Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

  • Upload
    others

  • View
    10

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series Instruction Manual

Page 2: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,
Page 3: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-1

1. INTRODUCTION This manual contains installation, operation and maintenance instructions for the Valor™ 4000W Series. Please read the manual completely before using the scale.

1.1 Safety Precautions Please follow these safety precautions:

Verify that the AC Adapter input voltage matches the local AC power supply. Do not drop loads on the platform. Do not place the scale upside down on the platform. Disconnect the scale from power supply when cleaning. Operate the scale only under ambient conditions specified in these

instructions. Service should be performed only by authorized personnel. Only use weights within the scale’s capacity as specified in these instructions. Do not operate the scale in hostile environments. Do not carry the scale by the pan or sub-platform. Use the handholds on the

side of the scale housing.

2. INSTALLATION 2.1 Package Contents

Scale Stainless Steel Pan

Power Adapter & Plug Instruction Manual

Warranty Card Capacity Label

2.2 Installing Components Place the stainless steel pan onto the weighing platform before turning the scale on.

Figure 2-1. Install the Stainless Steel Pan onto the weighing platform

Page 4: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-2 Valor™ 4000W Series

2.3 Selecting the Location Use the scale on a clean, firm and flat surface. Avoid locations with excessive air current, vibrations, heat sources, or rapid temperature changes. 2.4 Leveling the Equipment

Adjust the leveling feet so the bubble is centered in the circle. Be sure the equipment is level each time its location is changed.

Figure 2-2. Level Indicator

2.5 PowerThe AC Adapter is used to power the scale when battery power is not needed. First connect the AC Adapter plug to the scale input jack located at the bottom of the scale according to the description below.

Figure 2-3. Slide to plug in

Then connect the AC adapter to the proper AC supply.

Figure 2-4. Connecting the AC adapter to AC supply

Caution: The scale can only be used in a dry environment when powered by the AC adapter.

Level Bubble

Adjustable Feet

Page 5: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-3

2.5.1 Battery Power: The scale can be used on AC power immediately. Allow the battery to charge for 12 hours before using the scale on battery power. The scale will automatically switch to battery operation if there is a power failure or the power cord is removed. With AC power, the scale is constantly charging, so the battery charge indicator (see item 11 in table 3-2) will remain lit. The scale can be operated during charging, and the battery is protected against overcharging. For maximum operating time, the battery should be charged at room temperature. During battery operation, the battery symbol indicates the battery charging status. When charging, the symbol will blink slowly and when fully charged the symbol will stop blinking.

TABLE 2-1 Symbol Charge Level

Battery in use: Symbol displayed

Notes: When battery symbol blinks fast, approximately 30 minutes working time is left. When [Lo.bAt] is displayed, the scale will shut off. Charging the scale must be performed in a dry environment.

CAUTION: Battery is to be replaced only by an authorized OHAUS service dealer. Risk of explosion can occur if the rechargeable battery is replaced with the wrong type or if it is not properly connected. Dispose of the lead acid battery according to local laws and regulations.

Page 6: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-4 Valor™ 4000W Series

3. OPERATION 3.1 Controls

Figure 3-1. Front Control Panel Valor 4000W with display

TABLE 3-1

Button Functions

Short Press1 (when on): Sets display to zero Short Press (when off): Turns scale on Long Press2 (when on): Turns the scale off Short Press (in Menu): Selects / accepts displayed setting

Short Press: Accumulates the weight or displays the accumulated information at 0 load. Long Press: Toggles through active Units Short Press (in Menu): Toggles through available settings

Short Press: Initiates an application mode specific response Long Press: Selects active Mode Short Press (in Menu): returns to previous settings

Short Press: Enter / clear a Tare value Long Press: Enters User Menu Short Press (in Menu): Quickly exit User Menu

IR Sensor3 can be programmed to act as “touchless” button. See the User Menu section 4.3 for the available settings.

Notes: 1 Short Press: Press less than 2.5 seconds. 2 Long Press: Press and hold for more than 2.5 seconds. 3 The IR sensor can be activated by a hand or other object that is placed at a specified height (see

section 4.3) from the sensor. The sensor activation distance will vary based on the reflective nature of the object. If unwanted activations occur due to unique situations the sensor can be turned off.

