89
Vacuum Techniques •Gauges

6 Vacuum Gauges

Embed Size (px)

Citation preview

Page 1: 6 Vacuum Gauges

Vacuum Techniques

•Gauges

Page 2: 6 Vacuum Gauges

Vacuum Gauges• Thermal Conductivity Gauges: (Pirani, TC, Convection)

– Describe the operating principle for each gauge– Explain the use of the Wheatstone bridge– Compare the constant current and constant temperature modes of

operation– Explain why the TC gauges typically have an upper limit of 1 torr– Relate the lower pressure limit to thermal radiative heat transfer

• Ionization Gauges: (Hot Cathode, Cold Cathode)– Describe the operating principle and electrical configuration for each

gauge– Relate the sensitivity of the HCG to the geometry and gas properties– Explain what is meant by the ‘soft x-ray’ limit– Describe the role of magnetic field for the CCG

Page 3: 6 Vacuum Gauges

• Absollout gauges??/

Page 4: 6 Vacuum Gauges

Measurement of pressure

• Mechanical phenomena gauges: measure actual force exerted by gas (e.g. manometer).

• Transport phenomena: measuring gaseous drag on moving body (e.g. spinning rotor gauge) or thermal conductivity of gas (e.g. thermocouple gauge).

• Ionization phenomena gauges: ionize gas and measure total ion current (e.g. ion gauge).

• Partial pressure residual gas analyzers:mass spectrometers.

Page 5: 6 Vacuum Gauges

• Mechanical guagesMechanical: liquid column, diaphragm, etcReal pressure from force measurement

Page 6: 6 Vacuum Gauges

Mechanical phenomena Pressure Measurement

Page 7: 6 Vacuum Gauges

Mechanical phenomena Pressure Gauge

Page 8: 6 Vacuum Gauges

Bourdon Gauge

Page 9: 6 Vacuum Gauges

Mechanical phenomena How the gauge works

Page 10: 6 Vacuum Gauges

Capacitance (diaphragm) gauge: measure diaphragm bending by capacitance change

d

d

AC

pressure

Absolute pressurereference to a vacuum cell

Full range: 0.02 torr, 1 torr, 1000 torr,

10000 torr

Accuracy ~ 0.1 %

real species independent

linear, accurate

zero point is not absolute

Page 11: 6 Vacuum Gauges

• Transport phenomena

Page 12: 6 Vacuum Gauges

Transport phenomena(1)

• Gas molecules colliding with the heated filament take kinetic energy away from the filament’s atoms. This cools the filament down.

• A higher pressure means that none molecules collide with the filament per second, so a larger cooling effect occurs.

Page 13: 6 Vacuum Gauges

Transport phenomena(2)

• When the filament is cooler its resistance is lower.

• The resistance can be measured automatically.

• One way of doing this is to use a Wheatstone Bridge.

Page 14: 6 Vacuum Gauges

Transport phenomena(3)

Pirani Gauge

Filament

-+

Page 15: 6 Vacuum Gauges

Thermal conductivity gauges mechanisms

– Four possible mechanisms to take heat away from sensing element:

• Radiative heat transfer to surroundings

• Conduction along the sensing element

• Conduction through the residual gas

• Convective heat flow to the residual gas

Hot wire

1/2 WC1/2 WC

Thermocouplejoint

WR

WG

WCNV

Heatingcurrent

Page 16: 6 Vacuum Gauges

Thermal conductivity gauges mechanisms

1-Radiative heat transfer transfer:

WR 1 2r1L T14 T2

4

L

T1

T2r1

Page 17: 6 Vacuum Gauges

• Conduction along the wire

1

2WC r1

2kwiredT

dl

L

T1 T2

Page 18: 6 Vacuum Gauges

• Conduction through the residual gas:

as properties determine variation in sensor response to pressure

WG 2r12L F 1

1

P

R '

8MT2

1/2

T1 T2 (M, )

R`: wire resistance,

P: pressure ,

L, r1: Wire length and diameter,

M, , F: Gas properties.

