45
CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009 PIV Cameras and time resolved PIV measurement B. Lecordier et M. Trinité CORIA UMR CNRS 6614 Université & INSA de Rouen (France) Mail : [email protected] Web: http://www.coria.fr

PIV cameras and time resolved PIV measurement

  • Upload
    builien

  • View
    244

  • Download
    1

Embed Size (px)

Citation preview

Page 1: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

PIV Cameras and time resolved PIV measurement

B. Lecordier et M. Trinité

CORIAUMR CNRS 6614Université & INSA de Rouen (France)Mail : [email protected]: http://www.coria.fr

Page 2: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Outlines

Introduction

Interline transfer CCD in the 1990’s for PIVFrame straddling techniqueTransition toward cross-correlation approachActual CCD cameras

Time Resolved PIV measurement for high speed flows (> 1000 Hz)High speed photographic camera for PIV in 90’sFirst high speed CMOS video camera for PIVtActual high speed CMOS cameras

Conclusions et perspectives

Page 3: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Introduction 1/3 - End of the 80’s : Video Camera for PIV measurement

At the end of 80’s more or less all the PIV systems were based double exposed imagesRecording support

Photographic filmHigh resolution full frame CCD

Use of video forDigitized the film or one step of the treatmentDirect recording of double exposed image

Support: photographic film

From : http://25-years-piv.dlr.de/

Support: Full frame CCD camera

Framen-2 n-1 n n+1

Exposure Read-out

Laser

n-2 n-1 n n+1

Page 4: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Introduction 2/3 - End of the 80’s : Video Camera for PIV measurement

Auto-correlationDirectional ambiguity (Image shift)Lower seeding density than CCSmall dynamic range than CCThe recording media permitted to investigate high speed flow

Cross-correlationThe mains advantages of CC technique were well established, but the flow velocity range was limited by the video frame rate: (water flow…)

Support: Full frame CCD camera

Framen-2 n-1 n n+1

Exposure Read-out

Laser

n-2 n-1 n n+1

Support: Full frame CCD cameran-2 n-1 n n+1

n-2 n-1 n n+1

Δt

= 1/F > 10 ms to 1 sΔt

Page 5: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Introduction 3/3 - Interline transfer CCD Transition towards CC technique

Common pixel storage for odd and even fieldsRapid transfer of odd or even fields into the pixel storage

Raffel, M., Willert, C., Wereley, S. and Kompenhans, J.-

PIV. A practical guide (2d edition), Spinger-verlag, ISBN 978-3-540-72307-3

Even Odd Even Odd Even Odd Even Odd Even Odd Even

Frame rate

Field rate

Elec. Shutter

Page 6: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Interline transfer CCD: frame Straddling synchronisation 1/2

Positioning of the two laser pulses on the transition between odd and even field

Odd field

Even field

Even Odd Even Odd Even Odd Even Odd Even Odd Even

Frame rate

Field rate

Laser

CCD

Wenert

et al. (1991)

Lecordier

et al. (1994) EiFHuang, H.T. & Fieldler

H.E. (1994) EiF

Page 7: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Interline transfer CCD: frame Straddling synchronisation 2/2

Even Odd Even Odd Even Odd Even Odd Even Odd Even

Frame rate

Field rate

Laser

Odd field (t) Even field (t+Δt)

Complete the two fields to avoid one pixel shift between images

Minimum Δt

fixed by the transfer of field into the storage area (< 10 μs)

Format : 2 * 768.256 pixel image size

Page 8: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

First acquisition with frame-straddling at CORIA laboratory : Pulsed jet

Camera : XC77CE-SONY (similar: Pulnix, JAI…)Square pixels, 11×11 µmResolution: 756×581 (CCIR)Minimum inter-frame delay 10 μsRestart Reset function

Frame gabber connected to a PCCopper vapour laser (only two pulses)

V

P

Unsteady injection

Combustion chamber

Solenoid valves

Pressurized tank

Page 9: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

First acquisition with frame-straddling at CORIA laboratory : Pulsed jet

Camera : XC77CE-SONY (similar: Pulnix, JAI…)Square pixels, 11×11 µmResolution: 756×581 (CCIR)Minimum inter-frame delay 10 μsRestart Reset function

Frame gabber connected to a PCCopper vapour laser (only two pulses)

Synchronisation on the injection

Page 10: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

First acquisition with frame-straddling at CORIA laboratory : Pulsed jet

Camera : XC77CE-SONY (similar: Pulnix, JAI,…)Frame gabber connected to a PCCopper vapour laser (only two pulses)

Original images : Lecordier, 1994 -

CORIA

Odd field: 768x256 pixels (t) Even field: 768x256 pixels (t+Δt)

