PURELAB®
Laboratory water purification systems
Page 3 of 46
Page72
We understand how important it is for scientists to obtain a choice of water qualities that range from primary grade, for simple routine washing and rinsing, through to ultrapure for the most critical science and analytical applications.
With this in mind, we applied our expertise gained from 50 years of innovation in water purification technologies, to design the unrivalled PURELAB range. Our reliable water purification systems are constructed from the highest quality components to ensure optimal purity, while a rapid and easy sanitization program contributes towards an uninterrupted workflow. Built-in economical processes result in the lowest consumable costs with the highest water quality and precision.
This brochure outlines the key features of all the products in our PURELAB range, which are designed to provide water purification systems covering the three different water types (Type I-III) required for research and testing applications.
PURELAB
“ Water is all we do. We know water.”
Your first step to pure water: the ELGA LabWater water analysis kitAn informed start for a streamlined solution
At ELGA we do not speculate or work on assumptions about your water quality. On our first visit to your laboratory we will carry out a test, on site, that analyzes your feed water quality.
Armed with data about your laboratory’s water quality, required applications, lab design and budget, our sales team will deliver an informed proposal about the best water purification solutions to suit your needs.
Page 4 of 46
Page73
1-2PU
RELAB
PURELAB product finder
Matching systems to applications supporting you all the wayWater purification is often a complex process that incorporates multiple technologies into a single purification system; therefore, we have developed the product finder to help you select the best water purification system suited to your application needs. Our team of technical and application specialists is always available to answer any questions, as well as providing friendly, expert advice in choosing the most suitable water purification system.
* For more information on applications and scientific water standards see our education guidebook “The Pure LabWater Guide”. A free copy is available at www.elgalabwater.com
1000
500
100
50
10
1
Opt
ion-
Q (P
age
10)
Clas
sic
(Pag
e 9)
flex
3&
4 (P
age
7 &
8)
flex
3&
4 (P
age
7 &
8)
flex
3&
4 (P
age
7 &
8)
Puls
e (P
age
11)
flex
1&
2 (P
age
7& 8
)
flex
1&
2 (P
age
7 &
8)
Prim
a (P
age
14)
Ult
ra (P
age
5&6)
Opt
ion-
S (P
age
14)
flex
1&
2 (P
age
7 &
8)
Opt
ion-
R (P
age
12)70
00 (P
age
13)
3000
(Pag
e 13
)
Water rate liters/day
Water quality Type I+ Type I Type II+ Type II Type III
Typical applications
ICPMS (Inductively Coupled Plasma Mass Spectrometry)
Molecular biology techniques
Ultra trace analysis
Electrochemistry
Electrophoresis
GFAAS (Graphite Furnace Atomic Absorption Spectrophotometry)
HPLC
IC (Ion chromatography)
ICPAES (Inductively Coupled Plasma Atomic Emission Spectrometry)
ICPMS (Inductively Coupled Plasma Mass Spectrometry)
Mammalian and bacterial cell culture
Molecular biology
Plant tissue culture
Qualitative analyses
AAS (Atomic Absorption Spectrophotometry)
Buffer and media preparation
Electrophysiology
FAAS (Flame Atomic Absorption Spectrophotometry)
Feed to ultrapure water systems
Glassware washing/rinsing
General chemistry
Histology
Microbiological analysis
RIA (Radioimmunoassay) / ELISA (Enzyme Linked Immunoabsorbant Assay)
Sample dilution and reagent preparation
Spectrophotometry
Water analysis
Autoclave feed
Feed to ultrapure water systems
Hydroponics
Plant growth cabinets
Stability chambers
Steam generators
Sterilizer feed
Page 5 of 46
Page74
Every PURELAB comes complete with one extra feature – first class service and support wherever you are in the world. ELGA has installed thousands of systems globally, and our service engineers will apply their expertise to the installation, validation and maintenance of your water purification systems in compliance with all the relevant codes. Visit www.elgalabwater.com to find your nearest contact.
Service excellence
• We aim to provide first-time-fix service
• Preventative maintenance contracts are structured to match your precise needs, thereby minimizing interruptions to your workflow and sustaining system reliability
• All our calibration equipment is maintained to traceable standards and operated by fully trained service technicians, ensuring your water quality is consistently maintained at the required standard
Training
• “Hands on” operation training is arranged on or off site to ensure optimal system performance and minimize the risk of an interrupted workflow
For total confidence in your pure water systems
Worldwide service and support
Technical assistance
• Our specialized local team will ensure you find the perfect product to match both your budget and applications
• Our dedicated local help desk is always available to provide advice, troubleshooting and parts identification
Online facility www.elgalabwater.com
• Locate your local service representative
• Learn more about water quality and standards by downloading a FREE Pure Labwater Guide
• Find a water purification system which meets your needs with our online product selector
Page 6 of 46
Page75
Validation support
Water systems that are employed within a validated laboratory should always operate within specification. This should be evidenced through documentation and trending should warn the user if performance is likely to fall below their requirements. ELGA offers a market-leading validation package with fully trained personnel using controlled equipment and documentation to support you throughout the validation process.
ELGA products are continually being developed to further support validation requirements. Developments include:
Quality assuredOur commitment to the highest quality control processes guarantees reliability and compliance with both international, environmental and user organization standardsDesigned and manufactured under an ISO9001:2000 total quality system. Tested to comply with CE, EMC, EN 61010 (UL CSA), PIRA, WEEE Directive and other standards as appropriate.
3-4Service and Support
• PIN-coded access to software set points prevents unauthorized changes to operation or system settings
• Cartridge Identification (CID) technology provides full traceability of each cartridge for GLP and other validation requirements
• RS232 data port to enable downloading of data to PC or printer
• Water quality data available with or without temperature compensation
Page 7 of 46
Page76
Designed to meet and exceed ultrapure (Type I) water quality
When your application requires the ultimate in water purity, PURELAB Ultra provides the perfect solution. Consistently delivering water purity to 18.2 MΩ-cm and underpinned by the unique PureSure system, the PURELAB Ultra enables you to focus on attaining accurate results as well as ensuring an uninterrupted work flow.
Major molecular biology applications
PURELAB removes the impurity of concern
Gel preparation Nucleases and bacteria
Hybridization mixes and PCR
Nucleases, bacteria and salt concentrations of Mg2+ and Na+
Sequencing Nucleases, bacteria, endotoxin, salt concentrations of Mg2+ and Na+, humic acid
DNA to cell transfer Endotoxins
The PURELAB solution for overcoming your water purification challenges
Challenge PURELAB Solution
Your application demands the ultimate in water purity, i.e. simply having a resistivity of 18.2 MΩ-cm is not good enough
PureSure system provides the ultimate inorganic purity (guaranteeing the retention of silcon and boron species)
Organic breakthrough must be prevented
Constant real-time Total Organic Carbon (TOC) monitoring gives you peace of mind by letting you know what is happening with organic purity as you dispense your water sample
You cannot afford to stop work to change a purification pack
PureSure system ensures that work is not interrupted
The highest level of microbial purity is required throughout the entire life of the water purification system
Ongoing control to minimize potential bacterial growth is guaranteed through simple and complete sanitization of all wetted parts
Delivering the ultimate in water purity for absolute confidence in your results
PURELAB
Why the demand for ultimate water purity?
For the most sophisticated research and testing applications, e.g. ICPMS (Inductively Coupled Plasma Mass Spectrometry), ultra trace analysis and molecular techniques, it is critical that the ultimate water purity is employed to prevent interference and, therefore, inaccurate results from contaminating impurities (see examples in the table below).
Ultra
Page 8 of 46
Page77
5-6
PureSure – your water purity guarantee
The PureSure system, exclusive to the PURELAB Ultra, consists of two purification cartridges and resistivity monitor units in series. This arrangement ensures purity at 18.2 MΩ-cm and gives advanced warning of purification cartridge exhaustion without compromising final water purity. The PureSure system allows each cartridge to be used to its full potential, resulting in savings and capacity gains.
Ultra
4
Unique PureSure system
• Ensures purity beyond just 18.2 MΩ-cm
• Extends the life of your purification cartridges
• Advance warning of cartridge change so you are free to continue your research
4
Automatic self calibration of electronic circuitry – takes just 4 seconds every time the system is in process mode
3
Unique volumetric profile dispense facility for easy recording, replication and speedy dispensing of multiple set volumes
2
Efficient fast rinse ultrafiltration cartridge
• Filters out endotoxin, proteins and nucleases
• Rinses up to 7 times faster than other endotoxin filters
• Significantly reduces costs of sanitization
1 Full spectrum UV treatment
• Low microbial and organic specification
• High transmittance synthetic quartz sleeve for higher efficiency
• Lamp failure alarm
10
New Labpure purification cartridges
• High capacity cartridges
• Longer life in system
9
Cartridge Identification technology
• Provides validatable traceability
• Your guarantee of capacity, quality and safety
8
PIN-coded protection of critical operating parameters
7
Sensitive electronic dispense offers accurate flow control from single droplets up to 2 liters/min
6Real-time TOC monitoring
• System reports on TOC every 2 seconds
• Ensures confidence in organic quality
5
10
9
9
8
7
6
5
1
2
3
Page 9 of 46
Page78
flexPURELAB
The PURELAB flex is designed to deliver accuracy, flexibility and ease of use. This award winning system provides perfect water purity for analytical and lifescience applications which require RO Type III water up to ultrapure type I (18.2 MΩ-cm) water. It allows focus on routine test work without concern about the water quality affecting test results.
