers), be able to interface with RF-generating devices and cardiac cath lab imaging systems, and have

    multitasking software with user-configurable parameters that allows real-time signal analysis.

    The EP lab system should be capable of networking and interfacing with a hospital information system; such

    networking should be bidirectional and Health Level 7 (HL7) compliant. The EP lab system should also have the

    ability to interface with a national cardiology database (e.g., American College of Cardiology/National

    Cardiovascular Data Registry [ACC/NCDR]) and cardiology data management system.

    Other considerations

    Cardiac EP systems are typically customized to meet the facilitys needs. A basic system is configured with

    amplifiers and filters for signal acquisition and processing, a display screen, and a recorder. It is possible to use

    preexisting amplifiers and interface them with a corresponding display screen and recorder.

    Clinicians should determine the maximum number of ECG channels desired; the number of available ECG

    channels varies among systems. Varying sweep speeds up to 100 mm/sec must be available when performing

    EPSs. Display screens vary in the number of waveforms that are displayed; correspondingly, the recorders vary

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 5

    in the number of traces recorded. Some EP systems do not offer recorders as standard equipment because all the

    waveforms can be retrieved from storage and viewed onscreen for manipulation. However, if a recorder is

    required, it can be interfaced with these systems. Semiautomated or automated input switching systems allow for

    the selection of various electrode combinations. Some systems allow the addition of amplifiers as needed.

    Monitors used during EPSs are modular, allowing additional capabilities to be added later as the needs of the

    lab change. In addition, although each manufacturers system is configured differently, most allow some

    flexibility in adapting their configurations to the particular needs of the hospital. Therefore, in choosing a system,

    it is important that the hospital consider its present and future needs, space limitations, and patient volume.

    The number of channels provided by the recording system should correspond to the institutions applications.

    Current systems vary in the type of recorder used (i.e., thermal, fiberoptic, ink-jet, or laser).

    Data communications and data storage capabilities vary among systems. Communications capabilities can

    provide links to other EP labs, central workstations, personal computers, or hospital data management systems.

    Storage capabilities include various hard disk capacities and can also include a variety of digital media for

    archival purposes and easy retrieval of data. Hospitals should consider their communications and storage needs

    before making a purchase decision. EP monitoring and recording systems also come with a variety of analysis

    and data management software packages, which allow user-programmable or customized report generation.

    Stimulators are not offered as standard components by most EP monitoring and recording system

    manufacturers and must be purchased from a stimulator supplier. This must be factored into the cost of the EP

    system.

    Facilities should also consider whether they require separate rooms for EPSs only or catheterization/EPSs.

    When planning, the time required for each test, type and number of tests, and patient volume should be

    considered. Some cardiac cath lab monitoring systems are also used for EPSs. For more information on equipment

    planning, facilities should contact ECRI Institutes Applied Solutions Group (ASG).

    Cost containment

    Because EP is a dynamic technology, manufacturers frequently make software changes to enhance

    performance or to meet current clinical needs; therefore, facilities should negotiate for software upgrades and

    enhancements for no cost for the life of the equipment. Service contract costs should be specified in all purchase

    contracts; they are usually negotiable. If a hospitals clinical engineering department plans to service the

    equipment after the warranty period, the cost of service training should be included in the purchase contract, and

    the hospital should negotiate for an inventory of frequently needed parts to be placed in the facility on

    consignment. All clinical training should be included at no charge.

    A life-cycle cost (LCC) analysis can be used to compare high-cost alternatives and/or to determine the positive

    or negative economic value of a single alternative. For example, hospitals can use LCC analysis techniques to

    examine the cost-effectiveness of leasing or renting equipment versus purchasing the equipment outright.

    Because it examines the cash-flow impact of initial acquisition costs and operating costs over a period of time,

    LCC analysis is most useful for comparing alternatives with different cash flows and for revealing the total costs

    of equipment ownership. One LCC techniquepresent value (PV) analysisis especially useful because it

    accounts for inflation and for the time value of money (i.e., money received today is worth more than money

    received at a later date). Conducting a PV/LCC analysis often demonstrates that the cost of ownership includes

    more than just the initial acquisition cost and that a small increase in initial acquisition cost may produce

    significant savings in operating costs. The PV is calculated using the annual cash flow, the dollar discount factor,

    and the lifetime of the equipment (in years) in a mathematical equation.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    6 2009 ECRI Institute. All Rights Reserved.

    The following represents a sample seven-year PV/LCC analysis for an EP system.

