41
LED SLIT LAMP MW50D Instructions February 2017 D046-2E

LED SLIT LAMP MW50D Instructions

Embed Size (px)

Citation preview

LED SLIT LAMP

MW50D

Instructions

February 2017

D046-2E

1

CAUTION Read this instruction manual before using the instrument.

This instruction manual was written to provide the user with information

about safe operation of the LED SLIT LAMP MW50D.

This manual contains operating instructions, safety precautions, and

specifications for the instrument.

For installation instructions, contact your local dealer.

The material in this manual complies with the ISO and IEC standards.

Read this manual thoroughly and familiarize yourself with the operation of

the instrument and safety precautions before using the instrument.

Keep this instruction manual handy and refer to it whenever necessary.

If you have any questions regarding this manual, contact your local dealer.

・No reproduction of this manual in any form, in whole or in part, may be

made without written permission from Right MFG. Co., Ltd.

・The information contained in this manual is subject to change without

notice.

2

Contents

1. GENERAL INFORMATION .................................................................................................. 10

1-1. PRODUCT DESCRIPTION, STANDARD COMPONENTS, AND OPTIONAL ACCESSORIES ........ 10

1-2. INSTRUMENT CLASSIFICATIONS ............................................................................................. 11

1-3. SYMBOLS FOUND ON THE INSTRUMENT ................................................................................ 12

1-4. SAFETY LABELS ....................................................................................................................... 13

2. PICTORIAL NOMENCLATURE AND FUNCTIONS ........................................................ 14

2-1. SYSTEM CONFIGURATION ....................................................................................................... 14

2-2. EYEPIECE BLOCK .................................................................................................................... 15

2-3. ILLUMINATOR BLOCK .............................................................................................................. 15

2-4. CROSS-SLIDE TABLE ............................................................................................................... 16

2-5. CHIN REST ................................................................................................................................ 17

2-6. POWER SUPPLY UNIT .............................................................................................................. 17

3. INSTALLATION ..................................................................................................................... 17

3-1. CHECKING BEFORE USE .......................................................................................................... 18

3-2. EYEPIECE LENS ....................................................................................................................... 18

3-3. INSTALLING THE MIRROR ....................................................................................................... 19

3-4. DIOPTER ADJUSTMENT ........................................................................................................... 19

3-5. BACKGROUND ILLUMINATION UNIT ...................................................................................... 20

4. OPTIONAL ACCESSORIES .................................................................................................. 21

4-1. TONOMETER ADAPTER ............................................................................................................ 21

4-2. GUIDE PLATE ........................................................................................................................... 22

4-3. BREATH GUARD ....................................................................................................................... 22

4-4. HOLDING BARS ........................................................................................................................ 22

4-5. 78D AND 90D HOLDERS .......................................................................................................... 23

4-6. IMAGING UNIT .......................................................................................................................... 23

4-7. INCLINATION TUBE .................................................................................................................. 23

4-8. BEAM SPLITTER, CCTV ATTACHMENT, C-MOUNT(1/2),C-MOUNT(1/3) .............................. 24

4-9. DETACHABLE BACKGROUND ILLUMINATION UNIT ............................................................... 24

5. MAINTENANCE ..................................................................................................................... 25

5-1. REPLACING FUSES ................................................................................................................... 25

5-2. INSTALLING THE CHIN REST PAPER ...................................................................................... 25

5-3. CLEANING THE FOREHEAD REST, CHIN REST, AND HOLDING BARS ................................. 26

5-4. CLEANING THE LENS............................................................................................................... 26

5-5. CLEANING THE EXTERIOR OF THE INSTRUMENT .................................................................. 26

5-6. LIST OF SPARES AND SERVICE PARTS .................................................................................... 26

5-7. ZERO-POSITION ADJUSTMENT OF THE SLIT WIDTH ADJUSTMENT KNOB ......................... 27

3

5-8. IF THE SLIT WIDTH GETS NARROWER DURING OBSERVATION: ......................................... 28

6. TROUBLESHOOTING........................................................................................................... 29

7. SPECIFICATIONS .................................................................................................................. 31

■CONNECTION WITH PC ...................................................................................................... 33

■PATIENT ENVIRONMENT .................................................................................................. 33

■INFORMATION OF EMC (ELECTRO MAGNETIC COMPATIBILITY) ......................... 34

■EMC (ELECTROMAGNETIC COMPATIBILITY) .............................................................. 35

4

■Read First For Your Safety

Safety symbols used in this manual

Righton products are designed to provide you utmost safety during use. However,

incorrect operation or disregard of the instructions can cause personal injury or

property damage. Read the instruction manual carefully and thoroughly before using

the instrument. Keep this manual handy for quick reference. The following safety alert

symbols are used in this manual. Make sure you understand the meaning of the

symbols and observe the instructions given with it.

Symbol Meaning

WARNING: This symbol shows that a death or personal injury might occur

if you disregard this warning and use the instrument

improperly.

CAUTION: This symbol shows a potentially hazardous situation which, if

not avoided, may result in minor or moderate injury, or

property damage.

WARNING: To avoid the risk of electric shock, the instrument must only be connected

to a supply mains with protective earth.

WARNING: Do not modify the instrument without authorization of the manufacturer.

WARNING: The light emitted from this instrument is potentially hazardous. The longer the duration of exposure, the greater the risk of ocular damage. Exposure to light from this instrument when operated at maximum intensity will exceed the safety guideline after 2 min.

Spectrum of light souce

5

CAUTION 1. Intended use

Use the MW50D only for ophthalmic observation and microphotography

purposes. Do not use this instrument for any other purpose.

CAUTION 2. Do not disassemble

Disassembling this instrument may lead to electric shock and/or malfunction of

the instrument. Never disassemble this instrument.

CAUTION 3. Precautions to take when installing or carrying the instrument

・Always rest the instrument on a stable surface. Do not put it on a tilted surface.

The instrument may fall or drop and cause personal injury.

・Recommended operating conditions are as follows: barometric pressure 700 to

1060 hPa; ambient temperature 10 to 40ºC; and humidity 30 to 90%.

・This instrument complies with the EMC standards (IEC60601-1-2:2014).

However, it emits weak radio waves. This means that the instrument may

cause interference with radio communications. If this instrument does cause

interference with radio or television reception, try to increase the distance

between the instrument and the device that is receiving the interference, or

reorient the instrument.

