There are three key parts of the Image Receptor for Conventional Radiography:◦Film to record the image◦Intensifying Screens to expose the film◦Cassette to protect the screens and film

Page 3: Intensifying screens

Less than 1% of the incident x-rays interact with the film to contribute to the latent image.

The intensifying screens converts the remnant radiation to light than produces the latent image. They act as an amplifier of the remnant radiation.

Page 4: Intensifying screens

About 30% of the x-rays striking the screens interact with the screens producing a large number of visible light photons.

The use of intensifying screens results in considerable lower radiation dose to the patient but has the disadvantage of causing a slight blurring of the image.

Page 5: Intensifying screens

Most conventional radiographic cassettes have a pair of screens that sandwich the film. This design used double emulsion film.

Page 6: Intensifying screens

Four Distinct Layers◦ Protective Coating◦ Phosphor◦ Reflective layer◦ Base

Page 7: Intensifying screens

Coating is transparent to light. Resistant to abrasion and damage from

handling. Resistant to static electricity Provide a surface for cleaning while

protecting the phosphors.

Page 8: Intensifying screens

The active layer of the screen is the phosphors.

The phosphors emit light when stimulated by x-rays.

Prior to 1970 the most common phosphor was a crystalline form of Calcium Tungstate.

Page 9: Intensifying screens

Modern screens use rare earth elements such as:◦Gadolinium◦Lanthanum◦Yttrium

Page 10: Intensifying screens

High atomic number so x-ray absorption will be high. Quantum Detective Efficiency (DQE)

Emit a large amount of light per x-ray absorption. Conversion Efficiency

Light must be of proper wavelength to match the sensitivity of the film Spectral Matching

Page 11: Intensifying screens

Phosphor Afterglow should be minimal.

Phosphor should not be affected by heat humidity or other environmental conditions

Page 12: Intensifying screens

Thickness of the phosphor Layer

Concentration of the crystals

Size of the crystals.

Page 13: Intensifying screens

The light from the phosphors is emitted isotropically.

Without a reflective layer, only half of the light would interact with the film.

The reflective layer redirects the light to the film.

Page 14: Intensifying screens

Some screens have special dyes that absorb the light photons coming at a large angles.

These photons would increase the image blur.

Only the photons perpendicular to the film are emitted. The dye increases spatial resolution but reduce speed.

Page 15: Intensifying screens

The base is the layer farthest from the film.

It is usually made of polyester. The base should be:

◦ Rugged and moisture resistant◦ Can not be damaged by radiation or discoloration◦ Chemically inert, flexible and free of impurities.

Page 16: Intensifying screens

The x-ray photon is absorbed by the target atom. The outer shell electron is raised to an excited state. It returns to a ground state with emission of a light

photon.

Page 17: Intensifying screens

Any material that gives of light in response to a stimulus is a luminescent material.

Two types of luminescent material. Fluorescent: gives off light only during

stimulus. Good for screens Phosphorescence: continues to give off

light after stimulus. Bad for screens called Lag or Afterglow.

Page 18: Intensifying screens

Phosphor composition: Rare earth screens are very efficient in conversion of x-ray to light.

Page 19: Intensifying screens

Phosphor thickness: The thicker the phosphor layer, the higher the number of x-rays converted to light.

High speed screens have thick layer. Detail screens have a thin layer.

Reflective layer will increase speed and blur

Page 20: Intensifying screens

Dye: Light controlling dyes are added to control the light spread to improve spatial resolution.

Crystal size: Larger crystal produce more light per interaction. Detail screens have small crystals.

Page 21: Intensifying screens

Concentration of crystals: ◦The higher the concentration of crystals, the higher the speed.

Page 22: Intensifying screens

Rare earth screens have increased speed for two reasons:

◦ Detective Quantum Efficiency (DQE) or the ability to absorb the photons ( High Z number)

◦ Conversion Efficiency: Amount of light emitted per x-ray.

Page 23: Intensifying screens

Conversion Efficiency: High conversion efficiency results in increases image noise.

Noise (any unwanted information) can be speckled appearance or grainy◦ It occurs with fast screens and use of high kVp.

The factors that make rare earth screens have greater speed also contribute to increased noise.

Increased conversion efficiency results in lower exposure. Less x-rays results in an increased quantum mottle.

Page 24: Intensifying screens

Image detail is the result of spatial resolution and contrast resolution.

Generally the conditions that increase speed reduce spatial resolution.

Page 25: Intensifying screens

When screen phosphors reacts with x-rays a larger area of the film is exposed than what would be exposed by radiation alone.

This results in reduced spatial resolution and more blur.

