Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
©2014 Waters Corporation 1
©2014 Waters Corporation 2
Troubleshooting Common Chromatography Problems
Richard DawField Technical Support Specialist
©2014 Waters Corporation 3
OutlineOutline
Diagnostic Flow Chart -Peak Problems-Retention Time Shift-Pressure Problems-Resolution Loss-Sensitivity-Ghost Peaks
Benchmarking using QCRM
Practice
©2014 Waters Corporation 4
Peak ProblemPeak Problem
Tailing Peak
Fronting Peak
What do the peaks look like?
Broad PeakSplitting Peak
©2014 Waters Corporation 5
Peak ProblemPeak Problem
Are All Peaks Affected?
Yes – Physical No - Chemical
©2014 Waters Corporation 6
Fronting PeakFronting Peak
Could be Caused by: Chemical•Mass Overload•Equilibration
Fronting Peak
Could be Caused by: Physical•Connections•Column Void•Clogged Frit•Volume Overload
©2014 Waters Corporation 7
ConnectionsConnections
Bad Connections can cause voids in the flow
path. Voids are bad because they will cause variable concentrations
across a band.
©2014 Waters Corporation 8
VoidingVoiding
Voided ColumnNormal Column
All Peaks Distorted
©2014 Waters Corporation 9
Clogged FritClogged Frit
As you can see from these two frits, bands will become distorted passing through the clogged frit.
Unblocked
Blocked
Unblocked
Remember Good Mobile Phase and Sample Preparation!
©2014 Waters Corporation 10
Column Volume in 4.6 x 50 mL =0.8 mL (800 µL)*
0.625 mg injected in 5 µL
% of column volume: 0.625
2 4Minutes 6
Ab
sorb
ance
0.625 mg injected in 500 µL
% of column volume: 62.5
2 4Minutes 6
Ab
sorb
ance
Volume Overload
*Guideline for Injection Volume: < 5% of column volume
• Wider peaks first observed at low retention
• Peak position shifts to higher retention in proportion to the injection volume
©2014 Waters Corporation 11
Encountered when mass injected onto column exceeds a certain limit—note earlier lift-off point
• Analytical load of 6µg yields efficient peak shape
• Preparative load of 25mg generates mass overload peak shape
Note that the back of the peaks of the analytical and prep loads are at the same retention (———-)
6.25 µg injected
2 Minutes 4 6
Ab
sorb
ance
Ab
sorb
ance
25 mg injected
2 4Minutes 6
Mass Overload-Concentration
©2014 Waters Corporation 12
Equilibration/WettingEquilibration/Wetting
In order for chromatography to take place, you need to have the right mobile phase within the pore. The pore pictured has elution strength solvent within it.
This pore contains a low elution strength mobile phase. Analytes can now be loaded into the pores and will retain.
©2014 Waters Corporation 13
Tailing PeakTailing Peak
Tailing Peak
Could be Caused by: Physical•Connections•Column Void•Clogged Frit•Volume Overload
Could be Caused by: Chemical•Mass Overload•Equilibration•Secondary Interactions•Ligand Loss
©2014 Waters Corporation 14
Minutes0 5 10 15 20 25
Secondary InteractionsSecondary Interactions
Acids and Neutrals have good peak shape --Basic Analyte “Tails” - Chemistry Problem - Cation Exchange
©2014 Waters Corporation 15
Secondary Interactions
Time (min) 50
Conventional C18
Modern C18
Neutral
Base
Neutral
Base
0
pH 7
Differences in Silanol Activity of Column Brands (Cation Exchange)
-End Capping? Tri-functionally Bonded? Bridged Carbon Chains?
High Silanol Activity
Low Silanol Activity
©2014 Waters Corporation 16
Hydrophobic Interaction with Bonded Phase
Hydrophobic Interaction andIon exchange Interaction
with Charged SitesO-SiO-SiO-O-SiO-SiO-O-SiO -O-SiO-SiO-SiO-O--Si
O-Si
O-SiO-SiOHO-SiO-SiOHO-SiOHO-SiO-SiO-SiOH
O-SiO-Si
(CH3)2HN+(CH3)2
HN+Mobile
Phase pH < 3
Si - OH
-Mobile Phase pH 6-8Si – O
BaseVery Early Elution—IonizedNo Cation Exchange
Base
Tailing and excessive retention in HPLC
Same “Conventional” C8 Column used in both experiments
Cation Exchange
Secondary Interactions
©2014 Waters Corporation 17
IonIon--Pairing ReagentsPairing Reagents
Additives that pair with charges on a molecule to dampen the negative effects of those charges on reversed phase retention are
called Ion-Pairing Reagents.
