BD Kiestra™ TLA (Total Lab Automation)Technical Specifications
Media Plate SpecificationsDiameter (bottom) 85–91 mmDiameter (lid) 89–93 mmHeight (including lid) 13–16.2 mm
Sample Container Specifications*Container Diameter 10–50 mmContainer Height 60–125 mmCap/Lid Diameter 10–54 mmCap/Lid Height 11–30 mm* Some exceptions apply. Please check with your BD representative for additional container compatibility information.
Biosafety Cabinet Requires additional electric power 208–230V, +/ – 10% 13A / 50–60Hz, Max 16A
IT Requirements 2x 1GBps CAT 6 - preferably load balanced.Remote access through BD Assurity Link™.Access to ports 443; 8531; 10050; 10051.*�All�Configurations�come�with�a�System�Control� Unit (SCU)
LIS InterfaceBidirectional interface based on ASTM protocol
ReadA Compact Incubation Specifications Incubation Atmosphere: 02 or CO2
Temperature Range: 30°C–40°C ± 1°C CO2 Consumption: 50 L/h per ReadA Compact Note: Air compressors are not included in equipment configurations�and�need�to�be�ordered�separately.
Environmental ConditionsInstallation Category II and Pollution Degree 2 as per IEC 664
Noise Level <67 dBA
Environmental Requirements Clearance for Operator Service: Floor: 800 mm / 31.5 in arround all sides� �Height:�minimal�floor�height�
2780 mm / 109.5 inTemperature Requirements: 18°C–27°C / 64.4°F–80.6°F Relative Humidity: 20%–80% non-condensing Operating Elevation: 0–2000 m
BD Kiestra™ TLA (Total Lab Automation)Technical Specifications
BD, Sparks, MD 21152-0999 USATel: 800.638.8663
bd.com© 2016 BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company. 3-223576 May 2017 Printed in the U.S.
SorterA/ BarcodA InoqulA+™ TLA Ergonomic A
ReadA Compact ProceedA
ReadingWorkstation SCU**
Max Heat Production (BTU/hr)
*Approximately
*1365 *2730 *1365 *2730 NA *1365 *3750
Electrical Requirements
208–240 Vac 50–60 Hz
1x16A
208–240 Vac 50–60 Hz
1x16A
208–240 Vac 50–60 Hz
1x16A
208–240 Vac 50–60 Hz
1x16A
Dependent on customizable track
configuration
208–240 Vac 50–60 Hz
1x16A
208–204 Vac50–60 Hz
2x group 16A
Power Consumption (W) 400 800 400 800 NA 400 1100
Pneumatic Requirements
Total pneumatic requirements dependent on
system�configuration
>8 bar minimum, 100 L /min minimum Moisture and oil-free
ISO8573-1 Preferably Class IV
>8 bar minimum, 250 L /min minimum Moisture and oil-free
ISO8573-1 Preferably Class IV
N/A >8 bar minimum, 80 L /min minimumMoisture and oil-free
ISO8573-1 Preferably Class IV
>8 bar minimum, 50 L /min minimum Moisture and oil-free
ISO8573-1 Preferably Class IV
N/A N/A
Space Requirements: Width
*Size dependent on�configuration
*SorterA�Configuration Range: 18–3 to 24–6 2281 mm–2720 mm
90 in–107 in
2659 mm / 105 in ErgonomicA 165: 1650 mm / 65 in 200: 2000 mm / 79 in 220: 2200 mm / 87 in
1000 mm / 40 in Dependent on customizable track
configuration
Stand-Alone ErgonomicA
165: 1650 mm / 65 in 200: 2000 mm / 79 in 220: 2200 mm / 87 in
750 mm / 29.5 in
Space Requirements:Depth
*Size dependent on�configuration
*SorterA�Configuration Range: 18–3 to 24–6 830 mm–1380 mm
33 in–54 in
933 mm / 37 in 850 mm / 33.5 in 1594 mm / 63 in Dependent on customizable track
configuration
850 mm / 33.5 in 1130 mm / 44.4 in
Space Requirements: Height
1950 mm / 76.77 in(with signaling lamp)
1950 mm / 76.77 in(with signaling lamp)2196.66 mm / 86.48 in with biosafety hood
1950 mm / 76.77 in(with signaling lamp)
2300 mm / 76.55 in 1950 mm / 76.77 in(with signaling lamp)
1950 mm / 76.77 in(with signaling lamp)
1015 mm / 39.9 in
Weight *Approximately
*300–500 kg / 660–1100 lbs
*700 kg / 1543 lbs *130–160 kg / 285–350 lbs
*500kg / 1100 lbs *500–700 kg / 1100–1540 lbs
*150 kg / 330 lbs *262 kg / 577 lbs
*�*All�Configurations�come�with�a�System�Control�Unit�(SCU).
Recommended