Flexport System Interface 904XX Operations Manual 070-1148-00 Rev. J ®

904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Embed Size (px)

Citation preview

Page 1: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface904XX

Operations Manual070-1148-00 Rev. J

®

Page 2: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

©2008 Spacelabs Healthcare

All rights reserved. Contents of this publication may not be reproduced in any form without the written permission of Spacelabs Healthcare. Products of Spacelabs Healthcare are covered by U.S. and foreign patents and/or pending patents. Printed in U.S.A. Specifications and price change privileges are reserved.

Spacelabs Healthcare considers itself responsible for the effects on safety, reliability and performance of the equipment only if:

• assembly operations, re-adjustments, modifications or repairs are carried out by persons authorized by Spacelabs Healthcare, and

• the electrical installation of the relevant room complies with the requirements of the standard in force, and

• the equipment is used in accordance with the operations manual.

Spacelabs Healthcare will make available, on request, such circuit diagrams, component part lists, descriptions, calibration instructions or other information which will assist appropriately qualified technical personnel to repair those parts of the equipment which are classified by Spacelabs Healthcare as field repairable.

Spacelabs Healthcare is committed to providing comprehensive customer support beginning with your initial inquiry through purchase, training, and service for the life of your Spacelabs Healthcare equipment.

CORPORATE OFFICES

U.S.A.

Spacelabs Healthcare, LLC5150 220th Ave SEIssaquah, WA 98029Telephone: 425-657-7200Telephone: 800-522-7025Fax: 425-657-7212

Authorized EC Representative UNITED KINGDOM

Spacelabs Healthcare, Ltd.Beech House, Chiltern CourtAsheridge Road, CheshamBuckinghamshire HP5 2PXTelephone: 44 (0) 1494 784422Fax: 44 (0) 1494 794414

BirthNet, Clinical Browser, Data Shuttle, Flexport, Intesys, Mermaid, MOM, Multiview, PCIS, PCMS, PrintMaster, Quicknet, Sensorwatch, TRU-CAP, TRU-CUFF, TruLink, Ultralite, Ultraview, Ultraview Care Network, Ultraview Clinical Messenger, Ultraview Digital Telemetry, Ultraview SL, Uni-Pouch, UCW, Varitrend and WinDNA are trademarks of Spacelabs Healthcare.

Other brands and product names are trademarks of their respective owners.

Caution:US Federal law restricts the devices documented herein to sale by, or on the order of, a physician.

Before use, carefully read the instructions, including all warnings and cautions.

Rx Only

!

Page 3: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Contents Page

Introduction

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1Flexport System Interface Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2Supporting Data Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3Help Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

Abbott Lifecare 4, 4P, or MicroDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19

Alaris IVAC 560/570/599Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20

Alaris 980C/PC-1/PC-2/PC-4Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6Pump address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20

Bard CritiCoreDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

Flexport System Interface Operations Manual i

Page 4: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9

Baxter-Edwards ExplorerDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8

Baxter Flo-Gard 6201/6301Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20

Bear 1000Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13

Biochem International BCI 9000Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8Selecting a Waveform Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10

Flexport System Interface Operations Manual ii

Page 5: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Braun Fluid Management SystemDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-17Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-17Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-18Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-20

Criticare 504Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9

Criticare POET 601Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-2Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-3Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-4Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-5Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-5Viewing Trends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-6Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-7

Criticare POET-II 602-3Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-10

Criticare POET IQ2Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-10

Flexport System Interface Operations Manual iii

Page 6: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Critikon Dinamap 1846 and 8100-SeriesDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-5Displaying Heart Rate Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-6Displaying New or Previous Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-6Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-7Printing NIBP Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-8

Datex Engström Erica/Elvira VentilatorDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-4Display Detail — Erica . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-6Display Detail — Elvira . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-7Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-8Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-9Viewing Ventilator Mode History (Elvira only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-10Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-10Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-12Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-12Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-12Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-13

Datex Normocap CD2-02/CD-200Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-11

Datex UltimaDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-8Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-11

Dräger CatoDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-6

Flexport System Interface Operations Manual iv

Page 7: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-8Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-11Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-11

Dräger Cicero BDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-9

Dräger Cicero EMDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21-11

Dräger Evita II, Evita II Dura, and Babylog 8000Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-9Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-10Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-11Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-11Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-11Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22-12

Dräger Evita 4Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-8Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-9Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-9Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-11

Flexport System Interface Operations Manual v

Page 8: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Dräger Evita 4/XLDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-8Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-9Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-9Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-11

Dräger Irisa/EvitaDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-7

Dräger Narkomed 2B and 2CDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-8Freezing the Waveform on the Display (Narkomed 2C Only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26-10

Dräger Narkomed 3, 4, and GSDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-10

Edwards Vigilance IDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-7

Flexport System Interface Operations Manual vi

Page 9: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Storing a CCO Value as a Bolus CO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-9

Edwards Vigilance IIDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-8Storing a CCO Value as a Bolus CO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-9

Edwards VigileoDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30-8

Engström EAS 9000Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31-11

GE Ohmeda 5200/5210Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-8

GE Ohmeda 5250Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-6

Flexport System Interface Operations Manual vii

Page 10: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-9

GE Ohmeda 7800/7810Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34-8

GE Ohmeda Care-Plus/Infant WarmerDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35-6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35-8

GE Ohmeda Rascal IIDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36-9

Hamilton Amadeus/Veolar/Galileo/Raphael XTCDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37-9

Infrasonics Adult Star 2000Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-7

Flexport System Interface Operations Manual viii

Page 11: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-7Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-8Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38-10

Maquet Siemens Servo 300Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-4Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-5Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-7Viewing the Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-7Viewing Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39-8

Maquet Siemens Servo 900Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-4Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-5Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-8

Masimo RadicalDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41-9

Mortara ELI 100/STMDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-5Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-6Adjusting the Waveform Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-7Selecting a Rhythm Lead Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42-9

Newport Medical Instruments E150/E200Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-3

Flexport System Interface Operations Manual ix

Page 12: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-7Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-8Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43-9

Oridion VitalCapDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-6Silencing Alarm Tones . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-7Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44-11

OSI OpticalCAMDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45-4Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45-5Selecting a Unit of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45-5Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45-5Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45-6

PhysioFlexDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46-9

PiCCODirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47-4Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47-5Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47-5Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47-6

Radiometer TCM3Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-4

Flexport System Interface Operations Manual x

Page 13: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48-7

Respironics Novametrix EspritDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-7Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-8Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49-9

Respironics-Novametrix 7300/7600Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-7Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-8Display Detail - FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-9Display Detail - PAW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-10Display Detail - CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-12Display Detail - NICO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-13Display Detail - SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-14Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-16Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-17Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-17Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-19Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-19

Respironics Novametrix 840/860Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51-7

Respironics Novametrix 1260/1265Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52-10

Flexport System Interface Operations Manual xi

Page 14: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Respironics Novametrix 2001Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-5Viewing the Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53-9

Respironics Novametrix 7000/7100Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-1Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-2Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-2Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-3Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-5Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-5Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54-8

SenTec Digital MonitorDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56-10

Smiths Graseby 3000Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57-4Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57-5Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57-15Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57-15Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57-16

Tyco Nellcor N-395/N-595/N-600Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58-9

Flexport System Interface Operations Manual xii

Page 15: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Tyco Nellcor N-1000Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59-1Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59-2Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59-2Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59-3Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59-4Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59-5Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59-5Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59-6

Tyco Nellcor N-2500Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-8

Tyco Nellcor N6000Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-6Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61-9

Tyco Nellcor Symphony N3000Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-6Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62-9

Tyco Puritan Bennett 840Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-7Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-8Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63-9

Flexport System Interface Operations Manual xiii

Page 16: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Table of Contents

Tyco Puritan Bennett 7200A/AEDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-7Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-8Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64-11

Transonics HT-107Directory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65-3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65-4Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65-5Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65-6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65-7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65-7

VIASYS Ventilator, VelaDirectory of Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-1Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-4Screen Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-7Alarm Recordings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-7Entering Setup Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-8Printing the Current Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66-11

Flexport System Interface Operations Manual xiv

Page 17: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Introduction

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Flexport System Interface Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Supporting Data Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Help Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

OverviewThe Flexport® system interface provides access to monitoring data at the patient bedside. It enables data from compatible third-party peripheral devices such as ventilators, multigas analyzers, pulse oximeters, NIBP monitors, IV pumps, incubators, and capnographs to be displayed on Spacelabs Healthcare monitors.

The Flexport system interface provides current numeric data, alarm information, and selected waveforms to the Spacelabs Healthcare monitor, enables you to print the current display, and can support data transfer via the Data Shuttle® option (refer to Supporting Data Transfer on page 1-3).

Once device data is in the Spacelabs Healthcare monitor, it becomes an integral part of the monitoring system and can be communicated over the network to other locations providing alarms, centralized display, trending, and documentation capabilities at remote locations.

Flexport System Interface BasicsThe interface uses RS-232 serial communications to collect data and then transmits the data, via synchronous data link control (SDLC) communications, to the Spacelabs Healthcare monitor.

The Flexport system interface has a male, 9-pin, SDLC connector at one end and either a female, 6-pin, modular jack or a female, 8-pin receptacle at the other end (refer to Figure 1-1). The interface contains no operator controls. Power to the interface is provided by the Spacelabs Healthcare monitor through the SDLC connection.

Figure 1-1: Flexport system interface

9-pin SDLC connector 6-pin modular jack 8-pin receptacle

Flexport System Interface Operations Manual 1-1

Page 18: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Introduction

SetupThe Flexport system interface uses two cables:

• The “L-shaped” SDLC cable connects the Spacelabs Healthcare monitor to the Flexport system interface.

• The modular cable has modular connectors (similar to U.S. telephone connectors) at each end and connects the Flexport system interface to the peripheral device.

Figure 1-2: Flexport system interface connectors

The transition connector adapts the modular cable to the peripheral device.

The SDLC terminator is used at one end of the SDLC cable when the SDLC cable is not used to connect additional Spacelabs Healthcare products to the Spacelabs Healthcare monitor.

Note:• After receiving your Flexport installation kit, contact your Field Service Engineer or your Biomedical

Department to install your SDLC cable and terminator on your monitor or module housing.

• Additional installation details are available in the 90485/86/91/99 Module Housings and Power Supplies Service Manual (P/N 070-0680-xx, located on CD-ROM 084-0700-xx).

SDLC cable SDLC terminatortransition connector

Flexport System Interface Operations Manual 1-2

Page 19: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Introduction

Connecting the Flexport System InterfaceThe Flexport system interface connects to a peripheral device such as a ventilator, IV pump, multigas analyzer, or capnograph as shown in Figure 1-3.

Note:Your device setup may differ from the graphic shown here. Refer to the chapter corresponding to your peripheral device for specific setup information.

Figure 1-3: Flexport system interface connections

Caution:Attach only Spacelabs Healthcare-approved accessories to RS-232 connectors.

Supporting Data TransferIf you use the interface with a Spacelabs Healthcare monitor that includes a multi-parameter module with the Data Shuttle option, you can use that module’s data transfer capability to pass data from one monitor to another. To ensure the integrity of your data, you must finish data collection before you begin the transfer process. To end data collection, disconnect the modular cable from the interface.

For further information on the Data Shuttle option, refer to the Bedside Monitors chapter in the Ultraview SL® Operations Manual (P/N 070-1150-XX, located on CD-ROM 084-1101-xx).

SpacelabsHealthcare

monitor

SDLCcable

SDLCterminator

transitionconnector

modularcable peripheral

device

systemFlexport

interface

Flexport System Interface Operations Manual 1-3

Page 20: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Introduction

Help MessagesIf you are in doubt about a key, touch the monitor HELP key, and then touch the key in question. The monitor will display a brief description of its function.

Flexport System Interface Operations Manual 1-4

Page 21: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Directory of Keys

IV

IV - MAIN MENU

HISTORY SETUP EVENT DISCONNECT PRINT

Refer to page 2-2

IV - DISCONNNECTPUMP

1PUMP

2PUMP

3PUMP

4 ALL

IV - DISCONNNECT PUMP 1?YES NO

IV - EVENTPUMP

1PUMP

2PUMP

3PUMP

4 ALL

IV - HISTORY

PAGE ↑ PAGE ↑ PRINT

Flexport System Interface Operations Manual 2-1

Page 22: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

IV - MAIN MENU

HISTORY SETUP EVENT DISCONNECT PRINT

Refer to page 2-1

Refer to page 2-1

Refer to page 2-1

IV - SETUPPUMP

1PUMP

2PUMP

3PUMP

4 WEIGHT CHANGE CONFIG

IV - PUMP XXPRIMARY SELECT

FLUIDSELECT DRUG

SELECT CONC.

SELECT DILUENT DOSE CALCS

ALARMS

SECONDARY ON OFF

IV - CHANGE CONFIGURATION - SELECT PARAMETER TO EDIT

EDIT DRUG EDIT FLUID

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = ___ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

Flexport System Interface Operations Manual 2-2

Page 23: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

OverviewThe 90437A-01 Flexport System Interface enables information from Abbott Lifecare Models 4, 4P, or Micro pumps to be displayed on Spacelabs Healthcare monitors.

Configure the infusion pump as shown in Table 1. Consult the pump manual for additional configuration setup procedures.

Pump AddressesThe Spacelabs Healthcare monitor display identifies each IV pump by its address. The Abbott pump address is determined by the transition connector provided by Spacelabs Healthcare. The last two digits of the part number molded into the transition connector specify its address. Addresses range from 01 to 16.

Note:If you use multiple pumps, each pump must have a different address.

Table 1: Infusion Pump Configuration

BaudRate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

1200 8-bit none 1 131-0798-xx (15-pin) 650-0293-xx* 235

xx equals the desired address from 01 to 16.*Multi-pump adapter.

Flexport System Interface Operations Manual 2-3

Page 24: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the pump(s), complete the steps in the following Quickstart.

Figure 2-1: Multiple pump configuration

The Spacelabs Healthcare monitor is now ready to display information from the IV pump(s). To control other interface functions, refer to the sections that follow.

To connect the Flexport system interface to the pump(s):

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface.• Plug one end of the modular cable into the modular jack on the Flexport system

interface.

Single pump (refer to Figure 1-3 on page 1-3):

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the connector on the pump.

Multi-pump (refer to Figure 2-1 on page 2-4):

• Plug the other end of the modular cable into the jack labeled FLEXPORT on the multi-pump adapter.

• Plug another modular cable into one of the four IV PUMP jacks on the multi-pump adapter.

• Plug the other end of that modular cable into a transition connector.• Plug the transition connector into the connector on the pump.

SDLCcable

SDLCterminator

modularcables

IVSDLC connector

transition connector

9-pin connector

pump

IVpump

IVpump

IVpump

multi-pumpadapter

SpacelabsHealthcare

monitor

systemFlexport

interface

Flexport System Interface Operations Manual 2-4

Page 25: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pump(s).

Figure 2-2: Bedside, remote, or central monitor, full screen

Figure 2-3: Central monitor, split screen

� Pump address.

� Selected medication or fluid. (Until you set up the drug or fluid for each pump, this field displays * * * * * * * *.)

� Selected pump or infusion rate. Initially, this field displays the pump rate. The calculated medication rate is displayed once you have selected the concentration or dose. Units for displays are based on those selected.

� The volume of fluids and the dose delivered since the last time one of the following conditions occurred:

• The pump started a new dose.

• The pump rate was titrated.

• The pump changed from secondary to primary or from primary to secondary.

� IV parameter key.

� Latest status message (for example, PUMPING). The displayed infusion may be either the primary or the secondary. The secondary infusion is displayed when the pump is pumping secondary (SEC PUMPING) or concurrently (CONCURRENT). Refer to Status Messages on page 2-19.

� � � � � �

PUMP DRUG/FLUID RATE VOL. INFUSED PUMP STATUS12 DOPAMINE 10.0 mcg/kg/min 122.1 ml I

V12 PUMPING

14 HEPARIN 1000.0 u/hr 233.4 ml 14 STOPPED9 NOREPINEPHRINE 3.0 mcg/min 10.9 ml 9 PUMPING4 D5W 250.0 ml/hr 10.0 ml 4 INFSN DONE

� � � �

PUMP DRUG/FLUID STATUS24 DOPAMINE PUMPING I

V88 HEPARIN STOPPED92 NOREPINEPHRINE SEC PUMPING44 D5W INFSN DONE

Flexport System Interface Operations Manual 2-5

Page 26: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Entering Setup InformationSetup information is necessary for the monitor to display the drug (or fluid) names and to compute the medication and calculated pump rates. You can control the following functions from the Setup menu:

• Turn alarms ON or OFF.

• Enter patient weight.

• Edit drug or fluid names.

• Set up primary and secondary infusion lines.

• Select the infusion fluid.

• Select the infusion drug.

Once you select the infusion drug, you can perform the following functions:

• Enter the infusion concentration.

• Select the diluent fluid.

• Enter the infusion dose.

Note:Setup information can be entered only at the bedside monitor.

Enabling AlarmsWhen you turn alarms OFF, infusion pump alarms will no longer trigger an alarm on the Spacelabs Healthcare monitor.

Alarm settings default to ON.

Note:Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the infusion pump(s).

To turn alarms ON or OFF:

• Touch IV.• Touch SETUP.• Select a PUMP key.• Touch ALARMS ON/OFF.

Flexport System Interface Operations Manual 2-6

Page 27: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Selecting an Infusion LineYou cannot set up the secondary line until you complete the primary infusion setup.

Note:The Abbott Lifecare Model 4P is the only pump that accommodates piggyback or secondary infusions.

Selecting the Infusion FluidThere are two pages of Select Fluid screens. The first page displays the list of fluids automatically supplied by the Flexport system interface. The second page enables you to select from a list of up to eight fluids that you can enter into the Flexport system interface’s memory. Refer to Editing Drug or Fluid Names on page 2-14 for information on entering these names.

Figure 2-4: Select Fluid screen

To select an infusion line:

• Touch IV.• Touch SETUP.• Select a PUMP key.• Touch PRIMARY/SECONDARY.

PUMP 12 - SELECT FLUIDLR D5 LR D5 NS D5 1/2 NS

D5 1/2 NS 20 KCL D5 1/4 NS D5W D10 NS

D10 1/2 NS D10 1/4 NS D10W NS

1/2 NS 1/4 NS TPN LIPIDS

PAGE ↑ PAGE ↓ ENTER

To select an infusion fluid:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT FLUID.• Select a fluid from the list (or touch PAGE ↓ and then select a fluid).• Touch ENTER.

Flexport System Interface Operations Manual 2-7

Page 28: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Table 2 defines the fluid/concentration abbreviations that correspond to the fluids listed on the Select Fluid screen.

Table 2: Fluid Abbreviation

Abbreviation Description

LR Lactated ringers

D5 LR 5% dextrose with lactated ringers

D5 NS 5% dextrose with 0.9% saline

D5 1/2NS 5% dextrose with 0.45% saline

D5 1/2NS 20 KCL 5% dextrose with 0.45% saline and 20 mEq potassium chloride/liter

D5 1/4NS 5% dextrose with 0.2% saline

D5W 5% dextrose with water

D10 NS 10% dextrose with 0.9% saline

D10 1/2NS 10% dextrose with 0.45% saline

D10 1/4NS 10% dextrose with 0.2% saline

D10W 10% dextrose with water

NS 0.9% saline

1/2 NS 0.45% saline

1/4 NS 0.2% saline

TPN Total parenteral nutrition

LIPIDS Fat emulsion delivered as part of parenteral nutrition

Flexport System Interface Operations Manual 2-8

Page 29: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Selecting a DrugThe Select Drug screen provides 16 default drug names. Refer to Editing Drug or Fluid Names on page 2-14 for information on editing the default drug names.

Figure 2-5: Select Drug screen

After selecting a drug, the monitor displays (in order) screens for the concentration and diluent.

Selecting the Drug ConcentrationThe drug concentration represents the amount of a drug dissolved in a volume of diluent. The Flexport system interface saves the last entered default concentration values and unit of measure associated with the selected drug.

Access to this display is automatic from the Select Drug screen.

To change a value after you have completed an initial drug/fluid entry, complete all the steps listed in the following Quickstart. Figure 2-6 illustrates the Concentration screen.

PUMP 12 - SELECT DRUGAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

ENTER

To select a drug:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DRUG.• Select a drug from the list.• Touch ENTER.

Flexport System Interface Operations Manual 2-9

Page 30: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Figure 2-6: Concentration screen

Caution:• The validity of the drug concentration or dose values that you enter is NOT checked by the

system.

• Calculated drug rates are displayed to one significant decimal place. Rates less than 0.05 will appear as 0.0.

PUMP 12 - ENTER DRUG QTY, UNITS, AND VOLUME. CONCENTRATION = _______

7 8 9 mcg mg /50 ml /100 ml

4 5 6 g ml /250 ml /500 ml

1 2 3 units mEq /1 L /2 L

0 . CLEAR ENTER

To select a drug concentration:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT CONC.• Enter the drug quantity.• Select the unit of measure.• Select the fluid volume.• Touch ENTER.

Flexport System Interface Operations Manual 2-10

Page 31: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Selecting the DiluentTo select a drug for infusion, you must first choose a drug from the Select Drug page. The selected drug becomes active upon touching the ENTER key. The Flexport system interface provides 16 default drug names. Refer to Editing Drug or Fluid Names on page 2-14 for information on editing those default drug names.

Access to this display is automatic from the Concentration screen.

To change your selection after you have completed an initial diluent entry, complete all the steps listed in the following Quickstart.

When you touch ENTER to complete the diluent entry, the monitor displays the Setup menu.

Selecting the DoseThe dose represents the amount of drug to be infused over a specified time period. Use this display to enter the quantity, units of measurement, and time units.

If you have completed an initial drug/fluid entry and wish to change a value, complete all the steps listed in the Quickstart on page 2-12.

Note:If the interface displays an initial VOLUME INFUSED for the pump you intend to use, you must clear that value at the pump before you start an infusion. Otherwise, the Flexport system interface will appear to add the volume infused from the previous infusion to the current infusion (either primary or secondary).

To select a diluent:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DILUENT.• Choose a diluent from the list (or touch PAGE↓ and then select a diluent).• Touch ENTER.

Flexport System Interface Operations Manual 2-11

Page 32: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Figure 2-7: Dose Calcs screen

As you enter the quantity, unit of measurement, and unit of time, the monitor screen displays the complete dose and the calculated pump rate. This calculated pump rate is based on the concentration and dose values you enter. To arrive at this value, the Flexport system interface performs the following calculation:

CALCULATED PUMP RATE = DOSE/CONCENTRATION

Note:The calculated pump rate is provided as information only. The Flexport system interface does not control the actual set pump rate.

When you touch ENTER to complete the dose entry, the interface saves your selection and the monitor displays the Setup menu.

PUMP 12 - ENTER DRUG QUANTITY, UNITS, AND TIME. DOSE = _______

7 8 9 mcg mg /min /hr

4 5 6 g ml CALCULATED PUMP RATE= xxxx.x ml/hr

1 2 3 units mEq

0 . CLEAR mcg/kg mg/kg ENTER

To select the drug dose:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch DOSE CALCS.• Enter the dose quantity.• Enter unit of measurement.• Enter the time unit.• Touch ENTER.

Flexport System Interface Operations Manual 2-12

Page 33: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Setting Patient WeightTouching the WEIGHT key displays a screen with a prompt containing the current weight of the patient (if you entered it previously using the Weight screen, a cardiac output module, or the monitor admit function). The Patient Weight screen uses the same units as the bedside monitor (kg or pounds).

Figure 2-8: Patient Weight screen

After you complete your editing and touch ENTER, the current weight will appear on the appropriate displays. You can enter up to six characters, including the decimal point. The valid weight range is from 1 to 250 kg (2 to 551 pounds). The interface saves the weight so you can use it in the Dose Calcs screen with the mcg/kg and mg/kg units of measurement.

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = __ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

To enter patient weight:

• Touch IV.• Touch SETUP.• Touch WEIGHT.• Enter the patient weight.• Touch ENTER.

Flexport System Interface Operations Manual 2-13

Page 34: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Editing Drug or Fluid NamesThe Change Configuration screen for drug names (refer to Figure 2-9) initially lists the 16 default drug names. You can change the name of any of these drugs.

The Change Configuration screen for fluid names initially shows a page of eight blank keys. You can select any of the eight keys for the fluid name.

Figure 2-9: Change Drug Configuration screen

After you select a drug or fluid key to relabel, touch CHANGE NAME. The monitor displays the Change Drug (or Fluid) Name screen (refer to Figure 2-10).

To edit a drug or fluid name:

• Touch IV.• Select SETUP.• Touch CHANGE CONFIG.• Select EDIT DRUG or EDIT FLUID.• Select the name you want to change.• Touch CHANGE NAME.• Use the alphanumeric keys to make necessary changes.• Touch ENTER.

IV - CHANGE CONFIGURATION - SELECT DRUG KEY TO RELABELAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

CHANGE NAME

Flexport System Interface Operations Manual 2-14

Page 35: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Figure 2-10: Change Drug (or Fluid) Name screen

• Enter each character or number using the alphanumeric keys.

• Touch DEL to eliminate the character highlighted by the cursor.

• Move through the name, character by character, using the arrow keys.

• Touch CLEAR to eliminate the entire name and start over. The screen displays blank fields in place of the name.

• Touch ENTER to complete the edit and store the name in memory.

• Enter up to fifteen characters.

After you complete your editing and touch ENTER, the name will appear in the appropriate display for selecting drugs or fluids. The interface saves the name so you can recall it the next time you connect the Flexport system interface.

IV-CHANGE DRUG (or fluid) NAME= _________________________

0 1 2 3 4 5 6 7 8 9

A B C D E F G H I J

K L M N O P Q R S T

U V W X Y Z / % .

DEL ← → CLEAR ENTER

Flexport System Interface Operations Manual 2-15

Page 36: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Disconnecting a PumpWhen you touch DISCONNECT, the Flexport system interface discontinues the infusion record. If the pump is operating when you touch this key, the infusion record stops, a new record starts, and all pump setup information is cleared from the interface. If you touch DISCONNECT when the pump is not operating, the setup information stored by the Flexport system interface is cleared and the line of pump data is removed from the display.

Note:• Touching the DISCONNECT key does NOT turn OFF the infusion pump.

• A pump can be disconnected only at the bedside monitor.

Do not turn the power OFF at the pump while connected to the Flexport system interface. Disconnect the module cable from the pump first.

When you turn an infusion pump OFF at the pump itself, the pump status listed on the monitor changes to DISCONNECTED. At this point the interface completes the history for the current infusion (HISTORY shows the time that the infusion ended). If you turn the pump ON again, the interface continues to display the drug/fluid information entered previously.

Noting an EventThe EVENT key stores the time and pump status in memory. There is no immediate feedback when you touch this key. When you review the pump history, there is an entry labeled EVENT at the time you noted.

Events you may want to record include procedures such as changing IV lines or bags without turning the pump OFF.

Note:An event can be noted only at a bedside monitor.

To disconnect a pump:

• Touch IV.• Touch DISCONNECT.• Select a PUMP key.• Touch YES.

To note an event:

• Touch IV.• Touch EVENT.• Select a PUMP key.

Flexport System Interface Operations Manual 2-16

Page 37: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Displaying Infusion HistoryInfusion history displays in a table format listing the last 40 infusion events for the patient on four pages (refer to Figure 2-11). This view-only table is available at the bedside monitor only. It is not available for remote viewing.

The data shown in the table does not automatically update while you are viewing it. To update the history data, re-enter the history table by touching PREVIOUS and then HISTORY.

Figure 2-11: Example of History table

The word ACTIVE appearing in the Ended column indicates that the pump is presently infusing the drug or fluid. The interface updates the log of active entries each minute.

The word EVENT appearing in the Ended column indicates the time at which the clinician touched the EVENT key to record a miscellaneous event into the pump history. Events that you might want to record include changing infusion lines or bags without turning the pump OFF.

To display infusion history:

• Touch IV.• Touch HISTORY.

DATE BEGAN ENDED PUMP DRUG/FLUID CONCENTRATION FLOW RATE

VOL. INFSD

02/20 10:30am 12:10pm 12 DOPAMINE 400mg/250ml 30.0 ml/hr 50.0 ml

02/20 12:15pm EVENT 12 DOPAMINE 400mg/250ml 20.0 ml/hr 3.5ml

02/20 12:50pm 05:00pm 12 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/20 06:00pm 11:00pm 14 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/21 12:30am 01:00am 12 NOREPINEPHRINE 1mg/250ml 125.0 ml/hr 62.5 ml

02/21 01:30am ACTIVE 12 NOREPINEPHRINE 400mg/250ml 30.0 ml/hr 210.0 ml

02/21 01:30am 03:00am 14 HEPARIN 2000u/500ml 100.0 ml/hr 750.0 ml

02/21 03:30am 07:00am 13 D5W 100.0 ml/hr 350.0 ml

02/21 07:30am ACTIVE 14 D10 1/2NS 100.0 ml/hr 100.0 ml

02/21 08:00am ACTIVE 9 DOPAMINE 400mg/250ml 30.0 ml/hr 15.0 ml

Flexport System Interface Operations Manual 2-17

Page 38: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display or infusion history table. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the infusion information:

• Touch IV.• Touch PRINT.

-OR-• Touch HISTORY.• Touch PRINT.

Flexport System Interface Operations Manual 2-18

Page 39: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Abbott Lifecare 4, 4P, or Micro

Status MessagesThe message DISCONNECTED indicates a problem with the pump or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pump is powered ON.

• Verify that all cables are secure.

Table 3 lists the status messages that the Flexport system interface reports for the pump during Flexport system interface operation.

Note:Not all messages will appear for each pump. Refer to the manual for your specific pump to determine which messages are used.

Table 3: Status Messages

Message Text Indication

DISCONNECTED Pump has been disconnected or turned OFF.

HIGH FLOW Pump has detected high flow condition.

INFSN. DONE Primary infusion is complete.

KVO Pump pumping at KVO rate.

LOW BATTERY Battery requires recharging.

LOW FLOW Pump has detected low flow condition.

NO SENSOR Flow detector is not connected and dose function is disabled.

OCCLUSION Closed clamp, kinked or occluded tubing, or occluded filter detected.

OK Appears at pump sign ON.

PUMPING Pump is pumping normally.

PUMP ALARM Pump has detected an alarm condition.

SEC PUMPING Pump is operating the secondary line.

STOPPED Pump is stopped but not turned OFF.

Flexport System Interface Operations Manual 2-19

Page 40: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 41: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Directory of Keys

IV

IV - MAIN MENU

HISTORY SETUP EVENT DISCONNECT PRINT

IV - DISCONNNECT MENUPUMP

1PUMP

2PUMP

3PUMP

4 ALL

IV - DISCONNNECT PUMP XX?YES NO

IV - EVENT MENUPUMP

1PUMP

2PUMP

3PUMP

4 ALL

IV - SETUP MENUPUMP

1PUMP

2PUMP

3PUMP

4 WEIGHT CHANGE CONFIG

IV - PUMP XX MENUPRIMARY SELECT

FLUIDSELECT DRUG

SELECT CONC.

SELECT DILUENT

DOSE CALCS

ALARMS

SECONDARY ON OFF

IV - HISTORY MENU IV - CHANGE CONFIGURATION - SELECT PARAMETER TO EDIT

PAGE ↑ PAGE ↓ PRINT EDIT DRUG

EDIT FLUID

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = ___ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

Flexport System Interface Operations Manual 3-1

Page 42: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 43: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

OverviewThe 90437A-03 Flexport system interface enables information from an Alaris IVAC 560, 570, or 599 infusion pump to be displayed on Spacelabs Healthcare monitors.

Configure the infusion pump as shown in Table 1. Consult the pump manual for additional configuration setup procedures.

Pump AddressesThe Spacelabs Healthcare monitor display identifies each IV pump by its address. This address is set at the pump. The addresses range from 01 to 16.

Note:If you use multiple pumps, each pump must have a different address.

The multi-pump adapter enables you to connect up to four pumps to a single Flexport system interface.

Table 1: Infusion Pump Configuration

Infusion Pump Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

Alaris IVAC 560/570 with CIM II 1200 7 bit odd 1

131-0688-xx(9 pin)

Model 560 rev 0.19Model 570 rev 0.09

Alaris IVAC 599 with DCM 1200 8 bit forced 0 1

131-0686-xx(9 pin)

rev 10.0

Flexport System Interface Operations Manual 3-3

Page 44: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the pump(s), complete the steps listed in the following Quickstart.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

Single pump (refer to Figure 1-3 on page 1-3):• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the connector on the pumpMulti-pump (refer to Figure 3-1 on page 3-5):• Plug the other end of the modular cable into the jack labeled FLEXPORT on the

multi-pump adaptor.• Plug another modular cable into one of the four IV PUMP jacks on the multi-pump

adapter.• Plug the other end of that modular cable into a transition connector.• Plug the transition connector into the connector on the pump.• Repeat steps step 2 through step 4 for each pump.

Flexport System Interface Operations Manual 3-4

Page 45: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Figure 3-1: Multiple pump configuration

The Spacelabs Healthcare monitor is now ready to display information from the IV pump(s). To control other interface functions, refer to the sections that follow.

SDLCcable

SDLCterminator

transitionconnector

modularcables

IVpump

IVpump

IVpump

IVpump

multi-pumpadapter

SpacelabsHealthcare

monitor

systemFlexport

interface

Flexport System Interface Operations Manual 3-5

Page 46: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pump(s).

Figure 3-2: Bedside, remote, or central monitor, full screen

Figure 3-3: Central monitor, split screen

� Pump address.

� Selected medication or fluid. (Until you set up the drug or fluid for each pump, this field displays * * * * * * * *).

� Selected pump or infusion rate. Initially, this field displays the pump rate. The calculated medication rate is displayed once you have selected the concentration or dose. Units for displays are based on those selected.

� The display of the volume of fluids infused varies depending on the pump model and the infusion mode:

• Alaris IVAC 560, 570 Primary and Secondary, and 599 Primary mode: Displays the volume of fluids infused since the last time you zeroed the “total volume infused” on the pump.

• Alaris IVAC 599 Secondary mode: Displays the volume of fluids infused since the secondary infusion was started. If the pump’s “total volume infused” is zeroed during a secondary infusion, the display will continue to accumulate. (This is distinguished from the PRIMARY mode in which the display starts from zero again if the “total volume infused” was zeroed.)

� IV parameter key.

� Latest status message (for example, PUMPING). The displayed infusion may be either the primary or the secondary. The secondary infusion is displayed when the pump is pumping secondary (SEC PUMPING) or concurrently (CONCURRENT). Refer to Status Messages on page 3-20.

� � � � � �

PUMP DRUG/FLUID RATE VOL. INFUSED PUMP STATUS12 DOPAMINE 10.0 mcg/kg/min 122.1 ml I

V12 PUMPING

14 HEPARIN 1000.0 u/hr 233.4 ml 14 STOPPED9 NOREPINEPHRINE 3.0 mcg/min 10.9 ml 9 SEC PUMPING4 D5W 250.0 ml/hr 10.0 ml 4 INFSN DONE

� � � �

PUMP DRUG/FLUID STATUS12 DOPAMINE PUMPING I

V14 HEPARIN STOPPED9 NOREPINEPHRINE SEC PUMPING4 D5W INFSN DONE

Flexport System Interface Operations Manual 3-6

Page 47: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Entering Setup InformationSetup information is necessary for the monitor to display the drug (or fluid) names and to compute the medication and calculated pump rates. The SETUP key controls the following functions and effects only the monitor at which these adjustments are made:

• Turn alarms ON or OFF.

• Enter patient weight.

• Edit drug or fluid names.

• Set up primary and secondary infusion lines.

• Select the infusion fluid.

• Select the infusion drug.

Once you select the infusion drug, you may perform the following functions:

• Enter the infusion concentration.

• Select the diluent fluid.

• Enter the infusion dose.

Enabling Alarms When you turn alarms OFF, infusion pump alarms will no longer trigger an alarm on the Spacelabs Healthcare monitor.

The alarm setting defaults to ON.

Note:Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the infusion pump.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch ALARMS ON/OFF.

Flexport System Interface Operations Manual 3-7

Page 48: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Selecting an Infusion LineYou cannot set up the secondary line until you complete the primary infusion setup. The Alaris IVAC model 570 and 599 pump’s controlled release infusion system (CRIS) mode is considered a modified SECONDARY mode. The drug that is infused using the CRIS mode adapter is entered into the Flexport’s SECONDARY setup.

Note:The model 560 pump does not support secondary infusion.

Selecting the Infusion FluidThere are two pages of Select Fluid screens. The first page displays the list of fluids automatically supplied by the Flexport system interface. The second page enables you to select from a list of up to eight fluids that you can enter into the Flexport system interface’s memory. Refer to Editing Drug or Fluid Names on page 3-15 for information on entering these names.

To select an infusion line:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch PRIMARY/ SECONDARY.

To select an infusion fluid:

• Touch IV.• Touch SETUP.• Select a PUMP key.• Touch SELECT FLUID.

-OR-• Touch PAGE ↓ and then select a fluid.• Touch ENTER.

Flexport System Interface Operations Manual 3-8

Page 49: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Figure 3-4: Select Fluid screen

Table 2 defines the fluid/concentration abbreviations that correspond to the fluids listed on the Select Fluid screen.

PUMP 12 - SELECT FLUIDLR D5 LR D5 NS D5 1/2 NS

D5 1/2 NS 20 KCL D5 1/4 NS D5W D10 NS

D10 1/2 NS D10 1/4 NS D10W NS

1/2 NS 1/4 NS TPN LIPIDS

PAGE ↑ PAGE ↓ ENTER

Table 2: Fluid Abbreviations

Abbreviation Description

LR lactated ringers

D5 LR 5% dextrose with lactated ringers

D5 NS 5% dextrose with 0.9% saline

D5 1/2NS 5% dextrose with 0.45% saline

D5 1/2NS 20KCL 5% dextrose with 0.45% saline and 20 mEq Potassium Chloride/Liter

D5 1/4NS 5% dextrose with 0.2% saline

D5W 5% dextrose with water

D10 NS 10% dextrose with 0.9% saline

D10 1/2NS 10% dextrose with 0.45% saline

D10 1/4NS 10% dextrose with 0.2% saline

D10 W 10% dextrose with water

NS 0.9% saline

1/2 NS 0.45% saline

1/4 NS 0.2% saline

TPN Total parenteral nutrition

LIPIDS Fat emulsion delivered as part of parenteral nutrition

Flexport System Interface Operations Manual 3-9

Page 50: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Selecting a DrugTo select a drug for infusion, you must first choose a drug from the Select Drug page. The selected drug becomes active upon touching the ENTER key. The Flexport system interface provides 16 default drug names. Refer to Editing Drug or Fluid Names on page 3-15 for information on editing those default drug names.

Figure 3-5: Select Drug screen

After selecting a drug, the monitor displays (in order) the concentration and diluent.

Selecting the Drug ConcentrationThe drug concentration represents the amount of a drug dissolved in a volume of diluent. The Flexport system interface saves the last entered default concentration values and unit of measure associated with the selected drug. (Refer to Figure 3-6.)

Access to this display is automatic from the Select Drug screen; continue to complete the steps from the fifth bullet in the following Quickstart.

To change a value after you have completed an initial drug/fluid entry, follow all of the steps listed in the following Quickstart.

PUMP 12 - SELECT DRUGAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

ENTER

To select a drug:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DRUG.• Select a drug from the list.• Touch ENTER.

Flexport System Interface Operations Manual 3-10

Page 51: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Figure 3-6: Concentration screen

Caution:• The validity of the drug concentration or dose values that you enter is NOT checked by the

system.

• Calculated drug rates are displayed to one significant decimal place. Rates less than 0.05 will appear as 0.0.

Selecting the DiluentA diluent is a fluid used to dilute a drug for infusion. The fluids you can choose as drug diluents are the same as those you can choose as infusion fluids. Refer to Selecting the Infusion Fluid on page 3-8 for details on selecting a diluent. The Flexport system interface saves the last entered diluent associated with a selected drug.

To select a drug concentration:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT CONC.• Enter the drug quantity.• Select the unit of measure.• Select the volume of fluid.• Touch ENTER.

PUMP 12 - ENTER DRUG QTY, UNITS, AND VOLUME. CONCENTRATION = _______

7 8 9 mcg mg /50 ml /100 ml

4 5 6 g ml /250 ml /500 ml

1 2 3 units mEq /1 L /2 L

0 . CLEAR ENTER

Flexport System Interface Operations Manual 3-11

Page 52: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Access to this display is automatic from the Concentration screen. Continue to complete the steps after -OR- in the following Quickstart.

To change your selection after you have completed an initial diluent entry, complete all of the steps listed in the following Quickstart.

When you touch ENTER to complete the diluent entry, the monitor displays the Setup menu.

Selecting the DoseThe dose represents the amount of drug to be infused over a specified time period. Use this display to enter the quantity, units of measurement, and time units.

If you have completed an initial drug/fluid entry and wish to change a value, complete all of the steps listed in the following Quickstart.

Note:If the Flexport system interface displays an initial VOLUME INFUSED for the pump you intend to use, you must clear that value at the pump before you start an infusion. Otherwise, the Flexport system interface will appear to add the volume infused from the previous infusion to the current infusion (either primary or secondary).

To select a diluent:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DILUENT.• Choose a diluent from the list.

-OR-• Touch PAGE ↓ and then select a diluent.• Touch ENTER.

To select the drug dose:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch DOSE CALCS.• Enter the dose quantity.• Enter the unit of measurement.• Enter the time unit.• Touch ENTER.

Flexport System Interface Operations Manual 3-12

Page 53: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Figure 3-7: Dose Calcs screen

As you enter the quantity, unit of measurement, and unit of time, the monitor screen displays the complete dose and the calculated pump rate. This calculated pump rate is based on the concentration and dose values you enter. To arrive at this value, the Flexport system interface performs the following calculation:

CALCULATED PUMP RATE = DOSE/CONCENTRATION

Note:The calculated pump rate is provided as information only. The Flexport system interface does not control the actual set pump rate.

When you touch ENTER to complete the dose entry, the interface saves your selection and the monitor displays the Setup menu.

PUMP 12 - ENTER DRUG QUANTITY, UNITS, AND TIME. DOSE = _______

7 8 9 mcg mg /min /hr

4 5 6 g ml CALCULATED PUMP RATE= xxxx.x ml/hr

1 2 3 units mEq

0 . CLEAR mcg/kg mg/kg ENTER

Flexport System Interface Operations Manual 3-13

Page 54: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Setting Patient WeightTouching the WEIGHT key displays a screen with a prompt that contains the current weight of the patient (if you entered it previously using the Weight screen, a cardiac output module, or the monitor admit function). The Patient Weight screen uses the same units as the bedside monitor (kg or pounds).

Figure 3-8: Patient Weight screen

After you complete your editing and touch ENTER, the current weight will appear in the appropriate displays. You can enter up to six characters, including the decimal point. The valid weight range is from 1 to 250 kg (2 to 551 pounds). The Flexport system interface saves the weight so you can use it in the Dose Calcs screen with the mcg/kg and mg/kg units of measurement.

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = __ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

To enter patient weight:

• Touch IV.• Touch SETUP.• Touch WEIGHT.• Enter the patient weight.• Touch ENTER.

Flexport System Interface Operations Manual 3-14

Page 55: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Editing Drug or Fluid NamesThe Change Configuration screen for drug names initially lists the 16 default drug names. You can change the name of any of these drugs (refer to Figure 3-9).

The Change Configuration screen for fluid names initially displays eight blank keys. You can select any of the eight keys for the fluid name.

Figure 3-9: Change Drug Configuration screen

After you select a drug or fluid key to relabel, touch CHANGE NAME. The monitor displays the Change Drug (or Fluid) Name screen (refer to Figure 3-10).

Note:Drug or fluid names can be edited only at the bedside monitor.

To edit a drug or fluid name:

• Touch IV.• Select SETUP.• Touch CHANGE CONFIG.• Select EDIT DRUG or EDIT FLUID.• Select the name you want to change.• Touch CHANGE NAME.• Use the alphanumeric keys to make necessary changes.• Touch ENTER.

IV - CHANGE CONFIGURATION - SELECT DRUG KEY TO RELABELAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

CHANGE NAME

Flexport System Interface Operations Manual 3-15

Page 56: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Figure 3-10: Change Drug (or Fluid) Name screen

• Enter each character or number using the alphanumeric keys.

• Touch DEL to eliminate the character highlighted by the cursor.

• Move through the name, character by character, using the arrow keys.

• Touch CLEAR to eliminate the entire name and start over. The screen displays blank fields in place of the name.

• Touch ENTER to complete the edit and store the name in memory.

• Enter up to fifteen characters.

After you complete your editing and touch ENTER, the name appears in the appropriate display for selecting drugs or fluids. The interface saves the name so you can recall it the next time you connect the Flexport system interface.

IV-CHANGE DRUG (or fluid) NAME= _________________________

0 1 2 3 4 5 6 7 8 9

A B C D E F G H I J

K L M N O P Q R S T

U V W X Y Z / % .

DEL ← → CLEAR ENTER

Flexport System Interface Operations Manual 3-16

Page 57: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Disconnecting a PumpWhen you touch DISCONNECT, the Flexport system interface discontinues the infusion record. If the pump is operating when you touch this key, the infusion record stops, a new record starts, and all pump setup information is cleared from the interface. If you touch DISCONNECT when the pump is not operating, the setup information stored by the Flexport system interface is cleared and the line of pump data is removed from the display.

Note:• Touching the DISCONNECT key does NOT turn OFF the infusion pump.

• A pump can be disconnected only at the bedside monitor.

When you turn an infusion pump OFF at the pump itself, the pump status listed on the monitor changes to DISCONNECTED. At this point the interface completes the history for the current infusion (HISTORY shows the time that the infusion ended). If you turn the pump ON again, the interface continues to display the drug/fluid information entered previously.

Noting an EventThe EVENT key enables the interface to store time and pump status in memory. There is no immediate feedback when you touch this key. When you review the pump history, there is an entry labeled Event at the time you noted.

Events you may want to record include procedures such as changing IV lines or bags without turning the pump OFF.

Note:An event can be noted only at a bedside monitor.

To disconnect a pump:

• Touch IV.• Touch DISCONNECT.• Select a pump key.• Touch YES.

To note an event:

• Touch IV.• Touch EVENT.• Select a pump key.

Flexport System Interface Operations Manual 3-17

Page 58: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Displaying Infusion HistoryInfusion history displays in a table format listing the last 40 infusion events for the patient on four pages (refer to Figure 3-11). This view-only table is available at the bedside monitor only and is not available for remote viewing.

The data shown in the table does not automatically update while you are viewing it. To update the history data, re-enter the history table by touching PREVIOUS and then HISTORY.

Figure 3-11: Example of History table

The word ACTIVE appearing in the Ended column indicates that the pump is presently infusing the drug or fluid. The interface updates the log of active entries each minute.

The word EVENT appearing in the Ended column indicates the time at which the clinician touched the EVENT key to record a miscellaneous event into the pump history. Events that you might want to record include changing infusion lines or bags without turning the pump OFF.

To display infusion history:

• Touch IV.• Touch HISTORY.

DATE BEGAN ENDED PUMP DRUG/FLUID CONCENTRATION FLOW RATE VOL. INFSD

02/20 10:30am 12:10pm 12 DOPAMINE 400mg/250ml 30.0 ml/hr 50.0 ml

02/20 12:15pm EVENT 12 DOPAMINE 400mg/250ml 20.0 ml/hr 3.5 ml

02/20 12:50pm 05:00pm 12 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/20 06:00pm 11:00pm 14 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/21 12:30am 01:00am 12 NOREPINEPHRINE 1mg/250ml 125.0 ml/hr 62.5 ml

02/21 01:30am ACTIVE 12 NOREPINEPHRINE 400mg/250ml 30.0 ml/hr 210.0 ml

02/21 01:30am 03:00am 14 HEPARIN 2000u/500ml 100.0 ml/hr 750.0 ml

02/21 03:30am 07:00am 13 D5W 100.0 ml/hr 350.0 ml

02/21 07:30am ACTIVE 14 D10 1/2NS 100.0 ml/hr 100.0 ml

02/21 08:00am ACTIVE 9 DOPAMINE 400mg/250ml 30.0 ml/hr 15.0 ml

Flexport System Interface Operations Manual 3-18

Page 59: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display or infusion history table. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the infusion information:

• Touch IV.• Touch PRINT.

-OR-• Touch HISTORY.• Touch PRINT.

Flexport System Interface Operations Manual 3-19

Page 60: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris IVAC 560/570/599

Status MessagesThe message DISCONNECTED indicates a problem with the pump or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pump is powered ON.

• Verify that all cables are secure.

Note:On the Alaris IVAC model 599 infusion pump, all serial communications are suspended while the pump rate is being adjusted at the pump. If the pump rate adjustment occurs continuously over an extended time (greater than 20 seconds), the Flexport will consider the communications to be interrupted and display a DISCONNECTED message.

Table 3 lists the status messages that the Flexport system interface reports for the pump during Flexport system interface operation.

Table 3: Status Messages

Message Text Indication

AIR IN LINE Pump has detected air in IV tubing.

CRIS MODE Pump is operating in CRIS MODE.

DISCONNECTED Pump has been disconnected or turned OFF.

DOOR OPEN Pump has detected an open door.

FLOW SENSOR Pump has detected a problem with the flow sensor.

INFSN. DONE Primary infusion is complete.

KVO Pump pumping at KVO rate.

LOW BATTERY Battery requires recharging.

LOW FLOW Low flow condition detected.

OCCLUSION Tubing occlusion detected.

OK Appears at pump sign ON.

PUMPING Pump is pumping normally.

PUMP ALARM Pump has detected an alarm condition.

SEC PUMPING Pump is operating the secondary line.

STOPPED Pump is stopped but not turned OFF.

Flexport System Interface Operations Manual 3-20

Page 61: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Directory of Keys

IV

IV - MAIN MENU

HISTORY SETUP EVENT DISCONNECT PRINT

Refer to page 4-2

IV - DISCONNNECTPUMP

1PUMP

2PUMP

3PUMP

4 ALL

IV - DISCONNNECT PUMP 1?YES NO

IV - EVENTPUMP

1PUMP

2PUMP

3PUMP

4 ALL

IV - HISTORY

PAGE ↑ PAGE ↓ PRINT

Flexport System Interface Operations Manual 4-1

Page 62: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

IV

IV - MAIN MENU

HISTORY SETUP EVENT DISCONNECT PRINT

Refer to page 4-1

Refer to page 4-1

Refer to page 4-1

IV - SETUPPUMP

1PUMP

2PUMP

3PUMP

4 WEIGHT CHANGE CONFIG

IV - CHANGE CONFIGURATION - SELECT PARAMETER TO EDITEDIT

DRUGEDIT

FLUID

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = ___ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

IV - PUMP XXPRIMARY SELECT

FLUIDSELECT DRUG

SELECT CONC.

SELECT DILUENT

DOSE CALCS

ALARMS

SECONDARY ON OFF

Flexport System Interface Operations Manual 4-2

Page 63: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

OverviewThe 90437A-02 Flexport system interface enables information from the Alaris Medical Systems 980C, PC-1, PC-2, or PC-4 infusion pump to be displayed on Spacelabs Healthcare monitors.

Configure the infusion pump as shown in Table 1. Consult the pump manual for additional configuration setup procedures.

Pump AddressesThe Spacelabs Healthcare monitor display identifies each IV pump by its address is set at the pump and ranges from 01 to 16. Alaris refers to the pump address as the serial number. Refer to the manufacturer’s manual for instructions on setting pump addresses.

Note:If you use multiple pumps, each pump must have a different address.

The multi-pump adapter enables you to connect up to four pumps to a single Flexport system interface. The dual-line pump (Gemini PC-2) is considered two pumps that can be connected to one of the following:

• Four single-line pumps (for example, 980C or Gemini PC-1)

• Two dual-line pumps (for example, Gemini PC-2)

• One dual-line pump and up to two single-line pumps

• One quadruple-line pump for Gemini PC-4

Table 1: Pump Configuration

Baud Rate

Character Length

Parity Setting Stop Bit

1200 8 bit none 1

Flexport System Interface Operations Manual 4-3

Page 64: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Connecting the Flexport System InterfaceTo connect the Flexport system interface to multiple pumps, complete the steps listed in the following Quickstart.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface.• Plug one end of the modular cable into the modular jack on the Flexport system

interface.

Single pump (refer to Figure 1-3 on page 1-3):

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the connector on the pump.

Multi-pump (refer to Figure 4-1 on page 4-5):

• Plug the other end of the modular cable into the jack labeled FLEXPORT on the multi-pump adaptor.

• Plug another modular cable into one of the four IV PUMP jacks on the multi-pump adapter.

• Plug the other end of that modular cable into a transition connector.• Plug the transition connector into the connector on the pump.

Flexport System Interface Operations Manual 4-4

Page 65: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Figure 4-1: Multiple pump configuration

The Spacelabs Healthcare monitor is now ready to display information from the IV pump(s). To control other interface functions, refer to the sections that follow.

cable

SDLCterminator

transitionconnector

modularcables

IVpump

IVpump

IVpump

IVpump

multi-pumpadapter

SpacelabsHealthcare

monitor

systemFlexport

interface

Flexport System Interface Operations Manual 4-5

Page 66: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pump(s).

Figure 4-2: Bedside, remote, or central monitor, full screen

Figure 4-3: Central monitor, split screen

Pump address� Selected medication or fluid. (Until you set up the drug or fluid for each pump, this field

displays * * * * * * * *.)

� Selected pump or infusion rate. Initially, this field displays the pump rate. The calculated medication rate is displayed once you have selected the concentration or dose. Units for displays are based on those selected.

� The volume of fluids and the dose delivered since the last time one of the following conditions occurred:

• The pump started a new dose.

• The pump rate was titrated.

• The pump changed from secondary to primary or from primary to secondary.

� IV parameter key.

� Latest status message (for example, PUMPING). The displayed infusion may be either the primary or the secondary. The secondary infusion is displayed when the pump is pumping secondary (SEC PUMPING) or concurrently (CONCURRENT). Refer to Status Messages on page 4-20.

� � � � � �

PUMP DRUG/FLUID RATE VOL. INFUSED PUMP STATUS12 DOPAMINE 10.0 mcg/kg/min 122.1 ml I

V12 PUMPING

14 HEPARIN 1000.0 u/hr 233.4 ml 14 STOPPED9 NOREPINEPHRINE 3.0 mcg/min 10.9 ml 9 SEC PUMPING4 D5W 250.0 ml/hr 10.0 ml 4 INFSN DONE

� � � �

PUMP DRUG/FLUID STATUS12 DOPAMINE PUMPING I

V14 HEPARIN STOPPED9 NOREPINEPHRINE SEC PUMPING4 D5W INFSN DONE

Flexport System Interface Operations Manual 4-6

Page 67: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Entering Setup InformationSetup information is necessary for the Spacelabs Healthcare monitor to display the drug (or fluid) names and to compute the medication and calculated pump rates. The SETUP key controls the following functions and affects only the monitor at which these adjustments are made:

• Turn alarms ON or OFF.

• Enter patient weight.

• Edit drug or fluid names.

• Set up primary and secondary infusion lines.

• Select the infusion fluid.

• Select the infusion drug.

Once you select the infusion drug, you can perform the following functions:

• Enter the infusion concentration.

• Select the diluent fluid.

• Enter the infusion dose.

Note:Setup information can be entered only at the bedside monitor.

Enabling AlarmsWhen you turn alarms OFF, infusion pump alarms will no longer trigger an alarm on the Spacelabs Healthcare monitor.

Alarm settings default to ON.

Note:Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the infusion pump.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch ALARMS ON/OFF.

Flexport System Interface Operations Manual 4-7

Page 68: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Selecting an Infusion LineYou cannot set up the secondary line until you complete the primary infusion setup.

Selecting the Infusion FluidThere are two pages of Select Fluid screens. The first page displays the list of fluids automatically supplied by the Flexport system interface. The second page enables you to select from a list of up to eight fluids that you can enter into the Flexport system interface’s memory. Refer to Editing Drug or Fluid Names on page 4-15 for information on entering these names.

Figure 4-4: Select Fluid screen

Table 2 defines the fluid/concentration abbreviations that correspond to the fluids listed on the Select Fluid screen.

To select an infusion line:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch PRIMARY/SECONDARY.

PUMP 12 - SELECT FLUID

LR D5 LR D5 NS D5 1/2 NS

D5 1/2 NS 20 KCL D5 1/4 NS D5W D10 NS

D10 1/2 NS D10 1/4 NS D10W NS

1/2 NS 1/4 NS TPN LIPIDS

PAGE ↑ PAGE ↓ ENTER

Flexport System Interface Operations Manual 4-8

Page 69: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Table 2: Fluid Abbreviations

Abbreviation Description

LR lactated ringers

D5 LR 5% dextrose with lactated ringers

D5 NS 5% dextrose with 0.9% normal saline

D5 1/2NS 5% dextrose with 0.45% normal saline

D5 1/2NS 20KCL 5% dextrose with 0.45% normal saline and 20 mEq potassium chloride/liter

D5 1/4NS 5% dextrose with 0.2% normal saline

D5W 5% dextrose with water

D10 NS 10% dextrose with 0.9% normal saline

D10 1/2NS 10% dextrose with 0.45% normal saline

D10 1/4NS 10% dextrose with 0.2% normal saline

D10 W 10% dextrose with water

NS 0.9% normal saline

1/2 NS 0.45% normal saline

1/4 NS 0.2% normal saline

TPN total parenteral nutrition

LIPIDS fat emulsion delivered as part of parenteral nutrition

To select an infusion fluid:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT FLUID.• Select a fluid from the list (or touch PAGE ↓ and then select a fluid).• Touch ENTER.

Flexport System Interface Operations Manual 4-9

Page 70: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Selecting a DrugThe Select Drug screen provides 16 default drug names. Refer to Editing Drug or Fluid Names on page 4-15 for information on editing the default drug names.

Figure 4-5: Select Drug screen

After selecting a drug, the monitor displays (in order) screens for the concentration and diluent.

