Quick Guide Component R12 / R6 Cat. No. : CP-R12 / CP-R6 Setting Temperature Setting temperature -10 to 40℃ (R12) / -20 to 40℃ (R6) Setting temp. limit Lid Open Stop Start Max. RPM : 4,800 rpm Max. RCF : 7,470 xg Max. RPM : 5,000 rpm Max. RCF : 7,261 xg Component R12 Component R6 Serial No. Installation Date Display Screen Status of RPM / RCF / Time / Temperature / Rotor ID / ACC / DEC / Program Setting Speed Press once : Setting RPM Press twice : Setting RCF Setting Time Up to < 100 hr Continuous run (00:00) View Settings Verifying the setting parameters during centrifugation Program Save / Call Saving a program (up to 100) Calling up a program Setting the Rotor ID manually Enter the rotor ID. manually, when rotor id is not identified automatically Setting ACC / DEC Rate ACC (0 ~ 9 steps, Max.9) DEC (0 ~ 9 steps, Natural DEC.0) Display & Control Panel Saving / Calling Up a Program Setting Run Parameters (RPM / RCF / Time / Temperature / ACC / DEC) Temperature range that stops operation when the temperature is lower or higher than the set temperature. Setting Temp. Limit Press a parameter to be changed Enter the desired values on the keypad Press the Enter button Press the Enter button Enter the desired program number on the keypad Press the Save or call button Press the Temp. limit button Press the Enter button Enter the desired temperature range on the keypad

Quick Guide Component R12 / R6 · Component R12 Component R6 30 cm 30 cm 30 cm Emergency Lid Open If the centrifuge lid cannot be opened, insert the provided T-wrench into the emergency

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Quick Guide Component R12 / R6 · Component R12 Component R6 30 cm 30 cm 30 cm Emergency Lid Open If the centrifuge lid cannot be opened, insert the provided T-wrench into the emergency

Quick Guide Component R12 / R6 Cat. No. : CP-R12 / CP-R6

Setting Temperature Setting temperature -10 to 40℃ (R12) / -20 to 40℃ (R6) Setting temp. limit

Lid OpenStopStart

Max. RPM : 4,800 rpm

Max. RCF : 7,470 xg

Max. RPM : 5,000 rpm

Max. RCF : 7,261 xg

ComponentR12

ComponentR6

Serial No. Installation Date

Display ScreenStatus of RPM / RCF / Time / Temperature / Rotor ID / ACC / DEC / Program

Setting SpeedPress once : Setting RPMPress twice : Setting RCF

Setting TimeUp to < 100 hr Continuous run (00:00)

View SettingsVerifying the setting parameters during centrifugation

Program Save / Call Saving a program (up to 100)

Calling up a program

Setting the Rotor ID manually Enter the rotor ID. manually, when rotor id is not identified automatically

Setting ACC / DEC Rate ACC (0 ~ 9 steps, Max.9)

DEC (0 ~ 9 steps, Natural DEC.0)

Display & Control Panel

Saving / Calling Upa Program

Setting Run Parameters(RPM / RCF / Time / Temperature / ACC / DEC)

Temperature range that stops operation when the temperature is lower or higher than the settemperature.

Setting Temp. Limit

Press a parameterto be changed

Enter the desired values on the keypad

Press theEnter button

Press theEnter button

Enter the desired program number on the keypad

Press theSave or call button

Press theTemp. limit button

Press theEnter button

Enter the desired temperature range on the keypad

Page 2: Quick Guide Component R12 / R6 · Component R12 Component R6 30 cm 30 cm 30 cm Emergency Lid Open If the centrifuge lid cannot be opened, insert the provided T-wrench into the emergency

Connect the device to voltage

sources, which correspond to the

electrical requirements on the label

attached to the device.

Check the wall outlet is earth-

grounded, surge-protected and

can supply enough power.

Connect the power cable to the

distribution box, or install the plug

and socket that are packed with

the device.

Switch off the device after use.

Unplug the power plug before

cleaning or unused for a long

period of time.

Check the Line Voltage Grounded Outlet Power Connection Switch Off the Device

Quick Guide Component R12 / R6 Cat. No. : CP-R12 / CP-R6

Allow a safety clearance of 30 cm around the centrifuge during operation.

Rotors

Spread the grease lightly on the bucket where it contacts the rotor and the pivot.

Remove any dust regularly from the ventilation slits on the left side of the device.

There is a drain hose at the bottom of the device. The condensation water is automatically drained through the hose. Check and remove it regularly.

Component R6Component R12

30 cm

30 cm30 cm

Emergency Lid Open

If the centrifuge lid cannot be opened,insert the provided T-wrench into the emergency release hole and turn it clockwise. (Before activating emergency open, switch off the device and wait for the rotor to stop.)

Wipe up the moisture in the chamber with a dry cloth after use. Keep the lid open when not in use.

Clean the chamber and rotor by wipingwith a cloth soaked in mild detergent and dry completely.

Keep the Chamber Dry

Cleaning

Check the rotor, buckets and tubes are free of scratchand crack before operation.

Check the Accessories for Abnormalities

Load tubes symmetrically into the rotorwith identical amount.

Load Tubes Symmetrically

Place the lid on the wind shield rotor before centrifugation.

Check the Lid

After relocating the device or loading the rotor, balance them by using a leveler. If necessary, level the device by adjusting the ‘Leveling foot’ on the bottom of the device.

Balancing the Centrifuge and Rotor

3.0KVA (Component R12)

5.0KVA (Component R6)

220~240V 50/60Hz

Maintenance

Electrical Safety

Cautions for Use

Rev.1

Proudly made by Jiyoung Shim, Hyewon Jeong

Order & Inquiry l +82-2-3472-0727 l [email protected] l www.ihanil.com

Service l +82-2-3452-8966 l [email protected]

Before Operation After Operation

Maintenance of Swing-Out Rotor

Remove Dust Empty the Condensation Water

Allow a Safety Clearance of 30 cm

WS1000-6WS2400-6 6 x 1,000 mL / Max. 5k 6 x 2,400 mL / Max. 4.8k