COLUMN EFFICIENCY PARAMETERS
PRESENTED BYSUBODH S SATHEESH
MPHARM PHARMACEUTICS
ECPS
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
2
It is expressed by the number of theoretical plates It is determined by the formula
The number of theoretical plates is a measure of the “goodness” of the column
If the retention time is high and peak width is narrow then it shows excellent chromatograms
Column efficiency
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
3
where tr is the retention time measured from the instant of injection
w is the peak width
W is determined by SD= σ ie w=4σ
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
4
Resolution is the ability to seperate two signals In chromatography its the ability to seperate two peaks.ie
seperation of constituents
where tr1 and tr2 and w1 and w2 are the times and widths, respectively, of the two immediately adjacent peaks.
If the peaks are sufficiently close w is nearly the same for both peaks and resolution may be expressed as
Resolution
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
6
The rates of migration of substances in chromatographic procedures depend on the relative affinity of the substances for the stationary and the mobile phases
Its the difference in time between the point of injection and the time of emergence of separation of component from the column.
It is actually the time required for 50% of the component to get eluted.
It is measure in minutes or seconds
Retention time
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
8
It is the volume of carrier gas required to elute components from the column to the time the peak maximum is obtained.
Retention volume depends upon flowrate and retention time
VR= tR-FC
Retention volume
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
9
It is the ratio of partition coefficient of two components to be separated.
S= Kb/Ka = K’a/k’b= (tb-to)/(ta-to)
If peaks are far apart ie there is more difference in partition coefficient between compounds hence more seperation factor and viceversa
Separation factor
Less seperation factor
More seperation factor
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
10
HETP is numerically equal to the column length divided by the number of theoretical plates in the column
It varies from to one column to another as well as one solute to other
The more efficient the column the better the resolution and the smaller the HETP.
HETP=Length of column / no of theoretical plates
HETP
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
11
Here we can see the effect of flowrate on HETP. An ideal flowrate happens in the minimal HETP.
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
12
A chromatographic peak should be symmetrical about its centre to follow gaussian distribution
Asymetric factor is the measure of peak tailing or fronting. It is defined as the distance from the centre line of the peak to
the back slope divided by the distance from the centre line of the peak to the front slope.
Asymetric factor
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
13
The chromatographic peak in (a) is an example of tailing, which occurs when some sites on the stationary phase retain the solute more strongly than other sites. The peak in (b) is an example of fronting, which most often is the result of overloading the column with sample. For both (a) and (b) the green chromatogram is the asymmetric peak and the red dashed chromatogram shows the ideal, Gaussian peak shape.
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
14
Its function is to separate sample components to discrete peaks
It should have reasonable chemical and thermal stability
There are a lot liquid stationary phases available for gas chromatography. But there is no solvent that meet all the requirements of a perfect stationary phase
LIQUID STATIONARY PHASES
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
15
Liquid phase should not permeate too deeply into the fine pores of the support structure as slow diffusion in and out of pores affects column efficiency
It should be chemically inert It should be a good solvent for sample component Liquid phase should have low volatility and high stability at
elevated temperatures otherwise they can contribute to interference in analysis
Characters of good Liquid StationeryPhase
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
16
Stationary Phase Trade Name Max Temp Common Applications
Dimethyl Polysiloxane
OV – 1, SE – 30 350oC Hydrocarbons, Polynuclear
aromatics, PCB’s
Poly(phenyl methyl) siloxane
OV – 17 250oC Steroids, Pesticides, Glycols
Poly (Trifluoro propyl dimethyl)
siloxane
OV – 210 200oC Chlorinated Aromatics, Nitro Aromatics, Alkyl
substituted Benzenes
Polyethylene Glycol Carbowax 20 M 250oC Free acids, Alcohols, Essential
Oils, Glycols
5% Diphenyl – 95% Dimethyl
polysiloxane
DB – 5 325oC Flavors, environmental samples and
aromatic hydrocarbons
Typical liquid stationary phases
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
17
Nonpolar solutes like pentane butane etc can be effectively separated by nonpolar solvents like squalene. And polar solutes can be easily separated by polar solvents eg; PEG.
Boiling point is also a factor of consideration. among solutes with similar polarity if there is sufficient difference in BP effective separation can take place.
Eg. Squalene min/max temp= 293/423 SE4 423/573 A solvent that could generate different partition ratios
among solvents can only be useful in GLC
Selection of solvents
PHARMACEUTICAL ANALYSIS-GAS CHROMATOGRAPHY
18
H KAUR Instrumental methods of chemical analysis ninth edition 2013 ;1091-1092