Download pptx - ECG interpretation postCRT

Transcript
Page 1: ECG interpretation postCRT

BY-Dr Sarita Choudhary

WHY THINK ABOUT CRT Increasing number of new cases of heart failure being

diagnosed

Recent studies show despite optimal medical drug treatment mortality remains gt 25 at three years

Hospital admissions and office visits are frequent following diagnosis

20-40 of patients with heart failure have conduction

disease and QRSd gt 120 msec

Cardiac resynchronization therapy (CRT) is a

recommended treatment for patients with moderate to severe (drug-refractory) heart failure with left ventricular (LV) systolic dysfunction and evidence of ventricular dyssynchrony defined by a QRS duration gt120 ms

It improves symptoms exercise tolerance quality of life morbidity and mortality

However the problem of non-response to CRT remains crucial with prevalence of 30 of non-responders for clinical response and 45 for echocardiographic response

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class I1 CRT is indicated for patients who have LVEF less than or equal to 35 sinus rhythm LBBB with a QRS duration greater than or equal to 150 ms and NYHA class II III or ambulatory IV symptoms on GDMT (Level of Evidence A for NYHA class IIIIV Level of Evidence B for NYHA class II)

Modified recommendation (specifying CRT in patients with LBBB of 150 ms expanded to include those with NYHA class II symptoms)

Class IIa1 CRT can be useful for patients who have LVEF less than or equal to 35 sinus rhythm LBBB with a QRS duration 120 to 149 ms and NYHA class II III or ambulatory IV symptoms on GDMT (Level of Evidence B)

New recommendation

2 CRT can be useful for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class IIIambulatory class IV symptoms on GDMT(Level of Evidence A)

New recommendation

3 CRT can be useful in patients with atrial fibrillation and LVEF less than or equal to 35 on GDMT if a) the patient requires ventricular pacing or otherwise meets CRT criteria and b) AV nodal ablation or pharmacologic rate control will allow near 100 ventricular pacing with CRT (Level of Evidence B)

Modified recommendation (wording changed to indicate benefit based on ejection fraction rather than NYHA class level of evidence changed from C to B)

4 CRT can be useful for patients on GDMT who have LVEF less than or equal to 35 and are undergoing new or replacement device placement with anticipated requirement for significant (40) ventricular pacing (Level of Evidence C)

Modified recommendation (wording changed to indicate benefit based on ejection fraction and need for pacing rather than NYHA class class changed from IIb to IIa)

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class IIb1 CRT may be considered for patients who have LVEF less than or equal to 30 ischemic etiology of heart failure sinus rhythm LBBB with a QRS duration of greater than or equal to 150 ms and NYHA class I symptoms on GDMT (Level of Evidence C)

New recommendation

2 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with QRS duration 120 to 149 ms and NYHA class IIIambulatory class IV on GDMT (Level of Evidence B)

New recommendation

3 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class II symptoms on GDMT (Level of Evidence B)

New recommendation

Class III No Benefit1 CRT is not recommended for patients with NYHA class I or II symptoms and non-LBBB pattern with QRS duration less than 150 ms (Level of Evidence B)

New recommendation

2 CRT is not indicated for patients whose comorbidities andor frailty limit survival with good functional capacity to less than 1 year (Level of Evidence C)

Modified recommendation (wording changed to include cardiac as well as noncardiac comorbidities)

Ventricular Dysynchrony and Cardiac Resynchronization

Ventricular Dysynchrony1 Electrical Inter- or

Intraventricular conduction delays typically manifested as left bundle branch block

Structural disruption of myocardial collagen matrix impairing electrical conduction and mechanical efficiency

Mechanical Regional wall motion abnormalities with increased workload and stressmdashcompromising ventricular mechanics