T1 and T2 Wire and gas temperature respectively

Page 19: 6 Vacuum Gauges

Thermal Conductivity Gauges• Pirani calibration (and low pressure limit):

• Platinum wire with r1 = 0.0127 mm, L = 150 mm

• At very low P, Vbr = 0.3907 V, and Rs = 43.61 WR+WC =8.75x10-4 W

• Subtracting WR+WC from WT yields WG for N2 as shown below:

WT WG WR Wc Vbr

2

4RS

Page 20: 6 Vacuum Gauges

Thermal Conductivity Measurment Constant current mode

• Constant current mode:– As pressure decreases, less heat is

removed by conduction through the gas from the sensing element, and temperature increases.

– Measure temperature dependent change of resistance (or TC voltage).

Page 21: 6 Vacuum Gauges

Thermal Conductivity MeasurmentConstant temperature mode:

• Constant temperature mode:– Adjust current to heater to maintain constant

sensor property (resistance or voltage)– Use Wheatstone bridge arrangement to

optimize sensitivity– Correlate current flow against pressure– Extends high pressure range (used in Pirani,

not common for TC)

Page 22: 6 Vacuum Gauges

Pressure GaugesGauge

Compensator to system

milliammeter

Power Supply

Page 23: 6 Vacuum Gauges

Thermocouple gauge

For roughing vacuum measurements

Page 24: 6 Vacuum Gauges

• Ionization gauges

Page 25: 6 Vacuum Gauges

Ionization phenomena(1)

• Electrons emitted from the heated filament are attracted to the positive grid. Many electrons follow long looped paths before striking the grid.

• During this time they collide with gas molecules, thus creating positive ions.

Page 26: 6 Vacuum Gauges

Ionization phenomena(2)

• These ions are attracted to the negative collector and constitute a current into the gauge circuit.

• A higher pressure results in a higher gauge current.

• The emission current must be kept within strict limits as it too affects the gauge current.

Page 27: 6 Vacuum Gauges

Ionization phenomena(3)

• Thermionic/hot cathode ionization gauges.• Energetic beam of electrons (constant I-) used

to ionize gas molecules and produce ion current.

• e- + M → M+ + 2e-

I+ = p KI -, K: ion gauge sensitivity• Upper pressure limit (10-3 Torr): secondary ion

ionization excitation, filament burn out.• Lower pressure limit (10-10 Torr): secondary

electron current from X-ray emission.

Page 28: 6 Vacuum Gauges

Hot Catode Ionization Gauges – HCG(1)

• Hot cathode gauge (HCG)– Thermionic source (electrons)

cause inelastic collisions with gas, producing ions

– Ions collected at (-) biased surface– Electrons travel to (+180 V) grid

• Pre-1950 lower limit of 10-6 Pa– Soft x-rays produced at grid result in

photo-electron flow away from collector (same as ion flow to collector)

– Bayard-Alpert modification: fine wire collector surrounded by grid range extended to 10-9 Pa

+

-

-

collector

grid

filament

Page 29: 6 Vacuum Gauges

HCG(2)

Ionization Gauge

Heated Filament

Emission Current

A.C.

+ve

Grid

Ion Collector

Pressure Gauge

Page 30: 6 Vacuum Gauges

HCG(3)

Page 31: 6 Vacuum Gauges

HCG(4)

• 10-11 ~ 10-3 torr• linear sensitivity• absolute zero point • species dependent

Th on Ir has a lower work function,

so works at a lower temperature

sensitivity ∝ ionization cross section

air, N2, O2 1.0

He 0.15

Ne 0.3

H2 0.4

CH4 1.4

Page 32: 6 Vacuum Gauges

HCG(5) Gauge sensitivity

• Gauge sensitivity:– Dependent on rate of ionizing

collisions

– Ion collision cross section given by• = cross section per molecule• n = molecule density = P/(kT)• L = length of ionizing space• A = cross sectional area of

electron beam

LAn

Page 33: 6 Vacuum Gauges

HCG(6) Ion current

– Ion current is then:

• N = arrival rate of electrons• q = charge per electron

– With Nq = ie, we have i+ = P K ie + ir,

– Gauge sensitivity S = K ie

– Typical range for K: 0.02 Pa-1 to 0.2 Pa-1 (bigger is better)