The most intuitive and innovative system for all your pure water needs
Designed for today’s laboratories
Customized settings to suit your application
1
185nm/254nm UV lamp for low TOC water production needed for organic sensitive application
2
Easy-access for routine maintenance
3
Designed for simple easy installation
5
Purification pack quick and easy to replace
6
Multiple dispense positioning, wall, bench, height adjustable arm, hand-held dispensing
4
1
24
6
7
5
3
Data capture via USB for system performance validation
7
Page 10 of 46
Page79
7-8
Find a PURELAB flex to match your application
PURELAB flex 2
Daily volume >10 Liters <10 Liters >10 Liters <10 Liters >10 Liters
Water quality Type I ultrapure water Type II/Type II+ pure water Type III RO water
Feedwater source Pre-purified water
Pre-purified or tap water
Pre-purified water
Pre-purified or tap water
Pre-purified water
Typical applications
Electrochemistry
Electrophoresis
GFAAS (Graphite Furnace Atomic Absorption Spectrophotometry)
HPLC
IC (Ion Chromatography)
ICPAES (Inductively Coupled Plasma Atomic Emission Spectrometry)
ICPMS (Inductively Coupled Plasma Mass Spectrometry)
Mammalian and bacterial cell culture
Molecular biology
Plant tissue culture
Qualitative analyses
AAS (Atomic Absorption Spectrophotometry)
Buffer and media preparation
Electrophysiology
FAAS (Flame Atomic Absorption Spectrophotometry)
Feed to ultrapure water systems
Glassware washing/rinsing
General chemistry
Histology
Microbiological analysis
RIA (Radioimmunoassay) / ELISA (Enzyme Linked Immunoabsorbant Assay)
Sample dilution and reagent preparation
Spectrophotometry
Autoclave feed
Feed to ultrapure water systems
Hydroponics
Plant growth cabinets
Stability chambers
Steam generators
Sterilizer feed
PURELAB flex 1 PURELAB flex 1PURELAB flex 3&4 PURELAB flex 3&4
1 Intuitive to use, ergonomic handset design
Clear water purity display for absolute confidence as you dispense
2
Realtime TOC monitoring for critical analytical applications
3
Prioritized information shown at all times (system status, TOC, alarm)
4
Flexible dispensing in four different ways:• Variable flow – drop by drop or up to 2 liters per minute
• Autovolume dispense from 50ml to 60 liters and repeat dispensing
• Hands free with optional foot pedal• Locked dispense for glassware filling
5
1
2
34
5
flex
Page 11 of 46
Page80
PURELAB ClassicEconomy without compromise for high performance, ultrapure water production
The high usage option for pre-treated water feed
An ideal choice for the lab requiring more than 20 liters/day of Type I ultrapure water. Adjustable dispensing provides a flow rate of up to 2.0 liters/min and continuous water quality monitoring ensures ultrapure water with low bacterial specification. This range also provides the option to upgrade from single pack to twin pack purification.
Automatic intermittent recirculation to maintain high water purity with a low bacterial specification
1
1
Efficient fast rinse ultrafiltration cartridge• Filters out endotoxins, proteins
and nucleases• Rinses up to 7 times faster than
other endotoxin filters• Significantly reduces costs of
sanitization
2
2
Full spectrum UV treatment• Low microbial and organic specification• High transmittance synthetic quartz
sleeve for higher efficiency• Lamp failure alarm
3
3
PIN-coded access to software set points to prevent unauthorized changes to operation or system settings
4
4
Single purification pack design, easily upgraded to dual pack
5
5
Page 12 of 46
Page81
Tap to Type I — the plug-and-go option for tap water feed
Providing ultrapure water direct from a drinking water supply without the need for a pre-treated feed, the PURELAB Option-Q can deliver 1.0 liter/min, for labs requiring up to 100 liters/day.
PURELAB Option-Q
9-10
Ultrapure water direct from potable tap water
“Ultrapure water systems to match your budget”
Classic and Option-Q
Recirculation of purified water through our wrap-around reservoir – to maintain consistent peak water purity
1
1
Fully recirculating UV ensures the highest microbial quality
22
Height adjustable point-of-use dispense tap for a choice of wall or bench mounting
3
3
Front-entry service doors for easy access to consumables
4
4
6
Optional ELGA Biofilter – when fitted to the Option-Q produces water which is free from biologically active impurities
Data collection capabilities through RS232 interface – for compliance with GLP guidelines
5
6
5
Page 13 of 46
Page82
The only Electrodeionization (EDI) choice for type II+/II pure water
PulsePURELAB
The PURELAB Pulse uses state-of-the-art patented ‘Pulse Technology’ to deliver Type II+ and III pure water at minimal running costs. This system is perfect for laboratories who need up to 160 liters of purified water for an 8 hour day. Within the system the unique intelligent recirculation maintains consistent water purity at point of use. The PURELAB Pulse also reduces maintenance time, with longer intervals between sanitization and quick and easy replacement of consumables.
Space-saving wrap-around reservoir with smooth sides and vented filtration for optimized water purity
1
1
Fully recirculating UV ensures the highest microbial quality
2 2
Height adjustable point-of-use dispense tap for choice of wall or bench mounting
3
3Full recirculation guarantees water quality at point-of-use
4
4
Pulse Module (EDI purification)
55
Pulse Technology The Pulse technology uses Electrodeionization (EDI) to purify the water instead of using deionization cartridges. EDI is an electrically-driven water purification process which involves the use of Ion Exchange Resins and Ion Permeable Membranes.
Resins are continuously regenerated by the electrical current and are never exhausted. The water quality is always maintained due the absence of the adverse effects given by exhaustion of a resin bed, such as silicon, boron or organic breakthrough.
Page 14 of 46
Page83
1
5
4
3
2
The PURELAB solution for overcoming your water purification challenges
Challenge PURELAB Solution
Your application requires the removal of most contaminants but highly purified or ultrapure water is not essential. Your application may require high volumes of water
Sophisticated integral recirculation technology ensures optimum organic and inorganic quality where you need it — clear quality displays allow you to monitor quality in real-time at point of dispense
Reliable, high purity water with good inorganic and microbial specifications
Full recirculation ensures a high level of microbial control and easy use, and maintenance keeps quality high throughout the life of the product.
Space-saving wrap-around reservoir with smooth sides and vented filtration for optimized water purity
1
Fully recirculating UV ensures the highest microbial quality
2
Deionization purification (ion-exchange) cartridge
3
Height adjustable point-of-use dispense tap for choice of wall or bench mounting
4
Full recirculation guarantees water quality at point-of-use
5
Designed to meet and exceed Type II+/II pure water quality
Option-RPURELAB
The PURELAB Option-R produces Type II+ grade water, better than double-distilled quality using ion-exchange resins. Integral full recirculation of purified water maintains constant peak water purity and photo-oxidation ensures low bacterial counts. The range offers a choice of models to give flexibility in flow rate settings: providing either 7, 15, 30, or 60 liters/hr.
Pulse
11-12Pulse and O
ption-R
Page 15 of 46
Page84
PURELAB 3000
The PURELAB 3000 supplies pure water at flow rates suitable for a high-usage laboratory. The system is easily upgraded to increase flow rates from 60 to 120 liters/hr, and the simple addition of a purification pack will take you from Type III/RO to Type II pure water, providing flexibility to adapt to changes in usage requirements.
Designed to meet Type II or Type III water quality
When general laboratory and primary grade water is needed at high flow rates the PURELAB 3000 & 7000 offer a reliable and consistent supply with the flexibility to meet your usage needs.
3000 & 7000High volume delivery of pure water for general laboratory use
PURELAB
“The compact solution for small
distribution loops”
The PURELAB 7000 dispenses up to 4 liters/min of Type II+ water (>10 MΩ-cm) and can be connected to small distribution loops to feed an ultra pure system and two or three dispense points, as well as your main application. The system has an in-built reservoir to ensure your required pure water capacity is met at peak times.
PURELAB 7000
1
Reliable water quality• Unique sanitization
design for purification of RO membranes as well as the local distribution loop
• Sanitization designed to run quickly and easily
• User-defined, electronically protected sanitization reminders
1
Economic design• Unique wrap-around
reservoir• Easy-access doors and
castors means system can be located under bench
2
3 Unique system operation protection• Easy-to-use Building Management System• Automated alarms
3
Economic Purity• Unique RO resin mix to
optimize economy, increases yield of purification packs
• Optional degassing system further increases capacity of purification packs
4
Purity• Easily switch between
ultrafiltration or sub-micron filtration for your application
5
5
44
3
2
Page 16 of 46
Page85
13-14
Designed to meet Type II or Type III water quality
Option-S and PrimaSimple and reliable delivery of purified water for the general laboratory
PURELAB
PURELAB Option-SThe flexible solution for general laboratory grade water
Using multiple purification technologies, the PURELAB Option-S produces Type II grade water, better than single-distilled quality. The range offers a choice of models to give flexibility in flow rate settings: providing either 7, 15, 30, or 60 liters/hr.
PURELAB PrimaThe ultimate in flexibility for primary grade water
The compact PURELAB Prima systems provide Type III water and are available in a choice of flow rates, 7, 15, 30, 60, 90, or 120 liters/hr, to meet your specific needs.
