Page 1: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Electron microscope techniques-

Scanning Electron Microscope (SEM)Transmission Electron Microscope (TEM)

Page 2: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Topics

• Electron sources• Electro-magnetic lenses• Interaction of the electron probe with the sample

– Secondary electrons– Backscattered electrons– X-rays– Auger electrons

• Detection of electrons• Various modes of operation

– Imaging– Analytical

Page 3: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Brief historic overview

• TEM: 1931 byErnst Ruska (1906-1988)Nobel prize 1986

• SEM: 1937-39 by• Manfred Baron von Ardenne

(1907-1997)

Page 4: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Electron sources

Electron guns:

• Various examples of gun design

– Thermionic– Schottky– Field emission

• Cathode material– Tungsten– Lanthanum

hexaboride (LaB6)

– Others…• Cathode material

determines emission current density Energy scheme of various gun types

Page 5: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Electron guns

• Wehnelt cup to concentrate emission on a small area of the cathode • Schottky emitter can be considered as field assisted thermionic gun• Field emission needs a better vacuum – otherwise the tip radius is

destroyed by Ion bombardment of residual gas

Page 6: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Electron guns

Page 7: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Electron guns

• Various methods to determine gun brightness

Page 8: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Electro-magnetic lens systems

sketch

Page 9: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Electro-magnetic lens systems

• Axial magnetic field with rotational symmetry

• Radial component of the magnetic field is related to the radial component (Maxwell)

• Electrons travel along screw trajectories due to Lorentz force

• 1/f = 1/p + 1/q (lens equation) holds true, therefore crossover is demagnified by M= q/p

BveF

×=

Page 10: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Lens aberrations

sketch

Page 11: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Resolution

• Rayleigh Criterion (a)– Current density distribution

overlaps at half of separation with an intensity drop ≤ 75%

• Edge Resolution (b)– Distance between points

corresponding to 25% and 75% of total step height

• Radial Intensity Distribution (c)– Diameter that contains a given

percentage of the total probe current

• Maximum Spatial Frequency (d)– Contrast transfer drops so much

that periodicity Λmin cannot be detected

Bild 2.27

Page 12: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Electron-sample interactions

• Elastic scattering– Classical scattering: Electron-Nucleus (Rutherford)– Relativistic scattering: Schrödinger and Pauli-Dirac, Quantum mechanics of

scattering, exact Mott cross-section

• Inelastic scattering – Electron Excitation Processes and Energy Loss– Plasmon interaction and inter- intraband transitions– Electron-electron binary collisions– Ionization of inner shells

Page 13: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Elastic scattering

• Classical scattering (a): Electron-Nucleus (Rutherford)

Coulomb force:

The ratio dσ/dΩ is called „differential cross section for scattering through an angle θ“

• Relativistic scattering (b) – will not be treated here, but results in…

r20

2

ur4

ZeF

επ−=

Page 14: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Elastic scattering

Page 15: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Inelastic scattering

• Plasmon interaction and inter- intraband transitions– Plasmon is a collective longitudinal charge density wave– Plasmon energies are on the order of 5 – 30 eV– Inter- and intraband transitions have a similar energy regime

→ Energy losses are in the range of 0 – 50 eV at scattering angles below 10 mrad

• Electron-electron binary collisions and inner shell ionization are usually treated in a quantum mechanics way

• The ratio ν of total inelastic to total elastic cross section decreases with increasing Z

• Modeling of multiple scattering is handled by complex Monte-Carlo simulations

Z20

el

inel ≅σσ

Page 16: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Elastic + Inelastic scattering

Page 17: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Elastic + Inelastic scattering

Al: Z=13 Au: Z=79

Page 18: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Elastic + Inelastic scattering

C: Z=6Cu: Z=29Au: Z=79

Page 19: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

What kind of species are generated?

Probe-sample interaction results in the „generation“ of

• Secondary electrons• Backscattered electrons• X-rays • Auger electrons• Plasmons

Page 20: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Secondary electrons (SE)

• SE (exit energies < 50 eV) are generated if the energy gain of these species is large enough to overcome the work function

• This process needs to be treated quantum mechanically as the scattering of an electron wave at a potential barrier

• SE are only able to escape from a small surface range (probability of escaping is based on their inelastic mean free path)

• Backscattered electrons contribute to the SE yield δ

Page 21: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Backscattered electrons (BSE)

• BSE are present in the whole energy range from 50 eV (definition) to the maximum acceleration energy of the primary electrons (PE)

• Their spectrum shows a broad peak overlapped by SE and Auger peaks as well as plasmon loss

• BSE and SE are the most important signals for imaging. Knowledge about the dependence of the backscattering coefficient and the SE yield on surface tilt, material and electron energy is essential for any interpretation.

Page 22: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

X-ray

• Acceleration of a charged particle (electron) in the screened Coulomb potential of the nucleus leads – with a low probability – to an emission of a X-ray quantum (usually elastic scattering is observed)

• Electron is decelerated by hν (energy of the X-ray quantum) → continuous X-ray spectrum

• This continuous spectrum is superposed on the characteristic X-ray spectrum generated by filling of inner shell vacancies

Page 23: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

X-ray

Page 24: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Auger electrons (AE)

• De-excitation energy produced when an inner shell vacancy is filled does not necessarily involve the emission of a X-ray quantum – its also possible that an AE is emitted

• The AEs are superposed on the low energy tail of the BSE

• Only AEs from an ultra thin surface layer contribute to these energy peaks (multiple-loss AEs are lost to the BSE background)

Page 25: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

How to detect SE and BSE

• SE detector: Scintillation & Photomultiplier combination (Everhart & Thornley)

Page 26: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

How to detect SE and BSE

• In many cases BSEs are detected by semiconductor detectors (p-n junction diodes)

Page 27: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

How to detect X-rays

• Wavelength dispersive X-ray spectrometer (WDX)Bragg:

High energy resolution

λ=θ nsind2 B

• Energy dispersive X-ray spectrometer (EDX)

Whole spectrum can be recorded simultaneously

Page 28: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Imaging modes

• SE mode / BSE modeSE BSE

Page 29: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Imaging modes – X-ray

• X-ray Line-Scan/mapping

– Limited resolution– Usually bad signal to noise ratio

→ Limited situations were this technique is applicable

Page 30: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Analytic modes - diffraction

• Diffraction mode

– By equipping the instrument with a camera or CCD detector system it is possible to use a SEM or TEM in the diffraction mode

– According to Bragg‘s equation one will find constructive and destructive interference from electrons of energy E scattered at adequate lattice planes

– This also works to analyze the texture or stresses in thin films– In amorphous materials it is possible to gain insight into the nearest neighbor

distance

The general technique of diffraction will be treated in a separate lecture

Page 31: Scanning Electron Microscope (SEM) Transmission …...Electron microscope techniques-Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) Topics • Electron sources

Analytic modes - Auger

• This mode will be discussed in the next section as it is usually used in the context of a sputtering process to gain insight into the 3dimensional element distribution