Ion Pairing Agent: Pairs With:TriEthyl Amine (TEA) Negative Charges
TriFluoroacetic Acid (TFA) Positive ChargesHexylamine Negative Charges
Octane Sulfonic Acid Positive Charges
These additives require long equilibration times to ensure consistent concentrations throughout the pores of a chromatographic material.
©2014 Waters Corporation 18
Ligand Loss: Hydrolysis of a Bonded Phase Material
Note: Sometimes called “Column Bleed”
+ HCl
+
+
Low pH (hydrolysis of ligand)
SiC
CC
CC
CC
CH3
CH3
H3C
Cl
OHSi
O
O
O
SiC
CC
CC
CC
CH3
CH3
H3C
OSi
O
O
O
SiC
CC
CC
CC
CH3
CH3
H3C
HOOH
SiO
O
O
Note: as hydrolysis of ligand continues, you will obtain less retentivity and potentially poor peak shape for bases.Cannot “regenerate”
BONDING
Mobile phase
©2014 Waters Corporation 19
Modern Stationary PhasesModern Stationary Phases--
Anal. Chem. 2003, 75, 6781-6788
U.S. Patent No. 6,686,035 B2 Bridged Ethanes within a silica matrix
©2014 Waters Corporation 20
Broad PeakBroad Peak
Could be Caused by: Chemical•Mass Overload•Secondary interactions•Equilibration•Sample Solvent•Temperature
Broad Peak
Could be Caused by: Physical•Connections•Column Void•Clogged Frit•Volume Overload•System Volume•Data Acquisition
©2014 Waters Corporation 21
System VolumeSystem Volume
Initial Band Band over Time
©2014 Waters Corporation 22
Data Acquisition RateData Acquisition Rate
10pt/1sec3pt/1sec
If the data rate is increased, more fine detail can be seen.
©2014 Waters Corporation 23
Sample SolventSample Solvent
Column: SymmetryShield™ RP18, 3.5 µm, 3.9 x 150 mm
Guard Column: Sentry™ Guard Column SymmetryShield RP18, 5µmTemperature: 30°CMobile Phase: 0.1% TFA:Methanol
(60:40)Detection: UV at 210 nmFlow Rate: 1 mL/minInj. Volume: 30 µL
Sample Identification1.EDDP2. Diphenhydramine (IS)3. Methadone
5 1510 Min.
Sample in Water
0.04AU
1
23
Sample in 80% MeOH, 2% HAc
15
12
30.02AU
Need to focus sample on head of the column!!
This is hindered by using a poor diluent choice
©2014 Waters Corporation 24
Splitting PeakSplitting Peak
Could be Caused by: Physical•Connections•Column Void•Clogged Frit
Splitting Peak
Could be Caused by: Chemical•Impurity/Coelution
©2014 Waters Corporation 25
Retention Time Shift Retention Time Shift
Loss of Retention
Gain of Retention
Which way are they shifting?
Peaks Move Both Ways
©2014 Waters Corporation 26
Retention LossRetention Loss
Could be Caused by:•De-Wetting•Ligand Loss•pH effects•Equilibration•Temperature•Mobile Phase Composition•Mobile Phase Delivery
Loss of Retention
©2014 Waters Corporation 27
De-wetting/ “Hydrophobic Collapse”
Change in pressure ejects aqueous mobile phase from the pore
©2014 Waters Corporation 28
1,500 psi
1,500 psi
Mobile phase: Aqueous 0.1% Acetic Acid
Note: Column is not broken—just stopped working—Rewet to restore performance
Minutes0 2 4 6 8 10
Initial(column was wetted first with organic)
Amoxicillin
Vo: No retentivity for analyte
Total Loss of Retention
De-wetting/ “Hydrophobic Collapse”
After Flow Stoppage Re-Start in Mobile Phase (Pores de-wet 100%)
x
©2014 Waters Corporation 29
pH effects pH effects -- Retention mapRetention map
pH Range for Silica ParticlespH
0
5
10
15
20
25
30
35
40
0 2 4 6 8 10 12
Rete
ntio
n Fa
ctor
(k)
Acid
Base
Neutral
Maximum acidic compound retention range
Maximum basic compound retention range
pH Range for Hybrid Particles
©2014 Waters Corporation 30
pH EffectspH Effects
You can see that by changing the pH you can have peaks change retention times. This is due to a change in their
charged states, and happens to acids and bases.
Base Acid
Neutral
Acid Base Acid Base
NeutralNeutral
pH = 2 pH = 6 pH = 10
©2014 Waters Corporation 31
TemperatureTemperature
The Rate of Mass Transfer in and out of the pores of the stationary phase increases with increasing temperature.
25oC
90oC
©2014 Waters Corporation 32
Mobile Phase CompositionMobile Phase Composition
If these are mixed such that the bottom liquid is added first, which way will give a 60/40 mixture?
