Transcript
Page 1: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

Surgical Technique

Transforaminal Posterior Atraumatic Lumbar Cage System

T-PAL

Page 2: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

Image intensifier control

This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended.

Processing, Reprocessing, Care and MaintenanceFor general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to:http://emea.depuysynthes.com/hcp/reprocessing-care-maintenanceFor general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Page 3: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 1

Table of Contents

Introduction T-PAL 2

AO Spine Principles 4

Indications and Contraindications 5

Surgical Technique Preoperative Planning and Preparation 6

Patient Positioning 7

Access and Exposure 8 – Minimally invasive transforaminal approach 8 – Open transforaminal approach 10

Discectomy 11

Disc Space Preparation 12

Trial for Implant Size 13

Implant Preparation 23

PEEK Implant Insertion 26

Titanium Implant Insertion 30

Posterior Support 34

Implant Removal 35

Product Information Implants 37

Instruments 39

Sets 43

Applicator Instructions 45

Function control 48

Filling Material 49

Bibliography 50

Page 4: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

2 DePuy Synthes T-PAL Surgical Technique

• Rails on top of the implant, designed to guide and turn the cage between the vertebral bodies into the desired position

• Three x-ray markers help to visualize the implant under radiographic control

T-PAL. Transforaminal Posterior Atraumatic Lumbar Cage System.

Implant with guiding rails

MaterialAvailable in two materials; PEEK (with Titanium alloy marker pins [TAN]); Titanium alloy (TAN)

Two anterior radiographic marker pinsEnable visualization of the anterior implant positionThe B1.4 pins are located approximately 2 mm from the anterior edge of the implant

One proximal radiographic marker pinEnable visualization of implant tip position while insertion

Connection cylinderPermits the pivoting mechanism in combination with the applicator

Axial windowAccommodates autogenous bone graft or bone graft substitute to allow fusion to occur through the cage

Rails on the surfaceDesigned to guide and turn the cage into the desired position

Pyramidal teethDesigned to provide resistance to implant migration

Self-distracting noseAllows for ease of insertion

Lordotic angle5° to restore the natural spine lordotic curve (except for the 7 mm height)

Page 5: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

AP trial marker Lateral trial marker

T-PAL Surgical Technique DePuy Synthes 3

• Trial implants can be placed at the same position, where after the implant will be placed

• Designed for guided insertion because of the pivoting applicator and the rails on the surface of the implants and trials

• Designed for secure insertion because of permanent connection to the implants and trials during insertion process

• OR technique based on one key instrument: the applicator

• Designed for minimally invasive surgical technique

Guided insertion technique• One key instrument for insertion

of implant and trials• Applicator is designed to allow

a controlled and guided insertion based on the pivoting option

• Security button designed to prevent implant disengagement

• Applicator is designed for minimally invasive surgeries

One surgery, one instrument

Page 6: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

coronalaxial

sagittal

1 DePuy Synthes T-PAL Surgical Technique

The four principles to be considered as the foundation for proper spine patient management underpin the design and delivery of the Curriculum: Stability – Alignment – Biology – Function.1,2

StabilityStabilization to achieve a specifi c therapeutic out-come

BiologyEtiology, pathogenesis, neural protection, and tissue healing

AlignmentBalancing the spine in three dimensions

FunctionPreservations and resto-ration of function to pre-vent disability

AO Spine Principles

Copyright © 2012 by AOSpine

1 Aebi et al (1998)2 Aebi et al (2007)

Page 7: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 5

Indications and Contraindications

Intended useThe T-PAL implant is intended to replace lumbar inter-vertebral discs and to fuse the adjacent vertebral bodies together at vertebral levels L1–S1. The T-PAL implant is designed for a transforaminal approach.

IndicationsIndications are lumbar and lumbosacral pathologies in which segmental spondylodesis is indicated, for example:• Degenerative disc diseases and spinal instabilities• Revision procedures for post-discectomy syndrome• Pseudarthrosis or failed spondylodesis• Degenerative spondylolisthesis• Isthmic spondylolisthesis

Contraindications• Vertebral body fractures• Spinal tumours• Major spinal instabilities• Primary spinal deformities• Osteoporosis

Warning: T-PAL must be applied in combination with posterior fixation.

Page 8: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

6 DePuy Synthes T-PAL Surgical Technique

Preparation

Set

01.812.001 Set T-PAL

Optional sets

01.809.011 Dilation Instrument Set

01.615.004 Insight Tubes Setor01.615.002 Insight Retractor Set,

Standard Configuration

01.612.110 Set for MIS Support System

01.605.903 Set for Minimally Invasive Posterior Instruments

Have all necessary imaging studies readily available to plan implant placement and visualize individual patient anatomy.

Have all sets readily available prior to surgery.

Preoperative Planning and Preparation

Page 9: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 7

Position the patient

Position the patient in a restored physiological lordosis and decompress the abdomen to reduce venous stasis.

Patient Positioning

Page 10: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

8 DePuy Synthes T-PAL Surgical Technique

1. Approach

Optional sets

01.809.011 Dilation Instrument Set

01.615.004 Insight Tubes Setor01.615.002 Insight Retractor Set,

Standard Configuration

01.612.110 MIS Support System Set

Locate the correct operative level with fluoroscopic views. Push the Kirschner wire into the desired facet joint. Separate the posterior soft tissue by inserting the smallest diameter dilator over the Kirschner wire. Repeat with next larger diameter dilator until required dilation is achieved. Use fluoroscopy to determine the location of dilator.