NqLAP

kT

K LAkT

Page 34: 6 Vacuum Gauges

Cold Cathode Ion Gauges: CCG(1)

• Cold Cathode Gauge (CCG) Invented by Penning in 1937 - electrical configuration like diode SIP

• 1952 design of wire-in-cylinder with reverse polarity (+ wire) provides key performance improvement

• “Inverted Magnetron” - Magnetic field produces long helical path for electrons & ions many collisions

• Ionizing electrons part of self-sustaining gas discharge

• No background current to mask the ion current

• Power into CCG ~ 0.1 W

Page 35: 6 Vacuum Gauges

Cold Cathode Ion Gauges: CCG(2) similar to hot cathode ion gauge

more robust

less accurate

hard to ignite at low pressure

magnet

Hot Cold

Page 36: 6 Vacuum Gauges

CCG(3)

• “Striking time” is delay before discharge starts. At 10-4 Pa, ts ~ seconds; at 10-8 Pa, ts ~ hours or days

• Nearly linear dependence of ion current with pressure:

• Typical anode voltage +3 kV

• Typical current 0.1 mA

ig KPn; n 1.05 1.2

Page 37: 6 Vacuum Gauges

Pressure Measurement

Vacuum Gauge Selection adapted from Lesker.com

Convectron Gauge:Initial pumpdown from

1 atm, and as a foreline monitor

Thermal Conductivity of Gas

Baratron:Insensitive to gas

composition,Good choice for

process pressures

True Pressure (diaphragm displacement)

Ion Gauge:Sensitive to gas composition, buta good choice for base pressures

Ionization of Gas

RGA:A simple mass spectrometer

Page 38: 6 Vacuum Gauges

Residual gas analyzers

• More compact with higher sensitivity.• Gaseous ions formed in ion source box by

electron bombardment, extracted with suitable fields, separated in analyzer and then collected and measured.

• Magnetic sector analyzer: masses separated by static magnetic and electric fields.

• Quadrupole mass analyzer: masses separated in oscillating quadrupolar electric field.

Page 39: 6 Vacuum Gauges

Mass Spectrometry

• Quadrupole Mass Spectrometry– Most commonly used in laboratories

• Least expensive of commercially available mass specs

• Detects up to 100 AMU commonly

• Can be extended easily to 300 AMU detection

• Downside: fragmentation patterns of molecules can be complex

• Upside: the simplest and least expensive mass spec (typically $15,000 to $35,000 depending on attachments; other types of mass specs can easily cost a few million)

Page 40: 6 Vacuum Gauges

Ion source: Tungsten filament

Gas inlet

Diaphragm Vacuum Pump

Turbomolecular Drag Pump

Analyzer: Quadrupole Mass Filter

Detector: Faraday/ Channeltron

Gas outlet

Page 41: 6 Vacuum Gauges

Mass Spectrometer Basics

Vacuum Requirements

Ion Detection

Ion Filtering

Ion Creation

Page 42: 6 Vacuum Gauges

Mass Spectrometer Basics

e-

Ion Source

Quadrupole Mass Filter

Ion Detector

Page 43: 6 Vacuum Gauges

Vacuum Requirements

• Filament Longevity

Typical Vacuum ~ 1 x 10-6 Torr

HH

e -

Gas Density ~ 1010 molecules /cm3

(@ 760 Torr (1 atm) ~ 1019 molecules /cm3 )

+

Ion Mobility

Detector Operation

Page 44: 6 Vacuum Gauges

The Atomic Model

12 C = 12 A.M.U.

1 AMU = 1.66 X 10 -27 kg

= Electron ~ 0 AMU, -1e charge

= Neutron ~ 1 AMU , no charge

= Proton ~ 1AMU , +1e charge

Page 45: 6 Vacuum Gauges

The Periodic Table

Page 46: 6 Vacuum Gauges

Ionization

12 C 12 C+ + 2e-

Atom Ion

+ 1e-

e-

Page 47: 6 Vacuum Gauges

Gas Atom

Gas Ion

Electron

Ionization Source - Open

Page 48: 6 Vacuum Gauges

Filaments

Ions Out

Neutral Gas Atom/Molecule

Electron

Ion

Gas In

Pressure (mBar)

Filaments

10-05 10-03 1010-04

Ionization Source - Closed or Gas Tight

Page 49: 6 Vacuum Gauges

Electron EmissionThe emission of electrons from the filament is carefully controlled to provide consistent ion formation.

e-IF

IE

IF =Filament Current ~ 3 Amps d.c.