400mL MeOH
600mL Water
or
40% MeOH
60% Water
600mL Water
400mL MeOH
©2014 Waters Corporation 33
Mobile Phase CompositionMobile Phase Composition
If the liquids are measured out separately and then mixed the percentages will be more accurate. In practice
it is important that regardless how this is done, it is always done the same to avoid errors.
60:40 MeOH/water = 960mL
400mL MeOH
600mL Water
+
40% MeOH
60% Water
©2014 Waters Corporation 34
Mobile Phase DeliveryMobile Phase Delivery
Pump
If the pump is not delivering the appropriate amount of solvent over time, you will not get reproducible retention times. This can happen if the pump is improperly serviced, or not serviced at all.
©2014 Waters Corporation 35
Retention GainRetention Gain
Gain of Retention Could be Caused by:
•De-Wetting•Ligand Loss•Equilibration•pH effects•Temperature•Mobile Phase Composition•Mobile Phase Delivery
©2014 Waters Corporation 36
Both WaysBoth Ways
Could be Caused by:•Temperature•Mobile Phase Delivery•Mobile Phase Composition•Mixing
Peaks Move Both Ways
©2014 Waters Corporation 37
Gradient Formation/MixingGradient Formation/Mixing
In order for a gradient to be effective, small changes in the gradient need to be consistently presented to the head of the column.
How well are you mixing different solvents together?
VS
©2014 Waters Corporation 38
Pressure ProblemsPressure Problems
Pressure Problem
What happens?
Low Pressure
High Pressure
Fluctuating Pressure
©2014 Waters Corporation 39
Pressure Problem Pressure Problem -- HighHigh
High Pressure
•Clogged Frit, column, check valve, tubing…etc…
•Pump Problem•Viscosity•Flow Rate•Temperature
©2014 Waters Corporation 40
Pressure Problem Pressure Problem -- FluctuatingFluctuating
Fluctuating Pressure
•Pump Problem•Check Valve Issue
©2014 Waters Corporation 41
Pressure Problem Pressure Problem -- LowLow
Low Pressure
•Leak
•Flow Rate•Temperature
©2014 Waters Corporation 42
Resolution LossResolution Loss
Could be Caused by: Chemical •Temperature•Mobile Phase Composition•pH effects
Could be Caused by: Mechanical •Changed system volume•Mobile Phase Delivery•Equilibration
©2014 Waters Corporation 43
SensitivitySensitivity
All Peaks decreased?
Yes – Chromatography System
No – Sample/Sample Prep
©2014 Waters Corporation 44
Sensitivity Sensitivity –– Chromatography System Chromatography System
Possible Causes:•Old Lamp•Dirty Flow Cell•Wrong Wavelength•Dirty Cone•MS Tuning•Autosampler
©2014 Waters Corporation 45
Sensitivity Sensitivity –– Sample/Sample PrepSample/Sample Prep
Causes:•Matrix Effects•Adsorption to Vial/Filter•Sample Prep•Solubility
©2014 Waters Corporation 46
Ghost PeakGhost Peak
Unknown Peak
Blank Injection
Unknown Peak
Blank Injection
©2014 Waters Corporation 47
Ghost PeakGhost Peak
Unknown Peak
Blank Injection
0 µL Injection
Unknown Peak
Blank Injection
0 µL Injection
©2014 Waters Corporation 48
Ghost PeakGhost Peak
Unknown Peak
Blank Injection
0 µL Injection Contamination of Sample during
prep?
©2014 Waters Corporation 49
Ghost PeakGhost Peak
Unknown Peak
0 µL Injection
Blank Injection
©2014 Waters Corporation 50
Ghost PeakGhost Peak
Unknown Peak
0 µL Injection
Blank Injection
Contamination of Chromatography System/Detector
©2014 Waters Corporation 51
OutlineOutline
Diagnostic Flow Chart -Peak Problems-Retention Time Shift-Pressure Problems-Resolution Loss-Sensitivity-Ghost Peaks
Benchmarking using QCRM
Practice
©2014 Waters Corporation 52
System Standards AvailableSystem Standards Available
System Performance
Kits andStandards
ApplicationSpecific
Standards
QualityControl
Reference Materials
©2014 Waters Corporation 53
Quality Control Reference Material Quality Control Reference Material OverviewOverview
Requirements for a QCRM Material– Reproducible Lot to Lot– Accurate– Appropriate to analysis type (LCMS, LC-UV)
The quality of the reference material is critical to the evaluation of analytical data and instrument performance
QCRM’s can be used to benchmark a system when it is known to be in good working order
The most powerful way to use QCRM’s is by analysing them routinely on the analytical system and control charting the data. The current and historical data will then allow for identification of areas of excess variation which may warrant concern and action
©2014 Waters Corporation 54
New Catalog for 2015/2016New Catalog for 2015/2016
©2014 Waters Corporation 55
New Catalog for 2015/2016New Catalog for 2015/2016
©2014 Waters Corporation 56
OutlineOutline
Diagnostic Flow Chart -Peak Problems-Retention Time Shift-Pressure Problems-Resolution Loss-Sensitivity-Ghost Peaks
Benchmarking using QCRM
Practice
©2014 Waters Corporation 57
Retention Time ShiftRetention Time Shift
Day 2Day 1 Conditions:
BEH C18, 186002350
• Col Dimension : 2.1x50mm• Temperature : 35C• Mobile Phase A : 100mM Na2HPO4• Mobile Phase B : Methanol• Flow Rate : 0.45mL/min.• Detection : PDA
©2014 Waters Corporation 58
Retention Time ShiftRetention Time Shift
Day 2Day 1
Mobile Phase Delivery – Check Pumps!