Page 7: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-5

Display

TABLE 3-2 Display Symbols Item Description Item Description

1 NET symbol 9 Calibration / Menu Mode symbol 2 Center of Zero symbol 10 Dynamic symbol* 3 Negative symbol 11 Battery charge symbol 4 Stable weight symbol 12 Pound, Ounce, Pound:Ounce symbol 5 Preset Tare symbol* 13 Gram, Kilogram symbols 6 Pointer symbols 14 Pieces symbol* 7 IR Sensor symbol 15 Percent symbol 8 Accumulation symbol

Note: * Not Used

Figure 3-3. Valor 4000W Rear display.

The colored LED indicators on the right side of the control panels are used in Checkweigh mode (section 3.6) and will light up according to the following rules:

(Red) Loads > Upper limit

(Green) Loads ≥ Lower limit and ≤ Upper limit

(Yellow) Loads < Lower limit

1

1

IR Sensor

Figure 3-2. Valor 4000W Display

1

2 1

3 1

4 1

5 1

6 1

7

1

8 1

9 1 10 1 11

1 12

1 13

1 14

1 15

LED Indicators

Page 8: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-6 Valor™ 4000W Series

3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test, momentarily displays the software version, and then enters the active weighing mode. To turn the scale off, press and hold the On/Zero Off button until OFF is displayed. 3.3 Initial Calibration When the scale is first installed it should be calibrated to ensure accurate results. Before performing the calibration, be sure to have the appropriate calibration weights as listed in table 3-3. Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released, the display will display [C.A.L]. Press Yes to accept, [SpaN] will then be shown. Press Yes again to begin the span calibration. [--C--] blinks while zero reading is stored. Next, the display shows the calibration weight value. Place the specified calibration mass on the pan and press Yes. [--C--] blinks while the reading is stored. The scale returns to the previous application mode and is ready for use.

The message [Err 3.0 CAl] will be displayed if these calibration steps were not followed or if the wrong weight was used.

The calibration process can be aborted by turning the scale off.

TABLE 3-3 Required Span Calibration Mass (sold separately)

Max Mass1 Max Mass1

1500g 1.5kg / 3lb 6000g 6kg / 15lb 3000g 3kg / 6lb 15000g 15kg / 30lb

Note: 1 Pound masses are used when calibrating in the lb unit.

3.4 Weigh Mode 1. Press and hold Mode until [wJeIGH] (Weigh) is displayed. 2. If required, place an empty container on the pan and press Tare. 3. Add sample to the pan or container. The display shows the weight of the

sample.

Page 9: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-7

3.5 Percent Mode This mode measures the weight of a sample as a percentage of a reference weight.

1. Press and hold Mode until [PerCNt] is displayed. [Clr.ref] (clear reference) will then display.

Notes: Press Function to view the current reference weight. If there is no reference data stored, [Set.ref] is displayed. Press Yes to

set reference data. 2. Press No to use the stored reference weight and proceed to step 6. 3. If required, place an empty container on the pan and press Tare. Press Yes to

establish a new reference. Scale will now display [PUt.ref]. 4. Add the desired reference material to the container. Press Yes to store the

reference weight. The display shows 100%. 5. Remove the reference material. If required, place an empty container on the

pan and press Tare. 6. Add the sample material. The display shows the percentage of the sample

compared to reference weight. 7. To clear the stored reference data press and hold Mode until [PerCNt] is

displayed. Press Yes when [Clr.ref] is displayed. 3.6 Checkweigh Mode This mode sets low and high weight limits for portion control processes.

1. Press and hold Mode until [CHeCk] (Check) is displayed. [CLr.ref] (clear

references) will then display. Notes: Press Function to view the low and high reference weight limits. If there is no reference data stored, [Set.ref] is displayed. Press Yes to

set reference data. 2. Press No to use the stored reference weight limits and proceed to step 5. 3. Press Yes to establish new reference values. The scale will then display [Set.

LO]. Press Yes to view the “Low” limit value. Press Yes to accept or No to edit the “Low” limit value. The stored value then displays with the first digit highlighted [000.000 kg]. Repeatedly press No until the desired number appears. Press Yes to accept and highlight the next digit. Repeat until all the digits are correct. Press Yes to accept the “low” limit value, [Set. HI] will be displayed.

