TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR ATOMIC SPECTROSCOPY BY SUMMER N. HANNA A Dissertation Submitted to the Graduate Faculty of WAKE FOREST UNIVERSITY GRADUATE SCHOOL OF ARTS AND SCIENCES in Partial Fulfillment of the Requirements for the Degree of DOCTOR OF PHILOSOPHY Chemistry May 2011 Winston-Salem, North Carolina Approved By: Bradley T. Jones, Ph.D., Advisor Clifton P. Calloway, Ph.D., Chair Ulrich Bierbach, Ph.D. Christa L. Colyer, Ph.D. Willie L. Hinze, Ph.D.

TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

  • Upload
    others

  • View
    10

  • Download
    0

Embed Size (px)

Citation preview

Page 1: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR ATOMIC SPECTROSCOPY

BY

SUMMER N. HANNA

A Dissertation Submitted to the Graduate Faculty of

WAKE FOREST UNIVERSITY GRADUATE SCHOOL OF ARTS AND SCIENCES

in Partial Fulfillment of the Requirements

for the Degree of

DOCTOR OF PHILOSOPHY

Chemistry

May 2011

Winston-Salem, North Carolina

Approved By:

Bradley T. Jones, Ph.D., Advisor

Clifton P. Calloway, Ph.D., Chair

Ulrich Bierbach, Ph.D.

Christa L. Colyer, Ph.D.

Willie L. Hinze, Ph.D.

Page 2: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

i

DEDICATION

This dissertation is dedicated to my grandmother, Geraldine Gibson, the first person to take me into a lab.

Page 3: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

ii

ACKNOWLEDGEMENTS

I would like to first off thank my parents and family for always being a wellspring of enthusiasm, encouragement, and belief in my abilities even and most importantly when I doubted myself. They have been a constant source of compassion and strength throughout my life, and I know that there is no way that I would have ever been capable of “eating other people’s lunch” without them.

To those who have contributed to my education, thank you for encouraging me to constantly challenge myself. To my committee members, Dr. Hinze, Dr. Bierbach, Dr. Calloway, and Dr. Colyer I appreciate all of the time you have taken to direct me through this extensive process. Thank you to Dr. Owens who convinced me that it was infinitely better to be a chemistry major than anything else I could have done in college and then that it would be silly of me to not go to graduate school. You’ve been a steadfast source of enthusiasm and encouragement. Dr. Calloway, thank you for your continued support over the years. You are an inspiring educator and it has been such a pleasure working for you as an undergraduate and now alongside you over the past 7 years. Dr. Bierbach, I appreciate all of the time you’ve taken to counsel me regarding chemistry, my career, and life . You’ve helped me become a better professional and your dedication to students is amazing. Dr. Jones, thank you for your guidance throughout graduate school. Your ability to always see the humor in any situation, enthusiasm for education, and analytical chemistry savvy continues to motivate me to be a better chemist. I appreciate all of the long hours you have dedicated to your students and I am so grateful for the wonderful experience you have provided me.

To my friends, at Wake Forest and beyond, you have each touched my life in a way that I will never forget and I want to take the opportunity to thank you for making my world a better place. Whether you are a friend from childhood, Winthrop, or Salem Hall, I appreciate the time that you have taken to be my friend and the constant support that you have provided. Thank you April and Scott for forcing me to show up for my first day at Wake, even when I wanted to just set up a new life in Rock Hill. A special thanks to my lab group—Carl, Jiyan, and George. I’m pretty certain that I’ve spent more time with you guys than anyone else I know in the last four years. It’s been fantastic to have you all to bounce ideas off of and I’m certain that you’ve helped me become a better scientist. To the lunch group—Chris, Julie, Junker, Erika, and Tanya, I’m so thankful for you all and all the laughs and stories we’ve shared. You all are a fantastic group of friends and I’m so lucky to have you. Amanda—I’m glad that we’ve had the opportunity to become close and thankful for all of our OG days to celebrate, commiserate, and sometimes just get away. Thank you Leigh Ann for being my first friend in Winston—I’ll never forget our special bond and the ability to become immediate friends. I’ll miss dinner nights, OG, accidental twin days, and having such an amazing support system throughout graduate school. Rachel, I’ll always be grateful that I was lucky enough to find a hetero-life mate at 8 years old and even when we’re hundreds of miles apart our relationship is still as close as it was when we were cruising around in your Escort drinking Cherry Cokes. Additionally, thank you to all of the chemistry department graduate students. You all add to a fantastic program that feels more like a family than colleagues and each have made every day in Salem a pleasurable experience.

A special thanks to Daniel for being capable of handling all of the craziness that graduate school has brought out in me. You’ve been the one daily who encourages me and also reminds me that I should take time to enjoy life. I’m amazed by everything that we’ve encountered over the past eight years and your constant talent, humor, and enthusiasm. I’m so thankful for your presence in my life and I am excited to begin a new journey with you.

To anyone I may have left out, thank you for your contribution to making me the person I am today. I love you each, and I am blessed to have you in my life.

Page 4: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

iii

TABLE OF CONTENTS

List of Illustrations and Tables ...................................................................................................... VI

List of Abbreviations .................................................................................................................. VIII

Abstract .......................................................................................................................................... IX

Chapter 1: Introduction: A review of tungsten coil electrothermal vaporization as a sample

introduction technique in atomic spectrometry ................................................................................ 1

Introduction .................................................................................................................................. 2

Atomization Theory ..................................................................................................................... 4

Atomic Absorption Spectrometry ................................................................................................ 7

Atomic Emission Spectrometry.................................................................................................... 9

Atomic Fluorescence Spectrometry ........................................................................................... 11

Conclusion .................................................................................................................................. 15

References .................................................................................................................................. 16

Chapter 2: Design of a portable electrothermal vaporization flame atomic emission spectrometry

device for field analysis ................................................................................................................. 28

Introduction ................................................................................................................................ 30

Experimental .............................................................................................................................. 32

Instrumental ............................................................................................................................ 32

Visual Basic Program ............................................................................................................. 37

Page 5: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

iv

Reference Solutions and Standards ........................................................................................ 44

Safety Considerations ............................................................................................................. 44

Results and discussion ................................................................................................................ 45

System Optimization .............................................................................................................. 45

Limits of Detection ................................................................................................................. 45

Conclusion .................................................................................................................................. 50

References .................................................................................................................................. 51

Chapter 3: An electrothermal vaporization flame atomic emission spectrometer ......................... 54

Introduction ................................................................................................................................ 56

Experimental .............................................................................................................................. 59

Instrumental ............................................................................................................................ 59

Reference Solutions and Standards ........................................................................................ 66

Safety Considerations ............................................................................................................. 66

Results and Discussion ............................................................................................................... 67

System Optimization .............................................................................................................. 67

Gas Phase Excitation Temperature Mapping of ETV-FAES Flame ...................................... 70

Analytical Figures of Merit .................................................................................................... 71

Conclusion .................................................................................................................................. 77

References .................................................................................................................................. 78

Chapter 4: Design of a compact, aluminum, tungsten-coil electrothermal vaporization device for

inductively coupled plasma-optical emission spectrometry .......................................................... 80

Introduction ................................................................................................................................ 82

Page 6: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

v

Materials and Methods ............................................................................................................... 84

Reagents.................................................................................................................................. 84

Instrumental ............................................................................................................................ 84

Results and Discussion ............................................................................................................... 91

Optimization of Experimental Conditions .............................................................................. 91

Optimization of Torch Coolant Gas ....................................................................................... 91

Optimization of Radio Frequency Power ............................................................................... 91

Optimization of Carrier Gas ................................................................................................... 91

Limits of Detection ................................................................................................................. 96

Validation of Method .............................................................................................................. 98

Conclusion ................................................................................................................................ 100

References ................................................................................................................................ 101

Conclusion ................................................................................................................................... 104

Curriculum Vitae ......................................................................................................................... 106

Page 7: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

vi

LIST OF ILLUSTRATIONS AND TABLES

Chapter 1 .......................................................................................................................................... 1

Table I. Limits of detection of various W-coil sample introduction methods ............................ 13

Chapter 2 ........................................................................................................................................ 28

Table I. Vaporization program for the ETV-FAES Device ....................................................... 35

Figure 1. Illustration of the ETV-FAES Device ......................................................................... 36

Figure 2. ETV-FAES control electronics ................................................................................... 39

Figure 3. Screen capture image of the Visual Basic Control Program ....................................... 41

Figure 4. ETV-FAES spectrum for 5.0 mg L-1 barium and 2.0 mg L-1 rubidium with sodium

present as a contaminant ........................................................................................................... 42

Figure 5. ETV-FAES emission profile for 5.0 mg L-1 cesium, 5.0 mg L-1 barium, and 2.0 mg L-1

rubidium ..................................................................................................................................... 43

Table II. Optimized operating parameters for ETV-FAES Device ............................................ 47

Figure 6. Illustration of 6-MTH-N Cutting Tip .......................................................................... 48

Table III. ETV-FAES Limits of Detection ................................................................................. 49

Chapter 3 ........................................................................................................................................ 54

Table I. Heating program for the tungsten-coil vaporizer .......................................................... 61

Figure 1. Photograph of the ETV-FAES Device ........................................................................ 64

Figure 2. End-on view of the cutting tip during the vaporization cycle ..................................... 65

Figure 3. Variation in emission signal level with orifice diameter for a 1 µg/mL Mn solution . 68

Page 8: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

vii

Figure 4. Variation in emission signal level with ETV gas flow rate for a 1 µg/mL Mn solution

.................................................................................................................................................... 69

Figure 5. Axial temperature map of the ETV-FAES flame ....................................................... 72

Figure 6. Radial temperature map of the ETV-FAES flame ...................................................... 73

Figure 7. ETV-FAES emission spectrum for NIST SRM 2711 ................................................. 74

Table II. Figures of merit for nineteen elements of interest ....................................................... 75

Table III. Elemental Recoveries for WPS-1 and SRM 2711 ...................................................... 76

Chapter 4 ........................................................................................................................................ 80

Figure 1. Illustration of ETV cell .............................................................................................. 85

Figure 2. Photograph of the ETV cell and ICP torch connection ............................................... 86

Table I. Heating program for the tungsten-coil vaporizer .......................................................... 89

Figure 3. Time resolved emission profile of barium, manganese, and strontium found in SRM

2643e .......................................................................................................................................... 90

Table II. Optimized operating parameters .................................................................................. 93

Figure 4. Effect of Ar coolant gas flow rate on the emission intensity of 0.50 mg/L iron ......... 94

Figure 5. Effect of H2/Ar carrier gas flow rate on the emission intensity of 0.50 mg/L iron ..... 95

Table III. Figures of merit for twelve elements of interest ......................................................... 97

Table IV. Recoveries for seven elements of interest .................................................................. 99

Page 9: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

viii

LIST OF ABBREVIATIONS

Abbreviation Definition AAS Atomic Absorption Spectrometry AES Atomic Emission Spectrometry AFS Atomic Fluorescence Spectrometry

APDC Ammonium pyrrolidine dithiocarbamate CCD Charge Coupled Device CPE Cloud Point Extraction ETV Electrothermal Vaporization

ETV-FAES Electrothermal Vaporization-Flame Atomic Emission

Spectrometry FAAS Flame Atomic Absorption Spectrometry FAES Flame Atomic Emission Spectrometry

FF-AAS Flame Furnace-Atomic Absorption Spectrometry FI-AAS Flow Injection-Atomic Absorption Spectrometry

FOM Figures of Merit GFAAS Graphite Furnace Atomic Absorption Spectrometry

I Current ICP-AES Inductively Coupled Plasma Atomic Emission Spectrometry ICP-MS Inductively Coupled Plasma Mass Spectrometry

K Kelvin LOD Limit of Detection

L-PAD Large Programmable Array Detector LPM Liters per Minute

m Slope MIBK Methyl Isobutyl Ketone NPT National Pipe Thread RF Radio Frequency

SRM Standard Reference Material TRA Time Resolved Analysis

V Voltage VB Visual Basic

W-coil Tungsten Coil

Page 10: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

ix

ABSTRACT

Electrothermal vaporization has served as an alternative sample introduction method in atomic

spectroscopy for some time. Typically, electrothermal vaporizers made of graphite, but these

devices do not lend well to mass production, are expensive, and require large power supplies for

operation. In addition to these structural and operational limitations, they encounter significant

matrix interferences when using samples that contain carbide-forming species or have high

salinity or organic concentrations.

Metal atomizers (in particular the tungsten coil) are a good alternative to these graphite

electrothermal vaporizers. Tungsten coils are commercially produced, easy to use, and do not

require the large power sources of their graphite counterparts. They also are much less porous,

which allows them to easily quantify solutions with both high salt and surfactant contents.

Tungsten coils permit a large degree of user flexibility, and they have found use in both

instrumentation and sample introduction techniques.

This dissertation will focus on the use of the tungsten coil electrothermal vaporizer as a sample

introduction technique in both novel and traditional instrumentation. Here the tungsten coil has

been used as a sample introduction technique in flame atomic emission spectrometry using a

novel flame source: a welder’s torch. The torch has been configured to accept samples introduced

as vapor into a high temperature flame where the resulting emission has been collected by a

charge coupled device spectrometer. This potentially portable instrument quantified metals in

both water and soil samples and species spanned a wavelength range of 324-852 nm.

Tungsten coil electrothermal vaporization sample introduction was also used in inductively

coupled plasma optical emission spectrometry by the introduction of a new cell design into the

space formerly occupied by a spray chamber and nebulizer. This arrangement allowed for the

Page 11: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

x

multielement determination of metals in the ultraviolet and visible region with an average

detection limit improvement factor of fourteen when compared to traditional nebulization sample

introduction techniques.

Page 12: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

1

CHAPTER 1

INTRODUCTION

A Review of Tungsten Coil Electrothermal Vaporization as a Sample Introduction Technique in Atomic

Spectrometry

Summer N. Hanna, Bradley T. Jones

Department of Chemistry, Wake Forest University,

Winston-Salem, NC, 27109 USA

Note: The following manuscript was published in Applied Spectroscopy Reviews

(DOI: 10.1080/05704928.2011.582659). Stylistic variations are due to the requirements of the journal.

The manuscript was prepared by Summer N. Hanna and edited by Bradley T. Jones.

Page 13: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

2

INTRODUCTION

Electrothermal vaporization (ETV) has had widespread use throughout spectroscopy for some time. ETVs

are often made of graphite, but these ETVs require large power supplies, often do not produce isothermal

heating, and suffer from the formation of refractory compounds during atomization. In the early 1970s,

Williams, Reid, and Piepmeier identified tungsten filaments for use as atomizers in atomic spectroscopy.

These studies used commercially available, 24-Watt, tungsten light bulbs, which were chosen for their

precise construction, simplicity, and ability to measure nanogram amounts of a single dissolved element.

They were positioned in X-Y optical benches for adjustment and then used in atomic absorption

spectrometry. All viewing was done directly above the coil in conjunction with a hollow cathode lamp,

and researchers were able to obtain relatively interference free spectra from these alternative ETVs. 1, 2

Over the forty years since Williams’ discovery, tungsten coils (W-coils) have been used in a wide variety

of applications. The attributes that caused the initial employment in spectroscopy (ease of use, simple

design, affordability, and commercial manufacture) has made them widely accessible to a variety of users.

In 2002, Hou and Jones cited 179 occurrences where tungsten devices had been successfully employed in

analytical atomic spectroscopy.3 In the decade since that review, there remains significant interest in the

area of tungsten coil ETVs with work spanning atomic absorption, emission, and fluorescence

spectrometry (AAS, AES, and AFS, respectively). Advances have included the development of new

atomization cells and power supplies, the construction of new instrumentation, and the evaluation of

chemical modifiers. Applications cover areas such as the determination of the lanthanides, which are

notoriously difficult to determine, biological and environmental samples, and even the analysis of

materials used in biomedical engineering. 4-46 Their use is wide-reaching across not only the area of

atomic spectroscopy but also with notable interest coming from the developing world, in particular Brazil

and China. 9, 10, 23, 26, 27, 31, 32, 39, 41-44, 46-53

Page 14: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

3

In addition, tungsten coil devices are ideal sample introduction sources. They have been hyphenated to

systems including flow injection AAS (FI-AAS), inductively coupled plasma optical emission

spectrometry (ICP-OES), and inductively coupled plasma-mass spectrometry (ICP-MS). 54-57 The

following review will highlight the theory and use of the tungsten coil as a sample introduction method

covering the areas of AAS, AES, and AFS over the past decade since the Hou and Jones review.

Page 15: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

4

ATOMIZATION THEORY

The mechanism for atomization in the tungsten coil is a process that is rarely addressed in the literature,

despite the amount of attention given to W-coil ETV techniques. There have only been a few

investigations into the atomization process, but all findings are supportive of one another.

Initial studies into the atomization of metallic surfaces occurred during the early 1980s. Tungsten tubes

were constructed with the design of graphite furnace cylinders in mind so that they could easily be

inserted into a traditional graphite furnace atomic absorption spectrometer with only the simplest

modifications required to the instrumentation. Two tungsten strips were fabricated to form a cylindrical

tube that could be connected to an electrode within the atomizer, which had been enclosed and shielded

with an inert gas (argon). Modifications were made to a traditional graphite furnace power supply so that

it could generate the lower voltages associated with metal atomizers. An optical pyrometer was then used

to determine the surface temperature of the tungsten tube. An atomization temperature of 3250 K was in

agreement with previous temperature studies of graphite furnaces, but a significant increase in the

sensitivity occurred when using the metal atomizer. 58, 59 This increase was due to a lag between the

temperatures associated with the surface temperature and the vapor phase in graphite furnaces, but when

metal atomizers were used this lag was less pronounced. 60, 61 This explanation of the enhancement seen in

tungsten atomizers also characterized the high heating rate, near isothermal atomization, and lack of

interference encountered with graphite-metal compounds.