Access and Exposure Minimally invasive transforaminal approach

2a. Retraction with insight tubes

Instrument set

01.615.004 Insight Tubes Set

Optional sets

01.612.110 MIS Support System Set

01.809.011 Dilation Instrument Set

Determine the appropriate tube length from the depth indicators on the dilators.

Slide the tube over the dilators until it contacts the facet joint.

Use the flex arm to stabilize the tube to the OR table. Remove the dilators and the Kirschner wire.

Page 11: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 9

2b. Retraction with insight retractor

Instrument set

01.615.002 Insight Retractor Set, Standard Configuration

Optional Sets

01.809.011 Dilation Instrument Set

01.612.110 MIS Support System Set

Determine the appropriate retractor lengths of the cranial/caudal and medial/lateral blades from the depth indicators on the dilators.

Slide the retractor with the cranial/caudal blades over the dilators until the blades contact the facet joints. Distract the blades and introduce the second retractor with the medial/lateral blades.

Use the flex arm to stabilize the retractor to the OR table. Remove the dilators and the Kirschner wire.

3. Cut transforaminal window

Instruments

03.605.508 Osteotome, straight, black

03.605.520 Laminectomy Punch, 40°, 4.0 mm, black

Prepare a window for the transforaminal approach using the osteotome to remove the inferior facet of the cranial vertebra and the superior facet of the caudal vertebra.

With the punch, additional bone or osteophytes can be removed.

Page 12: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

11 DePuy Synthes T-PAL Surgical Technique

1. Retraction with an open transforaminal approach

Instrument

03.812.040 Lamina Spreader for T-PAL

Make a standard open incision, retract the muscle layer to view the desired segment.

Distract the segment if desired. Position the lamina spreader for T-PAL at the base of the spinous processes. Distract carefully until required distraction is achieved.

Distraction opens the posterior disc space and promotes exposure both for decompression and delivery of the implant.

2. Cut transforaminal window

Instruments

03.605.508 Osteotome, straight, black

03.605.520 Laminectomy Punch, 40°, 4.0 mm, black

Prepare a window for the transforaminal approach using the osteotome to remove the inferior facet of the cranial vertebra and the superior facet of the caudal vertebra.

With the laminectomy punch, additional bone or osteophytes can be removed.

Access and Exposure

Open transforaminal approach

Page 13: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 11

Instruments

03.605.507 Rasp, dual-sided, bayoneted, black

03.605.510 Ring Curette, straight, bayoneted, black

03.605.514 Rongeur, curved, 4.0 mm, black

03.605.520 Laminectomy Punch, 40°, 4.0 mm, black

03.605.527 Rongeur, straight, 4.0 mm, black

03.605.529 Curette, rectangular, angled, right, bayoneted, black

03.605.530 Curette, rectangular, angled, left, bayoneted, black

03.803.054 Curette, rectangular, bayoneted, black

389.767– Shaver for Intervertebral Discs, 389.777 size 7–17 mm

394.951 T-Handle with Quick Coupling

Through an incision above the pedicle, access the fora-men and remove disc material, using any of the follow-ing instruments: box and ring curettes, rongeurs as well as disc shavers.

The annulus must be preserved to provide additional support for the T-PAL implant and prevent migration of bone graft and bone graft substitute into the spinal canal.

The shavers can initially be used to ream out disc m aterial or for final removal of the disc material and cartilaginous tissue.

For removal of the tissue in the far lateral disc space, use the left/right angled curettes and the curved rongeur.

Warning: Provide enough lateral exposure to the disc to reduce dural retraction.

Discectomy

Page 14: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

12 DePuy Synthes T-PAL Surgical Technique

1. Prepare endplates

Instrument

03.605.511 Rasp, dual-sided, angled, bayoneted, black

When the discectomy is completed, use the rasp to re-move the superficial cartilaginous layers of the endplates to expose the bleeding bone.

Warning: Excessive removal of the subchondral bone may weaken the vertebral endplate. The entire removal of the endplate may result in subsidence and a loss of segmental stability.

Disc Space Preparation

2. Pack disc space

Instrument

03.605.532 Impactor, curved, standard, bayoneted, black

Before the T-PAL cage is implanted, the anterior and far lateral disc space should be filled with bone graft or bone graft substitute.

Note: For more information about the filling mate-rial chronOS see page 49.

Page 15: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

1

2

3

T-PAL Surgical Technique DePuy Synthes 13

Trial for Implant Size

1. Assemble applicator and connect non detachable trial implant

Instruments

03.812.001 Applicator Outer Shaft

03.812.307– T-PAL Small Trial Implant, size 7–17 mm, 03.812.317 non detachable

03.812.507– T-PAL Large Trial Implant, size 7–17 mm, 03.812.517 non detachable

03.812.004 Applicator Knob

The applicator must be assembled before insertion of the trial.

Attach the applicator knob to the proximal end of the applicator outer shaft by turning the knob counterclock-wise until it stops (1).