    Present Value/Life-Cycle Cost Analysis

    Assumptions

    Dollar discount factor is 6%

    Inflation rate is 6% for a full-service contract

    Service contract is based on 7% of the purchase price

    Capital costs

    Basic system = $105,000

    Additional slave monitor = $4,000

    Fluoroscopy image storage = $10,000

    Total Capital Costs = $119,000

    Operating costs

    Service contract, years 2 through 7 = $8,330/year

    Total Operating Costs = $0 in year 1; $8,330/year for years 2 through 7

    PV = ($168,980)

    As illustrated by the above sample PV/LCC analysis, the initial acquisition cost is only a fraction of the total

    cost of operation over seven years. Therefore, before making a purchase decision based solely on the acquisition

    cost, buyers should consider operating costs over the lifetime of the equipment.

    Other costs not included in the above analysis, but important to consider before making a final purchasing

    decision, include the costs of purchasing and operating EP imaging equipment. The fluoroscopic imaging system

    used during the catheterization procedure should be, at minimum, a single-plane configuration with either a

    tilting table or a C-arm. A biplane configuration is preferable and is often necessary for larger patients.

    Maintenance costs for the imaging equipment (e.g., x-ray tube replacement) must also be considered.

    EPSs can be performed in a cardiac cath lab, a dedicated or mobile fluoroscopy suite, or a dedicated EP

    laboratory. Because an EPS can take three to six hours or more, hospitals should analyze an EP laboratorys

    impact on existing cardiac catheterization or fluoroscopy services. Scheduling problems in cath labs may result

    due to increasing numbers of EPSs being performed. As a result, hospitals may consider upgrading or adding a

    new lab facility dedicated to EPSs. Installing a dedicated EP lab can cost between $300,000 and $500,000, while a

    new cath lab will cost twice that. Depending on laboratory setup and utilization issues, hospitals must choose

    between a dedicated EP system or a cath lab system with an EP option.

    For further information on PV/LCC analysis, customized analyses, and purchase decision support, readers

    should contact ECRI Institutes SELECTplus Group.

    Stage of development

    Since the late 1960s, cardiac EPSs have come into wide clinical acceptance for evaluating cardiac arrhythmias.

    Over time, there has been increased computerization of the cath labs in which EPSs are typically performed. This

    enables automated calculations and storage of waveforms and data on digital media.

    Bibliography

    Bevilacqua LM, Berul CL. Advances in pediatric electrophysiology. Curr Opin Pediatr 2004 Oct;16(5):494-9.

    Darling EJ. Overview of cardiac electrophysiologic testing. Crit Care Nurs Clin North Am 1994 Mar;6(1):1-13.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 7

    Determinants of predicted efficacy of antiarrhythmic drugs in the electrophysiologic study versus

    electrocardiographic monitoring trial. The ESVEM investigators. Circulation 1993 Feb;87(3):323-9.

    Jalife J. Basic cardiac electrophysiology for the clinician. 2nd ed. Hoboken (NJ):Wiley-Blackwell; 2009.

    Mason JW. A comparison of electrophysiologic testing with Holter monitoring to predict antiarrhythmic-drug

    efficacy for ventricular tachyarrhythmias. N Engl J Med 1993 Aug 12;329(7):445-51.

    Scheinman MM, Keung E. The year in review of clinical cardiac electrophysiology. J Am Coll Cardiol 2008 May

    27;51(21):2075-81.

    Strong growth in EP boosts cardiac rhythm market. Clinica 1996 Feb 5;691:16.

    Ward DE, Camm AJ. Dangerous ventricular arrhythmiascan we predict drug efficacy? [letter]. N Engl J Med

    1993 Aug 12;329(7):498-9.

    Willems S, Weiss C, Ventura R, et al. Catheter ablation of atrial flutter guided by electroanatomic mapping

    (CARTO): a randomized comparison to the conventional approach. J Cardiovasc Electrophysiol 2000

    Nov;11(11):1223-30.

    Zickler P. Electrophysiology. Med Imaging 1995 Jul;10(7):44-51.

    Supplier information

    BARD

    Angiomed GmbH & Co Medizintechnik KG Sub C R Bard Inc [332055]

    Wachhausstrasse 6

    Karlsruhe D-76227

    Germany

    Phone: 49 (721) 94450 Fax: 49 (721) 9445111

    Internet: http://www.crbard.com

    E-mail: [email protected]

    Bard China Sub C R Bard Inc [343164]

    Swissotel Beijing Hong Kong Macau Center Dongsishitiao

    Beijing 100027

    People's Republic of China

    Phone: 86 (10) 65014225

    Bard Electrophysiology Div CR Bard Inc [435644]

    55 Technology Dr

    Lowell, MA 01851

    Phone: (978) 441-6202, (800) 824-8724 Fax: (978) 323-2260

    Internet: http://www.bardep.com

    Bard Nederland bv Sub C R Bard Inc [343215]

    Weverstedehof 10

    Nieuwegein NL-3430 BD

    The Netherlands

    Phone: 31 (30) 6000570 Fax: 31 (30) 6022443

    Internet: http://www.crbard.com

    E-mail: [email protected]

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    8 2009 ECRI Institute. All Rights Reserved.