In the late 1990s and early 2000s Krakovská addressed the theory of atomization 64, 65 and clearly

explained the different types of reactions that occur when using a tungsten atomizer. There are three basic

reaction types that occur when atomization from a tungsten surface occurs. First is the evaporation of the

metallic analyte (1 and 2):

Page 16: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

5

M(s) M(l) M(g) (1)

M(s) M(g) (2)

The second is the reduction of metallic oxides by hydrogen from the protecting gas according to (3).

Queiroz supported these findings 62 and also proved that certain metals including sodium, potassium,

calcium, barium, and magnesium depend on the presence of hydrogen for atomization.

MO(s, l, g) + H2(g) M(s, l, g) + H2O(g) (3)

Finally the third reaction is between metallic oxides with tungsten forming tungsten oxides and the

gaseous metal analyte (4-6):

MO(s, l, g) + W(s,g) M(s, l, g) + WO(g) (4)

2MO(s, l, g) + W(s,g) 2M(s, l, g) + WO2(s, g) (5)

3MO(s, l, g) + W(s,g) 3M(s, l, g) + WO3(s, g) (6)

When this third reaction type occurs, the tungsten oxides are immediately reduced by hydrogen to

regenerate the metallic tungsten surface of the atomizer (7-9):

WO(g) + H2(g) W(s, g) + H2O(g) (7)

WO2(g) + 2H2(g) W(s, g) + 2H2O(g) (8)

WO3(g) + 3H2(g) W(s, g) + 3H2O(g) (9)

In addition to the nine reactions seen above (which occur on the tungsten surface), there are also four

additional reaction types that may occur. Reactions that take place at the wall of the atomizer between

residue components and the metallic surface after evaporization; reactions of the metallic surface of the

Page 17: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

6

atomizer with gaseous components of the system; reactions of a condensed analyte with gaseous

components of the system, in particular hydrogen; and homogenous reactions between gaseous

components of the protective gas and evaporated analytes. 3, 63, 64

Page 18: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

7

ATOMIC ABSORPTION SPECTROMETRY

Using the tungsten coil as a sample introduction method has been popular in conjunction with hydride

trapping. In AAS, this technique takes advantage of the formation of hydrides by As, Se, Sb, Bi, Ge, Sn,

Te, or Pb with NaBH4 when in acidic solutions. The sample is trapped on the tungsten coil and upon

release from the coil, the analyte is freed from the associated matrix and transported to the atomizer as a

gas. The atomizer is often a silica tube in a furnace or flame where the hydride decomposes and the

resulting absorption or emission measurement is acquired. Analytes studied by this method are often very

volatile and extremely toxic, so they are of great scientific interest.65, 66

Cankur 67 utilized hydride generation AAS for the determination of bismuth in both water and geological

materials. Bismuth was studied because of its recently determined antibacterial effectiveness towards h.

pylori, which excretes acid as a waste product and is attributed to ulcer formation within the

gastrointestinal tract.68 On-line trapping of bismuth was completed using a tungsten coil placed within

the arm of a silica T-tube atomizer. This atomizer was heated using an air-acetylene flame with samples

carried by a mixture of HCl and NaBH4. The resulting bismuthine was collected on a heated tungsten coil,

and the trapped species was found to be stable on the coil for at least 15 minutes. In order to release the

analyte, the coil was quickly heated to a temperature of 1200 °C. The quick heating made possible

through the use of the tungsten atomizer produced sharp, reproducible signals which produced limits of

detection (LODs) with an improvement factor of 150 times that of traditional flow injection hydride

generation AAS.

Permanent modifiers were utilized by both de Souza12 and Xi43 in their determination of selenium, arsenic,

and tellurium. The benefits of permanent modifiers, also known as matrix or chemical modifiers, with

tungsten coils are argued to be the same as when using them in graphite furnaces: increased sensitivity for

volatile analytes with a decrease in matrix effects and an increased atomizer lifetime. In the selenium and

Page 19: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

8

arsenic studies a rhodium-coated coil was used to analyze samples of animal blood, bovine muscle, and

drinking water.12 The modifier improved the sensitivity of arsenic by a factor of 2. When compared to an

untreated coil, selenium absorbance dropped at temperatures higher than 250 °C, but when a modified

coil was used selenium was thermally stable up to a temperature of 1400 °C.12 Tellurium was quantified

using a platinum coated tungsten coil that vaporized samples of stream sediments and soil into an

electrically heated quartz tube atomizer. Again, improved trapping was seen when using the modified coil,

and gave rise to enhanced sensitivity and LODs (Table I) when a sample volume of only 1.5 mL was

used.43

Flame furnace AAS was studied for its ability to quantify 25 volatile and semi-volatile elements. 49 This

technique used a tungsten coil atomizer connected to a quartz flame furnace positioned above a flame

burner head for heating. Some analyte loss was encountered during the transport to the quartz furnace and

was overcome by reducing the length of transport tubing. This loss is thought to occur because once the

sample is released by the coil, it is carried through the “cool” transport tubing by the carrier gas where it

may condense on the walls of the transport material. Despite this challenge, 15 of the 25 elements in

Table I saw improvements in their limits of detection. Volatile and semi-volatile metals both were

enhanced through the use of the tungsten coil sample introduction system. Part of this enhancement was

that the flame furnace no longer used energy for desolvation of the sample. With the 10 refractory

elements, no enhancement was seen because the flame furnace was not hot enough to excite these

elements as they typically utilize a nitrous-oxide-acetylene flame that burns about 1000 degrees hotter

than the air-acetylene flame used in this experiment.

Page 20: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

9

ATOMIC EMISSION SPECTROMETRY

Over the past decade tungsten coil sample introduction in atomic emission spectrometry has focused on

inductively coupled plasma techniques. Historically, sample introduction methods for ICP have been

limited to nebulization, but when using this introduction method approximately 88-97 % of the sample is

lost to waste.69 Multiple techniques have been hyphenated to ICP-OES including liquid chromatography,

ion chromatography, gas chromatography, and capillary electrophoresis, but these methods still use

nebulizers so the sample loss associated with nebulization remains. 70-73 In order to overcome this loss,

electrothermal vaporization has been used as a successful alternative. Laser ablation and graphite furnaces

ETV sample introduction sources allow nearly 100 % of the sample to be introduced to the plasma.74, 75 In

addition to these methods, tungsten coils have been successfully used as alternative sample introduction

systems for ICP analysis. Users often turn to W-coils because they do not require large power supplies or

laser systems needed by other techniques and are easy to use by comparison.

Davis focused her research of W-coil ICP-OES on the urinary excretion of cadmium as a biomarker for

diseases including osteoporosis, prostate cancer, or pancreatic cancer. 76 Three approaches to quantifying

urinary Cd were used: direct determination, chelation using ammonium pyrrolidine dithio-carbamate

(APDC) with a methyl isobutyl ketone (MIBK) extraction, and the use of Pd as a permanent modifier on

the tungsten coil. When using the direct determination method, cadmium content was quantified using

the standard addition method with and without a bismuth internal standard. Analysis was completed

using a glass tungsten coil vaporization cell with three ports; one of which fit to the ICP with an 8-inch

joint socket connected with a Teflon bushing and o-ring. The tungsten coil was heated with a laboratory

constructed power supply that was controlled with a home-written Visual Basic program. All subsequent

studies mentioned here completed by Davis used this instrumental setup. This method produced results

that agreed well with those determined by graphite furnace atomic absorption spectrometry.

Page 21: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

10

The use of APDC chelation with a MIBK extraction was then studied for its ability to chelate a variety of

elements in conjunction with the MIBK extraction as a preconcentration step. This method successfully

liberated the cadmium from the complicated matrix of urine and a bismuth internal standard was once

again used to improve precision and accuracy. The emission spectra gave rise to a Cd line at 228.8 nm

and a secondary emission line at 226. 5 nm, which allowed for increased confidence in measurements

since two locations were available to distinguish the signal from background. Chelation and extraction of

the urine samples produced relatively interference free spectra from an otherwise complicated matrix. 77

Davis continued to investigate methods of Cd determination, but focus shifted towards permanent

chemical modifiers for the tungsten coil. 78 Here palladium was used as the modifier and was adhered to

the coil by depositing small volumes of Pd solution onto the coil and drying at low currents. Samples

could then be pipetted onto the coil and the Pd modifier remained effective after 150 coil firings.

Rodríguez analyzed silicon based minerals using ultrasonic slurry sampling with a tungsten coil using

ICP-MS. 79 A quartz ETV cell was constructed and coupled to a quadrupole based ICP-MS. Samples

were sonicated for uniformity and then pipetted onto the tungsten coil and the gaseous analyte was

transported to the plasma via a glass tube. Samples of interest included micronized talc and quartz

samples. Slurry analysis allowed the user to avoid dangerous chemicals such as hydrofluoric acid in

addition to reducing sample preparation time and avoiding possible contamination. Slurry W-coil ICP-MS

did suffer from several drawbacks when compared to nebulization. LODs were 2-10 times higher than

when nebulization was used, which was thought to be because of the reduced sample volumes. Significant

matrix signal suppression also occurred, which contributed to poor LODs making standard additions

necessary for analysis. Limits of detection ranged from 1.3 to 0.002 ng g-1 and were low enough to

determine the elements of interest in the study. Concentrations found in the slurry samples were in

agreement when compared to conventional nebulization ICP-MS. Benefits of avoiding the hydrofluoric

acid through direct slurry analysis were deemed to outweigh the lack of enhancement.

Page 22: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

11

ATOMIC FLUORESCENCE SPECTROMETRY

Hydride generating analytes were again of interest in atomic fluorescence spectrometry using a tungsten

coil sample introduction system that had been coupled to an argon/hydrogen flame. As, Bi, Cd, Ge, Pb,

Sb, Se, Sn, and Te in water and rice were quantified by depositing aqueous samples onto a tungsten coil,

which then sent the vaporous sample via carrier gas to an electrically heated wire coiled around a quartz

tube atomizer. 51 This heated wire was used to ignite the sample-containing Ar/H2 carrier gas.

Fluorescence signals were then collected using a photomultiplier tube and transported to a personal

computer for processing. Some scattering interferences were seen in samples that had a naturally high salt

concentration or very complex matrix. These NaCl crystals deposited on the surface of the coil and

scattered incident light. Chemical modifiers could be used to overcome the salinity in the samples, but in

this instance simple dilution was chosen to alleviate the interferences. LODs were superior to both flame

AAS (FAAS) and ICP-OES for all elements of interest except for germanium, as shown in Table I.

Using cloud point extraction (CPE) cadmium was quantified by a W-coil coupled to an Ar/H2 flame for

fluorescence studies. Cloud point extraction is a liquid-liquid extraction and preconcentration technique

that uses surfactants to eliminate matrix effects found in complex samples. This technique is not often

used with FAAS or ICP-OES because the surfactants deposit carbon on surfaces as they are atomized,

damaging the instrumentation. The low porosity of the W-coil makes this carbon deposition less

detrimental to the instrumentation, making it an ideal sample introduction source. CPE eliminates sample

salinity, so scattering of the signal is no longer of concern when the W-coil is coupled to the Ar/H2 flame.

Water and rice samples were studied for their cadmium content, associated with concerns in China about

exposure to Cd through food and water sources from Cd-Ni battery production found within the country.80

When using CPE, faster desolvation times occurred due to the use of the tetrahydrofuran solvent, but

longer pyrolysis steps were needed to remove the high concentration of organics from the surfactant rich

separation. In addition, a Cd-dithizone complex formed during the extraction, making the atomization

Page 23: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

12

process different from mechanisms previously mentioned (atomization theory). The complex is broken by

thermal dissociation and free Cd atoms were formed through a sulfide intermediate on the atomizer in the

presence of the reducing atmosphere around the coil. This dithiozone complex enhanced Cd

measurements because it acted as a chemical modifier in conjunction with the Triton X-114 surfactant on

the surface of the tungsten coil. Through the use of this method, absolute LODs for cadmium were found

to be more sensitive compared to both vapor generation AFS and electrothermal AAS.40

A tungsten coil trapping system was coupled to a graphite fiber felt electrothermal AFS system for the

quantification of cadmium in several food materials: cabbage, spinach, tea, chicken, and rice powder. 81

The proposed graphite fiber felt ETV coupled to the W-coil trap allowed the user to place a solid sample

on the graphite ETV and then trap the vaporized sample on the tungsten coil as in hydride trapping. When

this trapping occurred, it released the cadmium free of almost all matrix interferences at a temperature of

2000 °C. The organic matrices were removed through an ashing step on the graphite and afterwards the

inorganic matrix was trapped on a room temperature coil. The complicated matrix was separated from the

analyte, which was swept into the AFS spectrometer by the carrier gas. This method allowed for sub-

picogram levels of cadmium to be determined using AFS.

Page 24: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

13

Table I. Limits of detection of various W-coil sample introduction methods

Element(s) LOD Technique Vaporization Temperatures (°C)

Application Ref.

Te 0.08 ng mL-1

Hydride trapping AAS

1630 geological samples 43

Bi 0.0027 ng mL-1

Hydride trapping AAS

1200 geological samples and water

69

Cd 0.5 µg L-1

Flame furnace AAS

2000 aqueous single element standards

52

Ag 0.9 Pb 25 Tl 10 Zn 3 Cu 2 Sb 15 Bi 15 Te 10 Sn 1000 In 10 As 30 Se 100 Au 50 Fe 250 Co 300 Ni 200 Cr 250 Mn 10 Pd 200 As 10 µg

L-1 Ar/H2 flame AFS without chemical vapor generation

1730 water and rice 51

Bi 2 Cd 0.08 Ge 50 Pb 3 Sb 3 Se 30 Sn 20 Te 3 Cd 0.01 µg

L-1 Ar/H2 flame AFS with cloud point

extraction

1900 water and rice 40

Cd 30 ng L-1 Graphite fiber felt vaporizer with

tungsten coil trap AFS

1000~2300 cabbage, spinach, tea, chicken, rice

powder

81

As 0.5 ng g-1

Ultrasonic slurry sampling ETV-

ICP-MS

NR Micronized talc, quartz

80

Page 25: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

14

Ce 0.8 Cd 0.9 U 0.3 V 0.5 Pb 1.3 Li 0.9 Rb 0.5 Sb 0.6 Ba 3.3 Cd 0.15 µg

L-1 ICP-AES NR urine 77

Cd 228.8 nm: 0.04

226.5 nm: 0.2 µg L-1

ICP-AES with ammonium

pyrrolidine dithio-carbamate chelation

NR urine 78

Cd 0.2 µg L-1

ICP-AES using Pd as a chemical

modifier

NR urine 79

NR= not reported

Page 26: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

15

CONCLUSION

Tungsten coil electrothermal vaporization as a sample introduction method has undergone a variety of

uses since its conception forty years ago. Table I has highlighted the limits of detection found in sample

introduction systems since 2002, and oftentimes W-coil methods are comparable with instrumentation

that is routinely used in the laboratory. Samples span a wide range with both biological and industrial

significance, illustrating the versatility of the technique. This body of work will focus on the use of a

tungsten coil electrothermal vaporizer as a sample introduction system for both flame atomic emission

spectrometry and inductively coupled plasma optical emission spectrometry. The accessibility of the

tungsten coil ETV allows continued advances in atomic spectroscopy, a field which some argue has

reached maturity. However, the use of this particular device illustrates that innovation from relatively

benign sources can easily open doors in scientific advancement that others assumed closed.

Page 27: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

16

REFERENCES

1. Williams, M.; Piepmeier, E. H., Commercial tungsten filament atomizer for analytical atomic

spectrometry. Analytical Chemistry 1972, 44, (7), 1342-1344.

2. Reid, R. D.; Peipmeier, E. H., Horizontal tungsten coil atomizer and integrated absorbance

readout for atomic spectrometry. Analytical Chemistry 1976, 48, (2), 338-341.

3. Hou, X. D.; Jones, B. T., Tungsten devices in analytical atomic spectrometry. Spectrochimica

Acta Part B-Atomic Spectroscopy 2002, 57, (4), 659-688.

4. Alp, O.; Ertas, N., In situ trapping of antimony hydride on iridium-coated tungten coil and

interference studies. Journal of Analytical Atomic Spectrometry 2008, 23, 976-980.

5. Alp, O.; Ertas, N., Determination of inorganic and total mercury by flow injection vapor

generation atomic absorption spectrometry using a W-coil atomizer. Journal of Analytical Atomic

Spectrometry 2009, 24, 93-96.

6. Alp, O.; Ertas, N., Determination of tin by in situ trapping of stannane on a resistively heated

iridium treated tungsten coil surface and interference studies. Talanta 2010, 81, 516-520.

7. Ayers, R.; Ferguson, V.; Belk, D.; Moore, J., Self-propagating high-temperature synthesis of

porou nickel-titanium. Materials Science Forum 2007, 561-565.

8. Bings, N. H.; Stefanka, Z., Development of a tungsten filament electrothermal vaporizer for

inductively coupled plasma time-of-flight mass spectrometry and its possibilities for the analysis of

human whole blood and serum. Journal of Analytical Atomic Spectroscopy 2003, 18, 1088-1096.

9. Bruhn, C. G.; Bustos, C. J.; Saez, K. L.; Neira, J. Y.; Alvarez, S. E., A comparative study of

chemical modifiers in the determination of total arsenic in marine food by tungsten coil electrothermal

atomic absorption spectrometry. Talanta 2007, 71, 81-89.

10. Bruhn, C. G.; Huerta, V. N.; Neira, J. Y., Chemical modifiers in arsenic determination in

biological materials by tungsten coil electrothermal atomic absorption spectrometry. Analytical and

Bioanalytical Chemistry 2004, 378, 447-455.

Page 28: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

17

11. Cankur, O.; Yavuz, A. O., Chemical vapor generation of Cd and on-line preconcentration on a

resistively heated W-coil prior to determination by atomic absorption spectrometry using an unheated

quartz absorption cell. Journal of Analytical Atomic Spectrometry 2007, 22, 791-799.