Cardiac Resynchronization

Therapeutic intent of atrial synchronized biventricular pacing Modification of interventricular intraventricular and atrial-

ventricular activation sequences in patients with ventricular dysynchrony

Complement to optimal medical therapy

1 Tavazzi L Eur Heart J 2000211211-1214

Clinical Consequences of Ventricular Dysynchrony

Abnormal interventricular septal wall motion

Reduced dPdt

Reduced pulse pressure

Reduced EF and CO

Reduced diastolic filling time

Prolonged MR duration

Proposed Mechanisms

IntraventricularSynchrony

AtrioventricularSynchrony

InterventricularSynchrony

LAPressure

LV DiastolicFilling

RV StrokeVolume

LVESV LVEDV

Reverse Remodeling

Cardiac Resynchronization

MR dPdt EF CO( Pulse Pressure)

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 2: ECG interpretation postCRT

WHY THINK ABOUT CRT Increasing number of new cases of heart failure being

diagnosed

Recent studies show despite optimal medical drug treatment mortality remains gt 25 at three years

Hospital admissions and office visits are frequent following diagnosis

20-40 of patients with heart failure have conduction

disease and QRSd gt 120 msec

Cardiac resynchronization therapy (CRT) is a

recommended treatment for patients with moderate to severe (drug-refractory) heart failure with left ventricular (LV) systolic dysfunction and evidence of ventricular dyssynchrony defined by a QRS duration gt120 ms

It improves symptoms exercise tolerance quality of life morbidity and mortality

However the problem of non-response to CRT remains crucial with prevalence of 30 of non-responders for clinical response and 45 for echocardiographic response

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class I1 CRT is indicated for patients who have LVEF less than or equal to 35 sinus rhythm LBBB with a QRS duration greater than or equal to 150 ms and NYHA class II III or ambulatory IV symptoms on GDMT (Level of Evidence A for NYHA class IIIIV Level of Evidence B for NYHA class II)

Modified recommendation (specifying CRT in patients with LBBB of 150 ms expanded to include those with NYHA class II symptoms)

Class IIa1 CRT can be useful for patients who have LVEF less than or equal to 35 sinus rhythm LBBB with a QRS duration 120 to 149 ms and NYHA class II III or ambulatory IV symptoms on GDMT (Level of Evidence B)

New recommendation

2 CRT can be useful for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class IIIambulatory class IV symptoms on GDMT(Level of Evidence A)

New recommendation

3 CRT can be useful in patients with atrial fibrillation and LVEF less than or equal to 35 on GDMT if a) the patient requires ventricular pacing or otherwise meets CRT criteria and b) AV nodal ablation or pharmacologic rate control will allow near 100 ventricular pacing with CRT (Level of Evidence B)

Modified recommendation (wording changed to indicate benefit based on ejection fraction rather than NYHA class level of evidence changed from C to B)

4 CRT can be useful for patients on GDMT who have LVEF less than or equal to 35 and are undergoing new or replacement device placement with anticipated requirement for significant (40) ventricular pacing (Level of Evidence C)

Modified recommendation (wording changed to indicate benefit based on ejection fraction and need for pacing rather than NYHA class class changed from IIb to IIa)

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class IIb1 CRT may be considered for patients who have LVEF less than or equal to 30 ischemic etiology of heart failure sinus rhythm LBBB with a QRS duration of greater than or equal to 150 ms and NYHA class I symptoms on GDMT (Level of Evidence C)

New recommendation

2 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with QRS duration 120 to 149 ms and NYHA class IIIambulatory class IV on GDMT (Level of Evidence B)

New recommendation

3 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class II symptoms on GDMT (Level of Evidence B)

New recommendation

Class III No Benefit1 CRT is not recommended for patients with NYHA class I or II symptoms and non-LBBB pattern with QRS duration less than 150 ms (Level of Evidence B)

New recommendation

2 CRT is not indicated for patients whose comorbidities andor frailty limit survival with good functional capacity to less than 1 year (Level of Evidence C)

Modified recommendation (wording changed to include cardiac as well as noncardiac comorbidities)

Ventricular Dysynchrony and Cardiac Resynchronization

Ventricular Dysynchrony1 Electrical Inter- or

Intraventricular conduction delays typically manifested as left bundle branch block

Structural disruption of myocardial collagen matrix impairing electrical conduction and mechanical efficiency