Selecting the Drug ConcentrationThe drug concentration represents the amount of a drug dissolved in a volume of diluent. The Flexport system interface saves the last entered default concentration values and unit of measurement associated with the selected drug.

Access to this display is automatic from the Select Drug screen.

To change a value after you have completed an initial drug/fluid entry, complete the steps listed in the following Quickstart. Figure 4-6 illustrates the Concentration screen.

PUMP 12 - SELECT DRUGAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

ENTER

To select a drug:

• Touch IV.• Touch SETUP.• Touch a pump key.• Touch SELECT DRUG.• Select a drug from the list.• Touch ENTER.

Flexport System Interface Operations Manual 4-10

Page 71: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Figure 4-6: Concentration screen

Caution:• The validity of the drug concentration or dose values that you enter is NOT checked by the

system.

• Calculated drug rates are displayed to one significant decimal place. Rates less than 0.05 will appear as 0.0.

To select a drug concentration:

• Touch IV.• Touch SETUP.• Touch a pump key.• Touch SELECT CONC.• Enter the drug quantity.• Select the unit of measurement.• Select the volume of fluid.• Touch ENTER.

PUMP 4 - ENTER DRUG QTY, UNITS, AND VOLUME. CONCENTRATION = _______

7 8 9 mcg mg /50 ml /100 ml

4 5 6 g ml /250 ml /500 ml

1 2 3 units mEq /1 L /2 L

0 . CLEAR ENTER

Flexport System Interface Operations Manual 4-11

Page 72: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Selecting the DiluentA diluent is a fluid used to dilute a drug for infusion. The fluids you can choose as drug diluents are the same as those you can choose as infusion fluids. Refer to Selecting the Infusion Fluid on page 4-8 for details on selecting a diluent. The Flexport system interface saves the last entered diluent associated with a selected drug.

Access to this display is automatic from the Concentration screen.

To change your selection after you have completed an initial diluent entry, complete all the steps listed in the following Quickstart.

When you touch ENTER to complete the diluent entry, the monitor displays the Setup menu.

Selecting the DoseThe dose represents the amount of drug to be infused over a specified time period. Use this display to enter the quantity, units of measurement, and time units.

If you have completed an initial drug/fluid entry and wish to change a value, complete all the steps listed in the following Quickstart.

Note:If the interface displays an initial VOLUME INFUSED for the pump you intend to use, you must clear that value at the pump before you start an infusion. Otherwise, the Flexport system interface will appear to add the volume infused from the previous infusion to the current infusion (either primary or secondary).

To select a diluent:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DILUENT.• Select a diluent from the list (or touch PAGE ↓ and then select a diluent).• Touch ENTER.

To select the drug dose:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch DOSE CALCS.• Enter the dose quantity.• Enter unit of measurement.• Enter the time unit.• Touch ENTER.

Flexport System Interface Operations Manual 4-12

Page 73: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Figure 4-7: Dose Calcs screen

As you enter the quantity, unit of measurement, and unit of time, the monitor displays the complete dose and the calculated pump rate. This calculated pump rate is based on the concentration and dose values you enter. To arrive at this value, the Flexport system interface performs the following calculation:

CALCULATED PUMP RATE = DOSE/CONCENTRATION

Note:The calculated pump rate is provided as information only. The Flexport system interface does not control the actual set pump rate.

When you touch ENTER to complete the dose entry, the interface saves your selection and the monitor displays the Setup menu.

PUMP A - ENTER DRUG QUANTITY, UNITS, AND TIME. DOSE = _______

7 8 9 mcg mg /min /hr

4 5 6 g ml CALCULATED PUMP RATE= xxxx.x ml/hr

1 2 3 units mEq

0 . CLEAR mcg/kg mg/kg ENTER

Flexport System Interface Operations Manual 4-13

Page 74: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Setting Patient WeightTouching the WEIGHT key displays a screen with a prompt containing the current weight of the patient (if you entered it previously using the Weight screen, a cardiac output module, or the monitor admit function). The Patient Weight screen uses the same units as the bedside monitor (kg or pounds).

Figure 4-8: Patient Weight screen

After you complete your editing and touch ENTER, the current weight will appear in the appropriate screens. You can enter up to six characters, including the decimal point. The valid weight range is from 1 to 250 kg (2 to 551 pounds). The interface saves the weight so you can use it in the Dose Calcs screen with the mcg/kg and mg/kg units of measurement.

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = __ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

To enter patient weight:

• Touch IV.• Touch SETUP.• Touch WEIGHT.• Enter the patient weight.• Touch ENTER.

Flexport System Interface Operations Manual 4-14

Page 75: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Editing Drug or Fluid NamesThe Change Configuration screen for drug names initially lists the 16 default drug names. You can change the name of any of these drugs.

The Change Configuration screen for fluid names initially displays eight blank keys. You can select any of the eight keys for the fluid name. Figure 4-9 shows the Change Drug Configuration display.

Figure 4-9: Change Drug Configuration display

After selecting a drug or fluid key to relabel, touch CHANGE NAME. The monitor displays the Change Drug (or Fluid) Name screen (refer to Figure 4-10).

To edit a drug or fluid name:

• Touch IV.• Touch SETUP.• Touch CHANGE CONFIG.• Touch EDIT DRUG or EDIT FLUID.• Select the name you wish to change.• Touch CHANGE NAME.• Use alphanumeric keys to make necessary changes.• Touch ENTER.

IV - CHANGE CONFIGURATION - SELECT DRUG KEY TO RELABELAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

CHANGE NAME

Flexport System Interface Operations Manual 4-15

Page 76: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Figure 4-10: Change Drug (or Fluid) Name screen

• Enter each character or number using the alphanumeric keys.

• Touch DEL to eliminate the character highlighted by the cursor.

• Move through the name, character by character, using the arrow keys.

• Touch CLEAR to eliminate the entire name and start over. The screen displays blank fields in place of the name.

• Touch ENTER to complete the edit and store the name in memory.

• Enter up to fifteen characters.

After you complete your editing and touch ENTER, the name will appear in the appropriate display for selecting drugs or fluids. The interface saves the name so you can recall it the next time you connect the Flexport system interface.

IV-CHANGE DRUG (or fluid) NAME= _________________________

0 1 2 3 4 5 6 7 8 9

A B C D E F G H I J

K L M N O P Q R S T

U V W X Y Z / % .

DEL ← → CLEAR ENTER

Flexport System Interface Operations Manual 4-16

Page 77: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Disconnecting a PumpWhen you touch DISCONNECT, the Flexport system interface discontinues the infusion record. If the pump is operating when you touch this key, the infusion record stops, a new record starts, and all pump setup information is cleared from the interface. If you touch DISCONNECT when the pump is not operating, the setup information stored by the Flexport system interface is cleared and the line of pump data is removed from the display.

Note:• Touching the DISCONNECT key does NOT turn OFF the infusion pump.

• A pump can be disconnected only at the bedside monitor.

Do not turn the power OFF on the pump while connected to the Flexport system interface. Disconnect the module cable from the pump first.

When you turn an infusion pump OFF at the pump itself, the pump status listed on the monitor changes to DISCONNECTED. At this point the interface completes the history for the current infusion (HISTORY shows the time that the infusion ended). If you turn the pump ON again, the interface continues to display the drug/fluid information entered previously.

Noting an EventThe EVENT key stores the time and pump status in memory. There is no immediate feedback when you touch this key. When you review the pump history, there is an entry labeled EVENT at the time you noted.

Events you may want to record include procedures such as changing IV lines or bags without turning the pump OFF.

Note:An event can be noted only at a bedside monitor.

To disconnect a pump:

• Touch IV.• Touch DISCONNECT.• Select a pump key.• Touch YES.

To note an event:

• Touch IV.• Touch EVENT.• Select a pump key.

Flexport System Interface Operations Manual 4-17

Page 78: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Displaying Infusion HistoryInfusion history displays as a table listing the last 40 infusion events for the patient on four pages (refer to Figure 4-11). This view-only table is available at the bedside monitor only and is not available for remote viewing.

The data shown in the table does not automatically update while you are viewing it. To update the history data, re-enter the history table by touching PREVIOUS and then HISTORY.

Figure 4-11: Example of History table

The word ACTIVE appearing in the Ended column indicates that the pump is presently infusing the drug or fluid. The interface updates the log of active entries each minute.

The word EVENT appearing in the Ended column indicates the time at which the clinician touched the EVENT key to record a miscellaneous event into the pump history. Events that you might want to record include changing infusion lines or bags without turning the pump OFF.

To display infusion history:

• Touch IV.• Touch HISTORY.

DATE BEGAN ENDED PUMP DRUG/FLUID CONCENTRATION FLOW RATE

VOL. INFSD

02/20 10:30am 12:10pm 12 DOPAMINE 400mg/250ml 30.0 ml/hr 50.0 ml

02/20 12:15pm EVENT 12 DOPAMINE 400mg/250ml 20.0 ml/hr 3.5 ml

02/20 12:50pm 05:00pm 12 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/20 06:00pm 11:00pm 14 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/21 12:30am 01:00am 12 NOREPINEPHRINE 1mg/250ml 125.0 ml/hr 62.5 ml

02/21 01:30am ACTIVE 12 NOREPINEPHRINE 400mg/250ml 30.0 ml/hr 210.0 ml

02/21 01:30am 03:00am 14 HEPARIN 2000u/500ml 100.0 ml/hr 750.0 ml

02/21 03:30am 07:00am 13 D5W 100.0 ml/hr 350.0 ml

02/21 07:30am ACTIVE 14 D10 1/2NS 100.0 ml/hr 100.0 ml

02/21 08:00am ACTIVE 9 DOPAMINE 400mg/250ml 30.0 ml/hr 15.0 ml

Flexport System Interface Operations Manual 4-18

Page 79: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display or infusion history table. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the infusion information:

• Touch IV.• Touch PRINT.

-OR-• Touch HISTORY.• Touch PRINT.

Flexport System Interface Operations Manual 4-19

Page 80: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Alaris 980C/PC-1/PC-2/PC-4

Status MessagesThe message DISCONNECTED indicates a problem with the pump or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pump is powered ON.

• Verify that all cables are secure.

Table 3 lists the status messages that the Flexport system interface reports for the pump during Flexport system interface operation.

Table 3: Status Messages

Message Text Indication

AIR IN LINE Pump has detected air in IV tubing

DISCONNECTED Pump has been disconnected or turned OFF

DOOR OPEN Pump has detected an open door

INFSN. DONE Primary infusion is complete

KVO Pump pumping at KVO rate

LOW BATTERY Battery requires recharging

OCCLUSION Closed clamp, kinked or occluded tubing, or occluded filter detected

OK Pump is signing ON

PRIMARY MODE Pump is operating the primary line

PUMP ALARM Pump has detected an alarm condition

PUMPING Pump is pumping normally

SEC PUMPING Pump is operating the secondary line

STOPPED Pump is stopped but not turned OFF

Flexport System Interface Operations Manual 4-20

Page 81: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bard CritiCore

Directory of Keys

URINE

URINE - MENUALARM LIMITS SETUP VIEW

ALARMS PRINT

URINE - SETUPLARGE TXT

SMALL TXT

URINE - ALARM LIMITS

TEMP CNTNR FLOW

URINE - ALARM LIMITS - FlowFLOW HI =

XXXXLO=

XXXXON OFF

URINE - ALARM LIMITS - CntnrCNTNR HI =

XXXXON OFF

URINE - ALARM LIMITS - TempTEMP HI =

XXXXLO=

XXXXON OFF

Flexport System Interface Operations Manual 5-1

Page 82: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 83: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bard CritiCore

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a Bard CritiCore monitor to be displayed on Spacelabs Healthcare monitors.

Configure the CritiCore monitor as shown in Table 1. Consult the Bard CritiCore manual for additional configuration setup procedures

* The Bard CritiCore monitor must contain the optional CritiCore communications module.

Table 1: Bard CritiCore Configuration

BaudRate

Minimum Software Revision

Installation Kit

Part Number

Modular Cable

Part Number

9600 N/A* 045-0036-xx 012-0510-xx

Flexport System Interface Operations Manual 5-3

Page 84: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bard CritiCore

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the CritiCore monitor, complete the steps in the following Quickstart.

Figure 5-1: Flexport system interface connections

Figure 5-2: CritiCore communications module side insert

To connect the Flexport System Interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 5-1.

• Plug the communications module power supply into a wall outlet and into the side of the CritiCore monitor (refer to Figure 5-2).

• Under the remote switch panel on the back of the CritiCore monitor, set the communications format switch to the Flexport Protocol position.

• Turn on the CritiCore monitor and complete the setup procedure on the back panel.• Plug one end of a modular cable into the modular jack on the Flexport system interface.• Plug the other end of the modular cable into the 4-pin female mini-DIN jack on the

CritiCore communications module.

communicationsmodulepower supply

SDLCcable

SDLCterminator

modularcable

Flexport

interfaceBard

CritiCore

SpacelabsHealthcare

monitor

system

indicator LED4-pin mini-DIN jack

power supply connector

Flexport System Interface Operations Manual 5-4

Page 85: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bard CritiCore

The Spacelabs Healthcare monitor is now ready to display information from the CritiCore monitor. The indicator LED on the CritiCore monitor’s side insert will pulse with each data burst. For further information on the indicator LED, refer to the appropriate CritiCore monitor documentation.

To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to the Flexport system interface, the monitor provides the following information from the CritiCore monitor.

Figure 5-3: Bedside monitor, large text screen

Figure 5-4: Bedside monitor, small text screen

� Present interval output value (ml)

� Flow Rate value (ml/hr)

� Minutes value from the CritiCore monitor time clock (min)

� URINE parameter key

� Bladder temperature value (°F or °C)

� Alarm status bell appears when Flexport alarms are turned ON. Flow, Cntnr, and Temp alarms must also be enabled at the CritiCore monitor.

� CritiCore monitor interval length value (minute or hour)

� � � � �

49 Prsnt 294 Flow 23 Time URINE

101.2 Temp

minml ml/hr F

27 1012 15Prior

ml

Cntnr

ml

Int

min

� �

� � �

URINE

Temp = 101.2 F

Prsnt = 49 ml

Prior

Time

=

=

27 ml

23 min

� �

Flexport System Interface Operations Manual 5-5

Page 86: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bard CritiCore

Container volume value (ml)

Prior interval output value (ml)

Note:The units for temperature (°F or °C) and interval length (minute or hour) are programmable via the CritiCore monitor. Refer to the appropriate CritiCore documentation for instructions. The communications link will be re-established after changing the units.

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter (FLOW, TEMP, and CNTNR). When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the URINE key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the CritiCore

monitor.

• Alarm limits can only be adjusted at the CritiCore monitor. Alarm violation is detected at the CritiCore monitor, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the CritiCore monitor can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch URINE.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 5-6

Page 87: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bard CritiCore

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the CritiCore monitor’s parameters (refer to Figure 5-5). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays URINE ALM OFF to the right of the URINE key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 5-5: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 5-5).

The default setting is LARGE TXT.

To view alarm settings:

• Touch URINE.• Touch VIEW ALARMS.

To select a display format:

• Touch URINE.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Temp 102.0 Flow 4500 URINE

101.2 Temp92.0 200

F

Cntnr 2000

Flexport System Interface Operations Manual 5-7

Page 88: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bard CritiCore

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the CritiCore display:

• Touch URINE.• Touch PRINT.

Flexport System Interface Operations Manual 5-8

Page 89: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bard CritiCore

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the CritiCore or the cabling. If the Spacelabs Healthcare monitor displays this message, perform the following troubleshooting steps:

• Verify that the CritiCore monitor is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the CritiCore monitor during Flexport operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

The communication between the Flexport system interface and the CritiCore monitor is interrupted. Yes

CHECK MONITORThe CritiCore monitor is OFF, or the CritiCore communications module is not receiving data internally from the CritiCore monitor.

No

CHECK “SET VALUE” SWITCH

The Set Value switch on the back of the CritiCore monitor was not returned to the normal operation position. No

FULL CONTAINER The Bard CritiCore collection container is full. No

NO CONTAINER No container is in place, or the collection container is not seated properly within the CritiCore monitor. No

MONITOR NOT LEVEL The CritiCore monitor is not level (volume measurements cannot be taken under this condition). No

CHECK TEMPERATURE PROBE

A valid temperature cannot be detected, or no temperature probe is connected to the CritiCore monitor. No

Flexport System Interface Operations Manual 5-9

Page 90: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 91: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 6-1

Baxter-Edwards Explorer

Directory of Keys

CO

CO - MAIN MENUALARM LIMITS SETUP VIEW

ALARMS PRINT

These keys appear on the bedside monitor only.

LARGE TXT

SMALL TXT

CO - ALARM LIMITS

SPO2 SVO2

CO - ALARM LIMITS - SVO2SVO2 HI =

XXXXLO=

XXXXSPO2

ON OFF SVO2

CO - ALARM LIMITS - SPO2SPO2 HI =

XXXXLO=

XXXXSPO2

ON OFF SVO2

Page 92: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 93: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter-Edwards Explorer

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90442A Flexport system interface allows information from a Baxter-Edwards Critical Care Division Explorer cardiac output (CO) monitor to be displayed on a Spacelabs Healthcare monitor.

Configure the CO monitor as shown in Table 1. Consult the manufacturer’s manual for additional configuration setup procedures.

Table 1: CO Monitor Configuration

Baud Rate

Minimum Software Revision

Baxter Interface Module

Installation Kit Part Number

Transition Connector

Part Number

9600 06.37 60-RS-SL 045-0010-xx 131-0686-xx

Flexport System Interface Operations Manual 6-3

Page 94: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter-Edwards Explorer

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the CO monitor, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the CO monitor. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the CO monitor.

Figure 6-1: Bedside monitor, large text

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Connect the 60-RS-SL interface (Baxter provides the interface) module to the

CO monitor.• Plug the transition connector into the 9-pin connector on the 60-RS-SL interface

module.

� � � � �

9.8 CO 398 EDV 163 SV

CO

96 SpO2

%l/min ml ml

--.- Cl 235 ESV 41 REF 77 SvO2

%ml %

� � �

Flexport System Interface Operations Manual 6-4

Page 95: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter-Edwards Explorer

Figure 6-2: Bedside monitor, small-text screen

Figure 6-3: Central monitor, split screen

� Cardiac output average value

� End-diastolic volume average value

� Stroke volume average value

� CO parameter key

� Pulse oximeter SpO2 value

� Alarm status bell (appears when alarms are turned ON)

� Venous oximeter SvO2 value

Ejection fraction average value

End-systolic volume average value

� Cardiac index average value

� Patient/room ID

Note:Monitors can display either cardiac output or cardiac index, but not both at the same time.

� �

CO

SpO2= 96%�SvO2= 77%

CO= 9.8 l/minCl= --.-

� � �

� �

� SpO2= 96% EDV=398ml

CO

� SvO2= 77% ESV=235ml

� CO= 9.8 l/min SV=163ml �� Cl=--.- REF= 41%

ICU 14�

Flexport System Interface Operations Manual 6-5

Page 96: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter-Edwards Explorer

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for SpO2 or SvO2. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the CO key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When both parameter alarms are turned OFF, alarm status messages will still cause the CO key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarms on the CO monitor.

• Alarm limits can only be adjusted at the CO monitor. Alarm violation is detected at the monitor, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the CO monitor can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON or OFF from the Spacelabs Healthcare monitor:

• Touch CO.• Touch ALARM LIMITS.• Select SPO2 or SVO2.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 6-6

Page 97: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter-Edwards Explorer

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the CO monitor’s parameters (refer to Figure 6-4). If alarms for SpO2 and SvO2 parameters are turned OFF, the Spacelabs Healthcare monitor displays CO ALM OFF to the right of the CO key. If alarms are turned ON for any parameter, a bell is displayed instead. The alarm bell flashes when an alarm limit is violated.

Figure 6-4: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 6-4).

The default setting is LARGE TXT.

To view alarms settings:

• Touch CO.• Touch VIEW ALARMS.

To select a display format:

• Touch CO.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

SpO2SvO2

100908040 C

O96 SpO2

%

80 SvO2

%

Flexport System Interface Operations Manual 6-7

Page 98: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter-Edwards Explorer

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the CO monitor or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the CO monitor is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the CO monitor during Flexport system interface operation.

To print the CO monitor display:

• Touch CO.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the CO monitor and the Spacelabs Healthcare monitor has been interrupted. Yes

EXPLORER DISCONNECTED

Communication problem has been detected between the CO monitor and 60-RS-SL. No

SVO2 PROBLEM — CHECK EXPLORER

The SvO2 catheter is not attached, not calibrated, or is defective. No

SPO2 PROBLEM — CHECK EXPLORER The SpO2 probe is not attached or is defective. No

CO/REF PROBLEM — CHECK EXPLORER

The cardiac output catheter is not connected or is not configured properly. No

Flexport System Interface Operations Manual 6-8

Page 99: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Directory of Keys

IV

IV - MAIN MENU

HISTORY SETUP EVENT DISCONNECT PRINT

IV - DISCONNNECTPUMP

APUMP

B ALL

IV - EVENT IV - DISCONNNECT PUMP XX?PUMP

APUMP

B ALL YES NO

IV - SETUPPUMP

APUMP

B WEIGHT CHANGE CONFIG

IV - PUMP XXPRIMARY SELECT

FLUIDSELECT DRUG

SELECT CONC.

SELECT DILUENT

DOSE CALCS

ALARMS

SECONDARY ON OFF

IV - CHANGE CONFIGURATION - SELECT PARAMETER TO EDIT

EDIT DRUG EDIT FLUID

IV - HISTORY

PAGE ↑ PAGE ↓ PRINT

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = ___ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

Flexport System Interface Operations Manual 7-1

Page 100: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 101: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

OverviewThe 90437A-06 Flexport system interface enables information from Baxter Flo-Gard 6201 and 6301 volumetric infusion pumps to be displayed on Spacelabs Healthcare monitors.

Configure the infusion pump as shown in Table 1. Consult the pump manual for additional configuration setup procedures.

Table 1: Infusion Pump Configuration

Baud Rate

Character Length

Parity Setting Stop Bit

9600 8-bit none 1

Flexport System Interface Operations Manual 7-3

Page 102: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the IV pump(s), complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the IV pump(s). To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pump(s).

Figure 7-1: Bedside, remote, or central monitor, full screen

Figure 7-2: Central monitor, split screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface.• Plug one end of the modular cable into the modular jack on the Flexport system

interface.• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the connector on the infusion pump “communications

port.”

� � � � � �

PUMP DRUG/FLUID RATE VOL. INFUSED PUMP STATUSA DOPAMINE 10.0 mcg/kg/min 122.1 ml I

VA PUMPING

B HEPARIN 1000.0 u/hr 233.4 ml B STOPPED

� � � �

PUMP DRUG/FLUID STATUSA DOPAMINE PUMPING I

VB HEPARIN STOPPED

Flexport System Interface Operations Manual 7-4

Page 103: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

� Pump address.

� Selected medication or fluid. (Until you set up the drug or fluid for each pump, this field displays * * * * * * * *.)

� Selected pump or infusion rate. Initially, this field displays the pump rate. The calculated medication rate is displayed once you have selected the concentration or dose. Units for displays are based on those selected.

� The volume of fluids and the dose delivered since the last time one of the following conditions occurred:

• The pump started a new dose.

• The pump rate was titrated.

• The pump changed from secondary to primary or from primary to secondary.

� IV parameter key.

� Latest status message (for example, PUMPING). The displayed infusion may be either the primary or the secondary. The secondary infusion is displayed when the pump is pumping secondary (SEC PUMPING). Refer to Status Messages on page 7-20.

Entering Setup InformationSetup information is necessary for the monitor to display the drug (or fluid) names and to compute the medication and calculated pump rates. The SETUP key controls the following functions and affects only the monitor at which these adjustments are made:

• Turn alarms ON or OFF.

• Enter patient weight.

• Edit drug or fluid names.

• Set up primary and secondary infusion lines.

• Select the infusion fluid.

• Select the infusion drug.

Once you select the infusion drug, you can perform the following functions:

• Enter the infusion concentration.

• Select the diluent fluid.

• Enter the infusion dose.

Note:Setup information can be entered only at the bedside monitor.

Flexport System Interface Operations Manual 7-5

Page 104: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Enabling AlarmsWhen you turn alarms OFF, infusion pump alarms will no longer trigger an alarm on the Spacelabs Healthcare monitor.

The alarm setting defaults to ON.

Note:Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the infusion pump(s).

Selecting an Infusion LineYou cannot set up the secondary line until you complete the primary infusion setup.

To turn alarms ON or OFF:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch ALARMS ON/OFF.

To select an infusion line:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch PRIMARY/SECONDARY.

Flexport System Interface Operations Manual 7-6

Page 105: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Selecting the Infusion FluidThere are two pages of Select Fluid screens. The first page displays the list of fluids automatically supplied by the Flexport system interface. The second page enables you to select from a list of up to eight fluids that you can enter into the Flexport system interface’s memory. Refer to Editing Drug or Fluid Names on page 7-14 for information on entering these names.

Figure 7-3: Select Fluid screen

Table 2 defines the fluid/concentration abbreviations that correspond to the fluids listed on the Select Fluid screen.

PUMP 12 - SELECT FLUIDLR D5 LR D5 NS D5 1/2 NS

D5 1/2 NS 20 KCL D5 1/4 NS D5W D10 NS

D10 1/2 NS D10 1/4 NS D10W NS

1/2 NS 1/4 NS TPN LIPIDS

PAGE ↑ PAGE ↓ ENTER

Table 2: Fluid Abbreviations

Abbreviation Description

LR Lactated ringers

D5 LR 5% dextrose with lactated ringers

D5 NS 5% dextrose with 0.9% saline

D5 1/2NS 5% dextrose with 0.45% saline

D5 1/2NS 20 KCL 5% dextrose with 0.45% saline and 20 mEq potassium chloride/liter

D5 1/4NS 5% dextrose with 0.2% saline

D5W 5% dextrose with water

D10 NS 10% dextrose with 0.9% saline

D10 1/2NS 10% dextrose with 0.45% saline

D10 1/4NS 10% dextrose with 0.2% saline

D10W 10% dextrose with water

Flexport System Interface Operations Manual 7-7

Page 106: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Selecting a DrugTo select a drug for infusion, you must first choose a drug from the Select Drug page. The selected drug becomes active upon touching the ENTER key. The Flexport system interface provides 16 default drug names. Refer to Editing Drug or Fluid Names on page 7-14 for information on editing those default drug names.

Figure 7-4: Select Drug screen

After selecting a drug, the monitor displays (in order) displays for the concentration and diluent.

NS 0.9% saline

1/2 NS 0.45% saline

1/4 NS 0.2% saline

TPN Total parenteral nutrition

LIPIDS Fat emulsion delivered as part of parenteral nutrition

To select an infusion fluid:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT FLUID.• Select a fluid from the list (or touch PAGE ↓ and then select a fluid).• Touch ENTER.

PUMP 12 - SELECT DRUGAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

ENTER

Table 2: Fluid Abbreviations (continued)

Abbreviation Description

Flexport System Interface Operations Manual 7-8

Page 107: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Selecting the Drug ConcentrationThe drug concentration represents the amount of a drug dissolved in a volume of diluent. The Flexport system interface saves the last entered default concentration values and unit of measurement associated with the selected drug.

Access to this display is automatic from the Select Drug screen.

To change a value after you have completed an initial drug/fluid entry, complete all the steps listed in the following Quickstart. Figure 7-5 illustrates the Concentration screen.

To select a drug:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DRUG.• Select a drug from the list.• Touch ENTER.

To select a drug concentration:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT CONC.• Enter the drug quantity.• Select the unit of measurement.• Select the fluid volume.• Touch ENTER.

Flexport System Interface Operations Manual 7-9

Page 108: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Figure 7-5: Concentration screen

Caution:• The validity of the drug concentration or dose values that you enter is NOT checked by the

system.

• Calculated drug rates are displayed to one significant decimal place. Rates less than 0.05 will appear as 0.0.

PUMP 12 - ENTER DRUG QTY, UNITS, AND VOLUME. CONCENTRATION = _______

7 8 9 mcg mg /50 ml /100 ml

4 5 6 g ml /250 ml /500 ml

1 2 3 units mEq /1 L /2 L

0 . CLEAR ENTER

Flexport System Interface Operations Manual 7-10

Page 109: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Selecting the DiluentA diluent is a fluid used to dilute a drug for infusion. The fluids you can choose as drug diluents are the same as those you can choose as infusion fluids. Refer to Selecting the Infusion Fluid on page 7-7 for details on selecting a diluent. The Flexport system interface saves the last entered diluent associated with a selected drug.

Access to this display is automatic from the Concentration screen.

To change your selection after you have completed an initial diluent entry, complete all the steps listed in the following Quickstart.

When you touch ENTER to complete the diluent entry, the monitor displays the Setup menu.

Selecting the DoseThe dose represents the amount of drug to be infused over a specified time period. Use this display to enter the quantity, units of measurement, and time units.

If you have completed an initial drug/fluid entry and wish to change a value, complete all the steps listed in the following Quickstart.

Note:If the interface displays an initial VOLUME INFUSED for the pump you intend to use, you must clear that value at the pump before you start an infusion. Otherwise, the Flexport system interface will appear to add the volume infused from the previous infusion to the current infusion (either primary or secondary).

To select a diluent:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DILUENT.• Choose a diluent from the list (or touch PAGE ↓ and then select a diluent).• Touch ENTER.

To select the drug dose:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch DOSE CALCS.• Enter the dose quantity.• Enter unit of measurement.• Enter the time unit.• Touch ENTER.

Flexport System Interface Operations Manual 7-11

Page 110: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Figure 7-6: Dose Calcs screen

As you enter the quantity, unit of measurement, and unit of time, the monitor screen displays the complete dose and the calculated pump rate. This calculated pump rate is based on the concentration and dose values you enter. To arrive at this value, the Flexport system interface performs the following calculation:

CALCULATED PUMP RATE = DOSE/CONCENTRATION

Note:The calculated pump rate is provided as information only. The Flexport system interface does not control the actual set pump rate.

When you touch ENTER to complete the dose entry, the interface saves your selection and the monitor displays the Setup menu.

PUMP 12 - ENTER DRUG QUANTITY, UNITS, AND TIME. DOSE = _______

7 8 9 mcg mg /min /hr

4 5 6 g ml CALCULATED PUMP RATE= xxxx.x ml/hr

1 2 3 units mEq

0 . CLEAR mcg/kg mg/kg ENTER

Flexport System Interface Operations Manual 7-12

Page 111: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Setting Patient WeightTouching the WEIGHT key displays a screen with a prompt containing the current weight of the patient (if you entered it previously using the Weight screen, a cardiac output module, or the monitor admit function). The Patient Weight screen uses the same units as the bedside monitor (kg or pounds).

Figure 7-7: Patient Weight screen

After you complete your editing and touch ENTER, the current weight appears on the appropriate displays. Enter up to six characters, including the decimal point. The valid weight range is from 1 to 250 kg (2 to 551 pounds). The interface saves the weight so you can use it in the Dose Calcs screen with the mcg/kg and mg/kg units of measurement.

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = __ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

To enter patient weight:

• Touch IV.• Touch SETUP.• Touch WEIGHT.• Enter the patient weight.• Touch ENTER.

Flexport System Interface Operations Manual 7-13

Page 112: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Editing Drug or Fluid NamesThe Change Configuration screen for drug names initially lists the 16 default drug names. You can change the name of any of these drugs (refer to Figure 7-8).

The Change Configuration screen for fluid names initially shows a page of eight blank keys. You can select any of the eight keys for the fluid name.

Figure 7-8: Change Drug Configuration screen

After you select a drug or fluid key to relabel and touch CHANGE NAME, the monitor displays the Change Drug (or Fluid) Name screen (refer to Figure 7-9).

IV - CHANGE CONFIGURATION - SELECT DRUG KEY TO RELABELAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

CHANGE NAME

Flexport System Interface Operations Manual 7-14

Page 113: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Figure 7-9: Change Drug (or Fluid) Name screen

• Enter each character or number using the alphanumeric keys.

• Touch DEL to eliminate the character highlighted by the cursor.

• Move through the name, character by character, using the arrow keys.

• Touch CLEAR to eliminate the entire name and start over. The screen displays blank fields in place of the name.

• Touch ENTER to complete the edit and store the name in memory.

• Enter up to fifteen characters.

After you complete your editing and touch ENTER, the name will appear in the appropriate display for selecting drugs or fluids. The interface saves the name enabling you to recall it the next time you connect the Flexport system interface.

IV-CHANGE DRUG (or fluid) NAME= _________________________

0 1 2 3 4 5 6 7 8 9

A B C D E F G H I J

K L M N O P Q R S T

U V W X Y Z / % .

DEL ← → CLEAR ENTER

Flexport System Interface Operations Manual 7-15

Page 114: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Disconnecting a PumpWhen you touch DISCONNECT, the Flexport system interface discontinues the infusion record. If the pump is operating when you touch this key, the infusion record stops, a new record starts, and all pump setup information is cleared from the interface. If you touch DISCONNECT when the pump is not operating, the setup information stored by the Flexport system interface is cleared and the line of pump data is removed from the display.

Note:• Touching the DISCONNECT key does NOT turn OFF the infusion pump.

• A pump can be disconnected only at the bedside monitor.

Do not turn the power OFF at the pump while connected to the Flexport system interface. Disconnect the module cable from the pump first.

When you turn an infusion pump OFF at the pump itself, the pump status listed on the monitor changes to DISCONNECTED. At this point the interface completes the history for the current infusion (HISTORY shows the time that the infusion ended). If you turn the pump ON again, the interface continues to display the drug/fluid information entered previously.

To edit a drug or fluid name:

• Touch IV.• Select SETUP.• Touch CHANGE CONFIG.• Select EDIT DRUG or EDIT FLUID.• Select the name you want to change.• Touch CHANGE NAME.• Use the alphanumeric keys to make necessary changes.• Touch ENTER.

To disconnect a pump:

• Touch IV.• Touch DISCONNECT.• Select a pump key.• Touch YES.

Flexport System Interface Operations Manual 7-16

Page 115: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Noting an EventWhen you note an event, the interface stores the time and pump status in memory. There is no immediate feedback when you touch this key. When you review the pump history, there is an entry labeled EVENT at the time you noted.

Events you may want to record include procedures such as changing IV lines or bags without turning the pump OFF.

Note:An event can be noted only at a bedside monitor.

To note an event:

• Touch IV.• Touch EVENT.• Select a pump key.

Flexport System Interface Operations Manual 7-17

Page 116: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Displaying Infusion HistoryInfusion history displays as a table listing the last 40 infusion events for the patient on four pages (refer to Figure 7-10). This view-only table is available at the bedside monitor only and is not available for remote viewing.

The data shown in the table does not automatically update while you are viewing it. To update the history data, re-enter the history table by touching PREVIOUS and then HISTORY.

Figure 7-10: Example of History table

The word ACTIVE appearing in the Ended column indicates that the pump is presently infusing the drug or fluid. The interface updates the log of active entries each minute.

The word EVENT appearing in the Ended column indicates the time at which the clinician touched the EVENT key to record a miscellaneous event into the pump history. Events that you might want to record include changing infusion lines or bags without turning the pump OFF.

DATE BEGAN ENDED PUMP DRUG/FLUID CONCENTRATION FLOW RATE

VOL. INFSD

02/20 10:30am 12:10pm 12 DOPAMINE 400mg/250ml 30.0 ml/hr 50.0 ml

02/20 12:15pm EVENT 12 DOPAMINE 400mg/250ml 20.0 ml/hr 3.5ml

02/20 12:50pm 05:00pm 12 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/20 06:00pm 11:00pm 14 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/21 12:30am 01:00am 12 NOREPINEPHRINE 1mg/250ml 125.0 ml/hr 62.5 ml

02/21 01:30am ACTIVE 12 NOREPINEPHRINE 400mg/250ml 30.0 ml/hr 210.0 ml

02/21 01:30am 03:00am 14 HEPARIN 2000u/500ml 100.0 ml/hr 750.0 ml

02/21 03:30am 07:00am 13 D5W 100.0 ml/hr 350.0 ml

02/21 07:30am ACTIVE 14 D10 1/2NS 100.0 ml/hr 100.0 ml

02/21 08:00am ACTIVE 9 DOPAMINE 400mg/250ml 30.0 ml/hr 15.0 ml

To display infusion history:

• Touch IV.• Touch HISTORY.

Flexport System Interface Operations Manual 7-18

Page 117: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display or infusion history table. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the infusion information:

• Touch IV.• Touch PRINT.

-OR-• Touch HISTORY.• Touch PRINT.

Flexport System Interface Operations Manual 7-19

Page 118: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Baxter Flo-Gard 6201/6301

Status MessagesThe message DISCONNECTED indicates a problem with the pump or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pump is powered ON.

• Verify that all cables are secure.

Table 3 lists the status messages that the Flexport system interface reports for the pump during Flexport system interface operation.

Table 3: Status Messages

Message Text Indication

DISCONNECTED Pump has been disconnected or turned OFF.

AIR IN LINE Pump has detected air in IV tubing.

DOOR OPEN Pump has detected an open door.

INFSN. DONE Primary infusion is complete.

KVO Pump pumping at KVO rate.

LOW BATTERY Battery requires recharging.

OCCLUSION Closed clamp, kinked or occluded tubing, or occluded filter detected.

OK Appears at pump sign ON.

PUMPING Pump is pumping normally.

PUMP ALARM Pump has detected an alarm condition.

SEC PUMPING Pump is operating the secondary line.

STOPPED Pump is stopped but not turned OFF.

Flexport System Interface Operations Manual 7-20

Page 119: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 8-1

Bear 1000

Directory of Keys

VENT

VENT - MAIN MENU

ALARM LIMITS SETUP

FREEZE LARGE TXT VIEW MODE

VIEW STATUS

VIEW ALARMS PRINT

ON OFF SMALL TXT

VENT - SETUPWAVE SIZE

↑SIZE

↓SWEEP SPEED

WAVE TYPE

ON OFF PRS FLOW

25 mm/sec

12.5 mm/sec

6.25 mm/sec

3.12 mm/sec

1.56 mm/sec

VENT - ALARM LIMITS

PIP Vmin RR

VENT - ALARM LIMITS - RRRR HI =

XXXLO=XXX

RR

ON OFF PIP

VENT - ALARM LIMITS - VminVmin HI =

XXXLO=XXX

Vmin

ON OFF RR

VENT - ALARM LIMITS - PIPPIP HI =

XXXXLO=XXX

PIP

ON OFF Vmin

Page 120: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 121: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

OverviewThe 90436A-12 Flexport system interface enables information from a Bear 1000 ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting Stop Bit Minimum Software

Revision

19,200 8-bit none 1 9.7 or A.3

Flexport System Interface Operations Manual 8-3

Page 122: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

Figure 8-1: Flexport system interface configuration without graphical display unit

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 8-1 and Figure 8-2.

• If no graphical display is connected to the ventilator, connect the transition connector (P/N 131-0686-xx) to the ventilator’s 9-pin connector labeled RS-232 INTERFACE (refer to Figure 8-1).If a graphical display is connected to the ventilator, connect the transition connector (P/N 131-1039-xx) to the graphical display unit connector labeled DIGITAL OUT (refer to Figure 8-2).

• Connect the 15-pin connector on the modular cable into the 15-pin connector on the ventilator labeled ANALOG OUTPUT and the 6-pin connector into the transition connector.

• Plug the other end of the modular cable into the 8-pin receptacle on the Flexport system interface.

SDLCcable SDLC

terminator

modularcable analog output

transition

RS-232 interface

connector ventilator

SpacelabsHealthcare

monitor

systemFlexport

interface

Flexport System Interface Operations Manual 8-4

Page 123: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Figure 8-2: Flexport system interface configuration with graphical display unit

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 8-3: Bedside monitor, large-text screen

Figure 8-4: Bedside monitor, small-text screen

SDLCcable SDLC

terminator

modularcable

ventilator

transitionconnector

graphicaldisplay unit

analogoutput

digital out

SpacelabsHealthcare

monitor

systemFlexport

interface

� � � � � �

21 PIP 6 PLT 22 FiO2 VENT

1.22 Vt WAVE=PRESScmH2O % L

cmH2O Vmin RR

1.8 I:E 8 MAP 14.6 12cmH2O L BPM

� � �

� � �

VENT

Vt = 1.22L WAVE=PRESSRR =12BPM

PIP = 21cmH2O

FiO2 = 22%

� � �

Flexport System Interface Operations Manual 8-5

Page 124: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Figure 8-5: Central monitor, split screen

� Peak inspiratory pressure

� Plateau pressure

� Fractional inspired O2 control setting

� VENT parameter key

� Tidal volume

� Waveform identification label

� Alarm status bell (appears when alarms are turned ON)

Respiratory rate

Minute volume

� Mean airway pressure

� Ratio of inspiration time to expiration time (I:E ratio is typically expressed as 1:X.X, where X.X is the value displayed and equals E/I) *

* The displayed I:E ratio may differ slightly from the value shown on the Bear 1000. The Bear 1000 does not send the displayed data as it does for other parameters. Instead, it sends a value requiring calculations that may result in a variance as shown in Table 2.

Table 2: Display Variance

I:E Ratio Range Maximum Possible Variance

0.25 to 1.2 0

1.3 to 4.6 ±0.1

4.7 to 6.5 ±0.2

6.6 to 8.2 ±0.3

>8.3 is not displayed

� � �

VENT

� Vt=1.22L PIP=21cmH2O

Vmin=14.6L MAP=8cmH2O

RR=12BPM PLT=6cmH2O

� FiO2=22% I:E=1.8

� � �

Flexport System Interface Operations Manual 8-6

Page 125: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Vmin alarm limits are displayed as question marks when insufficient data is provided by the ventilator. The ventilator determines when the limits have been exceeded and the Flexport will report the alarm condition.

• The ventilator provides a feature that prevents a high PIP alarm. With this feature, the ventilator will alarm, but will immediately force the PIP value to an acceptable range. The Flexport will display the generated alarm as well as the current, adjusted value of the PIP as shown on the ventilator.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 8-7

Page 126: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 8-6). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed instead. The alarm bell flashes when an alarm limit is violated. If an individual alarm is OFF, no limit values are displayed next to the parameter label.

Figure 8-6: View Alarms screen

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

PIP 21 RR 27VENT

1.22 VT WAVE=PRESS0 3 L

12 RR ??.? BPM

Vmin ??.?

Flexport System Interface Operations Manual 8-8

Page 127: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Viewing Ventilator Mode HistoryChronological history of the last eight ventilator mode settings is displayed in the waveform zone, with the most recent value at the top of the left column and the oldest value at the bottom of the right column. Possible modes include ASSIST, SIMV/CPAP, and PRESSURE.

Figure 8-7: View Mode History screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be displayed on the bedside monitor, and three on the remote monitor (refer to Status Messages on page 8-13).

To view ventilator mode:

• Touch VENT.• Touch VIEW MODE.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

17:10 ASSIST 11:15 ASSIST VENT

1.22VT WAVE=PRESS15:47 SIMV/CPAP 9:31 PRESSURE L

14:23 PRESSURE 7:02 SIMV/CPAP 12RR

12:49 SIMV/CPAP 4:56 ASSIST BPM

Flexport System Interface Operations Manual 8-9

Page 128: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVE ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVE ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch VENT.• Touch SETUP.• Touch WAVE ON/OFF.

To adjust waveform size:

• Touch VENT.• Touch SETUP.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 8-10

Page 129: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, 12.5, and 25 mm/second.

The default setting is 12.5 mm/second.

Selecting a Waveform TypeYou can select either pressure or flow waveforms for display. Select PRS to display the pressure waveform, or select FLO to display the flow waveform. The two waveform displays are mutually exclusive.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch VENT.• Touch SETUP.• Select SWEEP SPEED.• Select the desired speed.

To select a waveform type:

• Touch VENT.• Touch SETUP.• Touch WAVE TYPE.• Select PRS or FLOW.

To freeze the waveform on the display:

• Touch VENT.• Touch FREEZE ON.

Flexport System Interface Operations Manual 8-11

Page 130: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 8-5).

The default setting is LARGE TXT.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select a display format:

• Touch VENT.• Select LARGE TXT or SMALL TXT.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 8-12

Page 131: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Bear 1000

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

• Verify that the ventilator’s baud rate is set at 19,200.

Table 3 lists the status messages that the Flexport system interface reports for the ventilator during Flexport system interface operation.

Table 3: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

GAS FAIL Gas supply (O2) pressure less than 27.5 psi. Yes

CYCLE FAIL Ventilator is not providing any mechanical breaths. Yes

Flexport System Interface Operations Manual 8-13

Page 132: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 133: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Directory of KeysPO+ET

PO+ET - MAIN MENUALARM LIMITS SETUP

FREEZE SWITCH WAVES

VIEW ALARMS PRINT

ON OFFRefer to page 9-2

These keys are displayed only on the bedside monitor.

PO+ET - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

PO+ET - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec0.39

mm/sec

PO+ET - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

Flexport System Interface Operations Manual 9-1

Page 134: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

PO+ET

PO+ET - MAIN MENUALARM LIMITS SETUP

FREEZE SWITCH WAVES

VIEW ALARMS PRINT

ON OFFRefer to page 9-1

Refer to page 9-1

Refer to page 9-1

PO+ET - ALARM LIMITS

ETCO2 MINCO2 RESP I N2O SPO2 PULSE

PO+ET - ALARM LIMITS - PULSEPULSE HI =

XXLO = XX

ETCO2ON OFF PULSE

PO+ET - ALARM LIMITS - SPO2SPO2 HI =

XXLO = XX

SPO2ON OFF PULSE

PO+ET - ALARM LIMITS - I N2OI N2O HI =

XXLO = XX

I N2OON OFF SPO2

PO+ET - ALARM LIMITS - RESPRESP HI =

XXLO = XX

RESPON OFF I N2O

PO+ET - ALARM LIMITS - MINCO2MINCO2 HI =

XXMINCO2

ON OFF RESP

PO+ET - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XX

ETCO2ON OFF MINCO2

Flexport System Interface Operations Manual 9-2

Page 135: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Selecting a Waveform Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90442A Flexport system interface enables information from a Biochem International BCI 9000 Capnograph-Oximeter to be displayed on Spacelabs Healthcare monitors.

Configure the oximeter as shown in Table 1. Consult the oximeter manual for additional configuration setup procedures

Table 1: Analyzer Configuration

Baud Rate

Minimum Software Revision

Installation Kit

Part Number

Transition Connector

Part Number

9600 4.95G 045-0020-xx 131-0697-xx(25-pin)

Flexport System Interface Operations Manual 9-3

Page 136: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the oximeter, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the oximeter. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the oximeter.

Figure 9-1: Bedside monitor, large text screen

Figure 9-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 25-pin connector on the back of the oximeter.

� � � �

39 ETCO2 21 RESP PO+ET

96 SpO2

mmHg BPM %

1 MINCO2 0 I N2O 90 PULSE

mmHg % bpm

� � �

� � � �

PO+ET

SpO2= 96%PULSE= 90bpmETCO2= 39 mmHg

�MINCO2= 1 mmHg

Flexport System Interface Operations Manual 9-4

Page 137: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Figure 9-3: Central monitor, split screen

� Current end tidal CO2 value

� Current respiration rate value

� PO+ET parameter key

� Current oxygen saturation value from pulse oximeter

� Alarm status bell (appears when alarms are turned ON)

� Current pulse rate value from pulse oximeter

� Current inspired N2O value

Current minimum CO2 value

Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the PO+ET key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the PO+ET key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the oximeter.

• Alarm tones on the oximeter can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

� � � �

� SpO2= 96% RESP= 21BPM PO+ET

� PULSE= 90bpm I N2O= 0%

� ETCO2= 38mmHg

MINCO2= 1mmHg

ICU 14

Flexport System Interface Operations Manual 9-5

Page 138: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the oximeter’s parameters (refer to Figure 9-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays PO+ET ALM OFF to the right of the PO+ET key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 9-4: View Alarms screen

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch PO+ET.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.To adjust alarm limits:• Select HI= or LO=.• Use arrow keys to adjust.

To view alarm settings:

• Touch PO+ET.• Touch VIEW ALARMS.

PO+ET

96 SpO2

ETCO2 50 RESP 50 SpO2 1000 10 85 %

MINCO2 6 I N20 80 PULSE 120 88 PULSE0 40

bpm

Flexport System Interface Operations Manual 9-6

Page 139: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveform ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch PO+ET.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch PO+ET.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 9-7

Page 140: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 50, 25, 12.5, 6.25, 3.12, 1.56, 0.78, and 0.39 mm/second. The minimum recorder sweep speed is 1.56 mm/second.

The default setting is 25 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 9-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch PO+ET.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch PO+ET.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch PO+ET.• Touch FREEZE ON.

Flexport System Interface Operations Manual 9-8

Page 141: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Selecting a Waveform TypeYou can select either CO2 waveform or PULSE waveform for display. When the waveform is changed, a status message will briefly display stating the new type of waveform being displayed.

Note:• You can only change the waveform display at the bedside monitor.

• Changing the displayed waveform on the Spacelabs Healthcare monitor has no affect on the oximeter.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select a waveform type:

• Touch PO+ET.• Touch SWITCH WAVES.

To print the oximeter display:

• Touch PO+ET.• Touch PRINT.

Flexport System Interface Operations Manual 9-9

Page 142: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Biochem International BCI 9000

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the oximeter or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the oximeter is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the oximeter during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOSTCommunication between the oximeter and the Spacelabs Healthcare monitor has been interrupted.

Yes

DISPLAYING CO2 WAVE You have changed the displayed waveform to CO2 by touching SWITCH WAVES. No

DISPLAYING PLETH WAVE You have changed the displayed waveform to PLETH by touching SWITCH WAVES. No

CHECK SPO2 SENSOR The SpO2 sensor is disconnected or is not attached to the patient. No

OXIMETER MALFUNCTION Service required. No

APNEA An APNEA condition has been detected. Yes

OCCLUSION The CO2 sample line is blocked. Yes

WATER TRAP FULL The CO2 water trap is full; replace water trap. Yes

THERAPY PAUSEYou have pressed “Therapy Pause” on the oximeter. Monitoring of patient data has been suspended.

No

LOW CAL REQUIRED The oximeter requires calibration. Yes

HIGH CAL REQUIRED The oximeter requires calibration. Yes

Flexport System Interface Operations Manual 9-10

Page 143: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Directory of Keys

IV

IV - MAIN MENU

HISTORY SETUP EVENT DISCONNECT PRINT

Refer to page 10-2

Refer to page 10-2

IV - DISCONNNECTPUMP

XXPUMP

XXPUMP

XXPUMP

XXPUMP

XXPUMP

XX ALL

PUMP XX

PUMP XX

PUMP XX

PUMP XX

PUMP XX

PUMP XX

IV - DISCONNNECT PUMP XX?YES NO

IV - EVENT MENUPUMP

XXPUMP

XXPUMP

XXPUMP

XXPUMP

XXPUMP

XX ALL

PUMP XX

PUMP XX

PUMP XX

PUMP XX

PUMP XX

PUMP XX

Flexport System Interface Operations Manual 10-1

Page 144: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

IV

IV - MAIN MENU

HISTORY SETUP EVENT DISCONNECT PRINT

Refer to page 10-1

Refer to page 10-1

IV - SETUP MENUPUMP

XXPUMP

XXPUMP

XXPUMP

XXPUMP

XXPUMP

XX WEIGHT CHANGE CONFIG

PUMP XX

PUMP XX

PUMP XX

PUMP XX

PUMP XX

PUMP XX

IV - PUMP XX MENU

SELECT FLUID

SELECT DRUG

SELECT CONC.

SELECT DILUENT

DOSE CALCS

ALARMS

ON OFF

IV - HISTORY MENU IV - CHANGE CONFIGURATION - SELECT PARAMETER TO EDIT

PAGE ↑ PAGE ↓ PRINT EDIT DRUG EDIT FLUID

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = ___ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

Flexport System Interface Operations Manual 10-2

Page 145: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Noting an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Displaying Infusion History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

OverviewThe 90437A-05 and 90437A-07 Flexport system interfaces enable information from a Braun Fluid Management System to be displayed on Spacelabs Healthcare monitors.

Caution:• 90437A-05 provides direct connection to the Braun Pump system. Connecting this unit to the

Braun Bedside Communication Controller (BCC) system may cause failure of the Braun BCC device.

• 90437A-07 provides connection to the Braun BCC system. This unit will not operate with the Braun Pump system.

Configure the infusion pump as shown in Table 1. Consult the appropriate infusion pump manual for additional configuration setup procedures.

Pump AddressesConnecting directly to Braun pumps or a Fluid Management System without a BCC:

The Spacelabs Healthcare monitor display identifies each IV pump by its address, which can range from 01 and 99. Braun refers to the pump address as the computer control (CC) address. Complete the steps provided in the following Quickstart to set the unique CC addresses, or refer to the Braun pump’s programmer’s guide.

Note:If you use multiple pumps, each pump must have a different address.

Table 1: Infusion Pump Configuration

Baud Rate Character Length

Parity Setting Stop Bit

9600 8 bit none 1

Flexport System Interface Operations Manual 10-3

Page 146: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Connecting to a Braun Fluid Management System with a Bedside Communication Controller (data concentrator):

The Spacelabs Healthcare monitor display identifies each IV pump by its location on the BCC. The address uses the relative panel number as the first digit and the relative pump number as the second digit. The computer control address (CC) on each pump must be set to 01 in order to communicate properly with the BCC device. Complete the steps provided in the above Quickstart to set the CC addresses or refer to the Braun pump’s programmer’s guide.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the infusion pump, complete the steps in the following Quickstart.

To change the CC address on the Braun pump:

• Touch the SPECIAL FUNCTION key on the Infusomat fmS or Perfusor Compact until the CC-MODE screen appears.

• The current CC address is shown in the LED screen as Cxx, and in line 1 of the LCD as #xx (where xx is the current address).

• Touch C to clear the current address.• Select the desired CC address using the numeric keypad.• Touch END to permanently enter the new address.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 10-1.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

Single pump or BCC:• Plug the other end of the modular cable into the appropriate transition connector.• Plug the transition connector into the Braun cable #871166/1.• Attach the other end of the Braun cable to the connector on the pump or the BCC.Multi-pump:• Plug the other end of the modular cable into the jack labeled FLEXPORT on the

multi-pump adaptor.• Plug another modular cable into one of the four IV pump jacks on the multi-pump

adapter.• Plug the other end into the appropriate transition connector.• Plug the transition connector into the Braun cable #871166/1.• Attach the other end of the Braun cable to the connector on the pump.• Repeat these steps for each additional pump.

Flexport System Interface Operations Manual 10-4

Page 147: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Figure 10-1: Single pump configuration

Figure 10-2: Multiple pump configuration

The Spacelabs Healthcare monitor is now ready to display information from the infusion pump(s). To control other interface functions, refer to the sections that follow.

IVpumpor BCC

transitionconnector

modularcable

SDLCterminator

SDLCcable

P/N 871166/1Braun cable

SpacelabsHealthcare

monitor

systemFlexport

interface

IVpump

IVpump

IVpump

IVpump

modularcables

cableSDLC

SDLCterminator

transitionconnectors

multi-pumpadapter

Braun cablesP/N 871166/1

SpacelabsHealthcare

monitor

systemFlexport

interface

Flexport System Interface Operations Manual 10-5

Page 148: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the infusion pump(s).

Figure 10-3: Bedside monitor screen

Note:The bedside monitor screen will expand to accommodate up to 12 pumps.

Figure 10-4: Remote or central monitor, full screen

Figure 10-5: Central monitor, split screen

� Pump address.

� Selected medication or fluid. (Until you set up the drug or fluid for each pump, this field displays * * * * * * * *.)

� Selected pump or infusion rate. Initially, this field displays the pump rate. The calculated medication rate is displayed once you have selected the concentration or dose. Units for displays are based on those selected.

� � � � � � �

PUMP DRUG/FLUID RATE VOL. INFUSED PUMP STATUS12 DOPAMINE 10.0 mcg/kg/min 122.1 ml I

V12 PUMPING

14 HEPARIN 1000.0 u/hr 233.4 ml 14 STOPPED9 NOREPINEPHRINE 3.0 mcg/min 10.9 ml 9 SEC PUMPING4 D5W 250.0 ml/hr 10.0 ml 4 INFSN DONE

� � � � � � �

P. STATUS RATE P. STATUS RATE P. STATUS RATE24 PUMP 52.3 g/hr 62 DISC 200.0 ml/hr I

V12 PUMP 22.1 mg/hr

88 PUMP 100.0 ml/hr 22 DONE 11.5 mcg/min 14 STOP 4.0 mcg/kg/min92 PUMP 24.3 mg/kg/hr 33 PUMP 210.0 mEq/hr 13 PUMP 250.0 ml/hr44 PUMP 21.3 mg/hr 63 PUMP 2.5 ml/kg/min 4 DONE 900.0 u/hr

� � � � � � �

PUMP STATUS PUMP STATUS PUMP STATUS24 PUMPING 62 DISCONN 12 PUMPING I

V88 PUMPING 22 INFUSN. D. 14 STOPPED92 PUMPING 33 PUMPING 13 PUMPING44 PUMPING 63 PUMPING 4 INFSN. D.

Flexport System Interface Operations Manual 10-6

Page 149: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

� Volume of fluids and the dose delivered since the last time one of the following conditions occurred:

• Pump starts a new dose.

• Pump rate is titrated.

� IV parameter key.

� Pump’s latest status message (for example, PUMPING).

� Pump address and status.

Entering Setup InformationSetup information is necessary for the Spacelabs Healthcare monitor to display the drug (or fluid) names and to compute the medication and calculated pump rates. The SETUP key controls the following functions and affects only the monitor at which these adjustments are made:

• Turn alarms ON or OFF.