Page 26: Intensifying screens

Direct exposure can resolve 50 lp/mm with a very small focal spot.

High speed screens can resolve 7 lp/mm.

Detail screens can resolve 15 lp/mm

The unaided eye can resolve 10 lp/mm.

Page 27: Intensifying screens

High speed screens have thick layers of crystal and /or large crystals.

High detail screens have a thin layer of small crystals.

Page 28: Intensifying screens

Screens in pairs and double emulsion film is the standard of the industry. Less than 1% of the image is produced by the x-ray photons.

Each screen contributes relatively evenly in

the production of the image.

Page 29: Intensifying screens

The cassette is a rigid holder for the film and screens.

It will contain some form of compression to push the film in close contact with the screens.

The front of the cassette is made of a radiolucent material with low absorption characteristics.

Page 30: Intensifying screens

The back of the cassette may contain some form of metal that can absorb x-rays that are not absorbed by the screens.

Sometime with cassettes that do not adequately absorb the rays, back scatter will result from scatter radiation from the cassette holder or near by wall.

Page 31: Intensifying screens

For the screen to work at maximum efficiency, the light absorption characteristics of the film must be matched to the light emitted from the screens.

This is called spectrum matching.

Page 32: Intensifying screens

Calcium Tungstate emits a broad blue spectrum.

Rare earth emits a green spectrum.

The film, screens and safelight must match.

Page 33: Intensifying screens

Screens in the cassette can be of two types or speeds. Some people use two different speeds in cassette for full spine radiography.

When types of screens are different, they are referred to as Asymmetric screens. One side may be high contrast and the other side wide latitude. The combined image is superior.

Page 34: Intensifying screens

High quality radiography requires that the screens be clean and free of artifacts.

Avoid touching the screens with your hands.

Clean the screens with screen cleaner.

Do not slide the film in or out when loading the cassette.

Page 35: Intensifying screens

Keeping the dark room clean will help reduce dirt or dust getting into the cassette.

Don’t stack the cassette on top of each other as the weight can damage the cassette.

Load the film completely in the cassette.

Page 36: Intensifying screens

Clean the screens at least quarterly

Use only specially formulated screen cleaner with anti static properties

Never use alcohol to clean screens

Make sure they are dry before reloading with film.

Page 37: Intensifying screens

The hinge of the cassette has failed, resulting in a light leak.

Page 38: Intensifying screens

Card inside cassette

Page 39: Intensifying screens

This cassette popped partly open.

With cassette artifacts, think about how the cassette opens.

If the cassette pops open do not use the film.

Page 40: Intensifying screens

Dirty screens will appear as white spots on the film.

This film also has some static electricity artifacts.

Page 41: Intensifying screens

Dirty or damaged screens will cause white spots on the image.

Page 42: Intensifying screens

Dirty or damaged screen will cause white spots on the image.

Page 43: Intensifying screens

The white spots on this film are the result of damaged or worn out screens.

Never use alcohol or detergents to clean screens.

Page 44: Intensifying screens

Poor screen contact will cause an area of the image to appear cloudy and blurry.

Common reasons for poor contact include:◦ Worn contact felt◦ Loose, bent or broken hinges◦ Loose bent or broken latches◦ Warped screen

Page 45: Intensifying screens

Common reasons for poor contact include:◦ Warped cassette front or frame.◦ Sprung or cracked cassette frame.◦ Foreign matter in the cassette.

Screen contact is tested using a wire mesh test tool.

The wire mesh is placed on top of the cassette.

Page 46: Intensifying screens

A radiograph is taken and the film processed.

The image is viewed from 2 to 3 meters from the view box.

Poor contact will appear as a cloudy and blurry area on the film.

Page 47: Intensifying screens

Test the cassette when they are purchased and then twice yearly.

Page 48: Intensifying screens

Procedure: Clean screens and let

them dry. Use screen cleaner design for the screen used.

With a felt tip pen, write an identification number on the screen next to the I.D. and on the back of the cassette.

Load cassettes.

Page 49: Intensifying screens

Procedure: Set SID to 40” Table

Top Place cassette on

table. Place wire mesh tool

on cassette. Set collimation to film

size. Make exposure and

process film.

Page 50: Intensifying screens

Procedure:

Hang film on view box.

Step back 72” from view box and view film.

Areas of increased density or loss of resolution indicates poor contact or stained screens.

Page 51: Intensifying screens

There is a loss of detail in the thoracic and lumbar spine due to poor screen contact.

This was a new cassette.

Page 52: Intensifying screens

Note the blurry image in the spine but sharp image of the ribs.

The screens were not in proper contact in the middle of the cassette due to a bow in the cassette back.