Comparison of Single and Triple Quadrupole GC-MS Instruments for Full Scan Analysis of Semivolatile Organic CompoundsAmit Gujar,1 Paul Silcock,2 and Alexander Semyonov1 1Thermo Fisher Scientific, Austin, TX, USA; 2Thermo Fisher Scientific, Runcorn, UK
Techn
ical No
te 104
31
Key WordsFull Scan Acquisition, LOD, Semivolatile Organic Compounds, Spectral Quality
IntroductionSingle quadrupole mass spectrometers, such as the Thermo Scientific™ ISQ™ Series GC-MS system, are a popular choice among analysts using GC-MS for compound identification and quantitation in relatively simple matrices. However, triple quadrupole mass spectrometers such as the Thermo Scientific™ TSQ™ Duo GC-MS/MS and TSQ™ 8000 Evo GC-MS/MS systems have been gaining popularity due to their high selectivity and lower detection limits, especially in complex matrices.
The higher cost of triple quadrupole instruments is one of the main barriers to their adoption by analytical labs, however, the cost differential between single and triple quadrupole mass spectrometers has decreased in recent years such that today, triple quadrupole instruments are a viable alternative to their single quadrupole counterparts.1
An obvious concern for analysts considering using triple quadrupole GC-MS systems to perform single quadrupole GC-MS workflows is how the full scan performance of the triple quadrupole GC-MS/MS compares to that of the single quadrupole system. In this work, full scan performance of the TSQ Duo GC-MS/MS system is compared to that of the ISQ GC-MS system for full spectrum acquisition of semivolatile organic compounds. Limits of Detection (LOD) and spectral quality (as defined by search-match results of NIST 14 Mass Spectral Library) were chosen as the criteria for comparison.
Goal
To compare the full scan acquisition performance of the TSQ Duo triple quadrupole instrument with that of the ISQ single quadrupole instrument with respect to limits of detection (LOD) and spectral quality of semivolatile organic compounds
ExperimentalSample and sample preparation: Use Restek 8270 MegaMix® as the source of semivolatile organic compounds for analysis. This mixture contains 76 compounds. All but two of these compounds are at 1000 µg/mL concentrations. The other two compounds, 3-methylphenol and 4-methylphenol, are at 500 µg/mL. More details of these compounds can be found at http://www.restek.com/catalog/view/5744.
Use dichloromethane (DCM) extract of the soil matrix to dilute the MegaMix® to 1 µg/mL. Use the DCM extract of soil to take advantage of the matrix-induced chromatographic response enhancement phenomena.2
Prepare the extract as follows: Place 10 g of clean loam soil (Fluka mfr. no. CLNLOAM6-100g) in a screw-top vial with a PTFE liner on the cap and add 10 mL DCM.
2 Shake the vial using a wrist-action shaker (Burrell Scientific Model 75) for 30 minutes. Allow the mixture to settle for 30 minutes, and then remove the DCM extract from the soil using a glass pipette. Add 10 mL DCM to the same soil matrix and repeat the shaking and DCM extract removal process two more times.
Dry the DCM extract over sodium sulfate. Use Thermo Scientific™ eVol™ Sample Dispensing System to dilute MegaMix to 1 µg/mL using the DCM extract prepared above. Store the sample in a refrigerator at 1.5 °C in a National Scientific Target DP® vial (C4000-2W) until it is ready to be injected into the GC-MS.
GC-MS conditions: Use a TSQ Duo GC-MS/MS system and an ISQ GC-MS system, each mated to a Thermo Scientific™ TRACE™ 1310 GC and an AI/AS 1310 autosampler. Table 1 shows the autosampler, GC, and MS parameters used for the experiment.
Table 1. Autosampler, GC, and MS parameters used for the experiment.
Perform 10 sample injections on both the TSQ Duo GC-MS/MS and the ISQ GC-MS using the prepared 1 µg/mL MegaMix mixture. Choose a suitable quantitative ion that is unique to the compound and not included in the coeluting compounds. In most cases, the quantitative ion is the base peak for the compound. Calculate the LOD of the compounds by statistical analysis using the two-tailed student t-distribution at a 99% confidence interval for 10 back-to-back injections. Use the Thermo Scientific™ Dionex™ Chromeleon™ 7.2 SR1 CDS software for qualitative and quantitative data analysis.