    GALIX

    Galix Biomedical Instrumentation Inc [176265]

    2555 Collins Ave Suite C-5

    Miami Beach, FL 33140

    Phone: (305) 534-5905 Fax: (305) 534-8222

    Internet: http://www.galix-gbi.com

    E-mail: [email protected]

    GE HEALTHCARE

    GE Healthcare Asia (Japan) [300443]

    4-7-127 Asahigaoka Hino-shi

    Tokyo 191-8503

    Japan

    Phone: 81 (3) 425826820 Fax: 81 (3) 425826830

    Internet: http://japan.gehealthcare.com/cwcjapan

    E-mail: [email protected]

    GE Healthcare France [171319]

    11 avenue Morane Saulnier

    Velizy Cedex F-78457

    France

    Phone: 33 (6) 07453493 Fax: 33 (1) 34495202

    Internet: http://www.gehealthcare.com

    E-mail: [email protected]

    GE Healthcare Technologies Diagnostic Cardiology [441472]

    8200 W Tower Ave PO Box 414

    Milwaukee, WI 53223-3219

    Phone: (414) 355-5000, (800) 643-6439 Fax: (414) 355-3790

    Internet: http://www.gehealthcare.com

    GE Healthcare UK Ltd [401906]

    Amersham Place

    Little Chalfont HP 94SP

    England

    Phone: 44 (870) 6061920 Fax: 44 (1494) 544350

    Internet: http://www.gehealthcareeurope.com

    MENNEN MEDICAL

    Mennen Medical Corp [101498]

    950 Industrial Blvd

    Southampton, PA 18966

    Phone: (215) 259-1020, (800) 223-2201 Fax: (215) 675-6212

    Internet: http://www.mennenmedical.com

    E-mail: [email protected]

    Mennen Medical Ltd [171234]

    4 Ha-Yarden Street PO Box 102

    Rehovot IL-76100

    Israel

    Phone: 972 (8) 9323333 Fax: 972 (8) 9328510

    Internet: http://www.mennenmedical.com

    E-mail: [email protected]

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 9

    SIEMENS MEDICAL

    Siemens Canada Ltd [174735]

    2185 Derry Rd W

    Mississauga, ON L5N 7A6

    Canada

    Phone: (905) 819-8000, (888) 303-3353 Fax: (905) 819-5777

    Internet: http://www.siemens.ca

    E-mail: [email protected]

    Siemens Medical Solutions USA Inc Health Service Corp [399199]

    51 Valley Stream Pkwy

    Malvern, PA 19355

    Phone: (610) 219-6300, (888) 826-9702 Fax: (610) 219-3124

    Internet: http://www.siemensmedical.com

    E-mail: [email protected]

    Siemens SA de CV [339105]

    Poniente 116 No 590

    Cd de Mexico 02300

    Mexico

    Phone: 52 (5) 3282000 Fax: 52 (5) 3282017

    Internet: http://www.siemens.com.mx

    E-mail: [email protected]

    ST JUDE MEDICAL

    St Jude Medical UK Ltd [301608]

    Stratfrod Business & Technology Parl Banbury Road

    Stratford upon Avon CV37 7GY

    England

    Phone: 44 (1789) 207600 Fax: 44 (1789) 207601

    Internet: http://www.sjm.com

    E-mail: [email protected]

    St Jude Medical Japan KK [382507]

    F-Nisei Ebisu Building 6/Fl 3-16-3 Higashi Shibuya-ku

    Tokyo 150-0011

    Japan

    Phone: 81 (3) 57780702 Fax: 81 (3) 57780704

    Internet: http://www.sjm.com

    E-mail: [email protected]

    St Jude Medical Inc Cardiology/Atrial Fibrillation Div [452237]

    14901 DeVeau Pl

    Minnetonka, MN 55345-2126

    Phone: (952) 933-4700, (800) 328-3873 Fax: (952) 933-0307

    Internet: http://www.sjm.com

    E-mail: [email protected]

    TEB

    TEB (Tecnologia Eletronica Brasileira Ltda) [174429]

    Avenida Diederichsen 1057-1059 Vila Guarani

    Sao Paulo-SP 04310-000

    Brazil

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    10 2009 ECRI Institute. All Rights Reserved.

    Phone: 55 (11) 50178555 Fax: 55 (11) 50176472

    Internet: http://www.teb.com.br

    E-mail: [email protected]

    Note: The data in the charts derive from suppliers specifications and have not been verified through

    independent testing by ECRI Institute or any other agency. Because test methods vary, different products

    specifications are not always comparable. Moreover, products and specifications are subject to frequent changes.

    ECRI Institute is not responsible for the quality or validity of the information presented or for any adverse

    consequences of acting on such information.

    When reading the charts, keep in mind that, unless otherwise noted, the list price does not reflect supplier

    discounts. And although we try to indicate which features and characteristics are standard and which are not,

    some may be optional, at additional cost.

