SpecificationsSpecifications & features are subject to change without notice. All specifications at ambient of 25 °C, unless specified otherwise.
Application areas Performance
CAT#367R2/P1/3/
www.radix.co.in
Head Mount, 2-wire, Isolated,with HART protocol
PERFORMANCE CHARACTERISTICSResponse time 1 sReference operating Calibration temperature : conditions 23 °C (73.4 °F) ±5KLong term stability £ 0.05% / yearSwitch on delay £ 5sAmbient temperature Negligible effect Load effect NegligiblePower supply effect NegligibleResolution 0.3 µAMaximum measured error See Table 2 ENVIRONMENTAL CONDITIONSAmbient temperature -40 to 85 °C (-40 to 185 °F) limits -20 to 70 °C (-4 to 158 °F)Storage temperature -40 to 100 °C (-40 to 212 °F)Condensation AllowableDegree of protection IP20Shock and vibration 4g/2 to 150 Hz as per IEC 60 068-26 resistanceElectromagnetic Interference immunity and interference compatibility(EMC) emission according to GB/T17626.2-
1998), compliance with IEC 61000-4-3:1995
MOC PlasticWeight Approx. 150g
OTHER
INPUT
OUTPUTS
ISOLATION
POWER SUPPLY
Input type Thermocouple B, E, J, K, N, R, S, T
RTD Pt100, Pt500, Pt1000, Cu50, Cu100Resistance 0~400 W, 0~2000 W, 0~10000 W Linear voltage -10~75 mV, -100~100 mV,
-100~500 mV, -100~2000 mV Range limits See Table 1
Output signal 4~20 mALoad
Input/output relation Temperature linear, resistance linear, voltage linear
Mutual isolation between 2000 V AC RMS, 50 hz / 1 minute input, output
Supply voltage U = 7.5 to 45 VDCb
(reverse polarity protected) U = 10.5 to 45 VDCb
MONITORINGSensor break detection, Upscale ~ 20.8 mA, selectable Downscale ~ 3 .8 mA
Max. (U - 7.5V)/0.0208A b
Max. (U - 10.5V)/0.0208A b
· Inputs : Resistance thermometers (RTD),thermocouples (TC), resistance and voltage
· Input selection : Via HART protocol
· High performance, high reliability
· Digital communication
· Self-diagnostics function
· Backlit LCD display with bargraph
SCC6312-wire, HART Temperature Transmitter
Head mounted temperature transmitter with HART protocol for converting various input signals into a scalable 4 to 20 mA analogue output signal
Inputs : Resistance thermometers(RTD)thermocouples (TC)ResistanceVoltage
Ÿ Universal settings with HART protocol for various input signalsŸ 2-wire technology, 4~20 mA analog outputŸ High accuracy in total ambient temperature rangeŸ Galvanic isolationŸ An internal temperature sensor for active temperature compensationŸ Wide voltage supply rangeŸ Customer specific measurement range settingsŸ Multiparametric backlight rotatable LCD displayŸ Expanded resistance input (max 10 KW)Ÿ Expanded voltage input (max 2000 mV)Ÿ No external power supply for configuration with USB interfaceŸ Simple and user friendly software
Inputs TypeMinimum
Measurement Ranges
Pt100Pt500Pt1000
*Ni100*Ni500*Ni1000
B ( P tRh30-PtRh6)E ( N iCr-CuNi)J ( F e-CuNi)K ( N iCr-Ni)N ( N iCrSi-NiSi)R ( P tRh13-Pt)S ( P tRh10-Pt)T ( C u-CuNi)
Resistance
Resistancethermometer(RTD)
Thermocouples(TC)
Voltage Millivolt
Cu50Cu100
Measurement ranges
10K10K10K
10K10K
10K10K10K
500K50K50K50K50K500K500K50K
-10 to 75mV -100 to 100mV -100 to 500mV -100 to 2000mV
5mV5mV6mV20mV
Maximum measured error
Type Measurement accuracy
Resistancethermometer RTD
Thermocouple TC
Pt100, Ni100Pt500, Ni500Pt1000, Ni1000Cu50Cu100
typ.0.5K or 0.08%typ.1.0K or 0.08%typ.2.0K or 0.08%
K, J, T, ENS, B, R
0.2K or 0.08%0.5K or 0.20%0.3K or 0.12%0.2K or 0.08%0.3K or 0.12%
Resistance
Voltage
-10 to 75mV -100 to 100mV -100 to 500mV-100 to 2000mV
SCC6312-wire, HART Temperature Transmitter
Table 1
Table 2
CAT#367R2/P2/3/
www.radix.co.in
-200 °C to 850 °C (-328 °F to 1562 °F)-200 °C to 250 °C (-328 °F to 482 °F)-200 °C to 250 °C (-328 °F to 482 °F)
-60 °C to 180 °C (-76 °F to 356 °F)-60 °C to 180 °C (-76 °F to 356 °F)-60 °C to 150 °C (-76 °F to 302 °F)
-50 °C to 150 °C (-58 °F to 302 °F)-50 °C to 150 °C (-58 °F to 302 °F)
0 to 400 W0 to 2000 W
± 0.1 W or 0.08%± 1.5 W or 0.12%± 7.5 W or 0.20%
0 to 1820 °C (32 to 3308 °F)-270 to 1000 °C (-454 to 1832 °F)-210 to 1200 °C (-346 to 2192 °F)-270 to 1372 °C (-454 to 2501 °F)-270 to 1300 °C (-454 to 2372 °F)-50 to 1768 °C (-58 to 3214.4 °F)-50 to 1768 °C (-58 to 3214.4 °F)-270 to 400 °C (-454 to 752 °F)
10 W20 W100 W
Resistance W
*a= 5000 ppm / K or 6180 ppm / KConnection type : 2/3/4-wire connectionSensor current : 0.5 mA
± 20 mV or 0.08%± 20 mV or 0.08%± 30 mV or 0.08%± 50 mV or 0.08%
www.radix.co.inRadix Electrosystems Pvt Ltd, B-14, 2nd Floor, Ghanshyam Indl Estate, Veera Desai Road, Andheri (West), Mumbai - 400 053, IndiaTel : + 91 22 42537777 · Fax : + 91 22 42537700 · Email : [email protected] · www.radix.co.in
AN ISO 9001:2008 COMPANY
INSTRUMENTS T : + 91 22 42537777 x 701 F : + 91 22 E : [email protected]
SENSORS T : + 91 22 x 732 F : + 91 22 E : [email protected]
GAUGES T : + 91 22 x 733 F : + 91 22 E : [email protected]
AUTOMATION C : 0-9322405471 E : [email protected]
42537700
42537777 42537700
42537777 42537700
C : 0-9324319150
www.radix.co.in
2-wire, HART Temperature Transmitter SCC631
CAT#367R2/P3/3/
Dimensions
Electrical connections
A. Power supply, HART programmer
iPAQ
PC
PDA PDA
A
7.5 to 45 V10.5 to 45V
1
2
HART modem
ORDER CODE 2281
B. How to connect a sensor