1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Sourc e hn Sample Detecto r Cloc k

1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Embed Size (px)

Citation preview

Page 1: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

1

Ken HansonMWF 9:00 – 9:50 am

Office Hours MWF 10:00-11:00

CHM 5175: Part 2.6Time-resolved emission

Source

hn

Sample Detector

Clock

Page 2: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Steady-state Emission

550 600 650 700 750 800 8500

50000

100000

150000

200000

250000

Inte

nsi

ty

Wavelength (nm)

Intensity vs. WavelengthSource Samplehn

Information about emission intensity (yield) and wavelength.

S0

S1

EnergyConstant Excitation

Constant Emission

Equilibrium between absorption, non-emissive decay and emission.

Non-emissive

decay

hn

Page 3: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time-resolved Emission

Intensity vs. Time

Information about emission lifetimes.

S0

S1

EnergyPulsed

Excitationkrknr

0 200 400 600 800 10000

1000

2000

3000

4000

5000

Inte

ns

ity

Time (ns)

Short Burst of Light

Competition between non-emissive decay and emissive rates.

Source Samplehn

hn

Page 4: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Single Molecule Emission

Excited State Lifetime of an individual molecule: 0 – infinity

Anthracene Excited state Lifetime:Time spent in the excited state (S1) prior to radiative (kr) or non-radiative decay. (kr)

Ex Em

S0

S1

Energy

Time

Ex Em Ex

Page 5: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Ensemble Emission

Time-resolved EmissionIntensity vs. Time

0 200 400 600 800 10000

1000

2000

3000

4000

5000

Inte

ns

ity

Time (ns)

Single Molecule Emission

Excited State Lifetime of an individual molecule: 0 – infinity

Observe many single molecule emission events!

Ex Em

S0

S1

Energy

Time

Ex Em Ex

Page 6: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Ensemble Emission

hn Time 1

64 excited states32 excited states

+ 32 photons

Time 2

Time 3

16 excited states + 16 photons

Time 4

8 excited states + 8 photons

4 excited states + 4 photons

Time 5

etc.

Page 7: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Ensemble Emission

hn Time 1

64 excited states32 excited states

+ 32 photons

Time 2

Time 3

16 excited states + 16 photons

Time 4

8 excited states + 8 photons

4 excited states + 4 photons

Time 5

etc.

0 2 4 6 8 100

10

20

30

40

50

60

70

# E

xcit

ed S

tate

s

Time

0 1 2 3 4 5 6 7 8 9 100

10

20

30

40

Em

issi

on

Inte

nsi

ty

Time

32 photons

16 photons

8 photons

Page 8: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

kr + knr

Excited State Decay Curve

/t*

*

e)0(n

)t(n

0 2 4 6 8 100

10

20

30

40

50

60

70

# E

xcit

ed S

tate

s

Time

n*(0) is the # of the excited state at time 0

n*(t) is the # of the excited state at time t

t is the lifetime of the excited state

S0

S1

EnergyPulsed

Excitationkrknr

t =1

We don’t get to count the number of excited state molecules!

Page 9: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Intensity Decay Curve

I(0) is the initial intensity at time zero

I(t) is the intensity at time t

t is the lifetime of the excited state0 1 2 3 4 5 6 7 8 9 10

0

10

20

30

40

Em

issi

on

Inte

nsi

ty

Time

= e-t/t

kr + knr

t =1

t = time it takes for 63.2 % of excited states to decay

t should always be the same for a given molecule under the same conditions

0 200 400 600 800 10000

1000

2000

3000

4000

5000

Inte

nsi

ty

Time (ns)

I(t)

I(0)

Page 10: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

time

intensity

1.00 --

1/e

Exciting pulse

Emission

time

Log intensity

Exciting pulse

Emission

Intensity Decay Curve

Linear Scale Log Scale

= e-t/tI(t)

I(0)

Page 11: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Spectra Decay

= e-t/tI(t)