Select an appropriately size trial implant. Insert the trial implant shaft into the applicator outer shaft making sure that the arrow on the outer shaft is aligned with the dis-tal opening of the trial implant shaft (2). The trial implant shaft should now be trapped inside the applicator outer shaft (3).

Turn the applicator knob clockwise to secure the trial im-plant. During this attaching procedure the security ring moves upwards, so that the green color band is visible. Continue to turn the knob until it is tightened.

Warning: Ensure the arrows on the end of the applicator align with those on the trial implant. The con-tact surfaces between the trial and the applicator should have no gap (3).

Note: For disassembly pull the security ring down, turn the applicator knob counterclockwise until it stops. Push the small button on the applicator knob and simultaneously pull the trial implant shaft out of the applicator outer shaft. Turn the applicator knob clockwise. For detailed disassembly instruc-tions please refer to page 47.

Warning: Please read fi rst the applicator instruc-tions on page 45–46.

Page 16: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

1

2

3

11 DePuy Synthes T-PAL Surgical Technique

Trial for Implant Size

Optional: Assemble applicator and connect detachable trial implant

Instruments

03.812.001 Applicator Outer Shaft

03.812.003 Applicator Inner Shaft

03.812.004 Applicator Knob

The applicator must be assembled before insertion of the trial.

Attach the applicator knob to the proximal end of the applicator outer shaft by turning the knob counterclock-wise until it stops (1).

Insert the applicator inner shaft into the applicator outer shaft making sure that the arrow on the outer shaft is aligned with the distal opening of the inner shaft (2). The applicator inner shaft should now be trapped inside the outer shaft (3).

Note: For disassembly pull the security ring down, turn the applicator knob counterclockwise until it stops. Push the small button on the applicator knob and simultaneously pull the inner shaft out of the outer shaft. Turn the applicator knob clockwise. For detailed disassembly instructions please refer to page 47.

Warning: Please read first the applicator instruc-tions on page 45–46.

Page 17: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

1 2

3

T-PAL Surgical Technique DePuy Synthes 15

Connect detachable trial implant to the applicator

Instruments

03.812.007– T-PAL Small Trial Implant, size 7–17 mm03.812.017

03.812.207– T-PAL Large Trial Implant, size 7–17 mm03.812.217

Connect an appropriately sized trial implant to the appli-cator. Pull the security ring down and simultaneously turn the knob at the proximal end of the applicator counterclockwise. The applicator jaws open (1). Place the jaws over the proximal end of the trial implant making sure to align the arrows on the end of the applicator with those on the trial implant (2).

Turn the applicator knob clockwise to close the jaws. During this closing procedure the security ring moves upwards, so that the green color band is visible (3). Continue to turn the knob until it is tightened (2).

Note: When the applicator knob remains tightened, the trial implant can not pivot or detach.

Warning: Ensure the arrows on the end of the appli-cator align with those on the trial implant. The con-tact surfaces between the trial implant and the applicator should have no gap (2).

Page 18: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

1

2

3

16 DePuy Synthes T-PAL Surgical Technique

Trial for Implant Size

2. Insert trial implant

Optional instrument

SFW691R Prodisc-L Combined Hammer

Recheck the firm connection of trial implant to applica-tor. Insert the trial implant into the disc space, ensuring that the orientation of the trial implant is correct. The trial implant tip should be orientated medial. Maintain 10–15° between the applicator handle and the sagittal plane during trial implant insertion (1).Controlled and light hammering on the applicator may be required to advance the trial implant into the interver-tebral disc space. Use fluoroscopy to confirm position and fit of the trial implant. The tip should be positioned near the anterior edge of the adjacent vertebral bodies (2).

Notes:• Firm connection of trial implant to applicator can

be checked manually by applying pressure on the lateral side of the trial implant with the thumb. Trial implant should not pivot.

• Use soft tissue retractor 389.857–389.859 to protect soft tissue.

• Use fluoroscopy during the insertion to confirm anterior positioning of the trial implant.

Warnings: • The trial tip indicates approximate final anterior

position of trial implant (3).• Maintain 10–15° between the applicator handle

and the sagittal plane during trial implant insertion.

Page 19: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

10–15°

2

1

T-PAL Surgical Technique DePuy Synthes 17

3. Position trial implant

Optional Instrument

SFW691R Prodisc-L Combined Hammer

Turn the applicator knob counterclockwise until it stops (1)

Note: Ensure applicator knob is turned counter-clockwise until it stops to avoid trial or applicator outershaft deformation.

Controlled and light hammering on the applicator may be required to pivot the trial implant into final position (2).Use fluoroscopy during the pivoting procedure and confirm fit and position of the trial implant. Each trial has a medial/lateral and an anterior/posterior opening for position control. If the trial implant appears too small or too tight, try the next larger or smaller size height until the most secure fit is achieved.

Note: Ensure that the trial implant is positioned where the implant will be placed.

Warnings:• Maintain 10–15° between the applicator handle

and the sagittal plane during trial implant insertion.

• Do not detach the trial implant in the disc space.

Page 20: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

1 2

18 DePuy Synthes T-PAL Surgical Technique

Optional: Position trial implant

If trial implant does not pivot automatically, turn the applicator handle medially to initiate pivoting upon impaction (1). After pivoting is initiated the applicator handle must be turned back to an angle of 10–15° from the sagittal plane to pivot the trial implant into final position (2).