• Enter patient weight.

• Edit drug or fluid names.

• Select the infusion fluid.

• Select the infusion drug.

Once you select the infusion drug, you can perform the following functions:

• Enter the infusion concentration.

• Select the diluent fluid.

• Enter the infusion dose.

Note:Setup information can be entered only at the bedside monitor.

Enabling AlarmsWhen alarms are turned OFF, the infusion pump alarms will no longer trigger an alarm on the Spacelabs Healthcare monitor.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the infusion

pump.

• Alarm limits on the Flexport system interface can only be adjusted at the infusion pump. Alarm violation is detected at the infusion pump and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the infusion pump can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Flexport System Interface Operations Manual 10-7

Page 150: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Selecting the Infusion FluidThere are two pages of Select Fluid screens. The first page displays the list of fluids automatically supplied by the Flexport system interface. The second page enables you to select from a list of up to eight fluids that you can enter into the Flexport system interface’s memory. Refer to Editing Drug or Fluid Names on page 10-15 for information on entering these names.

Figure 10-6: Select Fluid screen

Table 2 defines the fluid/concentration abbreviations that correspond to the fluids listed on the Select Fluid screen.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch ALARMS ON/OFF.

PUMP 12 - SELECT FLUIDLR D5 LR D5 NS D5 1/2 NS

D5 1/2 NS 20 KCL D5 1/4 NS D5W D10 NS

D10 1/2 NS D10 1/4 NS D10W NS

1/2 NS 1/4 NS TPN LIPIDS

PAGE ↑ PAGE ↓ ENTER

Table 2: Fluid Abbreviations

Abbreviation Description

LR lactated ringers

D5 LR 5% dextrose with lactated ringers

D5 NS 5% dextrose with 0.9% saline

D5 1/2NS 5% dextrose with 0.45% saline

D5 1/2NS 20KCL 5% dextrose with 0.45% saline and 20 mEq Potassium Chloride/Liter

D5 1/4NS 5% dextrose with 0.2% saline

D5W 5% dextrose with water

Flexport System Interface Operations Manual 10-8

Page 151: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

D10 NS 10% dextrose with 0.9% saline

D10 1/2NS 10% dextrose with 0.45% saline

D10 1/4NS 10% dextrose with 0.2% saline

D10 W 10% dextrose with water

NS 0.9% saline

1/2 NS 0.45% saline

1/4 NS 0.2% saline

TPN total parenteral nutrition

LIPIDS fat emulsion delivered as part of parenteral nutrition

To select an infusion fluid:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT FLUID.• Select a fluid from the list (or touch PAGE↓ and then select a fluid).• Touch ENTER.

Table 2: Fluid Abbreviations (continued)

Abbreviation Description

Flexport System Interface Operations Manual 10-9

Page 152: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Selecting a DrugTo select a drug for infusion, you must first choose a drug from the Select Drug page. The selected drug becomes active upon touching the ENTER key. The Flexport system interface provides 16 default drug names. Refer to Editing Drug or Fluid Names on page 10-15 for information on editing those default drug names.

Figure 10-7: Select Drug screen

After selecting a drug, the monitor displays (in order) displays for the concentration and diluent.

PUMP 12 - SELECT DRUGAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

ENTER

To select a drug:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DRUG.• Select a drug from the list.• Touch ENTER.

Flexport System Interface Operations Manual 10-10

Page 153: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Selecting the Drug ConcentrationThe drug concentration represents the amount of a drug dissolved in a volume of diluent. The Flexport system interface saves the last entered default concentration values and unit of measurement associated with the selected drug.

Access to this display is automatic from the Select Drug screen.

To change a value after you have completed an initial drug/fluid entry, complete all of the steps listed in the following Quickstart. Figure 10-8 illustrates the Concentration screen.

Figure 10-8: Concentration screen

Caution:• The validity of the drug concentration or dose values that you enter is NOT checked by the

system.

• Calculated drug rates are displayed to one significant decimal place. Rates less than 0.05 will appear as 0.0.

PUMP 12 - ENTER DRUG QTY, UNITS, AND VOLUME. CONCENTRATION = _______

7 8 9 mcg mg /50 ml /100 ml

4 5 6 g ml /250 ml /500 ml

1 2 3 units mEq /1 L /2 L

0 . CLEAR ENTER

To select a drug concentration:

• Touch IV.• Select SETUP.• Select a pump key.• Touch SELECT CONC.• Enter the drug quantity.• Select the unit of measurement.• Select the volume of fluid.• Touch ENTER.

Flexport System Interface Operations Manual 10-11

Page 154: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Selecting the DiluentA diluent is a fluid used to dilute a drug for infusion. The fluids you can choose as drug diluents are the same as those you can choose as infusion fluids. Refer to Selecting the Infusion Fluid on page 10-8 for details on selecting a diluent. The Flexport system interface saves the last entered diluent associated with a selected drug.

Access to this display is automatic from the Concentration screen.

To change your selection after you have completed an initial diluent entry, complete all of the steps listed in the following Quickstart.

When you touch ENTER to complete the diluent entry, the monitor displays the Setup menu.

Selecting the DoseThe dose represents the amount of drug to be infused over a specified time period. Use this display to enter the quantity, units of measurement, and time units.

If you have completed an initial drug/fluid entry and wish to change a value, complete all of the steps listed in the following Quickstart.

To select a diluent:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch SELECT DILUENT.• Choose a diluent from the list (or touch PAGE↓ and then select a diluent).• Touch ENTER.

To select the drug dose:

• Touch IV.• Touch SETUP.• Select a pump key.• Touch DOSE CALCS.• Enter the dose quantity.• Enter unit of measurement.• Enter the time unit.• Touch ENTER.

Flexport System Interface Operations Manual 10-12

Page 155: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Note:If the interface displays an initial VOLUME INFUSED for the pump you intend to use, you must clear that value at the pump before you start an infusion. Otherwise, the Flexport system interface will appear to add the volume infused from the previous infusion to the current infusion.

Figure 10-9: Dose Calcs screen

As you enter the quantity, unit of measurement, and unit of time, the monitor displays the complete dose and the calculated pump rate. This calculated pump rate is based on the concentration and dose values you enter. To arrive at this value, the Flexport system interface performs the following calculation:

CALCULATED PUMP RATE = DOSE/CONCENTRATION

Note:The calculated pump rate is provided as information only. The Flexport system interface does not control the actual set pump rate.

When you touch ENTER to complete the dose entry, the interface saves your selection and the monitor displays the Setup menu.

PUMP 12 - ENTER DRUG QUANTITY, UNITS, AND TIME. DOSE = _______

7 8 9 mcg mg /min /hr

4 5 6 g ml CALCULATED PUMP RATE= xxxx.x ml/hr

1 2 3 units mEq

0 . CLEAR mcg/kg mg/kg ENTER

Flexport System Interface Operations Manual 10-13

Page 156: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Setting Patient WeightTouching the WEIGHT key displays a screen with a prompt containing the current weight of the patient (if you entered it previously using the Weight screen, a cardiac output module, or the monitor admit function). The Patient Weight screen uses the same units as the bedside monitor (kg or pounds).

Figure 10-10: Patient Weight screen

After you complete your editing and touch ENTER, the current weight will appear on the appropriate displays. You can enter up to six characters, including the decimal point. The valid weight range is from 1 to 250 kg (2 to 55 pounds). The interface saves the weight so you can use it in the Dose Calcs screen with the mcg/kg and mg/kg units of measurement.

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = __ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

To enter patient weight:

• Touch IV.• Touch SETUP.• Touch WEIGHT.• Enter the patient weight.• Touch ENTER.

Flexport System Interface Operations Manual 10-14

Page 157: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Editing Drug or Fluid NamesThe Change Configuration screen for drug names initially lists the 16 default drug names. You can change the name of any of these drugs (refer to Figure 10-11).

The Change Configuration screen for fluid names initially shows a page of eight blank keys. You can select any of the eight keys for the fluid name.

Figure 10-11: Change Drug Configuration screen

To edit a drug or fluid name:

• Touch IV.• Select SETUP.• Touch CHANGE CONFIG.• Select EDIT DRUG or EDIT FLUID.• Select the name you want to change.• Touch CHANGE NAME.• Use alphanumeric keys to make necessary changes.• Touch ENTER.

IV - CHANGE CONFIGURATION - SELECT DRUG KEY TO RELABELAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

CHANGE NAME

Flexport System Interface Operations Manual 10-15

Page 158: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

After selecting a drug or fluid key to relabel, touch CHANGE NAME. The monitor displays the Change Drug (or Fluid) Name screen (refer to Figure 10-12).

Figure 10-12: Change Drug Name screen

• Enter each character or number using the alphanumeric keys.

• Touch DEL to eliminate the character highlighted by the cursor.

• Move through the name, character by character using the arrow keys.

• Touch CLEAR to eliminate the entire name and start over. The screen displays blank fields in place of the name.

• Touch ENTER to complete the edit and store the name in memory.

• Enter up to fifteen characters.

After you complete your edit and touch ENTER, the name will appear in the appropriate display for selecting drugs or fluids. The interface saves the name so you can recall it the next time you connect the Flexport system interface.

IV-CHANGE DRUG (or fluid) NAME= _________________________

0 1 2 3 4 5 6 7 8 9

A B C D E F G H I J

K L M N O P Q R S T

U V W X Y Z / % .

DEL ← → CLEAR ENTER

Flexport System Interface Operations Manual 10-16

Page 159: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Disconnecting a PumpWhen you touch DISCONNECT, the Flexport system interface discontinues the infusion record. If the pump is operating when you touch this key, the infusion record stops, a new record starts, and all pump setup information is cleared from the interface. If you touch DISCONNECT when the pump displays the status DISCONNECTED on the monitor, the setup information stored by the Flexport system interface is cleared and the line of pump data is removed from the display.

Note:• Touching the DISCONNECT key does NOT turn OFF the infusion pump.

• A pump can be disconnected only at the bedside monitor.

When you turn an infusion pump OFF at the pump itself, the pump status listed on the monitor changes to DISCONNECTED. At this point the interface completes the history for the current infusion (HISTORY shows the time that the infusion ended). If you turn ON the pump again, the interface continues to display the drug/fluid information entered previously.

Noting an EventThe EVENT key stores the time and pump status in memory. There is no immediate feedback when you touch this key. When you review the pump history, there is an entry labeled EVENT at the time you noted.

Events you may want to record include procedures such as changing IV lines or bags without turning the pump OFF.

Note:An event can be noted only at a bedside monitor.

To disconnect a pump:

• Touch IV.• Touch DISCONNECT.• Select a pump key.• Touch YES.

To note an event:

• Touch IV.• Touch EVENT.• Select a pump key.

Flexport System Interface Operations Manual 10-17

Page 160: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Displaying Infusion HistoryInfusion history displays in a table format listing the last 40 infusion events for the patient on four pages (refer to Figure 10-13). This view-only table is available at the bedside monitor only and is not available for remote viewing.

The data shown in the table does not automatically update while you are viewing it. To update the history data, re-enter the history table by touching PREVIOUS and then HISTORY.

Figure 10-13: Example of History table

The word ACTIVE appearing in the Ended column indicates that the pump is presently infusing the drug or fluid. The interface updates the log of active entries each minute.

The word EVENT appearing in the Ended column indicates the time at which the clinician touched the EVENT key to record a miscellaneous event into the pump history. Events that you might want to record include changing infusion lines or bags without turning the pump OFF.

To display infusion history:

• Touch IV.• Touch HISTORY.

DATE BEGAN ENDED PUMP DRUG/FLUID CONCENTRATION FLOW RATE VOL. INFSD

02/20 10:30am 12:10pm 12 DOPAMINE 400mg/250ml 30.0 ml/hr 50.0 ml

02/20 12:15pm EVENT 12 DOPAMINE 400mg/250ml 20.0 ml/hr 3.5ml

02/20 12:50pm 05:00pm 12 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/20 06:00pm 11:00pm 14 HEPARIN 2000u/500ml 100.0 ml/hr 500.0 ml

02/21 12:30am 01:00am 12 NOREPINEPHRINE 1mg/250ml 125.0 ml/hr 62.5 ml

02/21 01:30am ACTIVE 12 NOREPINEPHRINE 400mg/250ml 30.0 ml/hr 210.0 ml

02/21 01:30am 03:00am 14 HEPARIN 2000u/500ml 100.0 ml/hr 750.0 ml

02/21 03:30am 07:00am 13 D5W 100.0 ml/hr 350.0 ml

02/21 07:30am ACTIVE 14 D10 1/2NS 100.0 ml/hr 100.0 ml

02/21 08:00am ACTIVE 9 DOPAMINE 400mg/250ml 30.0 ml/hr 15.0 ml

Flexport System Interface Operations Manual 10-18

Page 161: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display or infusion history table. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the infusion information:

• Touch IV.• Touch PRINT.

-OR-• Touch HISTORY.• Touch PRINT.

Flexport System Interface Operations Manual 10-19

Page 162: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Status MessagesThe message DISCONNECTED indicates a problem with the pump or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pump is powered ON.

• Verify that all cables are secure.

Table 3 lists the status messages that the Flexport system interface reports for the pump on a central monitor, split screen.

Note:Not all messages will appear for each pump. Refer to the manual for your specific pump to determine which messages are used.

Table 3: Status Messages — Central Monitor, Split Screen

Message Text Indication

DISCONN Pump has been disconnected or turned OFF.

AIR ALARM Pump has detected air in IV tubing.

DOOR OPEN Pump has detected an open door.

INFSN. D. Primary infusion is complete.

KVO Pump pumping at KVO rate.

LOW BATT Battery requires recharging.

OCCLUSN Closed clamp, kinked, occluded tubing, or occluded filter detected.

OK Appears at pump sign ON.

PUMPING Pump is pumping normally.

PUMP OFF Pump on BCC is turned OFF.

STOPPED Pump is stopped but not turned OFF.

IN STAND Pump is in a standby mode.

BOLUS RUN Bolus injection in progress.

STANDBY Pump’s standby time has expired.

DROP ALM Drop sensor problem detected.

CHK. SYR Syringe alarm has been detected.

COMM FAILURE Communication with Flexport failed.

Flexport System Interface Operations Manual 10-20

Page 163: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Table 4 lists the status messages that the Flexport system interface reports for the pump on a remote monitor.

Table 4: Status Messages — Remote Monitor

Message Text Indication

DISC Pump has been disconnected or turned OFF.

AIR Pump has detected air in IV tubing.

DOPN Pump has detected an open door.

DONE Primary infusion is complete.

KVO Pump pumping at KVO rate.

BATT Battery requires recharging.

OCCL Closed clamp, kinked, occluded tubing, or occluded filter detected.

OK Appears at pump sign ON.

PUMP Pump is pumping normally.

OFF Pump on BCC is turned OFF.

STOP Pump is stopped but not turned OFF.

STND Pump is in a standby mode or standby time expired.

BOLS Bolus injection in progress.

DROP Drop sensor problem detected.

SYRG Syringe alarm has been detected.

Flexport System Interface Operations Manual 10-21

Page 164: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Braun Fluid Management System

Table 5 lists the status messages that the Flexport system interface reports for the pump on a bedside monitor.

Table 5: Status Messages — Bedside Monitor

Message Text Indication

DISCONNECTED Pump has been disconnected or turned OFF.

AIR ALARM Pump has detected air in IV tubing.

DOOR OPEN Pump has detected an open door.

INFSN. DONE Primary infusion is complete.

KVO Pump pumping at KVO rate.

LOW BATTERY Battery requires recharging.

OCCLUSION Closed clamp, kinked, occluded tubing, or occluded filter detected.

OK Appears at pump sign ON.

PUMPING Pump is pumping normally.

PUMP OFF Pump on BCC is turned OFF.

STOPPED Pump is stopped but not turned OFF.

IN STANDBY Pump is in a standby mode.

BOLUS RUN Bolus injection in progress.

STANDBY TIME Pump’s standby time has expired.

DROP ALARM Drop sensor problem has been detected.

CHK. SYRINGE Syringe alarm has been detected.

COMM FAILURE Communication with Flexport failed.

Flexport System Interface Operations Manual 10-22

Page 165: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 11-1

Criticare 504

Directory of Keys

SPO2

SpO2 - MAIN MENU

ALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFF

SpO2 - SETUP

SIZE SWEEP SPEED

LARGE TXT

SMALL TXT

SpO2 - SETUP - SIZEWAVEFORM

SIZE ↑ SIZE ↓ON OFF

SpO2 - ALARM LIMITS SpO2 - SETUP - SWEEP SPEED

SpO2 HR 50 mm/sec

25 mm/sec

12.5 mm/sec

6.25 mm/sec

SpO2 - ALARM LIMITS - HRHR HI =

XXLO=XX

SPO2

ON OFF HR

SpO2 - ALARM LIMITS - SpO2SPO2 HI =

XXLO=XX

SPO2

ON OFF HR

These keys appear on the bedside monitor only.

Page 166: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 167: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare 504

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a Criticare 504 Pulse Oximeter to be displayed on Spacelabs Healthcare monitors.

Configure the pulse oximeter as shown in Table 1. Consult the pulse oximeter manual for additional configuration setup procedures.

Table 1: Pulse Oximeter Configuration

Baud Rate Installation Kit Part Number

Transition Connector

Part Number

Minimum Software Revision

9600 045-0001-xx 131-0685-xx(25-pin) 3.5B6

Flexport System Interface Operations Manual 11-3

Page 168: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare 504

Connecting the Flexport System InterfaceTo connect the Flexport system interface to multiple pumps, complete the steps in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the pulse oximeter. To control other interface functions, refer to the sections that follow.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the pulse oximeter.

Flexport System Interface Operations Manual 11-4

Page 169: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare 504

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pulse oximeter.

Figure 11-1: Bedside monitor, large text screen

Figure 11-2: Bedside monitor, small text screen

Figure 11-3: Central monitor, split screen

� SPO2 parameter key

� Current SpO2 (%) value

� Alarm status bell (appears when alarms are turned ON)

� SpO2 pulse rate

� Patient/room ID

� �

SPO2

97SpO2

%

68HR

bpm

� �

� �

SPO2

SpO2= 97%

HR= 68bpm

� �

� �

� SpO2= 97% SPO2

� HR = 68bpm

� ICU 14

Flexport System Interface Operations Manual 11-5

Page 170: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare 504

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the SPO2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When both parameter alarms are turned OFF, status alarms will not cause the SPO2 key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the pulse

oximeter.

• Alarm limits on can only be adjusted at the pulse oximeter. Alarm violation is detected at the pulse oximeter, and the Flexport system interface reports the alarm if it is turned on for that parameter.

• Alarm tones on the pulse oximeter can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the pulse oximeter’s parameters. If alarms for both parameters are turned OFF, the Spacelabs Healthcare monitor displays SPO2 ALM OFF to the right of the SPO2 key. If alarms are turned ON for either parameter, a bell is displayed instead. The alarm bell flashes when an alarm limit is violated.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch SPO2.• Touch ALARM LIMITS.• Touch SPO2 or HR.• Touch the parameter alarm ON/OFF key.

To view alarm settings:

• Touch SPO2.• Touch VIEW ALARMS.

Flexport System Interface Operations Manual 11-6

Page 171: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare 504

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch SPO2.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch SPO2.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 11-7

Page 172: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare 504

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 6.25, 12.5, 25, and 50 mm/second.

The default setting is 25 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 11-5).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch SPO2.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch SPO2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch SPO2.• Touch FREEZE ON.

Flexport System Interface Operations Manual 11-8

Page 173: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare 504

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the oximeter or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pulse oximeter is powered ON.

• Verify that all cables are secure.

To print the pulse oximeter display:

• Touch SPO2.• Touch PRINT.

Flexport System Interface Operations Manual 11-9

Page 174: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 175: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET 601

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

OverviewThe 90434A-01 Flexport system interface enables information from a Criticare POET 601 analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the analyzer as shown in Table 1. Consult the analyzer manual for additional configuration setup procedures.

To complete the configuration for the Criticare POET 601, use the analyzer’s configuration display to set the PO+ET serial interval to 1 second and the serial output format to ASCII printer.

Table 1: Analyzer Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

19.2k 8-bit none 1 131-0685-xx (25-pin)

POET V1.3K, screen V1.6G

Flexport System Interface Operations Manual 12-1

Page 176: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET 601

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the analyzer.

Flexport System Interface Operations Manual 12-2

Page 177: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET 601

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the analyzer.

Figure 12-1: Bedside monitor, large text screen

Figure 12-2: Bedside monitor, small text sceen

Figure 12-3: Central monitor, split screen

� Current respiration rate value

� PO+ET parameter key

� Current EtCO2 (mmHg) value

� Alarm status bell (appears when alarms are turned ON)

� Current SaO2 (%) value

� Apnea monitoring status

� Patient/room ID

� � � �

21 RR PO+ET

24 ETCO2

BPM mmHg

97 SAO2

%

� � �

PO+ET

ETCO2 = 24 mmHgSAO2 = 97%RR = 21 BPM

� �

� �

� ETCO2 = 242 mmHg PO+ET

� SAO2 = 97%� RR = 21 BPM� APNEA = ON� ICU - 14

Flexport System Interface Operations Manual 12-3

Page 178: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET 601

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the PO+ET key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the PO+ET key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Table 2: Alarm Limit Ranges

Parameter Valid Range

ETCO2 (mmHg)Low 0 to 79 mmHg

High 1 to 80 mmHg

Respiration (BPM)Low 0 to 199 BPM

High 1 to 200 BPM

Apnea (seconds) 5 to 55 seconds

SAO2 (%)Low 0 to 95%

High 5 to 100%

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch PO+ET.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.To adjust alarm limits:• Select HI= or LO= (if applicable).• Use arrow keys to adjust.

Flexport System Interface Operations Manual 12-4

Page 179: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET 601

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 12-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays PO+ET ALM OFF to the right of the PO+ET key. If alarms are turned ON for any parameter, a bell is displayed instead. The alarm bell flashes when an alarm limit is violated.

Figure 12-4: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 12-3).

The default setting is LARGE TXT.

To view alarm settings:

• Touch PO+ET.• Touch VIEW ALARMS.

To select a display format:

• Touch PO+ET.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

RR 45 ETCO2 56 PO+ET

24 ETCO2

15 16 mmHg

APNEA ON SAO2 100 97 SAO2

90 %

Flexport System Interface Operations Manual 12-5

Page 180: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET 601

Viewing TrendsThe TABULAR TRENDS key enables you to display trend data in tabular format, adjust the display interval from 1 to 60 minutes, and print the trend information.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To view trends:

• Touch PO+ET.• Touch TABULAR TREND.• Touch TIME INTERVAL.• Use the arrow keys to adjust. • Select PREV READING or NEXT READING to scroll through the data.

To print the analyzer display:

• Touch PO+ET.• Touch PRINT.

Flexport System Interface Operations Manual 12-6

Page 181: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET 601

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 3 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 3: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the analyzer and the monitor has been interrupted. Yes

APNEA No CO2 waveform for more than 20 seconds. Yes

CAPNOMETER OFF Capnometer is not turned ON. No

HIGH AMBIENT Ambient light or noise is being detected by the pulse oximeter sensor. Yes

MINIMUM INSPIRED Low flow condition, check airway adapter. Yes

OCCLUDED LINE Sample line is occluded or the pump is defective. Yes

OXIMETER OFF Oximeter is not turned ON. No

SEARCH Pulse is too weak. Yes

SENSOR OFF PATIENT

Too much light is being transmitted through the pulse oximeter monitoring site. Yes

SENSOR SIGNAL Sensor is not attached, needs to be moved, or is malfunctioning. No

WARMING UP Sensor is warming up or the analyzer has not completed calibrating the capnography sensor. No

WATER TRAP FULL Water trap needs to be replaced. Yes

Flexport System Interface Operations Manual 12-7

Page 182: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 183: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Directory of KeysPO+ET

PO+ET - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFFRefer to

page 13-2 These keys display on the

bedside monitor only

PO+ET - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

PO+ET - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec

PO+ET - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

Flexport System Interface Operations Manual 13-1

Page 184: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

PO+ET - MAIN MENUALARM LIMITS SETUP

FREEZE SWITCH WAVES

VIEW ALARMS PRINT

ON OFFRefer to

page 13-1Refer to

page 13-1Refer to

page 13-1

PO+ET - ALARM LIMITS

SPO2 ETCO2 RESP I O2 I AGT E AGT

PO+ET - ALARM LIMITS - E AGTE AGT HI =

XXLO = XX

ETCO2ON OFF E AGT

PO+ET - ALARM LIMITS - I AGTI AGT HI =

XXLO = XX

I AGTON OFF E AGT

PO+ET - ALARM LIMITS - I O2I O2 HI =

XXLO = XX

I O2ON OFF I AGT

PO+ET - ALARM LIMITS - RESPRESP HI =

XXLO = XX

RESPON OFF I O2

PO+ET - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XX

ETCO2ON OFF RESP

PO+ET - ALARM LIMITS - SPO2SPO2 HI =

XXLO = XXON OFF

Flexport System Interface Operations Manual 13-2

Page 185: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90442A Flexport system interface enables information from a Criticare 602-3 POET-II analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the analyzer as shown in Table 1. Consult the analyzer manual for additional configuration setup procedures.

To complete the configuration for the analyzer, use the analyzer’s Configuration screen to set the POET serial interval to one second and the serial output format to “UNIV FLEXPORT” at 9600 baud.

Table 1: Analyzer Configuration

Baud Rate Minimum Software Revision

Installation Kit

Part Number

Transition Connector

Part Number

9600M4.2ES4.9

G4.70045-0002-xx 131-0685-xx (25-pin)

Flexport System Interface Operations Manual 13-3

Page 186: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the analyzer.

Figure 13-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 25-pin connector on the back of the analyzer.

� � � � �

35 ETCO2 21 I O2 1.1 I AGTPO+ET

98 SpO2

mmHg % % %

16 RESP 0 I NO2 1.0 E AGT 70 PULSE

BPM % % bpm

� � �

Flexport System Interface Operations Manual 13-4

Page 187: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Figure 13-2: Bedside monitor, small text screen

Figure 13-3: Central monitor, split screen

� Current end tidal CO2 value

� Current inspired oxygen value

� Current inspired agent value

� PO+ET parameter key

� Current SpO2 value

� Alarm status bell (appears when alarms are turned ON)

� Current pulse rate value

Current expired agent value

Current inspired N2O value

� Current respiration rate value

� Patient/room ID

� � �

PO+ET

SpO2= 98%PULSE= 70bpmETCO2= 35mmHgI O2= 21%

� �

� �� �

� SpO2= 98% I O2= 21% PO+ET

� PULSE= 70bpm I N2O= 0%

� ETCO2= 35mmHg I AGT=1.1%

� RESP= 16BPM E AGT=1.0%

� ICU 14

Flexport System Interface Operations Manual 13-5

Page 188: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the PO+ET key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the PO+ET key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• Alarm limits can only be adjusted at the analyzer. Alarm violation is detected at the analyzer, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch PO+ET.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 13-6

Page 189: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 13-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays PO+ET ALM OFF to the right of the PO+ET key. If alarms are turned ON for any parameter, a bell is displayed instead. The alarm bell flashes when an alarm limit is violated.

Figure 13-4: View Alarms screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

To view alarm settings:

• Touch PO+ET.• Touch VIEW ALARMS.

To turn the waveform display ON or OFF:

• Touch PO+ET.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

SpO2 1009264

0

RESP 660

5018

I AGT 5.00.05.00.0

PO+ET

98 SpO2

ETCO2 I O2 E AGT %

70 PULSE

bpm

Flexport System Interface Operations Manual 13-7

Page 190: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, 12.5, 25 and 50 mm/second.

The default setting is 12.5 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 13-4).

The default setting is LARGE TXT.

To adjust waveform size:

• Touch PO+ET.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch PO+ET.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch PO+ET.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 13-8

Page 191: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process and update the numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To freeze the waveform on the display:

• Touch PO+ET.• Touch FREEZE ON.

To print the analyzer display:

• Touch PO+ET.• Touch PRINT.

Flexport System Interface Operations Manual 13-9

Page 192: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET-II 602-3

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

The communication between the Flexport system interface and the analyzer has been interrupted. Yes

GAS WARMING UP The CO2 and AGENT sensors are warming up in the PO+ET analyzer. No

GAS AUTOCAL The PO+ET device is performing a periodic automatic calibration. No

OCCLUDED The sample line is occluded or the pump is defective. Yes

APNEA No CO2 waveform has been detected for more than the set APNEA period. Yes

INSERT TRAP The water trap is not attached. Yes

REPLACE TRAP The water trap needs to be replaced. Yes

MIXED AGENT More than one agent type has been detected in measurable quantities. Yes

WRONG AGENT The agent detected does not match settings. Yes

SENSOR Too much light is being transmitted through the analyzer site or the SpO2 sensor is not attached to a patient. Yes

SENSOR SIGNAL The SpO2 sensor is not plugged into the PO+ET or is malfunctioning. Yes

HIGH AMBIENT Ambient light or noise is being detected by the analyzer. Yes

SPO2 SEARCH Pulse is too weak. Yes

Flexport System Interface Operations Manual 13-10

Page 193: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Directory of KeysPOET2

POET2 - MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFFRefer to

page 14-2These keys are

displayed only on the bedside

monitor.

POET2 - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

POET2 - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec

POET2 - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

Flexport System Interface Operations Manual 14-1

Page 194: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

POET2 - MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFFRefer to

page 14-1Refer to

page 14-1

POET2 - ALARM LIMITS

ETCO2 RESP I O2 I AGT E AGT

POET2 - ALARM LIMITS - E AGTE AGT HI =

XXLO = XXON OFF

POET2 - ALARM LIMITS - I AGTI AGT HI =

XXLO = XXON OFF

POET2 - ALARM LIMITS - I O2I O2 HI =

XXLO = XXON OFF

POET2 - ALARM LIMITS - RESPRESP HI =

XXLO = XXON OFF

POET2 - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XXON OFF

Flexport System Interface Operations Manual 14-2

Page 195: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90442A-03 Flexport system interface enables information from a Criticare POET IQ2 analyzer to be displayed on Spacelabs Healthcare monitors.

Note:If the factory default setting is selected all user-configured settings will be erased. You will have to reconfigure the output format settings for the POET IQ2 to function properly.

Configuring the POET IQ21 Configure the analyzer as shown in Table 1. Consult the analyzer manual for additional configuration setup

procedures.

2 Set the serial output format to “FLEX PRT” at 9600 baud using the Configuration screen.

3 Cycle the power to implement the configuration changes.

Table 1: Output Format Settings

Baud Rate Software RevisionInstallation

Kit Part Number

Transition Connector

Part Number

9600 2.2 045-0160-xx 131-0686-xx

Flexport System Interface Operations Manual 14-3

Page 196: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the analyzer.

Figure 14-1: Bedside monitor, large text screen

Figure 14-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface.• Plug one end of the modular cable into the modular jack on the Flexport system

interface.• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 25-pin connector on the back of the analyzer.

� � � �

35 ETCO2 21 I O2 1.1 I AGTPOET2

mmHg % %

16RESP

BPM 0I NO2

% 1.0E AGT

%

� � �

� � �

POET2

ETCO2= 35mmHgRESP= 16BPMI O2= 21%E AGT= 1.0%

� �

Flexport System Interface Operations Manual 14-4

Page 197: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Figure 14-3: Central monitor, split screen

� Current end tidal CO2 value

� Current inspired oxygen value

� Current inspired agent value

� POET2 parameter key

� Alarm status bell (appears when alarms are turned ON)

� Current expired agent value

� Current inspired N2O value

Current respiration rate value

Patient/room ID

� � � �� �

I O2= 21% POET2

I N2O= 0%

� ETCO2= 35mmHg I AGT=1.1%

RESP= 16BPM E AGT=1.0%

ICU 14

Flexport System Interface Operations Manual 14-5

Page 198: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the POET2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• Alarm limits can only be adjusted at the analyzer. Alarm violation is detected at the analyzer, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch POET2.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 14-6

Page 199: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 14-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays POET2 ALM OFF to the right of the POET2 key. If alarms are turned ON for any parameter, a bell is displayed instead. The alarm bell flashes when an alarm limit is violated.

Figure 14-4: View Alarms screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

To view alarm settings:

• Touch POET2.• Touch VIEW ALARMS.

To turn the waveform display ON or OFF:

• Touch POET2.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

ETCO2 64 I O2 66 E AGT 5.0 POET2

0 0 0.0

RESP 66 I AGT 500 18

Flexport System Interface Operations Manual 14-7

Page 200: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, 12.5, 25 and 50 mm/second.

The default setting is 12.5 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 14-4).

The default setting is LARGE TXT.

To adjust waveform size:

• Touch POET2.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch POET2.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch POET2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 14-8

Page 201: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process and update the numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To freeze the waveform on the display:

• Touch POET2.• Touch FREEZE ON.

To print the analyzer display:

• Touch POET2.• Touch PRINT.

Flexport System Interface Operations Manual 14-9

Page 202: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Criticare POET IQ2

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

The communication between the Flexport system interface and the analyzer has been interrupted. Yes

GAS WARMING UP The CO2 and AGENT sensors are warming up in the POET2 analyzer. No

GAS AUTOCAL The POET2 device is performing a periodic automatic calibration. No

OCCLUDED The sample line is occluded or the pump is defective. No

APNEA No CO2 waveform has been detected for more than the set APNEA period. No

INSERT TRAP The water trap is not attached. No

REPLACE TRAP The water trap needs to be replaced. No

MIXED AGENT More than one agent type has been detected in measurable quantities. No

WRONG AGENT The agent detected does not match settings. No

Flexport System Interface Operations Manual 14-10

Page 203: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 15-1

Critikon Dinamap 1846 and 8100-Series

Directory of Keys

NIBP

NIBP - MAIN MENUALARM LIMITS PRINT PREV

READINGNEXT

READINGCHANGE CONFIG

NIBP - CHANGE CONFIGURATIONDISPLAY HR

ON OFF

NIBP - ALARM LIMITSALARMS HI=

150LO=100 ↑ ↓ SYS DIA MEAN

ON OFF

Page 204: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 205: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Critikon Dinamap 1846 and 8100-Series

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Displaying Heart Rate Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Displaying New or Previous Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing NIBP Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90435A-01 and 90435A-02 Flexport system interfaces enable information from Critikon 1800 and 8100-series non-invasive blood pressure (NIBP) monitors to be displayed on Spacelabs Healthcare monitors.

Configure the NIBP monitor as shown in Table 1. Consult the NIBP monitor manual for additional configuration setup procedures.

Table 1: NIBP Monitor Configuration

Flexport Part Number and NIBP

Monitor

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector &Cable Part

Number

Minimum Software Revision

90435A-01; use with 1846-series Critikon NIBP devices

600 8 NO 1 131-0685-xx 26300 Rev. E or later

90435A-02; use with 8100-series Critikon NIBP devices

600 8 NO 1 131-1051-xx 8100RAH Rev. C or later

Flexport System Interface Operations Manual 15-3

Page 206: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Critikon Dinamap 1846 and 8100-Series

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the NIBP monitor, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the NIBP monitor. To control other interface functions, refer to the sections that follow.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the NIBP monitor.

Flexport System Interface Operations Manual 15-4

Page 207: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Critikon Dinamap 1846 and 8100-Series

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the NIBP monitor.

Figure 15-1: Bedside monitor with heart rate

Figure 15-2: Central monitor, split screen

� NIBP data table

� Current intervaI for automatic readings

� NIBP parameter key

� Last systolic and diastolic reading

� Last mean reading

� Alarm limits for systolic (HI=150; LO=100) and diastolic (HI=100; LO=60); asterisk (*) indicates alarm limits have been set for mean pressure

� X indicates alarm limits set on the NIBP monitor are being monitored by the Spacelabs Healthcare monitor

Date and time of last reading

Alarm status bell (appears when alarms are turned ON)

� Patient/room ID

� � � � �

INTERVAL q15 min

(109)22:03 140/ 72 (89) mmHg HR= 88 (NIBP) N

IBP

13322:18 128/ 67 (80) mmHg HR= 86 (NIBP) m 15022:33 132/ 67 (83) mmHg HR= 86 (NIBP) LAST BP = m X S * 10022:48 134/ 69 (85) mmHg HR= 86 (NIBP) 22:03 86 H D 10023:03 133/ 86 (109) mmHg HR= 82 (NIBP) 02/07 g 60

� �

� � � �

X NIBP

LAST BP =133/86 (109)mmHg 22:03

02/07� ICU 14

Flexport System Interface Operations Manual 15-5

Page 208: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Critikon Dinamap 1846 and 8100-Series

Displaying Heart Rate DataYou can display NIBP values with or without heart rate.

If you select DISPLAY HR OFF, 10 numerical readings are displayed. These 10 readings contain the following data:

• time of the reading

• systolic, diastolic, and mean blood pressure values

Displaying New or Previous ReadingsThe monitor can display up to 10 NIBP readings at one time and can store up to 30 readings. Touch PREV READING to scroll up one line. Touch NEXT READING to scroll down one line.

To display heart rate on NIBP table:

• Touch NIBP.• Touch CHANGE CONFIG.• Select DISPLAY HR ON.

To view NIBP readings not currently on the display:

• Touch NIBP.• Touch PREV READING or NEXT READING.

Flexport System Interface Operations Manual 15-6

Page 209: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Critikon Dinamap 1846 and 8100-Series

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the NIBP key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When alarms are turned OFF, alarm status messages will still cause the NIBP key to flash.

The default setting for alarms is OFF.

If alarms are turned OFF, the Spacelabs Healthcare monitor displays NIBP ALM OFF to the right of the NIBP key. If alarms are turned ON, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Note:Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the NIBP monitor.

Printing NIBP Readings You can print either the complete NIBP screen or only the numeric values of the last reading.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch NIBP.• Touch ALARM LIMITS.• Select SYS, DIA, or MEAN.• Touch the parameter alarm ON/OFF key.To adjust alarm limits:• Select HI= or LO=.• Use arrow keys to adjust.

To print the NIBP screen:

• Touch NIBP.• Touch PRINT.

To print the most recent reading only:

• Touch RECORD.• Touch NIBP.

Flexport System Interface Operations Manual 15-7

Page 210: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Critikon Dinamap 1846 and 8100-Series

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the NIBP monitor or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the NIBP monitor is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the NIBP monitor during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOSTCommunication between the NIBP monitor and the Spacelabs Healthcare monitor has been interrupted.

Yes

800 EXCESS PRESSURE

Excessive pressure has been caused by a kinked or pinched hose or cuff lead. Can also be caused by a quick patient motion, squeezing of the cuff, or an internal pneumatic system leak or failure.

Yes

802 CHECK INTERNAL DINAMAP SWITCHES

One or more of the internal dip switches is set wrong or has failed. Yes

811 DINAMAP VOLTAGE OUT OF LIMITS

One or more internal DC power supply voltages are out of tolerance. Yes

822 DINAMAP NEEDS PRESSURE CALIBRATION

Dinamap microprocessor is unable to set the proper zero offset at the beginning of a determination.

Yes

833 EXCESS TIME: INFLATION

Pneumatic pump cannot reach the target pressure within 40 seconds, usually due to a leak in the cuff or hose or due to an internal pneumatic system leak or failure.

Yes

844 EXCESS TIME: TOTAL READING

A measurement has not been completed and 120 seconds has elapsed since the reading was initiated, usually due to a leak in the cuff, hose, or connectors; excessive motion artifact; or an internal pneumatic system leak or failure.

Yes

855 EXCESS TIME: ONE PRESSURE LEVEL

Monitor senses the system pressure has remained at one pressure level either because no matched pulses were found or despite efforts to step down the pressure, usually caused by excessive motion artifact, arrhythmias, or a kinked or pinched cuff or hose.

Yes

Flexport System Interface Operations Manual 15-8

Page 211: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Critikon Dinamap 1846 and 8100-Series

866 DINAMAP RAM FAILUREIf the PULSE screen flashes 866, random access memory (RAM) has failed. If the display is blank, the microprocessor has failed.

No

877 DINAMAP ROM FAILUREIf the PULSE screen flashes 877, read-only memory (ROM) has failed. If the display is blank, the microprocessor has failed.

No

899 DETERMINATION UNSUCCESSFUL

NIBP monitor is unable to detect a strong enough signal to take a reading. Yes

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 15-9

Page 212: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 213: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 16-1

Datex Engström Erica/Elvira Ventilator

Directory of KeysVENT

VENT - MAIN MENU *

ALARM LIMITS SETUP

FREEZEVIEW MODE VIEW

STATUS PRINTON OFF

VENT - SETUPWAVE SIZE

↑SIZE

↓SWEEP SPEED

WAVETYPEON OFF

VENT - SETUP - SWEEP SPEED VENT - SETUP - WAVE TYPE25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec PRESS FLOW VOL

VENT - ALARM LIMITS * *

PIPPIP

VminVmin

RRRR

FiO2FiO2

VENT - ALARM LIMITS - FiO2FiO2 HI =

XXXLO = XXX

FiO2ON OFF PIP

VENT - ALARM LIMITS - RRRR HI =

XXXLO = XXX

RRON OFF FiO2

VENT - ALARM LIMITS - VminVmin HI =

XXXLO = XXX

VminON OFF RR

VENT - ALARM LIMITS - PIPPIP HI =

XXXLO = XXX

PIPON OFF Vmin

*Available on the Elvira only.

Page 214: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 215: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail — Erica . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Display Detail — Elvira . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Viewing Ventilator Mode History (Elvira only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

OverviewThe 90436A-14 Flexport system interface enables information from a Datex Engström Erica or Elvira ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the appropriate ventilator manual for additional configuration setup procedures.

Note:• On the Erica ventilator, set the PRINTOUT INTERVAL knob to 10 seconds.

• On the Elvira ventilator, set the PRINTOUT INTERVAL knob to OFF and, in the setup program, set PRINT to TTY.

• To connect to the Erica ventilator, you must use the current loop-V24 converter option.

Table 1: Ventilator Configuration

Ventilator Baud Rate

Character Length

Parity Setting

Stop Bit

Engström Elvira 9600 8-bit none 1

Engström Erica 300 7-bit odd 1

Flexport System Interface Operations Manual 16-3

Page 216: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Figure 16-1: Modular cables

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

To connect the Flexport system interface:

For Erica:• Connect the short branch of the SDLC cable to the 9-pin male connector on the

Flexport system interface.• Connect the 15-pin connector of the current loop-V24 converter to the printer

connection on the ventilator.• Connect the 9-pin connector of the modular cable (P/N 012-0544-xx) to the analog

recorder connection.• Attach one end of the modular cable to the 25-pin connector of the current

loop-V24 converter.• Plug the other end of the modular cable into the 8-pin receptacle on the Flexport system

interface.For Elvira:• Connect the short branch of the SDLC cable to the 9-pin male connector on the

Flexport system interface.• Connect the male 9-pin connector of the modular cable (P/N 012-0545-xx) to the

“analog out” connection.• Connect the female 9-pin connector of the modular cable to the printer

RS-232 connection.• Plug the other end of the modular cable into the 8-pin receptacle on the Flexport system

interface.

25-pin connector to Ericaventilator current loop-V24 converter

modular cable to Erica

modular cable to Elvira

9-pin connector to Elviraventilator analog out

9-pin connector to Erica ventilator analog recorder

9-pin connector to Elviraventilator printer RS-232

Flexport System Interface Operations Manual 16-4

Page 217: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Figure 16-2: Elvira and Erica Flexport system interface connections

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

SDLCcable

SDLCterminator

modularcable

Erica

analogrecorder

Configuration for Erica

current loop-V24converter

SDLCcable

SDLCterminator

modularcable

Elvira

printer RS 232 connection

analog out connection

Configuration for Elvira

printer

SDLCconnector

SDLCconnector

connection

connection

SpacelabsHealthcare

monitor

SpacelabsHealthcare

monitor

systemFlexport

interface

systemFlexport

interface

Flexport System Interface Operations Manual 16-5

Page 218: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Display Detail — EricaWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the Erica ventilator.

Figure 16-3: Erica — bedside monitor, large text screen

Figure 16-4: Erica — bedside monitor, small text screen

Figure 16-5: Erica — central monitor, split screen

� Peak inspiratory pressure

� Mean airway pressure

� Spontaneous breath rate

� VENT parameter key

� Tidal volume

� � � � � �

22 PIP 8 MAP 12 SRTVENT

1.22 VtWAVE=PRESSBPM L

cmH2O cmH2O SMV Vmin

20 PEEP 21 FiO2 14.6 14.6cmH2O % L L

� � �

� � �

VENT

Vt = 1.22 L WAVE=PRESSVmin = 14.6 L

PIP = 22 cmH2OPEEP = 20 cmH2O

� � �

� � �

� Vt = 1.22 L MAP = 8 cmH2O VENT

Vmin = 14.6 L SMV = 14.6 L

� PIP = 22 cmH2O SRT = 12 BPM

� PEEP = 20 cmH2O FiO2 = 21%

� �

Flexport System Interface Operations Manual 16-6

Page 219: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

� Waveform identification label

� Alarm status bell (appears when alarms are turned ON)

Minute volume

Spontaneous minute volume

� Fractional inspired O2

� Positive end expiratory pressure

Display Detail — ElviraWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the Elvira ventilator.

Figure 16-6: Elvira — bedside monitor, large text screen

Figure 16-7: Elvira — bedside monitor, small text screen

� � � � � �

1.4 I:E 21 PIP 14.6 Vmin

VENT

1.22 VtWAVE=PRESSL

cmH2O L RR

22 FiO2 20 PEEP 18 MAP 12.2% cmH2O cmH2O BPM

� � �

� � �

VENT

Vt = 1.22 L WAVE=PRESSVmin = 12.2 BPM

PIP = 21 cmH2OI:E = 1.4

� � �

Flexport System Interface Operations Manual 16-7

Page 220: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Figure 16-8: Elvira — central monitor, split screen

� Ratio of inspiration time to expiration time (I:E ratio is typically expressed as 1:X.X, where X.X is the value displayed and equals E/I) *

� Peak airway pressure

� Minute volume

� VENT parameter key

� Tidal volume

� Waveform identification label

� Alarm status bell (appears when alarms are turned ON)

Breath rate

Mean airway pressure

� Positive end expiratory pressure

� Fractional inspired O2 **

* Values less than 1:1.0 are displayed as undefined.** The ventilator may display FiO2 but delay transmission to the Flexport system interface system for up to

seven minutes after turning ON the ventilator, exiting STAND BY mode, or adjusting the FiO2 setting. During this time, the Flexport system interface will show FiO2 as undefined.

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all Flexport system interface alarms are turned OFF, the status alarms will still cause the VENT key to flash.

Alarm settings default to OFF.

� � � �

RR = 12.2 BPM I:E = 1.4 VENT

� FiO2 = 22% PIP= 21 cmH2O

� Vt = 1.22 L PEEP = 20 cmH2O

� Vmin = 14.6 L MAP = 8 cmH2O

Flexport System Interface Operations Manual 16-8

Page 221: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• The alarm limits are displayed as question marks when insufficient data is provided by the ventilator. The ventilator determines when the limits have been exceeded and the Flexport system interface will report the alarm condition.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

• When connected to the Elvira, the RR alarm follows the definition given in the Elvira manual:

- Low RR corresponds to Apnea.

and

- High RR corresponds to a high spontaneous breathing frequency.

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be displayed on the bedside monitor, and a maximum of three messages can be displayed on the remote monitor (refer to Status Messages on page 16-13).

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

Flexport System Interface Operations Manual 16-9

Page 222: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Viewing Ventilator Mode History (Elvira only)Chronological history of the last eight ventilator mode settings are displayed in the waveform zone, with the most recent setting at the top of the left column and the oldest setting at the bottom of the right column. Possible modes include SIMV, EMMV, and CMV.

Figure 16-9: Ventilator mode history

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVE ON.

To view ventilator mode:

• Touch VENT.• Touch VIEW MODE.

To turn the waveform display ON or OFF:

• Touch VENT.• Touch SETUP.• Touch WAVE ON/OFF.

11:54 SIMV 11:41 CMV+SIGHVENT

Vt = 1.22L WAVE=PRESS11:53 EMMV 11:40 STAND BY RR= 12.2 BPM

11:51 SPONT. 11:27 CMV PIP = 21cmH2O11:50 TRIG. CMV I:E = 1.4

Flexport System Interface Operations Manual 16-10

Page 223: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVE ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, 12.5, and 25 mm/second.

The default setting is 12.5 mm/second.

Selecting a Waveform TypeYou can select either pressure (PRESS), flow (FLOW), or volume (VOL) waveforms for display. The three waveform displays are mutually exclusive.

To adjust waveform size:

• Touch VENT.• Touch SETUP.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch VENT.• Touch SETUP.• Select SWEEP SPEED.• Select the desired speed.

To select a waveform type:

• Touch VENT.• Touch SETUP.• Touch WAVE TYPE.• Select desired wave type.

Flexport System Interface Operations Manual 16-11

Page 224: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail — Erica on page 16-6 and Display Detail — Elvira on page 16-7).

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To freeze the waveform on the display:

• Touch VENT.• Select FREEZE ON.

To select a display format:

• Touch VENT.• Select LARGE TXT or SMALL TXT.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 16-12

Page 225: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Engström Erica/Elvira Ventilator

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

• Verify that the printer interval knob is set to 10 seconds on the Erica or OFF on the Elvira.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

AIR PRESS LO The supply air pressure has fallen below 265 kPa (2.7 kp/cm2). Yes

O2 PRESS LO The oxygen supply pressure has fallen below 265 kPa (2.7 kp/cm2). Yes

O2 ALARMEither the oxygen supply has fallen below 265 kPa (2.7 kp/cm2) or the FiO2 reading is outside the ventilator’s tolerance from the setting.

Yes

APNEA Apnea alarm. Yes

CHECK VENT Unknown alarm. Check the ventilator for greater detail. Yes

Flexport System Interface Operations Manual 16-13

Page 226: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 227: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Directory of KeysGAS

GAS - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 17-2

GAS - SETUP MENU

WAVEFORM SIZE SWEEP SPEED

LARGE TXT PERCENT PERCENTSMALL TXT mmHg kPa

GAS - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec

GAS - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

GAS - SETUP - WAVEFORM

ETCO2 ETN2O ETO2

Flexport System Interface Operations Manual 17-1

Page 228: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

GAS

GAS - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 17-1

GAS - ALARM LIMITS

RR ETCO2 FiCO2 ETN2O FINO2 ETO2 FiO2 APNEA

GAS - ALARM LIMITS - APNEAALARM HI =

20APNEA

ON OFF RR

GAS - ALARM LIMITS - FiO2ALARM HI =

XXLO=XX ↑ ↓

FiO2ON OFF APNEA

GAS - ALARM LIMITS - ETO2ALARM HI =

XXLO=XX ↑ ↓

ETO2ON OFF FiO2

GAS - ALARM LIMITS - FINO2ALARM HI =

XXLO=XX ↑ ↓

FINO2ON OFF ETO2

GAS - ALARM LIMITS - ETN2OALARM HI =

XXLO=XX ↑ ↓

ETN2OON OFF FINO2

GAS - ALARM LIMITS - FICO2ALARM HI =

XXLO=XX ↑ ↓

FICO2ON OFF ETN2O

GAS - ALARM LIMITS - ETCO2ALARM HI =

XXLO=XX ↑ ↓

ETCO2ON OFF FICO2

GAS - ALARM LIMITS - RRALARM HI =

XXLO=XX ↑ ↓

RRON OFF ETCO2

Flexport System Interface Operations Manual 17-2

Page 229: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90433A-07 Flexport system interface enables information from a Datex Normocap CD-200 or CD2-02 gas analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the gas analyzer as shown in Table 1. Consult the appropriate analyzer manual for additional configuration setup procedures.

Table 1: Analyzer Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

1200 8-bit none 1 131-1067-xx(25-pin)

Any version thatincorporates Protocol 1

Flexport System Interface Operations Manual 17-3

Page 230: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the gas analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the gas analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the gas analyzer.

Figure 17-1: Bedside monitor, large text screen

Figure 17-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the gas analyzer.

� � � � � �

[CO2]19 EtO2 39 EtN2O 21 RR

GAS

2.4 EtCO2

% % BPM %

21 FiO2 30 FiN2O 9.7 FiCO2

% % %

� �

� � �

[CO2]

GAS

EtCO2 = 24 mmHgFiCO2 = 9.7 mmHgRR = 21 BPM �FiN2O = 30%

Flexport System Interface Operations Manual 17-4

Page 231: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Figure 17-3: Central monitor, split screen

Note:EtO2 and FiO2 values are not displayed on the Datex Normocap CD-200.

� Current expired oxygen (O2) value

� Current expired nitrous oxide (N2O) value

� Current respiration rate value

� GAS parameter key

� Current EtCO2 (mmHg or %) value

� Current waveform type

� Alarm status bell (appears when alarms are turned ON)

Current FiCO2 (mmHg or %) value

Current inspired nitrous oxide (N2O) value

� Current inspired oxygen (O2) value

� Apnea monitoring status

Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the GAS key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the GAS key to flash.

Alarm settings default to OFF.

� � � �

� EtCO2 = 2.4% RR = 21 BPMGAS

FiO2 = 9.7% APNEA = 20s

� EtN2O = 39% EtO2 = 19%

FiN2O = 30% FiO2 = 21%

ICU - 14

� �

Flexport System Interface Operations Manual 17-5

Page 232: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the gas

analyzer.

• Apnea alarm limits can only be adjusted at the gas analyzer. Alarm violation is detected at the gas analyzer, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the gas analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

The Apnea alarm limit is fixed at 20 seconds and is determined by the gas analyzer. The EtCO2 and FiCO2 limits depend barometric pressure.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch GAS.• Touch ALARMS LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.To adjust alarm limits:• Select HI= or LO=.• Use the arrow keys to adjust.

Flexport System Interface Operations Manual 17-6

Page 233: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the gas analyzer’s parameters (refer to Figure 17-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays GAS ALM OFF to the right of the GAS key. If alarms are turned ON for any parameter, a bell is displayed instead. The alarm bell flashes when an alarm limit is violated.

Figure 17-4: View Alarms screen

Note:EtO2 and FiO2 values are not displayed on the Datex Normocap CD-200.

Viewing StatusThe VIEW STATUS key enables you to display recent status messages sent by the gas analyzer. A maximum of four messages can be simultaneously displayed on the bedside monitor. When more than four messages are received from the gas analyzer, they appear in the order they are received (refer to Status Messages on page 17-11). A maximum of three messages can be simultaneously displayed on the remote monitor.

To view alarm settings:

• Touch GAS.• Touch VIEW ALARMS.

To view analyzer status:

• Touch GAS.• Touch VIEW STATUS.

[CO2]EtO2

21EtN2O

40RR

90EtCO2

5.2GAS

EtCO2 = 24 mmHg12 5 1 2.0 FiCO2 = 35 mmHg

RR = 21 BPM

FiO221

FiN2O80

APNEA20

FiCO26.0 FiN2O = 30%

18 0 4.0

Flexport System Interface Operations Manual 17-7

Page 234: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Entering Setup InformationChanging setup information is helpful in optimizing the bedside monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Selecting a Waveform TypeYou can select EtCO2, EtN2O, or EtO2 waveforms for display.

The default waveform is EtCO2.

Note:EtO2 is only available only with the Normocap CD2-02.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

To select a waveform type:

• Touch GAS.• Touch SETUP.• Touch WAVEFORM.• Select a waveform type.

To turn the waveform display ON or OFF:

• Touch GAS.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

Flexport System Interface Operations Manual 17-8

Page 235: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:• The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

• When both the CD2-02 and CD-200 are “zeroing,” the waveform data will alternate between maximum and minimum values.

• The waveform will assume a tabular appearance that extends outside normal ranges.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 25, 12.5, 6.25 and 3.12 mm/second.

The default setting is 12.5 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 17-4).

The default setting is LARGE TXT.

To adjust waveform size:

• Touch GAS.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch GAS.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch GAS.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 17-9

Page 236: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Selecting a Unit of MeasurementYou can monitor gas values in units of percent (%), millimeters of mercury (mmHg), or kilo Pascals (kPa).

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select a unit of measurement:

• Touch GAS.• Touch SETUP.• Select PERCENT or mmHg or kPa.

To freeze the waveform on the display:

• Touch GAS.• Touch FREEZE ON.

To print the analyzer display:

• Touch GAS.• Touch PRINT.