12. de Souza, S. S.; Santos, D.; Krug, F. J.; Barbosa, F., Exploiting in situ hydride trapping in

tungsten coil atomizer for Se and As determination in biological and water samples. Talanta 2007, 73,

451-457.

13. Donati, G. L.; Calloway, C. P.; Jones, B. T., Double tungsten coil atomic emission spectrometry:

signal enhancement and a new gas phase temperature probe. Journal of Analytical Atomic Spectrometry

2009, 24, (8), 1105-1110.

14. Donati, G. L.; Gu, J.; Nobrega, J. A.; Calloway, C. P.; Jones, B. T., Simultaneous determination

of the Lanthanides by tungsten coil atomic emission spectrometry. Journal of Analytical Atomic

Spectrometry 2008, 23, (3), 361-366.

15. Donati, G. L.; Kron, B. E.; Jones, B. T., Simultaneous determination of Cr, Ga, In and V in soil

and water samples by tungsten coil atomic emission spectrometry. Spectrochimica Acta Part B-Atomic

Spectroscopy 2009, 64, (6), 559-564.

16. Donati, G. L.; Nobrega, J. A.; Nascentes, C. C.; Jones, B. T., Indirect determination of iodide by

tungsten coil atomic emission spectrometry. Microchemical Journal 2009, 93, (2), 242-246.

17. Donati, G. L.; Pharr, K. E.; Calloway, C. P.; Nobrega, J. A.; Jones, B. T., Determination of Cd in

urine by cloud point extraction-tungsten coil atomic absorption spectrometry. Talanta 2008, 76, (5),

1252-1255.

18. Donati, G. L.; Gonzales, M. H.; Nobrega, J. A.; Jones, B. T., Multi-wavelength determination of

cobalt by tungsten coil atomic emssion spectrometry. Analytical Letters 2010, 43, 1723-1733.

19. Donati, G. L. K., B.E.; Jones, B.T., Simultaneous determination of Cr, Ga, In, and V in soil and

water samples by tungsten coil atomic emission spectrometry. Spectrochimica Acta Part B 2009, 64B,

559-564.

Page 29: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

18

20. Donati, G. L. N., J.A.; Nascentes, C.C.; Jones, B.T., Indirect determination of iodide by tungsten

coil atomic emission spectrometry. Microchemical Journal 2009, 93, 242-246.

21. Donati, G. L.; Wildman, R. B.; Jones, B. T., A new atomization cell for trace metal

determinations by tungsten coil atomic spectrometry. Analytica Chimica Acta 2011, 688, 36-42.

22. Ezer, M.; Elwood, S. A.; Jones, B. T.; Simeonsson, J. B., Evaluation of a tungsten coil

atomization-laser-induced fluorescence detection approach for trace elemental analysis. Analytica

Chimica Acta 2006, 571, (1), 136-141.

23. Gonzalez, M. H.; Santos, L. N.; Nobrega, J. A.; Rocha, F. R. P., Construction of a tungsten coil

atomic emission spectrometer (WCAES). Quimica Nova 2010, 33, 2266-2271.

24. Gu, J.; Donati, G. L.; Young, C. G.; Jones, B. T., Continuum source tungsten coil atomic

fluorescence spectrometry. Applied Spectroscopy 2011, 65, 382-385.

25. Gu, J.; Oliveira, S. R.; Donati, G. L.; Gomes Neto, A. G.; Jones, B. T., Rugged, Portable

Tungsten Coil Atomic Emission Spectrometer. Analytical Chemistry 2011.

26. He, S. P.; Fan, G. Y.; Jiang, X. M.; Hou, X. D., Determination of trace zinc by W-coil

electrothermal atomic absorption spectrometry. Fenxi Huaxue 2010, 38, 707-710.

27. Jiang, D.; Wen, X.; Xu, J.; Hou, X.; Huang, H., The power supply for miniature tungsten coil

electrothermal atoic absorption spectrometer. Jisuanji Yu Yingyong Huaxue 2009, 26, 873-877.

28. Kula, I.; Arslan, Y.; Bakirdere, S.; Ataman, O. Y., A novel analytical system involving hydride

generation and gold-coated W-coil trapping atomic absorption spectrometry for selenium determination at

ng l-1 level. Spectrochimica Acta Part B 2008, 63B, 856-860.

29. Liu, R.; Wu, P.; Maoyang, M.; Xu, K.; Lv, Y., Inorganic Arsenic Speciation Analysis of Water

Samples by Trapping Arsine on Tungsten Coil for Atomic Fluorescence Spectrometric Determination.

Talanta 2009, 78, 885-890.

30. Liu, R.; Wu, P.; Xi, M.; Xu, K.; Lv, Y., Inorganic arsenic speciation analysis of water samples by

trapping arsine on tungsten coil for atomic fluorescence spectrometric determinatoin. Talanta 2009, 78,

885-890.

Page 30: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

19

31. Neira, J. Y.; Mendoza, J.; Bruhn, C. G., Flow analysis by using slenoid valves for As (III)

determination in natural waters by an on-line separation and pre-concentration system coupled to a

tungsten coil atomizer. Quimica Nova 2005, 28, 217-223.

32. Nobrega, J. A.; Rust, J.; Calloway, C. P.; Jones, B. T., Evaluation of a continuum source

Tungsten coil atomic absorption spectrometer: A study of Zn behavior. Journal of the Brazilian Chemical

Society 2005, 16, (3B), 639-642.

33. Rust, J. A.; Nobrega, J. A.; Calloway, C. P.; Jones, B. T., Fraunhofer effect atomic absorption

spectrometry. Analytical Chemistry 2005, 77, (4), 1060-1067.

34. Rust, J. A.; Nobrega, J. A.; Calloway, C. P.; Jones, B. T., Analytical characteristics of a

continuum-source tungsten coil atomic absorption spectrometer. Analytical Sciences 2005, 21, (8), 1009-

1013.

35. Rust, J. A.; Nobrega, J. A.; Calloway, C. P.; Jones, B. T., Advances with tungsten coil atomizers:

Continuum source atomic absorption and emission spectrometry. Spectrochimica Acta Part B-Atomic

Spectroscopy 2005, 60, (5), 589-598.

36. Rust, J. A.; Nobrega, J. A.; Calloway, C. P.; Jones, B. T., Tungsten coil atomic emission

spectrometry. Spectrochimica Acta Part B-Atomic Spectroscopy 2006, 61, (2), 225-229.

37. Simeonsson, J. B.; Elwood, S. A.; Ezer, M.; Swart, D. J.; Beach, H. D.; Thomas, D. J.,

Development of ultratrace laser spectrometry techniques for measurements of arsenic. Talanta 2002, 58,

189-199.

38. Titretir, S.; Kenduezler, E.; Arslan, Y.; Kula, I.; Bakirdere, S.; Ataman, O. Y., Determination of

antimony by using tungsten trap atomic absorption spectrometry. Spectrochimica Acta Part B 2008, 63B,

875-879.

39. Wang, J. W.; He, Y. H.; Hu, J.; Lu, Y.; Hou, X. D., Ultrasonic assisted-acid leaching as a sample

preparation technique for determination of cadmium in rice by tungsten coil electrothermal atomic

absorption spectrometry. Huaxue Yanjiu Yu Yingyong 2007, 19, 430-432.

Page 31: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

20

40. Wen, X.; Wu, P.; Chen, L.; Hou, X., Determination of cadmium in rice and water by tungsten coil

electrothermal vaporization-atomic fluorescence spectrometry and tungsten coil electrothermal atomic

absorption spectrometry after coud point extraction. Analytica Chimica Acta 2009, 650, 33-38.

41. Wen, X.; Wu, P.; He, Y. H.; Xu, K. L.; Lu, Y.; Hou, X., Portable atomic absorption spectrometer

for determination of copper, chromium, lead, and cadmium in environmental water. Fenxi Huaxue 2009,

37, 772-775.

42. Wu, P.; He, Y.; Xie, W.; Li, M.; Hou, X., Simultaneous determination of calcium and strontium

in brine by tungsten coil electrothermal atomic emission spectometry. Sichuan Daxue Xuebao, Ziran

Kexueban 2006, 43, 1157-1161.

43. Xi, M.; Liu, R.; Wu, P.; Xu, K.; Hou, X.; Lv, Y., Atomic absorption spectrometric determination

of trace tellurium after hydride trapping on platinum-coated tungsten coil. Microchemical Journal 2010,

95, 320-325.

44. Xu, J. Y.; Lin, Y. C.; Zeng, B.; Hu, Z. W.; Ge, L. Q., Study on tungsten coil atomic absprption

spectrometer based on national instruments. Fenxi Shiyanshi 2007, 26, 103-106.

45. Zeng, C.; Wen, X.; Tan, Z.; Cai, P.; Hou, X., Hollow Fiber Supported Liquid Membrane

Extraction for Ultrasensitive Determination of Trace Lead by Portable Tungsten Coil Electrothermal

Atomic Absorption Spectrometry. Microchemical Journal 2010, 96, 238-242.

46. Zhang, L. P.; Wu, P. l.; He, L.; Chen, H.; Hou, X. D., Determination of As (III) and As (V) in

water samples by ETV-Ar/H2 flame-AFS after cloud point extraction. Xiran Kexueban 2007, 44, 645-648.

47. Pinto, F. G.; Lepri, F. G.; Saint'Pierre, T. D.; da Silva, J. B. B.; Costa, L. M.; Curtius, A. J., Direct

determination of Dy, Sm, Eu, Tm, and Yb in geological samples by slurry electrothermal vaporization

inductively coupled plasma-mass spectrometry. Analytical Letters 2010, 43, (6), 949-959.

48. Wu, P., Wen, X., He, L., He, Y., Chen, M., Hou, X., Evaluation of Tungsten Coil electrothermal

vaporization-Ar/H2 flame atomic fluorescence spectrometry for determination of eight traditional hydride-

forming elements and cadmium without chemical vaporization generation. Talanta 2008, 74, 505-511.

Page 32: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

21

49. Wu, P.; Zhang, Y.; Liu, R.; Lv, Y.; Hou, X., Comparison of tungsten coil electrothermal

vaporization and thermospray sample introduction methods for flame furnace atomic absorption

spectrometry. Talanta 2009, 77, 1778-1782.

50. Hu, Z. W.; Wang, T.; Zhao, J.; Xu, J. Y., Determination of lead in water by electrodeposition-

tungsten coil electrothermal atomic absorption spectrometry. Fenxi Shiyanshi 2010, 29, 42-44.

51. Wu, P.; Wen, X.; He, L.; He, Y.; Chen, M.; Hou, X., Evaluation of Tungsten Coil electrothermal

vaporization-Ar/H2 flame atomic fluorescence spectrometry for determination of eight traditional hydride-

forming elements and cadmium without chemical vaporization generation. Talanta 2008, 74, 505-511.

52. Xu, J.; Zeng, G.; Ge, L., Determination of trace gold in geological samples by electrodeposition-

tungsten coil electrotermal atomic absprption spectrometry. Huaxue Yanjiu Yu Yingyong 2008, 20, 495-

498.

53. Zeng, C., Wen, X., Tan, Z., Cai, P., Hou, X., Hollow fiber supported liquid membrane extraction

for ultrasensitive determination of trace lead by portable tungsten coil electrothermal atomic absorption

spectrometry. Microchemical Journal 2010, 96, 238-242.

54. Alp, O. E., N., Determination of Inorganic and Total Mercury by Flow Injection Vapor

Generation Atomic Absorption Spectrometry Using a W-Coil Atomizer. Journal of Analytical Atomic

Spectrometry 2009, 24, 93-96.

55. Amberger, M. A., Broekaert, J.A.C., Direct mulitelement determination of trace elements in

boron carbide powders by slurry sampling ETV-ICP-OES. Journal of Analytical Atomic Spectrometry

2010, 35, 1308-1315.

56. Barth, P.; Hauptkorn, S.; Krivan, V., Analysis of silicon dioxide and silicon nitride powders by

electrothermal vaporization inductively coupled plasma atomic emission spectrometry using a tungsten

coil and slurry sampling. Journal of Analytical Atomic Spectroscopy 1997, 12, 1359-1365.

57. Bings, N. H. S., Z., Development of a tungsten filament electrothermal vaporizer for inductively

coupled plasma time-of-flight mass spectrometry and its possibilities for the analysis of human whole

blood and serum. Journal of Analytical Atomic Spectroscopy 2003, 18, 1088-1096.

Page 33: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

22

58. Puschel, P. F., Z., Electrothermal atomization from metallic sufaces. Analytica Chimica Acta

1981, 127, 109-120.

59. Berndt, H., Schaldach, G., Simple low-cost tungsten-coil atomizer for electrothermal atomic

absorption spectrometry. Journal of Analytical Atomic Spectrometry 1988, 3, 709-712.

60. Chakrabarti, C. L., Hamed, H.A.; Wan, C.C.; Li, W.C.; Bertels, P.C.; Gregoire, D.C.; Lee, S.,

Capacitive discharge heating in graphite furnace atomic absorption spectrometry. Analytical Chemistry

1980, 52, 167-176.

61. Gregoire, D. C. C., C.L.; Bertels,P.C., Effect of heating rates in graphite furnace atomic

absorption spectrometry. Analytical Chemistry 1978, 50, 1730-1737.

62. Queiroz, Z. F., Krug, F.; Oliveira, P.V., Silva, C.S., Nobrega, J.A., Electrothermal behavior of

sodium, potassium, calcium, and magnesium in a tungsten coil atomizer and review of interfering effects.

Spectrochimica Acta Part B 2002, 57B, 49-61.

63. Krakovska, E. R., D., Tungsten atomizer-- the theory of atomization reactions. Spectrochimica

Acta Part B 1997, 52, 1327-1332.

64. Krakovska, E.; Remeteiova, D., Tungsten atomizer-- theory of atomizatoin mechanism of some

volatile analytes. Fresenius Journal of Analytical Chemistry 2000, 366, 127-131.

65. Skoog, D. A.; Holler, F. J.; Nieman, T. A., Principles of Instrumental Analysis. In 5th ed. ed.;

Thompson Learning: 1998; p 203.

66. Ingle, J. D.; Crouch, S. R., Spectrochemical Analysis. Prentice Hall: Upper Saddle River, NJ,

USA, 1998.

67. Cankur, O.; Ertas, N.; Ataman, Y., Determination of bismuth using on-line preconcentration by

trapping on resistively heated W-coil and hydride generation atomic absorption spectrometry. Journal of

Analytical Atomic Spectroscopy 2002, 17, 603-609.

68. Cendron, L.; Zanotti, G., Structural and functional aspects of unique type IV secretory

components in the Helicobacter pylori cag-pathogenicity island. FEBS Journal 2011, 278, (8), 1223-1231.

Page 34: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

23

69. Okamoto, Y.; Komori, H.; Kataoka, H.; Tsukahara, S.; Fujiwara, T., Direct determination of lead

in biological samples by electrothermal vaporization inductively coupled plasma mass spectrometry

(ETV-ICP-MS) after furnace-fusion in the sample cuvette-tungsten boat furnace. Fresenius Journal of

Analytical Chemistry 2000, 367, 300-305.

70. Coetzee, P. P.; Fischer, J. L.; Hu, M., Simultaneous separation and determination of Tl (I) and Tl

(III) by IC-ICP-OES and IC-ICP-MS. Water SA 2003, 29, 17-22.

71. Whalen, R.; Wolff-Briche, C., Comparison of GC-ICP-MS and HPLC-ICP-MS for species-

specific isotope dilution analysis of tributyltin in sediment after accelerate solvent extraction. Analytical

and Bioanalytical Chemistry 2003, 377, 140-148.

72. Profrock, D.; Lenohard, P.; Prange, A., Determination of sulfur and selected trace elements in

metallothionein-like proteins using capillary electrophoresis hyphenated to inductively coupled plasma

mass spectrometry with an octopole reaction cell. analytical and Bioanalytical Chemistry 2003, 377, 132-

139.

73. Martinez-Bravo, Y.; Roig-Navarro, A. F.; Lopez, F. J.; Hernandez, F., Multielemental

determination of arsenic, selenium, and chromium (VI) species in water by high-performance liquid

chromatography-inductively coupled plasma mass spectrometry. Journal of Chromatography A 2001, 926,

265-274.

74. Giussani, B.; Monticelli, D.; Rampazzi, L., Role of laser ablation-inductively coupled plasma-

mass spectrometry in cultural heritage research: A review. Analytica Chimica Acta 2009, 635, 6-21.

75. Resano, M.; Vanhaecke, F.; de Loos-Vollebregt, M. T. C., Electrothermal vaporization for

sample introduction in atomic absorption, atomic emission, and plasma mass spectrometry-- a critical

review with focus on solid sampling and slurry analysis. Journal of Analytical Atomic Spectroscopy 2008,

23, 1450-1475.

76. Davis, A. C.; Alligood, B. W.; Calloway, C. P.; Jones, B. T., Direct determination of cadmium in

urine by electrothermal vaporizer-inductively coupled plasma analysis using a tungsten coil vaporizer.

Applied Spectroscopy 2005, 59, (10), 1300-1303.

Page 35: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

24

77. Davis, A. C.; Calloway, C. P.; Jones, B. T., Chelation of urinary cadmium with ammonium

pyrrolidine dithio-carbamate prior to determination by tungsten-coil inductively coupled plasma atomic

emission spectrometry. Microchemical Journal 2006, 84, (1-2), 31-37.

78. Davis, A. C.; Calloway, C. P.; Jones, B. T., Direct determination of cadmium in urine by

tungsten-coil inductively coupled plasma atomic emission spectrometry using palladium as a permanent

modifier. Talanta 2007, 71, (3), 1144-1149.

79. Rodriguez, P. F.; Marchante-Gayon, J. M.; Sanz-Medel, A., Elemental analysis of silicon based

minerals by ultrasonic slurry sampling electrothermal vaporization ICP-MS. Talanta 2006, 68, 869-875.