・Remember that this instrument is not water-proof. Do not ever use or install it in

a location where exposure to liquids (such as rainwater, drinks, chemicals) may

occur.

・If condensation (dew) forms, do not use the instrument until the condensation

disappears.

・When carrying this instrument, remove the cable from the power supply unit,

and carry it by holding the base of the instrument. Holding or lifting it at a point

somewhere other than the base may damage the instrument.

・In case of installation of the main unit and accessories, options, be performed

by the trained service personnel always.

6

CAUTION 4. Precautions to take before and during observation

・When positioning a patient’s head in the main unit, use caution not to allow any

moving parts of the instrument to hit the person’s face while preparing the

instrument for observation. Preparations include adjustment of the observation

position, focus, observation and illumination angles, etc.

・Do not use the instrument in a dusty room.

・The best place to install the instrument is in a dimly-lit room. Do not have the

patient’s side of the instrument face a bright window or a light source. This may

adversely affect observations.

・Replace the lamp bulb with a new one when the bulb starts to dim. Insufficient

light may lead to improper observations.

・The eyesight of the patient may be temporarily impaired after an examination

because of the intense light used. Allow the patient to rest a while after an

examination, until the patient’s eyesight returns to normal.

・Do not use an excessive amount of light during an examination, to avoid placing

an unnecessary burden on the patient. Improper illumination intensity may

prevent correct observations.

・Allow a patient who has trouble viewing stereoscopically or one who squints to

wear their prescription lenses. A patient whose diopter correction is ±8D or

more will also need the correction offered by their glasses.

・Be careful not to pinch your fingers between the moving parts of the instrument

during observation.

・Keep other people away from the instrument during an examination. The patient

may be distracted, resulting in incorrect results.

7

CAUTION 5. Precautions to the light hazard

Because prolonged intense light exposure can damage the retina, the use of the

device for observation should not be unnecessarily prolonged, and the brightness

setting should not exceed what is needed to provide clear visualization of the

target structures. This device should be used with filters that eliminate UV

radiation (< 400 nm) and, whenever possible, filters that eliminate

short-wavelength blue light (<420 nm).

The retinal exposure dose for the light hazard is a product of the radiance and the

exposure time. If the value of radiance were reduced in half, twice the time would

be needed to reach the maximum exposure limit.

While no acute optical radiation hazards have been identified for slit lamps, it is

recommended that the intensity of light directed into the patient’s eye be

limited to the minimum level which is necessary for diagnosis.

Infants, senior citizens, aphakes and persons with diseased eyes will be at greater

risk.

The risk may also be increased if the person being examined has had any

exposure with the same device or any other ophthalmic device using a visible

light source during the previous 24 hours. This will apply particularly if the eye

has been exposed to retinal photography.

CAUTION 6. Other precautions to take during handling

・When positioning a patient’s head in the main unit, use caution not to allow any

moving parts of the instrument to hit the person’s face while preparing the

instrument for observation. Preparations include adjustment of the observation

position, focus, observation and illumination angles, etc.

・Do not scratch, break or alter the power cord, or overly bend, pull, twist or bundle

it. Also, placing heavy object on the power cord or subjecting it to heat may cause

it to break, resulting in fire or electric shock. If the power cord becomes damaged,

replace it with a new one.

・Do not connect or disconnect the power cord with a wet hand.

・If there is any dust on the power plug blades or the surface they connect to, pull

out the power plug and remove the dust. The dust or grime can make the power

plug and AC power outlet to be out of contact and may result in fire.

・Connect the power supply unit to a grounded electric outlet that meets the power

supply requirements of the unit. Failure to do so may affect the performance of

the instrument adversely or result in malfunction of the instrument, electrical

leaks, electric shock, or fire.

・Connect the power cord plug securely. Improper connection may cause a fire.

・Do not place obstacles near the instrument which might interfere when

operating the power switch or connecting and disconnecting the power cord from

the wall outlet.

8

・When replacing the fuse in the instrument with a new one (See 5-1. Replacing

Fuses), turn off the power and disconnect the power cord from the electrical

outlet. Do not use any fuses other than those specified. A fire or electrical shock

may result.

Specified fuse:

LITTELFUSE 0215.400XP (T400mAH 250V φ5×20mm)

・This instrument is a precision optical device containing many electronic

components. Handle it carefully and do not subject it to strong physical shock.

・Do not drop the instrument or bump it against other objects.

・If the instrument fails, remove the power cord from the wall outlet and contact

your dealer. Do not attempt to repair it by yourself.

・Please use the power supply cord of the attachment.

Area Order number

100V area:DDB20207

120V area:DDB20201-UL (hospital grade)

220V area:DDB20202

240V area:DDB20203

240V area:DDB20209 (China)

・The operator must not touch the external instrument and the external monitor

connected with this equipment and the patient simultaneously. Electrical shock

can result.

・When the device is return for maintenance or repair, clean the surface of the

device (especially, forehead rest, chinrest) with a cloth dampened with rubbing

alcohol for disinfection.

・Accessory equipment connected to the analog and digital interfaces must be

certified according to the appropriate national standards.

Furthermore, all configurations shall comply with the IEC 60601-1. Anyone

who connects additional equipment to the signal input part or signal output

part is responsible for making sure that the system complies with the

requirements of the IEC 60601-1.

If in doubt, consult the technical service department or your local

representative.

CAUTION

7. Precautions to take when transporting and storing the instrument

・The recommended transportation condition is as follows: ambient temperature

-40 to +70ºC , humidity 10 to 95% and atmospheric pressure 500 to 1060hPa.

・The recommended storage condition is as follows: ambient temperature -10 to

+55ºC , humidity 10 to 95% and atmospheric pressure 700 to 1060hPa..

・Store the instrument in a dust free location not subject to vibration or mechanical

shock.

9

・Be careful not to scratch or break the lens when cleaning it. (See5-4. Cleaning the

Lens)

・After examining each patient, clean and disinfect the areas that are in contact

with the patient. Failure to do so may result in spreading an infection to other

patients. (See5-3. Cleaning the Forehead Rest, Chin Rest, and Holding Bars)

・After using the instrument, turn off the power switch on the power supply unit

and install the dust cover.

・Remove the power cord from the wall outlet if the instrument will be out of

service for a long period of time. Any accumulation of dust between the

power cord plug and wall outlet may cause a fire.