Flexport System Interface Operations Manual 17-10

Page 237: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Normocap CD2-02/CD-200

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the gas analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the gas analyzer and the monitor has been interrupted. Yes

O2 INOPER O2 sensor is inoperative. Yes

OCCLUSION An occlusion is detected. Yes

AIR LEAK An air leak is detected. Yes

REBREAT ALM A rebreathing alarm is detected. Yes

O2 ZERO ERROR Zeroing error. Yes

CO2 ZERO ERROR Zeroing error. Yes

N2O ZERO ERROR Zeroing error. Yes

ZERO VALUE Zero value error condition detected. Yes

Flexport System Interface Operations Manual 17-11

Page 238: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 239: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Directory of KeysGAS

GAS - MENUALARM LIMITS SETUP

FREEZE VIEW STATUS PRINT

ON OFFRefer to

page 18-2

GAS - SETUP MENU

SIZE SWEEPSPEED WAVEFORM

LARGE TXT DISPLAY SETUPSMALL TXT

GAS - DISPLAY SETUP

SET 1 SET 2 SAVE SET

GAS - WAVEFORM SETUP

CO2 O2 Pleth Pressure

GAS - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec0.39

mm/sec

GAS - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

Flexport System Interface Operations Manual 18-1

Page 240: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

GAS

GAS - MENUALARM LIMITS SETUP

FREEZE VIEW STATUS PRINT

ON OFFRefer to

page 18-1Refer to

page 18-1Refer to

page 18-1Refer to

page 18-1

GAS - ALARM LIMITS

APNEA EtCO2 FiO2 PR MV Ppeak FiHAL SpO2

GAS - ALARM LIMITS - SPO2SpO2 HI =

ON/OFFLO =

ON/OFFAPNEA

ON OFF SpO2

GAS - ALARM LIMITS - FiHAL **FiHAL HI =

ON/OFFLO =

ON/OFFFiHAL

ON OFF SpO2

GAS - ALARM LIMITS - PpeakPpeak HI =

ON/OFFLO =

ON/OFFPpeak

ON OFF FiHAL

GAS - ALARM LIMITS - MVMV HI =

ON/OFFLO =

ON/OFFMV

ON OFF Ppeak

GAS - ALARM LIMITS - PRPR HI =

ON/OFFLO =

ON/OFFPR

ON OFF MV

GAS - ALARM LIMITS - FiO2FiO2 HI =

ON/OFFLO =

ON/OFFFiO2

ON OFF PR

GAS - ALARM LIMITS - EtCO2EtCO2 HI =

ON/OFFLO =

ON/OFFEtCO2

ON OFF FiO2

GAS - ALARM LIMITS - APNEAAPNEA APNEA

ON OFF EtCO2

Flexport System Interface Operations Manual 18-2

Page 241: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90443A-04 Flexport system interface enables information from a Datex Ultima gas monitor to be displayed on Spacelabs Healthcare monitors.

Configure the gas monitor as shown in Table 1. Consult the gas monitor manual for additional configuration setup procedures.

Table 1: Gas Monitor Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

InterfaceCable

Minimum Software Revision

1200 8 bit N 1 25 pin879016-2.2 (3 agents)882916-3.1 (5 agents)

Flexport System Interface Operations Manual 18-3

Page 242: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the gas monitor, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the gas monitor. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a bedside monitor that is properly connected to a Flexport system interface, the monitor displays data from the gas monitor (if powered ON) in one of four displays:

• Large text, Set 1 (factory default)

• Large text, Set 2

• Small text, Set 1

• Small text, Set 2

The default can be changed by touching SAVE SET while the desired data set is active.

Note:Information from the gas monitor is sent to the Flexport system interface at 10-second intervals. It may take up to 10 seconds for the data from the gas monitor to appear on the Spacelabs Healthcare monitor.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the 8-pin receptacle on the Flexport system interface.

• Plug the 25-pin connector on the other end of the modular cable into the RS-232 connector on the gas monitor labeled DIGITAL OUTPUT.

• Plug the 9-pin connector into the analog output connector on the side of the gas monitor.

Flexport System Interface Operations Manual 18-4

Page 243: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Figure 18-1: Set 1, large text screen

Figure 18-2: Set 1, small text screen

Figure 18-3: Set 1, split screen, central monitor

� � � � � � �

20 FeN2O 1.7 FeAGT 44FeO2 12 RR

GAS

4.9EtCO2 WAVE=

CO2% % % BPM %

14 FiN2O 2.5 FiAGT 50FiO2 99 SpO2 0.7 FiCO2% % % % %

� � �

� � �

GAS

EtCO2= 4.9% WAVE=CO2RR = 12 BPM �

FiO2 = 50% �FiAGT = 2.5%FiN2O = 14%

� � �

� EtCO2 = 4.9% FiCO2 = 0.7%GAS

� RR = 12 BPM SpO2 = 99%

� FeO2 = 44% FiO2 = 50%

� FeAGT = 1.7% FiAGT = 2.5%

� FeN2O = 20% FiN2O = 14%

� ICU 14

Flexport System Interface Operations Manual 18-5

Page 244: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Figure 18-4: Set 2, large text screen

Figure 18-5: Set 2, small text screen

Figure 18-6: Set 2, split screen, central monitor

� Current expired nitrous oxide value

� Current expired anesthetic agent value

� Current expired fraction O2 value

� Current respiratory rate

� GAS parameter key

� Current end tidal CO2 value

� Current waveform mode

Alarm status bell (appears when alarms are turned ON)

� � � � �

21 Ppeak 99 SpO2

GAS

4.9EtCO2

WAVE=CO2

cmH2O % %

1.22 MV 72 PR 0.7 FiCO2

L/min bpm %

� �

� � �

WAVE= CO2

GAS

EtCO2 = 4.9%

SpO2 = 99% �Ppeak = 21 CmH20MV = 1.22 L

� �

� �

� EtCO2= 4.9% FiCO2 = 0.7%GAS

� Ppeak = 21 cmH2O MV = 1.22 L

� SpO2 = 99%

� ICU - 14

Flexport System Interface Operations Manual 18-6

Page 245: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Current inspired fraction CO2 value

� Current SpO2 value

� Current inspired fraction O2 value

Current inspired anesthetic agent value (with mixed agent, secondary agent displays below primary agent)

� Current inspired nitrous oxide value

� Patient/room ID

� Peak inspiratory pressure

� Current minute volume

� Current pulse rate

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the GAS key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When an alarm is OFF, the high and low alarm limits are displayed as OFF. When an alarm is turned ON, the high and low limits are displayed as ON. When an alarm is active, the actual alarm limit is displayed in either the HI= limit or the LO= limit key.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the gas monitor.

• Alarm violation is detected at the gas monitor, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the gas monitor can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

• Alarm limit values are not available for the Minute Volume (MV) alarm. When this alarm is active, the HI or LO key flashes and the key continues to display ON instead of the alarm limit. The alarm limit values can be viewed on the gas monitor.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch GAS.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 18-7

Page 246: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the gas monitor. A maximum of four messages can be simultaneously displayed on the bedside monitor. When more than four messages are received from the gas monitor, they appear in the order they were received (refer to Status Messages on page 18-11). A maximum of three messages can be simultaneously displayed on the remote monitor.

Entering Setup InformationChanging setup information is helpful in optimizing the Spacelabs Healthcare monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

To view status:

• Touch GAS.• Touch VIEW STATUS.

To turn the waveform display ON or OFF:

• Touch GAS.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

Flexport System Interface Operations Manual 18-8

Page 247: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 18-4).

The default setting is LARGE TXT.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 0.39, 0.78, 1.56, 3.12, 6.25, 12.5, and 25 mm/second.

The default setting is 12.5 mm/second.

Selecting a Waveform TypeYou can select CO2, O2, PLETH, or PRESSURE waveforms for display. You can set PRESSURE from the gas monitor Setup menu.

To select a display format:

• Touch GAS.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To select a sweep speed:

• Touch GAS.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a waveform type:

• Touch GAS.• Touch SETUP.• Touch WAVEFORM.• Select a waveform type.

Flexport System Interface Operations Manual 18-9

Page 248: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Selecting a Display SetupTwo sets of patient data (SET 1 or SET 2) can be displayed on Spacelabs Healthcare monitors (refer to Display Detail on page 18-4). Bedside and remote monitors can be independently set. Touching SAVE SET will save the currently displayed set as the power ON default, including the LARGE TXT/SMALL TXT setting. (SMALL TXT is not available on a remote monitor.)

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select a display setup:

• Touch GAS.• Touch SETUP.• Touch DISPLAY SETUP.• Touch SET 1 or SET 2.• Touch SAVE SET to save display set 1 or 2.

To freeze the waveform on the display:

• Touch GAS.• Touch FREEZE ON.

Flexport System Interface Operations Manual 18-10

Page 249: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the gas monitor or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the gas monitor is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the gas monitor during Flexport system interface operation.

To print the gas monitor display:

• Touch GAS.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the gas monitor and the Spacelabs Healthcare monitor has been interrupted. Yes

CO2 ZERO ERROR Condensation or residual gases are affecting zero measurement. No

O2 ZERO ERROR Condensation or residual gases are affecting zero measurement. No

N2O ZERO ERROR Condensation or residual gases are affecting zero measurement. No

AGNT ZERO ERROR Condensation or residual gases are affecting zero measurement. No

REBREATH ALARM Inhaled (FICO2) concentration is higher than rebreathing limit. No

AIR LEAK Water trap is not connected, gas outlet is blocked, or leak in line. No

OCCLUSION Sampling line or the internal tubing is blocked. No

AGENT EXPIRE LO Expiratory anesthetic agent exceeds lower alarm limits. No

AGENT EXPIRE HI Expiratory anesthetic agent exceeds high alarm limits. No

LO PLETH SIGNAL Pulse wave amplitude is low and irregular. No

PULSE SEARCH Monitor is searching for a pulse. No

Flexport System Interface Operations Manual 18-11

Page 250: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Datex Ultima

PROBE OFF SpO2 probe is not attached to patient, or contact is loose. No

NO SPO2 PROBE SpO2 probe is not connected to monitor. No

OBSTRUCTION Tubes or airway is obstructed. No

DISCONNECTION Tubes or double lumen tube disconnected. No

LEAK Leak alarm is activated if there is a leak distal to D-LITE adapter. No

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 18-12

Page 251: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Directory of KeysANES

ANES - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 19-2

ANES - SETUP

WAVEFORM SIZE SWEEP SPEED

LARGE TXT PERCENT PERCENTSMALL TXT mmHg kPa

ANES - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec

ANES - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

ANES - SETUP - WAVEFORM TYPE

EtCO2 FLOW

Flexport System Interface Operations Manual 19-1

Page 252: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

ANES

ANES - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 19-1

ANES - ALARM LIMITS

EtCO2 FiAGT* FiO2 PIP Vmin

ANES - ALARM LIMITS - VminALARM HI =

XX.XLO = XX.X

VminON OFF EtCO2

ANES - ALARM LIMITS - PIPALARM HI =

XXXLO = XXX

PIPON OFF Vmin

ANES - ALARM LIMITS - FiO2ALARM HI =

XXXLO = XXX

FiO2ON OFF PIP

ANES - ALARM LIMITS - FiAGT*ALARM HI =

XXXLO = XXX

FiAGT*ON OFF FiO2

ANES - ALARM LIMITS - EtCO2ALARM HI =

XXLO = XX

EtCO2ON OFF FiAGT*

* FiAGT can also be FiDES, FiENF, FiHAL, FiISO, or FiSEV.

Flexport System Interface Operations Manual 19-2

Page 253: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90443A-08 Flexport system interface enables information from a Dräger Cato anesthesia machine to be displayed on Spacelabs Healthcare monitors.

Configure the anesthesia machine as shown in Table 1. Consult the anesthesia machine manual for additional configuration setup procedures.

Table 1: Anesthesia Machine Configuration

Baud Rate

Character Length

Parity Setting

StopBit

Minimum Medibus Protocol

9600 8 bit even 1 3.00

Flexport System Interface Operations Manual 19-3

Page 254: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the anesthesia machine, complete the steps listed in the following Quickstart.

The monitor is now ready to display information from the anesthesia machine. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the anesthesia machine.

Figure 19-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the anesthesia machine.

� � � � � �

0.89FiAGT

22RR

11.2Vmin

ANES

0.49 Vt [FLOW]

% BPM L L

35EtCO2

39PIP 30 FiO2

mmHg mbar %

� �

Flexport System Interface Operations Manual 19-4

Page 255: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Figure 19-2: Bedside monitor, small text

Figure 19-3: Central monitor, split screen

� Fractional inspired agent value can be FiHAL, FiDES, FiISO, FiENF, FiSEV, or FiAGT (if the agent is halothane, desflurane, isoflurane, enflurane, sevoflurane, or mixed/unidentified, respectively)

� Respiratory rate value

� Minute volume value

� ANES parameter key

� Expired tidal volume value

� Waveform identification label

� Alarm status bell (appears when alarms are turned ON)

Fraction inspired O2 value

Peak inspiratory pressure value

� End tidal CO2 value

� � �

[FLOW]ANES

FiO2=30%RR= 22BPMEtCO2=35mmHgVt= 0.49L

� � �

� � �

� EtCO2=35mmHg FiO2=30%%ANES

� RR=22BPM PIP=39mbar

� Vt=0.49L FiAGT=0.89%

� Vmin=11.2L

Flexport System Interface Operations Manual 19-5

Page 256: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the ANES key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the ANES key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the anesthesia

machine.

• Alarm limits can only be adjusted at the anesthesia machine. Alarm violation is detected at the anesthesia machine and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the anesthesia machine can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch ANES.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 19-6

Page 257: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the anesthesia machine’s parameters (refer to Figure 19-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays ANES ALM OFF to the right of the ANES key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 19-4: View Alarms screen

To view alarm settings:

• Touch ANES.• Touch VIEW ALARMS.

FiAGT 2.00 PIP 40 Vmin20.

0 ANES

FiO2=21%

0.50 4 4.0 RR= 22BPMEtCO2=35mmHg

EtCO2 75 FiO2 45 Vt= 0.49L00 18

Flexport System Interface Operations Manual 19-7

Page 258: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Viewing StatusThe VIEW STATUS key enables you to display recent status messages sent by the anesthesia machine. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three can be simultaneously displayed on the remote monitor (refer to Status Messages on page 19-11).

Figure 19-5: View Status screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Selecting a Waveform TypeYou can select either expired CO2 (EtCO2) or flow waveforms for display.

To view anesthesia machine status:

• Touch ANES.• Touch VIEW STATUS.

To select a waveform type:

• Touch ANES.• Touch SETUP.• Touch WAVEFORM.• Select EtCO2 or FLOW.

ANES

FiO2=30%

-- APNEA CO2 RR= 22BPM

-- CAL % O2 EtCO2=35mmHg

-- PRESSURE LTD Vt= 0.49L

Flexport System Interface Operations Manual 19-8

Page 259: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 3.12, 6.25, 12.5, and 25 mm/second.

The default setting is 12.5 mm/second.

To turn the waveform display ON or OFF:

• Touch ANES.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To change waveform size:

• Touch ANES.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch ANES.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

Flexport System Interface Operations Manual 19-9

Page 260: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 19-4).

The default setting is LARGE TXT.

Selecting Units for Expired CO2Expired CO2 can be displayed in units of percent (%), millimeters of mercury (mmHg), or kilo Pascals (kPa). The selection of mmHg or kPa is a function of the monitor’s setting. The units can be switched between the original setting and percent.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

To select a display format:

• Touch ANES.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To select EtCO2 units:

• Touch ANES.• Touch SETUP.• Select units.

To freeze the waveform on the display:

• Touch ANES.• Select FREEZE ON.

Flexport System Interface Operations Manual 19-10

Page 261: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the anesthesia machine or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the anesthesia machine is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the anesthesia machine during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the anesthesia machine and the monitor has been interrupted. Yes

CO2/AGT ERR Multigas monitor device failure. Yes

N2O ERR N2O measurement inoperable. Yes

AGT ERR Agent measurement inoperable. Yes

AGT WARM UP Agent monitor in low accuracy mode (warm up). No

AGT NOT SEL Agent not selected. No

APNEA VOL Apnea — no volume exhaled for 30 seconds. Yes

VOL CAL ? Flow calibration necessary. Yes

PAW NEGATIVE Mean airway pressure < -2 mbar. Yes

PRESS ERR Pressure measurement inoperable. Yes

AW-TMP INOP Airway temperature measurement inoperable. Yes

Flexport System Interface Operations Manual 19-11

Page 262: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cato

AW-TEMP HIGH Airway temperature > high limit. Yes

APNEA CO2 Apnea — no CO2 fluctuations for 30 seconds. Yes

INSP CO2 HI Inspiratory CO2 >high limit. Yes

CO2 LINE BLK CO2 patient sensor line blocked. Yes

CO2 ALRM OFF CO2 alarm disabled. Yes

CO2 WARM UP CO2 monitor in low accuracy mode (warm up). No

CO2 ERR CO2 device failure. Yes

SPEAKER FAIL Primary speaker failure. Yes

PM8050 ERR Hardware problem PM8050. Yes

COOLING PM8050 Check PM8050 cooling. Yes

CAL % O2? Oxygen analyzer not calibrated. No

% O2 $& <21 Inspiratory oxygen low limit <21%. No

% O2 ERR Inspiratory O2 measurement inoperable (alarm). Yes

FRESH GAS ? Check gas supply. Yes

PRESS EXP HI Expiratory pressure > high limit. Yes

VENT ERR Problems with anesthesia machine. Yes

COM VENT ERR Anesthesia machine communication lost. Yes

EXP-VALVE ? Check expiration-valve. Yes

PRESSURE LTD Pressure limited respiratory volume. No

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 19-12

Page 263: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 20-1

Dräger Cicero B

Directory of KeysVENT

VENT - MAIN MENUALARM LIMITS

LARGE TXT VIEW STATUS

VIEW ALARMS PRINT

SMALL TXT

VENT - ALARM LIMITS

EtCO2 SpO2 NIBP I.AGT FiO2 PIP Vmin

VENT - ALARM LIMITS - VminALARM HI =

XX.XLO = XX.X

VminON OFF EtCO2

VENT - ALARM LIMITS - PIPALARM HI =

XXXLO = XXX

PIPON OFF Vmin

VENT - ALARM LIMITS - FiO2ALARM HI =

XXXLO = XXX

FiO2ON OFF PIP

VENT - ALARM LIMITS - I.AGTALARM HI =

X.XXLO = X.XX

I.AGTON OFF FiO2

VENT - ALARM LIMITS - NIBPALARM HI =

XXXLO = XXX

NIBPON OFF I.AGT

VENT - ALARM LIMITS - SpO2ALARM HI =

XXXLO = XXX

SpO2ON OFF NIBP

VENT - ALARM LIMITS - EtCO2ALARM HI =

XXLO = XX

EtCO2ON OFF SpO2

Page 264: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 265: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90443A Flexport system interface enables information from a Dräger Cicero B ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Minimum Medibus Protocol

9600 8 bit even 1 3.00

Flexport System Interface Operations Manual 20-3

Page 266: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 20-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the ventilator.

� � � � � �

120 SYS 0.89I.AGT

22RR

11.2Vmin

VEET

0.49Vtexp

99 MEAN % BPM L L

60 DIA

59EtCO2

39PIP

98SpO2 30

FiO2

%NIBP mmHg mmHg mbar %

� � �

Flexport System Interface Operations Manual 20-4

Page 267: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

Figure 20-2: Bedside monitor, small text screen

Figure 20-3: Central monitor, split screen

� NIBP systolic, mean and diastolic blood pressure values

� Inspired agent value

� Respiratory rate value

� Minute volume value

� VENT parameter key

� Expired tidal volume value

� Alarm status bell (appears when alarms are turned ON)

Fraction inspired O2 value

Pulse oximetry O2 saturation value

� Peak inspiratory pressure value

� End tidal CO2 value

� �

VEET

EtCO2=59mmHgRR= 22BPM �Vtexp=0.49L �SpO2=98%

� � � �

� EtCO2=59mmHg FiO2=30%%

VENT

� RR=22BPM SpO2 = 98%

� Vtexp=0.49L PIP=39mbar

� Vmin=11.2L NIBP= 120/60mmHg

Flexport System Interface Operations Manual 20-5

Page 268: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 20-6

Page 269: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 20-4). If alarms for all ventilator parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 20-4: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three can be simultaneously displayed on the remote monitor (refer to Status Messages on page 20-9).

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

NIBP 150 I.AGT 2.00 PIP 40 Vmin 20.0 ANES

EtCO2=22mmHg70 0.50 4 4.0 RR= 20BPM

Vtexp=0.49L

EtCO2 75 SpO2 100 FiO2 45 SpO2= 96%00 85 18

Flexport System Interface Operations Manual 20-7

Page 270: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 20-4).

The default setting is LARGE TXT.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch VENT.• Select LARGE TXT or SMALL TXT.

To print a ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 20-8

Page 271: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the ventilator and the monitor has been interrupted. Yes

CO2/AGT ERR Multigas monitor device failure. Yes

N2O ERR N2O measurement inoperable. Yes

AGT ERR Agent measurement inoperable. Yes

A-VAP OFF Iris function disabled. Yes

AGT ALRM OFF Agent alarm disabled. Yes

AGT WARM UP Agent monitor in low accuracy mode (warm up). Yes

AGT NOT SEL Agent not selected. Yes

A-VAP CAL? Iris calibration necessary. Yes

A-VAP ERR Iris function inoperable. Yes

CHECK VAPOR Vapor not connected, known, or allowed. Yes

APNEA VOL Apnea — no volume exhaled for 30 seconds. Yes

APNEA PRES Apnea — pressure absent for 15 seconds. Yes

VOL ALRM OFF Volume alarm disabled. Yes

VOL CAL? Flow calibration necessary. Yes

VOL ERR Volume measurement inoperable (alarm). Yes

PRESS ERR Pressure measurement inoperable. Yes

AW-TMP INOP Airway temperature measurement inoperable. Yes

AW-TEMP HIGH Airway temperature > high limit. Yes

Flexport System Interface Operations Manual 20-9

Page 272: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

APNEA CO2 Apnea — no CO2 fluctuations for 30 seconds. Yes

INSP CO2 HI Inspiratory CO2 >high limit. Yes

CO2 LINE BLK CO2 patient sensor line blocked. Yes

APN.CO2’@OFF Apnea CO2 alarm disabled. Yes

CO2 ALRM OFF CO2 alarm disabled. Yes

CO2 WARM UP CO2 monitor in low accuracy mode (warm up). Yes

CO2 ERR CO2 device failure. Yes

MAN/SP’@MODE Manual spontaneous alarm mode ON. Yes

VENTMON COM? Communication error ventilator port. Yes

ECG COM ERR Communication error ECG-monitor port. Yes

INT COM ERR Internal communication error. Yes

PM8035 ERR Hardware problem PM 8035. Yes

PM8035 STBY PM 8035 standby. Yes

PM8035 TEST Self check PM 8035. Yes

BP CUFF ERR NIBP cuff improperly placed. Yes

BP CUFF DISC NIBP cuff disconnected or leaking. Yes

NIBP STANDBY NIBP monitor in standby mode. Yes

NIBP ERR NIBP device failure. Yes

CAL% O2? Oxygen analyzer not calibrated. Yes

% O2 ERR Inspiratory O2 measurement inoperable (alarm). Yes

NO SAO2 PULS No SAO2 pulse for 10 seconds. Yes

SAO2SEN DISC SaO2 sensor disconnected or fault. Yes

SAO2 ALRM OF Oximeter alarm disabled. Yes

SAO2 ERR Oximeter device failure. Yes

FRESH GAS? Check gas supply. Yes

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 20-10

Page 273: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero B

PRESS EXP HI Expiratory pressure > high limit. Yes

VENT ERR Problems with ventilator. Yes

COM VENT ERR Ventilator communication lost. Yes

EXP-VALVE? Check expiration-valve. Yes

PRESSURE LTD Pressure limited respiratory volume. Yes

MIN VOL LOW Minute volume less than low limit. Yes

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 20-11

Page 274: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 275: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Directory of KeysANES

ANES - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 21-2

ANES - SETUP

WAVEFORM SIZE SWEEP SPEED

LARGE TXT PERCENT PERCENTSMALL TXT mmHg kPa

ANES - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec

ANES - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

ANES - SETUP - WAVEFORM

EtCO2 PRESSURE

Flexport System Interface Operations Manual 21-1

Page 276: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

ANES

ANES - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 21-1

ANES - ALARM LIMITS

EtCO2 SpO2 NIBP FiAGT* FiO2 PIP Vmin

These keys appear only if data has been received.

ANES - ALARM LIMITS - VminALARM HI =

XX.XLO = XX.X

VminON OFF EtCO2

ANES - ALARM LIMITS - PIPALARM HI =

XXXLO = XXX

PIPON OFF Vmin

ANES - ALARM LIMITS - FiO2ALARM HI =

XXXLO = XXX

FiO2ON OFF PIP

ANES - ALARM LIMITS - FiAGT*ALARM HI =

X.XXLO = X.XX

FiAGT*ON OFF FiO2

ANES - ALARM LIMITS - NIBPALARM HI =

XXXLO = XXX

NIBPON OFF FiAGT

ANES - ALARM LIMITS - SpO2ALARM HI =

XXXLO = XXX

SpO2ON OFF NIBP

ANES - ALARM LIMITS - EtCO2ALARM HI =

XXLO = XX

EtCO2ON OFF SpO2

*FiAGT can also be FiDES, FiENF, FiHAL, FiISO, or FiSEV

Flexport System Interface Operations Manual 21-2

Page 277: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90443A-09 Flexport system interface enables information from a Dräger Cicero EM anesthesia machine to be displayed on Spacelabs Healthcare monitors.

Configure the anesthesia machine as shown in Table 1. Consult the anesthesia machine manual for additional configuration setup procedures.

Table 1: Anesthesia Machine Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Minimum Medibus Protocol

9600 8-bit even 1 3.00

Flexport System Interface Operations Manual 21-3

Page 278: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the anesthesia machine, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the anesthesia machine. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the anesthesia machine.

Figure 21-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the anesthesia machine.

� � � � � � �

120 SYS 0.89FiAGT

22 RR 11.2Vmin

ANES

0.49Vt PRES

99 MEAN % BPM L L

60 DIA

59EtCO2

39PIP

98SpO2 30

FiO2

%NIBP mmHg mmHg mbar %

� �

Flexport System Interface Operations Manual 21-4

Page 279: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Figure 21-2: Bedside monitor, small text

Figure 21-3: Central monitor, split screen

SpO2 and NIBP are only displayed if data is available.

� NIBP systolic, mean, and diastolic blood pressure values

� Inspired agent value

� Respiratory rate value

� Minute volume value

� ANES parameter key

� Expired tidal volume value

� PRES or ETCO2, depending upon waveform type selected

Alarm status bell (appears when alarms are turned ON)

Fraction inspired O2 value

� Pulse oximetry O2 saturation value

� Peak inspiratory pressure value

End tidal CO2 value

ANES

FiO2=30%RR= 22BPM �EtCO2=59mmHg Vt= 0.49L

� �� � �

EtCO2=59mmHg FiO2=30%%

ANES

� RR=22BPM SpO2 = 98%

� Vtexp=0.49L PIP=39mbar

� Vmin=11.2L NIBP= 120/60mmHg

Flexport System Interface Operations Manual 21-5

Page 280: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the ANES key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the ANES key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the anesthesia

machine.

• Alarm limits can only be adjusted at the anesthesia machine. Alarm violation is detected at the anesthesia machine and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the anesthesia machine can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch ANES.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 21-6

Page 281: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the anesthesia machine’s parameters (refer to Figure 21-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays ANES ALM OFF to the right of the ANES key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 21-4: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the anesthesia machine. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 21-11).

To view alarm settings:

• Touch ANES.• Touch VIEW ALARMS.

To view the anesthesia machine status:

• Touch ANES.• Touch VIEW STATUS.

FiAGT 2.00 PIP 40 Vmin 20.0 ANES

EtCO2=22%0.50 4 4.0 RR= 20BPM

EtCO2=49mmHg

NIBP 150 EtCO275 SpO2

100 FiO245 Vt= L

70 00 85 18

Flexport System Interface Operations Manual 21-7

Page 282: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Selecting a Waveform TypeYou can select either expired CO2 (EtCO2) or pressure waveforms for display.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only numeric values. When waveforms are OFF, the FREEZE ON/OFF keys, the SWEEP SPEED key, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

To select a waveform:

• Touch ANES.• Touch SETUP.• Touch WAVEFORM.• Select ETCO2 or PRESSURE.

To turn the waveform display ON or OFF:

• Touch ANES.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

Flexport System Interface Operations Manual 21-8

Page 283: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 3.12, 6.25, 12.5, and 25 mm/second.

The default setting is 12.5 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 21-4).

The default setting is LARGE TXT.

To adjust waveform size:

• Touch ANES.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch ANES.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch ANES.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 21-9

Page 284: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Selecting Units for Expired CO2Expired CO2 can be displayed units of percent (%), millimeters of mercury (mmHg), or kilo Pascals (kPa). The selection of mmHg or kPa is a function of the monitor’s setting. The units can be switched between the original setting and percent (%).

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select expired CO2 units:

• Touch ANES.• Touch SETUP.• Select units.

To freeze the waveform on the display:

• Touch ANES.• Select FREEZE ON.

To print the anesthesia machine display:

• Touch ANES.• Touch PRINT.

Flexport System Interface Operations Manual 21-10

Page 285: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the anesthesia machine or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the anesthesia machine is turned ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the anesthesia machine during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the anesthesia machine and the monitor has been interrupted. Yes

CO2/AGT ERR Multigas monitor device failure. Yes

N2O ERR N2O measurement is inoperable. Yes

AGT ERR Agent measurement is inoperable. Yes

AGT WARM UP Agent monitor in low accuracy mode (warm up). No

AGT NOT SEL Agent not selected. No

APNEA VOL Apnea — no volume exhaled for 30 seconds. Yes

VOL CAL? Flow calibration necessary. No

PAW NEGATIVE Mean airway pressure < -2 mbar. Yes

PRESS ERR Pressure measurement is inoperable. Yes

AW-TMP INOP Airway temperature measurement is inoperable. Yes

AW-TEMP HIGH Airway temperature > high limit. Yes

APNEA CO2 Apnea — no CO2 fluctuation for 30 seconds. Yes

INSP CO2 HI Inspiratory CO2 > high limit. Yes

CO2 LINE BLK CO2 patient sensor line blocked. Yes

CO2 ALRM OFF CO2 alarm disabled. No

CO2 WARM UP CO2 monitor in low accuracy mode (warm up). No

CO2 ERR CO2 device failure. Yes

SPEAKER FAIL Primary speaker failure. Yes

Flexport System Interface Operations Manual 21-11

Page 286: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

PM8060 ERR Hardware problem PM8060. Yes

COOLING PM8060 Check PM8060 cooling. Yes

CAL % O2? Oxygen analyzer not calibrated. No

% O2 $& <21 Inspired oxygen low limit < 21%. No

% O2 ERR Inspired O2 measurement inoperable (alarm). No

FRESH GAS? Check gas supply. Yes

PRESS EXP HI Expiratory pressure > high limit. Yes

VENT ERR Problems with anesthesia machine. Yes

COM VENT ERR Anesthesia machine communication lost. Yes

EXP-VALVE? Check expiration valve. Yes

PRESSURE LTD Pressure limited respiratory volume. No

APNEA Apnea by different or undefined source. Yes

VOL ERR Flow measurement inoperable. No

FLOW SENSOR? Check flow sensor. No

CHECK PAW Check PAW. No

HRT RATE FLOW Heart rate pulse < low limit. Yes

HRT RATE HI Heart rate > high limit. Yes

ECG ELEC. LL ECG left leg electrode is disconnected. No

ECG ELEC. LA ECG left arm electrode is disconnected. No

ECG ELEC. RA ECG right arm electrode is disconnected. No

ECG ELEC. N ECG neutral arm electrode is disconnected. No

NO S-T ECG S-T analysis impossible. No

ECG ELECTROD ECG electrode disconnected. No

S-T MV”# S-T segment-level deviation > limit. No

ASYSTOLY Asystole for 6 seconds. Yes

ECG ELEC. V ECG lead V disconnected. No

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 21-12

Page 287: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

ECG SMALL ECG signal too small. No

VENTR. FIBR. ECG ventricular fibrillation. Yes

PULSE DEFIC. Pulse deficit > high limit. No

NO IBP PULSE No invasive pulse for 20 seconds. No

SYS IBP1 LOW Invasive systolic 1 < low limit. No

SYS IBP1 HI Invasive systolic 1 > high limit. No

IBP1PULSE HI Invasive pulse 1 > high limit. No

CHECK IBP1 Check IBP1 sensor. No

IBP1PULSELO Invasive pulse 1 < low limit. No

NO IBP1 PULS No invasive 1 pulse. No

MEAN IBP1 LO Invasive mean 1 < low limit. No

MEAN IBP1 HI Invasive mean 1 > high limit. No

DIAS IBP1 LO Invasive diastolic 1 < low limit. No

DIAS IBP1 HI Invasive diastolic 1 > high limit. No

SYS IBP2 LOW Invasive systolic 2 < low limit. No

SYS IBP2 HI Invasive systolic 2 > high limit. No

IBP2PULSEHI Invasive pulse 2 > high limit. No

CHECK IBP2 Check IBP2 sensor. No

MEAN IBP2 LO Invasive mean 2 < low limit. No

MEAN IBP 2 HI Invasive mean 2 > high limit. No

IBP2PULSEHI Invasive pulse 2 > high limit. No

IBP2PULSELO Invasive pulse 2 < low limit. No

NO IBP2PULS No invasive 2 pulse. No

DIAS IBP2 LO Invasive diastolic 2 < low limit. No

DIAS IBP2 HI Invasive diastolic 2 > high limit. No

P-BOX ALAR OF Parameter-Box alarms disabled. No

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 21-13

Page 288: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Cicero EM

WATER TRAP? Check watertrap. No

NIBP SYS LOW NIBP systolic < low limit. No

NIBP DIAS LOW NIBP diastolic < low limit. No

NIBP SYS HI NIBP systolic > high limit. No

NIBP DIA HI NIBP diastolic > high limit. No

NO NIBP PULS No NIBP pulse. No

NIBP MEAN LO MIBP mean < low limit. No

NIBP MEAN HI NIBP mean > high limit. No

BP CUFF ERR NIBP cuff is improperly placed. No

BP CURR DISC NIBP cuff is disconnected or leaking. No

NIBP ALRM OF NIBP alarm is disabled. No

NIBP ERR NIBP device failure. No

NIBP MOTIONS NIBP motions detected. No

TEMP 1 LOW Temperature 1 < low limit. No

TEMP 2 LOW Temperature 2 < low limit. No

TMP 1 NOT CAL Temperature 1 is not calibrated or poorly calibrated. No

TMP 2 NOT CAL Temperature 2 is not calibrated or poorly calibrated. No

TEMP 1 ERR Temperature 1 probe is disconnected or fault. No

TEMP 2 ERR Temperature 2 probe is disconnected or fault. No

TEMP 1 HIGH Temperature1 > high limit. No

TEMP 2 HIGH Temperature 2 > high limit. No

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 21-14

Page 289: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 22-1

Dräger Evita II, Evita II Dura, and Babylog 8000

Directory of KeysVENT

VENT - MAIN MENUALARM LIMITS SETUP*

LARGE TXT VIEW STATUS

VIEW ALARMS PRINT

SMALL TXT

VENT - SETUP

SET 1 SET 2 SET 3 SAVE SET

VENT - ALARM LIMITS

VMIN VE* ETCO2* f-spo**Available on Evita II and Evita II Dura only.

VENT - ALARM LIMITS - f-spof-spo HI =

XXf-spo

ON OFF VE

VENT - ALARM LIMITS - EtCO2ETCO2 HI =

XX.XLO = XX.X

ETCO2ON OFF f-spo

VENT - ALARM LIMITS - VEVE HI =

X.XLO = X.X

VEON OFF ETCO2

VENT - ALARM LIMITS - VMINVMIN HI =

X.XXLO = X.XX ↑ ↓

ON OFF

Page 290: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 291: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

OverviewThe 90436A-10B Flexport system interface enables information from a Dräger Babylog 8000 ventilator to be displayed on Spacelabs Healthcare monitors.

The 90436A-10E Flexport system interface enables information from a Dräger Evita II and the Evita II Dura ventilators to be displayed on Spacelabs Healthcare monitors.

Note:This interface does not apply to the Evita IV ventilator. In this document, all references to the Evita II apply to both the Evita II and the Evita II Dura.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Note:To start communication with the Flexport system interface connected to the Babylog, select BabyLink from the Special Functions/Print menu and select EVEN from the Special Functions/Configuration/Com menu.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Minimum Medibus Protocol

Transition Connector

Part Number

9600 8 bit even 1 3.00 131-0686-xx

Flexport System Interface Operations Manual 22-3

Page 292: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator (depending on the ventilator model and the settings chosen).

Evita II — Set 1

Figure 22-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the cable into the transition connector.• Plug the transition connector into the RS-232 connector on the ventilator.

� � � � � �

22 MAP 19 FiO2 0.09 VT VENT

23PIP

mbar % L mbar

16 EtCO2 5.0EtCO2

21 f2.1

VE

L/minmbar % BPM

� �

Flexport System Interface Operations Manual 22-4

Page 293: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Figure 22-2: Bedside monitor, small text screen

Figure 22-3: Central monitor, split screen

� Mean airway pressure value

� Fraction inspired O2 value

� Tidal volume value

� VENT parameter key

� Peak inspiratory pressure value

� Alarm status bell (appears when alarms are turned ON)

� Expiratory minute volume value

Respiratory frequency value

End tidal CO2 value

� Positive end expiratory pressure value

� � � �

VENT

VT = 0.09LVE = 2.1L/minPIP = 23mbar VENT

ALMOFF

EtCO2 = 5.0%

� � � � � �

EtCO2=5% PIP=23mbar VENT

� FiO2=19% PEEP=16mbar

� MAP=22mbar VT=0.09L

f=21BPM VE=2.1L/min

Flexport System Interface Operations Manual 22-5

Page 294: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Evita II — Set 2

Figure 22-4: Bedside monitor, large text screen

Figure 22-5: Bedside monitor, small text

Figure 22-6: Central monitor, split screen

� Resistance value

� Intrinsic positive end expiratory pressure value

� VENT parameter key

� Spontaneous minute volume value

� Alarm status bell (appears when alarms are turned ON)

� � � � �

11 R 0 iPEEP VENT

00.5MVspo

mbar/L/s mbar L/min

13 C 0.0 OccPrs 10 Pmin36

f-spo

BPML/bar mbar mbar

� �

� � �

VENT

MVspo=00.5L/minR=11mbar/L/sC=13L/bar VENT

ALMOFF

f-spo=36BPM

� R=11mbar/L/s MVspo=00.5L/min VENT

C=13L/bar Pmin=10mbar

� iPEEP=0mbar f-spo=36BPM

OccPrs=0.0mbar

� � � �

Flexport System Interface Operations Manual 22-6

Page 295: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

� Spontaneous respiratory frequency value

� Minimal airway pressure value

Occlusion pressure value

Compliance value

Evita II — Set 3

Figure 22-7: Bedside monitor, large text screen

Figure 22-8: Bedside monitor, small text screen

� � � �

0 VCO2 VENT

22Pplat

ml/min mbar

25 Vds 10T

Cml

� �

� �

VENT

VCO2 = 0ml/minVds = 25ml �Pplat = 22mbar VENT

ALMOFF

T = 10C

� �

Flexport System Interface Operations Manual 22-7

Page 296: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Figure 22-9: Central monitor, split screen

� CO2 production value

� VENT parameter key

� Plateau pressure value

� Alarm status bell (appears when alarms are turned ON)

� Airway temperature value

� Dead space value

Babylog 8000

Figure 22-10: Bedside monitor, large text screen

Figure 22-11: Bedside monitor, small text screen

� VCO2=0ml/min VENT

� Vds=25ml

� pPlat=22mbar

� T=10C

� � � � � �

23PIP 18 MAP 0 SpVmin

VENT

0.63Vmin

mbar mbar % L

21FiO2

16PEEP

0LEAK 38

RR

% mbar % BPM

� �

� �

VENT

Vmin = 0.63 LRR = 38 BPM �PIP = 23 mbar VENT

ALMOFF

FiO2 = 21%

� �

Flexport System Interface Operations Manual 22-8

Page 297: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Figure 22-12: Central monitor, split screen

� Peak inspiratory pressure value

� Mean airway pressure value

� Spontaneous fraction minute volume value

� VENT parameter key

� Respiratory minute volume value

� Alarm status bell (appears when alarms are turned ON)

� Spontaneous respiratory rate value

Leakage value

Positive end expiratory pressure value

� Fraction inspired O2 value

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarms settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarms settings on the ventilator.

• The Evita II alarm limits can only be adjusted at the Evita II. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• The Babylog 8000 alarm limits are adjustable at the Flexport system interface.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

� � � �

� PIP = 23 mbar MAP = 18 mbar VENT

FPEEP = 16 mbar Vmin = 0.63L

� FiO2 = 21% SpVmin = 0%

� RR = 38BPM VLEAK = 0%

Flexport System Interface Operations Manual 22-9

Page 298: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 22-13 and Figure 22-14). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 22-13: View Alarms screen (Evita II)

Figure 22-14: View Alarms screen (Babylog 8000)

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Touch ALARM LIMITS.• Select alarm parameter.• Touch the parameter alarm ON/OFF key.To adjust alarm limits for the Babylog 8000:• Select HI= or LO=.• Use arrow keys to adjust.

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

EtCO2 6.0 f-spo 60 VENT

23PIP

0.0 mbar

VE 5.0 2.1VE

0.0 L/min

Vmin 1.00 VENT

1.00Vmin

0.10 L

20RR

BPM

Flexport System Interface Operations Manual 22-10

Page 299: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 22-12).

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 22-4).

The default setting is LARGE TXT.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

To select a display format:

• Touch VENT.• Select LARGE TXT or SMALL TXT.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 22-11

Page 300: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the Evita II ventilator. Messages marked with an “†” are also available for the Babylog 8000.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST†

Communication between the ventilator and the monitor has been interrupted. Yes

AIR SUPPLY? Check air supply. Yes

VOLUME INCONST Volume not constant. Yes

ASB>4sec Assisted spontaneous breathing > 4 seconds Yes

CO2 NOT CAL CO2 not calibrated. Yes

TACHYPN’@OFF Tachy/apnea alarm disabled. No

FLOW SENSOR? Check flow sensor. Yes

CO2 ALRM OFF CO2 alarm disabled. No

VOL ALRM OFF Volume alarm disabled. No

CO2 WARM UP CO2 monitor in low accuracy mode (warm up). No

CLEAN CO2 CO2 window occluded. Yes

CO2 Err CO2 device failure. Yes

RS232 INOP RS232 PORT. Yes

APNEA EVITA Apnea detected by Evita. Yes

PAW LOW Disconnection ventilator. Yes

EVITA ERR Problems with respirator (Evita). Yes

%O2 ERR† O2 measurement inoperable. Yes

Vol Err† Volume measurement inoperable. Yes

MV ALRM OFF MIN VOL alarm disabled. No

Flexport System Interface Operations Manual 22-12

Page 301: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita II, Evita II Dura, and Babylog 8000

PRESS ERR Pressure measurement inoperable. Yes

EXP-VALUE? Check expiration value. Yes

COOLING INOP Too high respirator device temperature. Yes

AW-TEMP INOP Airway temp measurement inoperable. Yes

AW-TEMP SENS Check airway temp sensor. Yes

AW-TEMP HIGH Airway temperature > high limit. Yes

SYSTEM OPEN Breathing system vented. No

%O2 HIGH % Oxygen > high limit. Yes

%O2 LOW % Oxygen < low limit. Yes

MIXER INOP Gas mixer inoperable. Yes

TIME LIMITED Time limited respiratory volume. No

PRESSURE LTD Pressure limited respiratory volume. No

COOLING INOP High respirator device temperature. Yes

SYNCHRO INOP Respirator synchronization inoperable. Yes

CYCLE FAILED Fail to cycle. Yes

CO2 SENS? CO2 sensor disconnected or fault. Yes

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 22-13

Page 302: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 303: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 23-1

Dräger Evita 4

Directory of KeysVENT

VENT - MAIN MENUALARM LIMITS SETUP

LARGE TXT VIEW STATUS

VIEW ALARMS PRINT

SMALL TXT

VENT - SETUP

SET 1 SET 2 SET 3 SAVE SET

VENT - ALARM LIMITS

MV ETCO2 f-spn

VENT - ALARM LIMITS - f-spnf-spn HI =

XXf-spn

ON OFF MV

VENT - ALARM LIMITS - EtCO2ETCO2 HI =

XX.XLO = XX.X

ETCO2ON OFF f-spn

VENT - ALARM LIMITS - MVMV HI =

X.XLO = X.X

MVON OFF ETCO2

Page 304: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 305: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90436A-16 Flexport system interface enables information from a Dräger Evita 4 ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Minimum Medibus Protocol

Transition Connector

Part Number

9600 8 bit even 1 3.00 131-0686-xx

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface.• Plug one end of the modular cable into the modular jack on the Flexport system

interface.• Plug the other end of the cable into the transition connector.• Plug the transition connector into the RS-232 connector on the ventilator.

Flexport System Interface Operations Manual 23-3

Page 306: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator. EtCO2 units of measure are determined by the ventilator configuration. They may display as percentages (refer to Figure 23-1), mmHg, or kPa. Pressures display as mbar, and temperatures display as ° C.

Evita 4 — Set 1

Figure 23-1: Bedside monitor, large text screen

Figure 23-2: Bedside monitor, small text screen

Figure 23-3: Central monitor, split screen

� � � � � �

22Pmean 19 FiO2 0.09 VTe

VENT

23Ppeak

mbar % L mbar

16PEEP

5.0EtCO2

21f 2.1

MV

mbar % BPM L/min

� �

� � � �

VENT

VT = 0.09LMV = 2.1L/minPpeak = 23mbar VENT

ALMOFF

EtCO2 = 5.0%

� � � � �

EtCO2=5% Ppeak=23mbar VENT

� FiO2=19% PEEP=16mbar

� Pmean=22mbar VTe=0.09L

f=21BPM MV=2.1L/min

Flexport System Interface Operations Manual 23-4

Page 307: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

� Mean airway pressure value

� Fraction inspired O2 value

� Expiratory tidal volume value

� VENT parameter key

� Peak inspiratory pressure value

� Alarm status bell (appears when alarms are turned ON)

� Minute volume value

Respiratory frequency value

End tidal CO2 value

� Positive end expiratory pressure value

Evita 4 — Set 2

Figure 23-4: Bedside monitor, large text screen

Figure 23-5: Bedside monitor, small text

� � � � �

11 R 0 PEEPi VENT

00.5 MVspn

mbar/L/s mbar L/min

13 C 0.0 PO.1 10 Pmin 36 f-spn

L/bar mbar mbar BPM

� �

� �

VENT

MVspn=00.5L/minR=11mbar/L/s �C=13L/bar VENT

ALMOFF

f-spn=36BPM

Flexport System Interface Operations Manual 23-5

Page 308: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

Figure 23-6: Central monitor, split screen

� resistance value

� intrinsic positive end expiratory pressure value

� VENT parameter key

� spontaneous breathed minute volume value

� alarm status bell (appears when alarms are turned ON)

� spontaneous respiratory portion of overall frequency value

� minimal airway pressure value

occlusion pressure value within 100 ms

compliance value

� R=11mbar/L/s MVspn=00.5L/min VENT

C=13L/bar Pmin=10mbar

� PEEPi=0mbar f-spn=36BPM

� PO.1=0.0mbar

� � � �

Flexport System Interface Operations Manual 23-6

Page 309: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

Evita 4 — Set 3

Figure 23-7: Bedside monitor, large text screen

Figure 23-8: Bedside monitor, small text screen

Figure 23-9: Central monitor, split screen

� CO2 production value

� VENT parameter key

� End inspiratory airway pressure

� Alarm status bell (appears when alarms are turned ON)

� Inspiratory breathing gas temperature value

� Serial dead space value

� � � �

0 VCO2 VENT

22 Pplat

ml/min mbar

25 Vds 10 T

ml C

� �

� �

VENT

VCO2 = 0ml/minVds = 25ml �Pplat = 22mbar VENT

ALMOFF

T = 10C

� �

� VCO2=0ml/min VENT

� Vds=25ml

� Pplat=22mbar

� T=10C

Flexport System Interface Operations Manual 23-7

Page 310: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarms settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarms settings on the ventilator.

• The Evita 4 alarm limits can only be adjusted at the Evita 4. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF for the Spacelabs Healthcare monitor:

• Touch VENT.• Touch ALARM LIMITS.• Select alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 23-8

Page 311: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 23-10). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 23-10: View Alarms screen (Evita 4)

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 23-11).

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

EtCO2 6.0 f-spn 60 VENT

23Ppeak

0.0 mbar

MV 5.0 2.1MV

0.0 L/min

Flexport System Interface Operations Manual 23-9

Page 312: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 23-4).

The default setting is LARGE TXT.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch VENT.• Select LARGE TXT or SMALL TXT.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 23-10

Page 313: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the Evita 4 ventilator.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the ventilator and the monitor has been interrupted. Yes

AIR SUPPLY? Check air supply. Yes

VOLUME INCONST Volume not constant. Yes

ASB>4sec Assisted spontaneous breathing > 4 seconds. Yes

CO2 NOT CAL CO2 not calibrated. Yes

TACHYPN’@OFF Tachy/apnea alarm disabled. No

FLOW SENSOR? Check flow sensor. Yes

CO2 ALRM OFF CO2 alarm disabled. No

VOL ALRM OFF Volume alarm disabled. No

CO2 WARM UP CO2 monitor in low accuracy mode (warm-up). No

CLEAN CO2 CO2 window occluded. Yes

CO2 Err CO2 device failure. Yes

RS232 INOP RS232 PORT. Yes

APNEA EVITA Apnea detected by Evita. Yes

PAW HIGH Airway pressure > high limit. Yes

PAW LOW Disconnection ventilator. Yes

EVITA ERR Problems with respirator (Evita). Yes

%O2 ERR O2 measurement inoperable. Yes

Vol Err Volume measurement inoperable. Yes

Flexport System Interface Operations Manual 23-11

Page 314: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4

MV ALRM OFF MIN VOL alarm disabled. No

PRESS ERR Pressure measurement inoperable. Yes

EXP-VALUE? Check expiration value. Yes

COOLING INOP Too high respirator device temperature. Yes

AW-TEMP INOP Airway temp measurement inoperable. Yes

AW-TEMP SENS Check airway temp sensor. Yes

AW-TEMP HIGH Airway temperature > high limit. Yes

SYSTEM OPEN Breathing system vented. No

%O2 HIGH % Oxygen > high limit. Yes

%O2 LOW % Oxygen < low limit. Yes

TIDVOL HIGH Tidal volume > high limit. Yes

MIXER INOP Gas mixer inoperable. Yes

TIME LIMITED Time limited respiratory volume. No

PRESSURE LTD Pressure limited respiratory volume. No

COOLING INOP High respirator device temperature. Yes

SYNCHRO INOP Respirator synchronization inoperable. Yes

CYCLE FAILED Fail to cycle. Yes

CO2 SENS? CO2 sensor disconnected or fault. Yes

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 23-12

Page 315: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 24-1

Dräger Evita 4/XL

Directory of Keys

VENT

VENT - MAIN MENU

ALARM LIMITS SETUP

LARGE TXT VIEW STATUS

VIEW ALARMS PRINT

SMALL TXT

VENT - SETUP

SET 1 SET 2 SET 3 SAVE SET

VENT - ALARM LIMITS

MV ETCO2 f-spn

VENT - ALARM LIMITS - f-spnf-spn HI =

XXf-spn

ON OFF MV

VENT - ALARM LIMITS - EtCO2ETCO2 HI =

XX.XLO = XX.X

ETCO2

ON OFF f-spn

VENT - ALARM LIMITS - MVMV HI =

X.XLO = X.X

MV

ON OFF ETCO2

Page 316: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 317: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90436A-16 Flexport system interface enables information from a Dräger Evita 4 or Evita XL ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Minimum Medibus Protocol

Transition Connector

Part Number

9600 8 bit even 1 3.00 131-0686-xx

Flexport System Interface Operations Manual 24-3

Page 318: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator. EtCO2 units of measure are determined by the ventilator configuration. They may display as percentages (refer to Figure 24-1), mmHg, or kPa. Pressures display as mbar, and temperatures display as ° C.

Evita 4/XL — Set 1

Figure 24-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface.• Plug one end of the modular cable into the modular jack on the Flexport system

interface.• Plug the other end of the cable into the transition connector.• Plug the transition connector into the RS-232 connector on the ventilator.• Connect the SDLC cable to the Ultraview monitor/housing.

� � � � � �

22 Pmean 19 FiO2 0.09 VTe VENT

23 Ppeak

mbar % L mbar

16 PEEP 5.0 EtCO2 21 f 2.1 MV

mbar % BPM L/min

� �

Flexport System Interface Operations Manual 24-4

Page 319: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

Figure 24-2: Bedside monitor, small text screen

Figure 24-3: Central monitor, split screen

� Mean airway pressure value

� Fraction inspired O2 value

� Expiratory tidal volume value

� VENT parameter key

� Peak inspiratory pressure value

� Alarm status bell (appears when alarms are turned ON)

� Minute volume value

Respiratory frequency value

End tidal CO2 value

� Positive end expiratory pressure value

� � � �

VENT

VT = 0.09LMV = 2.1L/minPpeak = 23mbar VENT

ALMOFF

EtCO2 = 5.0%

� � � � �

EtCO2=5% Ppeak=23mbar VENT

� FiO2=19% PEEP=16mbar

� Pmean=22mbar VTe=0.09L

f=21BPM MV=2.1L/min

Flexport System Interface Operations Manual 24-5

Page 320: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

Evita 4/XL — Set 2

Figure 24-4: Bedside monitor, large text screen

Figure 24-5: Bedside monitor, small text

Figure 24-6: Central monitor, split screen

� Resistance value

� Intrinsic positive end expiratory pressure value

� VENT parameter key

� Spontaneous breathed minute volume value

� Alarm status bell (appears when alarms are turned ON)

� Spontaneous respiratory portion of overall frequency value

� � � � �

11 R 0 PEEPi VENT

00.5 MVspn

mbar/L/s mbar L/min

13 C 0.0 PO.1 10 Pmin 36 f-spn

L/bar mbar mbar BPM

� �

� �

VENT

MVspn=00.5L/minR=11mbar/L/s �C=13L/bar VENT

ALMOFF

f-spn=36BPM

� R=11mbar/L/s MVspn=00.5L/min VENT

C=13L/bar Pmin=10mbar

� PEEPi=0mbar f-spn=36BPM

PO.1=0.0mbar

� � � �

Flexport System Interface Operations Manual 24-6

Page 321: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

� Minimal airway pressure value

Occlusion pressure value within 100 ms

Compliance value

Evita 4/XL — Set 3

Figure 24-7: Bedside monitor, large text screen

Figure 24-8: Bedside monitor, small text screen

Figure 24-9: Central monitor, split screen

� CO2 production value

� VENT parameter key

� End inspiratory airway pressure

� � � �

0 VCO2 VENT

22 Pplat

ml/min mbar

25 Vds 10 T

ml C

� �

� �

VENT

VCO2 = 0ml/minVds = 25ml �Pplat = 22mbar VENT

ALMOFF

T = 10C

� �

� VCO2=0ml/min VENT

� Vds=25ml

� Pplat=22mbar

� T=10C

Flexport System Interface Operations Manual 24-7

Page 322: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

� Alarm status bell (appears when alarms are turned ON)

� Inspiratory breathing gas temperature value

� Serial dead space value

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarms settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarms settings on the ventilator.

• The Evita 4 and Evita XL alarm limits can only be adjusted at the Evita 4/XL. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF for the Spacelabs Healthcare monitor:

• Touch VENT.• Touch ALARM LIMITS.• Select alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 24-8

Page 323: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 24-10). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 24-10: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 24-11).

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

EtCO2 6.0 f-spn 60 VENT

23Ppeak

0.0 mbar

MV 5.0 2.1MV

0.0 L/min

Flexport System Interface Operations Manual 24-9

Page 324: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 24-4).

The default setting is LARGE TXT.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch VENT.• Select LARGE TXT or SMALL TXT.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 24-10

Page 325: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the Evita 4 or Evita XL ventilator.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the ventilator and the monitor has been interrupted. Yes

AIR SUPPLY? Check air supply. Yes

VOLUME INCONST Volume not constant. Yes

ASB>4sec Assisted spontaneous breathing > 4 seconds. Yes

CO2 NOT CAL CO2 not calibrated. Yes

TACHYPN’@OFF Tachy/apnea alarm disabled. No

FLOW SENSOR? Check flow sensor. Yes

CO2 ALRM OFF CO2 alarm disabled. No

VOL ALRM OFF Volume alarm disabled. No

CO2 WARM UP CO2 monitor in low accuracy mode (warm-up). No

CLEAN CO2 CO2 window occluded. Yes

CO2 Err CO2 device failure. Yes

APNEA EVITA Apnea detected by Evita. Yes

PAW HIGH Airway pressure > high limit. Yes

PAW LOW Ventilator disconnection. Yes

EVITA ERR Problems with respirator (Evita). Yes

%O2 ERR O2 measurement inoperable. Yes

Vol Err Volume measurement inoperable. Yes

MV ALRM OFF MIN VOL alarm disabled. No

Flexport System Interface Operations Manual 24-11

Page 326: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Evita 4/XL

PRESS ERR Pressure measurement inoperable. Yes

EXP-VALUE? Check expiration value. Yes

COOLING INOP Too high respirator device temperature. Yes

AW-TEMP INOP Airway temp measurement inoperable. Yes

AW-TEMP SENS Check airway temp sensor. Yes

AW-TEMP HIGH Airway temperature > high limit. Yes

SYSTEM OPEN Breathing system vented. No

%O2 HIGH % Oxygen > high limit. Yes

%O2 LOW % Oxygen < low limit. Yes

TIDVOL HIGH Tidal volume > high limit. Yes

MIXER INOP Gas mixer inoperable. Yes

TIME LIMITED Time limited respiratory volume. No

PRESSURE LTD Pressure limited respiratory volume. No

COOLING INOP High respirator device temperature. Yes

SYNCHRO INOP Respirator synchronization inoperable. Yes

CYCLE FAILED Fail to cycle. Yes

CO2 SENS? CO2 sensor disconnected or fault. Yes

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 24-12

Page 327: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 25-1

Dräger Irisa/Evita

Directory of Keys

VENT

Note: The grayed keys are not function keys and are only used to display clinical data. (This data display is delayed for one or two breaths.)