4. Repeat the same procedure to accept or edit the “high” value. 5. If required, place an empty container on the pan and press Tare. Place

sample material on the pan or in the container. If the sample weight is under the target weight range, the yellow LED will light. If the sample is within the target weight range, the green LED will light. If the sample is over the target weight range, the red LED will light.

6. To clear the stored reference values press and hold Mode until [CHeCk] is displayed. Press Yes when [Clr.ref] is displayed.

Page 10: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-8 Valor™ 4000W Series

3.7 Accumulation Accumulation works together with each application mode. This function allows the user to store the total of a series of weight measurements.

1. Press the M+ key to add the weight to the accumulation data. The Σ icon will keep flashing until the weight is removed.

Note: When Accumulate is set to AUTO in the setup menu, it is not necessary to press the M+ key.

2. When the pan is cleared, press the M+ key to display the statistical information of the accumulation data.

3. To clear the accumulation data press the Tare key when the display is showing the statistical information of the accumulation data. The display shows [Clr.aCC]. Press the Yes key to clear the stored data and return to current mode.

Notes: Only stable weights are stored to accumulation total. To prevent the same load from being added to the total multiple times, the load on the pan must be returned to 0 before the next load can be added. When Legal for Trade is OFF, the display must return to 0 gross, 0 net or a negative net value; when Legal for Trade is ON, the display must return to 0 gross. Otherwise, the Σ icon will continue flashing. Gross loads and net loads cannot be added to the same total. If the first load is a gross weight, future loads must also be gross weights. If the first load is a net weight, future loads must also be net weights. The max of accumulation times is 9999. Changing the mode will clear the accumulation data. Accumulation Example:

In weighing mode; sub menu ACCUmM setting, select mMaNU: If required, place an empty container on the pan and press Tare. Step i. Put the weight (0.04 kg) on the pan and press M+ key. The Σ icon indicator will keep flashing until the weight is removed. Step ii. Remove the weight from the pan. Put another weight (0.03 kg) and press M+ key: Step iii. Remove the weight from the pan: Step iv. Press M+ key to display the statistical information: Step v. If required, follow step 3 above to clear the accumulation data.

N 2 tOtal 0.070 kg

mMIN 0.030 kg mMax 0.040 kg

mMaNU

* 0.040 kg

* 0.030 kg

* 0.000 kg

Page 11: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-9

4. MENU SETTINGS The User Menu allows the customizing of scale settings.

4.1 Menu Navigation User Menu:

Menu: C.A.L S.E.t.U.P r.E.A.d M.O.d.E U.n.i.t L.O.C.k E.n.d

Menu Items:

Span Lin

GEO End

Reset Pwr.Un A.Tare Ir.Func Ir.Adj

Accum End

Reset Stable Filter AZT Light Sleep A.Off End

Reset Percnt Check End

kg g oz lb

lb:oz End

Reset L.Cal

L.Setup L.Read L.Mode L.Unit End

Notes: Some Units/Modes may not be available in all models. When LEGAL FOR TRADE is set to ON (see section 4.9), the menu settings are affected. To Enter the Menu Mode Press and hold Menu until [mMeNU] (Menu) is displayed. When released the first sub-menu [C.a.l] (Cal) will be shown. Press Yes to enter the displayed sub-menu or press No to advance to the next. Selecting a sub-menu will display the first menu item. Press Yes to view the menu item setting or press No to move to the next menu item. When viewing the setting, press Yes to accept the setting, or press No to change the setting. When [End] is displayed, press Yes to return to the sub-menu selections or No to return to the first item in the current menu. Bold indicates factory default setting. Note: The Calibration / Menu Mode indicator is displayed when in the Menu Mode. 4.2 Cal Menu Enter this menu to perform calibrations. Span [SpaN] (yes, no)

Initiates a span calibration procedure (zero and span). Lin [LIN] (yes, no)

Initiates a linearity calibration procedure (zero, mid-point and span). GEO [GEO]

Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current location. Settings from 0 to 31 are available with 12 being the default. Refer to table 4-1 to determine the GEO factor that corresponds to your location.

