22
A Crystal Shashlik Electromagnetic Calorimeter for Future HEP Experiments Ren-Yuan Zhu California Institute of Technology May 17, 2016 Talk in the XVII International Conference on Calorimetry in Particle Physics, Daegu

A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

A Crystal Shashlik Electromagnetic Calorimeter for Future HEP Experiments

Ren-Yuan ZhuCalifornia Institute of Technology

May 17, 2016

Talk in the XVII International Conference on Calorimetry in Particle Physics, Daegu

Page 2: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Collaborators

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 2

Page 3: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Why Crystal Calorimetry in HEP?• Photons and electrons are fundamental particles.

Precision e/γ measurements enhance physics discovery potential for HEP experiments.

• Performance of crystal calorimeter in e/γmeasurements is well understood:– The best possible energy resolution;– Good position resolution;– Good e/γ identification and reconstruction efficiency.

• A crystal Shashlik ECAL concept was developed to preserve precision e/γ measurements and face the challenge of the severe radiation environment expected by the future HEP experiments at the energy frontier.

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 3

Page 4: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Physics with Crystal Calorimetry Charmonium system observed

by CB through Inclusive photonsHiggs -> γγ by CMS through reconstructing photon pairs

→ γγ

Η → γγ

CMS PWOCB NaI(Tl)

γγ bumps at 750 GeV reported in this conference hint the importance of the EM resolution at HL-LHCMay 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 4

Page 5: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Degradation of CMS PWO of 25 X0

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 5

Ionization dose induced dose rate dependent damage in PWO is well understood

R. T. de Lima, Talk in this conference, 5/16/2016

Page 6: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Reduce Light Path Length

Reduced Crystal Cost & Damage

Issues: Radiation hardness of the photo-detector and Cost

CMS ECAL endcap: Single Crystal: 160 cm3

Total number: 16,000 Total Volume: 3 m3

Issue: Radiation hardness of the photo-detector

Issues: Radiation hardness of WLS fibers & photo-detector

114 mm

14 mmW 2.5 mm

LYSO 1.5 mm

One option for CMS FCAL Upgrade

R.-Y. Zhu, Talk in CMS Forward Calorimeter Taskforce Meeting, CERN, 6/17/2010

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 6

Page 7: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

An Option for CMS FCAL Upgrade

A Front View

A Side Cut View

Typical Shashlik resolution: 10%/√E + 1%

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 7

Page 8: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

A 4 x 4 LYSO/W/Y-11 Array for 2014 BT

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 8

Page 9: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

A LYSO/W/Y-11 Tower for 2014 BTCoupled to PMT LYSO Plates

(14×14×1.5 mm)W Plates

(14×14×2.5 mm)

Monitoring fiber Monitoring fiber beam dump

4 Y-11 WLS fibers

Aluminum foil plus Tyvek paper wrapping

Aluminum Foil(14×14×0.015 mm)

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 9

Page 10: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Resolution in CERN H4 Beam

∆(Ε)/Ε : ∼1.5 %

Good resolution achieved at high energies

Preliminary

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 10

Page 11: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

LYSO/W/Quartz Capillary ShashlikQuartz Capillary

14 x 14 x 114 mm2

1/2 Franc LYSO Plate

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 11

Page 12: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

A LYSO/W/Capillary Tower for 2015 BT

A 4 x 4 matrix will be tested at

CERN in 2016

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 12

Page 13: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Bright, Fast Scintillator: LSO/LYSO Crystal NaI(Tl) CsI(Tl) CsI BaF2 BGO LYSO(Ce) PWO PbF2

Density (g/cm3) 3.67 4.51 4.51 4.89 7.13 7.40 8.3 7.77

Melting Point (ºC) 651 621 621 1280 1050 2050 1123 824

Radiation Length (cm) 2.59 1.86 1.86 2.03 1.12 1.14 0.89 0.93

Molière Radius (cm) 4.13 3.57 3.57 3.10 2.23 2.07 2.00 2.21

Interaction Length (cm) 42.9 39.3 39.3 30.7 22.8 20.9 20.7 21.0

Refractive Index a 1.85 1.79 1.95 1.50 2.15 1.82 2.20 1.82

Hygroscopicity Yes Slight Slight No No No No No

Luminescence b (nm) (at peak)

410 550 310 300220

480 402 425420

?

Decay Time b (ns) 245 1220 26 6500.9

300 40 3010

?

Light Yield b,c (%) 100 165 4.7 364.1

21 85 0.30.1

?

d(LY)/dT b (%/ ºC) -0.2 0.4 -1.4 -1.90.1

-0.9 -0.2 -2.5 ?