AI/AS 1310
Syringe 10 µL, 25 gauge, 50 mm length, cone tip (P/N 36500525)
Injection Volume 1 µL
Plunger Strokes 3
Viscous Sample No
Sampling Depth Bottom
Pre-/Post-Injection Dwell Time 0.0
Pre-Injection Solvent, Cycles Dichloromethane, 2
Sample Rinses 2
Post-Injection Solvent, Cycles Dichloromethane, 3
TRACE 1310 GC
Column TG-5SILMS 30 m × 0.25 mm × 0.25 µm
Liner Restek Sky® Inlet liner, splitless single taper gooseneck with wool, 4 mm × 6.5 mm × 76.5 mm
SSL Mode Splitless with surge
Inlet Temperature 270 °C
Split Flow 100 mL/min
Splitless Time 0.6 min
Surge Pressure 207 kPa
Surge Duration 0.6 min
Septum Purge Flow Constant flow of 5.0 mL/min
Carrier Flow Constant He flow of 1.2 mL/min
Oven Program 40 °C (1 min), 25 °C/min to 280 °C, 5 °C/min to 320 °C (1 min)
TSQ Duo GC-MS/MS and ISQ GC-MS
Method Type Acquisition – General
MS Transfer Line Temperature 280 °C
Ion Source Temperature 280 °C
Ionization Mode EI, 70 eV
Emission Current 50 µA
Scan Start At 2.4 min
Scan Range 35–550 amu
Scan Time 0.2 s
Chrom. Filter Peak Width 1.0 s
Use Last Tuned Detector Gained? Yes
Tuning Used EI Standard tune for TSQ Duo GC-MS/MS and EI Full tune for ISQ GC-MS (with 219 intensity set to 20 million)
3
Figure 3. Limits of detection comparison on selected compounds on the TSQ Duo GC-MS/MS and the ISQ GC-MS.
Figure 2. Comparing mass spectra of Benzyl butyl phthalate on the TSQ Duo GC-MS/MS and the ISQ GC-MS.
Figure 1. Comparing chromatograms for injections done on the TSQ Duo GC-MS/MS and the ISQ GC-MS.
Results and DiscussionFigure 1 shows the chromatograms for the full scan analyses run on the TSQ Duo GC-MS/MS and the ISQ GC-MS. The chromatograms look similar both in terms of the order of elution of components and their relative intensities. Figure 2 shows the comparison of mass spectra of Benzyl butyl phthalate on the TSQ Duo GC-MS/MS and the ISQ Series instrument. The fragmentation patterns and the relative intensities of the mass fragments are similar on both the systems.
LOD results: Figure 3 compares the results for the LODs for selected semivolatile organic compounds on the TSQ Duo GC-MS/MS and the ISQ GC-MS systems. The LODs for both are similar with 39 compounds showing lower LODs for the TSQ Duo GC-MS/MS than the ISQ GC-MS. The average % RSDs for the quantitative ion areas of 10 consecutive injections for the TSQ Duo system was 4.2% compared to 3.6% for the ISQ instrument.
2.4 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 19.7
Minutes
–5.0e70.0e0
5.0e8
Inten
sity [
coun
ts]
1 µg/mL Restek 8270 MegaMix® on TSQ Duo GC-MS/MS Duo
Inten
sity [
coun
ts]2.6 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 19.7
Minutes
–1.0e80.0e0
1.0e9
1 µg/mL Restek 8270 MegaMix® on ISQ GC-MS
35 100 200 300 400 500 550Mass [m/z]
−20
0
20
40
60
80
100
120
Rel.
Inten
sity [
%]
91.1
149.1
206.1
Rel.
Inten
sity [
%]
35 100 200 300 400 500 550Mass [m/z]
−20
0
20
40
60
80
100
120
91
149
206
0
50
100
150
200
250
Phen
ol
Bis(
2-ch
loro
ethyl)
ether
1,
3-Di
chlo
robe
nzen
e 1,
4-Di
chlo
robe
nzen
e Be
nzyl
alcoh
ol
1,2-
Dich
loro
benz
ene
2,2'
-oxy
bis(
1-ch
loro
prop
ane)
2-M
ethyl
phen
ol
Hexa
chlo
roeth
ane
Nitro
benz
ene
Isoph
oron
e Bi
s(2-
chlo
roeth
oxy)
meth
ane
1,2,
4-Tr
ichlo
robe
nzen
e Na
phth
alene
4-
Chlo
roan
iline
He
xach
loro
butad
iene
4-Ch
loro
-3-m
ethylp
heno
l 2-
Meth
ylnap
htha
lene
1-M
ethyln
apht
halen
e 2-
Chlo
rona
phth
alene
Di
meth
ylpht
halat
e 2,
6-Di
nitro
tolu
ene
Acen
apht
hylen
e 3-
Nitro
anili
ne
Acen
apht
hene
4-
Nitro
phen
ol
2,4-
Dini
troto
luen
e Di
benz
ofur
an
Dieth
ylpht
halat
e Di
phen
ylam
ine
Azob
enze
ne
4-Br
omop
heny
l phe
nyl e
ther
He
xach
loro
benz
ene
Phen
anth
rene
An
thra
cene
Ca
rbaz
ole
Di-n
-but
ylpht
halat
e Fl
uora
nthe
ne
Pyre
ne
Bis(
2-eth
ylhex
yl)ad
ipate
Be
nz(a
)ant
hrac
ene
Benz
o(b)
fluor
anth
ene
Benz
o(k)
fluor
anth
ene
Benz
o(a)
pyre
ne
Dibe
nz(a
,h)a
nthr
acen
e Be
nzo(
g,h,
i)per
ylene
TSQ Duo GC-MS/MS
ISQ GC-MS
LOD,
pg/
µL
Tech
nica
l No
te 1
04
31
TN10431-EN 0115S
Spectral quality results: Figure 4 shows the results for the search-match for chromatographically resolved components of the mixture with the NIST 14 mass spectral library. The TSQ Duo GC-MS/MS and the ISQ GC-MS have a match factor greater than 900 for 59 and 64 compounds, respectively. Both the TSQ Duo instrument and the ISQ system show 66 number one hits. These results show that the spectral quality for the TSQ Duo and the ISQ systems are similar, with both having excellent matching factors with the NIST 14 Mass Spectral Library.