VENT - MAIN MENUALARM LIMITS SETUP MECHANICS MODE

CMV*VIEW

ALARMS PRINT

*This could also be: CMV/A, S/MV, V/MV/PS, SPONT, PS, CPAP, CPAP/PS, MMV, MMV/PS,

APRV, MASTER, SLAVE, AP VENT, or UNKNOWN MODE

LARGE TXT

SMALL TXT

VENT - C in ml/cmH2O VENT - R in cmH2O/L/SC = XXX R = XXX

VENT - ALARM LIMITS

VMIN PIP

VENT - ALARM LIMITS - PIPPIP

ON OFF

VENT - ALARM LIMITS - VMINVMIN

ON OFF

Page 328: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 329: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Irisa/Evita

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

OverviewThe 90436A-09 Flexport system interface enables information from a Dräger Irisa/Evita ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the appropriate ventilator manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

1200 7-bit even 2 131-0697-xx (25-pin)

9.0, 9.5, 9.6 (Irisa)9.07 (Evita)

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector of the ventilator.

Flexport System Interface Operations Manual 25-3

Page 330: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Irisa/Evita

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 25-1: Bedside monitor, large text screen

Figure 25-2: Bedside monitor, small text

Figure 25-3: Central monitor, split screen

� Measured peak inspiratory pressure

� Fractional-inspired O2 value

� � � � �

30 PIP 20 FiO2 4.1 Vmin VENT

0.40 Vt

cmH2O % L L

1 PEEP 29 PLT 9 MAP 10 RR

cmH2O cmH2O cmH2O BPM

� � �

� �

VENT

VT = 0.40 LRR = 10 BPM �PIP = 30 cmH2O �PEEP = 1 cmH2O

� �

� � �

� RR = 10BPM FiO2 = 20%VENT

MAP = 9cmH2O PIP = 30cmH2O

� VT = 0.40L PLT = 29cmH2O

� VMIN = 4.1L PEEP = 1 cmH2O

� ICU 14

� �

Flexport System Interface Operations Manual 25-4

Page 331: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Irisa/Evita

� Measured minute volume

� Measured positive end expiratory pressure

� Measured inspiratory plateau pressure

� Mean airway pressure

� VENT parameter key

Measured tidal volume

Measured respiratory rate

� Alarm status bell (appears when alarms are turned ON)

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms still cause the VENT key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM ON/OFF.• Select an alarm parameter.• Select the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 25-5

Page 332: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Irisa/Evita

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 25-4). If alarms for all ventilator parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 25-4: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 25-4).

The default setting is LARGE TXT.

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To select a display format:

• Touch VENT.• Touch SETUP.• Touch LARGE TXT or SMALL TXT.

VENT

VT = 0.40 LVMIN FIO2 PIP RR = 10 BPM

PIP = 30 cmH2O

PEEP = 1 cmH2O

ICU 14

Flexport System Interface Operations Manual 25-6

Page 333: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Irisa/Evita

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport system interface operation.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

SERVICE NEEDED Contact Dräger Irisa service representative. No

COMPRESSES AIR LOW Compressed air supply less than 36 psi. No

O2 PRESSURE LOW Oxygen supply less than 36 psi. No

PRESSURE MEAS INOP Pressure sensor defective. No

O2 MEAS INOP O2 sensor defective. No

FLOW MEAS INOP Flow sensor defective. No

MIXER INOP HPSV defective. No

EXP VALVE INOP Flow in expiratory line during inspiratory phase. No

Flexport System Interface Operations Manual 25-7

Page 334: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Irisa/Evita

FAN DEFECT Contact Dräger Irisa service representative. No

TEMPERATURE MEAS INOP Temperature sensor defective. No

TEMPERATURE SENSOR? Temperature sensor was disconnected while machine was turned ON. No

TEMPERATURE HIGH Inspiratory gas temperature less than 40 ºC. No

APNEA OR FAIL TO CYCLE

Leak in the breathing system.Loss of volume due to pressure or time limitation.Spirolog is not calibrated.Machine malfunction; call biomed.

Yes

HIGH RATE Patient’s respiratory rate exceeds the respiratory rate limit. Yes

SYNCHRONIZATION INOP Contact Dräger Irisa service representative. Yes

PRESSURE LIMITED Set pressure control limit exceeded. Yes

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 25-8

Page 335: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Directory of KeysVENT

VENT - MAIN MENUALARM LIMITS SETUP

FREEZE * VIEW STATUS

VIEW ALARMS PRINT

ON OFF

VENT - SETUP *Narkomed 2C only.

SIZE* SWEEP

SPEED*

WAVEFORM* LARGE TXT

SMALL TXT

VENT - SETUP - WAVEFORM TYPE

FLOW PRESSURE

VENT - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec

VENT - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

VENT - ALARM LIMITS

FiO2 PIP Vmin

VENT - ALARM LIMITS - VminVmin HI =

XXXLO = XXX

FiO2ON OFF Vmin

VENT - ALARM LIMITS - PIPPIP HI =

XXXLO = XXX

PIPON OFF Vmin

VENT - ALARM LIMITS - FiO2FiO2 HI =

XXXLO = XXX

FiO2ON OFF PIP

Flexport System Interface Operations Manual 26-1

Page 336: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 337: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display (Narkomed 2C Only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90436A-11 Flexport system interface enables information from the Dräger Narkomed 2B and 2C ventilators to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. In the ventilator’s Serial Port Configuration menu, choose VITALINK as the Protocol selection. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Ventilator Baud Rate

Character Length

Parity Setting Stop Bit

Minimum Vitalink Protocol

Narkomed 2C 9,600 8 bit Even 1 1.20

Narkomed 2B 1,200 8 bit Even 1 1.15

Flexport System Interface Operations Manual 26-3

Page 338: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 26-1: Bedside monitor, large text screen (Narkomed 2C)

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector (P/N 131-1039-00).

• Plug the transition connector into the RS-232 connector on the ventilator.

� � � � � �

7MAWP 19 PIP 27 FiO2 V

ENT

0.25Vtexp

WAVE=FLOW

cmH2O cmH2O % L

3 PEEP 1.3 Vmin 36 RR

cmH2O L BPM

� �

Flexport System Interface Operations Manual 26-4

Page 339: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Figure 26-2: Bedside monitor, small text screen (Narkomed 2C)

Figure 26-3: Central monitor, split screen (Narkomed 2C)

� Mean airway pressure value

� Peak inspiratory pressure value

� Fraction inspired O2 value

� VENT parameter key

� Expired tidal volume value

� Current waveform mode

� Alarm status bell (appears when alarms are turned ON)

Respiratory rate value

Minute volume value

� Positive end expiratory pressure

� Patient/room ID

� � � �

VENT

FiO2= 27% WAVE=FLOWPIP= 19cmH2O

Vmin= 1.3LRR= 36BPM

� � � � �

RR=36BPM PIP=19cmH2OVENT

� FiO2=27% PEEP= 3cmH2O

Vmin=1.3L MAWP=7cmH2O

� Vtexp=0.25L

� ICU 14

Flexport System Interface Operations Manual 26-5

Page 340: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Touch ALARMS LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 26-6

Page 341: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Viewing Alarm SettingsThe VIEW ALARM key enables you to display the current alarm settings for all parameters (refer to Figure 26-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 26-4: View Alarms screen (Narkomed 2C)

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 26-10).

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

PIP 50 FiO2 70 VENT

1.62 Vtexp WAVE=FLOW30 L

Vmin 90 RR

0.1 BPM

Flexport System Interface Operations Manual 26-7

Page 342: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFF (Narkomed 2C Only)You can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform Size (Narkomed 2C Only)If the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch VENT.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch VENT.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 26-8

Page 343: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Selecting a Sweep Speed (Narkomed 2C Only)The sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 25, 12.5, 6.25, and 3.12 mm/second.

The default setting is 12.5 mm/second.

Selecting a Waveform Type (Narkomed 2C Only)You can select either FLOW or PRESSURE waveforms for display.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 26-4).

To select a sweep speed:

• Touch VENT.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a waveform type:

• Touch VENT.• Touch SETUP.• Touch WAVEFORM.• Select FLOW or PRESSURE waveform display.

To select a display format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 26-9

Page 344: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 2B and 2C

Freezing the Waveform on the Display (Narkomed 2C Only)The FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process and update the numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording on the Narkomed 2C is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

The Flexport system interface reports approximately 250 status messages for the ventilator during Flexport system interface operation, depending on the configuration of the ventilator and components on the Dräger Vitalink bus. Refer to your ventilator manual for the cause of the status messages and appropriate action.

To freeze the waveform on the display:

• Touch VENT.• Touch FREEZE ON.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 26-10

Page 345: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Directory of KeysANES

ANES - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 27-2

ANES - SETUP

SIZE SWEEP SPEED WAVEFORM

LARGE TXTSMALL TXT

ANES - SETUP - WAVEFORM TYPE

CO2 FLOW PRESSURE

ANES - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec0.39

mm/sec

ANES - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

Flexport System Interface Operations Manual 27-1

Page 346: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

ANES

ANES - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 27-1

ANES - ALARM LIMITS

ETCO2 SpO2 SYS I AGT FiO2 PIP Vmin

ANES - ALARM LIMITS - VminVmin HI =

XX.XLO = XX.X

ETCO2ON OFF Vmin

ANES - ALARM LIMITS - PIPPIP HI =

XXXLO = XXX

PIPON OFF Vmin

ANES - ALARM LIMITS - FiO2FiO2 HI =

XXXLO = XXX

FiO2ON OFF PIP

ANES - ALARM LIMITS - I AGTI AGT HI =

X.XXLO = X.XX

I AGTON OFF FiO2

ANES - ALARM LIMITS - SYSSYS HI =

XXXLO = XXX

SYSON OFF I AGT

ANES - ALARM LIMITS - SpO2SPO2 HI =

XXXLO = XXX

SpO2ON OFF SYS

ANES - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XX

ETCO2ON OFF SPO2

Flexport System Interface Operations Manual 27-2

Page 347: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90443A-02 Flexport system interface enables information from a Dräger Narkomed 3, 4, or GS anesthesia machine to be displayed on Spacelabs Healthcare monitors.

Configure the anesthesia machine as shown in Table 1. Consult the appropriate anesthesia machine manual for additional configuration setup procedures.

Table 1: Anesthesia Machine Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Vitalink Protocol

9600 8-bit none 1131-1039-xx

(9-pin)1.15

Flexport System Interface Operations Manual 27-3

Page 348: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the anesthesia machine, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the anesthesia machine. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the anesthesia machine.

Figure 27-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the cable into the transition connector.• Plug the transition connector into the RS-232 connector on the anesthesia machine.

� � � � � �

96 SPO2 27 PIP 0.90 I AGT ANES

162SYS WAVE=

CO2

% cmH2O % mmHg

35 ETCO2 12.7 Vmin 59 FiO2 90 DIA

mmHg L % mmHg

� � �

Flexport System Interface Operations Manual 27-4

Page 349: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Figure 27-2: Bedside monitor, small text

Figure 27-3: Central monitor, split screen

� Pulse oximetry value

� Measured inspiratory plateau pressure

� Inspired agent

� ANES parameter key

� Systolic blood pressure

� Current waveform type (Narkomed 4 only)

� Alarm status bell (appears when alarms are turned ON)

Diastolic blood pressure

Measured inspired oxygen level

� Minute volume

� Current end tidal CO2 (mmHg) value

Patient/room ID

� � �

ANES

SYS = 162 mmHg WAVE=FLOWDIA = 90 mmHg

SPO2 = 96 % �RR= 36BPMETCO2 = 35 mmHg

� �

� �

� ETCO2= 35 mmHg SYS= 162 mmHg ANES

� SPO2= 96 % DIA= 90 mmHg

FiO2= 59 % I AGT= 0.90 %

� Vmin= 12.7 L PIP= 27 cmH2O

ICU 14

� � �

Flexport System Interface Operations Manual 27-5

Page 350: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the ANES key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the ANES key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the anesthesia

machine.

• Alarm limits can only be adjusted at the anesthesia machine. Alarm violation is detected at the anesthesia machine, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the anesthesia machine can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch ANES.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 27-6

Page 351: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the anesthesia machine’s parameters. If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays ANES ALM OFF to the right of the ANES key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 27-4: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the anesthesia machine. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 27-10).

To view alarm settings:

• Touch ANES.• Touch VIEW ALARMS.

To view the anesthesia machine status:

• Touch ANES.• Touch VIEW STATUS.

I AGT4.09

ETCO245

SYS180 A

NES

162 SYS WAVE=CO2

0.00 25 85 mmHg

SPO2 100 PIP 50 Vmin FiO2 90 DIA

95 2.7 mmHg

Flexport System Interface Operations Manual 27-7

Page 352: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 0.39, 0.78, 1.56, 3.12, 6.25, 12.5, and 25 mm/second.

The default setting is 12.5 mm/second.

To adjust waveform size:

• Touch ANES.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch ANES.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

Flexport System Interface Operations Manual 27-8

Page 353: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Selecting a Waveform Type (Narkomed 4 Only)You can select CO2, FLOW, or PRESSURE waveforms for display.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 27-4).

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process and update the numeric information.

The default setting is FREEZE OFF.

To select a waveform type for Narkomed 4 only:

• Touch ANES.• Touch SETUP.• Touch WAVEFORM.• Select CO2, FLOW, or PRESSURE waveform display.

To select a display format:

• Touch ANES.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform display:

• Touch ANES.• Select FREEZE ON.

Flexport System Interface Operations Manual 27-9

Page 354: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Dräger Narkomed 3, 4, and GS

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the anesthesia machine or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the anesthesia machine is powered ON.

• Verify that all cables are secure.

To print the anesthesia machine display:

• Touch ANES.• Touch PRINT.

Flexport System Interface Operations Manual 27-10

Page 355: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 28-1

Edwards Vigilance I

Directory of Keys

CO

CO - MAIN MENUALARM LIMITS SETUP SEND

CCOVIEW

ALARMS PRINT

CO - SETUPLARGE TXT

SMALL TXT

CO - ALARM LIMITS

SvO2 CCO CCI

CO - ALARM LIMITS - CCICCI HI =

XXLO=XXON OFF

CO - ALARM LIMITS - CCOCCO HI =

XXLO=XXON OFF

CO - ALARM LIMITS - SvO2SvO2 HI =

XXLO=XXON OFF

These keys appear on the bedside monitor only.

Page 356: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 357: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance I

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Storing a CCO Value as a Bolus CO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a Baxter-Edwards Critical Care Division Vigilance I continuous cardiac output (CCO) monitor to be displayed on Spacelabs Healthcare monitors.

Configure the CCO monitor as shown in Table 1. Complete the steps in the following Quickstart to select the communication port and the Flexport system interface from the CCO monitor. Consult the manufacturer’s manual for additional configuration setup procedures.

To select the Flexport system interface from the CCO monitor:

• Power ON the CCO monitor.• Touch SETUP.• Touch SYSTEM CONFIG.• Touch DIGITAL PORTS.• Cursor to the device line and select COM1 or COM2.• Touch CHANGE.• Select the Flexport system interface.

Flexport System Interface Operations Manual 28-3

Page 358: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance I

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the CCO monitor, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the CCO monitor. To control other interface functions, refer to the sections that follow.

Table 1: CCO Monitor Configuration

Monitor Name and Model Number

Baud Rate

Minimum Software Revision

Installation Kit

Part Number

Transition Connector

Part Number

Baxter Edwards Critical Care Division Vigilance I Monitor VGS (100/115V) 9600 04.42 045-0033-xx 131-1039-xx

Baxter Edwards Critical Care Division Vigilance I Monitor VGS1 (220V) 9600 04.42 045-0034-xx 131-0686-xx*

* P/N 131-0686-xx requires a gender changer.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 9-pin connector (COM1 or COM2) on the

CCO monitor.

Flexport System Interface Operations Manual 28-4

Page 359: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance I

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the CCO monitor.

Figure 28-1: Bedside monitor, large-text screen

Figure 28-2: Bedside monitor, small-text screen

Figure 28-3: Central monitor, split screen

� Continuous cardiac output value

� Continuous systemic vascular resistance

� Venous O2 saturation

� CO parameter key

� � � � �

4.8 CCO 1200 SVR 80 SvO2

CO

4.2 CO CALC

%

2.4 2400 37.0 2.1 CI CALCCCI SVRI BT

°C

� � �

� �

CO

SvO2= 80%CCO= 4.8 �CC I = 2.4BT=37.0°C

� �

� �

� CO CALC= 4.2 SVR =1200

CO

� CC I CALC= 2.1 SVRI =2400

� CCO= 4.8 SvO2= 80% �� CC I = 2.4 BT=37.0 °C

� ICU 14 �

Flexport System Interface Operations Manual 28-5

Page 360: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance I

� Cardiac output CALC value

� Alarm status bell (appears when alarms are turned ON)

� Cardiac index CALC value

Blood temperature value

Continuous systemic vascular resistance index

� Continuous cardiac index value

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the CO key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the CO key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the CCO

monitor.

• Alarm limits can only be adjusted at the CCO monitor. Alarm violation is detected at the CCO monitor, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the CCO monitor can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch CO.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 28-6

Page 361: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance I

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the CCO parameters (refer to Figure 28-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays CO ALM OFF to the right of the CO key. If alarms are turned ON for any parameter, a bell displays. The alarm bell flashes when an alarm limit is violated.

Figure 28-4: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 28-5).

The default setting is LARGE TXT.

To view alarm settings:

• Touch CO.• Touch VIEW ALARMS.

To select a display format:

• Touch CO.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

SvO29910 CCI

7.01.0

CO

SvO2= 80%

CCO20.01.0 CCO = 4.8

CC I = 2.4BT=37.0°C

Flexport System Interface Operations Manual 28-7

Page 362: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance I

Storing a CCO Value as a Bolus COEach time the SEND CCO key is touched, the current CCO value is displayed as CO CALC. This CO CALC value is stored in TRENDS and is available for use in the Spacelabs Healthcare monitor’s CALCS package.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To store a CO CALC value:

• Touch CO.• Touch SEND CCO.

To print the CCO monitor display:

• Touch CO.• Touch PRINT.

Flexport System Interface Operations Manual 28-8

Page 363: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance I

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the CCO monitor or the cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the CCO monitor is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the CCO monitor during Flexport operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the Spacelabs Healthcare monitor and the CCO monitor has been interrupted. Yes

SvO2 PROBLEM — CHECK VIGILANCE

SvO2 catheter is not attached, not calibrated, or is defective. No

CCO PROBLEM — CHECK VIGILANCE

CCO catheter is not attached, is defective, or is not providing data. No

Flexport System Interface Operations Manual 28-9

Page 364: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 365: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 29-1

Edwards Vigilance II

Directory of Keys

CO

CO - MAIN MENUALARM LIMITS SETUP SEND

CCOVIEW

ALARMS PRINT

CO - SETUPLARGE TXT

SMALL TXT

CO - ALARM LIMITS

CCO CCI SVR CEDVI SvO2 RVEF

CO - ALARM LIMITS - RVEFRVEF HI =

XXLO=XXON OFF

CO - ALARM LIMITS - SvO2SvO2 HI =

XXLO=XXON OFF

CO - ALARM LIMITS - CEDVICEDVI HI =

XXLO=XXON OFF

CO - ALARM LIMITS - SVRSVR HI =

XXLO=XXON OFF

CO - ALARM LIMITS - CCICCI HI =

XXLO=XXON OFF

CO - ALARM LIMITS - CCOCCO HI =

XXLO=XXON OFF

These keys appear on the bedside monitor only.

Page 366: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 367: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance II

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Storing a CCO Value as a Bolus CO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A-03 Flexport system interface enables information from an Edwards Critical Care Division Vigilance II continuous cardiac output (CCO) monitor to be displayed on Spacelabs Healthcare monitors.

Configure the CCO monitor as shown in Table 1. Complete the steps in the following Quickstart to select the communication port and the Flexport system interface from the CCO monitor. Consult the manufacturer’s manual for additional configuration setup procedures.

To select the Flexport system interface from the CCO monitor:

• Power ON the CCO monitor.• Select SETUP.• Select SERIAL PORT SETUP.• Select DEVICE for either COM1 or COM2.• Select ‘Flexport’.

Table 1: CCO Monitor Configuration

Monitor Name and Model Number

Baud Rate

Minimum Software Revision

Installation Kit

Part Number

Transition Connector

Part Number

Edwards Critical Care Division Vigilance II Monitor VGS (100/115V)

9600 00.56 045-0161-xx 131-0686-xx*

* P/N 131-0686-xx requires a gender changer.

Flexport System Interface Operations Manual 29-3

Page 368: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance II

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the CCO monitor, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the CCO monitor. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the CCO monitor.

Figure 29-1: Bedside monitor, large-text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 9-pin connector (COM1 or COM2) on the

CCO monitor.

� � � � �

3.9 CCO 1185 SVR 61 SvO2

CO

4.1 CO CALC

%

1.6 37 56 1.7 CI CALCCCI CEDVI RVEF

%

� � �

Flexport System Interface Operations Manual 29-4

Page 369: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance II

Figure 29-2: Bedside monitor, small-text screen

Figure 29-3: Central monitor, split screen

� Continuous cardiac output value

� Continuous systemic vascular resistance

� Venous O2 saturation

� CO parameter key

� Cardiac output CALC value

� Alarm status bell (appears when alarms are turned ON)

� Cardiac index CALC value

Right ventricular ejection fraction

Continuous end diastolic volume indexed

� Continuous cardiac index value

� Patient/room ID

� �

CO

SvO2 = 61%CCO = 3.9 �CCI = 1.6RVEF = 56%

� �

� �

� CO CALC= 4.1 SVR = 1185

CO

� CI CALC= 1.7 CEDVI = 37

� CCO= 3.9 SvO2 = 61% �� CCI = 1.6 RVEF = 56%

� ICU 14 �

Flexport System Interface Operations Manual 29-5

Page 370: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance II

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the CO key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the CO key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the CCO

monitor.

• Alarm limits can only be adjusted at the CCO monitor. Alarm violation is detected at the CCO monitor, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the CCO monitor can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch CO.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 29-6

Page 371: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance II

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the CCO parameters (refer to Figure 29-5). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays CO ALM OFF to the right of the CO key. If alarms are turned ON for any parameter, a bell displays. The alarm bell flashes when an alarm limit is violated.

Figure 29-4: View Alarms screen, large text

Figure 29-5: View Alarms screen, small text

To view alarm settings:

• Touch CO.• Touch VIEW ALARMS.

CCO14.03.0 SVR

1500500 SvO2

9910

CO

4.1 CO CALC

CCI CEDVI RVEF 1.7 CI CALC7.01.0

15060

500

CCO14.0

3.0 SVR1500500 SvO2

9910

CO

SvO2 = 61%CCO = 3.9CCI = 1.6

CCI CEDVI RVEFRVEF = 56%

7.01.0

15060

500

Flexport System Interface Operations Manual 29-7

Page 372: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance II

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 29-4).

The default setting is LARGE TXT.

Storing a CCO Value as a Bolus COEach time the SEND CCO key is touched, the current CCO value is displayed as CO CALC. This CO CALC value is stored in TRENDS and is available for use in the Spacelabs Healthcare monitor’s CALCS package.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch CO.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To store a CO CALC value:

• Touch CO.• Touch SEND CCO.

To print the CCO monitor display:

• Touch CO.• Touch PRINT.

Flexport System Interface Operations Manual 29-8

Page 373: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigilance II

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the CCO monitor or the cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the CCO monitor is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the CCO monitor during Flexport operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the Spacelabs Healthcare monitor and the CCO monitor has been interrupted. Yes

SVO2 PROBLEM — CHECK VIGILANCE

SvO2 catheter is not attached, not calibrated, or is defective. No

CCO PROBLEM — CHECK VIGILANCE II

CCO catheter is not attached, is defective, or is not providing data. No

Flexport System Interface Operations Manual 29-9

Page 374: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 375: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 30-1

Edwards Vigileo

Directory of Keys

CO

CO - MENUALARM LIMITS SETUP VIEW

ALARMS PRINT

CO - SETUPLARGE TXT

SMALL TXT

CO - ALARM LIMITS

CO CI ScvO2 -OR- SvO2

CO - ALARM LIMITS - SvO2SvO2 HI =

XXLO=XXON OFF

CO - ALARM LIMITS - ScvO2ScvO2 HI =

XXLO=XXON OFF

CO - ALARM LIMITS - CICI HI =

XXLO=XXON OFF

CO - ALARM LIMITS - COCO HI =

XXLO=XXON OFF

Page 376: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 377: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigileo

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewNote:

The Edwards Vigileo monitor does not support blood temperature TB(CO). This value is obtained from the Vigileo monitor in trends as a zero value, even though it is not supported.

The 90442A-03 Flexport system interface enables information from a Edwards Critical Care Division Vigileo continuous cardiac output (CCO) monitor to be displayed on Spacelabs Healthcare monitors.

Configure the CCO monitor as shown in Table 1. Complete the steps in the following Quickstart to select the communication port and the Flexport system interface from the CCO monitor. Consult the manufacturer’s manual for additional configuration setup procedures.

To select the Flexport system interface from the CCO monitor:

• Power ON the CCO monitor.• Select the rectangular box at the bottom of the screen.• Select Serial Port Setup.• Change the device to Flexport.

Table 1: CCO Monitor Configuration

Monitor Name and Model Number

Baud Rate

Minimum Software Revision

Installation Kit

Part Number

Transition Connector

Part Number

Edwards Critical Care Division Vigileo Monitor VGS (100/115V) 9600 04.42 045-0034-xx 131-0686-xx

Flexport System Interface Operations Manual 30-3

Page 378: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigileo

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the CCO monitor, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the CCO monitor. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the CCO monitor.

Figure 30-1: Bedside monitor, large-text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 9-pin connector on the CCO monitor.

� � � � �

2.9 CI 88 SVml/b 1200 SVR

CO

4.2 CO

2 SVV 39 SVI 1943 SVRI 61 ScvO2

% %

� � �

Flexport System Interface Operations Manual 30-4

Page 379: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigileo

Figure 30-2: Bedside monitor, small-text screen

Figure 30-3: Central monitor, split screen

� Cardiac index

� Stroke volume

� Systemic vascular resistance

� CO parameter key

� Cardiac output value

� Alarm status bell (appears when alarms are turned ON)

� Central venous oxygen saturation (ScvO2) value

Systemic vascular resistance index

Stroke volume index

� Stroke volume variation

� Patient/room ID

� �

CO

CO = 7.1CI = 3.0 �ScvO2 = 62% �SV = 92 ml/b

� �

� CO= 7.2 SV = 95 ml/b

CO

� ScvO2= 61% SVI = 40

� CI= 3.0 SVR = 1200 �� SVV = 2% SVRI = 1896

� ICU 14 �

Flexport System Interface Operations Manual 30-5

Page 380: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigileo

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the CO key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the CO key to flash. Alarm settings default to ON.

Note:• Alarms that are generated on the CCO monitor will be displayed on the Spacelabs monitor within

three seconds.

• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the CCO monitor.

• Alarm limits can only be adjusted at the CCO monitor. An alarm violation is detected at the CCO monitor, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the CCO monitor can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch CO.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 30-6

Page 381: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigileo

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the CCO parameters (refer to Figure 30-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays CO ALM OFF to the right of the CO key. If alarms are turned ON for any parameter, a bell displays. The alarm bell flashes when an alarm limit is violated.

Figure 30-4: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 30-4). The default setting is LARGE TXT.

To view alarm settings:

• Touch CO.• Touch VIEW ALARMS.

To select a display format:

• Touch CO.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

CO 142.0 ScvO2 99

10

CO

6.6 CO

CI 7.01.0

62 ScvO2

%

Flexport System Interface Operations Manual 30-7

Page 382: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Edwards Vigileo

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the CCO monitor or the cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the CCO monitor is powered ON.

• Verify that all cables are secure.

To print the CCO monitor display:

• Touch CO.• Touch PRINT.

Flexport System Interface Operations Manual 30-8

Page 383: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Directory of KeysANES

ANES - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW STATUS

VIEW ALARMS PRINT

ON OFF

ANES - SETUP

SIZE SWEEP SPEED WAVEFORM

LARGE TXTSMALL TXT

ANES - SETUP - WAVEFORM

CO2 FLOW PRESSURE

ANES - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec

ANES - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

ANES - ALARM LIMITS

FiO2 PIP FiENF EtCO2

ANES - ALARM LIMITS - EtCO2EtCO2 HI =

X.XXLO = X.XX

EtCO2ON OFF FiO2

ANES - ALARM LIMITS - FiENFFiENF HI =

XXXLO = XXX

FiENFON OFF EtCO2

ANES - ALARM LIMITS - PIPPIP HI =

XXXLO = XXX

PIPON OFF FiENF

ANES - ALARM LIMITS - FiO2FiO2 HI =

XXLO = XX

FiO2ON OFF PIP

Flexport System Interface Operations Manual 31-1

Page 384: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 385: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90443A-06 Flexport system interface enables information from an Engström EAS 9000 anesthesia machine to be displayed on Spacelabs Healthcare monitors.

Configure the anesthesia machine as shown in Table 1. Consult the anesthesia machine manual for additional configuration setup procedures.

Table 1: Anesthesia Machine Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Modular cable

Interface Cable

Part Number

9600 8-bit even 1 25-pin 012-0512-xx

Flexport System Interface Operations Manual 31-3

Page 386: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the anesthesia machine, complete the steps listed in the following Quickstart.

Figure 31-1: Flexport system interface connections

The Spacelabs Healthcare monitor is now ready to display information from the anesthesia machine. To control other interface functions, refer to the sections that follow.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 31-1.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the RS-232 connector on the anesthesia machine labeled COMPUTER.

• Plug the 25-pin female connector into the ANALOG RECORDER connector on the side of the anesthesia machine.

SDLCcable

SDLCterminator

interfacecable

anesthesia machine

SpacelabsHealthcare

monitor

systemFlexport

interface

Flexport System Interface Operations Manual 31-4

Page 387: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the anesthesia machine.

Figure 31-2: Bedside monitor, large text screen

Figure 31-3: Bedside monitor, small text

Figure 31-4: Central monitor, split screen

� Fractional inspired agent value can be FiHAL, FiENF, FiISO, FiDES, FiSEV, or FiAGT (if the agent is halothane, enflurane, isoflurane, desflurane, sevoflurane, or no selected agent, respectively)

� Peak inspiratory pressure value

� Respiration rate

� ANES parameter key

� Expired tidal volume data

� � � � � �

0.89 FiENF 58 PIP 11 RR ANES

0.49 Vt [CO2]% cmH2O BPM L

5.9 FiN2O 9.0 PEEP 4.24 EtCO2 30 FiO2

% cmH2O % %

� � �

� � �

ANES

FiENF=0.89% [CO2]FiO2= 30% Vt= 0.49L �EtCO2=4.24%

� � � � � �

� FiENF=0.89% Vt = 0.49L ANES

FiO2=30% RR= 11BPM

EtCO2=4.24% FiN2O= 5.9%

� PIP=58cmH2O PEEP= 9.0cmH2O

Flexport System Interface Operations Manual 31-5

Page 388: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

� Waveform identification label

� Alarm status bell (appears when alarms are turned ON)

Fractional inspired O2 value

End tidal CO2 value; units may be % or mmHg (selected at Engström unit)

� Positive end expiration pressure setting

� Fractional inspired N2O value

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the ANES key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the ANES key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the anesthesia

machine.

• Alarm limits can only be adjusted at the anesthesia machine. Alarm violation is detected at the anesthesia machine and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the anesthesia machine can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch ANES.• Touch ALARMS LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 31-6

Page 389: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the anesthesia machine’s parameters (refer to Figure 31-5). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays ANES ALM OFF to the right of the ANES key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 31-5: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be displayed on the bedside monitor, and three can be displayed on the remote monitor (refer to Status Messages on page 31-11).

To view alarm settings:

• Touch ANES.• Touch VIEW ALARMS.

To view anesthesia status:

• Touch ANES.• Touch VIEW STATUS.

FiENF 2.00 PIP 40 ANES

FiENF=0.89%

0.50 4 FiO2= 30%

Vt= 0.49L

FiO2 45 EtCO2 7.50 EtCO2=4.24%18 2.00

Flexport System Interface Operations Manual 31-7

Page 390: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveforms ON or OFF:

• Touch ANES.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch ANES.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 31-8

Page 391: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, 12.5, and 25 mm/second.

The default sweep speed is 25 mm/second.

Selecting a Waveform TypeYou can select either CO2, FLOW, or PRESSURE waveforms for display.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 31-5).

The default setting is LARGE TXT.

To select a sweep speed:

• Touch ANES.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a waveform type:

• Touch ANES.• Touch SETUP.• Touch WAVEFORM.• Select the desired waveform.

To select a display format:

• Touch ANES.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 31-9

Page 392: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Freezing the Waveform on the DisplayThe FREEZE ON key allows you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To freeze the waveform on the display:

• Touch ANES.• Touch FREEZE ON.

To print the anesthesia machine display:

• Touch ANES.• Touch PRINT.

Flexport System Interface Operations Manual 31-10

Page 393: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the anesthesia machine or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the anesthesia machine is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the anesthesia machine during Flexport system interface operation.

Table 2: Status Messages

Status Message Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the ventilator and the monitor has been interrupted. Yes

FLOW SENSOR BAD Fresh gas sensor is faulty. No

VAPORIZER FAULT Vaporizer output is zero. No

ALARM SYS OFF No data comm between ventilator and volume monitor. No

CHECK O2 CONC O2 sensor not calibrated. No

SENSOR FAULT # (1-10) Repair appropriate sensor. No

VENT FAULT # (1-10) Repair ventilator. No

VOL MON FAULT # (1-10) Repair volume monitor. No

CALIB SYSTEM # (1-10) Calibrate the appropriate system. No

MINUTE VOL LOW Alarm is active if expiratory minute volume is low. Yes

MINUTE VOL HIGH Alarm is active if expiratory minute volume is high. Yes

N2O SUPPLY OFF N2O supply pressure is under limit. No

O2 SUPPLY OFF O2 supply pressure is under limit. No

HYPOXIC MIXTURE O2 concentration too low. No

N2O OFF Fresh gas N2O cut off due to hypoxic mixture. No

VAPORIZER HIGH Vaporizer setting is above capacity. No

CO2 NOT DETECT No CO2 is detected in sampling tube. No

PATIENT BAG EMP Preset tidal volume is not being delivered. No

N2O RESTORED Settings have been corrected — N2O restored. No

Flexport System Interface Operations Manual 31-11

Page 394: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Engström EAS 9000

ANES LIQUID LOW Anesthestic bottle nearly empty. No

BATTERY POWERED Main wall power is OFF. No

BACKUP BATT LOW Battery not fully charged. No

Table 2: Status Messages (continued)

Status Message Indication Alarm Sounds

Flexport System Interface Operations Manual 31-12

Page 395: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 32-1

GE Ohmeda 5200/5210

Directory of KeysCO2

CO2 - MAIN MENUALARM LIMITS SETUP VIEW

ALARMS PRINT

CO2 - SETUP LARGE TXT PERCENTSMALL TXT mmHg

CO2 - ALARM LIMITS

APNEA RR ETCO2 N2O MINCO2

CO2 - ALARM LIMITS - MINCO2MINCO2 HI =

XX ↑ ↓ETCO2

ON OFF MINCO2

CO2 - ALARM LIMITS - N2ON2O HI =

XXLO = XX ↑ ↓ON OFF

CO2 - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XX ↑ ↓

ETCO2ON OFF MINCO2

CO2 - ALARM LIMITS - RRRR HI =

XXLO = XX ↑ ↓

APNEAON OFF RR

CO2 - ALARM LIMITS - APNEAAPNEA APNEA

ON OFF RR

Page 396: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 397: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5200/5210

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90433A-02 Flexport system interface enables information from a GE Ohmeda 5200 or 5210 analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the analyzer as shown in Table 1. Consult the analyzer manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

Table 1: Analyzer Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

1200 7-bit odd 1 131-0685-xx (25-pin) 6.0

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the analyzer.

Flexport System Interface Operations Manual 32-3

Page 398: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5200/5210

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the analyzer.

Figure 32-1: Bedside monitor, large text screen

Figure 32-2: Bedside monitor, small text screen

Figure 32-3: Central monitor, split screen

� Current N2O value (present only when reading exceeds 5% or N2O alarm monitoring is enabled)

� Current minimum CO2 value

� CO2 parameter key

� Current EtCO2 value

� Alarm status bell (appears when alarms are turned ON)

� � � �

2 N2O 0.7MINCO2 CO2

5.6 ETCO2

% % %

12 RR

BPM

� �

� � �

CO2

ETCO2 = 5.6%

RR = 12 BPM

MINCO2= 0.7%

N2O = 2%

� � �

� � � �

� ETCO2 = 5.6% N2O = 2%CO2

� RR = 12BPM APNEA = 20s

� MINCO2 = 0.7%

ICU 14

Flexport System Interface Operations Manual 32-4

Page 399: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5200/5210

� Current respiration rate value

� Apnea monitoring status

Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the CO2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the CO2 key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

• The Apnea alarm limit on the analyzer is fixed at 20 seconds.

• The EtCO2 and minimum CO2 alarm limit ranges depend on the selected unit of measurement.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch CO2.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To adjust alarm limits:

• Touch CO2.• Select ALARM LIMITS.• Select an alarm parameter.• Select HI= or LO=.• Use the arrow keys to adjust.

Flexport System Interface Operations Manual 32-5

Page 400: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5200/5210

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 32-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays CO2 ALM OFF to the right of the CO2 key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 32-4: View Alarms screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 32-4).

To view alarm settings:

• Touch CO2.• Touch VIEW ALARMS.

To select a display format:

• Touch CO2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

N2O 80 MINCO2 0.7 ETCO2 CO2

5.6 ETCO2

0 %

APNEA 20 RR 12 RR

BPM

Flexport System Interface Operations Manual 32-6

Page 401: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5200/5210

Selecting a Unit of MeasurementExpired CO2 can be displayed in percent (%) or millimeters of mercury (mmHg).

The default setting is PERCENT.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a measurement unit:

• Touch CO2.• Touch SETUP.• Select PERCENT or mmHg.

To print the current display:

• Touch CO2.• Touch PRINT.

Flexport System Interface Operations Manual 32-7

Page 402: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5200/5210

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the analyzer and the monitor has been interrupted. Yes

OCCLUSION The sample line is blocked — check the sample line for blocking or crimps. No

SERVICE REQUIRED The sensor needs to be calibrated. No

SYSTEM PURGING... The monitor is purging moisture from the sample tube. No

Flexport System Interface Operations Manual 32-8

Page 403: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 33-1

GE Ohmeda 5250

Directory of KeysRGM

RGM - MAIN MENUALARM LIMITS SETUP VIEW

STATUSVIEW

ALARMS PRINT

RGM - SETUP MENULARGE TXT PERCENTSMALL TXT mmHg

RGM - ALARM LIMITS

FiAX FiN2O EtCO2 SpO2 FiO2 HR Vt PIP APNEA

RGM - ALARM LIMITS - APNEAAPNEA HI =

XXAPNEA

ON OFF FiAX

RGM - ALARM LIMITS - PIPPIP HI =

XX.XLO =XX.X

PIPON OFF APNEA

RGM - ALARM LIMITS - VtVt HI =

XX.XLO =XX.X

VtON OFF PIP

RGM - ALARM LIMITS - HRHR HI =

XXXLO =XXX

HRON OFF Vt

RGM - ALARM LIMITS - FiO2FiO2 HI =

XXXLO =XXX

FiO2ON OFF HR

RGM - ALARM LIMITS - SpO2SpO2 HI =

X.XXLO =X.XX

SpO2ON OFF FiO2

RGM - ALARM LIMITS - EtCO2EtCO2 HI =

XXXLO =XXX

EtCO2ON OFF SpO2

RGM - ALARM LIMITS - FiN2OFiN2O HI =

XXXLO =XXX

FiN2OON OFF EtCO2

RGM - ALARM LIMITS - FiAXFiAX HI =

XXLO =XX

FiAXON OFF FiN2O

Page 404: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 405: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5250

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90443A-03 Flexport system interface enables information from a GE Ohmeda 5250 Respiratory Gas Monitor to be displayed on Spacelabs Healthcare monitors.

Configure the gas monitor as shown in Table 1. Consult the gas monitor manual for additional configuration setup procedures.

Table 1: Gas Monitor Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

1200 7 bit odd 1131-1059-xx

(9-pin)

Flexport System Interface Operations Manual 33-3

Page 406: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5250

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the gas monitor, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the gas monitor. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the gas monitor.

Figure 33-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the gas monitor.

� � � � � �

7.0 FiAX

% 3.0 FiCO2

% 21 FiO2

% 77 HR

bpm RGM

3.33 VT

L

SpO2 EtCO2 FIN2O RR PIP

98 2.5 50 20 10% % % BPM cmH2O

� � �

Flexport System Interface Operations Manual 33-4

Page 407: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5250

Figure 33-2: Bedside monitor, small text

Figure 33-3: Central monitor, split screen

� Current inspired anesthetic agent value (agents include halothane, isoflurane, enflurane, desflurane, or sevoflurane)

� Current fraction inspired CO2 value

� Current fraction inspired O2 value

� Current heart rate

� RGM parameter key

� Current tidal volume value

� Alarm status bell (appears when alarms are turned ON)

Current peak inspiratory pressure

Current respiratory rate

� Current inspired nitrous oxide value

� Current end tidal CO2 value

Current SpO2 value

� Patient/room ID

� � � �

RGM

FiAX = 7.0%

FiCO2 = 3.0%

SPO2= 98%%

RR = 20BPM

� � � � �

� FiAX= 7% Fin2O= 50%RGM

SPO2= 98% FiO2=21%

RR= 20BPM HR = 77bpm

� Vt= 3.33L EtCO2=2.5%

� ICU 14

Flexport System Interface Operations Manual 33-5

Page 408: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5250

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter (except FiCO2). When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the RGM key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm limits for anesthetic agents are set at the gas monitor. When an agent parameter value exceeds an alarm limit, the monitor sends alarm data to the recorder.

When all parameter alarms are turned OFF, RGM status alarms do not cause the RGM key to flash. In particular, when the Apnea alarm is OFF and CO2 apnea occurs, the status message is displayed but does not initiate a flashing of the RGM key.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the gas monitor.

• Alarm limits can only be adjusted at the gas monitor. Alarm violation is detected at the gas monitor, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the gas monitor can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch RGM.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 33-6

Page 409: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5250

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the gas monitor’s parameters (refer to Figure 33-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays RGM ALM OFF to the right of the RGM key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 33-4: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the gas monitor. A maximum of three messages appear on any monitor. When more than three messages are received from the gas monitor, they appear according to a prioritized order (refer to Status Messages on page 33-9).

FiAX 8.0 FiO2 34 HR 150 Vt 1.10RGM

FiAX=7.0%6.6 18 121 0.86 FiCO2=3.0%

SPO2=98%

SpO2 80 EtCO2 9.3 FiN2O 80 Apnea 20 PIP 80 RR=20BPM?? 8.9 ?? ??

To view alarm settings:

• Touch RGM.• Touch VIEW ALARMS.

To view status:

• Touch RGM.• Touch VIEW STATUS.

Flexport System Interface Operations Manual 33-7

Page 410: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5250

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 33-4).

Selecting a Unit of MeasurementThe units of measurement can be toggled between percent (%) and millimeters of mercury (mmHg), or between percent (%) and kilo Pascals (kPa), depending on the bedside monitor setup.

Warning:When using this product on a network, make sure all monitors viewing this channel are set up for the same units.

To select a display format:

• Touch RGM.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To select a display format:

• Touch RGM.• Touch SETUP.• Select PERCENT or mmHg.

Flexport System Interface Operations Manual 33-8

Page 411: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5250

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the gas monitor or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the gas monitor is powered ON.

• Verify that all cables are secure.

Table 2 lists some of the other status messages that the Flexport system interface reports for the gas monitor during Flexport system interface operation.

To print the current display:

• Touch RGM.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOSTCommunication between the respiratory gas monitor and the Spacelabs Healthcare monitor has been interrupted.

Yes

CO2 APNEA The apnea alarm limit has been violated. No

HIGH SUSTAINED PAW The high-sustained PAW alarm limit has been violated. No

SUB-ATMOSPHERIC PAW The sub-atmospheric alarm limit has been violated. No

LOW CIRCUIT O2 The low circuit alarm limit has been violated. No

HIGH FICO2 The high FiCO2 alarm limit has been violated. No

LOW EXPIRED AGENT The low expired agent alarm limit has been violated. No

Flexport System Interface Operations Manual 33-9

Page 412: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 5250

HIGH EXPIRED AGENT The high expired agent alarm limit has been violated. No

AGENT DETECTED Anesthetic agent has been detected by the gas monitor. No

HIGH MINUTE VOLUME The high minute volume alarm limit has been violated. No

LOW MINUTE VOLUME The low minute volume alarm limit has been violated. No

LOW SUSTAINED PAW The low sustained PAW alarm limit has been violated. No

LINE/FILTER BLOCK The sample line or filter is blocked. Yes

REVERSE FLOW The sample flow sensor has detected a reverse one-directional airway flow. Yes

WATER TRAP FULL, PUMP OFF The fluid collection bottle is full. No

GAS ANALYZER SATURATED There is moisture in the measurement chamber or the calibration is incorrect. No

HIGH CIRCUIT O2 The high circuit alarm limit has been violated. No

SPO2 INOPERATIVE The SpO2 circuits have failed. No

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 33-10

Page 413: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 34-1

GE Ohmeda 7800/7810

Directory of KeysVENT

VENT - MAIN MENUALARM LIMITS SETUP VIEW

ALARMS PRINT

VENT - SETUP MENULARGE TXTSMALL TXT

VENT - ALARM LIMITS

APNEA VMIN O2 PIP

VENT - ALARM LIMITS - PIPPIP

ON OFF

VENT - ALARM LIMITS - O2O2

ON OFF

VENT - ALARM LIMITS - VMINVMIN

ON OFF

VENT - ALARM LIMITS - APNEAAPNEA

ON OFF

Page 414: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 415: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 7800/7810

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90436A-08 Flexport system interface enables information from a GE Ohmeda 7800/7810 ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

1200 7-bit odd 1 131-0685-xx (25-pin)

1.x or higher (7800)3.x or higher (7810)

Flexport System Interface Operations Manual 34-3

Page 416: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 7800/7810

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 34-1: Bedside monitor, large text screen

Figure 34-2: Bedside monitor, small text

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the ventilator.

� � � � �

2 PIP

cmH2O 21 O2

% 9.9 Vmin

LVENT

98 Vt

ml

1 PEEP

cmH2O0 PLT

cmH2O

MODE:MECH VENT ON 29 RR

BPM

� � �

� � � �

VENT

VT = 98 ml

RR = 29 BPM

PIP = 2 cmH2O

PEEP = 1 cmH2O

� �

Flexport System Interface Operations Manual 34-4

Page 417: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 7800/7810

Figure 34-3: Central monitor, split screen

� Measured peak inspiratory pressure

� Measured oxygen level

� Measured minute volume

� VENT parameter key

� Measured tidal volume

� Alarm status bell (appears when alarms are turned ON)

� Measured respiratory rate

Mechanical ventilator — ON/OFF status

Measured inspiratory plateau pressure

� Measured positive end expiratory pressure

� Apnea monitoring status

Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms still cause the VENT key to flash.

Alarm settings default to OFF.

� � � � �

� RR = 29 BPM O2 = 21% VENT

� APNEA = 30s PIP = 2 cmH2O

� VT = 98 ml PLT = 0 cmH2O

� VMIN = 9.9 L PEEP = 1 cmH2O

ICU 14

Flexport System Interface Operations Manual 34-5

Page 418: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 7800/7810

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• The Apnea alarm limit is fixed at 30 seconds.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 34-4). If alarms for all ventilator parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 34-4: View Alarms screen

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

APNEA 30 VMIN PIP 20 O2 73 VENT

VT = 98 ml3.0 18 RR = 28 BPM

PIP = 2 cmH2O

ICU 14PEEP = 1 cmH2O

Flexport System Interface Operations Manual 34-6

Page 419: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 7800/7810

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 34-4).

The default setting is LARGE TXT.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To print the current display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 34-7

Page 420: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda 7800/7810

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the ventilator and the monitor has been interrupted. Yes

VENT VOL MON STANDBY The ventilator is waiting for first breath to activate volume monitoring and apnea timer. No

CHECK VENT

The ventilator detects and reports one or more of the following alarm conditions:

• Sustained pressure alarm• Sub-atmospheric pressure alarm• AC fail (primary supply voltage low)• Low battery alarm• O2 limit set error• Vent setting range error• O2 sensor failure alarm• volume sensor failure• Maximum pressure > 60 cmH2O• Reverse flow• Low gas supply pressure alarm• Vent fail 0–12• Volume sensor cartridge coasting

(end of breath not detected)

Yes

Flexport System Interface Operations Manual 34-8

Page 421: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 35-1

GE Ohmeda Care-Plus/Infant Warmer

Directory of Keys

Note:WARMR will replace INCUB in the display when the unit is connected to the GE Ohmeda infant warmer.

INCUB

INCUB - MAIN MENUALARM LIMITS SETUP

*PRINT

INCUB - SETUP MENULARGE TXTSMALL TXT

INCUB - ALARM LIMITS

Pt AIR**

*This key appears on bedside monitors only.** This key appears only when the interface is connected to an incubator

INCUB - ALARM LIMITS - AIRAIR

ON OFF

INCUB - ALARM LIMITS - PtPt

ON OFF

Page 422: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 423: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Care-Plus/Infant Warmer

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90444A-01 Flexport system interface allows information from a GE Ohmeda Care Plus incubator or infant warmer to be displayed on Spacelabs Healthcare monitors.

Configure the incubator/warmer as shown in Table 1. Consult the incubator/warmer manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the incubator/warmer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the incubator/warmer. To control other interface functions, refer to the sections that follow.

Table 1: Incubator/Warmer Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

1200 7-bit odd 1 131-0686-xx

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the incubator/warmer.

Flexport System Interface Operations Manual 35-3

Page 424: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Care-Plus/Infant Warmer

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the incubator/warmer.

Incubator

Figure 35-1: Bedside monitor, large text screen

Figure 35-2: Bedside monitor, small text screen

Figure 35-3: Central monitor, split screen

� P-A is the difference between the patient temperature and the air temperature

� Control setting temperature

� INCUB parameter key

� � � �

0.6PA

C 35.4 CTRL

CINCUB

35.6 Pt

C

PATIENT CTRL 35.0 AIR

C

� � �

� �

INCUB

Pt = 35.6 C �AIR = 35.0 C �P-A = 0.6 C

CTRL = 35.4 C

� �

� � � � �

Pt = 35.6 C CTRL = 35.4 C INCUB

� AIR = 35.0 C PATIENT CTRL

ICU 14

Flexport System Interface Operations Manual 35-4

Page 425: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Care-Plus/Infant Warmer

� Current patient temperature

� Alarm status bell (appears when alarms are turned ON)

� Current air temperature inside incubator

� Incubator control mode:

• PATIENT CTRL indicates incubator is in the Patient Control mode

• AIR CTRL indicates incubator is in the Air Control mode

Patient/room ID

Infant Warmer

Figure 35-4: Bedside monitor, large text screen

Figure 35-5: Bedside monitor, small text screen

Figure 35-6: Central monitor, split screen

� � �

36.5 CTRL

CWARMR

36.3 Pt

C

SERVO MODE 65 POWER

%

� � �

� � �

WARMR

Pt = 36.3 C

POWER=65%CTRL = 36.5 C

� �

� � � � �

Pt = 36.3 C CTRL = 36.5 C WARMR

� POWER = 65.0% SERVO MODE

� ICU 14

Flexport System Interface Operations Manual 35-5

Page 426: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Care-Plus/Infant Warmer

� Control setting temperature

� WARMR parameter key

� Current patient temperature

� Alarm status bell (appears when alarms are turned ON)

� Current percentage of total available heater power being delivered to warmer

� Infant warmer control mode:

• SERVO MODE indicates infant warmer is in the Servo mode

• MANUAL MODE indicates infant warmer is in the Manual mode

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the INCUB (or WARMR) key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When the interface is connected to an incubator:

• If the incubator is in Patient Control mode, you can turn ON the Patient alarm only.

• If the incubator is in Air Control mode, you can turn ON the Air alarm only.

• When both parameter alarms are turned OFF, status alarms will still cause the CHECK INCUB message to display.

When the interface is connected to an infant warmer:

• If the infant warmer is in Servo mode, you can turn ON the Patient alarm only.

• If the infant warmer is in Manual mode, no alarms are present.

Alarm settings default to OFF.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch INCUB (or WARMR).• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 35-6

Page 427: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Care-Plus/Infant Warmer

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the incubator/

warmer.

• Alarm limits can only be adjusted at the incubator/warmer. Alarm violation is detected at the incubator/warmer and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the incubator/warmer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Table 2 lists additional system alarms received from the incubator/warmer. Refer to the incubator/warmer manual for additional information.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 35-4).

The default setting is LARGE TXT.

Table 2: System Alarms

Incubator Infant Warmer

HIGH AIR TEMP HEATER POSITION

AIR CIRCULATION CHECK PATIENT

PROBE FAILURE PROBE FAILURE

SYSTEM FAILURE SYSTEM FAILURE

POWER FAILURE POWER FAILURE

To select a display format:

• Touch INCUB.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 35-7

Page 428: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Care-Plus/Infant Warmer

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the incubator/warmer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the incubator/warmer is powered ON.

• Verify that all cables are secure.

To print the current display:

• Touch INCUB (or WARMR).• Touch PRINT.

Flexport System Interface Operations Manual 35-8

Page 429: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 36-1

GE Ohmeda Rascal II

Directory of KeysGAS

GAS - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFF

GAS - SETUP

SIZE SWEEPSPEED

LARGE TXTSMALL TXT

GAS - SETUP - SIZE GAS - SETUP - SWEEP SPEEDWAVEFORM SIZE

↑SIZE

↓25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/secON OFF

GAS - ALARM LIMITS

FiO2 FiAGT RR EtCO2

GAS - ALARM LIMITS - EtCO2EtCO2 HI =

XXXLO = XXX ↑ ↓

EtCO2ON OFF FiO2

GAS - ALARM LIMITS - RRRR HI =

XXXLO = XXX ↑ ↓

RRON OFF EtCO2

GAS - ALARM LIMITS - FiAGTFiAGT HI =

XXXLO = XXX ↑ ↓

FiAGTON OFF RR

GAS - ALARM LIMITS - FiO2FiO2 HI =

XXXLO = XXX ↑ Ø

FiO2ON OFF FiAGT

Page 430: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 431: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Rascal II

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90443A-07 Flexport system interface enables information from a GE Ohmeda Rascal II gas monitor to be displayed on Spacelabs Healthcare monitors.

Configure the gas monitor as shown in Table 1. Consult the gas monitor manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the gas monitor, complete the steps listed in the following Quickstart.

Table 1: Gas Monitor Configuration

Baud Rate Character Length

Parity Setting Stop Bit

9600 8-bit none 1

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the 8-pin receptacle on the Flexport system interface.

• Plug the 25-pin female connector on the other end of the modular cable into the connector on the gas monitor labeled RS-232.

• Plug the 37-pin female connector into the ANALOG connector on the gas monitor.

Flexport System Interface Operations Manual 36-3

Page 432: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Rascal II

Note:If the gas monitor does not have the analog option installed, turn the waveform option OFF (refer to Turning Waveforms ON or OFF on page 36-7).

The Spacelabs Healthcare monitor is now ready to display information from the gas monitor. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the gas monitor.

Figure 36-1: Bedside monitor, large text screen

Figure 36-2: Bedside monitor, small text screen

� � � � � �

21 FiO2

% 0.89 FiENF

% 78 FiN2O

% 2 FiN2

% GAS

13 FiCO2

mmHg

12RR

BPM 0.75FeENF

%77 FeN2O

%1 FeN2

% 39EtCO2

mmHg

� � �

� � �

GAS

FiO2= 21%

FiENF= 0.89%

EtCO2= 39mmHg

RR= 12BPM

Flexport System Interface Operations Manual 36-4

Page 433: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Rascal II

Figure 36-3: Central monitor, split screen

� Fractional inspired O2 value

� Fractional inspired agent value (parameter name will be FiHAL, FiENF, FiISO, FiDES, FiSEV, or FiAGT for halothane, enflurane, isoflurane, desflurane, sevoflurane, or an unidentified agent, respectively)

� Fractional inspired N2O

� Fractional inspired N2

� Fractional inspired CO2

� Alarm status bell (appears when alarms are turned ON)

� End tidal CO2

GAS parameter key

Fractional expired N2

� Fractional expired N2O

� Fractional expired agent (agent indication changes as noted in � above)

Respiration rate

Note:• Data displayed is delayed by at least six seconds from the data shown on the gas analyzer.

• COMMUNICATION LINK LOST will be displayed when the gas analyzer goes into calibration (CAL). If the operator does not promptly exit the CAL screen, the Flexport system interface will drop the Flexport screen from the bedside or central monitor (refer to <Plain>Status Messages on page 36-9).

� �

� FiO2=21% FiENF = 0.89%GAS

RR=12BPM FeENF=0.75%

� FiCO2=13mmHg FiN2O=78%

� EtCO2=39mmHg FeN2O=77%

� � �

Flexport System Interface Operations Manual 36-5

Page 434: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Rascal II

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the GAS key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the gas

analyzer.

• Alarm limits on the Flexport system interface are independent from the gas analyzer.

• Alarm violations detected at the gas analyzer may not be reported by the Flexport system interface if the alarm limit levels are different.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch GAS.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To adjust alarm limits:

• Touch GAS.• Select ALARM LIMITS.• Select an alarm parameter.• Select HI= or LO=.• Use the arrow keys to adjust.

Flexport System Interface Operations Manual 36-6

Page 435: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Rascal II

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the gas analyzer’s parameters (refer to Figure 36-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays GAS ALM OFF to the right of the GAS key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated. If an individual alarm is OFF, no limit values are displayed next to the parameter label.