IE =Electron (emission) Current ~ 1 mAmp d.c.

SET

IonFormationCamber

Page 50: 6 Vacuum Gauges

Ionization

1 e- + 12C12C++ + 3e-

1 e- + 18H2O 1 e- + 17OH 1 e- + 17OH 17OH+ +2 e-

Double Ionization

Fragmentation - Ionization

m/e = 6

m/e = 17

Page 51: 6 Vacuum Gauges

Ionization Products

Ionization can produce more than

one ion type.

E.g. Argon , 40Ar

40Ar + 1e- 40Ar+ + 2e-

40Ar2+ + 3e-

40Ar3+ + 4e-

(m/e = 40)

(m/e = 13.3)

(m/e = 20)

Page 52: 6 Vacuum Gauges

Ionization - Isotopes

Argon Isotopic Masses

P=Proton, N=Neutron

18P+22N=40AMU

18P+20N=38AMU

18P+18N=36AMU M/Z

Log

(Inte

nsity

)

-1e

-1e

-1e

Page 53: 6 Vacuum Gauges

Mass scan of ultra pure argon - showing singly and doubly charged isotope peaks

Ionization Products – UHP Argon

Page 54: 6 Vacuum Gauges

Ionization Products

Molecules can fragment in the ionization process.

ABC + e- ABC+ + 2e-

AB+C+ + 2e-

AB+ + C + 2e-

A + BC+ + 2e-

A+ + BC + 2e-

AC+ + B + 2e-

AC + B+ + 2e-

In addition to the above it is possible for other combinations to be formed

Page 55: 6 Vacuum Gauges

Typical Ion Formation

Spectrum of CO2 showing the 11 mostintense ions

NaturalAbundances

18O = 0.2%

13C = 1.1%

Page 56: 6 Vacuum Gauges

Mass Filter

• Cylindrical Rods.• Stainless Steel or Molybdenum.•Opposite Rods are Connected Electrically.•Alignment is Critical not adjustable.

U + V cos wt

+

+

-

-

Page 57: 6 Vacuum Gauges

Mass Filter

++ +++

++

I i

e-++

++

+

y

x

z

SEM

ION

SO

UR

CE

QUADRUPOLE ROD

QUADRUPOLE ROD

QUADRUPOLE ROD

Selected m/e ion

Higher m/e ion

Lower m/e ion - deflected in y-axis

- deflected in z-axis

- reaches detector

Page 58: 6 Vacuum Gauges

Resolution

Over Resolution

10

1

1 AMU

10% Valley Resolution

Under Resolution

1 AMU 1 AMU

Valley height

Page 59: 6 Vacuum Gauges

Ion Detectors - Faraday

I i

e-

I i ~ 10-14....10-9 A

= Selected ion - positive charge

Indestructible Detector but gain = unity.

Cannot detect small ion currents <10-14 Amps.(Limit depends on electrometer only)

Page 60: 6 Vacuum Gauges

SEM Detector – Continuous Dynode

I i

GAIN ~ 100 106

set by SEM VOLTS

e-

e-

I i ~ 10-14....10-5 A

= Selected ion - positive charge-attracted into SEM by -ve dc volts.

SEMVOLTS ~ - 1500V dc

Can be destroyed by high currents>10-5 Amps , or by operation at highpressure.

MASS FILTER

Page 61: 6 Vacuum Gauges

SEM Detector – Discrete Dynode

MASS FILTER

Deflection Unit

Vdef

Faraday

signal

HV- Ie

Page 62: 6 Vacuum Gauges

Data acquisition modes - Scan Analog

Here a range of masses are scanned and the data appears as mass vs. intensity. This mode is mainly used for diagnosticsand checking mass spec performance, peak shape etc.