End Cal [End] Advance to the next menu or return to the top of the current menu.

Page 12: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-10 Valor™ 4000W Series

4.3 Setup Menu Enter this menu to set scale parameters. Reset [reset] (no, yes)

Reset the Setup menu to factory defaults. Power on unit [pwWr.UN] (auto, kg, g, lb, oz, lb:oz)

Set the unit of measure displayed at startup Auto Tare [A.tArE] (off, on, on-acc)

Set the automatic tare functionality. If “on” is selected the first stable gross weight is tared. If “on-acc” is selected, stable gross loads within the accept limits are tared (in Checkweigh mode)

IR Function [Ir.FUNC] (off, tare) Set the IR sensor functionality.

IR Adj [Ir.adj] (hi, low) Set the detecting level for the IR sensor. (For reference: Hi: ~ 100 mm / 4 inches; Low: ~ 50mm / 2 inches)

Accumulation [aCCUmM] (off, auto, manu) Set the accumulation functionality.

End Setup [End] Advance to the next menu or return to the top of the current menu.

4.4 Read Menu Enter this menu to set user preferences. Reset [reset] (no, yes)

Reset the Read menu to factory defaults. Stable Range [Stable] (0.5, 1, 2, 5)

Set the amount the reading can vary while the stability symbol remains on. Filter [FILtEr] (low, medium, high)

Set the amount of signal filtering. Auto-Zero Tracking [AZt] (off, 0.5, 1, 3)

Set the automatic zero tracking functionality. Light [LIGHt] (hi, med, low)

Sets backlight functionality. Sleep [SLEEP] (off, on)

Set the display shut off time. Auto off [A.OFF] (off, 1, 5, 10)

Set the automatic shut off functionality. End Readout [End]

Advance to the next menu or return to the top of the current menu.

Page 13: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-11

4.5 Mode Menu Enter this menu to activate modes so they will be available for use with the Mode button. Weigh mode is always active by default. Reset [reset] (no, yes)

Reset the Mode menu to factory defaults. Percent [PerCNt] (off, on)

Set the status. Check [CHECk] (off, on)

Set the sub-mode. End Mode [End]

Advance to the next menu or return to the top of the current menu.

4.6 Unit Menu Enter this menu to activate units so they will be accessible with the Units button. The units in the menu must be turned “on” to be active. Note: Available units vary by model and local regulations. 4.7 Lock Menu The Lock Menu is a software controlled option which can lock Menu settings to prevent tampering. Reset [reset] (no, yes)

Reset the Lock menu to factory defaults. Lock Cal [l.CAL] (off, on)

Set the status. Lock Setup [l.SEtUP] (off, on)

Set the status. Lock Read [l.rEAD] (off, on)

Set the status. Lock Mode [l.mMOdE] (off, on)

Set the status. Lock Unit [l.UNIt] (off, on)

Set the status. End Lock [End]

Advance to the next menu or return to the top of the current menu. 4.8 End Menu Press ‘Yes’ to advance to the Calibration menu. Press ‘No’ to exit the menu and return to the current application mode.

Page 14: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-12 Valor™ 4000W Series

TABLE 4-1. GEO CODES

0 325 650 975 1300 1625 1950 2275 2600 2925 3250325 650 975 1300 1625 1950 2275 2600 2925 3250 3575

0 1060 2130 3200 4260 5330 6400 7460 8530 9600 106601060 2130 3200 4260 5330 6400 7460 8530 9600 10660 11730

0°00' 5°46' 5 4 4 3 3 2 2 1 1 0 05°46' 9°52' 5 5 4 4 3 3 2 2 1 1 09°52' 12°44' 6 5 5 4 4 3 3 2 2 1 1

