A New Method for Measuring Ammonia and TKN in Wastewater

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Page 2: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

Table of Contents

2

▰ Current automated ammonia & TKN methods

▰Why a new method?

▰ Operation

▰ Performance

▰ Approval

▰ Conclusions and questions

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Page 3: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

HELLO!I am Dr. Ilkka Lahdesmaki

Chief ScientistYou can find me at [email protected]

3

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Page 4: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

CURRENT METHODSAutomated methods forammonia and TKN

4

1Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Page 5: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

CURRENT METHODS

▰ Flow methods (FIA, SFA) achieve a high degree of automation

5

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

PUMP

R1

R2

WC

SW

INJECTION

VALVE

REACTION

COIL

REACTION

COIL

HEATED

DETECTOR

Reagent 1 -

Reagent 2 -

Carrier -

Page 6: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

CURRENT METHODS

▰ Use of gas diffusion to replace distillation

6

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

▰ Detection of diffusion-separated ammonia▻ Phenolate / salicylate + photometric detection (EPA 351.2, SM 4500-Norg D.)

▻ Conductimetric detection (EPA Timberline-Ammonia-001)

SAMPLE

ALKALI

NH4+

NaOH / KOH

NH3

ACCEPTORNH3

GAS-PERMEABLE

MEMBRANE

Page 7: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

NEW METHODWhat is the new method?

Why would a new method be needed?7

2Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Page 8: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

NEW METHOD – HOW DOES IT WORK?

▰ Based on:▻ o-phthalaldehyde indicator

▻ fluorescence detection

8

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Felix et al.

J. Braz. Chem. Soc., 23 (1), 142-147, 2012

365 nm excitation

430 nm emission

Page 9: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

NEW METHOD – HOW DOES IT WORK?

9

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

NaOH

OPA

Page 10: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

NEW METHOD – WHY INTRODUCE IT?

▰ Extensive prior use in oceanography

10

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169.

CRC Press, Boca Raton, 2009.

Page 11: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

NEW METHOD – WHY INTRODUCE IT?

▰ Used by customers for ammonia and TKN measurements▻ Oceanography research groups

▻ International environmental laboratories

▻ Very positive feedback: no phenol, more sensitive, stable baseline, consistent

11

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

▰ Bring method available to accredited labs in the U.S.▻ EPA Alternative Test Procedure (ATP) application

▻ Validation for use in wastewater analysis

Page 12: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

NEW METHOD – APPROVAL PROCESS

12

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

▰ Recommendation letter from EPA – 10 Apr 2018

▰ Inclusion in 2019 Methods Update Rule proposal – 11 Jun 2019

Page 13: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

PRACTICAL USEHow to implement on FIA instrument?

Performance metrics13

3Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Page 14: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

PRACTICAL USE – INSTRUMENT CONFIGURATION

14

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

9 in

6 in 5.5 in

Pictured:

FIAlyzer-1000 with PMT Detector

Page 15: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

PRACTICAL USE – DATA OUTPUT

15

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Calibration 50 – 10,000 µg N / L Single injection 50 µg N / L

Page 16: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

PRACTICAL USE – DATA OUTPUT

16

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

2nd order fit 50 – 10,000 µg N / L

Relative Standard Error

1st order fit 50 – 2,000 µg N / L

Page 17: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

PRACTICAL USE – PERFORMANCE METRICS

17

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Detection Limit 12 µg N/L

Reporting Limit 50 µg N/L

Range Upper Limit 10 000 µg N/L

Spike recovery 99.3% POTW (Anaerobic digester sludge)

97.3% Industrial discharge (Food process)

109% Industrial discharge (Metal finish)

96.9% River water

102% POTW (Final effluent, pre-UV)

105% POTW (Primary clarifier effluent)

Throughput 50 samples / h

Page 18: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

PRACTICAL USE – PERFORMANCE METRICS

18

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

▰ Above performance with 35 µL sample loop▻ "Standard" size that works both for ammonia and TKN

▻ Ability to adjust range by varying sample loop size

Page 19: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

CONCLUSIONSWhat you should take home from this talk

19

4Copyright 2018. FIAlab Instruments, INC. All rights reserved.

Page 20: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

CONCLUSIONS

20

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

▰ New method for measuring ammonia and TKN

▻Applicable to wastewater and natural waters

▻Included in 2019 MUR, signed and being proposed

▻Pending inclusion in 40 CFR Part 136

▰ Method highlights▻High sensitivity

▻Quick calibration and analysis

▻Robust – gas diffusion

▻Simple - one-reagent chemistry

FIAlyzer-1000

Page 21: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

CREDITS

21

Copyright 2018. FIAlab Instruments, INC. All rights reserved.

▰ Askew Scientific Consulting

▻Dr. Ed Askew

▰ EPA Clean Water Act ATP Coordination Team

▻Lem Walker

▻Kevin Roberts

▰ Laboratories participating in the ATP validation experiments

Page 22: A New Method for Measuring Ammonia and TKN in Wastewater...Jun 11, 2019  · A. Aminot et al. in O. Wurl (ed.) “Practical Guidelines for the Analysis of Seawater”, p. 166-169

22

THANKS!Any questions?