10/11/2005 1 ENGINEERING RESEARCH CENTER FOR STRUCTURED ORGANIC PARTICULATE SYSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY UNIVERSITY OF PUERTO RICO AT MAYAGÜEZ Vibrational Spectroscopy Vibrational Spectroscopy for Pharmaceutical Analysis for Pharmaceutical Analysis Part V. Methods for Obtaining Mid-IR spectra of Samples. Rodolfo J. Romañach, Ph.D.

10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

Embed Size (px)

Citation preview

Page 1: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

10/11/2005

1

ENGINEERING RESEARCH CENTER FOR

STRUCTURED ORGANIC PARTICULATE SYSTEMS

RUTGERS UNIVERSITYPURDUE UNIVERSITYNEW JERSEY INSTITUTE OF TECHNOLOGYUNIVERSITY OF PUERTO RICO AT MAYAGÜEZ

Vibrational Spectroscopy for Vibrational Spectroscopy for Pharmaceutical Analysis Pharmaceutical Analysis

Part V. Methods for Obtaining Mid-IR spectra of Samples.

Rodolfo J. Romañach, Ph.D.

Page 2: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

2

Working with LiquidsWorking with Liquids

Spectra of Solute are not as simple as in Ultraviolet region because solvents absorb IR radiation.

Some solvents have spectral windows, CCl4 from 4000 - 1350 cm-1 and CS2 from 1350 - 400 cm-1.

Hexane and iso-octane may also be used but with narrower spectral windows.

Page 3: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

3

Working with LiquidsWorking with Liquids

Capillary film method (like closing a door). Adjustable pathlength cells are available. Fixed pathlength cells. For quantitative work, thickness should such that

the transmission is between 20 - 80% (absorbance is 0.1 - 0.7).

Page 4: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

4

Page 5: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

5

Sealed (Fixed Pathlength) CellsSealed (Fixed Pathlength) Cells

Sealed cells should not be opened, should be left together.

If cell thicker than 0.1 mm, hold the cell at a slightly inclined angle(pencil underneath) and press gently on the syringe to fill cell.

Page 6: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

6

Flow through cell to be used with luer lock Flow through cell to be used with luer lock syringe.syringe.

Sealed Transmission Cell

Page 7: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

7

Sealed Cells(thicker than 0.1 mm)Sealed Cells(thicker than 0.1 mm)

Hold at an Inclined Angle. Place empty syringe at upper port, and the syringe with

sample at lower port. Fill the cell by withdrawing liquid with the empty syringe. Excess liquid should be removed from the filling ports. The plugs should be placed in the ports and rotated to aid in

producing a tight seal between the plug and cell.

Page 8: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

8

Working with Solids (Mulls)Working with Solids (Mulls)

Powdered crystalline solids transmit very little IR radiation, for the light will be scattered by multiple reflections of the particles.

The technique of dispersing a solid in a liquid can be used to reduce the amount of light scattered from the solid.

Page 9: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

9

Working with Solids (Mulls) (2)Working with Solids (Mulls) (2)

Scattering will be reduced by reducing particle size and sorrounding particles with a medium whose refractive index closely matches theirs.

Want to reduce the particle size of the solids to less than 2 m, when preparing mulls.

Page 10: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

10

Mulls (Technique)Mulls (Technique)

Use a large mortar and pestle to put enough pressure on sample to grind it.

Place about 25 mg of material in mortar. After crushing, grind with a circular motion until it cakes on

side of mortar and forms a glossy surface reflecting visible light.

Add a small drop of mulling fluid to mortar, mix to form a paste. With rubber policeman to place as thin film between cells.

Page 11: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

11

Mulls (Technique) (2)Mulls (Technique) (2)

Use mineral oil (1350 - 400 cm-1) or fluorolube (4000 - 1350 cm-1).

The compound will not be reduced after addition of mulling agent, so it is important to grind well previously.

Page 12: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

12

Page 13: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

13

KBr disk-AdvantagesKBr disk-Advantages

Spectra are obtained which are free from interfering bands.

Less light is scattered. Pellets can be stored for some time.

Page 14: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

14

KBr disk-disadvantagesKBr disk-disadvantages

Some water always remains as an impurity. Spectra of some samples do not agree with those

measured in other media. Physical or chemical changes can occur in the

sample during the grinding or pressing process. For ionic samples, exchanges of the halide with

sample ions can occur.

Page 15: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

15

KBr disk-ProcedureKBr disk-Procedure

First grind the sample to about 2 - 3 m particle size.

About 3 mg of sample is mixed gently with 300 - 350 mg of spectral quality KBr.

KBr should be placed in drying oven at 1100C about once a week and dried overnight. A vacuum oven is even better.