80. Sun, L. N.; Zhang, Y. H.; Sun, T. H.; Gong, Q.; Lin, X.; Li, H. B., Temporal-spatial distribution

and variability of cadmium contamination in soils in Shenyang Zhangshi irrigation area. Journal of

Environmental Science 2006, 18, 1241-1246.

81. Feng, L.; Liu, J., Solid sampling graphite fibre felt electroterhmal atomic fluorescence

spectrometry with tungsten coil atomic tramp for the determination of cadmium in food samples. Journal

of Analytical Atomic Spectrometry 2010, 25, 1072-1078.

82. Klaic, P. M. A. N., A.M.; Moreira, A.S.; Vendruscolo, C.T.; Ribeiro, A.S., Determination of Na,

K, Ca and Mg in xanthan gum: Sample treatment by acid digestion. Carbohydrate Polymers 2011, 83,

1895-1900.

83. Sokolnikova, J. V. V., I.E.; Menshikov, V.I., Determination of Trace Alkaline Metals in Quartz

by Flame Atomic Emission and Atomic Absorption Spectrometry. Spectrochim. Acta, Part B 2003, 58,

387-391.

84. Miller-Ihli, N. J., Atomic Absorption and Atomic Emission Spectrometry for the Determination

of the Trace Element Content of Selected Fruits Consumed in the United States. J. Food Comp. Anal.

1996, 9, 301-311.

85. Ingle, J. D. C., S.R., Spectrochemical Analysis. Prentice Hall: Upper Saddle River, NJ, USA,

1998.

Page 36: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

25

86. Pinto, F. G. L., F.G.; Saint'Pierre, T.D.; da Silva, J.B.B.; Costa, L.M.; Curtius, A.J., Direct

Determination of Dy, Sm, Eu, Tm, and Yb in Geological Samples by Slurry Electrothermal Vaporization

Inductively Coupled Plasma Mass Spectrometry. Anal. Lett. 2010, 43, (6), 949-959.

87. Lee, Y. I. K., J.K.; Kim, K.H.; Yoo, Y.J.; Back, G.H.; Lee, S.C., Design and Critical Evaluation

of Improved Electrothermal Vaporization Flame Atomic Absorption/Emission Spectrometry for Direct

Determination of Trace Metals in Microliter Samples. Microchem. J 1998, 60, 231-241.

88. Okamoto, Y. K., H.; Kataoka, H.; Tsukahara, S.; Fujiwara, T., Direct determination of bromine in

plastics by electrothermal vaporization /inductively coupled plasma mass spectrometry using a tungsten

boat furnace vaporizer and an exchangeable sample cuvette system. Rapid Commun. Mass Spectrom.

2010, 24, (9), 1265-1270.

89. Williams, M. P., E.H., Commercial Tungsten Filament Atomizer for Analytical Atomic

Spectrometry. Anal Chem 1972, 44, (7), 1342-1344.

90. Reid, R. D. P., E.H., Horizontal Tungsten Coil Atomizer and Integrated Absorbance Readout for

Atomic Spectrometry. Anal. Chem. 1976, 48, (2), 338-341.

91. Hou, X.; Levine, K. E.; Salido, A.; Jones, B. T.; Ezer, M.; Elwood, S.; Simeonsson, J. B.,

Tungsten coil devices in atomic spectrometry: absorption, fluorescence, and emission. Analytical Sciences

2001, 17, 175-180.

92. Barth, P. H., S.; Krivan, V., Analysis of Silicon Dioxide and Silicon Nitride Powders by

Electrothermal Vaporization Inductively Coupled Plasma Atomic Emission Spectrometry Using a

Tungsten Coil and Slurry Sampling. J. Anal. At. Spectrom. 1997, 12, 1359-1365.

93. Bings, N. H. S., Z., Development of a tungsten filament electrothermal vaporizer for inductively

coupled plasma time-of-flight mass spectrometry and its possibilities for the analysis of human whole

blood and serum. J. Anal. At. Spectrom. 2003, 18, 1088-1096.

94. Spaargaren, D. H., Presence and Significance of Lithium in the Brown Shrimp, Crangon crangon.

Comparative Biochemistry and Physiology, Part A: Molecular and Integrative Physiology 1998, 91, 457-

461.

Page 37: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

26

95. Koji, M. K., T.; Yamaya, H.; Kobayashi, E.; Nogawa, K., Urinary Excretion Levels of Sodium

and Potassium in Envrionmental Cadmium-Exposed Subjects. Toxicology 1998, 127, 187-193.

96. Shiraishi, K. K., S.; Zamostyan, P.V.; Tsigankov, N.Y.; Los, I.P.; Koruzun, V.N., Dietary intakes

of alkaline metals, alkaline earths, and phosphorous in Ukrainians. Biomedical Research on Trace

Elements 2009, 20, 62-68.

97. Donati, G., Gu, J., Nobrega, J.A., Calloway, C.P. Jr., Jones, B.T., J. Anal. At. Spectrom. 2008, 23,

361-366.

98. Nakamura, S.; Kobayashi, Y.; Kubota, M., Temperature of a W Ribbon Furnace in

Electrothermal Atomic Absorption Spectrometry. Spectrochimica Acta Part B 1986, 41, 817-823.

99. Alavosus, T. J.; Murphy, R.; Schatzlein, D., Am. Lab. 1995, 23, 31-34.

100. Christian, G. D., Feldman, F.J., A comparison study of detection limits using flame-emission

spectroscopy with the nitrous oxide-acetylene flame and atomic-absoroption spectroscopy. Applied

Spectroscopy 1971, 25, 660-663.

101. Ramirez-Munoz, J., Spectrochemical Methods of Analysis. Wiley-Interscience: New York, NY,

1971.

102. Bruno, T. J., Svoronos, P.D.N., Handbook of Basic Table for Chemical Analysis. CRC Press:

Boca Raton, FL, 1989.

103. Robinson, J. W., Undergraduate Instrumental Analysis. 5th ed.; Marcel Dekker: New York, NY,

1995.

104. Gaines, P., Sample introduction methods for ICP-MS and ICP-OES. Spectroscopy 2005, (January

2005), 20-23.

105. Nickel, H., Zadgorska, Z., A strategy for calibrating direct ETV-ICP-OES analysis of industrial

ceramics in powder form. Fresenius' Journal of Analytical Chemistry 1995, 351, 158-163.

106. Detcheva, A., Barth, P., Hassler, Calibration possiblities and modifier use in ETV-ICP-OES

determination of trace and minor elements in plant materials. Analytical Bioanalytical Chemistry 2009,

394, 1485-1495.

Page 38: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

27

107. Li, S., Hu, B., Jiang, Z., Chen, R., Direct determination of trace refractory elements in human

serum by ETV-ICP-MS with in-situ matrix removal. analytical Bioanalytical Chemistry 2004, 379, 1076-

1082.

108. Nakamura, S. K., Y., Kubota, M., Temperature of a W ribbon furnace in electrothermal atomic

absorption spectrometry. Spectrochimica Acta Part B 1986, 41, 817-823.

109. Nakamura, Y., Takahashi, K., Kujirai, O., Okochi, H., Evaluation of electrothermal vaporization,

emission intensity-tie-wavelength measurement and time resolution combined with an axially viewed

horizontal inductively coupled plasma using an echelle spectrometer with wavelength modultaion.

Journal of Analytical Atomic Spectrometry 1997, 12, 349-354.

110. Shibata, N.; Fudagawa, N.; Kubota, M., Electrothermal vaporization using a tungsten furnace for

the determination of rare-earth elements by inductively coupled plasma mass spectrometry. Analytical

Chemistry 1991, 63, 636-640.

111. Hanna, S. N.; Keene, J.; Calloway, C. P.; Jones, B. T., Design of a Portable Eelctrothermal

Vaporization Flame Atomic Emission Spectrometry Device for Field Analysis. Instrumentation Science

and Technology 2011, (in press).

112. Gras, L. d. L.-V., M.T.C., Limiting effects on transport efficiency by matrix load in inductively

coupled plasma optical emission spectrometry with electrothermal vaporization sample introduction.

Spectrochimica Acta Part B 2000, 55, 37-47.

113. Maestre, S. d. L.-V., M.T.C., Plasma behavior during electrothermal vaporization sample

introduction in inductively coupled plasma atomic emission spectrometry. Spectrochimica Acta Part B

2001, 56, 1209-1217.

Page 39: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

28

CHAPTER 2

Design of a Portable Electrothermal Vaporization Flame Atomic Emission Spectrometry Device for Field

Analysis

Summer N. Hanna,a Joseph Keene,a Clifton P. Calloway, Jr.,b Bradley T. Jonesa

a Department of Chemistry, Wake Forest University, Winston-Salem, North Carolina, 27109 USA

b Department of Chemistry, Physics, and Geology, Winthrop University, Rock Hill, South Carolina, 29733 USA

Note: The following manuscript was written for publication in Instrumentation, Science, and Technology.

Stylistic variations are due to the requirements of the journal. All of the presented research was conducted

by Summer N. Hanna with the exception of the results for lithium which were generated by Joseph Keene

under the direction of Summer Hanna. Computer programming was completed by Clifton P. Calloway, Jr.

The manuscript was prepared by Summer N. Hanna and edited by Bradley T. Jones.

Page 40: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

29

ABSTRACT

A non-traditional, portable flame atomic emission spectrometry device has been constructed using a

welding industry metal cutting torch as the flame source with a tungsten coil electrothermal vaporization

device for sample introduction. Twenty-microliter sample aliquots were pipetted onto a commercially

available tungsten filament and a home-written Visual Basic program was used to automatically dry and

atomize the sample. The sample then traveled into the heart of the air-acetylene flame where it was

excited, and the resulting emission signal was focused onto the entrance of a miniature charge coupled

device spectrometer. Figures of merit were determined for five alkaline and alkali earth metals of

biological and environmental significance. Limits of detection were, on average, 100 times lower than

those for traditional flame emission spectrometers. The capabilities of this portable device make it an

ideal tool for biological field analysis, particularly in areas where traditional laboratories are inaccessible.

KEYWORDS

Electrothermal Vaporization, Portable Instrumentation, Atomic Emission Spectrometry, Welding Torch

Page 41: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

30

INTRODUCTION

Most flame atomic emission spectrometry (FAES) devices are inexpensive, require little user training,

and are convenient to use in a wide variety of settings. Even as a mature technique, FAES remains a

standard in trace elemental analysis.82 This method is well-suited to alkaline and alkali earth metals

(which have significant biological and environmental importance) and is capable of analyzing a wide

variety of samples.83, 84 However, conventional flame systems often do not produce the energy necessary

to excite most elements on the periodic table, they are highly reactive and prone to interferences, and they

are limited in their sample introduction capabilities.85 This makes flame atomic spectroscopy methods

well-suited for the investigation of alternative sample introduction approaches.

Electrothermal vaporization (ETV) is a relatively common sample introduction technique for atomic

spectroscopy. ETVs come in a myriad of designs, including tubes, boats, wires, and filaments. ETVs have

been fashioned from an array of materials such as tantalum, graphite, or tungsten, with graphite being the

most popular. 48, 86-88 However, graphite ETVs require large power supplies and often suffer from memory

effects. Metal ETVs are a good alternative because of their lower power requirements, high melting

points, and chemical inertness. Tungsten has the highest melting point of any element at 3422°C, can

support chemical modifiers, and is easily manipulated into different shapes and arrangements. Tungsten

filaments have been used as ETVs since the early 1970s. 89-91 These filaments are commercially available

as incandescent light bulbs, and hence are mass produced to strict specifications. As a result, they are low

in cost and simple to obtain. Similar ETV devices have been reported in numerous applications: flame

furnace atomic absorption spectrometry (FF-AAS), Inductively Coupled Plasma Atomic Emission

Spectrometry (ICP-AES), and Inductively Coupled Plasma Mass Spectrometry (ICP-MS). 48, 92, 93 One

advantage of the ETV is that it desolvates the aqueous samples, and provides a dry aerosol plug to the

excitation source. The flame, therefore, only provides energy to atomize the dry aerosol, and then excite

the emission. In the end, more energy is available to excite the emission, and an enhanced signal is

observed.

Page 42: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

31

In this work, a non-traditional FAES source, a welding-industry metal cutting torch, will be coupled to a

tungsten coil ETV. The traditional use of the flame source will be altered, and the ETV will act as the

sample introduction system. An ETV cell is coupled to the torch, which allows a steady stream of 10%

hydrogen / 90 % argon gas to introduce aerosol samples as discrete plugs into the heart of an acetylene-air

flame. This system will allow for the simultaneous determination of several elements, in this case alkali

and alkaline earth metals. These species have dietary and environmental significance, making them ideal

analytes for field analysis techniques.94-96

Page 43: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

32

EXPERIMENTAL

Instrumental

A Victor® oxygen-acetylene machine cutting torch is used as the flame source. The torch is designed to

create a heating flame to melt a metal surface. It is composed of a brass body with three gas inlet lines.

The heating flame is supplied via a flashback-protected fuel (acetylene) port and a flashback-protected

oxidant (air) port. These two gases are kept separate throughout the barrel length of the torch, mixing

instead in the cutting tip at the apex of the device. The tip divides the pre-mixed flame gases into ten

equal streams arranged in a circle approximately 1 cm in diameter. The streams produce ten high

temperature flame jets that merge together about 1 cm away from the tip surface. The third gas port is

intended for “cutting” oxygen. This high velocity stream of oxygen is not mixed with the flame gases in

the tip. Instead, it passes through a dedicated central channel, exits the torch at the end of the tip, and

pierces the flame. As it passes through the flame, it is heated and oxidizes the metal being cut, resulting

in a slag jet. Cutting tips vary by the size of their central channel orifice diameter and are industry

standardized. Typically, the smaller the inner channel diameter, the smaller the thickness of the metal that

is to be cut. Larger central channels also require higher cutting gas flow rates.

The ETV-FAES device described here does not use the cutting oxygen. Instead, the central channel is

connected to an aluminum laboratory-designed electrothermal vaporization cell. The tungsten filament

inside the cell is bathed by a stream of 10% hydrogen / 90% argon gas, which prevents oxidation of the

filament and acts as a carrier gas. Samples are deposited onto the filament and then dried automatically

using a variable low current. Once dryness is achieved, a previously optimized vaporization program is

applied (Table I).97 When the sample is vaporized, the detector is triggered, and the carrier gas introduces

the sample into the central channel of the torch and directly into the flame as a concentrated aerosol plug.

Here the atoms in the sample are thermally excited and upon their relaxation, emission occurs through the

release of a photon at a characteristic wavelength.

Page 44: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

33

The Victor® MT 310A three hose machine cutting torch (Victor Professional, Part Number 0380-0222, St.

Louis, MO, USA) is used as the flame source. This torch has a total length of 355.6 mm. The barrel,

which houses the three “hoses” for air, acetylene, and the sample carrier gas, has a length of 254.0 mm.98

Different fuel and oxidant gases may be employed depending upon the specific torch acquired. The torch

is secured on an adjustable rail, positioned axially with respect to the detector, and connected to air,

acetylene, and sample gases. Historically, axial viewing improves the figures of merit for ICP

spectrometry, so this viewing arrangement was chosen for the ETV-FAES device.99

The ETV is connected to a solid state constant current power supply (BatMod, Part Number VI-LU1-EU-

BM, Vicor, Andover, MA, USA). The supply provides up to 200 W of power (15 V, 13.3 A max), and

may be configured to run from 110 VAC or from 12 VDC (for portable applications). The supply is

small: 235 x 65 x 35 mm (including cooling fins). The supply is controlled with a home-written Visual

Basic program and a digital-to-analog converter (Measurement Advantage MiniLab 1008, Measurement

Computing, Norton, MA, USA). The tungsten filament is carefully extracted from a commercially

available 150 W, 15 V projector light bulb (Osram HLX 64633 Xenophot, Augsburg, Germany). The W-

coil filament is exposed by removing the fused silica envelope while leaving the glass base intact. The

filament is coiled in ten turns and when viewed from the side, the turns are 3 mm high and together they

occupy a width of 6 mm. Viewed end-on, the coil encircles an internal area approximately 1 mm in width

and just less than 3 mm in height. The prongs of the bulb base are 7 mm apart and 10 mm long, and fit

into a ceramic base. The 19 mm outer diameter ceramic socket is sunk into a 25 mm o.d. aluminum rod

and sealed with epoxy so that the coil aligns with an 8-32 tapped hole on the adjustable cap, centered with

the line of sight of the coil (Figure 1). The cap of the cell also has a 1/2 NPT thread fitted with a

connector that attaches to the central channel of the torch. The bottom of the 88 mm long rod is tapped

(1/4-20) for mounting. The electrical leads to the socket pass through an electrical feed-through on the

side of the rod. Purge and carrier gases enter through the gas ports on either side of the rod, pass through a

Page 45: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

34

mixing chamber below the tungsten filament, and then escape through a small orifice in the ceramic base.

This flow carries the analyte to the torch during vaporization.

Page 46: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

35

Table I. Vaporization Program for the ETV-FAES device. The sample has been dried prior to the first step. The detector is triggered 5 s prior to the beginning of Step 2. Step 3 allows sufficient cooling before the introduction of the next sample.

Current (Amps)

Time (seconds)

0 10 8 10 0 60

Page 47: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

36

Figure 1. Illustration of ETV-FAES Device a) ½ NPT b) 8-32 tapped hole centered and in the line of sight of coil c) opening in ceramic base for purge gas throughput d) two 1/16 NPT gas fittings [one front and one rear of the rod e) electrical line feed through f) ¼-20 tapped hole for mounting

25 mm

19mm

40 mm

c

e

f

d

19 mm

88 mm

b

a

Page 48: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

37

The optics are simple. The tip of the torch is positioned two focal lengths (f = 75 mm) from a fused silica

lens that is two focal lengths from the detector. This forms a one-to-one image of the flame on the

entrance slit of a 2048-element CCD array spectrometer (Ocean Optics, Dunedin, FL, USA).