CAUTION 8. Precautions to take when disposing of the instrument ・Observe any local codes or regulations when disposing of or recycling the

instrument. It is advisable to allow an authorized industrial waste disposal specialist to dispose of the unit.

・When disposing of the packaging for the instrument, sort them by material type and dispose of them properly, according to your local codes or regulations.

Symbol for separate collection in European countries

Products bearing this symbol must be collected separately. The following rule is applicable only to users in European countries. ●This product is designated as an item to be collected separately

at an appropriate spot. Do not dispose of it as household waste. For further information, contact the retailer or the local authorities responsible for waste management.

CAUTION 9. Maintenance

USER MAINTENANCE

To maintain the safety and performance of the instrument, unless done by trained service personnel, never attempt to do maintenance of items other than those specified here in. For details about maintenance, read the description of this instruction. REPLACING THE FUSE Fuses on the primary side can be replaced, if necessary. For specific instructions, see 5-1. Replacing Fuses.

10

1. General Information

1-1. Product Description, Standard Components, and Optional Accessories

The MW50D slit lamps are designed for ophthalmic observation and microphotography.

These slit lamps is 5 step magnification microscopes that can be used for ophthalmic

observation.

This equipment is intended for use by medical doctors and optometrists.

This equipment is intended to use in the hospital.

Each instrument consists primarily of the main unit and the power supply unit.

<Operating principles>

Illuminates the observed part such as cornea, lens and etc by the illumination light emitted

from the illumination optical system and allows enlargement observation by binocular

stereoscopic microscope.

<Standard Components>

●Main unit (1) ●Power supply unit (1) ●12.5X eyepiece (1)、

12.5X eyepiece with reticle (1)

●Dust blower (1) ●*Mirrors (2) ●*Target rod (1)

(Large) (Small)

●*Protective cap (1) ●Power cord(1)

●Dust cover (1)

●Fuse (2) 0215.400XP (T400mAH 250V φ5×20mm)

●Instruction manual (this book) (1)

●Basic instruction (1)

●Power supply mounting plate (2)

●Three kinds of fixing screw for the table (each 4)

●Hex-key wrench ( 3 mm) (1)

* Detachable accessory

3.0m

100V area:DDB20207

120V area:DDB20201-UL (hospital grade)

220V area:DDB20202

240V area:DDB20203

240V area:DDB20209 (China)

11

<Optional Accessories>

●90D holder ●78D holder ●Chin rest paper ●Holding bar ●Breath guard ●Guide plate

●Tonometer adapter ●Imaging unit ●Inclination tube●Beam splitter ●CCTV attachment

●C-mount (1/2), C-mount (1/3) ●Detachable background illumination unit ●Software

instruction manual for the imaging unit

●CD-ROM (imaging unit software program MW File) ●Security key

Refer to 4. Optional Accessories for detail.

1-2. Instrument Classifications

<Classification under 93/42 EEC(MDD)>

Class I

<Type of protection against electrical shock>

Class I

This instrument is classified as equipment whose protection against electric shock does

not rely on basic insulation only, but which includes an additional safety precaution, in

that means are provided for the connection of the equipment to a protective earth

conductor in the fixed wiring of the installation in such a way that accessible metal parts

cannot become hazardous electric potential in the event of a failure of the basic

insulation.

Be sure to use the power cord that is supplied with the instrument, and connect it to a

grounded electrical outlet.

<Degree of protection against electrical shock>

Type-B applied part

The chin rest and forehead rest of the instrument are classified as “type-B applied parts”.

This instrument has undergone the tests (a patient leakage current test and a patient

auxiliary current test) specified in the relevant standard, and is equipped with a reliable

means to protect against electrical shock.

<Protection against harmful contamination by water or small particles>

IPX0

This instrument is not protected against exposure to water or other liquids.

Keep liquids away from the instrument.

<Degree of safety when used in a flammable atmosphere>

This instrument is not suitable for use in a flammable atmosphere.

Do not use this instrument if any flammable gases are present.

<Sterilization and disinfection methods permitted by the manufacturer>

Clean the chin rest, forehead rest, and holding bars with a clean, soft cloth soaked with

rubbing alcohol.

<Operation mode>

This instrument can be operated continuously.

12

<ISO15004-2 classification>

Group 2

The light emitted from this instrument is potentially hazardous.

The longer the duration of exposure, the greater the risk of ocular damage. Exposure to

light from this instrument when operated at maximum intensity will exceed the safety

guideline after 2 min

1-3. Symbols Found on the Instrument

:Calls your attention to a caution.

:Indicates a type-B applied part that complies with the specified requirements to

provide protection against electrical shocks.

:Indicates a connection to be used for DC power (alternating current) only.

:This symbol indicates the necessity of referring to the relevant part of the

instruction manual before use.

:Output

:Marked on the power switch. The power is on when this side is pressed down.

:Marked on the power switch. The power is off when this side is pressed down.

:Fuse

:Manufacturer

:Year of Manufacture

:Mind your finger warning symbol. Beware of moving mechanical parts to avoid

clipping fingers.

13

1-4. Safety Labels

The safety labels found on the instrument contain important information for the safety

of the user.

● Main unit

●Power supply unit

●Type-B applied parts

・Forehead rest ・Chin rest

14

2. Pictorial Nomenclature and Functions

2-1. System Configuration

●MW50D Clinical set (without Imaging unit)

MW50D digital imaging set (with Imaging unit)

]

Power supply unit

Main unit

*PC not supplied.

The operator must not touch the power/communication connectors

and the patient simulataneously.Electrical shock can result.

Power supply unit

Main unit

Imaging unit

PC for recording*

15

2-2. Eyepiece Block

2-3. Illuminator Block

Turn this ring before starting observation to compensate for differences in eyesight between the left and right eyes of the examiner and ensure easy viewing.

Adjust the spacing between the left and right eyepiece tubes to the pupilary distance of the examiner.

Used to change the magnification 5 step (5x 10x 16x 25x 50x).

Magnification control knob

Eyepiece tube

Slit aperture window

Move this lever left or right to select the filter to be used. The filters are assigned to the lever positions as follows from left to right. :Transparent filter

:ND (12.5% transmission) filter

:Green filter

:Cobalt blue filter

Slit adjustment knob Swing this knob left or right to rotate the slit. Turning this knob changes the slit aperture within a range of 0.2, 1, 2, 5, 10, 14, and 16 mm dia. and the slit length from 1 to 12 mm.

