24
1 NaI calibration NaI calibration and neutron neutron observation observation during the charge exchange experiment 1. Improving the NaI energy resolution (as low as reasonably achievable!) Common noise reduction Intercalibration Clustering algorithm 2. Observation of the prompt signal of “high energy” neutrons (8.9 MeV) A “matter-of-fact” evidence, in Xe and in NaI Comparison with cross sections First look at and requirements of a MC for neutrons in LXe. Giovanni Signorelli, INFN Pisa MEG collaboration meeting, PSI 9 Feb 2004

1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

Embed Size (px)

Citation preview

Page 1: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

1

NaI calibrationNaI calibration and neutron observationneutron observation during the charge exchange experiment

1. Improving the NaI energy resolution (as low as reasonably achievable!)• Common noise reduction • Intercalibration• Clustering algorithm

2. Observation of the prompt signal of “high energy” neutrons (8.9 MeV)• A “matter-of-fact” evidence, in Xe and in NaI• Comparison with cross sections• First look at and requirements of a MC for neutrons in LXe.

Giovanni Signorelli, INFN Pisa MEG collaboration meeting, PSI 9 Feb 2004

Page 2: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

2

1. Common noise reduction

2. Crystal intercalibration

3. Clustering for energy summation

NaI calibration procedure

23% 11% FWHM @ 55 MeV

Page 3: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

3

Common noise reduction

Algorithm:Simplif. From E.Frlez, D. Pocanic, S.Ritt NIM A463 (2001)

1. Take the ADC of the channels which see pedestal

2. Make the average

3. Subtract it from all channels (second pedestal correction)

The pedestal ’s shrink from 56 to 23.It’s not perfect but compatible with the ALARA principle

Correlation between channels due to electronics, noise in cables, ADCs…

Page 4: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

4

Crystal intercalibration

ROUGH CALIBRATION

•Cosmic ray runs can be used to inter-calibrate crystals

•Muons triggered by crystal pairs

•Position of the Landau peak

FINE TUNING

•Problems for crystals at the center (the crystal are not uniformly spanned by cosmics?)

•Refined with monoenergetic gammas

Page 5: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

5

Energy clustering

The cluster C includes the element of the detector with the maximum energy plus all the fired elements connected to another member of the cluster by a side or a corner

E = EiiC

Page 6: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

6

Results

•The resolution is acceptable

•The peak position is well reproduced

54.8 MeV 83 MeV 129.8 MeV

5.5% 5.1% 4.9%

Reconstructed peak

Page 7: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

7

A better NaI helps

•A cleaner separation of the two NaI peaks helps in reducing the tails on the Lxe distributions

•An improved collinearity requirement shows the real performance

Page 8: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

8

Neutron observation during the experiment

•Evidence for a prompt signal from neutrons

•8.9 MeV neutron in coincidence with the 129 MeV gamma

•Neutrons from the Am/Be source (10 MeV)

•Comparison with cross sections (physics)

•Inelastic scattering

•Xe level excitation

•First look at and requirements of a MC for neutrons in LXe.

•Geant 3.21 + GCALOR

•Geant 4

•Possible use of neutrons for calibration/monitoring purposes (Angela)

•Availability – switchability

•Probe of the entire detector

Page 9: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

9

Evidence

•Runs triggered with one of the detectors only (&S1 &RF…)

•Emeasured> 110 MeV selection of the - p n events

•No timing cut (implies an energy/position cut!)

XeXe NaINaI

50% efficiency

Page 10: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

10

Neutron-induced prompt signal in Xe

For fast neutrons (110 MeV) the total and scattering cross sections are similar for all isotopes

= 1 barn = 72 cm in LXe

Page 11: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

11

Neutron cross section

Page 12: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

12

Processes

A COMPLETE MONTECARLO CALCULATION IS NEEDED FOR COMPUTING THE NEUTRON EFFECTS IN THE CALORIMETER :• efficiency for fast and thermal neutron detection• determination of the energy spectrum in the calorimeter• energy released as a function of time• energy density (x,y,z)• dependence on threshold and n-energy

ALL THE RELEVANT NEUTRON CROSS-SECTIONS CAN BE INCLUDED IN GEANT 3.21 AND ARE INCLUDED IN GEANT4

information from medical physics…….!KERMA COEFF. (Kinetic Energy Released per unit Mass) and

tr / (mass energy transfer coefficient)tabulated for neutrons

Page 13: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

13

MC for neutrons in liquid Xenon

Though the most reliable simulation today is GEANT4, some quick results were obtained with GEANT 3.21 + GCALOR

•8.9 MeV neutron simulated impinging a 10 x 10 cm2 window of the Lproto (time cut-off at 600ns) coming from the LH2 target

•GCALOR (MICAP, En < 20 MeV) takes care of n cross sections (ENDF VI B)

•N,n n,2n …

•If the residual nucleus is left in an excited state the deexcitation photon is generated (this is not done in the n,Xe n’Xe case. Bug? We generated these photons by hand)

•Some refinement still possible

•In GEANT4 the code for the neutron transportation is automatically embedded in the package and is “benchmarked” with a comparison to real data!

Page 14: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

14

Neutron Monte Carlo event sample

Incoming neutron

•8.9 MeV neutron

•10 x 10 cm2 window

•Coincidence with the 129 MeV photon

Page 15: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

15

MC spectrum

2.2 MeV capture on protons

Xe levels

•A neutron edge is present

•Low energy lines due to Xe and/or other nuclear levels

•High energy tails: n capture and isotope production

•The comparison with the data is good but not excellent

Page 16: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

16

Conclusion

• A calibration procedure for the NaI has been estabilished and coded in the (Pisa version of the) analyzer, obtaining a fairly good E resolution for this detector

• The neutron prompt signal was identified in Xe and NaI and the understanding of the process is under way. We’ll do our best to reproduce the experimental result…

• A new window is open, a new handle is present. To us the difficult task to exploit it (calibration, monitoring…)!

Page 17: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

17

…timing

Page 18: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

18

neutron

gamma

n2n

n 1 n 2

n

Page 19: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

19

Xe 129

TOT SC

Page 20: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

20

Xe 129

n2n n3n

Initial energy degradation and neutron duplication

Page 21: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

21

Xe 129 nuclear level excitations n1 n2 etc.Levels 0.039 0.236 0.318 MeV

energy degradation and kinetic energy into energy

Page 22: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

22

Xe 129 n

Page 23: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

23

Xe 132nuclear level excitations n1 n2 etc.

Levels 0.628 0.1.298 01.44 MeV

Page 24: 1 NaI calibrationneutron observation NaI calibration and neutron observation during the charge exchange experiment 1.Improving the NaI energy resolution

24

Xe 132 n