Visual Basic Program

A Visual Basic (VB) program was designed to automatically dry and vaporize solutions so that

they move into the flame as an aerosol plug. Previous iterations of this program required a painstaking

manual optimization of the drying parameters. This was often a cumbersome process and had to be

repeated throughout the lifetime of a coil (~400 firings) due to changes in the resistance of the dry coil

(either from buildup of oxides along its surface or from a decrease in the filament wire diameter with age).

Variations in sample matrix also caused similar changes in resistance. When a drop of liquid was applied

to the W-coil, it spanned the loops of the filament, and acted as a short circuit. Thus, as the drop

evaporated the resistance across the coil changed. All of these effects caused changes in coil temperature

upon the application of a constant current, resulting in large variations in the drying time. The present

version of the Visual Basic program eliminates the need for this manually optimized drying program.

Each of the effects mentioned above result in a variation of the potential that must be applied across the

W-coil in order to maintain a constant current. By monitoring this potential, one may predict when the

W-coil is dry by measuring the change in potential per unit time (slope, m) for a given constant current.

The coil is determined to be dry once an empirically determined threshold slope is attained. The amount

of time necessary to reach dryness is therefore variable, and need not be previously determined. The

approach corrects for sample volume, filament age, and sample matrix by allowing the drying to continue

only up to the threshold value.

When the Visual Basic program is initiated, a series of warning messages appears to ensure that the user

is following all safety guidelines associated with the equipment. Then a digital 5V is applied through the

minilab 1008 to close a DC controlled, solid-state, normally open, optical relay switch (Crydom, Part

Page 49: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

38

Number D1D12, San Diego, CA, USA). Current passes through the W-coil only when the relay is closed

(Figure 2). The constant current power supply is programmed via analog control voltages (0-5 V) sent

from the D/A output of the minilab 1008 to the power supply’s trim pin. The desired current is generated

from the input control voltage (V) using the formula:

4 1 (1)

where I is the desired constant current, and Imax is the maximum current generated by the power supply,

13.3 Amps.

The automated dry step is applied first at 3.4 A. The VB program runs a variable time loop, and the

voltage across the coil is read every 0.5 seconds. The slope (m) is calculated using the most recent

voltage (V2) and the previous voltage (V1):

m V – V

. (2)

The two latest slope values are averaged and displayed on the program screen. Once the minimum slope

value of 0.38 (the threshold value seen in Figure 3) has been exceeded, the program terminates the drying

loop and proceeds to the vaporization stage (Table I). First the coil is cooled for 10 s at 0 A. Then the

dry sample is vaporized at 8 A, followed by a 60-second cooling period prior to the next sample injection.

The currents and times are user selectable during this stage, and up to 8 steps may be programmed (Figure

3).

Page 50: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

39

Figure 2. ETV-FAES control electronics.

Page 51: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

40

The spectrometer is triggered for automatic data collection by the VB control program (Figure 3). In the

Visual Basic code, a command reads to the “Trigger Step?” entry and the associated amount of time to

“measure background before trigger step” (step 2 and 5 s in the Fig 3). In this case, a 5 V signal is sent to

the spectrometer’s external trigger pin 5 s prior to the beginning of the vaporization step (Step 2). The

detector then collects a user-selected number of spectra each having a specified integration time. In this

case, upon receiving the trigger signal, 50 consecutive spectra are collected each having a 0.2 s

integration time. These 50 spectra span a 10 s period, beginning 5 s prior to the vaporization step. This

allows for observation of the background signal prior to the high temperature vaporization stage, to ensure

that no sample is lost prematurely. Figure 4 shows a typical ETV FAES spectrum. In Figure 5, the

signals at the peak wavelengths for Rb, Cs, and Ba for each of the 50 spectra are plotted versus time. The

analytical signal may be recorded as the maximum value from this plot (peak height) or the sum of

several successive spectra (from time 6-9 s for example) in a peak area mode. In this instance, all

analytical signals were recorded as the maximum value from the plot.

Page 52: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

41

Figure 3. Screen Capture Image of the Visual Basic Control Program

Page 53: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

42

Figure 4. ETV-FAES spectrum for 5.0 mg L-1 cesium, 5.0 mg L-1 barium, and 2.0 mg L-1 rubidium with sodium present as a contaminant.

0

10

20

30

40

50

60

70

80

90

350 450 550 650 750 850

Rel

ativ

e E

mis

sion

In

ten

sity

Wavelength (nm)

Na

Cs Rb

Ba

Page 54: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

43

Figure 5. ETV-FAES simultaneous emission profile for 5.0 mg L-1 cesium, 5.0 mg L-1 barium, and 2.0 mg L-1 rubidium. Monitored at 852.1 nm, 553.6 nm, 794.8 nm, respectively. Time 0 (not shown) occurs 5 s prior to the vaporization stage.

-5

0

5

10

15

20

25

30

5 6 7 8 9 10

Rel

ativ

e E

mis

sion

In

ten

sity

Time (seconds)

Rb

Cs

Ba

Page 55: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

44

Reference Solutions and Standards

All standard solutions were prepared from their metal salts using Trace Metal Grade Nitric Acid (Fisher

Scientific, Fair Lawn, NJ, USA) digestion or distilled deionized water (Milli-Q Systems, Millipore,

Billerica, MA, USA). Standard solutions for calibration curves were prepared by serial dilution using

distilled deionized water.

Safety Considerations

Material safety data sheets were consulted prior to the use of all chemical reagents. All indicated safety

precautions were taken during each step of analysis, and all waste was stored in separate containers prior

to disposal. Necessary precautions were also taken with regards to working with a high voltage power

supply.

Page 56: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

45

RESULTS AND DISCUSSION

System Optimization

For all system optimization studies, 20.0 µL samples of 20.00 µg/L aqueous lithium standard solution

were distributed onto the tungsten coil using a micropipette (Eppendorf, Hauppauge, NY, USA), dried

using the Visual Basic program, and vaporized using previously optimized parameters (Table I). Viewing

of the flame image was initially done in a radial, side-on fashion, but after a comparison of lithium

emission intensities, axial, end-on viewing was chosen for all analyses.99 The peak height values were

recorded, and all optimization studies were repeated in triplicate. The parameter operating value that

gave rise to the highest signal was determined to be the optimal setting (Table II).

The cutting tip designation 6-MTH-N (Victor Equipment, Part Number VIC0333-0370, St. Louis, MO,

USA) is designed to cut through metal at a maximum depth of 252 mm using natural gas or propane as

the fuel source. The tip is 8.0 cm in length with a diameter of 1.5 cm, and is composed of two brass pieces:

a hollow cylindrical barrel and a solid insert with a 3 mm diameter bore through the length of the insert as

seen in Fig. 6 a and b. The end of the insert has six, 1.0 mm openings in a circular pattern that allow the

fuel and oxidant to pass through into the cutting tip body. These gases are then directed into ten 1.0 mm

openings that are arranged circularly around the front end of the insert illustrated in Fig. 6 part d. While

the fuel and oxidant are mixed through the length of the tip as they exit the front end, a flame occurs in a

“flower-like” pattern. In the center of this flower, the sample enters the flame.

Limits of Detection

Calibration curves for five group I and II elements of interest were constructed using the previously

prepared standard solutions: Ba, Cs, K, Li, and Rb. These elements were chosen because of their ease of

excitation, which was well suited for the lower temperatures of the air-acetylene flame. A similar device

using an oxygen-acetylene flame has been validated in a laboratory environment and was found to be

capable of the analysis of less volatile species such as copper and lead with great success. In Table III, all

Page 57: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

46

LODs were calculated by multiplying the standard deviation in the blank signal by three and dividing by

the slope of the element’s calibration curve. To produce a reasonable guide for the FAES LOD, the LODs

from five sources were averaged and the results are used as a basis of comparison for ETV-FAES85, 100-103.

The ETV sample introduction provided a 25-fold enhancement of the LOD of lithium compared to a

standard flame atomic emission device, while Ba saw a 250-fold improvement. On average, the

improvement in LOD was a factor of 100.

Page 58: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

47

Table II. Optimized Operating Parameters for ETV-FAES Device

Parameter Optimized

Value Integration Time (msec) 180

Cutting Tip 6-MTH-N Acetylene Gas Flow

(LPM) 1.4

Air Flow (LPM) 10.6 H2/Ar Flow (LPM) 1.0

Page 59: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

48

Figure 6. Illustrations of 6-MTH-N Cutting Tip. a) solid insert including 10, 1.0 mm gas flow channels around the tip end b) outer copper sheath of cutting tip c) inner and outer pieces joined together for use d) end-on view of cutting tip with flow openings shaded.

a)

b)

c)

d)

Page 60: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

49

Table III. ETV-FAES Limits of Detection

Element Wavelength

(nm)

ETV-FAES LOD

(µg/L)

FAES LOD

(µg/L)

Improvement Factor

Li 670.8 0.04 1 25 K 766.5 0.8 3 4 Rb 794.8 0.3 60 200 Cs 852.1 4 200 50 Ba 553.6 2 500 250

Page 61: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

50

CONCLUSION

The ETV-FAES device described here can be easily constructed using commercially available materials.

It requires little user training, is rugged, portable, and capable of achieving LODs comparable to modern

instrumentation such as the inductively coupled plasma. The device is ideal for mobile labs, teaching,

industrial applications, and laboratories that cannot make significant monetary investments in analytical

instrumentation.

Page 62: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

51

REFERENCES

1. Klaic, P. M. A. N., A.M.; Moreira, A.S.; Vendruscolo, C.T.; Ribeiro, A.S., Determination of Na,

K, Ca and Mg in xanthan gum: Sample treatment by acid digestion. Carbohydrate Polymers 2011, 83,

1895-1900.

2. Sokolnikova, J. V. V., I.E.; Menshikov, V.I., Determination of Trace Alkaline Metals in Quartz

by Flame Atomic Emission and Atomic Absorption Spectrometry. Spectrochim. Acta, Part B 2003, 58,

387-391.

3. Miller-Ihli, N. J., Atomic Absorption and Atomic Emission Spectrometry for the Determination

of the Trace Element Content of Selected Fruits Consumed in the United States. J. Food Comp. Anal.

1996, 9, 301-311.

4. Ingle, J. D. C., S.R., Spectrochemical Analysis. Prentice Hall: Upper Saddle River, NJ, USA,

1998.

5. Pinto, F. G. L., F.G.; Saint'Pierre, T.D.; da Silva, J.B.B.; Costa, L.M.; Curtius, A.J., Direct

Determination of Dy, Sm, Eu, Tm, and Yb in Geological Samples by Slurry Electrothermal Vaporization

Inductively Coupled Plasma Mass Spectrometry. Anal. Lett. 2010, 43, (6), 949-959.

6. Lee, Y. I. K., J.K.; Kim, K.H.; Yoo, Y.J.; Back, G.H.; Lee, S.C., Design and Critical Evaluation

of Improved Electrothermal Vaporization Flame Atomic Absorption/Emission Spectrometry for Direct

Determination of Trace Metals in Microliter Samples. Microchem. J 1998, 60, 231-241.

7. Okamoto, Y. K., H.; Kataoka, H.; Tsukahara, S.; Fujiwara, T., Direct determination of bromine in

plastics by electrothermal vaporization /inductively coupled plasma mass spectrometry using a tungsten

boat furnace vaporizer and an exchangeable sample cuvette system. Rapid Commun. Mass Spectrom.

2010, 24, (9), 1265-1270.

8. Wu, P., Wen, X., He, L., He, Y., Chen, M., Hou, X., Evaluation of Tungsten Coil electrothermal

vaporization-Ar/H2 flame atomic fluorescence spectrometry for determination of eight traditional hydride-

forming elements and cadmium without chemical vaporization generation. Talanta 2008, 74, 505-511.

Page 63: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

52

9. Williams, M. P., E.H., Commercial Tungsten Filament Atomizer for Analytical Atomic

Spectrometry. Anal Chem 1972, 44, (7), 1342-1344.

10. Reid, R. D. P., E.H., Horizontal Tungsten Coil Atomizer and Integrated Absorbance Readout for

Atomic Spectrometry. Anal. Chem. 1976, 48, (2), 338-341.

11. Hou, X.; Levine, K. E.; Salido, A.; Jones, B. T.; Ezer, M.; Elwood, S.; Simeonsson, J. B.,

Tungsten coil devices in atomic spectrometry: absorption, fluorescence, and emission. Analytical Sciences

2001, 17, 175-180.

12. Barth, P. H., S.; Krivan, V., Analysis of Silicon Dioxide and Silicon Nitride Powders by

Electrothermal Vaporization Inductively Coupled Plasma Atomic Emission Spectrometry Using a

Tungsten Coil and Slurry Sampling. J. Anal. At. Spectrom. 1997, 12, 1359-1365.

13. Bings, N. H. S., Z., Development of a tungsten filament electrothermal vaporizer for inductively

coupled plasma time-of-flight mass spectrometry and its possibilities for the analysis of human whole

blood and serum. J. Anal. At. Spectrom. 2003, 18, 1088-1096.

14. Spaargaren, D. H., Presence and Significance of Lithium in the Brown Shrimp, Crangon crangon.

Comparative Biochemistry and Physiology, Part A: Molecular and Integrative Physiology 1998, 91, 457-

461.

15. Koji, M. K., T.; Yamaya, H.; Kobayashi, E.; Nogawa, K., Urinary Excretion Levels of Sodium

and Potassium in Envrionmental Cadmium-Exposed Subjects. Toxicology 1998, 127, 187-193.

16. Shiraishi, K. K., S.; Zamostyan, P.V.; Tsigankov, N.Y.; Los, I.P.; Koruzun, V.N., Dietary intakes

of alkaline metals, alkaline earths, and phosphorous in Ukrainians. Biomedical Research on Trace

Elements 2009, 20, 62-68.

17. Donati, G., Gu, J., Nobrega, J.A., Calloway, C.P. Jr., Jones, B.T., J. Anal. At. Spectrom. 2008, 23,

361-366.

18. Nakamura, S.; Kobayashi, Y.; Kubota, M., Temperature of a W Ribbon Furnace in

Electrothermal Atomic Absorption Spectrometry. Spectrochimica Acta Part B 1986, 41, 817-823.

19. Alavosus, T. J.; Murphy, R.; Schatzlein, D., Am. Lab. 1995, 23, 31-34.

Page 64: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

53

20. Christian, G. D., Feldman, F.J., A comparison study of detection limits using flame-emission

spectroscopy with the nitrous oxide-acetylene flame and atomic-absoroption spectroscopy. Applied

Spectroscopy 1971, 25, 660-663.

21. Ramirez-Munoz, J., Spectrochemical Methods of Analysis. Wiley-Interscience: New York, NY,

1971.

22. Bruno, T. J., Svoronos, P.D.N., Handbook of Basic Table for Chemical Analysis. CRC Press:

Boca Raton, FL, 1989.

23. Robinson, J. W., Undergraduate Instrumental Analysis. 5th ed.; Marcel Dekker: New York, NY,

1995.

Page 65: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

54

CHAPTER 3

An Electrothermal Vaporization Flame Atomic Emission Spectrometer

Summer N. Hanna and Bradley T. Jones

Department of Chemistry, Wake Forest University,

Winston-Salem, NC, 27109, USA

Note: The following manuscript was published in the Journal of Analytical Atomic Spectroscopy (DOI:

10.1039/c1ja10060b). Stylistic variations are due to the requirements of the journal. All of the presented

research was conducted by Summer N. Hanna. The manuscript was prepared by Summer N. Hanna and

edited by Bradley T. Jones.

Page 66: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

55

ABSTRACT

An inexpensively built, light-weight, potentially transportable flame atomic emission spectrometry device

has been designed and developed. A tungsten coil is employed as an electrothermal vaporization source

coupled to an oxygen-acetylene flame. The tungsten coil is simple in design and commercially available

as a 150 W, 15 V light bulb, while the flame source is generated by a welder’s metal-cutting torch.

Nineteen elements have been determined with limits of detection as low as 0.9 ng L-1 for Ca and 8.0 ng L-

1 for In. The accuracy of the method is verified by determining the concentration of thirteen elements in

two reference materials: Montana soil and a water pollution standard solution. No statistically significant

difference is observed between the certified and determined values at a 95 % level of confidence.

Sample volumes as low as 10 µL may be analyzed. The best torch tip design is identified, and gas flow

rates are optimized. The temperature profile of the flame is also discussed. The electrothermal vaporizer

and cutting torch are designed to be portable. Incorporation of a simple hand-held CCD spectrometer

would render the entire system usable in the field.

KEYWORDS

Tungsten coil, electrothermal vaporization, atomic emission spectrometry, portable instrumentation

Page 67: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

56

INTRODUCTION

The state-of-the-art for analyzing complicated samples by atomic spectrometry includes graphite furnace

atomic absorption spectrometry (GFAAS),1 inductively coupled plasma optical emission spectrometry

(ICP-OES),2 and inductively coupled plasma mass spectrometry (ICP-MS).3 These techniques are quite

powerful, providing detection limits at the part per billion (ppb) level and below. For the most part,

however, they involve instrumentation that is large, fragile, and expensive. A single element GFAAS

device may cost $50,000 at the lower end, while multi-element ICP-MS systems may be priced at

$200,000 or more. Even the least expensive of these devices is out of reach for small businesses,

undergraduate teaching institutions, and laboratories in third world countries. Another shortcoming of

state-of-the-art equipment is its limited utility in the field. Portability and analytical performance are often

diametrically opposed in environmental field applications.4 An Electrothermal Vaporization Flame

Atomic Emission Spectrometry (ETV-FAES) device will address these concerns. The ETV-FAES device

is a portable flame atomic emission detector. The novelty of the device is in the design of the excitation

source and the sample introduction apparatus. The emission source will be the oxygen-acetylene flame

generated by a typical sheet metal cutting torch, mass produced at low cost for the welding industry, and

designed to be portable for use in scrap metal yards. The cutting flame will burn at a temperature in

excess of 3000 °C, hot enough to excite the emission of almost every metal and semimetal on the periodic

table.5 Since the torch is not produced specifically for analytical instrumentation, ETV-FAES will benefit

from the lower cost associated with the demand for the product in the welding industry. Almost every

small town will have a welding supplier who carries such torches.