ConclusionFull scan comparison of the TSQ Duo GC-MS/MS and the ISQ GC-MS systems for 1 µg/mL Restek 8270 MegaMix shows that LODs for both instruments are similar, with 39 compounds having lower LODs on the TSQ Duo system. Additionally, the % RSD on the areas of the quantitative ions for most of the analytes was less than 5%. The spectral quality of the analytes, as measured by matching with NIST 14 Mass Spectral Library, was excellent, with 66 compounds having number one hits on both the TSQ Duo GC-MS/MS system and the ISQ GC-MS system.
Figure 4. Spectral quality comparison using the NIST 14 library for the TSQ Duo GC-MS/MS and the ISQ GC-MS.
References1. Mass Spectrometry: Limitless Innovation in Analytical
Science; SDi; Los Angeles, CA; October 2013.
2. Hajšlová, J.; Zrostlíková, J. Matrix effects in (ultra)trace analysis of pesticide residue in food and biotic matrices. J. Chromatogr. A 2003, 181–197.
AcknowledgementsThe authors would like to acknowledge helpful discussions with David Steiniger and Jason Cole.
600
650
700
750
800
850
900
950
1000
Phen
ol
Bis(
2-ch
loro
ethyl)
ether
1,
3-Di
chlo
robe
nzen
e 1,
4-Di
chlo
robe
nzen
e Be
nzyl
alcoh
ol
1,2-
Dich
loro
benz
ene
2-M
ethyl
phen
ol
2,2'
-oxy
bis(
1-ch
loro
prop
ane)
He
xach
loro
ethan
e Ni
trobe
nzen
e Iso
phor
one
Bis(
2-ch
loro
ethox
y)m
ethan
e 1,
2,4-
Trich
loro
benz
ene
Naph
thale
ne
4-Ch
loro
anili
ne
Hexa
chlo
robu
tadien
e 4-
Chlo
ro-3
-meth
ylphe
nol
2-M
ethyln
apht
halen
e 1-
Meth
ylnap
htha
lene
2-Ch
loro
naph
thale
ne
Dim
ethylp
htha
late
2,6-
Dini
troto
luen
e Ac
enap
hthy
lene
3-Ni
troan
iline
Ac
enap
hthe
ne
4-Ni
troph
enol
2,
4-Di
nitro
tolu
ene
Dibe
nzof
uran
Di
ethylp
htha
late
Diph
enyla
min
e Az
oben
zene
4-
Brom
ophe
nyl p
heny
l eth
er
Hexa
chlo
robe
nzen
e Ph
enan
thre
ne
Anth
race
ne
Carb
azol
e Di
-n-b
utylp
htha
late
Fluo
rant
hene
Py
rene
Bi
s(2-
ethylh
exyl)
adip
ate
Benz
(a)a
nthr
acen
e Be
nzo(
b)flu
oran
then
e Be
nzo(
k)flu
oran
then
e Be
nzo(
a)py
rene
Di
benz
(a,h
)ant
hrac
ene
Benz
o(g,
h,i)p
eryle
ne
NIST
14
Libr
ary M
atch
TSQ Duo GC-MS/MS
ISQ GC-MS
Africa +43 1 333 50 34 0Australia +61 3 9757 4300Austria +43 810 282 206Belgium +32 53 73 42 41Canada +1 800 530 8447China 800 810 5118 (free call domestic)
400 650 5118
Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0Finland +358 9 3291 0200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591
Japan +81 45 453 9100Korea +82 2 3420 8600Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Norway +46 8 556 468 00
Russia/CIS +43 1 333 50 34 0Singapore +65 6289 1190Spain +34 914 845 965Sweden +46 8 556 468 00Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752
www.thermoscientific.com©2015 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. MegaMix and Sky are registered trademarks of Restek Corporation. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
Thermo Fisher Scientific, Austin, TX USA is ISO 9001:2008 Certified.
Recommended