Summary: NOVAERUS was considered by Ward Staff to be generally acceptable and easy to use.For Environmental Surface TVC, values were 49% lower with the device switched on when compared to the internal control (device off). For Environmental Surface MRSA, the odds of a MRSA occurrence with the device onwas 3% (i.e. 97% reduction) of its internal control (device off). For Environmental Air MRSA, the odds of MRSA occurrence with NOVAERUS switched on were a quarter of that observed for the external control.
Citation preview
INDEPENDENT
VALIDATION
REPORT FOR
NOVAERUS AIR TREATMENT SYSTEMS
KEY FINDINGS
NOVAERUS was considered by Ward Staff to be generally acceptable and easy to use.
For Environmental Surface TVC, values were 49% lower with the device switched on when compared to the internal control (device off).
For Environmental Surface MRSA, the odds of a MRSA occurrence with the device on was 3% (i.e. 97% reduction) of its internal control (device off).
For Environmental Air MRSA, the odds of MRSA occurrence with NOVAERUS switched on were a quarter of that observed for the external control.