The cutting torch, on the other hand, is not designed for liquid sample introduction. The ETV-FAES

device, therefore, will use an electrothermal vaporizer (ETV) to introduce dry analyte-containing aerosol

into the oxygen “cutting” central channel of the torch. Again, the ETV will employ a heating element

designed and produced for a different purpose: the tungsten coil (W-coil) or filament contained in a

Page 68: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

57

common bulb, such as an automobile headlamp.6 The bulb is portable, mass produced to strict

specifications, costs less than $5, and may be powered by a 12 V battery, such as a car battery. Liquid

sample aliquots of 10 μL or more will be placed on the coil, dried at low current, and subsequently

vaporized at higher current. The low mass of the filament will eliminate the need for water cooling. A

simple flow of Ar with 10% hydrogen, a common welding gas, will protect the coil from oxidation, and

rapidly cool the system once the current is removed. During the vaporization step, a dry aerosol

containing the analyte atoms will be carried to the flame, where excitation resulting in atomic emission

will occur. For true portability, the emission signal could easily be collected with a hand-held charge

coupled device spectrometer. This will allow for trace level, simultaneous, multi-element analyses with

very small sample volumes.

Electrothermal vaporizers have been used only sparingly with flame systems. In 1993, a graphite furnace

ETV was employed with an air-acetylene flame in a laser enhanced ionization experiment.7 Ar was the

carrier gas between the two devices. Later, the same system was employed for the determination of Pb in

blood.8 In each case, the ETV provided sensitivities approaching the theoretical maximum. In 1995 a

tantalum filament was employed as an ETV for flame atomic absorption spectrometry.9 Again, Ar gas

transported the sample vapor to the air-acetylene flame. More recently, a W-coil ETV device was used to

provide sample vapor for a hydrogen-air flame in an atomic fluorescence experiment.10 Detection limits

were in the low ppb range. Each of these experiments probed analyte atoms in the ground state, so high

temperature emission sources were not necessary.

Most ETV applications for atomic emission spectrometry involve inductively coupled or microwave

plasmas. In fact, a very recent review with more than 300 references focuses largely on these

applications.11 Most ETV devices were composed of graphite, but some employed metal tubes, cups,

loops, and filaments. In 2002, a review article was published with 179 references to the use of W devices

in AAS and ETV systems.12

Page 69: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

58

The specific W-coil proposed for ETV-FAES has also been employed as an ETV sample introduction

system for ICP-OES. 13 14 15 16 On average, replacement of the traditional nebulizer with the W-coil ETV

device resulted in an LOD reduction factor near 300. The conventional reasoning for this improvement

involves three factors: introduction of the sample into the ICP as a dry aerosol (thus requiring no energy

for desolvation in the plasma), production of very fine particulate matter (or even atomic vapor) from the

ETV, and introduction of the analyte as a concentrated plug in a brief time period. These processes should

translate equally to the flame, so similar improvement factors are anticipated.

In the work described below, sample aliquots as low as ten microliters will be deposited onto a tungsten

filament commonly used in a 150 watt projector bulb. The filament will then be dried at low currents

supplied by a constant current power supply. The dried sample will be vaporized at a slightly higher

current and the resulting aerosol carried through the central channel of a standard metal-cutting torch.

This will guide the vapor into the heart of a flame, where the metal atoms will be excited at high

temperatures and emission will occur.17 This atomic emission will be collected using a spectrometer

connected to a personal computer. Multiple elements will be detected simultaneously at the parts per

billion level. The sensitivity and analysis time will rival modern commercial instrumentation but at a

fraction of the cost. The device will be small and robust enough to transport for field applications.

Page 70: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

59

EXPERIMENTAL

Instrumental

A Victor® oxygen-acetylene machine cutting torch is used as the flame source. The torch is designed to

create a heating flame to melt a metal surface. It is composed of a brass body with three gas inlet lines.

The heating flame is supplied via a flashback-protected fuel port and a flashback-protected oxidant port.

These two gases are mixed within the body of the torch. The third gas port is for “cutting” oxygen. This

high velocity stream of oxygen is not mixed with the flame gases in the torch. Instead, it passes straight

through the torch, pierces through the flame where it is heated and used to oxidize the metal in the plate

forming a slag jet. The ETV-FAES device will not produce the slag jet, rather the ETV gas will be excited

by the heating flame formed at the end of the cutting tip. The tip divides the pre-mixed flame gases into

six equal streams arranged in a circle approximately 1 cm in diameter. The six streams produce six high

temperature flame jets that merge together about 1 cm away from the tip surface. The ETV stream is

injected into these merging flame jets, through a seventh central channel. Cutting tips vary by the size of

their inner channel orifice diameter and are industry standardized. Typically, the smaller the inner channel

diameter, the smaller the thickness of the metal that is to be cut and likewise larger diameters are used to

cut thicker pieces of metal. This difference in size also requires a change in gas flow rates, with larger tips

requiring higher flow rates.

In the case of ETV-FAES, the design has been changed so that acetylene and oxygen continue to form a

flame in the six concentric circles of the cutting tip, but the center channel has undergone much change.

Connected to this outlet is a laboratory-designed electrothermal vaporization cell. Inside the vaporization

cell is a tungsten filament bathed by a stream of 10% hydrogen 90% argon gas, which prevents oxidation

of the filament and simultaneously acts as a carrier gas. The carrier gas may be mixed with auxiliary

oxidant (oxygen or nitrous oxide) outside the ETV cell, but prior to introduction to the torch. This allows

for higher flow rates through the central channel. Samples are pipetted onto the filament and then dried

Page 71: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

60

using a low current supplied by constant current power supply. The optimization method for the heating

program was developed elsewhere and used here to simultaneously vaporize the sample and trigger the

detector at the beginning of the vaporization cycle. The progressively lower drying currents prevented the

coil from heating to a red glow, mitigating the chance for a loss of analyte prior to vaporization.18 When

the sample is vaporized, the carrier gas introduces the sample into the center channel of the torch and

directly into the flame as a concentrated aerosol plug. In Figure 2 this sample plug can easily be seen

flowing through the center of the flame. Here the atoms in the sample are thermally excited and upon

their relaxation emission occurs through the release of a photon at a characteristic wavelength.

Page 72: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

61

Table I. Heating Program for the Tungsten-Coil Vaporizer.

Applied Current (Amps) Process Time (Seconds) Signal Read 4.1 Dry 30 No 3.6 Dry 25 No 3.3 Dry 20 No 3.1 Dry 20 No 0 Cool 10 No 8 Vaporize 10 Yes 0 Cool 20 No

Page 73: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

62

The Victor® MT 1065A three hose machine cutting torch (Victor Professional, Part Number 0399-1121)

is secured using an aluminum base onto an adjustable rail system. The torch is positioned axially and the

oxygen, acetylene, auxiliary oxidant, and sample gas inlet lines are securely connected as seen in Figure 1.

Axial viewing increases the sensitivity for inductively coupled plasma emission, so this viewing method

was chosen for ETV-FAES as well. The end-on viewing method reduces the background emission of the

oxygen-acetylene flame by focusing on the center channel where analyte emission occurs, rather than on

the bright blue peripheral flame jets (Figure 2). 19 The aluminum electrothermal vaporization cell is

connected to a solid state constant current power supply (BatMod, Victor, Andover, MA) and the power

supply is controlled with a Visual Basic program and a digital to analog converter (Measurement

Advantage MiniLab 1008, Measurement Computing, Norton, MA). The tungsten filament is carefully

extracted from a commercially available 150 W, 15 V projector light bulb (Osram HLX 64633 Xenophot,

Augsburg, Germany). The bulbs are mass produced to strict specifications. The W-coil filament is

exposed by removing the fused silica envelope while leaving the base intact. The filament is mounted in

the ETV housing using the two-pronged base and placed in the vaporization cell’s ceramic base. The 19

mm o.d. ceramic socket is countersunk into a 25 mm o.d. aluminum rod so that the upper surface is flush.

The bottom of the 7 cm long rod is tapped for mounting. The electrical leads to the socket pass through

feedthrough on the side of the rod. Purge/carrier gases enter through the other side of the rod, pass

through a mixing chamber, and then escape through a small orifice in the ceramic base. The torch is

positioned two focal lengths (f=75 mm) from a fused silica lens that is two focal lengths from the detector.

This forms a one-to-one image of the flame onto the 25 µm entrance slit of a small Czerny-Turner

monochromator (MonoSpec 18, Scientific Measurement Systems, Inc. Grand Junction, CO). The

monochromator has a 2400 gr mm-1 grating (110 x 100 mm) giving a reciprocal linear dispersion of 2 nm

mm-1 at 400 nm with a 156 mm focal length and optics of f/3.8. The analytical signal is detected by a

thermoelectrically cooled CCD spectrometer (Spec-10, Princeton Instruments, Roper Scientific, Trenton,

NJ) with a linear 2-dimensional array of 1340 x 100 pixels. Each pixel has a dimension of 20 x 20µm

which translates to a image area of 26.8 x 2 mm. This provides an approximately 55 nm spectral window

Page 74: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

63

depending on the central wavelength. Elements are observed using an integration time of 0.025 seconds

while collecting 20 spectra during a ten-second exposure period as triggered by the Visual-Basic coil

control program.

Page 75: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

64

Figure 1. Photograph of the ETV-FAES Instrument.

Flame Cutting Tip

Torch Flashback

Arrestor

Acetylene

Oxygen

Nitrous

Oxide

ETV

16.50 cm

Page 76: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

65

Figure 2. End on view of the cutting tip during the vaporization cycle of a) distilled deionized water b) tap water.

a b

Page 77: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

66

Reference Solutions and Standards

All standard solutions were prepared from their metal salts using the appropriate digestion method. 20

Standard solutions for calibration curves were prepared by serial dilution using distilled deionized water

(Milli-Q Systems, Millipore). Water Pollution Standard 1 (WPS-1) (VHG Labs, Manchester, NH) was

analyzed using the standard addition method and was also diluted before being spiked with the element of

interest. NIST standard reference material (SRM) 2711—Montana Soil B (NIST, Washington, D.C) was

prepared by weighing out 1.9935 g and digesting in 4.00 mL of 14 mol L-1 trace metal grade nitric acid

(Fisher Scientific, Pittsburg, PA) and 4.00 mL distilled deionized water (to prevent evaporation) on a

heating block at 100 º C for two hours. This solution was allowed to cool, filtered using a Büchner Funnel,

and diluted to 25.00 mL with distilled deionized water.

Safety considerations

Material safety data sheets were consulted before using each chemical reagent. Necessary safety

precautions were taken in each step of the analysis, and all aqueous waste was stored separately prior to

disposal. All personnel working with the flame source were educated in the potential hazards associated

with flammable gases, and were also trained in the safe handling and transportation of compressed gases

by National Welders Supply Company.

Page 78: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

67

RESULTS AND DISCUSSION

System Optimization

For optimization studies, 40.0 µL samples of an aqueous manganese standard solution were deposited

onto the tungsten filament using a micropipette (Eppendorf, Germany) and dried using a previously

determined heating program (Table I). The samples were then vaporized and passed through the torch into

the flame. The optimization parameters were then confirmed by repeating each experiment with a

different element, copper. Gas flow rates were optimized by altering the gas flow rates for each inner

channel orifice diameter and measuring the resulting emission intensities. Each gas flow was increased in

0.5 L/min increments from 0-2 L/min and then in 1.0 L/min increments until a flow rate of 10.0 L/min

was reached.

When gas flow rates were optimized, an inner channel orifice diameter of 3.58 mm (cutting tip size 7)

was the most suitable for the largest variety of elements. The optimal flow rates for acetylene, oxygen,

and auxiliary gas were 2.5, 4.5, and 3.0 L/min, respectively. 21The optimal H2/Ar ETV flow rate was

determined once the flame gases were set. The resulting emission intensities were higher when a flow

rate of 1.5-3 L/min was used for the 10 % H2, 90 % Ar carrier gas. For all reported work, an ETV carrier

gas flow of 2.0 L/min was used.

Page 79: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

68

Figure 3. Variation in emission signal level with orifice diameter for a 1 µg/mL single-element Mn standard solution.

0

10

20

30

40

50

60

70

80

90

100

0 1 2 3 4 5

Rel

ativ

e E

mis

sion

Sig

nal

Inner Orifice Diameter (mm)

Page 80: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

69

Figure 4. Variation in emission signal level with ETV gas (10% H2/Ar) flow rate for a 1 µg/mL Mn solution.

0

10

20

30

40

50

60

70

80

90

0 2 4 6 8 10 12

Rel

ativ

e E

mis

sion

Sig

nal

Carrier Flas Flow Rate (LPM)

Page 81: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

70

Once a cutting tip size was determined and operational variables were optimized, viewing parameters

were identified by adjusting the height and width of the viewing area in a 1x1 mm grid pattern. The flame

image was measured and found to have a diameter of 5 mm with the center channel at a diameter of 1mm.

The ideal viewing position occurred when the 1.0 mm inner center channel diameter image was focused

on the pinhole monochromator entrance. The torch was positioned at a distance of 2f from the end of the

tip of the inner cones of the flame to the focusing lens, which was positioned 2f from the entrance of the

monochromator. 22

Gas Phase Excitation Temperature Mapping of ETV-FAES Flame

In order to determine the temperature gradient across the excitation flame, an alternative method of gas

phase temperature analysis was explored. A derivation of the Boltzman Distribution developed by Donati

23 was utilized to generate gas phase temperature “maps” where the emission intensities of rubidium and

lithium are simultaneously monitored and using the mathematical relationship:

(1)

a gas phase temperature can be generated. 24 The emission intensities of the 780.0 and 670.8 nm lines of

rubidium and lithium, respectively, were measured using a fiber optic connected to the detector. The fiber

optic attachment was used so that a 5x5 mm grid pattern could easily be generated without moving the

apparatus. The results were then used to create a visual “map” of resulting flame temperatures. Using

these “maps” the highest temperature zones of the flame could easily be identified and calculated to be

about 3000ºC at the optimal viewing area. In Figures 5 and 6 there is a lack of symmetry in the

temperature gradient across the flame image that is thought to be due to the pull generated from the

exhaust hood located above the flame. In Figure 5, this drag and resulting lower temperatures were

located from 0-4 mm across and 8-10 mm in height of the flame image, while in Figure 6, the length of

the flame is shown. The area of the flame adjacent to the intake of the exhaust hood is located along the

Page 82: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

71

area located along the top of the image at a width of 10 mm. An effect identical to that in Figure 5 can be

seen spanning a width of 6-10 mm and along the length of the flame from the torch tip at 0 mm to about

25 mm down the flame.

Analytical Figures of Merit

Calibration curves were obtained for nineteen elements of interest using previously prepared standard

solutions: Ag, Al, Ba, Ca, Co, Cr, Cs, Cu, Eu, Fe, In, La, Mn, Ni, Pb, Pd, Sr, Tl, and Yb. These elements

were chosen because of either their ubiquity in natural samples and their range of specific properties.

Each was observed at the primary emission wavelength.25 Limits of detection (LODs) are listed in Table

II and were calculated by multiplying the standard deviation in the blank signal by three and dividing the

product by the slope of the element’s calibration curve. A typical emission spectrum is given in Figure 7.

The spectrum demonstrates the relatively low background emission encountered with the ETV-FAES

device. The ETV-FAES figures of merit were compared with those of traditional flame atomic emission

spectrometry (FAES) by connecting a nebulizer and spray chamber to the welding torch to collect

emission signals using continuous sample introduction. A significant enhancement in sensitivity for many

elements was observed. Detection limits for traditional FAES methods are well known, and usually vary

by no more than a 10.5,26,27,28,29The average improvement using the ETV-FAES was 43 and six of the

nineteen elements had improvement greater than fifty fold. 30

Element recovery for WPS-1, Water Pollution Standard, and NIST SRM 2711, Montana Soil, were also

completed using the method of standard additions and all data was reported at a 95 % significance level

(Table III).

Page 83: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

72

CEMD Temperature Profile-- Axial View

Flame Image Width (mm)

0 2 4 6 8 10

Fla

me

Imag

e H

eigh

t (m

m)

0

2

4

6

8

10

2200 2300 2400 2500 2600 2700

Figure 5. Axial Temperature Map of the ETV-FAES Flame with torch center located at 5 mm x 5 mm, indicated by the cross-hatched center circle. The entire cutting tip is indicated by the white circular outline as though it was being viewed end-on

Page 84: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

73

CEMD Flame Temperature Profile-- Radial Side View

Flame Length (mm)

0 10 20 30 40

Fla

me

Wid

th (

mm

)

0

2

4

6

8

10

0 500 1000 1500 2000 2500 3000 3500

Figure 6. Radial Temperature Map of the ETV-FAES Flame. In this arrangement, the edge of the torch tip spans the width of the flame and is located at the 0 mm length position, indicated by the white line with the center of the torch located at 0 mm x 5 mm as indicated by the white void.

Page 85: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

74

Figure 7. ETV-FAES Emission Spectrum for NIST SRM 2711 (Montana Soil).