Mechanical Regional wall motion abnormalities with increased workload and stressmdashcompromising ventricular mechanics

Cardiac Resynchronization

Therapeutic intent of atrial synchronized biventricular pacing Modification of interventricular intraventricular and atrial-

ventricular activation sequences in patients with ventricular dysynchrony

Complement to optimal medical therapy

1 Tavazzi L Eur Heart J 2000211211-1214

Clinical Consequences of Ventricular Dysynchrony

Abnormal interventricular septal wall motion

Reduced dPdt

Reduced pulse pressure

Reduced EF and CO

Reduced diastolic filling time

Prolonged MR duration

Proposed Mechanisms

IntraventricularSynchrony

AtrioventricularSynchrony

InterventricularSynchrony

LAPressure

LV DiastolicFilling

RV StrokeVolume

LVESV LVEDV

Reverse Remodeling

Cardiac Resynchronization

MR dPdt EF CO( Pulse Pressure)

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 3: ECG interpretation postCRT

Cardiac resynchronization therapy (CRT) is a

recommended treatment for patients with moderate to severe (drug-refractory) heart failure with left ventricular (LV) systolic dysfunction and evidence of ventricular dyssynchrony defined by a QRS duration gt120 ms

It improves symptoms exercise tolerance quality of life morbidity and mortality

However the problem of non-response to CRT remains crucial with prevalence of 30 of non-responders for clinical response and 45 for echocardiographic response

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class I1 CRT is indicated for patients who have LVEF less than or equal to 35 sinus rhythm LBBB with a QRS duration greater than or equal to 150 ms and NYHA class II III or ambulatory IV symptoms on GDMT (Level of Evidence A for NYHA class IIIIV Level of Evidence B for NYHA class II)

Modified recommendation (specifying CRT in patients with LBBB of 150 ms expanded to include those with NYHA class II symptoms)

Class IIa1 CRT can be useful for patients who have LVEF less than or equal to 35 sinus rhythm LBBB with a QRS duration 120 to 149 ms and NYHA class II III or ambulatory IV symptoms on GDMT (Level of Evidence B)

New recommendation

2 CRT can be useful for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class IIIambulatory class IV symptoms on GDMT(Level of Evidence A)

New recommendation

3 CRT can be useful in patients with atrial fibrillation and LVEF less than or equal to 35 on GDMT if a) the patient requires ventricular pacing or otherwise meets CRT criteria and b) AV nodal ablation or pharmacologic rate control will allow near 100 ventricular pacing with CRT (Level of Evidence B)

Modified recommendation (wording changed to indicate benefit based on ejection fraction rather than NYHA class level of evidence changed from C to B)

4 CRT can be useful for patients on GDMT who have LVEF less than or equal to 35 and are undergoing new or replacement device placement with anticipated requirement for significant (40) ventricular pacing (Level of Evidence C)

Modified recommendation (wording changed to indicate benefit based on ejection fraction and need for pacing rather than NYHA class class changed from IIb to IIa)

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class IIb1 CRT may be considered for patients who have LVEF less than or equal to 30 ischemic etiology of heart failure sinus rhythm LBBB with a QRS duration of greater than or equal to 150 ms and NYHA class I symptoms on GDMT (Level of Evidence C)

New recommendation

2 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with QRS duration 120 to 149 ms and NYHA class IIIambulatory class IV on GDMT (Level of Evidence B)

New recommendation

3 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class II symptoms on GDMT (Level of Evidence B)

New recommendation

Class III No Benefit1 CRT is not recommended for patients with NYHA class I or II symptoms and non-LBBB pattern with QRS duration less than 150 ms (Level of Evidence B)

New recommendation

2 CRT is not indicated for patients whose comorbidities andor frailty limit survival with good functional capacity to less than 1 year (Level of Evidence C)

Modified recommendation (wording changed to include cardiac as well as noncardiac comorbidities)

Ventricular Dysynchrony and Cardiac Resynchronization

Ventricular Dysynchrony1 Electrical Inter- or

Intraventricular conduction delays typically manifested as left bundle branch block