I(0)

inte

nsity

Page 12: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Why do we care about lifetimes?• Electron transfer rates

• Energy transfer rates

• Distance dependence

• Distinguish static and dynamic quenching

• Fluorescence resonance energy transfer (FRET)

• Track solvation dynamics

• Rotational dynamics

• Measure local friction (microviscosity)

• Track chemical reactions

• kr and knr (if you know F)

• GFP- Nobel prize, expression studies

• Sensing

Page 13: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Lifetime Measurements

Inte

nsity

time

Light source

Time Domain

Pulsed MethodHarmonic or phase-modulation method

Frequency Domain

time

Inte

nsity

Light source

Source Samplehn

hn

Page 14: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Samplehn

hn

hn h

n

Low I0 Excitation

High I0 Excitation

hnhn

Low I0 Excitation

Time

Frequency-domain Method

I0

Measure Events with Respect to Frequency

Page 15: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Frequency-domain Method

Page 16: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Frequency-domain Method

Excitation Modulation = ab

a = average intensity

b = average-to-peak intensity

Emission Modulation = AB

A = average intensity

B = average-to-peak intensity

Modulation (m) =

(B/A)(b/a)

Phase Shift (f)

Page 17: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Frequency-domain Method

Modulation (m)

Phase Shift (f)

Ex Frequency ()

2/121 ]1)/1[( mm

tan1

Changing , measuring m and to calculate lifetime.

Phase (τφ) and modulation (τm) lifetimes

Page 18: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

tan12/121 ]1)/1[( mm

Frequency-domain Method

Page 19: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Frequency-domain Method

• Lifetimes as short as 10 picoseconds • Can be measured with a continuous source• Tunable from the UV to the near-IR• Frequency domain is usually faster than time domain (same source)

Page 20: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Frequency-domain Method

2/121 ]1)/1[( mm

tan1

f

m

Modulation (m)

Phase Shift (f)

Ex Frequency ()

Page 21: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Frequency-domain Instrument

Page 22: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Frequency-domain Method

List of Commercially Available Frequency-domain Instruments

Page 23: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Lifetime Measurements

Inte

nsity

time

Light source

Time Domain

Pulsed MethodHarmonic or phase-modulation method

Frequency Domain

time

Inte

nsity

Light source

Source Samplehn

hn

Page 24: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Inte

nsity

time

Light source

Time-Domain Method

• Pulsed method• Lifetimes as short as 50 fs• Multiple measurement techniques• Sources typically not as tunable as frequency domain

Emission

Emission intensity is measured following a short excitation pulse

Measure Events with Respect to Time

Page 25: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time-domain Techniques

1 s1 ms1 ms

1 ns1 ps1 fs

secondsmillimicronanopicofemto0.001 s0.000001 s0.000 000 001 s

0.000 000 000 001 s0.000 000 000 000 001 s

1 s

Excitation

PhosphorescenceFluorescence

Internal Conversion

Intersystem Crossing

Page 26: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

TCSPC

Time-domain Techniques

1 s1 ms1 ms

1 ns1 ps1 fs

Streak Camera MCSStrobeUp-conversion

Real-time Measurement

Page 27: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time-domain Techniques

1. Real-Time lifetime measurement (t > 200 ps)

2. Multi-channel scaler/photon counter (t > 1 ns)

3. Strobe –Technique (t > 250 ps)

4. Time-correlated single-photon counting ( t > 20 ps)

5. Streak-camera measurements (t > 2 ps)

6. Fluorescence up-conversion (t > 150 fs)

Page 28: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Real-Time Lifetime

hn

Page 29: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Real-Time Lifetime

Source

hn

Sample Monochromator

Detector

Clock

(1) (2)

(3)

(4)