Warning: Maintain 10–15° between the applicator handle and the sagittal plane for final trial implant insertion.

Trial for Implant Size

Page 21: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

1 2

T-PAL Surgical Technique DePuy Synthes 19

4. Remove non detachable trial implant

Instruments

03.809.972 Oracle Slide Hammer

Optional instrument

SFW691R Prodisc-L Combined Hammer

Warning: The applicator must be in the pivoting position to remove the trial implant.

Slide the slide hammer onto the end of the applicator knob with quick coupling (1). While holding the handle with one hand, apply an upward force to the slide hammer with the other hand. Repeat this procedure until the trial implant is removed (2).

Optionally the combination hammer may also be used to remove the trial implant.

Remove the slide hammer from the handle by pushing on the end of the slide hammer.

Page 22: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

43

5

21 DePuy Synthes T-PAL Surgical Technique

Trial for Implant Size

To detach the trial implant from the applicator pull the security ring down simultaneously turn the knob coun-terclockwise until it stops (3). Push the small button on the applicator knob and remove the trial implant (4).

Insert the applicator inner shaft into the applicator outer shaft making sure that the arrow on the outer shaft is aligned with the distal opening of the inner shaft (5). The applicator inner shaft should now be trapped inside the outer shaft. The applicator is now ready to accept the implant.

Note: If the security ring cannot be pulled down, turn the knob clockwise a quarter turn. The ring can now be pulled down.

Page 23: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

1 2

T-PAL Surgical Technique DePuy Synthes 21

Optional: Remove detachable trial implant

Instrument

03.809.972 Oracle Slide Hammer

Optional instrument

SFW691R Prodisc-L Combined Hammer

Warning: The applicator must be in the pivoting position to remove the trial implant.

Slide the slide hammer onto the end of the applicator knob with quick coupling (1). While holding the handle with one hand, apply an upward force to the slide hammer with the other hand. Repeat this procedure until the trial implant is removed (2).

Optionally the combination hammer may also be used to remove the trial implant.

Remove the slide hammer from the handle by pushing on the end of the slide hammer.

Page 24: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

43

22 DePuy Synthes T-PAL Surgical Technique

Trial for Implant Size

To detach the trial implant, pull the security ring down and simultaneously turn the applicator knob counter-clockwise until it stops (3). The trial implant can now be removed from the applicator (4).

Note: If the security ring cannot be pulled down, turn the knob clockwise a quarter turn. The ring can now be pulled down.

Page 25: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 23

Implant Preparation

1. Select Implant

Implants and Instruments

08.812.007S– T-PAL Small, Cage, height 7–17 mm, 08.812.017S PEEK, sterile

08.812.207S– T-PAL Large, Cage, height 7–17 mm, 08.812.217S PEEK, sterile

04.812.007S– T-PAL Small, Cage, height 7–17 mm, 04.812.017S Titanium, sterile

04.812.207S– T-PAL Large, Cage, height 7–17 mm, 04.812.217S Titanium, sterile

03.812.044 Packing Block for T-PAL

Select the T-PAL implant that corresponds to the height and size measured using the trial implant in the previous steps.

Insert the selected implant into the appropriate packing block place.

Page 26: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

1

2

3

21 DePuy Synthes T-PAL Surgical Technique

2. Pack implant with bone graft

Instrument

03.812.043 Cancellous Bone Impactor for T-PAL

Turn the packing block on its side and use the cancellous bone impactor to firmly pack the filling material into the implant cavities (1).

Make sure the implant is well placed in the packing block to avoid implant damage while bone graft filling (2).

It is important to fill the implant until the filling material protrudes from its perforations in order to ensure optimal contact with the vertebral endplates (3).

For more information about the filling volumes, see page 37 in this surgical technique.

Note: For more information about the filling mate-rial chronOS see page 49.

Implant Preparation

Page 27: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

1

3

2

T-PAL Surgical Technique DePuy Synthes 25

3. Connect implant to the applicator

Instruments

03.812.001 Applicator Outer Shaft

03.812.003 Applicator Inner Shaft

03.812.004 Applicator Knob

To connect the implant to the applicator turn the pack-ing block upwards again. Pull the security ring down and simultaneously turn the knob at the proximal end of the applicator counterclockwise. The applicator jaws open (1). Place the jaws over the proximal end of the implant making sure to align the arrows on the end of the appli-cator with those on the implant (2).

Turn the applicator knob clockwise to close the jaws. During this closing procedure the security ring moves upwards, so that the green color band is visible. Con-tinue to turn the knob until it is tightened (3).

Note: When the applicator knob is tightened, the implant can not pivot or detach.

Warning: Ensure the arrows on the end of the appli-cator align with those on the implant. The contact surfaces between the implant and the applicator should have no gap (2).

Page 28: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

1

2

3

26 DePuy Synthes T-PAL Surgical Technique

1. Insert implant

Optional Instrument

SFW691R Prodisc-L Combined Hammer

Recheck the firm connection of implant to applicator. Insert the implant into the disc space, ensuring that the orientation of the implant is correct. The implant tip should be orientated medial. Maintain 10–15° between the applicator handle and the sagittal plane during implant insertion (1). Controlled and light hammering on the applicator may be required to advance the implant into the inter-vertebral disc space.Use fluoroscopy to confirm position and fit of the implant.The tip should be positioned near the anterior edge of the adjacent vertebral bodies (2).

