Simple Maintenance Procedures Extend the Life of Your Handpieces Call us to purchase maintenance products or to schedule staff training! Between-Patient Manual Maintenance for HIGH-SPEED HANDPIECES 1 2 3 3 Clean and lubricate Clean and lubricate When applying the lubricant into a 4/5-hole handpiece with dropper bottle, use the drive air tube only. Place 1-3 drops from the dropper bottle at all points as indicated below. For motors, perform weekly, as needed, or as specified by the manufacturer. For attachments and heads, perform between patients. Clean threads monthly with 2x2 and alcohol or Tidy Whities wipes. LOW-SPEED MOTORS & AIR SCALERS Run the handpiece on a flush station Bag and sterilize Bag and sterilize Run the handpiece Run for 20 seconds. Handpieces requiring a bur tool must be run with the bur installed. If flush station is not installed, run handpiece chairside. Wipe handpiece exterior with paper towel. Always let the handpieces go through the complete drying cycle (never use dry heat). Handpieces are ready to use and DO NOT NEED TO BE RELUBRICATED. Never sterilize with a bur installed. Always let the handpieces go through the complete drying cycle (never use dry heat). Handpieces are ready to use and DO NOT NEED TO BE RELUBRICATED. Never sterilize with a bur installed. Run for 20 seconds. Wipe handpiece exterior with paper towel. • Longer turbine life! • Easy & efficient • Quickly lube 2 high-speeds and 1 E-type handpiece Available in aerosol, dropper bottle or bottle for automatic lubrication station Tidy Whities TM Hayes Automatic Lubrication Station OneStep TM Handpiece Lubricant LOW-SPEED ATTACHMENTS & HEADS ELECTRIC, ENDO & IMPLANT HANDPIECES 1 1 3 3 1 3 3 1 1 1 1 1 1 1 1 1 1 3 1 2 3 1 1 1 1 1 1 1 1 1 1 2 2 1 1 3 1 1 1 1 1 3 Drive Air Optional on lube-free Omit on single speed Aerosol with nozzle adapter Aerosol with nozzle adapter E-Type nozzle Omit on single speed Aerosol with nozzle adapter 3 We sell the DCI Flush Station! Call Hayes for more information. www.hayeshandpiece.com Star NSK Lares W&H/Adec less bulb Kavo E-Type attach 5-Hole 3-Hole Midwest XGT/Stylus/ Bien-Air W&H/Adec w/bulb/ Schein FACTORY-TRAINED Original Handpiece & Instrument Wipes Cleans & shines • Convenient • Non-corrosive Scrub and rinse handpieces with mild soap and water or wipe with Tidy Whities original handpiece & instru- ment wipes before lubricating. (Do not wipe down with cleaning/disinfecting wipes or other harsh chemicals.) AEROSOL INSTRUCTIONS: A. Remove bur B. Select the appropriate nozzle and insert into swivel back-end or back of 4/5-hole handpiece Selecting the proper nozzle will ensure a tight seal to propel lubricant throughout the handpiece C. Spray one second when using Hayes OneStep aerosol (Refer to manufacturer instructions for J. Morita high-speeds) D. Clean chuck weekly with lube and proxy brush Between-Patient/Weekly Manual Maintenance for LOW-SPEEDS, ATTACHMENTS & SCALERS 1 Lubricating and flushing electric handpieces between patients is crucial. However, never lubricate an electric motor. Please consult your Hayes office for detailed instructions. 2 Remove microbial contamination from waterlines and bottles Protect staff, patients, and handpieces! Waterlines! Treat Your

Hayes Canada - Dental Handpiece Maintenance

Embed Size (px)

DESCRIPTION

Hayes Dental Handpiece Repair provides a big variety of professional dental handpiece repair tools for brands like Midwest, Star, Kavo Handpiece Repairs, NSK/Brasseler, and Bien-Air. Our skilled tools area unit top quality and precise that area unit designed for technicians and additionally for dentists World Health Organization have the information and want to interchange by themselves dental turbines, headgears, middle shafts, etc.