    For those models whose prices were supplied to us in currencies other than U.S. dollars, we have also listed the

    conversion to U.S. dollars to facilitate comparison among models. However, keep in mind that exchange rates change

    often.

    Need to know more?

    For further information about the contents of this Product Comparison, contact the HPCS Hotline at +1 (610)

    825-6000, ext. 5265; +1 (610) 834-1275 (fax); or [email protected] (e-mail).

    Last updated June 2009

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 11

    Policy Statement

    The Healthcare Product Comparison System (HPCS) is published by ECRI Institute, a nonprofit organization.

    HPCS provides comprehensive information to help healthcare professionals select and purchase diagnostic and

    therapeutic capital equipment more effectively in support of improved patient care.

    The information in Product Comparisons comes from a number of sources: medical and biomedical

    engineering literature, correspondence and discussion with manufacturers and distributors, specifications from

    product literature, and ECRI Institutes Problem Reporting System. While these data are reviewed by qualified

    health professionals, they have not been tested by ECRI Institutes clinical and engineering personnel and are

    largely unconfirmed. The Healthcare Product Comparison System and ECRI Institute are not responsible for the

    quality or validity of information derived from outside sources or for any adverse consequences of acting on such

    information.

    The appearance or listing of any item, or the use of a photograph thereof, in the Healthcare Product Comparison

    System does not constitute the endorsement or approval of the products quality, performance, or value, or of

    claims made for it by the manufacturer. The information and photographs published in Product Comparisons

    appear at no charge to manufacturers.

    Many of the words or model descriptions appearing in the Healthcare Product Comparison System are

    proprietary names (e.g., trademarks), even though no reference to this fact may be made. The appearance of any

    name without designation as proprietary should not be regarded as a representation that is not the subject of

    proprietary rights.

    ECRI Institute respects and is impartial to all ethical medical device companies and practices. The Healthcare

    Product Comparison System accepts no advertising and has no obligations to any commercial interests. ECRI

    Institute and its employees accept no royalties, gifts, finders fees, or commissions from the medical device

    industry, nor do they own stock in medical device companies. Employees engage in no private consulting work

    for the medical device industry.

    About ECRI Institute

    ECRI Institute, a nonprofit organization, dedicates itself to bringing the discipline of applied scientific research

    in healthcare to uncover the best approaches to improving patient care. As pioneers in this science for over 40

    years, ECRI Institute marries experience and independence with the objectivity of evidence-based research.

    More than 5,000 healthcare organizations worldwide rely on ECRI Institutes expertise in patient safety

    improvement, risk and quality management, healthcare processes, devices, procedures, and drug technology.

    ECRI Institute is one of only a handful of organizations designated as both a Collaborating Center of the World

    Health Organization and an Evidence-based Practice Center by the U.S. Agency for Healthcare Research and

    Quality. For more information, visit http://www.ecri.org.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    12 2009 ECRI Institute. All Rights Reserved.

    Product Comparison Chart

    MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1

    BARD GALIX GE HEALTHCARE

    Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT

    WHERE MARKETED Worldwide Asia, Latin America Worldwide

    FDA CLEARANCE Yes No Yes

    CE MARK (MDD) Yes No Yes

    CONFIGURATION Mobile (e.g., cart mounted), tabletop, wall mount

    Cart mounted, horizontal, portable

    PC-based software, fixed desk, mobile cart

    Stationary, mobile

    MAXIMUM CHANNELS By user requirements 16-80 16 IECG, 12 SECG Up to 128 with 224 inputs, 12-lead ECG, SpO2, up to 4 IBP, NIBP, TDCO, optional ETCO2, analog output synchronization

    PLUG-IN MODULES SECG, IBP, device interface(s), IECG; others as required

    SECG plus 2 IBP, V lead plus 2 IBP, IECG, STIM interface for cardiac stimulator, HLII1

    Integrated 12-lead ECG, 16 HIS ECG, built-in multiprogrammable stimulator

    CLab II plus amplifier, TRAM 451, ETCO2

    SECG CHANNELS 12 12 12 Up to 12

    IECG CHANNELS 24 Up to 72, 144 inputs 16 Up to 224

    STIMULATORS INCLUDED

    Optional No Yes Optional MicroPace

    PROGRAMMED LEAD SWITCHING

    Yes Yes Yes Yes

    DISPLAY SCREEN Yes Type High-resolution, 16 colors

    (1280 x 1024) High-resolution SVGA color LCD

    Two or three 20" flat-panel, 1600 x 1200 resolution

    Diag size, cm (in) 50.8 (20) 43.2 (17), 48.3 (19), 53.3 (21)

    50.8 (20)