Filter selector lever Used to observe or photograph the anterior segment of the patient’s eye. Keep the diffuser in the down position when not in use.

Diffuser

Barrier filter knob

Switches the barrier filter ON and OFF. When performing observations using fluorescent light, use the filter selector lever in the illuminator block to select a cobalt blue filter. Then switch the barrier filter on. Observation through the barrier filter (a yellow filter) in the eyepiece block allows high-contrast imaging.

Displays the slit aperture that has been adjusted using the slit adjustment knob.

Diopter adjustment ring

G

B

16

2-4. Cross-Slide Table

Slit width adjustment knob Turned to change the slit width continuously.

Centering knob

Turning the knob counterclockwise loosens the unit so that the slit image can swing horizontally. Turning the knob clockwise tightens the unit so that slit image returns to the center position. Installed to protect the

positioning rod (that holds the guide plate in place) from injuring the patient.

Illumination lock knob Turned to lock the angle between the illumination and observation systems.

Tilting stopper

Tilting lever

Shutter button

Observation lock knob

Turned to clamp the observation system.

Turned to clamp the cross-slide.

Turning this knob clockwise increases the brightness and turning it counterclockwise decreases it. The brightness can also be decreased by using the ND filter.

Tilting the joystick finely adjusts the position of the main unit (illuminator, microscope, and base). For coarse adjustment, slide the base while holding the joystick. Turning the joystick adjusts the height of the main unit (illuminator and microscope).

Cross-slide lock knob

Illumination control knob

Pressed to make the tilting of the illumination lamp free. The tilting angles are 0, 5, 10, 15, and 20 degrees.

Protective cap

Press the button to take a static image. This switch is used when observations are being made using the imaging unit.

Joystick Pilot lamp

Comes on when power is present and the power supply unit is in operation.

17

2-5. Chin Rest

2-6. Power Supply Unit

Patient eye level index The upper index of the two indexes grooved on each of the chin rest columns serves as a guide for the patient eye level.

Turning this handle moves the chin rest up or down.

Chin rest elevation handle

Forehead rest

Chin rest

Power switch

Flexible fixation lamp

Used to fix the patient’s line of sight. The flexible arm allows you to select the desired position easily.

Fuse is built.

Fuse holder

Connect to the power supply

cable from the main unit.

Power supply output connector

for main unit. AC inlet

This connector is used to

connect the power cord.

Connect to the power supply

cable from the table.

Power supply output connector

for fixation lamp

Turns the power on or off.

Power ON:

Power OFF:

18

3. Installation

3-1. Checking before use

(1) Switch on the power supply unit and make sure that the pilot lamp lights in green.

Also, open the slit and check if the illumination lamp emits light.

(2) Do preparation according to 5-2. Installing the Chin Rest Paper , if you use the chin

rest paper (spare parts).

3-2. Eyepiece Lens

The eyepiece is fixed with the index line facing up.

Make sure to insert each eyepiece lens fully. Failure to do so may prevent the

microscope from focusing properly.

Insert the eyepiece with the reticle into the right eyepiece tube sleeve.

Reticle View in the right eyepiece.

Pilot lamp. Check if green

light is on.

Power supply unit

Check if emits light

Eyepiece lens (Set the lens with reticle in the right sleeve.)

Eyepiece tube

Index line

Eyepiece lens fixation screw

・ Change a new chin rest paper for each patient.

・ Tear the paper towards operator side for faster changing of paper.

・ Beware of pulling off the pin while tearing the paper upwards.

Power switch

Flip to the [ ] side

19

3-3. Installing the Mirror

Insert the mirror as shown below.

Two types of mirror are available: a large one with a projection and a small one

without a projection.

When using the instrument in the tilt-up mode, use a mirror without a projection to

prevent eclipse from occurring in the view field.

3-4. Diopter Adjustment

(1) Insert the end of the target rod into the center hole, which is the center of rotation, in

the microscope block and illuminator block and turn it so that its black flat surface

faces the microscope mirror.

(2) Set the illuminator block to the 0-degree position.

(3) Turn on the power supply unit. Set the slit width to an appropriate value for easy

viewing.

(4) Turn the magnification control knob to the 16x magnification position.

(5) Set the pupil distance by adjusting the spacing between the left and right eyepiece

tubes.

(6) Turn the diopter adjustment ring of the right-side eyepiece fully counterclockwise.

(7) While observing through the right eyepiece tube, turn the diopter adjustment ring

Perform a diopter adjustment before starting an observation or photography session.

Skipping this step may result in out-of-focus observation or picture, or in improper

slit movement.

(Large) (Small) Mirror for tilting

Diopter adjustment ring

Target rod

20

clockwise until the reticle in the eyepiece is seen most sharply.

(If the ring is turned too far, turn the ring counterclockwise again to its end and

restart the adjustment.)

(8) Turn the diopter adjustment ring of the left eyepiece fully counterclockwise.

(9) While observing through the left eyepiece tube, turn the diopter adjustment ring

clockwise until the flat surface of the target rod is seen most sharply.

(If the ring is turned too far, turn the ring counterclockwise again to its end and

restart the adjustment.)

(10) Turn the right eyepiece without moving the diopter adjustment ring so that the

reticle in the right eyepiece rests horizontally in the view field.

(11) Turn the eyepiece lens fixation screw and fix the eyepiece.

3-5. Background Illumination Unit

The background illumination unit allows you to observe and photograph the surroundings together with the object itself.

Back ground illumination brightness can be adjusted by background illumination

knob

Background illumination unit

Background illumination knob

An examiner who wear spectacles should collapse the rubber eyecup.

Rubber eyecup

21

4. Optional Accessories

The following are available as optional accessories.

4-1. Tonometer adapter

Adapter for hanging type tonometer. Attach to the top of microscope unit.

For information on compatible tonometers, contact your dealer.

Tonometer adapter

Installing tonometer

Holding bar

78D holder

Imaging unit

Breath guard

Tonometer adapter

Inclination tube

Guide plate

Beam splitter

CCTV attachment

C-mount (1/2),

C-mount (1/3)

Detachable background

illumination unit

90D holder

22

4-2. Guide Plate

Positioning plate for a stationary tonometer.

As shown below, secure the guide plate so that it engages the positioning pin. Then

install a stationary tonometer, aligning it with the curved area ① and the hole ②.

For information on compatible tonometers, contact your dealer.