Notes: • Firm connection of implant to applicator can be

checked manually by applying pressure on the lateral side of the implant with the thumb. Implant should not pivot.

• Use soft tissue retractor 389.857–389.859 to reduce soft tissue damage/injury.

• Use fluoroscopy during the insertion to confirm anterior position of the implant.

• The anterior marker pins of the implant are located approximately 2 mm from the edge of the implant.

Warnings: • The implant tip marker pin indicates approximate

final anterior position of implant (3).• Maintain 10–15° between the applicator handle

and the sagittal plane during implant insertion.

PEEK Implant Insertion

Page 29: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

10–15°1

2

T-PAL Surgical Technique DePuy Synthes 27

2. Position implant

Optional Instrument

SFW691R Prodisc-L Combined Hammer

Turn the applicator knob counterclockwise until it stops (1).

Note: Ensure applicator knob is turned counter-clockwise until it stops to avoid deformation of the applicator outershaft.

Controlled and light hammering on the applicator may be required to pivot the implant into final position (2). Use fluoroscopy during the pivoting procedure and confirm fit and position of the implant. With a medial/lateral fluoroscopic image of the cage in the final position, the two anterior pins of the implant should appear as one line.

In an anterior/posterior fluoroscopic image, the two anterior pins should be equidistant to the pedicles. The tip pin indicates the lateral edge of the implant.

Note: If bone graft or bone graft substitutes are placed into the disc space after trialing, the implant may not reach the same position as the trial.

Warning: Maintain 10–15° between the applicator handle and the sagittal plane during implant insertion.

Page 30: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

1 2

28 DePuy Synthes T-PAL Surgical Technique

Optional: Position implant

If implant does not pivot automatically, turn the applica-tor handle medially to initiate pivoting upon impaction (1). After pivoting is initiated the applicator handle must be turned back to an angle of 10–15° from the sagittal plane to pivot the implant into final position (2).

Warning: Maintain 10–15° between the applicator handle and the sagittal plane for final implant insertion.

PEEK Implant Insertion

Page 31: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 29

3. Detach implant

To detach the implant, pull the security ring down and simultaneously turn the applicator knob counterclock-wise until it stops. The applicator can now be removed from the implant.

Use fluoroscopy to verify final position of the implant. With a medial/lateral fluoroscopic image, the two anterior pins of the implant should appear as one line and the tip marker as a dot.

Note: If the security ring cannot be pulled down, turn the knob clockwise a quarter turn. The ring can now be pulled down.

Note: If the applicator does not disengage from the implant turn the applicator handle laterally to free the instrument.

Page 32: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

2

3

31 DePuy Synthes T-PAL Surgical Technique

Titanium Implant Insertion

1. Insert implant

Optional instrument

SFW691R Prodisc-L Combined Hammer

Recheck the firm connection of implant to applicator. Insert the implant into the disc space, ensuring that the orientation of the implant is correct. The implant tip should be orientated medial. Maintain 10–15° between the applicator handle and the sagittal plane during im-plant insertion (1). Controlled and light hammering on the applicator may be required to advance the implant into the interverte-bral disc space.Use fluoroscopy to confirm position and fit of the implant. The tip should be positioned near the anterior edge of the adjacent vertebral bodies (2).

Notes:• Firm connection of implant to applicator can be

checked manually by applying pressure on the lat-eral side of the implant with the thumb. Implant should not pivot.

• Use soft tissue retractor 389.857–389.859 to protect soft tissue.

• Use fluoroscopy during the insertion to confirm anterior position of the implant.

Warnings: • The implant tip indicates approximate final

anterior position of implant (3).• Maintain 10–15° between the applicator handle

and the sagittal plane during implant insertion.

Page 33: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

10–15°

1

2

T-PAL Surgical Technique DePuy Synthes 31

2. Position implant

Optional instrument

SFW691R Prodisc-L Combined Hammer

Turn the applicator knob counterclockwise until it stops (1).

Note: Ensure applicator knob is turned counter-clockwise until it stops to avoid deformation of the applicator outershaft.

Controlled and light hammering on the applicator may be required to pivot the implant into final position (2). Use fluoroscopy during the pivoting procedure and con-firm fit and position of the implant. Each implant has a medial/lateral and an anterior/posterior opening for posi-tion control.

Note: If bone graft or bone graft substitutes are placed into the disc space after trialing, the implant may not reach the same position as the trial.

Warning: Maintain 10–15° between the applicator handle and the sagittal plane during implant insertion.

Page 34: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

10–15°

1 2

32 DePuy Synthes T-PAL Surgical Technique

Optional: Position implant

If implant does not pivot automatically, turn the applicator handle medially to initiate pivoting upon impaction (1). After pivoting is initiated the applicator handle must be turned back to an angle of 10–15° from the sagittal plane to pivot the implant into final position (2).

Warning: Maintain 10–15° between the applicator handle and the sagittal plane for final implant insertion.