Structural disruption of myocardial collagen matrix impairing electrical conduction and mechanical efficiency

Mechanical Regional wall motion abnormalities with increased workload and stressmdashcompromising ventricular mechanics

Cardiac Resynchronization

Therapeutic intent of atrial synchronized biventricular pacing Modification of interventricular intraventricular and atrial-

ventricular activation sequences in patients with ventricular dysynchrony

Complement to optimal medical therapy

1 Tavazzi L Eur Heart J 2000211211-1214

Clinical Consequences of Ventricular Dysynchrony

Abnormal interventricular septal wall motion

Reduced dPdt

Reduced pulse pressure

Reduced EF and CO

Reduced diastolic filling time

Prolonged MR duration

Proposed Mechanisms

IntraventricularSynchrony

AtrioventricularSynchrony

InterventricularSynchrony

LAPressure

LV DiastolicFilling

RV StrokeVolume

LVESV LVEDV

Reverse Remodeling

Cardiac Resynchronization

MR dPdt EF CO( Pulse Pressure)

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 4: ECG interpretation postCRT

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class I1 CRT is indicated for patients who have LVEF less than or equal to 35 sinus rhythm LBBB with a QRS duration greater than or equal to 150 ms and NYHA class II III or ambulatory IV symptoms on GDMT (Level of Evidence A for NYHA class IIIIV Level of Evidence B for NYHA class II)

Modified recommendation (specifying CRT in patients with LBBB of 150 ms expanded to include those with NYHA class II symptoms)

Class IIa1 CRT can be useful for patients who have LVEF less than or equal to 35 sinus rhythm LBBB with a QRS duration 120 to 149 ms and NYHA class II III or ambulatory IV symptoms on GDMT (Level of Evidence B)

New recommendation

2 CRT can be useful for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class IIIambulatory class IV symptoms on GDMT(Level of Evidence A)

New recommendation

3 CRT can be useful in patients with atrial fibrillation and LVEF less than or equal to 35 on GDMT if a) the patient requires ventricular pacing or otherwise meets CRT criteria and b) AV nodal ablation or pharmacologic rate control will allow near 100 ventricular pacing with CRT (Level of Evidence B)

Modified recommendation (wording changed to indicate benefit based on ejection fraction rather than NYHA class level of evidence changed from C to B)

4 CRT can be useful for patients on GDMT who have LVEF less than or equal to 35 and are undergoing new or replacement device placement with anticipated requirement for significant (40) ventricular pacing (Level of Evidence C)

Modified recommendation (wording changed to indicate benefit based on ejection fraction and need for pacing rather than NYHA class class changed from IIb to IIa)

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class IIb1 CRT may be considered for patients who have LVEF less than or equal to 30 ischemic etiology of heart failure sinus rhythm LBBB with a QRS duration of greater than or equal to 150 ms and NYHA class I symptoms on GDMT (Level of Evidence C)

New recommendation

2 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with QRS duration 120 to 149 ms and NYHA class IIIambulatory class IV on GDMT (Level of Evidence B)

New recommendation

3 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class II symptoms on GDMT (Level of Evidence B)

New recommendation

Class III No Benefit1 CRT is not recommended for patients with NYHA class I or II symptoms and non-LBBB pattern with QRS duration less than 150 ms (Level of Evidence B)

New recommendation

2 CRT is not indicated for patients whose comorbidities andor frailty limit survival with good functional capacity to less than 1 year (Level of Evidence C)

Modified recommendation (wording changed to include cardiac as well as noncardiac comorbidities)

Ventricular Dysynchrony and Cardiac Resynchronization

Ventricular Dysynchrony1 Electrical Inter- or

Intraventricular conduction delays typically manifested as left bundle branch block

Structural disruption of myocardial collagen matrix impairing electrical conduction and mechanical efficiency

Mechanical Regional wall motion abnormalities with increased workload and stressmdashcompromising ventricular mechanics

