Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com
The NIRvana: 640* from Princeton Instruments is the world’s first scientific grade, deep-cooled, large format InGaAs camera for low-light scientific SWIR imaging and spectroscopy applications. The camera uses a 640 x 512 InGaAs array with response from 0.9 µm to 1.7 µm. The detector is Peltier cooled to -85°C to minimize thermally generated noise and to improve signal-to-noise ratio for the most demanding SWIR applications. It offers 16-bit digitization and low read noise for outstanding dynamic range.
FEATURES BENEFITS640 x 512 InGaAs array High resolution imaging in the SWIR region;
Offers 4X more pixels over 320 x 256 sensor20 mm x 20 mm pixels High spatial resolutionResponse from 0.9 mm to 1.7 mm with 85% (typical) peak quantum efficiency
Excellent SWIR sensitivity for demanding imaging applications
Thermo-electric (Peltier) cooling Minimizes dark noise and allows extended integration times without the need for liquid nitrogen; Built-in air or liquid cooling option for vibration sensitive environments
Integrated cold-shield Special “cold shield” limits the ambient thermal background Electronic shutter Provides integration times from 2 ms to many minutesC-mount Standard lens interface compatible with numerous lenses and microscopes
(spectrometer adapters available)Fused silica window AR coated, fused silica window for high transmission in SWIR rangeGigE interface Industry standard for fast data transfer over long distances, up to 50 mOptional: LightField® (for Windows 8/7, 64-bit) Or WinView/Spec (for Windows 8/7/XP, 32-bit)
Flexible software packages for data acquisition, display and analysis; LightField offers intuitive, cutting edge user interface, IntelliCal® and more
SITK® for LabVIEW™ Easy integration into complex experimental setupPICAM (64-bit)/PVCAM (32-bit) software development kits (SDKs)
Compatible with Windows 8/7/XP, and Linux;Universal programming interfaces for easy custom programming.
NIRvana: 640
Applications: Nanotube fluorescence, emission, absorption, non-destructive testing and singlet oxygen detection
* Formerly known as the PIoNIR: 640
NOTE: Export of this camera outside of the United States is prohibited by law unless accompanied by a valid Export License as issued by the United States Department of Commerce.
NIRvana: 640 shown with lens, sold separately.
Page 2 of 8NIRvana: 640 Rev. P1
Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com
SPECIFICATIONS
NIRvana: 640
CCD image sensor 2D InGaAs focal plane array
CCD format 640 x 512 imaging pixels
CCD image area 12.8 mm x 10.24 mm
Pixel size 20 µm x 20 µm
Pixel well capacity (e-)
High Gain ≥ 40,000
High Capacity≥ 600,000
Nominal gain (e-/count) 1 14
Typical system read noise (e-) < 120* (High Gain)
Response nonlinearity: High gain < 2** %
Cooling temp. @ +20°C ambient -80°C (guaranteed), -85°C (typical with Air +15°C liquid circulation)
Cooling method Air only, liquid only, or a combination of air and liquid
Dark current (e-/p/sec) @ -80°C ~300***
Blemish specification Grade A: <2% defects For detailed blemish specifications, contact Princeton Instruments
Digitization 16 bits
Scan rate 2 MHz, 5 MHz, 10 MHz
Frame rate 22 fps @ 2 MHz55 fps @ 5 MHz110 fps @ 10 MHz
Binning and ROI Software only
Cold shield f#/1.5
Exposure time < 2 ms to >1 min
Window material Fused Silica (AR coated)
Thermostating precision ±0.05°C across entire temperature range
Operating temperature 0°C to +30°C
Certification CE
Weight / Dimensions (C-mount) 9.5 lbs (4.3 kg) / 7.40” x 5.8” x 5.8” (187.96 mm x 147.32 mm x 147.32 mm) (L x W x H)
All specifications are typical. Subject to change without notice.* Measured with 1 µs exposure time** Linearity for exposure > 20 ms *** Measured with a cold target at -174°C (+99°K)
Page 3 of 8NIRvana: 640 Rev. P1
Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com
SENSOR QUANTUM EFFICIENCY CURVE
0
10
20
30
40
50
60
70
80
90
100
0.80 0.90 1.00 1.10 1.20 1.30 1.40 1.50 1.60 1.70 1.80
Qua
ntum
Effi
cien
cy (%
)
Wavelength (µm)
NOTE: GraphshowstypicalQuantumEfficiency(QE)datameasuredat+25°C.Longercut-offwavelengthshiftstowardsblueby8nmperevery10°Cofcooling.