Titanium Implant Insertion

Page 35: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 33

3. Detach implant

To detach the implant, pull the security ring down and simultaneously turn the applicator knob counterclock-wise until it stops. The applicator can now be removed from the implant.

Use fluoroscopy to verify final position of the implant. With a medial/lateral fluoroscopic image, lateral opening of the titanium T-PAL implant should be visible.

Note: If the security ring cannot be pulled down, turn the knob clockwise a quarter turn. The ring can now be pulled down.

Note: If the applicator does not disengage from the implant turn the applicator handle laterally to free the instrument.

Page 36: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

31 DePuy Synthes T-PAL Surgical Technique

1. Pack disc space

Instrument

03.605.532 Impactor, curved, standard, bayoneted, black

After the T-PAL cage is implanted, fill the posterior disc space and the lateral disc space with bone graft or bone graft substitute to create desired conditions for fusion.

Note: For more information about the filling mate-rial chronOS see page 49.

Posterior Support

2. Supplemental fixation

The T-PAL cage is intended to be used in combination with posterior fixation (e.g. Click’X).

Page 37: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 35

Implant Removal

Implant removal with the applicator

Instruments

03.812.001 Applicator Outer Shaft

03.812.003 Applicator Inner Shaft

03.812.004 Applicator Knob

Ensure that the applicator is in the fully open position. Locate the implant and close the applicator by turning the knob clockwise until the security ring is moving upwards. There should be no gap between the applica-tor knob and the security ring. To ensure that the knob is in contact with the security ring, turn the knob coun-terclockwise until it stops, in this position the implant can pivot but not detach from the ap plicator. The im-plant can now be removed. The slap hammer may be required to facilitate removal.

Note: Distraction of the segment may facilitate implant removal. However, if possible, do not distract before ensuring a firm connection between the implant and the applicator.

Warning: The applicator must be in the pivoting position to remove the implant.

Page 38: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

36 DePuy Synthes T-PAL Surgical Technique

Implant Removal

Implant removal with the removal tool

Instrument

03.812.005 Removal Tool for T-PAL

Optional instrument

03.809.972 Oracle Slide Hammer

Ensure that the removal tool for T-PAL is in the fully open position. Locate the implant and squeeze the han-dle firmly. Advance the speed nut to lock the handle. The implant can now be removed. The slap hammer may be required to facilitate removal.

Note: When the removal tool handle is squeezed, the implant can pivot but not detach from the removal tool.

Note: Distraction of the segment may facilitate implant removal. However, if possible, do not distract before ensuring a firm connection between the implant and the removal tool.

Page 39: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 37

Graft volume

The table below shows the approximate graft volume that T-PAL implants will hold, depending on the foot-print and heights.