0

20

40

60

80

100

120

140

160

180

440 445 450 455 460 465

Rel

ativ

e E

mis

sion

Int

ensi

ty

Wavelength (nm)

Cs

Cr Eu

Sr

Cs

Page 86: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

75

Table II. Figures of Merit for Nineteen Elements of Interest

Element Wavelength

(nm)

FAES LOD

(ng/mL)

ETV-FAES LOD

(ng/mL)

LDR%

RSD

Improvement

Factor

Cu 324.7 0.1 0.2 1.8 2.7 0.5 Ag 328.1 0.08 0.01 2.1 2.9 8 Co 345.4 0.9 0.009 2.2 3.6 100 Ni 352.4 0.2 0.04 1.9 3.1 5 Pd 363.5 0.3 0.02 2.2 2.1 15 Fe 371.0 0.6 0.007 2.4 1.8 86 Tl 377.6 0.1 0.006 3 5.2 17 Al 396.2 0.06 0.003 1.9 6.9 20 Yb 398.8 0.2 0.003 2.5 2.6 67 Mn 403.3 0.03 0.007 2.8 4.3 4 Pb 405.8 0.8 0.003 1.7 2.3 267 Ca 422.7 0.01 0.0009 3 4.5 11 Cr 425.4 0.9 0.009 2.8 3.0 100 La 441.8 1 0.01 2.4 2.3 100 In 451.1 0.04 0.008 2.1 0.9 5 Cs 455.5 0.2 0.2 2.5 4.4 1 Eu 459.4 0.0005 0.0006 2.4 4.1 0.8 Sr 460.7 0.003 0.001 3 8.4 3 Ba 553.6 0.5 0.02 2.4 3.7 25

Page 87: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

76

Table III. Elemental Recoveries for WPS-1 and SRM 2711 VHG WPS-100 SRM 2711

Concentration

Element Wavelength

(nm) Reported (µg/mL)

Found (µg/mL)

% Recovery

Reported (µg/mL)

Found (µg/mL)

% Recovery

Cu 324.7 0.2 0.2 ± 0.2 100 0.8 ± .02 0.9 ± 0.2 112

Ag 328.1 - - - 0.19

± .003 0.2 ± 1.1 95

Co 345.4 0.08 0.08 ± 0.1 100 - - - Ni 352.4 0.2 0.3 ± 0.3 150 - - - Fe 371 0.2 0.2 ± 0.3 100 - - -

Tl 377.6 - - - 0.2 ± .001 0.2 ± 0.5 90

Al 396.2 0.2 0.4± 0.2 200 - - -

Yb 398.8 - - - 0.02* 0.03 ±

1.3 150

Mn 403.3 0.003 0.003 ±

0.3 100 1.27 ± 0.2 1.2 ±0.5 95

Pb 405.8 0.2 0.2 ± 0.6 100 0.94 ± 0.3 0.99 ±

0.2 105 Cr 425.4 0.2 0.1 ± 0.4 50 - - -

Cs 455.5 - - - 0.25* 0.22 ±

0.4 88

Eu 459.4 - - - 0.09* 0.12 ±

0.4 133

*Noncertified Values

Page 88: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

77

CONCLUSION

The ETV-FAES system is capable of achieving limits of detection comparable to traditional AES

instrumentation while using significantly smaller sample volumes. The instrument’s design allows for a

high degree of user accessibility and requires little training for use, which has been illustrated by a similar

ETV-FAES instrument used exclusively by undergraduate researchers. It has large selectivity coupled

with compact size, portability, and low cost. This makes ETV-FAES ideal for mobile labs, teaching

applications, and increases accessibility to instrumentation for those incapable of purchasing expensive,

complex equipment.

Page 89: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

78

REFERENCES

1 Sturgeon, R.E. Spectrochim. Acta Part B 1997, 52 1451. 2 Hou, X., Jones, B.T., “Inductively Coupled Plasma Optical Emission Spectroscopy,” in Encyclopedia of Analytical Chemistry, Robert A. Meyers, Ed., John Wiley & Sons Ltd., Chichester, 2008. 3 Becker, J.S., Can. J. Anal. Sci. and Spectrosc. 2002, 47, 98. 4 Hou, X. Jones, B.T, Microchem. J., 2000, 66, 115. 5 Ramirez-Munoz, J., “Flame Emission Spectrometry,” in Spectrochemical Methods of Analysis (J.D. Winefordner, Ed.), Wiley-Interscience, New York, NY, 1971. 6 Sanford, C.L.; Thomas, S.E.; Jones, B.T., Appl. Spectrosc. 1996, 50, 174. 7 Smith, B.W.; Petrucci, G.A.; Badini, R.G.; Winefordner, J.D., Anal. Chem. 1993, 65, 118. 8 Riter, K.L.; Matveev, O.I.; Smith, B.W.; Winefordner, J.D., Anal. Chim. Acta 1996, 333, 187. 9 Duxbury, M.; de Mora, S.J., Microchem. J. 1995, 51, 337. 10 Wu, P.; Wen,X.; He, L.; He, Y.; Chen, M.; Hou, X., Talanta 2008, 74, 505. 11 Resano, M.; Vanhaecke, F.; de Loos-Vollebregt, M. T. C., J. Anal. At. Spectrom. 2008, 23,1450. 12 Hou, X. Jones, B.T., Spectrochim. Acta Part B 2002, 57, 659. 13 Davis, A.C.; Alligood, B.W.; Calloway, C.P.; Jones, B.T., Appl. Spectrosc. 2005, 59, 1300. 14 Davis, A.C.; Calloway, C.P.; Jones, B.T., Micorchem. J. 2006, 84, 31. 15 Davis, A.C.; Calloway, C.P.; Jones, B.T., Talanta 2007, 71, 1144. 16 Skoog, D; Holler, F.; Nieman, T; Principles of Instrumental Analysis, Thomson Learning: USA, 1998; Vol 5. 17 Donati, G.L., Gu, J., Nobrega, J.A., Calloway, C.P. Jr., Jones, B.T. J. Anal. At. Spec., 2008, 23, 361. 18 Alavosus, T.J., Murphy, R.; Schatzlein, D., Am. Lab., 1995, 27, 31. 19 Winefordner, J.D., Masfield, C.T., Vickers, T.J. Anal. Chem., 1963, 35, 1611. 20 Victor Professional Industrial Gas Apparatus Catalog, 2008 21 Christian, G.D., Feldman, F.J. Anal. Chem., 1971, 43, 611. 22 Donati, G.L., Jones, B.T. J. Anal. At. Spectrom., 2011, 26, 838-344.

23 Donati, G.L., Calloway, C.P. Jr., Jones, B.T. J. Anal. At. Spectrom., 2009, 24, 1105

Page 90: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

79

24 Harrison, G.; Massachusetts Institute of Technology Wavelength Tables, The MIT Press: Cambridge, MA, 1969 25 Ingle, J.D.C.; Spectrochemical Analysis. Prentice Hall: Upper Saddle River, NJ, USA, 1998.

26 Bruno, T.J., Svornos, P.D.N. Handbook of Basic Tables for Chemical Analysis. CRC Press: Boca Raton, FL, 1989.

27 Christian, G.D.; Feldman, F.J. Appl. Spectrosc. 1971, 25, 660.

28 Robinson, J.W. Undergraduate Instrumental Analysis. 5th Ed. New York, NY: Marcel Dekker, 1995.

29 Smith, B.W., Parsons, M.L. J. Chem. Ed., 1973, 50, 679.

Page 91: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

80

CHAPTER 4

Design of a Compact, Aluminum, Tungsten-Coil Electrothermal Vaporization Device for Inductively Coupled Plasma-Atomic Emission Spectrometry

Summer N. Hanna,a Clifton P. Calloway, Jr.,b Jason D. Sanders,b Ronald A. Neslon,b Jamaal Cox,b Bradley T. Jonesa

a Department of Chemistry, Wake Forest University, Winston-Salem, North Carolina, 27109 USA

b Department of Chemistry, Physics, and Geology, Winthrop University, Rock Hill, South Carolina, 29733 USA

Note: The following manuscript was published in Microchemical Journal (DOI:

10.1016/j.microc.2011.04.009). Stylistic variations are due to the requirements of the journal. All of the

presented research was conducted by Summer N. Hanna with the exception of the computer programming

completed by Clifton P. Calloway and the autodry program developed by Jason Sanders, Jamaal Cox, and

Ronald Nelson. The manuscript was prepared by Summer N. Hanna and edited by Bradley T. Jones.

Page 92: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

81

ABSTRACT

A new compact, aluminum electrothermal vaporization cell was constructed for inductively coupled

plasma atomic emission spectrometry analysis. This cell is compact enough to fit within the space

occupied by the spray chamber and fits directly to the quartz torch without extraneous tubing through the

use of a simple Compression fitting. Sample volumes as low as 10 µL were analyzed with an automated

control program for efficient vaporization. Twelve elements were analyzed utilizing a time resolved

acquisition method so that real-time data could be generated over a period of ten seconds with an average

improvement factor of 14 for elements over a wavelength range of 193-445 nm.

KEYWORDS

Inductively coupled plasma atomic emission spectrometry, tungsten coil, electrothermal vaporization

Page 93: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

82

INTRODUCTION

Sample introduction for inductively coupled plasma-atomic emission spectrometry (ICP-AES) typically

relies on large volumes of liquid analytes. Methods historically fall within two categories: pneumatic and

ultrasonic nebulization, both of which depend on the use of a spray chamber to prevent large liquid

droplets from entering the plasma. When using nebulization, approximately only 3-12% of the sample is

introduced into the plasma, with the remainder being lost to waste. 69 This method often encounters

problems due to the autosampler and copious amounts of tubing necessary for sample transport, which

can introduce air bubbles into samples. Poor connections between components can also create problems

in otherwise routine methods. 104

Because of these limitations, all sample introduction techniques will focus on electrothermal vaporization

(ETV). This method has been used extensively for the introduction of slurries, powders, solid materials,

and even complex biological matrices.55, 105-107 ETVs are often made from graphite, which can be

expensive because of the material’s required purity, inability to tolerate mass production, and resulting

low demand for production. 59 A large thermal gradient often occurs and interferences may be seen when

determining carbide-forming elements. Samples with a large concentration of surfactants may also

interfere with measurements by depositing carbon on the torch.40 Because of these disadvantages, metal

atomizers are becoming increasingly popular since they are nonporous, can provide quick and efficient

vaporization of a wide variety of materials, and do not require the elaborate power supplies that are

necessary when using graphite ETVs.

Tungsten is particularly well suited for these types of applications since it has the highest melting point of

any element and is found as a boat, ribbon, or coil. 76, 108, 109 Extensive work has been done with tungsten

ETVs which are regarded as a reliable vaporization source. Rare earth elements, which are notoriously

difficult to analyze because of their chemical similarities and tendency to form oxides, have been well

characterized using tungsten as an ETV in inductively coupled plasma mass spectrometry (ICP-MS)

Page 94: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

83

analysis. 110 Complex matrices, such as human urine, have also been successfully quantified using

tungsten coil ETVs when paired with permanent modifiers or extraction techniques. 76, 77

Tungsten coils such as the ones used in the determination of cadmium in urine samples are a great

alternative ETV source. These coils are made for use in projector light bulbs and can be purchased for as

little as two dollars each since they are commercially mass produced, which ensures adherence to strict

structural specifications. Their small size of less than 20 mm3 allows for the use of sample volumes on a

microliter scale. This reduction in volume provides users the opportunity to analyze extremely small-

volume or precious samples while utilizing the multielement capabilities of ICP-OES. Tungsten coil

ETVs in the past have used cells that did not connect directly to the torch of the ICP, but instead relied on

centimeters of tubing. Tubing makes analyte loss possible through condensation of the “hot” analyte

vapor on the walls of “cool” transport tubing.51 To overcome potential loss, a new ETV cell design that

connects directly to the quartz torch of the ICP through a simple Compression fitting has been used. The

reduction in cell size and elimination of tubing provides an ETV sample introduction source that can fit

within the space formerly occupied by a spray chamber and nebulizer. The ETV components can be

replaced with a traditional sample introduction system in less than a minute, which allows the user to

quickly alternate between sample introduction methods.

Page 95: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

84

MATERIALS AND METHODS

2.1 Reagents

All standard solutions were prepared from their metal salts using 15.8 N trace metal grade nitric acid

(Fischer Scientific, Pittsburg, PA, USA) and 18 MΩ distilled, deionized water (Milli-Q Systems,

Millipore). Working solutions were prepared from NIST Standard Reference Material (SRM) 1643e,

Trace Elements in Water, by simple dilution using distilled, deionized water and also utilized the standard

additions method to mitigate matrix effects typically encountered with ICP-AES.

2.2 Instrumental

W-coil ETV-ICP-AES is composed of an aluminum electrothermal vaporization cell and the ICP-AES

system. Figure 1 shows a schematic diagram of the ETV-ICP-AES cell. In this system, a tungsten coil is

extracted from a commercially available 150 W * 15 V projector light bulb (Osram HLX 64633 Xenophot,

Augsburg, Germany). The filament is exposed by removing the silica envelope while leaving the base

intact and is then mounted in the ETV cell’s ceramic base. The ceramic base is secured using two screws

to an aluminum mount. This mount has a 1” outer diameter and is 7/8” in length. The mount and socket

fit flushed into an aluminum rod and use a screw to create an airtight seal. This rod is 3” in length with an

11/16” inner diameter to accommodate the socket and mount. The arrangement of the socket on the

aluminum mount allows for the W-coil to align with a 1/8” hole which is used for sample introduction.

This opening is plugged with a screw and rubber o-ring to prevent oxidizing gases to enter the ETV cell.

A 1/16” NPT for gas fittings is located on the back side of the cell and immediately following is a ¼”

NPT at the base of the rod which allows for the connection of a glass outer Compression joint fitting

union (SS-4-UT-6-400, Swagelok, Solon, OH, USA). The compression fitting provides an immediate

glass-to-glass connection to the ICP-OES quartz torch without use of extraneous tubing normally

necessary in ETV-ICP-AES.

Page 96: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

85

Figure 1. Illustration of ETV cell. a) tapped hole for riser screw b) 2 screws to attach socket to mount c) 1” aluminum rod with 11/16” inner diameter to fit ceramic socket d) 1/8” hole centered and in line of sight of coil e) 1/16” NPT for gas fitting in back near bottom f) ¼” NPT for glass outer joint compression fitting

a

b

c

d

e

f

Page 97: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

86

Figure 2. Photograph of the ETV cell and ICP torch connection

Page 98: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

87

Figure 2 depicts the joining of these two parts in a manner identical to the glass to glass connection that

would be used for a spray chamber. This connection allows for the quick and easy replacement of a spray

chamber with the ETV cell in less than a minute.

A Visual Basic program controls the coil drying and vaporization cycle while simultaneously initiating

ICP spectra collection. The program automatically dries the coil by measuring the potential across the coil

at a constant current (3.4 Amps). As the liquid sample evaporates the potential across the coil increases.

The Visual Basic program runs a variable time loop, and every 0.5 seconds reads the voltage across the

coil. Using this measured voltage, a slope is calculated using the formula:

. (1)

The two most recent slope values are averaged and displayed on the screen. Once the voltage potential

slope reaches 0.38 (value determined empirically), the coil is dry. The program then continues to a set of

user defined values which vaporize the sample and trigger data collection. Step 2 in Table I at 8 amps is

vaporization and designated trigger step, which signals the ICP data software to begin collection of the

analytical signal through the time resolved analysis feature.111

The ETV connects directly to the ICP-AES torch. The axially arranged quartz torch, (ML155064,

Meinhard Glass, Golden, CO, USA) is surrounded by three water-cooled radio frequency (RF) coils that

use a frequency of 40.68 MHz to generate an argon plasma. The analytical zone of the plasma is aligned

with the 40 µm x 100 µm entrance slit of the argon-purged optical system. The Prodigy High Dispersion

ICP-OES (Teledyne Leeman Labs, Hudson, NH, USA) uses an 110 mm x 110 mm Echelle grating of

52.13 gr/mm to focus (focal length (f), f=800mm) the analytical signal on a solid-state, Large Format,

Programmable Array Detector (L-PAD). This enables continuous coverage over a wavelength range of

165-1100 nm. The analytical signal is collected using the Prodigy’s Time Resolved Analysis (TRA)

feature, which collects all analytical lines in a method simultaneously over a defined time period (Elapsed

Time) at a defined frequency (Time Slice). The real-time data is displayed graphically, and allows the

Page 99: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

88

user to immediately observe analyte emission in the plasma. This capability is advantageous to previously

mentioned ETV-ICP analyses where data acquisition was only capable in 1 second intervals, causing

approximately 90 % of the signal to be lost as dead time.112, 113 TRA reduces the dead time by collecting

100 millisecond time slices acquired over the ten second vaporization period. Each of the wavelengths is

simultaneously observed using the LPAD camera, and a desired wavelength(s) can be exported

individually or as a group of wavelengths for data manipulation (Fig. 4).

Page 100: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

89

Table I. Heating program for the tungsten-coil vaporizer.

Step Current (Amps)

Time (seconds)

Signal Read

Drying Autodry variable No Cooling 0 10 No

Vaporization 8 10 Yes Cooling 0 60 No

Page 101: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

90

Figure 3. Time resolved emission profile of barium, manganese, and strontium found in SRM 1643e

-5

0

5

10

15

20

25

30

35

0 1 2 3 4 5 6 7 8 9 10

Rel

ativ

e E

mis

sion

In

ten

sity

Time (seconds)

Sr

Ba

Mn

Page 102: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

91

RESULTS AND DISCUSSION

3.1 Optimization of Experimental Conditions

Traditional sample introduction was used to optimize the viewing position of the spectrometer by the

connection of a double-pass spray chamber and nebulizer to the torch. The generation of a constant signal,

as opposed to a discrete plug, allowed the viewing of the plasma to be optimized using the manufacturer’s

suggested method. Upon optimization, the spray chamber and nebulizer were removed and the ETV cell

was reattached with special care to avoid jarring the torch, ensuring optimal viewing conditions for the

analytical signal.

To identify the maximum sample introduction efficiency, the operating parameters of the ETV-ICP

system were assessed. 40.0 µL samples of a 0.50 mg/L solution of iron were pipetted onto the W-coil and

subjected to the previously mentioned vaporization cycle. Five emission measurements at each setting

were averaged. Each averaged value within a parameter was plotted to determine the maximum. The

setting which produced the maximum emission signal was determined to be the optimal setting within

that particular parameter. The optimization of the auxiliary flow is not discussed here, as it did not

contribute to signal or plasma stability.