    Data displayed Not specified Not specified Not specified User selectable Number of traces 32, user-configurable Up to 80 simultaneously 16 64/page Sweep speeds, mm/sec 5, must be user-selectable 25, 50, 100, 200, 400, 800 25, 50, 100, 200, 400, 800 5, 10, 25, 50, 100, 200, 400,

    user-entered value

    This is the first of five pages covering the above model(s). These specifications continue onto the next four pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 13

    Product Comparison Chart

    MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1

    BARD GALIX GE HEALTHCARE

    Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT

    RECORDER Type Thermal array, laser printer High-resolution laser Any Windows-compatible

    laser printer NA (optional laser printer)

    Number of traces 12, user-configurable 80 16 Laser printer, based on screen layout

    Paper speed, mm/sec 5, 12.5, 25, 50; variable Not specified 25, 50, 100, 200, 400, 800 NA Annotations Name, ID, date, time;

    clinical values as required Yes Name, operator, date/time,

    speed, ECG grid, position, comments

    Name, ID, paper speed, ECG and waveform grids, text header, date/time, free drawing

    Line selections Timing, millimeter grid/ruler Grid for amplitude and timing reference measurements, 12-lead ECG available on demand to print

    Amplitude and time grids Laser printer, based on screen layout

    INTEGRAL COMPUTER Any IBM Pentium Best available PC 2.33 GHz dual-core Xeon, Windows XP SP2 operating system

    Applications Not specified Not specified Not specified Data management Yes Yes Yes Yes Statistical analysis Yes See Other Specifications No Yes, Centricity CVIS Report generator Yes Yes Microsoft Word-based MS Word-based Other None specified None specified Orders, scheduling, ADT,

    charge capture, physician reporting, Holter, macros

    Storage Not specified (DVD) 500 GB hard drive, CD-R/DVD long-term

    Dual 80 GB HDD, 9.4 GB DVD/RAM, 144-900 GB RAID server

    NETWORKING BETWEEN LABS

    Yes No Windows Yes

    INTERFACES Biosense Webster's CARTO XP, ablation generators and stimulators

    HIS Yes Optional No Optional (ADT, orders, billing/charge capture, material management, results)

    Suppliers Not specified NA Optional

    This is the second of five pages covering the above model(s). These specifications continue onto the next three pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    14 2009 ECRI Institute. All Rights Reserved.

    Product Comparison Chart

    MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1

    BARD GALIX GE HEALTHCARE

    Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT

    Cardiology data management system

    Yes Optional No Integrated with GE Centricity Cardiology CVIS, optional interface to third-party CVIS

    Suppliers Not specified NA Optional Cath lab imaging system Not specified Not specified Not specified Optional, GE and third party

    via DICOM National Cardiology Database

    Not specified Not specified Not specified GE and third party

    This is the third of five pages covering the above model(s). These specifications continue onto the next two pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 15

    Product Comparison Chart

    MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1

    BARD GALIX GE HEALTHCARE

    Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT

    POWER REQUIREMENTS VAC, Hz 100/120/220/240, 50/60 110/220, 50/60 Hz 100-240, 50/60 Hz Consumption, W Not specified 500 1,800 maximum

    H x W x D, cm (in) 86.3 x 74 x 160 (34 x 29 x 63)

    9.6 x 30.2 x 27.2 (3.8 x 11.9 x 10.7), console without PC

    82.5 x 81 x 76 (32.5 x 32.1 x 30)

    WEIGHT, kg (lb) 272 (600) 2 (4.4) without PC 113 (249.2)

    PURCHASE INFORMATION

    Price range $50,000-150,000 $16,000-25,000 Not specified Warranty 1 year, parts, service, and

    installation 18 months, parts and labor 1 year, parts and labor

    Delivery time, ARO 2-4 weeks 4 weeks 45 days Service contract ~8% of purchase price Yes Optional Year first sold 2003 1998 2003, 2000 (CardioLab

    2000, 4000, and 7000), 1989 (DOS-based)

    Number installed USA Not specified Not specified >1,000 (combined IT

    platform) Worldwide Not specified Not specified >1,500

    Fiscal year January to December April to March January to December

    GREEN FEATURES None specified None specified None specified

    This is the fourth of five pages covering the above model(s). These specifications continue onto the next page.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    16 2009 ECRI Institute. All Rights Reserved.

    Product Comparison Chart

    MODEL ECRI INSTITUTE'S RECOMMENDED SPECIFICATIONS1

    BARD GALIX GE HEALTHCARE

    Basic EP Systems LabSystem Pro PHYSIO-28 CardioLab IT

    OTHER SPECIFICATIONS Windows 2000 operating system; customizable user interface; modular STAMP amplifier; template matching; T-wave subtraction; waveform analysis; data centralization (stimulator, RF generator, imaging); MS Word reports; DVD archiving; MicroPace stimulator interface; optional image capture.