1) Pulsed excitation

2) Sample excitation/emission

3) Monochromator

4) Detector signal

5) Plot Signal vs. Time

Page 30: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Real-Time Lifetime

Light source D

etec

tor C

urre

nt

time

Emission

SourcesFlashlampLaserPulsed LED

Page 31: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Real-Time Lifetime

• Make excitation pulse width as short as possible• Time resolution is usually detector dependent• Excited-state lifetime > IRF• Lifetimes > 200 ps

Instrument Response Function (IRF)

Det

ecto

r Cur

rent

time

Emission

Page 32: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Real-Time Lifetime

0 200 400 600 800 10000

1000

2000

3000

4000

5000

Inte

nsi

ty

Time (ns)

100 averages

Page 33: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Strobe-Technique

25 images per second

Page 34: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Strobe-Technique

Photon Technology International (PTI)

Page 35: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Strobe-Technique

time

Light PulseMeasurement Window

time

Light PulseMeasurement Window

Page 36: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Strobe-Technique

time

Light Pulse

Measurement Window

time

time

Det

ecto

r Si

gnal

Page 37: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Strobe-TechniqueStrobe-Technique TCSPC

“Full decay curve is attainable after just one sweep (100 pulses)”

“TCSPC: for every 100 pulses, you get only up to three useful points”

“The Strobe technique is much faster than the TCSPC technique for generating the decay curve. This is particularly important in the life science area. Whereas the chemist can take hours or days to measure an inert chemical very accurately, the life scientists’ cell samples are long dead. “

Lower Time Resolution

Page 38: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

(1) (2)

(3) (4)

(5)

Strobe-Technique

1) Trigger Signal

2) Excitation Flash

3) Detector Signal Delay

4) Detect

5) Output t > 250 ps

Page 39: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time-Correlated Single-Photon Counting (TCSPC)

Excited State Lifetime of an individual molecule: 0 – infinity

The sum an individual molecule lifetimes = t

Ex Em

S0

S1

Energy

Time

Ex Em Ex

Page 40: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Low excitation intensity:

- Low number of excited state

- 20-100 pulses before emission is detected

- Only one or 0 photons detected per pulse

- Simulated single molecule imaging

Time-Correlated Single-Photon Counting (TCSPC)

Time

Page 41: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

1) Pulsed source “starts” the timing electronics

2) Timer “stopped” by a signal from the detector

3) The difference between start and stop is sorted into “bins.” -Bins are defined by a Dt after pulse at t = 0

Time

Detector Bins

Time-Correlated Single-Photon Counting (TCSPC)

Page 42: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time-Correlated Single-Photon Counting (TCSPC)

Time

Detector Bins

Sum the Photons per Bin

Page 43: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time-Correlated Single-Photon Counting (TCSPC)

Probability Distribution

Repeat

Page 44: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Excitation Pulse

Time-Correlated Single-Photon Counting (TCSPC)

Page 45: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Repeat: 10,000 counts in the peak channel

Time-Correlated Single-Photon Counting (TCSPC)

Page 46: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time-Correlated Single-Photon CountingSource:

Flash lampsolid state LED laser

Start PMT

Stop PMTsample

exc. monochromator

emis

sio

n

mo

no

chro

mat

or

pulsed source

t

1) Pulsed excitation (10kHz)

2) Monochromator

3) Beam Splitter

1) to trigger PMT

2) to sample

4) Excite Sample

5) Sample emits into monochromator

6) Emission hits PMT and timer stops

7) Repeat a million times

(1)

(3)

(2)

(4)(5)

(6)

Page 47: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

constant function discriminator (CFD)time-to-amplitude converter (TAC)programmable gain amplifier (PGA)analog-to-digital converter (ADC)

TCSPC

1) Pulsed excitation2) Ex CFD triggers TAC3) TAC voltage rises4) Em CFD stops TAC5) TAC discharges to PGA6) PGA siganl to ADC for a single data point

Page 48: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

48

TCSPC

Page 49: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Advantages:– High sensitivity

– Large dynamic range (3-5 decades)

– Well defined statistics

– Temporal resolution down to 20 ps

– Very sensitive (low emission materials)