3.1.1 Optimization of Torch Coolant Gas

The torch coolant gas provides a cooling mechanism to the quartz torch while also creating a more stable

plasma. Gas flows from 13.0-20.0 L/min were used to determine the effect on emission signal. 13.0 L/min

of coolant gas caused peak broadening, occasional splitting of emission peaks, and significant plasma

instability. Figure 4 shows that flows in excess of 18.0 L/min provided no observable enhancement in

emission. The highest emission intensities occurred when coolant flow rates of 17.0 L/min were used.

3.1.2 Optimization of Radio Frequency (RF) Power

Page 103: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

92

The wattage of the RF coils of the ICP-OES system creates the plasma by generation of a 40.68 MHz

radio frequency field. This field accelerates free electrons in the argon gas provided by the coolant flow.

When these free electrons collide, they generate extensive heating to produce the plasma used for

excitation of the vaporized sample. RF Powers of 0.9, 1.0, 1.1, 1.2, 1.3, and 1.4 kW were studied for their

ability to produce the highest emission signals while also creating a very stable plasma. The

manufacturer’s suggested setting is 1.1 kW, and optimization studies indicated that there was no

significant difference in the emission intensity with respect to the differing radio frequency powers. This

lack of difference allowed all studies to continue at the manufacturer’s suggested setting of 1.1 kW.

3.1.3 Optimization of Carrier Gas

0.5, 1.0, 1.5, and 2.0 L/min flows of the 10 % hydrogen/ 90 % argon carrier gas were studied for their

ability to transport the aerosol plug into the plasma. In addition to transport, this carrier gas creates a

reducing atmosphere around the tungsten coil while simultaneously providing quick cooling to the coil

after the application of high currents. If the flow of the carrier gas falls below 0.5 L/min it is unable to

prevent oxidation of the tungsten coil at high applied currents. This oxidation causes the coil to quickly

become brittle and introduces tungsten oxides into the plasma. The tungsten oxides create a large

emission band that severely interferes with signals of interest. Figure 5 shows virtually no emission signal

when this low carrier gas flow rate is used. Flow rates that are higher than 2.0 L/min create very weak

emission signals. The 1.0 L/min flow of the 10 % hydrogen/ 90 % argon carrier gas provided the best

signals while also preventing the coil from oxidizing at higher vaporization currents.

Page 104: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

93

Table II. Optimized Operating Parameters

Parameter Optimized

Value RF Power (kW) 1.1

Coolant Gas Flow (L/min)

17

H2/Ar Flow (L/min) 1.0 Auxiliary Flow (L/min) 0

Page 105: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

94

Figure 4. Effect of Ar coolant gas flow rate on the emission intensity of 0.50 mg/L iron

0

5

10

15

20

25

30

35

14 15 16 17 18 19 20

Rel

ativ

e E

mis

sion

Sig

nal

Coolant Flow (L/min)

Page 106: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

95

Figure 5. Effect of H2/Ar carrier gas flow rate on the emission intensity of 0.50 mg/L iron

0

50

100

150

200

250

0.5 1 1.5 2

Rel

ativ

e E

mis

sion

Sig

nal

Gas Flow (L/min)

Page 107: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

96

3.2 Limits of Detection

Using the optimized conditions in Table II, analytical figures of merit were determined for arsenic,

barium, cadmium, copper, iron, lead, manganese, palladium, platinum, selenium, silver, and strontium.

The standard solutions were prepared as indicated in Section 2.1. Five replicates of each concentration

were taken, and limits of detection (LODs) were calculated by multiplying the standard deviation of the

blank (n=10) by three and dividing by the slope of the calibration curve. For comparison to traditional

ICP-OES systems figures of merit (FOM), the ETV cell was replaced by a standard Scott-style double-

pass spray chamber (Part Number: ML155055, Meinhard Glass, Golden, CO, USA) and a conical

nebulizer (Model Number: TR-30-K2, Meinhard Glass, Golden, CO, USA). Results are listed in Table III

and an average improvement factor of 14 was observed for the twelve elements studied.

Page 108: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

97

Table III. Figures of merit for twelve elements of interest

Element Wavelength

(nm)

ETV LOD

(µg/L)

Nebulizer LOD

(µg/L)

Improvement Factor

Arsenic 193.7 70 200 3 Barium 445.4 0.1 4 40

Cadmium 226.5 2 2 1 Copper 324.8 0.5 10 20

Iron 259.9 4 4 1 Lead 220.4 2 90 45

Manganese 257.6 0.2 1 5 Palladium 340.5 10 30 3 Platinum 214.4 10 300 30 Selenium 196.1 8 100 13

Silver 328.1 1 3 3 Strontium 407.7 2 3 2

Page 109: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

98

3.3 Validation of Method

The method of ETV-ICP-AES was validated through the use of NIST SRM 1643e, trace metals in water,

using the standard additions method to reduce matrix effects often encountered when using ICP-OES.

Table IV lists the recovered values of seven elements of interest in the standard reference material. There

was a 103 % average recovery for these elements.

Page 110: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

99

Table IV. Recoveries for Seven Elements of Interest

Element Reported Value (µg/L)

Recovered Value (µg/L) % Recovery

Ag 0.11 0.11 100 Ba 0.55 0.61 110 Cd 0.7 0.7 100 Cu 0.02 0.03 108 Mn 0.42 0.42 100 Se 1.2 1.1 92 Sr 3.2 3.6 112

Page 111: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

100

CONCLUSION

A new electrothermal vaporization device was successfully constructed and used to quantify twelve

elements spanning the ultraviolet region. These elements are often difficult to excite, and an ETV source

was used to allow for improved limits of detection. This ETV allows the user to reduce necessary sample

volumes from several milliliters to only a few microliters. The ETV cell is also easily replaced by a

standard spray chamber, providing a rapid interchange between sample introduction methods. Figures of

merit were improved on average by a factor of 14. Particular success was seen with platinum, which often

relies on inductively coupled plasma-mass spectrometry because of the limited ability to analyze reduced-

volume samples associated with inductively coupled plasma atomic emission spectrometry. Future

outlooks include slurries and solids with anticipation that materials will perform well on the tungsten coil

due to the platform shape of the horizontally oriented tungsten coil.

Page 112: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

101

REFERENCES

1. Okamoto, Y.; Komori, H.; Kataoka, H.; Tsukahara, S.; Fujiwara, T., Direct determination of lead

in biological samples by electrothermal vaporization inductively coupled plasma mass spectrometry

(ETV-ICP-MS) after furnace-fusion in the sample cuvette-tungsten boat furnace. Fresenius Journal of

Analytical Chemistry 2000, 367, 300-305.

2. Gaines, P., Sample introduction methods for ICP-MS and ICP-OES. Spectroscopy 2005, (January

2005), 20-23.

3. Amberger, M. A., Broekaert, J.A.C., Direct mulitelement determination of trace elements in

boron carbide powders by slurry sampling ETV-ICP-OES. Journal of Analytical Atomic Spectrometry

2010, 35, 1308-1315.

4. Nickel, H., Zadgorska, Z., A strategy for calibrating direct ETV-ICP-OES analysis of industrial

ceramics in powder form. Fresenius' Journal of Analytical Chemistry 1995, 351, 158-163.

5. Detcheva, A., Barth, P., Hassler, Calibration possiblities and modifier use in ETV-ICP-OES

determination of trace and minor elements in plant materials. Analytical Bioanalytical Chemistry 2009,

394, 1485-1495.

6. Li, S., Hu, B., Jiang, Z., Chen, R., Direct determination of trace refractory elements in human

serum by ETV-ICP-MS with in-situ matrix removal. Analytical Bioanalytical Chemistry 2004, 379,

1076-1082.

7. Berndt, H., Schaldach, G., Simple low-cost tungsten-coil atomizer for electrothermal atomic

absorption spectrometry. Journal of Analytical Atomic Spectrometry 1988, 3, 709-712.

8. Wen, X.; Wu, P.; Chen, L.; Hou, X., Determination of cadmium in rice and water by tungsten coil

electrothermal vaporization-atomic fluorescence spectrometry and tungsten coil electrothermal atomic

absorption spectrometry after coud point extraction. Analytica Chimica Acta 2009, 650, 33-38.

Page 113: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

102

9. Davis, A. C.; Alligood, B. W.; Calloway, C. P.; Jones, B. T., Direct determination of cadmium in

urine by electrothermal vaporizer-inductively coupled plasma analysis using a tungsten coil vaporizer.

Applied Spectroscopy 2005, 59, (10), 1300-1303.

10. Nakamura, S. K., Y., Kubota, M., Temperature of a W ribbon furnace in electrothermal atomic

absorption spectrometry. Spectrochimica Acta Part B 1986, 41, 817-823.

11. Nakamura, Y., Takahashi, K., Kujirai, O., Okochi, H., Evaluation of electrothermal vaporization,

emission intensity-tie-wavelength measurement and time resolution combined with an axially viewed

horizontal inductively coupled plasma using an echelle spectrometer with wavelength modultaion.

Journal of Analytical Atomic Spectrometry 1997, 12, 349-354.

12. Shibata, N.; Fudagawa, N.; Kubota, M., Electrothermal vaporization using a tungsten furnace for

the determination of rare-earth elements by inductively coupled plasma mass spectrometry. Analytical

Chemistry 1991, 63, 636-640.

13. Davis, A. C.; Calloway, C. P.; Jones, B. T., Chelation of urinary cadmium with ammonium

pyrrolidine dithio-carbamate prior to determination by tungsten-coil inductively coupled plasma atomic

emission spectrometry. Microchemical Journal 2006, 84, (1-2), 31-37.

14. Wu, P.; Wen, X.; He, L.; He, Y.; Chen, M.; Hou, X., Evaluation of Tungsten Coil electrothermal

vaporization-Ar/H2 flame atomic fluorescence spectrometry for determination of eight traditional hydride-

forming elements and cadmium without chemical vaporization generation. Talanta 2008, 74, 505-511.

15. Hanna, S. N.; Keene, J.; Calloway, C. P.; Jones, B. T., Design of a Portable Eelctrothermal

Vaporization Flame Atomic Emission Spectrometry Device for Field Analysis. Instrumentation Science

and Technology 2011, (in press).

16. Gras, L. d. L.-V., M.T.C., Limiting effects on transport efficiency by matrix load in inductively

coupled plasma optical emission spectrometry with electrothermal vaporization sample introduction.

Spectrochimica Acta Part B 2000, 55, 37-47.

Page 114: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

103

17. Maestre, S. d. L.-V., M.T.C., Plasma behavior during electrothermal vaporization sample

introduction in inductively coupled plasma atomic emission spectrometry. Spectrochimica Acta Part B

2001, 56, 1209-1217.

Page 115: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

104

CONCLUSION

Atomic spectroscopy has been the standard for trace metal analysis for some time, and the incorporation

of novel sources into otherwise routine methods continues to allow for the development of breakthroughs

in spectroscopic techniques. The tungsten coil, despite its forty years of use, continues to produce

improvements in instrumentation development and applications.

Tungsten coil electrothermal vaporization as a sample introduction method continues to be a promising

technique. It generally outperforms its graphite counterparts by being operationally less demanding while

generating improved figures of merit. The electrothermal vaporization flame atomic emission

spectrometer discussed here is constructed for a total of less than five hundred dollars and can be operated

independently by an undergraduate researcher. Limits of detection ranged from 0.6 to 200 ng L-1, which

are on average 100 times better than figures of merit associated with traditional flame atomic emission

spectrometry. This technique is comparable to inductively coupled plasma optical emission spectrometry

but at a fraction of the cost and with the potential for portability.

The tungsten coil electrothermal vaporization sample introduction method for inductively coupled plasma

optical emission spectrometry eliminates the use of peristaltic tubing, nebulizer, and spray chamber

accessories of the sample introduction system. The newly designed aluminum electrothermal vaporization

cell fits directly onto the quartz torch via a glass to glass connection by using a simple compression fitting.

The likelihood of condensation on the sides of the vapor transport material is reduced and the device

introduces the analyte as a discreet plug into the heart of the argon plasma. Sample volumes are reduced

from the milliliter scale associated with nebulization to a microliter scale. The multielement analysis of

small-volume or precious samples is possible in real-time. The time resolved analysis feature allows user

Page 116: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

105

flexibility in data processing without the loss of emission data to dead time caused by a lagging

spectrometer. Using the tungsten coil electrothermal vaporization system, an average improvement factor

of fourteen was seen for twelve elements of interest spanning the ultraviolet and visible regions.

While this body of work has only focused on microliter scale, liquid samples, the logical next step would

be to focus on solid and slurry analysis. The ability to analyze slurries and solids eliminates the need for

extensive sample pretreatment—a step that is often the most time-consuming component to experimental

methods in atomic spectroscopy while also reducing the risk of analyte loss or contamination. The

tungsten coil would make an ideal solid sampling technique because of its rapid heating to temperatures

in excess of 3000 °C degrees. The tungsten coil’s design is already similar to that of a L’Vov platform, so

that solid or slurry samples could easily be placed on a horizontally oriented coil for rapid vaporization

into nearly any instrumental system for trace metal analysis. The cell housing for the tungsten coil would

need to be altered so that solid or slurry samples could easily be placed on the surface while ensuring that

samples did not fall off of the horizontally-oriented coil. Stability could be achieved by always

maintaining a sample cell that is parallel to the surface on which the instrument is placed.

Tungsten coil electrothermal vaporization sample introduction provides alternative methodology to

instrumentation routinely used in laboratories. This technique may also be coupled to systems that are not

normally considered for scientific use such as the welder’s torch encountered here. This method will

continue to allow for advances in atomic spectrometry by providing users with creative and flexible

techniques for a virtually infinite assortment of samples.

Page 117: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

106

CIRRICULUM VITAE

Birthplace

April 22, 1985 Marion, South Carolina

Education

Wake Forest University Winston-Salem, NC Doctor of Philosophy in Chemistry 2011 Winthrop University Rock Hill, SC Bachelor of Science in Chemistry 2007

Scholastic and Professional Experience

Research Assistant 2009-2010 Wake Forest University Winston-Salem, NC Teaching Assistant 2007-2009, 2010- 2011 Wake Forest University Winston-Salem, NC Graduate Student Instructor July 19-30, 2010 CERTL Summer Outreach Winston-Salem, NC Technology Transfer Graduate Student Intern August 2010-May 2011 Wake Forest University Baptist Medical Center Office of Technology Asset Management Winston-Salem, NC

Publications

Jiyan Gu, Summer N. Hanna, Bradley T. Jones. A Portable Tungsten Coil Atomic Emission Spectrometer for the Simultaneous Determination of Metals in Water and Soil Samples. Analytical Sciences. 2011. (in press) Summer N. Hanna, Joseph Keene, Clifton P. Calloway, Jr., Bradley T. Jones. Design of a Portable Electrothermal Vaporization Flame Atomic Emission Spectrometry Device for Field Analysis. Instrumentation, Science, and Technology. 2011. (in press) Summer N. Hanna, Bradley T. Jones. An Electrothermal Vaporization Flame Atomic Emission Spectrometer. Journal of Analytical Atomic Spectroscopy. 2011, DOI: 10.1039/c1ja10060b

Page 118: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

107

Summer N. Hanna, Bradley T. Jones. An Electrothermal Vaporization Source for Inductively Coupled Plasma Optical Emission Spectrometry. Microchemical Journal. 2011, DOI: 10.1016/j.microc.2011.04.009. Summer N. Hanna, Bradley T. Jones. A Review of Tungsten Coil Eelectrothermal Vaporization as a Sample Introduction Technique in AtomicSpectrometry. Applied Spectroscopy Reviews. 2011, DOI: 10.1080/05704928.2011.582659

Poster Presentations

Summer Hanna, Joseph Keene, Bradley T. Jones. 2009. “Cutting Edge Metal Detector:” Presented at the Wake Forest University Graduate Student Research Day. Summer Hanna, Joseph Kene, Bradley T. Jones. 2009. “Cutting Edge Metal Detector:” Presented at the Federation of Analytical Chemistry and Spectroscopy Societies Annual Conference. Summer Hanna, Bradley T. Jones. 2010. “Increasing Accessibility to Instrumentation: A “Cutting Edge Metal Detector” for Field Analysis:” Presented at the Wake Forest University Graduate Student Research Day. Summer Hanna, Bradley T. Jones. 2010. “Increasing Accessibility to Instrumentation: A “Cutting Edge Metal Detector” for Field Analysis:” Presented at the Federation of Analytical Chemistry and Spectroscopy Societies Annual Conference.

Oral Presentations

Summer Hanna. “Chemistry Beyond the Periodic Table.” 2010. Marion High School. Marion, SC Summer Hanna, Christopher MacNeil, Bradley T. Jones, Robert Coffin. 2010. “Quantification of Palladium in Low Bandgap Polymers Using Inductively Coupled Plasma Optical Emission Spectrometry.” Winthrop University, Rock Hill, SC. Summer Hanna, Bradley T. Jones. 2011. “Design of a Compact, Aluminum Electrothermal Vaporization Device for Inductively Coupled Plasma Optical Emission Spectrometry.” Presented at the Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy Annual Meeting.

Awards Dean’s Fellowship 2007-2008 Wake Forest University Winston-Salem, NC

Professional and Academic Organizations Society for Applied Spectroscopy 2008-present Wake Forest University Graduate Student Association 2007-2010

Reynolda Campus Chair 2009-2010

Page 119: TUNGSTEN COIL ELECTROTHERMAL VAPORIZATION FOR …

108

Association of University Technology Managers 2011-present

Related Professional Experience

Panelist for 2008 “Getting into College Alone: The Barriers of a First Generation College Student,” held at Winthrop University, sponsored by Pi Sigma Alpha Guest speaker and lab facilitator for “Opportunities in Science” held in 2009 at Marion High School, Marion, SC