    Programmable and preconfigured protocols; selection of stimulation channel by software; real-time and capture screens; triggering by any channel and stimulus synchronization; automatic rate measurement; amplitude and time measurement with unlimited number of calipers; event markers; storage of long-term studies; multiprogrammable stimulator with up to 3 extrastimuli; 3 methods of tachycardia termination; backup pacing and capability to work with independent automatic increment/decrement of coupling extrastimuli intervals.

    EP-only application; 3-D mapping interface option with Biosense Webster CARTO XP system; bidirectional interface; auto storage of mapping points; transfer of color snapshot and comprehensive report generation; auto control of CardioLab from CARTO XP system; CardioImage fluoroscopy image management system with up to 4 inputs of imaging; live and frozen snapshots display; centralized waveform archive with near real-time review of raw waveforms from any office PC connected to server; custom forms; interactive statistic, user-defined macros to automate tasks; full-disclosure storage of waveforms; remote modem support; dual hemo/EP capabilities; Nellcor OxySmart or Masimo SpO2; interactive administrative reporting; inventory management; billing/charge capture; online structured physician reporting.

    UMDNS CODE(S) 17898 17898 17898 17898

    LAST UPDATED May 2005 June 2009 June 2008

    Supplier Footnotes 1These recommendations are the opinions of ECRI Institute's technology experts. ECRI Institute assumes no liability for decisions made based on this data.

    Model Footnotes

    Data Footnotes

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 17

    Product Comparison Chart

    MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP

    System1 EP WorkMate Recording System

    WHERE MARKETED Worldwide Worldwide Worldwide Worldwide

    FDA CLEARANCE Yes Yes Yes Yes

    CE MARK (MDD) Yes Yes Yes Yes

    CONFIGURATION Stationary, mobile PC-based software, movable console, mobile cart, combo with Horizon SE

    Cabinet, table Cart or permanent install, 2 primary monitors, slave monitors, 120 or 56 catheter inputs

    MAXIMUM CHANNELS Up to 128 with 224 inputs, 12-lead ECG, SpO2, up to 4 IBP, NIBP, TDCO, optional ETCO2, analog output synchronization

    32 or 64, 12 SECG, 2 IBP, 18/50 IECG

    128 448 stored from up to 140 total inputs

    PLUG-IN MODULES CLab II plus amplifier, TRAM 451, ETCO2

    Integrated 12-lead ECG, 18/50 IECG, dual output multiprogrammable stimulator

    Inputs available in integrated input box and Hemopod (12-channel ECG, HR, cycle length, 4 IBP, thermal dilution CO, SpO2, NIBP, 1 QRS sync, respiration/CO2 module)

    Up to 4 pressure-monitoring channels, multimodality image acquisition, remote nurses charting station, navigation interfaces, FFT analysis

    SECG CHANNELS Up to 12 12 12 12

    IECG CHANNELS Up to 224 18/50 0, 32 (not available in USA), 64, or 128

    56, 120

    STIMULATORS INCLUDED

    Optional MicroPace Dual-output multiprogrammable

    Stimulator interface Yes

    PROGRAMMED LEAD SWITCHING

    Yes Yes Yes Yes

    DISPLAY SCREEN Type Two or three 20" flat-panel,

    1600 x 1200 resolution 2 high-resolution SVGA LCD flat-panel color

    21" (1600 x 1200) flat-panel color (EP and combo unit)

    High-resolution LCD flat-panel

    Diag size, cm (in) 50.8 (20) 50.8 (20) 53.3 (21), Sensis XP configuration dependent

    53.3 (21)

    Data displayed User selectable HR, systolic, diastolic, mean pressure

    User-configurable data parameters, including 12-lead ECG, 4 IBP, NIBP, SpO2, respiration, HR, cycle length, ablation data

    SECG, IECG pressure tracings, heart rate, ablation parameters

    Number of traces 64/page 2 panels, 32 traces each Unlimited, user configured, theoretical end at 80 traces

    32/page, 6 pages

    Sweep speeds, mm/sec 5, 10, 20, 25, 50, 100, 200, 400, user-entered value

    25, 50, 100, 150, 200, 300 12.5, 25, 50, 100, 200, 400 (real-time and review mode)

    10-600

    This is the first of five pages covering the above model(s). These specifications continue onto the next four pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    18 2009 ECRI Institute. All Rights Reserved.