4-3. Breath Guard

Breath guard is the plate prevent from blowing to the patient.

4-4. Holding Bars

Holding bars are designed for the patient to grip, to permit stable observation.

Positioning plate for stationary tonometer

Breath guard

Holding bars

23

4-5. 78D and 90D Holders

The 78D or 90D holder, when used with 78D or 90D lens, enable stable fundus

observations. See the figure shown below to install the 78D or 90D holder.

For instructions on the 78D and 90D lenses, refer to the instruction manual attached

to each lens.

4-6. Imaging unit

The Imaging unit allows you to observe video and record still images, video on a PC. Video : 1296 x 972 Still images : 2592 x1944

* see the instruction manual for the filing soft. The R/L sensor can automatically distinguish between the left and right eyes, and send that information to the filing program.

4-7. Inclination tube

It is possible to tilt the optical path of 15 ° by attaching the device.

User can make easy observation by attitude such as looking down.

Height is adjustable while eyepiece tube is being installed in rotated direction. Make sure

to place the eyepiece with reticle to right side after installed the eyepiece tube in rotated

direction.

Imaging unit Connect to PC to review

movie or capture image.

Normal Installation Rotate Installation

90D lens

24

4-8. Beam splitter, CCTV attachment, C-mount(1/2),C-mount(1/3)

External camera unit can be installed to the unit by adopting beam splitter and CCTV

device. When using external camera unit, make sure to turn the beam splitter to IN.

By turning the depth of field switch, operator can adjust the intensity and depth of field.

4-9. Detachable background illumination unit

Detachable background illumination unit enables to light the objects in any angle.

Examiner can hold and use the detached background illumination.

* If detachable background illumination unit is used, standard background

illumination unit is not able to use, however, light adjustment can be made by the

standard background light control knob.

External Camera Unit

C Mount(1/2) or C Mount (1/3)

Select according to the type of

external camera unit

Aperture lever Adjust the intensity and depth of

field.

Focus adjustment knob If the image is out of

focus, turn the knob while

looking at the image to

make adjustment.

Detahchable background

illumination unit It is possible to illuminate

any position. (upper and

lower, right and left, and

rotating.)

Beam splitter The image can be seen by

making beam splitter IN.

25

5. Maintenance

This device does not require periodic calibration.

Adjust the eyepiece diopter every time before using the device. (Refer to 3-4. Diopter

Adjustment)

5-1. Replacing Fuses

CAUTION

Insert the tip of the screwdriver into each of the fuse holder latches and pull out the

fuse holder toward you.

5-2. Installing the Chin Rest Paper

Remove the two chin rest paper pins from the chin rest. Place the chin rest paper (one

piece or more as needed) on the chin rest. Then install the pins to secure the paper.

Used chin rest paper can be removed from the chin rest without removing those pins.

Turn off the power and disconnect the power cord from the electrical outlet before checking

or replacing the fuses.

Use the following types of fuses. Do not use any fuses other than those specified. To order

spare fuses, contact your dealer.

LITTELFUSE time lag fuse:

0215.400XP (T400mAH 250V φ5×20mm)

Chin rest paper

Chin rest paper pins

Chin rest

Power supply unit

AC inlet

Fuse holder

26

5-3. Cleaning the Forehead Rest, Chin Rest, and Holding Bars

Cleaning should be done before the observation. Clean the Chin rest, Forehead and

Holding Bars with soft cloth soaked with rubbing alcohol.

ex)“Schnelldesinfektion”, ” Sixtan® Desinfektion”:They are registered in the VAH list.

When using the chin rest paper, use a new sheet for each patient. For the replacement

method of the chin rest paper, Refer 5-2 “Installing the Chin Rest Paper.”

5-4. Cleaning the Lens

If the lens gets dirty, blow the dust off the lens three or four times using the blower

that came with the instrument. If the dust is still not removed, gently wipe the lens

using a lens cleaning solution or absolute alcohol (commercially available).

5-5. Cleaning the Exterior of the Instrument

Do not use any organic solvents (alcohol, ether, thinner, etc.) when cleaning the coated,

plastic, or printed areas on the instrument. Discoloration or removal of the printing

may result. In cases of stubborn dirt, gently wipe the surface with gauze slightly

moistened with a diluted mix of water and mild detergent.

5-6. List of Spares and Service Parts

Part Name Part No. Remarks

Fuse

LITTELFUSE time lag fuse

0215.400XP

T400mAH 250V

φ5×20mm

Chin rest paper RXA30101

(Order No.) 500 pcs. per package

Be careful not to scratch or break the lens when cleaning it.

Chin rest

Forehead rest

Holding bars

27

5-7. Zero-Position Adjustment of the Slit Width Adjustment Knob

If there is any difference between the scale on the slit width adjustment knob and the

actual slit length, use the following procedure to adjust the zero position of the slit

width adjustment knob.

(1) While observing through the eyepiece lens, turn the slit width adjustment knob

clockwise and stop when the slit is fully closed.

(2) Loosen the slotted screw (located between the seals on the knob bottom) on the

scale ring to set the scale ring free.

Turn the scale ring until the “0” index on the seal on the scale ring is in alignment with

the reference index and tighten the slotted screw you loosened in step (2).

Never turn the slit width adjustment knob counterclockwise.

Do not turn the slit width adjustment knob at this moment.

Do not turn the slit width adjustment knob at this moment.

Scale ring

Hex-head wrench

Slit width adjustment knob

Viewed from bottom

Hexagon socket setscrew

28

(3) Turn the slit width adjustment knob one turn clockwise and make sure that the

slit is closed when the index line is set to “0”. Repeat steps (1) to (3) if the slit is not

closed.

5-8. If the Slit Width Gets Narrower During Observation:

The problem of the slit width becoming narrower during observation can occur if the

slotted screw in the center of the slit width adjustment knob has become loose.

Turn this screw clockwise a little with an Allen wrench (Hex-head wrench).

● Excessive tightening of the slotted screw clockwise will make the movement of the

slit width adjustment knob slower.

● Do not turn the slotted screw fully counterclockwise. The screw will come off. Once

this screw is removed, you may have to call your dealer for repair.

Hex-head wrench

Torque adjusted knob Viewed from bottom

Hexagon socket setscrew

29

6. Troubleshooting

If you have any problem with the instrument, check the following before asking your

dealer for assistance.

Problem Checks Remedy

Pilot lamp on the cross

slide table does not

come on, nor does the

illumination lamp.