12°44' 15°06' 6 6 5 5 4 4 3 3 2 2 115°06' 17°10' 7 6 6 5 5 4 4 3 3 2 217°10' 19°02' 7 7 6 6 5 5 4 4 3 3 219°02' 20°45' 8 7 7 6 6 5 5 4 4 3 320°45' 22°22' 8 8 7 7 6 6 5 5 4 4 322°22' 23°54' 9 8 8 7 7 6 6 5 5 4 423°54' 25°21' 9 9 8 8 7 7 6 6 5 5 425°21' 26°45' 10 9 9 8 8 7 7 6 6 5 526°45' 28°06' 10 10 9 9 8 8 7 7 6 6 528°06' 29°25' 11 10 10 9 9 8 8 7 7 6 629°25' 30°41' 11 11 10 10 9 9 8 8 7 7 630°41' 31°56' 12 11 11 10 10 9 9 8 8 7 731°56' 33°09' 12 12 11 11 10 10 9 9 8 8 733°09' 34°21' 13 12 12 11 11 10 10 9 9 8 834°21' 35°31' 13 13 12 12 11 11 10 10 9 9 835°31' 36°41' 14 13 13 12 12 11 11 10 10 9 936°41' 37°50' 14 14 13 13 12 12 11 11 10 10 937°50' 38°58' 15 14 14 13 13 12 12 11 11 10 1038°58' 40°05' 15 15 14 14 13 13 12 12 11 11 1040°05' 41°12' 16 15 15 14 14 13 13 12 12 11 1141°12' 42°19' 16 16 15 15 14 14 13 13 12 12 1142°19' 43°26' 17 16 16 15 15 14 14 13 13 12 1243°26' 44°32' 17 17 16 16 15 15 14 14 13 13 1244°32' 45°38' 18 17 17 16 16 15 15 14 14 13 1345°38' 46°45' 18 18 17 17 16 16 15 15 14 14 1346°45' 47°51' 19 18 18 17 17 16 16 15 15 14 1447°51' 48°58' 19 19 18 18 17 17 16 16 15 15 1448°58' 50°06' 20 19 19 18 18 17 17 16 16 15 1550°06' 51°13' 20 20 19 19 18 18 17 17 16 16 1551°13' 52°22' 21 20 20 19 19 18 18 17 17 16 1652°22' 53°31' 21 21 20 20 19 19 18 18 17 17 1653°31' 54°41' 22 21 21 20 20 19 19 18 18 17 1754°41' 55°52' 22 22 21 21 20 20 19 19 18 18 1755°52' 57°04' 23 22 22 21 21 20 20 19 19 18 1857°04' 58°17' 23 23 22 22 21 21 20 20 19 19 1858°17' 59°32' 24 23 23 22 22 21 21 20 20 19 1959°32' 60°49' 24 24 23 23 22 22 21 21 20 20 1960°49' 62°90' 25 24 24 23 23 22 22 21 21 20 2062°90' 63°30' 25 25 24 24 23 23 22 22 21 21 2063°30' 64°55' 26 25 25 24 24 23 23 22 22 21 2164°55' 66°24' 26 26 25 25 24 24 23 23 22 22 2166°24' 67°57' 27 26 26 25 25 24 24 23 23 22 2267°57' 69°35' 27 27 26 26 25 25 24 24 23 23 2269°35' 71°21' 28 27 27 26 26 25 25 24 24 23 2371°21' 73°16' 28 28 27 27 26 26 25 25 24 24 2373°16' 75°24' 29 28 28 27 27 26 26 25 25 24 2475°24' 77°52' 29 29 28 28 27 27 26 26 25 25 2477°52' 80°56' 30 29 29 28 28 27 27 26 26 25 2580°56' 85°45' 30 30 29 29 28 28 27 27 26 26 2585°45' 90°00' 31 30 30 29 29 28 28 27 27 26 26

Elevation in meters

Elevation in feet

GEO valueLatitude

Page 15: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-13

4.9 Legal For Trade When the scale is used in trade or a legally controlled application it must be set up, verified and sealed in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met.

4.9.1 Capacity Label A label showing the capacity and readability of the scale must be installed near each display. If the Capacity Labels were installed prior to delivery, no further action is needed. If the Capacity Labels were not installed, they have been placed in the packaging material. Affix the labels above the displays as shown in Figure 4-1. Note: The Capacity Labels will be destroyed upon removal, so only attempt to install them once.

Figure 4-1. Front and rear display capacity label locations 4.9.2 Settings Before verification and sealing, perform the following steps: 1. Set Legal For Trade to OFF. 2. Verify that the menu settings meet the local weights and measures regulations. 3. Perform a calibration as explained in Section 3.3. 4. Set Legal For Trade to ON. To change the scale’s Legal For Trade (LFT) setting, follow this procedure:

Caution: When accessing the bottom of the scale, avoid placing the scale upside down on the pan or sub-platform. Place the scale on its side.

