ICSAD: Imperial College Surgical Assessment Device Objective assessment of surgical dexterity Dept...

Preview:

Citation preview

ICSAD: Imperial College Surgical Assessment Device

Objective assessment of surgical dexterity

Dept of Surgical Oncology and Technology, Imperial College of Science Technology and Medicine, London, UK

ICSAD

Department of Surgical Oncology and Technology

Imperial College, London

• Commercially available motion tracking system (Isotrak II, Polhemus, USA)

• Signal generator (A) and two sensors (B) on the surgeon’s hands

A

B

Bespoke software:

‘Track 3D’- retrieves Cartesian positional data-x,y and z co-ordinates

‘Track’- positional data to dexterity measures Path length- economy of movement

Number of movements

Time taken

Speed

Department of Surgical Oncology and Technology

Imperial College, London

ICSAD

Department of Surgical Oncology and Technology

Imperial College, London

Validation of ICSAD-Lap suturing

8973N =

Groups

4321

Pa

th le

ng

th (

cm)

20000

10000

0

-10000

P= 0.009

1-experts

2- skilled

3- lap skills, knowledge

4- no lap skills, knowledge

Department of Surgical Oncology and Technology

Imperial College, London

Validation of ICSAD

N =

EXPERIENCE CODE

EDCBA

MO

VE

ME

NT

S B

OW

EL

3000

2000

1000

0 11111212N =

EXPERIENCE CODE

CONSULTANTSEN SPRJUN SPRSHO

NO

OF

MO

VE

ME

NT

S

3000

2000

1000

0

124

17

34

13

25

Department of Surgical Oncology and Technology

Imperial College, London

Correlation of motion analysis with qualitative assessment- OSATS

Department of Surgical Oncology and Technology

Imperial College, London

For small bowel anastomosis

N =

EXPERIENCE CODE

EDCBA

MO

VE

ME

NT

S B

OW

EL

3000

2000

1000

0 11111212N =

EXPERIENCE CODE

CONSULTANTSEN SPRJUN SPRSHO

NO

OF

MO

VE

ME

NT

S

3000

2000

1000

0

124

17

34

13

25

11111212N =

EXPERIENCE CODE

CONSULTANTSEN SPRJUN SPRSHO

GLO

BA

L S

CO

RE

40

30

20

10

18

38

20

19

Spearman correlation= -0.71, p<0.001

Department of Surgical Oncology and Technology

Imperial College, London

Correlation of motion analysis with qualitative assessment- Laparoscopic suturing

8973N =

Groups

4321

Pa

th le

ng

th (

cm)

20000

10000

0

-10000

8973N =

4.003.002.001.00

25

po

int

che

ck-l

ist

30

20

10

0

Spearman’s co-efficient: - 0.70, p<0.001

Department of Surgical Oncology and Technology

Imperial College, London

ICSAD- Training

Motion data of experts- standard for trainees

Provide objective feedback

Department of Surgical Oncology and Technology

Imperial College, London

Department of Surgical Oncology and Technology

Imperial College, London

Department of Surgical Oncology and Technology

Imperial College, London

ROVIMAS (Robotic Video Motion Analysis Software):

Department of Surgical Oncology and Technology

Imperial College, London

ACKNOWLEDGEMENTS

N Taffinder, A Wan, S Smith, J Torkington, T Brown, V Datta, S Bann, J Shah, S Mackay, Y Munz, A Dosis, A Chang, C Sutton, Prof Guang Zhong Yang, Dr. D. Gillies, Dr. F. Bello, Professor A Darzi

Recommended