Low NO/NO 2 Non-Dispersive UltraViolet Analyzer Sensors’ Low NO/NO2, uses Non-Dispersive Ultra-Violet (NDUV) technology to measure the levels of NO and NO 2 in the exhaust of both spark ignited and compression ignited engines. The analyzer meets the updated 2010 US EPA standards, as well as measurement standards worldwide. Non- dispersive ultraviolet analysis is an absorpon spectroscopy technique used for gas analysis. The ultraviolet wavelength is used for the measurement of NOx because it is not cross sensive to CO 2 and H 2 O, which do not absorb well in the UV region. Sensors’ latest design NDUV is built from the latest technologies, offering improved features and benefits over our previous design. Emissions Measurement Solutions

Semtech NO NO2 Analyser

Embed Size (px)

DESCRIPTION

Sensors NO-NO2 Analyser with dual channel option for pre/post catalyst.

Citation preview

Page 1: Semtech NO NO2 Analyser

Low NO/NO2Non-Dispersive UltraViolet Analyzer

Sensors’ Low NO/NO2, uses Non-Dispersive Ultra-Violet (NDUV) technology to measure the levels of NO and NO2 in the exhaust of both spark ignited and compression ignited engines. The analyzer meets the updated 2010 US EPA standards, as well as measurement standards worldwide. Non-dispersive ultraviolet analysis is an absorption spectroscopy technique used for gas analysis. The ultraviolet wavelength is used for the measurement of NOx because it is not cross sensitive to CO2 and H2O, which do not absorb well in the UV region. Sensors’ latest design NDUV is built from the latest technologies, offering improved features and benefits over our previous design.

Emissions Measurement

Solutions

Page 2: Semtech NO NO2 Analyser

Sensors, Inc. • 6812 State Rd. • Saline, Michigan 48176 • PH:(734) 429-2100 • FX:(734) 429-4080 • www.sensors-inc.comSensors Europe GmbH • Feldheider Str. 60 • 40699 Erkrath • Germany • PH: +49 (0) 2104-14188-0 • FX: +49 (0) 2104-14188-14

NOTE: Specifications are subject to change without no-tice. While due caution has been exercised in the produc-tion of this document, possible errors and omissions can occur.

R 06/10

Low NO/NO2

Non-Dispersive UltraViolet Analyzer

Features

LED Light Source: The LED can be operated at very high frequencies, enabling an excellent signal-to-noise ratio.

Shock Resistance: The mechanical design has been optimized for resistance to shock and vibration, ensuring accurate data in the most rugged of in-use environments.

24-bit A/D Converter: This converter prevents DACs from railing.

6 Temperature Controlled Zones: Temperature controlled zones for the detectors, the pressure transducer, the LED and the EDL ensure accurate data throughout a wide range of ambient temperatures, and minimize drift.

Backward Compatibility: The 2010 Low NO/NO2 is compatible with our previous NDUV.

2010 Low NO and NO2 Specifications

Gas NO NO2

Range of measurement 0 to 3,000 ppm 0 to 500 ppm

Accuracy± 5 ppm or

± 1% of reading± 5 ppm or

± 1% of reading

Resolution 0.1 ppm 0.1 ppm

Linearity

Intercept ≤ 0.5% of range0.990 ≤ Slope ≤ 1.010SEE ≤ 1.0% of range

R2 ≥ 0.998

Intercept ≤ 1.0% of range0.985≤ Slope ≤ 1.015SEE ≤ 1.0% of range

R2 ≥ 0.998

Repeatability± 5 ppm or ± 2% of read-

ing± 5 ppm or ± 2% of read-

ing

Noise ± 2 ppm ± 2 ppm

Span Drift (over 8 hours)± 2% of reading or

20 ppm, whichever is greater

± 10 ppm

Zero Drift (over 2 hours) ± 10 ppm ± 10 ppm

Response time < 2 sec < 2 sec