Page 11: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

First acquisition with frame-straddling at CORIA laboratory : Pulsed jet

Camera : XC77CE-SONY (similar: Pulnix, JAI…)Frame gabber connected to a PCCopper vapour laser (only two pulses)

Velocity field

Page 12: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

EUROPIV 1 : Comparison of stereoscopic PIV (translation conf.) to holography

Collaboration : CORIA and ISL (H. Royer) in a swirling flow (EUROPIV 1 Project)

Page 13: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Growing popularity of CC PIV method in the 90’s with low resolution interline CCD

Simplest way to resolve the directional ambiguityRange of particle density larger than auto-correlationLarge range of velocity, including zero displacementEasier to developed advanced PIV algorithms….Other aspects which contributed to the transition towards CC method

Sub-pixel accuracy to compensate low resolution of images (768x256 pixels)

Increasing of computer power for image processingDevelopment of full digital PIV system for recording and processing

u ≈

v ≈

3 pixels

Without sub-pixel With sub-pixel

Page 14: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

And Today ? CCD Camera for PIV : Full-Frame Interline Transfer CCD

Introduce in the first half of the 1990’sDerivate of the interline transfer CCDEach pixel has its own storage site

Fast transfer of the entire exposed images (< 1 μs)

Interline transfer

Full frame Interline transfer

CCD structure of Full frame interline transfer

Raffel, M., Willert, C., Wereley, S. and Kompenhans, J.-

PIV. A practical guide (2d edition),

Spinger-verlag, ISBN 978-3-540-72307-3

Page 15: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

And Today ? CCD Camera for PIV : Full-Frame Interline Transfer CCD

Introduce in the first half of the 1990’sDerivate of the interline transfer CCDEach pixel has its own storage site

Fast transfer of the entire exposed images Inter-frame delay < 1 μsReset/restart capabilityResolution 1k.1k up to 4k.4kPixel size 5~10 μmMicrolenses to improve fill factor

Time resolved measurement confine only for slow flow due to low frame rate (1 à 100 Hz).Integration of the signal in the second image

Page 16: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Time Resolved PIV in the 1990’s for high speed flow

High speed photographic camera for PIV in 90’sFirst high speed CMOS video camera for PIVtActual high speed CMOS cameras

Page 17: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Principle HS photographic camera

Maximum linear speed of the film : 0.135 mm/ms, namely 135 m/s (≈10 000 rpm)Drum size : 80 cm to 100 cmMinimal recording duration: 6 msImage size on film : 20x25 mm² - 40 à 150 images at each run

Lens

Photographic film

Rotating Drum

Mirror

Cording camera

Light

source

Coppe

r vapo

ur

laser

(1 000

to 20

000 p

ulses/s

)

Page 18: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Acquisition cycle with a drum camera

Load the film on the drum (1 min)Start the rotation and wait stabilization of drum speed (≈

30 s)

Acquisition (1 drum rotation) (6 to 100 ms)Chemical process of the film (1/2 h)Film digitalization for treatment (1/2h to 1h)

Signal to open the camera shutter

1 drum rotation

Copper vapour laser

Opening of Camera shutter

Page 19: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Film digitalization:8 bits in 2000 dpi2000x1500 pixels²/image

High resolution film scanner

High speed photographic PIV: Film digitalization with high resolution scanner

2000x3000 pixels in 24x36 mm film format

Page 20: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Unsteady laminar and turbulent combustion

Pressure : 1 à 5 barsSeeding : olive oil Duration of recording : ≈

6 msVariable turbulence levelCentral mixture ignition

Acquisition devicesCording camera ( 80 cm)

Photographic film : Kodak N&B - TMAX 400

Copper vapor laser (Oxford)Power : 45 W, soit : 5 to 10 mJ/pulseRepetitive rate : 2 à 20 kH

Page 21: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Unsteady laminar and turbulent combustion

CORIA, 1993

Page 22: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Example of flame propagation5000 fr/s – Laminar case 6800 images/s : turbulent case

Ignition

5.8 ms

Page 23: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Time evolution of the flame contour

Page 24: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Computation of PIV velocity field

Laminar

flame

(5000 images/s)

Turbulent flame

(8000 image/s)

Use two

successive images for PIV measurement

using

cross-correlation

Page 25: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Film positioning difficulties for CC (1/2)

Film rotation between two successive frame

Effect

of film rotation Effect

of film translation

Film displacement

Page 26: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Film positioning difficulties for CC (1/2)

Fixed pattern of particle in the corner of each image for positioning

Avoid film deformation during digitalization step RMS

Mean flow

Time

Page 27: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Error introduced by film positioning