3000 and 7000, Option-S and Prim
a
• Quick and easy sanitization and replacement of consumables to reduce maintenance time
• Features ‘Reverse Osmosis feed optimized’ resin mixes to increase the ion exchange capacity of consumables and minimize running costs
• PURELAB Option systems are designed to be easy to access whether wall or bench mounted with a convenient dispense tap. They can be used with our wrap-around reservoirs to save space whilst optimizing purity
Page 17 of 46
Page86
The LabWater Specialists ELGA is an integral part of Veolia Water Solutions and Technologies, the world’s leading water service company. Veolia Water Solutions and Technologies has a worldwide team of over 9,500 people and is renowned for its capabilities in providing water solutions to customers throughout the entire water cycle.
The ELGA team focuses exclusively on water and its purification. It continually contributes to the unique technical and scientific applications expertise developed during the past 50 years. We are experienced in meeting the challenges that arise during the development, installation and servicing of single point-of-use water purification systems as well as large projects involving consultation with architects, consultants and clients.
Contact us:
ELGA offices and distributors are located in more than 60 countries and are fully trained in all ELGA systems.
To find your nearest ELGA representative, go to www.elgalabwater.com and select your country for contact details.
ELGA Global Operations Centre
Tel: +44 1494 887 500 Fax: +44 1494 887 505
Email: info@elgalabwater.com Web: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA®, PURELAB®, ADEPT® Labpure® and PureSure® are registered trademarks of VWS (UK) Ltd.
Your local ELGA representative
Commitment to global
citizenship
The Veolia Environment Foundation supports worldwide projects contributing to sustainable development, with a special focus on outreach, workforce development and the environment.
Since 2004, with an annual budget of €5 million, the Foundation has supported more than 450 projects.
(see www.fondation.veolia.com for details)
LITR
387
67-0
4
www.elgalabwater.com
Page 18 of 46
Page87
PURELAB AnAlyticAl ReseARch
• Simple to use and maintain with a clear display of water quality
• Auto rinse facility maintains purity of water after periods of low use
• Minimal consumables delivers a lower cost of ownership
• Provides for future changes in demand with easy upgradeability to higher flow rates
• Optional integral feedwater boost pump reduces costs
the PURelAB Prima is ideal for feeding ultra-pure water systems and for a wide range of laboratory applications including glassware washing, autoclave feed and for use in environmental cabinets.
PURelAB Prima 7/15/30
The cost-effective choice for glassware washing
Outlet Outlet
FeedwaterInlet
Boost Pump(optional)
Sanitization Port
Pre-Treatment
ReverseOsmosis
ReverseOsmosis
Outlet
ReverseOsmosis
Drain
Water Quality Sensor
Outlet toDocking Vessel
or Reservoir
Process Flow PURELAB Prima 7/15/30Process Flow PURELAB Prima 7/15/30
Prima 7-15-30 Datasheet 2010.indd 1 23/01/2012 11:17Page 19 of 46
Page88
PURELAB AnAlyticAl ReseARch
TreatedWaterSpecifications
Model Prima7 Prima15 Prima30Make up rate1 7.5 l/hr 15 l/hr 30 l/hrDaily output (nominal max)1 180 l/24 hour day 360 l/24 hour day 720 l/24 hour dayOutput reverse pressure (max) 0.1 bar (1 psi) 0.1 bar (1 psi) 0.1 bar (1 psi)Purity:Inorganic - minimum >90% rejection >90% rejection >90% rejectionInorganic - typical up to 98% rejection up to 98% rejection up to 98% rejectionOrganics (MW>100 Dalton) >99% rejection >99% rejection >99% rejectionTOC2 <50 ppb <50 ppb <50 ppbBacteria2 <5 CFU/ml <5 CFU/ml <5 CFU/mlParticles >99% rejection >99% rejection >99% rejection
1Standard conditions are 4 bar inlet pressure, <0.1 bar back pressure at 15 degrees centigrade, fed with potable water and a clean pre-treatment
cartridge. Refer to flow tables outside these conditions. 2Subject to correct operating and maintenance procedures.
DimensionsandweightsHeight 460mm (18.1in) 460mm (18.1in) 460mm (18.1in)Width 410mm (16.2in) 410mm (16.2in) 410mm (16.2in)Depth 270mm (10.6in) 270mm (10.6in) 270mm (10.6in)Weight with internal boost pump 13.5kg (30lb) 14.5kg (32lb) 15.0kg (33lb)Weight without internal boost pump
11.5kg (25lb) 12.5kg (28lb) 13.0kg (29lb)
FeedwaterRequirementsSource quality Potable mains water supply Potable mains water supply Potable mains water supplyFouling index - maximum 10 10 10Conductivity 2000 µS/cm* 2000 µS/cm* 2000 µS/cm*Free chlorine - maximum 0.5 ppm 0.5 ppm 0.5 ppmHeavy metals - maximum 0.05 ppm 0.05 ppm 0.05 ppmSilica - maximum 30 ppm 30 ppm 30 ppmTemperature 1 - 35°C 1 - 35°C 1 - 35°CFlowrate (maximum requirement) 145 l/hr 150 l/hr 155 l/hrDrain requirements (gravity fall with air gap). Maximum during service
70 l/hr 70 l/hr 80 l/hr
FeedwaterPressureMaximum - without internal boost pump 6.0 bar (90 psi) 6.0 bar (90 psi) 6.0 bar (90 psi)Minimum - without internal boost pump 4.0 bar (60 psi) 4.0 bar (60 psi) 4.0 bar (60 psi)Maximum - with internal boost pump 2.0 bar (30 psi) 2.0 bar (30 psi) 2.0 bar (30 psi)Minimum - with internal boost pump Flooded Suction Flooded Suction Flooded Suction
* Deionization cartridge life may vary with feedwaters >1400 µS/cm
ElectricalRequirementsMains input 100-240V ac, 50-60Hz all models 100-240V ac, 50-60Hz all models 100-240V ac, 50-60Hz all modelsSystem voltage 24V dc 24V dc 24V dcPower consumption with boost pump 52VA 52VA 52VAPower consumption without boost pump
28VA 28VA 28VA
Fuses 2 x T3.15 Amp 2 x T3.15 Amp 2 x T3.15 AmpReservoir level connection Jack Plug 3.5mm Jack Plug 3.5mm Jack Plug 3.5mmNoise level <45dBA <45dBA <45dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2012 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38751-05
Prima 7-15-30 Datasheet 2010.indd 2 23/01/2012 11:17Page 20 of 46
Page89
PURELAB AnAlyticAl ReseARch
• Simple to use – clear display of quality
• Auto rinse facility maintains purity of water after periods of low use
• Minimal purification consumables (reminders for these on the display) means low cost of ownership
• Provides for future changes in demand: upgrade easily from one flow rate to another
the PURelAB Prima is ideal for feeding type i ultrapure water systems and for laboratory applications including glassware washing, autoclave feed, and for use in environmental cabinets. sophisticated microprocessor monitoring helps ensure optimum performance at all times – and you can rely on easy, cost-effective and convenient operation, with user-friendly controls, informative status displays, and simple maintenance routines.
PURelAB Prima 60/90/120
The cost-effective choice for higher volumes
Drain
BoostPump
ReverseOsmosis
ReverseOsmosis
Water Quality Sensor
Water Quality Sensor
Outletto Reservoir
Process Flow PURELAB Prima 30/60/90/120
Break Tank
Feedwater Inlet Pre-Filter
Process Flow PURELAB Prima 60/90/120
Page 21 of 46
Page90
PURELAB AnAlyticAl ReseARch
TreatedWaterSpecifications
Model Prima60 Prima90 Prima120
Inorganics – min >90% rejection
Inorganics – typical Up to 98% rejection
Organics 1 >99% rejection
TOC – typical 2 <100 ppb
Bacteria 3 <5 CFU/ml
Particles >99% rejection
1 MW > 100 Dalton 2 With suitable pretreatment, if required, depending on the feedwater 3 Subject to correct operating and maintenance procedures
Dimensionsandweights
Height 740mm
Width 570mm
Depth 320mm
Weight (operational) 47kg 48kg 54kg
FeedwaterRequirements
Sourcequality Potablemainswatersupply
Fouling index – maximum 10
Conductivity – maximum 2000 µS/cm
Free chlorine – maximum <0.5 ppm
Heavy metals – maximum 0.05 ppm
Temperature 1 – 35°C
Flowrate (maximum requirement) 9 l/hr
Feedwater pressure maximum 6.0 bar (90 psi)
Feedwater pressure minimum 1.4 bar (20 psi)
ElectricalRequirements
Mains input 230V ac, 50Hz115V ac, 60Hz
Power consumption 700VA
Fuses 2 x 3.15 Amp – 230V ac 50Hz2 x 6.3 Amp – 115V ac 60Hz
Reservoir level connection Din plug 6 way
Noise level <52dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38752-04 Page 22 of 46
Page91
PURELAB AnAlyticAl ReseARch
Delivers market leading microbial performance. In addition to simple sanitization • and composite vent filtration, the PURELAB Option-R system is the only fully recirculating Type II+ pure water system on the market, ensuring high specification water
Features ‘Reverse Osmosis feed optimized’ resin mixes to increase the ion exchange • capacity of consumables and minimize running costs
Quick and easy sanitization and replacement of consumables to reduce • maintenance time
PURELAB Option systems are designed to be easy to access whether wall or bench • mounted with a convenient dispense tap. They can be used with our wrap-around reservoirs to save space whilst optimizing purity
the PURelAB Option provides type ii+/ii pure water on demand with a wide range of flow-rates to suit your needs. Applications range from glassware washing and the feeding of ultrapure water systems to cell culture and media preparation. the PURelAB Option-R provides additional bacteria and inorganic quality for your more critical applications.
