Countersink
Depth Gage
Calibration
Instructions
Indicator model: 543-342BTL
Turn the indicator on by pressing the [ON/OFF] button.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Note: The Gage used in this example is a SR901 series Gage
with a radius type contact point. These instructions apply to the
903 Series and 905 Series Gages as well.
While the gage is fully compressed into the Calibration Set
Block, press the START/ZERO/ABS button.
The Gage is now set to zero.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Note: Ensure that the
bushing footprint is
resting completely onto
the block face.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Remove the gage from the block.
This value doesn’t meaning anything when
the gage is disengaged from the block.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Place the Gage back onto the Calibration Set Block to
ensure the gage is reading 0.0000in. This will complete the
self-calibration process and your gage is now ready for use.
Trulok recommends to repeat the calibration
process before every use and periodically
between measurements to ensure the most
accurate readings.
Note: If the Gage is not repeating, repeat steps 2-5.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
MHO-100-260 MHO-100-460
Countersink Angle: 100°
Countersink Diameter: 0.2600”
Thru Hole: .1250”
Countersink Depth: 0.0566”
Difference from Zero Preset: 0.0420”
MHO-100-360
Countersink Angle: 100°
Countersink Diameter: 0.3600”
Thru Hole: .1250”
Countersink Depth: 0.0986”
Difference from Zero Preset: 0.0000”
Countersink Angle: 100°
Countersink Diameter: 0.4600”
Thru Hole: .1250”
Countersink Depth: 0.1406”
Difference from Zero Preset: - 0.0420”
If you have purchased the Master Calibration Block Set, check the Gage on each
Master Block to verify the Gage is reading the full range.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
If an actual depth reading is preferred rather than a
deviation from zero, read the following steps to set up the
Gage to read an actual depth measurement.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Press the PEAK/PRESET button twice to move the cursor
down to CALC.
This screen will show with
the cursor flashing on TOL.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Once the cursor is on CALC, press the START/ZERO button
edit the calculation.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
While “ON” is flashing, press the START/ZERO button to set
the calculation.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Press the DATA/HOLD button while the “A” is flashing.
The “A” will be flashing.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Press the PEAK/PRESET button to change the “+” to a “-”.
The “+” will be flashing.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Press the MENU/ORIGIN button to exit.
The value should read -01.0000, as shown.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
From this screen press the MENU/ORGIN button. This will
return you to measurement mode.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Press the DATA/HOLD button to edit the preset.
The “P1” will be flashing.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Use the [START/ZERO] button to move the cursor to the right.
When the desired digit is flashing, use the [PEAK/PRESET]
button to decrease the digit and the [DATA/HOLD] button to
increase the digit.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Once the desired preset is entered, press the [MENU/ORIGIN]
button to exit the screen.
Note: This example is set to the depth of the MDO-100-460
Calibration Set Block as shown in step #6.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Insert the Gage into the Calibration Set Block. While the
gage is fully compressed into the block, press the
START/ZERO/ABS button. This will set the preset.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
Remove the gage from the block and re-insert to ensure the
gage is repeating. This will complete the self-calibration
process and your gage is now ready for use.
Note: If the Gage is not repeating, remove the Gage from the
Block, press and hold the PEAK/PRESET button for about 3
seconds until the P1 is flashing, and repeat step #22.
Trulok recommends to repeat the calibration
process before every use and periodically
between measurements to ensure the most
accurate readings.
COUNTERSINK DEPTH GAGE GAGE CALIBRATION
MHO-100-260 MHO-100-460
Countersink Angle: 100°
Countersink Diameter: 0.2600”
Thru Hole: .1250”
Countersink Depth: 0.0566”
Difference from Zero Preset: 0.0420”
MHO-100-360
Countersink Angle: 100°
Countersink Diameter: 0.3600”
Thru Hole: .1250”
Countersink Depth: 0.0986”
Difference from Zero Preset: 0.0000”
Countersink Angle: 100°
Countersink Diameter: 0.4600”
Thru Hole: .1250”
Countersink Depth: 0.1406”
Difference from Zero Preset: - 0.0420”
If you have purchased the Master Calibration Block Set, check the Gage on each
Master Block to verify the Gage is reading the full range.