1. Turn the scale OFF. 2. Remove the Security Cover under the scale to access

the two pins located in a slot (see figure 4-2). Short these pins (a slotted screwdriver may be used), while powering the scale ON. The scale will perform the start-up procedure, and then the display will show the current status [lft OFF] (LFT OFF, unlocked) or [lft ON] (LFT ON, locked); press the ‘No’ key to change this setting, or press the ‘Yes’ key to confirm it.

Note: When Legal For Trade is set to On, the menu settings are affected as follows: Calibration (C.A.L) menu is not accessible IR Function, Filter and Unit menus are locked at the current settings. Stable Range setting is locked at 1d Auto-Zero Tracking setting is locked at 0.5d 3. Replace the Security Cover.

Capacity Label

Figure 4-2. LFT pins.

Page 16: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-14 Valor™ 4000W Series

4.9.3 Verification and Sealing The local weights and measures official or authorized service agent must perform the verification procedure.

4.9.3.1 Physical Seals For jurisdictions that use the physical sealing method, the local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings. Refer to the illustrations below for sealing methods.

Unlocked Locked

Figure 4-3. Sealing

4.9.3.2 Audit Trail Seal (USA and Canada only) For jurisdictions that use the audit trail sealing method, the local weights and measures official or authorized service agent must record the current configuration and calibration event counter values at the time of sealing. These values will be compared to values found during a future inspection.

Note: A change to an event counter value is equivalent to breaking a physical seal.

The audit trail uses two event counters to record changes to configuration and calibration settings.

The configuration event counter (CFG) will index by 1 under the following conditions:

- When the LFT setting is changed from ON to OFF.

- When exiting the menu if one or more of the following menu settings are changed: Stable Range, Auto Zero Tracking (AZT), Filter, IR Function and Units (kg, g, oz, lb or lb:oz).

The calibration event counter (CAL) will index by 1 when exiting the menu if a Span Calibration or GEO setting change is made.

Note: The counter only indexes once, even if several settings are changed.

The event counters can be viewed by pressing and holding the MENU button. While the button is held, the display will show MENU followed by Audit. Release the button when Audit is displayed to view the audit trail information.

Wire Seal Paper Seal Security Cover

Page 17: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-15

The audit trail information is displayed in the format CFGxxx and CALxxx. The scale then returns to normal operation.

5. MAINTENANCE

5.1 Cleaning The housing may be cleaned with a cloth dampened with a mild detergent if necessary. Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panels.

5.2 Cleaning the Plastic Pan 5.2.1 Uninstalling and cleaning the plastic pan Please follow the steps below to clean the plastic pan: 1. Remove the stainless steel pan 2. Remove the four thumb screws holding the battery cover. 3. Unplug the two battery clips and remove the battery. 4. Using a Phillips screwdriver, remove the two screws located at the bottom of the battery compartment. 5. Remove the plastic pan. 6. Clean the plastic pan.

6

3

1

2

4

5

Figure 5-1. Pan cleaning

5.2.2 Reinstalling the pan after cleaning Please follow the steps below to reinstall the plastic pan (see also the steps in figure 5-1, in reversed order): 1. Using a Phillips screwdriver, attach the plastic pan to the housing with the two screws. 2. Place the battery in the battery compartment and attach the two battery clips. Attach the red wire to the positive (red) battery terminal and the black wire to the negative (black) battery terminal. 3. Attach the battery cover with the four thumb screws. 4. Place the scale upright and install the stainless steel pan.

Caution: Risk of explosion can occur if the battery is not properly connected.

Page 18: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-16 Valor™ 4000W Series

5.3 Troubleshooting The following table lists common problems and possible causes and remedies. If the problem persists, contact OHAUS or your authorized dealer.

TABLE 5-1 Symptom Possible Cause Remedy

Cannot turn on No power to scale Battery discharged

Verify connections, power source and battery charge status.

