P2131 

Automated Field Steam Sterilizer

John Hodges   |   jhodges@fortdefianceind.com |   865‐437‐6181

The P2131 System

Sterilizer Main Unit (SMU)

Water Recovery System (WRS)

Sterilizer Water Softener (SWS)

AccessoriesBag

Rugged

Lightweight and rugged all‐welded, high‐strength aluminum alloy construction 

AutomatedPre‐programmed FDA‐cleared cycles include:

Immediate Use Textiles Wrapped Instruments (4‐ & 10‐minute cycles)Bowie Dick & Vacuum Leak Tests

Effective

The only FDA‐Clearedpre‐vacuum steam sterilizer designed for field use

Safe

Chamber is designed, fabricated and tested to ASME Section VIII Division I 

standards

Reliable

Rigorously tested to ensure consistent 

performance in a variety of austere environments

Conserves Water

Capable of running 100 cycles with only 10 gallons of water, reducing overall water usage

Compact

Endcaps enclose the main unit and create a self‐contained shipping crate

Once in position, the endcaps fold down and lock into place to create a stand for the sterilizer

Efficient

The P2131 is capable of sterilizing up to 36 lbs. of wrapped instruments per load 

Transportable

Designed to be carried by 4 people, the P2131 is lightweight and easy to transport

Unit WeightSterilizer Main Unit 304 lbs (138 kg)Water Recovery System 151 lbs (68.5 kg)Sterilizer Water Softener 35 lbs (16 kg)Accessories Bag 17 lbs (7.7 kg) 

Modernized

The WRS not only recovers water from each cycle, but also provides the mechanical means to create the pre/post cycle vacuum

Maintainable

The SWS produces deionized water to remove scale build‐up on the sterilizer heating elements and reduce corrosion, reducing maintenance and improving performance

PreparedThe SWS is equipped with a manual hand pump in case a pressurized potable water source is not available

Markets Served

Military Field Hospitals

Disaster Relief Operations

Remote Area Medical Clinics

Robust

The P2131 sterilizer offers consistent and dependable operation whether at

40°F at 8,000 feet altitude (4°C at 2,438 m) or130°F (54°C ) at sea level

FDA‐Cleared

Meets applicable requirements of electrical safety standards UL‐61010‐1 and IEC‐61010‐2‐040

Meets applicable requirements of ANSI/AAMI ST8:2013

Approved

Latest in Technology

Offers the best in steam sterilization technology with a Pre/Post vacuum dynamic air removal process

Sterilizer Main Unit (SMU):• Electrical ratings 208‐230 VAC, 3 phase, 30 amps, 50/60 Hz• Sterilizing temperature 270°F (132°C)• Pre‐vacuum phase  3 pressure/vacuum pulses• Maximum working pressure/vacuum 40 psig/full vacuum• Electrical receptacle MIL‐C‐22992 cord connecting class L• Length (endcaps closed for transport) 46.00” (1,168 mm)• Length (endcaps open for operation) 38.00” (965 mm)• Width 26.75” (679 mm)• Height (endcaps closed for transport) 31.00” (787 mm)• Height (endcaps open for operation) 62.00” (1,575 mm)• Chamber size 16” (406 mm) diam x 36” (914 mm) long

4.2 ft3 (119 liters)• SMU weight (dry) 304 lbs. (138 kg)• SAB weight 17 lbs. (7.7 kg)• Jacket (boiler) water capacity 4.25 gallons (16 liters)• NEMA Classification 2 – dry location• Operating altitude 0 – 8,000 feet (0 – 2,438 m)• Operating ambient temperature 40°F – 130°F (4°C – 54°C)

Specifications and Physical Data

Specifications and Physical Data

Water Recovery System (WRS):• Length x Width x Height 30” x 23.75” x 24.5” (762 x 603 x 622 mm)• Weight (dry) 151 lbs. (68.5 kg)• Water capacity 6 gallons (22.7 liters)• Vacuum eductor capacity 25” Hg• Electrical ratings 115 VAC, 1 phase (fed from SMU)• Sound level 78 dBA• NEMA Classification 2 – dry location

Sterilizer Water Softener (SWS):• Length x Width x Height 15” x 15” x 38.75” (381 x 381 x 984 mm)• Weight (dry) 35 lbs. (16 kg)• Electrical Power N/A

Cycle Recommended UseMaximum 

Load

Exposure Temperature1

(°F/°C)

Exposure Time 

(minutes)Dry Time (minutes)

Immediate Use(IUSS)

Unwrapped nonporous(e.g., instruments) and porous items in mixed 

loads

36 lbs.(16.3 kg)

270/132 4 0

Textiles Textile packs3 textile packs2

270/132 4 53

Wrapped Instruments

Wrapped instruments/utensils

36 lbs.(16.3 kg)

270/132 4 203

Wrapped Instruments

Wrapped instruments/utensils

36 lbs.(16.3 kg)

270/132 104 203

Bowie‐Dick Test Test NA 273/134 3.5 0

Vacuum Leak Test Test NA NA Test time 20 0

1 The AAMI minimum temperature is 270°F (132°C) for all cycles except Bowie‐Dick which is 273°F (134°C).  The P2131 sterilizer setpoint temperature is 273°F (134°C) for all cycles except Bowie‐Dick which is 274°F.2  AAMI standard 16‐towel pack (9”x9”x6”).3  Dry times are pre‐set to established standard conditions, but can be manually increased.4  The FDA‐cleared Process Challenge Device (PCD) for this 10‐minute cycle can be purchased from FDI at (865) 408‐0100 or via email at sales@fortdefianceind.com.  Reference Appendix A.3 for additional information on this PCD.

Sterilization Cycles – FDA Cleared

Questions?

John Hodgesjhodges@fortdefianceind.com

865‐437‐6181

www.fortdefianceind.com