Citation preview

  • Simple Maintenance ProceduresExtend the Life of Your Handpieces

    Call us to purchase maintenance products or to schedule staff training!

    Between-Patient Manual Maintenance for

    HIGH-SPEED HANDPIECES

    1

    2

    3 3

    Clean and lubricate Clean and lubricate

    When applying the lubricant into a 4/5-hole handpiece with dropper bottle, use the drive air tube only. Place 1-3 drops from the dropper bottle at all points as indicated below. For motors, perform weekly, as needed, or as specified by the manufacturer. For attachments and heads, perform between patients. Clean threads monthly with 2x2 and alcohol or Tidy Whities wipes.

    LOW-SPEED MOTORS & AIR SCALERS

    Run the handpiece on a flush station

    Bag and sterilize Bag and sterilize

    Run the handpiece

    Run for 20 seconds.Handpieces requiring a bur tool must be run with the bur installed. If flush station is not installed, run handpiece chairside. Wipe handpiece exterior with paper towel.

    Always let the handpieces go through the completedrying cycle (never use dry heat). Handpieces areready to use and DO NOT NEED TO BERELUBRICATED. Never sterilize with a bur installed.

    Always let the handpieces go through the completedrying cycle (never use dry heat). Handpieces areready to use and DO NOT NEED TO BERELUBRICATED. Never sterilize with a bur installed.

    Run for 20 seconds. Wipe handpiece exterior with paper towel.

    • Longer turbine life!

    • Easy & efficient

    • Quickly lube 2 high-speeds and 1 E-type handpiece

    Available in aerosol, dropper bottle or bottle for automatic lubrication station

    Tidy WhitiesTM

    Hayes Automatic Lubrication Station

    OneStepTM

    Handpiece Lubricant

    LOW-SPEED ATTACHMENTS & HEADS

    ELECTRIC, ENDO & IMPLANT HANDPIECES

    11

    33

    1

    3

    3

    1

    1

    11

    1

    11

    1

    1

    1 3

    1

    2

    3

    1

    1

    1 1

    1

    1

    111

    1

    2

    2

    1

    13

    11

    11

    1

    3

    Drive Air

    Optional on lube-free

    Omit on single speed

    Aerosol with nozzle adapter

    Aerosol with nozzle adapter

    E-Type nozzle

    Omit on single speed

    Aerosol with nozzle adapter

    3

    We sell the DCI Flush Station! Call Hayes for more information.

    www.hayeshandpiece.com

    Star NSK Lares W&H/Adec less bulb

    Kavo

    E-Type attach 5-Hole 3-Hole

    Midwest XGT/Stylus/Bien-Air

    W&H/Adec w/bulb/Schein

    FACTORY-TRAINED

    Original Handpiece & Instrument Wipes

    • Cleans & shines• Convenient • Non-corrosive

    Scrub and rinse handpieces with mild soap and water or wipe with Tidy Whities™ original handpiece & instru-ment wipes before lubricating. (Do not wipe down with cleaning/disinfecting wipes or other harsh chemicals.)

    AEROSOL INSTRUCTIONS:A. Remove burB. Select the appropriate nozzle and insert into

    swivel back-end or back of 4/5-hole handpiece Selecting the proper nozzle will ensure a tight seal to propel lubricant throughout the handpiece

    C. Spray one second when using Hayes OneStep aerosol (Refer to manufacturer instructions for J. Morita high-speeds)

    D. Clean chuck weekly with lube and proxy brush

    Between-Patient/Weekly Manual Maintenance for

    LOW-SPEEDS, ATTACHMENTS & SCALERS

    1

    Lubricating and flushing electric handpieces between patients is crucial. However, never lubricate an electric motor. Please consult your Hayes office for detailed instructions.

    2

    • Remove microbial

    contamination from waterlines and bottles

    • Protect staff, patients, and handpieces!

    Waterlines!Treat Your