3D Echocardiography:

Principles & Applications

Roberto M Lang, MD

@RobertoMLang

1DM-mode

3DReconstruction Real-time

2DCut plane

Evolution of EchocardiographyEvolution of Echocardiography

Page 2: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

3D Reconstructive Methods

SpatialLocators

Mechanically Driven transducers

Gated Sequential Acquisition

Echo Lab

U of C

Off-line 3D Image Processing and

Reconstruction

Page 3: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Real-Time 3D Acquisition: Sparse Array

Sparse Array(~300 elements)

Volumetrics

Courtesy of Dr. Masood Ahmad, UTMB Galveston

C-scan

Orthogonal

Real-Time 3D Acquisition: Matrix Array

Echo Lab

U of C

Page 4: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Matrix Array: Advanced Technologies

Compressed into

Ergonomic Housing

Compressed into

Ergonomic Housing

Electronics needed

for 3000 Elements

Electronics needed

for 3000 Elements

>15

0B

oard

s

Compressed into

Application Specific Ics

(Next Generation ASICs)

Compressed into

Application Specific Ics

(Next Generation ASICs)

Microelectronics in the Transducer

MTEE

Matrix TTE&TEE Probe: 2007Matrix TTE&TEE Probe: 2007

Sugeng L, Shernan SK, Salgo IS, Weinert L, Shook D, Raman J, Jeevanandam V, DuPont F, Settlemier S, Savord B, Fox J, Mor-Avi V, Lang RM. J Am Coll Cardiol

2008 August 5;52(6):446-449.

Full Array

Page 5: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Courtesy of J. MaaloufCourtesy of J. Maalouf

2D vs 3D Echocardiography

Good 2D Equals Good 3D

Page 6: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Modes of Acquisition

30o

30o

85o

90o

30o

60o

Different Modes of Acquisition

Live 3D Full Volume 3D 3D Zoom

NY Style Pizza Chicago Style Pizza

Page 7: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Narrow Volume

30o

60o

Zoom Mode

30o

30o

Page 8: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Zoom Mode

• Indications:Prolapse

Flail

Perforation

Stenosis

• Less prone to artifacts

• Beware of losing spatial orientation

3D Full Volume Acquisition

Page 9: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Low Frame Rate

High Frame Rate

Page 10: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Number of Sub-volumes = Frame Rate

Depth = Frame Rate

Depth13 cm

Depth16 cm

Page 11: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Uses of 3D Echocardiography

Uses of 3D Echocardiography

Page 12: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Subjective

Experience dependent

Lack of standardization

Large inter- and intra-observer variability

Qualitative

Assessment

Eye ball How do we Assess LV Function ?

Hand tracing

Quantitation of 2D Echocardiography: 2015

Correct view? Foreshortening?

Tracing errors? Correct trace?

Correct shape? Geometry dependent?

Page 13: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

A4C

7

8

9

10

11

2D 3D

long

axi

s (c

m)

*

Mor-Avi V, Lang RM et al., Circulation 2004. 110: 1814-1818.

Why is 3D More Accurate?

Why is 3D More Accurate?

RT3DE volume measurements:

Validation by MRI

• Jacobs LD, et al. Eur Heart J 2005; 27:460-8• Sugeng L, et al. Circulation 2006; 114:654-61• Jenkins C, et al. J Am Soc Echocardiogr 2007; 20:962-8• Soliman OI, et al. Am Soc Echocardiogr 2007; 20:1042-9

EDV, ESV

Excellent correlation(r²>0.85)

but RT3DE underestimates

volumes

Page 14: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Sources of error: Tracing? Latex balloon:

• Mor-Avi V. et al, JACC Cardiovasc Img 2008: 1: 413-423

Human ventricles:

Patient A Patient B

?