Conditions:
BEH C18, 186002350
• Col Dimension : 2.1x50mm• Temperature : 35C• Mobile Phase A : 100mM Na2HPO4• Mobile Phase B : Methanol• Flow Rate : 0.45mL/min.• Detection : PDA
-Temperature-pH effects
©2014 Waters Corporation 59
Split PeaksSplit Peaks
Original Peak Bad Peak
©2014 Waters Corporation 60
Split PeaksSplit Peaks
Original Peak Bad Peak Chromatography
Physical?
©2014 Waters Corporation 61
A new column was tried and the poor peak shape remained.
Split PeaksSplit Peaks
Original Peak Bad Peak Chromatography
Physical?
©2014 Waters Corporation 62
A new column was tried and the poor peak shape remained.
Split PeaksSplit Peaks
ChromatographyOriginal Peak Bad Peak
Physical?
©2014 Waters Corporation 63
Retention Time ShiftRetention Time Shift
At Start1 Day Conditions:
SymmetryShield RP18; 186000174
• Col Dimension : 2.1x150mm• Temperature : 35C• Mobile Phase A : 100mM H3PO4• Mobile Phase B : Methanol• Flow Rate : 0.45mL/min.• Detection : Mass Spec ESI+
©2014 Waters Corporation 64
Retention Time ShiftRetention Time Shift
At Start1 Day
Ligand Hydrolysis-Dewetting
-Mobile Phase composition, mobile phase delivery, pH
Conditions:
SymmetryShield RP18; 186000174
• Col Dimension : 2.1x150mm• Temperature : 35C• Mobile Phase A : 100mM H3PO4• Mobile Phase B : Methanol• Flow Rate : 0.45mL/min.• Detection : Mass Spec ESI+
©2014 Waters Corporation 65
Tailing PeakTailing Peak
Does not meet Spec
©2014 Waters Corporation 66
Tailing PeakTailing Peak
Does not meet Spec
Not All Peaks TailStrong Base
©2014 Waters Corporation 67
Tailing PeakTailing Peak
Does not meet Spec
Not All Peaks TailStrong Base
Conditions: •Symmetry C18 3.5um 2.1x50mm•Temperature 25C•Flow Rate : 0.3mL/min.•Mobile Phase A : Water•Mobile Phase B : ACN
©2014 Waters Corporation 68
Tailing PeakTailing Peak
Does not meet Spec
Not All Peaks TailStrong Base
Conditions: •Symmetry C18 3.5um 2.1x50mm•Temperature 25C•Flow Rate : 0.3mL/min.•Mobile Phase A : Water•Mobile Phase B : ACN
-Add some buffer!
©2014 Waters Corporation 69
Fronting PeakFronting Peak
Why is this peak fronting?
©2014 Waters Corporation 70
Fronting PeakFronting Peak
Why is this peak fronting?
Why are all the peaks fronting?
©2014 Waters Corporation 71
Fronting PeakFronting Peak
Conditions:•Column : Atlantis T3 1x50mm 3µm•Injection : 20µL @ 5mg/mL Std.•Flow Rate : 0.1mL/min.•Mobile Phase A : 20mM Phosphate pH 3•Mobile Phase B : ACN
Why is this peak fronting?
Why are all the peaks fronting?
©2014 Waters Corporation 72
Fronting PeakFronting Peak
Conditions:•Column : Atlantis T3 1x50mm 3µm•Injection : 20µL @ 5mg/mL Std.•Flow Rate : 0.1mL/min.•Mobile Phase A : 20mM Phosphate pH 3•Mobile Phase B : ACN
Overload!
Too much Injection volume!
Why is this peak fronting?
Why are all the peaks fronting?
©2014 Waters Corporation 73
QUESTIONS???
Zach KoontzField Chemistry [email protected]
Richard Daw and Steven ParkerField Technical Support [email protected][email protected]
©2014 Waters Corporation 74