Experiment Crystal Ball

BaBar BELLEBES III

KTeVS. BELLE

Mu2e

(GEM)TAPS

Mu2e-II

L3BELLEEIC?

COMET{Mu2e,SuperB)

CMS?

CMSALICEPANDA

A4g-2

HHCAL?

a. at peak of emission; b. up/low row: slow/fast component; c. QE of readout device taken out.May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 13

Page 14: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Bright, Fast & Rad Hard LYSO• LYSO is a bright (200 times of PWO),

fast (40 ns) crystal scintillator widely used in the medical industry with mass production capability. It is also the most radiation hard crystal.

• Damage in LYSO does not recover, leading to a stable calorimeter.

• ɣ-ray and proton induced absorption coefficient is about 3 m-1 for 150 Mrad or 3 x 1014 p/cm2, leading to a robust calorimeter with few percent light output loss at the HL-LHC.

623 SIC LYSO Platesof 14 x 14 x 1.5 mm

with Five Holes

2.5 x 2.5 x 20 cm

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 14

See two talks on May 19, 2016, in this conference for the details

Page 15: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

RIAC at 430 nm in LYSO

Consistent RIAC at 430 nm is

observed in LYSO and LFS plates

irradiated by 24 GeV protons up to 8.19 x 1015p/cm2

at CERN in 2014 and 2015.

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 15

3 m-1 @ 3E14

Page 16: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

LO: Direct & Y-11 Couplings

Data consistent with average

light path length of 1.1 and 2.4 cm at

430 nm for direct and

Y-11 readout respectively.

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 16

3 m-1 @ 3E14

6% light output loss in LYSO plates @ 3E14 with Y-11 readout

Page 17: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

LFS/W/Capillary: 800 MeV ProtonsA Shashlik tower irradiated to 1.2×1015 p/cm2 in 3 steps with

degradation of 20%/50% after 4.3×1014/1.24×1015 p/cm2

0.940

0.801

0.506

0.00

0.10

0.20

0.30

0.40

0.50

0.60

0.70

0.80

0.90

1.00

1.10

0.00E+00 2.00E+14 4.00E+14 6.00E+14 8.00E+14 1.00E+15 1.20E+15 1.40E+15

Nor

mal

ized

Mon

itorin

g Si

gnal

800 MeV Proton Fluence (p/cm2)

LYSO-W-Capillary Shashlik CellMonitoring Source: 420 nm LEDReadout: quartz fiber and PMT.

~10% light output loss in a LYSO/capillary based Shashlik tower @ 3E14

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 17

Page 18: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Radiation Hardness: Quartz CapillaryRe

lativ

e Li

ght Y

ield

0

0.

2

0.4

0.6

0

.8

Distance from LED to Photosensor (mm)

0 20 40 60 80 100 120 140

50Mrad/0Mrad

100Mrad/0Mrad

150Mrad/0Mrad

200Mrad/0Mrad

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 18

Page 19: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Temp = 23c

Small Bias Shift of 0.5V

Low PDE change

Dark Current 1mA/mm2

Gain reduction due to heating of the device

FBK 1mm2 device with 12 micron pixels

Radiation Hardness: SiPM Irradiation to 1.3 x 1014 p/cm2

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 19

Page 20: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Radiation hardness: GaInPGeneration 3” first GaInP devices compared to GaAs

Generation 5 followed these with device structure further optimized for increased radiation tolerance

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 20

Page 21: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Alternative: LuAG:Ce Ceramics

Sample ID Dimension (mm3)

LuAG S1 25×25×0.4

LuAG S2 25×25×0.4

LuAG S1 LuAG S2 LuAG R1

Rad hard up to 220 Mrad

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 21

Page 22: A Crystal Shashlik Electromagnetic Calorimeter for Future HEP …zhu/talks/ryz_160517_calor16... · 2016. 5. 16. · A Crystal Shashlik Electromagnetic Calorimeter for Future HEP

Summary • Precision e/γ by total absorption crystal calorimeters

provide excellent physics discovery potential. • To preserve precision e/γ and face the challenge of the

radiation environment expected by future HEP experiments at the energy frontier, a crystal based Shashlik calorimeter reduces the light path length in crystals and thus enhances the radiation hardness.

• Bright, fast and radiation hard LSO/LYSO crystals provide a solid foundation for a stable and robust Shashlik calorimeter. Developments on quartz capillary based WLS and photo-detectors are very encouraging.

• Scintillating ceramics and glasses, such as LuAG:Ce ceramics, may play an important role for a cost-effective Shashlik calorimeter concept.

May 17, 2016 Presented by Ren-Yuan Zhu of Caltech in 2016 Calor Conference at Daegu, Korea 22