True volume: 150 ml

• Tracing error is the most important factor contributing to LV volume underestimation

RV ASSESSMENT: 3D Echocardiography

MuraruD…BadanoLPet.al.EHJCardiovascularImaging2015

MedvedofsyD,AddetiaK…LangRMet.al.JASE2015

r Bias LOA r Bias LOA

EDV 0.92 ‐15mL ±45mL 0.95 ‐11mL ±40mL

ESV 0.93 ‐4mL ±28mL 0.96 ‐0.3mL ±31mL

RVEF 0.86 1.4% ±9.7% 0.83 ‐3.3% ±15%

Generally, an RVEF of <45% represents abnormal RV systolic function

Page 15: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

RV Volumes and EF

<45

RV EDV/BSA, Men 61.3±13 35-87RV EDV/BSA, Women 53±10.5 32-74

RV ESV/BSA, Men 27±8.5 10-44RV ESV/BSA, Women 22±7 8-36

3D of the right ventricle is not easy

Good 

dataset

Poor

dataset

Page 16: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Salgo, Tsang, Lang et al. JASE 2011: 25(1):80-8

LV Curvature

11

1

Rk

22

1

Rk

R1

Curvature: the amount by which a geometric object deviates from being flat.

k =1/r

r

Page 17: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Salgo I, Tsang W, Lang RM et al.,

J Am Soc Echocardiogr 2011: 25(1):80-8

LV Shape AnalysisLV Shape Analysis

Mor-Avi V, Lang RM et al., JACC Cardiovasc Imaging. 2012 Aug;5(8):769-77.

LA VolumesLA Volumes

Page 18: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

3DE2DE vs. 3DE for LA Volume Quantification 

Mor-Avi V ,Lang RM et al.: Real-time 3D echocardiographic quantification of left atrial volume:Multicenter study for validation with magnetic resonance imaging. JACC Imaging 2012.

Standard viewsA4C

3DE-derived viewsA2C

Biplane volume: 82 ml

Atrial-focused views

Biplane volume: 87 ml

A4C A2C

3D volume: 88 ml

Left atrial volume on 2DE

Page 19: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

• Direct measurement without geometrical assumptions about cavity shape and hypertrophy distribution

• More accurate that the linear or the 2D measurements

• Higher inter-measurement and test/retest reproducibility

• Better discriminates small changes within a patient

Normal values less well established

Dependent on image quality

Patients cooperation required

RT3DE measurements of LV mass

RT3DE measurements of LV mass

• Mor-Avi V, et al. Circulation 2004; 110, 1814-1818

A4CA2C

A4CA2C

2D biplane

3D-guided biplane

MRI reference

0

40

80

120

160

LV

mas

s (g

)

2DECMR RT3DE

Page 20: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Uses of 3D EchocardiographyMitral regurgitationMitral 

stenosisMitral prosthesis

AC

B

D

E

F

G

H

I

J

K

The Mitre typically worn by popes and cardinals

Andreas VesaliusDe Humani Corporis

Fabrica1543

Page 21: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Surgeon’s View of the MVSurgeon’s View of the MV

Lang RM, Tsang W, Weinert L, Mor-Avi V, Chandra S. J Am CollCardiol 2011 November 1;5 8(19):1933-1944.

Mitral Stenosis

Page 22: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Mitral Stenosis

PBMV

inM

itralStenosis

Page 23: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

PBMV in Mitral Stenosis

PRE PBMV POST PBMV

Mitral Stenosis

Page 24: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

A2

P1 P2P3 A2

A2 A1A3

P2

P3 P1

APCC

Challenging the Hemispheric Assumption of Flow Convergence

Challenging the Hemispheric Assumption of Flow Convergence

3DE Evaluation of MR

Vena contracta (VC)

Proximal isovelocitysurface area (PISA)

Anatomic regurgitantorifice area (AROA)

Page 25: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

3DE Vena Contracta

• True VC cross-sectional area can be measured on 3D allowing recognition of the non-hemispheric shape of PISA and VC in the majority of patients

• Improves classification of MR in patients with eccentric MR and functional MR

Yosefy et al. Am J. Cardiol 2009:104;978, Kahlert et al. JASE 2008:21;912

Good Agreement: AROA and PISAGood Agreement: AROA and PISA

Page 26: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Worse Agreement: MROA and PISAWorse Agreement: MROA and PISA