PURelAB Option-s/R 7/15
The only fully deionization recirculating Type II pure water system
Outlet Outlet
FeedwaterInlet
Boost Pump(optional)
Sanitization Port
Pre-Treatment
ReverseOsmosis
ReverseOsmosis
Drain
Water Quality Sensor
Recirculation Pump
Outlet toDocking Vessel
or Reservoir
Pump feed fromDocking Vessel
or Reservoir
Dispense Tap
DeionizationCartridge
UV Lamp
WaterQualitySensor
Process Flow PURELAB Option-R 7/15 Process Flow PURELAB Option-R 7/15
Page 23 of 46
Page92
PURELAB AnAlyticAl ReseARch
TreatedWaterSpecificationsModel Option-S 7 Option-S15 Option-R 7 Option-R15Make up rate 1 7.5 l/hr 15 l/hr 7.5 l/hr 15 l/hrDispense rate from tap 1.0 l/min-nominal
(less with POU filter)1.0 l/min-nominal
(less with POU filter)Daily output (nominal max) 1 180 l/24 hour day 360 l/24 hour day 180 l/24 hour day 360 l/24 hour dayOutput reverse pressure (max) 2 0.1 bar (1 psi) 0.1 bar (1 psi) 0.1 bar (1 psi) 0.1 bar (1 psi)Purity:Inorganics@ 25°C 1 to >10 MΩ-cm 1 to >10 MΩ-cm 10 to >15 MΩ-cm 10 to >15 MΩ-cmTotal organic carbon (TOC) <30 ppb <30 ppb <20 ppb <20 ppbBacteria 2 n/a n/a <1 CFU/ml <1 CFU/mlpH Effectively neutral Effectively neutral Effectively neutral Effectively neutralParticles n/a n/a Optional 0.2µm
POU filterOptional 0.2µm
POU filter1 Standard conditions are 4 bar inlet pressure, 0 bar back pressure at 15 degrees centigrade, fed with potable water and a clean pre-treatment cartridge. 2 Subject to suitable feedwater
DimensionsandweightsHeight 460mm (18.1in) 460mm (18.1in) 460mm (18.1in) 460mm (18.1in)Width 410mm (16.1in) 410mm (16.1in) 550mm (21.7in) 550mm (21.7in)Depth 270mm (10.6in) 270mm (10.6in) 270mm (10.6in) 270mm (10.6in)Weight with internal boost pump 16kg (35lb) 16.5kg (36lb) 20kg (44lb) 21kg (46lb)Weight without internal boost pump
13.5kg (30lb) 14.5kg (32lb) 18kg (40lb) 19kg (42lb)
FeedwaterRequirementsSource Quality Potable mains water
supplyPotable mains water
supplyPotable mains water
supplyPotable mains water
supplyFouling index - maximum 10 10 10 10
Total dissolved solids - maximum <2000 µS/cm* <2000 µS/cm* <2000 µS/cm* <2000 µS/cm*Free chlorine - maximum 0.5 ppm 0.5 ppm 0.5 ppm 0.5 ppmHeavy metals - maximum 0.05 ppm 0.05 ppm 0.05 ppm 0.05 ppmSilica - maximum 30 ppm 30 ppm 30 ppm 30 ppmTemperature 1 - 35°C 1 - 35°C 1 - 35°C 1 - 35°CFlowrate (maximum requirement) 78 l/hr 85 l/hr 78 l/hr 85 l/hrDrain requirements (gravity fall with air gap). Maximum during service
70 l/hr
FeedwaterPressureMaximum - without internal boost pump 6.0 bar (90 psi) 6.0 bar (90 psi) 6.0 bar (90 psi) 6.0 bar (90 psi)Minimum - without internal boost pump 4.0 bar (60 psi) 4.0 bar (60 psi) 4.0 bar (60 psi) 4.0 bar (60 psi)Maximum - with internal boost pump 2.0 bar (30 psi) 2.0 bar (30 psi) 2.0 bar (30 psi) 2.0 bar (30 psi)Minimum - with internal boost pump Flooded Suction Flooded Suction Flooded Suction Flooded Suction
* Deionization cartridge life may vary with feedwaters >1400 µS/cm
ElectricalRequirementsMains input 100-240V ac, 50-60Hz 100-240V ac, 50-60Hz 100-240V ac, 50-60Hz 100-240V ac, 50-60HzSystem voltage 24V dc 24V dc 24V dc 24V dcPower consumption with boost pump 43VA 43VA 80VA 80VAPower consumption without boost pump
19VA 19VA 50VA 50VA
Fuses 2 x T3.15 Amp 2 x T3.15 Amp 2 x T6.3 Amp 2 x T6.3 AmpReservoir level connection Jack Plug 3.5mm Jack Plug 3.5mm Jack Plug 3.5mm Jack Plug 3.5mmNoise level <45dBA <45dBA <45dBA <45dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.comELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38753-03/1 Page 24 of 46
Page93
PURELAB AnAlyticAl ReseARch
• Delivers market leading microbial performance. In addition to simple sanitization and composite vent filtration, the PURELAB Option-R system is the only fully recirculating Type II system on the market, ensuring high specification water
• Features ‘RO feed optimized’ resin mixes to increase the ion exchange capacity of consumables and minimize running costs
• Quick and easy sanitization and replacement of consumables to reduce maintenance time
• PURELAB Option systems are designed to be easy to access whether wall or bench mounted with a convenient dispense tap
the PURelAB Option provides type ii pure water on demand at higher flow-rates of up to 60 liters per hour. Applications range from glassware washing to tissue and cell culture and media preparation. the PURelAB Option-R provides additional bacteria and inorganic quality for your more critical applications.
PURelAB Option-s/R 30/60
Type II pure water on demand at higher flow-rates
Drain
Returnto Break Tank
DeionizationCartridge
Pump
Break Tank
RecirculationPump
ReverseOsmosis
Water Quality Sensor
Water Quality Sensor
Feedwater Inlet
Outletto Reservoir
Pump Feedfrom Reservoir
UV Lamp
DispenseTap
Process Flow PURELAB Option-R 30/60
Pre-Filter
Process Flow PURELAB Option-R 30/60
Page 25 of 46
Page94
PURELAB AnAlyticAl ReseARch
TreatedWaterSpecifications
Model Option-S30 Option-S60 Option-R30 Option-R60Make up rate 1 30 l/hr 60 l/hr 30 l/hr 60 l/hr
Dispense rate from tap 1.0 l/min-nominal (less with POU filter)
1.0 l/min-nominal (less with POU filter)
Purity:
Inorganics @ 25°C 1 – 15 MΩ-cm 1 – 15 MΩ-cm 5 – 15 MΩ-cm 5 – 15 MΩ-cm
Total organic carbon (TOC) <50 ppb <50 ppb <20 ppb <20 ppb
Bacteria 2 <1 CFU/ml <1 CFU/ml <1 CFU/ml <1 CFU/ml
pH Effectively neutral Effectively neutral Effectively neutral Effectively neutral
Ion/organic cartridge capacity 175,000 liters >1 MΩ-cm µS/cm at pH7 3
175,000 liters >1 MΩ-cm µS/cm at pH7 3
165,000 liters >5 MΩ-cmµS/cm at pH7 3
165,000 liters >5 MΩ-cmµS/cm at pH7 3
1 Subject to suitable feedwater 2 Subject to suitable operating procedures 3 Conductivity of feed to Deionization Cartridge (RO Quality)
Dimensionsandweights
Height 740mm 740mm 740mm 740mm
Width 570mm 570mm 570mm 570mm
Depth 320mm 320mm 320mm 320mm
Weight (operational) 52kg 53kg 53kg 54kg
FeedwaterRequirements
Source quality Potable mains water supply
Fouling index - maximum 10
Total dissolved solids - maximum 2000µS/cm*
Free chlorine - maximum <0.5 ppm
Heavy metals - maximum 0.05 ppm
Temperature 1 - 35°C
Flowrate (maximum requirement) 9 l/min
Feedwater pressure maximum 6.0 bar (90 psi)
Feedwater pressure minimum 1.4 bar (20 psi)
* Deionization cartridge life may vary with feedwaters >1400 µS/cm
ElectricalRequirements
Mains input 230V ac, 50Hz115V ac, 60Hz
100V ac, 50/60Hz
Power consumption with boost pump
700VA
Fuses 2 x 3.15 Amp – 230V ac, 50Hz2 x 6.3 Amp – 115V ac, 60Hz
Reservoir level connection Din plug 6 way
Noise level <52dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38755-04 Page 26 of 46
Page95
PURELAB AnAlyticAl ReseARch
• Recirculation of purified water through our wrap-around reservoir – to maintain consistent peak water purity
• Optional ELGA Biofilter - when fitted to the Option-Q produces water which is free from biologically active impurities. This makes it suitable for use with applications which require endotoxin free ultrapure water, bacteria free water, and nuclease free ultrapure water
• Type III / RO water available from reservoir
• Front-entry service doors for easy and quick access to consumables
• Data collection capabilities through RS232 interface – for compliance with GLP guidelines
• Perfect single system solution for analytical and lifescience applications
tap to type i ultrapure water direct from a potable tap water supply. the PURelAB Option-Q is ideal for laboratories who need up to 100 liters of 18.2 MΩ.cm ultrapure water. With its simple design and ease of use, water can be dispensed at a flow rate of 1.0 l/min.