    Product Comparison Chart

    MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP

    System1 EP WorkMate Recording System

    RECORDER Type NA (optional laser printer) Any network laser printer Laser printer

    (network/parallel port) LaserJet

    Number of traces Laser printer, based on screen layout

    32 Unlimited, user configured 64

    Paper speed, mm/sec NA 25, 50, 100, 150, 200, 250, 300

    12.5-400 real-time mode 10-300

    Annotations Name, ID, paper speed, ECG and waveform grids, text header, date/time, free drawing

    Name, operator, date/time, speed, ECG grid, position, comments

    Yes Annotations are user-configurable; text, characters, or numbers

    Line selections Laser printer, based on screen layout

    Amplitude and time grids Millimeter grid, raster is user configurable

    Lines or grid format

    INTEGRAL COMPUTER 2.33 GHz dual-core Xeon, Windows XP SP2 operating system

    2.8 GHz Pentium 4 Pentium 4, 3.0 GHz Dual-core Xeon processors

    Applications Not specified Not specified Microsoft XP Professional on ACQ and workstations, Microsoft Server 2003 on high-end server

    Not specified

    Data management Yes Optional Yes (with information system)

    EP monitoring and stimulation

    Statistical analysis Yes, Centricity CVIS Yes (with optional data management)

    Yes (with information system)

    No

    Report generator MS Word-based MS Word-based MS Word-based online and offline

    Yes

    Other Orders, scheduling, ADT, charge capture, physician reporting, Holter, ST segment, macros

    None specified Dedicated AXIOM Sensis server

    None specified

    Storage Dual 80 GB HDD, 9.4 GB DVD/RAM, 144-900 GB RAID server

    80 GB hard drive, optional CD-R/DVD-RW

    160 GB hard disk (RAID 5 available)

    160 GB RAID redundant backup

    NETWORKING BETWEEN LABS

    Yes Windows 2000 Ethernet, TCP/IP, DICOM, HL7, open architecture

    Optional

    INTERFACES Biosense Webster's CARTO XP, ablation generators and stimulators

    Not specified Not specified Biosense Webster CARTO system, Phillips Allura system, most approved ablation generators, HL7, LUMEDX, Xcelera

    HIS Optional (ADT, orders, billing/charge capture, material management, results)

    Optional DICOM, HL7, ASCII, bidirectional

    Yes

    Suppliers Optional Not specified; standard HL7 protocol for ADT, HL7 results or XML, RTF, and FTP

    Proprietary Optional

    This is the second of five pages covering the above model(s). These specifications continue onto the next three pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 19

    Product Comparison Chart

    MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP System1 EP WorkMate Recording

    System

    Cardiology data management system

    Integrated with GE Centricity Cardiology CVIS, optional interface to third-party CVIS

    Optional Yes Yes

    Suppliers Optional Mennen Medical All Optional Cath lab imaging system

    Optional, GE and third party via DICOM

    Not specified Siemens proprietary Not specified

    National Cardiology Database

    GE and third party Not specified ACC-NCDR modules (data collected for submission to third party), Heartlab/LUMEDX/Apollo/AXIS/Goodroe

    Not specified

    This is the third of five pages covering the above model(s). These specifications continue onto the next two pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    20 2009 ECRI Institute. All Rights Reserved.

    Product Comparison Chart

    MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP

    System1 EP WorkMate Recording System

    POWER REQUIREMENTS VAC, Hz 100-240, 50/60 Hz 110/220, 50/60 Hz 100-120, 230-240, 50/60 Hz 85-265, 47-63 Consumption, W 1,800 maximum 400 1,200 VA maximum 1,500

    H x W x D, cm (in) 82.5 x 81 x 76 (32.5 x 32.1 x 30)

    74.5 x [80.5-96] x 82 (29.3 x [31.7-37.8] x 32.3) console with monitor; 103 x 60 x 60 (40.6 x 23.6 x 23.6) mobile cart

    72 x 120 x 80 (28.3 x 47.2 x 31.5) with cabinet

    183 x 121.9 x 91.4 (72 x 48 x 36)

    WEIGHT, kg (lb) 113 (249.2) 80 (176) console without transformer, 50 (110) mobile cart without transformer, 29 (63.9) transformer

    ~85 (187.4), including monitors

    272 (600)

    PURCHASE INFORMATION

    Price range Not specified $66,750-82,000 Not specified $208,215-238,140 Warranty 1 year, parts and labor 1 year, parts and labor 1 year 1 year, parts and labor Delivery time, ARO 45 days 90 days FOB 6-8 weeks 30-60 days Service contract Optional Yes Optional 1-year increments, full

    service and biomedical second call-in

    Year first sold 2003, 2000 (ComboLab 2000 and 7000), 1989 (DOS-based)

    2003 2003 1996

    Number installed USA >1,000 (combined IT

    platform) Not specified >700 700

    Worldwide >1,500 Not specified >1,000 400 Fiscal year January to December January to December October to September January to December

    GREEN FEATURES None specified None specified None specified None specified

    This is the fourth of five pages covering the above model(s). These specifications continue onto the next page.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 21

    Product Comparison Chart

    MODEL GE HEALTHCARE MENNEN MEDICAL SIEMENS MEDICAL ST JUDE MEDICAL1 ComboLab IT EMS-XL 32/64 AXIOM Sensis XP EP

    System1 EP WorkMate Recording System

    OTHER SPECIFICATIONS EP-only application; 3-D mapping interface option with Biosense Webster CARTO XP system; bidirectional interface; auto storage of mapping points; transfer of color snapshot and comprehensive report generation; auto control of CardioLab from CARTO XP system; CardioImage fluoroscopy image management system with up to 4 inputs of imaging; live and frozen snapshots display; centralized waveform archive with near real-time review of raw waveforms from any office PC connected to server; custom forms; interactive statistic, user-defined macros to automate tasks; full-disclosure storage of waveforms; remote modem support; Nellcor OxySmart or Masimo SpO2; interactive administrative reporting; inventory management; billing/charge capture; online structured physician reporting.

