Transcript
Page 1: Prismaflex TPE 1000 and TPE 2000 Sets - Baxter Sverige. · POWERED BY Prismaflex TPE 1000 and TPE 2000 Sets Pre-connected sets for plasmafiltration POWERED BY TPE 1000-2000 2016_Nordic.indd

P O W E R E D B Y

Prismaflex TPE 1000 and

TPE 2000 Sets

Pre-connected sets for plasmafiltration

P O W E R E D B Y

TPE 1000-2000 2016_Nordic.indd 1 2016-07-04 11:54:05

Page 2: Prismaflex TPE 1000 and TPE 2000 Sets - Baxter Sverige. · POWERED BY Prismaflex TPE 1000 and TPE 2000 Sets Pre-connected sets for plasmafiltration POWERED BY TPE 1000-2000 2016_Nordic.indd

Prismaflex TPE 1000 and TPE 2000 Sets

SE/M

G120

/16-

0002

, NO/

MG1

20/1

6-00

03, F

I/MG1

20/1

6-00

02, D

K/M

G120

/16-

0002

General DataPrismaflex

TPE 1000 setPrismaflex

TPE 2000 set

Weight 680 g 760 g

Blood volume in set ± 10 % 71 ml 125 ml

Blood volume in plasmafilter 23 ml ±10% 41 ml ±10%

Minimal patient weight 9 kg adult

Prismaflex TPE 1000 set

Prismaflex TPE 2000 set

QB** = 50 ml/min 14 mmHg - 1.9 kPa –

QB** = 100 ml/min 29 mmHg - 3.9 kPa 25 mmHg - 3.3 kPa

QB** = 180 ml/min 52 mmHg - 6.9 kPa –

QB** = 200 ml/min – 51 mmHg - 6.8 kPa

QB** = 250 ml/min – 64 mmHg - 8.5 kPa

QB** = 400 ml/min – 101 mmHg - 13.5 kPa

*Arterial transmembrane pressure**Access blood flow rate***Hematocrit

Sieving coefficient

(In vivo data, 19 treatments) Albumin = 0.97, IgG = 1, Apolipoprotein B = 0.95, IgM = 0.92

Plasma filtration with blood

In vitro, values ±20% (Bovine blood at 37°C, Htc 32%, Protein concentration 60 g/l)Plasma filtration is controlled by the Prismaflex system.

00 20 40 60 80 100 120 140 160 180

10

20

30

40

50

60

Pla

sma

filtr

atio

n ra

te (

ml/m

in)

TMP (mmHg)

QF max = 49 ml/minQB = 180 ml/min

QF max = 29 ml/minQB = 100 ml/min

QF max = 16 ml/minQB = 50 ml/min

Prismaflex TPE 1000 set

0 50 100 150 200 2500

20

40

60

80

100

120

140

160

Pla

sma

filtr

atio

n ra

te (

ml/m

in)

TMP (mmHg)

QF max = 147 ml/min

QF max = 114 ml/min

QF max = 79 ml/min

QF 42 = 114 ml/min

QB = 400 ml/min

QB = 300 ml/min

QB = 200 ml/min

QB = 100 ml/min

Prismaflex TPE 2000 set

Prismaflex TPE 1000 set

Prismaflex TPE 2000 set

QB** = 50 ml/min 100 mmHg - 13.3 kPa –

QB** = 100 ml/min 140 mmHg - 18.7 kPa 120 mmHg - 16 kPa

QB** = 180 ml/min 190 mmHg - 25.3 kPa –

QB** = 200 ml/min – 171 mmHg - 22.8 kPa

QB** = 250 ml/min – 193 mmHg - 25.7 kPa

QB** = 400 ml/min – 246 mmHg - 32.8 kPa

Ordering InformationCode N° N° units/box

Prismaflex TPE 1000 setPrismaflex TPE 1000 set CKT

107143115314

4

Prismaflex TPE 2000 setPrismaflex TPE 2000 set CKT

107144115315

4

Materials

Plasmafilter hollow fiber: Polypropylene

Filter housing and headers: Polycarbonate

Filter potting compound: Polyurethane

Tubing material: Plasticized polyvinyl chloride (PVC)

Cartridge: PETG

Sterilization mode: EtO (ethylene oxide)

Filter operating specifications

Maximum blood flow rate: Prismaflex TPE 1000 set: 180 ml/min Prismaflex TPE 2000 set: 400 ml/min

Minimum blood flow rate: Prismaflex TPE 1000 set: 50 ml/min Prismaflex TPE 2000 set: 100 ml/min

Maximum TMPa* (mmHg/kPa):

Filter data

Nominal physical characteristics

Effective surface area: Prismaflex TPE 1000 set: 0.15 m² Prismaflex TPE 2000 set: 0.35 m²

Fiber internal diameter (wet): 330 μm

Fiber wall thickness: 150 μm

In Vitro PerformancesBlood pressure drop ±30%

(Bovine blood, Htc***32%, Protein concentration 60 g/l, 37°C)

Sweden:Baxter Medical AB Box 63 164 94 KistaT: 08-632 64 00 www.baxter.se

Denmark:Baxter A/STobaksvejen 2A 2860 SøborgT: 4816 6400www.baxter.dk

Finland:Baxter OyPL 119 00181 Helsinki T: (09) 8621 111 www.baxter.fi

Norway:Baxter ASGjerdrumsvei 110484 OsloT: 22 58 48 00www.baxter.no

TPE 1000-2000 2016_Nordic.indd 2 2016-07-04 11:54:09


Recommended