PURelAB Option-Q
Ultrapure type I water direct from a tapwater supply
Outlet Outlet
FeedwaterInlet
Boost Pump(optional)
Sanitization Port
Pre-Treatment
ReverseOsmosis
ReverseOsmosis
Drain
Water Quality Sensor
Recirculation Pump
Outlet toDocking Vessel
or Reservoir
Pump feed fromDocking Vessel
or Reservoir
DispenseTap
DeionizationCartridge
185/254 nm UV Lamp
WaterQualitySensor
Process Flow PURELAB Option-Q Process Flow PURELAB Option-Q
Option Q Datasheet 2011.indd 1 23/01/2012 11:16Page 27 of 46
Page96
PURELAB AnAlyticAl ReseARch
TreatedWaterSpecifications
Model Option-Q7 Option-Q15 WithELGABiofilterInorganics (resistivity @ 25°C) Up to 18.2 MΩ-cm Up to 18.2 MΩ-cm Up to 18.2 MΩ-cmOrganics (TOC) – typical 1 – 3 ppb 1 – 3 ppb 1 – 3 ppbDirect from reservoir Type III / RO water 1 Type III / RO water 1 Type III / RO water 1
Bacteria <1 CFU/ml <1 CFU/ml <0.1 CFU/mlEndotoxin - - <0.001 EU/mlRNase - - <0.002 ng/mlDNase - - <20 pg/mlpH Effectively neutral Effectively neutral Effectively neutral
1 (Equal to or >) Type III / RO water
Dimensionsandweights
Height 460mm (18.1in) 460mm (18.1in)Width 550mm (21.7in) 550mm (21.7in)Depth 270mm (10.6in) 270mm (10.6in)Weight with internal boost pump 20kg (44lb) 21kg (46lb)Weight without internal boost pump
18kg (40lb) 19kg (42lb)
FeedwaterRequirements
Source quality Potable mains water supply Potable mains water supplyFouling index - maximum 10 10Conductivity* <2000 μS/cm <2000 μS/cmTotal chlorine 0.5 ppm 0.5 ppmHeavy metals - maximum 0.05ppm 0.05ppmSilica - maximum 30 ppm 30 ppmTemperature 1 - 35°C 1 - 35°CFlowrate (maximum requirement) 78 l/hr 85 l/hrDrain requirements (gravity fall with air gap). Maximum during Service
70 l/hr 70 l/hr
FeedwaterPressureMaximum - without internal boost pump 6 bar (90 psi) 6 bar (90 psi)Minimum - without internal boost pump 4 bar (60 psi) 4 bar (60 psi)Maximum - with internal boost pump 2 bar (30 psi) 2 bar (30 psi)Minimum - with internal boost pump Flooded suction Flooded suction
* A restriction on the daily output may be necessary for feedwater >1400 µS/cm
ElectricalRequirements
Mains input 100 - 240V ac, 50-60Hz all models
100 - 240V ac, 50 - 60Hz all models
System voltage 24V dc 24V dcPower consumption with boost pump
90VA 90VA
Power consumption without boost pump
60VA 60VA
Fuses 2 x T6.3 Amp 2 x T6.3 AmpReservoir level connection Jack Plug 3.5mm Jack Plug 3.5mmNoise level <45dBA <45dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38757-04
Option Q Datasheet 2011.indd 2 23/01/2012 11:16Page 28 of 46
Page97
PURELAB AnAlyticAl ReseARch
Ultra-pure water at very economic costs for the equipment and • the cost of ownership
Complete sanitization of all wetted parts ensures optimum • microbial performance
Automatic intermittent recirculation minimizes temperature • build-up and optimizes microbial performance
Very easy to maintain – incorporates ‘fast rinse’ ultra filter•Upgradable from single pack to twin pack purification•
the PURelAB classic system combines high performance with economy to deliver 18.2 MΩ-cm water at a very cost-effective price. Despite its budget price the PURelAB classic contains many market leading features.
PURelAB classic
Ultra-pure water at a very economic cost
Feedwater Inlet
(Pre-purified Water) PumpDeionization
Cartridge
DrainPurified Water
Outlet
Dispenser
UltrafiltrationMembrane
Water Quality Sensor
0.2µm Point-of-use Filter(optional)
185/254 nm UV Lamp
Process Flow PURELAB Classic UVF
Process Flow PURELAB Classic UVF
Page 29 of 46
Page98
PURELAB AnAlyticAl ReseARch
TreatedWaterSpecifications
Model ClassicDI ClassicUV ClassicUF ClassicUVFFlowrate 2.0 l/min max 2.0 l/min max 2.0 l/min max 2.0 l/min maxInorganics 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cmTOC 3 – 10 ppb 1 – 3 ppb 3 – 10 ppb 1 – 3 ppbBacteria <1 CFU/ml1 <0.1 CFU /ml1,2 <0.1 CFU /ml11,2 <0.1 CFU /ml11,2
Bacterial endotoxin - - <0.001 EU/ml <0.001 EU/mlpH Effectively neutral Effectively neutral Effectively neutral Effectively neutralParticles 0.2 µm1 0.2 µm1 Ultrafiltration UltrafiltrationRNase and DNase - - Removed RemovedCartridge capacity (LC186) 45,000 liters >18MΩ-cm per single purification pack/µS at pH 7.0
70,000 liters >1MΩ-cm per single purification pack/µS at pH 7.0
1 With POU filter fitted. 2 <1 CFU/ml without point-of-use filter.
DimensionsandWeights
Height 490mm (19.3in) 490mm (19.3in) 490mm (19.3in) 490mm (19.3in)Width 410mm (16.2in) 410mm (16.2in) 410mm (16.2in) 410mm (16.2in)Depth 365mm (14.4in) 365mm (14.4in) 365mm (14.4in) 365mm (14.4in)Weight 14.0kg (30.8 lb) 14.5kg (32.0 lb) 14.5kg (32.0 lb) 15.0kg (33.1 lb)
FeedwaterRequirements
Parameter LimitsSource - originally from potable supply, then pre-treated
Preferably reverse osmosis (RO) or filtered service deionization (SDI) or distilled. Note: mixed bed or twin bed deionized supplies should be cation limited at exhaustion.
Fouling index (max) 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds.Service deionization (SDI) - MΩ-cm 1 MΩ-cm minimum resistivity at exhaustion.Reverse osmosis (RO) - µS/cm Recommended < 30 µS/cmFree chlorine 0.05 ppm max.TOC Recommended 50 ppb max.Carbon dioxide 30 ppm max.Silica 2 ppm max.Particulates Filtration down to 0.2 micron advisable to protect internal and/or point of use filters.Temperature 1 - 40°C Recommended 10 - 15°CFlowrate (maximum requirement) 130 l/hrDrain requirements (gravity fall with air gap). Maximum during service
Up to 2 l/min
Feedwater pressure 0.7 bar (10 psi) maximum, 0.07 bar (1 psi) minimum
ElectricalRequirementsMains input 100 - 240V ac, 50 - 60Hz all modelsSystem voltage 24V dcPower consumption during recirculation
60VA
Power consumption during dispense
75VA
Fuses 2 x T6.3 AmpReservoir level connection Jack Plug 3.5mmNoise level during recirculation <40dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.comELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2010 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38748-03 Page 30 of 46
Page99
PURELAB AnAlyticAl ReseARch
• Twin purification pack “PureSure” system gurantees water purity. It stops organics leaching in to the final water purity
• Real-time TOC monitoring ensures confidence in organic quality (Genetic and Analytic models only)
• PureSure system ensures that work is not interrupted by cartridge replacement
• Ongoing bacterial performance guaranteed through complete sanitization of all wetted parts
• Suitable for applications such as ICP-MS, LC-MS, HPLC, AAS, GC-MS, DNA sequencing, PCR, Molecular Biology and Plant Tissue
the PURelAB Ultra is the intelligent type i ultrapure water purification system for your most critical applications. it incorporates many unique and innovative features that set it apart from other systems and guarantees high water purity of 18.2 MΩ-cm.
PURelAB Ultra
Guaranteed water purity for your critical applications
Feedwater Inlet
(Pre-purified Water)Pump
Drain
Primary Cartridge
185/254 nmUV Lamp
Polishing Cartridge
IntermediateWater Quality
Sensor
Post UVQualitySensor
DrainFeedwater
Outlet
Dispenser
UltrafiltrationMembrane
WaterQuality Sensor
0.2µm Point-of-use Filter(optional)
Process Flow PURELAB Ultra GeneticProcess Flow PURELAB Ultra Genetic
Page 31 of 46
Page100
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA®, PURELAB® and PureSure® are registered trademarks of VWS (UK) Ltd.
