View
276
Download
6
Tags:
Embed Size (px)
Citation preview
Scanning Electron Microscope
القادر. عبد عمر د
القادر. عبد عمر د
The basic premise of an SEM is that signal produced from a scanned area of the specimen is displayed as an image with the exact same scan pattern on a CRT
القادر. عبد عمر د
The scan pattern on the specimen is created by a set of deflection coils in the column that move the beam in a coordinated X/Y pattern. This is referred to as a scan or “raster” pattern
القادر. عبد عمر د
Cathode Ray Tube accelerates electrons towards the phosphor coated screen where they produce flashes of light upon hitting the phosphor. Deflection coils create a scan pattern forming an image in a point by point manner
القادر. عبد عمر د
Color CRTs usually have three separate e-guns, one each for red, green, and blue (RGB)
القادر. عبد عمر د
القادر. عبد عمر د
The scan generator coordinates the movement of the primary beam with the movement of the e-gun in the back of the CRT
القادر. عبد عمر د
Magnification is accomplished by scanning a progressively smaller portion of the specimen and displaying the image on the CRT. Thus total magnification is square area of CRT divided by area scanned.
القادر. عبد عمر د
In contrast focus is accomplished by bringing the beam to its crossover point on the surface of the specimen. In this way focus and magnification are completely separate from one another in the SEM.
القادر. عبد عمر د
In the TEM the specimen lies very close to the objective lens resulting in a relatively large half angle of illumination. In SEM since the image is not formed by an objective lens the half angle can be very small resulting in a large depth of field.
القادر. عبد عمر د
10X
An SEM focused at high magnification will still be in focus at low magnification
القادر. عبد عمر د
110X
القادر. عبد عمر د
200X
القادر. عبد عمر د
400X
القادر. عبد عمر د
4K
القادر. عبد عمر د
16K
القادر. عبد عمر د
45K .القادر عبد عمر د
Strong Lens:Small probe size, high resolution, short working distance and shallow depth of field
Weak Lens:Larger probe size, low resolution, long working distance, and larger depth of field .القادر عبد عمر د
A smaller final lens aperture can reduce the half angle and therefore increase the depth of field. This is true on a relatively strong lens which has a fairly short working distance and therefore high resolution.
القادر. عبد عمر د
The SEM forms an image by generating a number of signals as a result of the beam interacting with the specimen.
القادر. عبد عمر د
The SEM is a probe forming (e- beam) and signal detecting device. By developing an image created in a point by point fashion an important factor is the signal to noise (S/N) ratio. The signal being the result of the beam interacting with the specimen and the noise being the result of imperfections in the electronics of the detector and display systems as well as spurious signal.
القادر. عبد عمر د
Signal can be increased by:Creating more beam specimen interactions
Noise can be reduced by:Cooling electronicsKeeping detectors settings to a minimum
Signal/noise ratio can be increased by:Placing detector closer to source of signalSlowing down the scan (collect more signal per unit time)
القادر. عبد عمر د
Although the same amount of signal is produced throughout the specimen the topography of the surface will allow differing amounts of signal to reach a detector placed off to the side.
القادر. عبد عمر د
A number of different detectors can be incorporated into the chamber surrounding the specimen.
القادر. عبد عمر د
The shadow produced in an SEM is determined by the position of the detector but the view is a “beam’s eye” view as ifcolumn
one were looking down theالقادر. عبد عمر د
LEO Gemini Column
A detector placed within the column is known as
an “in-lens” detector and produces a very different image compared to a conventionally located detector القادر. عبد عمر د
Secondary Electron Detector
Side Mounted In-Lens
القادر. عبد عمر د
Secondary Electron Detector
Side Mounted In-Lens
MIRA SEM
القادر. عبد عمر د
MIRA SEM
القادر. عبد عمر د