Page 16: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

16

Making a KBr PelletMaking a KBr Pellet

Page 17: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

17

KBr pellet holderKBr pellet holder

Page 18: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

18

Modern ATR AccesoriesModern ATR Accesories

MIRacleTM Single Reflection ATR (left), HATR Multi-Reflection ATR (right). Available from Pike Techonology, (www.piketech.com)

Page 19: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

19

Attenuated Total Reflectance (ATR)Attenuated Total Reflectance (ATR)

• An ATR crystal (crystal with a high refractive index) is in contact with sample.

• The radiation is focused on the crystal and total internal reflection occurs at the crystal, but part of the electric field (evanescent wave) penetrates into sample at the point of reflection.

• The sample will absorb some of the radiation, and thus a transmission spectrum is obtained.

Page 20: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

20

ATR CrystalATR Crystal

• The ATR Crystal (also called internal reflection element) is made with materials that have a high refractive index, such as ZnSe, Si, Ge, and diamond.

• Some of these accessories are configured to allow the radiation to bounce several times against crystal (increase sensitivity as radiation penetrates sample several times).

Page 21: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

21

ATR - AdvantagesATR - Advantages

• Non-destructive method of analysis. • May be used for samples that are too thick and are not

suitable for transmission analysis (the radiation cannot pass through a sample that is too thick). Allows analysis of fibers, pastes, powders, viscous liquids, papers, rubber, coatings, adhesives.

• May be used to obtain spectra in aqueous solutions (short pathlength limits the absorbance of O-H bands).

• Good method for analysis of surfaces as the radiation penetrates only a few micrometers into sample.

Page 22: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

22

ATR Spectra vs. Transmission SpectraATR Spectra vs. Transmission Spectra

• ATR and transmission spectra resemble each other, but there will be some differences because the pathlength is not uniform in ATR.

Page 23: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

23

Gas CellsGas Cells

• The molecules are dispersed, but the radiation is reflected several times through the cell.

• A 20 m cell path-length may be obtained with a cell cavity of about 0.5 m in length.

Pike Technologies

J.M. Chalmers and G. Dent, Industrial Analysis with Vibrational Spectroscopy, The Royal Society of Chemistry 1997, page 141 – 142.

Page 24: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

24

Diffuse ReflectanceDiffuse Reflectance (DRIFTS) - (DRIFTS) - Reflectance is termed diffuse where the Reflectance is termed diffuse where the

angle of reflected light is independent of the incident angleangle of reflected light is independent of the incident angle

Spectra Affected by:

Particle size of sample.

Packing density of sample, and pressure on sample.

Refractive index of sample.

Crystalline form of sample.

Absorption coefficients of sample.

Characteristics of the sample’s surface.

J.M. Chalmers and G. Dent, “Industrial Analysis with Vibrational Spectroscopy”, Royal Society of Chemistry, 1997, pages 153 -162.

Page 25: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

25

Diffuse Reflectance Diffuse Reflectance

• For mid-IR spectroscopy diffuse reflectance spectra normally require grinding of the sample and KCl powder. The analyte is usually diluted to about 10% (wt/wt) in KCl.

• Dilution is not necessary in NIR spectroscopy.

Pike Technologies

J.M. Chalmers and G. Dent, “Industrial Analysis with Vibrational Spectroscopy”, Royal Society of Chemistry, 1997, pages 153 -162.

Page 26: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

26

Infrared MicroscopyInfrared Microscopy

• Infrared Microscopy may be used to focus beam on a small sample area.

• A spectrum may be obtained for a sample as small as 10 μm. Slide Courtesy Smiths Detection

Page 27: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

27

OpticsOptics

ATR objective

Slide Courtesy Smiths Detection

Page 28: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

28

Spectrum obtained may be compared to spectra in library.

Slide Courtesy Smiths Detection

Page 29: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

29

Validation of an IR MethodValidation of an IR Method

Validated a method to identify desiccant raw material as an alternate method to XRF spectroscopy (USP method).

Selected several lots of the desiccant which had been previously identified as desiccant according USP procedure.

Page 30: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

30

Validation of an IR MethodValidation of an IR Method

Obtained 3 spectra from each batch of the desiccant. Removed desiccant from capsule, milled with mortar and

pestle. Prepared KBr pellet. Checked interbatch variation of spectra. Prepared procedure for KBr pellet sample preparation. Evaluated analyst to analyst variation by having two other

analysts prepare samples. Checked method with other batches (validation batches).

Page 31: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

31

Caution with KBrCaution with KBr

Very hygroscopic. If you do not protect the pellet, the OH stretch

band will grow. OH band could become the strongest band in the

spectrum.

Page 32: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

32Desiccant spectrum – IR method developed.

Page 33: 10/11/2005 1 ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY

33

Assay Validation RequirementsAssay Validation Requirements

Studies ID Impurities Final Product Dissolution, CU

Accuracy - + +Repeatability + + +

Interm. Precision + + +Specificity (2) + + +

Detection Limit - + -Quantitation Limit - + -

Linearity - + +Range - + +