LITR 38750-04
PURELAB AnAlyticAl ReseARch
TreatedWaterSpecificationsLifeScience Analysis GeneralScience
Model Genetic Bioscience Analytic Ionic ScientificFlowrate 2.0 l/min max 2.0 l/min max 2.0 l/min max 2.0 l/min max 2.0 l/min maxInorganics 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cmTOC <1 ppb 2 3 – 10 ppb 1 <1 ppb 2 3 – 10 ppb 1 3 – 10 ppb 1
Bacteria <0.1 CFU/ml 3 <0.1 CFU/ml 3 <0.1 CFU/ml 3 <0.1 CFU/ml 3 <1 CFU/ml 3
Bacterial endotoxin <0.001 EU/ml <0.001 EU/ml – – –pH Effectively neutral Effectively neutral Effectively neutral Effectively neutral Effectively neutralParticles Ultrafiltration Ultrafiltration 0.05 µm 0.05 µm 0.2 µm 2
RNase <0.002 ng/ml <0.002 ng/ml – – –DNase <20 pg/ml <20 pg/ml – – –Labpure cartridge capacity (LC182) Liters at 18.2 MΩ-cm = 80,000 / µS/cm + ( 2.3 x ppm CO2 )
1 Dependant on feedwater 2 With LC185 Low TOC Purification Cartridge fitted and fed with purified water from PURELAB Option-R/Pulse or other feeds with <20 ppb TOC. Otherwise TOC will be 1-3 ppb. 3 With POU filter fitted.
Dimensionsandweights
Dimensions Height 490mm (19.3”), Width 410mm (16.2”), Depth 365mm (14.4”)
Weight 15.0kg (33.1 lb) 14.5kg (32.0 lb) 15.0kg (33.1 lb) 14.5kg (32.0 lb) 14.0kg (30.8 lb)
FeedwaterRequirementsParameter LimitsSource - originally from potable supply, then pre-treated
Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled. Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion.
Fouling Index (max) 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds
Service Deionization (SDI) - MΩ-cm 1 MΩ-cm minimum resistivity at exhaustion
Reverse Osmosis (RO) - µS/cm Recommended <30 µS/cm
Free chlorine 0.05 ppm max
TOC Recommended 50 ppb max
Carbon dioxide 30 ppm max
Silica 2 ppm max
Particulates Filtration down to 0.2 micron advisable to protect internal and/or point of use filters
Temperature 1 - 40°C - Recommended 10 - 15°C
Flowrate (maximum requirement) 130 l/hr
Drain requirements (gravity fall with air gap). Maximum during service
Up to 2 l/min
Feedwater Pressure 0.7 bar (10 psi) maximum, 0.07 bar (1 psi) minimum
ElectricalRequirementsMains input 100 - 240V ac, 50 - 60Hz all modelsSystem voltage 24V dcPower consumption during recirculation
60VA
Power consumption during dispense
75VA
Fuses 2 x T6.3 AmpReservoir level connection Jack Plug 3.5mmNoise level during recirculation <40dBA
Page 32 of 46
Page101
Pure and ultrapure laboratory water from a single system
Treated Water SpecificationsMODEL PURELAB flex 1 PURELAB flex 1
With Purification PackPURELAB flex 2 PURELAB flex 3 PURELAB flex 4
Daily volume >10 liters >10 liters >10 liters <10 liters <10 litersDelivery flow rate – maximum Up to 2 l/min Up to 2 l/min Up to 2 l/min Up to 2 l/min Up to 2 l/min Inorganics (resistivity @25°C) As per feedwater 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cmOrganics (TOC) – typical Dependant on
feedwaterDependant on
feedwater<5ppb 1 <5ppb 1 <5ppb 1
Direct from internal reservoir N/A N/A N/A Type III / RO water 2 Type III / RO water 2
Bacteria – typical (when fitted with POU Filter) <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10mlBacteria – typical (when fitted with Biofilter) <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml Endotoxin (when fitted with Biofilter) N/A <0.001 EU/ml <0.001 EU/ml <0.001 EU/ml <0.001 EU/mlDNase (when fitted with Biofilter) N/A N/A <20 pg/ml <20 pg/ml <20 pg/mlRNase (when fitted with Biofilter) N/A N/A <0.002 ng/ml <0.002 ng/ml <0.002 ng/ml
Dimensions and weightsDimensions Width 236mm, Depth 374mm, Height minimum 900mm,
Height maximum 1020mm Width 236mm, Depth 470mm, Height
minimum 900mm, Height maximum 1020mm
Operational weight 10kg (22lb) 10.5kg (23.1lb) 11kg (24.2lb) 23kg (57.3lb) 23kg (57.3lb)Installation Bench/wall Bench/wall Bench/wall Bench/wall Bench/wall
Feedwater requirementSource Originally from potable supply, then pre treated.
Preferably reverse osmosis (RO) or filtered service deionization (SDI) or distilled.
Potable Tap Water Originally from potable supply, then pre treated.
Preferably reverse osmosis (RO) or filtered
service deionization (SDI) or distilled.
Conductivity <1 µS/cm 3 <1 µS/cm 3 <1 µS/cm 3 <2000 µS/cm 4 <30 µS/cm ContaminantHardness N/A N/A N/A <350 ppm as CaCO3 TBCFree Chlorine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.5 ppm Cl2 <0.05 ppm Cl2
Chloramine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.2 ppm Cl2 <0.05 ppm Cl2
Total Chlorine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.5 ppm Cl2 <0.05 ppm Cl2
Silica <2 ppm <2 ppm <2 ppm <30 ppm SiO2 <2 ppm SiO2
Carbon Dioxide Maximum <0.1 ppm <0.1 ppm <0.1 ppm <30 ppm (recommended <20 ppm)
<30 ppm (recommended <20 ppm)
Fouling index <1 <1 <1 <10 <1Iron / Manganese N/A N/A N/A <0.05 ppm N/AOrganics (TOC) N/A N/A N/A <2 ppm <50 ppb
recommendedParticulates A 0.2 micron membrane pre filter is recommended for all non-RO feeds to extend
point-of-use filter lifeN/A
Temperature 4 - 40°C (Recommended 10 - 25°C)Flowrate (requirement at 15°C) <2 l/min <2 l/min <2 l/min Up to 75 l/hr Up to 60 l/hrDrain requirements (gravity fall with air gap) None required None required None required >90L/hr >70L/hr
3 If feedwater is in the range of 1 µS/cm – 30 µS/cm you will need a pre-conditioning cartridge 4 Purification pack life may vary with feedwaters >1400 µS/cm
Feedwater pressureMaximum 22psi (1.5bar) 22psi (1.5bar) 22psi (1.5bar) 90psi (6bar) 5 90psi (6bar)Minimum Flooded suction Flooded suction Flooded suction 30psi (2bar) 5 1psi (0.0 7bar)
5 If <60psi (4 bar) a separate boost pump is recommended
Electrical requirementsMains Input 100-240V ac, 50-60Hz
System control voltage (not including pumps and UV)
24V dc
Power consumption during peak demand 120VA 120VA 120VA 100VA 100VA
Noise level <40dBa
Page 33 of 46
Page102
PURELAB flex 2
ELGA has been a trusted name in water purification for over 50 years, pioneering innovative technologies and award winning product design for our customers. Part of the world’s leading water services company Veolia Water Solutions and Technologies, ELGA have the most comprehensive range of water purification systems available today for laboratory research, healthcare and diagnostics.
Water is the most common reagent used in today’s laboratories for research and testing applications and we understand how important accurate consistent test results are. The PURELAB flex range is designed around your needs, delivering accuracy, flexibility and ease of use with an innovative and ergonomic design.
Daily volume >10 Liters <10 Liters >10 Liters <10 Liters >10 Liters
Water quality Type I ultrapure water Type II/Type II+ pure water Type III RO water
Feedwater source Pre-purified water
Pre-purified or tap water
Pre-purified water
Pre-purified or tap water
Pre-purified water
Typical applications
Electrochemistry
Electrophoresis
GFAAS (Graphite Furnace Atomic Absorption Spectrophotometry)
HPLC
IC (Ion Chromatography)
ICPAES (Inductively Coupled Plasma Atomic Emission Spectrometry)
ICPMS (Inductively Coupled Plasma Mass Spectrometry)
Mammalian and bacterial cell culture
Molecular biology
Plant tissue culture
Qualitative analyses
AAS (Atomic Absorption Spectrophotometry)
Buffer and media preparation
Electrophysiology
FAAS (Flame Atomic Absorption Spectrophotometry)
Feed to ultrapure water systems
Glassware washing/rinsing
General chemistry
Histology
Microbiological analysis
RIA (Radioimmunoassay) / ELISA (Enzyme Linked Immunoabsorbant Assay)
Sample dilution and reagent preparation
Spectrophotometry
Autoclave feed
Feed to ultrapure water systems
Hydroponics
Plant growth cabinets
Stability chambers
Steam generators
Sterilizer feed
Find a PURELAB flex to match your application
PURELAB flex 1 PURELAB flex 1PURELAB flex 3&4 PURELAB flex 3&4
Your laboratory water specialists
Page 34 of 46
Page103
POU filters for multiple applications• Remove endotoxins, DNase, RNase and bacteria
• Conforms to international water standards e.g. CLSI, CLRW, ISO 3696: Grade 1,2,3, ASTM D1193-06, Pharmacopeia USP, EP and JP
Optional accessories• Wall mounting bracket
• Foot pedal
• Thermal printer
• Point of Use filters
• Leak sensor
Flexible dispensing in four different ways:
• Variable flow – drop by drop or up to 2 liters per minute
• Autovolume dispense from 50ml to 60 liters and repeat dispensing
• Hands free with optional foot pedal
• Locked dispense for glassware filling
Clear water purity display for absolute confidence as you dispense
Prioritized information shown at all times (system status, TOC, alarm)
Intuitive to use, ergonomic handset design
Realtime TOC monitoring for critical analytical applications
Water purification made easy
Page 35 of 46
Page104
Quick and easy sanitizationMinimize microbial growth to give you absolute confidence in the water purity. The simple procedure involves one single sanitization pack so no harsh chemicals need to be ordered, handled or mixed. No additional consumables need to be replaced during this short process.