– Time resolution limited by detector

– Price as low as $15 K

Disadvantages:– “Long” time to acquire data

– Complicated electronics

– Stray light

– Lifetimes < 10 ms

– Resolution vs. acquisition time

TCSPC

Molecule with a 10 ms lifetime• 10,000 peak counts• 1024 bins for a 20 ms window• Total counts = 4,422,800• 20 ms rep rate• 1 count per 20 reps= 20.5 day measurement

Page 50: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Acquisition Time

Time

Detector Bins

Resolution vs. Acquisition Time

5 ns wide bin = 5 ns resolution10 minutes to acquire 10,000 counts

Time

Detector Bins

1 ns wide bin = 1 ns resolution50 minutes to acquire 10,000 counts

Resolution

Acquisition TimeResolution

Page 51: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time

Repetition Rate to High

hn

hn

Real start-stop-time

Page 52: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Sign

al

time

Repetition Rate to High

If the rep rate is too high the histogram is biased to shorter times!

Measured t < Real t

Keep rep rate at least 10 times slower than your t

Page 53: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Stop count rate < 2% of the excitation rate.

Limited number of emitted photons. Failure to do so can lead to a biasing towards detection of photons arriving at shorter times, a phenomenon known as pulse pile up.

Intensity to High

Single Photon Counting only counts the first photon!

Page 54: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Photoelectric Effect

Photon Energy - binding energy = electron kinetic energy

Side Note: PMT Lifetime

Page 55: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Side Note: PMT LifetimePhotoelectric Effect

Photon Energy - binding energy = electron kinetic energy

Higher Energy Photons = Faster Signal

Measured Lifetime < Real Lifetime

Page 56: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Streak-CameraTemporal profile from Spatial profile

Laser Pointer Duty Cycle Calculating Duty Cycle

Pointer Motionm/s

Dist

ance Length

(spatial)

Use length to calculate time

Page 57: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Streak-CameraCathode Ray Tube

e-

+

-

Page 58: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Streak-Camera

(2)(1)

1) Light hits cathode (ejects e-)

2) Voltage sweep from low to high

3) e- hits MCP-Phosphor Screen

4) Emitted photos hit CCD detector

Source

hn

Sample Monochromator

(3)

(4)

Page 59: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Calculating Duty Cycle

Pointer Motionm/s

Dist

ance Length

(spatial)

Use length to calculate time

Streak-Camera

Sweep Ratem/s

Length

Use length and intensity to calculate lifetimee-

time(0) time(t)

0 200 400 600 800 10000

1000

2000

3000

4000

5000

Inte

ns

ity

Time (ns)

+-

Intensity

Page 60: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Streak-Camera

(2)(1)

1) Light hits cathode (ejects e-)

2) Voltage sweep from low to high

3) e- hits MCP-Phosphor Screen

4) Emitted photos hit CCD detector

Source

hn

Sample Monochromator

(3)

(4)

Page 61: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Electrons that arrive first hit the detector at a different position compared to electrons that arrive later.

Streak-Camera

Page 62: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Streak-Camera

Page 63: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

http://www.youtube.com/watch?v=rA6A7haKFwI

Streak-Camera

Page 64: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

• Advantages:– Direct two-dimensional resolution– Sensitivity down to single photon– Very productive– Not detector limited (like TCSPC)

• Disadvantage: – Depends on high stability of laser– Limited time resolution: 2-10 ps– Needs careful and frequent calibration– Expensive

Streak-Camera

Page 65: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Streak-Camera

Time resolution down to 2ps or even 100s of femtoseconds.