Poor accuracy Improper calibration Unstable environment

Perform calibration Move scale to suitable location

Cannot calibrate Unstable environment Incorrect calibration weight

Move the scale to suitable location Use correct calibration weight

Cannot access mode Mode not enabled Enter menu and enable mode Cannot access unit Unit not enabled Enter menu and enable unit LO ref Reference weight is too low Increase reference weight. Err 3.0 Cal Incorrect calibration weight See table 3-3 for correct weights

Err 8.1 ˉLOadˉ Power on zero range exceeded Clear pan

Err 8.2 _LOad_ Power on zero under range Install pan

Err 8.3 ˉLOadˉ Overload Load exceeds scale maximum capacity

Err 8.4 _LOad_ Under load Reading below min. range - Re-install pan.

Err 8.5 ˉtareˉ Tare out of range Tare value exceeds maximum. Err 9 Data Internal data error. Contact an authorized service agent Err13 mMEmM Fail to write EEPROM. Contact an authorized service agent Err 53 C.SUmM Invalid checksum data Contact an authorized service agent

Lo.Bat Battery is discharged Connect the power and charge the battery

NO.ACC Gross and net weights cannot be accumulated together

Accumulate only gross or net weights.

Battery fails to charge fully Battery is defective Have battery replaced by OHAUS

authorized service dealer.

5.4 Service Information If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service agent. Please visit our web site, www.ohaus.com to locate the OHAUS office nearest you. An OHAUS product service specialist will be available to provide assistance. 6. TECHNICAL DATA The technical data is valid under the following ambient conditions: Indoor use only Operating temperature: -10ºC (14ºF) to 40ºC (104ºF) Relative humidity: 10% to 90% relative humidity, non-condensing Altitude: Up to 2000 m Power: AC Adapter (supplied) - 12 VDC 0.84 A output, internal rechargeable sealed lead-acid battery Mains supply voltage fluctuations: up to ±10% of the nominal voltage Installation Category: II Pollution Degree: 2

Page 19: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-17

6.1 Specifications TABLE 6-1

MODEL V41PWE1501T V41XWE1501T

V41PWE3T V41XWE3T

V41PWE6T V41XWE6T

V41PWE15T V41XWE15T

Capacity x Readability (Max x d non-approved)

1.5 kg x 0.0002 kg 1500 g x 0.2 g 3 lb x 0.0005 lb 48 oz x 0.01 oz

3 kg x 0.0005 kg 3000 g x 0.5 g 6 lb x 0.001 lb

96 oz x 0.02 oz

6 kg x 0.001 kg 6000 g x 1 g

15 lb x 0.002 lb 240 oz x 0.05 oz

15 kg x 0.002 kg 15000 g x 2 g

30 lb x 0.005 lb 480 oz x 0.1 oz

Maximum Displayed Resolution 7500 6000 6000 7500

Capacity X Readability (Max x e approved)

1.5 kg x 0.0005 kg 1500 g x 0.5 g 3 lb x 0.001 lb 48 oz x 0.02 oz

3 kg x 0.001 kg 3000 g x 1 g

6 lb x 0.002 lb 96 oz x 0.05 oz

6 kg x 0.002 kg 6000 g x 2 g

15 lb x 0.005 lb 240 oz x 0.1 oz

15 kg x 0.005 kg 15000 g x 5 g 30 lb x 0.01 lb

480 oz x 0.2 oz

Approved Resolution 3000 3000 3000 3000

Repeatability 0.0005 kg 0.001 kg 0.002 kg 0.005 kg

Linearity ± 0.0005kg ± 0.001kg ± 0.002kg ± 0.005kg

Weighing Units* Non-Approved models: g, kg, lb, oz, lb:oz

EC and OIML Approved models: g, kg Measurement Canada and NTEP Approved models: g, kg, lb, oz

Tare Range To capacity by subtraction

Stabilization Time ≤ 0.5 seconds

Safe Overload Protection 150 % of scale capacity

Weight Display 2 x Red LED (front and rear)

6-digit 7-segment, 20.5 mm / 0.8 in characters

Keyboard Four buttons

Application Modes Weighing, Percent, Check Weighing (each with Accumulate function)