Global network, local supportOur highly trained service engineers will apply their expertise to the installation, validation and maintenance of your water purification system. We aim to provide a first-time-fix service.
Designed for today’s laboratories
Customized settings to suit your application
Designed for simple easy installation
185nm/254nm UV lamp for low TOC water production needed for organic sensitive application
Easy-access for routine maintenance
Multiple dispense positioning, wall, bench, height adjustable arm, hand-held dispensing
Data capture via USB for system performance validation
Purification pack quick and easy to replace
Page 36 of 46
Page105
Feedwater Inlet
(Pre-purified Water)
RecirculationPump
Dispense Handset 0.2µm Point of Use Filter or Biofilter (optional)
WaterPurity and
Temperature Sensor
Water Purity Sensor
Water Purity and Temperature
Sensor
Purification Pack
185/254 nm UV Lamp
Feedwater Inlet(potable tap water)
Recirculation Pump
ManualOutlet
Dispense Handset 0.2µm Point of Use Filter or Biofilter (optional)
Water Purity and
Temperature Sensor
Water Purity and Temperature
Sensor
Purification Pack185/254 nm
UV Lamp
CVF
Reservoir
Composite Vent Filter
ReverseOsmosis
Water Purity Sensor
Inside the PURELAB flex
PURELAB flex 2 Process Flow Diagram
Want to know more?Find out more or book your demonstration at www.purelabflex.com
Commitment to quality PURELAB flex systems are supplied with a Certificate of Conformity, ensuring they have been manufactured and tested at ELGA LabWater Global Operations, whose Quality Management System is approved by Lloyds Register Quality Assurance (LRQA) and complies with BS EN ISO 9001:2008.ELGA’s manufacturing facility has been approved by LRQA and complies with the environmental management system ISO 14001:2004.The PURELAB flex has been designed to conform to the requirements of the Waste Electrical & Electronic Equipment (WEEE) Directive and the Restriction of Hazardous Substance (RoHS) Directive.
Assured safetyThe PURELAB flex is compliant with CE directives associated with safety and electromagnetic compatibility and is approved and tested by an accredited external company. The PURELAB flex is also compliant with the requirements of IEC/EN61010-1 and UL61010-1 and is marked with an ETL label showing compliance.
PURELAB flex 3 Process Flow Diagram
Page 37 of 46
Page106
Feedwater Inlet
(Pre-purified Water)
RecirculationPump
Dispense Handset 0.2µm Point of Use Filter or Biofilter (optional)
WaterPurity and
Temperature Sensor
Water Purity Sensor
Water Purity and Temperature
Sensor
Purification Pack
185/254 nm UV Lamp
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2010 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR38975-03
Treated Water SpecificationsMODEL PURELAB flex 1 PURELAB flex 1
With Purification PackPURELAB flex 2 PURELAB flex 3 PURELAB flex 4
Daily volume >10 liters >10 liters >10 liters <10 liters <10 litersDelivery flow rate – maximum Up to 2 l/min Up to 2 l/min Up to 2 l/min Up to 2 l/min Up to 2 l/min Inorganics (resistivity @25°C) As per feedwater 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cmOrganics (TOC) – typical Dependant on
feedwaterDependant on
feedwater<5ppb 1 <5ppb 1 <5ppb 1
Direct from internal reservoir N/A N/A N/A Type III / RO water 2 Type III / RO water 2
Bacteria – typical (when fitted with POU Filter) <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10mlBacteria – typical (when fitted with Biofilter) <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml Endotoxin (when fitted with Biofilter) N/A <0.001 EU/ml <0.001 EU/ml <0.001 EU/ml <0.001 EU/mlDNase (when fitted with Biofilter) N/A N/A <20 pg/ml <20 pg/ml <20 pg/mlRNase (when fitted with Biofilter) N/A N/A <0.002 ng/ml <0.002 ng/ml <0.002 ng/ml
Dimensions and weightsDimensions Width 236mm, Depth 374mm, Height minimum 900mm,
Height maximum 1020mm Width 236mm, Depth 470mm, Height
minimum 900mm, Height maximum 1020mm
Operational weight 10kg (22lb) 10.5kg (23.1lb) 11kg (24.2lb) 23kg (57.3lb) 23kg (57.3lb)Installation Bench/wall Bench/wall Bench/wall Bench/wall Bench/wall
Feedwater requirementSource Originally from potable supply, then pre treated.
Preferably reverse osmosis (RO) or filtered service deionization (SDI) or distilled.
Potable Tap Water Originally from potable supply, then pre treated.
Preferably reverse osmosis (RO) or filtered
service deionization (SDI) or distilled.
Conductivity <1 µS/cm 3 <1 µS/cm 3 <1 µS/cm 3 <2000 µS/cm 4 <30 µS/cm ContaminantHardness N/A N/A N/A <350 ppm as CaCO3 TBCFree Chlorine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.5 ppm Cl2 <0.05 ppm Cl2
Chloramine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.2 ppm Cl2 <0.05 ppm Cl2
Total Chlorine <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <0.5 ppm Cl2 <0.05 ppm Cl2
Silica <2 ppm <2 ppm <2 ppm <30 ppm SiO2 <2 ppm SiO2
Carbon Dioxide Maximum <0.1 ppm <0.1 ppm <0.1 ppm <30 ppm (recommended <20 ppm)
<30 ppm (recommended <20 ppm)
Fouling index <1 <1 <1 <10 <1Iron / Manganese N/A N/A N/A <0.05 ppm N/AOrganics (TOC) N/A N/A N/A <2 ppm <50 ppb
recommendedParticulates A 0.2 micron membrane pre filter is recommended for all non-RO feeds to extend
point-of-use filter lifeN/A
Temperature 4 - 40°C (Recommended 10 - 25°C)Flowrate (requirement at 15°C) <2 l/min <2 l/min <2 l/min Up to 75 l/hr Up to 60 l/hrDrain requirements (gravity fall with air gap) None required None required None required >90L/hr >70L/hr
3 If feedwater is in the range of 1 µS/cm – 30 µS/cm you will need a pre-conditioning cartridge 4 Purification pack life may vary with feedwaters >1400 µS/cm
Feedwater pressureMaximum 22psi (1.5bar) 22psi (1.5bar) 22psi (1.5bar) 90psi (6bar) 5 90psi (6bar)Minimum Flooded suction Flooded suction Flooded suction 30psi (2bar) 5 1psi (0.0 7bar)
5 If <60psi (4 bar) a separate boost pump is recommended
Electrical requirementsMains Input 100-240V ac, 50-60Hz
System control voltage (not including pumps and UV)
24V dc
Power consumption during peak demand 120VA 120VA 120VA 100VA 100VA
Noise level <40dBa
1 Dependant on feedwater 2 (Equal to or >) Type III RO Water
Page 38 of 46
Page107
ProtekPretreatment
Cartridge
FeedwaterInlet
UV LampCategory 5Air Gap
BoostPump Drain
ReverseOsmosisModules
(Optional)E-Cartridge
PumpCVF
Reservoir
Composite Vent Filter
Delivery/ Recirculation
Pump
LabpureDeionization
Cartridge
(Optional) Labpure
DeionizationCartridge
Ultra Microfiltration
Water Quality Sensor
Water Quality Sensor
Process Flow PURELAB 7000 Series
ReturnFrom Loop
To Distribution
Loop
PURELAB AnAlyticAl ReseARch
• Auto rinse and simple sanitization procedures maintain the purity of water for the life of the system
• A variety of connections and distribution pumps are available to cater for a diverse range of applications
the PURelAB 3000 series is ideal for all high usage applications requiring type iii water. the flexibility of the system means that it can be easily upgraded to provide type ii pure water or increased capacity should requirements change. Pure water can be used direct from the internal 50 liter reservoir or fed to much larger vessels as required.
the PURelAB 7000 series is designed for applications requiring higher flows of type ii pure water or as a pure water feed system for small distribution loops. the system can supply water from the internal 50 liter reservoir at varying flow rates or can feed into much larger vessels as required.