TCSPC

Instrument Response Functions

Page 66: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Fluorescence up-conversion

Sum Frequency Method

ωsum = ω1 + ω2

Page 67: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Fluorescence up-conversion

1) Excitation pulse/gate pulse

2) Sample is excited

3) Sample Emission

4) Emission and Gate are collinear

5) NLO crystal sums Emission and Gate

6) Only Summed Light is measured

(1)

(4)

(2)

excitation beamgate beam

(1)

(3)

(5)(6)

Page 68: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Fluorescence up-conversionExcitation pulse

Emission

Inte

nsity

Excitation pulse

Inte

nsity

Gatepulse

td1

Inte

nsity

Summed Light at time 1

time

time timeExcitation

pulse

Inte

nsity

Gatepulse

td2

Inte

nsity Summed Light

at time 2

time time

Inte

nsity

time

Control td and measure only summed light

Graph of td vs intensity

Page 69: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Fluorescence up-conversion

1) Excitation pulse/gate pulse

2) Sample is excited

3) Sample Emission

4) Emission and Gate are collinear

5) NLO crystal sums Emission and Gate

6) Only Summed Light is measured

(1)

(4)

(2)

excitation beamgate beam

(1)

(3)

(5)(6)

Signal is only measured when gate is pulsed

td is controlled by the delay track

Light Travels 0.9 m in 1 ns

Page 70: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

ComparisonIn

tens

ity

time

Control td and measure only summed light

Detector is not time resolved (left open).Not limited by detector speed.Data point limited by pulse width (fs)

Sum Frequency Generation TCSPC

Detector Bins

Inte

nsity

time

Limited by detector response.Data point limited by PMT (10 ps)

Control excitation measure td

Page 71: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Fluorescence up-conversion

Page 72: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Phys . Chem. Chem. Phys. 2005, 7, 1716 – 1725.

Fluorescence up-conversion

Page 73: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Fluorescence up-conversion

Page 74: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

74

• Advantage: – (very) high time resolution, limited mainly by laser pulse duration

• Disadvantages:– Demanding in alignment– Limited sensitivity, decreasing with increasing time resolution

(crystal thickness)– Required signal calibration

Fluorescence up-conversion

Page 75: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Decay Fitting

Exponential decay

= e-t/tI(t)

I(0)

Non-exponential decay

Page 76: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Non-exponential Decay (Log)

Exponential decay Non-exponential decay

Time Time

Inte

nsity

= e-t/tI(t)

I(0)

Inte

nsity

Page 77: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Non-exponential

Possible explanations:

- Two or more emitters

- In homogeneous samples (QDs)

- Dual Emission

- Multiple emissive sites

On surfaces

Polymer Films

Peptides

Dual Emission

Page 78: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Intensity

t / ns

Log

I

t / ns

5 ns

50 ns

Non-exponential Decay

= A1e-t/ 1t + A2e-t/ 2tI(t)

I(0)

A1 = amplitude of component 1t1 = lifetime of component 1A2 = amplitude of component 2t2 = lifetime of component 2

Linear Scale

Log Scale

Biexponential Fit

Page 79: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Non-exponential Decay

= A1e-t/ 1t + A2e-t/ 2tI(t)

I(0)

Page 80: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Limitations of Multi-exponential Fits

Linear Scale: No differenceLog Scale: minor differences at 30–50 ns

At 50 ns there are only about 3 photons per channel with a 1-ns width. The difference between the two decays at long times is just 1–2 photons.

Biexponential Fits

t1 = 5.5 ns and t2 = 8.0 nsor

t1 = 4.5 ns and t2 = 6.7 ns

Page 81: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Fitting Data

Page 82: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

y = A1e-k1t + A e-k2t + A3e-k3t

c2 = 26.466

y = A1e-k1t + A e-k2t

c2 = 2.133

The Data Exponential

Bi-exponential Tri-exponential

c2 = 1.194

Multi-exponential Fits

y = A1e-k1t

Page 83: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

J. of Political Economy 2005, 113, 949

It could be worse!

Page 84: 1 Ken Hanson MWF 9:00 – 9:50 am Office Hours MWF 10:00-11:00 CHM 5175: Part 2.6 Time-resolved emission Source h Sample Detector Clock

Time-resolved Emission End

Any Questions?