Battery Operating Time (at 20ºC) Typically 50 hours with 12-hour full charge

Construction V41PW: ABS housing with 304 Stainless Steel Platform

V41XW: ABS bottom housing with 304 stainless steel top housing and platform

Ingress Protection IPX8

Pan Dimensions 190 x 242 mm / 7.5 x 9.5 in

Approval Class III

Net Weight V41PW: 3.0 kg / 6.6 lb; V41XW: 3.9 kg / 8.6 lb

Shipping Weight V41PW: 4.0 kg / 8.8 lb; V41XW: 4.9 kg / 10.8 lb

Shipping Dimensions 410 x 370 x 220 mm / 16.1 x 14.6 x 8.7 in

Note: Turn off lb:oz unit with LFT ON

6.2 Drawings and Dimensions

V41PW A 256 mm / 10.1 in. B 280 mm / 11.0 in. C 121 mm / 4.8 in.

V41XW A 256 mm / 10.1 in. B 288 mm / 11.3 in. C 124 mm / 4.9 in.

A C

B

TABLE 6-2

Figure 6-1. Dimensions

Page 20: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-18 Valor™ 4000W Series

6.3 Compliance Compliance to the following standards is indicated by the corresponding mark on the product.

Mark Standard

This product conforms to the EMC Directive 2004/108/EC, the Low Voltage Directive 2006/95/EC and the Non-Automatic Weighing Instrument Directive 2009/23/EC. The Declaration of Conformity is available online at europe.ohaus.com/europe/en/home/support/compliance.aspx.

AS/NZS CISPR 11

CAN/CSA-C22.2 No. 61010-1-12 UL Std. No. 61010-1 (3rd edition)

NSF/ANSI 169–2009

NSF/ANSI/3-A 14159-1-2010

Important notice for verified weighing instruments

Weighing Instruments verified at the place of manufacture bear one of the preceding marks on the packing label and the green ‘M’ (metrology) sticker on the descriptive data plate. They may be put into service immediately.

Weighing Instruments to be verified in two stages have no green ‘M’ (metrology) on the descriptive data plate and bear one of the preceding identification marks on the packing label. The second stage of the initial verification must be carried out by an authorized and certified service organization established within the European Community or by the National Notified Body.

The first stage of the initial verification has been carried out at the manufacturers work. It comprises all tests according to the adopted European standard EN 45501:1992, paragraph 8.2.2. If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the re-verification period and inform the respective weights and measures authorities.

Page 21: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

Valor™ 4000W Series EN-19

Disposal

In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. The Batteries Directive 2006/66/EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU Member States. To comply with this Directive, this device has been designed for safe removal of the batteries at end-of-life by a waste treatment facility.

Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. Disposal instructions in Europe are available online at europe.ohaus.com/europe/en/home/support/weee.aspx. Thank you for your contribution to environmental protection.

FCC Note This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Industry Canada Note This Class B digital apparatus complies with Canadian ICES-003. ISO 9001 Registration In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality management system is compliant with the ISO 9001 standard’s requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard. Product Registration Protect your investment. Register your product with your local OHAUS dealer or online at www.ohaus.com.

Page 22: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

EN-20 Valor™ 4000W Series

Limited Warranty

OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period OHAUS will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to OHAUS. This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than OHAUS. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by OHAUS Corporation. OHAUS Corporation shall not be liable for any consequential damages. As warranty legislation differs from state to state and country to country, please contact OHAUS or your local OHAUS dealer for further details.

Page 23: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,
Page 24: Valor™ 4000W Series Instruction Manual · EN-6 Valor™ 4000W Series 3.2 Turning Scale On/Off To turn the scale on, press the On/Zero Off button. The scale performs a display test,

OHAUS Corporation 7 Campus Drive Suite 310 Parsippany, NJ 07054 USA Tel: +1 973 377 9000 Fax: +1 973 944 7177 With offices worldwide / Con oficinas alrededor del mundo / Avec des bureaux dans le monde entier / Weltweite Geshäftsstellen / Con uffici in tutto il mondo. www.ohaus.com

*30026168* P/N 30026168B © 2013 OHAUS Corporation, all rights reserved / todos los derechos reservados / tous droits réservés / Alle Rechte vorbehalten / tutti i diritti riservati. Printed in China / Impreso en la China / Imprimé en Chine / Gedruckt in China / Stampato in Cina