Cardiac Resynchronization

Therapeutic intent of atrial synchronized biventricular pacing Modification of interventricular intraventricular and atrial-

ventricular activation sequences in patients with ventricular dysynchrony

Complement to optimal medical therapy

1 Tavazzi L Eur Heart J 2000211211-1214

Clinical Consequences of Ventricular Dysynchrony

Abnormal interventricular septal wall motion

Reduced dPdt

Reduced pulse pressure

Reduced EF and CO

Reduced diastolic filling time

Prolonged MR duration

Proposed Mechanisms

IntraventricularSynchrony

AtrioventricularSynchrony

InterventricularSynchrony

LAPressure

LV DiastolicFilling

RV StrokeVolume

LVESV LVEDV

Reverse Remodeling

Cardiac Resynchronization

MR dPdt EF CO( Pulse Pressure)

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 5: ECG interpretation postCRT

Notable Recommendation Changes in 2012 ACCFAHAHRS Focused Update

2012 DBT Focused Update Recommendations Comments

Class IIb1 CRT may be considered for patients who have LVEF less than or equal to 30 ischemic etiology of heart failure sinus rhythm LBBB with a QRS duration of greater than or equal to 150 ms and NYHA class I symptoms on GDMT (Level of Evidence C)

New recommendation

2 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with QRS duration 120 to 149 ms and NYHA class IIIambulatory class IV on GDMT (Level of Evidence B)

New recommendation

3 CRT may be considered for patients who have LVEF less than or equal to 35 sinus rhythm a non-LBBB pattern with a QRS duration greater than or equal to 150 ms and NYHA class II symptoms on GDMT (Level of Evidence B)

New recommendation

Class III No Benefit1 CRT is not recommended for patients with NYHA class I or II symptoms and non-LBBB pattern with QRS duration less than 150 ms (Level of Evidence B)

New recommendation

2 CRT is not indicated for patients whose comorbidities andor frailty limit survival with good functional capacity to less than 1 year (Level of Evidence C)

Modified recommendation (wording changed to include cardiac as well as noncardiac comorbidities)

Ventricular Dysynchrony and Cardiac Resynchronization

Ventricular Dysynchrony1 Electrical Inter- or

Intraventricular conduction delays typically manifested as left bundle branch block

Structural disruption of myocardial collagen matrix impairing electrical conduction and mechanical efficiency

Mechanical Regional wall motion abnormalities with increased workload and stressmdashcompromising ventricular mechanics

Cardiac Resynchronization

Therapeutic intent of atrial synchronized biventricular pacing Modification of interventricular intraventricular and atrial-

ventricular activation sequences in patients with ventricular dysynchrony

Complement to optimal medical therapy

1 Tavazzi L Eur Heart J 2000211211-1214

Clinical Consequences of Ventricular Dysynchrony

Abnormal interventricular septal wall motion

Reduced dPdt

Reduced pulse pressure

Reduced EF and CO

Reduced diastolic filling time

Prolonged MR duration

Proposed Mechanisms

IntraventricularSynchrony

AtrioventricularSynchrony

InterventricularSynchrony

LAPressure

LV DiastolicFilling

RV StrokeVolume

LVESV LVEDV

Reverse Remodeling

Cardiac Resynchronization

MR dPdt EF CO( Pulse Pressure)

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 6: ECG interpretation postCRT

Ventricular Dysynchrony and Cardiac Resynchronization

Ventricular Dysynchrony1 Electrical Inter- or

Intraventricular conduction delays typically manifested as left bundle branch block

Structural disruption of myocardial collagen matrix impairing electrical conduction and mechanical efficiency

Mechanical Regional wall motion abnormalities with increased workload and stressmdashcompromising ventricular mechanics

Cardiac Resynchronization

Therapeutic intent of atrial synchronized biventricular pacing Modification of interventricular intraventricular and atrial-

ventricular activation sequences in patients with ventricular dysynchrony

Complement to optimal medical therapy

1 Tavazzi L Eur Heart J 2000211211-1214

Clinical Consequences of Ventricular Dysynchrony

Abnormal interventricular septal wall motion

Reduced dPdt

Reduced pulse pressure

Reduced EF and CO

Reduced diastolic filling time

Prolonged MR duration

Proposed Mechanisms

IntraventricularSynchrony

AtrioventricularSynchrony

InterventricularSynchrony

LAPressure

LV DiastolicFilling

RV StrokeVolume

LVESV LVEDV

Reverse Remodeling

Cardiac Resynchronization

MR dPdt EF CO( Pulse Pressure)

