2005_different Approaches Measurement of Hemodynamic by Electrical Impedance Plethysmography Method

  • Upload
    suknats

  • View
    213

  • Download
    0

Embed Size (px)

Citation preview

  • 7/31/2019 2005_different Approaches Measurement of Hemodynamic by Electrical Impedance Plethysmography Method

    1/2

    DIFFERENT APPROACHES MEASUREMENT OF HEMODYNAMIC BY

    ELECTRICAL IMPEDANCE PLETHYSMOGRAPHY METHOD

    Albinas Stankus

    KMU, Institute of Psychophysiology and Rehabilitation, Palanga, Lithuania

    [email protected]

    Abstract: The goal of presented study is an elaboration of methodology for direct measurement ofcirculation by means of volume units. Particular hardware has been elaborated for this reason. A newprinciple in measurement of relative changes of the pulse wave amplitude allows registering byelectroplethysmogram. It enabled directly to measure changes in the volume of tissues and to assesshemodynamics of various body parts by relative (ml/100ml) volume units and do this method more precise.This study enabled to improve this method for measurement of the blood circulation in legs. This methodcould be used for diagnostic purposes in vascular pathology.

    Introduction

    Popularity of the impedance plethysmography method is based on its simplicity. However, somebiophysical and methodological problems occur in the impedance cardiography, performing measurementsof the stroke volume. During the last 30 years, main attention of the investigators was focused on thecomparison of the absolute stroke volume values, measured by this method and found by other methods.Among investigated persons correlation values in 23 studies were from 0.49 to 0.97. Because of physicalreasons, results from earlier mentioned dependences might not be absolutely accurate. Usage of thismethod for investigations of the blood circulation in lower limbs is not doubtful. Values used in limbsvolume computing, are the following: specific tissue impedance (, *cm), distance between electrodes (l,cm) and constant impedance (Z0).Specific impedance of tissues is considered constant and equal to 135-150*cm. However experimentshowed that it depends on blood hematocryte changes from 22 to 66%. Adapting the specific impedance inthe concrete instances makes this method more difficult. In real conditions it is very difficult to keep

    parallelism of the placed band electrodes to precisely measure them and then to keep a distance. Constantimpedance in many cases is measured incorrectly, also from indicator. Generally, there are too manyapproximate calculations and not enough precise measurements. Measurement of the specific impedance,dependent on tissues and blood properties (in the other words homogeneity of the object to electricalconductance), is most difficult.

    Materials and Methods

    The new approaches measurement of hemodynamic is based on the premise, that we must measuretetrapolar means directly the ratio, as it is shown on equation: V/V0=- ko (Z/Z0). It is easy to realize it byusing a computer or/and electronic method [1]. Supposing, that the specific impedance of the measuredarea is constant, a direct ratio is found between the alternating volume changes and the total volumemeasured by the electrical method [1]. The blood filling of the measured segment, whose pulse curveindicates how it changes in time, averages the result, measured by this way. V/V0= ml/ml *100 %. Thisallows one to come to the measurement units, used in physiology ml/100ml. The device is calibrated by0.1 % change of the main impedance by a parallel connection of the resistor. The maximum amplitudevalues found, by using this method, indicate how many ml volume increases in each 100 ml segmentsbetween electrodes. The amplitude changes without additional calculations that are found relative changesof the volume in time. Circumferential electrodes are placed over a segment of the thigh, knee-joint, shinand foot.

    Results

    I investigated 43 healthy subjects (HSs) 30 - 69 years old and 42 patients (Pts) with differently stage ofendarteritis obliterans, from 40 to 69 year age. Table 1 shows the results of measuring the magnitude of

    ratio resistances. Magnitude of maximal pulse waves amplitude ratio to constant impedance was atThe maximume in knee-joint. In other region this ratio was lesser. The statistic analysis of the mainparameters shows very distinct and reliable differences between the HSs and the Pts (p

  • 7/31/2019 2005_different Approaches Measurement of Hemodynamic by Electrical Impedance Plethysmography Method

    2/2

    Table 1: Distribution of ratio resistance in leg

    Subjects Region Thigh Knee-joint Shin Foot

    HealthyLeft, % 0,115 0,00

    60,165 0,010

    0,115 0,073

    0,103 0,010

    Right, % 0,117 0,007

    0,169 0,011 0,116 0,083

    0,106 0,08

    PatientsLeft, % 0.099 0.007* 0.121 0.008* 0.079 0.006* 0.073 0.006*

    Right, % 0.095 0.007* 0.124 0.008* 0.086 0.007* 0.074 0.008*

    Discussion

    Applying of this method, had solved several serious problems. 1. Registering amplitude changes withoutadditional recalculations depict volumetric changes. The changes of the amplitude are valuated in percentvalues. For example, registered in the leg area 0.1 % the impedance changes mean, that for each 100 mlof the leg volume during the systole, comes 0.1 ml of blood. Its allow directly evaluate circulation of thevarious regions of the organism by measurement units, which are acceptable in medicine, e.g. ml/100ml of

    material. 2. The direct Z/Zmeasurement is performed more precisely, than in other methods, where theZvalue was found from the electropletysmographic sensor. 3. There is no need for calculations of thecross section area or the length of the conductor, because obtained results are not dependent on thesevalues. Pulse wave volume changes are measured in each 100 ml. volume as much as in big or in smallvolumes. There is another dependence: physiological and anatomical. When changes in length or in thecross section area of the conductor occur because of changes in electrode position, and in the spacebetween electrodes from another tissue type with different blood circulation can get in. The distancebetween the electrodes influences other physiological dependence, e.g. blood flow measurements isaveraged more in higher volume because of the diversity of the tissues. 4. There is no need for themeasurement of the specific impedance of the tissues, because during calculations the ratio of theimpedances disappears.This method of measurement reveals some discussable problems, but does not avoid conflicts becausenon-homogeneity of the tissues to the electrical conductance. This problem may be solved only by

    searching for tissues homogeneity, e.g. finding the best places for generating and measuring electrodes.The problem of homogeneity is related to one problem coefficients ko show, that absolute compliancedoes not always exists between impedance and volume. Current flow could occupy only a part of thevolume seen or measured by physical methods. Adapting the practice values of the specific impedance [2]for different organs is difficult and inaccurate. It is better to define this value as specific impedance,corresponding to the body area and to measure it every time. That would allow closer a measurement ofthe blood circulation, not calculation. This theme has many unanswered questions and needs furtherinvestigating.

    Conclusions.

    The method of impedance plethysmogrephy with direct Z/Z measurement of hemodynamic is moreprecisely. Relative amplitude of the leg artery pulse wave was decreased in patients with endarteritis

    obliterans of leg.

    References

    [1] STANKUS A., LAURINAVIIUS A.(1995): Elektopletizmografas. Patentas Nr. LT3906 BA61B 5/02:6[2] FAES T. J., VAN DER MEIJ H. A., DE MUNCK J. C., HEETHAAR R. M. (1999): The electric resistivityof human tissues (100 Hz-10 MHz): a meta-analysis of review studies, Physiol. Meas., 20(4): pp. R1-10.