PURelAB 3000
PURelAB 7000
Process Flow PURELAB 7000
• Small high purity distribution system with variable flow rates
• Auto rinse and simple sanitization procedures maintain the purity of water for the life of the system
Page 39 of 46
Page108
PURELAB AnAlyticAl ReseARch
TreatedWaterSpecificationsPURELAB3000 PURELAB7000
Model 3060 3120 7060 7120Make-up rate @ 15°C 60 l/hr 120 l/hr 60 l/hr 120 l/hrDaily usage – typical 480 liters 960 liters 480 liters 960 litersDaily usage – maximum 1440 liters 2880 liters 1440 liters 2880 litersDelivery flowrate - - Up to 4 l/min @
12 psi (0.8 bar) 3Up to 4 l/min @ 12 psi (0.8 bar) 3
Inorganics rejection >95% >95% - -Inorganics (resistivity @ 25OC) >10 MΩ-cm 1 , 2 >10 MΩ-cm 1 , 2 >15 MΩ-cm >15 MΩ-cmBacteria – typical - - <1 CFU/ml <1 CFU/mlOrganics (TOC) – typical <50 ppb 1 <50 ppb 1 <10 ppb <10 ppbParticles >99% Rejection 1 >99% Rejection 1 0.05 µm 0.05 µm
Specifications are for a system fed with a suitable water supply and installed, operated and sanitized according to the operator manuals.1 Quality of water feeding t he reservoir. Quality out of reservoir depends on use and design.2 With optional deionization cartridge fitted. 3 With High Flow Kit (LA723) or 2 l/min @ 12 psi (0.8 bar) without high flow kit
DimensionsandweightsDimensions Height 820mm (32.8in), 834mm (33.4in) including castors, Width 794mm (31.8in), Depth 470mm (18.8in)Supply weight 52kg 60kg 52kg 53kgOperational weight 109kg 117kg 109kg 110kgInstallation Floor/bench Floor/bench Floor/bench Floor/bench
FeedwaterRequirements
Source Tap water as detailed below
Contaminant
Conductivity <2000 µS/cm*
Hardness <250 ppm as CaCO3
Free chlorine <4 ppm Cl2
Chloramine <1 ppm as Cl2
Silica <30 ppm SiO2
Fouling Index <10 FI
Iron/manganese <0.1 ppm Fe/Mn
Organics <3 ppm TOC
Temperature 4 – 40°C (Recommended 15 – 25°C)
Flowrate (maximum requirement @ 15°C)
9 l/min
Drain requirements (gravity fall with air gap)
20 l/min
Feedwater pressure 6 bar (90 psi) maximum, 2 bar (30 psi) minimum
Failure to comply with the feedwater pretreatment recommendations will affect the life and performance of key components and may invalidate the warranty. Feedwater requirements are specified for units fitted with a Protek L2 pretreatment cartridge. If feedwater purity is variable or values are outside any of the above ranges, additional pretreatment is recommended to be installed in the feedwater supply to the unit. * Deionization cartridge life may vary with feedwater >1400 µS/cm If in doubt seek advice from Technical Support at ELGA LabWater.
ElectricalRequirementsMains input 230V ac, 50Hz 115V ac, 60HzSystem control voltage (not including pumps and UV)
24V dc
Power consumption (peak demand) 650VAElectrical protection rating 10 amps
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38749-04 Page 40 of 46
Page109
PURELAB ReseaRch & TesTing
• Incorporates patented ‘Pulse Technology’ using Electrodeionization (EDI) to provide a constant supply of high purity water at economical running costs. No costly resin replacement
• Unique integral recirculation ensures optimum pure water quality at point of dispense
• Improved laboratory productivity – longer periods between sanitization along with quick and easy replacement of consumables reduces maintenance time
• Easy to access whether wall or bench mounted, with a convenient dispense tap. The systems can be used with our wrap-around reservoir to minimize space whilst optimizing purity
The PUReLaB Pulse is the cost-effective choice for laboratories that require higher output volumes especially in hard water areas. applications range from sample dilution and reagent preparation, cell culture, media preparation and glassware washing.
PUReLaB Pulse
The only EDI Type II water system that fully recirculates
Outlet Outlet
FeedwaterInlet
Boost Pump(optional)
Sanitization Port
Pre-Treatment
ReverseOsmosis
ReverseOsmosis
Drain Drain
Recirculation Pump
Conditioning Cartridge
Outlet toDocking Vessel
or Reservoir
Pump feed fromDocking Vessel
or Reservoir
Pulse Module
DispenseTap
WaterQualitySensor
Water QualitySensor
Flow Switch
Process Flow PURELAB Pulse
UV Lamp
Process Flow PURELAB Pulse
Pulse Datasheet 2010.indd 1 23/01/2012 11:13Page 41 of 46
Page110
TreatedWaterSpecifications
Model Pulse 1 Pulse 2Make up rate @ 15oC 1 up to 10 l /hr up to 20 l /hrDaily output 1 up to 216 l /24 hour day up to 216 l /24 hour dayDispense rate from tap (max) 1.0 l / min-nominal (less with POU filter) 1.0 l / min-nominal (less with POU filter)Output back pressure (max) 2 0.1 bar (1 psi) 0.1 bar (1 psi)Inorganics @25°C 3 10 to >15 MΩ-cm 10 to >15 MΩ-cmTotal organic carbon (TOC) <20 ppb <20 ppbBacteria 2 <1 CFU/10ml <1 CFU/10mlpH Effectively neutral Effectively neutralParticles Optional 0.2 µm POU filter Optional 0.2 µm POU filter
1 Standard conditions are 4 bar inlet pressure, 0 bar back pressure, fed with potable water and a clean pre-treatment cartridge. 2 Subject to correct operating and maintenance procedures and use of POU filter. 3 Optimum performance will be achieved with moderate use on moderate feedwaters. At high usage, (>100 l/day) with high Total Conductivity and CO2 feedwaters (>700μS/cm, 20 ppm CO2) some reduction in product water resistivity may occur.
Dimensionsandweights
Height, Width, Depth Height 460mm (18.1in), Width 550mm (21.7in), Depth 270mm (10.6in)Operational weight with internal boost pump
20kg (44lb) 21kg (46lb)
Operational weight without internal boost pump
18kg (40lb) 19kg (42lb)
Installation Bench/wall Bench/wall
FeedwaterRequirements
Source Potable Mains water supply as detailed belowConductivity* <2000 µS/cm <2000 µS/cmTemperature 1 – 35oC 1 – 35oCFlow rate required (maximum) 80 l/hr 80 l/hrDrain requirements (minimum) 80 l/hr 80 l/hrContaminantHardness <350 ppm as CaCO3 <350 ppm as CaCO3
Free Chlorine <0.5 ppm <0.5 ppmChloramine <0.2 ppm <0.2 ppmTOC <3 ppm <3 ppmSilica <30 ppm <30 ppmFouling Index <10 <10Iron/Manganese <0.2 ppm <0.2 ppmCO2 – maximum** <30 ppm <30 ppmFeedwaterPressureWithout internal boost pump 90 psi (6.0 bar) maximum, 60 psi (4.0 bar) minimumWith internal boost pump 30 psi (2.0 bar) maximum, flooded suction minimum
* A restriction on the daily output may be necessary for feedwater >1400 μS/cm. ** Contact ELGA LabWater for feed water >30ppm
ElectricalRequirements
Mains input 100 - 240V ac, 50 - 60Hz 100 - 240V ac, 50 - 60HzSystem voltage 24V dc 24V dcPower consumption with boost pump 110VA 110VAPower consumption without boost pump
85VA 85VA
Fuses 2 x T6.3 Amp 2 x T6.3 AmpReservoir level connection Jack Plug 3.5mm Jack Plug 3.5mmNoise level <45dBA <45dBA
PURELAB ReseaRch & TesTing
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2011 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR 38754-03
Pulse Datasheet 2010.indd 2 23/01/2012 11:13Page 42 of 46
Page111
PURELAB ACCESSORIESRemote Dispensing anD monitoRing
Remote Dispense GunELGA Dispense Guns offer fingertip control over the flow rate, of up to 2 liters per minute. Water purity is maintained at all times by recirculation through the Dispense Gun.
Dispense Gun clips securely into a wall-mounting bracket for fixed dispensing or can • be unclipped for extended reach (1 meter range from the bracket)
The wall-mounting bracket can be positioned up to 2 meters from the water • purification system
Bench-top Dispensers and Control stationsRemote dispensing and monitoring enables you to position your pUReLaB water purification system under the bench or on a wall to free up valuable space on the bench-top.
Integral Dispense GunIncreases the dispensing range by 1 meter (3ft) •
Remote Control StationEnables the user to monitor water quality and check system status from a convenient location without needing to make a visual check on the main system. The Remote Control replicates the display on the water purification system and can be positioned up to 3 meters away from the unit.
Provides a choice of display modes for operational status • via push button controls
Wall mountable for flexibility and convenience•RS232 port for connection to a printer for easy data •
collection for compliance with GLP guidelines
Page 43 of 46
Page112
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.comELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2010 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.
LITR38918-02
PURELAB ACCESSORIESRemote Dispensing anD monitoRing
SpecificationsRemoteDispensersTube length Approx 2 metersFlow Rate(Depending on water system used) 0.8 – 2 l/min
OrderingInformationRemoteDispensersDescription PartNumberRemote Dispense Gun – PURELAB Option-R 7/15 and Option-E 5/10 LA629Remote Dispense Station (Including Dispense Gun) – PURELAB Ultra, Classic, Option-Q 7/15, Option-R 7/15 and Option-E 5/10
LA643
Integral Dispense Gun – PURELAB Ultra and Classic LA644Open Circuit Ring Main Kit – PURELAB Ultra and Classic LA646Closed Circuit Ring Main Kit – PURELAB Ultra and Classic LA647
DimensionsandWeightRemoteControlsHeight 120mmWidth 200mmDepth 70mmWeight 1kg
OrderingInformationRemoteControlsModel PartnumberPURELAB Ultra and Classic LA645PURELAB Option-Q 7/15 LA675PURELAB Option-S 7/15 LA624PURELAB Option-R 7/15 LA625PURELAB Option-E 5/10 LA626PURELAB Prima 7/15/30 LA628
Page 44 of 46