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 7: ECG interpretation postCRT

Clinical Consequences of Ventricular Dysynchrony

Abnormal interventricular septal wall motion

Reduced dPdt

Reduced pulse pressure

Reduced EF and CO

Reduced diastolic filling time

Prolonged MR duration

Proposed Mechanisms

IntraventricularSynchrony

AtrioventricularSynchrony

InterventricularSynchrony

LAPressure

LV DiastolicFilling

RV StrokeVolume

LVESV LVEDV

Reverse Remodeling

Cardiac Resynchronization

MR dPdt EF CO( Pulse Pressure)

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 8: ECG interpretation postCRT

Proposed Mechanisms

IntraventricularSynchrony

AtrioventricularSynchrony

InterventricularSynchrony

LAPressure

LV DiastolicFilling

RV StrokeVolume

LVESV LVEDV

Reverse Remodeling

Cardiac Resynchronization

MR dPdt EF CO( Pulse Pressure)

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 9: ECG interpretation postCRT

Achieving Cardiac ResynchronizationMechanical Goal Atrial-synchronized bi-ventricular pacing

Transvenous Approach Standard pacing lead in RA Standard pacing or defibrillation lead in RV Specially designed left heart lead placed in a left

ventricular cardiac vein via the coronary sinus

Right AtrialLead

Right VentricularLead

Left VentricularLead

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 10: ECG interpretation postCRT

Step 1 Cannulate CS

bull Use extreme care when passing the guide catheter through vessels

bull Due to the relative stiffness of the catheter damage to the walls of the vessels may include dissections or perforations

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 11: ECG interpretation postCRT

Step 2 Perform Venograms Varying Patient Anatomy 123

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 12: ECG interpretation postCRT

Cardiac Venous Anatomy

CS Os

Middle Posterior

Postero-lateral

Great

Lateral

Antero-lateral

Anterior

Step 2 Perform Venograms

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 13: ECG interpretation postCRT

Lead in Lateral Cardiac Vein

Step 2 Perform Venograms

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 14: ECG interpretation postCRT

Step 4 Place Leads

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 15: ECG interpretation postCRT

Clinical characteristics of CRT responders and non-responders

Response more likely Response less likely

QRS duration gt150ms lt150ms

Heart disease Non-ischemic Ischemic

Dyssynchrony Present Absent

Bundle branch block Left Right

Scar burden(MRI) Low burden High burden

Non-transmural Transmural

Posterolateral segments spared

Posterolateral segments involved

Severity of mitral regurgitation

Mild-moderate Severe

Lead position Posterior-lateral Anterior or inferior

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 16: ECG interpretation postCRT

CRT Responders functional improvement -NYHA functional class(1) Quality of life(10 pt) Distance covered in 6mins hall walk test(60 m) Change of peak oxygen uptake VO2 (1-2

mlkgmin) during exercise Reduction in LV diameter LVEDV(10)LVESV(15)

Main reasons for non-response to CRT

Improper patient selection Suboptimal lead placement Inappropriate device programming

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 17: ECG interpretation postCRT

Device interrogation RV and LV capture AV and VV optimization

Capture threshold test-o To assess appropriate capture in both RV and LVo Atrial capture testing is not crucial unless

suspect atrial capture problemso Capture threshold varies Normal daily variation

posture and meal times drugs cardiac ischemia disease progression

o The diagnosis of loss of biventricular capture is documented with analysis of the QRS complexes on the surface ECG

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 18: ECG interpretation postCRT

Threshold tests in the era of CRT devices that had a single ventricular channel with an internal Y-connector to the RV and LV (as opposed to current devices that have separate ventricular channels that allow measurement of RV and LV thresholds individually)

The threshold test is initiated with a high voltage output that results in BV capture with gradual reduction in amplitude until one of the ventricles fails to be entrained resulting in a change in QRS morphology