Figure 36-4: Views Alarms screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

To view alarm settings:

• Touch GAS.• Touch VIEW ALARMS.

To turn the waveform display ON or OFF:

• Touch GAS.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

FiO2 45 FiENF 9.50 GAS

18 1.50

EtCO2 75 RR 4020 5

Flexport System Interface Operations Manual 36-7

Page 436: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Rascal II

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, 12.5 and 25 mm/second.

The default setting is 12.5 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitors (refer to Display Detail on page 36-4).

To adjust waveform size:

• Touch GAS.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch GAS.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch GAS.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 36-8

Page 437: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

GE Ohmeda Rascal II

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the gas analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the gas analyzer is powered ON.

• Verify that all cables are secure.

• If the gas analyzer is calibrating, wait until it is finished, then return to normal monitoring.

To freeze the waveform on the display:

• Touch GAS.• Select FREEZE ON.

To print the current display:

• Touch GAS.• Touch PRINT.

Flexport System Interface Operations Manual 36-9

Page 438: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 439: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 37-1

Hamilton Amadeus/Veolar/Galileo/Raphael XTC

Directory of Keys

VENT

VENT - MAIN MENU

ALARM LIMITS SETUP

FREEZEMECHANICS MODE

XXX*UNKINOWN

MODEVIEW

ALARMS PRINTON OFF

*This could also be: UNKNOWN MODEMODE A/CONTRL

MODE SIMVMODE SPONT

MODE MMVMODE AMBIENTMODE PCV-CMVMODE PCV-SIMV

VENT - C in ml/cmH2O VENT - R in cmH2O/L/SC = XXX R = XXX

VENT - SETUP

SIZE SWEEP SPEED

WAVE TYPE

LARGE TXT

SMALL TXT

PATIENT PRESSURE

OPTIONAL PRESSURE FLOW VOLUME

50mm/sec

25mm/sec

12.5mm/sec

6.25mm/sec

3.12mm/sec

1.58mm/sec

WAVEFORM↑ ↓

ON OFF

VENT - ALARM LIMITS

RR VMIN FiO2 PIP PEEPPEEP alarms are not available from the Galileo/Raphael XTC ventilators.

RR VMIN FiO2 PIP PEEP

ON OFF ON OFF ON OFF ON OFF ON OFF

Page 440: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 441: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Hamilton Amadeus/Veolar/Galileo/Raphael XTC

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90436A-07 Flexport system interface enables information from a Hamilton Amadeus/Veolar/ Galileo/Raphael XTC ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Vent Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

Amadeus/Veolar 9600 7-bit even 2 131-0685-xx(25-pin) NIK01S.1

Galileo/Raphael XTC 9600 7-bit even 2 131-0686-xx

(9-pin) 3.2

Flexport System Interface Operations Manual 37-3

Page 442: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Hamilton Amadeus/Veolar/Galileo/Raphael XTC

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Note:PLT pressures are not sent by ventilators in a non-pressure mode.

Figure 37-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the ventilator.

� � � �

30 PIP 20 FIO2 4.1 Vmin VENT

340 Vt

cmH2O % L mL

1 PEEP 29 PLT 9 MAP 10 RR

cmH2O cmH2O cmH2O BPM

� � � �

Flexport System Interface Operations Manual 37-4

Page 443: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Hamilton Amadeus/Veolar/Galileo/Raphael XTC

Figure 37-2: Bedside monitor, small text

Figure 37-3: Central monitor, split screen

� Measured peak inspiratory pressure

� Measured oxygen level

� Measured minute volume

� Measured tidal volume

� Alarm status bell (appears when alarms are turned ON)

� Measured respiratory rate

� VENT parameter key

Mean airway pressure

Measured inspiratory plateau pressure

� Measured positive end expiratory pressure

� Patient/room ID

� �

VENT

VT = 340 mL

RR = 10 BPM �

PIP = 30 cmH2O �PEEP = 1 cmH2O

� �

� � � � �

� RR = 10 BPM FIO2 = 20% VENT

MAP = 9 cmH2O PIP = 30 cmH2O

� VT = 340 mL PLT = 29 cmH2O

� VMIN = 4.1 L PEEP = 1 cmH2O

� ICU 14

Flexport System Interface Operations Manual 37-5

Page 444: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Hamilton Amadeus/Veolar/Galileo/Raphael XTC

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms still cause the VENT key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 37-6

Page 445: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Hamilton Amadeus/Veolar/Galileo/Raphael XTC

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 37-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Note:PEEP alarms are not sent by the Galileo and the Raphael ventilators.

Figure 37-4: View Alarms screen

Selecting a Display FormatTwo screen formats are available for the bedside monitor (refer to Display Detail on page 37-4).

The default setting is LARGE TXT.

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To select a screen format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

RR 29 VMIN 4.1 PIP 79 VENT

VT = 340 mL0.2 RR = 29 BPM

ICU 14 FIO2 20 PEEP ON PIP = 30 cmH2O10 PEEP = 1 cmH2O

Flexport System Interface Operations Manual 37-7

Page 446: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Hamilton Amadeus/Veolar/Galileo/Raphael XTC

Printing the Current DisplayTouch the PRINT key to print the information shown on the current screen. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 37-8

Page 447: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Hamilton Amadeus/Veolar/Galileo/Raphael XTC

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 displays the status messages that the Flexport system interface reports for the ventilator during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

GENERAL ALARM Loss of power or inappropriate voltage levels internally. Yes

DISCONNECT PATIENT Patient coughs, patient fights against the ventilator, an acute change in lung compliance/airway resistance. Yes

APNEACessation of breathing during spontaneous modes. Patient has not exhaled at least 50 ml/second flow through the flow sensor during the last 20 seconds.

Yes

DISCONNECTED VENTTubing disconnected, large air leaks, hiccups, defective expiratory membrane, or incorrect flow sensor measurement.

Yes

OPERATOR Fan filter occluded, flow sensor backward, or incorrect control setting. Yes

GAS SUPPLY Gas supply failure. Yes

Flexport System Interface Operations Manual 37-9

Page 448: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 449: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 38-1

Infrasonics Adult Star 2000

Directory of KeysVENT

VENT - MAIN MENUALARM LIMITS SETUP VIEW

MODEVIEW

STATUSVIEW

ALARMS PRINT

VENT - SETUPLARGE TXTSMALL TXT

ANES - ALARM LIMITS

RR VT VMIN PIP PEEP I:E APNEA

VENT - ALARM LIMITS - APNEAAPNEA HI =

XXRR

ON OFF APNEA

VENT - ALARM LIMITS - I:EI:E LO =

X.XI:E

ON OFF APNEA

VENT - ALARM LIMITS - PEEPPEEP LO =

XXXPEEP

ON OFF I:E

VENT - ALARM LIMITS - PIPPIP HI =

XXLO = XXX

PIPON OFF PEEP

VENT - ALARM LIMITS - VMINVMIN LO =

XXXVMIN

ON OFF PIP

VENT - ALARM LIMITS - VTVT HI =

XXXLO = XX.X

VTON OFF VMIN

VENT - ALARM LIMITS - RRRR HI =

XXRR

ON OFF VT

Page 450: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 451: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Infrasonics Adult Star 2000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90436A-05 Flexport system interface enables information from an Infrasonics Adult Star 2000 ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

TransitionConnector

Part Number

Minimum Software Revision

2400 8 bit none 1 131-1155-xx 216 or higher

Flexport System Interface Operations Manual 38-3

Page 452: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Infrasonics Adult Star 2000

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 38-1: Bedside monitor, large text screen

Figure 38-2: Bedside monitor, small text

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 jack on the ventilator.

� � � � � �

68 PIP 65 PLT 21 FiO2 VENT

405 VT

I:E cmH2O cmH2O % mL

1.8 5 PEEP 20 MAP 4.0 VMIN 10 RR

cmH2O cmH2O L BPM

� � �

� � �

VENT

VT= 405 mL

RR= 10 BPM

PIP= 68 cmH2O

PEEP= 5 cmH2O

� �

Flexport System Interface Operations Manual 38-4

Page 453: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Infrasonics Adult Star 2000

Figure 38-3: Central monitor, split screen

� Ratio of inspiration time to expiration time (I:E ratio is typically expressed as 1:X.X where X.X is the value displayed and equals E/I)

� Current peak inspiratory pressure value

� Current inspiratory plateau pressure value

� Current fractional inspired O2 value

� VENT parameter key

� Current tidal volume value

� Alarm status bell (appears when alarms are turned ON)

Current respiratory rate value

Current minute volume value

� Current mean airway pressure value

� Current positive end expiratory pressure value

Patient/room ID

� � � � � �

RR = 10 BPM VT = 405 mL VENT

� FiO2 = 21% PIP = 68 cmH2O

� PEEP = 5 cmH2O MAP = 20 cmH2O

VMIN = 4.0 L I : E = 1.8

ICU - 4

Flexport System Interface Operations Manual 38-5

Page 454: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Infrasonics Adult Star 2000

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Touch ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 38-6

Page 455: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Infrasonics Adult Star 2000

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 38-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

VT HI LIMIT will be present if the alarm is ON and the ventilator is in PCV/Time Cycled or PCV/I:E Cycled mode, but will not be present in Volume Cycled mode.

Figure 38-4: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 38-10). When more than four messages are received from the ventilator, they are displayed in receiving order.

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

PIP 120 VT 650 APNEA 20 VENT

405VT

3 160 mL

I : E PEEP VMIN RR 70 10 RR

1.0 2 0.3 BPM

Flexport System Interface Operations Manual 38-7

Page 456: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Infrasonics Adult Star 2000

Viewing Ventilator Mode HistoryChronological history of the last eight ventilator mode settings are displayed with the oldest value at the top of the left column and the most recent value at the bottom of the right column. Possible modes include ASST/CTL, SIMV, CPAP, SIMV+PPS, CPAP+PPS, and APNEA.

Figure 38-5: View Mode screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 38-4).

The default setting is LARGE TXT.

To view ventilator mode history:

• Touch VENT.• Touch VIEW MODE.

To select a display format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

15:47 ASST/CTL 15:49 ASST/CTL VENT

405 VT

15:47 SIMV 15:49 CPAP+PPS mL

15:47 SIMV+PPS 15:51 CPAP 10 RR

15:49 CPAP 15:56 SIMV BPM

Flexport System Interface Operations Manual 38-8

Page 457: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Infrasonics Adult Star 2000

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the ventilator display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 38-9

Page 458: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Infrasonics Adult Star 2000

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

LOW PRS O2 INLET Pressure at the oxygen inlet fitting is less than 30 psig and the O2% is 22 or greater. No

LOW PRS AIR INLET Pressure at the air inlet is equal to or less than 30 psig. No

LOW BATTERY Available power life is less than 4 minutes. No

OBSTRUCTED TUBE Airway is obstructed or airway pressure line is obstructed. No

AIRWAY LEAK Leak of 6 LPM or more has been detected. No

EXHAL VALVE LEAK Exhalation flow is greater than 10% of inhaled set flow. Leak is greater than 4 LPM for 60 milliseconds. No

HIGH PEEP/CPAP The baseline pressure has fluctuated between 6 to 10 cmH2O above set parameter for more than 5 seconds. No

SHORT INSP TIME Inspiratory time is less than 240 milliseconds for mechanical breaths. No

SHORT EXHAL TIME Exhalation time is less than 200 milliseconds for mechanical breaths or 260 milliseconds for PPS breaths. No

Flexport System Interface Operations Manual 38-10

Page 459: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 300

Directory of KeysVENT

VENT - MAIN MENU

SETUPFREEZE LARGE TXT TABULAR

MODEVIEW

ALARMS PRINTON OFF SMALL TXT

VENT - SETUP WAVE SIZE

↑SIZE

↓SWEEP SPEED

WAVE TYPEON OFF PRS FLOW

FLOW - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec

Flexport System Interface Operations Manual 39-1

Page 460: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 461: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 300

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing the Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90436A-03 Flexport system interface enables information from a Maquet Siemens Servo 300 ventilator to be displayed on Spacelabs Healthcare monitors.

Warning:This product does not support alarms.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Ventilator Baud Rate

Character Length

Parity Setting Stop Bit

Transition Connector

Part Number

Minimum Software Revision

Servo 300 9600 8 bit even 1 131-1067-xx(26-pin) COM x1.01

Flexport System Interface Operations Manual 39-3

Page 462: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 300

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 39-1: Bedside monitor, large text screen

Figure 39-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the data communication connector (N82) of the

ventilator.

� � � � � �

13 PIP 57 PLT 20 FiO2 VENT

0.33 Vt WAVE=FLOWcmH2O cmH2O % L

4 PEEP 98 MAP 7.6 Vmin 22 RRcmH2O cmH2O L BPM

� �

� � �

VENT

Vt = 0.33 L WAVE=PRESSRR = 72 BPM �

PIP = 13 cmH2O �FiO2 = 20 % �

Flexport System Interface Operations Manual 39-4

Page 463: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 300

Figure 39-3: Central monitor, split screen

� Measured peak inspiratory pressure (PIP)

� Plateau pressure (PLT)

� Measured oxygen level (FiO2)

� VENT parameter key

� Measured tidal volume (Vt)

� Waveform type (PRESSURE or FLOW)

� Measured respiratory rate (RR)

Measured minute volume (Vmin)

Mean airway pressure (MAP)

� Positive end expiratory pressure (PEEP)

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVE ON.

To turn the waveform display ON or OFF:

• Touch VENT.• Touch SETUP.• Select WAVEFORM ON/OFF.

� �

� Vt = 0.33 L PIP = 13 cmH2O VENT

Vmin = 7.5 L MAP = 5 cmH2O

� RR = 22BPM PLT = 81 cmH2O

� FiO2 = 20% PEEP = 4 cmH2O

� � �

Flexport System Interface Operations Manual 39-5

Page 464: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 300

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVE ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 25, 12.5, 6.25, and 3.12 mm/second.

The default setting is 12.5 mm/second.

Selecting a Waveform TypeYou can select either pressure or flow waveforms for display. Select PRS to display the pressure waveform, or select FLO to display the flow waveform. The two waveform displays are mutually exclusive.

To adjust waveform size:

• Touch VENT.• Touch SETUP.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch VENT.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a waveform type:

• Touch VENT.• Touch SETUP.• Touch WAVE.• Select PRS or FLO.

Flexport System Interface Operations Manual 39-6

Page 465: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 300

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 39-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Viewing the Ventilator Mode HistoryThe last eight ventilator mode settings that were used are displayed in the waveform zone. The oldest value is at the top of the left column. The most recent value at the bottom of the right column.

To select a display format:

• Touch VENT.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch VENT.• Select FREEZE ON.

To view ventilator mode history:

• Touch VENT.• Touch VIEW MODE.

Flexport System Interface Operations Manual 39-7

Page 466: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 300

Viewing Status MessagesThe VIEW STATUS key allows you to display the recent status messages sent by the ventilator.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThis section includes information on how to recognize and respond to a status message.

The message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

To view ventilator mode history:

• Touch VENT.• Touch VIEW STATUS.

To print the current display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 39-8

Page 467: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 900

Directory of KeysVENT

VENT - MAIN MENU

SETUPFREEZE LARGE TXT VIEW

STATUS PRINTON OFF SMALL TXT

VENT - SETUP WAVE SIZE

↑SIZE

↓SWEEP SPEED

WAVE TYPEON OFF PRS FLOW

FLOW - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec

Flexport System Interface Operations Manual 40-1

Page 468: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 469: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 900

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90436A-02 Flexport system interface enables information from a Maquet Siemens Servo 900 ventilator to be displayed on Spacelabs Healthcare monitors.

Warning:This product does not support alarms.

Note:• The Maquet Siemens Servo 900 ventilator updates the patient’s parameter data with each patient

breath cycle and sends this information through the Maquet Siemens 990 digital interface to the Flexport. In Ventilator Assist, the ventilator updates the respiration rate value only after each detected breath. The updated respiration rate is correctly displayed on the bedside monitor.

• If a patient does not take a subsequent breath, the data displayed will remain unchanged indefinitely because the 990 digital interface will not update the displayed values.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting Stop Bit

Transition Connector

Part Number

Minimum Software Revision

9600 8 bit even 1 131-0685-xx(25-pin) COM x1.01

Flexport System Interface Operations Manual 40-3

Page 470: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 900

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 40-1: Bedside monitor, large text screen

Figure 40-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the data communication connector (N82) of the

ventilator.

� � � � � �

13 PIP 57 PLT 20 FiO2 VENT

0.33 Vt WAVE=FLOWcmH2O cmH2O % L

4 98 MAP 7.6 Vmin 22 RRcmH2O L BPM

� �

� � �

VENT

Vt = 0.33 L WAVE=PRESSRR = 72 BPM �

PIP = 13 cmH2O �FiO2 = 20 % �

Flexport System Interface Operations Manual 40-4

Page 471: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 900

Figure 40-3: Central monitor, split screen

� Measured peak inspiratory pressure (PIP)

� Plateau pressure (PLT)

� Measured oxygen level (FiO2)

� VENT parameter key

� Measured tidal volume (Vt)

� Waveform type (PRESSURE or FLOW)

� Measured respiratory rate (RR)

Measured minute volume (Vmin)

Mean airway pressure (MAP)

� Positive end expiratory pressure (PEEP)

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVE ON.

To turn the waveform display ON or OFF:

• Touch VENT.• Touch SETUP.• Select WAVEFORM ON/OFF.

� �

� Vt = 0.33 L PIP = 13 cmH2O VENT

Vmin = 7.5 L MAP = 5 cmH2O

� RR = 22BPM PLT = 81 cmH2O

� FiO2 = 20% PEEP = 4 cmH2O

� � �

Flexport System Interface Operations Manual 40-5

Page 472: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 900

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVE ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 25, 12.5, 6.25, and 3.12 mm/second.

The default setting is 12.5 mm/second.

Selecting a Waveform TypeYou can select either pressure or flow waveforms for display. Select PRS to display the pressure waveform, or select FLO to display the flow waveform. The two waveform displays are mutually exclusive.

To adjust waveform size:

• Touch VENT.• Touch SETUP.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch VENT.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a waveform type:

• Touch VENT.• Touch SETUP.• Touch WAVE.• Select PRS or FLO.

Flexport System Interface Operations Manual 40-6

Page 473: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 900

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 40-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select a display format:

• Touch VENT.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch VENT.• Select FREEZE ON.

To print the current display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 40-7

Page 474: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Maquet Siemens Servo 900

Status MessagesThis section includes information on how to recognize and respond to a status message.

The message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Flexport System Interface Operations Manual 40-8

Page 475: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 41-1

Masimo Radical

Directory of KeysLINK

LINK - MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFF

These keys appear on the bedside monitor only.

LINK - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

LINK - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

LINK - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec0.39

mm/sec

LINK - ALARM LIMITS

SPO2 PR

LINK - ALARM LIMITS - PRPR HI =

XXLO = XXON OFF

LINK - ALARM LIMITS - SpO2SPO2 HI =

XXLO = XXON OFF

Page 476: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 477: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Masimo Radical

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a Masimo Radical pulse oximeter to be displayed on Spacelabs Healthcare monitors.

Configure the pulse oximeter as shown in Table 1. Consult the pulse oximeter manual for additional configuration setup procedures

Table 1: Pulse Oximeter Configuration

Baud Rate Minimum Software Revision Installation Kit Part Number

Transition Connector

Part Number

9600Masimo SET V3.0.2.2Handheld R2.1.0.5D-Station R2.1.0.7

045-0152-xx 131-1755-xx (9-pin)

Flexport System Interface Operations Manual 41-3

Page 478: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Masimo Radical

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the pulse oximeter, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the pulse oximeter. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pulse oximeter.

Figure 41-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Power on the pulse oximeter and set it up to communicate with the

Spacelabs Healthcare protocol.• Plug the transition connector into the RS-232 connector on the pulse oximeter serial

port.

� �

LINK

98SpO2

%

70 PR

bpm

� �

Flexport System Interface Operations Manual 41-4

Page 479: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Masimo Radical

Figure 41-2: Bedside monitor, small text screen

Figure 41-3: Central monitor, split screen

� LINK parameter key

� Current SpO2 value

� Alarm status bell (appears when alarms are turned ON)

� Current pulse rate value

� Patient/room ID

� �

LINK

SpO2= 98%

PR= 70bpm

� �

� �

� SpO2 = 98% LINK

� PR = 70bpm

� ICU - 4

Flexport System Interface Operations Manual 41-5

Page 480: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Masimo Radical

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the LINK key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to ON.

Note:• Turning interface alarms ON or OFF does not affect alarm settings on the pulse oximeter.

• Alarm limits can only be adjusted at the pulse oximeter. Alarm violation is detected at the pulse oximeter, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• The SpO2 high limit can be turned OFF at the pulse oximeter. When this high limit is OFF, the value 101 is displayed as the SpO2 high limit on the Spacelabs Healthcare monitor(s).

• Alarm tones on the pulse oximeter can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch LINK.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 41-6

Page 481: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Masimo Radical

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the pulse oximeter’s parameters (refer to Figure 41-4). If alarms for both parameters are turned OFF, the Spacelabs Healthcare monitor displays LINK ALM OFF to the right of the LINK key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 41-4: View Alarms screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

To view alarm settings:

• Touch LINK.• Touch VIEW ALARMS.

To turn the waveform display ON or OFF:

• Touch LINK.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

SpO2 100 LINK

98SpO2

85 %

PR 180 70 PR

40 bpm

Flexport System Interface Operations Manual 41-7

Page 482: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Masimo Radical

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 50, 25, 12.5, 6.25, 3.12, 1.56, 0.78, and 0.39 mm/second. The minimum recorder sweep speed is 1.56 mm/second.

The default setting is 25 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 41-4).

The default setting is LARGE TXT.

To change waveform size:

• Touch LINK.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch LINK.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch LINK.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 41-8

Page 483: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Masimo Radical

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the oximeter or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pulse oximeter is powered ON.

• Verify that all cables are secure.

To freeze the waveform on the display:

• Touch LINK.• Select FREEZE ON.

To print the pulse oximeter display:

• Touch LINK.• Touch PRINT.

Flexport System Interface Operations Manual 41-9

Page 484: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 485: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 42-1

Mortara ELI 100/STM

Directory of Keys

ST

ST - MAIN MENUALARM LIMITS

WAVEFORM SIZE SETUP RHYTHM

LEAD PRINT

ST - RHYTHM LEADI II III V1 V2 V3 V4 V5 V6 AVF AVL AVR

ST - SETUP

SWEEP SPEED QRS TONE

WAVEFORM ST VALUES

ON OFF ON OFF

ST - SETUP - QRS TONETONE VOLUME

↑VOLUME

↓ON OFF

ST - SETUP - SWEEP SPEED

50mm/sec

25mm/sec

12.5mm/sec

ST - WAVEFORM SIZE

SIZE ↑

SIZE ↓

ST - ALARM LIMITSALARMS SINGLE

ST = XXXMULTI

ST = XXXON OFF

Page 486: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 487: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Mortara ELI 100/STM

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Adjusting the Waveform Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Selecting a Rhythm Lead Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90440A-01 Flexport system interface enables information from a Mortara Instrument ELI 100/STM ST monitor to be displayed on Spacelabs Healthcare monitors.

The ST monitor must have a monitoring session in progress before data will be sent to the Flexport system interface. To begin a monitoring session on the ST monitor, complete the steps listed in the following Quickstart.

Configure the ST monitor as shown in Table 1. Consult the ST monitor manual for additional configuration setup procedures.

Table 1: ST Monitor Configuration

Baud Rate Character Length

Parity Setting

Stop Bit

19,200 8 bit even 1

To start an ST monitoring session from the Mortara ST monitor:

• Touch ON/OFF/DC (function key) to access the main menu.• Touch MON to access ST monitoring.• Touch YES to start a new monitoring session.• Touch YES to clear previous patient data.• Touch YES for New Patient.• Enter patient ID information.

NOTE: The ST monitor acquires a template ECG, prints it, and begins monitoring. The ST monitoring menu appears.

Flexport System Interface Operations Manual 42-3

Page 488: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Mortara ELI 100/STM

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ST monitor, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ST monitor. To control other interface functions refer to the sections that follow.

Display DetailWhen you power ON an Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ST monitor.

Figure 42-1: Bedside/remote monitor, full screen

Figure 42-2: Central monitor, split screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the 8-pin receptacle on the Flexport system interface.

• Plug the 9-pin male connector on the other end of the modular cable into the RS-232 connector on the ST monitor labeled SERIAL.

� � � �

I 20 AVR -10 V1 05 V4 10 LEAD = V3II 00 AVL 20 V2 00 V5 00 S

TST = -20uV

III -20 AVF -10 V3 -20 V6 -15

ICU - 5

� � �

� �

I 20 V2 00 V6 -15II 00 V3 20 AVR -10 S

TIII -20 V4 10 AVL 20V1 05 V5 00 AVF -10

� ICU - 5

Flexport System Interface Operations Manual 42-4

Page 489: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Mortara ELI 100/STM

� ST values for 12 leads

� ST parameter key

� Selected rhythm lead designator

� QRS detect indicator

� Alarm status bell (appears when alarms are turned ON)

� ST value for selected lead

� Patient/room ID (appears on remote and central monitors only)

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for SINGLE ST and MULTI ST alarms. The SINGLE ST segment level alarm is violated when the magnitude of a single ST segment level exceeds the single lead threshold. The MULTI level alarm is violated when the magnitude of any two or more leads exceeds the two-lead threshold.

When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• the alarm limits key for that parameter flashes,

• the parameter value flashes,

• the ST key flashes, and

• the alarm bell flashes.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ST monitor.

• Alarm limits can only be adjusted at the ST monitor. Alarm violation is detected at the ST monitor, and the Flexport system interface reports the alarm if the alarm key is turned ON.

• Alarm reporting may not occur simultaneously at the ST monitor and the Spacelabs Healthcare monitor. The alarm may be reported at the Spacelabs Healthcare monitor before it is reported at the ST monitor.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch ST.• Select ALARM LIMITS.• Select ALARMS ON or OFF.

Flexport System Interface Operations Manual 42-5

Page 490: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Mortara ELI 100/STM

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 50, 25, and 12.5 mm/second.

The default setting is 25 mm/second.

Adjusting the QRS ToneThe QRS tone can be turned ON or OFF. When it is ON, you can adjust the volume level.

The default setting is TONE OFF.

To select a sweep speed:

• Touch ST.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To adjust the QRS tone:

• Touch ST.• Touch SETUP.• Touch QRS TONE.• Touch TONE ON.• Adjust the volume using VOLUME ↑ or VOLUME ↓

Flexport System Interface Operations Manual 42-6

Page 491: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Mortara ELI 100/STM

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the WAVEFORM SIZE key, the SWEEP SPEED key, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Enabling the ST Values on the DisplayThe ST Values screen can be enabled or disabled. Use the ST VALUES ON/OFF key to enable or disable the ST Values screen.

The default setting is ST VALUES ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch ST.• Touch SETUP.• Touch WAVEFORM ON/OFF.

To enable the ST values screen:

• Touch ST.• Touch SETUP.• Touch ST VALUES ON/OFF.

To change waveform size:

• Touch ST.• Touch WAVEFORM SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 42-7

Page 492: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Mortara ELI 100/STM

Selecting a Rhythm Lead WaveformYou can select one of the twelve waveforms for display.

The default depends upon the settings on the ST monitor.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select a rhythm lead:

• Touch ST.• Touch RHYTHM LEAD.• Select desired lead.

To print the ST monitor display:

• Touch ST.• Touch PRINT.

Flexport System Interface Operations Manual 42-8

Page 493: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Mortara ELI 100/STM

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ST monitor or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ST monitor is powered ON.

• Verify that all cables are secure.

• Verify that the ST monitor has monitoring enabled.

Table 2 lists the status messages that the Flexport system interface reports for the ST monitor during Flexport system interface operation.

Note:If the ST monitor’s directory fills to capacity, the ST monitor will stop sending data to the Flexport system interface and the ST monitor’s parameter zone will drop from the Spacelabs Healthcare monitor.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ST monitor and the Spacelabs Healthcare monitor has been interrupted. Yes

LEAD FAULT A lead(s) failure in the standard lead set — correct the faulty leads. Yes

NOISY SIGNALExcessive high frequency noise on the baseline (muscle tremor) — check on patient preparation and/or patient movement.

Yes

NO QRS DETECTION

No QRS complexes were detected in the 10 second record — correct the condition causing the invalid ECG. Yes

Flexport System Interface Operations Manual 42-9

Page 494: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 495: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Newport Medical Instruments E150/E200

Directory of Keys

VENT

VENT - MAIN MENU

ALARM LIMITS SETUP VIEW

ALARMS PRINT

VENT - SETUP LARGE TXT

SMALL TXT

VENT - ALARM LIMITS

VMIN RR PIP MAP

VENT - ALARM LIMITS - MAPMAP HI =

XXLO = XXON OFF

VENT - ALARM LIMITS - PIPPIP HI =

XXLO = XXON OFF

VENT - ALARM LIMITS - RRRR HI =

XXLO = XXON OFF

VENT - ALARM LIMITS - VMINVMIN HI =

XXLO = XXON OFF

Flexport System Interface Operations Manual 43-1

Page 496: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 497: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Newport Medical Instruments E150/E200

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from Newport Medical Instruments E150 and E200 ventilators to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate TransitionConnector Part Number Minimum Software Revision

9600 Provided by Newport Medical Instruments

E150: V1.5E200: Main - R5.17B; Panel - R5.10BSL 300: V1.3

Flexport System Interface Operations Manual 43-3

Page 498: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Newport Medical Instruments E150/E200

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of a modular cable into the modular jack on the Flexport system interface.• Plug the other end of the modular cable into the modular jack labeled “FLEXPORT” on

the SL 300 interface adapter.• Plug one end of another modular cable into the modular jack labeled “VENT” on the

SL 300 interface adapter.• Plug the other end of the modular cable into the transition connector on either the E150

or E200 ventilator.• Plug the transition connector into the connector labeled “RS232” on the ventilator.• Connect the wire from the Newport Medical AC adapter to the connector labeled

“POWER” on the SL 300 interface adapter and plug the AC adapter into a wall outlet.• Turn ON the ventilator, and then turn ON the SL 300 interface adapter. Note that the

SL 300 will sound a continuous alarm for up to one minute after the power is turned ON (until it receives a data packet from the ventilator, which comes in one-minute intervals).

Flexport System Interface Operations Manual 43-4

Page 499: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Newport Medical Instruments E150/E200

Figure 43-1: Flexport system interface connections

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON an Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 43-2: Bedside monitor, large text screen; Central monitor, full screen

SDLCcable

SDLCterminator

Newport

Newport

WAVEE200

Ventilator

BREEZEE150

Ventilator

NewportMedicalInstrumentsSL 300

adapter

OR

modularcables

25-pin transitionconnector

9-pin transitionconnector

connectionlabeled“VENT”

connectionlabeled“FLEXPORT”

connectionlabeled“POWER”

Newport MedicalInstruments AC adapter

interface

SpacelabsHealthcare monitor

systemFlexport

interface

� � � � �

5 PIP 40 O2 4.2 VMIN VENT

0.28 VT

cmH2O % L L

2 MAP 1 PEEP 15 RR

cmH2O cmH2O BPM

� �

Flexport System Interface Operations Manual 43-5

Page 500: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Newport Medical Instruments E150/E200

Figure 43-3: Bedside monitor, small text screen

Figure 43-4: Central monitor, split screen

� Peak inspiratory pressure — displays high pressure alarm

� FiO2 — sets oxygen concentration

� Inspiratory minute volume — displays low and high minute volume alarms (Vmin is not displayed on the BREEZE 150)

� VENT parameter key

� E200: inspiratory tidal volume shows the larger of mandatory or voluntary tidal volumeE150: set tidal volume is shown

� Alarm status bell (appears when alarms are turned ON)

� Respiratory rate — displays apnea alarm

Positive end expiratory pressure — displays the baseline pressure

Mean airway pressure — displays low pressure alarm

� Patient/room ID

� � �� �

VENT

VT= 0.28L

RR= 15BPM

PIP= 5cmH2O

MAP= 2cmH2O

� � � �

� VT=0.28L O2= 40% VENT

� RR = 15bpm PEEP= 1cmH2O

� PIP= 5 cmH2O VMIN= 4.2L

MAP= 2cmH2O

� ICU 14

Flexport System Interface Operations Manual 43-6

Page 501: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Newport Medical Instruments E150/E200

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 43-7

Page 502: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Newport Medical Instruments E150/E200

Viewing Alarm SettingsThe VIEW ALARMS key normally enables you to display the current alarm settings for all parameters. However, this data is not currently available from the ventilator, and dashes are shown instead.

If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 43-5: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 43-5).

The default setting is LARGE TXT.

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To select a display format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

VENT

0.25 VT

VMIN PIP L

26 RR

RR MAP BPM

..

Flexport System Interface Operations Manual 43-8

Page 503: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Newport Medical Instruments E150/E200

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem between the Flexport system interface and the Spacelabs Healthcare interface adapter. The message VENTILATOR LINK LOST indicates a problem between the SL 300 and the ventilator. If the monitor displays either message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

UFP Parameter KeyIf the letters UFP are displayed on the parameter key instead of VENT, the Flexport system interface has entered a self-test mode. This can happen if power is applied to the Flexport system interface (via the SDLC cable) before power is applied to the Newport Medical Instruments SL 300 interface adapter. This can be corrected as follows:

• Apply power to the SL 300 interface adapter.

• Cycle the power to the Flexport system interface by unplugging and plugging in the SDLC cable.

To print the current display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 43-9

Page 504: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 505: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 44-1

Oridion VitalCap

Directory of KeysCO2

CO2 - MAIN MENUALARM LIMITS SETUP

FREEZE TOGGLESILENCE

VIEW ALARMS PRINT

ON OFF

CO2 - SETUP

SIZE SWEEPSPEED

LARGE TXTSMALL TXT

CO2 - SETUP - SIZE CO2 - SETUP - SWEEP SPEEDWAVEFORM SIZE

↑SIZE

↓25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/secON OFF

CO2 - ALARM LIMITS

ETCO2 RR FlCO2

CO2 - ALARM LIMITS - FlCO2FlCO2 HI =

XXXON OFF

CO2 - ALARM LIMITS - RRRR HI =

XXXLO = XXXON OFF

CO2 - ALARM LIMITS - ETCO2ETCO2 HI =

XXXLO = XXXON OFF

These keys appear on the bedside monitor only.

Page 506: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 507: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Silencing Alarm Tones . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90442A Flexport system interface enables information from an Oridion VitalCap portable bedside capnograph to be displayed on Spacelabs Healthcare monitors.

Configure the capnograph as shown in Table 1. Consult the capnograph manual for additional configuration setup procedures.

Table 1: Capnograph Configuration

Baud Rate Minimum Software Revision Installation Kit

Part Number

9600 2.75 045-0155-xx

Flexport System Interface Operations Manual 44-3

Page 508: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the capnograph, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the capnograph. To control other interface functions, refer to the sections that follow.

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Connect the other end of the modular cable (DB25 male) to the VitalCap Interface Adapter.NOTE: In the package, one end of the RJ45 monitor cable is connected to the VitalCap Interface Adapter.

• Connect the loose end of the RJ45 cable to the capnograph.• Turn on the capnograph.

Flexport System Interface Operations Manual 44-4

Page 509: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the capnograph.

Figure 44-1: Bedside monitor, large text screen

Figure 44-2: Bedside monitor, small text screen

Figure 44-3: Central monitor, split screen

� CO2 parameter key

� Current end-tidal CO2 (EtCO2) value

� Alarm status bell (appears when alarms are turned ON)

� Current respiratory rate (RR) value

� Current fractional inspired CO2 (FiCO2) value

� �

CO2

30 EtCO2

mmHg

20RR

bpm

� �

� �

CO2

EtCO2 = 30mmHgRR = 20bpm �FiCO2 = 0.0mmHg �

� �

EtCO25015 FiCO2

8CO2

EtCO2 = 30mmHgRR = 20bpm �

RR30 FiCO2 = 0.0mmHg �6

Flexport System Interface Operations Manual 44-5

Page 510: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the CO2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the CO2 key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the capnograph.

• Alarm limits can only be adjusted at the capnograph. Alarm violation is detected at the capnograph, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• The No Breath alarm replaces all other alarms and is indicated by low RR (RR flashes).

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch CO2.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 44-6

Page 511: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Silencing Alarm TonesThere are two types of alarm silence for the capnograph: temporary (lasting one or two seconds) and constant.

Temporary silence is enabled or disabled by touching the capnograph’s SILENCE key or by touching the TOGGLE SILENCE key on the Spacelabs Healthcare monitor.

Constant silence is set from the Alarm Silence menu on the capnograph. When the constant silence is enabled, the temporary silence has no effect because the constant silence has a higher priority. If you cancel the constant silence while the temporary silence is active, the alarms will remain silent until the time-out of the temporary silence.

Regardless of how the alarms are silenced:

• Audible tones on the capnograph and the Spacelabs Healthcare monitors are silenced.

• Alarm recordings and alarm watches are disabled on the Spacelabs Healthcare monitors.

• The flashing of the CO2 key and the large bell on the Spacelabs Healthcare monitors ceases.

• If the parameter is displayed as a numeric key, no indication of an alarm is displayed.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the capnograph’s parameters (refer to Figure 44-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays CO2 ALM OFF to the right of the CO2 key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 44-4: View Alarms screen

Note:The VIEW ALARMS key is not available on the central monitor, split screen.

To view alarm settings:

• Touch CO2.• Touch VIEW ALARMS.

EtCO25015 FiCO2 8 C

O2

30 EtCO2

mmHg

RR 306 20 RR

bpm

Flexport System Interface Operations Manual 44-7

Page 512: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:• The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

• If the capnograph is set up for “Small CO2 Scale” (either 0 to 50 mmHg or 0 to 7 kPa/Vol%), two sizes are available.

• If the capnograph is set up for “Large CO2 Scale” (either 0 to 99 mmHg or 0 to 14 kPa/Vol%), only one size is available.

To turn the waveform display ON or OFF:

• Touch CO2.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch CO2.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 44-8

Page 513: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 25, 12.5, 6.25 or 3.12 mm/second.

The default setting is 12.5 mm/second for neonatal patients and 6.25 mm/second for adult patients.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 44-5).

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch CO2.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch CO2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch CO2.• Select FREEZE ON.

Flexport System Interface Operations Manual 44-9

Page 514: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter on the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To print the current display:

• Touch CO2.• Touch PRINT.

Flexport System Interface Operations Manual 44-10

Page 515: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Oridion VitalCap

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the capnograph or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the capnograph is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the capnograph during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

NO BREATH No valid breath has been detected.

BATTERY EMPTYBattery pack is discharged. Replace or recharge the battery, or connect the VitalCap monitor to AC power.

STANDBY Standby mode. To monitor patient, exit standby mode.

CO2 FAILURE Instrument fault. Contact authorized service representative.

BATTERY LOWBattery charge level is low. Replace or recharge the battery, or connect the VitalCap monitor to AC power.

FILTERLINEFilter line is disconnected or is not securely connected. Connect the filter line to the input connector on the VitalCap or tighten the connection.

BLOCKAGE

Filter line is twisted or clogged. The message appears after 30 seconds of unsuccessful clearing. Check the filter line and, if necessary, replace it.

Microstream airway adapter is clogged. Check the Microstream airway adapter and, if necessary, replace it.

RECALIBRATE CO2 Needs calibration. Recalibrate the VitalCap.

CLEARING FILTERLINEFilter line tube is twisted or clogged with water. Check the filter line and, if necessary, untwist or replace it.

Flexport System Interface Operations Manual 44-11

Page 516: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 517: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 45-1

OSI OpticalCAM

Directory of KeysABG

ABG - MAIN MENU

SETUP PRINT

CO2 - SETUP LARGE TXTSMALL TXT

Page 518: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 519: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

OSI OpticalCAM

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Selecting a Unit of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

OverviewThe 90442A Flexport system interface allows information from an Optical Sensors Incorporated (OSI) OpticalCAM Blood Measurement System to be displayed on Spacelabs Healthcare monitors.

Configure the blood gas analyzer as shown in Table 1. Consult the analyzer manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the blood gas analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the blood gas analyzer. To control other interface functions, refer to the sections that follow.

Table 1: Analyzer Configuration

Baud Rate Transition Connector Part Number

9600 131-0685-xx

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the monitoring port on the back of the blood gas

analyzer.

Flexport System Interface Operations Manual 45-3

Page 520: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

OSI OpticalCAM

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the blood gas analyzer.

Figure 45-1: Bedside monitor, large text screen

Figure 45-2: Bedside monitor, small text screen

� Time value from the OpticalCAM monitor time clock (hours and minutes) — indicates time at which the ABG was measured.

Note:The times displayed in the tabular or graphic trend displays of the Spacelabs Healthcare monitor are set by the Spacelabs Healthcare clock.

� Base excess

� PH

� ABG parameter key

� Partial pressure of dissolved carbon dioxide in arterial blood

� Partial pressure of dissolved oxygen in arterial blood

� Bicarbonate

Oxygen saturation

Total carbon dioxide

� � � � �

1642 Time -1 BE

mEq/L 7.38 pH

ABG

40 PCO2

mmHg

25.2TCO2

mmol 97SaO2

%24 HCO3

mEq/L 98PO2

mmHg

� �

� �

ABG

pH = 7.38

PCO2 = 40mmHg �PO2 = 98mmHg �Time = 1642 �

Flexport System Interface Operations Manual 45-4

Page 521: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

OSI OpticalCAM

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 45-4).

The default setting is LARGE TXT.

Selecting a Unit of MeasurementThe PCO2 and PO2 values can be displayed in units of millimeters of mercury (mmHg) or kilo Pascals (kPa). Complete the steps in the following Quickstart.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch ABG.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To select the unit of measurement from the OpticalCAM monitor:

• Power up the blood gas analyzer.• Touch SETUP MENU.• Select SERVICE.• Enter your password.• Select UNITS OF MEASURE.• Use the control knob to select mmHg or kPa.

To print the current display:

• Touch ABG.• Touch PRINT.

Flexport System Interface Operations Manual 45-5

Page 522: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

OSI OpticalCAM

Status MessagesNote:

If an OpticalCAM monitor error occurs, a general error message will be displayed on the Spacelabs Healthcare monitor. The error message will refer you to the blood gas analyzer for more specific information.

The message COMMUNICATION LINK LOST or ABG LINK LOST indicates a problem with the blood gas analyzer or the cabling. If the monitor displays either of these messages, perform the following troubleshooting steps:

• Verify that the blood gas analyzer is powered ON.

• Verify that all cables are secure.

Flexport System Interface Operations Manual 45-6

Page 523: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 46-1

PhysioFlex

Directory of KeysVENT

VENT - MAIN MENUALARM LIMITS SETUP

FREEZE SWITCHWAVES

VIEW ALARMS

*PRINT

ON OFF

VENT - SETUP * These keys are displayed on the bedside monitor only.

SIZE SWEEPSPEED

LARGE TXT *SMALL TXT

VENT - SETUP - SIZE VENT - SETUP - SWEEP SPEEDWAVEFORM SIZE

↑SIZE

↓12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/secON OFF

VENT - ALARM LIMITS

FiAn FiCO2

VENT - ALARM LIMITS - FiO2FiO2 HI =

XXLO = XX

FiO2ON OFF FiAn

VENT - ALARM LIMITS - FiAnFiAn HI =

XXXFiAn

ON OFF FiO2

Page 524: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 525: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PhysioFlex

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a PhysioFlex anesthesia ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

In order to communicate with Spacelabs Healthcare equipment, the ventilator should be equipped with a Spacelabs Healthcare communications option. Check the Options screen on the ventilator monitor to verify that the Spacelabs Healthcare communications option is present. The System Information box on the ventilator monitor should contain the following line:

Data protocol: SL X,YY (where X, the major revision number, should be 1 or higher and YY can be any number)

The ventilator needs no further configuration.

Table 1: Ventilator Configuration

Baud Rate Minimum Software Revision

InstallationKit Part Number

9600 6.04.01 045-0035-xx (Spacelabs)5501 5001 (Physio Medical)

Flexport System Interface Operations Manual 46-3

Page 526: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PhysioFlex

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 46-1: Bedside monitor, large text screen

Figure 46-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the ventilator, at the backside of the monitor DOME.

� � � � � �

446 Vt

ml 11 RR

BPM 1.1 FiAn

% VENT

31 FiO2

%

218VO2

ml pm 22Ppeak

cmH2O 1.0 FetAn

% 7.9etCO2

%

� �

� � � �

VENT

FiO2 = 31%etCO2 = 7.9%FetAn = 1.0%

VO2 = 218 ml pm �

Flexport System Interface Operations Manual 46-4

Page 527: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PhysioFlex

Figure 46-3: Central monitor, split screen

� Current tidal volume in milliliters

� Current respiration rate in breaths per minute

� Current inspired anesthetic agent concentration in %

� VENT parameter key

� Current inspired oxygen concentration in %

� Alarm status bell (appears when alarms are turned ON)

� Current end tidal CO2 concentration (units as selected on ventilator)

Current end tidal anesthetic agent concentration

Current peak airway pressure in cmH2O

� Current oxygen fresh gas flow in milliliter per min

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the VENT key to flash.

Alarm settings default to ON.

� � �

� FiO2 = 31% RR = 11BPM VENT

� etCO2 = 7.9% Ppeak = 22cmH2O

� Vt = 446ml FiAN = 1.1%

� VO2 = 218ml pm FetAn = 1.0%

� ICU 14

Flexport System Interface Operations Manual 46-5

Page 528: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PhysioFlex

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits can only be adjusted at the ventilator. Alarm violation is detected at the ventilator and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 46-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 46-4: View Alarms screen

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

FiAn 1.1 VENT

FiO2 = 31%etCO2 = 7.9%

FiO2 100 FetAn = 1.0%18.0 VO2 = 218 ml pm

ICU 14

Flexport System Interface Operations Manual 46-6

Page 529: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PhysioFlex

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key accesses the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch VENT.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch VENT.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 46-7

Page 530: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PhysioFlex

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 46-4).

The default setting is LARGE TXT.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, and 12.5 mm/second.

The default sweep speed is 6.25 mm/second.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a display format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To select a sweep speed:

• Touch VENT.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To freeze the waveform on the display:

• Touch VENT.• Select FREEZE ON.

Flexport System Interface Operations Manual 46-8

Page 531: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PhysioFlex

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport system interface operation.

To print the current display:

• Touch VENT.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the Flexport system interface and the ventilator is interrupted. Yes

CO2 WAVE SELECTED The SWITCH WAVES key was touched and the CO2 waveform is presently displayed. No

PAW WAVE SELECTED The SWITCH WAVES key was touched and the airway pressure waveform is presently displayed. No

Flexport System Interface Operations Manual 46-9

Page 532: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 533: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PiCCO

Directory of Keys

PiCCO

PiCCO - MAIN MENULARGE NUM

PRINTSMALL NUM

Flexport System Interface Operations Manual 47-1

Page 534: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 535: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PiCCO

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

OverviewThe 90439 Flexport system interface enables continuous cardiac output (CCO) information from a Pulsion PiCCO plus to be displayed on Spacelabs Healthcare monitors. Consult the manufacturer's manual for configuration of the PiCCO device.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the PiCCO device, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the PiCCO device. To control other interface functions, refer to the sections that follow.

Table 1: PiCCO Configuration

Baud Rate

Minimum Software Revision

Transition Connector Part Number

19200 PiCCO 7.x 131-1399-xx

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 9-pin connector on the PiCCO monitor.

Flexport System Interface Operations Manual 47-3

Page 536: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PiCCO

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the PiCCO device.

Figure 47-1: Bedside monitor, small-text screen

Figure 47-2: Bedside monitor, large-text screen

Figure 47-3: Central monitor, split screen

� PiCCO parameter key

� Pulse contour cardiac index

� Global end-diastolic volume index

� � � �

PiCCO

PCCI = 1.66GDVI = 715SVRI = 1563ELWI = 9

� � � � �

3 SVV 1185 dPmx 1563 SVRI PiCCO

1.66 PCCI

TD Values17:07 08.01.2007 37 9 715GEF ELWI GDVI

� � � �

� �

� PCCI = 1.66 GDVI = 715 PiCCO

� SVRI = 1563 ELWI = 9� dPmx = 1185 GEF = 37� SVV = 8 17:07 08.01.2007

� �

Flexport System Interface Operations Manual 47-4

Page 537: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PiCCO

� Systemic vascular resistance index

� Extravascular lung water index

� Index of left ventricular contractility

� Stroke volume variation

Timestamp of last TD measurement

Global ejection fraction

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 47-4). The default setting is SMALL NUM.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch PiCCO.• Select LARGE NUM or SMALL NUM.

To Print the display:

• Touch PiCCO.• Touch PRINT.

Flexport System Interface Operations Manual 47-5

Page 538: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

PiCCO

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the PiCCO device or the cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the PiCCO device is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the PiCCO device during Flexport operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the Spacelabs Healthcare monitor and the CCO monitor has been interrupted.

YES

INJECTION ERROR Injection error on PiCCO device, check PiCCO device NO

INJECTATE TEMPERATURE SENSOR ERROR

Injectate temperature sensor error on PiCCO device, check PiCCO device

NO

INJECT FASTER THAN 10SEC Inject faster than 10 sec on PiCCO device, check PiCCO device NO

Dst (DOWNSLOPE TIME ERROR) Downslope time error on PiCCO device, check PiCCO device NO

BASELINE ERROR Baseline error on PiCCO device, check PiCCO device NO

MTt (MEAN TRANSIT TIME) CALCULATION ERROR

Mean time calculation error on PiCCO device, check PiCCO device

NO

TIME OUT (THERMODILUTION CURVE LONGER THAN 90S)

Time out on PiCCO device, check PiCCO device NO

TINJECTION > TBLOOD Tinjection > Tblood on PiCCO device, check PiCCO device NO

INVALID PCCO CALIBRATION Invalid PCCO calibration on PiCCO device, check PiCCO device NO

Flexport System Interface Operations Manual 47-6

Page 539: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 48-1

Radiometer TCM3

Directory of KeysTCP

TCP - MAIN MENUALARM LIMITS SETUP VIEW

ALARMS PRINT

TCP - SETUP *LARGE TXT *These keys appear on the bedside monitor only.SMALL TXT

TCP - ALARM LIMITS

PO2 PCO2 ETEMP TIMER

TCP - ALARM LIMITS - TIMERTIMER LO =

XX ↑ ↓ETEMP

ON OFF TIMER

TCP - ALARM LIMITS - ETEMPETEMP HI =

XXLO = XX ↑ ↓

ETEMPON OFF TIMER

TCP - ALARM LIMITS - PCO2PCO2 HI =

XXLO = XX ↑ ↓

PO2ON OFF PCO2

TCP - ALARM LIMITS - PO2PO2 HI =

XXLO = XX ↑ ↓

PO2ON OFF PCO2

Page 540: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 541: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Radiometer TCM3

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

OverviewThe 90438A-01 Flexport system interface enables information from a Radiometer TCM3 Transcutaneous Pressure Analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the analyzer as shown in Table 1. Consult the analyzer manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

Table 1: Analyzer Configuration

BaudRate

Character Length Parity Setting Stop Bit

Transition Connector

Part Number

2700 7-bit odd 1 131-0789-xx (25-pin)

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the analyzer.

Flexport System Interface Operations Manual 48-3

Page 542: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Radiometer TCM3

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to the Flexport system interface, the monitor provides the following information from the analyzer.

Figure 48-1: Bedside monitor, large text screen

Figure 48-2: Bedside monitor, small text screen

Figure 48-3: Central monitor, split screen

� � � �

42.8 ETEMP

ºC4.00 TIMER

H.MIN TCP

157 PO2

mmHg

8 PCO2

mmHg

� �

� � � �

TCP

PO2 = 157mmHgPCO2 = 8 mmHgETEMP = 42.8 ºCTIMER = 4.00 H.MIN

� �

� �

� PO2 = 157mmHg TCP

� PCO2 = 8 mmHg

� ETEMP = 42.8 ºC

� TIMER = 4.00 H.MIN

� ICU - 4

Flexport System Interface Operations Manual 48-4

Page 543: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Radiometer TCM3

� Electrode (sensor) temperature measured in °C

� Site timer (time remaining since the user set the analyzer up for the length of time to remain on site)

� TCP parameter key

� Transcutaneous oxygen tension (PtcO2)

� Alarm status bell (appears when alarms are turned ON)

� Transcutaneous carbon dioxide tension (PtcCO2)

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the TCP key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch TCP.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To adjust alarm limits:

• Touch TCP.• Select ALARM LIMITS.• Select an alarm parameter.• Select HI= or LO=.• Use the arrow keys to adjust.

Flexport System Interface Operations Manual 48-5

Page 544: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Radiometer TCM3

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 48-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays TCP ALM OFF to the right of the TCP key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 48-4: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 48-4).

The default setting is LARGE TXT.

To view alarm settings:

• Touch TCP.• Touch VIEW ALARMS.

To select a display format:

• Touch TCP.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

PO2169

ETEMP43.3

TCP

157 PO2

mmHg60 42.7

PCO245 TIMER 8 PCO2

mmHg5 0.01

Flexport System Interface Operations Manual 48-6

Page 545: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Radiometer TCM3

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or the cabling. If the Spacelabs Healthcare monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport operation.

To print the current display:

• Touch TCP.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOSTCommunication between the Flexport system interface and the analyzer has been interrupted.

Yes

SERVICE REQUIRED TCM3 electrode is defective or rejected by the analyzer. Monitor is defective. No

CALIBRATING TCM3 calibration is occurring. No

Flexport System Interface Operations Manual 48-7

Page 546: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 547: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

Directory of KeysVENT

VENT - MENUALARM LIMITS SETUP VIEW

MODEVIEW

STATUSVIEW

ALARMS PRINT

Refer to page 49-2

VENT - SETUPLARGE TXTSMALL TXT

Flexport System Interface Operations Manual 49-1

Page 548: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

VENT

VENT - MENUALARM LIMITS SETUP VIEW

MODEVIEW

STATUSVIEW

ALARMS PRINT

Refer to page 49-1

VENT - ALARM LIMITS

RR VT Vmin PIP PEEP I : E APNEA SpO2*

VENT - ALARM LIMITS - SpO2SpO2 HI =

XXLO = XX

RRON OFF SpO2

VENT - ALARM LIMITS - APNEAAPNEA HI =

XXAPNEA

ON OFF SpO2

VENT - ALARM LIMITS - I : EI : E LO =

X.XI : E

ON OFF APNEA

VENT - ALARM LIMITS - PEEPPEEP LO =

XXPEEP

ON OFF I : E

VENT - ALARM LIMITS - PIPPIP HI =

XXXLO = XXX

PIPON OFF PEEP

VENT - ALARM LIMITS - VminVmin LO =

XX.XVmin

ON OFF PIP

VENT - ALARM LIMITS - VTVT LO =

X.XXVT

ON OFF Vmin

VENT - ALARM LIMITS - RRRR HI =

XXRR

ON OFF VT

*Only SpO2 alarm limit is displayed when activated.

Flexport System Interface Operations Manual 49-2

Page 549: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90436A-01 Flexport system interface enables information from a Respironics Novametrix Esprit ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Note:SpO2 monitoring is not supported on the Respironics Esprit ventilator. The SpO2 parameter will display with question marks (???) when the Flexport is configured to display large text.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting Stop Bit

TransitionConnector Part

Number

Minimum Software Revision

9600 8-bit none 1 131-1589-00 26300 Rev. E or higher

Flexport System Interface Operations Manual 49-3

Page 550: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Figure 49-1: Flexport system interface connections

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the DIGITAL OUTPUT connector on the ventilator.

SDLCcable

SDLCterminator

modularcable

Flexport

interface

ventilator

9-pin connectorto DIGITAL OUTPUT(back of ventilator)

SpacelabsHealthcare

monitor

system

Flexport System Interface Operations Manual 49-4

Page 551: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 49-2: Bedside monitor, large text screen

Figure 49-3: Bedside monitor, small text screen

Figure 49-4: Central monitor, split screen

� � � � � �

1.8I:E

21PIP

cmH2O 0PLT

cmH2O 21FiO2

% VENT

1.22VT

L

???SpO2

% 1PEEP

cmH2O8 MAP

cmH2O30.0 Vmin

L 27RR

BPM

� � � � �

� � � �

VENT

VT=1.22LRR= 27BPMPIP= 21cmH2OPEEP= 1cmH2O

� �

� � � � �

� RR= 27BPM VT= 1.22L VENT

� FiO2 = 21% PIP= 21cmH2O

� PEEP= 1cmH2O SpO2= ???

� Vmin= 30.0L MAP= 8cmH2O

� ICU 14

Flexport System Interface Operations Manual 49-5

Page 552: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

� Ratio of exhalation time to inspiratory time

� Peak inspiratory pressure

� Inspiratory plateau pressure

� Fractional inspired O2

� VENT parameter key

� Tidal volume

� Alarm status bell (appears when alarms are turned ON)

Respiratory rate

Minute volume

� Mean airway pressure

� Positive and expiratory pressure

O2 saturation

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to OFF.

Note:Alarms that are generated on the Respironics Esprit ventilator will be displayed on the Spacelabs monitor within seven seconds.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 49-6

Page 553: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits on the Flexport system interface can only be adjusted at the ventilator. Alarm violation is detected at the ventilator, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 49-5). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 49-5: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 49-9).