Anterior

Posterior

Implants

Total depthSmall implants14 mmLarge implants16 mm

Insertion depthSmall implants 10 mmLarge implants 12 mm

LengthSmall implants 28 mmLarge implants 31 mm

* 7 mm height = 0°

T-PAL Implant, 5° angle*, small footprint 10 × 28 mm

Heights Filling volume PEEK Implants Titanium Implants

7 mm 0.4 cc 08.812.007S 04.812.007S

8 mm 0.4 cc 08.812.008S 04.812.008S

9 mm 0.5 cc 08.812.009S 04.812.009S

10 mm 0.6 cc 08.812.010S 04.812.010S

11 mm 0.7 cc 08.812.011S 04.812.011S

12 mm 0.8 cc 08.812.012S 04.812.012S

13 mm 0.9 cc 08.812.013S 04.812.013S

15 mm 1.0 cc 08.812.015S 04.812.015S

17 mm 1.3 cc 08.812.017S 04.812.017S

T-PAL Implant, 5° angle*, large footprint 12 × 31 mm

Heights Filling volume PEEK Implants Titanium Implants

7 mm 0.6 cc 08.812.207S 04.812.207S

8 mm 0.7 cc 08.812.208S 04.812.208S

9 mm 0.8 cc 08.812.209S 04.812.209S

10 mm 0.9 cc 08.812.210S 04.812.210S

11 mm 1.1 cc 08.812.211S 04.812.211S

12 mm 1.2 cc 08.812.212S 04.812.212S

13 mm 1.4 cc 08.812.213S 04.812.213S

15 mm 1.7 cc 08.812.215S 04.812.215S

17 mm 2.0 cc 08.812.217S 04.812.217S

Page 40: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

38 DePuy Synthes T-PAL Surgical Technique

Height

Height Posterior Height

* 7 mm height = 0°

T-PAL Implant, 5° angle*, small footprint 10 × 28 mm

Heights Posterior height PEEK Implants Titanium Implants

7 mm 7.0 mm 08.812.007S 04.812.007S

8 mm 7.2 mm 08.812.008S 04.812.008S

9 mm 8.2 mm 08.812.009S 04.812.009S

10 mm 9.2 mm 08.812.010S 04.812.010S

11 mm 10.2 mm 08.812.011S 04.812.011S

12 mm 11.2 mm 08.812.012S 04.812.012S

13 mm 12.2 mm 08.812.013S 04.812.013S

15 mm 14.2 mm 08.812.015S 04.812.015S

17 mm 16.2 mm 08.812.017S 04.812.017S

T-PAL Implant, 5° angle*, large footprint 12 × 31 mm

Heights Posterior height PEEK Implants Titanium Implants

7 mm 7.0 mm 08.812.207S 04.812.207S

8 mm 7.0 mm 08.812.208S 04.812.208S

9 mm 8.0 mm 08.812.209S 04.812.209S

10 mm 9.0 mm 08.812.210S 04.812.210S

11 mm 10.0 mm 08.812.211S 04.812.211S

12 mm 11.0 mm 08.812.212S 04.812.212S

13 mm 12.0 mm 08.812.213S 04.812.213S

15 mm 14.0 mm 08.812.215S 04.812.215S

17 mm 16.0 mm 08.812.217S 04.812.217S

Implants

Page 41: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 39

03.812.001 Applicator Outer Shaft

03.812.003 Applicator Inner Shaft

03.812.004 Applicator Knob

03.812.005 Removal Tool for T-PAL

03.812.043 Cancellous Bone Impactor for T-PAL

Instruments

03.812.307– T-PAL Small Trial Implant size 7–17 mm, 03.812.317 non detachable

03.812.507– T-PAL Large Trial Implant, size 7–17 mm, 03.812.517 non detachable

Page 42: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

11 DePuy Synthes T-PAL Surgical Technique

03.809.972 Oracle Slide Hammer

SFW691R Prodisc-L Combined Hammer

03.812.044 Packing Block for T-PAL

03.605.514 Rongeur, curved, 4.0 mm, black

03.605.520 Laminectomy Punch, 40°, 4.0 mm, black

03.605.527 Rongeur, straight, 4.0 mm, black

Instruments

Page 43: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 11

03.605.529 Curette, rectangular, angled, right, bayoneted, black

03.605.530 Curette, rectangular, angled, left, bayoneted, black

03.803.054 Curette, rectangular, bayoneted, black

03.605.532 Impactor, curved, standard, bayoneted, black

03.605.508 Osteotome, straight, black

03.605.507 Rasp, dual-sided, bayoneted, black

03.605.510 Ring Curette, straight, bayoneted, black

03.605.511 Rasp, dual-sided, angled, bayoneted, black

Page 44: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

12 DePuy Synthes T-PAL Surgical Technique

389.857 – Soft Tissue Retractor, width 6, 8, 10 mm389.859

389.767– Shaver for Intervertebral Discs, 389.777 size 7–17 mm

394.951 T-Handle with Quick Coupling

03.812.040 Lamina Spreader for T-PAL

03.812.007– T-PAL Small Trial Implant, size 7–17 mm03.812.017

03.812.207– T-PAL Large Trial Implant, size 7–17 mm03.812.217

Optional Instruments

Instruments

Page 45: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 13

Sets

Vario Case

68.812.001 Vario Case for T-PAL

Instruments

03.812.001 Applicator Outer Shaft

03.812.003 Applicator Inner Shaft

03.812.004 Applicator Knob

03.812.005 Removal Tool for T-PAL

03.812.307– T-PAL Small Trial Implant, size 7–17 mm, 03.812.317 non detachable

03.812.507– T-PAL Large Trial Implant, size 7–17 mm, 03.812.517 non detachable

03.812.043 Cancellous Bone Impactor for T-PAL

03.812.044 Packing Block for T-PAL

03.809.972 Oracle Slide Hammer

SFW691R Prodisc-L Combined Hammer

Page 46: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

11 DePuy Synthes T-PAL Surgical Technique

Vario Case

68.812.002 Disc Removal Set for T-PAL

Instruments

03.605.507 Rasp, dual-sided, bayoneted, black

03.605.508 Osteotome, straight, black

03.605.510 Ring Curette, straight, bayoneted, black

03.605.511 Rasp, dual-sided, angled, bayoneted, black

03.605.514 Rongeur, curved, 4.0 mm, black

03.605.520 Laminectomy Punch, 40°, 4.0 mm, black

03.605.527 Rongeur, straight, 4.0 mm, black

03.605.529 Curette, rectangular, angled, right, bayoneted, black

03.605.530 Curette, rectangular, angled, left, bayoneted, black

03.803.054 Curette, rectangular, bayoneted, black

03.605.532 Impactor, curved, standard, bayoneted, black

03.812.040 Lamina Spreader for T-PAL

389.857– Soft Tissue Retractor,389.859 width 6, 8 and 10 mm

389.767– Shaver for Intervertebral Discs, 389.777 size 7–17 mm

394.951 T-Handle with Quick Coupling

Sets

Page 47: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 15

Attach positionPull the security ring down and simultaneously turn the knob counterclockwise. No gap between the handle, security ring and the applicator knob should be present. The green color band should not be visible.

The implant or trial can be attached.

Insertion positionTurn the applicator knob clockwise to close the jaws. During this closing procedure the security ring moves upwards, so that the green color band is visible. Continue to turn the knob until it is tightened.

In the insertion position; the implant or trial is fixed. The implant or trial can not pivot or detach.

Applicator Instructions

Page 48: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

16 DePuy Synthes T-PAL Surgical Technique

Pivoting positionTurn the applicator knob counterclockwise until it stops. The applicator knob and the security ring will now be in contact.