3-D Surface

Area: 5.44 cm2

Automatic 3D PISA Surface Area Detection Methods

Courtesy: Mani Vannan

Page 27: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

M-mode

2D Echocardiography

Anterior

Septal

Anterior

Septal

SeptalPosterior

The TV on 3D Echo

S

A PA

S

P

RV perspective RA perspective

Transthoracic 3D EchocardiographyTransthoracic 3D

Echocardiography

Page 28: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

P S Commissure: CorrectPosition

P-S Commissure: Correct Position

Mediratta A, Addetia K, Moss JD, Nayak HM, Yamat M, Weinert L, Mor-Avi V, Lang RM:3D echocardiographic location of implantable device leads and mechanism of associatedtricuspid regurgitation. J Am Coll Cardiol Cardiovasc Img 2014; 7:337-347

Page 29: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

LVOT

AoA SOV

STJ

Elliptical Aortic AnnulusElliptical Aortic Annulus

Page 30: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

BA

C D

Fig

Page 31: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Stenotic Bioprosthetic MVF

rom

LA

Fro

mLV

Mechanical Bileaflet Aortic Valve

Page 32: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

• 38 year old female with a history of mitral stenosis post mechanical MV replacement

Thrombosed Mechanical Prosthesis

Uses of 3D Echocardiography

Page 33: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Dehisced Mitral Valve

Dehisced Mitral Valve

Page 34: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Number of Leaks

Percutaneous Repair of Prosthetic Mitral Valve Dehiscence

Page 35: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Percutaneous Repair of Prosthetic Mitral Valve Dehiscence

Percutaneous Repair of Prosthetic Mitral Valve Dehiscence

Page 36: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Percutaneous Repair of Prosthetic Mitral Valve Dehiscence

Percutaneous Repair of Prosthetic Mitral Valve Dehiscence

Page 37: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Percutaneous Repair of Prosthetic Mitral Valve Dehiscence

Percutaneous Repair of Prosthetic Mitral Valve Dehiscence

Page 38: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Mitral Clip

Steinberg D H et al. Eur Heart J Suppl 2010;12:E2-E9

Inter-atrial Septal

Puncture

Page 39: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Position and Orient Clip

Grasping the Leaflet

Ventricular Side Atrial Side

Page 40: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Alfieri Stich

Little et al. JACC Intervention 2016

Procedural Planning of Mitral Valve Repair with Clip and Plug

3D Printed modeling of MV 3D Printed modeling of MV Procedural PlanningProcedural PlanningClinical

ProcedureClinical

Procedure

MitraClip @ A2/P2ASD occluder within P2

Page 41: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Valve in Valve

Tricuspid Valve in

Ring

Tricuspid Valve in

Ring

Page 42: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Tricuspid Valve in Ring

Tricuspid Valve in Ring

• Displays anatomy intuitively

• Pre-procedure assessment

• Intra-procedure guidance

• Post-procedure follow-up

Bhave and Lang, Atlas of 3D Echocardiography, ed. Gill,

Interventional EchocardiographyInterventional Echocardiography

Page 43: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

FUSION IMAGING

3D Printing

Page 44: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Virtual RealityComputer-generated simulation of a three-dimensional image

that can be interacted with by a person using

special electronic equipment (helmet

with a screen inside or gloves fitted with

sensors.

Virtual Reality Set-Up

Room SensorsHeadset and ControllersComputer

Page 45: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Courtesy RealView Imaging .

3D Holography: Photographic technique that records the light scattered from an object, and then

presents it in a way that appears 3D.

Page 46: 0507-0530-Lang - 3D Echocardiography - Principles …...Real-Time 3D Acquisition: Sparse Array Sparse Array (~300 elements) Volumetrics Courtesy of Dr. Masood Ahmad, UTMB Galveston

Artificial Intelligence

Fusion Imaging

3D Printing

Virtual Reality

Holography

@RobertoMLang