Is the fuse blown? Replace the fuse.

Is the power cord firmly

connected to the wall outlet and

to the power supply unit?

Connect firmly.

Pilot lamp on the cross

slide table comes on,

but the illumination

lamp does not come on.

Is the slit closed? Open the slit.

Is the illumination knob being

turned off?

Rotate the knob to adjust the

brightness.

Is the illumination lamp blown? Replace the lamp.

Contact your dealer for exchange

service.

Uneven illumination Check the slit width adjust

knob position.

Turn the knob back to the click

position.

Is the illumination lamp blown? Replace the lamp.

Contact your dealer for exchange

service.

Insufficient

illumination

Is the ND filter installed? Remove ND filter using the filter

selector lever.

Is the illumination control knob

set too dark?

Turn the knob to adjust the

amount of light.

Microscope does not

focus.

Is the eyepiece lens fully

inserted?

Fully insert the eyepiece lens

and fix..

Did you perform a diopter

adjustment before starting the

observation?

Perform a diopter adjustment.

Refer to section 3-4. Diopter

Adjustment.

Slit does not return to

the center position.

Is the centering knob loose

enough that the slit position has

moved accidentally?

Turn the knob clockwise to

tighten it.

Slit aperture is not

centered.

Is the slit adjustment knob not

sitting in a detent?

Turn the knob until you feel a

click at the desired setting.

Slit width gets

narrower during

observation.

Is the set screw in the center of

the slit width adjustment knob

loose?

Tighten the screw. Refer to

section 5-8. If the Slit Width

Gets Narrower During

Observation

30

The position of the

main unit cannot be

changed with the

joystick.

Is the main unit locked with the

cross-slide lock knob?

Loosen the knob.

The angle between the

illuminator block and

the microscope cannot

be changed.

Is the angle locked with the

illumination lock knob?

Loosen the knob.

Microscope is stuck. Is it locked with the observation

lock knob?

Loosen the knob.

Background

illumination is not

bright enough.

Is the background illumination

knob being turn off?

Adjust the brightness by

turning the knob.

No display while

using external camera

unit.

Is the beam splitter turn to IN? Turn the beam splitter to IN to

make sure light goes into

external camera unit.

No image appears on a

PC during an

observation using the

Imaging unit.

Are the USB cable and PC

properly connected?

Connect properly.

Is the necessary program

installed on PC?

Install the program.

The image seen

through the eyepiece

during observations

using the Imaging unit

does not match the one

shown on the PC.

Are you looking at the image

through the right eyepiece?

Look through the right eyepiece.

Remember that the Imaging

unit camera is installed on the

same side as the right eyepiece.

This means the image seen

through the right eyepiece will

appear on the PC.

31

7. Specifications

<Instrument Name> LED SLIT LAMP MW50D

Main Unit

■ Microscope

Type 5 step magnifications microscope

Objective angle 13.2 º

Total magnification 5x, 10x, 16x, 25x, 50x

Actual field of view φ44.5,φ22.3, φ14.2, φ9, φ4.5

Eyepiece lens magnification

12.5x

Eyepiece diopter adjustment range

-8D to +8D

PD adjustment range 55 mm to 80 mm

Binocular tube

convergence angle.

Working distance 100.5 mm

Reaching distance 314.0 mm

■ Illumination system

Light source White LED 5W

Slit width 0 to 16 mm (continuous)

Slit rotation angle 90º to the right and left

Slit aperture φ0.2, 1, 2, 5, 10, 14, 16 mm; 1 to 12 mm (continuous)

Slit tilting angle 0, 5, 10, 15, 20º (from bottom)

Slit swing 8 mm to the left or right (on the target face)

Filters Transparent, ND (12.5%), green and cobalt blue filters

■ Arm

Rotation angle 90º to the left or right

■ Cross-slide table

Horizontal motion (Forward/Back and Left/Right)

100 x 110 mm (via joystick) (Forward: 30 mm; Back: 70 mm; Left/Right: 55 mm)

Vertical motion 30 mm (15 mm up or down via joystick)

■ Chin rest

Vertical motion 30 mm up or down from the reference position which is 274 mm above the table top.

■ Wight

Main unit without Imaging unit

13 kg

Main unit with Imaging unit

14 kg

■ Dimensions

Main unit with metal table

350 x 387 x678.5 mm (width x depth x height)

32

Height from table top to patient’s eye

374 mm

Power supply unit

Input AC100V- 240V 50/60Hz

Output DC12V

Power consumption 35VA

Rated fuse T400mAH 250V / φ5×20mm

Dimensions 105 x 130 x 48.5mm (width x depth x height)

Wight 450 g

Imaging unit

Optical path division ratio

(T/R)

30:70

Image size 2592(H) x 1944(V)

Movie size 1296(H) x 972(V)

Aspect ratio 4:3

Interface USB 3.0

Inclination tube

Inclination angle. 15°

Beam splitter

Optical path division ratio

(T/R)

30:70

CCTV attachment

Aperture diameter φ9-φ1.6 mm (Linear Variable)

Click position φ9, 4.5, 2.3 mm

Adjustment range of focus ±3 mm

■ ISO10939 : 2007 Compliance

33

■Connection with PC

CAUTION

Accessory equipment connected to the analog and digital interfaces must be certified

according to the appropriate national standards.

Furthermore, all configurations shall comply with the system standard IEC 60601-1.

Anyone who connects additional equipment to the signal input part or signal output part is

responsible for making sure that the system complies with the requirements of the system

standard IEC 60601-1.

If in doubt, consult the technical service department or your local representative.

■Patient environment

The patient environment represents a space where there is a possibility of direct contact

between the patient or the operator and third person.

When another type of device is used in the patient environment, use a device that complies

with IEC 60601-1. If the devices that do not comply with IEC 60601-1 are used, it is

necessary to use an isolating transformer to power the device or to connect the devices to

additional protective grounding.

Radius of 1.5m

1.5m 1.5m

2.5m

34

■Information of EMC (Electro Magnetic Compatibility)

CAUTION

1. Reorient or relocate the receiving device. Reorient or relocate the receiving device.

2. Increase the distance to the device.

3. Correct the device into an outlet on a circuit different from that to which the other

device(s) are connected.