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

PIP 120 3 SpO2

0 90 VT 0.00 APNEA

20 VENT

1.22 VT

L

I : E 1.0 PEEP 0 Vmin 0.3 RR 70 27 RR

BPM

Flexport System Interface Operations Manual 49-7

Page 554: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

Viewing Ventilator Mode HistoryChronological history of the last eight ventilator mode settings are displayed in the waveform zone with the oldest value at the top of the left column and the most recent value at the bottom of the right column. Possible modes include CMV, SIMV, CPAP, SIMV/PS, CPAP/PS, SIMV/FB, CPAP/FB, or UNKNOWN MODE.

Figure 49-6: View Mode screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 49-5).

The default setting is LARGE TXT.

To view ventilator mode history:

• Touch VENT.• Touch VIEW MODE.

To select a display format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

15:4715:47

CMVSIMV/PS

15:4915:49

SIMV/PSCMV V

ENT

1.21 VT

L15:47 CMV 15:51 CPAP/PS 25 RR

15:49 CMV 15:56 SIMV/PS BPM

Flexport System Interface Operations Manual 49-8

Page 555: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix Esprit

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport operation.

To print the current display:

• Touch VENT.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

LOW PRS O2 INLET Pressure at the oxygen inlet fitting is less than 35 psig, and the O2% is 22 or greater. No

LOW PRS AIR INLET Pressure at the air inlet is equal to or less than 35 psig. No

LOW BATTERY Voltage of the internal batteries is too low to sustain one hour of operation. No

EXHAL VALVE LEAK Volume of exhalation gas is less than 10% of inspiration gas. No

DISCONNECT VENT Patient service system is disconnected or plugged. No

Flexport System Interface Operations Manual 49-9

Page 556: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 557: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Directory of KeysFLOW

FLOW - MAIN MENUALARM LIMITS SETUP

FREEZE NEXTDISPLAY

VIEW ALARMS PRINT

ON OFF

These keys appear on the bedside monitor only.

FLOW - SETUP - SIZE

SIZE SWEEPSPEED

LARGE TXTSMALL TXT

FLOW - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

FLOW - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec

FLOW - ALARM LIMITS

RR

FLOW - ALARM LIMITS - RRRR HI =

XXLO = XXON OFF

Flexport System Interface Operations Manual 50-1

Page 558: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

PAW

PAW - MAIN MENU

SETUPFREEZE NEXT

DISPLAY PRINTON OFF

These keys appear on the

bedside monitor only.

PAW - SETUP - SIZE

SIZE SWEEPSPEED

LARGE TXTSMALL TXT

PAW - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

PAW - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec

Flexport System Interface Operations Manual 50-2

Page 559: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

CO2

CO2 - MAIN MENUALARM LIMITS SETUP

FREEZE NEXTDISPLAY

VIEW ALARMS PRINT

ON OFF

These keys appear on the bedside monitor only.

CO2 - SETUP - SIZE

SIZE SWEEPSPEED

LARGE TXTSMALL TXT

CO2 - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

CO2 - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec

CO2 - ALARM LIMITS

ETCO2 RR VCO2

CO2 - ALARM LIMITS - VCO2VCO2 HI =

XXLO = XXON OFF

CO2 - ALARM LIMITS - RRRR HI =

XXLO = XXON OFF

CO2 - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XXON OFF

Flexport System Interface Operations Manual 50-3

Page 560: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

NICO

NICO - MAIN MENUALARM LIMITS SETUP NEXT

DISPLAYVIEW

ALARMS PRINT

These keys appear on the bedside monitor only.

NICO - SETUPLARGE TXTSMALL TXT

NICO - ALARM LIMITS

C.O.

NICO - ALARM LIMITS - C.O.C.O. HI =

XXLO = XXON OFF

Flexport System Interface Operations Manual 50-4

Page 561: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

SPO2

SPO2 - MAIN MENUALARM LIMITS SETUP

FREEZE NEXTDISPLAY

VIEW ALARMS PRINT

ON OFF

These keys appear on the bedside monitor only.

SPO2 - SETUP - SIZE

SIZE SWEEPSPEED

LARGE TXTSMALL TXT

SPO2 - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

SPO2 - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec

SPO2 - ALARM LIMITS

SPO2 PR

SPO2 - ALARM LIMITS - PRPR HI =

XXLO = XXON OFF

SPO2 - ALARM LIMITS - SPO2SPO2 HI =

XXLO = XXON OFF

Flexport System Interface Operations Manual 50-5

Page 562: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 563: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Display Detail - FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Display Detail - PAW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Display Detail - CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Display Detail - NICO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Display Detail - SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

OverviewThe 90442A-03 Flexport system interface enables cardiorespiratory information from a Respironics-Novametrix Inc. Model 7300 Cardiopulmonary Management System (NICO) to be displayed on Spacelabs Healthcare monitors.

The displayed information is organized into four or five parameter sets (depending upon the mode of operation). The user can transition between these parameter sets by touching the NEXT DISPLAY key at the bedside monitor.

Note:Variables not included in the currently displayed parameter set are not trended, and their alarms are not available at the Spacelabs Healthcare monitor(s).

Configure the Novametrix Model 7300 as shown in Table 1.

Table 1: Model 7300 Configuration

Model Number

Baud Rate Minimum Software Revision

Installation Kit

Part Number

Transition Connector

Part Number

Respironics-Novametrix

73009600 NICO firmware version 5.0 045-0158-xx 131-0686-xx

Respironics-Novametrix

76009600 NICO 2 version 10.0 045-0158-xx 131-0686-xx

Flexport System Interface Operations Manual 50-7

Page 564: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Connecting the Flexport System InterfaceTo connect the Flexport system to the SDLC cable, complete the steps listed in the following Quickstart.

For information on configuring the Novametrix Model 7300 to communicate with the Flexport, refer to the Novametrix User’s Manual.

Figure 50-1: Flexport System Interface connections

The Spacelabs Healthcare monitor is now ready to display numerics. To control other interface functions, refer to the sections that follow.

To connect the Flexport system interface:

• Ensure that the Spacelabs Healthcare monitoring system is shut down (power is turned OFF).

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 50-1.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 9-pin serial port connector number 2 or 3 on the

Novametrix Model 7300 rear panel.• Power on the Novametrix Model 7300.

On the Novametrix Model 7300:• Press the MENU button and use the knob to select the SETUP menu and then the

INPUT/OUTPUT menu.• Select either RS232-2 or RS232-3 to match the port connection on the rear panel.• Use the knob to select FLEXPORT.• Press the knob to save.

• Configure the Novametrix Model 7300 for its intended use (for example, units of measurement or adult/neonatal mode of operations).

• Turn the power to the Spacelabs Healthcare monitoring system ON.

SDLC cable

SDLCterminator

modular cable

transition connector

Novametrix Model 730serial port 2 or 3

SpacelabsHealthcare

monitor

systemFlexport

interface

Flexport System Interface Operations Manual 50-8

Page 565: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Ideally, the Novametrix Model 7300/7600 should be powered ON and fully configured prior to connecting it to the Flexport system interface.

Once the Model 7300/7600 has been connected to the Flexport system interface and is providing data to the Spacelabs Healthcare monitoring system, if the configuration needs to change, complete the following procedure:

1 Turn OFF the Spacelabs Healthcare monitoring system.

2 Reconfigure the Novametrix Model 7300/7600 monitor as needed.

3 Turn the power to the Spacelabs Healthcare monitoring system ON.

Display Detail - FLOWWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitors display the displays shown below.

Figure 50-2: Bedside monitor, large text screen

Figure 50-3: Bedside monitor, small text screen

� � � �

6.1 MV

L FLOW

420Vti

mL

417 VtemL15 RR

br/m

� � � �

� � �

FLOW

Vti= 420mLVte= 417mLMV=6.1LRR= 15br/m

� � � �

Flexport System Interface Operations Manual 50-9

Page 566: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Figure 50-4: Central monitor, split screen

� Current minute volume expired (MV) value

� Alarm status bell (appears when alarms are turned ON)

� FLOW parameter key

� Current tidal volume inspired (Vti) value

� Current tidal volume expired (Vte) value

� Flow waveform

� Current respiratory rate (RR) value

Patient/room ID

Display Detail - PAWWhen you touch the NEXT DISPLAY key on the bedside monitor from the FLOW Parameter screen, the monitors display the screens shown below.

Figure 50-5: Bedside monitor, large text screen

� Vti= 420mL FLOW

� Vte= 417mL �� MV= 6.1L

� RR= 15br/m

ICU - 4

� � � �

27 Cdyn 9 MAP

cmH2O PAW

23PIP

cmH2O

4 PEEP

cmH2O15 Raw

� � �

Flexport System Interface Operations Manual 50-10

Page 567: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Figure 50-6: Bedside monitor, small text screen

Figure 50-7: Central monitor, split screen

� Current dynamic compliance (Cdyn) value

� Current mean airway pressure (MAP) value

� PAW parameter key

� Current peak inspiratory pressure (PIP) value

� Current positive end expiratory pressure (PEEP) value

� PAW waveform

� Current resistance expired (Raw) value

Patient/room ID

� �

PAW

PIP= 23cmH2OPEEP= 4cmH2O �Cdyn= 27 �Raw= 15 �

� �

� PIP= 23cmH2O MAP= 9cmH2O PAW

� PEEP= 4cmH2O� Cdyn= 27� Raw= 15 ICU - 4

Flexport System Interface Operations Manual 50-11

Page 568: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Display Detail - CO2

When you touch the NEXT DISPLAY key on the bedside monitor from the PAW Parameter screen, the monitors display the screens shown below.

Figure 50-8: Bedside monitor, large text screen

Figure 50-9: Bedside monitor, small text screen

Figure 50-10: Central monitor, split screen

� Current volumetric CO2 (VCO2) value

� Current dead space to tidal volume ratio (Vd/Vt) value

� CO2 parameter key

� Current end-tidal CO2 (EtCO2) value

� � � � �

170VCO2

mL/m1.5 Vd/Vt

CO2

34 ETCO2

mL

23PeCO2

mmHg15 RR

br/m

� � �

� �

CO2

ETCO2= 34mmHgRR= 15br/m �

VCO2= 170mL/m �

PeCO2= 23mmHg

� �

� � �

� ETCO2= 34mmHg Vd/Vt= 1.5 CO2

� RR= 15br/m� VCO2= 170mL/m PeCO2= 23mmHg ICU - 4

Flexport System Interface Operations Manual 50-12

Page 569: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

� Alarm status bell (appears when alarms are turned ON)

� Current respiratory rate (RR) value

� CO2 waveform

Current mixed expired CO2 (PeCO2) value

Patient/room ID

Note:• The VCO2 alarm is active only in Adult Respiratory Mechanics mode.

• The PeCO2 label changes to FeCO2 when units are changed to percent (%).

Display Detail - NICOWhen you touch the NEXT DISPLAY key on the bedside monitor from the CO2 Parameter screen, the monitors display the screens shown below. This screen will not display if the Novametrix Model 7300 is in the respiratory mechanics mode. The SPO2 screens will display instead.

Figure 50-11: Bedside monitor, large text screen

Figure 50-12: Bedside monitor, small text screen

� � � � � �

82SV 2.1Cl 4.6PCBF NICO

5.3 C.O.

L/minmL L/min

� � �� �

NICO

C.O. =5.3L/minPCBF=4.6L/minCl=2.1 �

SV=82mL

Flexport System Interface Operations Manual 50-13

Page 570: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Figure 50-13: Central monitor, split screen

� Current stroke volume (SV) value

� Current cardiac index (CI) value

� Current pulmonary capillary blood flow (PCBF) value

� NICO parameter key

� Current cardiac output (C.O.) value

Note:This is the average cardiac output, irrespective of C.O. Mode on the NICO monitor.

� Alarm status bell (appears when alarms are turned ON)

� Patient/room ID

Display Detail - SpO2

When you touch NEXT DISPLAY on the bedside monitor from the CO2 (if in respiratory mechanics mode) or NICO Parameter screen, the monitors display the screens shown below.

Figure 50-14: Bedside monitor, large text screen

� � � �

� C.O.= 5.3L/min Cl= 2.1 NICO

� SV= 82mL PCBF= 4.6L/min

� ICU - 4

� � �

SPO2

98 SpO2

%

70 PR

bpm

� � �

Flexport System Interface Operations Manual 50-14

Page 571: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Figure 50-15: Bedside monitor, small text screen

Figure 50-16: Central monitor, split screen display

� SPO2 parameter key

� Current oxygen saturation (SpO2) value

� Alarm status bell (appears when alarms are turned ON)

� Current pulse rate (PR) value

� SpO2 waveform

� Patient/room ID

� �

SPO2

SpO2= 98%

PR= 70bpm

� � �

� �

� SpO2 = 98% SPO2

� PR = 70bpm

� ICU - 4�

Flexport System Interface Operations Manual 50-15

Page 572: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the main parameter key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to ON.

Note:• Alarm limits on the Flexport system interface can only be adjusted at the Novametrix Model 7300.

• Turning Flexport system interface alarms ON or OFF at the Spacelabs Healthcare monitor(s) does not affect alarms on the Novametrix Model 7300.

• Alarm violation is detected at the Novametrix Model 7300, and the Flexport system interface reports the alarm if it is turned ON for that parameter. When the alarming parameter is enabled and displayed on the currently active display or channel, the alarm response described above takes place. When the alarming parameter is enabled and NOT displayed on the currently active display or channel, the Spacelabs Healthcare monitor(s) displays an ALARM VIOLATION status message in the waveform zone.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch the main parameter key.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 50-16

Page 573: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for respiration rate (RR), EtCO2, VCO2, CO, SpO2, and pulse rate (refer to Figure 50-17). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays ALM OFF to the right of the parameter key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 50-17: View Alarms screen, SpO2 example

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON/OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SIZE ↑ and SIZE ↓ keys, the SWEEP SPEED key, and the FREEZE ON/OFF keys are disabled.

The default setting is WAVEFORM ON.

To view alarm settings:

• Touch the main parameter key.• Touch VIEW ALARMS.

To turn the waveform display ON or OFF:

• Touch the main parameter key.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

SpO2100 S

PO2

98 SpO2

%85

PR180

40 70 PR

bpm

Flexport System Interface Operations Manual 50-17

Page 574: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 25, 12.5, and 6.25 mm/second.

The default setting is 25 mm/second for SpO2 and 12.5 mm/second for all others.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to the Display Detail sections on pages page 50-9 through page 50-14).

The default setting is LARGE TEXT.

To adjust waveform size:

• Touch the main parameter key.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch the main parameter key.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch the main parameter key.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 50-18

Page 575: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the Novametrix Model 7300 or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the Novametrix Model 7300 is powered ON.

• Verify that all cables are secure.

The Novametrix Model 7300 sends six different status messages to the Flexport, depending upon the parameter set displayed. There are no visible or audible alarm indications associated with the display of these messages.

To freeze the waveform on the display:

• Touch the main parameter key.• Select FREEZE ON.

To print the current display:

• Touch the main parameter key.• Touch PRINT.

Flexport System Interface Operations Manual 50-19

Page 576: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics-Novametrix 7300/7600

Table 2 lists the status messages that the Flexport system interface reports for the Novametrix Model 7300 during Flexport system interface operation.

Note:Additional messages display on the Novametrix Model 7300 that are not sent to the Flexport for display on the Spacelabs Healthcare monitor(s).

Table 2: Status Messages

Message Text Parameter Set Display Zone Indication

FLOW DISCONNECT FLOW, PAW, NICO Waveform Flow sensor is disconnected.

CO2 DISCONNECT CO2, NICO Waveform CO2 sensor is disconnected.

SPO2 DISCONNECT SpO2 Waveform SpO2 sensor is disconnected.

CHECK NICO MESSAGE All Prompt

An informational message (other than Alert Limit Violations) is active at the Model 7300.

DEMO MODE DATA All Waveform Model 7300 is in DEMO mode.

ALARM VIOLATION All WaveformA parameter not in the current parameter set is in Alarm state.

Flexport System Interface Operations Manual 50-20

Page 577: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 840/860

Directory of Keys

TCP

TCP - MAIN MENU

ALARM LIMITS SETUP VIEW

ALARMS PRINT

TCP - SETUP LARGE TXT

SMALL TXT

TCP - ALARM LIMITS

PCO2 PO2 ETEMP

TCP - ALARM LIMITS - ETEMPETEMP HI =

X.XLO = X.X ↑ ↓

ON OFF

TCP - ALARM LIMITS - PO2PO2 HI =

X.XLO = X.X ↑ ↓

PCO2

ON OFF PO2

TCP - ALARM LIMITS - PCO2PCO2 HI =

X.XLO = X.X ↑ ↓

PCO2

ON OFF PO2

Flexport System Interface Operations Manual 51-1

Page 578: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 579: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 840/860

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

OverviewThe 90438A-02 Flexport system interface enables information from a Respironics Novametrix Transcutaneous Pressure (TCP) Analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the TCP analyzer as shown in Table 1. Verify that the RS-232 serial port of the TCP analyzer is configured for Terminal Mode operation. Once in this mode, MODE 0 must be selected. Refer to the TCP analyzer manual for additional configuration setup procedures.

Table 1: TCP Analyzer Configuration

Novametrix Model Number

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Respironics Novametrix 840 9600 8 bit none 1 131-0687-xx

(25-pin)

Respironics Novametrix 860 9600 8 bit none 1 131-0946-xx

(25-pin)

Flexport System Interface Operations Manual 51-3

Page 580: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 840/860

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the TCP analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the TCP analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the TCP analyzer.

Figure 51-1: Bedside monitor, large text screen

Figure 51-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the TCP analyzer.

� � �

42.8 ETEMPTCP

157 PCO2

oC mmHg

8 PO2

mmHg

� �

� � �

TCP

PCO2 = 157 mmHgPO2 = 8 mmHgETEMP = 42.8oC

� �

Flexport System Interface Operations Manual 51-4

Page 581: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 840/860

Figure 51-3: Central monitor, split screen

� Electrode (sensor) temperature measured in °C

� TCP parameter key

� Transcutaneous carbon dioxide tension (PtcCO2)

� Alarm status bell (appears when alarms are turned ON)

� Transcutaneous oxygen tension (PtcO2)

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the TCP key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms will still cause the TCP key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the TCP

analyzer.

• Alarm tones on the TCP analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch TCP.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

� �

� PCO2 = 157 mmHgTCP

� PO2 = 8 mmHg

� ETEMP = 42.8oC

� ICU 14

Flexport System Interface Operations Manual 51-5

Page 582: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 840/860

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the TCP analyzer’s parameters (refer to Figure 51-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays TCP ALM OFF to the right of the TCP key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 51-4: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 51-4).

To adjust alarm limits:

• Touch TCP.• Select ALARM LIMITS.• Select an alarm parameter.• Touch HI= or LO=.• Use arrow keys to adjust.

To adjust alarm limits:

• Touch TCP.• Touch VIEW ALARMS.

To select a display format:

• Touch TCP.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

PCO2 199 ETEMP 50.0 TCP

PCO2 = 157 mmHg0 0 PO2 = 8 mmHg

PO2 990 ETEMP = 42.8oC0

Flexport System Interface Operations Manual 51-6

Page 583: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 840/860

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the TCP analyzer during Flexport system interface operation.

To print the current display:

• Touch TCP.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the TCP analyzer and the monitor has been interrupted. Yes

CALIBRATION REQUIRED Analyzer requires calibration. No

Flexport System Interface Operations Manual 51-7

Page 584: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 585: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 52-1

Respironics Novametrix 1260/1265

Directory of KeysCO2

CO2 - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFF

CO2 - SETUP

SIZE SWEEPSPEED

LARGE TXTSMALL TXT

CO2 - SETUP - SIZE CO2 - SETUP - SWEEP SPEEDWAVEFORM

↑ ↓25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/secON OFF

CO2 - ALARM LIMITS

APNEA RR ETCO2

CO2 - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XX ↑ ↓ON OFF

CO2 - ALARM LIMITS - RRRR HI =

XXLO = XX ↑ ↓

APNEAON OFF RR

CO2 - ALARM LIMITS - APNEAAPNEA HI =

XXLO = XX ↑ ↓

APNEAON OFF RR

Page 586: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 587: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 1260/1265

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90433A-01 Flexport system interface enables information from a Respironics Novametrix 1260/1265 analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the analyzer as shown in Table 1. Consult the appropriate analyzer manual for additional configuration setup procedures.

Table 1: Analyzer Configuration

Analyzer Name and Model Number

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

Respironics Novametrix 1260 9600 8-bit none 1 131-0684-xx (25-pin) H:eng-1260-86

Respironics Novametrix 1265 9600 8-bit none 1 131-0684-xx (25-pin) eng-1265-18

Flexport System Interface Operations Manual 52-3

Page 588: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 1260/1265

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the analyzer.

Figure 52-1: Bedside monitor, large text screen

Figure 52-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the analyzer.

� �

CO2

51 ETCO2

mmHg

14 RR

BPM

� �

� � �

CO2

ETCO2 = 51 mmHgRR = 14 BPM

Flexport System Interface Operations Manual 52-4

Page 589: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 1260/1265

Figure 52-3: Central monitor, split screen

� ETCO2 parameter key

� Current EtCO2 value

� Alarm status bell (appears when alarms are turned ON)

� Current respiration rate value

� Apnea monitoring status

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the CO2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the CO2 key to flash.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch CO2.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

� � �

� ETCO2 = 51 mmHg APNEA = 20s C

O2

� RR = 14 BPM

� ICU 14

Flexport System Interface Operations Manual 52-5

Page 590: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 1260/1265

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 52-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays CO2 ALM OFF to the right of the CO2 key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 52-4: View Alarms screen

To adjust alarm limits:

• Touch CO2.• Select ALARM LIMITS.• Select an alarm parameter.• Select HI= or LO=.• Use the arrow keys to adjust.

To view alarm settings:

• Touch CO2.• Touch VIEW ALARMS.

ETCO2 5530 C

O2

49 ETCO2

mmHg

APNEA 20 RR 28 12 RR

8 BPM

Flexport System Interface Operations Manual 52-6

Page 591: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 1260/1265

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn the waveform OFF and display only numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the size keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch CO2.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch CO2.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 52-7

Page 592: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 1260/1265

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, 12.5, and 25 mm/second.

The default setting is 12.5 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 52-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch CO2.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch CO2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch CO2.• Select FREEZE ON.

Flexport System Interface Operations Manual 52-8

Page 593: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 1260/1265

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To print the current display:

• Touch CO2.• Touch PRINT.

Flexport System Interface Operations Manual 52-9

Page 594: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 1260/1265

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

AIR CAL ERROR Airway zero calibration error — retry airway zero. Yes

CAPNOMETER FAULTY Monitor hardware error, monitor faulty — do not use. Yes

CHECK AIRWAY ADAPTER Sensor not properly connected to airway adapter or calibration cells. No

COMMUNICATION LINK LOST Communication with the gas analyzer is interrupted. Yes

MONITOR ERROR 1 Monitor faulty, do not use — contact qualified service personnel. Yes

MONITOR ERROR 2 Monitor faulty, do not use — contact qualified service personnel. Yes

NO UP TO TEMP The analyzer has not completed its warm-up. No

RS232 ERROR Communication between the analyzer and the analyzer’s RS232 port is faulty. Yes

RECALIBRATE SENSORMonitor has detected that the sensor signals have changed since the analyzer was calibrated — recalibrate the sensor.

No

SENSOR FAULTY Sensor faulty — do not use, replace sensor. Yes

SENSOR OVER-TEMP Sensor faulty — do not use, replace sensor. No

SENSOR UNPLUGGED Sensor not connected to monitor — connect it.Sensor or sensor cable faulty — replace sensor. No

SPAN CAL ERROR Calibration error — retry calibration. Yes

ZERO CAL ERROR Calibration error — retry calibration. Yes

Flexport System Interface Operations Manual 52-10

Page 595: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 53-1

Respironics Novametrix 2001

Directory of Keys

SPO2

SPO2 - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFF

SPO2 - SETUP

SIZE SWEEPSPEED

LARGE TXTSMALL TXT

SPO2 - SETUP - SIZE SPO2 - SETUP - SWEEP SPEEDWAVEFORM SIZE

↑SIZE

↓50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/secON OFF

SPO2 - ALARM LIMITS

SPO2 PR

SPO2 - ALARM LIMITS - PRRR HI =

XXLO = XXON OFF

SPO2 - ALARM LIMITS - SPO2SPO2 HI =

XXLO = XXON OFF

These keys appear on the bedside monitor only.

Page 596: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 597: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 2001

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing the Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a Respironics Novametrix 2001 Pulse Oximeter to be displayed on Spacelabs Healthcare monitors.

Configure the pulse oximeter as shown in Table 1. Consult the pulse oximeter manual for additional configuration setup procedures.

Table 1: Pulse Oximeter Configuration

Baud Rate

Minimum Software Revision

Installation Kit Part Number

Transition Connector

Part Number

9600 Respironics Novametrix 2001; V 2.1 045-0156-xx 131-0946-xx

(25 pin)

Flexport System Interface Operations Manual 53-3

Page 598: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 2001

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the pulse oximeter, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the pulse oximeter. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pulse oximeter.

Figure 53-1: Bedside monitor, large text screen

Figure 53-2: Bedside monitor, small text

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 25-pin connector on the pulse oximeter serial port.

� � �

5.6 SIG SpO2

98 SPO2

%

70 RR

bpm

� �

� �

SpO2

SPO2= 98%

PR= 70bpm

SIG= 5.6

� � �

Flexport System Interface Operations Manual 53-4

Page 599: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 2001

Figure 53-3: Central monitor, split screen

� Current signal strength value

� SPO2 parameter key

� Current SpO2 value

� Alarm status bell (appears when alarms are turned ON)

� Current pulse rate value

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for SpO2. When an alarm is turned ON and the parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the SPO2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When both parameter alarms are turned OFF, alarm status messages will still cause the SPO2 key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the pulse

oximeter.

• Alarm limits can only be adjusted at the pulse oximeter. Alarm violation is detected at the pulse oximeter, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the pulse oximeter can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

� �

� SPO2 = 98% SPO2

� PR = 70bpm� SIG = 5.6

� ICU 14

Flexport System Interface Operations Manual 53-5

Page 600: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 2001

Viewing the Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the pulse oximeter’s parameters (refer to Figure 53-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays SPO2 ALM OFF to the right of the SPO2 key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 53-4: View Alarms screen

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch SPO2.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To view alarm settings:

• Touch SPO2.• Touch VIEW ALARMS.

SPO2 100 SpO2

98 SPO2

85 %

PR 180 70 RR

40 bpm

Flexport System Interface Operations Manual 53-6

Page 601: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 2001

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SIZE ↑ and SIZE ↓ keys, the SWEEP SPEED key, and the FREEZE ON/OFF keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch SPO2.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch SPO2.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 53-7

Page 602: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 2001

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 50, 25, 12.5, and 6.25 mm/second.

The default setting is 25 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 53-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch SPO2.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch SPO2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch SPO2.• Select FREEZE ON.

Flexport System Interface Operations Manual 53-8

Page 603: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 2001

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the oximeter or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the oximeter is powered ON.

• Verify that all cables are secure.

Note:When powering OFF the pulse oximeter with all cables attached, do NOT power the pulse oximeter back ON until after the Flexport channel has disappeared from the Spacelabs Healthcare monitor (approximately one minute).

To print the current display:

• Touch SPO2.• Touch PRINT.

Flexport System Interface Operations Manual 53-9

Page 604: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 605: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90434A-03 Flexport system interface enables information from a Respironics Novametrix 7000/7100 analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the analyzer as shown in Table 1. Consult the appropriate analyzer manual for additional configuration setup procedures.

To complete the configuration, set the analyzer to send CO2 waveform data (Tele-Cap/Spacelabs) from the RS-232 port. If the analyzer is configured to send SpO2 waveform data (Tele-Sat/Spacelabs), the Spacelabs Healthcare monitor will display the message COMMUNICATION ERROR – CALL SERVICE and cease processing analyzer information.

Table 1: Analyzer Configuration

Analyzer Name and Model Number

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

Respironics Novametrix Model 7000 9600 8-bit none 1 131-0684-xx

(25-pin) H: eng-1260-83

Respironics Novametrix Model 7100: CO2SMO 9600 8-bit none 1 131-0684-xx

(25-pin) Eng-7100-1699

Flexport System Interface Operations Manual 54-1

Page 606: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the analyzer.

Figure 54-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the analyzer.

� � � �

21 RR PO+ET

24 ETCO2

BPM mmHg

97SPO2

%

Flexport System Interface Operations Manual 54-2

Page 607: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

Figure 54-2: Bedside monitor, small text screen

Figure 54-3: Central monitor, split screen

� Current respiration rate value

� PO+ET parameter key

� Current EtCO2 (mmHg) value

� Alarm status bell (appears when alarms are turned ON)

� Current SpO2 (%) value

� Apnea monitoring status

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the PO+ET key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, status alarms still cause the PO+ET key to flash.

Alarm settings default to OFF.

� �

PO+ET

ETCO2 = 24 mmHg

SPO2 = 97%

RR = 21 BPM

� � �

� �

� ETCO2 = 24mmHg PO+ET

� SPO2 = 97%

� RR = 21 BPM

� APNEA = 20 sec

� ICU 14

Flexport System Interface Operations Manual 54-3

Page 608: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch PO+ET.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To adjust alarm limits:

• Touch PO+ET.• Select ALARM LIMITS.• Select an alarm parameter.• Select HI= or LO=.• Use the arrow keys to adjust.

Flexport System Interface Operations Manual 54-4

Page 609: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 54-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays PO+ET ALM OFF to the right of the PO+ET key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 54-4: View Alarms screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the size keys are disabled.

To view alarm settings:

• Touch PO+ET.• Touch VIEW ALARMS.

To turn the waveform display ON or OFF:

• Touch PO+ET.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

RR 40 EtCO275 P

O+ET

24 ETCO2

10 50 mmHg

APNEA 20 SpO2100 97

SPO2

85 %

Flexport System Interface Operations Manual 54-5

Page 610: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 1.56, 3.12, 6.25, 12.5, and 25 mm/second.

The default setting is 6.25 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 54-2).

To adjust waveform size:

• Touch PO+ET.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch PO+ET.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch PO+ET.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 54-6

Page 611: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To freeze the waveform on the display:

• Touch PO+ET.• Select FREEZE ON.

To print the current display:

• Touch PO+ET.• Touch PRINT.

Flexport System Interface Operations Manual 54-7

Page 612: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

AIR CAL ERROR Airway zero calibration error — retry airway zero. Yes

CAPNOMETER FAULTY Monitor hardware error, monitor faulty — do not use. Yes

CHECK AIRWAY ADAPTER Sensor not properly connected to airway adapter or calibration cells. No

COMMUNICATION ERROR — CALL SERVICE

The analyzer RS-232 port output is incorrectly configured. The analyzer is set to send SpO2 waveform data (Tele-Sat/Spacelabs) instead of CO2 waveform data (Tele-Cap/Spacelabs) from the RS-232 port. The monitor displays this message for approximately 20 seconds, then displays COMMUNICATION LINK LOST.

No

COMMUNICATION LINK LOST Communication between the analyzer and the monitor has been interrupted. Yes

CONNECT 87XX PROBE Connect sensor — if sensor is connected, replace, do not use. Yes

INSUFFICIENT LIGHTTissue between sensor LEDs and photo-diode is too thick or opaque to allow adequate light transmission.

Yes

LIGHT INTERFERENCE Ambient light interference (i.e., warming lights, direct sunlight). Yes

LOW SIGNAL STRENGTH Pulsatile signal too weak for proper sensor operation. Yes

MONITOR ERROR 1 Monitor faulty, do not use — contact qualified service personnel. Yes

Flexport System Interface Operations Manual 54-8

Page 613: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Respironics Novametrix 7000/7100

MONITOR ERROR 2 Monitor faulty, do not use — contact qualified service personnel. Yes

PROBE ERROR 1 Sensor faulty, do not use — replace sensor. Yes

PROBE ERROR 2 Sensor faulty, do not use — replace sensor. Yes

PROBE FAILURE Sensor faulty, do not use — replace sensor; also refer to UNIDENTIFIED PROBE. Yes

PROBE OFF PATIENT

Sensor dislodged from patient, improperly applied, or placed on an area too translucent for proper sensor operation — reposition sensor.

Yes

PULSE OUT OF RANGE Detected pulse rate is less than 30 bpm or greater than 250 bpm. No

RS232 ERROR Communication between the analyzer and the monitor has been interrupted. Yes

RECALIBRATE SENSORMonitor has detected that the sensor signals have changed since the analyzer was calibrated — recalibrate the sensor.

No

SENSOR FAULTY Sensor faulty — do not use, replace sensor. Yes

SENSOR OVER-TEMP Sensor faulty — do not use, replace sensor. No

SENSOR UNPLUGGEDSensor not connected to monitor — connect it.Sensor or sensor cable faulty — replace sensor.

No

SPAN CAL ERROR Calibration error — retry calibration. Yes

UNIDENTIFIED PROBE

Monitor cannot determine if sensor is of correct type (87xx series). May appear if attached to a thick tissue site when sensor is first plugged into monitor.

Yes

ZERO CAL ERROR Calibration error — retry calibration. Yes

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface Operations Manual 54-9

Page 614: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 615: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Directory of Keys (with PCO2/SpO2/PR parameters enabled on SenTec device)

SDM

SDM - MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFF

These keys appear on the bedside monitor only.

SDM - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

SDM - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

SDM - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec0.39

mm/sec

SDM - ALARM LIMITS

PCO2 SPO2 PR

SDM - ALARM LIMITS - PRPR HI =

XXLO = XXON OFF

SDM - ALARM LIMITS - SpO2SPO2 HI =

XXLO = XXON OFF

SDM - ALARM LIMITS - PCO2PCO2 HI =

XXLO = XXON OFF

Flexport System Interface SenTec Digital Monitor 55-1

Page 616: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Directory of Keys(with only PCO2 parameter enabled on SenTec device)

SDM

SDM - MENUALARM LIMITS SETUP VIEW

ALARMS PRINT

These keys appear on the bedside monitor only.

SDM - SETUPLARGE TXTSMALL TXT

SDM - ALARM LIMITS

PCO2

SDM - ALARM LIMITS - PCO2PCO2 HI =

XXLO = XXON OFF

Flexport System Interface SenTec Digital Monitor 55-2

Page 617: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

OverviewThe 90442A-03 Flexport system interface enables information from a SenTec Digital Monitor (SDM) to be displayed on Spacelabs Healthcare monitors.

Configure the SDM as shown in Table 1. Consult the SDM manual for additional configuration setup procedures.

Table 1: SenTec Digital Monitor Configuration

Baud Rate

Minimum Software Revision

Installation Kit Part Number

Transition Connector Part Number

9600 v06.10 045-0162-xx 131-1972-xx(9 pin)

Flexport System Interface SenTec Digital Monitor 56-3

Page 618: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the SDM, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the SDM. To control other interface functions, refer to the sections that follow.

Note:During the warm-up phase, the PCO2 values display in grey on the device and the monitor displays question marks (???). Once the device is sufficiently warmed up, PCO2 values will display normally.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the SDM.

Figure 56-1: Bedside monitor, large text screen with PCO2/SpO2/PR enabled

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 9-pin connector on the back of the SDM serial

port.

� � � �

41.9STempºC 85

PR

bpm SDM

31.8 PCO2

mmHg

99 SPO2

%7.8 STimehours

� � �

Flexport System Interface SenTec Digital Monitor 56-4

Page 619: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Figure 56-2: Bedside monitor, large text screen with only PCO2 enabled

Figure 56-3: Bedside monitor, small text screen with PCO2/SpO2/PR enabled

Figure 56-4: Bedside monitor, small text screen with only PCO2 enabled

Figure 56-5: Central monitor, split screen with PCO2/SpO2/PR enabled

� � �

41.9 STempºC S

DM

31.8 PCO2

mmHg

7.8 STimehours

� �

� � �

SDM

PCO2= 31.8mmHg

SPO2= 99%

PR= 85bpm

STemp= 41.9ºC

� � �

� �

SDM

PCO2= 31.8mmHg

STemp= 41.9ºC

� �

� � �

� PCO2 = 31.8mmHg SDM

� SPO2 = 99%

� STemp = 41.9ºC PR = 85bpm

� STime = 7.8 hours

SL370

Flexport System Interface SenTec Digital Monitor 56-5

Page 620: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Figure 56-6: Central monitor, split screen with only PCO2 enabled

� SDM parameter key

� Current PCO2 (transcutaneous carbon dioxide tension) value

� Current SpO2 (oxygen saturation) value

� Current PR (pulse rate) value

� Current STemp (sensor temperature measured in ºC) value

� Current STime (site timer countdown clock) value

� Alarm status bell (appears when alarms are turned ON)

Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the SDM key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the SDM key to flash.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the SDM.

• Alarm limits can only be adjusted at the SDM. Alarm violation is detected at the SDM, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the SDM can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

� PCO2 = 31.8mmHg SDM

� STemp = 41.9ºC

� STime = 7.8 hours

SAT1

Flexport System Interface SenTec Digital Monitor 56-6

Page 621: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the SDM’s parameters (refer to Figure 56-7, Figure 56-8, Figure 56-9, and Figure 56-10). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays SDM ALM OFF to the right of the SDM key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 56-7: View Alarms screen, large text, PCO2/SpO2/PR enabled

Figure 56-8: View Alarms screen, small text, PCO2/SpO2/PR enabled

To turn alarms ON or OFF from the Spacelabs Healthcare monitor:

• Touch SDM.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To view alarm settings:

• Touch SDM.• Touch VIEW ALARMS.

PCO2 45.02.0 PR

14540 S

DM

31.8PCO2

mmHg

99SPO2 10070

SpO2

%

PCO2 45.02.0

PR 14540 S

DM

PCO2= 31.8mmHg

SPO2= 99%

SPO2 10070

PR= 46bpm

STemp= 42.0ºC

Flexport System Interface SenTec Digital Monitor 56-7

Page 622: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Figure 56-9: View Alarms screen, large text, only PCO2 enabled

Figure 56-10: View Alarms screen, small text, only PCO2 enabled

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SIZE ↑ and SIZE ↓ keys, the SWEEP SPEED key, and the FREEZE ON/OFF keys are disabled.

The default setting is WAVEFORM ON.

When PCO2 is the only parameter enabled on the SDM device, the WAVEFORM ON/OFF key, the SWEEP SPEED key, and the FREEZE ON/OFF keys are not displayed on Spacelabs Healthcare monitors.

To turn the waveform display ON or OFF:

• Touch SDM.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

PCO2 45.0 SDM

31.8 PCO2

2.0 mmHg

PCO2 45.02.0 S

DM

PCO2= 31.8mmHg

STemp= 42.0ºC

Flexport System Interface SenTec Digital Monitor 56-8

Page 623: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 50, 25, 12.5, 6.25, 3.12, 1.56, 0.78, and 0.39 mm/second.

The default setting is 12.5 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 56-4).

The default setting is LARGE TXT.

To adjust waveform size:

• Touch SDM.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch SDM.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch SDM.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface SenTec Digital Monitor 56-9

Page 624: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the SDM or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the SDM is powered ON.

• Verify that all cables are secure.

Note:• When powering OFF the SDM with all cables attached, do NOT power the SDM back ON until after the

Flexport channel has disappeared from the Spacelabs Healthcare monitor (approximately one minute).

To freeze the waveform on the display:

• Touch SDM.• Select FREEZE ON.

To print the SDM display:

• Touch SDM.• Touch PRINT.

Flexport System Interface SenTec Digital Monitor 56-10

Page 625: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

Table 2 lists the status messages that the Flexport system interface reports for the SenTec Digital Monitor during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

MONITOR FAULT Monitor faulty, do not use — contact qualified service personnel. Yes

CONNECT SENSOR Sensor is not detected — if sensor is connected, replace, do not use. Yes

SENSOR FAULTThe system detected a severe sensor fault. Sensor is shut down — do not use, replace sensor.

Yes

TEMP. LIMITER ACTIVE Temperature surveillance detected a temperature problem. Yes

GAS LEAK IN DOCKING STATIONDocking station (DS) surveillance detected a leak in the DS chamber (for example, polluted by sensor gel).

Yes

SENSOR OFF PATIENT The sensor was dropped or is removed from the patient and calibration is needed. Yes

CALIBRATE SENSOR Sensor is removed from the docking station and calibration is needed. Yes

DOCKING STATION FAULTDocking station surveillance detected a severe DS fault (for example, the gas pressure too high).

Yes

GAS BOTTLE EMPTY Gas bottle on the docking station is empty. Yes

SPO2 STABILIZING SpO2 and pulse rate are not stable. No

LOW SIGNAL Low pulse signal. No

MOTION ARTIFACT Motion artifact detected. No

SITE TIME ELAPSED Check measurement site, and if necessary, relocate the sensor to prevent skin irritation. Yes

REMEMBRANCE SENSOR The sensor needs to be remembrance. Yes

BATTERY LOWBattery capacity is low (<10%) while the power supply cable is unplugged or the power supply is plugged in and charging.

Yes

PCO2 STABILIZINGPCO2 measurement is not yet stable after sensor application. PCO2 values do not yet reflect patient data.

No

Flexport System Interface SenTec Digital Monitor 56-11

Page 626: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

SenTec Digital Monitor

READY FOR USE Ths sensor is removed from the docking station and calibration is NOT needed No

Table 2: Status Messages (continued)

Message Text Indication Alarm Sounds

Flexport System Interface SenTec Digital Monitor 56-12

Page 627: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 57-1

Smiths Graseby 3000

Directory of Keys

IV

IV - MAIN MENU

SETUP PRINT

IV - SETUP MENU

PUMP A WEIGHT CHANGE CONFIG

IV - CHANGE CONFIGURATION - SELECT PARAMETER TO EDIT

EDIT DRUG EDIT FLUID

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = ___ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

IV - PUMP A MENUPRIMARY SELECT

DRUGSELECT CONC.

SELECT DILUENT

DOSE CALCS

ALARMS

SECONDARY ON OFF

Page 628: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 629: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Disconnecting a Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

OverviewThe 90437A-08 Flexport system interface enables information from a Smiths Graseby 3000 volumetric infusion pump to be displayed on Spacelabs Healthcare monitors.

Configure the infusion pump as shown in Table 1. Consult the infusion pump manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the infusion pump, complete the steps listing in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the pump. To control other interface functions, refer to the sections that follow.

Table 1: Infusion Pump Configuration

Baud Rate

Character Length

Parity Setting Stop Bit

9600 7-bit odd 1

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the connector on the infusion pump communications

port.

Flexport System Interface Operations Manual 57-3

Page 630: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pump.

Figure 57-1: Bedside, remote, or central monitor, full screen

Figure 57-2: Central monitor, split screen

� Pump address

� Selected medication or fluid (until you set up the drug or fluid for each pump, this field displays * * * * * * * *)

� Selected pump or infusion rate. Initially, this field displays the pump rate. The calculated medication rate is displayed once you have selected the concentration or dose. Units for displays are based on those selected.

� The volume of fluids and the dose delivered since the last time one of the following conditions occurred:

• The pump started a new dose

• The pump rate was titrated

• The pump changed from secondary to primary or from primary to secondary

� IV parameter key.

� Latest status message (for example, PUMPING). The displayed infusion may be either the primary or the secondary. The secondary infusion is displayed when the pump is pumping secondary (SEC PUMPING) or concurrently (CONCURRENT). Refer to Status Messages on page 57-16.

� � � � � �

PUMP DRUG/FLUID RATE VOL. INFUSED PUMP STATUSA DOPAMINE 10.0 mcg/kg/min 122.1 ml I

VA PUMPING

� � � �

PUMP DRUG/FLUID STATUSA DOPAMINE PUMPING I

V

Flexport System Interface Operations Manual 57-4

Page 631: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Entering Setup InformationSetup information is necessary for the monitor to display the drug (or fluid) names and to compute the medication and calculated pump rates. You can control the following functions from the Setup menu:

• Turn alarms ON or OFF.

• Enter patient weight.

• Edit drug or fluid names.

• Set up primary and secondary infusion lines.

• Select the infusion fluid.

• Select the infusion drug.

Once you select the infusion drug, you can perform the following functions:

• Enter the infusion concentration.

• Select the diluent fluid.

• Enter the infusion dose.

Note:Setup information can be entered only at the bedside monitor.

Enabling AlarmsWhen you turn alarms OFF, infusion pump alarms will no longer trigger an alarm on the Spacelabs Healthcare monitor.

Alarm settings default to ON.

Note:Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the infusion pump.

To turn alarms ON or OFF:

• Touch IV.• Select SETUP.• Touch PUMP.• Touch ALARMS ON/OFF.

Flexport System Interface Operations Manual 57-5

Page 632: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Selecting an Infusion LineYou cannot set up the secondary line until you complete the primary infusion setup.

Selecting the Infusion FluidThere are two pages of Select Fluid screens. The first page displays the list of fluids automatically supplied by the Flexport system interface. The second page enables you to select from a list of up to eight fluids that you can enter into the Flexport system interface’s memory. Refer to Editing Drug or Fluid Names on page 57-13 for information on entering these names.

Figure 57-3: Select Fluid screen

Table 2 defines the fluid/concentration abbreviations that correspond to the fluids listed on the Select Fluid screen.

To select an infusion line:

• Touch IV.• Select SETUP.• Touch PUMP.• Touch PRIMARY/SECONDARY.

Table 2: Fluid Abbreviations

Abbreviation Description

LR lactated ringers

D5 LR 5% dextrose with lactated ringers

D5 NS 5% dextrose with 0.9% saline

D5 1/2NS 5% dextrose with 0.45% saline

D5 1/2NS 20KCL 5% dextrose with 0.45% saline and 20 mEq potassium chloride/liter

D5 1/4NS 5% dextrose with 0.2% saline

PUMP A - SELECT FLUID

PAGE ↑ PAGE ↓ ENTER

LR

D5 1/2NS 20KCL

D10 1/2NS

1/2 NS

D5 LR

D5 1/4NS

D10 1/4NS

1/4NS

D5 NS

D5W

D10W

TPN

D5 1/2NS

D10 NS

NS

LIPIDS

Flexport System Interface Operations Manual 57-6

Page 633: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

D5W 5% dextrose with water

D10 NS 10% dextrose with 0.9% saline

D10 1/2NS 10% dextrose with 0.45% saline

D10 1/4NS 10% dextrose with 0.2% saline

D10 W 10% dextrose with water

NS 0.9% saline

1/2 NS 0.45% saline

1/4 NS 0.2% saline

TPN Total parenteral nutrition

LIPIDS Fat emulsion delivered as part of parenteral nutrition

To select an infusion fluid:

• Touch IV.• Select SETUP.• Touch PUMP.• Touch SELECT FLUID.• Select a fluid from the list (or touch PAGE↓, then select a fluid).• Touch ENTER.

Table 2: Fluid Abbreviations (continued)

Abbreviation Description

Flexport System Interface Operations Manual 57-7

Page 634: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Selecting a DrugTo select a drug for infusion, you must first choose a drug from the Select Drug page. The selected drug becomes active upon touching the ENTER key. The Flexport system interface provides 16 default drug names. Refer to Editing Drug or Fluid Names on page 57-13 for information on editing those default drug names.

Figure 57-4: Select Drug screen

After selecting a drug, the monitor displays (in order) displays for the concentration and diluent.

PUMP A - SELECT DRUGAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

ENTER

To select a drug:

• Touch IV.• Select SETUP.• Touch PUMP.• Touch SELECT DRUG.• Select a drug from the list.• Touch ENTER.

Flexport System Interface Operations Manual 57-8

Page 635: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Select the Drug ConcentrationThe drug concentration represents the amount of a drug dissolved in a volume of diluent. The Flexport system interface saves the last entered default concentration values and unit of measurement associated with the selected drug.

Access to this display is automatic from the Select Drug screen.

To change a value after you have completed an initial drug/fluid entry, complete all the steps listed in the Quickstart on page 57-10. Figure 57-5 illustrates the Concentration screen.

Figure 57-5: Concentration screen

Caution:• The validity of the drug concentration or dose values that you enter is NOT checked by the

system.

• Calculated drug rates are displayed to one significant decimal place. Rates less than 0.05 will appear as 0.0.

PUMP A - ENTER DRUG QTY, UNITS, AND VOLUME. CONCENTRATION = _______

7 8 9 mcg mg /50 ml /100 ml

4 5 6 g ml /250 ml /500 ml

1 2 3 units mEq /1 L /2 L

0 . CLEAR ENTER

Flexport System Interface Operations Manual 57-9

Page 636: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Selecting the DiluentA diluent is a fluid used to dilute a drug for infusion. The fluids you can choose as drug diluents are the same as those you can choose as infusion fluids. Refer to Selecting the Infusion Fluid on page 57-6 for details on selecting a diluent. The Flexport system interface saves the last entered diluent associated with a selected drug.

Access to this display is automatic from the Concentration screen.

To change your selection after you have completed an initial diluent entry, complete all the steps listed in the following Quickstart.

When you touch ENTER to complete the diluent entry, the monitor displays the Setup menu.

Selecting the DoseThe dose represents the amount of drug to be infused over a specified time period. Use this display to enter the quantity, units of measurement, and time units.

If you have completed an initial drug/fluid entry and wish to change a value, complete all of the steps listed in the Quickstart on page 57-11.

Note:If the Flexport system interface displays an initial VOLUME INFUSED for the pump you intend to use, you must clear that value at the pump before you start an infusion. Otherwise, the Flexport system interface will appear to add the volume infused from the previous infusion to the current infusion (either primary or secondary).

To select a drug concentration:

• Touch IV.• Select SETUP.• Touch PUMP.• Touch SELECT CONC.• Enter the drug quantity.• Select the unit of measurement.• Select the volume of fluid.• Touch ENTER.

To select a diluent:

• Touch IV.• Select SETUP.• Touch PUMP.• Touch SELECT DILUENT.• Choose a diluent from the list (or touch PAGE , then select a diluent).• Touch ENTER.

Flexport System Interface Operations Manual 57-10

Page 637: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Figure 57-6: Dose Calcs screen

As you enter the quantity, unit of measurement, and unit of time, the monitor displays the complete dose and the calculated pump rate. This calculated pump rate is based on the concentration and dose values you enter. To arrive at this value, the Flexport system interface performs the following calculation:

CALCULATED PUMP RATE = DOSE/CONCENTRATION

Note:The calculated pump rate is provided as information only. The Flexport system interface does not control the actual set pump rate.

When you touch ENTER to complete the dose entry, the interface saves your selection and the monitor displays the Setup menu.

PUMP A - ENTER DRUG QUANTITY, UNITS, AND TIME. DOSE = _______

7 8 9 mcg mg /min /hr

4 5 6 g ml CALCULATED PUMP RATE= xxxx.x ml/hr

1 2 3 units mEq

0 . CLEAR mcg/kg mg/kg ENTER

To select the drug dose:

• Touch IV.• Select SETUP.• Touch PUMP.• Touch DOSE CALCS.• Enter the dose quantity.• Enter unit of measurement.• Enter the time unit.• Touch ENTER.

Flexport System Interface Operations Manual 57-11

Page 638: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Setting Patient WeightTouching the WEIGHT key displays a screen with a prompt containing the current weight of the patient (if you entered it previously using the Weight screen, a cardiac output module, or the monitor admit function). The Patient Weight screen uses the same units as the bedside monitor (kg or pounds).

Figure 57-7: Patient Weight screen

After you complete your editing and touch ENTER, the current weight will appear in the appropriate displays. You can enter up to six characters, including the decimal point. The valid weight range is from 1 to 250 kg (2 to 551 pounds). The interface saves the weight so you can use it in the Dose Calcs screen with the mcg/kg and mg/kg units of measurement.

IV - ENTER PATIENT WEIGHT FOR CALCULATIONS. WEIGHT = __ lbs/kg

7 8 9

4 5 6

1 2 3

0 . CLEAR ENTER

To enter patient weight:

• Touch IV.• Touch SETUP.• Touch WEIGHT.• Enter the patient weight.• Touch ENTER.

Flexport System Interface Operations Manual 57-12

Page 639: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Editing Drug or Fluid NamesThe Change Configuration screen for drug names initially lists the 16 default drug names. You can change the name of any of these drugs (refer to Figure 57-8).

The Change Configuration screen for fluid names initially shows a page of eight blank keys. You can select any of the eight keys for the fluid name.

Figure 57-8: Change Drug Configuration screen

After selecting a drug or fluid key to relabel, touch CHANGE NAME. The monitor displays the Change Drug (or Fluid) Name screen (refer to Figure 57-9).

IV - CHANGE CONFIGURATION - SELECT DRUG KEY TO RELABELAMINOPHYLLINE AMRINONE BRETYLIUM DOBUTAMINE

DOPAMINE ESMOLOL HEPARIN ISOPROTERENOL

LIDOCAINE METARAMINOL NITROGLYCERIN NITROPRUSSIDE

NOREPHINEPHRINE PHENTOLAMINE PHENYLEPHRINE PROCAINAMIDE

CHANGE NAME

Flexport System Interface Operations Manual 57-13

Page 640: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Figure 57-9: Change Drug Name screen

• Enter each character or number using the alphanumeric keys.

• Touch DEL to eliminate the character highlighted by the cursor.

• Move through the name, character by character, using the arrow keys.

• Touch CLEAR to eliminate the entire name and start over. The screen displays blank fields in place of the name.

• Touch ENTER to complete the edit and store the name in memory.

• Enter up to fifteen characters.

After you complete your editing and touch ENTER, the name will appear in the appropriate display for selecting drugs or fluids. The interface saves the name so you can recall it the next time you connect the Flexport system interface.

IV-CHANGE DRUG (or fluid) NAME= _________________________

0 1 2 3 4 5 6 7 8 9

A B C D E F G H I J

K L M N O P Q R S T

U V W X Y Z / % .

DEL ← → CLEAR ENTER

Flexport System Interface Operations Manual 57-14

Page 641: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Disconnecting a PumpWhen you disconnect the pump from the Flexport system, the Flexport system interface discontinues the infusion record.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To edit a drug or fluid name:

• Touch IV.• Select SETUP• Touch CHANGE CONFIG.• Touch EDIT DRUG or EDIT FLUID.• Select the name you want to change.• Touch CHANGE NAME.• Use the alphanumeric keys to make necessary changes.• Touch ENTER.

To disconnect a pump, detach the modular cable connecting the Flexport to the pump at either end.

To print the infusion information:

• Touch IV.• Touch PRINT.

Flexport System Interface Operations Manual 57-15

Page 642: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Smiths Graseby 3000

Status MessagesThe message DISCONNECTED indicates a problem with the pump or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pump is powered ON.

• Verify that all cables are secure.

Table 3 lists the status messages that the Flexport system interface reports for the pump during Flexport system interface operation.

Note:When the pump is in alarm for On Hold, the displayed status message is PUMP ALARM. Touching the SILENCE key on the pump causes the pump to transition from the alarm state to the hold state. The monitor reflects this by displaying a status of STOPPED.

Table 3: Status Messages

Message Text Indication

DISCONNECTED Pump has been disconnected.

PUMPING Pump is pumping normally.

SEC PUMPING Pump is operating the secondary line.

STOPPED Pump is in hold state or turned OFF.

AIR IN LINE Pump has detected air in IV tubing.

DOOR OPEN Pump has detected an open door.

KVO Pump pumping at KVO rate.

LOW BATTERY Pump battery requires recharging.

OCCLUSION Closed clamp, kinked, occluded tubing, or occluded filter detected.

PUMP ALARM Pump has detected an alarm condition.

Flexport System Interface Operations Manual 57-16

Page 643: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 58-1

Tyco Nellcor N-395/N-595/N-600

Directory of Keys

SPO2

SpO2 - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFF

These keys appear on the bedside monitor only.

SpO2 - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

SpO2 - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

SpO2 - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec0.39

mm/sec

SpO2 - ALARM LIMITS

SPO2 PR

SpO2 - ALARM LIMITS - PRPR HI =

XXLO = XXON OFF

SpO2 - ALARM LIMITS - SpO2SPO2 HI =

XXLO = XXON OFF

Page 644: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 645: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-395/N-595/N-600

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a Tyco Nellcor N-395, N-595, or N-600 pulse oximeter to be displayed on Spacelabs Healthcare monitors.

Configure the pulse oximeter as shown in Table 1. Consult the pulse oximeter manual for additional configuration setup procedures.

Table 1: Pulse Oximeter Configuration

Baud Rate

Minimum Software Revision

Installation Kit Part Number

Transition Connector Part Number

9600N-395: V 1.8.0.0N-595: V 3.3.0.0N-600: V 1.1.2.0

045-0151-xx 131-1725-xx (15 pin)

Flexport System Interface Operations Manual 58-3

Page 646: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-395/N-595/N-600

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the pulse oximeter, complete the steps listed in the following Quickstart.

Note:When powering OFF the equipment, power OFF the Nellcor pulse oximeter or disconnect the Flexport before powering OFF the Spacelabs Healthcare monitor. Failure to do so may cause the Nellcor pulse oximeter to reset. If, after a monitor reset or power sequence, the Flexport does not start updating information, disconnect the Flexport from the back of the monitor and reconnect to establish communications.

The Spacelabs Healthcare monitor is now ready to display information from the pulse oximeter. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the pulse oximeter.

Figure 58-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 15-pin connector on the pulse oximeter serial

port.

� �

SpO2

98SpO2

%

70 PR

bpm

� �

Flexport System Interface Operations Manual 58-4

Page 647: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-395/N-595/N-600

Figure 58-2: Bedside monitor, small text screen

Figure 58-3: Central monitor, split screen

� SpO2 parameter key

� Current SpO2 value

� Alarm status bell (appears when alarms are turned ON)

� Current pulse rate value

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for SpO2 and pulse rate. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the SpO2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the SpO2 key to flash.

Alarm settings default to ON.

� �

SpO2

SpO2= 98%

PR= 70bpm

� �

� �

� SpO2 = 98% SpO2

� PR = 70bpm

� ICU - 4

Flexport System Interface Operations Manual 58-5

Page 648: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-395/N-595/N-600

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarms settings on the pulse

oximeter.