The algorithm was designed to identify the ventricle which had lost capture by evaluating

changes in QRS axis

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 19: ECG interpretation postCRT

When to suspect loss of Biventricular capture

1048713The most frequent situation is a hemodynamic deterioration occurring after a period of significant clinical improvement

1048713In some cases can cause acute or sub‐acute pulmonary edema

1048713 The loss of biventricular capture may be asymptomatic and diagnosed during a scheduled follow‐up with the analysis of QRS complexes on surface ECG or it may be suspected from data from the device

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 20: ECG interpretation postCRT

Algorithms to confirm LV capture on a standard ECG

Ammann et al described loss of LV capture as RS ratiolt1 in V1 and gt1 in lead I sensitivity of 94 and specificity of 93 Yong and Duby Loss of LV capture is indicated by increasing

QRS positivity in lead I and loss of RV capture by increasing positivity in lead III

sensitivity of 97ndash100 and a specificity of 92ndash97

In Biventricular pacing loss of q or Q wave in lead 1 is 100 predictive of loss of LV capture

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 21: ECG interpretation postCRT

Geneva algorithmFirst step- evaluates QRS width in the limb

leads whose widening points to a change from BV capture to univentricular (LV or RV) capture

The second step evaluates net QRS amplitude in lead I with greater negativity indicating increasing participation of LV capture (ie RV-BV or BV-LV capture)

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 22: ECG interpretation postCRT

Causes of loss of permanent or temporarybiventricular pacing

Left ventricular lead dislodgement Increase in LV or RV pacing thresholds1048713 Right ventricular lead dislodgement1048713 Non-optimal AV delay1048713 Atrial tachyarrhythmias with rapid ventricular rate1048713 Low maximal tracking rate1048713 Frequent ventricular premature beats1048713 Atrial undersensing1048713 T Wave oversensing1048713 Far-field atrial sensing1048713 Ventricular double counting

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 23: ECG interpretation postCRT

A-V and V-V delays adjustment

Two major approaches are currently used to guide their adjustment echocardiography and intracardiac electrograms (IEGMs)

Echocardiography-directed adjustment can be by many different methods (eg the Ritters formula the iterative technique) but LV outflow tract velocityndashtime integral (VTI) (a surrogate measure of stroke measure) maximization is the most popular

IEGM-directed adjustment is by proprietary algorithms developed by different CIED manufacturers recommending values for the A-V and V-V delays on the basis of the intracardiac atrial RV and LV signals

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 24: ECG interpretation postCRT

AV OPTIMIZATION The AV delay is the time between the

atrial beat and the corresponding ventricular paced event Usually programmed empirically at 80ndash120 ms

A long AV delay gives the ventricle a lot of ldquoopportunityrdquo to beat on its own before the ventricular output pulse is delivered

Shorten the AV delay as much as is reasonable Paced AV delay Sensed AV delay

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 25: ECG interpretation postCRT

AV delay- significant impact on systolic function

Rule of thumb in AV delay optimization - program AV delay to about 75of the native PR interval(interval from intrinsic atrial contraction to intrinsic ventricular contraction) - for continuous ventricular pacing

Too short AV delay- undermine CRT stimulation and increase symptoms

Too long AV delay- increases MR

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 26: ECG interpretation postCRT

Rate-responsive AV delay (RRAVD) is the automatic shortening of the AV delay as the patientrsquos heart rate increases

Program this ON Sensed AV delay is from AS to VP (the sensed

AV delay starts the timer at the moment the atrial device is sensed)

Paced AV delay is from AP to VP (the paced AV delay starts the time at the moment the atrial output pulse is delivered)

As a result it is generally prudent to program a sensed AV delay that is about 25 ms shorter than the paced AV delay

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 27: ECG interpretation postCRT

VV OPTIMIZATION VV optimization- harmonize activities of RV and

LV Optimal VV timing delay is the one that

produces the greatest VTI value using echo Echo remains the gold standard of VV timing

optimization Timing optimization is that proper CRT depends

on precise timing of the ventricular contractions Timing must allow for

Adequate time for the passive filling of the ventricles Proper contraction of the right and left ventricles with

respect to each other

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 28: ECG interpretation postCRT