In this position the implant or trial can pivot 80°. Implant or trial cannot detach from applicator.

Detach positionPull the security ring down and simultaneously turn the knob counterclockwise. No gap between the handle, security ring and the Applicator knob should be present. The green color band should not be visible.

The implant or trial can be detached.

Note: If the security ring cannot be pulled down, turn the knob clockwise a quarter turn. The ring can now be pulled down.

Applicator Instructions

Page 49: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

1

2

12

1

T-PAL Surgical Technique DePuy Synthes 17

2

1

3

Applicator Outer ShaftInner ShaftKnobT-PAL Small Trial Implant, size 7–17 mm, non detachableT-PAL Large Trial Implant, size 7–17 mm, non detachable

03.812.00103.812.00303.812.004

03.812.307–31703.812.507–517

4

Page 50: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

18 DePuy Synthes T-PAL Surgical Technique

Possible damage• Overlaying deformation at the

edge of the groove

Prevention• none

Recommendation• Exchange instrument

Function Control

03.812.003 Applicator Inner Shaft

Page 51: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

T-PAL Surgical Technique DePuy Synthes 19

Filling Material

chronOS is a bone graft substitute consisting of pure b-tricalcium phosphate. Its compressive strength is simi-lar to that of cancellous bone once it has been incorpo-rated and remodeled.1 Based on literature, the use of b-tricalcium phosphate in the spinal column is a valuable alternative to allografts and autografts, even when larger amounts are required.2

Resorbable It is being replaced in the human body by host bone in 6 to 18 months; depending on the indication and the patient’s conditions.2,3–5

Synthetic Having a synthetic origin, chronOS offers the advantage of uniform quality and unlimited availability.3

Osteoconductive Interconnected macropores of defined size (100–500 μm) facilitate bone formation throughout the entire implant. Interconnected micropores (<10 μm) allow supply of nutrients.1,6

Osteoinductive with bone marrow The patient’s blood, blood platelet concentrate or bone marrow aspirate enhances the properties of chronOS required for fusion.4,5

chronOS Granules

Art. no. B (mm) cc

710.000S 0.5–0.7 0.5

710.001S 0.7–1.4 0.5

710.002S 0.7–1.4 1

710.003S 0.7–1.4 2.5

710.011S 1.4–2.8 2.5

710.014S 1.4–2.8 5

710.019S 1.4–2.8 10

710.021S 1.4–2.8 20

710.024S 2.8–5.6 2.5

710.025S 2.8–5.6 5

710.026S 2.8–5.6 10

710.027S 2.8–5.6 20

1 Gazdag et al. 1995 ² Muschik et al. 2001 ³ Stoll et al. 2004 4 Becker et al. 2006 5 Wheeler et al. 2005 6 Lu et al. 1999

chronOS – synthetic b-tricalcium phosphate cancellous bone substitute

Page 52: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

51 DePuy Synthes T-PAL Surgical Technique

Bibliography

Aebi M, Arlet V, Webb JK, (2007): AOSPINE Manual(2 vols), Stuttgart, New York: Thieme.

Aebi M, Thalgott JS, Webb JK (1998): AO ASIF Principlesin Spine Surgery. Berlin: Springer.

Becker S, Maissen O, Ponomarev I, Stoll T, Rahn B, Wilke I (2006) Osteopromotion by a b-Tricalcium Phosphate/Bone Marrow Hybrid Implant for Use in Spine Surgery. Spine 31 (1): 11-17 Gazdag AR, Lane JM, Glaser D, Forster RA (1995) Alter-natives to autogenous bone graft: efficacy and indica-tions. J Am Acad Orthop Surg 3: 1-8. Lu J. X. , et al. Role of interconnections in porous bio ceramics on bone recolonization in vitro and in vivo. J Mater Sci Mater Med. 1999; 10(2): 111–120. Muschik M, Ludwig R, Halbhubner S, Bursche K, Stoll T (2001) Beta-tricalcium phosphate as a bone substitute for dorsal spinal fusion in adolescent idiopathic scoliosis: preliminary results of a prospective clinical study. Eur Spine J. 10 (2):178–84. Stoll T, Maissen O, Meury T, Becker S (2004) New as-pects in osteoinduction. Mat.-wiss. u. Werk stofftech 35: 198–202 Wheeler D. L., et al. Grafting of Massive Tibial Subchon-dral Bone Defects in a Caprine Module Using beta-Trical-cium Phosphate versus Autograft. J Orthop Trauma 2005; 19(2): 85–91.

Page 53: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care
Page 54: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care
Page 55: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care
Page 56: Transforaminal Posterior Atraumatic Lumbar Cage System T-PALsynthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · For general information about reprocessing, care

0123

Synthes GmbHEimattstrasse 34436 OberdorfSwitzerlandTel: +41 61 965 61 11Fax: +41 61 965 66 00www.depuysynthes.com

Not all products are currently available in all markets.

This publication is not intended for distribution in the USA.

All surgical techniques are available as PDF files at www.depuysynthes.com/ifu ©

DeP

uy S

ynth

es S

pine

, a d

ivis

ion

of S

ynth

es G

mbH

. 201

7.

All

right

s re

serv

ed.

036.

001.

088

DS

EM

/SP

N/0

714/

0145

(2)

11/1

7