4. Consult the manufacturer or field service technician for assistance.

・In installation and operation of the device, observe the following instructions about EMC

(electromagnetic compatibility):

1. Do not use the device simultaneously with other electronic equipment to avoid

electromagnetic interference with the operation of the device.

2. Do not use the device near, on, or under other electronic equipment to avoid

electromagnetic interference with the operation of the device.

3. Do not use the device in the same room with other equipment such as life-support

equipment, other equipment that has major affects on the life of the patient and

results of treatment, or other measurement or treatment equipment that involves

small electric current.

4. Do not use the device simultaneously with portable and mobile radio frequency

communication systems because it may have an adverse effect on operation of the

device.

5. Do not use cables and accessories that are not specified for the device because that

may increase the emission of electromagnetic waves from the device or the system

and decrease the immunity of the device to electromagnetic disturbance.

35

■EMC (Electromagnetic Compatibility)

The Electromagnetic Compatibility Directive sets the essential requirements for electrical and electronic equipment that

may disturb or even be disturbed by other equipment. The MW50D complies with these requirements as tabled below.

Follow the guidance on the tables for use of the device in the electromagnetic environment.

WARNING: The use of this equipment adjacent to or stacked with other equipment

should be avoided because it could result improper operation.

If such use is necessary, this equipment and the other equipment should

be observed to verify that they are operating normally.

WARNING: The use of accessories and cables other than those specified or provided by

the RIGHT MFG. CO., LTD. of MW50D could result in increased

electromagnetic emissions or decreased electromagnetic immunity of

MW50D and result in improper operation.

(A):Environments for MW50D

This device is applied for use in professional healthcare facility environment.

-Doctor's offices, clinics, medical facilities, hospital, etc.

(B)List of cables, Length of cables

Model name Model No. Length

Power cord 100V area DDB20207 3.0m

Power cord 120V area (Hospital grade) DDB20201-UL 3.0m

Power cord 120V area DDB20202 3.0m

Power cord 240V area DDB20203 3.0m

Power cord 240V area(China) DDC20209 3.0m

36

(C)EMC (IEC60601-1-2:2014)

(1)Emission guidance MW50D is intended for use in the electromagnetic environment specified below. The customer or the user of the

MW50D should assure that they are used in such an environment.

Emissions test Basic EMC standard or test method※1

Conducted and radiated RF

EMISSION CISPRE 11 Group 1 ClassB

Harmonic EMISSION IEC61000-3-2※2

Voltage fluctuations and flicker IEC61000-3-3※2

※1:See “(A) Environments for MW50D” for information about the environments of INTENED USE.

※2:This test is not applicable in this environment unless the ME EQUIPMENT and ME SYSTEMS used there will be connected to the

PUBLIC MAINS NETWORK and the power input is otherwise within the scope of the Bacis EMC standard.

(2)Immunity guidance MW50D is intended for use in the electromagnetic environment specified below. The customer or the MW50D

should assure that they are used in such an environment.

・Enclosure port

Phenomenon

Basic EMC

standard or test

method

IMMUNITY TEST LEVELS

Professional healthcare facility environment

Electrostatic discharge IEC 61000-4-2 ±8 kV contact

±2 kV, ±4 kV, ±8 kV, ±15 kV air

Radiated RF EM fields※1 IEC 61000-4-3

3 V/m※6

80 MHz – 2,7 GHz※2

80% AM at 1 kHz※3

Proximity fields from RF

wireless communications

equipment

IEC 61000-4-3

See (C)-(3)

Test Specifications for ENCLOSURE PORT IMMUNITY

to RF wireless communications equipment

Rated power frequency

magnetic fields※4※5 IEC 61000-4-8

30 A/m※7

50 Hz or 60 Hz

※1:The interface between the PATIENT physiological signal simulation, if used, and the ME EQUIPMENT or ME

SYSTEM shall be located within 0,1 m of the vertical plane of the uniform field area in one orientation of the ME

EQUIPMENT or ME SYSTEM.

※2:ME EQUIPMENT and ME SYSTEMS that intentionally receive RF electromagnetic energy for the purpose of their

operation shall be tested at the frequency of reception. Testing may be performed at other modulation frequencies

identified by the RISK MANAGEMENT PROCESS. This test assesses the BASIC SAFETY and ESSENTIAL

PERFORMANCE of an intentional receiver when an ambient signal is in the passband.It is understood that the

receiver might not achieve normal reception during the test.

※3:Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS.

※4:Applies only to ME EQUIPMENT and ME SYSTEMS with magnetically sensitive components or circuitry.

※5:During the test, the ME EQUIPMENT or ME SYSTEM may be powered at any NOMINAL input voltage, but with

the same frequency as the test signal

※6:Before modulation is applied.

※7:This test level assumes a minimum distance between the ME EQUIPMENT or ME SYSTEM and sources of power

frequency magnetic field of at least 15 cm. if the RISK ANALYSIS shows that the ME EQUIPMENT or ME

SYSTEM will be used closer than 15 cm to sources of power frequency magnetic field, the IMMUNITY TEST

LEVEL shall be adjusted as appropriate for the minimum expected distance.

37

・Input a.c. power port

Phenomenon

Basic EMC

standard or test

method

IMMUNITY TEST LEVELS

Professional healthcare facility environment

Electrical fast transients /

bursts※1※12※15 IEC 61000-4-4

±2 kV

100 kHz repetition frequency

Surges ※1※2※10※15

Line-to-line IEC 61000-4-5 ±0,5 kV, ±1 kV

Surges※1※2※10※11※15

Line-to-ground IEC 61000-4-5 ±0,5 kV, ±1 kV, ±2 kV

Conducted disturbances

induced by RF fields※3※4※15 IEC 61000-4-6

3 V ※13

0,15 MHz – 80 MHz

6 V※13 in ISM bands between 0,15 MHz and 80 MHz※14

80% AM at 1 kHz※5

Voltage dips ※6※16※18 IEC 61000-4-11

0% UT; 0,5 cycle※7

At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°※17

0% UT; 1 cycle and 70% UT; 25/30 cycles※8

Single phase: at 0°

Voltage interruptions※6※9※15※18 IEC 61000-4-11 0% UT; 250/300 cycle※8

※1:The test may be performed at any one power input voltage within the ME EQUIPMENT or ME SYSTEM RATED

voltage range. If the ME EQUIPMENT or ME SYSTEM is tested at one power input voltage, it is not necessary to

re-test at additional voltages.

※2: All ME EQUIPMENT and ME SYSTEM cables are attached during the test.