In up to 8ndash10 of the patients undergoing biventricular pacemaker implantation LV lead placement is not possible for a variety of reasons namely inability to cannulate the CS absence of suitable branches lack of lead stability phrenic nerve capture etc

Surgical LV epicardial lead placement using left anterior or lateral mini-thoracotomy video-assisted thoracoscopy approach and robotically enhanced systems is an option in these patients

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 29: ECG interpretation postCRT

FAILURE RESPONSE TO CRT

YES

CORRECT

NO

LOSS OF LV PACING

NO

OPTIMIZE AV

DELAYVV

DELAYNO RESPONSE

DYSSYNCHRONY PRESENT

POOR LEAD POSITION

YES

REPOSITION LV LEAD

YES

CORRECT

ATRIAL

FIBRILLATIO

NVENTRICULA

R DOUB

LE COUNTING

OVERSENSIN

GLOSS OF LV CAPTU

RELOW URL

LONG AV

DELAYPVARP EXTENSION

ATRIAL UNDERSENS

INGFREQU

ENT PVC

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 30: ECG interpretation postCRT

Pacing strategies can we do better

Typically LBBB is linked with an electrical activation sequence that courses the apex with delayed activation of the lateral and posterolateral portion of the LV

Biventricular pacing improves LV synchrony via stimulation of the late-activated regions of the LV

There is still controversy regarding the best lead positioning strategy and the choice between an optimal anatomical position targeting either the segment with maximal mechanical dyssynchrony or a region with maximal electrical delay is still up for debate

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 31: ECG interpretation postCRT

The current trend is that the LV lead be placed at an optimal anatomic pacing site (usually defined as the lateral and posterolateral LV wall)

However the lack of a favourable clinical response in nearly a third of the patients receiving CRT suggests limitations in this approach to pacing site selection

Since CRT is a form of electrical therapy for disorderly electrical activation of the heart it makes sense to attempt to target the region with the maximal electrical delay

  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69
Page 32: ECG interpretation postCRT
  • Slide 1
  • WHY THINK ABOUT CRT
  • Slide 3
  • Slide 4
  • Slide 5
  • Ventricular Dysynchrony and Cardiac Resynchronization
  • Clinical Consequences of Ventricular Dysynchrony
  • Proposed Mechanisms
  • Achieving Cardiac Resynchronization Mechanical Goal Atrial-sy
  • Step 1 Cannulate CS
  • Step 2 Perform Venograms
  • Cardiac Venous Anatomy
  • Lead in Lateral Cardiac Vein
  • Step 4 Place Leads
  • Slide 15
  • Slide 16
  • Slide 17
  • Slide 18
  • Slide 19
  • Slide 20
  • Slide 21
  • Slide 22
  • Slide 23
  • Slide 24
  • Slide 25
  • Slide 26
  • Slide 27
  • Slide 28
  • Slide 29
  • Slide 30
  • Slide 31
  • Slide 32
  • Slide 33
  • Slide 34
  • Slide 35
  • Slide 36
  • Slide 37
  • Slide 38
  • Slide 39
  • Slide 40
  • Slide 41
  • Slide 42
  • Slide 43
  • Slide 44
  • Slide 45
  • Slide 46
  • Slide 47
  • Clinical characteristics of CRT responders and non-responders
  • Slide 49
  • Device interrogation
  • Slide 51
  • Slide 52
  • Slide 53
  • When to suspect loss of Biventricular capture
  • Algorithms to confirm LV capture on a standard ECG
  • Slide 56
  • Causes of loss of permanent or temporary biventricular pacing
  • A-V and V-V delays adjustment
  • AV OPTIMIZATION
  • Slide 60
  • Slide 61
  • Slide 62
  • VV OPTIMIZATION
  • Slide 64
  • Slide 65
  • Slide 66
  • Pacing strategies can we do better
  • Slide 68
  • Slide 69