• Alarm limits can only be adjusted at the pulse oximeter. Alarm violation is detected at the pulse oximeter and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the pulse oximeter can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the pulse oximeter’s parameters (refer to Figure 58-4). If alarms for both parameters are turned OFF, the Spacelabs Healthcare monitor displays SPO2 ALM OFF to the right of the SpO2 key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 58-4: View Alarms screen

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch SpO2.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To view alarm settings:

• Touch SpO2.• Touch VIEW ALARMS.

SpO2 100 SpO2

98SpO2

85 %

PR 180 70 PR

40 bpm

Flexport System Interface Operations Manual 58-6

Page 649: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-395/N-595/N-600

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SIZE ↑ and SIZE ↓ keys, the SWEEP SPEED key, and the FREEZE ON/OFF keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch SpO2.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To change waveform size:

• Touch SpO2.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 58-7

Page 650: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-395/N-595/N-600

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 50, 25, 12.5, 6.25, 3.12, 1.56, 0.78, and 0.39 mm/second. The minimum recorder sweep speed is 1.56 mm/second.

The default setting is 25 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 58-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch SpO2.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch SpO2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch SpO2.• Select FREEZE ON.

Flexport System Interface Operations Manual 58-8

Page 651: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-395/N-595/N-600

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the pulse oximeter or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the pulse oximeter is powered ON.

• Verify that all cables are secure.

Note:When powering OFF the pulse oximeter with all cables attached, do NOT power the pulse oximeter back ON until after the Flexport channel has disappeared from the Spacelabs Healthcare monitor (approximately one minute).

To print the pulse oximeter display:

• Touch SpO2.• Touch PRINT.

Flexport System Interface Operations Manual 58-9

Page 652: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 653: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-1000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

OverviewThe 90434A-02 Flexport system interface enables information from a Tyco Nellcor N-1000 analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the analyzer as shown in Table 1. Consult the analyzer manual for additional configuration setup procedures.

Table 1: Analyzer Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

Transition Connector

Part Number

Minimum Software Revision

9600 8-bit none 1 131-0686-xx (9-pin) 02.03.03

Flexport System Interface Operations Manual 59-1

Page 654: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-1000

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the analyzer.

Figure 59-1: Bedside monitor, large text screen

Figure 59-2: Bedside monitor, small text

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the analyzer.

� � � � �

40 N2O 21 RR PO+ET

24 ETCO2

% BPM mmHg

83 PR 97 SPO2

bpm %

� �

� � �

PO+ET

ETCO2= 24mmHg

SPO2 = 97%RR = 21BPMN2O= 40%

� � �

Flexport System Interface Operations Manual 59-2

Page 655: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-1000

Figure 59-3: Central monitor, split screen

� Current nitrous oxide value*

� Current respiration rate value

� PO+ET parameter key

� Current ETCO2 (mmHg) value

� Alarm status bell (appears when alarms are turned ON)

� Current SpO2 (%) value

� Current pulse rate value

Apnea monitoring status (displays ON or OFF)

Patient/room ID

*The N2O display is ON when the N2O value from the analyzer is greater than 2% or when the N2O alarm limit key is turned ON.

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the PO+ET key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the PO+ET key to flash.

Alarm settings default to OFF.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch PO+ET.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

� � �

� ETCO2= 24mmHg N2O = 40% PO+ET

� RR = 21BPM APNEA = 20S

� PR = 83bpm

� SPO2 = 97%

ICU 14

Flexport System Interface Operations Manual 59-3

Page 656: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-1000

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• When the Apnea alarm is turned ON, it is fixed at 20 seconds.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 59-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays PO+ET ALM OFF to the right of the PO+ET key. If alarms are turned ON for any parameter, a bell is displayed instead. The alarm bell flashes when an alarm limit is violated.

Figure 59-4: View Alarms screen

To adjust alarm limits:

• Select HI= or LO=.• Use the arrow keys to adjust.

To view alarm settings:

• Touch PO+ET.• Touch VIEW ALARMS.

N2O 70 RR 40 ETCO2 55 PO+ET

24 ETCO2

20 10 25 mmHg

APNEA 20 PR 110 BPM 100 97 SPO2

50 70 %

Flexport System Interface Operations Manual 59-4

Page 657: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-1000

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 59-2).

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch PO+ET.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To print the analyzer display:

• Touch PO+ET.• Touch PRINT.

Flexport System Interface Operations Manual 59-5

Page 658: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-1000

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST Communication between the analyzer and the monitor has been interrupted. Yes

CO2-ZERO CALIBRATIONThe analyzer is calibrating (occurs after you initiate analyzer calibration function or periodically during automatic self-calibration).

No

GAS SYSTEM DISABLED Gas system is disabled (occurs after you touch the analyzer DISABLE button). No

GAS SYSTEM PURGING Gas system is purging (occurs when you touch the analyzer START PURGE button). No

Flexport System Interface Operations Manual 59-6

Page 659: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-2500

Directory of KeysGAS

GAS - MAIN MENUALARM LIMITS SETUP VIEW

ALARMS PRINT

GAS - SETUP MENULARGE TXTSMALL TXT

GAS - ALARM LIMITS

RR ETCO2 SPO2 PR N2O

GAS - ALARM LIMITS - N2ON2O HI =

XXLO = XX ↑ ↓ON OFF

GAS - ALARM LIMITS - PRPR HI =

XXLO = XX ↑ ↓

SPO2ON OFF PR

GAS - ALARM LIMITS - SPO2SPO2 HI =

XXLO = XX ↑ ↓

SPO2ON OFF PR

GAS - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XX ↑ ↓

RRON OFF ETCO2

GAS - ALARM LIMITS - RRRR HI =

XXLO = XX ↑ ↓

RRON OFF ETCO2

Flexport System Interface Operations Manual 60-1

Page 660: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 661: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-2500

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

OverviewThe 90433A-05 Flexport system interface enables information from a Tyco Nellcor N-2500 gas analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the gas analyzer as shown in Table 1. Consult the analyzer manual for additional configuration setup procedures.

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the gas analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the gas analyzer. To control other interface functions, refer to the sections that follow.

Table 1: Gas Analyzer Configuration

Baud Rate

Character Length

Parity Setting Stop Bit Transition Connector

9600 8-bit none 1 131-0686-xx (9-pin)

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector in the gas analyzer.

Flexport System Interface Operations Manual 60-3

Page 662: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-2500

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the gas analyzer.

Figure 60-1: Bedside monitor, large text screen

Figure 60-2: Bedside monitor, small text

Figure 60-3: Central monitor, split screen

� Current isoflurane value

� Current nitrous oxide value

� Current respiration rate

� GAS parameter key

� Current end tidal CO2 (mmHg) value

� Alarm status bell (appears when alarms are turned ON)

� � � � �

2.9 ISO 40 N2O 22 RRGAS

24 ETCO2

% % BPM mmHg

0.1 ENF 0.4 HAL 4.2 PR 96 SPO2

% % BPM %

� � �

� �

GAS

ETCO2 = 24mmHg

SPO2 = 96%

RR = 22 BPM

N2O= 40%

� � � �

� � � � �

� ETCO2= 24mmHg N2O = 40%GAS

� RR = 22 BPM ISO = 2.9%

PR = 4.2 BPM HAL = 0.4%

� SPO2 = 96% ENF = 0.1%

� ICU 14

Flexport System Interface Operations Manual 60-4

Page 663: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-2500

� Current pulse oximetry value

Current pulse rate

Current halothane value

� Current enflurane value

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the GAS key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm limits for anesthetic agents can only be adjusted at the analyzer. When an agent parameter value exceeds an alarm limit, the Spacelabs Healthcare monitor does not send alarm data to the recorder, but the GAS parameter key flashes and a message is displayed on the monitor, indicating which parameter is in alarm.

Alarm settings default to OFF.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the gas

analyzer.

• Alarm tones on the gas analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch GAS.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To adjust alarms limits:

• Touch GAS.• Select ALARM LIMITS.• Select an alarm parameter.• Select HI= or LO=.• Use the arrow keys to adjust.

Flexport System Interface Operations Manual 60-5

Page 664: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-2500

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 60-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays GAS ALM OFF to the right of the GAS key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 60-4: View Alarms screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 60-4).

To view alarm settings:

• Touch GAS.• Touch VIEW ALARMS.

To select a display format:

• Touch GAS.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

N2O60

PR 90 ETCO2 45GAS

ETCO2 = 24 mmHg20 05 01 SPO2 = 96%

PR 120 SPO2 100 RR = 22 BPM

50 90 N2O = 40%

Flexport System Interface Operations Manual 60-6

Page 665: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-2500

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To print the analyzer display:

• Touch GAS.• Touch PRINT.

Flexport System Interface Operations Manual 60-7

Page 666: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N-2500

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

AGENT — ZERO CALIBRATIONThe analyzer is performing a calibration. Agent processing is suspended while this message is displayed.

No

APNEA VIOLATION The apnea alarm limit has been violated. The limit is 20 seconds. Yes

COMMUNICATION LINK LOST Communication between the gas analyzer and the monitor has been interrupted. Yes

CO2 — ZERO CALIBRATIONThe analyzer is performing a calibration. EtCO2 and RR processing are suspended while this message is displayed.

No

GAS SYSTEM DISABLE

Displayed when you touch the DISABLE button on the analyzer. Agent, EtCO2, and RR processing are suspended while this message is displayed.

No

GAS SYSTEM PURGING

Displayed when you touch the START PURGE button on the analyzer. Agent, EtCO2, and RR processing are suspended while this message is displayed.

No

XXX LO LIMIT VIOLATION The XXX alarm limit has been violated (where XXX = ENF, ISO, or HAL). Yes

XXX LO LIMIT VIOLATION The XXX alarm limit has been violated (where XXX = ENF, ISO, or HAL). Yes

Flexport System Interface Operations Manual 60-8

Page 667: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

Directory of KeysCO2

CO2 - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFFRefer to

page 39-2These keys appear on

the bedside monitor only.

CO2 - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

CO2 - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

CO2 - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec

Flexport System Interface Operations Manual 61-1

Page 668: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

CO2

CO2 - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFFRefer to

page 39-1

CO2 - ALARM LIMITS

SPO2 PULSE ETCO2 RESP I CO2

CO2 - ALARM LIMITS - I CO2I CO2 HI =

XXLO = XX

I CO2ON OFF SpO2

CO2 - ALARM LIMITS - RESPRESP HI =

XXLO = XX

RESPON OFF I CO2

CO2 - ALARM LIMITS - ETCO2ETCO2 HI =

XXLO = XX

ETCO2ON OFF RESP

CO2 - ALARM LIMITS - PULSEPULSE HI =

XXLO = XX

PULSEON OFF ETCO2

CO2 - ALARM LIMITS - SpO2SPO2 HI =

XXLO = XX

SPO2ON OFF PULSE

Flexport System Interface Operations Manual 61-2

Page 669: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a Tyco Nellcor N6000 analyzer to be displayed on Spacelabs Healthcare monitors.

Configure the analyzer as shown Table 1. Consult the analyzer manual for additional configuration setup procedures.

To complete the configuration and set the serial output format to FLEXPORT, complete the following steps on the analyzer:

1 Press the FREEZE hard key on the analyzer.

2 Simultaneously touch soft keys 2 and 4 (they are unlabeled).

3 Touch the SET CUSTOM DEFAULTS soft key.

4 Touch the SYSTEM soft key.

5 Touch the SYSTEM VARIABLES soft key.

6 Touch the COMM MODE soft key until FLEXPORT is highlighted.

7 Press the MENU hard key on the analyzer twice to return to the Main Monitoring screen.

Table 1: Analyzer Configuration

Baud Rate

Minimum Software Revision

Installation Kit Part Number

Transition Connector Part Number

9600 2.0 045-0031-xx 131-1039-xx(9 pin)

Flexport System Interface Operations Manual 61-3

Page 670: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the analyzer, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the analyzer. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the analyzer.

Figure 61-1: Bedside monitor, large text screen

Figure 61-2: Bedside monitor, small text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 9-pin connector on the back of the analyzer.

� � � �

35 ETCO2 1 I CO2CO2

98 SpO2

mmHg mmHg %

16 RESP 70 PULSE

BPM bpm

� � �

� � ��

CO2

SpO2= 98%

PULSE= 70bpm

ETCO2= 35mmHg

RESP= 16BPM

� �

Flexport System Interface Operations Manual 61-4

Page 671: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

Figure 61-3: Central monitor, split screen

� Current end tidal CO2 value

� Current inspired CO2 value

� CO2 parameter key

� Current SpO2 value

� Alarm status bell (appears when alarms are turned ON)

� Current pulse rate value

� Current respiration rate value

Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the CO2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the CO2 key to flash.

Alarm settings default to OFF.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch CO2.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

� �

� SpO2 = 98% I CO2 = 1mmHgCO2

� PULSE = 70bpm

� ETCO2 = 35mmHg

� RESP = 16BPM

ICU 4

Flexport System Interface Operations Manual 61-5

Page 672: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the analyzer.

• Alarm limits can only be adjusted at the analyzer. Alarm violation is detected at the analyzer, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the analyzer can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the analyzer’s parameters (refer to Figure 61-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays CO2 ALM OFF to the right of the CO2 key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 61-4: View Alarms screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

To view alarm settings:

• Touch CO2.• Touch VIEW ALARMS.

SpO2100 ETCO2 66 I CO2 8 C

O2

98 SpO2

92 0 0 %

PULSE 80 RESP 30 70 PULSE

0 6 bpm

Flexport System Interface Operations Manual 61-6

Page 673: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 6.25, 12.5, 25, and 50 mm/second.

The default setting is 12.5 mm/second.

To turn the waveform display ON or OFF:

• Touch CO2.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch CO2.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch CO2.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

Flexport System Interface Operations Manual 61-7

Page 674: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 61-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select a display format:

• Touch CO2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch CO2.• Select FREEZE ON.

To print the analyzer display:

• Touch CO2.• Touch PRINT.

Flexport System Interface Operations Manual 61-8

Page 675: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor N6000

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the analyzer or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the analyzer is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the analyzer during Flexport system interface operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOSTThe communication between the Flexport system interface and the 60-RS-SL interface is interrupted.

Yes

CO2 SENSOR WARMING The CO2 and AGENT sensors are warming up in the CO2 analyzer. No

SAMPLING LINE OCCLUDED The sample line is occluded or the pump is defective. Yes

APNEA No CO2 waveform has been detected for more than the set APNEA period. Yes

PULSE SEARCH Pulse is too weak. No

AIRWAY OCCLUDED The airway adapter is occluded. No

CO2 SENSOR NOT CONNECTED The CO2 sensor is not plugged into the Tyco Nellcor N6000. No

SPO2 SENSOR NOT CONNECTED The SpO2 sensor is not plugged into the Tyco Nellcor N6000. No

CO2 DISABLED CO2 monitoring has been disabled on the Tyco Nellcor N6000. No

SPO2 DISABLED SpO2 monitoring has been disabled on the Tyco Nellcor N6000. No

Flexport System Interface Operations Manual 61-9

Page 676: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 677: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 62-1

Tyco Nellcor Symphony N3000

Directory of KeysSPO2

SpO2 - MAIN MENUALARM LIMITS SETUP

FREEZE VIEW ALARMS PRINT

ON OFF

These keys appear on the bedside monitor only.

SpO2 - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

SpO2 - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

SpO2 - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec

SpO2 - ALARM LIMITS

SPO2 PR

SpO2 - ALARM LIMITS - PRPR HI =

XXLO = XXON OFF

SpO2 - ALARM LIMITS - SpO2SPO2 HI =

XXLO = XXON OFF

Page 678: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 679: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor Symphony N3000

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90442A Flexport system interface enables information from a Tyco Nellcor Symphony N3000 pulse oximeter to be displayed on Spacelabs Healthcare monitors.

Configure the pulse oximeter as shown Table 1.

The Flexport system interface is fully compatible with N3000 oximeters (N3000-N20, N3000-N21, and N3000-22). Other N3000 oximeters (for example, international oximeters) may require modifications. Refer to Directions For Use supplied with the GPSI adapter or contact Tyco Nellcor technical service for instructions.

Table 1: Pulse Oximeter Configuration

Baud Rate

Minimum Software Revision

Installation Kit

Part Number

Transition Connector

Part Number

9600 N3000: UIF 3.4.0 045-0037-xx 131-0686-xx (9 pin)

Flexport System Interface Operations Manual 62-3

Page 680: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor Symphony N3000

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the N3000 oximeter, complete the steps listed in the following Quickstart.

Refer to Directions For Use supplied with the GPSI adapter for instructions on connecting the N3000 oximeter to the GPSI adapter.

The Spacelabs Healthcare monitor is now ready to display information from the pulse oximeter. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the patient monitor.

Figure 62-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the RS-232 connector on the Nellcor GPSI

communications adapter.• Plug the round DIN connector of the GPSI cable into the patient monitor serial port.

� �

SpO2

98 SPO2

%

70 PR

bpm

� �

Flexport System Interface Operations Manual 62-4

Page 681: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor Symphony N3000

Figure 62-2: Bedside monitor, small text screen

Figure 62-3: Central monitor, split screen

� SPO2 parameter key

� Current SpO2 value

� Alarm status bell (appears when alarms are turned ON)

� Current pulse rate value

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the SPO2 key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

When all parameter alarms are turned OFF, alarm status messages will still cause the SPO2 key to flash.

Alarm settings default to ON.

� � �

SpO2

SPO2= 98%

PR= 70bpm

� �

� SpO2 = 98% SpO2

� PR = 70bpm

� ICU 14

Flexport System Interface Operations Manual 62-5

Page 682: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor Symphony N3000

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarms settings on the pulse

oximeter.

• Alarm limits can only be adjusted at the pulse oximeter. Alarm violation is detected at the oximeter and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the oximeter can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the pulse oximeter’s parameters (refer to Figure 62-4). If alarms for both parameters are turned OFF, the Spacelabs Healthcare monitor displays SPO2 ALM OFF to the right of the SPO2 key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 62-4: View Alarms screen

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch SPO2.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To view alarm settings:

• Touch SPO2.• Touch VIEW ALARMS.

SPO2 100 SPO2

98 SPO2

92 %

PR 80 70 PR

0 bpm

Flexport System Interface Operations Manual 62-6

Page 683: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor Symphony N3000

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch SPO2.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To change waveform size:

• Touch SPO2.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 62-7

Page 684: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor Symphony N3000

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 50, 25, 12.5, 6.25, 3.12, 1.56, 0.78, and 0.39 mm/second. The minimum recorder sweep speed is 1.56 mm/second.

The default setting is 25 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 62-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch SPO2.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch SPO2.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

Flexport System Interface Operations Manual 62-8

Page 685: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Nellcor Symphony N3000

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the pulse oximeter or cabling. If the Spacelabs Healthcare monitor displays this message, perform the following troubleshooting steps:

• Verify that the pulse oximeter is powered ON.

• Verify that all cables are secure.

The pulse oximeter serial port is not operational when the oximeter is operating on batteries.

GPSI Communications Adapter LED Status

The Nellcor GPSI adapter has two LED indicators labeled N3000 and HOST. They indicate the status of communications with the Flexport system interface as follows:

• If both LED indicators are ON, then the GPSI communications adapter, the pulse oximeter, and the Flexport system interface are all successfully communicating.

• If the HOST indicator is flashing and the pulse oximeter indicator is OFF, check the cabling to the Flexport system interface and the Spacelabs Healthcare monitor.

• If the N3000 indicator is flashing and the HOST indicator is ON, check the configuration on the pulse oximeter (for example, software version, RS-232 mode settings, or BAUD rate).

• If both LED indicators are flashing, the GPSI adapter has failed its self test and must be replaced. Contact Nellcor technical service for replacement.

To print the patient monitor display:

• Touch SPO2.• Touch PRINT.

Flexport System Interface Operations Manual 62-9

Page 686: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 687: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Flexport System Interface Operations Manual 63-1

Tyco Puritan Bennett 840

Directory of KeysVENT

VENT - MAIN MENUALARM LIMITS SETUP VIEW

MODE VIEW STATUS VIEW ALARMS PRINT

VENT - SETUP LARGE TXTSMALL TXT

VENT - ALARM LIMITS

RR Vte Vmin Ppeak APNEA O2 Src.

VENT - ALARM LIMITS - O2 Src.O2 Src. O2 Src.

ON OFF RR

VENT - ALARM LIMITS - APNEAAPNEA HI =

XXAPNEA

ON OFF O2 Src.

VENT - ALARM LIMITS - PpeakPpeak HI =

XXXPpeak

ON OFF APNEA

VENT - ALARM LIMITS - VminVmin LO =

XX.XVmin

ON OFF Ppeak

VENT - ALARM LIMITS - VteVte LO =

X.XXVte

ON OFF Vmin

VENT - ALARM LIMITS - RRRR HI =

XXRR

ON OFF Vte

Page 688: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 689: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 840

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Selecting a Display Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

OverviewThe 90436A-15 Flexport system interface allows information from a Tyco Puritan Bennett 840 ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting Stop Bit

TransitionConnector

Part Number

9600 8-bit none 1 131-1039-xx

Flexport System Interface Operations Manual 63-3

Page 690: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 840

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 63-1: Bedside monitor, large text screen

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the DIGITAL OUTPUT connector on the ventilator.

� � � � � �

1.8I:E

21Ppeak

cmH2O 0Pplat

cmH2O 21O2

% VENT

1.22Vte

L

10.5 PEEP

cmH2O8 Pmean

cmH2O30.0 Vmin

L 27RR

BPM

� � �

Flexport System Interface Operations Manual 63-4

Page 691: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 840

Figure 63-2: Bedside monitor, small text screen

Figure 63-3: Central monitor, split screen

� Ratio of inspiratory time to expiratory time

� Peak inspiratory pressure

� Inspiratory plateau pressure

� Oxygen level

� VENT parameter key

� Tidal volume

� Alarm status bell (appears when alarms are turned ON)

Respiratory rate

Minute volume

� Mean airway pressure

� Positive end expiratory pressure

Patient/room ID

� �

VENT

Ppeak= 21cmH2OPEEP= 10.5cmH2O �Vte= 1.22L �RR= 27BPM

� � � � �

RR= 27BPM Vte= 1.22L VENT

� O2 = 21% Vmin= 30.0L

� Ppeak= 21cmH2O I:E = 1.8

� PEEP= 10.5cmH2O Pmean= 8cmH2O

ICU 14

Flexport System Interface Operations Manual 63-5

Page 692: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 840

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to ON.

Note:• The expired tidal volume (Vte) alarm displayed on the Spacelabs Healthcare monitor reflects the

spontaneous low Vte alarm on the ventilator.

• The Spacelabs Healthcare monitor does not display the mandatory high Vte, mandatory low Vte, and spontaneous high Vte alarms from the ventilator.

• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits on the Flexport system interface can only be adjusted at the ventilator. Alarm violation is detected at the ventilator, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be temporarily silenced. If alarm tones are also OFF at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 63-6

Page 693: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 840

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 63-4). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 63-4: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 63-9).

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

Ppeak120

Vte 0.00 APNEA20

VENT

1.22 Vte

LVmin 0.3 RR 70

27 RR

BPM

Flexport System Interface Operations Manual 63-7

Page 694: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 840

Viewing Ventilator Mode HistoryChronological history of the last eight ventilator mode settings are displayed in the waveform zone with the oldest value at the top of the left column and the most recent value at the bottom of the right column. Possible modes include A/C, SPONT, SIMV, and BILEVEL.

Figure 63-5: View Mode screen

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 63-4).

The default setting is LARGE TXT.

To view ventilator mode history:

• Touch VENT.• Touch VIEW MODE.

To select a display format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

15:4715:47

A/CSIMV/PS

15:4915:49

SIMV/PSSPONT V

ENT

1.21 Vte

L15:47 A/C 15:51 A/C/PS15:49 A/C 15:56 SIMV/PS 27 RR

BPM

Flexport System Interface Operations Manual 63-8

Page 695: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 840

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

Status MessagesThe message CIRCUIT DISCONNECT indicates that the ventilator has detected a disconnected circuit (no ventilation). If the monitor displays this message, check the patient and reconnect the circuit.

The message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport operation.

To print the current display:

• Touch VENT.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

CIRCUIT DISCONNECT Ventilator detects circuit disconnect; patient ventilation has stopped. Yes

NO AIR/O2 SUPPLY Ventilator is not delivering breaths to the patient. Yes

LOW PRS O2 INLET

Pressure at the oxygen inlet fitting is less than 35 psig. An O2 Source alarm is also triggered. The ventilator generates an alert when O2 equals 21% and generates an alarm when O2 is 22% or greater.

Yes

LOW PRS AIR INLET Pressure at the air inlet is equal to or less than 35 psig. No

Flexport System Interface Operations Manual 63-9

Page 696: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 697: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Directory of KeysVENT

VENT - MENUALARM LIMITS SETUP

FREEZE VIEWSTATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 64-2

VENT - SETUP

SIZE SWEEP SPEED

WAVE LARGE TXTPRS FLO SMALL TXT

VENT - SETUP - SWEEP SPEED25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec0.39

mm/sec

VENT - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

Flexport System Interface Operations Manual 64-1

Page 698: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

VENT

VENT - MENUALARM LIMITS SETUP

FREEZE VIEWSTATUS

VIEW ALARMS PRINT

ON OFFRefer to

page 64-1

VENT - ALARM LIMITS

RR VT Vmin PIP PEEP I : E APNEA SpO2*

VENT - ALARM LIMITS - SpO2SpO2 HI =

XXLO = XX

RRON OFF SpO2

VENT - ALARM LIMITS - APNEAAPNEA HI =

XXAPNEA

ON OFF SpO2

VENT - ALARM LIMITS - I : EI : E LO =

X.XI : E

ON OFF APNEA

VENT - ALARM LIMITS - PEEPPEEP LO =

XXPEEP

ON OFF I : E

VENT - ALARM LIMITS - PIPPIP HI =

XXXLO = XXX

PIPON OFF PEEP

VENT - ALARM LIMITS - VminVmin LO =

XX.XVmin

ON OFF PIP

VENT - ALARM LIMITS - VTVT LO =

X.XXVT

ON OFF Vmin

VENT - ALARM LIMITS - RRRR HI =

XXRR

ON OFF VT

*Only SpO2 alarm limit is displayed when activated.

Flexport System Interface Operations Manual 64-2

Page 699: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Viewing Ventilator Mode History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90436A-01 and 90436A-41 Flexport system interfaces enable information from a Tyco Puritan Bennett 7200A/AE ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Baud Rate

Character Length

Parity Setting

Stop Bit

TransitionConnector

Part Number

Minimum Software Revision

9600 8-bit none 1 131-0685-xx 26300 Rev. E or higher

Flexport System Interface Operations Manual 64-3

Page 700: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Figure 64-1: Flexport system interface connections

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the DIGITAL OUTPUT connector on the ventilator.• If you are using the 90436A-41 Flexport system interface, plug the

9-pin connector into the ANALOG OUTPUT connector on the side of the ventilator.

SDLCcable

SDLCterminator

modularcable

Flexport

interface

ventilator

25-pin connectorto DIGITAL OUTPUT(back of ventilator)

9-pin connector

(with 90436A-41 Flexport system interface only)to ANALOG OUTPUT

SpacelabsHealthcare

monitor

system

Flexport System Interface Operations Manual 64-4

Page 701: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Note:Pressure or flow waveforms are available only with the 90436A-41 Flexport system interface.

Figure 64-2: Bedside monitor, large text screen

Figure 64-3: Bedside monitor, small text screen

Figure 64-4: Central monitor, split screen

� � � � � � �

1.8I:E

21PIP

cmH2O 0PLT

cmH2O 21FiO2

% VENT

1.22VT

L

WAVE=PRES

97SpO2

% 1PEEP

cmH2O8 MAP

cmH2O30.0 Vmin

L 27RR

BPM

� � �

� � � �

VENT

VT=1.22L WAVE=PRES

RR= 27BPMPIP= 21cmH2OPEEP= 1cmH2O

� � � � �

RR= 27BPM VT= 1.22L VENT

� FiO2 = 21% PIP= 21cmH2O

PEEP= 1cmH2O SpO2= 97%

� Vmin= 30.0L MAP= 8cmH2O

� ICU 14

Flexport System Interface Operations Manual 64-5

Page 702: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

� Ratio of exhalation time to inspiratory time

� Peak inspiratory pressure

� Inspiratory plateau pressure

� Fractional inspired O2

� VENT parameter key

� Tidal volume

� Current waveform mode (Flexport 90346A-41 only)

Alarm status bell (appears when alarms are turned ON)

Respiratory rate

� Minute volume

� Mean airway pressure

Positive end expiratory pressure

� O2 saturation

� Patient/room ID

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to OFF.

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

Flexport System Interface Operations Manual 64-6

Page 703: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits on the Flexport system interface can only be adjusted at the ventilator. Alarm violation is detected at the ventilator, and the Flexport system interface reports the alarm if it is turned ON for that parameter.

• Alarm tones on the ventilator can be selectively enabled or disabled. If alarm tones are also disabled at the Spacelabs Healthcare monitor(s), alarm violations will be reported visually only.

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters (refer to Figure 64-5). If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. The alarm bell flashes when an alarm limit is violated.

Figure 64-5: View Alarms screen

Viewing StatusThe VIEW STATUS key enables you to display the recent status messages sent by the ventilator. A maximum of four messages can be simultaneously displayed on the bedside monitor, and three messages can be simultaneously displayed on the remote monitor (refer to Status Messages on page 64-11).

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

To view ventilator status:

• Touch VENT.• Touch VIEW STATUS.

PIP120

3 SpO20

90 VT 0.00 APNEA20 V

ENT

1.22 VT

L

I : E 1.0 PEEP 0 Vmin 0.3 RR 70 27 RR

BPM

Flexport System Interface Operations Manual 64-7

Page 704: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Viewing Ventilator Mode HistoryChronological history of the last eight ventilator mode settings are displayed in the waveform zone with the oldest value at the top of the left column and the most recent value at the bottom of the right column. Possible modes include CMV, SIMV, CPAP, SIMV/PS, CPAP/PS, SIMV/FB, CPAP/FB, or UNKNOWN MODE.

Figure 64-6: View Mode screen

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Note:Pressure or flow waveforms are available only with the 90436A-41 Flexport system interface.

Turning Waveforms ON or OFFYou can turn the waveform display OFF and display only the numeric values. When the waveform is OFF, the SWEEP SPEED key, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

To view ventilator mode history:

• Touch VENT.• Touch VIEW MODE.

To turn the waveform display ON or OFF:

• Touch VENT.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

15:4715:47

CMVSIMV/PS

15:4915:49

SIMV/PSCMV V

ENT

1.21 VT

L15:47 CMV 15:51 CPAP/PS 25 RR

15:49 CMV 15:56 SIMV/PS BPM

Flexport System Interface Operations Manual 64-8

Page 705: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 25, 12.5, 6.25, 3.12, 1.56, 0.78, and 0.39 mm/second.

The default setting is 12.5 mm/second.

Selecting a Waveform TypeYou can select either pressure or flow waveforms for display. Select PRS to display the pressure waveform, or select FLO to display the flow waveform. The two waveform displays are mutually exclusive.

To adjust waveform size:

• Touch VENT.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

To select a sweep speed:

• Touch VENT.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a waveform type:

• Touch VENT.• Touch SETUP.• Touch WAVEFORM.• Select PRS or FLO.

Flexport System Interface Operations Manual 64-9

Page 706: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 64-5).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform on the display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the printout.

To select a display format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch LINK.• Select FREEZE ON.

To print the current display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 64-10

Page 707: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Tyco Puritan Bennett 7200A/AE

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the ventilator during Flexport operation.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the ventilator and the monitor has been interrupted. Yes

LOW PRS O2 INLET Pressure at the oxygen inlet fitting is less than 35 psig, and the O2% is 22 or greater. No

LOW PRS AIR INLET Pressure at the air inlet is equal to or less than 35 psig. No

LOW BATTERY Voltage of the internal batteries is too low to sustain one hour of operation. No

EXHAL VALVE LEAK Volume of exhalation gas is less than 10% of inspiration gas. No

LOW PULSE RATE Pulse rate is less than the set low pulse rate limit. No

HIGH PULSE RATE Pulse rate is greater than the set high pulse rate limit. No

DISCONNECT VENT Patient service system is disconnected or plugged. No

SPO2 PROBE OFF SpO2 probe is off the patient’s finger. No

POOR SPO2 SIGNAL SpO2 probe is dirty or probe is positioned incorrectly. No

SPO2 WRONG PLACE Placement of SpO2 probe is incorrect. No

Flexport System Interface Operations Manual 64-11

Page 708: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 709: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Transonics HT-107

Directory of KeysFLOW

FLOW - MAIN MENU

SETUPFREEZE

PRINTON OFF

FLOW - SETUP

SIZE SWEEP SPEED

LARGE TXT This key is displayed on the bedside monitor only.SMALL TXT

FLOW - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.79

mm/sec0.39

mm/sec

FLOW - SETUP - SIZEWAVE SIZE

↑SIZE

↓ON OFF

Flexport System Interface Operations Manual 65-1

Page 710: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 711: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Transonics HT-107

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Display Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Freezing the Waveform on the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Printing the Current Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

OverviewThe 90442A Flexport system interface enables information from a Transonics Systems HT-107 Medical Flowmeter to be displayed on Spacelabs Healthcare monitors.

Configure the flowmeter as shown in Table 1. Consult the flowmeter manual for additional configuration setup procedures

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the flowmeter, complete the steps listed in the following Quickstart.

The Spacelabs Healthcare monitor is now ready to display information from the flowmeter. To control other interface functions, refer to the sections that follow.

Table 1: Flowmeter Configuration

Baud Rate

Minimum Software Revision

Installation Kit Part Number

Transition Connector Part Number

9600 1.25 045-0030-xx 131-0684-xx (25-pin)

To connect the Flexport system interface:

• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface as shown in Figure 1-3 on page 1-3.

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the transition connector.• Plug the transition connector into the 25-pin connector on the back of the flowmeter.

Flexport System Interface Operations Manual 65-3

Page 712: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Transonics HT-107

Display DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the flowmeter.

Figure 65-1: Bedside monitor, large text screen

Figure 65-2: Bedside monitor, small text screen

Figure 65-3: Central monitor, split screen

� FLOW parameter key

� Current flow value

� Patient/room ID

� �

FLOW

1.20 BL FLOl/m

� �

FLOW

BLFLO = 1.20 l/m

� �

BL FLO = 1.20 l/m FLOW

� ICU 14

Flexport System Interface Operations Manual 65-4

Page 713: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Transonics HT-107

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the display size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to appear and function.

To turn the waveform display ON or OFF:

• Touch FLOW.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch FLOW.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 65-5

Page 714: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Transonics HT-107

Selecting a Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the display. Available sweep speeds are: 50, 25, 12.5, 6.25, 3.12, 1.56, 0.78, and 0.39 mm/second. The minimum recorder sweep speed is 1.56 mm/second.

The default setting is 25 mm/second.

Selecting a Display FormatTwo display formats are available for the bedside monitor (refer to Display Detail on page 65-4).

The default setting is LARGE TXT.

Freezing the Waveform on the DisplayThe FREEZE ON key enables you to freeze the waveform display. When you freeze the waveform, the Flexport system interface continues to process, update, and display numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch FLOW.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a display format:

• Touch FLOW.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform on the display:

• Touch FLOW.• Select FREEZE ON.

Flexport System Interface Operations Manual 65-6

Page 715: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Transonics HT-107

Printing the Current DisplayTouch the PRINT key to print the information shown on the current display. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

Status MessagesThe message COMMUNICATION LINK LOST indicates a problem with the flowmeter or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the flowmeter is powered ON.

• Verify that all cables are secure.

Table 2 lists the status messages that the Flexport system interface reports for the flowmeter during Flexport system interface operation.

To print the current display:

• Touch FLOW.• Touch PRINT.

Table 2: Status Messages

Message Text Indication Alarm Sounds

COMMUNICATION LINK LOST

Communication between the flowmeter and the monitor has been interrupted. Yes

ACOUSTIC ERROR Indicates a break in either the ultrasonic or electrical signal path. No

Flexport System Interface Operations Manual 65-7

Page 716: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 717: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Directory of KeysVENT

VENT - MENUALARM LIMITS SETUP

FREEZEWAVEFORMS VIEW

ALARMS PRINTON OFF

VENT - SETUP

SIZE SWEEP SPEED

LARGE TXTSMALL TXT

VENT - SETUP - SWEEP SPEED50

mm/sec25

mm/sec12.5

mm/sec6.25

mm/sec3.12

mm/sec1.56

mm/sec0.78

mm/sec0.39

mm/sec

VENT - SETUP - SIZEWAVEFORM SIZE

↑SIZE

↓ON OFF

VENT - ALARM LIMITS

Apnea Ppeak Ve

VENT - ALARM LIMITS - VeVe LO =

XXXXON OFF

VENT - ALARM LIMITS - PpeakPpeak HI =

XXXXLO = XXXXON OFF

VENT - ALARM LIMITS - ApneaApnea HI =

XXXXON OFF

These keysappear on thebedsidemonitor only.

Flexport System Interface Operations Manual 66-1

Page 718: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 719: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Contents

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Connecting the Flexport System Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Screen Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Enabling Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Viewing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Alarm Recordings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Entering Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Printing the Current Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

OverviewThe 90436A-18 Flexport system interface enables information from a VIASYS Vela Ventilator to be displayed on Spacelabs Healthcare monitors.

Configure the ventilator as shown in Table 1. Consult the ventilator manual for additional configuration setup procedures.

Table 1: Ventilator Configuration

Spacelabs Healthcare Model Number/Ventilator Name and Model Number

Baud Rate Cabling Adapter Minimum

Software Version

Spacelabs Healthcare 90436A-18 9600 9-pin modular adapter

required (P/N 131-0686-03) (or 25-pin modular adapter with 25-pin/9-pin converter)

V1.00.01EN

Bird Monitor Interface15674 9600 V3.3

Flexport System Interface Operations Manual 66-3

Page 720: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Connecting the Flexport System InterfaceTo connect the Flexport system interface to the ventilator, complete the steps listed in the following Quickstart.

Figure 66-1: Flexport system interface connections

To connect the Flexport system interface:

• Plug one end of the modular cable into the modular jack on the Flexport system interface.

• Plug the other end of the modular cable into the 9-pin modular adapter, then into the jack labeled TO MONITOR on the BMI.

• Plug one end of a fiber optic cable into the fiber optic jack labeled DATA IN on the BMI.• Plug the other end of the fiber optic cable into the fiber optic jack labeled DATA LINK

on the ventilator.• Ensure that all screw locks are tight.• Connect the wire from the power adapter to the jack labeled INPUT 9/15 VDC on the

BMI.• Turn ON the ventilator, then plug in the BMI interface power adapter.• Connect the long branch of the SDLC cable to the SDLC connector on the back of the

monitor.• Attach the SDLC terminator to the SDLC cable (refer to Figure 66-1).• Connect the SDLC cable to the 9-pin male connector on the Flexport system interface.

SDLCcable

SDLCterminator

Flexport

interface

Bird

SpacelabsHealthcare

monitor

system

monitor interface

VIASYS Ventilator, Vela

power adapter

modularcable

9-pinadapter

data indata link

optional data out connection

powerconnector

Flexport System Interface Operations Manual 66-4

Page 721: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Caution:Attach only Spacelabs Healthcare-approved accessories to RS-232 connectors.

Note:• The maximum total SDLC cable length between any Flexport system interface and its power source

should not exceed 6 feet (1.8 m). The power source is the monitor or the module housing.

• If you use SDLC links to connect other devices (such as expansion housings or printers) to the monitor, the Flexport system interface must be the device closest to the monitor.

The Spacelabs Healthcare monitor is now ready to display information from the ventilator. To control other interface functions, refer to the sections that follow.

Screen DetailWhen you power ON a Spacelabs Healthcare monitor that is properly connected to a Flexport system interface, the monitor provides the following information from the ventilator.

Figure 66-2: Bedside monitor, large text

Figure 66-3: Bedside monitor, small text

� Peak inspiratory pressure — displays high pressure alarm

� Mean airway pressure

� Respiratory rate

� VENT parameter key

� Inspiratory tidal volume

� � � � � �

999 Ppeak 99.9 Pmean 999 Rate VENT

9999 Vti

cmH2O cmH2O bpm ml

99 PEEP 99.9 Ti 99.9 Ve 9999 Vte

cmH2O Sec L ml

� �

� �

VENT

Ppeak=999cmH2O

PEEP=99cmH2O �

Vte=9999ml

Ve=99.9L

� �

Flexport System Interface Operations Manual 66-5

Page 722: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

� Alarm status bell (any)

� Exhaled tidal volume

Exhaled minute volume

Inspiratory time

� Positive end expiratory pressure

Enabling AlarmsThe Alarm Limits menu allows you to enable alarm limits for each parameter. When an alarm is turned ON and that parameter value exceeds an alarm limit, all of the following occur:

• an alarm tone sounds,

• the VENT key flashes,

• the alarm limits key for that parameter flashes, and

• the alarm bell flashes.

Alarm settings default to ON.

Note:• Turning Flexport system interface alarms ON or OFF does not affect alarm settings on the ventilator.

• Alarm limits on the Flexport system interface can only be adjusted at the ventilator and are only shown on the Spacelabs Healthcare monitor for reference.

Apnea AlarmThe Apnea alarm is a high limit alarm, which displays if a breath isn’t detected within a preset period of time. The alarm limit is adjustable, and may be set at the Vela ventilator. Figure 66-4 displays a recording of an Apnea HI Limit alarm at 20 seconds. Notice that the unit of measure (seconds) is not included in the recorder strip.

Figure 66-4: Apnea HI Limit alarm

Caution:• The Apnea HI Limit alarm doesn’t display an associated parameter on the bedside monitor.

• The breaths per minute rate may appear abnormally high when an Apnea HI Limit alarm occurs.

Refer to the Vela Ventilator System, Operator’s Manual for further information.

Flexport System Interface Operations Manual 66-6

Page 723: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Viewing Alarm SettingsThe VIEW ALARMS key enables you to display the current alarm settings for the ventilator’s parameters. If alarms for all parameters are turned OFF, the Spacelabs Healthcare monitor displays VENT ALM OFF to the right of the VENT key. If alarms are turned ON for any parameter, a bell is displayed. (An alarm bell also is displayed, on the large text view, next to the parameters for which alarms are turned ON.) The alarm bell flashes when an alarm limit is violated.

The alarm limits on the VIEW ALARMS display correspond to the limits which are set for each of the parameters on the ventilator.

• If the ventilator displays the low Ppeak limit as OFF, the low Ppeak limit in VIEW ALARMS displays as zero (0).

• If the ventilator displays the low VE limit as OFF, the low VE limit in VIEW ALARMS displays as 0.1.

Alarm RecordingsAlarm recordings that are sent to a printer will display only one parameter at a time. If multiple parameters are simultaneously in alarm, the precedence on the display is:

• Apnea

• Ppeak

• VE

To turn alarms ON/OFF from the Spacelabs Healthcare monitor:

• Touch VENT.• Select ALARM LIMITS.• Select an alarm parameter.• Touch the parameter alarm ON/OFF key.

To view alarm settings:

• Touch VENT.• Touch VIEW ALARMS.

Flexport System Interface Operations Manual 66-7

Page 724: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Entering Setup InformationChanging setup information is helpful in optimizing the monitor display. The SETUP key controls the functions described below and affects only the monitor at which these adjustments are made.

Turning Waveforms ON or OFFYou can turn waveforms OFF and display only the numeric values. When waveforms are OFF, the SWEEP SPEED key, the FREEZE ON/OFF keys, and the SIZE ↑ and SIZE ↓ keys are disabled.

The default setting is WAVEFORM ON.

Adjusting the Waveform SizeIf the waveform is too large to fit within the display zone, use the waveform size keys to adjust the screen size.

Note:The WAVEFORM ON/OFF key must be set to ON for the size keys to function.

To turn the waveform screen ON or OFF:

• Touch VENT.• Touch SETUP.• Touch SIZE.• Select WAVEFORM ON/OFF.

To adjust waveform size:

• Touch VENT.• Touch SETUP.• Touch SIZE.• Touch SIZE ↑ or SIZE ↓ to adjust the waveform size.

Flexport System Interface Operations Manual 66-8

Page 725: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Selecting a Screen Sweep SpeedThe sweep speed determines the speed at which the waveform trace moves across the screen. Available sweep speeds are: 0.39, 0.78, 1.56, 3.12, 6.25, 12.5, 25, and 50 mm/second.

The default setting is 12.5 mm/second.

Note:Sweep speeds below 1.56 mm/sec are supported on the display, but are not supported on the recordings.

Selecting a Screen FormatTwo screen formats are available for the bedside monitor (refer to Screen Detail on page 66-5).

The default setting is LARGE TXT.

Freezing the Waveform DisplayThe FREEZE ON/OFF key enables you to freeze the waveform display. When you freeze the waveform, the interface continues to process and update the numeric information.

The default setting is FREEZE OFF.

To select a sweep speed:

• Touch VENT.• Touch SETUP.• Touch SWEEP SPEED.• Select the desired speed.

To select a screen format:

• Touch VENT.• Touch SETUP.• Select LARGE TXT or SMALL TXT.

To freeze the waveform display:

• Touch VENT.• Touch FREEZE ON/OFF.

Flexport System Interface Operations Manual 66-9

Page 726: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Selecting a WaveformThe WAVEFORMS key enables you to select a pressure, a flow, or a volume waveform. Touch the WAVEFORMS key until the desired type of waveform displays. The default setting displays the pressure waveform.

Note:This key is only available at the bedside monitor.

Volume Waveform

The volume waveform displays on a Spacelabs Healthcare monitor at 10% of the value displayed on the VIASYS Ventilator. If the value on the VIASYS Ventilator is 800, it will display at 80 on the monitor.

All recordings generated using the RECORD key or as a result of an alarm condition are scaled in this manner.

Printing the Current ScreenTouch the PRINT key to print the information shown on the current screen. Each recording is identified by the bed name, patient name, time, and date.

Note:• The patient name you enter in the Spacelabs Healthcare monitor is the name that appears on the

printout.

• The PRINT key does not print the waveform. A waveform recording is obtained by touching the monitor RECORD key.

To select a waveform:

• Touch VENT.• Touch WAVEFORMS.

To print the anesthesia machine display:

• Touch VENT.• Touch PRINT.

Flexport System Interface Operations Manual 66-10

Page 727: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

VIASYS Ventilator, Vela

Status MessagesTable 2 lists the status messages that the Flexport interface reports for the ventilator during operation.

Status Alarms

Status Alarms are status messages, which have “alarm-like” characteristics such as:

• A flashing open door key.

• An audible tone.

The COMMUNICATION LINK LOST message is an example of a Status Alarm.

The message COMMUNICATION LINK LOST indicates a problem with the ventilator or cabling. If the monitor displays this message, perform the following troubleshooting steps:

• Verify that the ventilator is turned ON.

• Verify that all cables are secure.

Note:The COMMUNICATION LINK LOST message displays for approximately 40 seconds prior to the channel disappearing from the screen.

Table 2: Status Messages

Message Text Indication Alarm Sounds

PRESSURE WAVEFORM SELECTED The current waveform is a pressure waveform. No

FLOW WAVEFORM SELECTED The current waveform is a flow waveform. No

VOLUME WAVEFORM SELECTED The current waveform is a volume waveform. No

Flexport System Interface Operations Manual 66-11

Page 728: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,
Page 729: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

The following list of international and safety symbols describes all symbols used on Spacelabs Healthcare products. No one product contains every symbol.

HELP Key Keyboard Connection

SPECIAL FUNCTIONS Key Mouse Connection

RECORD Key START/STOP Key

NORMAL SCREEN Key START/STOP

MONITOR SETUP Key STOP or CANCEL Key

ALARMS Key CONTINUE Key

PREVIOUS MENU Key ENTER Key

ON — Power Connection to Mains OFF — Power Disconnection from Mains

ON Position for Push Button Power Switch

OFF Position for Push Button Power Switch

On Direction ON/OFF

ON — Part of the Instrument Only OFF — Part of the Instrument Only

Standby STANDBY KeyPower ON/OFF Key

Book Title Variable A-1

Page 730: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

PAUSE or INTERRUPT Slow Run

Alarm Reset Power Indicator LED

Alarm Audio ON Alarm Audio OFF

Alarm Audio Paused Activate Telemetry Recorder

Indicator — Remote Control Indicator — Local Control

PRINT REPORT Key Indicator — Out of Paper

Partial ON/OFF Recorder Paper

Normal Screen Return to Prior Menu

Clock/Time Setting Key TREND/TIMER Key

HELP (Explain Prior Screen) Key Keypad

Activate Recorder for Graphics Indoor Use Only

START (NIBP) Key Auto Mode (NIBP)

Television; Video Display Video Output

Output (Non-terminated) No Output (Terminated)

123 1

23

?

Book Title Variable A-2

Page 731: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

Data Input/Output Input/Output

Input Reset

Menu Keys Waveform/Parameter Keys

Monitor SetupSelect Program Options Set Initial Conditions Menu

Access Special Function Menu Return Unit to Monitor Mode

Serial Port 1 Serial Port 2

External Marker Push Button Connection SDLC Port

Arterial Pulse Electrocardiograph or Defibrillator Synchronization

Gas Exhaust Foot Switch

Enlarge, Zoom Delete

PCMCIA Card Event

Keep Dry Fragile; Handle with Care

Environmental Shipping/Storage Altitude Limitations This Way Up

Environmental Shipping/Storage Temperature Limitations

Environmental Shipping/Storage Humidity Limitations

123

123

A

123

B123

1 2

SDLC

x

12,200 m

Book Title Variable A-3

Page 732: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

Open Padlock Closed Padlock

Down Arrow Up Arrow

Hard Drive Power Indicator LED

Antenna Mermaid Connector

Microphone Omnidirectional Microphone

Audio Output, Speaker Universal Serial Bus

Network Connection Oxygen reference gas port

Gas Sampling Port Gas Return Port

Low Priority Alarm Nurse Call

High Priority Alarm Medium Priority Alarm

Alarms Paused Nurse Alert Interface

Battery Status Alarm OFF

BatteryReplace only with the appropriate battery.

Low Battery

Ref.

Book Title Variable A-4

Page 733: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

All batteries should be disposed of properly to protect the environment. Lithium batteries should be fully discharged before disposal. Batteries such as lead-acid (Pb) and nickel-cadmium (Ni-Cd) must be recycled. Please follow your internal procedures and or local (provincial) laws regarding disposal or recycling.

Replace only with the appropriate battery.(+ / - signs may be reversed)

Caution - hazardous voltages. To reduce risk of electric shock, do not remove the cover or back. Refer servicing to a qualified field service engineer (U.S.A.).DANGER - High Voltage (International)

This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.

Protective Earth Ground Functional Earth Ground

Replace Fuse Only as Marked Fuse

Power supply jack polarity. (+ / - signs may be reversed) Equipotentiality Terminal

Alternating Current Direct Current

Both Direct and Alternating Current AC/DC Input

Amperes Hertz

Volts Watts

IEC 60601-1 Type B equipment. The unit displaying this symbol contains an adequate degree of protection against electric shock.

IEC 60601-1 Class II equipment, double-isolated. The unit displaying this symbol does not require a grounded outlet.

A Hz

V W

Book Title Variable A-5

Page 734: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

IEC 60601-1 Type BF equipment which is defibrillator-proof. The unit displaying this symbol is an F-type isolated (floating) patient-applied part which contains an adequate degree of protection against electric shock, and is defibrillator-proof.

IEC 60601-1 Type BF equipment. The unit displaying this symbol is an F-type isolated (floating) patient-applied part providing an adequate degree of protection against electric shock.

IEC 60601-1 Type CF equipment. The unit displaying this symbol is an F-type isolated (floating) patient-applied part providing a high degree of protection against electric shock, and is defibrillator-proof.

IEC 60601-1 Type CF equipment. The unit displaying this symbol is an F-type isolated (floating) patient-applied part providing a high degree of protection against electric shock.

Loop Filter Adult Noninvasive Blood Pressure (NIBP)

ETL Laboratory Approved Canadian Standards Association Approved

Risk of Explosion if Used in the Presence of Flammable Anesthetics

Operates on Non-Harmonized Radio Frequencies in Europe

Note Note Caution - Consult Accompanying Documents

Warning Warning About Potential Danger to Human Beings Caution Caution About Potential Danger to a

Device

Noninvasive Blood Pressure (NIBP), Neonate Fetal Monitor Connection (Analog)

Fetal Monitor ConnectionRS-232 (Digital)

Physiological Monitor Connection RS-232 (Digital)

Symbol Set, Adult/Pediatric Cuff Sizes

Symbol Set, Neonatal Cuff Sizes

® U S

!!

Book Title Variable A-6

Page 735: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

NIBP Cuff, Neonatal 1 NIBP Cuff, Neonatal 2

NIBP Cuff, Neonatal 3 NIBP Cuff, Neonatal 4

NIBP Cuff, Neonatal 5 NIBP Cuff, Single Hose

NIBP Cuff, Dual Hose Consult Instructions For Use

NIBP Cuff, Surface Applied to Patient

NIBP Cuff, Child Size, Long (12 to 19 cm)

NIBP Cuff, Small Adult Size, Long (17 to 25 cm)

NIBP Cuff, Small Adult Size (17 to 25 cm)

NIBP Cuff, Adult Size, Long (23 to 33 cm)

NIBP Cuff, Large Adult Size, Long (31 to 40 cm)

NIBP Cuff, Large Adult Size (31 to 40 cm)

NIBP Cuff, Adult Size (23 to 33 cm)

NIBP Cuff, Child Size (12 to 19 cm)

NIBP Cuff, Infant Size (8 to 13 cm)

NIBP Cuff, Neonatal 1 Size (3 to 6 cm)

THIS SIDE TO PATIENT

CHILD, LONG

SMALL ADULT, LONG

SMALL ADULT

ADULT, LONG

LARGE ADULT, LONG

LARGE ADULT

ADULT CHILD

INFANT NEONATAL 1

Book Title Variable A-7

Page 736: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

NIBP Cuff, Neonatal 2 Size (4 to 8 cm)

NIBP Cuff, Neonatal 3 Size (6 to 11 cm)

NIBP Cuff, Neonatal 4 Size(7 to 13 cm)

NIBP Cuff, Neonatal 5 Size(8 to 15 cm)

NIBP Cuff, Thigh Size (38-50 cm) NIBP Cuff, Soft Material

NIBP Cuff, Nylon Material NIBP Cuff, Vinyl Material

Place Artery Symbol and Arrow over Brachial or Femoral Artery Quantity

Happy Face Sad Face

Magnifying Glass Compression

File Cabinet List of Rooms

Arrows Printer

Recycle Service Message

Non Sterile PVC-Free (Polyvinyl Chloride)

Latex-Free Do Not Reuse; Single Use Only

Radio transmitting device; elevated levels of non-ionizing radiation Reusable

Batch Code Reference Number or Order Number

NEONATAL 2 NEONATAL 3

NEONATAL 4 NEONATAL 5

THIGH SOFT

NYLON VINYL

ARTERY QTY

PVC

LATEX 2

LOT REF

Book Title Variable A-8

Page 737: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

Abbreviations used as symbols are shown below.

Date of Manufacture Nellcor Oxisensor II Compatible

UL recognized component in Canada and United States Novametrix Compatible

Nellcor OxiMax Compatible Spacelabs TruLink Compatible

Masimo SET Compatible Nellcor OxiMax Compatible

Spacelabs Compatible

1 - 32 Access Codes 1 Through 32 AIR Air

ANT 1ANT 2

Diversity Antenna System 1Diversity Antenna System 2

Arr1ArrNet2

Arrhythmia Net 1Arrhythmia Net 2

avDO2 Arterial/Venous Oxygen Difference CaO2 Arterial Oxygen

CHch

EEG, EMG, or ECG ChannelEEG Channels - CH1, CH2, CH3, CH4EMG Channel - CH5

cmH2O Centimeters of Water

C.O.COco

Cardiac Output CvO2 Venous Oxygen

DIAdia Diastolic ECG

ecg Electrocardiogram

EEGeeg Electroencephalogram EMG

emg Electromyogram

ESIS Electrosurgical Interference Suppression EXT External

NE2

NVX

R

Book Title Variable A-9

Page 738: 904XX Operations Manual - Spacelabs Healthcaremanuals.spacelabshealthcare.com/wp-content/uploads/2013/10/070... · ... Clinical Browser, Data Shuttle, Flexport, Inte sys, Mermaid,

Appendix A — Symbols

FECG Fetal Electrocardiogram FHR1FHR2

Fetal Heart Rate, Channel 1Fetal Heart Rate, Channel 2

GNDgnd Ground Hgb Hemoglobin

HLOhlo High-Level Output Multiview Multi-Lead Electrocardiogram

NIBPnibp Noninvasive Blood Pressure N2O Nitrous Oxide

O2 Oxygen O2AV Oxygen Availability

PaO2 Partial Pressure of Arterial OxygenPRESSpressPRS

Pressure

PvO2Partial Pressure of Mixed Venous Oxygen

RESPresp Respiration

SDLC Synchronous Data Link Control

SPO2SpO2SpO2SaO2

Arterial Oxygen Saturation as Measured by Pulse Oximetry

SVO2SvO2SvO2

Mixed Venous Oxygen Saturation SYSsys Systolic

T1T2T3T4

Temperature 1Temperature 2Temperature 3Temperature 4

TEMPtemp Temperature

UA Uterine Activity or Umbilical Artery VO2 Oxygen Consumption

VAC Vacuum Connection UV Umbilical Venous

Book Title Variable A-10