    12-lead ECG and IECG interface for third-party mapping systems; interface to RF ablation systems.

    Multitasking, multiuser system; DICOM, HL7, ASCII communication optional. Meets requirements of CAN/CSA C22.2 No. 601-1M90; EN 865; IEC 60601-1, 60601-1-2, 60601-1-25, 60601-2-30, 60601-2-34, and 60601-2-49; and UL 2601-1.

    None specified.

    UMDNS CODE(S) 17898 17898 17898 17898

    LAST UPDATED June 2008 June 2009 June 2009 June 2009

    Supplier Footnotes 1St Jude Medical recently purchased EP MedSystems.

    Model Footnotes 1Hemo, EP, and combo system.

    Data Footnotes

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    22 2009 ECRI Institute. All Rights Reserved.

    Product Comparison Chart

    MODEL TEB SP12/32

    WHERE MARKETED Mexico, South America

    FDA CLEARANCE No

    CE MARK (MDD) No

    CONFIGURATION Cart mounted

    MAXIMUM CHANNELS 38

    PLUG-IN MODULES None, all functions are integral; 12 SECG, 18 IECG, 4 IBP, 1 thermodilution, 4 external inputs, 3 outputs

    SECG CHANNELS 12

    IECG CHANNELS 18

    STIMULATORS INCLUDED

    Yes

    PROGRAMMED LEAD SWITCHING

    Yes

    DISPLAY SCREEN Type 2 high-resolution color LCDs Diag size, cm (in) 43 (17) Data displayed Not specified Number of traces 32 Sweep speeds, mm/sec 12.5, 25, 50, 100, 200, 300

    This is the first of five pages covering the above model(s). These specifications continue onto the next four pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 23

    Product Comparison Chart

    MODEL TEB SP12/32

    RECORDER Type Laser printer Number of traces 32 Paper speed, mm/sec 12.5, 25, 50, 200, 300, 400 Annotations Date, time, lead, patient ID,

    gain, paper speed Line selections Time grid

    INTEGRAL COMPUTER AMD SEMPROM 64 3200 Applications Not specified

    Data management No Statistical analysis No Report generator Yes Other No

    Storage 80 GB

    NETWORKING BETWEEN LABS

    No

    INTERFACES HIS No

    Suppliers NA

    This is the second of five pages covering the above model(s). These specifications continue onto the next three pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    24 2009 ECRI Institute. All Rights Reserved.

    Product Comparison Chart

    MODEL TEB SP12/32

    Cardiology data management system

    No

    Suppliers NA Cath lab imaging system Not specified National Cardiology Database

    Not specified

    This is the third of five pages covering the above model(s). These specifications continue onto the next two pages.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    2009 ECRI Institute. All Rights Reserved 25

    Product Comparison Chart

    MODEL TEB SP12/32

    POWER REQUIREMENTS VAC, Hz 110/220, 50/60 Consumption, W 380

    H x W x D, cm (in) 140 x 90 x 63 (55 x 35 x 25)

    WEIGHT, kg (lb) 82 (180)

    PURCHASE INFORMATION

    Price range $44,000 Warranty 1 year, parts and labor Delivery time, ARO 4-6 weeks Service contract No Year first sold 1994 Number installed

    USA NA Worldwide 452

    Fiscal year January to December

    GREEN FEATURES None specified

    This is the fourth of five pages covering the above model(s). These specifications continue onto the next page.

  • Physiologic Monitoring Systems, Cardiac Electrophysiology

    26 2009 ECRI Institute. All Rights Reserved.

    Product Comparison Chart

    MODEL TEB SP12/32

    OTHER SPECIFICATIONS Continuous storage in hard disk; DVD recorder included to save stored data at end of procedure; USB port; Windows-based software included on PC. Registered by the Brazilian Ministry of Health. TEB is ISO 9001 and ISO 13485 certified; IEC 601 certified.

    UMDNS CODE(S) 17898

    LAST UPDATED June 2009

    Supplier Footnotes

    Model Footnotes

    Data Footnotes

    Scope of this Product ComparisonPurposePrinciples of operationSecurityExternal databases and registries

    Reported problemsPurchase considerationsECRI Institute recommendationsOther considerationsCost containmentPresent Value/Life-Cycle Cost Analysis

    Stage of developmentBibliographySupplier informationPolicy StatementAbout ECRI InstituteProduct Comparison Chart