Page 4 of 8NIRvana: 640 Rev. P1
Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com
DARK CHARGE VS. FPA TEMPERATURE
10
100
1,000
10,000
100,000
1,000,000
10,000,000
- 100 - 80 - 60 - 40 - 20 0.0 20 40
Dar
k C
harg
e (e
-/p/
s)
FPA Temperature (°C)
LN Temperature Target
Room Temperature Target
Page5of8NIRvana: 640 Rev. P1
Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com
MAX. EXPOSURE TIME VS. FPA TEMPERATURE
0.01
0.1
1
10
100
-100 -80 -60 -40 -20 0 20 40
Expo
sure
Tim
e (s
ec)
FPA Temperature (°C)
Page 6 of 8NIRvana: 640 Rev. P1
Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com
APPLICATION IMAGES
Photoluminescence of a textured solar cell
Single walled nanotube (SWNT)
Topographical rendering of energy-momentum-resolved spectroscopy
Image courtesy of Dr. Steven Johnson, National Renewable Energy Laboratory, Golden, CO USA
Images courtesy of Dr. R. Bruce Weisman, Professor of Chemistry, The Richard E. Smalley Institute for Nanoscale Science and Technology, Rice University, Houston, TX USA
Image courtesy of Dr. Rashid Zia, Manning Assistant Professor, School of Engineering, Brown University, Providence, RI USA
Page 7 of 8NIRvana: 640 Rev. P1
Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com
OUTLINE DRAWING WITH C MOUNT
Spectrometer adapter is available upon request.
0
0
0.75019.05
2.35059.69
2.90073.66
0.75019.05
2.35059.69
2.90073.66
2.
350
59.6
9
2.
900
73.6
6
3.
906
99.2
0
2.
350
59.6
9
2.
900
73.6
6
00.
351
8.91
WIN
DO
W F
RO
NT
SU
RFA
CE
0.47
612
.08
WIN
DO
W IN
VA
CU
UM
SU
RFA
CE
0.69
FP
A L
OC
ATI
ON
1.
150
29.2
1
3.13
79.5
3
7.40
187.
84
7.62
193.
66
0
0
0.62
515
.88
0.62
015
.74
1.72
543
.82
1.75
544
.58
0.50512.83
1.25531.88
1.79645.62
0.3168.03
NOTES:
WEIGHT: 9.5 LBS / 4.3 KGS2.
WINDOW THICKNESS = 0.125" [3.18MM]3.
MECHANICAL SHUTTER NOT AVAILABLE4.
MAX. POWER CONSUMPTION = 135W5.
C-MOUNT THREAD1.00-20 UN
4.063 20 UN THREAD
LIQUID COOLANTINLET/OUTLET
ALL DIMENSIONS IN INCHES [MM]1.
Page 8 of 8NIRvana: 640 Rev. P1
Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com Princeton Instruments | 3660 Quakerbridge Rd | Trenton, NJ 08619 USA | +1 609-631-4000 | www.princetoninstruments.com | info@princetoninstruments.com
OUTLINE DRAWING WITH SPECTROMETER MOUNT
35°
17.5° 3.600
91.45 B.C.
0
0
0.75019.05
2.90073.66
0.75019.05
2.90073.66
2.
900
73.6
6
3.
906
99.2
0
2.
900
73.6
6 0
0
0.25
26.
41FR
ON
T S
UR
FAC
EO
F V
AC
UU
M W
IND
OW
0.59
014
.99
FPA
DIS
TAN
CE
1.
050
26.6
7
3.
032
77.0
2
7.
297
185.
33
7.
526
191.
16
0.0
63
1.12528.58
1.12528.58
0
0
0.50512.83
1.25531.88
1.79645.62
0.3168.03
0.
625
15.8
8
0.
620
15.7
4
1.
725
43.8
2
1.
755
44.5
8
NOTES:
WEIGHT: 9.5 LBS / 4.3 KGS2.
WINDOW THICKNESS = 0.125" [3.18MM]3.
MECHANICAL SHUTTER NOT AVAILABLE4.
MAX. POWER CONSUMPTION = 135W5.
4.200106.68
COOLANT INLET/OUTLET
2X 10-32 UNF TYP. ALL FOUR SIDES
ALL DIMENSIONS IN INCHES (MM)1.
C-mount adapter is available upon request.