Error : 0.05 to 0.1 pixel

RMS

Mean Flow

Time (ms)

Measurement in a flow at rest (compared to the acquisition rate : 5000 Hz)Pixe

l

Page 28: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Laminar flame – 5000 fr/s – 2000x1500 pixels

propane/air : φ=1.0

Page 29: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Laminar flame – 5000 fr/s – 2000x1500 pixels

Page 30: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Turbulent flame – 6800 fr/s – 2000x1500 pixels

(propane/air : φ=1.0, u’/Sl=1,5)

Page 31: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

(propane/air : φ=1.0, u’/Sl=1,5)

High speed photographic PIV: Turbulent flame – 6800 fr/s – 2000x1500 pixels

Page 32: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

(propane/air : φ=1.0, u’/Sl=1,5)

High speed photographic PIV: Turbulent flame – 6800 fr/s – 2000x1500 pixels

Page 33: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

(propane/air : φ=1.0, u’/Sl=1,5)

High speed photographic PIV: Turbulent flame – 6800 fr/s – 2000x1500 pixels

Page 34: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Time evolution of flame contour

Page 35: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Avantages et drawbacks

High repetitive rateHigh resolution of image (> 2000x1500 pixels at 6000 fr/s)

No statisticFixed film length limited duration of recording

+

-

Lens

Photographic film

Rotating Drum

Mirror

Cording camera

Light

source

Coppe

r vapo

ur

laser

(1 000

to 20

000 p

ulses/s

)

Page 36: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed photographic PIV: Increasing of recording duration

DrumRotation

Laser

Camera opening

D1

D1

D1

D1

D2 D3

Dn

Dn-1

D2 D3D4

Dn-1 Dn

D4

D1

Dn-1 Dn

D1

D2

D2

D2

1 or more drum

rotation

Recording duration ≈

100 ms

N drum

rotations

Honore, D.; Lecordier, B.; Susset, A.; Jaffre, D.; Perrin, M.; Most, J. M. & Trinite, M. (2000), Experiments In Fluids 29

Page 37: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

First time resolved PIV with high speed CMOS video camera

KODAK EKTAPRO HS Motion Analyzer, Model 4540

Page 38: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed CMOS camera : KODAK EKTAPRO HS - Model 4540

Resolution : 256x256 at 4500 fr/sFrame rate : Records up to 4,500 full frames

per second or up to 40,500 partial frames per secondPixel size : 22 μmNo frame straddling synchronisation 256 pixels

Page 39: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed CMOS camera : KODAK EKTAPRO HS - Model 4540

msVideo

Exposure

Laser

Δt = 1/flaserNo frame straddling shychronisation

Page 40: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed CMOS camera : KODAK EKTAPRO HS - Model 4540

Camera : KODAK EKTAPRO HS - Model 4540Laser : Copper vapour Laser at 9kHzSeeding particle : Olive oil

Demarre

(93) CORIA

Lifted turbulent flame

Page 41: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed CMOS camera : KODAK EKTAPRO HS - Model 4540

Demarre

(93) CORIA

9000 fr/s

resolution : 256x128 pixel

Page 42: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed CMOS camera : KODAK EKTAPRO HS - Model 4540

Demarre

(93) CORIA

9000 fr/s

256x128 pixels

Page 43: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed CMOS camera: “New” generation

Resolution: larger than 2k.2kAcquisition rate: 1 to 6kHz at full sizeDepth : 8 to 12 bitsPixel size 10 to 20 μmFrame straddling feature (Δt < 5 μs)

Page 44: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

High speed CMOS camera: New generation – frame straddling≈

ms

Frame rate

Exposure

Laser< 5 μs

Page 45: PIV cameras and time resolved PIV measurement

CNRS – UNIVERSITE et INSA de Rouen Symposium 25 Years of PIV in aerodynamics, Göttingen, September 23 - 25, 2009

Conclusions

Low repetitive rate cameraHigh resolution (up to 4k.4k)Δt < 300 nsHigh sensitivity8 to 12 bits imagesFrame rate 1 to 100 Hz

Integration of the signal in the second imageReduction of pixel size to reduce peak locking

High speed CMOS video cameraResolution up to 2k.2k at more than 4kHz8 to 12 bits imagesPixels size (10 to 22 μm) – To be decreasedSensitivity has to be improvedStraddling synchronization

Increasing of the laser power50 to 100 mJ/pulse @ 5kHz with 10 ns pulse duration would be perfect

Unfortunately, PIV is a “marginal” market for camera manufacturersIn the early stage of development, custom cameras were possibleNow, we have to wait new models on the market