Page 63: 6 Vacuum Gauges

Data acquisition modes - Scan Bargraph

This mode is similar to Analog but it only reports data when a masspeak is found - this reduces the data for simplicity. This mode is mainly used for scanning for unknown components.

Page 64: 6 Vacuum Gauges

Data acquisition modes - MID, MCD

MID/MDC - here the masses to be monitored are known and thesystem will only report back these points. It is the fastest mode andvery useful for trending data over time

Multiple Ion Current Determination.Ion current vs. time

Multiple Concentration Determination.Concentration vs. time

NOTE:Requires calibration gas mixes

MCD =

MID =

Page 65: 6 Vacuum Gauges

Residual Gas Analyzer

QUADRUPOLEHEAD

CONTROL UNIT

Page 66: 6 Vacuum Gauges

How the RGA works

Page 67: 6 Vacuum Gauges

MASS NUMBER (A.M.U.)

RE

LA

TIV

E I

NT

EN

SIT

Y

NORMAL (UNBAKED) SYSTEM

H2

H2O

N2,, CO

CO2

(A)

RGA SPECTRUM

Page 68: 6 Vacuum Gauges

MASS NUMBER (A.M.U.)

RE

LA

TIV

E I

NT

EN

SIT

Y

SYSTEM WITHAIR LEAK

H2

H2O

N2

CO2

(B)

O2

RGA SPECTRUM

Page 69: 6 Vacuum Gauges

The RGA 100 Residual Gas Analyzer

Page 70: 6 Vacuum Gauges

Quadrupole Mass Filter Components

Page 71: 6 Vacuum Gauges

Principles of Filter Operation

Page 72: 6 Vacuum Gauges

Residue Gas Analyzer (RGA)

A small quadrupole mass analyzer with an electron impact ionizer

With electron multiplier, sensitivity ~ 10-15 torr

Page 73: 6 Vacuum Gauges

Residual Gas Analysis

Page 74: 6 Vacuum Gauges

Residual Gas Analyzers

Page 75: 6 Vacuum Gauges

Learning Objectives

• Identify the 3 primary components of an RGA and describe the available options for each component

• Describe three performance parameters of an RGA and factors that influence the performance

• Describe the operation of the RF quadrupole

Page 76: 6 Vacuum Gauges

RGA Components

• Ion source– Open ion source– Closed ion source

• Mass filter (analyzer)– Magnet sector– RF quadrupole

• Ion current detection system– Faraday cup– Electron multiplier

Page 77: 6 Vacuum Gauges

Ionization• Ion production of each

species is proportional to its density or partial pressure – but sensitivity and gain

are not independent of mass

– Linear dependence on partial pressure holds true up to total pressure of ~ 10-3 Pa

• Ionization voltage typically 70 volts

• Ionization products can be single or double ionized molecules, or fragment ions

Page 78: 6 Vacuum Gauges

Ionization Sources

• Open Ion Source– Like the Bayard-Alpert gauge– High conductance of ion formation region to analyzer region for

composition consistency– Grid at (+) potential w.r.t. analyzer: defines ion energy

• 5-15 V quadrupole, 100-1000 V magnet sector

Page 79: 6 Vacuum Gauges

Ionization Sources• Closed Ion Source

– Process gas flows through ionization region

– Higher pressure in enclosure enables increased sensitivity (10 to 100 times higher pressure in ionization vs. analyzer region)

– Avoids effects of electron stimulated desorption (ESD) from walls when sampling high-pressure chambers

Page 80: 6 Vacuum Gauges

Mass Analyzers• Separate ions according to ratio of M/q

40Ar++ and 20Ne+ produce identical signal• Magnetic Sectors

– As seen in the helium mass spectrometer– Separation angles of 60º, 90º, 180º are common– Mass sweep via acceleration voltage change limited to x25– Electromagnetic change (0-0.25T) enables 1-100 amu

sweep

Rel

ati v

e In

t en

sity

M/q

Page 81: 6 Vacuum Gauges

Mass Analyzers• RF Quadrupole

– Two pairs of rods, one (+) pair, one (-) pair– DC plus AC(rf) potentials applied to all rods:

• (+) pair: • (-) pair:

– (+) pair creates ion valley most of the time: high mass, high inertia molecules pass through

– (-) pair creates ion hill most of the time: low mass, low inertia molecules pass through

U V cost (U V cost)

Page 82: 6 Vacuum Gauges

Mass Analyzers

• RF Quadrupole (cont.)– The width of the band pass,

or resolving power (M/M), is adjusted via ratio of U/V, length of the rods, or ion energy

– Resolving power measured by

– Detector on axis at the end of the filter counts the transmitted ions

– Linear sweep of rf & dc potential produces linear sweep of amu’s

peak width at 10% height

Page 83: 6 Vacuum Gauges

Ion Detectors

• Ion current, in, sensitivity S’n and partial pressure Pn are related by in = S’n Pn

– Sensitivity determines smallest detectable signal (peak height)– Typical sensitivity for nitrogen:

• Faraday cup– Measures electric current flowing to neutralize ion arrival– Incorporates means to avoid secondary electron release– Should be followed by stable, low-noise, high-gain FET amplifier– Commercial detector limit of 5x10-14 A (10-8 Pa, or 106 ions/sec)– Higher sensitivity (down to 10-12 Pa) requires an electron

multiplier

S’n = 5x10-6 A/Pa

Page 84: 6 Vacuum Gauges

Ion Detectors: Electron Multiplier• How does it work?

– Secondary electrons from ion impact generate multiple electrons upon each successive impact as they are accelerated toward the anode

• Gain

– G1 =

– G2 =

• Inlet voltage typically -2000 V• Glass tube is curved to prohibit (+) ions generated at anode from

traveling back full length of tube and generating out-of-phase secondary electrons

G G1G2n

# electrons generated by ion impact

# electrons generated by electron impact at each stage

Page 85: 6 Vacuum Gauges

Operation

• Mount a valve between RGA and chamber• Mount the RGA in a non-remote location• If pressure > 10-3 Pa:

– reduce electron multiplier voltage – use simple Faraday cup– use differentially pumped sampling configuration

• If spectrum changes when ion energy is varied, then ESD is present (reduce ion energy)

• Select appropriate filament (ThO2, W, Re) and precondition it by heating above normal operating temp.

• Turn off any Bayard-Alpert gauge• Save background spectrum for later comparison

Page 86: 6 Vacuum Gauges

Calibration• Do not accept partial pressure results

unless instrument has recent calibration for several useful gas mixtures

• Calibration procedure standards available from AVS

• Pulse injection method (quick, in-situ method)

– Isolate small quantity of gas in pipe (volume Vi, pressure Pi ) between two valves

– Calculate instantaneous pressure rise in chamber Pc = (PiVi/Vc)

– Determine sensitivity from peak ion current at t=0

S ' I peakPc

I peakVcPiVi

Page 87: 6 Vacuum Gauges

Did you catch it?

• Identify at least 3 important features of an RGA

» Range (1-800 amu)– Sensitivity (10-5 A/Pa, G ~ 106, 10-12 Pa)– Resolving power (1 amu)

• Identify the approximate voltage between each of these stages:– Electron filament and grid– Grid and quadrupole analyzer– Ion detector inlet and ground

70 volts

5 - 15 volts

-2000 volts

Page 88: 6 Vacuum Gauges

Gauge Operating Ranges

10-12 10-10 10-8 10-6 10-4 10-2 1 10+2

P (mbar)

Rough VacuumHigh VacuumUltra High Vacuum

Bourdon Gauge

Thermocouple Gauge

Cold Cathode Gauge

Capacitance Manometer

Hot Fil. Ion Gauge

Residual Gas Analyzer

Pirani Gauge

Spinning Rotor Gauge

McLeod Gauge

Page 89: 6 Vacuum Gauges

Vacuum gauges must calibrated by

• Comparison with absolute standard calibrated from its own physical properties.

• Attachment to calibrated vacuum system.

• Comparison with calibrated reference

gauge.