※3:Calibration for current injection clamps shall be performed in a 150 Ω system.

※4:If the frequency stepping skips over an ISM or amateur band, as applicable, an additional test frequency shall be

used in the ISM or amateur radio band. This applies to each ISM and amateur radio band within the specified

frequency range.

※5:Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS.

※6:ME EQUIPMENT and ME SYSTEMS with a d.c. power input intended for use with a.c.-to-d.c. converters shall be

tested using a converter that meets the specifications of the MANUFACTURER of the ME EQUIPMENT orME

SYSTEM. The IMMUNITY TEST LEVELS are applied to the a.c. power input of the converter.

※7:Applicable only to ME EQUIPMENT and ME SYSTEMS connected to single-phase a.c. mains.

※8:E.g. 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz.

※9:ME EQUIPMENT and ME SYSTEMS with RATED input current greater than 16 A / phase shall be interrupted

once for 250/300 cycles at any angle and at all phases at the same time (if applicable). ME EQUIPMENT and ME

SYSTEMS with battery backup shall resume line power operation after the test. For ME EQUIPMENT and ME

SYSTEMS with RATED input current not exceeding 16 A, all phases shall be interrupted simultaneously.

※10:ME EQUIPMENT and ME SYSTEMS that do not have a surge protection device in the primary power circuit may

be tested only at ± 2 kV line(s) to earth and ± 1 kV line(s) to line(s).

※11:Not applicable to CLASS II ME EQUIPMENT and ME SYSTEMS.

※12:Direct coupling shall be used.

※13:r.m.s., before modulation is applied.

※14:The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz;

13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands

between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3

MHz, 10,1 MHz to 10,15MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to

24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.

※15:Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A /

phase and ME EQUIPMENT and ME SYSTEMS with RATED input current greater than 16 A / phase.

38

※16:Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal to 16 A / phase.

※17:At some phase angles, applying this test to ME EQUIPMENT with transformer mains power input might cause an

overcurrent protection device to open. This can occur due to magnetic flux saturation of the transformer core after

the voltage dip. If this occurs, the ME EQUIPMENT or ME SYSTEM shall provide BASIC SAFETY during and

after the test.

※18:For ME EQUIPMENT and ME SYSTEMS that have multiple voltage settings or auto ranging voltage capability,

the test shall be performed at the minimum and maximum RATED input voltage. ME EQUIPMENT and ME

SYSTEMS with a RATED input voltage range of less than 25% of the highest RATED input voltage shall be

tested at one RATED input voltage within the rage.

・Signal input/output parts port

Phenomenon

Basic EMC

standard or test

method

IMMUNITY TEST LEVELS

Professional healthcare facility environment

Electrostatic discharge※4 IEC61000-4-2 ± 8 kV contact

± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air

Conducted disturbance by RF

field※1※3※5 IEC61000-4-6

3 V※6

0,15 MHz – 80 MHz

6 V※6 in ISM bands between 0,15 MHz and 80 MHz※7

80% AM at 1 kHz※2

※1: SIP/SOPS whose maximum cable length is less than 3 m in length are excluded.

※2: Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS.

※3: Calibration for current injection clamps shall be performed in a 150 Ω system.

※4: Connectors shall be tested per 8.3.2 and Table 4 of IEC 61000-4-2:2008. For insulated connector shells, perform air

discharge testing to the connector shell and the pins using the rounded tip finger of the ESD generator, with the

exception that the only connector pins that are tested are those that can be contacted or touched, under conditions of

INTENDED USE, by the standard test finger shown in Figure 6 of the general standard, applied in a bent or

straight position.

※5: If the frequency stepping skips over an ISM or amateur radio band, as applicable, an additional test frequency

shall be used in the ISM or amateur radio band. This applies to each ISM and amateur radio band within the

specified frequency range.

※6: r.m.s., before modulation is applied.

※7: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz;

13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.

The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to

5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to

21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.

39

(3)Test Specifications for ENCLOSURE PORT IMMUNITY to RF wireless

communications equipment Test

Frequency

(MHz)

Band※1

(MHz)

Service※1 Modulation※2

Maximum

power

(W)

Distance

(m)

IMMUNITY

TEST LEVEL

(V/m)

385 380-390 TETRA400

Pulse

Modulation※2

18Hz

1.8 0,3 27

450 430-470 GMRS460

FRS460

FM※3

±5kHz deviation

1kHz sine

2 0,3 28

710

745

780

704-787 LTE Band 13,

17

Pulse

modulation※2

217Hz

0,2 0,3 9

810

870

930

800-960

GSM800/900,

TETRA800,

iDEN820,

CDMA850,

LTE Band 5

Pulse

modulation※2

18Hz

2 0,3 28

1720

1845

1970

1700-1990

GSM1800;

CDMA1900;

GSM1900;

DECT;

LTE Band 1, 3,

4, 25;

UMTS

Pulse

modulation※2

217Hz

2 0,3 28

2450 2400-2570

Bluetooth,

WLAN,

802.11 b/g/n,

FRID 2450,

LTE Band 7

Pulse

modulation※2

217Hz

2 0,3 28

5240

5500

5785

5100-5800 WLAN 802.11 a/n

Pulse

modulation※2

217Hz

0,2 0,3 9

NOTE:If necessary to achive the IMMUNITY TEST LEVEL,the distance between the transmitting antenna and the

ME EQUIPMENT or ME SYSTEM may be reduced to 1m. The 1m test distance is permitted by IEC61000-4-3.

※1:For some services, only the uplink frequencies are included.

※2:The carrier shall be modulated using a 50% duty cycle square wave signal.

※3:As an alternative to FM modulation, 50% pulse modulation at 18Hz may be used because while it does not

represent actual modulation, it would be worst case

RIGHT MFG. CO., LTD.

1‐47‐3, Maeno-cho, Itabashi-ku, Tokyo, 174‐8633 Japan.

Tel : +81‐3‐3960‐2275 Fax : +81‐3‐3960‐2285

TOHOKU RIGHT MFG. CO., LTD.

Ophthalmic service, Factory

45‐1, Aza Yashikimae, Nakamura Osato-cho, Kurokawa-gun, Miyagi

981‐3521 Japan.

Tel : +81‐22‐359‐3113 Fax : +81‐22‐359‐3213

Medical Device Safety Service GmbH

Schiffgraben 41

30175 Hannover, Germany